Nihon Kohden ZM-941PA Medical Telemetry Transmitter User Manual 2
Nihon Kohden Corporation Medical Telemetry Transmitter Users Manual 2
Contents
- 1. Users Manual 1
- 2. Users Manual 2
Users Manual 2
When Using Disposable Cuffs or YP-960T series Reusable Cuffs To use these NIBP cuffs, an optional YN-990P extension hose (1.5 m) is required. NOTE Connect the joints properly. If there is an air leak, NIBP cannot be measured properly. 1. Connect the NIBP cuff to the extension hose. 2. Connect the other end of the extension hose to the NIBP socket on the transmitter. Turn the joint clockwise until it clicks. To disconnect the cuff from the transmitter, turn the hose joint counterclockwise. Attaching the NIBP Cuff to the Patient WARNING Be careful when measuring NIBP on a patient with known bleeding disorders or coagulation. After NIBP measurement, there may be dot hemorrhage, or circulatory disorder by thrombus where the cuff is attached. CAUTION Do not wrap the cuff on an arm or thigh which is used for injection. NIBP measurement on an arm or thigh which is used for injection may cause reflux of blood and stop injection. CAUTION Do not wrap the cuff too tight. It may cause poor blood circulation and congestion. If the cuff is wrapped too loosely, the NIBP value may increase. Operator’s Manual ZM-940PA/941PA CAUTION Do not attach the cuff to the site where there is injury or inflammation. If the skin gets irritated or redness appears on the skin from the cuff, change the attachment site or stop using the cuff. Take extreme care on the patients with delicate skin. CAUTION Do not reuse disposable parts and accessories. 41 CAUTION NIBP and SpO2 can be measured on the same limb, but the SpO2 monitoring might not be accurate during NIBP measurement. Be careful when reading the SpO2 values.* * Monitoring SpO2 during NIBP Measurement When the SpO2 probe is attached to the same limb as the NIBP cuff, the blood flow decreases during NIBP measurement and pulse wave cannot be detected and SpO2 cannot be monitored properly. When “INHIBIT SpO2 DURING NIBP” on the PARAMETER SETUP screen is set to ON (factory default setting), SpO2 monitoring is paused during NIBP measurement to avoid SpO2 alarm occurrence. However, when monitoring SpO2 on the same limb as the NIBP, be careful when reading SpO2 values. NOTE • Measuring NIBP at a site other than the upper arm gives different values from those measured at the upper arm. When making diagnosis based on the NIBP values, measure NIBP on an upper arm. • To accurately detect the pulsatile flow of the artery, the cuff should be wrapped around a bare upper arm. • Do not use an abnormal cuff. The cuff deteriorates from use and cleaning. Before use, check the cuff and confirm that there is no flaw, crack or hole in it. Be careful not to damage the inflation bag. If the inflation bag has a hole or a flaw, it may burst during use. Dispose of an abnormal cuff and replace it with a new one. • Refer to the NIBP cuff manual for details. Cuff Position When placing the transmitter on a bed, make sure that the hose is not bent. Heart 42 Place the cuffed upper arm (brachium) at the same height as the patient’s heart. If the cuff is not at the same level as the heart, the weight of the blood affects the blood pressure reading. The pressure difference per unit height is 0.7 mmHg/cm. The blood pressure reading decreases when the arm is higher than the heart and increases when lower. The best measuring condition is when the patient is lying on his/her back with arms and legs relaxed. If the cuff position cannot be on the same level as the heart, the displayed blood pressure reading must be mathematically adjusted. Operator’s Manual ZM-940PA/941PA Attaching the Transmitter on an Arm (Using the YP-943P/944P NIBP Cuff) 1. Attach the NIBP cuff to the transmitter. Refer to the “Connecting the NIBP Cuff to the Transmitter” section. 2. Insert the end of the cuff into the belt and then through the D ring as shown at left. Belt D ring End of cuff 3. Fold back the cuff at the D ring and fasten it using the velcro tape. Make sure that the cuff is not attached on a joint. NOTE The cuff must not wrap around the elbow. Operator’s Manual ZM-940PA/941PA 43 Attaching the Strap to the Transmitter NOTE • Use the strap to prevent the transmitter from falling. • Do not attach the clip to hard objects such as thick cloth or zipper. It will break the clip. Attach a strap provided with the transmitter to the NIBP cuff and patient clothes. To open the clip, firmly pull out the tab in direction of the arrow. Belt for the strap on the NIBP cuff To adjust the strap length, push down the tab on the adjuster and slide. 1. Adjust the length of the strap. 2. Clip one end of the strap to the belt for the strap on the NIBP cuff. 3. Clip the other end of the strap to the patient’s clothes as shown left. 44 Operator’s Manual ZM-940PA/941PA Placing the Transmitter on the Bed (Using the Disposable Cuffs or YP-960T series Reusable Cuffs) ARTERY ▼ 1. Put the cuff on the upper arm so that the ▼ mark of “ARTERY ▼” aligns with the artery of the patient. 2. Wrap the cuff so that “INDEX within the “ RANGE ”. RANGE ” comes If “Index ” is not within the “ ”, change the cuff size. RANGE Attaching Electrodes Selecting Electrode Lead CAUTION Only use Nihon Kohden specified electrodes and electrode leads. When other type of electrodes or electrode leads are used, the “CHECK ELECTRODES” message may be displayed and monitoring may stop. The following electrode leads can be used on the transmitter (option). BR-903PA, 3 electrodes, clip type BR-913PA, 3 electrodes, snap type Operator’s Manual ZM-940PA/941PA BR-906PA, 6 electrodes, clip type BR-916PA, 6 electrodes, snap type 45 Connecting the Electrode Lead to the Transmitter Connect the electrode lead to the ECG/RESP socket on the transmitter. When the transmitter is attached on an arm CAUTION CAUTION Do not shake or swing the transmitter while holding the leads or cables connected to the transmitter. The transmitter may come off and injure someone or damage surrounding instruments. Hold the connector of the electrode lead when connecting/disconnecting the electrode lead. If you disconnect the electrode lead by pulling the lead, it damages the electrode lead. Selecting the Electrode Position Follow the physician’s instructions for electrode placement when available. For ECG monitoring, electrodes are attached only on the chest to allow patient movement and obtain continuous stable ECG. Following leads are examples. When also monitoring respiration, refer to the “Electrode Position for Respiration Monitoring” section. NOTE The optimum electrode positions for ECG measurement of a patient are not always optimum for respiration measurement of the patient. Select positions suitable for both ECG and respiration measurements, or positions which have priority for one measurement. 46 Operator’s Manual ZM-940PA/941PA Six Electrodes The 6-electrode method with lead II and lead V5 is effective for monitoring myocardial ischemia. You can improve monitoring accuracy considerably by adding lead V4 to this combination. Va and Vb can be at any position of the standard 12 leads V1 to V6, but V4 and V5 are most appropriate for myocardial ischemic monitoring. LA/L RA/R Va/Ca Vb/Cb RL/RF LL/F Electrode Position Symbol AHA IEC Lead Color AHA IEC Left infraclavicular fossa LA Black Yellow Right infraclavicular fossa RA White Red Below lowest rib on the left anterior axillary line LL Red Green Right anterior axillary line at the same level as LL/F RL RF Green Black Fifth intercostal space on the left midclavicular line. (V4 position of standard 12 leads) Left anterior axillary line at the same level as Va. (V5 position of standard 12 leads) Operator’s Manual ZM-940PA/941PA Va Ca Brown-blue Whitebrown Vb Cb Brownorange White-black 47 Lead Position Standard limb leads Lead I RA Lead II LA Lead III RA LL N (RL) Monopolar limb leads aVR lead N (RL) aVL lead aVF lead RA LA RA LA N (RL) N (RL) LA LL LL LL LA LL LL N (RL) RA RA LA N (RL) Monopolar chest leads V1 to V6 leads to RA LA LL N (RL) 48 Operator’s Manual ZM-940PA/941PA Three Electrodes • Lead MII, which is similar to standard lead II, used when ECG measurement has priority Electrode Position Left infraclavicular fossa Right infraclavicular fossa Below lowest rib on the left anterior axillary line • Lead MI, which is similar to standard lead I Change F/LL and L/LA of the lead MII. Symbol AHA IEC LA RA LL Lead Color AHA IEC Black Yellow White Red Red Green • Lead MIII, which is similar to standard lead III. Change R/RA and L/LA of the lead MII. If the electrode position shown above is not available due to chest surgery, attach the electrodes to the root of the limbs or below the clavicles for stable ECG monitoring. Operator’s Manual ZM-940PA/941PA 49 Attaching Electrodes to the Patient and Connecting the Electrode Leads to Disposable Electrodes Prepare the Patient Skin Shave off excessive body hair. To reduce skin impedance, clean the electrode site with cream or with a cotton pad moistened with alcohol. Thoroughly dry the skin with a clean cotton pad. NOTE • For a patient with frequent body movement, rub the sites with Skinpure skin preparation gel. However, do not use Skinpure skin preparation gel on sensitive skin. • Do not place electrodes on a wound or on an inflamed, wrinkled or uneven skin surface. Attaching Electrodes to the Patient CAUTION Do not reuse disposable parts and accessories. NOTE • To maintain good contact between the electrode and skin, check that the paste of the disposable electrode is not dry. • When contact between the disposable electrode and skin becomes poor, replace electrodes with new ones immediately. Otherwise, contact impedance between the skin and the electrode increases and the correct ECG cannot be obtained. Refer to the electrode operator’s manual for details. 1. Carefully remove the backing paper from the electrode. Avoid touching the adhesive surface. 2. Place the electrode on the previously cleaned skin. Pay attention to the electrode lead color and symbol. 3. Clip the electrode lead to the electrode. 4. Fasten the electrode lead wire with surgical tape with an extra length of wire between the tape and the electrode. This lessens the movement of electrode leads by body movement and helps stable monitoring. 50 Operator’s Manual ZM-940PA/941PA Electrode Position for Respiration Monitoring Place the R/RA and F/LL electrodes so that the lungs are between the electrodes. NOTE The optimum electrode positions for ECG measurement of a patient are not always optimum for respiration measurement of the patient. Select positions suitable for both ECG and respiration measurements, or positions which have priority for one measurement. Electrode Position Examples NOTE The following examples are when monitoring with 3 electrodes. ECG cannot be monitored correctly when electrodes are attached as the following examples when monitoring with 6 electrodes. Position 1 In this position, respiration measurement is available; however, there is a difference in amplitude between different patients. R or RA F or LL Fifth intercostal space on the Right infraclavicular fossa left midclavicular line, V4 Position 2 In this position, the waveform amplitude is usually large and the ECG lead is similar to Lead MII. This position can be generally recommended. R or RA F or LL Fifth intercostal space on the Right infraclavicular fossa left midaxillary line, V6 Operator’s Manual ZM-940PA/941PA 51 Position 3 In this position, the respiration waveform is optimum, but the ECG lead is unusual. R or RA Right midaxillary at the horizontal level of V4 F or LL Fifth intercostal space on the left midaxillary line, V6 Position 4 In this position, the respiration measurement is influenced by the impedance variation of the abdomen, so the cardiac pulse wave included in the respiration wave is reduced. Note that the waveform is inverted in phase compared with the chest movement (the waveform goes down during inspiration). It is difficult to measure the ECG at the same time. R or RA Lowest rib on the right� anterior axillary line 52 F or LL Lowest rib on the left anterior axillary line Operator’s Manual ZM-940PA/941PA Attaching