BeneVision TMS60 Telemetry Monitoring System Operator’s Manual Introduction © Copyright 2015 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved. For this Operator’s Manual, the issue date is 2015-06. TMS60 Operator’s Manual Intellectual Property Statement Introduction Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this Mindray product and this manual. This manual may refer to information protected by copyrights or patents and does not convey any license under the patent rights of Mindray, nor the rights of others. Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden. Release, amendment, reproduction, distribution, rental, adaption and translation of this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden. , and are the registered trademarks or trademarks owned by Mindray in China and other countries. All other trademarks that appear in this manual are used only for editorial purposes without the intention of improperly using them. They are the property of their respective owners. Manufacturer’s Responsibility Contents of this manual are subject to changes without prior notice. All information contained in this manual is believed to be correct. Mindray shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing, performance, or use of this manual. Mindray is responsible for the effects on safety, reliability and performance of this product, only if: II ■ all installation operations, expansions, changes, modifications and repairs of this product are conducted by Mindray authorized personnel; ■ the electrical installation of the relevant room complies with the applicable national and local requirements; ■ the product is used in accordance with the instructions for use. TMS60 Operator’s Manual Introduction Warranty WARNING • • Only skilled/trained clinical professionals should operate this equipment. It is important for the hospital or organization that employs this equipment to carry out a reasonable service/maintenance plan. Neglect of this may result in machine breakdown or personal injury. Warranty Mindray warrants that components within its products will be free from defects in workmanship and materials for a period of three years from the date of purchase except that disposable or one-time use products are warranted to be free from defects in workmanship and materials up to a date one year from the date of purchase or the date of first use, whichever is sooner. This warranty does not cover consumable items such as, but not limited to, batteries, external cables, and sensors. Mindray shall not be liable for any incidental, special, or consequential loss, damage, or expense directly or indirectly arising from the use of its products. Liability under this warranty and the buyer’s exclusive remedy under this warranty is limited to servicing or replacing the affected products, at Mindray option, at the factory or at an authorized distributor, for any product which shall under normal use and service appear to Mindray to have been defective in material or workmanship. Recommended preventative maintenance, as prescribed in the service manual, is the responsibility of the user and is not covered by this warranty. No agent, employee, or representative of Mindray has any authority to bind Mindray to any affirmation, representation, or warranty concerning its products, and any affirmation, representation or warranty made by any agent, employee, or representative shall not be enforceable by buyer or user. THIS WARRANTY IS EXPRESSLY IN LIEU OF, AND MINDRAY EXPRESSLY DISCLAIMS, ANY OTHER EXPRESS OR IMPLIED WARRANTIES, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY, NON-INFRINGEMENT, OR FITNESS FOR A PARTICULAR PURPOSE, AND OF ANY OTHER OBLIGATION ON THE PART OF MINDRAY. Damage to any product or parts through misuse, neglect, accident, or by affixing any non-standard accessory attachments, or by any customer modification voids this warranty. Mindray makes no warranty whatsoever in regard to trade accessories, such being subject to the warranty of their respective manufacturers. A condition of this warranty is that the equipment or accessories which are claimed to be defective be returned when authorized, freight prepaid to Mindray DS USA, Inc., Mahwah, New Jersey 07430 or its authorized representative. Mindray shall not have any responsibility in the event of loss or damage in transit. TMS60 Operator’s Manual III Company Contact Introduction Exemptions Mindray's obligation or liability under this warranty does not include any transportation or other charges or liability for direct, indirect or consequential damages or delay resulting from the improper use or application of the product or the use of parts or accessories not approved by Mindray or repairs by people other than Mindray authorized personnel. This warranty does not extend to: ■ Malfunction or damage caused by improper use or man-made failure. ■ Malfunction or damage caused by unstable or out-of-range power input. ■ Malfunction or damage caused by force majeure events, such as (i) flood, fire and earthquake or other similar elements of nature or acts of God; (ii) riots, war, civil disorders, rebellions, or revolutions in any country; or (iii) any other cause beyond the reasonable control of Mindray. ■ Malfunction or damage caused by improper operation or repair by unqualified or unauthorized service people. ■ Malfunction of the instrument or part whose serial number is not legible. ■ Others not caused by instrument or part itself. Company Contact Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Address: Mindray Building, Keji 12th Road South, Hi-tech industrial park, Nanshan, Shenzhen 518057, P.R.China Website: www.mindray.com E-mail Address: service@mindray.com Tel: +86 755 81888998 Fax: +86 755 26582680 Distributor: Mindray DS USA, Inc. Address: 800 MacArthur Boulevard, Mahwah, New Jersey 07430 USA Tel: 1.800.288.2121, 1.201.995.8000 Website: www.mindray.com EC-Representative: Shanghai International Holding Corp. GmbH (Europe) Address: Eiffestraβe 80, 20537 Hamburg, Germany Tel: 0049-40-2513175 IV TMS60 Operator’s Manual Introduction Preface Fax: 0049-40-255726 Preface Manual Purpose This manual contains the instructions necessary to operate the product safely and in accordance with its function and intended use. Observance of this manual is a prerequisite for proper product performance and correct operation and ensures patient and operator safety. This manual is based on the maximum configuration and therefore some contents may not apply to your product. If you have any questions, please contact Mindray. This manual is an integral part of the product. It should always be kept close to the equipment so that it can be obtained conveniently when needed. Intended Audience This manual is geared for clinical professionals who are expected to have a working knowledge of medical procedures, practices and terminology as required for monitoring of critically ill patients. Illustrations All illustrations in this manual serve as examples only. They may not necessarily reflect the setup or data displayed on your patient monitor. Conventions ■ Italic text is used in this manual to quote the referenced chapters or sections. ■ [ ] is used to enclose screen texts. ■ → is used to indicate operational procedures. TMS60 Operator’s Manual Preface Introduction This page intentionally left blank. VI TMS60 Operator’s Manual Contents 1 Safety ........................................................................................................................................ 1 - 1 1.1 Safety Information ..................................................................................................................................... 1 - 2 1.1.1 Warnings ........................................................................................................................................ 1 - 3 1.1.2 Cautions .......................................................................................................................................... 1 - 4 1.1.3 Notes ................................................................................................................................................ 1 - 5 1.2 Equipment Symbols .................................................................................................................................. 1 - 5 2 General Product Description ................................................................................................... 2 - 1 2.1 Intended Use ................................................................................................................................................ 2 - 2 2.2 Applied Parts ................................................................................................................................................ 2 - 2 2.3 Key Features ................................................................................................................................................. 2 - 2 2.4 System Components ................................................................................................................................. 2 - 3 2.5 TD60 Physical View .................................................................................................................................... 2 - 3 2.6 Antenna Array .............................................................................................................................................. 2 - 5 2.7 Telemetry Receiver (RC60) ...................................................................................................................... 2 - 5 2.8 Touch Screen Display ................................................................................................................................ 2 - 5 2.8.1 Display Screen .............................................................................................................................. 2 - 6 2.8.2 On-Screen Keyboard .................................................................................................................. 2 - 7 3 Getting Started ......................................................................................................................... 3 - 1 3.1 Unpacking and Checking ........................................................................................................................ 3 - 2 3.2 Environmental Requirements ................................................................................................................ 3 - 3 3.3 Connecting the ECG Leadwire ............................................................................................................... 3 - 4 3.4 Installing the Batteries .............................................................................................................................. 3 - 5 3.4.1 Installing the Lithium-ion Rechargeable Battery ............................................................. 3 - 6 3.4.2 Installing the AA Batteries ........................................................................................................ 3 - 7 3.5 Powering On the Unit ............................................................................................................................... 3 - 8 3.6 Understanding Touch Gestures ............................................................................................................ 3 - 9 3.7 Basic Operations ......................................................................................................................................... 3 - 9 3.7.1 Understanding the Screen Display Orientation .............................................................3 - 10 3.7.2 Browsing the Screen Display ................................................................................................3 - 10 3.7.3 Switching the Screen Display Orientation .......................................................................3 - 10 3.7.4 Flipping the Landscape Display ...........................................................................................3 - 10 3.7.5 Displaying the Quick Keys Area ...........................................................................................3 - 11 3.7.6 Entering the Main Menu .........................................................................................................3 - 11 3.7.7 Turning the Display Off ...........................................................................................................3 - 12 3.7.8 Turning the Display On ...........................................................................................................3 - 13 3.7.9 Unlocking the Screen ..............................................................................................................3 - 13 TMS60 Operator’s Manual 3.7.10 Acknowledging the Nurse Call ..........................................................................................3 - 13 3.8 Using the Pouch ........................................................................................................................................3 - 14 3.8.1 Securing the Pouch ..................................................................................................................3 - 14 4 User Configurations ................................................................................................................. 4 - 1 4.1 Introduction .................................................................................................................................................. 4 - 2 4.2 Configuring the Display ........................................................................................................................... 4 - 2 4.2.1 Setting the Default Display Orientation ............................................................................. 4 - 2 4.2.2 Understanding Portrait Orientation Display Rules ......................................................... 4 - 2 4.2.3 Setting the Portrait Display ..................................................................................................... 4 - 2 4.2.4 Understanding Landscape Orientation Display Rules ................................................... 4 - 3 4.2.5 Setting the Landscape Display ............................................................................................... 4 - 4 4.2.6 Setting the Display Brightness ............................................................................................... 4 - 4 4.3 Configuring the Audio Volume ............................................................................................................. 4 - 5 5 Patient Management ............................................................................................................... 5 - 1 5.1 Introduction .................................................................................................................................................. 5 - 2 5.2 Admitting a Patient .................................................................................................................................... 5 - 2 5.3 Changing the Patient Category ............................................................................................................. 5 - 2 5.4 Placing a Device in Standby ................................................................................................................... 5 - 3 5.5 Resume Monitoring ................................................................................................................................... 5 - 4 5.6 Discharging the Patient ............................................................................................................................ 5 - 4 5.6.1 Selecting the [Discharge Patient] menu ............................................................................. 5 - 4 5.6.2 Restarting the TD60 ................................................................................................................... 5 - 5 6 Alarms ....................................................................................................................................... 6 - 1 6.1 Introduction .................................................................................................................................................. 6 - 2 6.2 Alarm Categories ........................................................................................................................................ 6 - 2 6.3 Alarm Levels ................................................................................................................................................. 6 - 2 6.4 Alarm Indicators .......................................................................................................................................... 6 - 3 6.4.1 Alarm Light .................................................................................................................................... 6 - 3 6.4.2 Alarm Tones .................................................................................................................................. 6 - 3 6.4.3 Alarm Messages ........................................................................................................................... 6 - 4 6.4.4 Alarm Status Symbols ................................................................................................................ 6 - 4 6.5 Configuring the Alarms ............................................................................................................................ 6 - 5 6.6 Resetting the Alarms ................................................................................................................................. 6 - 5 7 Monitoring ECG ........................................................................................................................ 7 - 1 7.1 Introduction .................................................................................................................................................. 7 - 2 7.2 Safety .............................................................................................................................................................. 7 - 2 7.3 Preparation for Monitoring ECG ........................................................................................................... 7 - 3 7.3.1 Preparing the Patient’s Skin .................................................................................................... 7 - 3 7.3.2 Positioning the Electrodes ....................................................................................................... 7 - 3 TMS60 Operator’s Manual 7.3.3 Setting ECG Lead Labeling ...................................................................................................... 7 - 5 7.3.4 Placing the Electrodes ............................................................................................................... 7 - 5 7.3.5 Checking the Lead Placement ................................................................................................ 7 - 9 7.3.6 Checking the Paced Status ....................................................................................................7 - 10 7.4 Changing the ECG Settings ...................................................................................................................7 - 11 7.4.1 Configuring the ECG Setup ...................................................................................................7 - 11 7.4.2 ECG Leadwire Types .................................................................................................................7 - 12 7.4.3 Configuring the ECG Waveforms ........................................................................................7 - 12 7.4.4 Configuring the Pacer .............................................................................................................7 - 13 7.4.5 Configuring the ECG Waveform Size .................................................................................7 - 14 7.5 Understanding the ECG Display ..........................................................................................................7 - 14 7.5.1 HR Digital Area ...........................................................................................................................7 - 14 7.5.2 About the HR Digital Area ......................................................................................................7 - 14 7.5.3 ECG Waveform Area .................................................................................................................7 - 15 7.5.4 About the ECG Waveform Area ............................................................................................7 - 15 8 Monitoring SpO2 (Optional) .................................................................................................... 8 - 1 8.1 Introduction .................................................................................................................................................. 8 - 2 8.2 Measurement Limitations ....................................................................................................................... 8 - 2 8.3 Safety .............................................................................................................................................................. 8 - 3 8.4 Connecting the SpO2 Module ............................................................................................................... 8 - 4 8.5 Changing the SpO2 Settings .................................................................................................................. 8 - 5 8.5.1 Configuring the SpO2 Setup ................................................................................................... 8 - 5 8.5.2 Configuring the SpO2 Waveform .......................................................................................... 8 - 8 8.6 SpO2 Measurement ................................................................................................................................... 8 - 8 8.6.1 Identifying SpO2 Modules ....................................................................................................... 8 - 8 8.6.2 Applying the Sensor ................................................................................................................... 8 - 8 8.7 Understanding the SpO2 Display .......................................................................................................8 - 10 8.7.1 SpO2 Digital Area ......................................................................................................................8 - 10 8.7.2 About the SpO2 Digital Area ................................................................................................8 - 11 8.7.3 SpO2 Waveform Area ..............................................................................................................8 - 11 8.7.4 About the SpO2 Waveform Area .........................................................................................8 - 11 8.8 Masimo Information ................................................................................................................................8 - 12 9 Monitoring with the TD60 at the CS ....................................................................................... 9 - 1 9.1 Introduction .................................................................................................................................................. 9 - 2 9.2 Physiological Alarms ................................................................................................................................. 9 - 2 9.2.1 Factory Default Parameter Alarm Limits ............................................................................. 9 - 3 9.2.2 Parameter Alarm Responses ................................................................................................... 9 - 4 9.2.3 Factory Default Parameter Alarm Settings ........................................................................ 9 - 5 9.2.4 Factory Default Arrhythmia Alarm Settings ...................................................................... 9 - 6 9.2.5 Arrhythmia Threshold Settings ............................................................................................9 - 10 9.3 ECG Monitoring .........................................................................................................................................9 - 12 9.3.1 HR Settings ..................................................................................................................................9 - 12 TMS60 Operator’s Manual 9.3.2 Waveform Setup ........................................................................................................................9 - 13 9.3.3 Other Settings ............................................................................................................................9 - 15 9.3.4 ECG Display .................................................................................................................................9 - 16 9.4 QT Monitoring ...........................................................................................................................................9 - 18 9.4.1 Measurement Limitations ......................................................................................................9 - 18 9.4.2 QT Settings ..................................................................................................................................9 - 18 9.4.3 QT View Menu (Only for Mindray ECG Algorithm) ........................................................9 - 20 9.4.4 QT Display ....................................................................................................................................9 - 21 9.5 ST Monitoring ............................................................................................................................................9 - 22 9.5.1 Measurement Limitations ......................................................................................................9 - 23 9.5.2 ST Settings ...................................................................................................................................9 - 23 9.5.3 Adjusting ST Measurement Points .....................................................................................9 - 24 9.5.4 ST Display .....................................................................................................................................9 - 25 9.5.5 ST Segment Display ..................................................................................................................9 - 26 9.6 Arrhythmia Monitoring ..........................................................................................................................9 - 27 9.6.1 Measurement Limitations ......................................................................................................9 - 27 9.6.2 Understanding the Arrhythmia Events .............................................................................9 - 27 9.6.3 Arrhythmia Settings .................................................................................................................9 - 30 9.6.4 Relearning ....................................................................................................................................9 - 30 9.6.5 Configuring the QRS Threshold ...........................................................................................9 - 31 9.7 SpO2 Monitoring ......................................................................................................................................9 - 33 9.7.1 Measurement Limitations ......................................................................................................9 - 33 9.7.2 SpO2 Settings .............................................................................................................................9 - 33 9.7.3 SpO2 Display ...............................................................................................................................9 - 34 10 Configuring the TD60 .......................................................................................................... 10 - 1 10.1 Introduction .............................................................................................................................................10 - 2 10.2 Maintenance Menu ...............................................................................................................................10 - 2 10.2.1 Entering the Maintenance menu ......................................................................................10 - 2 10.2.2 Configuring the General Menu ..........................................................................................10 - 2 10.2.3 Configuring the Alarms Menu ...........................................................................................10 - 3 10.2.4 Quick Keys Menu ....................................................................................................................10 - 5 10.2.5 Configuring the Defaults Menu .........................................................................................10 - 6 10.2.6 Transferring a Configuration ..............................................................................................10 - 7 10.2.7 Screen Lock Menu ..................................................................................................................10 - 8 10.2.8 Changing the Passwords .....................................................................................................10 - 9 10.2.9 Changing the Device Name ................................................................................................10 - 9 10.2.10 Demo Mode ...........................................................................................................................10 - 9 10.2.11 Service Menu .........................................................................................................................10 - 9 11 Battery ................................................................................................................................... 