Life Sensing Instrument TM8D Body worn medical telemetry transmitter User Manual TM 8 DUAL LEAD TRANSMITTER

Life Sensing Instrument Company Inc. Body worn medical telemetry transmitter TM 8 DUAL LEAD TRANSMITTER

Manual

Download: Life Sensing Instrument TM8D Body worn medical telemetry transmitter User Manual TM 8 DUAL LEAD TRANSMITTER
Mirror Download [FCC.gov]Life Sensing Instrument TM8D Body worn medical telemetry transmitter User Manual TM 8 DUAL LEAD TRANSMITTER
Document ID2992402
Application IDj4MN4+5nFkHAKzcruGWoFA==
Document DescriptionManual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize8.73kB (109108 bits)
Date Submitted2016-05-16 00:00:00
Date Available2016-05-16 00:00:00
Creation Date2017-12-05 18:15:58
Producing SoftwareGPL Ghostscript 9.18
Document Lastmod2017-12-05 18:15:58
Document TitleTM 8 DUAL LEAD TRANSMITTER
Document CreatorWriter
Document Author: Life Sensing

LIFE SENSING
INSTRUMENT CO., INC.
ENGINEERING APPROVAL
WORK INSTRUCTION
Document no. I-2018
LATEST REVISION DATE
10/22/13
DATE
10/22/13
TM8D DUAL ECG TRANSMITTER
FCC ID: OF8-TM8D
LIFE SENSING INSTRUMENTS, Inc.
329 WEST LINCOLN STREET
TULLAHOMA, TENNESSEE 37388
(931- 455-0262)
declares that the device manufactured as TeleMaster, model no. TM8D
does conform to the regulations as outlined in Title 47 of the US code of federal
regulations, Part 15 covering Class B personal computers and peripherals, and
Part 95, subpart H covering Wireless Medical Telemetry Service (WMTS).
Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause
undesired operation.
NOTE: This equipment has been tested and found to comply with the limits for a class B digital device,
pursuant to Parts 15 and 95 of the FCC rules. These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio
or television communications. However, there is no guarantee that the interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or television reception, which can be
determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one
or more of the following measures:
Reorient or relocate the receiving antenna of the interfered device. Increase the separation between the
equipment and interfered device.
 Connect the equipment to a different outlet on a circuit other than the one the interfered device is
connected to.
 Consult the dealer or an experienced radio/TV technician for help.

