National Instruments Pxi Ni 5401 Users Manual User

NI 5401 to the manual 039b4da2-64a8-4ea3-af6b-28cdc43061e1

2015-02-05

: National-Instruments National-Instruments-Pxi-Ni-5401-Users-Manual-493601 national-instruments-pxi-ni-5401-users-manual-493601 national-instruments pdf

Open the PDF directly: View PDF PDF.
Page Count: 60

DownloadNational-Instruments National-Instruments-Pxi-Ni-5401-Users-Manual- NI 5401 User Manual  National-instruments-pxi-ni-5401-users-manual
Open PDF In BrowserView PDF
Computer-Based
Instruments
NI 5401 User Manual
PXI ™/PCI Arbitrary Function Generator
NI 5401 User Manual

March 1999 Edition
Part Number 322419A-01

Worldwide Technical Support and Product Information
www.natinst.com
National Instruments Corporate Headquarters
11500 North Mopac Expressway

Austin, Texas 78759-3504

USA Tel: 512 794 0100

Worldwide Offices
Australia 03 9879 5166, Austria 0662 45 79 90 0, Belgium 02 757 00 20, Brazil 011 284 5011,
Canada (Ontario) 905 785 0085, Canada (Québec) 514 694 8521, China 0755 3904939, Denmark 45 76 26 00,
Finland 09 725 725 11, France 01 48 14 24 24, Germany 089 741 31 30, Hong Kong 2645 3186,
India 91805275406, Israel 03 6120092, Italy 02 413091, Japan 03 5472 2970, Korea 02 596 7456,
Mexico (D.F.) 5 280 7625, Mexico (Monterrey) 8 357 7695, Netherlands 0348 433466, Norway 32 27 73 00,
Singapore 2265886, Spain (Madrid) 91 640 0085, Spain (Barcelona) 93 582 0251, Sweden 08 587 895 00,
Switzerland 056 200 51 51, Taiwan 02 2377 1200, United Kingdom 01635 523545
For further support information, see the Technical Support Resources appendix. To comment on the
documentation, send e-mail to techpubs@natinst.com.
© Copyright 1999 National Instruments Corporation. All rights reserved.

Important Information
Warranty
The NI 5401 is warranted against defects in materials and workmanship for a period of one year from the date of shipment,
as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that
proves to be defective during the warranty period. This warranty includes parts and labor.
The media on which you receive National Instruments software are warranted not to fail to execute programming
instructions, due to defects in materials and workmanship, for a period of 90 days from date of shipment, as evidenced
by receipts or other documentation. National Instruments will, at its option, repair or replace software media that do not
execute programming instructions if National Instruments receives notice of such defects during the warranty period.
National Instruments does not warrant that the operation of the software shall be uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside
of the package before any equipment will be accepted for warranty work. National Instruments will pay the shipping costs
of returning to the owner parts which are covered by warranty.
National Instruments believes that the information in this document is accurate. The document has been carefully
reviewed for technical accuracy. In the event that technical or typographical errors exist, National Instruments reserves
the right to make changes to subsequent editions of this document without prior notice to holders of this edition. The
reader should consult National Instruments if errors are suspected. In no event shall National Instruments be liable for
any damages arising out of or related to this document or the information contained in it.
EXCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS
ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. CUSTOMER’S RIGHT TO RECOVER DAMAGES CAUSED
BY FAULT OR NEGLIGENCE ON THE PART OF NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE
CUSTOMER. NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR DAMAGES RESULTING FROM LOSS OF DATA, PROFITS, USE OF PRODUCTS,
OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY THEREOF. This limitation of the liability of
National Instruments will apply regardless of the form of action, whether in contract or tort, including negligence.
Any action against National Instruments must be brought within one year after the cause of action accrues. National
Instruments shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty
provided herein does not cover damages, defects, malfunctions, or service failures caused by owner’s failure to follow
the National Instruments installation, operation, or maintenance instructions; owner’s modification of the product;
owner’s abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties,
or other events outside reasonable control.

Copyright
Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical,
including photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without
the prior written consent of National Instruments Corporation.

Trademarks
ComponentWorks™, CVI™, LabVIEW™, natinst.com™, PXI™, RTSI™, and VirtualBench™ are trademarks of
National Instruments Corporation.
Product and company names mentioned herein are trademarks or trade names of their respective companies.

WARNING REGARDING MEDICAL AND CLINICAL USE OF NATIONAL INSTRUMENTS PRODUCTS
National Instruments products are not designed with components and testing for a level of reliability suitable for use in
or in connection with surgical implants or as critical components in any life support systems whose failure to perform
can reasonably be expected to cause significant injury to a human. Applications of National Instruments products
involving medical or clinical treatment can create a potential for death or bodily injury caused by product failure, or by
errors on the part of the user or application designer. Because each end-user system is customized and differs from
National Instruments testing platforms and because a user or application designer may use National Instruments products
in combination with other products in a manner not evaluated or contemplated by National Instruments, the user or
application designer is ultimately responsible for verifying and validating the suitability of National Instruments products
whenever National Instruments products are incorporated in a system or application, including, without limitation,
the appropriate design, process and safety level of such system or application.

Conventions
The following conventions are used in this manual:
»

The » symbol leads you through nested menu items and dialog box options
to a final action. The sequence File»Page Setup»Options directs you to
pull down the File menu, select the Page Setup item, and select Options
from the last dialog box.

♦

The ♦ symbol indicates that the following text applies only to a specific
product, a specific operating system, or a specific software version.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to
avoid injury, data loss, or a system crash.
This icon denotes a warning, which advises you of precautions to take to
avoid being electrically shocked.

bold

Bold text denotes items that you must select or click on in the software,
such as menu items and dialog box options. Bold text also denotes
parameter names.

italic

Italic text denotes variables, emphasis, a cross reference, or an introduction
to a key concept.

monospace

Text in this font denotes text or characters that you should enter from the
keyboard. This font is also used for the proper names of functions,
variables, and filenames and extensions.

monospace italic

Italic text in this font denotes text that is a placeholder for a word or value
that you must supply.

Contents
Chapter 1
Generating Functions
with the NI 5401
About Your NI 5401 ......................................................................................................1-1
Connecting Signals ........................................................................................................1-2
ARB Connector ...............................................................................................1-3
SYNC Connector.............................................................................................1-3
PLL Ref Connector..........................................................................................1-4
Pattern Out Connector (PCI Only) ..................................................................1-5
Connector Pin Assignments ..............................................................1-5
Signal Descriptions ...........................................................................1-6
SHC50-68 50-Pin Cable Connector ................................................................1-6
Software Options for Your NI 5401 ..............................................................................1-8
Software Included with Your NI 5401 ............................................................1-8
VirtualBench .....................................................................................1-8
NI-FGEN Instrument Driver .............................................................1-9
Additional National Instruments Development Tools.....................................1-9
LabVIEW ..........................................................................................1-9
LabWindows/CVI .............................................................................1-10
ComponentWorks .............................................................................1-10
Using the Soft Front Panels to Generate Waveforms ....................................................1-11
Generating Standard Functions .......................................................................1-11
Generating Multiple Frequencies in a Sequence .............................................1-13
Waveform Editor .............................................................................................1-15
Power-Up and Reset Conditions....................................................................................1-16

Chapter 2
Function Generator Operation
Generating Waveforms ..................................................................................................2-2
Direct Digital Synthesis (DDS) .....................................................................................2-3
Frequency Hopping and Sweeping..................................................................2-4
Triggering ......................................................................................................................2-4
Trigger Sources ...............................................................................................2-4
Modes of Operation.........................................................................................2-5
Single Trigger Mode .........................................................................2-5
Continuous Trigger Mode.................................................................2-6
Stepped Trigger Mode ......................................................................2-7
Analog Output................................................................................................................2-7

© National Instruments Corporation

v

NI 5401 User Manual

Contents

SYNC Output and Duty Cycle........................................................................ 2-9
Output Attenuation.......................................................................................... 2-9
Output Impedance ........................................................................................... 2-10
Output Enable ................................................................................................. 2-10
Pre-Attenuation Offset .................................................................................... 2-11
Phase-Locked Loops and Board Synchronization......................................................... 2-11
Analog Filter Correction................................................................................................ 2-13
RTSI/PXI Trigger Lines ................................................................................................ 2-14
Calibration ..................................................................................................................... 2-15

Appendix A
Specifications
Appendix B
Optional Accessories
Appendix C
Frequency Resolution and Lookup Memory
Appendix D
Technical Support Resources
Glossary
Index
Figures
Figure 1-1.
Figure 1-2.
Figure 1-3.
Figure 1-4.
Figure 1-5.
Figure 1-6.
Figure 1-7.
Figure 1-8.
Figure 1-9.
Figure 1-10.
Figure 1-11.

NI 5401 User Manual

NI 5401 I/O Connectors........................................................................ 1-2
Output Levels and Load Termination Using a
50 Ω Output Impedance........................................................................ 1-3
SYNC Output and Duty Cycle.............................................................. 1-4
NI 5401 50-Pin Digital Connector Pin Assignments ............................ 1-5
SHC50-68 68-Pin Connector Pin Assignments .................................... 1-7
VirtualBench-FG Soft Front Panel for Function Generation................ 1-11
VirtualBench-FG General Settings Dialog Box for the NI 5401.......... 1-12
VirtualBench-FG Signals Settings Dialog Box for the NI 5401........... 1-12
VirtualBench-FG Load Waveform Dialog Box.................................... 1-13
VirtualBench-FG Frequency List Editor Dialog Box ........................... 1-14
Waveform Editor Soft Front Panel ....................................................... 1-15

vi

www.natinst.com

Contents

Figure 2-1.
Figure 2-2.
Figure 2-3.
Figure 2-4.
Figure 2-5.
Figure 2-6.
Figure 2-7.
Figure 2-8.
Figure 2-9.
Figure 2-10.
Figure 2-11.
Figure 2-12.
Figure 2-13.
Figure 2-14.
Figure 2-15.