11 - 1 11.1 Introduction .............................................................................................................................................11 - 2 11.2 Safety ..........................................................................................................................................................11 - 2 11.3 Installing the Battery .............................................................................................................................11 - 3 11.4 Checking the Battery Charge Status ...............................................................................................11 - 4 TMS60 Operator’s Manual 11.5 Removing the Battery ...........................................................................................................................11 - 4 11.6 Charging the Rechargeable Lithium-ion Battery ........................................................................11 - 5 11.7 Storing the Batteries .............................................................................................................................11 - 6 11.7.1 Storing Rechargeable Lithium-ion Battery ....................................................................11 - 6 11.7.2 Storing AA Batteries ...............................................................................................................11 - 6 11.8 Maintaining the Rechargeable Lithium-ion Battery ..................................................................11 - 7 11.9 Disposing of the Batteries ...................................................................................................................11 - 8 11.9.1 Disposing of the Rechargeable Lithium-ion Battery .................................................11 - 8 11.9.2 Disposing of the AA Batteries ............................................................................................11 - 8 12 Troubleshooting ................................................................................................................... 12 - 1 12.1 General Problems ...................................................................................................................................12 - 2 12.1.1 Troubleshooting Tools .........................................................................................................12 - 2 12.1.2 Problem List ..............................................................................................................................12 - 2 12.2 Physiological Alarm Messages at the CS .......................................................................................12 - 3 12.3 Technical Alarm Messages at the TD60 .........................................................................................12 - 5 13 Cleaning and Disinfecting ................................................................................................... 13 - 1 13.1 Introduction .............................................................................................................................................13 - 2 13.2 Safety Information .................................................................................................................................13 - 2 13.3 Cleaning of the TD60 ............................................................................................................................13 - 3 13.4 Cleaning the Reusable ECG Leadwires, SpO2 Modules and Sensors ..................................13 - 4 13.5 Cleaning the Battery and Battery Compartment ........................................................................13 - 5 13.6 Disinfection ..............................................................................................................................................13 - 5 13.7 Sterilization ..............................................................................................................................................13 - 5 14 Maintenance ......................................................................................................................... 14 - 1 14.1 Introduction .............................................................................................................................................14 - 2 14.2 Safety ..........................................................................................................................................................14 - 2 14.3 Regular Check ..........................................................................................................................................14 - 3 14.3.1 Power-on Test ..........................................................................................................................14 - 3 14.3.2 Battery Check ...........................................................................................................................14 - 3 14.4 Maintenance and Testing Schedule ................................................................................................14 - 4 14.5 Checking the System Information ...................................................................................................14 - 4 14.6 Verifying the ECG at the TD60 ...........................................................................................................14 - 4 15 Accessories ............................................................................................................................ 15 - 1 15.1 ECG Accessories ......................................................................................................................................15 - 2 15.1.1 ECG Electrodes .........................................................................................................................15 - 2 15.1.2 ECG Leadsets ............................................................................................................................15 - 3 15.2 SpO2 Accessories ...................................................................................................................................15 - 3 15.2.1 Masimo SpO2 Module ..........................................................................................................15 - 3 TMS60 Operator’s Manual 15.2.2 Masimo SpO2 Sensor ............................................................................................................15 - 4 15.2.3 Nonin SpO2 Module ..............................................................................................................15 - 4 15.2.4 Nonin SpO2 Sensor ................................................................................................................15 - 4 15.3 Miscellaneous ..........................................................................................................................................15 - 5 A Product Specifications ............................................................................................................. A - 1 A.1 Classifications ..............................................................................................................................................A - 2 A.2 Environmental Specifications ................................................................................................................A - 2 A.3 Power Supply Specifications ..................................................................................................................A - 3 A.3.1 TD60 ................................................................................................................................................A - 3 A.3.2 RC60 .................................................................................................................................................A - 3 A.3.3 Central Charger ...........................................................................................................................A - 4 A.4 Physical Specifications .............................................................................................................................A - 4 A.4.1 TD60 ................................................................................................................................................A - 4 A.4.2 RC60 .................................................................................................................................................A - 4 A.4.3 Central Charger ...........................................................................................................................A - 4 A.5 Hardware Specifications ..........................................................................................................................A - 5 A.5.1 TD60 ................................................................................................................................................A - 5 A.5.2 RC60 .................................................................................................................................................A - 6 A.5.3 Central Charger ...........................................................................................................................A - 6 A.6 Wireless Specification ...............................................................................................................................A - 7 A.6.1 Technique Specification ...........................................................................................................A - 7 A.6.2 Implemented Functions ...........................................................................................................A - 7 A.6.3 Function Specification ..............................................................................................................A - 8 A.7 Mindray Patient Area Network (PAN) Specification ......................................................................A - 9 A.7.1 Technique Specification ...........................................................................................................A - 9 A.7.2 Implemented Function .............................................................................................................A - 9 A.7.3 Function Specification ..............................................................................................................A - 9 A.8 Measurement Specifications ............................................................................................................... A - 10 A.8.1 ECG ................................................................................................................................................ A - 10 A.8.2 SpO2 ............................................................................................................................................. A - 14 B EMC ............................................................................................................................................ B - 1 C FCC Compliance ........................................................................................................................ C - 1 D Symbols and Abbreviations ....................................................................................................D - 1 D.1 Units ............................................................................................................................................................... D - 2 D.2 Symbols ........................................................................................................................................................ D - 3 D.3 Abbreviations ............................................................................................................................................. D - 3 E Anomaly .....................................................................................................................................E - 1 E.1 Anomaly Description ................................................................................................................................. E - 1 TMS60 Operator’s Manual 1 Safety Safety Information...................................................................................................1-2 Equipment Symbols................................................................................................1-5 TMS60 Operator’s Manual 1-1 Safety Information 1.1 Safety Safety Information WARNING • Indicates a potential hazard or unsafe practice that, if not avoided, could result in death or serious injury. CAUTION • Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage. NOTE • 1-2 Provides application tips or other useful information to ensure that you get the most from your product. TMS60 Operator’s Manual Safety 1.1.1 Safety Information Warnings WARNING • • • • • • • • • • • The TD60 is intended to be used for a single patient at a time. The Telemetry Monitoring System (TMS60) must be operated by medical personnel in hospitals or medical institutions. To avoid explosion hazard, do not use the equipment in the presence of oxygen-rich atmospheres, flammable anesthetics, or other flammable agents. Do not use this equipment in conjunction with Electro Surgical Unit (ESU). Do not expose the equipment to a Magnetic Resonance (MR) environment. ◆ Thermal injury and burns may occur due to the metal components of the equipment which can heat during MR scanning. ◆ The equipment may present a risk of projectile injury due to the presence of ferromagnetic materials which can be attracted by the MR magnet core. ◆ The leadwires and electrodes will generate artifacts in the MR image. ◆ The equipment will not function properly due to the strong magnetic and radio frequency fields generated by the MR scanner. Before putting the system into operation, the operator must verify that the equipment, connecting cables and accessories are in correct working order and operating condition. Do not come into contact with the patient during defibrillation. Otherwise serious injury or death could result. Do not touch the patient and live parts simultaneously. Do not open the equipment housings. All servicing and future upgrades must be carried out by trained and authorized personnel. Do not rely exclusively on the audible alarm system for monitoring. Adjustment of alarm volume to a low level may result in a hazard to the patient. Always keep the patient under close surveillance. The physiological data and alarm messages displayed on the system are for reference only and cannot be directly used for diagnostic interpretation. TMS60 Operator’s Manual 1-3 Safety Information Safety WARNING • • • 1.1.2 Do not operate the touch screen with water on the hand. Only use parts and accessories specified in this manual. Route, wrap and secure the cables to avoid inadvertent disconnection, stumbling and entanglement. Cautions CAUTION • • • • • • • • • 1-4 Do not let the display directly touch the patient when the display is on. When the central station presents the alarm “No RF Signal”, the setting being performed on the TD60 may not be transferred to the central station. Check the patient condition and the settings on the central station. When disposing of the packaging material, be sure to observe the applicable waste control regulations and keep it out of children’s reach. After the configurations, such as the patient category, paced status, are changed at the TD60, the medical personnel shall be check those configurations at the CS to make sure both of the configurations are consistent. Magnetic and electrical fields are capable of interfering with the proper performance of the equipment. For this reason make sure that all external equipment operated in the vicinity of the equipment comply with the relevant EMC requirements. Mobile phone, X-ray equipment or MRI equipment are a possible source of interference as they may emit higher levels of electromagnetic radiation. Always install or carry the equipment properly to avoid damage caused by drop, impact, strong vibration or other mechanical force. Dry the equipment immediately in case of rain or water spray. The system generates and uses the Radio Frequency (RF) energy. If it is not installed correctly or not used as per the manual, RF interference to other equipment could result. Signal quality can be impacted on an ambulatory patient by the construction materials used within the hospital. TMS60 Operator’s Manual Safety Equipment Symbols CAUTION • • 1.1.3 At the end of its service life, the equipment, and its accessories, must be disposed of in compliance with the guidelines regulating the disposal of such products. If you have any questions concerning disposal of the equipment, please contact Mindray. When programming the frequency for a transmitter,the frequencies allocated to all other transmitters need to be considered to ensure that no two transmitters are programmed with the same frequency. Notes NOTE • • • • 1.2 Put the equipment in a location where you can easily see the screen, and access the operating controls. The software was developed in compliance with IEC60601-1-4. The possibility of hazards arising from software errors is minimized. This manual describes all features and options. Your equipment may not have all of them. Keep this manual in the vicinity of the equipment so that it can be obtained conveniently when needed. Equipment Symbols Symbol TMS60 Operator’s Manual Description Symbol Description Power On/Off key Main menu key Nurse call key Alternating current (AC) Defibrillation-proof Type CF applied part Serial number Date of Manufacture Symbol for “MANUFACTURER” 1-5 Equipment Symbols Safety Symbol MR Description Symbol Description MR Unsafe – do not subject to magnetic resonance imaging (MRI) IPX7 Protection against fluid ingress Interference may occur in the vicinity of equipment marked with this symbol General warning sign Refer to instruction manual/booklet The presence of this label indicates the machine was certified by ETL with the statement: Conforms to AAMI Std ES 60601-1, IEC 60601-1-6, IEC Std 60601-1-8, IEC Std 606012-27, IEC Std 60601-2-49, ISO Std 80601-2-61 Certified to CSA Std C22.2 NO. 60601-1, NO. 60601-1-6, NO. 60601-1-8, NO.60601-227, NO. 60601-2-49, NO.80601-2-61 ETL CLASSIFIED Intertek 3191955 NOTE • 1-6 Some symbols may not appear on your equipment. TMS60 Operator’s Manual 2 General Product Description Intended Use .............................................................................................................2-2 Applied Parts .............................................................................................................2-2 Key Features...............................................................................................................2-2 System Components ..............................................................................................2-3 TD60 Physical View..................................................................................................2-3 Antenna Array ...........................................................................................................2-5 Telemetry Receiver (RC60)....................................................................................2-5 Touch Screen Display..............................................................................................2-5 TMS60 Operator’s Manual 2-1 Intended Use 2.1 General Product Description Intended Use The TMS60 transmitter is intended for use on Adult and Pediatric patients to monitor ECG and SpO2 physiological data. The physiological data can be reviewed locally on the display of the transmitter. The CentralStation will support ECG, Heart Rate, SpO2, Pulse Rate, Arrhythmia analysis, QT monitoring, and ST Segment Analysis for the TMS60. It must be operated by trained medical personnel in hospitals or medical institutions. WARNING • • • • 2.2 Only skilled/trained clinical professionals should operate this equipment. The equipment is not designed for monitoring critically ill patients. If the accuracy of any value displayed at the CS or Telemetry transmitter (TD60) is questionable, determine the patient’s vital signs by alternative means and verify that the TMS is working correctly. The system transmits the data through the wireless connection. There might be a risk of data loss. Applied Parts The equipment has the following applied parts: 2.3 2-2 ■ ECG leadwires ■ SpO2 cables ■ SpO2 sensors Key Features ■ 3.5" color PTC touch screen display is easy for clinicians to use. ■ Small, portable, and lightweight for patients to wear. ■ Supports 3/5 lead ECG. ■ Supports Masimo and Nonin SpO2 modules. ■ Communication to the CS utilizes the protected WMTS 608-614 band. ■ Displays the battery status and supports the multiple levels of battery alarms. ■ Displays Heart Rate (HR) and SpO2 parameters, ECG and SpO2 waveforms. ■ Battery options of two AA, three AA, or lithium-ion battery pack are available. TMS60 Operator’s Manual General Product Description ■ 2.4 System Components Display supports Parameter auto-sizing. System Components The telemetry monitoring system (TMS60) includes the following components: 2.5 ■ Telemetry transmitter (TD60) ■ Telemetry antenna array ■ Telemetry receiver (RC60) TD60 Physical View 10 11 1. Display Activation (Power On/Off ) key When the TD60 is powered off ◆ Pressing this key will turn the TD60 on. When the TD60 is powered on ◆ If the screen display is on, pressing this key will turn the display off. ◆ If the screen display is off, pressing this key will turn the display on. ◆ Press and hold this key for two seconds to display the power off confirmation menu. TMS60 Operator’s Manual 2-3 TD60 Physical View 2. General Product Description Nurse Call key Pressing this key will send a nurse call request to the CS. The alarm light/indicator will illuminate cyan, and a “Nurse Call Initiated” message will display in the message area if the display is on. 3. 4. Main Menu key ◆ Pressing this key when on the main screen will open the main menu. ◆ Pressing this key when a menu is open will return to the main screen. ◆ Pressing this key when the display is off will turn the display on. Pressing this key when the screen lock mode is configured for View Only will display the [Screen Locked] menu. Display Touch screen display for viewing patient information and adjusting patient settings. 5. Alarm light/indicator Flashes in different color and frequency corresponding to the alarm level. 6. ECG connector ECG lead connector. 7. SpO2 cap Covers SpO2 connector when SpO2 is not in use. 8. SpO2 connector Connects the SpO2 module. 9. Speaker 10. USB connector It is only available for authorized service personnel. 11. Battery compartment Contains the lithium-ion battery pack or AA battery tray. 2-4 TMS60 Operator’s Manual General Product Description 2.6 Antenna Array Antenna Array The antenna array must be installed and configured by Mindray authorized personnel. For more details about the antenna array installation, calibration, and validation, refer to the Telemetry Monitoring System Installation Guide (P/N 046-007624-00). WARNING • 2.7 Authorized Mindray personnel are required to confirm the coverage area of the antenna array in the following situations: ◆ When the antenna array is initially installed. ◆ When the antenna array is modified. ◆ When the building construction is modified. Telemetry Receiver (RC60) The RC60 receives the data from the TD60 via the antenna array, and sends the data to the CS for analysis and display. For details about the general wireless communication problems, refer to "General Problems" on page 12 - 2. For details about the frequency allocation and receiver connection, refer to the Telemetry Monitoring System Installation Guide (P/N 046-007624-00). For details about the RC60, refer to the TMS60 Service Manual (P/N 046-007057-00). 2.8 Touch Screen Display WARNING • Do not operate the touch screen with water on the hand. Move your finger on the touch screen display to operate the TD60. For details about the supported touch gestures, refer to "Understanding the Screen Display Orientation" on page 3 - 10. TMS60 Operator’s Manual 2-5 Touch Screen Display 2.8.1 General Product Description Display Screen The main screen displays patient parameters and waveforms. A typical display screen is shown below. 1. Patient information area This area shows the patient information such as patient category, device name, and department. Tapping this area displays the [Patient Info] menu. 2. 3. Alarm symbols ◆ indicates that the alarm system is reset. ◆ indicates that the technical alarm audio is turned off. Battery symbol This symbol indicates the battery charge status. Refer to "Checking the Battery Charge Status" on page 11 - 4 for details. Tapping the battery symbol opens the [System Info] menu to the battery section. 4. Message area This area shows technical alarm messages and informational messages, where there are multiple messages, the messages scroll. 5. Patient data area This user configurable area can display parameter/waveform data. The parameter/ waveform is labeled in the upper left corner. You may also tap this area to display the Setup menu for the corresponding parameter/waveform. For details about the touch screen operations, refer to "Basic Operations" on page 3 - 9. 2-6 TMS60 Operator’s Manual General Product Description 2.8.2 Touch Screen Display On-Screen Keyboard The TD60 uses an on-screen keyboard to enter alphanumeric information, such as the device name and passwords. 2.8.2.1 Alphabetic Keyboard 2.8.2.2 1. Alphabetic buttons: tap to input the desired alphabetic text. 2. Delete button: tap to erase the text to the left of the cursor. 3. Accept button: tap to save the settings and exit the keyboard. 4. Space button: tap to input a space. 5. Numeric switch button: tap to switch to the numeric layout. 6. Case shift button: tap to switch the case of the letter. This switch is active for one character entry. Numeric Keyboard 1. Numeric buttons: tap to input the desired numbers. 2. Punctuation buttons: tap to input the desired punctuation mark or symbol. 3. Delete button: tap to erase the text to the left of the cursor. 4. Accept button: tap to save the settings and exit the keyboard. 5. Space button: tap to input a space. 6. Alphabetic switch button: tap to switch to the alphabetic layout. TMS60 Operator’s Manual 2-7 Touch Screen Display 7. 2-8 General Product Description More punctuation buttons: tap to display the punctuation keyboard, as shown below. TMS60 Operator’s Manual 3 Getting Started Unpacking and Checking......................................................................................3-2 Environmental Requirements .............................................................................3-3 Connecting the ECG Leadwire ............................................................................3-4 Installing the Batteries ...........................................................................................3-5 Powering On the Unit.............................................................................................3-8 Understanding Touch Gestures ..........................................................................3-9 Basic Operations.......................................................................................................3-9 Using the Pouch .......................................................................................................3-14 TMS60 Operator’s Manual 3-1 Unpacking and Checking Getting Started WARNING • • • • • • 3.1 The telemetry monitoring system (TMS60) shall be installed by Mindray authorized personnel. The software equipment copyright is solely owned by Mindray. No organization or individual shall resort to altering, copying, or exchanging it or to any other infringement on it in any form or by any means without due permission. Connect only approved devices to this system. Devices connected to the equipment must meet the requirements of the applicable IEC standards (e.g. IEC 60950 safety standards for information technology equipment and IEC 60601-1 safety standards for medical electrical equipment). The system configuration must meet the requirements of the IEC 60601-1 medical electrical systems standard. Any personnel who connect devices to the equipment’s signal input/output port is responsible for providing evidence that the safety certification of the devices has been performed in accordance to the IEC 60601-1. If you have any questions, please contact Mindray. If it is not evident from the equipment specifications whether a particular combination with other devices is hazardous, for example, due to summation of leakage currents, please consult the manufacturers or else an expert in the field, to ensure the necessary safety of patients and all devices concerned will not be impaired by the proposed combination. Contact Mindray to relocate the TMS60. Only Mindray authorized personnel can update the TMS60. Unpacking and Checking Before unpacking, examine the packing case carefully for signs of damage. If any damage is detected, contact the carrier or Mindray. If the packing case is intact, open the package and remove the device and accessories carefully. Check all materials against the packing list and check for any mechanical damage. Contact Mindray in case of any problem. 