TM8D DUAL ECG TRANSMITTER
Page 1 of 4
LIFE SENSING
INSTRUMENT CO., INC.
ENGINEERING APPROVAL
WORK INSTRUCTION
Document no. I-2018
LATEST REVISION DATE
10/22/13
DATE
10/22/13
OPERATING INSTRUCTIONS
CAUTION: Changes or modifications not expressly approved by the manufacturer can void the user’s
authority to operate the equipment. It is the responsibility of the equipment user to maintain compliance
with the FCC rules. To ensure compliance, use only those cables and/or accessories prescribed by the
manufacturer.
IMPORTANT: To comply with the FCC RF exposure requirements, users should avoid grasping the
transmitter for any extended period of time while the device is in operation.
IMPORTANT - Operation of a TM8D requires prior coordination with the ASHE
frequency coordination designator. For further guidance visit the ASHE
website at:
www.wmtssearch.com
The TM8D Dual ECG Transmitter is a frequency synthesized transmitter operating in the UHF (608 614Mhz) range. The transmitter utilizes 5 independent patient lead wires with snap electrode
adapters for disposable electrodes.
The ECG viewed from ECG1 is dependent upon the placement of the White (-) and the Red (+)
electrodes. The ECG viewed from ECG2 is dependent upon the placement of the Black (-) and Brown
(+) electrodes. Both ECG leads are transmitted all the time. The Green lead serves as a reference to
all four ECG leads, which minimizes common mode interference.
Should an electrode become detached, a Lead Fault message will be displayed at the central station.
The central station will switch to the other lead.
A switch on the transmitter permits remote activation of the central station recorder.
The TM8D transmitter uses 2 each 1.5 volt “AAA” batteries. Average battery life is 2 to 3 days.
When the batteries reach an unsafe operating level, a Low Bat message will be displayed at the
central station. Approximately 1 hour battery life is left when the Low Bat message is first displayed.
Page 2 of 4
LIFE SENSING
INSTRUMENT CO., INC.
ENGINEERING APPROVAL
WORK INSTRUCTION
Document no. I-2018
LATEST REVISION DATE
10/22/13
DATE
10/22/13
The TM8D transmitter utilizes 5 electrodes to provide Lead II and a Modified Chest Lead (MCL) for
review at the central station. For optimum monitoring the electrodes should be placed in the following
locations:
WHITE: ECG1 negative electrode. RIGHT CLAVICLE OR RIGHT STERNAL BORDER.
RED: ECG1 positive electrode. LOWER LEFT RIB CAGE.
BLACK: ECG2 negative electrode. LEFT CLAVICLE OR LEFT STERNAL BORDER.
BROWN: ECG2 positive electrode. PLACE ON ANY OF 6 V-LEAD POSITIONS.
GREEN: Reference electrode. LOWER RIGHT RIB CAGE.
If only one lead of ECG is required remove the brown and black lead wires from the transmitter.
Refer to the section on MULTI-VIEW WAVEFORMS in the central station operating manual to disable
lead fault on second channel.
WARNING!
ONLY USE THE SAFETY LEAD WIRES SUPPLIED WITH THE TRANSMITTER.
Page 3 of 4
LIFE SENSING
INSTRUMENT CO., INC.
ENGINEERING APPROVAL
WORK INSTRUCTION
Document no. I-2018
LATEST REVISION DATE
10/22/13
DATE
10/22/13
Attaching the Transmitter to the Patient
Gently abrade the electrode sites with an alcohol prep pad until slight redness is observed. Dry the
site before applying the electrodes. Attach the electrodes to the lead wires.
Insert the batteries in the transmitter as follows:
Hold the transmitter in one hand. Using the other hand thumb and index finger, grasp the battery cap
and turn counter clockwise until the cap can be removed.
Insert both batteries with the positive ends up. See the battery direction indicator on the battery cap.
Repeat the above instructions for the battery cap, but rotate the cap clockwise.
CAUTION
MAKE SURE THE BATTERIES ARE INSTALLED CORRECTLY. IF THE BATTERIES
ARE INSTALLED INCORRECTLY, THE TRANSMITTER WILL NOT FUNCTION.
MAKE SURE THAT THE BATTERY CAP IS TIGHT. IF THE CAP IS NOT PROPERLY
TIGHTENED, THE TRANSMITTER MAY BE INTERMITTENT AND THE WATER
PROOFING SEAL MAY LEAK, CAUSING DAMAGE TO THE TRANSMITTER.
Attach the electrodes to the patient on the prepared sites. Apply pressure to the outer edges of the
electrode, not the center. Applying pressure to the center may cause the electrode conductive jelly to
ooze out resulting in a loose electrode.
REMOTE RECORD FUNCTION
If it is desirable to start the recorder from the transmitter, press and hold the Remote Record control
for approximately 2 seconds. This will generate a 20-second ECG recording at the central station.
Page 4 of 4

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : Yes
Encryption                      : Standard V4.4 (128-bit)
User Access                     : Print, Extract, Print high-res
Author                          : Life Sensing
Create Date                     : 2016:04:21 09:48:53-05:00
Modify Date                     : 2016:04:21 11:11:29-04:00
Language                        : en-US
XMP Toolkit                     : Adobe XMP Core 5.4-c005 78.147326, 2012/08/23-13:03:03
Format                          : application/pdf
Creator                         : Life Sensing
Title                           : TM 8 DUAL LEAD TRANSMITTER
Creator Tool                    : Writer
Metadata Date                   : 2016:04:21 11:11:29-04:00