NI 5401 Block Diagram ........................................................................2-1
Waveform Data Path Block Diagram....................................................2-2
DDS Building Blocks............................................................................2-3
Waveform Generation Trigger Sources ................................................2-5
Single Trigger Mode .............................................................................2-6
Continuous Trigger Mode .....................................................................2-6
Stepped Trigger Mode...........................................................................2-7
Analog Output and SYNC Out Block Diagram ....................................2-8
Waveform and Trigger Timings............................................................2-8
Output Attenuation Chain .....................................................................2-9
PLL Architecture for the NI 5401 for PCI ............................................2-12
PLL Architecture for the NI 5401 for PXI ............................................2-12
Analog Filter Correction .......................................................................2-13
RTSI Trigger Lines and Routing for the NI 5401 for PCI ....................2-14
PXI Trigger Lines, 10 MHz Backplane Oscillator, and
Routing for the NI 5401 for PXI ...........................................................2-14

Table 1-1.

Digital Connector Signal Descriptions..................................................1-6

Table

© National Instruments Corporation

vii

NI 5401 User Manual

1

Generating Functions
with the NI 5401

The NI 5401 User Manual describes the features, functions, and operation
of the NI 5401 arbitrary function generator. This device performs
comparably to standalone instruments while providing the flexibility of
computer-based operation.

About Your NI 5401
Thank you for buying a National Instruments NI 5401 arbitrary function
generator. The NI 5401 family consists of two different devices:
•

NI 5401 for PCI

•

NI 5401 for PXI

Your NI 5401 device has the following features:
•

One 12-bit resolution output channel

•

Up to 16 MHz sine and transistor-transistor logic (TTL) waveform
output

•

Software-selectable output impedances of 50 Ω and 75 Ω

•

Output attenuation levels from 0 to 73 dB

•

Phase-locked loop (PLL) synchronization to external clocks

•

Sampling rate of 40 MS/s

•

Digital and analog filter

•

32-bit direct digital synthesis (DDS) for standard function generation

•

External trigger input

•

Real-Time System Integration (RTSI) and PXI triggers

All NI 5401 devices follow industry-standard Plug and Play specifications
on both buses and offer seamless integration with compliant systems.
Detailed specifications for the NI 5401 are in Appendix A, Specifications.

© National Instruments Corporation

1-1

NI 5401 User Manual

Chapter 1

Generating Functions with the NI 5401

Connecting Signals
Figure 1-1 shows the front panels for the NI 5401 for the PXI and PCI
buses. The front panel contains three types of connectors: BNC, SMB, and
50-pin very high-density SCSI (VHDSCSI). The main waveform is
generated through the connector labeled ARB.

LOCK ACCESS

ARB OUT

EXT TRIG

PLL REF

ARB

SYNC
PLL IN

SYNC OUT

PATTERN OUT

PXI

PCI

Figure 1-1. NI 5401 I/O Connectors

NI 5401 User Manual

1-2

www.natinst.com

Chapter 1

Generating Functions with the NI 5401

ARB Connector
The ARB connector provides the waveform output. The maximum output
levels on this connector depend on the type of load termination. If the
output of your NI 5401 terminates into a 50 Ω load, the output levels are
±5 V, as shown in Figure 1-2. If the output of your NI 5401 terminates into
a high-impedance load (HiZ), the output levels are ±10 V. If the output
terminates into any other load, the levels are as follows:
RL
- × 10 V
V out = ± -----------------RL + RO
where Vout is the maximum output voltage level
RL is the load impedance in ohms, and
RO is the output impedance on the NI 5401.
By default, RO = 50 Ω, but you can use your software to set it to 75 Ω.
NI 5401

NI 5401
Load

ARB
RO =
50 Ω

RL =
50 Ω

50 Ω Load

ARB

±5 V

Load
RO =
50 Ω

RL =
HiZ

±10 V

High-Impedance Load

Figure 1-2. Output Levels and Load Termination Using a 50 Ω Output Impedance
Note Software sets the voltage output levels based on a 50 Ω load termination.

For more information on waveform generation and analog output
operation, refer to Chapter 2, Function Generator Operation. For
specifications on the waveform output signal, see Appendix A,
Specifications.

SYNC Connector
The SYNC connector provides a TTL version of the sine waveform being
generated at the output. You can think of the SYNC output as a very
high-frequency resolution, software-programmable clock source for many
applications. You can also vary the duty cycle of the SYNC output on the
fly by software control, as shown in Figure 1-3. tp is the time period of the
© National Instruments Corporation

1-3

NI 5401 User Manual

Chapter 1

Generating Functions with the NI 5401

sine wave being generated and tw is the pulse width of the SYNC output.
The duty cycle is (tw/tp) × 100%.
tp
ARB Output

SYNC Output
(50% Duty Cycle)

tw

SYNC Output
(33% Duty Cycle)

Figure 1-3. SYNC Output and Duty Cycle

For your NI 5401 for PCI, you can route the SYNC output to the RTSI lines
over the RTSI bus. For your NI 5401 for PXI, you can route the SYNC
output to the TTL trigger lines over the TTL trigger bus. The SYNC output
is derived from a comparator connected to the analog waveform and
provides a meaningful waveform only when you are generating a sine wave
on the ARB output. For more information on SYNC output, see Chapter 2,
Function Generator Operation.

PLL Ref Connector
The PLL Ref connector is a phase-locked loop (PLL) input connector that
can accept a reference clock from an external source and frequency lock the
NI 5401 internal clock to this external clock. The reference clock should
not deviate more than ±100 ppm from its nominal frequency. The minimum
amplitude levels of 1 Vpk-pk are required on this clock. You can lock
reference clock frequencies of 1 MHz and 5–20 MHz in 1 MHz steps.
Note You can frequency lock the NI 5401 for PCI to other National Instruments devices

over the RTSI bus using the 20 MHz RTSI clock signal. You can frequency lock the
NI 5401 for PXI to other National Instruments devices using the 10 MHz backplane clock.
If no external reference clock is available, the NI 5401 automatically tunes
the internal clock to the highest accuracy possible. For more information on
PLL operation, refer to Chapter 2, Function Generator Operation.

NI 5401 User Manual

1-4

www.natinst.com

Chapter 1

Generating Functions with the NI 5401

Pattern Out Connector (PCI Only)
This connector is used on the NI 5401 for PCI to supply the external trigger
input to the board.

Connector Pin Assignments
Figure 1-4 shows the NI 5401 50-pin digital connector. Refer to Table 1-1
for a description of the signals.

DGND
NC
DGND
NC
DGND
NC
DGND
RFU
DGND
RFU
DGND
RFU
DGND
RFU
DGND
RFU
DGND
RFU
DGND
RFU
DGND
RFU
DGND
RFU
DGND

50
49
48
47
46
45
44
43
42
41
40
39
38
37
36
35
34
33
32
31
30
29
28
27
26

25
24
23
22
21
20
19
18
17
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1

EXT_TRIG
NC
NC
NC
NC
NC
NC
RFU
RFU
RFU
RFU
RFU
RFU
RFU
RFU
RFU
RFU
RFU
RFU
RFU
RFU
RFU
RFU
RFU
RFU

Figure 1-4. NI 5401 50-Pin Digital Connector Pin Assignments

© National Instruments Corporation

1-5

NI 5401 User Manual

Chapter 1

Generating Functions with the NI 5401

Signal Descriptions
Table 1-1 shows the pin names and signal descriptions used on the NI 5401
digital output connector.
Table 1-1. Digital Connector Signal Descriptions

Signal Name

Type

Description

DGND

–

Digital ground

EXT_TRIG

Input

External trigger—The external trigger input signal is a
TTL-level signal that you can use to start or step through a
waveform generation. For more information on trigger sources
and trigger mode, see Chapter 2, Function Generator
Operation.

NC

–

Not connected.

RFU

–

Reserved for future use. Do not connect signals to this pin.

SHC50-68 50-Pin Cable Connector
You can use an optional SHC50-68 50-pin to 68-pin cable for external
trigger input. The cable connects to the digital connector on the NI 5401.
Figure 1-5 shows the 68-pin connector pin assignments on the SHC50-68
cable.
Note The SHC50-68 connector uses the same signals as the NI 5401 digital output

connector shown in Table 1-1.

NI 5401 User Manual

1-6

www.natinst.com

Chapter 1

Generating Functions with the NI 5401

RFU

1

35

DGND

RFU

2

36

DGND

RFU

3

37

DGND

RFU

4

38

DGND

RFU

5

39

DGND

RFU

6

40

DGND

RFU

7

41

DGND

RFU

8

42

DGND

RFU

9

43

DGND

RFU

10

44

DGND

RFU

11

45

DGND

RFU

12

46

DGND

RFU

13

47

DGND

RFU

14

48

DGND

RFU

15

49

DGND

RFU

16

50

DGND

RFU

17

51

DGND

RFU

18

52

DGND

RFU

19

53

DGND

RFU

20

54

DGND

RFU

21

55

DGND

RFU

22

56

DGND

RFU

23

57

DGND

RFU

24

58

RFU

NC

25

59

DGND

NC

26

60

DGND

NC

27

61

DGND

NC

28

62

DGND

NC

29

63

DGND

NC

30

64

DGND

NC

31

65

DGND

NC

32

66

DGND

NC

33

67

DGND

EXT_TRIG

34

68

DGND

Figure 1-5. SHC50-68 68-Pin Connector Pin Assignments

© National Instruments Corporation

1-7

NI 5401 User Manual

Chapter 1

Generating Functions with the NI 5401

Software Options for Your NI 5401
This section describes the NI-FGEN driver software and development tools
that you can use to create application software for your NI 5401.

