IMEDIPLUS MCE12L001 Cardiart Wireless 12-Lead Electrocardiograph Monitor User Manual

IMEDIPLUS INC. Cardiart Wireless 12-Lead Electrocardiograph Monitor

User Manual

             Cardiart Wireless 12-Lead Electrocardiograph Monitor Owner’s manual              Be sure to read the original manual and follow the instructions before use  2017.09.30 Ver. 1.0
1  Contents Warning and Precaution ........................................................................................ 2 Chapter 1 Outline .................................................................................................. 4 1. Product introduction ................................................................................ 4 2. Intended use ............................................................................................. 4 3. Principle ................................................................................................... 4 4. Features .................................................................................................... 4 Chapter 2 Composition .......................................................................................... 5 2.1 Appearance .............................................................................................. 5 2.2 Content ..................................................................................................... 6 Chapter 3 Electrode locations ............................................................................... 7 Chapter 4 Installation ............................................................................................ 9 Chapter 5 Software Operation ............................................................................. 11 5.4 VIEW/ Add/ Delete the patient data ............................................. 20 5.5 Select a patient profile to creative a profile for patient report ...... 21 5.6 View a patient report .................................................................... 21 Chapter 6 Environment and Maintenance ........................................................... 22 Chapter 7 Products Specification ........................................................................ 25 Chapter 8 Symbol Description and Manufacturer Information .......................... 26
2  Warning and Precaution Warnings for product using 1) The product is not intended to diagnose. 2) Please remove the battery when not using for a long time, to avoid corrosion of the spring if the battery holder. 3) The product can be used continuously for at least 19 hours under normal use. 4) Do not use the product near any flammable gas (e.g., high concentrations oxygen, hydrogen or anesthetics), which can cause an explosion or fire. 5) To avoid electric shock, do not disassemble or modify the product of one’s own. 6) Please contact the manufacturer before connecting the product to other equipment. 7) This product is a portable device, but the safety of the instrument is not able to be ensured in the transport process with a strong impact or fall. 8) Avoid the product from water, or the patient and the operator may involve in electric shock and other dangers. 9) To avoid infection, please follow the instruction bellow  Regularly clean the parts that contact with patients.  Patients of open type infectious disease do not use the product for testing. 10) When placing the wires or cables, be aware to avoid the possibility of tripping people or winding the patient's neck. 11) To ensure using the equipment safely, follow the maintenance procedures contained inthe original manual. 12) Improper battery disposal may result in explosion or contamination. Please recycle the battery in accordance with local regulations, do not throw the battery as common trash 13) Do not use the product with Pacemaker.  Precaution 1) The conductor parts of the electrode of ECG signal record box and its connector (including the neutral electrode) should not contact with other conductor parts, including the earth. 2) The factory will yearly check whether the ECG signal record box and the cable are damaged. 3) Do not use the product with high frequency surgical equipment. 4) The product is classified as CF type equipment, but not directly used for the patient's heart. 5) There will be potential risks when using the product along with installation of cardiac rhythm or other electrical stimulator at the same time. Pay full attention to changes of the ECG to ensure safety and to ensure that the leakage current is within the allowable range. 6) Do not use accessories that are not included in the product system. The use of non-original accessories may result in inaccurate patient data, damage to equipment, and invalid warranty.
3  7) Please use the original standard patient cable, otherwise it will reduce the minimum safety of the product. 8) The electrodes is suggested to be placed by professional medical personnel. 9) To avoid the possible damage, do not contact the product with sharp or hard objects. 10) Do not expose the patient cable to strong ultraviolet radiation. 11) Do not stretch the patient cable for this may cause mechanical or electrical malfunction. Before storing the patient cable should be wrapped as a loose circle. 12) Avoid placing the patient cable where it may be caught, stretched, or stamped. Otherwise the results may be inaccurate and may require repair. 13) Please use the original accessories. The use of non-designated accessories may result in degradation of equipment performance or unsafety. 14) Portable and mobile radio frequency (RF) communication equipment will affect the performance of the product. 15) Other medical devices (including but not limited to defibrillators, ultrasound instruments, pacemakers and other stimulators) can be used along with the product, but such equipment may interfere the signal.