the SpO2 Probe Selecting the SpO2 Probe Select an appropriate probe for the patient. CAUTION CAUTION Only use Nihon Kohden specified electrodes, electrode leads, SpO2 probes, and NIBP cuffs. Otherwise, the maximum performance from the transmitter cannot be guaranteed. Do not use a damaged or disassembled probe. It causes incorrect measurement and may injure the patient. Reusable Probes When using a TL-201T finger probe, choose the appropriate cable length for attachment. Probe Finger probe TL-201T Cable Length 0.6 m Patient Adult or child 20 kg or more Attachment Site Finger Adult or infant 3 kg or more Finger or toe Neonate 3 kg or less Instep and sole 1.6 m Multi-site probe TL-220T Attachment tape Finger probe TL-630T1/TL-630T3/TL-631T1/TL-631T3 Attachment tape TL-630T1/630T3: Finger or toe Adult or child 50 kg or more TL-631T1/631T3: Adult or child 20 kg or more Cable length TL-630T1/TL-631T1: 0.6 m TL-630T3/TL-631T3: 1.6 m Operator’s Manual ZM-940PA/941PA 53 Disposable Probes CAUTION The disposable probe is not sterilized. Use the disposable probe only for a single patient. Never reuse the disposable probe for another patient because it causes cross infection. Probe Patient Attachment Site Finger or toe TL-251T Adult 30 kg or more TL-252T Child 3 to 40 kg Finger or toe TL-253T Neonate 3 kg or less Instep and sole TL-051S/TL-052S Adult 50 kg or more Finger Neonate 3 kg or less Instep and sole Adult or child 15 to 50 kg Finger Infant 3 to 15 kg Toe 40 mm Cable length TL-051S: 0.8 m TL-052S: 1.6 m TL-061S/TL-062S 35 mm Cable length TL-061S: 0.8 m TL-062S: 1.6 m 54 Operator’s Manual ZM-940PA/941PA Probe TL-271T/TL-271T3 Cable length TL-271T: 0.8 m TL-271T3: 1.6 m TL-272T/TL-272T3 Cable length TL-272T: 0.8 m TL-272T3: 1.6 m TL-273T/TL-273T3 Patient Adult 30 kg or more Attachment Site Finger or toe Child 10 to 50 kg Neonate 3 kg or less Instep and sole Adult 40 kg or more Finger or toe Cable length TL-273T: 0.8 m TL-273T3: 1.6 m TL-274T/TL-274T3 Infant 3 to 20 kg Cable length TL-274T: 0.8 m TL-274T3: 1.6 m Operator’s Manual ZM-940PA/941PA 55 Connecting the SpO2 Probe to the Transmitter Connect the probe to the SpO2 socket on the transmitter. When the transmitter is attached on an arm 56 CAUTION CAUTION Do not shake or swing the transmitter while holding the leads or cables connected to the transmitter. The transmitter may come off and injure someone or damage surrounding instruments. Hold the connector when connecting/ disconnecting the SpO2 probe. If you disconnect the SpO2 probe by pulling the cable, it damages the cable. Operator’s Manual ZM-940PA/941PA Attaching the Probe to the Patient Attach the probe to the patient by referring to the probe’s manual. Make sure that the light emitter and photo detector of the probe face each other at the attachment site. WARNING WARNING • When using the TL-201T finger probe, do not fasten the probe and cable to the finger by wrapping with tape. This may cause burn, congestion or pressure necrosis from poor blood circulation. • When using probes other than the TL-201T finger probe, to avoid poor circulation, do not wrap the tape too tight. Check the blood circulation condition by observing the skin color and congestion at the skin peripheral to the probe attachment site. Even for short-term monitoring, there may be burn or pressure necrosis from poor blood circulation, especially on neonates or low birth weight infants whose skin is delicate. Accurate measurement cannot be performed on a site with poor peripheral circulation. Check the circulation condition by observing the skin color at the measurement site and pulse waveform. Change the measurement site every 8 hours for disposable probes and every 4 hours for reusable probes (every 8 hours for TL-630T/TL631T series probe). The skin temperature may increase at the attached site by 2 or 3°C (4 or 5°F) and cause a burn or pressure necrosis. When using the probe on the following patients, take extreme care and change the measurement site more frequently according to symptoms and degree. • Patient with a fever • Patient with peripheral circulation insufficiency • Neonate or low birth weight infant with delicate skin • Patient who is receiving photodynamic therapy* * Photodynamic therapy is a treatment to remove the affected tissue by using a photosensitizing agent and exposing the tissue to light. This treatment has a side effect of photosensitivity and the light from the finger probe sensor may cause a burn. This probe uses two light wavelengths in the range from 650 to 950 nm. The maximum light intensity is less than 5.5 mW/sr. Operator’s Manual ZM-940PA/941PA 57 CAUTION If the attachment site is dirty with blood or bodily fluids, clean the attachment site before attaching the probe. If there is nail polish on the attachment site, remove the polish. Otherwise, the amount of transmitted light decreases, and measured value may be incorrect or measurement cannot be performed. CAUTION If the skin gets irritated or redness appears on the skin from the probe, change the attachment site or stop using the probe. Take extreme care for the patients with delicate skin. CAUTION CAUTION When the probe is attached on an appropriate site with sufficient circulation and the error message confirming the probe attachment repeatedly appears, the probe may be deteriorated. Replace it with a new one. Do not immerse the disposable probe in detergents or water. If the probe adhesive surface gets wet, adhesiveness becomes weak and the probe cannot be attached to the skin. CAUTION Do not use a probe which is deteriorated by aging. Accurate measurement cannot be performed. CAUTION When removing a probe that is taped to the skin, do not pull the probe cable because this can damage the cable. CAUTION Neonatal skin is delicate. Remove the probe and tape carefully and slowly. CAUTION When removing the probe from the attachment tape, do not pull the sensor cable because this can damage the cable. CAUTION Refer to the probe instruction manual for details. 58 Operator’s Manual ZM-940PA/941PA Locking the Keys on the Transmitter To prevent the patient from pressing the keys on the transmitter during monitoring, you can lock the NIBP START/STOP and NIBP INTERVAL keys. 1. Press the NIBP INTERVAL key for about 3 seconds. 2. A “pip, pip” sounds and the “LOCK KEYS” message is displayed on the LCD. Pip, pip LOCK KEYS When the NIBP START/STOP key or NIBP INTERVAL key is pressed while the keys are locked, the “PRESS INT. KEY 3S TO UNLOCK” message appears. To unlock the keys: 1. Press the NIBP INTERVAL key for about 3 seconds. 2. A “pip, pip” sounds and the keys are unlocked. The “UNLOCK KEYS” message appears and the keys are unlocked. Operator’s Manual ZM-940PA/941PA 59 Monitoring When preparation is done, monitoring starts. NIBP Oscillometric Method NIBP is measured from the change in amplitude pattern of pulsatile oscillation in cuff pressure as the cuff pressure is reduced from above systolic to below diastolic pressure. The occlusiveoscillometry method uses this to determine the systolic, diastolic and mean arterial pressure. NIBP Monitoring Selecting the Initial Cuff Inflation Pressure The initial cuff inflation pressure can be changed on the PARAMETER SETUP screen. The default setting is 180 mmHg. To change the setting, refer to the “Changing Parameter Setup Settings” section. Selecting the Measurement Mode and Interval Measurement Modes There are three measurement modes: manual, auto and STAT. The selected mode or interval is displayed at the lower right of the screen. The measurement mode and interval can be changed by pressing the NIBP INTERVAL key. When the key is pressed, the measurement mode changes according to the modes selected at “SELECTABLE INTERVALS” on the PARAMETER SETUP screen. MANUAL mode is already selected for the mode selection. To select the modes for the mode selection, refer to the “Changing Parameter Setup Settings” section. Manual Measurement In Manual mode, a single NIBP measurement is performed when the NIBP START/STOP key is pressed. STAT (Continuous) Measurement In STAT mode, measurement is continuously repeated for 15 minutes after the NIBP START/ STOP key is pressed. When the STAT measurement for 15 minutes is completed, the measurement mode automatically changes to the Manual mode or Auto mode of selected interval depending on the “NIBP MODE AFTER STAT” setting on the PARAMETER SETUP screen. The default setting is Manual mode. 60 Operator’s Manual ZM-940PA/941PA Refer to the “Changing Parameter Setup Settings” section. The STAT measurement completes within 15 minutes. When more than 12 minutes elapse from the start of measurement, there will be no more measurement performed and the measurement mode changes to the mode selected for “NIBP MODE AFTER STAT” on the PARAMETER SETUP screen. Auto Measurement In Auto mode, measurement is performed automatically at the preset time intervals. In Auto mode, a single measurement can be performed by pressing the NIBP START/STOP key between auto measurements. Measuring NIBP WARNING Be careful when measuring NIBP on a patient with known bleeding disorders or coagulation. After NIBP measurement, there may be dot hemorrhage, or circulatory disorder by thrombus where the cuff is attached. WARNING When performing NIBP measurements in STAT mode or 5 minute intervals, periodically remove the cuff from the patient for ventilation. The skin temperature may increase at the cuff attachment site by 2 or 3°C (4 or 5°F). When measuring a patient with a fever or peripheral circulation insufficiency, it may cause a burn. WARNING NIBP measurement may be incorrect in the following cases. • When using an electrosurgical unit • When there is body movement • When the pulse wave is small (insufficient peripheral circulation) • Too many arrhythmias • When there is vibration • When there is a rapid blood pressure change • During CPR CAUTION When performing NIBP measurement repeatedly, have a rest between measurements to recover adequate circulation. Operator’s Manual ZM-940PA/941PA 61 NOTE • When measuring patients who are conscious, help the patient to relax. Measurement may not be accurate if the patient’s arm is tense or if the patient talks. • The data for measurement on a leg tends to be higher than measurement on the arm. When making diagnosis based on the NIBP values, measure NIBP on an upper arm. • Do not apply pressure to the cuff or air hose. NIBP may not be measured correctly because of noise or NIBP measurement may stop due to the NIBP safety circuit. • When the transmitter is attached to the patient arm and the NIBP measurement is performed when moving, tell the patient to relax and keep quiet. Otherwise, measurement may be stopped or remeasurement is repeated due to body movement. • If there is an abnormal noise generated during measurement, stop using the transmitter and contact your Nihon Kohden representative. • Do not measure NIBP of a patient on whom an IABP is being used. Measurement may be incorrect due to the mixing of the patient’s own pulse and IABP pulse. • NIBP cannot be measured on a neonate using this transmitter. 1. Select the measurement mode by pressing the NIBP INTERVAL key. 2. Press the NIBP START/STOP key to perform measurement. 