Instruction Manual of WatchBP Pro
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
operational sequence
Actions
LCD display
* Power –Up the WatchBP
Pro
*Power –Up the Bluetooth
*First time switch on
Description
z Locate the black power switch
on the back side.
z Turn it on
z The green indicator light on the
top right front of the WatchBP
Pro will illuminate and remain
lit.
z Locate the silver BT switch
on the back side.
z Turn it on
z “Antenna” icon is flashing and
waiting for connection
Please refer to User’s guideline
for WatchBP Pro Middleware
Switch on the O/I button at the
back side of the unit, LCD full
segments display for 1 second
with backlight on
Auscultative Mode (one time measurement, on left arm).
*Measuring mode selection
Mode switch to AUSCULT.,
device ready for measuring.
Plug icon shows on LCD if
7.5V, 2.0A adaptor is used.
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
Remark
*Cuff connection
*Press Start/Stop button
*Auto zero reset
* Fuzzy inflation
* Deflation
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
Binding left-arm cuff on your
left arm and place chest-piece
of stethoscope on your brachial
artery correctly.
LCD full segment displays for
2 seconds with backlight on
Display shows “0” when device z
in auto zero reset.
Pumping starts, and cuff
pressure increases to working
pressure (for fuzzy inflation).
Pumping stops, pressure in
deflation by 2~4 mmHg/sec.
5 short “B”s
*Catch the SYS pressure.
*Catch the DIA pressure
* Measurement finishes
Press “M” button when a faint,
rhythmic tapping or thumping
sound is heard, which stands
for your SYS reading.
Press “M” button when your
diastolic reading reaches and
thumping sound stops, this is
DIA reading.
When measuring finishes, cuff z A long beep
deflates completely and a long
beep sounds.
AUT (auto-measured result),
AUS (Aus-measured result)
display for about 2 sec with
backlight on simultaneously.
*Display measured values
Device first displays of SP/DP, z
followed by MAP & PP values. z
SP/DP and MAP/PP values
display one after the other.
Backlight remains on 40 sec.
*Under auscult. mode, device auto
measures patient’s BP & pulse and
shows the results on the upper side
of LCD (AUT).
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
PP=SP-DP
MAP=
DP+1/3 PP
z
*Memory reading
*Withdraw from readings
Device auto memorizes one set z
of measured values. Press M
button, LCD shows AUE/AUS,
then memory readings, with
backlight on simultaneously
Memory SP/DP and MAP/PP
displays one after the other.
Backlight remains on 40 sec.
Press Start/Stop button or put z
the device untouched for 5’ to
terminate measurement display.
First Visit mode (3 times measurement in succession, on both arms).
*Measuring mode selection
*Cuff and
connection
blue
tooth
Mode switch to 1st and device
ready for measuring.
Binding left & right arm cuffs
on your both arms correctly.
Push blue-tooth switch on the
back side of device (ON/OFF
sculpted) to “ON” for computer
connection if remote control.
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
*Press Start/Stop button or
remote control.
*60 sec count down before
measurement.
* Auto zero reset
* First turn fuzzy inflation.
*First turn deflation.
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
LCD full segment displays for
1 second with backlight on
Beeper sounds 5 seconds before z
the 1st measurement.
Press Start/Stop button or
remote control to skip over
count down and begin 1st turn
measuring immediately.
5 short “B”s
Display shows “0” for 2 sec z
when device in auto zero reset.
5 short “Bs”
Pumping starts & cuff pressure
increases to working pressure
(for fuzzy inflation).
Pumping stops, pressure in
deflation by 4~6 mmHg/sec.
*Pulse detected
Detect the pulse, “♥” icon start z
to flash and a beep is heard for
every heartbeat.
Beeper each
pulse
*First turn measurement
finishes
After first turn measuring, cuff z
deflates completely, and a long
beeper sounds
After another two close beeps,
device comes into 60 sec count
down.
A long “B”,
and 2 short
Beeper sounds 5 seconds before z
the 2nd measurements.
5 short “Bs”
Press Start/Stop button or
remote control to skip over
count down and begin 2nd turn
measuring immediately.
* Auto zero reset
Display shows “0” for 2 sec z
when device in auto zero reset.
* 2nd turn fuzzy inflation.
Pumping starts & cuff pressure z
increases to working pressure
(for fuzzy inflation).
*60 sec count down
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
5 short “Bs”
*2nd turn deflation.
Pumping stops, pressure in z
deflation by 4~6 mmHg/sec.
*Pulse detected
Detect the pulse, “♥” icon start z
to flash and a beep is heard for
every heartbeat.
A beep each
pulse
*2nd turn
finishes
After 2nd turn measuring, cuff z
deflates completely, and a long
beeper sounds
After another two close beeps,
device comes into 60 sec count
down.
