7.2 REPLACEMENT PROCEDURES
Technical Service Manual 7 - 9 430-95551-001-DRAFT
7.2.6
CE MODULE REPLACEMENT
The recommended tool for this procedure is a No. 2 Phillips screwdriver.
The replacement parts for this procedure are:
CE Module
Screw, 4-40 x 1/4, Pan Head, Square Cone
CAUTION: Use proper ESD grounding techniques when handling components.
Wear an antistatic wrist strap and use an ESD-protected workstation. Store the PWA
in an antistatic bag before placing it on any surface.
CAUTION: When replacing the CE module, carefully check the Ethernet MAC
address on the module label to assure it matches the infuser barcode.
Note: Replacing the CE module does not change the existing biomed settings.
To replace the CE module, see Figure 7-3, Rear View of the Infuser, and Figure 7-4,
CE Module Replacement, then proceed as follows:
1. Press [ON/OFF] to turn off the infuser, and disconnect the device from AC power.
2. Carefully set the infuser face down.
3. Using the Phillips screwdriver, remove the two 4-40 screws from the CE module.
4. Carefully pull the assembly away from the infuser.
Note: When removing the CE module, note the placement guides where the
peripheral PWA rests.
5. Install the replacement CE module in the exact reverse order of removal.
Note: Verify the peripheral PWA is placed properly between the guides and
fits correctly into the CPU PWA.
To verify successful CE module replacement, perform the PVT in Section 5.2.
SECTION 7 REPLACEABLE PARTS AND REPAIRS
430-95551-001-DRAFT 7 - 10 Plum A+ with Hospira MedNet
Figure 7-3. Rear View of the infuser
AC POWER CORD 6-32 X 3 1/4
PAN HEAD
SCREW
POLE CLAMP
ASSEMBLY
6-32 X 2 1/2
PAN HEAD
SCREW
05K01043
GREEN
LED YELLOW
LED
NURSE
CALL
JACK
RJ45
ETHERNET
CONNECTOR
VOLUME
CONTROL
KNOB
7.2 REPLACEMENT PROCEDURES
Technical Service Manual 7 - 11 430-95551-001-DRAFT
Figure 7-4. CE Module Replacement
CE MODULE
4-40 x 1/2
PAN HEAD
SCREW (2)
05K01040