4  Chapter 1 Outline 1. Product introduction Wireless ECG monitor of IMEDIPLUS (hereinafter referred as the product; model: MCE12L001) is a compact, lightweight, easy carry, and agile product. The data is transmitted through Bluetooth, avoided the inconvenience of the traditional data transmission; also making transmission space and distance more flexible, convenient and smooth. The software can be installed on iOS devices, users can view data on iPad / iPhone / iTouch; this kind of working mode is more flexible than the traditional system composed of personal, laptop, the record box (with data lines ), and the printer (with data transmission cable).  2. Intended use The product is mainly for checking abnormalities of the adults’ heart through recording ECG by medical staff in hospital or clinic. The product cannot be used as the only basis for diagnosis.  3. Principle The product collect the ECG signal through lead wire, initially amplifies the signal through preamplifier, and filter out the polarization voltage, baseline shift, and high frequency noise. Amplify the signal with the use of secondary amplification circuit, and convert the analog signal to digital signal with the A / D converter. Then transmit the signal to iOS mobile system through Bluetooth, at the same time, display on the screen for clinicians to diagnose and study.  4. Features  Compact, lightweight, simple operation, and easy carry.  Breakthrough the limitation of traditional ECG (workstations).  Timely information transmission.  Low power-consuming Bluetooth.
5  Chapter 2 Composition The product is composed of ECG record box, ECG lead wire, electrode patch/ suck ball and instruction manual.   2.1 Appearance   Power button  Bluetooth       Power Indicator Green light for sufficient power; Glint for power shortage  Lead wire port      ECG Signal Record box Front    Lead wire plug        2 AA 1.5 V batteries Battery compartment   Battery cover   Back   Caution: If there’s any damage between the connection of the lead wire and signal record box, please replace with the specified lead wire.
6  2.2 Content   No.  Item Name Amount 1 ECG Signal record box 1 2 ECG lead cable 1 3 Instruction Manual 1 4 ECG Arm clamp 4 5 ECG Suck ball 1 set
7  Chapter 3 Electrode locations  3.1 WILSON standard lead  Electrode loctions  Electrode AHA IEC Hookup A V1 red C1 red The fourth intercostal space to the right of the sternum B V2 yellow C2 yellow The fourth intercostal space to the left of the sternum C V3 green C3 green Midway between electrode B and D D V4 blue C4 brown The fifth intercostal space in the mid-clavicular line E V5 orange C5 black Horizontally even with D in the anterior axillary line F V6 purple C6 purple Horizontally even with D in the midaxillary line G LA black L yellow Left arm (resting ECG) or left shoulder (exercise test) J RA white R red Right arm (resting ECG) or right shoulder (exercise test) H LL red F green Left foot (resting ECG) or left leg (exercise test) I RL green N black Right foot (resting ECG) or right leg (exercise test)
8  3.2 Electrode connection and position  Precordial lead:Coat the patient’s chest with conductive paste or wipe with alcohol cotton balls in correct position, then attach the lead and suck ball to the patient in the order as below:  Precordial lead position: V1:The fourth intercostal space to the right of the sternum V2:The fourth intercostal space to the left of the sternum V3:Mid-point between V2 and V4 V4:The fifth intercostal space in the left clavicular midline V5:The fifth intercostal space in the axillary line  V6:The fifth intercostal space in the axillary midline      Electrode clip: Limb lead: 1) Clean the skin of the patient 2) Apply conductive paste or wipe with alcohol cotton balls 3) Attach the electrode clip, preferably on the inside of limbs 4) To prevent oxidation and poor connection, please clean the clips after use     Limb leas position: RA:Right arm LA:Left arm RL:Right leg LL:Left keg     Recommendation: 1) It is recommended that electrodes be placed by medical trained personnel. 2) To obtain correct ECG record, please previously clean the skin.
9  Chapter 4 Installation This chapter is divided into two parts. First, the components and the installation of hardware; second, the components and the installation of software. 4.1 Hardware installation The product is mobile device combines ECG record box (see 2.1 appearance) with iOS system, please follow the instructions before use. 4.2 Download vhECG from Apple App Sore Step 1: Turn on the ECG record box; press the power  ,  the  indicator  would light (green light for sufficient power while glint for power shortage). Step 2: Turn on the mobile device; click〝 Settings〞, select and turn on 〝Bluetooth〞.    Step 3: Open the Cardiart     , enter the home page, enter the settings     , then click "ECG device link" and select Bluetooth. As below:   Step 4: Back to the home page and click   to scan the device. When successfully connected, you will press  to start working.
10    Device connection diagram    Notice: The last five digits of the factory number of the ECG record box will be displayed at the bottom of the mobile device and you will need to select the same device number to connect.
