Mcube Technology BIOCON900S CUBEScan Charger User Manual

Mcube Technology Co.,Ltd CUBEScan Charger

User Manual

 BioCon -900S OPERATOR’S    MANUAL
CUBEScanTM BioCon-900S Operator’s Manual   Page 2 This page left blank intentionally.
CUBEScanTM BioCon-900S Operator’s Manual   Page 3  BioCon-900S  Bladder Volume Measurement System  Operator’s Manual   CAUTION : In the United States, federal Law restricts this device to sale by or on the order of a physician. “This product is not made with natural rubber latex.”   The information in this document is subject to change at any time without notice. No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the  express written permission of  Mcube Technology. Examples  or  images  in  this  document  are  fictitious  and  do  not  in  any  way  represent  real patient  data.  For  up-to-date  user  information,  contact  your  local  distributor  or mcube@mcubetech.co.kr.  All other product or service names are the property of their respective owerner..         Mcube Technology Co., Ltd. #803 123, Bonghwasan-ro, Jungnang-gu, Seoul, Korea  Tel: +82-2-3421-7780 Fax: +82-2-3421-7076 E-mail: mcube@mcubetech.co.kr Web site: www.mcubetech.co.kr           Copyright  ⓒ 2018 by Mcube Technology Co.,Ltd. All rights reserved.
CUBEScanTM BioCon-900S Operator’s Manual   Page 4 TABLE OF CONTENTS  1 GENERAL INFORMATION ........................................................................................ 6 1.1 PRODUCT DESCRIPTION ............................................................................................................. 6 1.2 INTENDED USE / INDICATIONS FOR USE ................................................................................... 6 1.3 PATIENT/USER CHARACTERISTICS .......................................................................................... 6 1.4 PRESCRIPTION STATEMENT....................................................................................................... 6 1.5 SERVICE ........................................................................................................................................ 6 2 SAFETY INFORMATION ........................................................................................... 8 2.1 NOTICE TO ALL USERS................................................................................................................ 8 2.2 CONTRAINDICATIONS ................................................................................................................. 8 2.3 BIOLOGICAL SAFETY .................................................................................................................. 8 2.4 ELECTRICAL SAFETY .................................................................................................................. 8 2.5 DEVICE SAFETY............................................................................................................................ 9 2.6 SAFE HANDLING PROCEDURES FOR TRANSPORTER ......................................................... 11 3 INTRODUCTION ...................................................................................................... 12 3.1 PRODUCT FEATURES ................................................................................................................ 12 3.2 SYSTEM COMPONENTS ............................................................................................................ 12 3.3 OUTER APPEARANCE – PROBE ............................................................................................... 13 3.4 OUTER APPEARANCE - CHARGER .......................................................................................... 14 3.5 DISPLAY AND ICONS .................................................................................................................. 15 3.6 ICON DIRECTORY ....................................................................................................................... 15 4. SETUP ........................................................................................................................ 18 4.1 ACCURACY OF MEASUREMENT .............................................................................................. 18 4.2 CHARGE THE BATTERY ............................................................................................................. 18 4.3 SCAN TYPE .................................................................................................................................. 19 5 HOW TO USE .......................................................................................................... 20 5.1 CHECKUP BEFORE USE ............................................................................................................ 20 5.2 TURN SYSTEM ON/OFF .............................................................................................................. 20 5.3 SELECT PATIENT TYPE .............................................................................................................. 21 5.4 RECORD INFORMATION FOR A PATIENT ................................................................................. 22 5.5 MEASURE BLADDER VOLUME ................................................................................................. 24 5.6 DISPLAY A SCAN RESULTS ....................................................................................................... 26 5.7 RE-AIM BLADDER ....................................................................................................................... 28 5.8 SAVE A SCAN RESULT ............................................................................................................... 28 5.9 FETCH A MISSING DATA ............................................................................................................ 29 5.10 FINISH THE MEASUREMENT ..................................................................................................... 29 5.11 MANAGE SAVED SCAN RESULTS ............................................................................................ 29 6 SCREENS ................................................................................................................ 32 6.1 HOME SCREEN ........................................................................................................................... 32 6.2 INFORMATION SCREEN ............................................................................................................. 32 6.3 BARCODE SCREEN (OPTIONAL) .............................................................................................. 33 6.4 VIRTUAL KEYBOARD SCREEN ................................................................................................. 33 6.5 NORMAL SCAN SCREEN ........................................................................................................... 34 6.6 PRE-SCAN SCREEN ................................................................................................................... 34 6.7 SCAN RESULT SCREEN ............................................................................................................. 35 6.8 REVIEW SCREEN ........................................................................................................................ 35 6.9 SETUP SCREEN .......................................................................................................................... 36 6.10 BATTERY STATUS ....................................................................................................................... 37 6.11 DIALOG BOXES ........................................................................................................................... 38 7 SETTINGS ............................................................................................................... 39 7.1 DATE FORMAT SETTINGS .......................................................................................................... 39 7.2 DATE SETTINGS .......................................................................................................................... 39
CUBEScanTM BioCon-900S Operator’s Manual   Page 5 7.3 TIME SETTINGS ........................................................................................................................... 39 7.4 SLEEP MODE SETTINGS ............................................................................................................ 40 7.5 DISPLAY BRIGHTNESS SETTINGS ........................................................................................... 40 7.6 PRE-SCAN SETTINGS ................................................................................................................ 41 7.7 DELETE ALL MEASUREMENTS ................................................................................................ 41 8 MAINTENANCE ....................................................................................................... 43 8.1 TROUBLESHOOTING ................................................................................................................. 43 8.2 DIALOG BOXES FOR MAINTENANCE....................................................................................... 43 8.3 CLEANING & DISINFECTION ..................................................................................................... 44 8.4 BATTERY MAINTENANCE .......................................................................................................... 46 8.5 WEEKLY INSPECTION ................................................................................................................ 46 8.6 DISPOSAL .................................................................................................................................... 46 8.7 DEVICE REPAIR .......................................................................................................................... 46 8.8 SELF TEST ................................................................................................................................... 47 8.9 CALIBRATION ............................................................................................................................. 48 9 SPECIFICATIONS ................................................................................................... 51 9.1 ACOUSTIC OUTPUT TABLE ....................................................................................................... 51 9.2 ELECTROMAGNETIC COMPATIBILITY ..................................................................................... 51 9.3 EXCESSIVE TEMPERATURE TEST RESULT ............................................................................ 54 9.4 SPECIFICATIONS OF COMPONENTS ....................................................................................... 54 9.5 ENVIRONMENTAL CONDITIONS ............................................................................................... 56 10 GLOSSARY ............................................................................................................. 58 11 SYMBOL DIRECTORY ............................................................................................ 59 12 REFERENCES ......................................................................................................... 61
CUBEScanTM BioCon-900S Operator’s Manual   Page 6 1  GENERAL INFORMATION  1.1  PRODUCT DESCRIPTION CUBEScanTM BioCon-900S is a safe and easy, non-invasive system to measure the bladder volume. The device  consists  of  a  probe,  CUBEScan  Charger,  and  various  components.  The  probe  is  a  B-mode instrument, hand-held, wireless and battery-operated. A 3D-mechanical sector transducer provides cross–sectional images of the bladder from up to 12 scan planes and bladder volume is calculated based upon those images and displays 12 scan planes on a screen. Furthermore, a live image of the bladder during Pre-scan  makes  it  easier  to  detect  the  bladder  before  scanning.  The  CUBEScan  Charger  is  used  for recharge of the probe’s internal battery.   1.2  INTENDED USE / INDICATIONS FOR USE CUBEScanTM  BioCon-900S  is  a  B-mode  pulsed-echo  ultrasound  device.  The  BioCon-900S  projects ultrasonic energy through the lower abdomen of a patient to obtain images of the bladder to calculate the urine volume non-invasively.  BioCon-900S  is intended to  be used  by a  qualified  medical professional to non-invasively measure the urine volume in the bladder. Contraindications for the BioCon-900S are fetal use and use on pregnant patients.   1.3  PATIENT/USER CHARACTERISTICS 1.3.1  PATIENT POPULATION 1) Male (All other patients) 2) Female patient - A female who has not had a hysterectomy - A female who has had a hysterectomy 3) Pediatric patient; A child with height less than 47 inches (120cm) and weight less than 55 lbs (25kg)  1.3.2  USERS 1) Physicians 2) Medical Professionals  1.3.3  MEDICAL CONDITIONS INCLUDING TARGET GROUP AND DISEAS Bladder volume measurement system measures bladder volume, and it provides the accurate data to aid in the diagnosis of common urological condition, to assess urinary retention, to help unnecessary catheterization and to reduce rates of catheter-associated urinary tract infection.  The benefits are as below; BioCon-900S assists to; a. Assess urinary retention b. Evaluate post-operative urinary retention (POUR) c. Reduce unnecessary catheterization d. Reduce catheter-associated urinary tract infections(CAUTIs) e. Identify a blocked Foley catheter f. Evaluate the need to catheterize after Foley catheter removal g. Evaluate the need to catheterize during intermittent catheterization h. Support to train urination and self-cathererization i. Measure PVR (Post Void Residual)  1.4  PRESCRIPTION STATEMENT Federal (United States) law restricts this system to sale by or on the order of a physician.  1.5  SERVICE
CUBEScanTM BioCon-900S Operator’s Manual   Page 7 If  you  encounter  difficulty  with  the  system,  please  contact  local  distributers  or  Mcube  Technology  at mcube@mcubetech.co.kr
CUBEScanTM BioCon-900S Operator’s Manual   Page 8 2  SAFETY INFORMATION  2.1  NOTICE TO ALL USERS This  guide  covers  components,  function,  maintenance,  storage,  and  precautions  needed  to  use  this system. All users must read and thoroughly understand this entire guide prior to using the  BioCon-900S. This section has information on safe use of the BioCon-900S (Biological Safety, Electrical Safety, Device Safety and Safe handling procedures for transporter).   Complies  with  FDA  performance  standards  for  laser  products  except  for  deviations  pursuant  to  Laser Notice No. 50, dated June 24, 2007.   2.2  CONTRAINDICATIONS   WARNING Do not use the BioCon-900S on following cases: 1) Fetal use or pregnant patients.   2) Patients with ascites. 3) Patients with open or damaged skin. Wounds in the suprapubic region.   2.3  BIOLOGICAL SAFETY To  date,  exposure  to  pulsed  diagnostic  ultrasound  has  not  been  shown  to  produce  adverse  effects. However,  ultrasound  should  be  used  only  by  medical  professionals  when  clinically  indicated,  using  the lowest possible exposure times possible commensurate with clinical utility. The ultrasonic output power of BioCon-900S  is  limited  to  the  minimum  level  necessary  for  performance  effectively.  Data  on  acoustic output levels is listed in section “9. Specifications”.     2.4  ELECTRICAL SAFETY This system meets IEC 60601-1, Class I, Type BF isolated patient-applied parts safety requirements. This system  complies  with the  applicable  medical  device  requirements  published  in  the  Canadian  Standards Association  (CSA),  European  Harmonized  Standards,  and  Underwriters  Laboratories  (UL)  safety standards. To avoid the risk of electrical shock or injury, observe the following warnings and cautions.   WARNING   To  avoid  the  risk  of  electric  shock,  this  device  must  only  be  connected  to  an  adapter  with  the protective  earth.  -  Grounding  reliability  can  only  be  achieved  when  device  is  connected  to  a receptacle marked “Hospital Only” or “Hospital Grade” or the equivalent.  Do not operate the system in the presence of flammable gases or anesthetics.  Do not use the device with any defibrillator at the same time.  Do not use the device with any HF surgical equipment at the same time.  Before using the system, inspect the enclosures, power cord and adapter.Do not use the system if these are damaged.  Do not use the probe if the entire probe is accidentally immersed in solution.  Connect only items that have been specified as part of the BioCon-900S or that have been specified as being compatible with the BioCon-900S.  Charge the probe only with the provided Charger.    Connect the Charger only to the provided adapter.