3-2 TMS60 Operator’s Manual Getting Started Environmental Requirements WARNING • • Before use, please verify whether the packages are intact, especially the packages of single use accessories. In case of any damage, do not apply it to patients. Do not let the display directly touch the patient when the display is on. NOTE • 3.2 Save the packing case and packaging material as they can be used if the device must be reshipped. Environmental Requirements The operating environment of the system must meet the requirements specified in this manual. The equipment operating environment should be reasonably free from noises, vibration, dust, corrosive, flammable and explosive substances. When the device is moved from one place to another, condensation may occur as a result of temperature or humidity difference. In this case, never start the system before the condensation disappears. WARNING • Make sure that the device operating environment meets the specifications. Otherwise unexpected consequences, e.g. damage to the device, could result. NOTE • The system transmits data through a wireless connection. External radio frequency interference may result in occasionally data dropout. Contact Mindray for any questions regarding the electromagnetic environment. TMS60 Operator’s Manual 3-3 Connecting the ECG Leadwire 3.3 3-4 Getting Started Connecting the ECG Leadwire 1. Align the ECG leadwire plug with the ECG connector as indicated by the arrow in the following figure. 2. Insert the ECG leadwire plug into the ECG connector as shown in the enlarged figure below. TMS60 Operator’s Manual Getting Started Installing the Batteries WARNING • • Insert the ECG lead set into the ECG connector. The following performance may be affected by a weak connection: ◆ ECG signal quality ◆ Wireless signal strength ◆ Water resistance Do not use the ECG leadwire to move or lift the TD60. This may cause the device to fall, which may damage the equipment or injure the patient. NOTE • • 3.4 ECG leadwires are used as the antenna for the TD60. To ensure good radio performance, always connect the ECG leadwires to the ECG connector while monitoring the patient. Insert the SpO2 cap in the SpO2 connector when SpO2 is not in use. Installing the Batteries You can use two AA, three AA batteries or a lithium-ion rechargeable battery pack to run the TD60. The runtime is dependent on the battery solution you chose. A lithium-ion battery pack will provide the longest runtime. For details about the recommended AA batteries, refer to "Miscellaneous" on page 15 - 5. NOTE • • Always keep the battery compartment dry. Never use brute force to install the lithium-ion battery pack or AA battery tray. Otherwise the waterproof ring surrounding the battery frame edge may be broken to affect the waterproof performance. TMS60 Operator’s Manual 3-5 Installing the Batteries 3.4.1 Getting Started Installing the Lithium-ion Rechargeable Battery WARNING • Only use specified lithium-ion rechargeable batteries. Use of other lithium-ion batteries will adversely affect the batteries: ◆ Level reporting ◆ Low battery alarms ◆ Life performance NOTE • 3-6 The lithium-ion rechargeable battery should be fully charged prior to first use. 1. Make sure the battery compartment is empty. 2. Align the hook on the upper part of the lithium-ion battery pack with the slot on the battery compartment, as indicated by the enlarged figure below. TMS60 Operator’s Manual Getting Started 3. Installing the Batteries Press down the battery pack until it is installed firmly, as indicated by the arrow in the following figure. The TD60 is automatically powered on after the battery is installed. 3.4.2 Installing the AA Batteries There are two types of AA battery trays, which are used for holding AA batteries: ■ TP-2AA battery tray can hold two AA batteries. ■ TP-3AA battery tray can hold three AA batteries. To install the AA batteries: 1. Make sure the battery compartment is empty. 2. Insert two or three 1.5V alkaline AA batteries according to the diagram in the bottom of the battery tray as shown in the images below. Installing two AA batteries TMS60 Operator’s Manual Installing three AA batteries 3-7 Powering On the Unit Getting Started 3. Align the hook on the upper part of the battery tray with the slot on the battery compartment, as indicated by the enlarged part in the following figure. 4. Press down the battery tray until it closes firmly, as indicated by the arrow in the following figure. The TD60 is automatically powered on after the batteries are installed. 3.5 Powering On the Unit Press the key to turn on the TD60. The cyan alarm light will momentarily turn on to indicate that the device is starting. The TD60 performs a self-test during startup. The device sounds a beep, and the alarm light serially turns red, yellow, cyan, and then off. This indicates that the alarm system functions correctly. Upon powering up, there are two situations: 3-8 ■ If the TD60 is turned on at first time, the device will request you to configure first time startup. Refer to the TMS60 Service Manual (P/N 046-007057-00) for details. ■ If the TD60 is turned on next time, the device will prompt whether it is a new patient. Select [Yes] or [No] as desired. If the device is a lock mode, a passcode is required. TMS60 Operator’s Manual Getting Started Understanding Touch Gestures WARNING • 3.6 Check that visual and auditory alarm signals are presented correctly when the equipment is powered on. Do not use the equipment for any monitoring procedure on a patient if you suspect the equipment is not working properly or if the equipment is mechanically damaged. Contact your service personnel or Mindray. Understanding Touch Gestures Before using the TD60, understand the supported touch screen gestures: Gesture Description Tap Briefly touch the surface with your fingertip to select a target. Press and hold Touch the surface for extended period of time. Drag Swipe 3.7 Move your fingertip over the surface without losing contact. Quickly brush the surface with your fingertip. Basic Operations This section describes the basic operations for the TD60. WARNING • Patients should be instructed not to interact with the display of the device and to not open the battery compartment while the TD60 is in use. TMS60 Operator’s Manual 3-9 Basic Operations 3.7.1 Getting Started Understanding the Screen Display Orientation The TD60 supports both the portrait and landscape display orientations. Example of portrait display 3.7.2 Example of landscape display ■ Portrait: both digital and waveform tiles take up the entire width of the screen. ■ Landscape: the digital tile takes up one half of the width of the screen; the waveform tile takes up the entire width of the screen. Browsing the Screen Display To scroll through the waveforms/parameters, swipe your finger up or down on the screen. 3.7.3 Switching the Screen Display Orientation 1. Swipe your finger down from the top of the main screen to display the drop-down menu. 2. Tap the desired option to switch the screen display orientation. For example, to switch from portrait display to landscape display: 3.7.4 3 - 10 1. Swipe your finger down from the top of the main screen to display the drop-down menu. 2. Tap [Landscape] to switch to landscape display. Flipping the Landscape Display 1. Swipe your finger down from the top of the main screen to display the drop-down menu. 2. Tap [Flip Display] to horizontally flip the landscape display. TMS60 Operator’s Manual Getting Started 3.7.5 Basic Operations Displaying the Quick Keys Area Swipe your finger up from the bottom of the main screen to display the quick keys area. The following table lists the six default quick keys: Quick keys Description Discharge Patient Tap the button to enter the [Discharge Patient] menu. Refer to "Discharging the Patient" on page 5 - 4 for details. Standby Tap the button to enter the [Standby] menu. Refer to "Placing a Device in Standby" on page 5 - 3 for details. Change Lead Tap the button to change the current first ECG lead waveform to the next ECG lead waveform that is available in sequential order. For example, if the current first ECG lead waveform is I lead, tap the button, the I lead waveform is changed to II lead waveform. Print Tap the button to notify the central station (CS) to start real-time print. The “Print Initiated” message displays on the screen. Manual Event Tap the button to notify the CS to save the event to the event database. The “Manual Event” message displays on the screen. Alarm Reset Tap the button to reset the alarm system. Refer to "Resetting the Alarms" on page 6 - 5 for details. You can customize the most frequently used functions to the quick keys. For details about setting the quick keys, refer to "Quick Keys Menu" on page 10 - 5. 3.7.6 Entering the Main Menu Press the key to enter the main menu. The main menu allows access to most of the system functions and settings. TMS60 Operator’s Manual 3 - 11 Basic Operations Getting Started All menus contain the following parts: 1. Heading: displays the current menu title. 2. Scroll bar: indicates the current scroll position within the menu. 3. Main body: contains menus, buttons, and other controls to configure and operate the device. Controls Description Accesses a submenu to reveal more options or information. Indicates that a password is required for access. Submenus Contains more operations or information related to the corresponding menu. Buttons Provides an option to operate a function. Switch Drag to right to enable the switch; drag to left to disable the switch. 4. 3.7.7 exits the current menu and return to the previous menu or the main screen. Turning the Display Off You can manually turn the display off, or let the display automatically turn off based on the configured timeout. Press the key to manually turn the display off. If the touch screen is not touched for the configured Display Auto Off time, then the screen will turn off after the configured Display Auto Off time. For details about configuring the time for Display Auto Off, refer to "Configuring the General Menu" on page 10 - 2. NOTE • 3 - 12 While the display is off, the TD60 enters the power saving mode, and does not provide audio and visual alarms. TMS60 Operator’s Manual Getting Started 3.7.8 Basic Operations Turning the Display On If the screen is off, press the or key to turn the display on. CAUTION • 3.7.9 Do not let the display directly touch the patient when the display is on. Unlocking the Screen If you set the screen lock, you need to input the correct passcode to unlock the screen after the display turns off. To unlock the screen in Locked mode: 1. If the screen is off, press the [Screen Locked] menu. or 2. Input the passcode to unlock the screen. key to turn the display on and access the Once the passcode is entered the screen is temporarily unlocked. If the is pressed or the device times out, the screen will lock again and a passcode must be entered. To unlock the screen in View Only mode: 1. If the screen is off, press the or 2. Press the 3. Input the passcode to unlock the screen. key to turn the display on. key to display the [Screen Locked] menu. Once the passcode is entered the screen is temporarily unlocked. If the is pressed or the device times out, the screen will lock again and a passcode must be entered. For details about setting the screen lock, refer to "Screen Lock Menu" on page 10 - 8. 3.7.10 Acknowledging the Nurse Call To acknowledge the triggered nurse call, tap [Attendant Present] in the main menu. The “Nurse Call Cancelled” message will display in the message area. For details about how to trigger a nurse call, refer to "TD60 Physical View" on page 2 - 3. TMS60 Operator’s Manual 3 - 13 Using the Pouch Getting Started WARNING • 3.8 Do not only rely on the nurse call function, the medical personnel should also pay close attention to the patient’s condition. Using the Pouch The TD60 is not intended for direct contact with the patient’s skin. During normal use, the TD60 could be worn over clothing, in a pocket, or in a pouch. The waterproof pouch with clear front is an appropriate means for holding the TD60. Both disposable and reusable pouches specified in this manual can be used for the TD60. For details about the pouch, refer to "Miscellaneous" on page 15 - 5. 3.8.1 Securing the Pouch To secure the pouch: 1. Place the TD60 into the pouch with the ECG leadwires and the SpO2 sensor cable, if used, exiting from the pouch opening, as shown in the following figures. For disposable pouch 3 - 14 For reusable pouch 2. Pinch the snap-fastener to close the pouch. 3. Secure the pouch on the patient with ties around the patient’s shoulder and under the arm, as shown in the following figure. TMS60 Operator’s Manual Getting Started Using the Pouch Wearing the disposable pouch Wearing the reusable pouch WARNING • While using a pouch with the TD60 on the patient, consider the patient’s condition. Be careful about the placement of the straps, as the straps could present a strangulation hazard. NOTE • The pouch is used only for the TD60. The pouch cannot be used for carrying other personal devices, such as a mobile phone. TMS60 Operator’s Manual 3 - 15 Using the Pouch Getting Started This page intentionally left blank. 3 - 16 TMS60 Operator’s Manual 4 User Configurations Introduction...............................................................................................................4-2 Configuring the Display.........................................................................................4-2 Configuring the Audio Volume...........................................................................4-5 TMS60 Operator’s Manual 4-1 Introduction 4.1 User Configurations Introduction This chapter describes the configurations available for users to do, such as configuring the Display Setup, and Audio Volume. 4.2 Configuring the Display You can configure the display by setting the display layout, display orientation, and screen brightness. In the main menu, tap [Display Setup] to enter the [Display Setup] menu. 4.2.1 Setting the Default Display Orientation For details about the display orientation, refer to "Understanding the Screen Display Orientation" on page 3 - 10. 1. In the [Setup] section of the [Display Setup] menu, tap [Default Orientation]. Two buttons display: [Portrait] and [Landscape]. 2. Tap a button to set the default orientation. The selected orientation displays to the right of [Default Orientation]. 3. 4.2.2 Restart the TD60 to apply the setting. Understanding Portrait Orientation Display Rules In portrait orientation, both digital and waveform areas take up the entire width of the screen. Therefore, these parameters will be displayed in the exact order of the [Display Setup] menu, provided the sensor is attached and monitoring data. 4.2.3 Setting the Portrait Display 1. In the [Portrait] section of the [Display Setup] menu, tap [Rows]. Three options display: [2], [3], and [4]. 2. Tap an option to set the row numbers. The selected option displays to the right of [Rows]. 3. Tap [Portrait Order] to enter the [Portrait Order] menu. 4. Tap a parameter or waveform to select it. The 4-2 icon displays to the right of the selected parameter or waveform. 5. Drag the selected parameter or waveform to the desired position, and then release it. 6. Repeat steps 4 and 5 until the desired order is configured. 7. Tap the icon to exit the [Portrait Order] menu. TMS60 Operator’s Manual User Configurations 4.2.4 Configuring the Display Understanding Landscape Orientation Display Rules In landscape orientation, waveform areas take up the entire width of the screen. Digital areas only take up one half of the width of the screen. The following rules define how the tiles will be laid out: 1. The areas shall be displayed in the order of the [Display Setup] menu except the digital area locations shall be optimized to reduce blank tiles. 2. A waveform area always takes up the entire width of the screen. 3. A digital area always takes up one half of the width of the screen. Therefore, a row with a digital tile in it shall be split into two half tiles. 4. A digital area shall not be the only parameter in a row unless an odd number of digital areas exist. In this case, the last digital parameter area shall have one tile on the left side and the right half will be blank. 5. Digital areas shall be paired with the next available digital area to satisfy rule 4. This means that a digital area may be moved ahead of a waveform area if a half of a row needs to be filled. For example, if the landscape display rows is set to [3] and the parameter order is as follows: HR ECG I ECG II ECG III ECG aVR ECG aVF ECG aVL ECG V SpO2 PLETH The landscape layout displays as follows: HR* SpO2 ECG I ECG II ECG III ECG aVR ECG aVF ECG aVL ECG V PLETH Bold is displaying on screen; nonbold data need to be scrolled to. TMS60 Operator’s Manual 4-3 Configuring the Display 4.2.5 User Configurations Setting the Landscape Display 1. In the [Landscape] section of the [Display Setup] menu, tap [Rows]. Three options display: [2], [3], and [4]. 2. Tap an option to set the row numbers. The selected option displays to the right of [Rows]. 3. Tap [Landscape Order] to enter the [Landscape Order] menu. 4. Tap a parameter or waveform option to select it. The 4.2.6 icon displays to the right side of the selected parameter or waveform. 5. Drag the selected parameter or waveform to the desired position, and then release it. 6. Repeat steps 4 and 5 until the desired order is configured. 7. Tap the icon to exit the [Landscape Order] menu. Setting the Display Brightness 1. In the [Setup] section of the [Display Setup] menu, tap [Display Brightness]. The [Display Brightness] menu displays. 4-4 2. Drag the slider to left or right to adjust the brightness. 3. Tap the icon to exit the [Display Brightness] menu. TMS60 Operator’s Manual User Configurations 4.3 Configuring the Audio Volume Configuring the Audio Volume You can independently set the technical alarm volume, touch screen click, and systole beep volume. The method for setting the three volumes are the same. To change the volume settings: 1. In the main menu, tap [Audio Volume]. 2. In the [Technical Alarm], [Touch Screen Click], or [Systole Beep] section, drag the slider to the left or right to adjust the volume. 3. Tap the icon to exit the [Audio Volume] menu. NOTE • • The icon indicates that the audio volume is turned off. The minimum value for the technical alarm volume depends on the minimum technical alarm volume, refer to "Configuring the Alarms Menu" on page 10 - 3 for details. TMS60 Operator’s Manual 4-5 Configuring the Audio Volume User Configurations This page intentionally left blank. 4-6 TMS60 Operator’s Manual 5 Patient Management Introduction...............................................................................................................5-2 Admitting a Patient .................................................................................................5-2 Changing the Patient Category ..........................................................................5-2 Placing a Device in Standby.................................................................................5-3 Resume Monitoring ................................................................................................5-4 Discharging the Patient .........................................................................................5-4 TMS60 Operator’s Manual 5-1 Introduction 5.1 Patient Management Introduction The chapter describes how to admit a patient, change the patient size, enter and exit the Standby mode, and discharge the patient. 5.2 Admitting a Patient When admitting a TM80 for the first time, the device must be admitted to the system through the CS. For details about admitting a patient through the CS, refer to the BeneVision Central Station Operator’s Manual (P/N 046-005011-00). After first admitting the device to the CS, you can directly admit the new patient at the TD60 by discharging the current patient, and then pressing the key to admit a new patient. Refer to "Discharging the Patient" on page 5 - 4 for details. 5.3 Changing the Patient Category NOTE • Ensure the patient category selection is appropriate for the patient before monitoring begins. 1. In the main menu, tap [Patient Info]. 2. In the [Patient Info] menu, tap [Patient Category] to select the desired patient category. The screen displays the “Are you sure you want to change patient category?” message. 3. Select [Yes] to confirm that the patient category is changed. The selected patient category displays to the right of [Patient Category]. 4. Tap the icon to exit the [Patient Info] menu. NOTE • • 5-2 The patient category can only be changed at the TD60. Adjusting patient category restores the TD60 to the default (preset) settings but does not clear patient information or data. TMS60 Operator’s Manual Patient Management Placing a Device in Standby NOTE • 5.4 When the device is connected to the CS, the patient category at the CS is updated if the patient category is changed at the TD60. Refer to the BeneVision Central Station Operator’s Manual (P/N 046-005011-00) for details. Placing a Device in Standby NOTE • When connected to the CS, and a device enters or exits Standby mode, the CS is also notified to enter or exit Standby mode. Refer to the BeneVision Central Station Operator’s Manual (P/N 046-005011-00) for details. To enter the Standby mode: 1. In the main menu, tap [Standby]. 2. In the [Standby] confirmation menu, select [Yes]. Placing a device into Standby mode does the following: ■ Suspends patient monitoring ■ Alarms are suspended ■ Displays [Standby] on the screen. ■ The screen display automatically turns off after the device enters the Standby mode for 30 seconds. NOTE • When connected to the CS, and a device enters or exits Standby mode, the CS is also notified to enter or exit Standby mode. TMS60 Operator’s Manual 5-3 Resume Monitoring 5.5 Patient Management Resume Monitoring Press the key to exit Standby mode. Resume monitoring: 5.6 ■ Restores patient’s settings, resumes alarm notification on the TD60 and the CS. ■ Alarm system is activated. ■ The TD60 notifies the CS of returning to the Monitoring mode. Discharging the Patient Discharging the patient will stop monitoring, clear patient information, and restore default (preset) settings on the TD60. When a new patient is admitted, the user configuration will be applied. If the user configuration has not been saved, the factory default configuration will be applied. A patient can be discharged by selecting the [Discharge Patient] menu, or restarting the TD60 and selecting that a new patient is on the TD60. NOTE • 5.6.1 Selecting the [Discharge Patient] menu 1. In the main menu, tap [Discharge Patient]. 2. In the [Discharge Patient] confirmation menu, select [Yes]. 3. 5-4 Discharging the patient on the TD60 discharges the patient from the CS. Refer to the BeneVision Central Station Operator’s Manual (P/N 046005011-00) for details. ◆ The patient is discharged from both the TD60 and the CS. ◆ The patient’s configuration is cleared and the configuration is restored to the saved user configuration or factory default configuration. ◆ The patient will be added to the [Discharged Pat.] list at the CS. Press the key to admit a new patient. TMS60 Operator’s Manual Patient Management 5.6.2 Discharging the Patient Restarting the TD60 1. If the TD60 is powered off, press the key to turn on the TD60. The device will prompt as to whether this is a new patient or not. 2. Select [Yes] if this is a new patient. Select [Yes] when asked to confirm that the discharge should begin. Refer to "Selecting the [Discharge Patient] menu" on page 5 - 4 for details on the discharge process. TMS60 Operator’s Manual 5-5 Discharging the Patient Patient Management This page intentionally left blank. 5-6 TMS60 Operator’s Manual 6 Alarms Introduction...............................................................................................................6-2 Alarm Categories......................................................................................................6-2 Alarm Levels...............................................................................................................6-2 Alarm Indicators .......................................................................................................6-3 Configuring the Alarms .........................................................................................6-5 Resetting the Alarms ..............................................................................................6-5 TMS60 Operator’s Manual 6-1 Introduction 6.1 Alarms Introduction Alarms, triggered by technical problems, are visually and audibly indicated to the user when the display is on. The TD60 provides a subset of the technical alarms, this chapter describes the technical alarms presented at the TD60 only. For details regarding full list of the technical alarms as well as the physiological alarms displayed at the central station (CS), refer to "Physiological Alarms" on page 9 - 2. WARNING • 6.2 The reception failure of alarm signals may occur in the distributed alarm system. Alarm Categories The TD60 alarm system only supports technical alarms, see the BeneVision Central Station Operator’s Manual (P/N 046-007960-00) for details on physiological alarms. Technical alarms are triggered by system status, patient status, a device malfunction or a patient data distortion due to improper operation or mechanical problems. Technical alarms are available both at the TD60 and the CS. In addition to the technical alarms, the TD60 also displays informational message to inform the user of patient/system status. The TD60 will display informational messages in the message area at the top of the display. 6.3 Alarm Levels The alarms can be classified into three severity levels: high level, medium level and low level. 6-2 Alarm Levels Technical alarms High level Indicates a severe device malfunction or an improper operation, which could make it possible that the monitor cannot detect critical patient status and thus threaten the patient’s life, such as low battery. Medium level Indicates a device malfunction or an improper operation, which may not threaten the patient’s life but may compromise the monitoring of vital physiological parameters. Low level Indicates a device malfunction or an improper operation, which may compromise a certain monitoring function but will not threaten the patient’s life. TMS60 Operator’s Manual Alarms 6.4 Alarm Indicators Alarm Indicators When a technical alarm occurs, the TD60 notifies the user through visual or audible alarm indications. ■ Alarm light ■ Audible alarm tones ■ Alarm message NOTE • 6.4.1 When the TD60 display is off, the user must activate the screen to view any local alarms. Alarm Light If a technical alarm occurs, the alarm light on the TD60 flashes. The color and flashing frequency correspond to the alarm level as follows: 6.4.2 ■ High level alarms: the lamp quickly flashes red. ■ Medium level alarms: the lamp slowly flashes yellow. ■ Low level alarms: the lamp lights cyan without flashing. Alarm Tones The TD60 has three alarm tone configurations: ISO, Mode 1 and Mode 2. For each configuration, the alarm tones enunciate the alarm levels as follows: ■ ■ ■ ISO pattern: ◆ High level alarms: triple+double+triple+double beep ◆ Medium level alarms: triple beep ◆ Low level alarms: single beep Mode 1: ◆ High level alarms: high-pitched single beep ◆ Medium level alarms: double beep ◆ Low level alarms: low-pitched single beep Mode 2: TMS60 Operator’s Manual 6-3 Alarm Indicators Alarms ◆ High level alarms: high-pitched triple beep ◆ Medium level alarms: double beep ◆ Low level alarms: low-pitched single beep NOTE • 6.4.3 When multiple technical alarms of different levels occur simultaneously, the TD60 selects the alarm of the highest level to light the alarm light and sound alarms accordingly, while all the alarm messages scroll in the message area on the top of the screen. Alarm Messages When a technical alarm occurs on the TD60 screen, the alarm message appears in the message area. The background color of the alarm message and the asterisk symbols (*) before the alarm message are designed to indicate the alarm level. 6.4.4 Alarms Background color Asterisk symbols (*) High level alarms red *** Medium level alarms yellow ** Low level alarms cyan Alarm Status Symbols The TD60 still uses the following symbols indicating the alarm status: : indicates the technical alarm audio is turned off. : indicates the alarm system is reset. 6-4 TMS60 Operator’s Manual Alarms 6.5 6.6 Configuring the Alarms Configuring the Alarms ■ For details on configuring the technical alarm volume, refer to "Configuring the Audio Volume" on page 4 - 5. ■ For details on configuring the TD60 technical alarm settings, refer to "Configuring the Alarms Menu" on page 10 - 3. ■ For the CS alarm configurations, refer to the BeneVision Central Station Operator’s Manual (P/N 046-007960-00). Resetting the Alarms You can acknowledge the on-going alarms by resetting the alarms. After being reset the alarm system can respond to a subsequent alarm condition. When a technical alarm occurs, follow this procedure to reset the TD60 alarm system. ■ Press the key to enter the main menu, and then tap [Alarm Reset] from the [Commands] section. OR 1. Swipe your finger up at the bottom of the main screen to display the quick keys area. 2. Tap the [Alarm Reset] quick key to reset the alarm system. When the alarm system is reset, depending upon the technical alarm there are several ways the alarm system may respond as follows: ■ The alarm sound will be silenced, the alarm light will continue to indicate the alarm, a √ will appear before the alarm message. The top of the main screen. symbol appears on the ■ The technical alarm will be changed to the prompt message, it will not longer make sound or be indicated by the alarm light. ■ The technical alarms are cleared, there will be no alarm indications. For details about the indications of technical alarms when the alarm system is reset, refer to "Technical Alarm Messages at the TD60" on page 12 - 5. TMS60 Operator’s Manual 6-5 Resetting the Alarms Alarms This page intentionally left blank. 