A long “B”,
and 2 short
Beeper sounds 5 seconds before z
the 2nd measurements.
5 short “Bs”
Press Start/Stop button or
remote control to skip over
count down and begin 3rd turn
measuring immediately.
Display shows “0” for 2 sec z
when device in auto zero reset.
measurement
*60 sec count down
*Auto zero reset
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
5 short “Bs”
*3rd turn fuzzy inflation.
Pumping starts & cuff pressure
increases to working pressure
(for fuzzy inflation).
*3rd turn deflation.
Pumping stops, pressure in
deflation by 4~6 mmHg/sec.
*Pulse detected
Detect the pulse, “♥” icon start z
to flash and a beep is heard for
every heartbeat.
A beep each
pulse
*The 3rd turn measuring
finishes
After 3rd turn measuring, cuff z
deflates completely, and a long
beeper sounds
A long “B”.
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
* Display of average values
LCD displays values averaged
from 3 times measurement with
backlight on simultaneously.
Left arm BP values will flash 5
times by default if difference
between L and R values are less
than 10 mmHg, otherwise, the
higher (L or R) flashes 5 times.
Display of SP/DP, following
which, MAP & PP shows on.
SP/DP and MAP/PP values
display one after the other till
device auto off.
Backlight remains on 40 sec.
*Data transmission.
Through blue tooth, measured
values are submitted to
computer for further analysis.
*Memory reading--average
values display.
Device auto memorizes 3 times
measured values and their
average readings.
Press M button, LCD shows an
“A” then average readings,
with
backlight
on
simultaneously
Averaged SP/DP and MAP/PP
displays one after the other.
Backlight remains on 40 sec.
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
* Memory reading--3rd turn
measured values display.
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
Press M button again, LCD
shows a “3” then displays of
the 3rd turn measured values.
SP/DP and MAP/PP displays
one after the other.
*Memory reading--2nd turn
measured values display.
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
Press M button again, LCD
shows a “2” then displays of
the 2nd turn measured values.
SP/DP and MAP/PP displays
one after the other.
*Memory reading--1st turn
measured values display.
*Withdraw from readings
Press M button again, LCD
shows a “1” then display of the
1st turn measured values.
SP/DP and MAP/PP displays
one after the other.
Press Start/Stop button or put
the device untouched for 5’ to
terminate measurement display.
Follow up mode (3 times measurement in succession, optional on left or right arm).
*Measuring mode selection
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
Mode switch to “FOLLOW UP”
and device ready for measuring.
*Cuff and
connection
blue
tooth
Binding left & right arm cuffs
on your both arms correctly.
Push blue-tooth switch on the
back side of device (ON/OFF
sculpted) to “ON” for computer
connection if remote control.
Hold down M button for 3 sec,
memory shows with backlight
on, till L or R flashes.
Press M button to choose that
preferred arm for measurement,
and press Start/Stop button for
confirmation.
*Press Start/Stop button or
remote control.
LCD full segment displays for
1 second with backlight on
*60 sec count down before
measurement.
Beeper sounds 5 seconds z
before the 1st measurement.
Press Start/Stop button or
remote control to skip over
count down and begin 1st turn
measuring immediately.
5 short “B”s
Display shows “0” for 2 sec z
when device in auto zero reset.
5 short “B”s
*Choose one preferred arm
for measurement.
*Auto zero reset
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
z
Pumping starts & cuff
pressure increases to
working pressure (for
fuzzy inflation).
Pumping stops, pressure in
deflation by 4~6 mmHg/sec.
*Pulse detected
Detect the pulse, “♥” icon start z
to flash and a beep is heard for
every heartbeat.
*First turn measurement
finishes
After first turn measuring, cuff
deflates completely and device
comes into 60 sec count down.
Beeper sounds 5 seconds z
before the 2nd measurements.
Press Start/Stop button
or remote control to
skip over count down
and
begin
2nd
turn
measuring immediately.
* First turn fuzzy
inflation.
*First turn deflation.
*60 sec count down
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
Beeper each
pulse
sh
ort
“B
”s
* Auto zero reset
Display shows “0” for 2 sec z
when device in auto zero reset.
* 2nd turn fuzzy inflation.
Pumping starts & cuff pressure
increases to working pressure
(for fuzzy inflation).
*2nd turn deflation.
Pumping stops, pressure in
deflation by 4~6 mmHg/sec.
*Pulse detected
Detect the pulse, “♥” icon start z
to flash and a beep is heard for
every heartbeat.