CUBEScanTM BioCon-900S Operator’s Manual   Page 9  Do not attempt to open the system components. This may cause serious  injury to the operator  or patients. All services must be made by a qualified technician only.    Use of components and cables other than those specified or provided by the manufacturer of  this device could result in increased electromagnetic emissions or decreased electromagnetic immunity of this device and result in improper operation.  Portable RF communications equipment (including peripherals such as antenna cables and external antennas)  should  be  used  no  closer  than  30  cm  (12  inches)  to  any  part  of  the  BioCon-900S, including cables specified  by the manufacturer.  Otherwise, degradation of the performance  of  this device could result.    CAUTION   The BioCon-900S has been tested and found to comply with the electromagnetic compatibility (EMC) limits for medical systems to IEC 60601-1-2. These limits are designed to provide reasonable protection against  harmful  interference  in  a  typical  medical  installation.  Medical  systems  must  be  installed  and operated  according  to  the  instruction  in  this  manual.  For  more  information,  see  the  section  “9.2 ELECTROMAGNETIC COMPATIBILITY”.    This device is suitable for use in the professional health care environment  Medical electric devices can be affected by portable or mobile RF communication devices. Turn off any portable or mobile RF device before operating your system.    Electrostatic discharge (ESD), or static shock, is a naturally occurring phenomenon. ESD is common in conditions of low humidity, which can be caused by heating or air conditioning.  Static shock is a discharge of electrical energy from a charged body to a lesser or non-charged body. The degree of discharge can be significant enough to cause damage to a transducer or an ultrasound system. The following precautions can help reduce ESD: anti-static spray on carpet, anti-static spray on linoleum, and anti-static mats.  The EMISSIONS characteristics of this device make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this device might not offer  adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the device.  Do not  use the system if  an error  message  appears  on the  display: note the  error code;  call Mcube technology or your local distributor, shut down the system   2.5  DEVICE SAFETY To protect your BioCon-900S, and components, follow these warnings and cautions.      WARNING   No modification of this device is allowed.  Do not use the system if the head of probe is damaged.    To avoid the risk of excessive heating or damage to the device, use the device in a well-ventilated environment.  To avoid damage to cables, do not excessively twist or bend cables associated with the device.  Turn off the device and unplug the charger if it will not be used for a long period of time.  Do not use the system if there is evidence of leakage of internal liquids. Wash hands immediately in warm, soapy water.    In the event that LCD is damaged, care should be taken to avoid contact with liquid crystal. Take the urgent action indicated should any of the following situations arise: -  If  liquid  crystal  comes  in  contact  with  your  skin,  clean  the  area  with  a  cloth  and  then  wash thoroughly with soap and running water.
CUBEScanTM BioCon-900S Operator’s Manual   Page 10 - If liquid crystal enters your eyes, flush the affected eye with clean water for at least 15 minutes and then seek medical assistance. - If liquid crystal is swallowed, rinse your mouth thoroughly with water. Drink large quantities of water and induce vomiting, then seek medical assistance.    CAUTION   Use the system indoors only.  Do not use the system if it exhibits erratic or inconsistent behavior. Shut down the system and contact Mcube Technology or your local distributor.  Do not spill liquid on the system.   2.6  COMPLIANCE STATEMENTS Federal Communication Commission Interference Statement   This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference  when  the  equipment  is  operated  in  a  commercial  environment.  This  equipment  generates, uses,  and  can  radiate  radio  frequency  energy  and,  if  not  installed  and  used  in  accordance  with  the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.”  This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:   (1) This device may not cause harmful interference, and (2)  this  device  must  accept  any  interference  received,  including  interference  that  may cause  undesired operation.  FCC Radiation Exposure Statement:   This equipment complied with FCC radiation exposure limits set forth for an uncontrolled environment. This equipment  should  be  installed  and  operated  with  minimum  distance  20cm  between  the  radiator  &  your body.   CAUTION   Any  changes  or  modifications  not  expressly approved by the  party responsible  for  compliance  could void the user’s authority to operate this equipment.   IC Statement  This Class A digital apparatus complies with Canadian ICES-003.  This device complies with Industry Canada license-exempt RSS standard(s).  Operation is subject to the following two conditions:   (1) this device may not cause interference, and   (2) this device must accept any interference, including interference that may cause undesired operation of the device.  Cet appareil numérique de la classe A est conforme á la norme NMB-003 du Canada.
CUBEScanTM BioCon-900S Operator’s Manual   Page 11  2.7  SAFE HANDLING PROCEDURES FOR TRANSPORTER Quarantine: Packages that are crushed, punctured or torn open to reveal contents should not be placed into  deployment.  Such  packages  should  be  isolated  until  the  shipper  has  been  contacted,  provides disposition instructions and, if appropriate, arranges to have the product inspected and repacked.  Spoiled  Product:  In  the  event  that  damage  to  packaging  results  in  damage  to  the  battery  causing released electrolyte, the spill should be contained and the shipper should be contacted for instructions.
CUBEScanTM BioCon-900S Operator’s Manual   Page 12 3  INTRODUCTION  3.1  PRODUCT FEATURES The main components of CUBEScanTM BioCon-900S are a Probe and the Charger.      BioCon-900S Probe & Charger    Hand-held, portable and lightweight  Touchscreen interface  Measures bladder volume with ultrasound    Provides a live image of bladder during Pre-scan  Takes scans and provides scan results in a short time    Easy to use: staff can easily learn how to scan patients in a short time    Battery-operated, wireless charging Figure 1 CUBEScanTM BioCon-900S     3.2  SYSTEM COMPONENTS  No. Picture Parts Q’ty Description 1  BioCon-900S (Probe) 1 Hand-held,  wireless,  battery-operated, touchscreen, ultrasound probe 2  Charger 1 Use  the  Charger  to  charge  BioCon-900S’s internal battery. 3  Mobile Cart (MRC-900S) 1 Mobile cart (Optional Component) 4  AC/DC Adapter 1 Use only with the Charger. 5  Power cord 1 The type of power cord depends on your country.
CUBEScanTM BioCon-900S Operator’s Manual   Page 13 6 - Barcode Module 1 Reading a barcode with information for a patient. (Optional Component) 7 - Operator’s manual 1 -  Users can replace only gel  The other parts are not allowed for the users to disassemble or replace except the gel. Please contact your local distributor or Mcube Technology.    Only use the ultrasonic gel approved by FDA, CE.   3.3  OUTER APPEARANCE – PROBE     Figure 2 Probe Appearance No. Part Function 1 Touchscreen   Display Displays bladder volume and other information.   2 Power Button Press to turn the probe on or off and release 3 Scan Button Press to take a scan 4 Probe Head Generates, transmits and receives ultrasound waves, produces a 3D image of bladder. Type BF applied part 5 Power Indicator Indicates power status. Orange solid – System is being booted Yellow solid – Power on status Off – Power off
CUBEScanTM BioCon-900S Operator’s Manual   Page 14   3.4  OUTER APPEARANCE - CHARGER   6 Battery indicator Indicates charging status. Orange solid - Battery is being charged.   Off – Battery is fully charged. 7 Barcode Module Reads printed a barcode 8 Laser Safety Label Explanatory label – Class 1 laser product 9 Label Put on this label when there is no optional barcode module 10 Back Label NOTE : Labels are subject to change without prior notice  NOTE: The appearance of these labels(No.8 - 9) depends on wheter the barcode option is selected or not.      Figure 3 Charger Front-View   Figure 4 Charger Bottom-View   No. Item Function 1 Battery indicator Indicates charging status. Green (Solid) – Power cord is connected or Charging is completed. Green (Blink) – Charging.in progress Yellow (Solid) – Error Status 2 Silicone pad Anti-skid pad 3 Insert Nut Insert  nut  to  secure  the  Charger  to  the  mobile  cart  (MRC-900S). It prevents movement or damage of Charger from shock. 4 Label NOTE : Labels are subject to change without prior notice 5 AC/DC Adapter Inlet AC/DC adapter inlet 6 Vents Ventilate the Charger.
CUBEScanTM BioCon-900S Operator’s Manual   Page 15 3.5  DISPLAY AND ICONS A touch-screen display allows you to perform scans, manage results and setup. The Home screen appears when the probe is turned on. It services as a starting point for all the main functions of the device. You can select the desired icons by tapping them directly on the touch-screen.         Figure 5 Home Screen   3.6  ICON DIRECTORY  Patient Type  Home screen  Information screen & Pre-scan screen No. Description 1 Status  –  Current  time,  the  number  of  saved  data and battery status. 2 Main Display – controls the user interface   Icon Description  Male (All other patients)  Female Patient – A female who has not had a hysterectomy  Female Patient – A female who has had a hysterectomy  Pediatric Patient – A child with height less than 47 inches (120cm) and weight less than 55 lbs(25kg). Icon Description  Review – Go to the Review screen.  Missing Data – Go to review the missing (unsaved) scan result.  Setup – Go to the Setup screen.  Battery status.