11  Chapter 5 Software Operation ECG software (Cardiart ECG) is an easy-to-use application that offers professional electrocardiographic tools and an excellent user experience. Most of the operation is the same as iOS system. If you are familiar with the Apple app and have the basic knowledge of ECG, there won’t be any problem of operation. This chapter portrays the basic operation of the software, please read carefully. When the connection between mobile device and the record box is completed, click the Cardiart icon    to run the software and enter the home page (as below):   System settings                    Home page Communication mode display
12  ● Software is mainly composed of settings, ECG monitoring, records, and network      5.1 Settings You can customize your vhECG with your own prefernece. Click the icon   to enter the settings which consists of four parts: 1) General  Hospital Name – Input your hospital or organization name.  Physician Name - Used for diagnosis signatures. 2) ECG Signal capture  Use demo data - ON / OFF: Data DEMO in monitoring  Fallback record mode - ON / OFF, Recording mode.  High-pass Filter Frequency- Cutoff frequency of high-pass filter: 0.05-0.5Hz, default: 70Hz.  Low-pass Filter Frequency - Cutoff frequency of low-pass filter: 30-250Hz, default: 70Hz.  Notch Filter Frequency: Eliminate power-line interference. Network: display the uploaded ECG report. ECG monitoring: ECG recording, new medical records, data acquisition (record), preview, edit, delete, print report. ECG records: to view the collected ECG, remove cases and so on. Settings: hospital basic information, recording time, filter, report format and other related parameters.
13   Anti-Baseline Drift - ON / OFF:Keep the baseline stable.  ENG filter - ON / OFF:Remove the EMG from the ECG signal 3) Communication  ECG Device Connectivity – Bluetooth 4) Report Custom the printing form:  Title – title on the printed report  Report format – PDF or JPG  Lead layout - 12×1\2×6\3*4+1  Paper orientation – A4 or B5, vertical or horizontal  Notice:  Anti-Baseline Drift: This function can effectively remove the baseline drift due to respiration or body movement without affecting the ECG frequency response, such as ST segment.  Muscle noise can be removed by setting LPF with low cutoff frequency but it will severely affect ECG high frequency component such as QRS complex. vhECG EMG filter can effectively remove muscle noise with little affection to ECG signal.  5.2 Monitoring and Recording This section shows the way to record ECGs. Step 1: click “ECG monitoring”   to enter the recording page. Make sure the ECG record box has successfully connected to mobile device, if not, please refer to chapter 4 for reconnection. Notice:
14  If successfully connected, enter the "recording" stage (as below)                  "Pre -recording stage" interface    Step 2: When the ECG waveform is stable, press "Start" button     to record.  "Formal recording" interface   Notice: Double-click will refresh the interface to restart recording. Lead display switching: 12*1、6*2、3*4。 Start recording (roll down with two fingers)
15  Icon description:              Step 3: After recording, the user can choose the operation according to the actual needs: view now, uploading, add the back mode, end recording.  “End recording”interface Description: shows ECG connection Filter selection:When the waveform drift, click to adjust the waveform. Recording time selection: 10s、20s、30s、60s. Lead Switch:3*4,6*2,12*1. Sound switch.
16  (1) View Now ECG report browsing interface, drag to view the complete ECG report.  ECG report browsing    (2) Background mode Add background continues to ECG (3) End recording Return to home page.
17  5.3 ECG records 5.3.1 View the record (1) Click    to view all recorded ECG.  (2) Icon Description  Click  to display the icon as the below    Compare with the previous data    Press    for QRS wave adjustment window, drag left and right to view QRS lead wave.
18     “QRS”window In "QRS" window, the bottom of the report shows single lead waveform. Click on a lead to view the waveform (as below). QRS window of Single Lead  In the operating area, slide with two fingers to call out ruler to measure the waveform (Figure: blue dotted lines and numbers); adjust the locating point of QRS wave by dragging the blue arrow Click on a lead to view thewaveform (as below)
19   Click     to display the icon as the below     Sending email: send the current record through email.  If you have not set up sender's mailbox, a prompt would bump out (as below). Click “Yes” to exit the software, return to desktop and set up the mailbox.   Print preview for current ECG report.   (3)   : uploading  Please link  http://www.cardiartcloud.net to register your data before uploading  The recorded ECG could be uploaded to Cloud or computer.   (4) Click  Signature for confirming diagnosis (physicians’ signature)
20      Select the diagnostic result     Delete the diagnostic result   Refresh the diagnostic results, the diagnosis and signature will all be covered.   5.4 VIEW/ Add/ Delete the patient data 1) Create a patient profile: Click    and enter a patient profile interface to review the record.  2) Create a patient report 2.1 Enter  〝ECG record〞 2.2 Click a  〝patient report〞 2.3 Click  〝quickly record〞
21      5.5 Select a patient profile to creative a profile for patient report       5.6 View a patient report  Enter ECG record to review a patient report If patient’s medical record has been added, enter the patient’s name or number to search for the patient.