Software Included with Your NI 5401
Your NI 5401 kit includes several VirtualBench soft front panels to help
you get up and running quickly with your waveform generator. These soft
front panels are an onscreen interface similar to standalone instruments. An
NI-FGEN instrument driver is also included, which you can use with a
wide variety of development tools to build applications for your NI 5401.
These software tools are discussed in the following sections.

VirtualBench
Similar to standalone instruments, VirtualBench acquires, controls,
analyzes, and presents data. However, since VirtualBench operates on your
PC, it provides additional processing, storage, and display capabilities.
VirtualBench loads and saves waveform data in a form that popular
spreadsheet programs and word processors can use. It can also generate
reports—a complement to the raw data storage—by adding timestamps,
measurements, user names, and comments. You can print the waveforms
and the settings of VirtualBench to a printer connected to the PC.
VirtualBench has two components—VirtualBench-FG and Waveform
Editor—that you can use with your NI 5401. These components are
described in the following sections.

VirtualBench-FG
VirtualBench-FG transforms your PC into a fully featured function
generator that rivals desktop models by using the DDS capabilities of your
NI 5401. VirtualBench-FG emulates benchtop function generators, so you
can quickly learn to use computer-based instruments.
With VirtualBench-FG, you can generate a variety of waveforms, including
five standard waveforms: sine, square, triangle, rising exponential, and
falling exponential. Using VirtualBench-FG, you load waveforms from an
ASCII text file and generate them repeatedly. You can generate these
waveforms with a resolution of approximately 10 mHz and perform
frequency sweeps and shift-keying. As with all VirtualBench instruments,
you can load and save instrument settings.

NI 5401 User Manual

1-8

www.natinst.com

Chapter 1

Generating Functions with the NI 5401

Waveform Editor
You use the Waveform Editor to create, sketch, and edit complex
waveforms that the VirtualBench-FG player can then generate. A library of
standard waveforms for creating complex waveforms is included, and you
can also write equations to create arbitrary waveforms and view the
waveforms in a time or frequency domain.

NI-FGEN Instrument Driver
To create your application, you need an industry-standard software driver
such as NI-FGEN to control your instrument. The NI-FGEN driver
includes a set of standard functions for configuring, creating, starting, and
stopping waveform generation. The instrument driver reduces your
program development time and simplifies instrument control by
eliminating the need to learn a complex programming protocol for your
instrument.
NI-FGEN is in a standard instrument driver format that works with
LabVIEW, LabWindows/CVI, and conventional programming languages
such as C, C++, and Visual Basic.
Refer to the NI-FGEN readme.txt file for more details on the NI-FGEN
instrument driver. This file can be launched from the
Start»Programs»National Instruments FGEN menu.
Note An NI-FGEN Instrument Driver Quick Reference Guide is included in your NI 5401
kit. This reference guide helps you program your NI 5401.

Additional National Instruments Development Tools
The following sections describe several additional tools that you can use to
develop complex applications for your NI 5401. The NI-FGEN instrument
driver exposes the Application Programming Interfaces (APIs) to these
development environments.

LabVIEW
LabVIEW is a graphical programming language for building
instrumentation systems. With LabVIEW, you quickly create front panel
user interfaces, giving you interactive control of your software system. To
specify the functionality, you assemble block diagrams—a natural design
notation for engineers and scientists. LabVIEW has all of the same
development tools and language capabilities of a standard language such

© National Instruments Corporation

1-9

NI 5401 User Manual

Chapter 1

Generating Functions with the NI 5401

as C, including looping and case structures, configuration management
tools, and compiled performance.
Note Use the NI-FGEN instrument driver to program and control your NI 5401 using

LabVIEW.

LabWindows/CVI
LabWindows/CVI is an interactive, ANSI C programming environment
designed for automated test applications.
LabWindows/CVI has an interactive drag-and-drop editor for building your
user interface and a complete ANSI C development environment for
building your test program logic. The LabWindows/CVI environment has
a wide collection of automatic code-generation tools and utilities that
accelerate your development process, without sacrificing any of the power
and flexibility of a language such as C. In addition, the LabWindows/CVI
run-time libraries are compatible with standard C/C++ compilers,
including Visual C++ and Borland C++ under Windows.
Note Use the NI-FGEN instrument driver to program and control your NI 5401 using

LabWindows/CVI.

ComponentWorks
ComponentWorks is a collection of 32-bit ActiveX controls for building
virtual instrumentation systems. ComponentWorks gives you the power
and flexibility of standard development tools, such as Microsoft Visual
Basic or Visual C++, with the instrumentation expertise of National
Instruments. Based on ActiveX technology, ComponentWorks controls are
easy to configure using property sheets and are easy to control from your
programs using high-level properties and methods. ComponentWorks
features instrumentation-based graphical user interface (GUI) tools,
including graphs, meters, gauges, knobs, dials, and switches.
Note Use the NI-FGEN instrument driver to program and control your NI 5401 using

ComponentWorks.

NI 5401 User Manual

1-10

www.natinst.com

Chapter 1

Generating Functions with the NI 5401

Using the Soft Front Panels to Generate Waveforms
You use the VirtualBench soft front panels to interactively control your
NI 5401 as you would a desktop function generator.

Generating Standard Functions
If you need to generate standard waveforms such as a sine, square, ramp,
or DC signal, you can use the VirtualBench-FG soft front panel shown in
Figure 1-6. Launch the front panel by selecting Start»Programs»National
Instruments FGEN»VirtualBench FG. You use this front panel to
control the frequency, amplitude, offset, and type of waveform generated.
The maximum sine frequency you can generate is 16 MHz. The maximum
amplitude is 5 Vpk into a 50 Ω load. If the load is a high-impedance load,
the actual levels will be twice that shown on the front panel.

Figure 1-6. VirtualBench-FG Soft Front Panel for Function Generation

© National Instruments Corporation

1-11

NI 5401 User Manual

Chapter 1

Generating Functions with the NI 5401

To control additional instrument parameters, select Edit»54xx Settings to
bring up the dialog box shown in Figures 1-7 and 1-8.

Figure 1-7. VirtualBench-FG General Settings Dialog Box for the NI 5401

Figure 1-8. VirtualBench-FG Signals Settings Dialog Box for the NI 5401
Note Refer to the online help for further information about the 54xx Settings dialog box.

NI 5401 User Manual

1-12

www.natinst.com

Chapter 1

Generating Functions with the NI 5401

You can also load a custom waveform pattern with VirtualBench-FG. This
waveform should be a text file and should contain exactly 16,384 samples.
If the defined waveform does not contain exactly 16,384 samples, you may
see undesired effects in your waveform output. Follow these steps to load a
custom waveform:
1.

Select File»Load Waveform to bring up the dialog box shown in
Figure 1-9.

Figure 1-9. VirtualBench-FG Load Waveform Dialog Box

2.

Specify the delimiter used in the text file, the number of columns, the
start line, and the number of samples.

3.

Click OK to return to the main VirtualBench-FG screen shown in
Figure 1-6.

4.

Click the User button to use the information in the text file as the
source for the waveform.

5.

Click the On button to generate the waveform.

Generating Multiple Frequencies in a Sequence
If desired, you can generate multiple frequencies in a sequence, which can
include frequency sweeping, hopping, and so on. You can list up to 512
different frequencies and specify the duration of generation for each of

© National Instruments Corporation

1-13

NI 5401 User Manual

Chapter 1

Generating Functions with the NI 5401

them using the VirtualBench-FG Frequency List Editor. Follow these steps
to create a list of multiple frequencies:
1.

Select Window»Frequency List Editor from the VirtualBench-FG
soft front panel to bring up the dialog box shown in Figure 1-10.

Figure 1-10. VirtualBench-FG Frequency List Editor Dialog Box

2.

Specify the frequency and duration of each function in the sequence.

3.

Save the sequence by selecting File»Save.

4.

To return to the main VirtualBench-FG screen shown in Figure 1-6,
select File»Close.

5.

Select File»Load Frequency List to load the frequency list.

You can combine the frequency list generation with different trigger modes
to get the desired frequency generation.

NI 5401 User Manual

1-14

www.natinst.com

Chapter 1

Generating Functions with the NI 5401

Waveform Editor
You can use the Waveform Editor shown in Figure 1-11 to create a custom
waveform. To launch the Waveform Editor, select Start»Programs»
National Instruments FGEN»Waveform Editor. You can select
waveforms from the function library, write equations, or draw them
manually. Each segment can have more than one waveform component in
it, and you can perform a variety of math functions on each component.

Figure 1-11. Waveform Editor Soft Front Panel

This soft front panel is resizable so you can view the waveforms you create
with as much precision as you wish. You can save the waveforms in the
following formats:
•
•
•

Voltage (.wfm)
Text (.txt)
Binary (.bin)

Text waveforms are the only format you can use with the NI 5401, and they
must contain exactly 16,384 samples to function properly.

© National Instruments Corporation

1-15

NI 5401 User Manual

Chapter 1

Generating Functions with the NI 5401

Power-Up and Reset Conditions
When you power up your computer, the NI 5401 is in the following state:
•

The output is disabled and set to 0 V.

•

The trigger mode is set to continuous.

•

The trigger source is set to automatic (the software provides the
triggers).

•

The digital filter is enabled.

•

Output attenuation remains unchanged from its previous setting.

•

The analog filter remains unchanged from its previous setting.

•

Output impedance remains unchanged from its previous setting.

When you reset the board using NI-FGEN or any other application
software, your NI 5401 is in the same state as shown at power up,
previously listed, with the following differences:

NI 5401 User Manual

•

Output attenuation is set to 0 dB.

•

The analog filter is enabled.

