Hulu Devices HULUDEV0001 Smartphone Auscultatory Blood Pressure Kit User Manual

Shanghai Hulu Devices Co., Ltd. Smartphone Auscultatory Blood Pressure Kit

User Manual

Download: Hulu Devices HULUDEV0001 Smartphone Auscultatory Blood Pressure Kit User Manual
Mirror Download [FCC.gov]Hulu Devices HULUDEV0001 Smartphone Auscultatory Blood Pressure Kit User Manual
Document ID3256668
Application IDp7XElKvRlgzslGv97QW55A==
Document DescriptionUser Manual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize20.77kB (259635 bits)
Date Submitted2017-01-12 00:00:00
Date Available2017-01-14 00:00:00
Creation Date2017-01-10 11:18:22
Producing SoftwareAcrobat Distiller 11.0 (Windows)
Document Lastmod2017-01-10 17:29:33
Document TitleMicrosoft PowerPoint - Accutension说明书.pptx
Document CreatorPScript5.dll Version 5.2.2
Document Author: 0357

Accutension
cuff
Smartphone Aided Ascultatory Blood Pressure Kit
stethoscope
head
Dear Customer,
Thank you for choosing Accutension. Before you use the product,
please read the manual carefully and follow the instructions. This
kit is not an automatic blood pressure monitor, so some practice
may be needed to make it work properly. Please contact us if you
encounter any problem.
problem
phone
1.
2.
Packing List
Item
Quantity
Cuff (22cm‐32cm)
(22cm 32cm)
Cuff (32cm‐42cm)
Pressure Sensor Module
Hand pump with quick release
Stethoscope with microphone
Micro USB charging cable
Headphone plug converter
Slow deflation vent
Short rubber tube
Setup the kit
Figure 1 shows how the parts are connected together and how
the kit is operated. The cuff tube is connected to the pressure
sensor module, then to the slow deflation vent and then to the
hand pump with quick release valve. You can use the short rubber
tube we provided to make the connection. Three slow deflation
vents with different deflation rates are provided.
provided You need to
choose the right one after you learned your rough blood pressure
value and the right cuff size with the aid of App that will be
introduced later. You need to choose the right cuff size. The cuff
has an index and please make sure it is within the OK range when
you wear it.
3.
artery
hand
pump
pressure
sensor
module
Get familiar with the App
Search “Accutension” in Apple App Store to install the app. Figure
slow
deflation
vent
quick
release
valve
Figure 1. The Kit Setup And Operation
2 shows the measurement user interface. No.1 is the pressure
sensor module
d l battery
indicator
and
d it shows
up when
h the
connection is formed. No. 2 is the connection button. Before
measurement, you need to touch this button to set connection after
the pressure sensor module is powered on. No. 3 is the current user
icon. Default is Guest. You can create new users in “more” tab. If
there are multiple users, you can touch this button to switch the
user. No.4 is the location to show cuff pressure during cuff inflation
and deflation. No. 5 and 6 are the systolic and diastolic BP marking
lines respectively. You can move No. 5 to the first sound of blood
flow pulse and No. 6 to the last. These two lines can be moved by
scrolling No. 8. When you scrolling up and down No.8, the numbers
change
g with BP markingg line locations. No. 7 shows the cuff deflation
speed after the measurement. Cuff deflation speed is important for
Auscultatory method. Please select right slow deflation vent to make
the speed within 2‐4 mmHg/s.
4.
Do a measurement
Power on module. Push the button for a while will turn on the
pressure sensor module. When it is on, the blue light will blink.
Plug the stethoscope into your phone. One end of the stethoscope
1. Battery indicator
2. Set connection
3. Current user
4. Cuff pressure
display
5. Systolic BP
marking line
6. Diastolic BP
marking line
7. Cuff deflation
speed
8. BP reading picker
Figure 2. Measure User Interface
is a standard 3.5mm headphone plug, please plug it into your
phone jack. If you want to hear the sounds in real time, you can
use the
h h
headphone
d h
plug
l converter. In this
h case, please
plug
l the
stethoscope and your headphone into the converter and then
plug the converter into the phone. You need to choose the
converter with the right color depending on your iDevice as
shown in the table below.
S it bl ffor iDevice
Suitable
iD i in
i China
Chi
Suitable for iDevice out of China
Start the App.
Set Bluetooth connection. Touch the connect icon (No.2 of
Figure
2)) and in the popup
p p p window choose Accutension. The
Bluetooth connection between the pressure sensor module and
the phone will be set. The blue light of the pressure sensor
module will stop blinking.
Choose the user. Follow instructions in Section 3 to change a
user.
Wear the cuff. The bottom cuff edge needs to be 1 inch above
the antecubital fossa. The Cuff can’t be too loose or two tight. It
i OK to have
is
some space to put into
two finger
fi
tips.
i The
h cuff
ff
INDEX needs to be within the OK range as printed in the cuff,
otherwise try another one.
Start measurement. Quickly pump the cuff to the pressure
30mmHg higher than the estimated blood pressure value. For
example, if your systolic blood pressure is 120mmHg, you need
to pump the cuff to at least 150mmHg. Then stop pumping, the
cuff will start self deflation. You can see the sound signals on the
phone screen. The Korotkoff sound signal will appear when the
systolic blood pressure is reached. Then the sound signal will
continue on the screen until the diastolic blood pressure is
reached. Then yyou need to p
push the q
quick release button on the
manual pump to quickly deflate the cuff. Please make sure the
cuff air is fully released when you release the button. Now the
phone should display the blood pressure value. Please be noted
that the blood pressure value is not the final value because this is
not an automatic blood pressure kit. You might need to playback
the sounds to double confirm.
confirm
Choose the right slow deflation vent. There are three slow
deflation vents in grey, white and blue respectively. The grey one