CUBEScanTM BioCon-900S Operator’s Manual   Page 16  Scan Result screen  Review screen  Setup screen Icon Description  Barcode - Record a barcode (Optional)  Home – Return to the Home screen.    Scan – Perform the Normal scan or Pre-scan.    Exit –Return to the Information screen.   Icon Description  Aiming Information – Rotate 15° clockwise, indicating the aiming point of the bladder.  Urine Volume – Measured urine volume. P1 Plane No.  –  The  1st  plane  of  12  planes  of  the  result  (PX  –  The  ordinal  number  of  12 planes)  Scan – Re-scan the bladder.  Home – Return to the Home screen.    Save – Save the scan result. Icon Description  Delete – On the Review screen, delete what is currently being displayed.  OK – Select the option or perform the action.  Cancel – Cancel the current function.  Previous – On the Review screen, move to the previous patient scan result.  Next – On the Review screen, move to the next patient scan result.  Exit –Return to the Home screen.   Icon Description  Date format (YY-MM-DD, DD-MM-YY,MM-DD-YY)    Date    Time (format : 24-hour)
CUBEScanTM BioCon-900S Operator’s Manual   Page 17      Sleep mode (2min, 3min, 5min)    Display brightness (0 to 100%, default= 50%))  Pre-scan On ( ) / Off ( )  Delete all saved results.    Calibration Reminder On ( ) / Off ( )  Self-test (YY/MM/DD)  Phantom Calibration (YY/MM/DD)  CalKit Calibration (YY/MM/DD)  On the self-test or calibration screen, perform self-test or calibration  Cease – On the self-test or calibration screen, cease self-test or calibration  Exit – Return from the self-test or calibration screen to the Setting screen.    Previous – Move to previous page.  Next – Move to next page.  Home – Return to Home screen.  OK – Select the option or perform the action.  Cancel – Cancel the current function.
CUBEScanTM BioCon-900S Operator’s Manual   Page 18 4. SETUP   See the section “4.1 ACCURACY OF MEASUREMENT” before use.    Check  the  probe’s  battery  has  sufficient  power  through  battery  icon.  If  the  battery  has  not  been sufficiently charged, recharge the battery before bladder measurement.    Wipe the probe head gently with a soft cloth dampened in cleaning and disinfection agents to clean and disinfect the probe. See the section “8.3 CLEANING & DISINFECTION”.   4.1  ACCURACY OF MEASUREMENT Following cases affect ultrasound transmission and accuracy of measurement.   When a patient has had supra-pubic or pelvic surgery.  A patient with catheter in his or her bladder.  A patient with scar, sutures, staples or incisions in his or her abdomen.  Use of unsuitable ultrasound gel .  Air bubbles between probe head and the skin of the patient.  An excessive obese patient.  Use of the probe with an unclean probe head.  Inappropriate patient position.   4.2  CHARGE THE BATTERY       CAUTION   The  Charger,  power  cord,  and  adapter  are  not  intended  for  patient  contact.  Ensure  6  ft  (2  m)  are maintained between the patient and these components.  Immediately discontinue use of  the battery,  if  the  battery emits an unusual smell, feels hot,  changes color  or  shape,  or  appears  abnormal  in  any  other  way.  Contact  Mcube  Technology  or  your  local distributors.  The probe is powered by a internal lithium ion battery. The battery is charged by Charger and the Battery   WARNING   The battery must be replaced by the authorized service provider only.    Use only the provided Charger and power cord to charge the battery for electric safety.    Do not disassemble, heat or incinerate the battery.  Keep battery out of reach of children and in original sealed package.  Dispose of used batteries promptly according to local recycling or waste regulations.    To reduce the risk of leakage, explosion, fire, or serious injury, note the following when handling the lithium-ion battery included in the system: - Never short-circuit the battery by either accidentally or intentionally bringing the battery terminals into contact with any other conductive objects. This could cause serious injury or fire and could also damage the battery and BioCon-900S.   -  Never  expose  the  battery  to  abnormal  shock,  vibration  or  pressure.  The  battery’s  internal protective  covering  could  fail,  causing  it  to  overheat  or  ignite,  resulting  in  caustic  liquid  leakage, explosion or fire.  Avoid charging near fire or in the sunlight.    Do not use the Charger for other equipment.
CUBEScanTM BioCon-900S Operator’s Manual   Page 19 Status  icon  (    )  is always  displayed  in  the  Status  at  the top  of  the screen,  indicating the  current capacity. Before using the probe for the first time, you must fully charge the battery. When you don’t use your  probe,  we  recommend  that  you  mount  and  store  it  on  the  Charger  to  ensure  that  the  probe  is sufficiently  charged.  The  battery  is  protected  by  protection  circuit  module  and  have  the  functions  of overcharge,  overdischarge  and  overcurrent  prevention  to  maintain  safety  and  prevent  significant deterioration of cell performances.    To conserve battery life, the probe is configured to go into sleep mode. The probe goes into sleep mode when you press the Power Button( ) twice or if the probe has not been touched in the designated time with power on. You can adjust the time for sleep mode in the setup(2min/3min/5min), if necessary. See the section“7.4 SLEEP MODE SETTING”. Shortly press the Power Button( ) to wake up the system.       Figure 6 Probe-mounting Figure 7 Probe Indicatior Figure 8 Charger Indicator  1) Connect the AC/DC adapter to the power cord. 2) Connect the AC/DC adapter to the Charger   3) Plug the AC/DC adapter into an electric outlet 4) The battery indicator (3) on the Charger lights on in steady green (Power on).   5) Mount the probe on the Charger. Make sure that the upper part of probe reaches the wall of Charger. See Figure 5. 6) As soon as the probe is mounted on the Charger, the beep and light indicators work as below; a) A short beep: A connection for the probe’s battery charging is completed.     b) The battery indicator (3) on the Charger lights on in blinking green(Charging in progress).   c) The battery indicator (2) on the probe lights on in steady orange (Charging in progress).  7) The Power indicator (1) is turned on in steady yellow when the probe is ready to use. 8) It takes approximately 6 hours for the discharged battery to fully charge. When fully charged, the battery indicator(2) on the probe turns off.   4.3  SCAN TYPE You can adjust one of the following options to measure bladder volume:  Pre-scan: General 2D real-time ultrasound scanning. 2D ultrasound images are displayed continuously in live image basis. This helps an operator locate bladder position and predicting the range of residual urine to detect the bladder before normal scan.  Normal  Scan:  Get  12-plane  ultrasound  images  and  calculate  the  residual  urine  in  the  bladder  (3D scanning).
CUBEScanTM BioCon-900S Operator’s Manual   Page 20 5  HOW TO USE 5.1  CHECKUP BEFORE USE Before using BioCon-900S, ensure you are familiar with the device and its parts. If you use it for the first time,  we  strongly  recommend  you  to  try  your  first  bladder  volume  measurement  on  a  patient  with moderately full bladder using Pre-scan. Pre-scan helps you detect the bladder and moderately full bladder can be located more easily.        CAUTION  If the oil leaks from housing and fluid gets in contact with your eyes, skin or clothing, flush the affected area immediately with clean water and seek medical attention or see a doctor.   5.2  TURN SYSTEM ON/OFF There  are  two power  modes  to system  on/off  the BioCon-900S. Loading  screen  depends  on  the  power mode and it shows during start-up.    Figure 9 Loading Screen (1)   Figure 10 Loading Screen (2)   Figure 11 Home Screen  5.2.1  SLEEP MODE   To preserve battery power, the probe goes into sleep mode by shutting itself down automatically when not in use.  You can set the time to go  into sleep mode in the setup (2min/3min/5min). See  the section “7.4 SLEEP MODE SETTING”.   However, if a battery needs to be recharged immediately, Dialog box ( ) appears shortly and then probe goes to sleep mode by shutting itself down. We strongly recommend that you should recharge the battery immediately. See the section “4.2 CHARGE THE BATTERY”.   Turn System on  WARNING   If you use BioCon-900S in the presence of flammable anesthetics, explosion hazard exists.    Do not use the BioCon-900S on following cases: 1) Fetal use or pregnant patients.   2) Patients with ascites. 3) Patients with open or damaged skin. 4) Wounds in the suprapubic region.
CUBEScanTM BioCon-900S Operator’s Manual   Page 21 1) Press the Power Button ( ) shortly to wake the probe from sleep mode. 2) Only Loading screen (2) appears on the display. 3) Home screen appears when the system is fully loaded at start up. Turn System off 1) Press the Power Button ( ) shortly twice to go into sleep mode.  5.2.2  SHUTDOWN MODE The probe draws little power when it is gone into sleep mode. To preserve battery power, the probe shuts down the system in approximately 24 hours after going into sleep mode, if there is no activity. However, if you do not plan to use the probe for more than a week, you should shut the system down completely to prevent it from discharging completely  Turn System on 1) Press the Power Button ( ) shortly to wake the probe from shutdown mode. 2) Loading screen (1) appears on the display and Loading screen (2) follows shortly. 3) Home screen appears when the system is fully loaded at start up  Turn System off 1) Press the Power Button( ) for 3~4 seconds. 2) A Shutdown dialog box appears on the Home screen. 3) Ths system shut down shortly after Shutdown screen appears.     Figure 12 Shutdown Dialog Box Figure 13 Shutdown Screen   5.3  SELECT PATIENT TYPE BioCon-900S has several patient types. Tap the Patient icon on the Home screen to select the patient type.    1)    Male (All other patients) 2)    Female patient – Select if you are scanning a female patient. If you select the female patient type on the Home screen, the BioCon-900S allows you more granularity in chosing an female patient type on the Information screen. Select female patient by tapping the (1) of Figure 23. - A female who has not had a hysterectomy ( ) - A female who has had a hysterectomy ( ) 3)    Pediatric patient – Select if you are scanning a child with height less than 47 inches (120cm) and weight less than 55 lbs. (25kg).
CUBEScanTM BioCon-900S Operator’s Manual   Page 22      Figure 14 Patient Type    Figure 15 Female Patient Type     5.4  RECORD INFORMATION FOR A PATIENT 5.4.1  INPUT INFORMATION – VIRTUAL KEYBOARD The Information screen appears when you tap one of the Patient icon on the Home screen. You can tap the  text  box  and  update  information  for  a  patient  by  typping  the  virtual  keyboard.  The  virtual  keyboard screen  appears  depend  on  the  text  box  you  select.  See  the  section  “6.4  VIRTUAL  KEYBOARD SCREEN”on for information about the function of the virtual keyboard.    You  can  scan  a  bladder  volume  without  entering  patient  information.  If  you  do  not  enter  a  patient information,  the  patinet  tag  content(2)  is  empty  on  the  Scan  Result  screen.  Scanned  date  and  time(1) automatically display and it is based on your probe’s time settings. Once a scan result is saved, you cannot add or modify patient information.    You can measure bladder volume by tapping the Scan icon( ) on the Information screen or pressing the Scan Button . If you want to return to the Home screen, tap the Home icon( )      Figure 16 Information Screen   Figure 17 Scan Result Screen   Patient Tag   Tap Patient Tag text box, and then enter the patient’s tag by typping the virtual keyboard
CUBEScanTM BioCon-900S Operator’s Manual   Page 23 First Name   Tap First Name text box, and then enter the patient’s first name by typping the virtual keyboard Sur Name   Tap Sur Name text box, and then enter the patient’s sur name by typping the virtual keyboard  If you would like to the change the virtual keyboard type, tap Keyboard icon ( ) in the left corner of the virtual keyboard. The icon allows you to select an numeric, uppercase or lowercase virtual keyboard.      Figure 18 Numeric Key   Figure 19 Letter Key (Lowcase)   Figure 20 Letter Key (Uppercase)     Date of Birth   Tap Date of Birth text box, and then enter the patient’s birthday by typping the virtual keyboard   Figure 21 Date Key  5.4.2  INPUT INFORMATION – OPTIONAL BARCODE The CUBEScanTM BioCon-900S can attach an optional barcode module. Installing the optional barcode module on the device allows you to record a patient tag as necessary. Skip Skip this section if you do not enter patient information through a barcode or if you do not have a barcode reader.