•

Output impedance is set to 50 Ω.

•

The PLL reference source is set to internal tuning.

•

The SYNC duty cycle is set to 50%.

1-16

www.natinst.com

2

Function Generator Operation
This chapter describes how to use your NI 5401.
Figure 2-1 shows the NI 5401 block diagram.

RTSI/PXI Trigger Bus

DDS +
Lookup
Memory

Instruction
FIFOs

Digital
Filter

IFIFO
Control

RTSI
Control

Clock
Controls

Waveform
Sequencer

DDS
Control

Filter
Controls

Trigger
Control

Analog
Control

DAC

Attenuators,
Filter, and
Amplifier

ARB

SYNC
Level
Crossing
Detector

Data Path

PLL Ref
PLL and
Clocking

Bus
Interface

PXI/PCI Channel

Figure 2-1. NI 5401 Block Diagram

© National Instruments Corporation

2-1

NI 5401 User Manual

Chapter 2

Function Generator Operation

The NI 5401 has several main components:
•

A PXI or PCI bus interface that handles Plug and Play protocols for
assigning resources to the device and providing drivers for the data and
address bus that are local to the device

•

A waveform sequencer that performs multiple functions such as
arbitrating the data buses and controlling the triggers, filters,
attenuators, clocks, PLL, RTSI switch, instruction FIFO, and DDS

•

The data from the memory is fed to a digital-to-analog converter
(DAC) through a half-band interpolating digital filter. The output from
the DAC goes through the filter to the amplifiers, attenuators, and,
finally, the I/O connector.

Generating Waveforms
The NI 5401 generates waveforms using DDS, which is used for generating
standard waveforms that are repetitive in nature, such as sine, TTL, square,
and triangular waveforms. DDS mode limits you to one buffer, and the
buffer size must be exactly equal to 16,384 samples.

A
Filter
MUX
B

12

DAC Register

12 Bits

Digital Filter

Figure 2-2 shows a block diagram of the data path for waveform
generation. The data for waveform generation comes from DDS lookup
memory. This data is interpolated by a half-band digital filter and then fed
to a high-speed DAC. The data has a pipeline delay of 26 update clocks
through this digital filter. Although the digital filter can be disabled through
software, there will still be a 26 update clock delay.

12

DAC

DDS Lookup
Memory
Digital Filter
Enable
DDS
Div/2

80 MHz Oscillator

Figure 2-2. Waveform Data Path Block Diagram

On the NI 5401, the high-speed DAC is always updated at 80 MHz, but the
update clock for memory is 40 MHz.

NI 5401 User Manual

2-2

www.natinst.com

Chapter 2

Function Generator Operation

Direct Digital Synthesis (DDS)
Direct digital synthesis (DDS) is a technique for deriving, under digital
control, an analog frequency source from a single reference clock
frequency. This technique produces high-frequency accuracy and
resolution, temperature stability, wideband tuning, and rapid and
phase-continuous frequency switching.
The NI 5401 uses a 32-bit, high-speed accumulator with a lookup memory
and a 12-bit DAC for DDS-based waveform generation. Figure 2-3 shows
the building blocks for DDS-based waveform generation.

Frequency

Time

Lookup Data Out (16)
Memory

(14)

DDS

Sequencer

Frequency
Time

16-Bit
Counter

Instruction FIFO

Div/2

80 MHz Oscillator

Figure 2-3. DDS Building Blocks

The lookup memory is dedicated to the DDS. You can store one cycle of a
repetitive waveform—a sine, triangular, square, or arbitrary wave—in the
lookup memory. Then, you can change the frequency of that waveform by
sending just one instruction. You can use DDS mode for very fine
frequency resolution function generation. You can generate sine waves of
up to 16 MHz with the NI 5401. Waveform generation always loops back
to the beginning of the lookup memory after passing through the end of the
lookup memory.
The NI 5401 uses a lookup waveform memory for storing the waveform
buffer and FIFO memory for storing the staging list, which contains
multiple frequency list information. This FIFO is referred to as an
instruction FIFO.

© National Instruments Corporation

2-3

NI 5401 User Manual

Chapter 2

Function Generator Operation

Each stage is made up of two instructions: the frequency, which specifies
the frequency of the waveform to be generated, and the time, which
specifies the time for which the frequency is to be generated.
Note You cannot specify the number of iterations for a waveform to be generated.

Frequency Hopping and Sweeping
You can define a staging list for performing frequency hops and sweeps.
The entire staging list uses the same buffer loaded into the lookup memory.
All stages differ in the frequency to be generated.
Note The minimum time that a frequency should be generated is 2 µs. Therefore, the

maximum hop rate from frequency to frequency is 500 kHz.
The maximum number of stages that can be stored in the instruction FIFO
for DDS mode is 512. For more information on the waveform generation
process, refer to your software documentation.

Triggering
You use triggering to start and step through a waveform generation. The
trigger sources and modes of operation are explained in the following
sections.

Trigger Sources
Trigger sources are software selectable. By default, the software produces
the trigger sources. You can also use an external trigger from a pin on the
digital I/O connector, the RTSI trigger lines on the RTSI bus, or the TTL
trigger lines on the PXI trigger bus on the backplane. Figure 2-4 shows the
trigger sources for the NI 5401.

NI 5401 User Manual

2-4

www.natinst.com

Function Generator Operation

RTSI Switch

Chapter 2

RTSI/PXI Trigger
Lines <0..6>
7

RTSI/PXI Trigger
Digital
MUX

External Trigger

Start Trigger

Software Trigger

Trigger Select

Figure 2-4. Waveform Generation Trigger Sources

If you need to automatically trigger the waveform generation, use software
to generate the triggers. A rising TTL edge is required for external
triggering. For more information on triggering over RTSI lines, see the
RTSI/PXI Trigger Lines section later in this chapter.

Modes of Operation
The NI 5401 has three triggering modes—single, continuous, and
stepped—described in the following sections.

Single Trigger Mode
The waveform you define in the staging list is generated only once by going
through the entire staging list. Only one trigger is required to start the
waveform generation.
In single trigger mode, after the NI 5401 receives a trigger, the waveform
generation starts at the first stage and continues through the last stage. The
last stage is generated repeatedly until you stop the waveform generation.
Figure 2-5 illustrates a single trigger mode of operation.

© National Instruments Corporation

2-5

NI 5401 User Manual

Chapter 2

Function Generator Operation

End of All Stages

Start Trigger

f1, ∆T1

f2, ∆T2

f3, ∆T3

Last Stage Generated
f4 Continuously Until Stopped

Figure 2-5. Single Trigger Mode

For example, assume that one cycle of a sine wave is stored in the DDS
lookup memory. For stage 1, f1 specifies the sine frequency to be generated
for time ∆T1, f2 and ∆T2 for stage 2, and so on. If there are four stages in
the staging list, f4 will be generated continuously until the waveform
generation is stopped.

Continuous Trigger Mode
The waveform you define in the staging list is generated infinitely by
continually cycling through the staging list. After a trigger is received, the
waveform generation starts at the first stage, continues through the last
stage, and loops back to the start of the first stage, continuing until you stop
the waveform generation. Only one trigger is required to start the waveform
generation.
Figure 2-6 illustrates a continuous trigger mode of operation.

Start Trigger

(f1, ∆T1)

End of All Stages

(f2, ∆T2)

(f1, ∆T1)
(f3, ∆T3)
(f4, ∆T4)

Repeat
Until Stopped

(f2, ∆T2)

Figure 2-6. Continuous Trigger Mode

NI 5401 User Manual

2-6

www.natinst.com

Chapter 2

Function Generator Operation

Stepped Trigger Mode
After a start trigger is received, the waveform defined by the first stage is
generated. Then, the device waits for the next trigger signal. On the next
trigger, the waveform described by the second stage is generated, and so on.
Once the staging list is exhausted, the waveform generation returns to the
first stage and continues in a cyclic fashion.
Figure 2-7 illustrates a stepped trigger mode of operation. Switching from
stage to stage is phase continuous. In this mode, the time instruction is not
used. The trigger paces the waveform generation from one frequency to the
other.
Start Trigger

Start Trigger

Start Trigger Start Trigger

f1

f2

f3

Start Trigger

f4

f1

End of All Stages

Figure 2-7. Stepped Trigger Mode

Analog Output
Analog waveforms are generated as follows:
1.

The 12-bit digital waveform data is fed to a high-speed DAC.

2.

A lowpass filter filters the DAC output.

3.

This filtered signal is amplified before it goes to a 10 dB attenuator.

Note The DAC output can be fine-tuned for gain and offset. Since the offset is adjusted
before the main attenuators and amplifier, it is referred to as pre-attenuation offset. This
fine-tuning of gain and offset is performed by separate DACs.

4.

The output from the 10 dB attenuator then goes to the main amplifier,
which can provide up to ±5 V levels into 50 Ω. An output relay can
switch between ground level and the main amplifier. Refer to the
Output Enable section of this document for additional information
about this relay.

5.

The output of this relay goes to a series of passive attenuators.

6.

The output of the attenuators goes through a selectable output
impedance of 50 or 75 Ω to the I/O connector.

© National Instruments Corporation

2-7

NI 5401 User Manual

Chapter 2

Function Generator Operation

Figure 2-8 shows the essential block diagram of analog waveform
generation.
Attenuators
(63 dB in 1 dB steps)

10 dB
Attenuator

Lowpass
Filter

Output
Enable
12

Arb

25 Ω

50 Ω
DAC
Pre Amp

Main Amp

Clock

50 Ω/75 Ω
Selector
Gain
DAC
Offset
DAC

SYNC

Comparator
50 Ω

+
-

Level
DAC

Figure 2-8. Analog Output and SYNC Out Block Diagram

Figure 2-9 shows the timing relationships of the trigger input and
waveform output. Td1 is the pulse width on the trigger signal. Td2 is the time
delay from trigger to output on Arb output. Refer to Appendix A,
Specifications, for more information on these timing parameters.