6-6 TMS60 Operator’s Manual 7 Monitoring ECG Introduction...............................................................................................................7-2 Safety............................................................................................................................7-2 Preparation for Monitoring ECG .........................................................................7-3 Changing the ECG Settings ..................................................................................7-11 Understanding the ECG Display .........................................................................7-14 TMS60 Operator’s Manual 7-1 Introduction 7.1 Monitoring ECG Introduction The chapter describes the ECG monitoring function of the TD60, including skin preparation and lead placement, ECG Setup, ECG Waveform, and Pacer. The TD60 can obtain an ECG value by using a 3/5 lead ECG leadwires in conjunction with the matching skin electrodes. For details about CS configuration of the ECG parameters, QT analysis, ST analysis, and Arrhythmia analysis, refer to Chapter 9 Monitoring with the TD60 at the CS. 7.2 Safety WARNING • • • • Use manufacturer specified electrodes and lead wires. Make sure the conductive parts of electrodes and associated connectors for applied parts, including the neutral electrode, do not contact any other conductive parts including earth. Periodically inspect the electrode application site to ensure skin quality. If the skin quality changes, replace the electrodes or change the application site. Do not touch the patient or any device connected to the patient, including the bed and gurney, during defibrillation. Otherwise serious injury or death could result. CAUTION • Interference from a non-grounded instrument near the patient and electro-surgery interference can cause problems with the waveform. NOTE • 7-2 After defibrillation, the waveform recovers within 10 seconds applied in accordance with the manufacturer’s instructions for use. TMS60 Operator’s Manual Monitoring ECG Preparation for Monitoring ECG 7.3 Preparation for Monitoring ECG 7.3.1 Preparing the Patient’s Skin Proper skin preparation is essential in obtaining an accurate ECG reading. Electrode sites should be clean and dry and should provide a smooth flat surface. Incidental electrical activity and inaccurate readings may arise from incorrect skin preparation. The following procedure is recommended for secure electrode application: 1. Shave the chest hair from the electrode sites in a circular area with a diameter of 2 to 4 inches. 2. Use a dry gauze pad to remove excess skin oils, skin cells and residue from the electrode sites. Never rub the skin until it is raw or bleeding. NOTE • 7.3.2 Prepare the electrode site with alcohol only if the skin is extremely greasy. If alcohol is used as a drying agent, always allow the skin to dry before placing the electrode on the skin. Positioning the Electrodes NOTE • • • • Store electrodes at room temperature and open just prior to use. Avoid more than one type of electrode on a patient because of variations in electrical resistance. Avoid placing electrodes directly over bone prominences or over any high activity movement areas such as shoulders or arms because muscle motion produces electrical activity. If an electrode is placed over a large muscle such as the pectorals, the device may detect this additional muscle activity and could lead to false arrhythmia calls. Using a Transcutaneous Electrical Nerve Stimulator (TENS): Since a TENS unit transmits electrical impulses, avoid placing ECG electrode near the TENS electrodes. ECG electrodes may need to be repositioned and the ECG lead viewed may need to be adjusted until the optimum ECG tracing is obtained. TMS60 Operator’s Manual 7-3 Preparation for Monitoring ECG 1. Monitoring ECG Peel the backing off of the electrode. Visually inspect the contact gel medium for moistness. If the gel medium is not moist, do not use the electrode patch. Dry electrode patches are not conductive. NOTE • • To prevent evaporation of the contact gel medium, peel the backing off of the electrode patch only when it is ready for use. If using the snap type lead wires, attach the electrode to the lead wire before placing the electrode on the patient. 2. Attach the electrode patch to the skin at the prepared site. Smooth the electrode patch down in a circular motion to ensure proper skin contact. If using soft gel electrodes, never push down directly over the contact gel medium as this may displace the gel and cause monitoring artifact. If using hard gel electrodes, it is recommended that during application, the center of the electrode should be slightly pressed onto the skin to ensure direct contact. Consult the electrode manufacturer’s instructions for specific use. 3. Secure the lead wires to the patient according to hospital practice. CAUTION • Route leadwires neatly. Ensure leadwires are kept away from patient’s neck to avoid strangulation. Keep floors and walkways free of cables to reduce risk to hospital personnel, patients and visitors. NOTE • 7-4 It is recommended that the electrodes be changed at least every 24 to 36 hours to maintain proper contact with the skin, although some patients may require more frequent changing. Do not reapply disposable electrode. Try to avoid reusing the exact same electrode site during reapplication. If an electrode becomes wet with fluid, change the electrode. TMS60 Operator’s Manual Monitoring ECG Preparation for Monitoring ECG 7.3.3 Setting ECG Lead Labeling 7.3.3.1 Lead Naming Standards This manual presents lead placement according to the guidelines of the American Heart Association (AHA) and the International Electro-Technical Commission (IEC). AHA IEC Lead position 7.3.3.2 Label Color Label Color Chest Brown White Left Leg LL Red Green Right Leg RL Green Black Left Arm LA Black Yellow Right Arm RA White Red Choosing Lead Labeling For details on choosing the lead labeling, refer to "Configuring the General Menu" on page 10 - 2. 7.3.4 Placing the Electrodes For lead placement, the ECG algorithm works best when the patient’s R wave is significantly larger than the P wave or the T wave. If the R wave is not significantly larger than other lower voltage waves on the ECG tracing, the monitor may have some difficulty in identifying the appropriate waves. On some patients, electrode placement and/or the viewed ECG lead may need to be adjusted in order to obtain a significant R wave. TMS60 Operator’s Manual 7-5 Preparation for Monitoring ECG 7.3.4.1 Monitoring ECG Standard 3-Leadwire Electrode Placement A 3-wire lead set can monitor one of three ECG vectors (I, II, or III). The recommended 3wire lead placement is as follows: RA LA LL 3-wire lead placement (IEC) 3-wire lead placement (AHA) ■ Place the RA (white) electrode under the patient’s right clavicle, at the midclavicular line within the rib cage frame. ■ Place the R (red) electrode under the patient’s right clavicle, at the midclavicular line within the rib cage frame. ■ Place the LA (black) electrode under the patient’s left clavicle, at the mid-clavicular line within the rib cage frame. ■ Place the L (yellow) electrode under the patient’s left clavicle, at the mid-clavicular line within the rib cage frame. ■ Place the LL (red) electrode on the patient’s lower left abdomen within the rib cage frame. ■ Place the F (green) electrode on the patient’s lower left abdomen within the rib cage frame. 7-6 TMS60 Operator’s Manual Monitoring ECG 7.3.4.2 Preparation for Monitoring ECG Standard 5-Leadwire Electrode Placement A 5-wire lead set can monitor seven ECG vectors (I, II, III, aVR, aVL, aVF, and V) simultaneously. The recommended 5-wire lead placement is as follows: 5-wire lead placement (IEC) 5-wire lead placement (AHA) ■ Place the RA (white) electrode under the patient’s right clavicle, at the midclavicular line within the rib cage frame. ■ Place the R (red) electrode under the patient’s right clavicle, at the midclavicular line within the rib cage frame. ■ Place the LA (black) electrode under the patient’s left clavicle, at the mid-clavicular line within the rib cage frame. ■ Place the L (yellow) electrode under the patient’s left clavicle, at the mid-clavicular line within the rib cage frame. ■ Place the LL (red) electrode on the patient’s lower left abdomen within the rib cage frame. ■ Place the F (green) electrode on the patient’s lower left abdomen within the rib cage frame. ■ Place the RL (green) electrode on the patient’s lower right abdomen within the rib cage frame. ■ Place the N (black) electrode on the patient’s lower right abdomen within the rib cage frame. ■ Place the V (brown) electrode in the Vlead position as shown in the figure or decided by the clinician. ■ Place the C (white) electrode in the C-lead position as shown in the figure or decided by the clinician. TMS60 Operator’s Manual 7-7 Preparation for Monitoring ECG 7.3.4.3 Monitoring ECG Lead Placement: Pacemaker Patients The recommended lead placement for monitoring a pacemaker patient is as follows. Pacer RA Pacer LA LL 3-wire lead placement for a pacemaker patient (AHA) Pacer RA LL 5-wire Lead Placement for a Pacemaker Patient (AHA) Pacer LA RL 3-wire Lead Placement for a Pacemaker Patient (IEC) 5-wire Lead Placement for a Pacemaker Patient (IEC) A pacemaker patient usually requires a different electrode patch placement configuration than a non-pacemaker patient. Do not place an ECG electrode directly over the pacemaker generator. Place the electrode patches 3 to 5 inches away from the pacemaker generator area. For example, if the pacemaker generator is located in the right subclavian area, relocate the Right Arm electrode closer in towards the center of the chest. 7-8 TMS60 Operator’s Manual Monitoring ECG 7.3.5 Preparation for Monitoring ECG Checking the Lead Placement With the Lead Placement function, you can check the lead status, information, and lead off messages. 7.3.5.1 Entering the Lead Placement Menu Enter the [Lead Placement] menu in either of the following ways: 7.3.5.2 ■ Tap the lead fault message in the message area of the main screen. ■ In the main menu, tap [Lead Placement]. Understanding the Lead Placement Instructions The [Lead Placement] window indicates the lead status. Information bar Lead off indicator Lead on indicator Example lead placement window When any of the leads are off, the indications are as follows: ■ The lead off message displays on the information bar. The background color of the information bar corresponds to the alarm level. ■ A flashing circle indicates the disconnected lead. The color of the flashing circle is based on the alarm level. TMS60 Operator’s Manual 7-9 Preparation for Monitoring ECG 7.3.6 Monitoring ECG Checking the Paced Status It is important to correctly set the patient’s paced status before you start monitoring ECG. To check the paced status: ■ On the main screen, tap the HR digital area or ECG waveform area to enter the [ECG] menu. OR 1. In the main menu, tap [Patient Info]. 2. In the [Pacer] field, check the setting of the paced status. The current paced status setting displays to the right of [Paced]. 3. If the paced status setting is not correct, tap [Paced] and select the correct paced status. ■ When [Paced] is set to [Yes] at the TD60, and the pacer pulse is detected, the symbol displays in the waveform area of the CS screen, and the pace pulse marks will display on the ECG waveform both at the TD60 and CS. ■ When [Paced] is set to [No] at the TD60, and the pacer pulse is detected, the symbol displays in the waveform area of the CS screen. WARNING • • • • • 7 - 10 For paced patients, you must set [Paced] to [Yes]. If it is incorrectly set to [No], the CS could mistake a pace pulse for a QRS and fail to alarm when the ECG signal is too weak. Do not rely entirely on rate meter alarms when monitoring patients with pacemakers. Always keep these patients under close surveillance. The pacer pulses may be counted as QRS complexes, hence leading to wrong HR readings or failure to diagnose certain arrhythmia symptoms. Be sure to keep a close eye on patient’s with pacemaker devices. For non-paced patients, you must set [Paced] to [No]. False low heart rate indicators or false asystole calls may result with certain pacemakers because of pacemaker artifact such as electrical overshoot of the pacemaker overlapping the true QRS complexes. In order to minimize the possibility of interference, place electrodes, leadwires and TD60s as far away from the pacemaker as possible. TMS60 Operator’s Manual Monitoring ECG Changing the ECG Settings NOTE • 7.4 When [Paced] is set to [Yes], the system does not detect PVC-related arrhythmia (including PVCs) resulting from pacemaker but still analyzes the normal QRS complex. Changing the ECG Settings You can change the ECG settings from the ECG menu. 7.4.1 Configuring the ECG Setup Enter the ECG menu in either of the following ways: ■ On the main screen, tap the HR digital area or ECG waveform area to enter the [ECG] menu. ■ In the main menu, tap [Parameter Setup] → [ECG] to enter the [ECG] menu. 1. In the [Setup] section of the [ECG] menu, select the options described in the following table. Options Description Settings* Lead Placement Enters the [Lead Placement] window. Refer to "Checking the Lead Placement" on page 7 - 9 for details. Cable Type Selects the current ECG leadwire type. Auto, 3 Lead, 5 Lead Refer to "ECG Leadwire Types" on page 7 - 12 for details. Smart Lead (Monitored Lead) When [Cable Type] is set to [Auto], the option displays [Smart Lead]. Drag the swtich to right or left to enable or disable the Smart Lead function. When [Cable Type] is set to [3 Lead], the option displays [Monitored Lead]. Refer to "ECG Leadwire Types" on page 7 - 12 for details. The factory default settings are in bold. TMS60 Operator’s Manual 7 - 11 Changing the ECG Settings Monitoring ECG Options Description Settings* Filter Selects the ECG filter. Monitor, ST Monitor Use under normal measurement conditions. ST Use when ST monitoring is applied. Color Selects the ECG waveform color. 16 colors The default color is green. The factory default settings are in bold. 2. 7.4.2 Tap to exit the [ECG] menu. ECG Leadwire Types ECG leadwire type has three options as follows: ■ [Auto]: the device automatically sets the leadwire type according to the leadset connected. ■ [3 Lead]: the leadwire type is set to 3-lead. If the leadwire type is set to 3-lead, the [Smart Lead] option becomes [Monitored Lead]. You can select the desired lead from the [Monitored Lead] option to set the first ECG waveform displayed on the main screen. ■ [5 Lead]: the leadwire types is set to 5-lead. All waveform leads display on the main screen. 7.4.3 Configuring the ECG Waveforms 1. In the [Waveform] section of the [ECG] menu, select the options described in the following table. Options Description Settings* All Lead Size Selects the waveform size for all the leads. To set the waveform size for a specific lead, select that lead from the [Waveform Size] section. 1.25 mm/mV, 2.5 mm/mV, 5 mm/ mV, 10 mm/mV, 20 mm/mV, 40 mm/mV, Auto This configuration will be applied for all ECG waveform size. 7 - 12 The factory default settings are in bold. TMS60 Operator’s Manual Monitoring ECG Changing the ECG Settings Options Description Settings* Speed Selects the waveform sweep speed. 6.25 mm/s, 12.5 mm/s, 25 mm/s The factory default settings are in bold. 2. 7.4.4 Tap to exit the [ECG] menu. Configuring the Pacer 1. In the [Pacer] section of the [ECG] menu, tap the options described in the following table. Options Description Settings* Paced Selects the paced status. Unspecified, No, Yes [Unspecified] is only available for the first time you set the paced status. Refer to "Checking the Paced Status" on page 7 - 10 for details. Markers Selects the pacer indicator. Line, Dot, Off Line A 1 cm line shows above each ECG waveform each time the pace pulse is detected. Dot A 2 mm dot shows above each ECG waveform each time the pace pulse is detected. Pacer Reject Selects whether or not to reject the pace pulses. On, Off The factory default settings are in bold. 2. Tap to exit the [ECG] menu. NOTE • When [Paced] is set to [Yes], the [Markers] and [Pacer Rejection] options can be available. TMS60 Operator’s Manual 7 - 13 Understanding the ECG Display 7.4.5 Monitoring ECG Configuring the ECG Waveform Size The [Waveform Size] section of the [ECG] menu lists all available leads. You can select the desired ECG lead to set the waveform size. 7.5 Understanding the ECG Display 7.5.1 HR Digital Area The HR digital area displays: 1. Parameter name 2. Measurement unit 3. Heart rate value 7.5.2 7 - 14 About the HR Digital Area ■ The HR area displays heart rate in the unit of bpm with a resolution of 1 bpm. ■ If the HR measurement is invalid, “- - -” displays in place of the HR value. ■ The HR value displays “0”, when the HR value is less than 15 bpm. TMS60 Operator’s Manual Monitoring ECG 7.5.3 Understanding the ECG Display ECG Waveform Area The ECG waveform area displays: 1. ECG Lead 2. ECG scale bar 3. ECG waveform 4. ECG filter setting 5. ECG scale 7.5.4 About the ECG Waveform Area ■ The ECG waveform, scale indicator, lead, and filter settings display in the configured ECG color. ■ The ECG waveform area provides scrolling, real-time waveform data and an erase bar to provide a time indicator of oldest and new data. ■ The ECG waveform area scrolls the waveform in the configured sweep speed. ■ The ECG waveform area rails the top most value when the waveform exceeds the upper scale limit for real-time waveforms. ■ The ECG waveform area rails the bottom most value when the waveform exceeds the lower scale limit for real-time waveforms. ■ The ECG waveform area indicates a Pacer indicator when a pace pulse is detected and Paced is enabled. TMS60 Operator’s Manual 7 - 15 Understanding the ECG Display Monitoring ECG This page intentionally left blank. 7 - 16 TMS60 Operator’s Manual 8 Monitoring SpO2 (Optional) Introduction...............................................................................................................8-2 Measurement Limitations.....................................................................................8-2 Safety............................................................................................................................8-3 Connecting the SpO2 Module ............................................................................8-4 Changing the SpO2 Settings ...............................................................................8-5 SpO2 Measurement ................................................................................................8-8 Understanding the SpO2 Display ......................................................................8-10 TMS60 Operator’s Manual 8-1 Introduction 8.1 Monitoring SpO2 (Optional) Introduction The chapter describes the SpO2 monitoring function at the TD60 in detail, including connecting the SpO2 module, configuring the SpO2, and monitoring the SpO2. For details about the SpO2 parameter setup at the central station (CS), refer to Chapter 9 Monitoring with the TD60 at the CS. SpO2 monitoring is a non-invasive technique used to measure the amount of oxygenated hemoglobin and pulse rate by measuring the absorption of selected wavelengths of light. The light generated in the probe passes through the tissue and is converted into electrical signals by the photo detector in the probe. The SpO2 module processes the electrical signal and displays a waveform and digital values for SpO2 and pulse rate. The TD60 can be configured with Masimo SpO2 or Nonin SpO2. NOTE • • • • • 8.2 A functional tester or SpO2 simulator cannot be used to assess the accuracy of a SpO2 module or a SpO2 sensor. This device is calibrated to display functional oxygen saturation. A pulse oximeter should not be used as an apnea monitor. Pulse rate measurement is based on the optical detection of a peripheral flow pulse and therefore may not detect certain arrhythmias. The pulse oximeter should not be used as a replacement or substitute for ECG based arrhythmia analysis. The MS board pulse oximeter can be used during defibrillation, but the readings may be inaccurate for a short time. Measurement Limitations If the SpO2 measurement seems out of range or inaccurate, check the patient’s vital signs. Then check the equipment and SpO2 sensor. The following factors may influence the accuracy of measurement: 8-2 ■ Ambient light ■ Physical movement ■ Low perfusion ■ Electromagnetic interference TMS60 Operator’s Manual Monitoring SpO2 (Optional) 8.3 Safety ■ Dysfunctional hemoglobin, such as carboxyhemoglobin (COHb) and methemoglobin (MetHb) ■ Presence of certain dyes, such as methylene and indigo carmine ■ Inappropriate positioning of the SpO2 sensor, or use of incorrect SpO2 sensor ■ Drop of arterial blood flow to immeaurable level caused by shock, anemia, low temperature or vasoconstrictor. ■ Inaccurate measurements may be caused by venous pulsations. ■ Placement of a sensor on an extremity that has a blood pressure cuff, arterial catheter, or intra-vascular line. ■ Loss of pulse signal can occur when the sensor is too tight. ■ Loss of pulse signal can occur when there is arterial occlusion proximal to the sensor. Safety WARNING • • • • • Only use SpO2 sensors specified in this manual. Follow the SpO2 sensor’s instructions for use and adhere to all warnings and cautions. The operator is responsible for checking the compatibility of the pulse oximetry monitor, sensor, and patient cable prior to use. Incompatible components can result in degraded performance and/or device malfunction. When a trend toward patient deoxygenation is indicated, blood samples should be analyzed by a laboratory co-oximeter to completely understand the patient’s condition. Do not use SpO2 sensors during magnetic resonance imaging (MRI). Induced current could potentially cause burns. The sensor may affect the MRI image, and the MRI unit may affect the accuracy of the oximetry measurements. Prolonged and continuous monitoring may increase the temperature of the sensor and cause the patient discomfort. It is especially important to check the sensor placement, and ensure proper attachment on patients suffering from poor perfusion or skin sensitivity. Check the sensor location every two to three hours and move to another location if the skin deteriorates. More frequent examinations may be required for different patients. TMS60 Operator’s Manual 8-3 Connecting the SpO2 Module 8.4 Monitoring SpO2 (Optional) Connecting the SpO2 Module Connect the SpO2 module to the TD60. The TD60 can auto detect the SpO2 module type when the SpO2 module is connected. SpO2 module SpO2 sensor con- SpO2 sensor Connecting the Masimo SpO2 module 8-4 TMS60 Operator’s Manual Monitoring SpO2 (Optional) Changing the SpO2 Settings SpO2 module SpO2 sensor connector SpO2 sensor Connecting the Nonin SpO2 module 8.5 Changing the SpO2 Settings You can change the SpO2 settings from the SpO2 menu. 8.5.1 Configuring the SpO2 Setup Enter the SPO2 menu in either of the following ways: ■ On the main screen, tap the SpO2 digital area or SpO2 waveform area to enter the [SpO2] menu. ■ In the main menu, tap [Parameter Setup] → [SpO2] to enter the [SpO2] menu. 1. In the [Setup] section of the [SpO2] menu, select the options described in the following table. TMS60 Operator’s Manual 8-5 Changing the SpO2 Settings Monitoring SpO2 (Optional) Options Description Settings* Display PI (Masimo only) Configures whether or not to show the perfusion index (PI) value in the digital area. The perfusion index allows clinicians to assess the pulse strength at the monitoring site for optimal sensor placement. PI gives the numerical value for the pulsatile to non-pulsatile portion of the measured signal caused by arterial pulsation. PI is an indicator of the pulsatile strength. On, Off Sensitivity (Masimo only) Selects the sensitivity mode depending upon signal quality and patient motion. High, Normal, and APOD High: This mode should be used for the sickest patients, where obtaining a reading is most difficult. High Sensitivity is designed to interpret and display data for even the weakest of signals. This mode is recommended during procedures and when clinician and patient contact is continuous. Normal: This mode provides the best combination of sensitivity and probe off detection performance. This mode is recommended for the majority of patients. APOD (Adaptive Probe Off Detection): This mode is the least sensitive in picking up a reading on patients with low perfusion but has the best detection for probe-off conditions. This mode is useful for patients that are at particular risk of the sensor becoming detached (pediatric, combative, etc.) 8-6 The factory default settings are in bold. TMS60 Operator’s Manual Monitoring SpO2 (Optional) Changing the SpO2 Settings Options Description Settings* Averaging (Masimo only) The user-selectable averaging feature allows the clinician to select the desired level of visibility to subtle variations in the measured value. Depending on the patient acuity and area of care, shorter averaging times are sometimes preferred (sleep testing) over longer averaging times (telemetry) and vice-versa. 8-second averaging is generally considered the most common averaging interval and recommended for most patients since it is short enough to provide visibility to subtle desaturations while also being long enough to minimize major changes in SpO2 due to quick, transitory desaturations. Although averaging times greater than 10 seconds are more likely to reduce visibility to rapid, brief desaturations, this may be desirable in care areas where brief desaturations that do not require clinician intervention occur more often (such as NICU). It is also recommended that this be enabled as a “sticky” configuration so as to hold the setting after power cycles. 2-4 sec, 4-6 sec, 8 sec, 10 sec, 12 sec, 14 sec, 16 sec Fast SAT (Masimo only) Selects whether or not to enable FastSat. FastSat enables rapid tracking of arterial oxygen saturation changes as may be required in urgent situations. On, Off Color Selects the SpO2 waveform color. 16 colors The default color is cyan. The factory default settings are in bold. 2. Tap TMS60 Operator’s Manual to exit the [SpO2] menu. 8-7 SpO2 Measurement 8.5.2 Monitoring SpO2 (Optional) Configuring the SpO2 Waveform 1. In the [Waveform] field of the [SpO2] menu, select the options described in the following table. Options Description Settings* Speed Selects the SpO2 pleth waveform speed. 6.25 mm/s, 12.5 mm/s, 25 mm/s Display SIQ (Masimo only) Selects whether or nor to show the Signal Indicator Quality (SIQ) in the SpO2 waveform area. The SIQ wave indicates the confidence associated with the saturation measurement and timing of the pulse. Higher pulse indicates a better signal. On, Off The factory default settings are in bold. 2. Tap to exit the [SpO2] menu. 8.6 SpO2 Measurement 8.6.1 Identifying SpO2 Modules To identify which SpO2 module you are using, see the company logo on the SpO2 module. 8.6.2 ■ Masimo SpO2 module: white, with a logo of Masimo SET. ■ Nonin SpO2 module: blue, with a logo of Nonin. Applying the Sensor 1. Select an appropriate sensor according to the module type, patient size, and weight. 2. Remove colored nail polish from the application site. 3. Apply the sensor to the patient. 4. Connect the sensor to the SpO2 module and the SpO2 module to the TD60. The SpO2 measurement displays when the TD60 detects that a sensor is connected to the patient. 8-8 TMS60 Operator’s Manual Monitoring SpO2 (Optional) SpO2 Measurement WARNING • • • • When equipped with Masimo SpO2 module, use only Masimo SpO2 sensors specified in this manual. Use of other SpO2 sensors may cause improper oximeter performance. When equipped with Nonin SpO2 module, use only Nonin SpO2 sensors specified in this manual. Use of other SpO2 sensors may cause improper oximeter performance. Do not disconnect the Nonin Spo2 sensor connector from the Nonin SpO2 module during defibrillation. If the sensor is too tight because the application site is too large or becomes too large due to edema, excessive pressure for prolonged periods may result in venous congestion distal from the application site, leading to interstitial edema and tissue ischemia. CAUTION • Many patients suffer from poor peripheral perfusion due to hypothermia, hypovolemia, severe vasoconstriction, reduced cardiac output, etc. These symptoms may cause a loss in vital sign readings. TMS60 Operator’s Manual 8-9 Understanding the SpO2 Display Monitoring SpO2 (Optional) 8.7 Understanding the SpO2 Display 8.7.1 SpO2 Digital Area The SpO2 digital area displays: 1. Parameter name 2. SpO2 value 3. Perfusion indicator 4. Perfusion index (PI) label 5. Perfusion index value 6. SpO2 unit of measure 7. Pulse rate (PR) label 8. PR measurement unit 9. PR value Masimo SpO2 digital area (for portrait display) Masimo SpO2 digital area (for landscape display) Nonin SpO2 digital area 8 - 10 TMS60 Operator’s Manual Monitoring SpO2 (Optional) 8.7.2 8.7.3 Understanding the SpO2 Display About the SpO2 Digital Area ■ The SpO2 digital area displays in units of % with a resolution of 1%. ■ The SpO2 PR value displays in units of bpm with a resolution of 1 bpm. ■ Displays Masimo PI resolution as 0.01 when the PI value is smaller than 10%. ■ Displays Masimo PI resolution as 0.1 when the PI value is greater than or equal to 10%. ■ If the SpO2 measurement or PR is invalid, “- - -” displays in place of digits. SpO2 Waveform Area The SpO2 waveform area displays: 1. Signal Indicator Quality (SIQ) 2. Area name 3. Pleth waveform Masimo SpO2 waveform area (SIQ enabled) Nonin SpO2 waveform area 8.7.4 About the SpO2 Waveform Area ■ Displays in the configured SpO2 color. ■ Provides scrolling, real-time waveform data. ■ Scrolls the waveform in the configured sweep speed. ■ Automatically scales the SpO2 waveform data area to maximize the vertical height of the Pleth waveform for the data range. ■ If using Masimo SpO2, the Signal Quality Index (SIQ) will display below the waveform if enabled. TMS60 Operator’s Manual 8 - 11 Masimo Information 8.8 Monitoring SpO2 (Optional) Masimo Information ■ Masimo Patents This device is covered under one or more the following U.S.A. patents: 5,758,644; 5,823,950; 6,011,986; 6157,850; 6,263,222; 6,501,975; 7,469,157 and other applicable patents listed at: www.masimo.com/patents.htm. ■ No Implied License Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized sensors or cables which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device. 