CUBEScanTM BioCon-900S Operator’s Manual   Page 24   WARNING  “LASER RADIATION – AVOID DIRECT EXPOSURE TO BEAM”  Do not capture human being with the barcode module.  Hold the probe with both hands tightly, when you read barcodes with patient information.  Recording barcodes is not available, after the scan result is saved.        Figure 22 Information Screen Figure 23 Barcode Scan Screen  How to read a barcode 1) Tap the Barcode icon( ) on the Information screen. 2) Move the aiming position to the center of the barcode and press the Scan Button   with thumb.   3) A red light guide is emitted.   4) Keep pressing the Scan Button   and do not release until getting barcodes.   5) The barcode reader decodes continuously for a good barcode scan.   6) You have a good barcode with a short beep. The barcode numbers is displayed on Patient Tag of the Information screen.   <good>  <bad>   5.5  MEASURE BLADDER VOLUME 5.5.1  NORMAL SCAN The normal scan is executed only when Pre-scan setup is ‘Off( )’ on the Setup screen. See the section “7.7 PRE-SCAN SETTING” 1) The patient should be lying in a supine position and lift up the patient garment exposing the abdominal region from the pubis to the navel. 2) Place  an  ample  amount  of  gel  without  air  bubbles  to  the  probe  head.  In  order  to  assure  optimal transmission of  energy  between the  patient  and the  probe,  a  conductive  gel  must  be  applied on  the probe head. 3) Place the probe head on the patient’s abdomen with the screen facing upwards and aim it to the location where the bladder is expected to be above 3~4cm the pubic bone.   4) Ensure that the probe head is pointing toward the patient’s foot.
CUBEScanTM BioCon-900S Operator’s Manual   Page 25   Figure 24 Positioning Bladder Figure 25 Probe Direction 1) Tap the Scan icon( ) on the Information screen or press the probe Scan Button . The scanning process begins. 2) The  progress bar  on the  screen  appears.  It takes 3  seconds  to complete  the  normal scan. Hold  the probe  steady  while  scanning.  Movement  of  the  probe  during  scanning  reduces  the  accuracy  of measurement. 3) When you hear the beeps, the scan is complete.    Figure 26 Normal Scan Process  5.5.2  PRE-SCAN The Pre-scan is executed only when Pre-scan setup is ‘ON( )’ on the Setup screen. When Pre-scan is ‘Off()’, a normal scan is executed. See the section “7.7 PRE-SCAN SETTING”  1) The patient should be lying in a supine position and lift up the patient garment exposing the abdominal region from the pubis to the navel. 2) Place  an  ample  amount  of  gel  without  air  bubbles  to  the  probe  head.  In  order  to  assure  optimal transmission of  energy  between the  patient  and the  probe,  a  conductive  gel  must  be applied  on  the probe head. 3) Place the probe head on the patient’s abdomen with the screen facing upwards and aim it to the location where the bladder is expected to be above 3~4cm the pubic bone.   4) Ensure that the probe head is pointing toward the patient’s foot.
CUBEScanTM BioCon-900S Operator’s Manual   Page 26    Figure 27 Positioning Bladder  Figure 28 Probe Direction 5) Tap the Scan  icon( ) on the  Information  screen  or  press the  probe  Scan  Button   to start  Pre-scan. A Pre-scan image such as 2D ultrasonic image appears on the probe screen.   No. Description 1 Center Line 2 Bladder  point:  It  helps  to easily  recognize  the  bladder. NOTE:  It  is  not  100% guaranteed  that  the  pointer shows  the  bladder  in  all cases 3 Depth  Line:  Max  detection depth to measure  a  padiatric patient’s bladder volume. 4 Exit Icon    Figure 29 Pre-scan Screen    Figure 30 Normal Screen  6) To locate the bladder, keep the probe contacted the abdomen. And tilt the probe slowly back and forth, side to side. The position where the bladder image is the largest and most centered is the optimal place to start the normal scan process. Locate the bladder along the centerline(1) on the Pre-scan screen. It helps detect the optimal position during Pre-scan. NOTE: Depth Line(3) is not displayed for male and female patient type. 7) If you want to stop the Pre-scan and return to the Information screen, tap the Exit icon( ).   8) Press the Scan Button   again or tap the Pre-scan screen. The progress bar on the screen appears. It takes 3 seconds to complete the normal scan Hold the probe steady while scanning. Movement of the probe during scanning reduces the accuracy of measurement. 9) When you hear the beeps, the scan is complete.   5.6  DISPLAY A SCAN RESULTS After a scan has been completed, the Scan Result screen appears. The Scan Result screen includes the following information:  Current Urine Volume  Maximum Urine Volume  Patient Type & Tag  Scan Date & Time  Bladder Planes  Aiming Information
CUBEScanTM BioCon-900S Operator’s Manual   Page 27     Figure 31 Scan Result Screen  5.6.1  BLADDER VOLUME   View the bladder volume displayed in milliliters (ml). The Urine Volume icon( ) shows the measured urine volume. (1) shows the most recent volume and (2) shows the maximum volume from multiple results for one patient.   5.6.2  PATIENT INFORMATION AND SCAN DATE & TIME (3) shows the date and time on the current scan. (4) shows the patient tag entered on the Information screen.     5.6.3  BLADDER PLANES Tap the bladder plane repeatedly with your finger to check other 12 planes of bladder on the Scan Result screen, if necessary. At the same time, the Aiming information (crosshair) is rotated clockwise by 15° and displayed the 2nd&8th planes, 3rd&9th, 4th&10th, 5th&11th, 6th&12th planes of bladder in order. If these plane’s bladder is not centerd in the target or too small, we recommend that you scan the bladder again. See the section “5.7 RE-AIM BLADDER”.      Figure 32 Planes & Aimming Information
CUBEScanTM BioCon-900S Operator’s Manual   Page 28 5.7  RE-AIM BLADDER If  the  bladder  is  not  centerd  in  the  target  or  too  small,  you  can  re-aim  and  re-scan  the  bladder  before saving  the  scan  result.  We  strongly  recommend  that  you  re-aim  and  re-scan  more  than  one  time  for accuracy.    Figure 33 Incorrectly Positioned  The crosshair helps how to re-aim and the guiding arrows show that the user has to move or tilt the probe for a more accurate scan. If the bladder appears in the center of the crosshair, press the Scan Button  or tap the Pre-scan part of the screen.     Figure 34 How To Re-aim   5.8  SAVE A SCAN RESULT    When you do not need to scan again, you can save the currently displayed scan result. And also missing scan result can be saved after fetching the missing scan result. See the section “5.9 FETCH A MISSING DATA”. To save the scan results, tap the Save icon ( ) below the display. The Home screen appears after the scan result is saved successfully.
CUBEScanTM BioCon-900S Operator’s Manual   Page 29 5.9  FETCH A MISSING DATA   If you did not save a scan result and return to the Home screen, the Missing Data icon ( ) appears on the Home  screen. Tap  the Missing  Data  icon  ( )  to fetch  the  unsaved  scan  result.  The  Scan  Result screen on the missing data appears. The performances below make the Missing Data icon( ) disappear.  1) Fetch and save the missing scan result. 2) A new normal scan is performed.   Figure 35 Home Screen     5.10  FINISH THE MEASUREMENT   After  finishing  scanning,  take the  probe  away  from  the patient  and  then  wipe  the  ultrasound  gel  off  the probe. See the section “8.3 CLEANING & DISINFECTION” When the probe is not in use, press the Power Button( ) twice to place the system in sleep mode and then mount it for charging internal battery in the Charger.   5.11  MANAGE SAVED SCAN RESULTS  1)  The  scan  results  are  saved  to  the  internal  memory  of  BioCon-900S  Probe.  If  you  scan  the  patient several times using the probe, the lagest result is saved. Information can be entered and modified before the scan result is saved, but once the scan result is saved, patient information can not be modified 2) If you want to delete the saved scan result, see the section “5.11.2 DELET A SAVED SCAN RESULT”  5.11.1  REVIEW A SCAN RESULT 1) The Review screen is displayed when you tap the Review icon ( ) on the Home screen.   2) The Review screen opens, it shows the urine volume, ultrasound images and other saved information.   3) Previous & Next icon ( ,  ) allow you to review the scan results of the other patients. 4) Delete & Exit icons ( ,  ) allow you to delete the current reviewing results or return to the Home screen. 5) Tap the (1) area to review the information you entered for the patient.
CUBEScanTM BioCon-900S Operator’s Manual   Page 30   Figure 36 Home Screen   Figure 37 Review Screen    5.11.2  DELETE A SAVED SCAN RESULT 1) The Review screen is displayed when you tap the Review icon ( ) on the Home screen.   2) Select the scan result you want to delete by tapping Previous & Next icon ( ,  ). 3) Tap the Deletion icon ( ) and a Delete Confirm dialog box ( ) appears. 4) Tap the OK icon ( ) to delete the scan result or Tap the Cancel icon ( ) to cancel.      Figure 38 Review Screen Figure 39 Delete Scan Result  6) If you delete the last scan result, Empty Foder dialog box ( ) appears and BioCon-900S returns to the Home screen. 7) If you tap the Review icon when there is no scan result, Empty Folder dialog box ( ) appears on the Home screen. 8) You can delete all  results  in  probe’s internal  memory at one  time,  if  you  want.  See  the section    “7.8 DELETE ALL MEASUREMENTS” for information on deleting all scan results  NOTE: The scan results are saved to the internal memory, which can store 50 data. If you exceed this limit, the system deletes the oldest data to accommodate the new data.
CUBEScanTM BioCon-900S Operator’s Manual   Page 31    Figure 40 Delete Last Scan Result  Figure 41 Empty Folder Dialog Box
CUBEScanTM BioCon-900S Operator’s Manual   Page 32 6  SCREENS 6.1  HOME SCREEN Turn the  probe  on  and  the  Home  screen  appears.  It  shows  the main functions  of  the  probe.  The  icons below on the screen allow you to select patient type, go to the Review screen or go to the Setup screen.      No. Description 1 Time 2 Number of Saved Data / Internal Storage (50) 3 Battery Status 4 Model Name 5 Patient Type Icons  6 Review Icon 7 Setup Icon  Figure 42 Home Screen    6.2  INFORMATION SCREEN The  Information  screen  is  used  to  enter  patient  data.  See  the  section  “5.4.1  INPUT  INFORMATION  – VIRTUAL KEYBOARD” for information on completing on the text box if you want.    Figure 43 Information Screen (All)   Figure 44 Information Screen (Female)   Figure 45 Information Screen (Pediatric)
CUBEScanTM BioCon-900S Operator’s Manual   Page 33 6.3  BARCODE SCREEN (OPTIONAL)  The Barcode screen is used to record data by scanning barcode with patient information. See the section “5.4.2 INPUT INFORMATION – OPTIONAL BARCODE” for more information if you want.         Figure 46 Information Screen (Optional Barcode) Figure 47 Barcode Screen  6.4    VIRTUAL KEYBOARD SCREEN BioCon-900S has three different virtual keyboard types; Numeric, Letter(Uppercase, Lowcase), and Date.    Figure 48 Numeric Keyboard   Figure 49 Letter Keyboard   Figure 50 Date Keyboard  Icon Description  Numeric and Letter Keyboard - Delete a character to the right of the cursor. Date Keyboard – Move the cursor to the left  Delete all characters.  Convert a string on a virtual keyboard.  Exit the Virtual Keyboard screen after updating edited characters.