Td1
Trigger Input Signal
(Slope: Positive, TTL)
Td2

Waveform Output
(±5 Vpp into 50 Ω)

Figure 2-9. Waveform and Trigger Timings
Note You can switch off the analog lowpass filter at any time during waveform generation.

When you change this setting, the bouncing of electromechanical relays on the NI 5401
distorts the output signal for about 10 ms.

NI 5401 User Manual

2-8

www.natinst.com

Chapter 2

Function Generator Operation

SYNC Output and Duty Cycle
The SYNC output is a TTL version of the sine waveform generated at the
output. The signal from the pre-amplifier is sent to a comparator, where it
is compared against a level set by the level DAC. The output of this
comparator is sent to the SYNC connector through a hysteresis buffer and
a 50 Ω series resistor to reverse terminate reflected pulses.
You can use the SYNC output as a very high-frequency resolution,
software-programmable clock source for many applications. You also can
vary the duty cycle of SYNC output, on the fly, by changing the output of
the level DAC. The SYNC output might not carry meaning for other types
of generated waveforms.
Note You can change the duty cycle of the SYNC output at any time during waveform

generation.

Output Attenuation
Figure 2-10 shows the NI 5401 output attenuator chain. The output
attenuators are made of resistor networks and may be switched in any
combination. The maximum attenuation possible on the NI 5401 is 73 dB.

1 dB

2 dB

8 dB

4 dB

16 dB

32 dB

Out

In

Figure 2-10. Output Attenuation Chain

By attenuating the output signal, you keep the dynamic range of the DAC;
that is, you do not lose any bits from the digital representation of the signal
because the attenuation is done after the DAC and not before it.
attenuation (in decibels) = –20 log10 (Vo /Vi)
where Vo = desired voltage level for the output signal
Vi = input voltage level.
Note For the NI 5401, Vi = ±5 V for a terminated load and ±10 V for an

unterminated load.

© National Instruments Corporation

2-9

NI 5401 User Manual

Chapter 2

Function Generator Operation

NI-FGEN calculates the value of the output attenuation chain, which you
can control by changing the peak-to-peak amplitude parameter. 0 dB
attenuation corresponds to an amplitude of 10 Vpk-pk. The maximum
attenuation of 73 dB corresponds to an amplitude of 2.24 mV pk-pk. Any
amplitude less then this is coerced to this value.
Note You can change the output attenuation at any time during waveform generation.

When you change this setting, the bouncing of electromechanical relays on the NI 5401
distorts the output signal for about 10 ms.

Output Impedance
As shown in Figure 2-10, before the signal reaches the output connector,
you can select an output impedance of 50 Ω or 75 Ω. If the load impedance
is 50 Ω and all the attenuators are off (an output attenuation of 0 dB), the
output levels are ±5 V.
Most applications use a load impedance of 50 Ω, but applications such as
video device testers require 75 Ω. If the load is a very high-input impedance
load (~1 MΩ), you will see output levels up to ±10 V.
Note You can change the output impedance at any time during waveform generation.
When you change this setting, the bouncing of electromechanical relays on the NI 5401
distorts the output signal for about 10 ms.

Output Enable
You can switch off the waveform generation at the output connector by
controlling the output enable relay, as shown in Figure 2-8. When the
output enable relay is off, the output signal level goes to ground level.
Note Even though the output enable relay is in the off position, the waveform generation

process continues internally on the NI 5401.
You can use this feature to disconnect and connect different devices to the
NI 5401 on the fly.
Note You can change the output enable state at any time during waveform generation.

When you change this setting, the bouncing of electromechanical relays on the NI 5401
distorts the output signal for about 10 ms.

NI 5401 User Manual

2-10

www.natinst.com

Chapter 2

Function Generator Operation

Pre-Attenuation Offset
The NI 5401 hardware supports a DC offset of up to ±2.5 V before the
attenuation chain. Unless the 10 dB attenuator is switched in, which occurs
when the amplitude is less then 3.16 Vpk-pk, the waveform maximum plus
the offset must not exceed ±5 V into 50 Ω. If it does, the waveform is
clipped. Refer to Figure 2-8 for a diagram showing the location of the
10 dB attenuator.
NI-FGEN automatically calculates the pre-attenuation offset value based
on the DC offset and amplitude values, so the allowable DC offset range is
dependent on the amplitude. For example, if you have an amplitude of
1 Vpk-pk, the maximum DC offset you can apply is 0.25 V, which
corresponds to a pre-attenuation offset of 2.5 V.
Note You can change the DC Offset at any time during waveform generation. Refer to

your software documentation for additional information.

Phase-Locked Loops and Board Synchronization
Figure 2-11 illustrates the block diagram for the NI 5401 for PCI PLL
circuit. Figure 2-12 illustrates the block diagram for the NI 5401 for PXI
PLL circuit. The PLL consists of a voltage-controlled crystal oscillator
(VCXO) with a tuning range of ±100 ppm. This VCXO generates the main
clock of 80 MHz.
The PLL can lock to a reference clock source from the external connector,
from a RTSI Osc line on the RTSI bus (NI 5401 for PCI), or from a
10 MHz Osc line on the PXI backplane bus (for NI 5401 for PXI). The PLL
can also be tuned internally using a calibration DAC (CalDAC). National
Instruments accurately performs this tuning during manufacturing. Refer to
the RTSI/PXI Trigger Lines section later in this manual for additional
information on using the RTSI and 10 MHz Osc lines.
The reference and VCXO clock are compared by a phase comparator
running at 1 MHz. The loop filters the error signal and sends it to the control
pin of the VCXO to complete the loop.

© National Instruments Corporation

2-11

NI 5401 User Manual

Chapter 2

Function Generator Operation

RTSI
Switch

Master/Slave

RTSI Bus

Board Clock (Master)
RTSI Osc

RTSI Clock (Slave)
(20 MHz)

Control
Voltage

AMUX

Source

80 MHz

Loop
Filter
Tune
DAC

Phase
Comp
14

20 MHz

Div/4

PLL Ref
(1 Vpk-pk min)
Board Clock

VCXO

10 MHz Osc

PXI Bus

Figure 2-11. PLL Architecture for the NI 5401 for PCI

Control
Voltage

AMUX

Source

80 MHz

Loop
Filter
CAL
DAC

Div/8

Phase
Comp

PLL Ref
(1 Vpk-pk min)

10 MHz

VCXO

Figure 2-12. PLL Architecture for the NI 5401 for PXI

You can frequency lock to an external reference clock source of 1 MHz and
from 5–20 MHz in 1 MHz increments. The PLL can lock to a signal level
of at least 1 Vpk-pk.
Do not increase the voltage level of the clock signal at the PLL reference input
connector by more than the specified limit, 5 V pk-pk.

Caution

Note If two or more NI 5401 devices are locked to each other using the same reference

clock, they are frequency locked, but the phase relationship is indeterminate.

NI 5401 User Manual

2-12

www.natinst.com

Chapter 2

Function Generator Operation

Analog Filter Correction
The NI 5401 can correct for slight deviations in the flatness of the
frequency characteristic of the analog lowpass filter in its passband, as
shown in Figure 2-13. Curve A shows a typical lowpass filter curve. The
response of the filter is stored in an onboard EEPROM in 1 MHz
increments up to 16 MHz. Curve C is the correction applied to the
frequency response. The resulting Curve B is a flat response over the entire
passband. If you want to generate a sine wave at a particular frequency with
filter correction applied, you have to specify that frequency through your
software.

A
B

Gain (dB)

C

Frequency (MHz)
A. Typical Analog Filter Characteristics
B. Corrected Filter Characteristics
C. Correction Applied

Figure 2-13. Analog Filter Correction
Note You can change the filter frequency correction at any time during waveform
generation.

© National Instruments Corporation

2-13

NI 5401 User Manual

Chapter 2

Function Generator Operation

RTSI/PXI Trigger Lines
The NI 5401 for PCI contains seven trigger lines and one RTSI clock line
available over the RTSI bus to send and receive NI 5401-specific
information to other boards that have RTSI connectors. Figure 2-14 shows
the RTSI trigger lines and routing of NI 5401 for PCI signals to the RTSI
switch.

SYNC

RTSI 0

Start Trigger

RTSI 1

RTSI Trigger

RTSI
Switch

RTSI 3
RTSI 4
RTSI 5

Board Clock

RTSI Bus

RTSI 2

RTSI 6
Master/Slave

RTSI Osc

RTSI Clock

Figure 2-14. RTSI Trigger Lines and Routing for the NI 5401 for PCI

Figure 2-15 shows the PXI trigger lines and routing of NI 5401 for PXI
signals to the RTSI switch.

SYNC

TRIG 0

Start Trigger

TRIG 1

RTSI Trigger
BOARD_SYNC

RTSI
Switch

TRIG 3
TRIG 4
TRIG 5

PXI Bus

TRIG 2

PXI STAR

PXI 10MHz Osc

Figure 2-15. PXI Trigger Lines, 10 MHz Backplane Oscillator, and Routing for the
NI 5401 for PXI

NI 5401 User Manual

2-14

www.natinst.com

Chapter 2

Function Generator Operation

The NI 5401 can receive a hardware trigger from another board as a RTSI
trigger signal on any of the RTSI/PXI trigger lines.
You can also route signals as follows:

♦

•

Route the Start Trigger signal generated on the NI 5401 to other boards
through any of the RTSI/PXI bus trigger lines.

•

Route the SYNC output generated on the NI 5401 to other boards
through any of the RTSI/PXI bus trigger lines.You can use this signal
to give other boards an accurate and fine frequency resolution clock.

