InBody INBODY-H20B Low Power Communication Device User Manual manual

InBody Co., Ltd. Low Power Communication Device manual

manual

       Thank you for purchasing InBody H20 or H20Ⓑ. H20Ⓑ has Bluetooth function but H20 doesn’t. Please read this manual carefully and use it with proper operation. Keep it handy for future reference. ✽This product is medical equipment.
    ⓒ 1996-2012 Biospace Co., Ltd. All rights reserved.  Notice This user’s manual  cannot be copied, translated or transformed into different media without a prior written  consent  from  Biospace  Co.,  Ltd,  according  to  the  copyright  laws.  This  user’s  manual  and product  pertinent  to  this  manual  may  have  misprints  or  technical  flaws,  which  can  be  modified without prior notice to the readers. Biospace is not liable for any damage caused by a failure to meet the requirements in this user’s manual.  Biospace Co., Ltd. [Head office] 518-10, Dogok 2-Dong, Gangnam-gu, Seoul 135-784, Korea TEL : 82-2-501-3939 FAX : 82-2-578-2716 Homepage: http://www.inbody.com E-mail: info@inbody.com  Acknowledgements Biospace, InBody, Lookin’ Body are the registered trademarks of Biospace Co., Ltd. The names of the companies and products in this manual other than those of Biospace Co., Ltd. are the trademarks of the companies. Stating the products of other companies is strictly for the purpose of  providing  information,  and  not  to  guarantee  or  recommend  the  products.  Biospace  is  not responsible for the performance or the use of these products.  Biospace reserves the right to modify the dimensions or exterior of InBody H20 to improve the quality of the products, without consent of the customer.    BM-ENG1C-A5-A-120127
I  How to use this manual  This user’s manual explains the functions of the InBody H20. Follow the instructions below for effective use this manual.  1.  Please read all the instructions in this manual thoroughly before operation. 2.  Fully  utilize  the  helpful  materials,  such  as  the  diagrams  and  illustrations,  to  get  a  clear understanding of the product. 3.    Before calling Biospace for assistance,   please refer to “ Chapter 11. One-minute check before calling for service ”. 4.    If  you  have  clinical  issues  while  using  the  InBody  H20,  please  contact  us  using  the  e-mail address as shown below. E-mail : info@inbody.com    TEL: 82-2-501-3939 5.    Most importantly, please read the instructions and become familiar with the following notations:      The important  information  to warn  you of  situations  which might  cause  major  injury and/or damage to property if instructions are not carefully followed.     The important  information  to warn  you of  situations  which might  cause  minor  injury and/or damage  to property if instructions are not carefully followed.       The important and helpful information for operation.
II  Safety Information     Make sure not to use this equipment with medical electrical devices, such as a pacemaker.   NO USAR USARLO CON EQUIPO MEDICO ELECTRONICO ASI COMO CON MARCAPASOS.   Keep this equipment away from liquid. MANTENER USARLO LEJOS DEL AGUA.    1. Do not use this equipment in combination with the following electronic medical devices. - Medical devices, such as a pacemaker - Electronic life support systems, such as an artificial heart/lung - Portable electronic medical devices, such as an electrocardiograph.  2.  Physically  disabled  persons  or  children  should  not  attempt  to  take  measurements  alone,  but instead should get help by having manager or guardian hold them from the side or the back not to slip or to measure incorrectly.    3. Do not dismantle the equipment or open the back cover. Internal parts are not for customer use. If the  equipment is  dismantled,  the  warranty is  void,  and  service  costs  will  be  charged.  If  service  is required, contact Biospace or the supplying agency.  4.  When  connecting  peripherals  (optional  devices)  to  the  InBody  H20,  turn  on  the  power  of  the peripherals before turning on the InBody H20. When turning the power off, turn off the InBody H20 before turning off the peripherals. This process will minimize the harm to the equipment caused by electrical shock.
III     5. You may slip off the device if you step on it with wet feet.  6. The arm consists of a hand electrode. Do not force the arm in the wrong direction. The resulting damage may affect the functioning of the internal cable and circuit board.  7. Individuals with any kind of contagious disease or any kind of injury on the palm of their hand or sole of their foot should not come in contact with this product.  8. Never start weight reduction or exercise therapy without instruction from a physician or a specialist. Self-diagnosis may damage your health. Consult with your physician first.    9. This equipment is specifically designed to analyze body composition. Use the equipment only for its intended use as described in this manual.    10.  Follow  local  governing  ordinances  and  recycling  plans  regarding  the  disposal  or  recycling  of device components.