62 Operator’s Manual ZM-940PA/941PA The cuff is inflated and the inflation pressure is displayed on the screen. Inflation pressure In manual mode: Measurement is performed once. In STAT mode: Measurement is performed repeatedly for 15 minutes. In auto mode: The first measurement is performed when the NIBP START/STOP key is pressed. The second measurement is performed when the current time in the transmitter reaches the selected time interval. To stop measurement during measurement, press the NIBP START/STOP key again. In STAT mode, after completing the STAT measurement, the measurement mode changes to the mode set for “NIBP MODE AFTER STAT” on the PARAMETER SETUP screen. In auto mode, to stop measurement in auto mode, change the mode to manual. To cancel one measurement, press the NIBP START/STOP key during measurement. After the measurement is complete, the measured data is displayed on the screen and is transmitted to the monitor. When SpO2 is not monitored, the pulse rate at the end of NIBP measurement is displayed. During auto mode measurement, the measurement mode can be changed. During the interval, press the NIBP INTERVAL key to change the mode. When “MANUAL” is displayed for more than one second, the measurement in auto mode is stopped. A buzzer can be set to sound at the start and end of NIBP measurement. Refer to the “Changing Parameter Setup Settings” section. Operator’s Manual ZM-940PA/941PA 63 Data Display After NIBP Measurement When the time set at “OLD NIBP DATA” on the PARAMETER SETUP screen elapses after the last measurement, the NIBP data is dimmed or hidden. Whether to dim or hide the old data can also be selected at “OLD NIBP DATA”. Refer to the “Changing Parameter Setup Settings” section. Data Display on the Receiving Monitor When the “Low battery ” message is displayed on the receiving monitor, NIBP might not have been measured according to the NIBP interval setting. Therefore, the NIBP data displayed on the receiving monitor might not be updated. In this case, check the measurement time of the NIBP data displayed on the receiving monitor. Monitoring SpO2 during NIBP Measurement When the SpO2 probe is attached to the same limb as the NIBP cuff, the blood flow decreases during NIBP measurement and pulse wave cannot be detected and SpO2 cannot be monitored properly. When “INHIBIT SpO2 DURING NIBP” on the PARAMETER SETUP screen is set to ON (factory default setting), SpO2 monitoring is paused during NIBP measurement to avoid SpO2 alarm occurrence. However, when monitoring SpO2 on the same limb as the NIBP, be careful when reading SpO2 values. ECG and Respiration Monitoring When the electrodes are attached and the ECG leads are connected to the electrodes, heart rate, ECG waveform, respiration rate and respiration waveform appear on the monitor. When 6 leads are used on this transmitter, up to 8 lead (I, II, III, aVR, aVL, aVF, Va and Vb) of ECG waveforms can be displayed on the receiving monitor. The heart rate is also measured. When 3 leads are used, one channel ECG waveform of lead II can be displayed on the receiving monitor. Refer to the operator’s manual of the monitor for details. Respiration is monitored by measuring changes in impedance between the RA and LL ECG electrodes. This transmitter sends the changes in impedance to the monitor as a respiration waveform. The monitor displays the respiration waveform and calculates respiration rate. Refer to the operator’s manual of the monitor for details. 64 Operator’s Manual ZM-940PA/941PA WARNING Interaction Between Minute Ventilation Rate-Adaptive Pacemakers and Cardiac Monitoring and Diagnostic Equipment* The bioelectric impedance measurement sensor of a minute ventilation rate-adaptive implantable pacemaker may be affected by transmitter which is connected to the same patient. If this occurs, the pacemaker may pace at its maximum rate and the transmitter may give incorrect data to the monitor. If this occurs, disconnect the electrode leads from the patient or change the setting on the pacemaker by referring to the pacemaker’s manual. For more details, contact your pacemaker representative or Nihon Kohden representative. * Minute ventilation is sensed in rate-adaptive pacemakers by a technology known as bioelectric impedance measurement (BIM). Many medical devices in addition to pacemakers use this technology. When one of these devices is used on a patient with an active, minute ventilation rateadaptive pacemaker, the pacemaker may erroneously interpret the mixture of BIM signals created in the patient, resulting in an elevated pacing rate. For more information, see the FDA web site. http://www.fda.gov/cdrh/safety.html WARNING When the transmitter is used with an electrosurgical unit (ESU), firmly attach the entire area of the ESU return plate. Otherwise, the current from the ESU flows into the electrodes of the transmitter, causing electrical burn where the electrodes are attached. For details, refer to the ESU manual. CAUTION Turn off the power of mobile phones, small wireless devices and other devices which produce strong electromagnetic interference around a patient (except for devices allowed by the hospital administrator). Radio waves from devices such as mobile phones or small wireless devices may be mistaken as pulse waves and the displayed data may be incorrect. NOTE • Noise generated from an electrosurgical unit may interfere on an ECG waveform, but will not damage it. • If an electric blanket is used and incorrect heart rate is displayed on the monitor, turn off the pacing spike detection on the monitor. • Turn the pacing spike detection to ON on the monitor when monitoring a pacemaker Operator’s Manual ZM-940PA/941PA 65 patient. Pacing pulse is detected by the transmitter and transmitted to the monitor. If the pacing spike detection is turned OFF, QRS and pacemaker spike may not be distinguished and pacemaker failure may not be recognized. Electrode Detachment In the following conditions, the check electrode indication is displayed on the LCD of the transmitter and the “CHECK ELECTRODE” message is displayed on the monitor. • Electrode is detached from skin. • Electrode lead is disconnected from the electrode. • Polarization voltage between the electrode and skin is excessively high. In these cases, check the cause and if necessary, replace electrodes with new ones. CAUTION When the “ELECTRODE OFF” or “CHECK ELECTRODE” message is displayed on the receiving monitor, ECG is not monitored properly and the ECG alarm does not function. Check the electrode, electrode leads, and if necessary, replace with new ones. Check Electrode Indication on the Transmitter when Monitoring with 3 Electrodes The “ ” mark is displayed. Check electrode symbol Check Electrode Indication on the Transmitter when Monitoring with 6 Electrodes The “ ” mark and either the detached lead or detached electrode position is indicated, depending on the LEADS OFF DISPLAY setting on the PARAMETER SETUP screen. Check electrode symbol Check electrode indication When LEADS OFF DISPLAY is set to CHAR, the detached lead is indicated 66 When LEADS OFF DISPLAY is set to IMAGE, the detached electrode position is indicated with X Operator’s Manual ZM-940PA/941PA SpO2 Monitoring When monitoring starts, SpO2 and pulse waveform are sent to the monitor and SpO2, pulse rate and pulse level bar graph are displayed on the transmitter LCD. WARNING WARNING SpO2 measurement may be incorrect in the following cases. • When the patient’s carboxyhemoglobin or methemoglobin increases abnormally. • When dye is injected in the blood. • When using an electrosurgical unit. • During CPR. • When measuring at a site with venous pulse. • When there is body movement. • When the pulse wave is small (insufficient peripheral circulation). Check the circulation condition by observing the skin color at the measurement site and pulse waveform. Change the measurement site every 8 hours for disposable probes and every 4 hours for reusable probes (every 8 hours for TL-630T/TL631T series probe). The skin temperature may increase at the attached site by 2 or 3°C (4 or 5°F) and cause a burn or pressure necrosis. When using the probe on the following patients, take extreme care and change the measurement site more frequently according to symptoms and degree. • Patient with a fever • Patient with peripheral circulation insufficiency • Neonate or low birth weight infant with delicate skin • Patient who is receiving photodynamic therapy* WARNING When not monitoring SpO2, disconnect the SpO2 cable from the transmitter. Otherwise, noise from the probe sensor may interfere and incorrect data is displayed on the screen. * Photodynamic therapy is a treatment to remove the affected tissue by using a photosensitizing agent and exposing the tissue to light. This treatment has a side effect of photosensitivity and the light from the finger probe sensor may cause a burn. This probe uses two light wavelengths in the range from 650 to 950 nm. The maximum light intensity is less than 5.5 mW/sr. Operator’s Manual ZM-940PA/941PA 67 CAUTION CAUTION Turn off the power of mobile phones, small wireless devices and other devices which produce strong electromagnetic interference around a patient (except for devices allowed by the hospital administrator). Radio waves from devices such as mobile phones or small wireless devices may be mistaken as pulse waves and the displayed data may be incorrect. Do not pull or bend the probe cable, and do not put caster feet on the probe cable. Do not immerse the probe cable in chemical solutions or water. Failure to follow these instructions may cause cable discontinuity, short circuit, skin burn on the patient and incorrect measurement data. Replace any broken probe with a new one. CAUTION CAUTION Normal external light does not affect monitoring but strong light such as a surgical light or sunlight may affect monitoring. If affected, cover the measuring site with a blanket. When the probe is attached on an appropriate site with sufficient circulation and the error message confirming the probe attachment repeatedly appears, the probe may be deteriorated. Replace it with a new one. CAUTION CAUTION When a message indicates a faulty probe, stop monitoring and replace the probe with a new one. While a patient is on medication which causes vasodilation, the pulse waveform may change and in rare cases the SpO2 value might not be displayed. NOTE In order to maintain sufficient blood circulation, keep the measurement site warm by covering it with a blanket or something similar. Warming the site is effective, especially for a patient with a small pulse amplitude. SpO2 and PR Display Order You can select the display order for SpO2 and PR (pulse rate) on the LCD. Refer to the “Changing Parameter Setup Settings” section. 68 Operator’s Manual ZM-940PA/941PA Monitoring SpO2 during NIBP Measurement When the SpO2 probe is attached to the same limb as the NIBP cuff, the blood flow decreases during NIBP measurement and pulse wave cannot be detected and SpO2 cannot be monitored properly. When “INHIBIT SpO2 DURING NIBP” on the PARAMETER SETUP screen is set to ON (factory default setting), SpO2 monitoring is paused during NIBP measurement to avoid SpO2 alarm occurrence. However, when monitoring SpO2 on the same limb as NIBP, be careful when reading SpO2 values. When monitoring SpO2 is important, attach the probe to the limb to which the NIBP cuff or catheter is not attached. When SpO2 monitoring is paused during NIBP measurement, the SpO2 value just before the start of NIBP measurement and an mark are displayed on the transmitter for 30 seconds. When NIBP measurement is not completed after 30 seconds, “– – –” is displayed for the SpO2 value. The same data also appears on the monitor screen. NOTE • When continuous SpO2 monitoring is necessary, attach the probe to the limb to which the NIBP cuff is not attached and set “INHIBIT SpO2 DURING NIBP” on the PARAMETER SETUP screen to OFF. • When the probe is attached to the same limb as the NIBP cuff, set the sync source to a parameter other than SpO2 on the receiving monitor. • When monitoring SpO2 during STAT NIBP measurement, attach the probe to the limb to which the NIBP cuff is not attached. Operator’s Manual ZM-940PA/941PA 69 Display and Message List Battery Indication Indication Cause Countermeasure Fully charged battery —— Batteries are low. Batteries are low. NIBP cannot be measured. No indication Replace batteries. Dead batteries ECG/Respiration Indication Cause Electrode lead is disconnected from the electrode. Electrode lead is disconnected from the transmitter. Electrode lead discontinuity. Electrode is not firmly attached to the skin. Polarization voltage is abnormally high. Countermeasure Firmly connect the electrode lead to the electrode. Firmly connect the electrode lead to the transmitter. Replace the electrode lead with a new one. Replace the electrode with a new one. When monitoring ECG with 6 electrodes, the electrode or lead detached position is indicated by either lead or electrode position. This is set at LEADS OFF DISPLAY on the PARAMETER SETUP screen. Refer to the “Changing Parameter Setup Settings” section. LEADS OFF DISPLAY set to CHAR ECG ELECTRODE set to AHA LEADS