NI 5401 for PCI
For frequency locking to other boards as a master, the NI 5401 sends an
onboard 20 MHz signal to the RTSI Osc line as a board clock signal. For
locking to other devices as a slave, the NI 5401 receives the RTSI Osc line
as a RTSI clock signal.

♦

NI 5401 for PXI
For frequency locking to other boards, the NI 5401 for PXI receives the
PXI backplane 10 MHz Osc as a reference clock signal. All the NI 5401s
for PXI use this common signal as the reference clock for frequency
locking.

Note Refer to your software documentation for selecting and routing signals to the

RTSI/PXI trigger bus.

Calibration
Calibration is the process of minimizing measurement errors by making
small circuit adjustments. On the NI 5401, NI-FGEN automatically makes
these adjustments by retrieving predetermined constants from the onboard
EEPROM, calculating correction values, and writing those values to the
CalDACs.
National Instruments calibrates all NI 5401 devices to the levels indicated
in Appendix A, Specifications. Factory calibration involves procedures
such as nulling the offset and gain errors. However, since offset and gain
errors may drift with time and temperature, you may need to recalibrate
your device. Contact National Instruments to recalibrate your NI 5401.

© National Instruments Corporation

2-15

NI 5401 User Manual

A

Specifications

This appendix lists the specifications for the NI 5401. These specifications
are typical at 25 °C unless otherwise stated. The operating temperature
range is 0 to 50 °C.

Analog Output
Number of channels ............................... 1
Resolution .............................................. 12 bits
Maximum update rate ............................ 40 MHz
DDS accumulator ................................... 32 bits
Frequency range
Sine ................................................. 16 MHz, max
SYNC (TTL)................................... 16 MHz, max
Square ............................................. 1 MHz, max
Ramp ............................................... 1 MHz, max
Triangle ........................................... 1 MHz, max
Frequency resolution.............................. 9.31 mHz

Voltage Output
Ranges ..................................................... ±5 V into a 50 Ω load

±10 V into a high-impedance load

Accuracy ................................................ ±0.1 dB
Output attenuation.................................. 0 to 73 dB
Resolution ....................................... 0.001 dB steps

© National Instruments Corporation

A-1

NI 5401 User Manual

Appendix A

Specifications

Pre-attenuation offset
Range...............................................±2.5 V into 50 Ω1
Accuracy..........................................±5 mV
Output coupling .....................................DC
Output impedance ..................................50 Ω or 75 Ω software selectable
Load impedance .....................................50 Ω or greater
Output enable..........................................Software switchable
Protection................................................Short-circuit protected

Sine Spectral Purity
Harmonic products and spurs
Up to 1 MHz....................................–60 dBc
Up to 16 MHz..................................–35 dBc
Phase noise .............................................–105 dBc/Hz at 10 kHz from
carrier

Filter Characteristics
Digital
Type.................................................Half-band interpolating
Selection ..........................................Software switchable (enable or
disable)
Taps ................................................67
Filter coefficients ............................Fixed 20-bit
Data interpolating frequency ..........80 MS/s
Pipeline signal delay .......................26 sampling periods
Analog
Type.................................................7th-order L-C lowpass filter
Passband ripple................................±2 dB

1

With less than 10 dB of attenuation, signal maximum plus offset (before attenuation) must not exceed ±5 V (into 50 Ω)

NI 5401 User Manual

A-2

www.natinst.com

Appendix A

Specifications

Waveform Specifications
Memory .................................................. 16,384 16-bit samples
Segment length....................................... 16,384 samples, exact
Segment linking (instruction FIFO) ....... 512 links

Timing I/O
Update clock .......................................... Internal, 40 MHz only
Frequency locking
External reference sources .............. Input connector, RTSI clock line,
or internal
Reference clock frequencies ........... 1 MHz, 5–20 MHz in 1 MHz steps
Frequency locking range................. ±100 ppm

Triggers
Digital Trigger
Compatibility ........................................ TTL
Response ............................................... Rising edge
Pulse width (Td1) .................................... 20 ns min
Trigger to waveform
output delay (Td2)................................... 28 sample clocks + 150 ns max

RTSI
Trigger lines .......................................... 7
Clock lines.............................................. 1

Bus Interface
Type ....................................................... Slave

Operational Modes
Type ....................................................... Single, continuous, stepped

© National Instruments Corporation

A-3

NI 5401 User Manual

Appendix A

Specifications

SYNC Out
Level ......................................................TTL
Duty cycle...............................................20% to 80%, software
controllable

External Clock Reference Input
Frequency ...............................................1 MHz or 5–20 MHz in 1 MHz
steps
Amplitude ...............................................1 Vpk-pk ≤ level ≤ 5 Vpk-pk

Internal Clock
Frequency ...............................................40 MHz
Initial accuracy .......................................±5 ppm
Temperature stability (0 to 5 °C)............±25 ppm
Aging (1 year).........................................±5 ppm

Mechanical
Connectors
ARB (output) ...................................SMB/BNC
SYNC (output).................................SMB/BNC
PLL reference (input) ......................SMB
External trigger in............................50-pin digital (PCI),
SMB (PXI)
Size .........................................................1 slot
Power requirements ...............................5 V, 3.5 A max
12 V, 125 mA

NI 5401 User Manual

A-4

www.natinst.com

Optional Accessories

B

National Instruments offers a variety of products to use with your NI 5401,
including probes, cables, and other accessories:
•

Shielded and unshielded I/O connector blocks (SCB-68, TBX-68,
CB-68)

•

RTSI bus cables

For more specific information about these products, refer to your National
Instruments catalogue or Web site, or call the office nearest you.

Cabling
The following list gives recommended part numbers for cables that you can
use with your NI 5401 device:
•

BNC male to BNC male, 50 Ω cable from ITT Pomona Electronics
(part number BNC-C-xx)

•

BNC male to BNC male, 75 Ω cable from ITT Pomona Electronics
(part number 2249-E-xx)

•

BNC female to RCA phono plug adapter, from ITT Pomona
Electronics (part number 5319)

•

BNC 50 Ω feed-through terminator adapter from ITT Pomona
Electronics (part number 4119-50)

•

BNC female-female adapter from ITT Pomona Electronics
(part number 3283)

© National Instruments Corporation

B-1

NI 5401 User Manual

C

Frequency Resolution and
Lookup Memory

For DDS-based waveform generation, you must first load one cycle of the
desired waveform into the lookup memory. The size of the DDS lookup
memory is 16,384 samples. Each sample is 16 bits wide.
Note One cycle of the waveform buffer loaded into the memory should be exactly equal

to the size of the DDS lookup memory.
Fc = update clock for the accumulator.
For the NI 5401, Fc = 40 MHz.
Fa = desired frequency of the output signal
N = accumulator size in bits
For the NI 5401, N = 32.
FCW = frequency control word to be loaded into the accumulator to
generate Fa.
The frequency control word is calculated using the formula:
FCW = (2N * Fa) / Fc
The frequency resolution is then given by:
frequency resolution = Fc / 2N = (40 × 106) / 232 = 9.31322 mHz
For example, if you need to generate a frequency of 10 MHz, then the FCW
is (232 * 10E6)/40E6, which equals 1,073,741,824. If you need to generate
a frequency of 1 Hz, then the FCW is (232 * 1)/40E6, which equals 107.
Note On the NI 5401, the maximum frequency of a sine wave you can generate reliably

is limited to 16 MHz. Other waveforms such as square or triangular waves are limited to
1 MHz.

© National Instruments Corporation

C-1

NI 5401 User Manual

Appendix C

Frequency Resolution and Lookup Memory

You can also synthesize arbitrary waveforms using DDS. Generating
arbitrary waveforms this way will be very limited; you are restricted to a
single buffer, and this buffer should be exactly equal to the size of the
lookup memory (16,384 samples).
To update every sample of an arbitrary waveform in lookup memory at the
maximum clock rate of 40 MHz, the software writes an FCW value of
2(N–L), where N is the size of the accumulator and L is the number of
address bits of lookup memory (L = 14 bits). Thus, the FCW value for the
NI 5401 equals 262,144. Since FCW = (2N * Fa) / Fc, Fa = (2(N–L) * Fc) / 2N,
so you would write a frequency value of (2(32–14) × (40 × 106)) / 232, which
equals 2.441 kHz
If you want to update every sample in lookup memory at an integral
subdivision, D, of the maximum clock rate, then you want an FCW value
of 2(N–L–D+1). In other words, for an effective update rate of every sample at
half the maximum clock rate, write a frequency value of
(2(32–14–2+1) × (40 × 106)) / 232, which equals 1.221 kHz.

NI 5401 User Manual

C-2

www.natinst.com

Technical Support Resources

D

This appendix describes the comprehensive resources available to you in
the Technical Support section of the National Instruments Web site and
provides technical support telephone numbers for you to use if you have
trouble connecting to our Web site or if you do not have internet access.

NI Web Support
To provide you with immediate answers and solutions 24 hours a day,
365 days a year, National Instruments maintains extensive online technical
support resources. They are available to you at no cost, are updated daily,
and can be found in the Technical Support section of our Web site at
www.natinst.com/support.

On-Line Problem-Solving and Diagnostic Resources
•

KnowledgeBase—A searchable database containing thousands of
frequently asked questions (FAQs) and their corresponding answers or
solutions, including special sections devoted to our newest products.
The database is updated daily in response to new customer experiences
and feedback.

•

Troubleshooting Wizards—Step-by-step guides lead you through
common problems and answer questions about our entire product line.
Wizards include screen shots that illustrate the steps being described
and provide detailed information ranging from simple getting started
instructions to advanced topics.

•

Product Manuals—A comprehensive, searchable library of the latest
editions of National Instruments hardware and software product
manuals.