8 - 12 TMS60 Operator’s Manual 9 Monitoring with the TD60 at the CS Introduction...............................................................................................................9-2 Physiological Alarms...............................................................................................9-2 ECG Monitoring ........................................................................................................9-12 QT Monitoring...........................................................................................................9-18 ST Monitoring............................................................................................................9-22 Arrhythmia Monitoring..........................................................................................9-27 SpO2 Monitoring .....................................................................................................9-33 TMS60 Operator’s Manual 9-1 Introduction 9.1 Monitoring with the TD60 at the CS Introduction The chapter describes the configurations and displays at the central station (CS) once the TD60 is connected to the CS. At the CS, Mindray ECG algorithm and Mortara ECG algorithm are available. You can select either algorithm as required. 9.2 Physiological Alarms At the CS, you can view and change the physiological alarm limits and alarm levels in the [Alarm Setup] menu. The [Alarm Setup] menu has three tabs: ■ [Parameter Alarm Settings]: view and change the parameter alarm limits, alarm levels and alarm responses. ■ [Arrhythmia Alarms]: view and change the arrhythmia alarms levels, and alarm responses. ■ [Arrh. Threshold Setup]: view and change the arrhythmia threshold settings for some arrhythmia alarms. For details about the [Alarm Setup] menu, refer to the BeneVision Central Station Operator’s Manual (P/N 046-005011-00). WARNING • • • • • 9-2 Be aware that the devices in your care area may each have different alarm settings to suit different patients. Always check that the alarm settings are appropriate for your patient before you start monitoring. A potential hazard can exist if different alarm presets are used for the same or similar equipment in any single area, such as an intensive care unit or cardiac operating room. Make sure that the alarm limits settings are appropriate for your patient before monitoring. When monitoring patients that are not continuously attended by a clinical operator, properly configure the alarm system and adjust alarm settings as per the patient's condition. Setting alarm limits to extreme values may cause the alarm system to become ineffective. For example, high oxygen levels may predispose a premature infant to retrolental fibroplasia. If this is a consideration do NOT set the high alarm limit to 100%, which is equivalent to switching the alarm off. TMS60 Operator’s Manual Monitoring with the TD60 at the CS Physiological Alarms WARNING • 9.2.1 If you switch off all arrhythmia alarms, the CS cannot give any arrhythmia alarms. Always keep the patient under close surveillance. Factory Default Parameter Alarm Limits The following table lists the factory default alarm limits for all parameters. Parameters Alarm limit Adult Pediatric HR Unit of measure: bpm Invalid data: “--” High limit (Low limit + 2) to 300 Default: 120 (Low limit + 2) to 350 Default: 160 Low limit 15 to (high limit - 2) Default: 50 15 to (high limit - 2) Default: 75 High limit (Low limit + 0.2) to 2.00 Default: 0.2 (Low limit + 0.2) to 2.00 Default: 0.2 Low limit -2.00 to (high limit – 0.2) Default: -0.2 -2.00 to (high limit – 0.2) Default: -0.2 High limit (Low limit + 0.2) to 2.00 Default: 0.2 (Low limit + 0.2) to 2.00 Default: 0.2 Low limit -2.00 to (high limit – 0.2) Default: -0.2 -2.00 to (high limit – 0.2) Default: -0.2 High limit 200 to 800 Default: 500 200 to 800 Default: 480 Low limit N/A N/A High limit 30 to 200 Default: 60 30 to 200 Default: 60 Low limit N/A N/A High limit (Low limit + 1) to 100 Default: 100 (Low limit + 1) to 100 Default: 100 Low limit 0 to (high limit - 1) Default: 90 0 to (high limit - 1) Default: 90 ST Single Unit of measure: mV Invalid data: “--” ST Dual Unit of measure: mV Invalid data: “--” QTc Unit of measure: ms Invalid data: “--” ∆QTc* Unit of measure: ms Invalid data: “--” SpO2 Unit of measure: % Invalid data: “--” ΔQTc is only available for Mindray ECG algorithm. TMS60 Operator’s Manual 9-3 Physiological Alarms Monitoring with the TD60 at the CS Parameters Alarm limit Adult Pediatric SpO2 Desat High limit N/A N/A Unit of measure: % Invalid data: “--” Low limit 0 to 100 Default: 80 0 to 100 Default: 80 High limit (Low limit + 2) to 300 Default: 120 (Low limit + 2) to 300 Default: 160 Low limit 18 to (high limit - 2) Default: 50 18 to (high limit - 2) Default: 75 PR Unit of measure: bpm Invalid data: “--” ΔQTc is only available for Mindray ECG algorithm. NOTE • 9.2.2 The SpO2 Desat alarm limit is restricted such that the alarm limit can not be higher than the SpO2 low limit. Parameter Alarm Responses At the CS, the following parameter alarm responses may occur when a parameter alarm occurs. 9-4 ■ Record on Alarm: directs the CS to send the alarm data to the configured recorder. ■ Print on Alarm: directs the CS to send the alarm data to the configured printer. ■ Paging Switch: directs the CS to send an alarm to the beeper to notify the clinician. TMS60 Operator’s Manual Monitoring with the TD60 at the CS 9.2.3 Physiological Alarms Factory Default Parameter Alarm Settings The following table lists the factory default alarm levels and responses for all parameters. Alarm levels* Factory default alarm responses Activation State Parameters Record on Alarm Print on Alarm Paging Switch On Off Off Off Off Off Off Off Off Off Off Off Off Off Off Off ∆QTc*** Off Off Off Off SpO2 — On Off Off Off SpO2 Desat**** — — On Off Off Off PR — On Off Off Off High Medium Low HR X** — ST Single ST Dual QTc X indicates available alarm level, — indicates alarm level not available ** The factory default settings are in bold. *** ΔQTc is only available for Mindray ECG algorithm. **** The alarm level option for SpO2 Desat is not configurable. [High] is the only alarm level and cannot be changed. TMS60 Operator’s Manual 9-5 Physiological Alarms Monitoring with the TD60 at the CS 9.2.4 Factory Default Arrhythmia Alarm Settings 9.2.4.1 Mindray Algorithm Factory default alarm responses Alarm levels* Activation State Parameters Record on Alarm Print on Alarm Paging Switch On Off Off Off — On Off Off Off — — On Off Off Off — — — On Off Off Off — — — On Off Off Off Extreme Brady — — — On Off Off Off PVCs/min On Off Off Off R on T On Off Off Off Run PVCs Off Off Off Off Couplet Off Off Off Off PVC Off Off Off Off Vent. Rhythm On Off Off Off Bigeminy On Off Off Off Trigeminy On Off Off Off Tachy Off Off Off Off Brady Off Off Off Off Pacer Not Pacing Off Off Off Off High Medium Low Message Asystole X** — — — VFib/VTac — — VTac — Vent. Brady Extreme Tachy X indicates available alarm level, — indicates alarm level not available ** The factory default settings are in bold. 9-6 TMS60 Operator’s Manual Monitoring with the TD60 at the CS Physiological Alarms Factory default alarm responses Alarm levels* Activation State Parameters Record on Alarm Print on Alarm Paging Switch Off Off Off Off Off Off Off Off Off Off Off Off On Off Off Off Off Off Off Off AFib Off Off Off Off Irr.Rhythm — — Off Off Off Off Pauses/min On Off Off Off High Medium Low Message Pacer Not Capture Missed Beat Multif. PVC Nonsus. Vtac Pause X indicates available alarm level, — indicates alarm level not available ** The factory default settings are in bold. NOTE • When [Paced] is set to [Yes], the Missed Beat (MIS) alarm is reported as the pacer not captured (PNC) or pacer not paced (PNP) alarm. TMS60 Operator’s Manual 9-7 Physiological Alarms 9.2.4.2 Monitoring with the TD60 at the CS For Mortara Algorithm Factory default alarm responses Alarm levels* Activation State Parameters Record on Alarm Print on Alarm Paging Switch On Off Off Off — On Off Off Off — — On Off Off Off — — — On Off Off Off — — — On Off Off Off PVCs/min On Off Off Off R on T On Off Off Off Run PVCs Off Off Off Off Couplet Off Off Off Off Vent. Rhythm On Off Off Off Bigeminy On Off Off Off Trigeminy On Off Off Off Tachy Off Off Off Off Brady Off Off Off Off Pacer Not Pacing Off Off Off Off Pacer Not Capture Off Off Off Off Multif. PVC Off Off Off Off Pause Off Off Off Off High Medium Low Message Asystole X** — — — VFib — — VTac — Extreme Tachy Extreme Brady X indicates available alarm level, — indicates alarm level not available ** The factory default settings are in bold. 9-8 TMS60 Operator’s Manual Monitoring with the TD60 at the CS Physiological Alarms Factory default alarm responses Alarm levels* Activation State Parameters High Medium Low Message Irr.Rhythm — — Pauses/min Record on Alarm Print on Alarm Paging Switch Off Off Off Off On Off Off Off X indicates available alarm level, — indicates alarm level not available ** The factory default settings are in bold. NOTE • The priority of lethal arrhythmia alarms is always high. It is unchangeable. In addition, the activation state of some arrhythmias can be set as a whole with the following buttons that are at the bottom of the [Arrhythmia Alarms] tab. Button Description Lethals Only Sets the lethal arrhythmia alarms to on and all non-lethal arrhythmia alarms to off. All Alarm On Sets all arrhythmia alarms to on. All Alarm Off Sets all arrhythmia alarms to off. This button is enabled when the [Lethal Arrh. OFF] option in the [Telemetry] tab from the [Admin Setup] menu is set to [Enable]. TMS60 Operator’s Manual 9-9 Physiological Alarms 9.2.5 Monitoring with the TD60 at the CS Arrhythmia Threshold Settings When an arrhythmia violates its threshold, an alarm is triggered. For the Mortara algorithm, the setting of asystole delay is related to arrhythmia relearn. When HR is less than 30 bpm, it is recommended to set asystole delay to 10 s. 9.2.5.1 Mindray ECG algorithm Arrh. event Range or Option Default Step Unit of measure PVCs High 1 to 100 10 minute Pauses/min 1 to 15 N/A Asys. Delay 3 to 10 second Tachy High 60 to 300 Adult: 120 Pediatric: 160 bpm Brady Low 15 to 120 Adult: 50 Pediatric: 75 bpm Extreme Tachy 60 to 300 Adult: 160 Pediatric: 180 bpm Extreme Brady 15 to 120 Adult: 35 Pediatric: 50 bpm Multif. PVC’s Window 3 to 31 15 /min Vtac Rate 100 to 200 130 bpm Vtac PVC 3 to 99 beat Pause Time 1.5, 2.0, and 2.5 2.0 N/A second Vbrd Rate 15 to 60 40 bpm Vbrd PVCs 3 to 99 beat 9 - 10 TMS60 Operator’s Manual Monitoring with the TD60 at the CS 9.2.5.2 Physiological Alarms Mortara ECG algorithm Arrh. event Range or Option Default Step Unit PVCs High 1 to 100 10 minute Pauses/min 1 to 15 N/A Asys. Delay 2 to 10 seconds Vtac Rate 100 to 200 130 bpm Vtac PVC 3 to 12 beat Extreme Tachy Adult: 100 to 300 Pediatric: 160 to 300 Adult: 160 Pediatric: 180 bpm Extreme Brady Adult: 15 to 60 Pediatric: 15 to 80 Adult: 35 Pediatric: 50 bpm Multif. PVC’s Window 3 to 31 15 /min Tachy High Adult: 100 to 300 Pediatric: 160 to 300 Adult: 120 Pediatric: 160 bpm Brady Low Adult: 15 to 60 Pediatric: 15 to 80 Adult: 50 Pediatric: 75 bpm Pause Threshold RR 1.2, 1.3, 1.4, 1.5, 1.6, 1.7, 1.8, 1.9, and 2.0 2.0 N/A TMS60 Operator’s Manual 9 - 11 ECG Monitoring 9.3 Monitoring with the TD60 at the CS ECG Monitoring At the CS, you can view and change the heart rate (HR), QT, ST, and arrhythmia settings in the [ECG] tab of the [Parameter Setup] menu. For details about the [Parameter Setup] menu, refer to the BeneVision Central Station Operator’s Manual (P/N 046-005011-00). 9.3.1 HR Settings The following table lists the HR settings in the [HR] section of the [ECG] tab. Options or Buttons Description Settings* HR Activation State Configures whether or not to enable the HR alarm. On, Off HR Alarm Priority Configures the HR alarm levels. High, Med HR High Limit (bpm) Configures the HR high alarm limit. Adult: (Low limit + 2) to 300 The default is 120. Pediatric: (Low limit + 2) to 350 The default is 160. HR Low Limit (bpm) Configures the HR low alarm limit. Adult: 15 to (high limit - 2) The default is 50. Pediatric: 15 to (high limit - 2) The default is 75. Pacer Rate (bpm) (only Mortara ECG algorithm) When some pacemaker pulses are difficult to reject, the pulses are counted as a QRS complex and could result in an incorrect HR and failure to detect some arrhythmias. Configures the pacemaker rate, the system can calculate HR and detect arrhythmias more accurately. 40 to 100 The default is 60. The option is unavailable when [Paced] is set to [No] at the TD60. For details about the paced status, refer to "Checking the Paced Status" on page 7 - 10. Sweep Speed Configures the ECG wave speed. 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s HR Source Configures the HR source. ECG, SpO2, Auto, Both 9 - 12 The factory default settings are in bold. TMS60 Operator’s Manual Monitoring with the TD60 at the CS ECG Monitoring Options or Buttons Description Settings* HR Record on Alarm Select whether or not to activate the option to direct the CS to send the HR alarm data to the configured recorder. On, Off Waveform Setup Select to display the [Waveform Setup] menu. Refer to "Waveform Setup" on page 9 - 13 for details. Other Settings Select to display the [Other Settings] menu. Refer to "Other Settings" on page 9 - 15 for details. 9.3.2 The factory default settings are in bold. Waveform Setup The ECG leads waveforms displayed in the waveform area are defined as the displaying lead. The CS uses information from two leads to detect beats and to compute HR. These two leads are referred to as the primary and secondary leads. In addition, for Mortara ECG algorithm, information from an additional lead (analysis lead) is made use of to classify the beats (normal, abnormal etc). The user can select any of the available leads (depending on whether a 3 or 5 lead cable is used) as primary, secondary or other analysis leads. For best results, the following guidelines should be used: 9.3.2.1 ■ The QRS complex should be tall, narrow and preferably either completely above or below the baseline. If at all possible, avoid selecting a lead where the QRS complex is biphasic. ■ The P-waves and T-waves should be small compared to the QRS. They should be less than about 0.2 mV. Changing the Setting for Analysis Leads and Displaying Leads The default system setting for the analysis leads and displaying leads is [On], which means that the analysis leads are consistent with the displaying leads. You can change the default setting as following steps, if necessary. 1. At the CS, click [Admin Setup] → input the password → [OK]. 2. In the [Admin Setup] menu, click the [Telemetry] tab. 3. On the left side of the [Telemetry] tab, click [Analysis Lead Setting]. The corresponding setting displays to the right of [Telemetry]. 4. Set [Consistent with Displayed Lead] to [On] or [Off ]. ◆ [On]: the analysis lead is the same as the displaying lead. ◆ [Off ]: the analysis lead is different from the displaying lead. TMS60 Operator’s Manual 9 - 13 ECG Monitoring 9.3.2.2 Monitoring with the TD60 at the CS Changing ECG Wave Settings In the [Waveform Setup] menu, you can configure the displaying leads and analysis leads as desired. The lead settings are dependent on the setting of [Analysis Lead Setting]. ■ When [Analysis Lead Setting] is set to [On], the ECG waveform configurations are applied for the displaying leads and analysis leads. Follow the steps below to configure the ECG waveform setting: 1. In the [HR] section of the [ECG] tab, click [Waveform Setup]. The [Waveform Setup] menu displays. 2. Select the options described in the following table to configure the ECG waveforms. Settings* Options Description Lead Waveform size ECG 1 Select the desired ECG lead and set the corresponding gain for ECG 1. I, II, III, aVR, aVL, aVF, V × 0.125, × 0.25, × 0.5, × 1, × 2, × 4 ECG 2 Select the desired ECG lead and set the corresponding gain for ECG 2. I, II, III, aVR, aVL, aVF, V × 0.125, × 0.25, × 0.5, × 1, × 2, × 4 ECG 3 Select the desired ECG lead and set the corresponding gain for ECG 3. I, II, III, aVR, aVL, aVF, V × 0.125, × 0.25, × 0.5, × 1, × 2, and × 4 The factory default settings are in bold. 3. ■ Click [Exit] to save the settings and close the [Waveform Setup] menu. When [Analysis Lead Setting] is set to [Off ], the displaying leads and analysis leads should be configured respectively. Follow the steps below to configure the ECG waveform setting: 1. In the [HR] section of the [ECG] tab, click [Waveform Setup]. The [Waveform Setup] menu displays. 2. 9 - 14 Select the options described in the following table to configure the ECG waveforms. TMS60 Operator’s Manual Monitoring with the TD60 at the CS ECG Monitoring Settings* Options Description Lead Waveform size ECG 1 Select the desired ECG lead and set the corresponding gain for ECG 1 to display in the waveform area. I, II, III, aVR, aVL, aVF, V × 0.125, × 0.25, × 0.5, × 1, × 2, × 4 ECG 2 Select the desired ECG lead and set the corresponding gain for ECG 2 to display in the waveform area. I, II, III, aVR, aVL, aVF, V × 0.125, × 0.25, × 0.5, × 1, × 2, × 4 ECG 3 Select the desired ECG lead and set the corresponding gain for ECG 3 to display in the waveform area. I, II, III, aVR, aVL, aVF, V × 0.125, × 0.25, × 0.5, × 1, × 2, × 4 Primary lead Configures the primary analysis lead. I, II, III, aVR, aVL, aVF, V None Secondary lead Configures the secondary analysis lead. I, II, III, aVR, aVL, aVF, V None Classification lead (only for Mortara ECG algorithm) Configures other analysis lead. I, II, III, aVR, aVL, aVF, V None The factory default settings are in bold. 3. 9.3.3 Click [Exit] to save the settings and close the [Waveform Setup] menu. Other Settings The following table lists all settings in the [Other Settings] menu. Options Description Paced The option is unavailable at the CS. The paced status can be set at the TD60 only. Refer to "Checking the Paced Status" on page 7 - 10 for details. Pacer Reject Configures whether or not to reject the pace pulses. ■ [On]: the pace pulses are not counted as extra QRS complexes. ■ [Off]: the pace pulses are not rejected. Settings* On, Off This [Pacer Reject] option is only available when [Paced] is set to [Yes] at the TD60. The factory default settings are in bold. TMS60 Operator’s Manual 9 - 15 ECG Monitoring Monitoring with the TD60 at the CS Options Description Settings* Filter Configures the ECG filter in all operating modes. Monitor, ST ■ [Monitor]: use under measurement conditions. ■ [ST]: use when the ST monitoring is applied. normal The factory default settings are in bold. 9.3.4 ECG Display 9.3.4.1 ECG Digital Area The ECG digital area displays: 1. Area name 2. HR value 3. High HR alarm limit 4. Low HR alarm limit 5. Pauses threshold 6. Pauses per minute value 7. Pauses per minute label 8. PVCs threshold 9. PVCs per minute value 10. PVCs per minute label 10 ECG digital area 9 - 16 TMS60 Operator’s Manual Monitoring with the TD60 at the CS ECG Monitoring NOTE • 9.3.4.2 If the Activation State for HR, PVCs or Pauses alarm is set to [Off], the symbol displays to the right of corresponding parameter. ECG Waveform Area The ECG waveform area displays: 1. ECG Lead 2. ECG waveform size 3. ECG filter setting 4. Notch filter setting 5. ECG waveform 6. ECG scale ECG waveform area TMS60 Operator’s Manual 9 - 17 QT Monitoring 9.4 Monitoring with the TD60 at the CS QT Monitoring A normal ECG waveform (as shown in the following figure) typically includes a sharp and well defined QRS complexes with consistent spacing between R waves, and an ECG baseline that is free of noise and artifact. ST Point J Point ISO Point QT Interval A normal ECG waveform (for QT monitoring) The QT interval in an ECG lead is the time interval from the onset of the earliest deflection in the QRS complex to the end of the T wave. QT monitoring can assist in the detection of prolonged QT interval syndrome. 9.4.1 Measurement Limitations 9.4.2 ■ At least one measured V-lead must be available in order for the algorithm to process QT. ■ QT/QTc values are calculated with 3-leadwire or 5-leadwire ECG cables. QT Settings The following table lists the QT settings in the [QT Analysis] section of the [ECG] tab. Options Description Settings* QT Analysis Enables or disables QT analysis. On, Off 9 - 18 The factory default settings are in bold. TMS60 Operator’s Manual Monitoring with the TD60 at the CS Options Description QT Computational Formula Configures the QTc formula used. QT Monitoring Settings* Bazett: QTHR- QTc = QT × -----------60 Fridericia: 1/2 QTHR- QTc = QT × -----------60 1/3 60 ------------- Framingham: QTc = QT + 154 × 1 – QTHR Hodges: QTc = QT + 1.75 × ( QTHR – 60 ) QT Alarm Setup QT View Configures the QT alarms. ■ For details about the QT alarms limits, refer to "Factory Default Parameter Alarm Limits" on page 9 - 3 for details. ■ For details about the QT alarms responses, refer to "Factory Default Parameter Alarm Settings" on page 9 - 5 for details. ■ For Mindray algorithm, you can configure the QTc and ∆QTc alarm. ■ For Mortara algorithm, you can configure the QTc alarms. Select to display the [QT View] menu. Refer to "QT View Menu (Only for Mindray ECG Algorithm)" on page 9 - 20 for details. This button is only available for the Mindray ECG algorithm. The factory default settings are in bold. TMS60 Operator’s Manual 9 - 19 QT Monitoring 9.4.3 Monitoring with the TD60 at the CS QT View Menu (Only for Mindray ECG Algorithm) In the [QT View] menu, you can view a snapshot of the real-time wave and to verify that the QT algorithm detects correct Q and T points. The [QT View] menu displays, as shown in the following figure: ◆ The current waveform and parameter values display in green. ◆ The template waveform and parameter values display in yellow. ◆ The Q and T points are marked with a vertical line. ◆ The ∆QTc value is equal to the current QTc value minus the template QTc value. QT View menu Using the buttons described in the following table as desired. Buttons Description Settings* Lead Select the desired lead to display on the [QT View] menu screen. II, I, III, aVR, aVL, aVF, V Set as Ref. Replaces the template waveform and QT/ QTc values with the current waveform and QT/QTc values. None The QT template updated time displays at the bottom of the screen. Print Prints the template and current waveforms, and QT/QTc values for all leads. None Exit Closes the [QT View] Menu. None 9 - 20 The factory default settings are in bold. TMS60 Operator’s Manual Monitoring with the TD60 at the CS 9.4.4 QT Monitoring QT Display When [QT Analysis] is enabled, the QT digital area displays: 1. Area name 2. QTc value 3. High QTc alarm limit 4. Activation State Off symbol for QTc alarm 5. ∆QTc label 6. ∆QTc value 7. High ∆QTc alarm limit 8. Activation State Off icon for ∆QTc alarm 9. QT value 10. QT label 11. QTc label 12. QT-HR label 13. QT-HR value 11 10 Example QT digital area (for Mindray ECG algorithm) 12 13 Example QT digital area (for Mortara algorithm) TMS60 Operator’s Manual 9 - 21 ST Monitoring 9.5 Monitoring with the TD60 at the CS ST Monitoring ST deviation Depression or Elevation ST Point J Point ISO Point 40 to 80 msec ST Segment A normal ECG waveform (for ST monitoring) The ST segment of an ECG waveform (as shown in the above figure) represents the period from the end of ventricular de-polarization, to the beginning of ventricular repolarization, or the end of the QRS complex (the J point) and the beginning of the Twave. ST Segment analysis is used to monitor the oxygen supply and the viability of the heart muscle. ST deviation is the vertical distance between the isoelectric (ISO) point level and signal level at ST point. The ISO point is located between the end of the P-wave and the onset of the QRS complex. The ISO point provides the baseline for this measurement. The ST point is a fixed distance from the J point at the end of the QRS complex. The ST point can be configured to 40, 60, or 80 milliseconds past the J-point, independent of the heart rate. By default, the ST point is positioned as follows: ■ at 80 milliseconds for heart rates less than or equal to 120 beats per minute ■ at 60 milliseconds for heart rates greater than 120 beats per minute All available ECG leads are analyzed to measure deviations in the ST segment. Selecting leads that contain the least amount of baseline flutter will improve measurement accuracy, but accurate ST deviation measurement is dependent on the correct location of the ISO and ST points. 9 - 22 ■ ST segment analysis calculates ST segment elevations and depressions for individual leads and then displays them as numerics in the ST digital area. ■ A positive value indicates ST segment elevation; a negative value indicates ST segment depression. TMS60 Operator’s Manual Monitoring with the TD60 at the CS ST Monitoring WARNING • 9.5.1 The ST algorithm has been tested for accuracy of the ST segment data. The significance of the ST segment changes need to be determined by a clinician. Measurement Limitations 9.5.2 ■ ST values may be affected by such factors as some drugs or metabolic and conduction disturbances. ■ Since ST is often calculated with a fixed delay from the J point, changes in heart rate may affect ST. ■ The ST algorithm has been tested for accuracy of the ST segment data. The significance of the ST segment changes needs to be determined by a physician. ST Settings The following table lists the ST settings in the [ST Analysis] section of the [ECG] tab. Options or Buttons Description Settings* ST Analysis Enables or disables ST analysis. On, Off If [ST Analysis] is set to [On], the [Filter] option from the [Other Settings] menu is automatically set to [ST]. Display ST Segments Select whether or not to display the ST segments in the waveform area. On, Off ST Alarm Setup Configures the ST alarm settings. ■ For details about the ST alarms limits, refer to "Factory Default Parameter Alarm Limits" on page 9 - 3 for details. ■ For details about the ST alarms responses, refer to "Factory Default Parameter Alarm Settings" on page 9 - 5 for details. Define ST Point Select to display the [Define ST Point] menu. Refer to "Adjusting ST Measurement Points" on page 9 - 24 for details. This button is only available when [ST Analysis] is set to [On]. The factory default settings are in bold. TMS60 Operator’s Manual 9 - 23 ST Monitoring 9.5.3 Monitoring with the TD60 at the CS Adjusting ST Measurement Points The ISO and ST points need to be adjusted when you start monitoring and if the patient’s heart rate or ECG morphology changes significantly. Exceptional QRS complexes are not considered for ST-segment analysis. WARNING • 1. Always make sure that the ST measurement points are appropriate for your patient. In the [ST Analysis] section of the [ECG] tab, click [Define ST Point]. The [Define ST Point] menu displays, as shown in the following figure. Define ST point menu 2. Adjust the parameter using the buttons described in the following table. Buttons Description Settings* Lead Select the desired ECG lead. I, II, III, aVR, aVL, aVF, V ST (ms) Based on the ms setting selected, moves the ST point further or closer to the J point in the ST template. J+40, J+60, J+80, J+60/80 9 - 24 The factory default settings are in bold. TMS60 Operator’s Manual Monitoring with the TD60 at the CS ST Monitoring Buttons Description Settings* ISO/J Selecting [Auto] fixes the ISO and J/ST points. Selecting [Manual] allows the clinician to manually adjust ISO and J/ST points. Auto, Manual For Mortara algorithm, the ISO and J/ST points can only be adjusted manually. ISO left arrow If [ISO/J] is set to manual, the button adjusts the ISO reference line to the left. ISO right arrow If [ISO/J] is set to manual, the button adjusts the ISO reference line to the right. J left arrow If [ISO/J] is set to manual, the button adjusts the J reference line to the left. J right arrow If [ISO/J] is set to manual, the button adjusts the J reference line to the right. The factory default settings are in bold. 3. 9.5.4 Select the [OK] or [Cancel] button. ◆ The [OK] button saves the settings and closes the menu. ◆ The [Cancel] button closes the menu without saving the settings. ST Display When [ST Analysis] is enabled, the ST digital area displays: 1. Area name 2. Activation State Off icon for ST alarm 3. ST values 4. Units of measure 5. Lead identifier Example 5-lead ST digital area TMS60 Operator’s Manual 9 - 25 ST Monitoring 9.5.5 Monitoring with the TD60 at the CS ST Segment Display When [Display ST Segments] is enabled, the ST segments display in the waveform area: 1. Lead identifier 2. ST markers (ISO, J/ST) 3. ST value 4. ECG scale 3-lead ST segment 5-lead ST segments 9 - 26 TMS60 Operator’s Manual Monitoring with the TD60 at the CS 9.6 Arrhythmia Monitoring Arrhythmia Monitoring WARNING • The arrhythmia analysis program is intended to detect ventricular arrhythmias. It is not designed to detect atrial or supraventricular arrhythmias. It may incorrectly identify the presence or absence of an arrhythmia. Therefore, a physician must analyze the arrhythmia information with other clinical findings. • 9.6.1 Heart-rate reading may be affected by cardiac arrhythmias. Do not rely entirely on heart rate alarms when monitoring patients with arrhythmia. Always keep these patients under close surveillance. Measurement Limitations ■ Since the arrhythmia detection algorithm sensitivity and specificity is less than 100%, sometimes there may be some false arrhythmias detected and also some true arrhythmia events may not be detected. This is especially true when the signal is noisy. ■ The ECG size and QRS width settings affect arrhythmia detection and heart rate calculation sensitivity. ■ If QRS amplitude is low, the monitor might not be able to calculate HR and false asystole may occur. ■ At the algorithm learning phase, arrhythmia detection may not be available. Therefore, closely monitor the patient’s condition during the learning phase and after the learning phase to allow the algorithm to reach optimal detection performance. 