IV     1. Cross Contamination Individuals with any kind of contagious disease or infection must not use or come in contact with the platform or this product. Please be sure to clean the platform with appropriate disinfectant after each use. Never pour any liquid directly on the platform, as it may leak and cause internal damage. Use a soft cloth and appropriate ethyl alcohol to wipe off the platform. Do not wipe the platform with strong chemicals.  2. While moving, installing or using this product, be sure to protect it against any physical shock or damage. Always use the packing material and the original shipping box when moving or transporting this product.    3.  Always  operate  this  product  within  prescribed  ranges  of  temperature,  humidity,  and  pressure. Operating  in  ranges  outside  of  those  specified  may  affect  the  operation  of  this  product  and  may cause malfunctions.  4.  Be  careful  not to  spill  or  drop  any  residues  of  food or  beverages  on  this  product.  It may cause serious damage to the electronic components.  5. Cleaning Keep  the  platform  and control panel  clean  by  wiping  them  with  disinfectant.  Never  pour  any liquid directly on the platform, Use a soft cloth and appropriate ethyl alcohol to wipe off the platform. Do not wipe the platform with strong chemicals.  6. Wipe up gently the platform and control panel of the product with a soft cloth for disinfectant.  7. Install or locate equipment only in accordance with the provided installation instructions.
V    8.  This  equipment  should  be  serviced  only  by  qualified  personnel.  Contact  Biospace  for  any Customer Service issues.  9. Please use your hands to operate buttons. Using feet may lead to malfunction of buttons.    1. This equipment may cause the above mentioned medical devices to malfunction. This  equipment may  cause  harmful interference to  other devices in the vicinity, if  not installed and used  in  accordance  with  the  installation  specifications.  However,  there  is  no  guarantee  that  the interference will not occur for a particular installation.    2. Do not dismantle the product. Internal parts are not for customer use. If the unit is dismantled, the warranty is void, and service costs will be charged to you. If service is required, contact Biospace or the supplying agency.    3. This  product  should  be  serviced  only  by  qualified  service  personnel.  Contact  Biospace  for examination, repair or adjustment.   4. Do not move the equipment with the power on.  5. Do not jump or shock on the platform, these may cause an incorrect measurement or malfunction.  6. Keep this equipment away from liquid, other contaminates. These will cause a loss of function or a critical damage on the product. Do not put this product in direct sunlight, close to heaters.
VI     Information 1.  This product  has  been  designed,  manufactured,  and  inspected  under  the  full  quality  assurance system of Biospace. Biospace fulfills European MDD(Medical Device Directive), and InBody H20 has acquired the CE conformity marking. 2. This equipment may cause the above mentioned medical devices to malfunction. 3. This equipment may cause harmful interference to other devices in the vicinity if not installed and used  in  accordance  with  the  installation  specifications.  However,  there  is  no  guarantee  that  the interference will not occur for a particular installation.   4. The InBody H20 has been designed, manufactured, and inspected under the full quality assurance system of  Biospace.  Biospace  fulfills the international standardization system, ISO 90001 and ISO 13485.   5.  The  InBody  H20  fulfills  the  Standards  of  IEC60601-1(EN60601-1),  Safety  of  Electric  Medical Equipment. In addition, the InBody H20 complies not only with Level A for Noise Immunity, but also with  Level  A  for  Noise  Emission  by  the  Standard  IEC60601-1-2(EN60601-1-2),  Electromagnetic Compatibility Requirements.