•

Hardware Reference Database—A searchable database containing
brief hardware descriptions, mechanical drawings, and helpful images
of jumper settings and connector pinouts.

•

Application Notes—A library with more than 100 short papers
addressing specific topics such as creating and calling DLLs,
developing your own instrument driver software, and porting
applications between platforms and operating systems.

© National Instruments Corporation

D-1

NI 5401 User Manual

Appendix D

Technical Support Resources

Software-Related Resources
•

Instrument Driver Network—A library with hundreds of instrument
drivers for control of standalone instruments via GPIB, VXI, or serial
interfaces. You also can submit a request for a particular instrument
driver if it does not already appear in the library.

•

Example Programs Database—A database with numerous,
non-shipping example programs for National Instruments
programming environments. You can use them to complement the
example programs that are already included with National Instruments
products.

•

Software Library—A library with updates and patches to application
software, links to the latest versions of driver software for National
Instruments hardware products, and utility routines.

Worldwide Support
National Instruments has offices located around the globe. Many branch
offices maintain a Web site to provide information on local services. You
can access these Web sites from www.natinst.com/worldwide.
If you have trouble connecting to our Web site, please contact your local
National Instruments office or the source from which you purchased your
National Instruments product(s) to obtain support.
For telephone support in the United States, dial 512 795 8248. For
telephone support outside the United States, contact your local branch
office:
Australia 03 9879 5166, Austria 0662 45 79 90 0, Belgium 02 757 00 20,
Brazil 011 284 5011, Canada (Ontario) 905 785 0085,
Canada (Québec) 514 694 8521, China 0755 3904939,
Denmark 45 76 26 00, Finland 09 725 725 11, France 01 48 14 24 24,
Germany 089 741 31 30, Hong Kong 2645 3186, India 91805275406,
Israel 03 6120092, Italy 02 413091, Japan 03 5472 2970,
Korea 02 596 7456, Mexico (D.F.) 5 280 7625,
Mexico (Monterrey) 8 357 7695, Netherlands 0348 433466,
Norway 32 27 73 00, Singapore 2265886, Spain (Madrid) 91 640 0085,
Spain (Barcelona) 93 582 0251, Sweden 08 587 895 00,
Switzerland 056 200 51 51, Taiwan 02 2377 1200,
United Kingdom 01635 523545

NI 5401 User Manual

D-2

www.natinst.com

Glossary
Prefix

Meaning

Value

µ-

micro-

10 – 6

m-

milli-

10 –3

k-

kilo-

10 3

M-

mega-

10 6

Numbers/Symbols
%

percent

+

positive of, or plus

–

negative of, or minus

±

plus or minus

/

per

°

degree

Ω

ohm

+5 V

+5 V output signal

A
A

amperes

amplification

method of scaling the signal level to a higher level

ARB

normal waveform output signal

attenuation

decreasing the amplitude of a signal

© National Instruments Corporation

G-1

NI 5401 User Manual

Glossary

B
b

bit—one binary digit, either 0 or 1

B

byte—eight related bits of data, an eight-bit binary number. Also used to
denote the amount of memory required to store one byte of data.

BNC

a type of coaxial signal connector

buffer

temporary storage for acquired or generated data

bus

the group of conductors that interconnect individual circuitry in a computer.
Typically, a bus is the expansion vehicle to which I/O or other devices are
connected. Examples of PC buses are the AT bus (also known as the ISA
bus) and the PCI bus.

C
C

Celsius

CalDAC

calibration DAC

clock

hardware component that controls timing for reading from or writing to
groups

CMOS

complementary metal-oxide semiconductor

continuous trigger mode

repeats a staging list until waveform generation is stopped

counter

a circuit that counts external pulses or clock pulses (timing)

coupling

the manner in which a signal is connected from one location to another

D
DAC

digital-to-analog converter—an electronic device, often an integrated
circuit, that converts a digital number into a corresponding analog voltage
or current

dB

decibel—the unit for expressing a logarithmic measure of the ratio of two
signal levels: dB = 20 log10 V1/V2, for signals in volts

dBc

decibel referred to carrier level

NI 5401 User Manual

G-2

www.natinst.com

Glossary

DC

direct current

DC coupled

allowing the transmission of both AC and DC signals

DDS

direct digital synthesis—a digital technique of frequency generation using
a numerically controlled oscillator (NCO), a dedicated lookup memory,
and a DAC

DDS mode

a method of waveform generation that uses built-in DDS functionality to
generate very high frequency resolution standard waveforms

DGND

digital ground signal

digital word

See word.

driver

software that controls a specific hardware device

dynamic range

the ratio of the largest signal level a circuit can handle to the smallest signal
level it can handle (usually taken to be the noise level), normally expressed
in dB

E
EEPROM

electrically erasable programmable read-only memory—ROM that can be
erased with an electrical signal and reprogrammed

external trigger

a voltage pulse from an external source that triggers an event such as A/D
conversion

EXT_TRIG

external trigger input signal

© National Instruments Corporation

G-3

NI 5401 User Manual

Glossary

F
FIFO

first-in first-out memory buffer—the first data stored is the first data sent to
the acceptor. FIFOs are often used on DAQ devices to temporarily store
incoming or outgoing data until that data can be retrieved or output. For
example, an analog input FIFO stores the results of A/D conversions until
the data can be retrieved into system memory, a process that requires the
servicing of interrupts and often the programming of the DMA controller.
This process can take several milliseconds in some cases. During this time,
data accumulates in the FIFO for future retrieval. With a larger FIFO,
longer latencies can be tolerated. In the case of analog output, a FIFO
permits faster update rates, because the waveform data can be stored on the
FIFO ahead of time. This again reduces the effect of latencies associated
with getting the data from system memory to the DAQ device.

filters

digital or analog circuits that change the frequency characteristics of a
waveform

frequency hop

change from one frequency to another

frequency resolution

the smallest frequency change that can be generated by a NI 5411/5431

frequency sweep

change the frequency of a waveform in a controlled manner

G
gain

the factor by which a signal is amplified, sometimes expressed in decibels

GUI

graphical user interface

H
hardware

the physical components of a computer system, such as the circuit boards,
plug-in boards, chassis, enclosures, peripherals, cables, and so on

HiZ

high impedance

Hz

hertz—the number of cycles or repetitions per second

NI 5401 User Manual

G-4

www.natinst.com

Glossary

I
I/O

input/output—the transfer of data to/from a computer system involving
communications channels, operator interface devices, and/or data
acquisition and control interfaces

IFIFO

instruction FIFO

instruction FIFO

the FIFO that stores the waveform generation staging list

K
k

kilo—the standard metric prefix for 1,000, or 103, used with units of
measure such as volts, hertz, and meters

K

kilo—the prefix for 1,024, or 210, used with B in quantifying data or
computer memory

kS

1,000 samples

Kword

1,024 words of memory

L
latch

a digital device that stores digital data based on a control signal

level DAC

the calibration DAC used to change the voltage levels to another device

linking

linking different buffers stored in the waveform memory

looping

repeating the same buffer in the waveform memory. This method of
waveform generation decreases memory requirements.

lowpass filter

a circuit used to smooth the waveform output and removed unwanted high
frequency contents form the signal

© National Instruments Corporation

G-5

NI 5401 User Manual

Glossary

M
m

meters

M

(1) Mega, the standard metric prefix for 1 million or 106, when used with
units of measure such as volts and hertz; (2) mega, the prefix for 1,048,576,
or 220, when used with B to quantify data or computer memory

master/slave

locking the NI 5401 clock in frequency to an external phase locking
reference clock source

MB

megabytes of memory

N
noise

an undesirable electrical signal—Noise comes from external sources such
as the AC power line, motors, generators, transformers, fluorescent lights,
soldering irons, CRT displays, computers, electrical storms, welders, radio
transmitters, and internal sources such as semiconductors, resistors, and
capacitors. Noise corrupts signals you are trying to send or receive.

O
output enable relay

a relay switch at the output of the NI 5401 that can enable the waveform
generation at any time or that can connect the output to ground

P
passband

the range of frequencies which a device can properly propagate or measure

PCI

Peripheral Component Interconnect—a high-performance expansion bus
architecture originally developed by Intel to replace ISA and EISA. It is
achieving widespread acceptance as a standard for PCs and work-stations;
it offers a theoretical maximum transfer rate of 132 Mbytes/s.

PCLK

digital pattern clock output

peak-to-peak

a measure of signal amplitude; the difference between the highest and
lowest excursions of the signal

NI 5401 User Manual

G-6

www.natinst.com

Glossary

pipeline

a high-performance processor structure in which the completion of an
instruction is broken into its elements so that several elements can be
processed simultaneously from different instructions

PLL

phase-locked loop—a circuit that synthesizes a signal whose frequency is
exactly proportional to the frequency of a reference signal

PLL Ref

a PLL input that accepts an external reference clock signal and phase locks
to it the NI 5401 internal clock

Plug and Play devices

devices that do not require dip switches or jumpers to configure resources
on the devices—also called switchless devices

ppm

parts per million

pre-attenuation offset

an offset provided to the signal before it reaches the attenuators

protocol

the exact sequence of bits, characters, and control codes used to transfer
data between computers and peripherals through a communications
channel, such as the GPIB bus

PXI

PCI eXtensions for Instrumentation

R
resolution

the smallest signal increment that can be detected by a measurement
system. Resolution can be expressed in bits, in proportions, or in percent of
full scale. For example, a system has 12-bit resolution, one part in 4,096
resolution, and 0.0244 percent of full scale.

