Nokia Solutions and Networks T5GX1 UBS CDMA XMI Transceiver at 800 MHz User Manual Exhibit 123d
Nokia Solutions and Networks UBS CDMA XMI Transceiver at 800 MHz Exhibit 123d
Contents
Exhibit 123d
1X UBS Macro BTS Optimization/ATP Acceptance Test Procedures - TX & RX Perform the procedure in Procedure 4-6 for all-inclusive RX test. Procedure 4-6 All RX ATP Test Set up the test equipment for abbreviated acceptance tests per Procedure 4-3 On LMF, select devices to be tested. To select multiple items, hold down the Shift or Ctrl key while making the selections. Click on Tests in the UBS Menu bar, and select All RX Test... from pull-down menu. Select the appropriate carrier from those displayed in the Channels/Carrier pick list. Select the appropriate RX Branch (Both, Main, or Diversity) from the drop-down menu. In the Rate Set box, select the appropriate data rate (1=9600, 2= 14400, 3 = 9600 1X) from the drop-down list. Click OK to display a status bar followed by a Directions pop-up window. Follow the cable connection directions as they are displayed, and click the Continue button to begin testing. As the ATP process is completed, results will be displayed in a status report window. Click the Save Results or Dismiss button. If Dismiss is used, the test results will not be saved in the test report file. 68P09283A63-5 4-13 FOA AUG 2007 Individual Tests Chapter 4: Acceptance Test Procedures Individual Tests ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ TX and RX Testing The following individual ATP tests can be used to evaluate specific aspects of UBS operation against individual performance requirements. All testing is performed using the LMF GUI environment. TX Testing TX tests verify transmit antenna paths and output power control. All tests are performed using the external, calibrated test equipment. All measurements are made at the appropriate UBS TX OUT connector(s). TX tests verify TX operation of the entire CDMA forward link using selected sector-carriers assigned to respective sector antennas. Each sector-carrier is keyed up to generate a CDMA carrier (using BLO) at the NEC file-specified carrier output power level. RX Testing RX testing verifies receive antenna paths for the sector-carriers selected for the test. All tests are performed using the external, calibrated test equipment to inject a CDMA RF carrier with all zero long code at the specified RX frequency at the appropriate UBS RX IN connector(s). RX tests verify RX operation of the entire CDMA reverse link. Individual Tests Standard Pattern (per the 3GPP2 standard) - For UBS Macro, the standard pattern is the total digital gain of the signal (383) that includes the following channels (9 in all) and their individual digital gain: • Pilot -171 • Page - 165 • Sync - 83 • 6xTCH - 117 In the UBS Macro, the pilot is set to 36dBm, which brings the total power of this pattern to 43dBm - 7dB higher. Spectral Purity TX Mask This test verifies that the transmitted CDMA carrier waveform generated on each sector meets the transmit spectral mask specification with respect to the assigned NEC file values. 4-14 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Individual Tests Waveform Quality (Rho) This test verifies the transmitted Pilot channel element digital waveform quality of each sector-carrier enabled at a specific frequency per the current NEC file assignment. All tests are performed using the external calibrated test set controlled by the same command. All measurements are via the appropriate TX OUT connector. Pilot Time Offset The Pilot Time Offset is the difference between the CDMA analyzer measurement interval (based on the UBS system time reference) and the incoming block of transmitted data from the UBS. Code Domain Power/Noise Floor This test verifies the Code Domain Power/Noise of each sector-carrier is enabled at a specific frequency per the current NEC file assignment. All tests are performed using the external calibrated test set controlled by the same command. All measurements are via the appropriate TX OUT connector. Frame Error Rate (FER) The Frame Error Rate (FER) test verifies RX operation of the entire CDMA Reverse Link using all equipped DMIs assigned to all respective sectors/antennas. This test verifies the UBS sensitivity on all traffic channel elements currently configured on all equipped DMIs at an RF input level of -123.0 dBm. TX Audit The TX Audit is a power out test that will pass within +/-2dB of the expected power. The standard power per S-C is 43dBm and the Audit test will pass, if the output power is within the limits listed above. Calibration It is optional, but supported. The calibration in UBS is always a negative number representing the LOSS between the output of the XMI to the top of the frame. In this frame, it is typically -1dB to –2dB depending on RF elements in the frame. 