CUBEScanTM BioCon-900S Operator’s Manual   Page 34   6.5  NORMAL SCAN SCREEN Normal  scan  -  Get  12-plane  ultrasound  images  and  calculate  the  residual  urine  in  the  bladder  (3D scanning). It takes 3 sec to complete the Normal scan. As bladder volume is calculated, the progress bar increases until the scan is completed.    Figure 51 Normal Scan Screen   6.6  PRE-SCAN SCREEN Pre-scan-  2D ultrasound  images  are  displayed  continuously in  live  image basis.  This  helps an  operator locate bladder position and predict the range of residual urine. And also the centerline is provided during Pre-scan. Locate the bladder along the centerline on the screen. It helps detect the optimal position during Pre-scan. The  Pre-scan  is  executed  only  when  ‘Pre-scan’  setup  is  ‘ON  ( )’.  When  ‘Pre-scan’  is  ‘OFF  ( )’,  a Normal scan is executed.     No. Description 1 Centerline. 2 Bladder point 3 Max detection depth  to measure  a pediatric urine volume. 4 Cease the  Pre-scan and  return to the Information screen.   5 Tap  the  Pre-scan  screen,  the  Pre-scan  is executed.     Figure 52 Pre-scan Screen (Default)   Add a space.  Exit the screen after recovering previous characters.
CUBEScanTM BioCon-900S Operator’s Manual   Page 35 6.7  SCAN RESULT SCREEN   The Scan Result screen appears automatically when a normal 3D measurement is complete. The aiming information (crosshair) helps you verify the Scan result. The icons below on the display allow you to review other planes of the same measurement, return to the Home screen, and save the measurement. Also if you want to recan the bladder, tap the Re-Scan icon( ).          Figure 53 Scan Result (All) Figure 54 Scan Result (Female) Figure 55 Scan Result (Pediatric)       Figure 56 Scan Result Screen  No. Description 1 Time 2 Number of Saved Data / Internal Storage (50) 3 Battery Status 4 Aiming Information(Crosshair) 5 Measured Date & Time 6 Patient Tag 7 The first plane of 12 planes 8 Save 9 Home 10 Re-Scan 11 Current Volume from Multiple Results 12 Maximum Volume from Multiple Results 13 Patient Types ( ) 14 Bladder Outline 15 Max detection depth to measure a urine volume (Only pediatric patient)   6.8  REVIEW SCREEN Tap the Review icon( ) on the Home screen, Review screen appears  The Review  screen shows the bladder volume, ultrasound images and  other saved  information  with the saved scan results. While reviewing the result, the menu icons below on the display allow you to delete the current  reviewing  result  or  return  to  the  Home  screen.  If  you  want  to  review  the  scan  results  of  other patients, tap the (9) icons.  (11) shows the chronological order of saved results and the total number of saved results. For example, 3/49 means the 3rd saved scan result of the 49 stored results.
CUBEScanTM BioCon-900S Operator’s Manual   Page 36   Figure 57 Review Screen     No. Description 1 Time 2 Number of Saved Data / Internal Storage (50) 3 Battery Status 4 Aiming /information(Crosshair) 5 Measured Date & Time 6 Patient Tag 7 The first plane of 12 planes 8 Delete Icon 9 Previous & Next Patient Icon 10 Exit Icon 11 Chronological Order of Saved Results / Total Order of Saved Results 12 Bladder volume 13 Patient Types ( ) 14 Bladder Outline 15 Centerline (helps detect the optimal position) 16 First Name & Sur Name 17 Date of Birth 18 Delete Dialog Box 19 OK Icon 20 Cancel Icon   Figure 58 Review Screen - Delete    6.9  SETUP SCREEN Tap the Setup icon ( ) on the Home screen. In the Setup screen the left column shows the Setup Items and the right column shows the Setup Values. For more information on the system setting, see the section “7. SETTINGS”
CUBEScanTM BioCon-900S Operator’s Manual   Page 37 1) Tap a Setup Value of the desired setup item. 2) Tap repeatedly a setup value untile desired setup value. 3) You can move to other Setup screen (1, 2, 3) by tapping the Previous or Next icon ( , ) 4) Tap the Home icon ( ) to finish adjusting the setup.        Figure 59 Setup Screen (1) Figure 60 Setup Screen (2) Figure 61 Setup Screen (3)   6.10  BATTERY STATUS     A fully charged battery   A battery about 80% charged.   A battery about 60% charged   A battery about 40% charged   A battery nearly depleted   Recharge the battery immediately  When the battery is nearly depleted, following dialog box appears asking you to charge the battery.    Figure 62 Dialog box - Needs to be recharged   To conserve battery power, the probe goes into sleep mode by turning off the screen when not in use for a specified time. You can adjust the time to go into sleep mode in the setup.  See the section “7.4 SLEEP MODE SETTINGS”. The probe also goes into sleep mode if you press the Power Button() briefly with power on.  To preserve battery power, the probe goes into shut down mode in approximately 24 hours after going into sleep mode, if there is no activity. However, if you do not plan to use the probe for more than a week, you should shut the probe down completely to prevent it from discharging completely.  It takes approximately 6 hours for the discharged battery to fully charge.    Fully charged battery can provide approximately 1,000 normal scans.  Contact your local distributor when the fully charged battery does not allow normal scan for 10min.
CUBEScanTM BioCon-900S Operator’s Manual   Page 38 6.11  DIALOG BOXES    Windows Description  Confirm Message: “Are you sure you want to delete this saved scan result?”  Confirm Message: “Are you sure you want to delete all saved measurements?”  Confirm Message: shutdown mode – To prevent the probe from discharging. “Are you sure you want to go into shutdown mode?”    Alert Message: The message prompts you when calibration is needs.  Alert Message: The battery is nearly depleted and needs to be recharged.    Alert  Message:  The  battery  needs  to  be  recharged  immediately,  so  the  probe goes to sleep mode soon by shutting itself down    Alert Message: The temperature of oil inside the probe is below 10°C.   So the thermostat  inside probe makes  the oil  warm and  the  temperature  of oil rises above 10°C. See the section “8.2 DIALOG BOXES FOR MAINTENANCE” for more information
CUBEScanTM BioCon-900S Operator’s Manual   Page 39 7  SETTINGS 7.1  DATE FORMAT SETTINGS 1) Tap the  Setup  icon ( ) on the  Home screen.  Tap the  Home  icon ( ) to  return to the Home screen, if you want. 2) Locate the Date Format option( ), and then tap the Date Format Items ( / /) repeatedly to adjust the desired date format.   3) A format of date values immediately updates depending on the date format.      Figure 63 Date Format (YY-MM-DD, MM-DD-YY, DD-MM-YY)   7.2  DATE SETTINGS 1) Tap the  Setup  icon ( )  on the  Home  screen.  Tap  the Home  icon  ( )  to  return to  the  Home screen, if you want. 2) Locate  Date  option( )  and  tap  Date  Value Item  (eg. ),  use  the  virtual  keyboard  to enter a date. 3) Tap the OK icon ( ) to finish adjusting the date after entering.        Figure 64 Date Setup Figure 65 Enter Date Values (YY-MM-DD, MM-DD-YY, DD-MM-YY)    7.3  TIME SETTINGS
CUBEScanTM BioCon-900S Operator’s Manual   Page 40 A 24-hour clock provided only.   1) Tap the  Setup  icon ( )  on the  Home  screen.  Tap  the Home  icon  ( )  to  return to  the  Home screen, if you want. 2) Locate  the  Time  option( )  and  tap  Time Item  (eg. )  to  adjust  the  time,  use  the  virtual keyboard to enter a time. 3) Tap the OK icon ( ) to finish adjusting the time.     Figure 66 Time Setup Figure 67 Enter Time Values (hour:minute)   7.4  SLEEP MODE SETTINGS A time setting to go into sleep mode: 2min, 3min, 5min   1) Tap  the  Setup  icon  ( )  on  the  Home screen.Tap  the  Home  icon  ( )  to  return  to  the Home screen, if you want. 2) Locate the Sleep Mode option( ) and tap Sleep Mode Item ( 2min / 3min / 5min ) repeatedly to adjust the time to go to sleep mode.        Figure 68 Time Setting of Sleep Mode (2min, 3min, 5min)    7.5  DISPLAY BRIGHTNESS SETTINGS
CUBEScanTM BioCon-900S Operator’s Manual   Page 41  1) Tap the  Setup icon ( )  on the  Home  screen.  Tap  the Home  icon ( )  to return  to  the Home screen, if you want. 2) Use the Next icon ( ) on the Setup screen, and look for Brightness option( ). 3) Tap the Brightness Item (eg.50%, 60%, 70%) repeatedly to adjust display brightness (0%-100%).    NOTE: The display brightness setting affects battery life. To conserve battery life, adjust brightness to a lower setting      Figure 69 Brightness Setup   7.6  PRE-SCAN SETTINGS   1) Tap the Setup icon ( ) on the Home screen to open the Setup screen. Tap the Home icon ( ) to return to the Home screen, if you want. 2) Use the Next icon ( ) on the Setup screen, and look for Pre-scan option( ). 3) Turn the Pre-scan on    ( ) or off ( ) by tapping the Light Item (   or )     Figure 70 Pre-scan On/Off Setting 7.7  DELETE ALL MEASUREMENTS
CUBEScanTM BioCon-900S Operator’s Manual   Page 42 1) Tap the Setup icon ( ) on the Home screen to open the Setup screen. Tap the Home icon ( ) to return to the Home screen, if you want. 2) Use the Next icon ( ) on the Setup screen, and look for the Delete option( ). 3) Tap the Ok Item( ) to delete all saved results.   4) And a dialog box ( ) to confirm appears before deletion.   5) Tap the Ok icon ( ) to delete or tap the Cancel icon ( ) to cancel.   6) Tap the Home icon ( ) to returen to the Home screen. 7) See the Home screen whether all saved results were deleted.      Figure 71 Delete All Data