has the fastest deflation while the blue one has the slowest
deflation. The deflation speed will be showed after each
measurement. The best deflation speed is 2 ‐ 4mmHg/s, so if it is
higher than that,
that please select a slower deflation valve.
valve The ideal
range of deflation speed is 2mmHg‐4mmHg to balance accuracy
and comfort.
Playback the sounds. Touch Playback in the App to playback the
sounds. You need to use headphone to listen clearly because the
iphone speaker is not good at playing low frequency (bass)
sounds.
Adjust the reading if necessary. You can scroll the blood
pressure number picker to adjust the blood pressure marking
line location.
Interpret the sounds curve. Figure 3 shows various sounds
strength curves. Please note noise such as friction can induce a
peak in the sounds curve, so the features of blood flow sounds
can help you recognize the noise sometimes.
(a) is the most common curve. It shows regular heart beat, there
is clear first sound and last sound of blood flow pulse. ① is
systolic BP and ② is diastolic BP.
((b)) shows there is a gap
g p between regular
sounds signals.
For this
case, you need to listen to confirm whether the first part ① ‐ ②
are the blood flow pulse sounds or noise. If they are the former,
the gap is called auscultatory gap and it happens to some people.
①
②
(a)
①
②
b.
③
④
d.
(b)
①
c.
e.
②
(c)
f.
①
②
(d)
Figure 3 Various Sounds Strength Curves
① is systolic BP and ④ is diastolic BP. If they are the latter, ③ is
systolic BP and ④ is diastolic BP.
(c) shows that the auscultatory sounds do not disappear.
disappear After ②,
②
the sounds become weak and the sounds strength keeps almost
unchanged. This also happens to some people. In this case, ② is
diastolic BP.
(d) shows irregular heart beat. For example Afib patients have this
kind of sounds strength curve. For this case, you need to playback
and listen carefully to find out the first and last sounds of blood flow.
flow
Accutension can help you find out this kind of irregular heart beat.
You need to consult your doctor when you find this phenomenon. It
could be linked to some diseases.
Turn off the pressure sensor module. Please do not forget to turn
off the pressure sensor module after use!
5.
Others:
a.
The sound quality determines the measurement quality. So
g.
please do your best to avoid noise. For example, find a quiet
place to do the measurement. Ask others not to talk during
measurement. Keep the tube of the stethoscope clear from
the cuff tube to avoid unnecessary noise. Don’t touch the
tube of the stethoscope during measurement.
Please do not use when you charge your phone because of
heavy electrical noise.
If you have questions on the sounds or the number, you can
send the sounds to your doctor or medical professionals
through
g email. You doctor can open
p the measurement data
in his/her phone to find the BP readings for you.
Follow all the normal blood pressure measurement
requirement. For example, take a rest before measurement,
be seated on a chair with back support, put two feet on the
floor, keep the cuff on the same level of the heart, etc.
The pressure sensor module can be cross‐calibrated with
most of the upper arm automatic blood pressure monitors
or traditional blood pressure kit, so if you find any abnormal
of the pressure number, you can do this by finding another
blood pressure monitor.
You can use the micro USB cable to charge the pressure
sensor module.
mod le The red light is on during
d ring charge.
charge The green
light is on when the charge is full.
You can use Accutension Stetho App to check the algorithm
accuracy of an automatic BPM. That App requires a QR code
to register. Here is the code to scan. In the future,
Accutension App may also need this code. So Please keep
this manual.
FCC Statement
This device complies with part 15 of the FCC Rules. Operation
is subject to the following two conditions: (1) This device may
not cause harmful interference, and (2) this device must
accept any interference received, including interference that
may cause undesired operation.
Any Changes or modifications not expressly approved by the
party responsible for compliance could void the user's
authority to operate the equipment.
equipment
Note: This equipment has been tested and found to comply
with the limits for a Class B digital device, pursuant to part 15
of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a
residential installation. This equipment generates uses and
can radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no
guarantee that interference will not occur in a particular
installation If this equipment does cause harmful
installation.
interference to radio or television reception, which can be
determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more
of the following measures:
‐Reorient or relocate the receiving antenna.
‐Increase the separation between the equipment and
receiver.
‐Connect the equipment into an outlet on a circuit different
from that to which the receiver is connected.
‐Consult the dealer or an experienced radio/TV technician for
help.
help

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : No
Author                          : 0357
Create Date                     : 2017:01:10 11:18:22+08:00
Modify Date                     : 2017:01:10 17:29:33+08:00
XMP Toolkit                     : Adobe XMP Core 5.4-c005 78.147326, 2012/08/23-13:03:03
Format                          : application/pdf
Title                           : Microsoft PowerPoint - Accutension说明书.pptx
Creator                         : 0357
Creator Tool                    : PScript5.dll Version 5.2.2
Metadata Date                   : 2017:01:10 17:29:33+08:00
Producer                        : Acrobat Distiller 11.0 (Windows)
Document ID                     : uuid:19d0d502-57c1-4d02-8b95-187f4915a535
Instance ID                     : uuid:e4225e50-bca2-47e7-bd99-5fed8d0354b7
Page Count                      : 3
EXIF Metadata provided by EXIF.tools
FCC ID Filing: 2AKSGHULUDEV0001

Navigation menu