OFF DISPLAY set to CHAR ECG ELECTRODE set to IEC LEADS OFF DISPLAY set to IMAGE 70 Operator’s Manual ZM-940PA/941PA SpO2 Message During NIBP measurement Cause SpO2 monitoring is paused for NIBP measurement. Considerable body movement. Detecting body The probe is not attached to the movement patient properly. The probe is not attached to the patient properly. SpO2 CHECK The probe is not attached at the PROBE appropriate site. Probe is expired. SpO2 DETECTING PULSE SpO2 LIGHT INTERFERENCE SpO2 PROBE FAILURE SpO2 WEAK PULSE Searching for the correct pulse wave. The SpO2 value cannot be obtained because the waveform is unstable. The probe is not attached to the patient properly. SpO2 measurement site is under fluorescent light, surgical light, sunlight, etc. Countermeasure Wait for NIBP measurement to finish. When the message is displayed frequently, check the patient condition and, if necessary, change the attachment site. Attach the probe to the patient properly. Attach the probe to a site 6 to 14 mm thick. Replace the probe with a new one. Wait until the pulse wave is detected. Attach the probe to the patient properly. Cover the measurement site with a blanket or cloth. Probe is expired. Replace the probe with a new one. Probe is damaged or shortcircuited. Replace the probe with a new one. Poor peripheral circulation. Check the patient condition and change the attachment site. The probe is attached too tightly and is obstructing the blood circulation. Check the probe attachment condition and if necessary, reattach the probe. Operator’s Manual ZM-940PA/941PA 71 NIBP Message Cause The cuff and extension hose are not properly connected. NIBP AIR LEAK The cuff hose (or extension hose) is not properly connected to the NIBP socket. Countermeasure Connect them properly. The cuff or extension hose is damaged. Replace with a new one. The patient’s pulse wave is small. Measure by palpation or auscultation. The cuff is not wrapped on the patient properly. Wrap the cuff on the patient properly. NIBP CHANGE BATTERIES NO NIBP NIBP cannot be measured due to low battery. Replace batteries with new ones. NIBP CUFF OCCLUSION Transmitter malfunction. Immediately remove the cuff from the patient and contact your Nihon Kohden representative. NIBP HIGH CUFF PRESS Enormous pressure was applied by the pressure of the cuff. Remove the cause. NIBP INFLATION PRESS LOW Insufficient cuff inflation pressure. Wait for the remeasurement to be performed with increased cuff inflation pressure. NIBP MEAS TIME-OUT The measuring time exceeded the Remove the cause if the cause is body specified time due to arrhythmia, body movement, vibration or squeezing of movement, vibration or, cuff or air hose cuff or hose. being squeezed. NIBP CANNOT DETECT PULSE NIBP MODULE FAILURE Module malfunction. NIBP REMEASURING NIBP is being remeasured due to arrhythmia, body movement, vibration or, cuff or air hose being squeezed. Contact your Nihon Kohden representative. If the message still appears after remeasurement, remove the cause if the cause is body movement, vibration or squeezing of cuff or hose. Check that the hose is not bent or squeezed. NIBP SAFETY CIRCUIT RUNNING (When this message is displayed, measurement cannot be performed for 40 seconds.) Measurement stopped by the safety circuit. NIBP SYS OUT OF RANGE The maximum blood pressure cannot be measured even when the cuff Measure by palpation or auscultation. inflation pressure exceeded 280 mmHg when using adult cuff. NIBP WEAK PULSE NIBP ZEROING 72 Wait 40 seconds, then perform remeasurement. If the message still appears, contact your Nihon Kohden representative. The patient’s pulse wave is too small. Measure by palpation or auscultation. The cuff is wrapped too loosely. Wrap the cuff properly. The cuff size is not appropriate. Use the appropriate cuff. NIBP zero balance is being adjusted. Do not touch the cuff during zeroing. Wait for the message to disappear. Operator’s Manual ZM-940PA/941PA Troubleshooting If the problem still remains after checking the following, contact your Nihon Kohden representative. Transmitter Problem Nothing is displayed on the LCD after turning the power on. Cause Batteries are not installed correctly. The battery polarity is wrong. Batteries are completely discharged. LCD is difficult to see LCD brightness is not (too dark or too light). appropriate. Nothing is displayed on the monitor after turning the transmitter power on. Signal receiving condition is poor. The channel of the transmitter and monitor does not match. The software version of the multiple patient receiver is old. Countermeasure Install the batteries correctly. Replace the batteries with new ones. Change the LCD brightness on the SYSTEM SETUP screen. Refer to the “Changing System Setup Settings” section. Set the correct channel on the monitor. Upgrade the multiple patient receiver software to receive signal from the transmitter. The software version must be 01-09 or later. Another transmitter of the same Turn the transmitter power off. If the channel is used nearby. monitor still receives a signal, there is a high probability that another transmitter of the same channel is used nearby. Follow the instruction of your channel administrator and use another transmitter of a different channel. Signals are mixing. Follow the instructions of your channel administrator and use another transmitter of a different channel. The transmitter is damaged. Contact your Nihon Kohden representative. Operator’s Manual ZM-940PA/941PA 73 ECG/Respiration Problem The heart rate is unstable. The “CHECK ELECTRODE” message appears on the receiving monitor. ECG baseline is thick. (Hum is overlapping) Respiration waveform measurement is unstable. Cause Countermeasure Pacing detection setting on the monitor is not correct. Turn off the pacing detection setting on the monitor. When monitoring a pacemaker patient, turn on pacing detection. Electrode lead is disconnected Firmly connect the electrode lead to the from the electrode. electrode. Electrode lead discontinuity Replace the electrode lead with a new one. Electrode is not firmly attached Replace the electrode with a new one. to the skin. Polarization voltage is Use Nihon Kohden specified electrodes. abnormally high. The gel on the electrode is dried Replace the electrode with a new one. out. The gel on the electrode is coming off. Electric blanket is used. Cover the blanket with a shield cover. Hum filter is set to OFF on the Set the filter to ON. monitor The gel on the electrode is dried Replace the electrode with a new one. out. The gel on the electrode is coming off. SpO2 Problem SpO2 data is unstable and not reliable. 74 Cause The probe size is not appropriate for the patient. Probe attachment condition is poor. Probe is partly detached from the skin. External light gets in. Measurement site is dirty. Patient is wearing nail polish. Probe is attached to the same limb that is used for NIBP measurement. Countermeasure Use the appropriate probe for the patient. Firmly attach the probe according to the procedure in the probe operator’s manual. Remove dirt and nail polish. When the probe and cuff are attached to the same limb, set “INHIBIT SpO2 DURING NIBP” setting on the PARAMETER SETUP screen to ON. Attach the probe to the opposite limb. Avoid a site where blood circulation condition changes greatly. Operator’s Manual ZM-940PA/941PA NIBP Problem Cause Cuff inflation pressure The cuff hose is not connected is less than 10 mmHg. to the NIBP socket properly. The cuff is not wrapped around the arm or is wrapped too loosely. The cuff does not The cuff hose is not connected inflate when the NIBP to the NIBP socket. START/STOP key is The cuff hose or extension hose pressed. may be folded or squeezed when the cuff pressure display on the screen increases quickly but the actual cuff does not inflate. Abnormal The cuff size is not correct. measurement results are displayed. The cuff is not wrapped around the arm correctly. NIBP data is not correct because of body movement. Vibration on the cuff. The cuff is suddenly deflated during inflation. Auto mode measurement does not start even when the time interval has passed. The cuff suddenly inflates. Cannot connect cuff to the air hose. The NIBP START/STOP key is pressed during inflation. The NIBP INTERVAL key is pressed and the measurement mode is changed. Countermeasure Connect the cuff hose to the socket properly. Wrap the cuff around the upper arm. Connect the cuff hose to the socket firmly. Check the cuff hose and air hose. Select the cuff which fits the patient’s limb circumference. Wrap the cuff around the upper arm, not too tightly or too loosely. Prevent the patient from moving during measurement. Check that nothing is touching the cuff during measurement. Change the measuring site. — Check the measurement mode and interval. The measurement mode is set to Check the time interval. If necessary, auto mode. stop measurement. Unspecified cuff is used. Use a cuff specified by Nihon Kohden. Operator’s Manual ZM-940PA/941PA 75 Problem Cause Cannot measure NIBP. Vibration on the cuff. Countermeasure Check that nothing is touching the cuff during measurement. The cuff hose or extension hose Remove the cause. is bent or squeezed. The cuff has worn out. Use a new cuff. Blood congestion Measuring over a long period of Increase the measuring interval. occurs. time at short intervals. Do not measure NIBP over a long time. Thrombus occurs. Measuring on a patient with Do not perform NIBP measurement on known bleeding disorders or such a patient. coagulation. NIBP data on the The time set for “OLD NIBP When NIBP is measured again, the data screen is --- or dark. DATA” on the PARAMETER is displayed in normal brightness. SETUP screen elapsed from the last measurement. Three loud pip sounds The cuff is not deflated enough Wait 30 seconds and measure again. indicting NIBP to start another measurement. measurement cannot be started. 76 Operator’s Manual ZM-940PA/941PA Maintenance To use the instrument in safe and optimum condition, perform maintenance check once every six months. CAUTION Never disassemble or repair the transmitter. Disassembly and repair must be performed by qualified service personnel. A maintenance check sheet is provided at the end of this section. Make a copy of this check sheet before performing maintenance check. 1. External Check • There are no damaged or dirty parts on the outside of the transmitter. • The battery case cover is not damaged, the spring is firmly fixed and the battery case cover can be closed firmly. • NIBP socket is not damaged. • Keys are not damaged. • Electrode leads are not damaged. • There is no blood or chemicals on the transmitter. 2. Transmitter Channel Check that the channel of the transmitter and the label match. 1. Check that the channel number label attached to the transmitter is not torn or removed. Channel label 2. Remove one battery. Operator’s Manual ZM-940PA/941PA 77 3. While pressing the NIBP START/STOP and NIBP INTERVAL keys, install the battery. The SETUP screen appears. 4. Press the NIBP INTERVAL key to move the cursor to “SYSTEM SETUP”. 5. Press the NIBP START/STOP key to enter the SYSTEM SETUP screen. The channel of this transmitter is displayed. Channel 78 6. Check that the channel displayed on the LCD matches the label on the transmitter. Operator’s Manual ZM-940PA/941PA 3. LCD Display Check that there are no dots missing on the LCD. 1. Remove one battery. 2. While pressing the NIBP START/STOP and NIBP INTERVAL keys, install the battery. The SETUP screen appears. 3. Press the NIBP INTERVAL key twice to move the cursor to “MANUAL CHECK”. 4. Press the NIBP START/STOP key to enter the MANUAL CHECK screen. 5. Check that the cursor is on “LCD TEST” and press the NIBP START/STOP key. Operator’s Manual ZM-940PA/941PA 79 6. Every time the NIBP INTERVAL key is pressed, the screen changes as below. Check that there are no dots missing. When the NIBP START/STOP key is pressed, the screen returns to the MANUAL CHECK screen. 4. Key Operation NIBP START/STOP Key 1. Attach the NIBP cuff to your upper arm. 2. Press the NIBP START/STOP key. Check that the cuff inflates and deflates properly. 