VII  Indicators & Safety Symbols     Disposal of old Electrical & Electronic Equipment   (Application  in  the  European  Union  and  other  European  countries  with  separate  collection system) This symbol indicates that this product shall not be treated as household waste. Instead, it shall be  handed  over  to  the  applicable  collection  point  for  the  recycling  of  electrical  and  electronic equipment.  By  ensuring  this  product  is  disposed  of  correctly,  you  will  help  prevent  potential negative  consequences  for  the  environment  and  human  health,  which  could  otherwise  be caused by inappropriate waste handling of this product. For more detailed information about             recycling this product, please refer to local governing ordinances and recycling plans.     Follow local government ordinances and recycling instructions regarding disposal or recycling of device components, including batteries.    10
VIII  Workplace Requirements         □ Operation Environment   Temperature range    : 10 – 40 ℃ Relative humidity    : 30 – 75 % Atmospheric pressure range  : 70 – 106 kPa  □ Storage Environment   Temperature range    : -20 – 70 ℃ Relative humidity        : 10 – 95 % (No condensation) Atmospheric pressure range  : 50 – 106 kPa  □ Battery     Ratting    : AA Type Battery(1.5V X 4ea)
IX  Introduction  Body  Composition  consists  of  4  major  components: Water,  Protein,  Minerals  and  Fat.  These  four elements are the fundamental ingredients the body is comprised of, and it is important for them to be in balance. Body composition analysis is expected to quantify and measure these ingredients.    Until recently, diagnosing obesity has focused on appearance, without considering a balanced body composition.  For  more  reasonable  healthcare,  accurate  body  composition  analysis  must  be performed first, to achieve the balance of the four major body components.  Biospace  has  earned  international  recognition  for  technical  expertise  demonstrated  through  the InBody  series.  Based  on  the  experience  and  technology  over  the  last  10  years,  Biospace  has released the body composition analyzer, the InBody H20.  InBody is highly recognized  for its world-best technological excellence by many customers in more than 40 different markets including the US, European countries, Japan, and more. The strong brand is well-known among medical professionals as well as academic communities, and now it is the time for you to meet InBody at your place. Enjoy healthy lifestyle with InBody!                                                                           Kichul Cha, CEO
X  Notice  Please read thoroughly and follow the instructions.  1. Please do not press            with other object on the platform.  2. Please use your hands to operate buttons. Using feet may lead to malfunction of buttons.  3.  Please  be  sure  to  avoid  keeping  or  using  InBody  in  wet  area,  i.e.  bathroom.  It  may  cause malfunction of the device.  4. You may slip off the device if you step on it with wet feet.  5. Please wrap InBody and keep it level while storing for a long period.    6.  Please  remove  the  batteries  from  the  device  while  not  in  use  for  a  long  time.  It  may  cause corrosion when left in contact with the device.   (Please follow the instruction in reverse order of ‘ Chapter 2. Insert Battery ’ )  7. When you change the batteries, please use only the new batteries. The 4 batteries that are used should be the same model from the same company.
  Contents  How to use this manual                                                          Ⅰ Safety Information                                                                  Ⅱ Indicators & Safety Symbols                                              Ⅶ Workplace Requirements            Ⅷ Introduction              Ⅸ Notice                Ⅹ   1.  Product Package              1 2.  Insert Battery              2 3.      Connecting InBody H20Ⓑ  and PC          3 4.      Weight Measurement Mode            4 5.  InBody Test Mode              5 6.      Maintenance              10 7.  Understanding of Results          11 8.  Error messages              14 9.  One-minute check before calling for service      15   Appendix              17  ✽  More about the InBody H20  ✽  Specifications  ✽  Customer Service Information
1  1. Product Package  When opening the box, check to make sure all the following items are included.    Included Items    ①  InBody H20 unit      ② Quick Installation Guide  ③  Battery (AA 1.5 V 4 ea) ④  Bluetooth dongle     To prevent physical shock, use Biospace’s packing material when shipping or transporting the equipment.
2  2. Insert Battery                                                                                                  < Rear part>  (1) Open the battery cover at the bottom of the device. (2) Insert the batteries (AA 1.5 V 4 ea) in right direction. (3) Close the battery cover until it clicks.    Battery is low when              is still on after you turn on InBody H20.   Please replace the batteries when              is blinking after you turned on InBody H20.
3  3. Connecting InBodyH20Ⓑ  and PC  (1) Connect Bluetooth dongle(Bluetooth dongle from product package) to PC.          (2) Access http://www.myinbody.com and register.   (3) Click “InBody PC Program Install Guide” and install the program. (4) Turn on H20Ⓑ  When the bluetooth is being connected to InBodyH20Ⓑ, the logo will display on the LCD screen.                               Please use the Bluetooth dongle provided by Biospace only.