68P09283A63-5 4-15 FOA AUG 2007 Individual Tests Chapter 4: Acceptance Test Procedures Receive Signal Strength Indication (RSSI) This test verifies the gain for UBS reverse (RX) paths for each carrier-sector selected is within requirements for correct operation for the operating band of the UBS as follows: • 1900 MHz: –80 dBm (-86 dBm or greater) • 800 MHz: –80 dBm (-86 dBm or greater) The LMF injects a -80 dBm signal (default) input to the UBS. The RSSI must be +/-6 dB. 4-16 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP TX Spectral Purity Transmit Mask Acceptance Test TX Spectral Purity Transmit Mask Acceptance Test ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ TX Mask Test This test verifies the spectral purity of each sector-carrier at a specific frequency, per the current NEC file assignment. All tests are performed using the external calibrated test set, controlled by the same command. All measurements are via the appropriate TX OUT connector. The standard pattern is used (pilot 171 and 43dBm) The parameters of the TX Mask are as follows: > -45.0 dBc / 30 kHz@>750 kHz offset > -60.0 dBc / 30 kHz@ 1.98 MHz offset The sector-carrier is keyed or dekeyed during the test. Figure 4-3 TX Mask Verication Spectrum Analyzer Display (1900 MHz) Mean CDMA Bandwidth Power Reference .5 MHz Span/Div Ampl 10 dB/Div Center Frequency Reference Attenuation level of all spurious and IM products with respect to the mean power of the CDMA channel +1.98 MHz - 1.98 MHz - 900 kHz - 750 kHz +750 kHz ti-cdma -00150-A.e ps 68P09283A63-5 4-17 FOA AUG 2007 TX Waveform Quality (Rho) Acceptance Test Chapter 4: Acceptance Test Procedures TX Waveform Quality (Rho) Acceptance Test ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Rho Test The Pilot Gain is set to 171 for each antenna. The sector-carrier is enabled using both rflvl and bay level offsets, to generate a CDMA carrier (with pilot channel element only, Walsh code 0). Sector-carrier power output is set to 36 dBm as measured at the TX OUT connector (on the UBS directional coupler). The calibrated communications test set measures and returns the Pilot channel element digital waveform quality (rho) in dB, verifying that the result meets system tolerances: • Waveform quality (rho) should be greater than or equal to 0.97 (-0.4 dB). Figure 4-4 Rho Signal ti-cdma-rho_p to.eps 4-18 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP TX Pilot Time Offset Acceptance Test TX Pilot Time Offset Acceptance Test ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ PTO Acceptance Test This test verifies the transmitted Pilot channel element Pilot Time Offset of each sector-carrier keyed up at a specific frequency per the current NEC file assignment. All tests are performed using the external calibrated test set controlled by the same command. All measurements are via the appropriate TX OUT connector. The Pilot Gain is set to 171 LSB for each antenna. The sector-carrier is enabled, using both rfLvl and bay level offsets, to generate a CDMA carrier (with pilot channel element only, Walsh code 0). Sector-carrier power output is set to 36 dBm as measured at the TX OUT connector (on the UBS directional coupler). The calibrated communications test set measures and returns the Pilot Time Offset in us, verifying results meet system tolerances: • Pilot Time Offset should be within +/– 3 us of the target PT Offset (0 us). 68P09283A63-5 4-19 FOA AUG 2007 TX Pilot Time Offset Acceptance Test Figure 4-5 Chapter 4: Acceptance Test Procedures Pilot Only Signal ti-cdma-pilot_only.eps 4-20 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP TX Code Domain Power/Noise Floor Acceptance Test TX Code Domain Power/Noise Floor Acceptance Test ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Code Domain Power Test For each sector/antenna under test, the Pilot Gain is set to 383 LSB. All channel elements under test are configured to generate Orthogonal Channel Noise Source (OCNS) on standard pattern only 9 channels where the 6 Traffic channels (TCH) can be anywhere across the code domain The maximum number of channel elements (CEs) to be tested at any one time is 32 (32 odd Walsh codes). If more than 32 CEs exist, then multiple sets of measurements are made; so all channel elements are verified on all sectors. Sector-carrier