22  Chapter 6 Environment and Maintenance 6.1 Environment Avoid sunlight and keep dry; avoid wet and dust to remain the cooling channel smooth; keep away from strong magnetic field.  Operating temperature: 5-40℃  Humidity: 25%-95%(no condensation)  Atmospheric pressure: 700hPa~1060hPa 6.2 Transportation and Storage  Temperature: -20℃~55℃  Humidity: ≤93%  6.3 Cleansing ECG record box  Wipe the surface with soft, damp cloth, soaked with neutral detergent.  Do not use strong detergent to avoid damaging the plastic surface.  ECG Cable  Disconnect the lead before cleaning or disinfecting.  Wipe the surface with soft, damp cloth, soaked with neutral detergent.  Do not immerse lead wires in liquid.   6.4 Maintenance  Make sure whether the record box and lead wires are damaged before using; if damaged, stop using and contact the manufacturers in time.  The ECG record box shall be verified every two years only by the original factory or recognized professionals.    Please remove the battery whenever the ECG will not be used for a long time to avoid corroding the spring in the battery seat.  Forbidden to use while charging. Warning:
23  6.5 Troubleshooting   Common problem Solution Unable to start the record box Please insert a battery with sufficient power.  Unable to record Check the connection between the mobile device and the record box. Restart the record box.  Collected wave form appears straight or noise 1. Check the connection between lead wire and record box. 2. Make sure there’s no strong magnetic field (e.g. MRI) in the room.  Above are the common fault for mobile ECG, if the fault repeatedly occurs or there’s any other problems, please contact the original factory.  6.6 Description of electromagnetic interference This section is the special tips for electromagnetic compatibility. The wireless ECG monitor of IMEDIPLUS should be installed and used according to this section. The basic function of this product is to record common ECG for doctors to diagnose; the error of sensitivity is ± 5%.   ★ Portable and mobile RF communication devices may influence the use of mobile ECG; it is recommended to keep away from or turn off portable and mobile RF communication devices while using ECG. ★ The lead wire must be provided by IMEDIPLUS. Among this, the length of the cable: lead wire: 1.7m with shelter. ★ To ensure the product can be used normally, the electromagnetic emission is not increased, and the immunity is not reduced, please use the cable and related accessories provided by IMEDIPLUS; the other accessories may result in increasing the electromagnetic emission and reducing immunity. ★ This product should not be used near or stacked with other equipment with the same or similar frequency of operation. If it is unavoidable, make sure the product works properly in its configuration. ★ Using the product along with accessories other than the provided ones may result in increasing equipment or system emissions or reducing immunity. ★ Bluetooth module parameters: emission frequency: 2400MHz-2483.5MHz; Modulation type: GFSK; Effective radiated power: -30dBM ~ -40dBM ★ WARNING: The equipment or system may still be interfered, even if other equipment complies with appropriate national standard emission requirements   Precaution:
24  FCC Warning  FCC Intentional Radiator Certification Cardiart Wireless 12-Lead Electrocardiograph Monitor (Model no.: MCE12L001) FCC ID: 2AM7N-MCE12L001 This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesirable operation. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part  15  of  the  FCC  rules.  These  limits  are  designed  to  provide  reasonable  protection  against  harmful interference in a residential installation. This  equipment  generates,  uses  and  can  radiate  radio frequency  energy  and,  if  not  installed and used  in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: -Reorient or relocate the receiving antenna. -Increase the separation between the equipment and receiver. -Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. -Consult the dealer or an experienced radio/TV technician for help. You  are  cautioned  that  changes  or  modifications  not  expressly  approved  by  the  part  responsible  for compliance could void the user’s authority to operate the equipment.     6.7 Environmental protection Disposing of the battery and equipment shall be done within the area approved by municipal unit, to avoid environmental pollution. The service life of the device is 5 years.
25  Chapter 7 Products Specification   Model MCE12L001  Intended Use The product is mainly for checking abnormalities of the adults’ heart through recording ECG by medical staff in hospital or clinic. Lead 12 lead Input circuit Built-in anti-defibrillation protection Recording mode Synchronous 12-lead recording Dynamic measurement range ± 5mVpp The time constant ≥ 3.2s Frequency response 0.05Hz~150Hz(-3dB) Calibration voltage 1mV± 3% Sensitivity Divided into three gear: 5mm / mv, 10mm / mv, 20mm / mv, the error range: ± 5%. Input resistance ≥ 50MΩ Noise voltage level No more than 30µ A Anti-baseline drift automatic / manual Patient leakage ≦ 10μA Common-mode rejection ratio ≥90dB Data transmission wireless Bluetooth transmission distance 10m Power supply Two AA 1.5 V Alkaline battery Sampling rate 500sps/channel Safety classification: Class II, CF type
26   Chapter 8 Symbol Description and Manufacturer Information     Manufacturer name & address  Manufacturing date (year/ month/ day)  Lot number   Serial number  Expiration (limited to year/ month/ day)  Only for single use  Do not use if package is damaged  Do not use if package is damaged  Notice, please check the warning / precautions   Anti-Automated External Defibrillator CF type  Blue tooth  Avoid high temperature and sunlight  Upper / lower temperature limitation   Electromagnetic wave warning      Manufacturer / Drug maker: IMEDIPLUS INC. Address: 2F, 12, ShengYi Rd. Sec. 2, Chupei City, Hsichu County