4  4. Weight Measurement Mode  A. Body Composition Analyzing (1) Step on the foot board to measure your weight. (2) When the measurement is over there will be a beep and you will see your weight on the screen.  (3) When the weight is not correct, press the foot plate to turn on H20 and wait for 5 seconds. Then you can re-measure your weight.   B. Power Off (1) Weight measurement is done, step down from the foot plate and the power will automatically be turned off. (2) If you step down the foot plate while measuring, the device will not be turned off for 20 seconds.
5  5. InBody Test Mode    A. Tips Before Body composition analyzing    (1) Make yourself as light as possible. Wear light clothes for measurement.     (2)  Please  go  to  the  bathroom  before  measurement.  Urine  and  feces  affect  your  weight  and consequently the body composition analysis. (3)  Please  conduct the measurement  before meals  or  at  least  two  hours after meals.  Food in  the stomach before digestion affects body weight, causing possible errors in the result.   (4)  Please  avoid  taking  measurement  right  after  bath,  sauna,  physical  exercise.  Body  water  is  in instable  condition  after  a  bath,  a  sauna,  or  physical  exercise,  which  could  lead  to  incorrect measurement data.   (5) It is recommended to have a test in the morning. Body composition is in the most stable condition before starting full activities of the day.   (6) In case of repeated measurement to compare the results, please maintain surrounding conditions same (weight of clothes, time, food intake, etc.).   (7) If you wear a pacemaker or other electronic medical devices inside your body, please do not use this body composition analyzer.
6    B. Body Composition Analyzing (1) Push the button in the center to turn on H20. (2) Rotate the button in the center to set your height. Wait for few seconds once it is set.  (3) The height will appear on the LCD screen.   (4) After entering the height hold the handle and step on the foot board to measure your weight.  (5) When weight is set, the measurement position icon will blink while checking if the hand and feet electrodes are connected.   (6) With measuring sound, InBody measurement will start. (The measurement will take 3~5 seconds)  (7) When the test is over there will be a beep and you will see END on the screen. And Weight, Body fat Mass, Muscle Mass will be displayed on the LCD.   (8) Measurement results will be shown twice and the END will appear on the LCD. After that you may release the handles and step down.  (9) After stepping down, the result can be seen using the dial button.  Please use your hands to operate buttons. Using feet may lead to malfunction of buttons.  InBody H20Ⓑ  can be  operated for the body composition analysis without connection to PC. For more results and data management, however, use of InBody H20Ⓑ  with PC connection is recommended..
7    C. How to Hold InBody Electrode  Take the hand grip and hold it with the palms and thumbs touching the hand electrodes. Step on the foot electrodes at the platform. Make sure your hands and feet on both sides are not touching each other. Please refer to next page for the right measurement posture.    The measurement begins automatically when your position is set.    (1) Hand Electrodes  Fingers must not leave the surface of the electrodes. Right hand and left hand must not touch each other.   (2) Foot Electrodes  Step on the foot electrode with the              Be careful not to stand with your heels heels first and then the entire soles of        on the pants hems. your feet.                                                        Make sure your bare hands and bare feet maintain contact with electrodes. Please be careful not to let the hems of your pants get stuck between your heels and the foot electrodes. In case your feet are relatively small, please step on the electrode with the heels touching them first.
8    D. Recommended measurement posture   ➊  Please keep your eyes to the front.   ➋  Grab the hand electrodes with bare hands and be careful not to let your hands touch each other.     ➌  Thighs should not touch each other if possible. ✽Tip:  For  those  whose  thighs  contact  each  other,  it  is  recommended  to  wear  shorts,  which  come  down  below  the  thighs  for accurate measurement.     ➍  No moisture or too much callus tissues on the palms or soles of feet may lower the accuracy of the measurement.   ✽Tip : Wipe your hands and feet with wet tissues before the measurement.   ➎  Take your socks or stockings off and measure with bare feet.   Step on the platform with your feet and the electrodes aligned on foot electrodes. Your feet must not contact each other.
9    E. Power Off (1) InBody test is done, step down from the foot plate and the power will automatically be turned off after 60 seconds. Turn the dial to see the measurement results again in 60 seconds.   (2) If you step down the foot plate while measuring, the device will not be turned off for 60 seconds. If you step on the foot plate within 60 seconds the measurement will restart.