9.6.2 Understanding the Arrhythmia Events 9.6.2.1 Mindray ECG Algorithm Arrhythmia message Description Category Asystole No QRS detected within the set time threshold in absence of ventricular fibrillation or chaotic signal. Lethal arrhythmia VFib/VTac A fibrillatory wave for 6 consecutive seconds. A dominant rhythm of adjacent Vs and a HR > the V-Tac HR limit. VTac The consecutive PVCs ≥ Vtac PVCs limit, and the HR ≥ the Vtac rate limit. TMS60 Operator’s Manual 9 - 27 Arrhythmia Monitoring Monitoring with the TD60 at the CS Arrhythmia message Description Category Vent. Brady The consecutive PVCs ≥ the Vbrd threshold and the ventricular HR < the Vbrd Rate threshold. Lethal arrhythmia Extreme Tachy The heart rate is equal to or greater than the extreme tachycardia limit. Extreme Brady The heart rate is equal to or less than the extreme bradycardia limit. PVCs/min PVCs/min exceeds high limit. R on T R on T detected in normal heartbeats. Run PVCs More than 2 consecutive PVCs. Couplet Paired PVCs detected in normal heartbeats. PVC One PVC detected in normal heartbeats. Vent. Rhythm The consecutive PVCs ≥ the Vbrd PVCs limit, and the HR ≥ Vbrd Rate limit but < the Vtac Rate limit. Bigeminy A dominant rhythm of N, V, N, V, N, V. Trigeminy A dominant rhythm of N, N, V,N, N, V, N, N, V. Tachy The average heart rate is equal to or greater than the tachycardia limit. Brady The average heart rate is equal to or less than the bradycardia limit. Pacer Not Pacing No pace pulse detected for 1.75 x average R-to-R intervals following a QRS complex (for paced patients only). Pacer Not Capture No QRS complex detected for 300 milliseconds following a pace pulse (for paced patients only). Missed Beat No beat detected for 1.75 x average R-R interval for HR <120, or No beat for 1 second with HR > 120 (for non-paced patients only), or No beat detected for more than the set pause threshold. Multif. PVC Multiform PVCs detected in Multif. PVC's Window (which is adjustable). Nonsus. Vtac The consecutive PVCs < the Vtac PVCs limit but > 2, and HR ≥ the Vtac Rate limit. Pause No QRS detected within the set time threshold of pause. AFib Continuously detect that the RR intervals of normal sinus beats irregular and no P wave of normal sinus beats. Irr. Rhythm Consistently irregular rhythm. Pauses/min The number of pauses detected per minute. 9 - 28 Nonlethal arrhythmia TMS60 Operator’s Manual Monitoring with the TD60 at the CS 9.6.2.2 Arrhythmia Monitoring Mortara ECG Algorithm Arrhythmia message Description Category Asystole No QRS complex detected within the set time threshold (in absence of ventricular fibrillation or chaotic signals). Lethal arrhythmia VFib Ventricular fibrillation occurs and persists for 6 seconds. VTac Ventricular HR is greater or equal to the preset threshold and the number of consecutive PVCs is greater than the preset threshold. Extreme Tachy The heart rate is equal to or greater than the extreme tachycardia limit. Extreme Brady The heart rate is equal to or less than the extreme bradycardia limit. PVCs/min PVCs/min exceeds high limit. R on T R on T is detected. Run PVCs More than 2 consecutive PVCs. Couplet Paired PVCs are detected. Vent. Rhythm Ventricular HR is less than the preset threshold and the number of PVCs is greater than or equal to 3. Bigeminy A dominant rhythm of N, V,N, V, N, V. Trigeminy A dominant rhythm of N, N, V,N, N, V, N, N, V. Tachy The HR is greater than the set tachycardia high limit. Brady The HR is less than the set bradycardia low limit. Pacer Not Pacing No pace pulse detected for (60×1000/pace rate +90) milliseconds following a QRS complex or a pacer pulse (for paced patients only). Pacer Not Capture No QRS complex detected for 300 milliseconds following a pace pulse (for paced patients only). Multif. PVC More than 2 PVCs of different forms occur in the predefined search window (3-31). Pause No beat detected for 1.75x average R-R interval for HR <120, or No beat for 1 second with HR >120 (for non-paced patients only), or No beat detected for more than the set pause threshold. Irr. Rhythm Consistently irregular rhythm Pauses/min The number of pauses detected per minute. TMS60 Operator’s Manual Nonlethal arrhythmia 9 - 29 Arrhythmia Monitoring 9.6.3 Monitoring with the TD60 at the CS Arrhythmia Settings The following table lists the arrhythmia settings in the [Arrhythmia Analysis] section of the [ECG] tab. Options or Buttons Description Settings* Arrhythmia Alarms Configures the arrhythmia alarm settings. Refer to "Factory Default Arrhythmia Alarm Settings" on page 9 - 6 for details. Relearn Enables an arrhythmia relearning. Refer to "Relearning" on page 9 30 for details. QRS Threshold Settings Configures the QRS threshold. Refer to "Configuring the QRS Threshold" on page 9 - 31 for details. 9.6.4 Relearning A relearn can be done for arrhythmia, ST analysis, or for both simultaneously. NOTE • ST Analysis must be turned on before it will relearn. The TD60 initiates the learning process for ST or Arrhythmia analysis after any of the following: ■ TD60 power-up ■ Return to normal monitoring from the Standby mode ■ Enabling ST or Arrhythmia analysis ■ The lead has been changed in ECG 1 waveform (3 lead only) ■ Patient size changes ■ Selecting the Relearn button in the [Arrhythmia Analysis] section Selecting the Relearn button in the [Arrhythmia Analysis] section is recommended after one or more of the following: 9 - 30 ■ ECG electrodes have been repositioned ■ Eight hours have passed since the last relearn ■ Significant changes occurred to the patient QRS complex TMS60 Operator’s Manual Monitoring with the TD60 at the CS Arrhythmia Monitoring ■ Significant changes occurred to the patient ECG rhythm ■ A clinician has observed clinically questionable arrhythmia calls ■ “Learning” occurred during a Leads Off condition CAUTION • 9.6.5 Initiate ECG relearning only during periods of normal rhythm and when the ECG signal is relatively noise-free. If ECG learning takes place during ventricular rhythm, the ectopics may be incorrectly learned as the normal QRS complex. This may result in missed detection of subsequent events of V-Tach and V-Fib. Configuring the QRS Threshold The minimum detection threshold is approximately 0.16 mV. In case the P waves are very tall, one might consider moving the minimum QRS detection threshold up to be above the level of the P waves, so events like ventricular standstill are not missed. Two horizontal lines (one below and one above the baseline) appear on the screen. These represent the current minimum detection threshold on the positive and negative sides of the baseline so manual adjustment for both positive and negative going QRS's can be made. Using the commands provided, move the minimum detection threshold up or down to ensure it is above the level of the P waves but below the peak of the Rwave. Since the P-wave height could vary a little from beat to beat, do not set the horizontal line representing the minimum detection threshold at or barely above the level of the peak of the P wave. Ensure that it is at least one or two millimeters above the peak of the P wave but below the peak of the R-wave. 1. In the [Arrhythmia Analysis] section of the [ECG] tab, click [Minimum QRS Threshold]. The [Minimum QRS Threshold] menu displays, as shown in the following figure. TMS60 Operator’s Manual 9 - 31 Arrhythmia Monitoring Monitoring with the TD60 at the CS Minimum QRS Threshold menu The current waveform displays the data of previous eight seconds. Use the buttons described in the following table as desired. Buttons Description Settings* Gain Select the desired ECG waveform size. X1, X2, X4 Refresh Displays the real-time waveform. Default Automatically sets the default threshold: 0.16 mV. Manually adjust the minimum QRS detection threshold. or Select the button to move the threshold line above the P wave, or select the button to move the threshold line down closer to the P wave. The factory default settings are in bold. 2. 9 - 32 Select the [OK] or [Cancel] button. ◆ Once the threshold is in the desired position, select the [OK] button to save the settings and close the menu. ◆ To quit adjusting the threshold, select the [Cancel] button to close the menu without saving the settings. TMS60 Operator’s Manual Monitoring with the TD60 at the CS 9.7 SpO2 Monitoring SpO2 Monitoring At the CS, you can view and change the SpO2 settings in the [SpO2] tab of the [Parameter Setup] menu. For details about the [Parameter Setup] menu, refer to the BeneVision Central Station Operator’s Manual (P/N 046-005011-00). 9.7.1 Measurement Limitations Refer to "Measurement Limitations" on page 8 - 2 for details. 9.7.2 SpO2 Settings The following table lists the SpO2 settings in the [SpO2] tab. Options Description Settings* Activation State Configures whether or not to enable the SpO2 alarm. On, Off Alarm Priority Configures the SpO2 alarm levels. High, Med Record on Alarm Selects whether or not to activate the recorder when an SpO2 alarm is triggered. On, Off Sweep Speed Configures the SpO2 waveform sweep speed. 6.25 mm/s, 12.5 mm/s, 25.0 mm/ s, 50.0 mm/s Desat Limit Activation State Configures whether or not to enable the SpO2 Desat alarm. On, Off Sensitivity (Masimo only) The option is not configurable. The option setting is synchronous with the setting at the TD60. Refer to "Configuring the SpO2 Setup" on page 8 - 5. SpO2 High Limit (%) Configures the SpO2 high alarm limit. (Low limit + 1) to 100 The default is 100. SpO2 Low Limit (%) Configures the SpO2 low alarm limit. 0 to (high limit - 1) The default is 90. PR High Limit (bpm) Configures the PR high alarm limit. (Low limit + 2) to 300 The default for adult is 120. The default for pediatric is 160. PR Low Limit (bpm) Configures the PR low alarm limit. 18 to (high limit - 2) The default for adult is 50. The default for pediatric is 75. The factory default settings are in bold. TMS60 Operator’s Manual 9 - 33 SpO2 Monitoring Monitoring with the TD60 at the CS Options Description Settings* Desat Limit (%) Configures the SpO2 Desat low limit. 0 to 100 The default is 80. Averaging (Masimo only) The option is not configurable. The option setting is synchronous with the setting at the TD60. Refer to "Configuring the SpO2 Setup" on page 8 - 5 for details. The factory default settings are in bold. 9.7.3 SpO2 Display 9.7.3.1 SpO2 Digital Area The SpO2 digital area displays: 1. Area name 2. High SpO2 alarm limit 3. PR label 4. PR value 5. High PR alarm limit 6. SpO2 unit of measure 7. Low PR alarm limit 8. Masimo perfusion index value 9. PI label 10. Low SpO2 alarm limit 11. SpO2 value 11 10 Masimo SpO2 digital area 9 - 34 TMS60 Operator’s Manual Monitoring with the TD60 at the CS SpO2 Monitoring NOTE • When [HR Source] of the [ECG] tab is set to [Both], the PR value displays on the SpO2 digital area. Nonin SpO2 digital area 9.7.3.2 SpO2 Waveform Area The SpO2 waveform area displays: 1. Area name 2. Pleth waveform SpO2 waveform area TMS60 Operator’s Manual 9 - 35 SpO2 Monitoring Monitoring with the TD60 at the CS This page intentionally left blank. 9 - 36 TMS60 Operator’s Manual 10 Configuring the TD60 Introduction...............................................................................................................10-2 Maintenance Menu .................................................................................................10-2 TMS60 Operator’s Manual 10 - 1 Introduction 10.1 Configuring the TD60 Introduction This TD60 Maintenance menu provides access to the system settings such as location, device name, alarm settings, quick keys, screen lock, and password updates. Entering this menu requires a password. 10.2 Maintenance Menu The [Maintenance] menu contains the following submenus: 10.2.1 10.2.2 ■ General ■ Alarms ■ Quick Keys ■ Defaults ■ Screen Lock ■ Edit Passcodes ■ Device Name ■ Demo Mode ■ Service Entering the Maintenance menu 1. In the main menu, tap [Maintenance]. 2. Input the maintenance password. 3. Tap [Accept] to enter the [Maintenance] menu. Configuring the General Menu Select [General] to configure the display auto off, language, location, notch filter, ECG lead labeling, SpO2 module, SpO2 tone, and enable or disable ECG calibration. 1. In the [Maintenance] menu, tap [General]. The current setting displays to the right of the option. 2. Select the options described in the following table. Options Description Settings* Display Auto Off Configures the time for display auto off. 1 min, 2 min, 5 min, 15 min 30 min, Off Language Configures the system language. ENGLISH, FRENCH 10 - 2 The factory default settings are in bold. TMS60 Operator’s Manual Configuring the TD60 Maintenance Menu Options Description Settings* Location Changes the hospital and department names. N/A Notch Filter Configures the ECG Notch filter. This option is used to filter out AC line noise from the ECG waveform. 50 Hz, 60 Hz, Off The default is 50 Hz when the device is not configured for the US. ECG Lead Labeling Changes the ECG lead labeling. AHA, IEC Calibrate ECG Enables or disables the ECG verification. On, Off Refer to "Verifying the ECG at the TD60" on page 14 - 4 for details. SpO2 Module Changes the SpO2 module. Masimo, Nonin SpO2 Tone Configures the SpO2 tone. Mode 1, Mode 2 The factory default settings are in bold. 3. Tap to exit the [General] menu. NOTE • 10.2.3 Mindray recommends the same SpO2 tone mode be used for the device within a monitoring area. Configuring the Alarms Menu Select [Alarms] to configure the alarm tone, reminder tone, reminder interval, minimum alarm volume, and technical alarm priority. 1. In the [Maintenance] menu, tap [Alarms]. The current setting displays to the right of the option. 2. Select the options described in the following table. Section & Options Description Settings* Allows an authorized user to set the alarm tone pattern. ISO, Mode 1, Mode 2 Sounds Style The factory default settings are in bold. TMS60 Operator’s Manual 10 - 3 Maintenance Menu Configuring the TD60 Section & Options Description Settings* Reminder Tone Allows an authorized user to enable or disable the reminder tone. If the alarm tone is turned off, enabling this setting can issue a periodic reminder tone. On, Off Reminder Interval Allows an authorized user to configure the intervals between the alarm tones. 1 min, 2 min, 3 min, 5 min, 10 min Timeout Minimum Alarm Volume Technical Allows an authorized user to set the minimum technical alarm volume. The minimum technical alarm volume defines the minimum value you can set for the technical alarm volume. For example: If the minimum technical alarm volume is set to 5, the minimum value you can set for the technical alarm volume in the [Audio Volume] menu is 5 (as shown in the following figure). Off, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10 Minimum value If the minimum technical alarm volume is set to Off, and the technical alarm volume can be set to 0, the alarm sound is turned off and the symbol appears on the screen. Technical Alarm Priority ECG Lead Off Allows an authorized user to configure the alarm level. Low, Medium, High SpO2 Sensor Off Allows an authorized user to configure the alarm level. Low, Medium, High The factory default settings are in bold. 3. 10 - 4 Tap to exit the [Alarms] menu. TMS60 Operator’s Manual Configuring the TD60 Maintenance Menu WARNING • • When the technical alarm audio volume is set to alarm sound and the alarm sound is turned off, the TD60 will not enunciate technical alarms when they occur. Be careful when turning off the alarm volume. Do not rely exclusively on the audible alarm system for monitoring. Adjustment of alarm volume to a low level may result in a hazard to the patient. Always keep the patient under close surveillance. 10.2.4 Quick Keys Menu 10.2.4.1 Changing the Quick Keys 1. In the [Maintenance] menu, tap [Quick Keys]. The [Quick Keys] configuration menu displays. 2. From the quick keys area at the bottom of the screen, tap a quick key you want to configure. A list of option displays. TMS60 Operator’s Manual 10 - 5 Maintenance Menu Configuring the TD60 A list of options 10.2.4.2 3. Tap the desired option from the list of options to configure the selected quick key. 4. Repeat steps 2 to 3 to configure other quick keys, if needed. 5. Tap to exit the [Quick Keys] menu. Deleting a Quick Key 1. From the quick keys area at the bottom of the [Quick Keys] menu, press and hold the desired quick key for two seconds, and then release it. The quick key background turns to red and displays [Delete]. 2. Tap [Delete]. The quick key is removed from the quick keys area, and the area displays [Not Used]. 3. 10.2.5 Tap to exit the [Quick Keys] menu. Configuring the Defaults Menu The [Defaults] menu allows an authorized user to manage the system configurations. 1. In the [Maintenance] menu, tap [Defaults]. 2. Select the options described in the following table. Options Description Save Departmental Defaults Allows an authorized user to save the current device settings for the selected patient category. Export Device Settings Copies the user configuration to the external device. 10 - 6 TMS60 Operator’s Manual Configuring the TD60 Maintenance Menu Options Description Import Device Settings Copies the settings from the external device to theTD60 unit. Refer to "Transferring a Configuration" on page 10 - 7 for details. Restore Factory Defaults Allows an authorized user to reestablish the original database power up settings to factory default values. 3. 10.2.6 Tap to exit the [Defaults] menu. Transferring a Configuration The TD60 is capable of configuring multiple devices through one export operation via a wireless interface. To transfer the configuration from the TD60 unit to an external device: 1. In the [Maintenance] menu of an external device, tap [Defaults] → [Import Device Settings] to enter the settings import mode. 2. In the [Maintenance] menu of the TD60 unit, tap [Defaults] → [Export Device Settings] to enter the [Export Device Settings] menu. The discovered external devices are listed in the [Export Device Settings] menu. 3. Select the desired external devices by tapping the check box. 4. Tap [Export] to start exporting the TD60 unit configuration. The selected external devices screen will shortly display the “Downloading device settings” message. WARNING • Do not power off the devices during the download process. ■ On the external devices if an import is successful, the external device will display the “Download complete.” message, and after 10 seconds return to the [Defaults] menu. On the TD60 unit, the status for the external device will display the “Complete” message. ■ On the external device if the import fails, the external device will display the “Import attempt failed.” message. On the TD60 unit, the status for the failed external device will display the “Failure” message and remain in the list. The user has two options when a failure occurs: ◆ TMS60 Operator’s Manual To stop the import from the external device, tap the [Import Device Settings]. icon to exit the 10 - 7 Maintenance Menu Configuring the TD60 ◆ To retry the import from the TD60 unit, tap the external device which is displaying the [Failure] message, tap the [Retry] button to restart the transfer. You may need to repeat the retry operation several times until the transfer is successful. 10.2.7 Screen Lock Menu 10.2.7.1 Understanding the Screen Lock Mode There are two modes of being able to lock the screen to assist in preventing unauthorized use. Each mode allows the user access to certain features of the product without entering a passcode. When the correct passcode is entered, all features are available. ■ ■ 10.2.7.2 Locked Mode Features: ◆ Main screen and main menu are not accessible without passcode entry. ◆ The message area is still viewable. ◆ Hardkeys are enabled. View Only Features: ◆ Upon powering up, the main screen will be displayed after the new patient choice is made. ◆ Main Screen is accessible without passcode entry. ◆ The [System Info] menu is accessible by tapping the battery symbol on the main screen. ◆ The [Lead Placement] menu is accessible by tapping an “ECG Lead Off” message in the message area. ◆ Ability to change display orientation. ◆ Hardkeys are enabled. Setting the Screen Lock The initial enabling of screen lock mode requires a passcode to be entered immediately as follows: 1. In the [Maintenance] menu, tap [Screen Lock] to select a screen lock mode. 2. Enter a new passcode for the screen lock. After the passcode is entered, the screen exits the passcode setup menu. The selected lock mode displays to the right of [Screen Lock]. 10 - 8 TMS60 Operator’s Manual Configuring the TD60 10.2.7.3 Changing the Current Screen Lock Passcode 1. 10.2.8 In the [Maintenance] menu, tap [Screen Lock]. 2. Tap [Screen Lock Passcode]. 3. Input the current password. 4. Input and verify the new password Changing the Passwords 1. In the [Maintenance] menu, tap [Edit Passwords]. 2. 10.2.9 Maintenance Menu ◆ Tap [Maintenance Password] and follow the on-screen instructions to change the maintenance password. ◆ Tap [Service Password] and follow the on-screen instructions to change the service password. Tap the icon to exit the [Edit Passwords] menu. Changing the Device Name 1. In the [Maintenance] menu, tap [Device Name]. 2. Use the on-screen keyboard to input the device name. 3. Tap [Accept] to save the setting and exit the [Device Name] menu. NOTE • Do not set the same device name for the TD60s. 10.2.10 Demo Mode Allows an authorized user to choose a demonstration mode for in-servicing staff or testing product features. 10.2.11 Service Menu Allows an authorized user access to the passcode protected Service menu. TMS60 Operator’s Manual 10 - 9 Maintenance Menu Configuring the TD60 This page intentionally left blank. 10 - 10 TMS60 Operator’s Manual 11 Battery Introduction...............................................................................................................11-2 Safety............................................................................................................................11-2 Installing the Battery ..............................................................................................11-3 Checking the Battery Charge Status .................................................................11-4 Removing the Battery ............................................................................................11-4 Charging the Rechargeable Lithium-ion Battery .........................................11-5 Storing the Batteries ...............................................................................................11-6 Maintaining the Rechargeable Lithium-ion Battery....................................11-7 Disposing of the Batteries.....................................................................................11-8 TMS60 Operator’s Manual 11 - 1 Introduction 11.1 Battery Introduction The TD60 can be powered by a lithium-ion rechargeable battery or AA batteries. This chapter provides instructions on how to use, maintain, and dispose of the batteries. 11.2 Safety SE WARNING • • • • • • • • • • 11 - 2 Keep the batteries out of children’s reach. Keep the batteries in their original package until you are ready to use them. The battery compartment should be closed during defibrillation. Only use specified AA batteries or rechargeable lithium-ion battery to power the TD60. Other power supply may cause damage to the equipment or lead to body injury. While installing AA batteries, do not apply reverse polarity. Only use specified fresh AA batteries. Using other AA batteries can give unacceptable performance. Do not mix old and new AA batteries. Only use specified rechargeable lithium-ion battery. Unspecified lithium-ion battery can give unacceptable performance. Use caution when handling the rechargeable lithium-ion battery. Misuse or abuse may cause bodily injury or device damage. ◆ Do not short circuit. Take care that the terminals do not contact metal or other conductive materials during transport and storage. ◆ Do not crush, drop or puncture the battery. Mechanical abuse can lead to internal damage and internal short circuits. If a battery has been dropped or banged against a hard surface, whether damage is externally visible or not, remove the battery from use and dispose of it properly. ◆ Do not incinerate batteries or expose them to temperatures above 60°C (140°F). The rechargeable lithium-ion batteries should be charged in the specified central charger. TMS60 Operator’s Manual Battery Installing the Battery WARNING • • • • If a battery shows signs of damage or signs of leakage, replace it immediately. Use caution in removing the battery. Avoid contact with skin. Refer to qualified service personnel. Some failure conditions, such as short circuits, can cause a battery to overheat during using. High temperature can cause burns to the patient or user. If the device becomes too hot to the touch, remove it from the patient and place aside until it cools. Then remove the battery from the device, and contact your service personnel to identify the cause of overheating. Replace the battery immediately once the “Critically Low Battery” alarm message displays. Replace the battery in time once the “Low Battery” alarm message displays. If those conditions are not corrected, device shutdown and cessation of monitoring will result. To eliminate the risk of electrical shock or burn, do not carry loose batteries on your person, such as placing the battery in clothing pockets. CAUTION • • 11.3 Remove the battery before transporting the device or if the device is not in use or is being stored. AA batteries should be removed from the device at the end of the battery’s useful life to prevent leakage. In case of battery leakage, use caution to remove the batteries and clean the battery compartment. Install fresh AA batteries and check if the TD60 can power on properly. If the TD60 fails to power on, contact your service personnel. Installing the Battery Refer to "Installing the Batteries" on page 3 - 5 for details. TMS60 Operator’s Manual 11 - 3 Checking the Battery Charge Status 11.4 Battery Checking the Battery Charge Status The battery symbol displaying on the top of main screen indicates the battery charge status. The white part ( ) indicates the remaining battery charge. NOTE • 11.5 If the “Low Battery” or “Critically Low Battery” alarm occurs, the TD60 turns off the audio, the screen display the “Local Audio Off” message, and the screen brightness turns dimmer. Removing the Battery CAUTION • • • Some failure conditions, such as short circuits, can cause a battery to overheat during using. High temperature can cause burns to the patient or user. If the device becomes too hot to the touch, remove it from the patient and place aside until it cools. Then remove the battery from the device, and contact your service personnel to identify the cause of overheating. Avoid scraping the metal contact in the battery compartment while removing the lithium-ion battery pack or AA battery frame. Otherwise, the broken contact will affect the power supply performance. Remove the battery before transporting the device or if the device is not in use or is being stored. NOTE • 1. 11 - 4 Retain the ECG cable and SpO2 module with the device while removing the battery. Lift up the lithium-ion battery pack or AA battery tray at the bottom of the TD60. TMS60 Operator’s Manual Battery Charging the Rechargeable Lithium-ion Battery 2. 11.6 Remove the lithium-ion battery pack or AA battery tray from the TD60. Charging the Rechargeable Lithium-ion Battery WARNING • • • • • • Only use the specified central charger to charge to the lithium-ion batteries designated by Mindray. Only use the approved power cord with the grounded mains plug to firmly connect the central charger to a grounded AC mains socket. Never refit the mains plug to fit an ungrounded AC mains socket. Do not use the Multiple Portable Socket Outlets (MPSO) or AC mains extension cords. Use an IEC 60601-1 approved isolation / separation transformer, otherwise, it may result in leakage current. Insure that the sum of the individual ground leakage currents does not exceed the allowable limits. Do not place any shield object (such as cloth or paper) to cover the central charger or batteries, and keep ventilated while charging the lithium-ion batteries. Do not connect other devices to the power supply system. Do not use the central charger to charge the lithium-ion batteries in high temperature above 40°C. Use the central charger to charge the lithium-ion batteries. The central charger can charge 10 lithium-ion batteries at one time. For details about the central charger, refer to the BeneVision Central Charger Operator’s Manual (P/N 046-007059-00). TMS60 Operator’s Manual 11 - 5 Storing the Batteries Battery 11.7 Storing the Batteries 11.7.1 Storing Rechargeable Lithium-ion Battery When storing batteries, make sure that the battery terminals do not come into contact with metallic objects. If you need to store the batteries for an extended period of time, place the batteries in a cool, dry place (ideally at 15°C or 60°F) with a partial charge of about 50% capacity (two LEDs illuminated). Storing batteries in a cool place can slow the aging process. Stored batteries should be charged to about 50% of their capacity every six months. The battery should be fully charged prior to first use. NOTE • • • 11.7.2 Remove the lithium-ion battery from the device if the device is not used for a prolonged time (for example, several weeks), and keep the device in clean place to avoid the dust or liquid entering the battery compartment. Storing batteries at high temperatures for an extended period of time will significantly shorten their life expectancy. Do not store the batteries in an environment above 60°C (122°F) or lower than -20°C (4°F). Storing AA Batteries If you remove undepleted AA batteries from the TD60 and need to store the batteries, keep the batteries together as a set for later re-use so that all batteries will have the same level of remaining power. Do not store disposable AA batteries by leaving the batteries in the incorrect polarity position in the TD60. 