CUBEScanTM BioCon-900S Operator’s Manual   Page 43 8  MAINTENANCE  8.1  TROUBLESHOOTING  Error message Description Actions  FPGA Configuration Error Contact your local distributor or Mcube Technology.    System Information Error Contact your local distributor or Mcube Technology.  Angle Motor Error Contact your local distributor or Mcube Technology.  Plane Motor Error Contact your local distributor or Mcube Technology..  High Temperature Error Contact your local distributor or Mcube Technology.  Low Temperature Error Contact your local distributor or Mcube Technology.  Data Deletion Error Contact your local distributor or Mcube Technology.   8.2  DIALOG BOXES FOR MAINTENANCE  popup Description  Indicates  that  the  temperature  of  oil  inside  the  probe  is  below  10°C.  So  the thermostat  inside  probe  makes  the  oil  warm  and  the  temperature  of  oil  rises above 10°C. The following dialog boxes appear during the progress of warming oil. Warming oil  lasts  (about  3min  or  less)  until  the  temperature  of  oil  reaches  10°C.  When finished, it goes to the Home screen.          (During warming oil)        (when finished)
CUBEScanTM BioCon-900S Operator’s Manual   Page 44 8.3  CLEANING & DISINFECTION      CAUTION  This product may only be  cleaned and disinfected by using the approved processes provided in this manual. Cleaning and disinfection methods listed are recommended by Mcube Technology based on compatibility with component materials.    Ensure that you follow the manufacturer’s instructions for handling and disposing of the cleaning and disinfection solutions provided in this manual.  BioCon-900S  probe  is  intended  to  contact  the  abdominal  skin  to  measure  the  bladder  volume.  For  this reason, cleaning and disinfecting the probe and  Charger  is an important procedure  in using the system. The following table shows the risk assessment for the probe including the Spauling’s/CDC classification for the minimum required disinfection level.  Component Packaged Use Spaulding’s/CDC classification Clean Disinfection level Low High Probe Head Nonsterile Reusable Noncritical √ √  Probe Except head Nonsterile Reusable Noncritical √   Charger Nonsterile Reusable Noncritical √    8.3.1  Compatibility  The  following  solutions  have  demonstrated  cleaning  or  disinfection  efficiency  and  material  compatibility with system components:   Metrex® CaviWipes™  Metrex® CaviCide™ Spray  8.3.2  Cleaning Ensure the power cord is not plugged into wall outlet. Using a dry paper towel or soft cloth, wipe any ultrasound gel, soils and dirt completely off the Charger or   WARNING   Cleaning is a necessary step to perform before disinfecting the device. If the device is not cleaned properly,  even  after  the  disinfection  procedure  is  completed,  there  may  be  contamination  of  the device.    Incorrect cleaning or disinfecting of any part of the system can cause permanent damage.    Do not use solvents such as thinner or benzene, or abrasive cleaner on any part of the system.  Do not subject the system to any method of sterilization.  Do not immerse the probe except the probe head in cleaning or disinfectant solution.    Do not immerse the Charger in cleaning or disinfectant solution.    Do not use metal or abrasive brushes. These may scratch the instrument causing permanent device damage.  Wear protective gloves when cleaning the device.  Availability of  cleaning  and  disinfection  products  varies  by country,  and  Mcube Technology  is not able  to  test  the  products  in  every  market.  Please  contact  your  local  service  provider  or  Mcube technology if you need more information.    Do not spray the disinfectant spray solution directly on the surface, recessed areas, crack,    etc. of the probe and the Charger.
CUBEScanTM BioCon-900S Operator’s Manual   Page 45 and probe.  1) Cleaning outer case (housing) of the Charger. a) Put on new gloves. b) Spray  a  soft-bristled  brush(M16  recommended  and  validated)  with  Metrex  CaviCide  spray  until thoroughly  wet.  Using  the  brush,  scrub  mated  surfaces,  recessed  areas,  cracks,  crevices  and other hard to reach areas until no visible contaminant is left on the Charger. c) Use a CaviWipe towelette to completely clean the surface on the Charger. d) Use  a  second  CaviWipe towelette  to  thoroughly  wet  all  surfaces  of  the  Charger  until  all  visible contaminants are removed. Allow the Charger to remain wet for 2 minutes. e) If using an additional Metrex CaviCide towelette, leave the solution on the wiped Charger surface for about 2 minutes. f) Visually inspect the Charger for visible contaminants.   g) Allow the Charger to air dry.   Figure 72 Charger  Please  ensure  that  the  electric  inlet  of  Charger  is  not  exposed  to  any  solution  during  the  cleaning procedure. If the inlet is exposed to the solution (such as water, isopropyl alchol or ethyl), dry the Charger completely before use.  2) Cleaning the probe   Since the surface of the probe affects the result of data, users should keep as follows:     a) Put on new gloves. b) Spray  a  soft-bristled  brush(M16  recommended  and  validated)  with  Metrex  CaviCide  spray  until thoroughly  wet.  Using  the  brush,  scrub  mated  surfaces,  recessed  areas,  cracks,  crevices  and other hard to reach areas until no visible contaminant is left on the probe. c) Use a CaviWipe towelette to completely clean the surface of the probe.   d) Use a second CaviWipe towelette to thoroughly wet all surfaces of the probe. Allow the probe to remain wet for 2 minutes. e) If using an additional Metrex CaviWipe towelette, leave the solution on the wiped probe surface for about 2 minutes. f) Visually inspect the probe for visible contaminants.   g) Allow the probe to air dry.      Figure 73 Probe Front-View Figure 74 Probe Back-View
CUBEScanTM BioCon-900S Operator’s Manual   Page 46 8.3.3  Disinfection Please note the that low-level disinfection of the probe head is required before a new measurement. a) Put on new gloves b) Spray the probe head until it is sufficiently wet with a Metrex CaviCide spray.   Figure 75 Probe Head  c) Leave the solution on the probe head surface for about 3 minutes.   d) If you spray the Metrex CaviCide spray additionally, leave the solution on the surface for at least 3 minutes. e) Allow the probe head to air dry.   f) Wipe the probe head thoroughly using the fresh CaviWipes. g) Keep the probe head wet for 3 minutes.   h) Additional wipes may be used to ensure the probe remaining wet for the 3 minutes, if necessary.   i) Allow the probe head to air dry.   j) Using sterile, lint-free cloths wetted with RO/DI water thoroughly wipe the probe for at least a minute k) Allow the probe head to air dry.     8.4  BATTERY MAINTENANCE BioCon-900S battery cannot be removed  by the user. If  you do not plan to use  BioCon-900S for a long period of time, you should switch the probe to shutdown mode to prevent its battery from fully discharging. The  BioCon-900S  should  be  charged  every  6  months  at  a  minimum,  store  it  in  accordance  with  the recommended conditions. See the section “9.5 ENVIRONMENTAL CONDITIONS”.     8.5  WEEKLY INSPECTION a) Inspect the probe and Charger thoroughly for cracks or leakages. b) Inspect the power cord for checking if there are any damaged parts. c) When scanning, check out any abnormal noise emanating from the probe head.   8.6  DISPOSAL The device and components may contain environmentally hazardous materials (mineral oil, lead, battery pack, etc.). When they have reached the end of its useful service life, return them to the Mcube Technology, or follow your local regulations for hazardous waste disposal.   8.7  DEVICE REPAIR Faults  not  described  in  section  “8.1  TROUBLESHOOTING”  are  intended  to  be  serviced  by  a  certified technician. When any troubles unlisted in section “8.1 TROUBLESHOOTING” occur, contact an authorized service provider or Mcube Technology.
CUBEScanTM BioCon-900S Operator’s Manual   Page 47 8.8  SELF TEST Self-test 1) Put the probe on the Charger   2) Turn the probe on and tap the Setup icon ( ) on the bottom right side of the Home screen. 3) Tap the Next icon ( ) on the Setup screen and select the Self-test icon ( )   4) Tap the Scan icon ( ) to start the calibration. You can cease the Self-test by tapping the Cease icon ( ). Tap the Exit icon ( ) to return to Setup screen, if you want. 5)  Self-test  progress  bar  appears,  when  the  calibration  is  started.  The  progress  bar  indicates  the progress of calibration. It takes about 3minutes to complete the calibration. When the calibration is completed, 100% sign is displayed.   Tap the Ok icon ( ) and the latest calibration date is updated. Tap the Exit icon ( ) to return to the Setup screen.
CUBEScanTM BioCon-900S Operator’s Manual   Page 48 8.9  CALIBRATION BioCon-900S must be calibrated every 12 months to ensure accurate results. Calibration ensures accurate alignment of BioCon-900S’s internal system. You can take scan, even if calibration is not performed by the designated date. But result can be compromised. There are following methods for the calibration.    CAUTION  Calibrate BioCon-900S when the battery is charged over 60%.    8.9.1 PHANTOM CALIBRATION Phantom Calibration 1) Place  the  CubeScan  phantom  on  a  flat surface and open the cover of the phantom.  2) Pour about 5ml of water or more on the center surface of the phantom  3) Place  the  phantom  cover  on  the  top  of  the CubeScan phantom. Check if the cover is in a stable and flat position.  4) Put the probe head into the probe holder firmly. 5) Turn the probe on and Tap the Setup icon ( ) on the bottom right side of the Home screen. 6) Tap the  Next  icon  ( )  on  the  Setup  screen and select Phantom Calibration icon ( ) 7) Tap the Scan icon ( ) to start the calibration. You  can  cease  the  phantom  calibration  by tapping  the  Cease  icon  ( ).  Tap  the  Exit  icon ( ) to return to Setup screen, if you want. 8) The calibration progress bar appears, when the calibration is started. The progress bar indicates the  progress  of  calibration.  It  takes  about 3minutes to complete the calibration. When the calibration is completed, 100% sign is displayed on the screen.   9) Tap the  Ok  icon ( ) and the latest  calibration date is updated. Tap the Exit icon ( ) to return to the Setup screen.