10  6. Maintenance      (1)  After  usage,  wipe  electrodes  with  a  wet  tissue  (also  known  as  wet  naps,  surface  wipes,  anti-bacterial tissues, etc.)    Wet tissues can be purchased at any local store.  (2)  Handle  the  arms  with  care.  Never  apply  excessive  force  near  the  hand  electrode  joint.  The damage caused by misuse may affect the function of the internal cable and electric board.  (3) Do not place anything on the platform or apply any pressure onto it when the InBody H20 is not in use.  (4) Do not drop any food or liquid on the equipment. It may affect the electrical parts in the equipment or cause damage.  (5) Once a week, wipe the exterior sides of the InBody H20 with a dry towel. In particular, clean the LCD monitor, using gentle care not to scratch the surface.  (6)  Follow  local  governing  ordinances  and  recycling  plans  regarding  the  disposal  or  recycling  of device components.
11  7. Understanding of Results    A. Result Screen During measurement, the InBody  H20  displays  information of  an  examinee’s  body composition on the LCD. The results are shown while you are on the foot plate.     ⓛWeight (kg)      Ideal weight is based on examinee’s height. Using the classical BMI method, the InBody 370 identifies the standard BMI as 22kg/m2 for males, 21.5kg/m2 for western females and 21kg/m2 for asian females.  Formula to get ideal weight Male Ideal Weight = Height2(㎡) × 22 Female Ideal Weight = Height2(㎡) × 21          (asian) Ideal Weight = Height2(㎡) × 21.5      (western) The standard range of weight is 85~115% based on the standard range of BMI.
12     Percentage of Body Fat (%)②  Percentage of Body Fat indicates the percentage of body fat to body weight. The ideal percentage of body fat is 15% for men and 23% for women, while the standard range of Body Fat Mass for men is 10-20% of the weight and 18-28% of the weight for women. In the case of children under the age of 18, children’s standard is used.        Ref. Samuel J. Fomon, et al. (1982): Body Composition of reference children from birth to age 10 years. The American Journal of Clinical Nutrition 35, 1169-1175.  ③ Skeletal Muscle Mass (kg)  Distinctively, skeletal muscle mass, which generally indicates the lean body mass of each arm and leg, can be controlled by exercise and dietary habits. Compare the bar graphs’ lengths of skeletal muscle mass with body fat mass. If the bar of skeletal muscle mass is relatively shorter and under the standard value, lean body mass lacks in the body, while the opposite case is proper. 100% signifies ideal lean body mass when examinee’s weight is normal. The standard range is 90~110% of standard skeletal muscle mass based on standard weight.
13    ④  BMI (Body Mass Index, kg/㎡) BMI is determined by using only weight and height and diagnoses superficial obesity. The standard values are 22kg/m2 for male and 21.5kg/m2 for western female and 21kg/m2 for asian female. Formula)    BMI = weight (kg) ÷ height2 (㎡)  Determination 1) WHO Standard     BMI(kg/㎡)  Classification  Diagnosis <18.5  Underweight  Under  Infectious disease, malnutrition related disease 18.5~24.9  Normal  Standard  Least risk at most disease 25.0~29.9  Overweight   Over  May cause health problem 30.0~34.9  Obese1 Increase of the risk of cardiac disease,   high blood pressure, diabetes, etc 35.0~39.9  Obese2 >40  Severely Obese Ref. WHO and the National Heart, Lung, and Blood Institute: clinical guidelines on the identification, evaluation, and treatment of overweight and obesity in adults, the evidence report. June 1998, xiv  Determination 2) Asian-Pacific Standard     BMI(kg/㎡)  Classification  Diagnosis <18.5  Underweight  Low (high risk of other clinical disease) 18.5~22.9  Normal  Average >23  Overweight   23~24.9  Risky Overweight  Increased 25.0~29.9  Obese step1  Moderate >30  Obese step2  Severe Ref. Korean Society for the Study of Obesity, chapter 2. Redefining and Evaluation, The Asian-Pacific perspective : Redefining Obesity and its Treatment:, 1st edition, Korean Society for the Study of Obesity, 2001, p10.