11 - 6 TMS60 Operator’s Manual Battery Maintaining the Rechargeable Lithium-ion Battery NOTE • 11.8 Replace the AA battery frame on the battery compartment after removing the AA batteries. Maintaining the Rechargeable Lithium-ion Battery Take care of the rechargeable lithium-ion battery once you receive a new battery for use. The following table describes the battery maintenance activities and recommended frequency. Activity Recommended Frequency Visual inspection Before installing a battery in the TD60. Charge the battery Upon receipt, after use, a “Low Battery” or “Critically Low Battery” alarm occurs. To optimize performance, a fully or almost fully discharged battery must be charged immediately. Clean the battery At each patient discharge, or in case that the battery is exposed to contaminants. Do not clean the battery connector during the cleaning. Charge stored battery to at least 40% of the battery capacity. Every six months if the TD60 is not in use for an extended period of time. Dispose of the battery When the “Battery Maintenance Required” alarm message displays on the TD60. The lifetime of a lithium-ion battery depends on the frequency and duration of use. With good maintenance, the useful life is approximately four years or 500 complete chargedischarge cycles. In addtion, experience indicates that the incidence of failure may increase with battery service life due to the accumulated stresses of daily use. Therefore, Mindray strongly recommend that lithium-ion battery should be replaced after two years or 300 complete charge-discharge cycles. Using the outdated battery may cause the device abnormity and unacceptable performance. The age of a lithium-ion battery begins at the date of manufacture. The date of manufacture is listed on the rear of the battery. TMS60 Operator’s Manual 11 - 7 Disposing of the Batteries Battery NOTE • The battery capacity degrades as using time and number of recharge cycles.Toward the end of its useful life, the battery capacity may be reduced by 20% to 25%. If the reduced battery life is unacceptable for your device, Mindray recommends the battery be replaced. 11.9 Disposing of the Batteries 11.9.1 Disposing of the Rechargeable Lithium-ion Battery Discard the lithium-ion battery in the following situations: ■ The battery has visual signs of damage. ■ The battery fails. ■ The battery is aged and its runtime significantly less than the specification. ■ The battery has been used for more than two years or 500 complete charge-discharge circles. Discharge the battery and insulate the terminals with tape before disposal. Properly dispose of the batteries according to local regulations. 11.9.2 Disposing of the AA Batteries The batteries may be subject to local regulations regarding disposal. Dispose of batteries in approved containers. Follow local regulations, if any, to recycle the batteries. 11 - 8 TMS60 Operator’s Manual 12 Troubleshooting General Problems ....................................................................................................12-2 Physiological Alarm Messages at the CS .........................................................12-3 Technical Alarm Messages at the TD60 ...........................................................12-5 TMS60 Operator’s Manual 12 - 1 General Problems 12.1 Troubleshooting General Problems The following table lists the problems that are likely to occur. If the problem persists after corrective actions have been taken, contact your service personnel. 12.1.1 12.1.2 Troubleshooting Tools ■ Telemetry Receiver (RC60) ■ TD60 ■ Central Station (CS) Problem List Symptom Possible cause Solution No RF Signal (The RC60 does not receive valid data for consecutive 5 seconds) 1. The TD60 battery charge is to be depleted. 1. Check if the TD60 battery charge is depleted. 2. The patient is out of the antenna array coverage area. 2. Check if the patient is out of the coverage area. 3. The TD60 is not connected with the ECG leadwire. 3. Check if the TD60 is connected with the ECG leadwire properly. RF Interference (The TD60 receives three consecutive wrong frames.) The TD60 signal is weak. Check if the Patient is at the edge of the coverage area or in an elevator. Remove behind a reinforced concrete wall. SpO2 No Pulse The SpO2 sensor failed to obtain pulse signal. Check the patient’s condition and change the sensor application site. If the error persists, replace the sensor. The CS does not display the SpO2 data. 1. The SpO2 module is not connected to the TD60. 1. Connect the SpO2 module to the TD60. 2. There may be error in the SpO2 module. 2. If there is an error in the SpO2 module, replace the SpO2 module with a new one. ECG noise ECG waveforms are overlapped with the noise interference. Check if the ECG leadwire is intertwined with cables of other devices. ECG Signal Saturated The TD60 detects that ECG signal saturation or overload. 1. Check if patient has excessive movement. 2. Check if the electrodes are in good contact with the skin. 3. Check if the electrodes operating time is over the electrode service life. Wrong Channel 12 - 2 The configured frequency of the transmitter does not match the channel. Reconfigure the frequency of the transmitter to match the channel. TMS60 Operator’s Manual Troubleshooting Physiological Alarm Messages at the CS Symptom Possible cause Solution The AC power indicator on the RC60 is off. 1. The RC60 is not connected to the power. 1. Check if the RC60 is connected to the power. 2. The RC60 is power off. 2. Check if the RC60 sounds a beep when turn the receiver on. The TD60 or SpO2 module restarts repeatedly. The TD60 battery charge is to be depleted. 12.2 Replace with new batteries. Physiological Alarm Messages at the CS The following table lists the major telemetry-related physiological alarm messages displayed on the central station (CS) screen. Measurement Alarm message Alarm level Possible cause Solution Heart rate (HR) HR High Medium* HR Low Medium* HR value has risen above the high alarm limit or fallen below the low alarm limit. Check the patient’s condition and make sure that the patient category and alarm limit settings are correct. The asterisk (*) means the alarm level is configurable. TMS60 Operator’s Manual 12 - 3 Physiological Alarm Messages at the CS Troubleshooting Measurement Alarm message Alarm level Possible cause Solution ECG ECG Weak Signal High The ECG signal is so weak that the monitor can’t perform ECG analysis. Check the patient’s condition and the ECG connections. Asystole High The patient is in arrhythmia. Vfib/Vtac High Check the patient’s condition and the ECG connections. VTac High Vent. Brady High Extreme Tachy High Extreme Brady High R on T Medium* VT > 2 Low* Couplet Message* PVC Message* PVCs/min Medium* Bigeminy Medium* Trigeminy Medium* Tachy Medium* Brady Medium* Missed Beat Message* Vent. Rhythm Medium* Pause Medium* PNP (Pacer Not Pacing) Message* The pacer appears abnormal. Check the pacer. PNC (Pacer Not Capture) Message* 12 - 4 The asterisk (*) means the alarm level is configurable. TMS60 Operator’s Manual Troubleshooting Technical Alarm Messages at the TD60 Measurement Alarm message Alarm level Possible cause Solution SpO2 SpO2 Desat High The SpO2 value has fallen below the desaturation alarm limit. Check the patient’s condition and check if the alarm limit settings are correct. No Pulse High The pulse signal was so weak that the monitor cannot perform pulse analysis. Check the patient’s condition, SpO2 sensor and measurement site. 12.3 The asterisk (*) means the alarm level is configurable. Technical Alarm Messages at the TD60 The following table lists the major technical alarm messages displayed in the message area of the TD60. The Alarm Indication column in the table below is capable of three different indication types: A, B, C. ■ A: The alarm sound will be silenced, the alarm light will continue to indicate the alarm, a √ will appear before the alarm message. The top of the main screen. ■ B: The technical alarm will be changed to the prompt message, it will not longer make sound or be indicated by the alarm light. ■ C: The technical alarms are cleared, there will be no alarm indications. Measurement ECG symbol appears on the Alarm message Alarm level Alarm Indication ECG Lead Off Low* ECG XX** Lead Off Low* ECG Module Error High Possible cause Solution The electrode has become detached from the patient or the lead wire has become disconnected from the adapter cable. Check the connections of the electrodes and leadwires. An error occurred to the ECG module. Restart the TD60. If the problem persists, contact your service personnel. There is a problem with the communications between the module and the TD60. The asterisk (*) means the alarm level is configurable. ** XX is the ECG lead name. TMS60 Operator’s Manual 12 - 5 Technical Alarm Messages at the TD60 Measurement ECG SpO2 Troubleshooting Alarm message Alarm level Alarm Indication Possible cause Solution ECG Noise Low The ECG signal is noisy. Check for any possible sources of signal noise around the cable and electrode, and check the patient for great motion. ECG Cable Type Error Low Connect wrong ECG leadwire. Reconnect the 3 or 5 lead ECG leadwire. HR Overrang Low HR exceeds the measurement limit. Contact your medical pesonnel. SpO2 Sensor Off Low* The SpO2 sensor has become detached from the patient or the module. SpO2 Sensor Fault Low Check the sensor application site and the sensor type, and make sure the sensor is not damaged. SpO2 No Sensor Low SpO2 Module Error High SpO2 Too Much Light Low There is too much light on the SpO2 sensor. Move the sensor to a place with lower level of ambient light or cover the sensor to minimize the ambient light. SpO2 No Pulse Low SpO2 sensor failed to obtain pulse signal. Move the sensor to a site with better perfusion. SpO2 Unplugg ed Low SpO2 module connector is disconnected from the TD60. Reconnect the SpO2 module to the TD60. PR Overrang Low The measured PR value exceeds the measurement range. Contact Mindray or your service personnel. There is a fault with the SpO2 sensor. The asterisk (*) means the alarm level is configurable. ** XX is the ECG lead name. 12 - 6 Reconnect the sensor or use a new sensor. An unspecified SpO2 sensor has been used. TMS60 Operator’s Manual Troubleshooting Measurement Power System Technical Alarm Messages at the TD60 Alarm message Alarm level Alarm Indication Possible cause Solution Low Battery Medium The battery charge is low. Replace with new batteries. Critically Low Battery High The battery charge is almost depleted. Battery Mainten ance Required Medium The lithium-ion battery is aging. Battery Error Medium The lithium-ion battery communication is error. Battery Type Error Medium The battery contact is not making adequate connection. Device Error High The selftest of the TD60 main board is error. The selftest of the parameter module or Mindray PAN module are error. Restart the TD60. If the problem persists, contact your service personnel. The parameter module communication is error or initialization is error. Real Time Clock Error High The real time clock initialization is error. Restorin g Last Defaults Failed Low Restoring the last default configuration is error. Loading Defaults Failed Low Loading the default configuration is error. The asterisk (*) means the alarm level is configurable. ** XX is the ECG lead name. TMS60 Operator’s Manual 12 - 7 Technical Alarm Messages at the TD60 Troubleshooting This page intentionally left blank. 12 - 8 TMS60 Operator’s Manual 13 Cleaning and Disinfecting Introduction...............................................................................................................13-2 Safety Information...................................................................................................13-2 Cleaning of the TD60..............................................................................................13-3 Cleaning the Reusable ECG Leadwires, SpO2 Modules and Sensors ....13-4 Cleaning the Battery and Battery Compartment .........................................13-5 Disinfection................................................................................................................13-5 Sterilization ................................................................................................................13-5 TMS60 Operator’s Manual 13 - 1 Introduction 13.1 Cleaning and Disinfecting Introduction Only use the substances approved by Mindray and methods listed in this chapter to clean or disinfect your device. Our warranty does not cover damage caused by unapproved substances or methods. Mindray makes no claims regarding the efficacy of the listed chemicals or methods as a means for controlling infection. For the method to control infection, consult your hospital’s Infection Control Officer or Epidemiologist. 13.2 Safety Information WARNING • • • • • 13 - 2 The responsible hospital or institution shall carry out all cleaning and disinfection procedures specified in this chapter. Be sure to shut down the system before cleaning the equipment. Non-medical equipment does not provide the same level of protection against electrical shock. Do not touch the patient and any part of nonmedical equipment at the same time. Some examples of non-medical equipment are laser printers and nonmedical computers. Avoid use of cleaners, materials or chemicals that may damage device surfaces, labels, or cause equipment failures. Keep your equipment and accessories free of dust and dirt. To avoid damage to the equipment, follow these guidelines: ◆ Always dilute according to the manufacturer’s instructions or use lowest possible concentration. ◆ Do not immerse any part of the device into liquid. Do not pour liquid onto the equipment or accessories. ◆ Do not allow liquid to enter the case and the device interior. ◆ Never use abrasive materials (such as steel wool or silver polish), or erosive cleaners (such as acetone or acetone-based cleaners). TMS60 Operator’s Manual Cleaning and Disinfecting Cleaning of the TD60 CAUTION • • • 13.3 If liquid has accidentally entered the system or its parts, shut down the system and have the device serviced by authorized service personnel. Remove the equipment from use if liquid is spilled on the equipment or accessories. Contact your service personnel. When cleaning, avoid the ECG leadwire connector and other connectors. Cleaning of the TD60 CAUTION • Only use the following approved cleaning solutions. The system may become inoperable or halted because of contamination or damage caused by use of unapproved cleaning solution. Clean your equipment on a regular basis. Before cleaning the equipment, consult your hospital’s regulations for cleaning the equipment. Recommended cleaning agent are: ■ Water ■ Mild soap Before cleaning your TD60, do the following preparations: ■ Install the battery pack or battery tray to firmly close the battery compartment. ■ Insert the SpO2 cover in the SpO2 connector when SpO2 is not in use. ■ Insert the ECG leadwire plug into the ECG connector. WARNING • Never allow the cleaning solutions to spill or enter the plug, connector and battery compartment. To clean your TD60, follow this procedure: 1. Shut down the device. TMS60 Operator’s Manual 13 - 3 Cleaning the Reusable ECG Leadwires, SpO2 Modules and Sensors 13.4 Cleaning and Disinfecting 2. Dilute the mild soap in water to make a cleaning solution. 3. Soak a clean and soft cloth in the solution and wring out excess solution. 4. Thoroughly wipe the display screen and the exterior surface of the device with the damp cloth, avoiding the connectors. 5. Wipe off all the cleaning solution with a dry cloth if necessary. 6. Dry your device in a ventilated, cool place. Cleaning the Reusable ECG Leadwires, SpO2 Modules and Sensors ■ Check reusable sensors and cables daily for signs of damage. Replace as required. ■ Clean the sensors before and after each new patient. ■ Wipe sensors and cables using a soft cloth with mild soap and water solution. ■ Allow the cables and sensors to completely dry before using. Refer to the cleaning procedure described in "Cleaning of the TD60" on page 13 - 3. CAUTION • • • • Never immerse cables and sensors in any fluids. Do not clean them with harsh chemicals such as acetone or non-diluted bleach. Clean the cables carefully to avoid breaking internal wires by excessive bending, strain, or flexing. Do not autoclave, radiation or steam sterilize the cables and sensors. Extended exposure to Ethylene Oxide gas may shorten life of the cables, leading to poor signal quality. NOTE • 13 - 4 Refer to the individual instruction sheets that are packaged with the accessories for additional information. TMS60 Operator’s Manual Cleaning and Disinfecting 13.5 Cleaning the Battery and Battery Compartment Cleaning the Battery and Battery Compartment ■ Clean the exterior surface of the lithium-ion battery pack and AA battery tray before and after each new patient. ■ Wipe the lithium-ion battery pack, AA battery tray and the battery compartment using a soft cloth with mild soap and water solution. Use caution to avoid the battery connector. ■ Allow the lithium-ion battery pack, AA battery tray and the battery compartment completely dry before using. Refer to the cleaning procedure described in "Cleaning of the TD60" on page 13 - 3. CAUTION • 13.6 Never immerse the lithium-ion battery pack, AA battery tray and AA batteries. Do not clean them with harsh chemicals such as acetone or nondiluted bleach. Disinfection Disinfect the TD60 and accessories as required in your hospital’s servicing schedule. Cleaning before disinfecting is recommended. The recommended disinfectants are: 13.7 ■ 70% isopropyl alcohol ■ 10% sodium hypocholride (bleach) solution ■ 3% hydrogen peroxide ■ Virkon ■ Super Sani-cloth (0.5% Quaternary ammonium chloride and 55% Isopropyl alcohol) ■ 50% propyl alcohol (1-propyl alcohol) ■ 70% ethanol Sterilization Sterilization is not recommended for this equipment, related products, accessories or supplies unless otherwise indicated in the Operating Instructions that accompany the accessories or supplies. TMS60 Operator’s Manual 13 - 5 Sterilization Cleaning and Disinfecting This page intentionally left blank. 13 - 6 TMS60 Operator’s Manual 14 Maintenance Introduction...............................................................................................................14-2 Safety............................................................................................................................14-2 Regular Check ...........................................................................................................14-3 Maintenance and Testing Schedule..................................................................14-4 Checking the System Information .....................................................................14-4 Verifying the ECG at the TD60 .............................................................................14-4 TMS60 Operator’s Manual 14 - 1 Introduction 14.1 Maintenance Introduction The chapter outlines the routine telemetry devices maintenance guidelines. The telemetry devices are designed for stable operation over long periods of time. Under normal circumstances the devices should not require technical maintenance beyond that described in this chapter. However, routine maintenance, calibration and safety checks are recommended at least once a year or more often as required by local statutory or hospital administration practice. 14.2 Safety WARNING • • • • • • • • 14 - 2 Failure on the part of the responsible individual hospital or institution employing the use of this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards. The safety checks or maintenance involving any disassembly of the equipment should be performed by service personnel. Otherwise, undue equipment failure and possible health hazards could result. No modification of this equipment is allowed. To avoid the electric shock, do not open the central charger housing, otherwise . All servicing and future upgrades must be carried out by the service personnel. All replaced components and accessories as well as consumables are provided or qualified by Mindray. If you discover a problem with any of the equipment, contact your service personnel or Mindray. The service personnel must be properly qualified and thoroughly familiar with the equipment operation. TMS60 Operator’s Manual Maintenance 14.3 Regular Check Regular Check Perform a visual inspection before the equipment is first used every day. Verify that the equipment meets the following requirements: ■ The housing and display screen are free from cracks or other damages. ■ All keys function properly. ■ Connectors are not loose, cracked, or bent and cables have no cuts, nicks, or fraying. ■ ECG leadwires are securely connected with the equipment. ■ Battery pack is installed and has sufficient charge. ■ Chest electrodes are free from cracks and limb electrodes can properly clamp. After your equipment has been used for 6 to 12 months, or whenever your equipment is repaired or upgraded, a thorough inspection should be performed by qualified service personnel to ensure the reliability. Follow these guidelines when inspecting the devices: ■ Make sure that the environment and power supply meet the requirements. ■ Inspect the devices and their accessories for mechanical damage. ■ Inspect all plugs, connectors, leadwires for damage, and make sure that their insulation is in good condition. ■ Make sure that only specified accessories are applied. ■ Make sure that the alarm system functions correctly. ■ Make sure that the battery meet the performance requirements. ■ Make sure that the devices are in good working condition. In case of any damage or abnormity, do not use the devices. Contact the hospital’s biomedical engineers or your service personnel immediately. 14.3.1 Power-on Test The TD60 performs a self-test during startup. You can refer to "Powering On the Unit" on page 3 - 8 for details. 14.3.2 Battery Check For details about the battery charge check and maintenance, refer to "Maintaining the Rechargeable Lithium-ion Battery" on page 11 - 7. TMS60 Operator’s Manual 14 - 3 Maintenance and Testing Schedule 14.4 Maintenance Maintenance and Testing Schedule The following maintenance and tests, except for visual inspection, power on test, and battery check, shall be carried out by the service personnel only. Contact your service personnel if any maintenance is required. Make sure to clean and disinfect the equipment before any test and maintenance. Check/Maintenance Item Recommended Frequency Visual inspection When first installed or reinstalled. ECG test and verification Performance test 1. If the user suspects that the measurement is incorrect. Verification 2. Following any repairs or replacement of relevant module. 3. Once every two years. SpO2 test Power on test Battery check Functionality test Performance test 14.5 1. When first installed or reinstalled. 2. Following any maintenance or the replacement of any main unit parts. 1. When first installed. 2. Whenever the battery is replaced. When the battery run time reduced significantly. Checking the System Information To view the information about the device, radio frequency (RF), battery, MPAN, and system statistics, you can go to the main menu → [System Info]. 14.6 Verifying the ECG at the TD60 The ECG signal may be inaccurate due to hardware or software problems. As a result, the ECG wave amplitude becomes greater or smaller. To verify the ECG waveform amplitude: 1. In the main menu, tap [Maintenance]. 2. Input the maintenance passcode. 3. Tap [Accept]. 4. In the [Maintenance] menu, tap [Others]. 5. Enable [Calibrate ECG]. A square wave appears on the screen and the message [ECG Calibrating] is displayed. 14 - 4 TMS60 Operator’s Manual Maintenance Verifying the ECG at the TD60 6. Compare the amplitude of the square wave with the wave scale. The difference should be within 5%. 7. After completing the verification, disable [Calibrate ECG]. TMS60 Operator’s Manual 14 - 5 Verifying the ECG at the TD60 Maintenance This page intentionally left blank. 14 - 6 TMS60 Operator’s Manual 15 Accessories ECG Accessories........................................................................................................15-2 SpO2 Accessories .....................................................................................................15-3 Miscellaneous............................................................................................................15-5 TMS60 Operator’s Manual 15 - 1 ECG Accessories Accessories The accessories listed in this chapter comply with the requirements of IEC 60601-1-2 when in use with the device. The accessory material that contacts the patients has undertaken the bio-compatibility test and is verified to be in compliance with ISO 10993-1. For details about the accessories, refer to the instructions for use provided with the accessory. WARNING • • • • • Use accessories specified in this chapter. Using other accessories may cause damage to the monitor or not meet the claimed specifications. Single-use accessories are not designed to be reused. Reuse may cause a risk of contamination and affect the measurement accuracy. Check the accessories and their packages for any sign of damage. Do not use them if any damage is detected. Use the accessories before the expiry date if indicated. The disposable accessories shall be disposed of according to the hospital's regulations. 15.1 ECG Accessories 15.1.1 ECG Electrodes PN Description Applicable property Applicable patient 0010-10-12304 Adult Electrode (Kendall, package of 10) Disposable Adult 9000-10-07469 Pediatric ECG electrode (3M, package of 50) Disposable Pediatric 15 - 2 TMS60 Operator’s Manual Accessories 15.1.2 SpO2 Accessories ECG Leadsets 3-Lead PN Description Applicable property Applicable patient 009-004765-00 3-Lead, New Telemetry, AHA, Snap, 24" Reusable Adult, Pediatric 009-004766-00 3-Lead, New Telemetry, AHA, Snap, 36" 009-004771-00 3-Lead, New Telemetry, AHA, Pinch, 24" 009-004772-00 3-Lead, New Telemetry, AHA, Pinch, 36" 009-004777-00 3-Lead, New Telemetry, Disp, AHA, Snap, 24" Disposable PN Description Applicable property Applicable patient 009-004782-00 5-Lead, New Telemetry, AHA, Snap, 24" Reusable Adult, Pediatric 009-004783-00 5-Lead, New Telemetry, AHA, Snap, 36" 009-004786-00 5-Lead, New Telemetry, AHA, Pinch, 24" 009-004787-00 5-Lead, New Telemetry, AHA, Pinch, 36" 009-004790-00 5-Lead, New Telemetry, AHA, Snap, 24" 5-Lead 15.2 Disposable SpO2 Accessories The SpO2 sensor material that contacts patients or other staff has undertaken the biocompatibility test and is verified to be in compliance with ISO 10993-1. 15.2.1 Masimo SpO2 Module PN Description Applicable property Applicable patient 009-004936-00 Masimo SpO2 module (SET uSpO2) Reusable Adult, Pediatric TMS60 Operator’s Manual 15 - 3 SpO2 Accessories 15.2.2 Accessories Masimo SpO2 Sensor PN Description 0600-00-0121 LNCS Adtx-Adult Single Patient Adhesive Sensors, >30 kg (20/box) 0600-00-0122 LNCS Pdtx-Pediatric Single Patient Adhesive Sensors,10-50 kg (20/box) 0600-00-0126 LNCS DCI Adult Reusable Finger Sensor, >30 kg 0600-00-0127 LNCS DCIP Pediatric Reusable Finger Sensor, 1050 kg Applicable property Applicable patient Disposable Adult Pediatric Reusable Adult Pediatric Wavelength emitted by the sensors is between 600 nm and 1000 nm. The maximum photic output consumption of the sensor is less than 18 mW. The information about the wavelength range and maximum photic output consumption can be especially useful to clinicians (for example, when photodynamic therapy is performed). 15.2.3 Nonin SpO2 Module PN Description Applicable property Applicable patient 009-004935-00 Nonin SpO2 module (XPOD 3012LP) Reusable Adult 15.2.4 Nonin SpO2 Sensor PN Description Applicable property Applicable patient 0600-00-0139-24 Nonin 7000AA SpO2 Sensor (box of 24) Disposable Adult 100-000077-00 Nonin 8000AA SpO2 Sensor Reusable 100-000134-00 Nonin 8000AP SpO2 Sensor Reusable 100-000135-00 Nonin 7000AP SpO2 Sensor (box of 24) Disposable 15 - 4 Pediatric TMS60 Operator’s Manual Accessories 15.3 Miscellaneous Miscellaneous PN Description 115-026852-00 Main unit of the charger 022-000196-00 Telemetry Rechargeable battery 045-001698-00 TP-2AA battery frame 045-001699-00 TP-3AA battery frame 0000-10-10902 Alkaline 1.5 V AA battery 0146-00-0077-10 L91 AA battery 048-005247-00 Disposable pouch 048-005246-00 Reusable pouch TMS60 Operator’s Manual 15 - 5 Miscellaneous Accessories This page intentionally left blank. 