3. Press the NIBP START/STOP key again. During inflation, press the NIBP START/STOP key to check that the cuff deflates properly. NIBP INTERVAL Key 1. Press the NIBP INTERVAL key and check that the NIBP measuring mode can be changed. 2. 80 Select any interval and press the NIBP START/STOP key to perform auto measurement. Check that the NIBP is measured at the selected interval. Operator’s Manual ZM-940PA/941PA 5. NIBP Cuff for Attaching Transmitter to Patient Arm The NIBP cuff is a consumable. Check the following and when necessary, replace it with a new one. Appearance • There are no dirty parts. • There are no broken stitches on the cuff. • The label on the cuff is readable. • The velcro tape on the cuff is not removed and there are no broken stitches. • The lock plate is not damaged and functions properly. Inflation bag • The inflation bag is not torn or damaged. • There is no water inside the inflation bag. • The connector on the inflation bag is not damaged. Operator’s Manual ZM-940PA/941PA 81 Maintenance Check Sheet Hospital/Organization: Service Personnel: Instrument Name: Transmitter Instrument Model: ZM-940PA/ZM-941PA Instrument Serial Number: Hardware Revision Number: Software Revision Number: 1. External Check OK No 2. Transmitter Channel OK No 3. LCD Display OK No 4. Key Operation OK No 5. NIBP Cuff for Attaching Transmitter to Patient Arm OK No Overall Judgement □ OK □ Can be used but needs maintenance □ Maintenance required. Cannot be used. 82 Operator’s Manual ZM-940PA/941PA Repair Parts Availability Policy Nihon Kohden Corporation (NKC) shall stock repair parts (parts necessary to maintain the performance of the instrument) for a period of 8 years from the date of delivery. In that period NKC or its authorized agents will repair the instrument. This period may be shorter than 8 years if the board or part necessary for the faulty section is not available. Operator’s Manual ZM-940PA/941PA 83 Lifetime and Disposal Disposing of Used Batteries Battery Lifetime Replace the batteries when the battery replacement indication appears on the transmitter. When using rechargeable batteries, recharge them. ZM-940PA Type NiMH secondary Alkaline primary Lifetime (Measuring parameters) ECG, SpO2, NIBP ECG, SpO2 ECG only 2 days 2.5 days 3 days 1 day 2.5 days 3 days ZM-941PA Type NiMH secondary Alkaline primary Lifetime (Measuring parameters) ECG, SpO2, NIBP ECG, SpO2 ECG only 1.5 days 2 days 2.5 days 1 day 2 days 2.5 days The above data is when the following batteries and battery charger which are recommended by Nihon Kohden are used. The measurement is performed at room temperature, NIBP is measured in auto mode at 60 minute intervals and SpO2 is measured on an index finger of a male patient with weight 60 kg. Operation time depends on the thickness of the SpO2 probe attachment site. NiMH secondary: SANYO HR-3UF (W) Battery charger: SANYO NC-M55 Alkaline primary: Nihon Kohden Medipower (equivalent to Panasonic LR6 (G)) Disposal NOTE Remove the batteries before disposing of the transmitter. Before disposing of the batteries, check with your local solid waste officials for details in your area for proper disposal. It may be illegal to dispose of these batteries in the municipal waste stream. Disposing of Electrodes, SpO2 Probes and NIBP Cuffs Refer to the manual of each item. 84 Operator’s Manual ZM-940PA/941PA Cleaning, Disinfection and Sterilization Transmitter and Electrode Leads CAUTION CAUTION This transmitter is not waterproof. If detergent or liquid spills into the transmitter, stop cleaning or disinfecting it and contact your Nihon Kohden representative. The transmitter needs to be checked for safety and function before use. Before cleaning or disinfection, remove the batteries from the transmitter. Failure to follow this instruction may result in electrical shock or transmitter malfunction. CAUTION CAUTION Dispose of the transmitter, options and accessories as specified by Nihon Kohden. Otherwise, it causes infection or environmental contamination. The transmitter cannot be sterilized. Sterilizing the transmitter may damage it. Before cleaning or disinfecting, remove the batteries from the transmitter. Be careful not to let any liquid get inside the transmitter. Cleaning Wipe the transmitter and electrode leads with a soft cloth moistened with disinfecting alcohol or neutral detergent diluted with water. After cleaning, dry them completely. Disinfection CAUTION • Do not immerse the electrode lead connector in liquid. • Do not disinfect with hypochlorous acid. • Use the recommended concentration. Wipe the outside surface of the transmitter and electrode lead with a non-abrasive cloth moistened with any of the disinfectants listed below. Use the recommended concentration. Operator’s Manual ZM-940PA/941PA 85 Disinfectant Glutaraldehyde solution Alkyldiaminoethylglycine hydrochloride Benzalkonium chloride Benzethonium chloride solution Chlorhexidine gluconate solution Concentration (%) 2.0 0.5 0.2 0.2 0.5 SpO2 Probe Refer to the probe manual. YP-943P/944P NIBP Cuffs CAUTION • • • • Do not autoclave. Use only glutaraldehyde solution. Never allow liquid to get inside the inflation bag. Do not sterilize or disinfect the cuff with ultraviolet light or ozone. Cleaning To clean the cuff, remove the lock plate and carefully pull out the inflation bag from the cloth cover. Cloth cover: Wash with neutral detergent and water. Thoroughly dry it. When washing in a washing machine, put it in a net. Inflation bag: Wipe with a soft cloth or cotton moistened with disinfecting alcohol. Thoroughly dry it. Disinfection To disinfect the cuff, use glutaraldehyde solution. Use the recommended concentration of the disinfectant. Refer to the disinfectant manual for details. After disinfection, clean the cuff as described above. 86 Operator’s Manual ZM-940PA/941PA Specifications ZM-940PA Measuring Parameters Measuring waveforms: Measuring numeric data: ECG, Respiration in impedance method, pulse SpO2, NIBP, pulse rate Transmitting Data Waveform data: Numeric data: Status information: ECG, respiration, pulse wave SpO2 and NIBP Battery replacement, channel ID, type of transmitter, check electrodes, abnormal polarization voltage, pacing data, SpO2 status, NIBP status Displayed Data SpO2, NIBP, pulse rate, pulse wave bar graph, check electrode, battery replacement, NIBP measurement mode and status information ECG Measurement Channels: Input range: DC offset: Input impedance: Pacing pulse detection: ±5 mV or more ±500 mV or more 5 MΩ or more (5 Hz) ANSI/AAMI EC13 Based upon Pacemaker pulse rejection Capability ECG recovery time after defibrillation: within 10 s Respiration Measurement Measuring method: Impedance range: Impedance method 0 to 2 kΩ or less SpO2 Measurement Display range: Measuring range: Minimum display range: Measuring accuracy Depends on the receiving monitor 0 to 100%, in 1% steps 1% When the measuring accuracy of the SpO2 probe is not considered: ±1 (80% ≤ SpO2 ≤ 100%) ±2 (50% ≤ SpO2 < 80%) Less than 50% is not specified Operator’s Manual ZM-940PA/941PA 87 When considering the measuring accuracy of the SpO2 probe: ±2 (80% ≤ SpO2 ≤ 100%) ±3 (70% ≤ SpO2 < 80%) Less than 70% is not specified NIBP Measurement Displayed items: Cuff pressure display range: Measurement modes: Measurement accuracy: Systolic, diastolic, mean 0 to 300 mmHg Manual, STAT, auto at 5, 10, 15, 30, 60, 120 or 240 minute interval ±3 mmHg (0 ≤ NIBP ≤ 200 mmHg) ±4 mmHg (200 < NIBP < 300 mmHg) Meets or exceeds AAMI Sp-10. 1992 standard (Maximum mean error: ±5 mmHg Maximum standard deviation: 8 mmHg) Pulse Rate Measuring range: 30 to 200 beats/min ±8 beats/min (NIBP) 30 to 250 beats/min ±3% ±1 beat/min (SpO2) Transmitter FCC regulation: Field strength limits: Undesired emission: Antenna: Transmission channel: Transmission frequency range: Channel spacing: Modulation: Type of emission: Occupied bandwidth: Effective radiated power: FCC part 95 Subpart H Wireless Medical Telemetry Service (WMTS) <200 mV/m (at 3 m) below 960 MHz: 200 μV/m (at 3 m) above 960 MHz: 500 μV/m (at 3 m) Internal Indicated on the transmitter 608.0250 to 613.9750 MHz 50 kHz or 37.5 kHz (12.5 kHz when interleaved) Frequency shift keying F1D <20 kHz 1.0 mW Power Requirements Operating voltage: Battery type: 88 3.2 to 4.8 V Three AA (R6) type NiMH secondary batteries Three AA (R6) type alkaline dry cell primary batteries Operator’s Manual ZM-940PA/941PA Battery lifetime: Type NiMH secondary Alkaline primary Lifetime (Measuring parameters) ECG, SpO2, NIBP ECG, SpO2 ECG only 2 days 2.5 days 3 days 1 day 2.5 days 3 days The above data is when the following batteries and battery charger which are recommended by Nihon Kohden are used. The measurement is performed at room temperature, NIBP is measured in auto mode at 60 minute intervals and SpO2 is measured on an index finger of a male patient with weight 60 kg. Operation time depends on the thickness of the SpO2 probe attachment site. NiMH secondary: SANYO HR-3UF (W) Battery charger: SANYO NC-M55 Alkaline primary: Nihon Kohden Medipower (equivalent to Panasonic LR6 (G)) Dimension and Weight Dimension: Weight: accessories) 114 W × 103 H × 58 D (mm) 280 g ±30 g (excluding batteries, NIBP cuff and other Environment Operating environment Operating temperature: Operating humidity: Operating atmospheric pressure: 5 to 40°C, 41 to 104°F When using NIBP cuff, 10 to 40°C, 50 to 104°F 30 to 85% (non-condensing) 70 to 106 kPa Storage environment Storage temperature: Storage humidity: Storage atmospheric pressure: −20 to 65°C, −4 to 149°F 10 to 95% 70 to 106 kPa Safety Standards Safety standard: Operator’s Manual ZM-940PA/941PA CAN/CSA-C22.2 No. 601-1 M90: 1990 CAN/CSA-C22.2 No. 601-1. 1S1-94: 1994 CAN/CSA-C22.2 No. 601-1. 1B-90: R2002 CAN/CSA-C22.2 No. 60601-2-49-04: 2004 CAN/CSA-C22.2 No. 60601-2-27: 1998 CAN/CSA-C22.2 No. 60601-2-30: 2002 IEC 60601-1:1988 IEC 60601-1 Amendment 1: 1991 89 IEC 60601-1 Amendment 2: 1995 IEC 60601-1-2: 2001 IEC 60601-2-27: 1994 IEC 60601-2-30: 1999 IEC 60601-2-49: 2001 According to the type of protection against electrical shock: INTERNALLY POWERED EQUIPMENT According to the degree of protection against electrical shock: ECG and impedance method respiration: DEFIBRILLATION-PROOF TYPE CF APPLIED PART SpO2 and NIBP: DEFIBRILLATION-PROOF TYPE BF APPLIED PART According to the degree of protection against harmful ingress of water: IPX0 (Ordinary equipment) According to the degree of safety of application in the presence of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE: Equipment not suitable for use in the presence of FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE According to the mode of operation: CONTINUOUS OPERATION Electromagnetic Compatibility IEC 60601-1-2: 2001 Electromagnetic Emissions This Model ZM-940PA is intended for use in the electromagnetic environment specified below. The customer or the user of the ZM-940PA should assure that it is used in such an environment. Emissions test RF emissions CISPR 11 RF emissions CISPR 11 Harmonic emissions IEC 61000-3-2 Voltage fluctuations/ flicker emissions IEC 61000-3-3 90 Compliance Electromagnetic environment guidance Group 1 The ZM-940PA uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. Class B The ZM-940PA is suitable for use in all establishments, including domestic establishments. Not applicable Not applicable Operator’s Manual ZM-940PA/941PA Electromagnetic Immunity This Model ZM-940PA is intended for use in the electromagnetic environment specified below. The customer or the user of the ZM-940PA should assure that it is used in such an environment. Immunity test Electrostatic discharge (ESD) IEC 61000-4-2 IEC 60601 test level ±6 kV contact ±8 kV air Compliance level ±6 kV contact ±8 kV air Electrical fast transient/ burst IEC 61000-4-4 Not applicable ±2 kV for power supply lines ±1 kV for input/output lines Surge Not applicable ±1 kV differential IEC 61000-4-5 mode ±2 kV common mode Voltage dips, short <5% UT (>95% dip in Not applicable interruptions and UT) for 0.5 cycle voltage variations on power supply 40% UT (60% dip in input lines UT) for 5 cycles IEC 61000-4-11 70% UT (30% dip in UT) for 25 cycles Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 <5% UT (>95% dip in UT) for 5 s 3 A/m Electromagnetic environment - guidance Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. — — — 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. NOTE: UT is the AC mains voltage prior to application of the test level Avoiding Electromagnetic Interference (Impedance Respiration) Impedance respiration measurement is very sensitive and affected by electromagnetic interference. Technological limitations do not allow immunity levels higher than 1 V/m for radiated RF electromagnetic fields. Electromagnetic fields with field strengths above 1 V/m may cause measurement error. Do not use electrically radiating equipment near the impedance respiration measurements. Operator’s Manual ZM-940PA/941PA 91 Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Portable and mobile RF communications equipment should be used no closer to any part of the ZM-940PA, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance Conducted RF IEC 61000-4-6 3 Vrms 3 Vrms Radiated RF IEC 61000-4-3 3 V/m 3 V/m 80 MHz to 80 MHz to 2.5 GHz 2.5 GHz (1 V/m 80 MHz to 2.5 GHz for respiration) d = 1.2√P d = 1.2√P 80 MHz to 800 MHz d = 2.3√P 800 MHz to 2.5 GHz (d = 3.5√P 80 MHz to 800 MHz for respiration d = 7.0√P 800 MHz to 2.5 GHz for respiration) where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as deter mined by an electromagnetic site survey*1, should be less than the compliance level in each frequency range*2. Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. *1 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the ZM-940PA is used exceeds the applicable RF compliance level above, the ZM-940PA should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the ZM-940PA. *2 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m for respiration and 3 V/m for all other functions. 92 Operator’s Manual ZM-940PA/941PA Recommended Separation Distances between Portable and Mobile RF Communications Equipment The ZM-940PA is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the ZM-940PA can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the ZM-940PA as recommended below, according to the maximum output power of the communications. Separation distance according to frequency of transmitter (m) Rated maximum output power of transmitter (W) 150 kHz to 80 MHz d = 1.2√P 80 MHz to 800 MHz d = 1.2√P 800 MHz to 2.5 GHz d = 2.3√P (For respiration: d = 3.5√P) (For respiration: d = 7.0√P) 0.01 0.12 0.12 (0.35*) 0.23 (0.7*) 0.1 0.38 0.38 (1.1*) 0.73 (2.2*) 1.2 1.2 (3.5*) 2.3 (7.0*) 10 3.8 3.8 (11*) 7.3 (22*) 100 12 12 (35*) 23 (70*) For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. (* For respiration) NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Recovery Time after Defibrillation The transmitter returns to the normal operating mode within 10 seconds after defibrillation. The stored settings are not affected. System Composition for EMC Test The ZM-940PA bedside monitor is tested to comply with IEC 60601-1-2: 2001 with the following composition. Units ZM-940PA transmitter YP-943P NIBP cuff BR-906P ECG electrode lead TL-201T finger probe Operator’s Manual ZM-940PA/941PA Cable length — 0.15 m 0.8 m 1.6 m 93 ZM-941PA Measuring Parameters Measuring waveforms: Measuring numeric data: ECG, Respiration in impedance method, pulse SpO2, NIBP, pulse rate Transmitting Data Waveform data: Numeric data: Status information: ECG, respiration, pulse wave SpO2 and NIBP Battery replacement, channel ID, type of transmitter, check electrodes, abnormal polarization voltage, pacing data, SpO2 status, NIBP status Displayed Data SpO2, NIBP, pulse rate, pulse wave bar graph, check electrode, battery replacement, NIBP measurement mode and status information ECG Measurement Channels: Input range: DC offset: Input impedance: Pacing pulse detection: ±5 mV or more ±500 mV or more 5 MΩ or more (5 Hz) ANSI/AAMI EC13 Based upon Pacemaker pulse rejection Capability ECG recovery time after defibrillation: within 10 s Respiration Measurement Measuring method: Impedance range: Impedance method 0 to 2 kΩ or less SpO2 Measurement Display range: Measuring range: Minimum display range: Measuring accuracy 94 Depends on the receiving monitor 0 to 100%, in 1% steps 1% When the measuring accuracy of the SpO2 probe is not considered: ±1 (80% ≤ SpO2 ≤ 100%) ±2 (50% ≤ SpO2 < 80%) Less than 50% is not specified When considering the measuring accuracy of the SpO2 probe: ±2 (80% ≤ SpO2 ≤ 100%) ±3 (70% ≤ SpO2 < 80%) Less than 70% is not specified Operator’s Manual ZM-940PA/941PA NIBP Measurement Displayed items: Cuff pressure display range: Measurement modes: Systolic, diastolic, mean 0 to 300 mmHg Manual, STAT, auto at 5, 10, 15, 30, 60, 120 or 240 minute interval ±3 mmHg (0 ≤ NIBP ≤ 200 mmHg) ±4 mmHg (200 < NIBP < 300 mmHg) Meets or exceeds AAMI Sp-10. 1992 standard (Maximum mean error: ±5 mmHg Maximum standard deviation: 8 mmHg) Measurement accuracy: Pulse Rate Measuring range: 30 to 200 beats/min ±8 beats/min (NIBP) 30 to 250 beats/min ±3% ±1 beat/min (SpO2) Transmitter FCC regulation: FCC part 95 Subpart H Wireless Medical Telemetry Service (WMTS) <740 mV/m (at 3 m) below 960 MHz: 200 μV/m (at 3 m) above 960 MHz: 500 μV/m (at 3 m) Internal Indicated on the transmitter 1395.0250 to 1399.9750 MHz 1427.0250 to 1431.9750 MHz 50 kHz or 37.5 kHz (12.5 kHz when interleaved) Frequency shift keying F1D <20 kHz 5.0 mW (factory default setting) Can be changed to 1.0 mW if required Field strength limits: Undesired emission: Antenna: Transmission channel: Transmission frequency range: Channel spacing: Modulation: Type of emission: Occupied bandwidth: Effective radiated power: Power Requirements Operating voltage: Battery type: 3.2 to 4.8 V Three AA (R6) type NiMH secondary batteries Three AA (R6) type alkaline dry cell primary batteries Battery lifetime: Type NiMH secondary Alkaline primary Lifetime (Measuring parameters) ECG, SpO2, NIBP ECG, SpO2 ECG only 1.5 days 2 days 2.5 days 1 day 2 days 2.5 days Operator’s Manual ZM-940PA/941PA 95 The above data is when the following batteries and battery charger which are recommended by Nihon Kohden are used. The measurement is performed at room temperature, NIBP is measured in auto mode at 60 minute intervals and SpO2 is measured on an index finger of a male patient with weight 60 kg. Operation time depends on the thickness of the SpO2 probe attachment site. NiMH secondary: SANYO HR-3UF (W) Battery charger: SANYO NC-M55 Alkaline primary: Nihon Kohden Medipower (equivalent to Panasonic LR6 (G)) Dimension and Weight Dimension: Weight: 114 W × 103 H × 58 D (mm) 280 g ±30 g (excluding batteries, NIBP cuff and other accessories) Environment Operating environment Operating temperature: Operating humidity: Operating atmospheric pressure: Storage environment Storage temperature: Storage humidity: Storage atmospheric pressure: 5 to 40°C, 41 to 104°F When using NIBP cuff, 10 to 40°C, 50 to 104°F 30 to 85% (non-condensing) 70 to 106 kPa −20 to 65°C, −4 to 149°F 10 to 95% 70 to 106 kPa Safety Standards Safety standard: 96 CAN/CSA-C22.2 No. 601-1 M90: 1990 CAN/CSA-C22.2 No. 601-1. 1S1-94: 1994 CAN/CSA-C22.2 No. 601-1. 1B-90: R2002 CAN/CSA-C22.2 No. 60601-2-49-04: 2004 CAN/CSA-C22.2 No. 601.2.27-98: 1998 CAN/CSA-C22.2 No. 60601-2-30-02: 2002 IEC 60601-1:1988 IEC 60601-1 Amendment 1: 1991 IEC 60601-1 Amendment 2: 1995 IEC 60601-1-2: 2001 IEC 60601-1-2 Amendment 1: 2004 IEC 60601-2-27: 2005 IEC 60601-2-30: 1999 IEC 60601-2-49: 2001 Operator’s Manual ZM-940PA/941PA According to the type of protection against electrical shock: INTERNALLY POWERED EQUIPMENT According to the degree of protection against electrical shock: ECG and impedance method respiration: DEFIBRILLATION-PROOF TYPE CF APPLIED PART SpO2 and NIBP: DEFIBRILLATION-PROOF TYPE BF APPLIED PART According to the degree of protection against harmful ingress of water: IPX0 (Ordinary equipment) According to the degree of safety of application in the presence of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE: Equipment not suitable for use in the presence of FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE According to the mode of operation: CONTINUOUS OPERATION Electromagnetic Compatibility IEC 60601-1-2: 2001 IEC 60601-1-2 Amendment 1: 2004 Electromagnetic Emissions This Model ZM-941PA is intended for use in the electromagnetic environment specified below. The customer or the user of the ZM-941PA should assure that it is used in such an environment. Emissions test RF emissions CISPR 11 RF emissions CISPR 11 Harmonic emissions IEC 61000-3-2 Voltage fluctuations/ flicker emissions IEC 61000-3-3 Compliance Electromagnetic environment guidance Group 1 The ZM-941PA uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. Class B The ZM-941PA is suitable for use in all establishments, including domestic establishments. Not applicable Not applicable Operator’s Manual ZM-940PA/941PA 97 Electromagnetic Immunity This Model ZM-941PA is intended for use in the electromagnetic environment specified below. The customer or the user of the ZM-941PA should assure that it is used in such an environment. Immunity test Electrostatic discharge (ESD) IEC 61000-4-2 IEC 60601 test level ±6 kV contact ±8 kV air Compliance level ±6 kV contact ±8 kV air Electrical fast transient/ burst IEC 61000-4-4 Not applicable ±2 kV for power supply lines ±1 kV for input/output lines Surge Not applicable ±1 kV differential IEC 61000-4-5 mode ±2 kV common mode Voltage dips, short <5% UT (>95% dip in Not applicable interruptions and UT) for 0.5 cycle voltage variations on power supply 40% UT (60% dip in input lines UT) for 5 cycles IEC 61000-4-11 70% UT (30% dip in UT) for 25 cycles Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 <5% UT (>95% dip in UT) for 5 s 3 A/m Electromagnetic environment - guidance Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. — — — 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. NOTE: UT is the AC mains voltage prior to application of the test level Avoiding Electromagnetic Interference (Impedance Respiration) Impedance respiration measurement is very sensitive and affected by electromagnetic interference. Technological limitations do not allow immunity levels higher than 1 V/m for radiated RF electromagnetic fields. Electromagnetic fields with field strengths above 1 V/m may cause measurement error. Do not use electrically radiating equipment near the impedance respiration measurements. 98 Operator’s Manual ZM-940PA/941PA Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Portable and mobile RF communications equipment should be used no closer to any part of the ZM-941PA, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance Conducted RF IEC 61000-4-6 3 Vrms 3 Vrms Radiated RF IEC 61000-4-3 3 V/m 3 V/m 80 MHz to 80 MHz to 2.5 GHz 2.5 GHz (1 V/m 80 MHz to 2.5 GHz for respiration) d = 1.2√P d = 1.2√P 80 MHz to 800 MHz d = 2.3√P 800 MHz to 2.5 GHz (d = 3.5√P 80 MHz to 800 MHz for respiration d = 7.0√P 800 MHz to 2.5 GHz for respiration) where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as deter mined by an electromagnetic site survey*1, should be less than the compliance level in each frequency range*2. Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. *1 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the ZM-941PA is used exceeds the applicable RF compliance level above, the ZM-941PA should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the ZM-941PA. *2 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m for respiration and 3 V/m for all other functions. Operator’s Manual ZM-940PA/941PA 99 Recommended Separation Distances between Portable and Mobile RF Communications Equipment The ZM-941PA is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the ZM-941PA can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the ZM-941PA as recommended below, according to the maximum output power of the communications. Separation distance according to frequency of transmitter (m) Rated maximum output power of transmitter (W) 150 kHz to 80 MHz d = 1.2√P 80 MHz to 800 MHz d = 1.2√P 800 MHz to 2.5 GHz d = 2.3√P (For respiration: d = 3.5√P) (For respiration: d = 7.0√P) 0.01 0.12 0.12 (0.35*) 0.23 (0.7*) 0.1 0.38 0.38 (1.1*) 0.73 (2.2*) 1.2 1.2 (3.5*) 2.3 (7.0*) 10 3.8 3.8 (11*) 7.3 (22*) 100 12 12 (35*) 23 (70*) For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. (* For respiration) NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Recovery Time after Defibrillation The transmitter returns to the normal operating mode within 10 seconds after defibrillation. The stored settings are not affected. System Composition for EMC Test The ZM-941PA bedside monitor is tested to comply with IEC 60601-1-2: 2001 and IEC 60601-1-2 Amendment 1: 2004 with the following composition. Units ZM-941PA transmitter YP-943P NIBP cuff BR-906P ECG electrode lead TL-201T finger probe 100 Cable length — 0.15 m 0.8 m 1.6 m Operator’s Manual ZM-940PA/941PA Standard Accessories No. Name 1 NIBP cuff for adult, standard 2 Strap Model YP-943P — Q’ty Supply Code No. S938B Y236 The following parts are available for replacement. Lock plate is a standard accessory of the YP943P/944P NIBP cuff. No. Name 3 Battery case cover 4 Lock plate Model — — Operator’s Manual ZM-940PA/941PA Q’ty Supply Code No. 6144-012004 6113-049585 101 Options CAUTION Only use Nihon Kohden specified electrodes, electrode leads, SpO2 probes, and NIBP cuffs. Otherwise, the maximum performance from the transmitter cannot be guaranteed. Transmitter Channel writer, QI-901PK ECG/RESP Name Electrode lead 102 Application 3 electrodes, clip type, lead length 80 cm 3 electrodes, snap type, lead length 80 cm 6 electrodes, clip type, lead length 80 cm 6 electrodes, snap type, lead length 80 cm Model Q’ty Supply Code No. BR-903PA K911A BR-913PA K910B BR-906PA K912A BR-916PA K915A Operator’s Manual ZM-940PA/941PA NIBP Width (cm) Name Cuff for adult, for attaching transmitter to patient arm Cuff for infant Cuff for child Cuff for adult Standard 13 Large 15 Small Standard Standard Large Disposable cuff for infant Disposable cuff for child Disposable cuff for adult Small Standard Medium large Large Extra large Extension hose 10 13 15 10 14 Air Hose Length (cm) 15 Model YP-943P* YP-944P* Q’ty Supply Code No. S938B S938C 17 17 17 20 YP-960T YP-961T YP-962T YP-963T YP-964T YP-810P YP-811P YP-812P YP-813P 15 20 YP-814P S946G 17 18 — 20 20 150 YP-815P YP-816P YN-990P S946H S946I S903 15 20 S943A S943B S943C S944B S944C S945C S945D S946E S946F * The lock plate is provided with these NIBP cuffs. Operator’s Manual ZM-940PA/941PA 103 SpO2 Name Finger probe (reusable) Multi-site probe (reusable) Finger probe (reusable) SpO2 probe (for adult, disposable) SpO2 probe (for child, disposable) SpO2 probe (for neonate, disposable) SpO2 probe (for adult, disposable) SpO2 probe (for child, disposable) SpO2 probe (for neonate/adult, disposable) SpO2 probe (for child/infant, disposable) SpO2 probe (for adult/neonate, disposable) SpO2 probe (for child/infant, disposable) COTTONY tape Foam tape for TL-051S/052S/ 061S/062S Attachment tape for TL-220T/ 251T/252T/253T/630T/631T Probe fastener 104 Cable length 0.6 m 1.6 m 1.6 m 0.6 m 1.6 m 0.6 m 1.6 m 1.6 m 0.8 m 1.6 m 0.8 m 1.6 m 0.8 m 1.6 m 0.8 m 1.6 m 0.8 m 1.6 m 0.8 m 1.6 m — Model/ Code No. 20 Supply Code No. P225H P225F P225G P310A P310C P311A P311C P201A P201B P201C P203A P203E P203B P203F P203C P203G P203D P203H P228A P228B P229A P229B P259 4 × 25 packages P260 3 × 30 packages P263 30 P267 Q’ty TL-201T TL-220T TL-630T1 TL-630T3 TL-631T1 TL-631T3 TL-251T TL-252T TL-253T TL-271T TL-271T3 TL-272T TL-272T3 TL-273T TL-273T3 TL-274T TL-274T3 TL-051S TL-052S TL-061S TL-062S 340703 24 — YS-093P2 Operator’s Manual ZM-940PA/941PA Transmission Frequencies Channel: 9002 to 9478 Transmission Transmission Transmission frequency Channel No. frequency Channel No. frequency Channel No. (MHz) (MHz) (MHz) 608.0250 608.0375 608.0500 608.0625 608.0750 608.0875 608.1000 608.1125 608.1250 608.1375 608.1500 608.1625 608.1750 608.1875 608.2000 608.2125 608.2250 608.2375 608.2500 608.2625 608.2750 608.2875 608.3000 608.3125 608.3250 608.3375 608.3500 608.3625 608.3750 608.3875 608.4000 608.4125 608.4250 9002 9003 9004 9005 9006 9007 9008 9009 9010 9011 9012 9013 9014 9015 9016 9017 9018 9019 9020 9021 9022 9023 9024 9025 9026 9027 9028 9029 9030 9031 9032 9033 9034 Operator’s Manual ZM-940PA/941PA 608.4375 608.4500 608.4625 608.4750 608.4875 608.5000 608.5125 608.5250 608.5375 608.5500 608.5625 608.5750 608.5875 608.6000 608.6125 608.6250 608.6375 608.6500 608.6625 608.6750 608.6875 608.7000 608.7125 608.7250 608.7375 608.7500 608.7625 608.7750 608.7875 608.8000 608.8125 608.8250 608.8375 9035 9036 9037 9038 9039 9040 9041 9042 9043 9044 9045 9046 9047 9048 9049 9050 9051 9052 9053 9054 9055 9056 9057 9058 9059 9060 9061 9062 9063 9064 9065 9066 9067 608.8500 608.8625 608.8750 608.8875 608.9000 608.9125 608.9250 608.9375 608.9500 608.9625 608.9750 608.9875 609.0000 609.0125 609.0250 609.0375 609.0500 609.0625 609.0750 609.0875 609.1000 609.1125 609.1250 609.1375 609.1500 609.1625 609.1750 609.1875 609.2000 609.2125 609.2250 609.2375 609.2500 9068 9069 9070 9071 9072 9073 9074 9075 9076 9077 9078 9079 9080 9081 9082 9083 9084 9085 9086 9087 9088 9089 9090 9091 9092 9093 9094 9095 9096 9097 9098 9099 9100 105 Transmission Transmission Transmission frequency Channel No. frequency Channel No. frequency Channel No. (MHz) (MHz) (MHz) 106 609.2625 609.2750 609.2875 609.3000 609.3125 609.3250 609.3375 609.3500 609.3625 609.3750 609.3875 609.4000 609.4125 609.4250 609.4375 609.4500 609.4625 609.4750 609.4875 609.5000 609.5125 609.5250 609.5375 609.5500 609.5625 609.5750 609.5875 609.6000 609.6125 609.6250 9101 9102 9103 9104 9105 9106 9107 9108 9109 9110 9111 9112 9113 9114 9115 9116 9117 9118 9119 9120 9121 9122 9123 9124 9125 9126 9127 9128 9129 9130 609.7125 609.7250 609.7375 609.7500 609.7625 609.7750 609.7875 609.8000 609.8125 609.8250 609.8375 609.8500 609.8625 609.8750 609.8875 609.9000 609.9125 609.9250 609.9375 609.9500 609.9625 609.9750 609.9875 610.0000 610.0125 610.0250 610.0375 610.0500 610.0625 610.0750 9137 9138 9139 9140 9141 9142 9143 9144 9145 9146 9147 9148 9149 9150 9151 9152 9153 9154 9155 9156 9157 9158 9159 9160 9161 9162 9163 9164 9165 9166 610.1625 610.1750 610.1875 610.2000 610.2125 610.2250 610.2375 610.2500 610.2625 610.2750 610.2875 610.3000 610.3125 610.3250 610.3375 610.3500 610.3625 610.3750 610.3875 610.4000 610.4125 610.4250 610.4375 610.4500 610.4625 610.4750 610.4875 610.5000 610.5125 610.5250 9173 9174 9175 9176 9177 9178 9179 9180 9181 9182 9183 9184 9185 9186 9187 9188 9189 9190 9191 9192 9193 9194 9195 9196 9197 9198 9199 9200 9201 9202 609.6375 9131 610.0875 9167 610.5375 9203 609.6500 9132 610.1000 9168 610.5500 9204 609.6625 9133 610.1125 9169 610.5625 9205 609.6750 9134 610.1250 9170 610.5750 9206 609.6875 9135 610.1375 9171 610.5875 9207 Operator’s Manual ZM-940PA/941PA Transmission Transmission Transmission frequency Channel No. frequency Channel No. frequency Channel No. (MHz) (MHz) (MHz) 609.7000 9136 610.1500 9172 610.6000 9208 610.6125 9209 611.0625 9245 611.5125 9281 610.6250 9210 611.0750 9246 611.5250 9282 610.6375 610.6500 610.6625 610.6750 610.6875 610.7000 610.7125 610.7250 610.7375 610.7500 610.7625 610.7750 610.7875 610.8000 610.8125 610.8250 610.8375 610.8500 610.8625 610.8750 610.8875 610.9000 610.9125 610.9250 610.9375 610.9500 610.9625 610.9750 610.9875 611.0000 611.0125 611.0250 611.0375 9211 9212 9213 9214 9215 9216 9217 9218 9219 9220 9221 9222 9223 9224 9225 9226 9227 9228 9229 9230 9231 9232 9233 9234 9235 9236 9237 9238 9239 9240 9241 9242 9243 611.0875 611.1000 611.1125 611.1250 611.1375 611.1500 611.1625 611.1750 611.1875 611.2000 611.2125 611.2250 611.2375 611.2500 611.2625 611.2750 611.2875 611.3000 611.3125 611.3250 611.3375 611.3500 611.3625 611.3750 611.3875 611.4000 611.4125 611.4250 611.4375 611.4500 611.4625 611.4750 611.4875 9247 9248 9249 9250 9251 9252 9253 9254 9255 9256 9257 9258 9259 9260 9261 9262 9263 9264 9265 9266 9267 9268 9269 9270 9271 9272 9273 9274 9275 9276 9277 9278 9279 611.5375 611.5500 611.5625 611.5750 611.5875 611.6000 611.6125 611.6250 611.6375 611.6500 611.6625 611.6750 611.6875 611.7000 611.7125 611.7250 611.7375 611.7500 611.7625 611.7750 611.7875 611.8000 611.8125 611.8250 611.8375 611.8500 611.8625 611.8750 611.8875 611.9000 611.9125 611.9250 611.9375 9283 9284 9285 9286 9287 9288 9289 9290 9291 9292 9293 9294 9295 9296 9297 9298 9299 9300 9301 9302 9303 9304 9305 9306 9307 9308 9309 9310 9311 9312 9313 9314 9315 Operator’s Manual ZM-940PA/941PA 107 Transmission Transmission Transmission frequency Channel No. frequency Channel No. frequency Channel No. (MHz) (MHz) (MHz) 108 611.0500 9244 611.5000 9280 611.9500 9316 611.9625 611.9750 611.9875 612.0000 612.0125 612.0250 612.0375 612.0500 612.0625 612.0750 612.0875 612.1000 612.1125 612.1250 612.1375 612.1500 612.1625 612.1750 612.1875 612.2000 612.2125 612.2250 612.2375 612.2500 612.2625 612.2750 612.2875 612.3000 612.3125 612.3250 612.3375 612.3500 612.3625 612.3750 612.3875 9317 9318 9319 9320 9321 9322 9323 9324 9325 9326 9327 9328 9329 9330 9331 9332 9333 9334 9335 9336 9337 9338 9339 9340 9341 9342 9343 9344 9345 9346 9347 9348 9349 9350 9351 612.4125 612.4250 612.4375 612.4500 612.4625 612.4750 612.4875 612.5000 612.5125 612.5250 612.5375 612.5500 612.5625 612.5750 612.5875 612.6000 612.6125 612.6250 612.6375 612.6500 612.6625 612.6750 612.6875 612.7000 612.7125 612.7250 612.7375 612.7500 612.7625 612.7750 612.7875 612.8000 612.8125 612.8250 612.8375 9353 9354 9355 9356 9357 9358 9359 9360 9361 9362 9363 9364 9365 9366 9367 9368 9369 9370 9371 9372 9373 9374 9375 9376 9377 9378 9379 9380 9381 9382 9383 9384 9385 9386 9387 612.8625 612.8750 612.8875 612.9000 612.9125 612.9250 612.9375 612.9500 612.9625 612.9750 612.9875 613.0000 613.0125 613.0250 613.0375 613.0500 613.0625 613.0750 613.0875 613.1000 613.1125 613.1250 613.1375 613.1500 613.1625 613.1750 613.1875 613.2000 613.2125 613.2250 613.2375 613.2500 613.2625 613.2750 613.2875 9389 9390 9391 9392 9393 9394 9395 9396 9397 9398 9399 9400 9401 9402 9403 9404 9405 9406 9407 9408 9409 9410 9411 9412 9413 9414 9415 9416 9417 9418 9419 9420 9421 9422 9423 Operator’s Manual ZM-940PA/941PA Transmission Transmission Transmission frequency Channel No. frequency Channel No. frequency Channel No. (MHz) (MHz) (MHz) 612.4000 9352 612.8500 9388 613.3000 9424 613.3125 9425 613.5375 9443 613.7625 9461 613.3250 9426 613.5500 9444 613.7750 9462 613.3375 9427 613.5625 9445 613.7875 9463 613.3500 9428 613.5750 9446 613.8000 9464 613.3625 9429 613.5875 9447 613.8125 9465 613.3750 9430 613.6000 9448 613.8250 9466 613.3875 9431 613.6125 9449 613.8375 9467 613.4000 9432 613.6250 9450 613.8500 9468 613.4125 9433 613.6375 9451 613.8625 9469 613.4250 9434 613.6500 9452 613.8750 9470 613.4375 9435 613.6625 9453 613.8875 9471 613.4500 9436 613.6750 9454 613.9000 9472 613.4625 9437 613.6875 9455 613.9125 9473 613.4750 9438 613.7000 9456 613.9250 9474 613.4875 9439 613.7125 9457 613.9375 9475 613.5000 9440 613.7250 9458 613.9500 9476 613.5125 9441 613.7375 9459 613.9625 9477 613.5250 9442 613.7500 9460 613.9750 9478 Operator’s Manual ZM-940PA/941PA 109 Channel: E002 to E398 Transmission frequency (MHz) 1395.0250 1395.0375 1395.0500 1395.0625 1395.0750 1395.0875 1395.1000 1395.1125 1395.1250 1395.1375 1395.1500 1395.1625 1395.1750 1395.1875 1395.2000 1395.2125 1395.2250 1395.2375 1395.2500 1395.2625 1395.2750 1395.2875 1395.3000 1395.3125 1395.3250 1395.3375 1395.3500 1395.3625 1395.3750 1395.3875 1395.4000 1395.4125 1395.4250 1395.4375 1395.4500 110 Channel No. E002 E003 E004 E005 E006 E007 E008 E009 E010 E011 E012 E013 E014 E015 E016 E017 E018 E019 E020 E021 E022 E023 E024 E025 E026 E027 E028 E029 E030 E031 E032 E033 E034 E035 E036 Transmission frequency (MHz) 1395.4625 Channel No. E037 1395.4750 1395.4875 1395.5000 1395.5125 1395.5250 1395.5375 1395.5500 E038 E039 E040 E041 E042 E043 E044 1395.5625 1395.5750 1395.5875 1395.6000 1395.6125 1395.6250 1395.6375 1395.6500 1395.6625 1395.6750 1395.6875 1395.7000 1395.7125 1395.7250 1395.7375 1395.7500 1395.7625 1395.7750 1395.7875 1395.8000 1395.8125 1395.8250 1395.8375 1395.8500 1395.8625 1395.8750 1395.8875 E045 E046 E047 E048 E049 E050 E051 E052 E053 E054 E055 E056 E057 E058 E059 E060 E061 E062 E063 E064 E065 E066 E067 E068 E069 E070 E071 Transmission frequency (MHz) Channel No. 1395.9000 1395.9125 1395.9250 1395.9375 1395.9500 1395.9625 1395.9750 1395.9875 1396.0000 1396.0125 1396.0250 1396.0375 1396.0500 1396.0625 1396.0750 