CUBEScanTM BioCon-900S Operator’s Manual   Page 49 8.9.2 CALKIT CALIBRATION Calkit Calibration 1) Open the cover of the calibration kit and pour the saline solution.  Ideal calibration condition You  can  pure  water  instead  of  the saline  solution. Ensure that water does not contain any air bubbles before starting the calibration.   2) There  is  a  mark  on  the  inside  of  the  Calkit showing  the  Water  Level.  The  Water  Level  is about 1cm below the top edge. Pour water to the Water Level.       3) Close the cover of the calibration Kit. Make sure that the arrow of the Calkit cover is aligned with the nipple of the Calkit body.  4) Align the probe’s Scan Button   with the arrow of  the cover  and  put  the  probe  head  firmly  into the cover. 5) Turn the probe on and tap the Setup icon ( ) on the botton right side of the Home screen.   6) Tap the Next icon ( ) on the Setup screen and select the CalKit Calibration icon ( ) 7) Tap the Scan  icon ( ) to  start the  calibration. You  can  cease  the  calkit  calibration  by  tapping the  Cease  icon  ( ).  Tap  the  Exit  icon  ( )  to return to the Setup screen, if you want.   8) The calibration progress bar appears, when the calibration is started. The progress bar indicates the  progress  of  calibration.  It  takes  about 3minutes  to complete  the  calibration. When the calibration  is  completed,  100%  sign  wil  be displayed on the screen.   Tap  the  Ok  icon  ( )  and  the  latest  calibration date is updated. Tap the Exit icon ( ) to return to the Setup screen.   8.9.3 CALIBRATION REMINDER 1) Calibration Reminder Dialog box   Calibration Reminder Dialog box appears on the top screen under the conditions as follows: -  After 1 year since the last calibration Tap the Ok icon ( ) to return to the Home screen. The Calibration Reminder Dialog box disappears.  2) Calibration Reminder icon The Calibration Reminder icon ( ) is displayed on the Home screen after the Calibration Reminder
CUBEScanTM BioCon-900S Operator’s Manual   Page 50 Dialog  box  disappears.  When  you  finish  the  calibration,  the  last  calibration  date  is  updated  and  the Calibration Reminder icon (    ) disappears.      Figure 76 Calibration Reminder Icon on Home Screen
CUBEScanTM BioCon-900S Operator’s Manual   Page 51 9  SPECIFICATIONS 9.1  ACOUSTIC OUTPUT TABLE   Transducer Model: BioCon-900S Transducer Operating Mode: B-mode The values in this table are the maximum readings obtained from each frequency.     TIS TIB TIC Index Label MI scan non-scan non-      Aaprt<=1 Aaprt >1 scan  Global Maximum Index Value 0.34 0.62      Pr,a   (MPa) 0.48       P (mW)  0.99      Min of [Pα(zs), Ita,α(zs)]] (mW)       Assoc zs (cm)       Acoustic zbp (cm)       Parameter zb (cm)        z at max. Ipi (cm) 2.7       deq(zb) (cm)        fawf (MHz) 2.04 2.04      Dim of Aaprt X (cm)  1.1      Y (cm)  1.1      td (microsec) 1.22       prr (Hz) 390       pr at max. Ipi (MPa) 0.61      Other deq at max. Ipi (cm)       Information Focal   FLx (cm)  5.5      Length FLy (cm)  5.5      Ipa.3 at MImax (W/cm2) 7.8               Operating        Control        Conditions                 9.2  ELECTROMAGNETIC COMPATIBILITY 9.2.1  Electromagnetic emissions  Guidance Manufacturer’s declaration - electromagnetic emissions The BioCon-900S is intended for use in the electromagnetic environment specified below. The customer or the user of the BioCon-900S should assure that it is used in such an environment. Phenomenon   Test level / requirement Electromagnetic environment - guidance   RF Emissions CISPR 11   Group 1 The  BioCon-900S  uses  RF  energy  only  for  its   WARNING   ALARA principle (As Low As Reasonably Achievable) should be employed for all medical ultrasound exposure. Users can not adjust the acoustic output energy of the BioCon-900S. The output level is below  than  the  levels  which  were  referred  in  AIUM  remarks  in  1992  and  there  were  no  reported biological effects under the level. However, to minimize exposure, measurements should be kept as short as possible. Refer to the acoustic output section for more information.
CUBEScanTM BioCon-900S Operator’s Manual   Page 52 internal function. Therefore, its RF emissions are very  low  and  are  not  likely  to  cause  any interference in nearby electronic equipment RF Emissions CISPR 11   Class A The  BioCon-900S  is  suitable  for  use  in  all establishments,  other  than  domestic  and  those directly  connected  to  the  public  low-voltage power  supply  network  that  supplies  buildings used for domestic purposes Harmonic Current Emissions   IEC 61000-3-2 Class A Voltage fluctuations/Flicker emission   IEC 61000-3-3 Complies   9.2.2  Electromagnetic Immunity  Guidance Manufacturer’s declaration - electromagnetic immunity The BioCon-900S is intended for use in the electromagnetic environment specified below. The customer or the user of the BioCon-900S should assure that it is used in such an environment.   Immunity test IEC 60601-1-2 Test level Compliance Electromagnetic environment - Guidance Electrostatic discharge (ESD)  IEC 61000-4-2 ±8kV Contact  ± 2, ± 4, ± 8, ±15kV Air In compliance Floors should be  wood, concrete  or ceramic  tile.  If  floors  are  covered with  synthetic  material,  the  relative humidity should be at least 30%. Electrical  fast transient/burst   IEC 61000-4-4 ± 2 kV 100 kHz repetition frequency In compliance Mains  power  quality  should  be  that of  a  typical  commercial  or  hospital environment.   Surge  IEC 61000-4-5 Line to Line   ± 0.5 kV, ± 1 kV  Line to Ground   ± 0.5 kV, ± 1 kV, ± 2 kV In compliance Mains  power  quality  should  be  that of  a  typical  commercial  or  hospital environment.   Voltage dips  IEC 61000-4-11 0 % UT: 0.5 cycle At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315° In compliance Mains  power  quality  should  be  that of  a  typical  commercial  or  hospital environment.  If  the  user  of  the BioCon-900S  requires  continued operation  during  power  mains interruptions, it is recommended that the BioCon-900S ultrasound system be  powered  from  an  uninterruptible power supply or a battery. 0 % UT; 1 cycle   and 70 % UT; 25/30 cycles Single phase: at 0º In compliance Voltage interruptions  IEC 61000-4-11 0 % UT for 250/300 cycle In compliance Power frequency magnetic  field Immunity  IEC 61000-4-8 30 A/m 50 Hz & 60 Hz In compliance Power  frequency  magnetic  fields should  be  at  levels  characteristic  of a  typical  location  in  a  typical commercial or hospital environment.  9.2.3  RF Immunity  Immunity test   IEC 60601 test level   Compliance   Electromagnetic environment
CUBEScanTM BioCon-900S Operator’s Manual   Page 53 Conducted RF   IEC 61000-4-6 3 V 0.15-80 MHz  6 V in ISM bands between 0.15 MHz and 80 MHz  80%  AM  at  1 kHz   3 V 0.15-80 MHz  6 V in ISM bands between 0.15 MHz and 80 MHz  80% AM at 1 kHz Portable  and  mobile  RF  communications equipment should be used no closer to any part of  the  BioCon-900S,  including  cables,  than  the recommended  separation  distance  calculated from the equation applicable to the frequency of the transmitter.      Recommended separation distance   d =1.2  √𝐏 Radiated RF  IEC 61000-4-3   3 V/m 80MHz-2.7GHz 80% AM at 1 lHz 3 V/m 80MHz-2.7GHz 80% AM at 1 lHz d = 1.2  √𝐏 80 MHz to 800 MHz d =2.3  √𝐏 800 MHz to 2.7 GHz  where P  is the maximum output power rating of the  transmitter  in  watts  (W)  according  to  the transmitter  manufacturer  and  d  is  the recommended separation distance in meters (m).  Field  strengths  from  fixed  RF  transmitters,  as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. b  Interference  may  occur  in  the  vicinity  of equipment marked with the following symbol :  NOTE 1) At 80MHz and 800MHz, the higher frequency range applies.   NOTE  2)  These  guidelines  may  not  apply  in  all  situations.  Electromagnetic  propagation  is  affected  by absorption and reflection from structures, objects and people. a.  Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and  land  mobile  radios,  amateur  radio,  AM  and  FM  radio  broadcast  and  TV  broadcast  cannot  be predicted  theoretically  with  accuracy.  To  assess  the  electromagnetic  environment  due  to  fixed  RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location  in  which the  BioCon-900S  is  used  exceeds  the  applicable  RF  compliance  level  above,  the BioCon-900S  should  be  observed  to  verify  normal  operation.  If  abnormal performance  is  observed, additional measures may be necessary, such as re-orienting or relocating the BioCon-900S. b.  Over the frequency range 0.5MHz to 80MHz, field strengths should be less than 3 V/m.  9.2.4  Recommended separation distance  Recommended Separation Distances between Portable and Mobile RF communications equipment and the BioCon-900S The  BioCon-900S  is  intended  for  use  in  an  electromagnetic  environment  in  which  radiated  RF disturbances  are  controlled.  The  customer  or  the  user  of  BioCon-900S  is  can  help  prevent electromagnetic  interference  by  maintaining  a  minimum  distance  between  portable  and  mobile  RF communications  equipment  (transmitters)  and  BioCon-900S  as  recommended  below,  according  to  the maximum output power of the communications equipment. RATED MAXIMUM OUTPUT POWER OF SEPARATION DISTANCE ACCORDING TO FREQUENCY OF TRANSMITTER (m) 150 kHz to 80 MH 80 MHz to 800 MHz 800 MHz to 2.5 GHz
CUBEScanTM BioCon-900S Operator’s Manual   Page 54 TRANSMITTER (W) d = 1.2√  𝐏 d = 1.2√  𝐏 d = 2.3√  𝐏 0.01 0.12 0.12 0.23 0.1 0.38 0.38 0.73 1 1.2 1.2 2.3 10 3.8 3.8 7.3 100 12 12 23 For  transmitters  rated  at  a  maximum  output  power  not  listed  above,  the  recommended  separation distance  d  in  meters  (m)  can  be  estimated  using  the  equation  applicable  to  the  frequency  of  the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.  9.2.5  Components conformance to standards  To  maintain  electromagnetic  interference  (EMI)  within  certified  limits,  the  device must  be  used  with  the cables, components specified or supplied by Mcube Technology. The use of cables or compoments other than those specified or supplied may result in increased emissions or decreased immunity of the system. EMC standards   Compoment Max length Power Cord 1.83 m(6ft) Adapter 1 m (3.3ft)   9.3  EXCESSIVE TEMPERATURE TEST RESULT  Transducer type    → External use Test to be applied ↓ Simulated use test Temperature rise The ambient temperature : 22.8  ℃ The temperature rise to:    24.6  ℃ Still air test (no gel) Temperature rise The ambient temperature : 23.1  ℃ The temperature rise to: 25.5  ℃   9.4  SPECIFICATIONS OF COMPONENTS    9.4.1  WIRELESS CHARGING  Main Chipset TI (BQ500210RGZ) Coil type 1 Coil (A10)
CUBEScanTM BioCon-900S Operator’s Manual   Page 55 Power Input 1A / 60-80VA Power Output <5W (5V, 1A) Charging efficiency 70%(Power max.) Temperature sensor   NTCG163JF103FT1 LED Battery indicator   2LED (Yellow:1 / Green:1) Operating frequency (wireless charging function) 149-156 kHz  NOTE: 1. The output power of a wireless charging is lower than 5 W.  2. The distance for wireless charging between Charger and a probe is less than 10mm.    3. The probe has to be properly mounted on the Charger for wireless charging. See the section “4.2 CHARGE THE BATTERY”, if the battery is not charging.   9.4.2  SPECIFICATION OF COMPONENTS      Probe Item Features Bladder Volume Range 0 - 999ml Accuracy 0-99ml±10ml, 100-999ml±10% (According  to  the  scanning  instruction,  and  scanning  on  a  Mcube Technology tissue-equivalent bladder phantom.) Scan time Less than 3 sec Power MC-BA-01 (Lithium-ion rechargeable battery) - Nominal Voltage: 3.6v - Nominal capacity:5200mAh Charging time: less than 6 hours - Charged by the Charger Number of scanning : approximately 1000 scans.   - For a new battery module fully charged. - Tested under Mcube Technology’s test conditions. Expected Service Life 7 years Display TFT-LCD 2.7 inch Format : 240 X RGB X 320 Stripe Patient ID Input Virtual Keyboard   1D/2D Barcode (Optional) User Interface - LCD Icons - 2 buttons - Touch screen Water Resistance Rated at IPX3   Probe Head -Transducer : Diameter: 14mm -Sector scan -Dual frequency (2MHz, 3.4MHz) -B-mode scan image     -Scan angle : 120°     -Penetration depth(normal patient): 23cm     -Applied part: Probe Cap -Type BF Mode of Operation - Continuous operation Weight 540g with battery
CUBEScanTM BioCon-900S Operator’s Manual   Page 56 Charger Item Features Use Indoor Power AC/DC Adapter: Manufacturer: MEAN WELL Model: GSM60A12 Input: 100- 240VAC, 50/60Hz, 1.4-0.7A   Output: 12V    5.0 A, 60W MAX Output Wireless Power Transfer Charger Insulation Class I with protective earth Water Resistance IPX0     Mode of Operation Continuous operation. Weight 480g (without adapter and AC cord)   9.5  ENVIRONMENTAL CONDITIONS      CAUTION  Do not place the heavy object on the device. The device can be damaged.  Keep away from the humid or dusty place.   CAUTION  Do not place the heavy object on the device. The davice can be damaged.  Keep away from the humid or dusty place.  Use the device indoors only under the environmental conditions.  For additional protection of the device during a lightning storm, or when it is left unattended and unused for long periods of time, unplug it from the wall outlet.  The immediate use of the probe which has been stored under 10℃  without warming may damage the probe. So take it into a room where the temperature is over 10℃  and let BioCon-900S warm up for a while. And then when the temperature of probe is over 10℃  you may use it.  If  you  are  not  using  BioCon-900S  more  than  3  months,  remove  the  battery  to  prevent  it  from discharging and store it in accordance with the recommended conditions, and charge the battery every 6 months.    Contact your local distributor or Mcube Technology, when you need to remove or replace the battery.  Enviromental Specifications Operating conditions Use Indoor Ambient temperature range +10 - +40℃  (+50 - +104℉) Relative humidity +30% - +75% non-condensing Atmospheric pressure range +700hPa - +1060hPa Storage and transport conditions Use Indoor Ambient temperature range -20 - +60℃  (+14 - +140℉)    WARNING  Do not leave this device in places subject to extremely high temperatures such as a sealed vehicle or in direct sunlight. This can cause a fire.  Do not place the device on an unstable  surface. This can cause the device to  fall or  tip over and cause injury.