RTSI bus

Real-Time System Integration bus—the National Instruments timing bus
that connects DAQ boards directly, by means of connectors on top of the
boards, for precise synchronization of functions

S
s

seconds

S

samples

sampling rate

the rate, in samples per second (S/s), at which each sample in the waveform
buffer is updated

© National Instruments Corporation

G-7

NI 5401 User Manual

Glossary

SCSI

Small Computer System Interface (bus)

sequence list

See staging list.

shift-keying

frequency shift keying (FSK)

single trigger mode

when the arbitrary waveform generator goes through the staging list only
once

SMB

Sub Miniature Type B connector that features a snap coupling for fast
connection

S/s

samples per second—used to express the rate at which a DAQ board
samples an analog signal

stage

in Arb mode, specifies the buffer to be generated, the number of loops on
that buffer, the marker position for that buffer, and the sample count for the
buffer; for DDS mode, specifies the frequency to be generated of the
waveform in the lookup memory and the time for which that frequency has
to be generated

staging list

a buffer that contains linking and looping information for multiple
waveforms; also known as a sequence list or waveform sequence

stepped trigger mode

a mode of waveform generation used when you want a trigger to advance
the waveforms specified by the stages in the staging list

SYNC

TTL version of the sine waveform output signal generated by the NI 5401

T
trigger

any event that causes or starts some form of data capture

TTL

transistor-transistor logic

U
update rate

NI 5401 User Manual

the rate at which a DAC is updated

G-8

www.natinst.com

Glossary

V
V

volts

VCXO

voltage controlled crystal oscillator

VHDSCSI

very high-density SCSI

W
waveform

multiple voltage readings taken at a specific sampling rate

waveform buffer

the collection of 16-bit data samples stored in the waveform memory that
represent a desired waveform. Also known as a waveform segment.

waveform linking and
looping

See linking, looping.

waveform memory

physical data storage on the NI 5401 for storing the waveform data samples

waveform segment

See waveform buffer.

waveform sequence

See staging list.

waveform staging

See linking, looping.

word

The standard number of bits that a processor or memory manipulates at one
time. Microprocessors typically use 8-, 16-, or 32-bit words.

© National Instruments Corporation

G-9

NI 5401 User Manual

Index
A

digital trigger specifications, A-3
direct digital synthesis (DDS)
building blocks for DDS (figure), 2-3
description, 2-3 to 2-4
frequency hopping and sweeping, 2-4
frequency resolution and lookup memory,
C-1 to C-2
duty cycle of SYNC output, 2-9

analog filter correction, 2-13
analog output, 2-7 to 2-11
analog output and SYNC out block
diagram, 2-8
output attenuation, 2-9 to 2-11
output enable, 2-10
output impedance, 2-10
pre-attenuation offset, 2-11
specifications, A-1
SYNC output and duty cycle, 2-9
waveform and trigger timings (figure), 2-8
ARB connector, 1-3
attenuation of output, 2-9 to 2-11

E
external clock reference input, A-4
EXT_TRIG signal (table), 1-6

F
filters
analog filter correction, 2-13
characteristics, A-2
frequency
direct digital synthesis, 2-4
frequency hops and sweeps, 2-4
frequency resolution and lookup memory,
C-1 to C-2
generating multiple frequencies in
sequence, 1-13 to 1-14
function generator operation
analog filter correction, 2-13
analog output, 2-7 to 2-11
output attenuation, 2-9 to 2-11
output enable, 2-10
output impedance, 2-10
pre-attenuation offset, 2-11
SYNC output and duty cycle, 2-9
block diagram, 2-1
calibration, 2-15
direct digital synthesis (DDS), 2-3 to 2-4
frequency hopping and sweeping, 2-4

B
block diagram for NI 5401, 2-1
bus interface specifications, A-3

C
cables, optional, B-1
calibration, 2-15
clocks
external clock reference input, A-4
internal clock, A-4
ComponentWorks software, 1-10
connectors. See I/O connectors.
continuous trigger mode, 2-6
conventions used in manual, iv

D
DDS. See direct digital synthesis (DDS).
DGND signal (table), 1-6
diagnostic resources, online, D-1

© National Instruments Corporation

I-1

NI 5401 User Manual

Index

M

phase-locked loops and board
synchronization, 2-11 to 2-12
RTSI/PXI trigger lines, 2-14 to 2-15
triggering, 2-4 to 2-7
continuous trigger mode, 2-6
single trigger mode, 2-5 to 2-6
stepped trigger mode, 2-7
trigger sources, 2-4 to 2-5
waveform generation, 2-2

mechanical specifications, A-4
multiple frequencies, generating in sequence,
1-13 to 1-14

N
National Instruments Web support, D-1 to D-2
NI 5401. See also function generator
operation.
block diagram, 2-1
components, 2-2
connecting signals, 1-2 to 1-7
ARB connector, 1-3
I/O connectors (figure), 1-2
Pattern Out connector, 1-5 to 1-6
PLL Ref connector, 1-4
SHC50-68 50-pin cable connector,
1-6 to 1-7
SYNC connector, 1-3 to 1-4
features, 1-1
PLL architecture (figures), 2-12
power-up and reset conditions, 1-16
software options, 1-8 to 1-10
NI-FGEN instrument driver, 1-9

I
impedance, output, 2-10
instruction FIFO, 2-3
internal clock specifications, A-4
I/O connectors, 1-2 to 1-7
ARB connector, 1-3
I/O connectors on front panel (figure), 1-2
Pattern Out connector, 1-5 to 1-6
pin assignments (figure), 1-5
signal descriptions (table), 1-6
PLL Ref connector, 1-4
SHC50-68 50-pin cable connector,
1-6 to 1-7
pin assignments (figure), 1-7
signals, 1-6
SYNC connector, 1-3 to 1-4
I/O connectors (figure), 1-2

O
online problem-solving and diagnostic
resources, D-1
operational mode specifications, A-3
optional accessories, B-1
output attenuation, 2-9 to 2-11

L
LabVIEW and LabWindows/CVI software,
1-9 to 1-10
lookup memory
description, 2-3
frequency resolution and lookup memory,
C-1 to C-2
lookup waveform memory, 2-3

NI 5401 User Manual

P
Pattern Out connector, 1-5 to 1-6
pin assignments (figure), 1-5
signal descriptions (table), 1-6
phase-locked loop (PLL) Ref connector, 1-4

I-2

www.natinst.com

Index

operational modes, A-3
sine spectral purity, A-2
SYNC out, A-4
timing I/O, A-3
triggers, A-3
voltage output, A-1 to A-2
waveform, A-3
staging list, 2-3
stepped trigger mode, 2-7
SYNC connector, 1-3 to 1-4
SYNC output
description, 2-9
duty cycle, 2-9
specifications, A-4

phase-locked loops and board
synchronization, 2-11 to 2-12
pin assignments (figure)
Pattern Out connector, 1-5
SHC50-68 50-pin cable connector, 1-7
PLL Ref connector, 1-4
power-up and reset conditions, 1-16
pre-attenuation offset, 2-11
problem-solving and diagnostic resources,
online, D-1

R
reset conditions, 1-16
RTSI trigger specifications, A-3
RTSI/PXI trigger lines, 2-14 to 2-15

T
technical support resources, D-1 to D-2
time, in direct digital synthesis, 2-4
timing I/O specifications, A-3
trigger specifications, A-3
digital trigger, A-3
RTSI, A-3
triggering, 2-4 to 2-7
continuous trigger mode, 2-6
single trigger mode, 2-5 to 2-6
stepped trigger mode, 2-7
trigger sources, 2-4 to 2-5

S
SHC50-68 50-pin cable connector, 1-6 to 1-7
pin assignments (figure), 1-7
signals, 1-6
sine spectral purity, A-2
single trigger mode, 2-5 to 2-6
soft front panels. See VirtualBench software.
software options, 1-8 to 1-10
ComponentWorks, 1-10
LabVIEW, 1-9 to 1-10
LabWindows/CVI, 1-10
NI-FGEN instrument driver, 1-9
VirtualBench, 1-8
VirtualBench-FG, 1-8
Waveform Editor, 1-9
software-related resources, D-2
specifications, A-1 to A-4
analog output, A-1
bus interface, A-3
external clock reference input, A-4
filter characteristics, A-2
internal clock, A-4
mechanical, A-4

© National Instruments Corporation

V
VirtualBench software
description, 1-8
generating multiple frequencies in
sequence, 1-13 to 1-14
generating standard functions, 1-11 to
1-13
VirtualBench-FG software
description, 1-8
Frequency List Editor, 1-14
General Settings dialog box (figure), 1-12

I-3

NI 5401 User Manual

Index

Load Waveform dialog box (figure), 1-13
Signal Settings dialog box (figure), 1-12
voltage output specifications, A-1 to A-2

W
Waveform Editor
creating custom waveforms, 1-15
description, 1-9
waveform generation. See also function
generator operation.
analog waveforms, 2-7 to 2-11
data path block diagram (figure), 2-2
overview, 2-2
using soft front panels
generating multiple frequencies in
sequence, 1-13 to 1-14
generating standard functions,
1-11 to 1-13
using Waveform Editor, 1-15
waveform specifications, A-3
Web support from National Instruments,
D-1 to D-2
online problem-solving and diagnostic
resources, D-1
software-related resources, D-2
Worldwide technical support, D-2

NI 5401 User Manual

I-4

www.natinst.com



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.3
Linearized                      : Yes
Encryption                      : Standard V1.2 (40-bit)
User Access                     : Print, Copy, Annotate, Fill forms, Extract, Assemble, Print high-res
Create Date                     : 1999:05:24 14:13:10
Producer                        : Acrobat Distiller 3.01 for Windows
Author                          : Document Production
Title                           : NI 5401 User Manual
Creator                         : FrameMaker 5.5.6p145
Modify Date                     : 1999:06:29 18:06:59-05:00
Subject                         : 322419A-01; March 1999
Page Count                      : 60
Page Mode                       : UseOutlines
Page Layout                     : SinglePage
EXIF Metadata provided by EXIF.tools

Navigation menu