14  8. Error Messages  InBody  H20  helps  users  respond  to  possible  malfunctions  that  may  occur  by  displaying  error messages described below. This is the list of typical error messages and adequate responses.    A. Error 1.  Cause 1    Bad body posture due to poor contacts at hand or foot electrodes. Actuion1    See “Chapter 6 - C. How to Hold InBody Electrode” and correct the body posture.  B. Error 2.  Cause 1    Fundamental problem of the device suspected. Actuion1    Please contact Customer Support . (info@inbody.com)
15  9. One-Minute Check Before Calling for Service  This section lays out the order of steps you have to take in case of malfunction, with the assumption that  you  have  some  basic  knowledge  about  how  to  operate  the  equipment.  If  you  still  have  the problem after taking the following steps, contact the local distributor or Biospace.    A. I cannot turn on the device even though I press InBody or Weight button.  Cause 1  The batteries might have been fully discharged. Action 1   Replace the batteries.    Cause 2  User might have pressed                    too shortly. Action 2  Press and hold the button until the screen lights up.   B. Measurement data looks strange.     (Percent body fat, muscle mass, or other parameters are too high or too low.)  Cause 1   Poor contact between the body and the electrodes or incorrect body posture. Action 1   Please maintain correct body posture throughout the test.  Cause 2   Measured weight looks strange. Action 2    Please check if the hand grip is placed at right position. or    Please check Weight Offset in the configuration.  Cause 3   Personal information of others – ID, height, etc. – might have changed. Action 3   Please check ID, age, height, gender, and other personal information. If height information is wrong, please correct it.
16    C. Measurement data is not transferred.   (When the Bluetooth is connected, the data will be automatically transferred after the measurement.)  Cause 1  Poor connection of the Bluetooth dongle. Its connection can be checked by the Bluetooth icon. Action 1  Please check if there is a Bluetooth icon on the LCD and reconnect the Bluetooth dongle. Action 2  Please find the Bluetooth icon from Control Panel and double click, then below screen will appear. Please set as below picture and click “OK” button to save the set value.
17  Appendix    More about the InBody H20  A. How does BIA work? The  Bioelectrical  Impedance  Analysis  (BIA)  method  is  based  on  the  fact  that  the  human  body consists of conductors and non-conductors. Generally, 50~70% of the human body consists of water which functions as a conductor, whereas body fat functions as a non-conductor. The  classic  BIA  method  measures  the  impedance  of  the  whole  body  on  the  assumption  that  the human body can be considered a cylinder for the application of this model. If A is the cross sectional area, and L is the length, the impedance of the cylinder can be expressed as follows.    If both sides are multiplied by L, We get the new expression as follows.    According to this expression, if we know the L and the impedance value, we get the volume. That is to say, if we know the height of the human body (acting as a conductor), and know the impedance value,  we  can  get  the  volume  of  body  water.  Here,  the  volume  represents  examinee’s  height. Therefore, the two directly used variables in body composition analysis are impedance and height.  The  principle  of  the  InBody  H20’s  body  composition  analysis  is  explained  by  the  following;  the volume of body water, an electrolyte, is calculated first with a measured impedance value. Then, we can get the value of fat free mass using the volume of body water. Body fat mass is determined by deducting the lean body mass from the measured weight. Height should be entered by the user. Weight can be directly measured on the InBody H20.
18    B. Core Technology  The following are key features that make the InBody H20 extremely convenient, timely, and accurate.  (1) Tetrapolar 8-Point Tactile Electrode It  was  a  complex  and  inconvenient  procedure  to  attach  and  detach  the electrodes  to  a  specific  spot  every  time.  Trained  technicians  were  needed  for each  measurement.  The  InBody  H20  uses  tactile  electrodes  to  avoid  the possibility  of  errors  and  inaccuracies.  The  8-point  Tactile  Electrode  method enables the InBody H20 to efficiently produce accurate data every time.   (2) Segmental Analysis   There  are  some  claims  to  be  able  to  estimate  the  body  composition  separately;  there  is  no technology which can really measure it separately other than Biospace. Segmental measurement is the  technology  that assumes the body as  five  cylinders  of four  limbs  and trunk and measures the impedance  of  these  parts  separately.  Segmental  body  composition  analysis  provides  segmental measurement  of  body water, muscle mass,  and  fat  free  mass.  Furthermore,  the  analysis is  highly accurate  because the measured  value of a certain  part does not  affect the measurements of other segments. It is because body composition analyzers lack accuracy in measuring body fat and cannot figure out the patient’s exact shape that they must rely on empirical references to correct inaccurate measured values. But, the InBody with the technology of segmental analysis can exactly figure out difference by gender, aging, disease and ethnic without any empirical estimation. Based on the fact that  fat  free  mass  (FFM)  consists  of  about  73.3%  of  body  fluid,  it  can  be  concluded  that  the distribution  of  body  fluid  reflects  the  distribution  of  FFM.  Because  the  InBody  can  analyze  the segmental  body  fluid  distribution  (each  arm,  trunk,  and  each  leg  separately),  it  can  as  a  result examine a patients segmental development.