CUBEScanTM BioCon-900S Operator’s Manual   Page 57 Relative humidity +20%    -  +80% non-condensing Atmospheric pressure range +600hPa - +1060hPa   9.6  European Union Declaration of Conformity   This device complies with the essential requirements of the Radio Equipment directive: 2014 / 53 / EU. The following  test  methods  have  been  applied  to  prove  presumption  of  conformity  with  the  essential requirements of the Radio Equipment directive: 2014/53/EU:  Manufacturer Mcube Technology Co.,Ltd. Products CUBEScan Charger / Charger-001 Bladder Volume Measurement System / BioCon-900S Product Description WPT System   EU Directives 2014/53/EU – Radio Equipment Directive (RED)  Reference standards used for presumption of conformity: Article Number Requirement External use 3.1a Medical electrical equipment IEC 60601-1 : 2005 + CORR. 1 (2006) + corr. 2 (2007) EN 60601-1 : 2006 + A1 : 2013 RF Exposure EN 62311:2008 3.1b ElectroMagnetic Compatibility (EMC) standard for radio equipment and services EN 301 489-1 V2.1.1 EN 301 489-3 V2.1.1 3.2 Wireless power transmission systems, using technologies other than radio frequency beam, in the 19 - 21 kHz, 59 - 61 kHz, 79 - 90 kHz, 100 - 300 kHz, 6 765 - 6 795 kHz ranges EN 303 417 V1.1.1
CUBEScanTM BioCon-900S Operator’s Manual   Page 58 10 GLOSSARY  Probe The main device with the LCD display. Charger CUBEScan Charger; Charge a lithium-ion battery with a probe.   Transducer Device  that  transforms  one  form  of  energy  into  another  form  of  energy. Ultrasound  transducer  transforms  electric  energy  into  acoustic  energy  and vice versa. Transducer in this guide means ultrasound transducer. B-mode A  kind  of  ultrasound  imaging  mode.  Displays  the  brightness  information corresponding to the amplitude of the signal. Pre-scan General  2D  real-time  ultrasound  scanning.  2D  ultrasound  images  are displayed continuously in live image basis.   This  helps  an  operator  locate  bladder  position  and  predicting  the  range  of residual urine. Normal Scan Gets  12-plane  ultrasound  images  and  calculates  the  residual  urine  in  the bladder (3D scanning). Self-Test Displays Self-test progress and results.   Phantom-Calibration Displays phantom-calibration progress and results.
CUBEScanTM BioCon-900S Operator’s Manual   Page 59 11 SYMBOL DIRECTORY  Symbol Description Symbol Description  Warning Failure to observe the warning could  result  in  personal  injury or  serious  damage  to  the system  Serial number  Caution Failure  to  observe  the recommendations  and precautions  may  result  in system damage or failure  Catalog(Part) number NOTE Provide information to help you use  your  device  more efficiently  Type BF applied part (B=  Body,  F=  Floating  applied part)  Operation instructions - Please refer to the operator’s manual.     Degree  of  protection  against harmful  ingress  of  water  as detailed in the IEC 60529:IPX3 -  Protected  against  spraying water  Manufactured date  Manufacturer  EC  REP  –  Authorized representative in the European community  CE marked in accordance with the Medical Device Directive  WEEE  –  Subject  to  waste electrical  and  electronic equipment regulations Refer  to  local  regulations  for disposal.  Tested  to  Federal Communications  Commission requirements  Stand-by  Statement of prescription  Explanatory  Sign  –  CLASS  1 LASER PRODUCT  UL  –  Underwriters Laboratories  certification  mark for  electrical  shock,  fire,  and mechanical hazards only  EU's  Restriction  of  Hazardous Substances  MEDLINE Logo  Fragile item, Handle carefully  Temperature limitation  This Way Up  Humiditylimitation
CUBEScanTM BioCon-900S Operator’s Manual   Page 60  Use no hook  Atmospheric  pressure limitation  Keep dry  Do not use blades to open
CUBEScanTM BioCon-900S Operator’s Manual   Page 61 12 REFERENCES  AIUM: Medical Ultrasound Safety, American Institute of Ultrasound in Medicine, Laurel, MD, 1994.  AIUM:  Acoustic  Output  Labeling  Standard  for  Diagnostic  Ultrasound  Equipment:  A  Standard  for  How Manufactures  Should  Specify  Acoustic  Output  Data,  Revision  1,  American  Institute  of  Ultrasound  in Medicine, Laurel, MD, 2008.  AIUM/NEMA:  Acoustic Output  Measurement  Standard for Diagnostic  Ultrasound  Equipment, Revision 3. NEMA  Standards  Publication  UD  2-2004;  American  Institute  of  Ultrasound  in  Medicine,  Laurel,  MD; National Electrical Association, Rosslyn, VA; 2004b.  Health  Canada:  “Guidelines  for  the  safe  use  of  diagnostic  ultrasound,”  Cat.  H46-2/01-255E,  Ministry  of Public Works and Government Services Canada, 2001.  Health Canada: Medical Devices Regulations SOR/98-282  MDD 93/42/EEC, Medical Device, Office for the Official Publications of the European Communities, 2003.  IEC 60601-1, Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance, International Electrotechnical Commission, 2012.  UL 60601-1, Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance, Underwriter Laboratories Inc, 2012.  CSA C22.2 No. 601.1B-90, Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance, Canadian Standards Association, 2012.  IEC  60601-2-37,  Medical  electrical  equipment  –  Part  2-37:  Particular  requirements  for  the  safety  of ultrasonic medical diagnostic and monitoring equipment, International Electrotechnical Commission, 2007.  ISO  10993-1,  Biological  Evaluation  of  Medical  Devices  Part  1:  Evaluation  and  Testing  with  in  a  risk management process, 2009.  FDA  Guidance:  Information  for  Manufacturers  Seeking  Marketing  Clearance  of  Diagnostic  Ultrasound Systems and Transducers, 2008  FDA Guidance: General Principles of Software Validation; Final Guidance for Industry and FDA Staff, 2002.  Woo Sung Hong,  Sun  Young Ham, Tong-Wook  Kim, Jeong-Seok Seo, Sang-Kuk  Yang. Usefulness  of a Sonographic  Bladder  Scan  for  Uroflowmetry  and  the  Evaluation  of  the  Anxiety  Level  Associated  with Uroflowmetry. The Korean Journal of Urology 2007:48(06):633-637  Chung B, Lee T, Yang J H. The Diagnostic Value of Portable Bladder Volume Measurement (BVMS) with Real  Bladder  Image  in  the  Measurement  of  Bladder  Volume  According  to  the  Different  Angling  of Transducer. The Korean Journal of Urology 2006:47(06):1320-1326  Bodker B, Lose G. Postoperative urinary retention in gynecologic patients. Int Urogenecol J Pelvic Floor Dsyfunct 2003;14:94-97  McNaughton-Collins  M,  Barry  MJ.  Managing  patients  with  lower  urinary  tract  symptoms  suggestive  of benign prostatic hyperplasia. Am J Med 2005;118:1331-9
CUBEScanTM BioCon-900S Operator’s Manual      U.S. Distributor Medline Industries, Inc. One Medline Place Mundelein, Illinois, 60060 USA Tel: 1-800-MEDLINE Website: www.medline.com   DongBang Acuprime 1 Forrest Units, Hennock Road East, Marsh Barton, Exeter EX2 8RU, U.K Tel: +44 1392 829500 Fax: +44 1392 823232  Local Distributor    Manufacturer Mcube Technology Co., Ltd. #803, 123, Bonghwasan-ro, Jungnang-gu, Seoul, Korea  Tel.: +82-2-3421-7780 Fax.: +82-2-3421-7076 E-mail: mcube@mcubetech.co.kr Web site: www.mcubetech.co.kr    Revision Date : AUG.21.2018 MUM-BioCon-900S EN(Rev.1.0)