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
5 short “Bs”
A beep each
pulse
After 2nd turn measuring, cuff
deflates completely.
z If a difference of 5mmHg exists
between 1st&2nd time measured
sys or dia pressure, device auto
comes into 60 sec count down
and takes the 3rd measurement.
*Measuring finishes with a long
beep and average values display if
difference between 1st & 2nd turn
measured sys or dia pressure is less
than 5 mmHg.
*2nd turn measurement
finishes
*60 sec count down
Beeper sounds 5 sec before the z
3rd turn measurement.
Press Start/Stop button
or remote control to
skip over count down
and
begin
3rd
turn
measuring immediately.
*Auto zero reset
Display shows “0” for 2 sec z
when device in auto zero reset.
* 3rd turn fuzzy inflation.
Pumping starts & cuff pressure
increases to working pressure
(for fuzzy inflation).
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
sh
ort
“B
”s
5 short “Bs”
*3rd turn deflation.
Pumping stops, pressure in
deflation by 2~4 mmHg/sec.
*Pulse detected
Detect the pulse, “♥” icon start z
to flash and a beep is heard for
every heartbeat.
A beep each
pulse
*3rd turn
finishes
measurement
After 3rd turn measuring, cuff z
deflates completely, and a long
beeper sounds
A long “B”.
* Display of average values
LCD displays values averaged
from 3 times measurement with
backlight on simultaneously.
Display of SP/DP, following
which, MAP & PP shows on.
SP/DP and MAP/PP values
display one after the other till
device auto off.
Backlight remains on 40 sec.
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
*Data transmission.
Through blue tooth, measured
values are submitted to
computer for further analysis.
*Memory reading--average
values display.
Device auto memorizes 3 times
measured values and their
average readings.
Press M button, LCD shows an
“A” then average reading, with
backlight on simultaneously.
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
Averaged SP/DP and MAP/PP
displays one after the other.
Backlight remains on 40 sec.
* Memory reading--3rd turn
measured values display.
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
Press M button again, LCD
shows a “3” then displays of
the 3rd turn measured values.
SP/DP and MAP/PP displays
one after the other.
*Memory reading--2nd turn
measured values display.
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
Press M button again, LCD
shows a “2” then displays of
the 2nd turn measured values.
SP/DP and MAP/PP displays
one after the other.
*Memory reading--1st turn
measured values display.
Press M button again, LCD
shows a “1” then display of the
1st turn measured values.
SP/DP and MAP/PP displays
one after the other.
*Withdraw from readings
Press Start/Stop button or put
the device untouched for 5’ to
terminate measurement display.
*Memories deletion.
Hold down M button for 7 sec z
till 3 short beepers are heard
and bars appear on LCD with
backlight on simultaneously,
which means memory readings
are deleted successfully.
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
3 short “B”
* Battery low indication
*Turn off the device.
When voltage inside batteries is
lower to 4.8V, LCD shows low
battery icon and batteries begin
to be charged if adaptor is
connected.
Once voltage inside batteries is
lower to 4.6V, LCD shows low
battery icon, then batteries must
be charged again.
Switch off O/I button at the
back side of the unit
Memories won’t be stored once
the device is turned off.
Federal Communications Commission (FCC) Statement
15.21
You are cautioned that changes or modifications not expressly
approved by the part responsible for compliance could void the user’s
authority to operate the equipment.
15.105(b)
This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause
harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to
radio or television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
-Reorient or relocate the receiving antenna.
-Increase the separation between the equipment and receiver.
-Connect the equipment into an outlet on a circuit different from that to which the receiver is
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro
connected.
-Consult the dealer or an experienced radio/TV technician for help.
Operation is subject to the following two conditions:
1) this device may not cause interference and
2) this device must accept any interference, including interference
that may cause undesired operation of the device.
FCC RF Radiation Exposure Statement:
This equipment complies with FCC radiation exposure limits set forth
for an uncontrolled environment. End users must follow the specific
operating instructions for satisfying RF exposure compliance. This
transmitter must not be co-located or operating in conjunction with
any other antenna or transmitter.
© 2007 Microlife Corporation
Document: Instruction Manual of WatchBP Pro

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : No
XMP Toolkit                     : 3.1-701
Producer                        : Acrobat Distiller 7.0 (Windows)
Modify Date                     : 2007:06:01 16:37:20+08:00
Create Date                     : 2007:06:01 16:37:20+08:00
Format                          : application/pdf
Title                           : untitled
Document ID                     : uuid:d0e9d32b-6be8-4d05-8fc3-e5feb43feb5b
Instance ID                     : uuid:e94ea0c5-40f3-4828-912c-d37380a18167
Page Count                      : 24
EXIF Metadata provided by EXIF.tools
FCC ID Filing: U7I-TWIN200

Navigation menu