19    C. Classifications Type of protection against electric shock: Internally Powered Equipment Degree of protection against electric shock: Type B Applied Parts Degree of protection against the ingress of water: Ordinary equipment: IPX0 Equipment  not  suitable  for  use  in  the  presence  of  a  flammable  anesthetic mixture  with  air  or  with oxygen or nitrous oxide.   Mode of operation: Continuous operation Any  changes  or  modifications  in  construction  of  this  device  which  are  not  expressly  approved  by  the  party responsible for compliance could void the user’s authority to operate the equipment.       Contains transmitter FCC ID : QOQWT11  This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment  is    operated  in  a  commercial    environment.  This  equipment  generates,  uses,  and  can  radiate  radio frequency  energy  and,  if  not  installed  and  used  in  accordance  with  the  instruction  manual,  may  cause  harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful       interference in which case the user will be required to correct the interference at his own expense.
20    D. Specifications  Items & Standard  BODY COMPOSITION ANALYZER of Direct Segmental Multi-frequency Bioelectrical Impedance Method Bioelectrical Impedance(BIA) Measurement Items Bioelectrical Impedance(Z) 10 Impedance Measurements by Using 2 Different Frequencies (20 ㎑  and 100 ㎑) at Each of 5 Segments (Right Arm, Left Arm, Trunk, Right Leg and Left Leg) Electrode Method  Tetrapolar 8-Point Tactile Electrodes Measurement Method  Direct Segmental Multi-frequency Bioelectrical Impedance Analysis Method; DSM-BIA Method Body Composition Calculation Method  No Empirical Estimation   Outputs  Weight, Skeletal Muscle Mass, Percent Body Fat, BMI Applied Rating Current  450 ㎂ Power  DC 6V (1.5V X 4ea) AA Type Battery Dimensions  310.3(W)  ×  356.4(L)  ×  58.3(H) : mm Machine Weight  2.7kg (6lb.) Measurement Duration  5 seconds   Color  Soft White Operation Environment  10 ~ 40℃, 30 ~ 75%RH, 70 ~ 106kPa Storage Environment  -20 ~ 70℃, 10 ~ 95%RH, 50 ~ 106kPa (No condensation) Weight Range  10 ~ 200kg (22 ~ 440.9 lb.) Height Range  50 ~ 300cm (19.68 ~ 118.1 in.)   * Specifications may change without prior notice.
21    E. Customer Service Information   Corporate agents’ addresses are listed below. Contact us for assistance or more information.  Biospace Co., Ltd. 518-10, Dogok 2-Dong, Gangnam-gu, Seoul 135-784, Korea TEL : 82-2-501-3939 FAX : 82-2-578-2716 Homepage://www.inbody.com E-mail : info@inbody.com  Biospace Japan Inc. [JAPAN] Second Floor Ayabe Bldg., 2-17-3 Sotokanda, Chiyoda-ku, Tokyo JAPAN   TEL : 81-03-5298-7667   FAX : 81-03-5298-7668,7669 Homepage : http://www.biospace.co.jp E-mail : biospace@biospace.co.jp  Biospace Inc. [U.S.A] 8820 Wilshire Blvd. Suite 310 Beverly Hills, CA 90211 U.S.A   TEL : 1-310-358-0360 FAX : 1-310-358-0370 Homepage : http://www.biospaceamerica.com E-mail : USA@biospaceamerica.com  EU Representative. [Europe] DongBang Acuprime. Gater House, Gater Lane, Palace Gate, Exeter, EX1 1JL, UK TEL : 44-1392- 273908 FAX : 44-1392- 273909 E-mail : info@acuprime.com  Biospace China. [CHINA] Block A, 26/F, Hai Xing plaza, No. 1 South Ruijin Road, Shanghai 200023, CHINA   TEL : 86-21-64439738, 9739, 9705   FAX : 86-21-6443906 Homepage : http://www.biospacechina.com E-mail : admin@ biospacechina.com