1396.0875 1396.1000 1396.1125 1396.1250 1396.1375 1396.1500 1396.1625 1396.1750 1396.1875 1396.2000 1396.2125 1396.2250 1396.2375 1396.2500 1396.2625 1396.2750 1396.2875 1396.3000 1396.3125 1396.3250 E072 E073 E074 E075 E076 E077 E078 E079 E080 E081 E082 E083 E084 E085 E086 E087 E088 E089 E090 E091 E092 E093 E094 E095 E096 E097 E098 E099 E100 E101 E102 E103 E104 E105 E106 Operator’s Manual ZM-940PA/941PA Transmission frequency (MHz) 1396.3375 1396.3500 1396.3625 1396.3750 1396.3875 1396.4000 1396.4125 1396.4250 1396.4375 1396.4500 1396.4625 1396.4750 1396.4875 1396.5000 1396.5125 1396.5250 1396.5375 1396.5500 1396.5625 1396.5750 1396.5875 1396.6000 1396.6125 1396.6250 1396.6375 1396.6500 1396.6625 1396.6750 1396.6875 1396.7000 1396.7125 1396.7250 1396.7375 1396.7500 1396.7625 1396.7750 E107 E108 Transmission frequency (MHz) 1396.7875 1396.8000 E109 E110 E111 E112 E113 E114 E115 E116 E117 E118 E119 E120 E121 E122 E123 E124 E125 E126 E127 E128 E129 E130 E131 E132 E133 E134 E135 E136 E137 E138 E139 E140 E141 E142 1396.8125 1396.8250 1396.8375 1396.8500 1396.8625 1396.8750 1396.8875 1396.9000 1396.9125 1396.9250 1396.9375 1396.9500 1396.9625 1396.9750 1396.9875 1397.0000 1397.0125 1397.0250 1397.0375 1397.0500 1397.0625 1397.0750 1397.0875 1397.1000 1397.1125 1397.1250 1397.1375 1397.1500 1397.1625 1397.1750 1397.1875 1397.2000 1397.2125 1397.2250 Channel No. Operator’s Manual ZM-940PA/941PA Channel No. E143 E144 E145 E146 E147 E148 E149 E150 E151 E152 E153 E154 E155 E156 E157 E158 E159 E160 E161 E162 E163 E164 E165 E166 E167 E168 E169 E170 E171 E172 E173 E174 E175 E176 E177 E178 Transmission frequency (MHz) 1397.2375 1397.2500 Channel No. E179 E180 1397.2625 1397.2750 1397.2875 1397.3000 1397.3125 1397.3250 1397.3375 1397.3500 1397.3625 1397.3750 1397.3875 1397.4000 1397.4125 1397.4250 1397.4375 1397.4500 1397.4625 1397.4750 1397.4875 1397.5000 1397.5125 1397.5250 1397.5375 1397.5500 1397.5625 1397.5750 1397.5875 1397.6000 E181 E182 E183 E184 E185 E186 E187 E188 E189 E190 E191 E192 E193 E194 E195 E196 E197 E198 E199 E200 E201 E202 E203 E204 E205 E206 E207 E208 1397.6125 1397.6250 1397.6375 1397.6500 1397.6625 1397.6750 E209 E210 E211 E212 E213 E214 111 Transmission frequency (MHz) Channel No. Transmission frequency (MHz) Channel No. 1397.6875 1397.7000 1397.7125 1397.7250 1397.7375 1397.7500 1397.7625 1397.7750 1397.7875 1397.8000 1397.8125 1397.8250 1397.8375 1397.8500 1397.8625 1397.8750 1397.8875 1397.9000 1397.9125 1397.9250 1397.9375 1397.9500 1397.9625 1397.9750 1397.9875 1398.0000 1398.0125 1398.0250 1398.0375 1398.0500 1398.0625 1398.0750 1398.0875 E215 E216 E217 E218 E219 E220 E221 E222 E223 E224 E225 E226 E227 E228 E229 E230 E231 E232 E233 E234 E235 E236 E237 E238 E239 E240 E241 E242 E243 E244 E245 E246 E247 1398.1375 1398.1500 1398.1625 1398.1750 1398.1875 1398.2000 1398.2125 1398.2250 1398.2375 1398.2500 1398.2625 1398.2750 1398.2875 1398.3000 1398.3125 1398.3250 1398.3375 1398.3500 1398.3625 1398.3750 1398.3875 1398.4000 1398.4125 1398.4250 1398.4375 1398.4500 1398.4625 1398.4750 1398.4875 1398.5000 1398.5125 1398.5250 1398.5375 E251 E252 E253 E254 E255 E256 E257 E258 E259 E260 E261 E262 E263 E264 E265 E266 E267 E268 E269 E270 E271 E272 E273 E274 E275 E276 E277 E278 E279 E280 E281 E282 E283 1398.1000 1398.1125 1398.1250 E248 E249 E250 1398.5500 1398.5625 1398.5750 E284 E285 E286 112 Transmission frequency (MHz) Channel No. 1398.5875 1398.6000 1398.6125 1398.6250 1398.6375 1398.6500 1398.6625 1398.6750 1398.6875 1398.7000 1398.7125 1398.7250 1398.7375 1398.7500 1398.7625 1398.7750 1398.7875 1398.8000 1398.8125 1398.8250 1398.8375 1398.8500 1398.8625 1398.8750 1398.8875 1398.9000 1398.9125 1398.9250 1398.9375 1398.9500 1398.9625 1398.9750 1398.9875 1399.0000 1399.0125 1399.0250 E287 E288 E289 E290 E291 E292 E293 E294 E295 E296 E297 E298 E299 E300 E301 E302 E303 E304 E305 E306 E307 E308 E309 E310 E311 E312 E313 E314 E315 E316 E317 E318 E319 E320 E321 E322 Operator’s Manual ZM-940PA/941PA Transmission frequency (MHz) 1399.0375 1399.0500 1399.0625 1399.0750 1399.0875 1399.1000 1399.1125 1399.1250 1399.1375 1399.1500 1399.1625 1399.1750 1399.1875 1399.2000 1399.2125 1399.2250 1399.2375 1399.2500 1399.2625 1399.2750 1399.2875 1399.3000 1399.3125 1399.3250 1399.3375 1399.3500 1399.3625 1399.3750 1399.3875 1399.4000 1399.4125 1399.4250 1399.4375 1399.4500 E323 E324 Transmission frequency (MHz) 1399.4625 1399.4750 E325 E326 E327 E328 E329 E330 E331 E332 E333 E334 E335 E336 E337 E338 E339 E340 E341 E342 E343 E344 E345 E346 E347 E348 E349 E350 E351 E352 E353 E354 E355 E356 1399.4875 1399.5000 1399.5125 1399.5250 1399.5375 1399.5500 1399.5625 1399.5750 1399.5875 1399.6000 1399.6125 1399.6250 1399.6375 1399.6500 1399.6625 1399.6750 1399.6875 1399.7000 1399.7125 1399.7250 1399.7375 1399.7500 1399.7625 1399.7750 1399.7875 1399.8000 1399.8125 1399.8250 1399.8375 1399.8500 1399.8625 1399.8750 Channel No. Operator’s Manual ZM-940PA/941PA Channel No. E357 E358 E359 E360 E361 E362 E363 E364 E365 E366 E367 E368 E369 E370 E371 E372 E373 E374 E375 E376 E377 E378 E379 E380 E381 E382 E383 E384 E385 E386 E387 E388 E389 E390 Transmission frequency (MHz) 1399.8875 1399.9000 1399.9125 1399.9250 1399.9375 1399.9500 1399.9625 1399.9750 Channel No. E391 E392 E393 E394 E395 E396 E397 E398 113 Channel: E502 to E898 Transmission frequency (MHz) 1427.0250 1427.0375 1427.0500 1427.0625 1427.0750 1427.0875 1427.1000 1427.1125 1427.1250 1427.1375 1427.1500 1427.1625 1427.1750 1427.1875 1427.2000 1427.2125 1427.2250 1427.2375 1427.2500 1427.2625 1427.2750 1427.2875 1427.3000 1427.3125 1427.3250 1427.3375 1427.3500 1427.3625 1427.3750 1427.3875 1427.4000 1427.4125 1427.4250 1427.4375 114 E502 E503 E504 E505 E506 E507 E508 E509 E510 E511 E512 E513 E514 E515 E516 E517 E518 E519 E520 E521 E522 E523 E524 E525 E526 E527 E528 E529 E530 E531 E532 E533 E534 Transmission frequency (MHz) 1427.4500 1427.4625 1427.4750 1427.4875 1427.5000 1427.5125 1427.5250 1427.5375 1427.5500 1427.5625 1427.5750 1427.5875 1427.6000 1427.6125 1427.6250 1427.6375 1427.6500 1427.6625 1427.6750 1427.6875 1427.7000 1427.7125 1427.7250 1427.7375 1427.7500 1427.7625 1427.7750 1427.7875 1427.8000 1427.8125 1427.8250 1427.8375 1427.8500 E535 1427.8625 Channel No. E536 E537 E538 E539 E540 E541 E542 E543 E544 E545 E546 E547 E548 E549 E550 E551 E552 E553 E554 E555 E556 E557 E558 E559 E560 E561 E562 E563 E564 E565 E566 E567 E568 Transmission frequency (MHz) 1427.8750 1427.8875 1427.9000 1427.9125 1427.9250 1427.9375 1427.9500 1427.9625 1427.9750 1427.9875 1428.0000 1428.0125 1428.0250 1428.0375 1428.0500 1428.0625 1428.0750 1428.0875 1428.1000 1428.1125 1428.1250 1428.1375 1428.1500 1428.1625 1428.1750 1428.1875 1428.2000 1428.2125 1428.2250 1428.2375 1428.2500 1428.2625 1428.2750 E569 1428.2875 Channel No. Channel No. E570 E571 E572 E573 E574 E575 E576 E577 E578 E579 E580 E581 E582 E583 E584 E585 E586 E587 E588 E589 E590 E591 E592 E593 E594 E595 E596 E597 E598 E599 E600 E601 E602 E603 Operator’s Manual ZM-940PA/941PA Transmission frequency (MHz) 1428.3000 1428.3125 1428.3250 1428.3375 1428.3500 1428.3625 1428.3750 1428.3875 1428.4000 1428.4125 1428.4250 1428.4375 1428.4500 1428.4625 1428.4750 1428.4875 1428.5000 1428.5125 1428.5250 1428.5375 1428.5500 1428.5625 1428.5750 1428.5875 1428.6000 1428.6125 1428.6250 1428.6375 1428.6500 1428.6625 1428.6750 1428.6875 1428.7000 1428.7125 1428.7250 1428.7375 E604 Transmission frequency (MHz) 1428.7500 E605 E606 E607 E608 E609 E610 E611 E612 E613 E614 E615 E616 E617 E618 E619 E620 E621 E622 E623 E624 E625 E626 E627 E628 E629 E630 E631 E632 E633 E634 E635 E636 E637 E638 E639 1428.7625 1428.7750 1428.7875 1428.8000 1428.8125 1428.8250 1428.8375 1428.8500 1428.8625 1428.8750 1428.8875 1428.9000 1428.9125 1428.9250 1428.9375 1428.9500 1428.9625 1428.9750 1428.9875 1429.0000 1429.0125 1429.0250 1429.0375 1429.0500 1429.0625 1429.0750 1429.0875 1429.1000 1429.1125 1429.1250 1429.1375 1429.1500 1429.1625 1429.1750 1429.1875 Channel No. Operator’s Manual ZM-940PA/941PA E640 Transmission frequency (MHz) 1429.2000 E641 E642 E643 E644 E645 E646 E647 E648 E649 E650 E651 E652 E653 E654 E655 E656 E657 E658 E659 E660 E661 E662 E663 E664 E665 E666 E667 E668 E669 E670 E671 E672 E673 E674 E675 1429.2125 1429.2250 1429.2375 1429.2500 1429.2625 1429.2750 1429.2875 1429.3000 1429.3125 1429.3250 1429.3375 1429.3500 1429.3625 1429.3750 1429.3875 1429.4000 1429.4125 1429.4250 1429.4375 1429.4500 1429.4625 1429.4750 1429.4875 1429.5000 1429.5125 1429.5250 1429.5375 1429.5500 1429.5625 1429.5750 1429.5875 1429.6000 1429.6125 1429.6250 1429.6375 Channel No. Channel No. E676 E677 E678 E679 E680 E681 E682 E683 E684 E685 E686 E687 E688 E689 E690 E691 E692 E693 E694 E695 E696 E697 E698 E699 E700 E701 E702 E703 E704 E705 E706 E707 E708 E709 E710 E711 115 Transmission frequency (MHz) 1429.6500 1429.6625 1429.6750 1429.6875 1429.7000 1429.7125 1429.7250 1429.7375 1429.7500 1429.7625 1429.7750 1429.7875 1429.8000 1429.8125 1429.8250 1429.8375 1429.8500 1429.8625 1429.8750 1429.8875 1429.9000 1429.9125 1429.9250 1429.9375 1429.9500 1429.9625 1429.9750 1429.9875 1430.0000 1430.0125 1430.0250 1430.0375 1430.0500 1430.0625 1430.0750 1430.0875 116 E712 E713 Transmission frequency (MHz) 1430.1000 1430.1125 E714 E715 E716 E717 E718 E719 E720 E721 E722 E723 E724 E725 E726 E727 E728 E729 E730 E731 E732 E733 E734 E735 E736 E737 E738 E739 E740 E741 E742 E743 E744 E745 E746 E747 1430.1250 1430.1375 1430.1500 1430.1625 1430.1750 1430.1875 1430.2000 1430.2125 1430.2250 1430.2375 1430.2500 1430.2625 1430.2750 1430.2875 1430.3000 1430.3125 1430.3250 1430.3375 1430.3500 1430.3625 1430.3750 1430.3875 1430.4000 1430.4125 1430.4250 1430.4375 1430.4500 1430.4625 1430.4750 1430.4875 1430.5000 1430.5125 1430.5250 1430.5375 Channel No. E748 E749 Transmission frequency (MHz) 1430.5500 1430.5625 E750 E751 E752 E753 E754 E755 E756 E757 E758 E759 E760 E761 E762 E763 E764 E765 E766 E767 E768 E769 E770 E771 E772 E773 E774 E775 E776 E777 E778 E779 E780 E781 E782 E783 1430.5750 1430.5875 1430.6000 1430.6125 1430.6250 1430.6375 1430.6500 1430.6625 1430.6750 1430.6875 1430.7000 1430.7125 1430.7250 1430.7375 1430.7500 1430.7625 1430.7750 1430.7875 1430.8000 1430.8125 1430.8250 1430.8375 1430.8500 1430.8625 1430.8750 1430.8875 1430.9000 1430.9125 1430.9250 1430.9375 1430.9500 1430.9625 1430.9750 1430.9875 Channel No. Channel No. E784 E785 E786 E787 E788 E789 E790 E791 E792 E793 E794 E795 E796 E797 E798 E799 E800 E801 E802 E803 E804 E805 E806 E807 E808 E809 E810 E811 E812 E813 E814 E815 E816 E817 E818 E819 Operator’s Manual ZM-940PA/941PA Transmission frequency (MHz) 1431.0000 1431.0125 1431.0250 1431.0375 1431.0500 1431.0625 1431.0750 1431.0875 1431.1000 1431.1125 1431.1250 1431.1375 1431.1500 1431.1625 1431.1750 1431.1875 1431.2000 1431.2125 1431.2250 1431.2375 1431.2500 1431.2625 1431.2750 1431.2875 1431.3000 1431.3125 1431.3250 1431.3375 1431.3500 1431.3625 1431.3750 1431.3875 1431.4000 1431.4125 1431.4250 E820 E821 Transmission frequency (MHz) 1431.4375 1431.4500 E822 E823 E824 E825 E826 E827 E828 E829 E830 E831 E832 E833 E834 E835 E836 E837 E838 E839 E840 E841 E842 E843 E844 E845 E846 E847 E848 E849 E850 E851 E852 E853 E854 1431.4625 1431.4750 1431.4875 1431.5000 1431.5125 1431.5250 1431.5375 1431.5500 1431.5625 1431.5750 1431.5875 1431.6000 1431.6125 1431.6250 1431.6375 1431.6500 1431.6625 1431.6750 1431.6875 1431.7000 1431.7125 1431.7250 1431.7375 1431.7500 1431.7625 1431.7750 1431.7875 1431.8000 1431.8125 1431.8250 1431.8375 1431.8500 1431.8625 Channel No. Operator’s Manual ZM-940PA/941PA Channel No. E855 E856 E857 E858 E859 E860 E861 E862 E863 E864 E865 E866 E867 E868 E869 E870 E871 E872 E873 E874 E875 E876 E877 E878 E879 E880 E881 E882 E883 E884 E885 E886 E887 E888 E889 Transmission frequency (MHz) 1431.8750 1431.8875 1431.9000 1431.9125 1431.9250 1431.9375 1431.9500 1431.9625 1431.9750 Channel No. E890 E891 E892 E893 E894 E895 E896 E897 E898 117
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.4 Linearized : No Modify Date : 2008:11:17 13:57:03-08:00 Create Date : 2008:11:17 13:56:44-08:00 Creator : Adobe InDesign CS2_J (4.0.5) Producer : Adobe PDF Library 7.0 Page Count : 77 Mod Date : 2008:11:17 13:57:03-08:00 Creation Date : 2008:11:17 13:56:44-08:00 Metadata Date : 2008:11:17 13:57:03-08:00EXIF Metadata provided by EXIF.tools