15 - 6 TMS60 Operator’s Manual A Product Specifications Classifications............................................................................................................A-2 Environmental Specifications..............................................................................A-2 Power Supply Specifications................................................................................A-3 Physical Specifications ...........................................................................................A-4 Hardware Specifications........................................................................................A-5 Wireless Specification.............................................................................................A-7 Mindray Patient Area Network (PAN) Specification.....................................A-9 Measurement Specifications ...............................................................................A-10 TMS60 Operator’s Manual A-1 Classifications A.1 Product Specifications Classifications The Telemetry Monitoring System is classified, according to IEC60601-1: TD60: energized from an internal electrical power source. RC60: class I. Central Charger: class I. Type of protection against electrical shock Degree of protection against electrical shock for the TD60 Type CF defibrillation proof for ECG, and SpO2. Mode of operation Continuous Degree of protection against harmful ingress of water TD60: IPX7 RC60: IPX0 Central Charger: IPX0 Degree of protection against hazards of explosion Not suitable: Equipment not suitable for use in the presence of a flammable anesthetic mixture with air with oxygen or nitrous oxide. Degree of protection against hazard of dropping for the TD60 A.2 No damage by dropping from a height of 1.5 m. Environmental Specifications WARNING • The devices may not meet the performance specifications if stored or used outside the specified temperature and humidity ranges. Item Operating conditions Storage conditions Temperature 0 °C to 37 °C -20 °C to 60 °C Relative humidity (noncondensing) 15% to 95% 10% to 95% Barometric 427.5 mmHg to 805.5 mmHg, or 57.0 kPa to 107.4 kPa 120 mmHg to 805.5 mmHg, or 16.0 kPa to 107.4 kPa A-2 TMS60 Operator’s Manual Product Specifications Power Supply Specifications A.3 Power Supply Specifications A.3.1 TD60 The TD60 is powered by batteries. Battery type AA batteries (two or three) Rechargeable lithium-ion battery (one) Capacity for lithium-ion battery ≥ 3500 mAh Safety and authentication for lithium-ion battery UL2054, IEC62133, UN38.3 Run time for (at the temperature 25 °C ±5 °C, the screen display is off, no sound and light as default, no alarms, at least including 15 minutes operating time when the screen display is on, AA battery with PN 014600-0077-10 are used) Two AA batteries Three AA batteries Lithium-ion battery 5-lead ECG: ≥ 48 hours ≥ 72 hours ≥ 72 hours 5-lead ECG + Nonin SpO2 ≥ 36 hours ≥ 48 hours ≥ 48 hours 5-lead ECG + Masimo SpO2 ≥ 36 hours ≥ 48 hours ≥ 48 hours Power On/Off The startup logo displaying time is less than or equal to three seconds. The time for entering the main screen to display the waveforms is less than or equal to 15 seconds. Shutdown delay at least 15 minutes after the low battery alarm first occurs at least five minutes after the critically low battery alarm first occurs A.3.2 RC60 The telemetry receiver is powered by external AC power. Input voltage 100 to 240 VAC (± 10%) Frequency 50 Hz/60 Hz (± 3 Hz) Input current 0.6 A to 0.3 A TMS60 Operator’s Manual A-3 Physical Specifications A.3.3 Product Specifications Central Charger The central charger is powered by external AC power. Input voltage 100 VAC to 240 VAC (± 10%) Frequency 50 Hz/60 Hz (±3 Hz) Input current 1.5 A to 0.75 A At the room temperature: ≤ 5 hours Charge time Overcharge protection function The charge time definition: the time for the battery from the exhausted status to 90% battery charge (within the operating temperature range declared for the unit). The charger automatically stop charging when the lithium-ion battery charge is full. A.4 Physical Specifications A.4.1 TD60 Size 125.8 mm × 63.7 mm × 23 mm (only for the transmitter, without the ECG leadwire, SpO2 module and any other accessories) Weight 210 g (with two AA batteries and 3-lead ECG cable, without the SpO2 module and any other accessories) A.4.2 RC60 Size 120 mm × 300 mm × 350 mm Weight 7000 g A.4.3 Central Charger Size 365 mm × 170.6 mm × 77.9 mm (without batteries and wallmount bracket) Weight 1130 g (without batteries and wall-mount bracket) A-4 TMS60 Operator’s Manual Product Specifications Hardware Specifications A.5 Hardware Specifications A.5.1 TD60 Display Screen type Color TFT LCD screen Screen size 3.5" Resolution 480 pixels × 320 pixels Display Activation (Power On/Off) key The switch time is less than or equal to two seconds. LED Alarm lamp 1 (three colors: red, yellow, and cyan) Audio Indicator Speaker Sound Pressure Range 45 dBA to 85 dBA Keys Nurse call Power On/Off Main menu External Connectors ECG connector SpO2 connector TMS60 Operator’s Manual A-5 Hardware Specifications A.5.2 Product Specifications RC60 LED Communications indicator 1 (green), indicates the ready and communications status. Power indicator 1 (green), indicates the power status. Audio Indicator 1, indicates internal abnormity. Buzzer External Connectors Ethernet connector 1, standard RJ45 connector Antenna connector A.5.3 Central Charger Charger slot 10 LED 10, which indicates the battery charge status. AC power indicator Installation mode Place on the desktop, or mount on GCX® wall channel. A-6 TMS60 Operator’s Manual Product Specifications Wireless Specification A.6 Wireless Specification A.6.1 Technique Specification Protocol standard Private protocol Debugging mode GFSK Work frequency 608 MHz to 614 MHz Channel interval 25 KHz Wireless baud rate 8 kbps ± 3% Output power < 10 mW Receiving sensitivity ≤-110 dBm (≤ 1% error code rate) Data security Private protocol A.6.2 Implemented Functions The transmitter sends physiological data (such as ECG waveform and SpO2 parameter) and the transmitter status data (such as the status about lead, button, battery voltage, and others) to the receiver via the wireless network, and then the receiver sends those data to the central station (CS). The ECG data precision sent to the CS from the transmitter is 200 Hz, 8 bit per channel. The transmitter automatically networks with the CS after the transmitter is turned on. The time from the boot to connecting with the CS network is less than or equal to 15 seconds. The receiver can detect the wireless signal strength of the transmitter, and send the information to the CS. TMS60 Operator’s Manual A-7 Wireless Specification A.6.3 Product Specifications Function Specification WARNING • Do perform all network functions of data communication within an enclosed network. Data integrity Code error ≤ 1% Data latency Total delay of data transmission from the transmitter to the CS: ≤ 3 seconds Priority All communication data type have the same priority. Transmission distance Distinct vision distance between the transmitter and the receiver antenna is greater than or equal to 50 m. Receiver capacity 16 beds System capacity 32 beds Dynamic networking stability The wireless functions of the test transmitter are normal while the transmitter is moving at the rate of no more than 3.75 m/s within the wireless transmission distance. Resistance to wireless interference When the distance between the wireless device at the frequency of 2.4 GHz (including cellular mobile communication networks, microwave ovens, cordless phones, and other devices) and the receiver antenna is one meter, and the receiver antenna exists a WMTS interference at the same frequency of 118 dBm, the wireless functions are normal. Communication interruption alarm After the communication between the transmitter and the receiver is interrupted, the CS displays the alarm information within five seconds. Signal strength detection RSSI detection scope: -110 dBm to -80 dBm A-8 TMS60 Operator’s Manual Product Specifications Mindray Patient Area Network (PAN) Specification A.7 Mindray Patient Area Network (PAN) Specification A.7.1 Technique Specification Protocol standard Mindray PAN low energy 4.0 Debugging mode GFSK Work frequency 2400 MHz to 2483.5 MHz Channel interval 2 MHz Wireless baud rate 1 Mbps Output power ≤ 1 mW Data Security 128 bit AES A.7.2 Implemented Function The function implemented by the Mindray PAN is transferring the configuration between the transmitters. A.7.3 Function Specification Data integrity Code error ≤ 1% Data latency Total delay of data transmission between the transmitters: ≤ 1 second Priority All communication data type shall have the same priority. Transmission distance Distinct vision distance between the transmitters is greater than or equal to 3 m. System capacity 10 m2 space can hold 10 transmitters at the same time. Resistance to wireless interference When the distance between the wireless device at the frequency of 2.4 GHz (including cellular mobile communication networks, microwave ovens, cordless phones, and other devices) and the transmitter is one meter, the Mindray PAN communication functions are normal. Communication interruption alarm After the Mindray PAN communication between the transmitters is interrupted, the transmitter prompts that the communication is failed. TMS60 Operator’s Manual A-9 Measurement Specifications Product Specifications A.8 Measurement Specifications A.8.1 ECG ECG Standard Meet standards of IEC60601-2-27 and IEC60601-2-25 Leadset 3-lead: I, II, III 5-lead: I, II, III, aVR, aVL, aVF, V Automatic 3/5 lead recognition ECG standard AHA, IEC Sweep speed 6.25 mm/s, 12.5 mm/s, 25 mm/s Accuracy: ± 10% Display Sensitivity 1.25 mm/mV (×0.125), 2.5 mm/mV (×0.25), 5 mm/mV (×0.5), 10 mm/ mV (×1), 20 mm/mV (×2), 40 mm/mV (×4) Accuracy: ± 5% Input signal range ±8 mV (peak-to-peak value) Calibration signal 1 mV (peak-to-peak value) Accuracy: ± 5% Differential input impedance ≥ 5 MΩ Bandwidth (-3dB) Monitor mode: 0.5 Hz to 40 Hz ST mode: 0.05 Hz to 40 Hz Common mode rejection ratio (with Notch off) Monitor mode: > 105 dB ST mode: > 105 dB Notch 50/60 Hz, rejection capacity ≥ 20 dB Polarizing voltage scope ± 500 mV Noise ≤ 30 μV (p-v RTI) Baseline recovery time < 5 s (after defibrillation) Direct current leakage Input electrode: < 0. 1μA Drive electrode: <1μA Electrode polarization recovery time < 10 s A - 10 TMS60 Operator’s Manual Product Specifications Defibrillation energy absorption Measurement Specifications ≤ 10% (100 Ω load) Pace Pulse Pace pulse markers Pace pulses meeting the following conditions are labelled with a PACE marker: Amplitude: ±2 mV to ±700 mV Width: 0.1 ms to 2 ms Rise time: 10 μs to 100 μs Amplitude: ≥ 0.2mV RTI Pace pulse rejection When tested in accordance with the IEC60601-2-27: 201.12.1.101.13, the heart rate meter rejects all pulses meeting the following conditions. Amplitude: ±2 mV to ±700 mV Width: 0.1 ms to 2 ms Rise time: 10 μs to 100 μs About 50% pulse of ANSI/AAMI EC13 (5d) can trigger the pulse detector, the least changing rate is 20 V/s RTI. HR Measurement range Adult: 0, 15 bpm to 300 bpm Pediatric: 15 bpm to 350 bpm Resolution 1 bpm Accuracy ± 1 bpm or ± 1%, whichever is greater. Mindray ECG algorithm Mortara ECG algorithm 200 μV 160 μV to 480 μV Sensitivity TMS60 Operator’s Manual A - 11 Measurement Specifications HR averaging method Product Specifications Mindray ECG algorithm Mortara ECG algorithm In compliance with the requirements in Clause 201.7.9.2.9.101 b) 3) of IEC60601-2-27, the following method is used: If the last 3 consecutive RR intervals are greater than 1200 ms, the 4 most recent RR intervals are averaged to compute the HR. Otherwise, heart rate is computed by subtracting the maximum and minimum ones from the most recent 12 RR intervals and then averaging them. The HR value displayed on the monitor screen is updated every second. In compliance with the requirements in Clause 201.7.9.2.9.101 b) 3) of IEC60601-2-27, the following method is used: Heart rate is computed by averaging the most recent 16 RR intervals, unless the HR by averaging the most recent 4 heart beats is less than or equal to 48 bpm. The HR value displayed on the monitor screen is updated every second. Tall T-wave rejection capability When the test is performed based on Clause 201.7.9.2.9.101 b) 2)of IEC60601-2-27, the heart rate meter will reject all 100 ms QRS complexes with less than 1.2 mV of amplitude, and T waves with Twave interval of 180 ms and those with Q-T interval of 350 ms. Response time to heart rate change Meets the requirements of IEC60601-2-27: Clause 201.7.9.2.9.101 b) 5). From 80 bpm to 120 bpm: less than 11 s From 80 bpm to 40 bpm: less than 11 s Response to irregular rhythm In compliance with the requirements in Clause 201.7.9.2.9.101 b) 4) of IEC60601-2-27, the heart rate after 20 seconds of stabilization is displayed as follows: Ventricular bigeminy (3a): 80 ± 1 bpm Slow alternating ventricular bigeminy (3b): 60 ± 1 bpm Rapid alternating ventricular bigeminy (3c): 120 ± 1 bpm Bidirectional systoles (3d): 90 ± 2 bpm Time to alarm for tachycardia Meets the requirements of IEC60601-2-27: Clause 201.7.9.2.9.101 b) 6). Waveform 4ah - range: < 11 s 4a - range: < 11 s 4ad - range: < 11 s 4bh - range: < 11 s 4b - range: < 11 s 4bd - range: < 11 s ST Segment Analysis Measurement range A - 12 -2.0 mV to +2.0 mV RTI TMS60 Operator’s Manual Product Specifications Measurement Specifications Resolution 0.01 mV Accuracy -0.8 mV to +0.8 mV: ± 0.02 mV or ± 10%, whichever is greater Beyond this range: Not specified Arrhythmia Analysis Arrhythmia analysis classifications Mindray ECG algorithm Mortara ECG algorithm Asystole, VFib/VTac, Vtac, Vent. Brady, Extreme Tachy, Extreme Brady, PVCs/min, Pauses/min, R on T, Run PVCs, Couplet, Multif. PVC, PVC, Bigeminy, Trigeminy, Tachy, Brady, Pacer Not Pacing, Pacer Not Capture, Missed Beat, Nonsus. Vtac, Vent. Rhythm, Pause, Irr.Rhythm, Afib Asystole, VFib, Vtac, Extreme Tachy, Extreme Brady, PVCs/ min, Pauses/min, R on T, Run PVCs, Couplet, Multif. PVC, Bigeminy, Trigeminy, Tachy, Brady, Pacer Not Pacing, Pacer Not Capture, Vent. Rhythm, Pause, Irr.Rhythm QT Analysis QTc formula Bazett, Fridericia, Framingham, and Hodges Mindray ECG algorithm Mortara ECG algorithm QT measurement range [200, 800] ms [300, 600] ms QT accuracy [200, 800] ms: ± 30 ms Beyond this range: Not specified [300, 600] ms: ± 30 ms Beyond this range: Not specified QT resolution [200, 800] ms: 4 ms Beyond this range: Not specified [300, 600] ms: 2 ms Beyond this range: Not specified QTc measurement range [200, 800] ms [300, 600] ms QTc resolution [200, 800] ms: 1 ms Beyond this range: Not specified [300, 600] ms: 1 ms Beyond this range: Not specified QT-HR measurement range Adult: [15, 150] bpm Pediatric: [15, 180] bpm N/A TMS60 Operator’s Manual A - 13 Measurement Specifications A.8.2 Product Specifications SpO2 NOTE • A.8.2.1 A functional tester or SpO2 simulator can be used to determine the pulse rate accuracy. Nonin SpO2 Module SpO2 Measurement range 0% to 100% Resolution 1% Accuracy 70-100%: ±3 digits 0-69%: Not specified Data update rate ≤2s *Studies were performed to validate the accuracy of Pulse Oximeter with Nonin SpO2 sensors by contrast with a CO-Oximeter. In total, 12 adults (7 male & 5 female) aged from 19 to 35 years old were voluntarily involved in this study. Five are Caucasian, two are Indian, three are Asian, and two are African-American. All of them conform the clinical study requirements for age, gender, skin and health. The following table shows the accuracy (Arms) for the SpO2 sensors in four kinds of SaO2 ranges. For the Fitting Curve of the SpO2 sensors, refer to the "Fitting Curve for Nonin SpO2 Sensors" on page A - 15. SaO2 range Measured Arms Value (7000A, 7000P, 8000AA, 8000AP) 70% to 100% 1.54 70% to 80% 1.41 80% to 90% 1.97 90% to 100% 1.28 PR Measurement range 20 bpm to 300 pm Resolution 1 bpm A - 14 TMS60 Operator’s Manual Product Specifications Measurement Specifications ±3 bpm Accuracy Response Time ■ ≤ 25 s (PR 75 bpm, average time 4 heart beats, no disturbance, SpO2 value rises from 70% to 100%) ■ ≤ 35 s (SpO2 value 98%, average time 4 heart beats, no disturbance, PR value rises from 60 bpm to 150 bpm) A.8.2.2 Fitting Curve for Nonin SpO2 Sensors 7000A, 7000P, 8000AA, 8000AP SpO2 = 1.01 × SaO2 – 2.6 Scatter Plot of the SpO2 versus the SaO2 TMS60 Operator’s Manual A - 15 Measurement Specifications Product Specifications Bland-Altman Plots of the Bias versus the Mean A.8.2.3 Masimo SpO2 Module SpO2 Measurement range 1% to 100% Resolution 1% Accuracy 70 to 100%: ±2% (measured without motion in adult/pediatric mode) 70 to 100%: ±3% (measured with motion) 1% to 69%: Not specified. Low Perfusion Conditions Pulse amplitude: > 0.02% Light penetration: > 5% Low Perfusion Accuracy 2% *SpO2 accuracy was determined by testing on healthy adult volunteers in the range of 60% to 100% SpO2 against a laboratory CO-Oximeter. SpO2 accuracy was determined on 16 neonatal NICU patients ranging in age from 7-135 days old and weighing between 0.5-4.25 kg. Seventy-nine (79) data samples were collected over range of 70-100% SaO2 with a resultant accuracy of 2.9% SpO2. The following table shows the accuracy (Arms) for the SpO2 sensors in four kinds of SaO2 ranges. For the Fitting Curve of the SpO2 sensors, refer to the "Fitting Curve for Masimo SpO2 Sensors" on page A - 18. A - 16 TMS60 Operator’s Manual Product Specifications Measurement Specifications Measured Arms Values SaO2 range LNCS Adtx, LNCS Pdtx LNCS DCI, LNCS DCIP 70% to 100% ± 2% 2% 70% to 80% 1.55% 0.60% 80% to 90% 1.07% 0.54% 90% to 100% 1.64% 0.60% PR Measurement range 25 bpm to 240 bpm Resolution 1 bpm Accuracy ± 3 bpm (without motion) ± 5 bpm (with motion) Response Time ≤ 25 s (PR 75 bpm, average time 8 s, no disturbance, SpO2 value rises from 50% to 100%) ≤ 20 s (SpO2 value 98%, average time 8 s, no disturbance, PR value rises from 60 bpm to 150 bpm) PI Measurement range 0.02% to 20% Least resolution 0.01, use three valid digits Response Time ■ ≤ 25 s (PR 75 bpm, average time 8 s, no disturbance, SpO2 value rises from 50% to 100%) ■ ≤ 20 s (SpO2 value 98%, average time 8 s, no disturbance, PR value rises from 60 bpm to 150 bpm) TMS60 Operator’s Manual A - 17 Measurement Specifications A.8.2.4 A - 18 Product Specifications Fitting Curve for Masimo SpO2 Sensors TMS60 Operator’s Manual B EMC TMS60 Operator’s Manual B-1 EMC The telemetry monitoring system (TMS) meets the requirements of IEC 60601-1-2. WARNING • • • • • • Use of accessories and cables other than those specified may result in increased emission and/or decreased immunity of the system. Devices too close or stacked may interfere with each other. Do not put devices too close or stack them together. Keep a close eye on the system in case there are other devices around it. Neighboring frequencies of transmitters too close may interfere with each other, and the CS cannot receive data or receive erroneous data. Therefore it is recommended that the interval of neighboring frequencies is not smaller than 25kHz. Devices even in compliance with CISPR transmitting requirements may interfere with the system. If the input signal is lower than the specified threshold, measurements may be inaccurate. Other equipment that have RF transmit or source may affect this device (for example, cell phones, PDAs, PCs with wireless function). Guidance and declaration — electromagnetic emissions The system is intended for use in the electromagnetic environment specified below. The user of the system should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment — guidance RF emissions CISPR 11 Group 1 The system uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. RF emissions CISPR 11 Class A Harmonic Emissions IEC61000-3-2 Class A The system is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Voltage Fluctuations/Flicker Emissions IEC 61000-3-3 Compliance B-2 TMS60 Operator’s Manual EMC Guidance and declaration — electromagnetic immunity The system is intended for use in the electromagnetic environment specified below. The user of the system should assure that it is used in such an environment. Immunity test IEC 60601 Test level Compliance level Electromagnetic environment - guidance Electrostatic Discharge (ESD) IEC 61000-4-2 ± 6 kV contact ± 8 kV air Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. Electrical fast Transient/burst IEC 61000-4-4 ± 2 kV for power cord ± 1 kV for I/O cables Mains power quality should be that of a typical commercial or hospital environment. Surge IEC 61000-4-5 ± 1 kV different mode ± 2 kV common mode Voltage dips, Short interruptions and voltage variation on power supply input lines IEC 61000-4-11 <5% UT (>95% dip in UT ) for 0.5 cycle 40% UT (60% dip in UT ) for 5 cycle 70% UT (30% dip in UT ) for 25 cycle <5% UT (>95% dip in UT ) Mains power quality should be that of a typical commercial or hospital environment. If the user of our product requires continued operation during power mains interruptions, it is recommended that our product be powered from an uninterruptable power supply or a battery. for 5 sec Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. Note: UT is the A.C. mains voltage prior to application of the test level. TMS60 Operator’s Manual B-3 EMC Guidance and declaration — electromagnetic immunity The system is intended for use in the electromagnetic environment specified below. The customer or the user of the system should assure that it is used in such an environment. Immunity test IEC 60601 Test level Compliance level Conduced RF IEC 61000-4-6 3 Vrms 150kHz to 80MHz 3 Vrms Radiated RF IEC 61000-4-3 3 V/m 80MHz to 2.5GHz 3 V/m Electromagnetic environment - guidance Portable and mobile RF communications equipment should be used no closer to any part of the system, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance d = 1.2 d = 1.2 80 M to 800 MHz d = 2.3 800 M to 2.5 GHz where P is the maximum output power rating of the transmitter in watts (W ) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey a, should be less than the compliance level in each frequency range b. Interference may occur in the vicinity of equipment marked with the following symbol: Note: At 80 MHz and 800 MHz, the higher frequency range applies. Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Note: The TMS that intentionally receives RF electromagnetic energy at the exclusion band (2395.825MHz2487.645MHz) is exempt from the ESSENTIAL PERFORMANCE requirements, but remains safe. a: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the system is used exceeds the applicable RF compliance level above, the system should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the system. b: Over the frequency ranges 150kHz to 80MHz, field strengths should be less than 3V/m. B-4 TMS60 Operator’s Manual EMC If the system is operated within the electromagnetic environment listed in Table Guidance and declaration — electromagnetic immunity, the system will remain safe and provide the following essential performance: ■ Operating mode ■ Accuracy ■ Function ■ Accessories identification ■ Data stored ■ Alarm ■ Detect for connection Recommended separation distances between portable and mobile RF communication and the system The system is intended for use in an electromagnetic environment in which radiated RF disturbance are controlled. The customer or the user of the system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communication equipment (transmitters) and the system as recommended below, according to the maximum output power of the communication equipment. Separation Distance According to Frequency of Transmitter M (Meters) Rated Maximum Output power of Transmitter W (Watts) 150kHz -80MHz d= 3.5 --------3 80MHz to 800MHz d= 3.5 --------3 800MHz to 2.5GHz d= 0.01 0.12 0.12 0.23 0.1 0.37 0.37 0.74 1.17 1.17 2.34 10 3.69 3.69 7.38 100 11.67 11.67 23.34 7--3 For transmitters at a maximum output power not listed above, the recommended separation distanced in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. TMS60 Operator’s Manual B-5 EMC This page intentionally left blank. B-6 TMS60 Operator’s Manual C FCC Compliance The telemetry monitoring system (TMS) complies with the requirements of FCC Part 95: Authorized health care providers, in conjunction with the equipment manufacturers, must cooperate in the selection and use of frequencies in order to reduce the potential for interference with other wireless medical telemetry devices, or other co-primary users. Operations in the 608–614 MHz band (television channel 37) are not protected from adjacent band interference from broadcast television operating on channels 36 and 38. As the RF range of the system is 608-614Mhz, if located near the radio astronomy observatories the two parties will interfere with each other. Therefore, we don't suggest that the equipment can be installed or operated within 80 kilometers of: 1. National Astronomy and Ionosphere Center, Arecibo, Puerto Rico: 18°20′38.28″ North Latitude, 66°45′09.42″ West Longitude. 2. National Radio Astronomy Observatory, Socorro, New Mexico: 34°04′43″ North Latitude, 107°37′04″ West Longitude. 3. National Radio Astronomy Observatory, Green Bank, West Virginia: 38°26′08″ North Latitude, 79°49′42″ West Longitude. This device and its antenna must not be located or operating in conjunction with any other antenna and transmitter. This device complies with part 15 of the FCC rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. TMS60 Operator’s Manual C-1 FCC Compliance NOTE ● This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Radio Frequency Exposure This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. This transmitter must not be co‐located or operating in conjunction with any other antenna or transmitter. Operation of this equipment requires the prior coordination with a frequency coordinator designated by the FCC for the Wireless Medical Telemetry Service. Radio frequency Radiation Exposure Information For body worn operation, this equipment has been tested and meets the FCC and CE RF exposure guidelines when used with the accessories supplied or those approved for use with this product. Use of other accessories may not ensure compliance with FCC and CERF exposure guidelines within 32 kilometers of the National Radio Astronomy Observatory centered on: TMS60 Operator’s Manual C‐3 FCC Compliance Very long baseline array stations Latitude (north) Longitude (west) Pie Town, NM 34°18′ 108°07′ Kitt Peak, AZ 31°57′ 111°37′ Los Alamos, NM 35°47′ 106°15′ Fort Davis, TX 30°38′ 103°57′ North Liberty, IA 41°46′ 91°34′ Brewster, WA 48°08′ 119°41′ Owens Valley, CA 37°14′ 118°17′ Saint Croix, VI 17°46′ 64°35′ Mauna Kea, HI 19°49′ 155°28′ Hancock, NH 42°56′ 71°59′ If the installation distance is not enough, obtain the written concurrence of the Director of the affected radio astronomy station before the equipment can be installed or operated. RF Parameters TMS60 Operator’s Manual C‐3 FCC Compliance C-4 Item Parameter Operating Frequency Band (MHz) 608 to 614 Modulation GFSK Transmitter Output Power(e.i.r.p) (dBm) < 10 Channel Space (KHz) 25 TMS60 Operator’s Manual D Symbols and Abbreviations Units..............................................................................................................................D-2 Symbols .......................................................................................................................D-3 Abbreviations ............................................................................................................D-3 TMS60 Operator’s Manual D-1 Units D.1 Symbols and Abbreviations Units D-2 ampere Ah ampere hour bpm beats per minute °C centigrade cc cubic centimeter cm centimeter dB decibel DS dyne. second °F fahrenheit gram hr hour hPa hundred pascal Hz hertz in inch kilo kg kilogram kPa kilopascal litre lb pound meter mg milligrams min minute ml milliliter mm millimeters ms millisecond mV millivolt mW milliwatt nm nanometer ppm part per million second volt VA volt ampere Ω ohm μA microampere μm micron TMS60 Operator’s Manual Symbols and Abbreviations D.2 D.3 Symbols μV microvolt watt Symbols minus percent per; divide; or power plus equal to less than greater than ≤ less than or equal to ≥ greater than or equal to ± plus or minus × multiply © copyright Abbreviations AAMI Association for Advancement of Medical Instrumentation AC alternating current AHA American Heart Association ANSI American National Standard Institute ARR arrhythmia ART arterial AUX Auxiliary output aVF left foot augmented lead aVL left arm augmented lead aVR right arm augmented lead CH channel CISPR International Special Committee on Radio Interference CS central station ECG electrocardiograph EMC electromagnetic compatibility TMS60 Operator’s Manual D-3 Abbreviations D-4 Symbols and Abbreviations err error ES electrosurgical ESU electrosurgical unit HR heart rate HT height IEC International Electrotechnical Commission ISO International organization for standardization MRI magnetic resonance imaging LA(L) left arm LAP left atria pressure LED light emitting diode LL(F) left leg Loop loop read-write test fail M, MEAN mean pressure O2 oxygen power PAN Patient Area Network PR pulse rate QRS interval of ventricular depolarization (QRS complex) RA(R) right arm RL(N) right leg ROM read-only memory SpO2 arterial oxygen saturation from pulse oximetry VGA Video Graphics Array TMS60 Operator’s Manual E E.1 Anomaly Anomaly Description This product does not have any anomalies. TMS60 Operator’s Manual E-1 Anomaly Description Anomaly This page intentionally left blank. E-2 TMS60 Operator’s Manual P/N: 046-007056-00 (1.0)
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.6 Linearized : No Author : 50212800 Create Date : 2015:06:29 11:17:16Z Modify Date : 2015:07:13 16:42:07+08:00 Has XFA : No XMP Toolkit : Adobe XMP Core 4.2.1-c041 52.342996, 2008/05/07-20:48:00 Creator Tool : FrameMaker 8.0 Metadata Date : 2015:07:13 16:42:07+08:00 Format : application/pdf Title : 608M_FDA.book Creator : 50212800 Producer : Acrobat Distiller 9.0.0 (Windows) Document ID : uuid:a5f1e963-15a4-4776-af08-727e0b0c9219 Instance ID : uuid:7b4d04f4-6b4d-4aed-9b11-a0745afedf22 Page Mode : UseOutlines Page Count : 206EXIF Metadata provided by EXIF.tools