power output is set to 43 dBm as measured at the TX OUT connector. • CD_PILOT_RATIO: -6.5dB to -7.5dB • CD_PAGING_RATIO:-6.8dB to -7.8dB • CD_SYNC_RATIO: -12.8dB to -13.8dB • CD_TRAFFIC_ON_MAX_RATIO: -9.8dB to -10.8dB • CD_TRAFFIC_ON_MIN_RATIO: -9.8dB to -10.8dB • CD_TRAFFIC_OFF_MAX_RATIO: -27dB (none) Refer to Figure 4-6. If using Advantest test equipment, Code Domain Test MUST be configured in RC-1 mode. See Procedure 4-5 to perform this test. 68P09283A63-5 4-21 FOA AUG 2007 TX Code Domain Power/Noise Floor Acceptance Test Chapter 4: Acceptance Test Procedures Figure 4-6 Code Domain Power and Noise Floor Levels PILOT PAGING TRAFFIC SYNC TRAFFIC ti-cdma-standa rd_pa ttern.eps 4-22 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP RX Frame Error Rate (FER) Acceptance Test RX Frame Error Rate (FER) Acceptance Test ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ FER Test This test verifies the UBS FER on all traffic channel elements is currently configured (full rate at 1% FER) at an RF input level of -123 dBm. All tests are performed using the external calibrated test set as the signal source controlled by the same command. All measurements are via the LMF. The sector-carrier is enabled, using only rflvl level offsets, to generate a CDMA carrier (with pilot channel element only). Sector-carrier power output is set to +27.5 dBm as measured at the TX OUT connector. The UBS must be keyed to enable the RX receive circuitry. The LMF prompts the CE under test to measure all zero long code and provide the FER report on the selected reverse link for both the main and diversity RX antenna paths, verifying that results meet the following specification: • FER returned <1% @ –123 dBm and total frames measured is 1500 All CEs selected are tested on the specified RX antenna path. See Procedure 4-6 to perform this test. 68P09283A63-5 4-23 FOA AUG 2007 Continuous Waveform Mode Chapter 4: Acceptance Test Procedures Continuous Waveform Mode ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Objective This test verifies that the UBS frames can generate an unmodulated carrier tone on a single sector at any single channel number within the supported carrier frequency range of the frame. Unlock Continuous Waveform Mode WinLMF must be connected to the UBS frame to be tested and logged in. Procedure 4-7 Procedure to Unlock Continuous Waveform Mode On the menu, click on UBS -> Unlock Continuous Waveform Mode Click the Yes button in the confirmation dialog box. A status report window will appear that shows the result of the action request. WinLMF will fail to set up Continuous Waveform Mode for Modem devices provisioned in the conguration les on DMI, but not physically installed in the frame, and for Modem devices under reset. Select tested Sector/Carrier and click Device->XCVR->Key menu item. In Key window: Enter the desired power level [dBm] in the XCVR Gain box Enter the desired channel in the XCVR Channel box Click the OK button in Key window. A status report window will appear that shows the result of the action request. Perform the required testing on the selected Sector/Carrier. Select Sector/Carrier keyed in step 4 and click Device->XCVR->Dekey menu item. A status report window will appear that shows the result of the action request. To perform unmodulated carrier mode test on other Sector/Carrier repeat procedure from step 4. If testing using unmodulated carrier mode is completed follow Lock Continuous Waveform Mode Procedure 4-24 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Continuous Waveform Mode Lock Continuous Waveform Mode Procedure 4-8 Procedure to Lock Continuous Waveform Mode On the menu, click on UBS -> Lock Continuous Waveform Mode A status report window will appear that shows the result of the action request. WinLMF will fail to disable Continuous Waveform Mode for Modem devices provisioned in the conguration les on DMI, but not physically installed in the frame and for Modem devices under reset. 68P09283A63-5 4-25 FOA AUG 2007 Generate ATP Report Chapter 4: Acceptance Test Procedures Generate ATP Report ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Background Each time an Acceptance Test Procedure (ATP) report is updated to include the results of the most recent ATP tests if the Save Results button is used to close the status report window. The ATP report is not updated if the status reports window is closed using the Dismiss button. The ATP report for a UBS will contain test results saved with Save Results button. Results will be saved for the last test performed on the same: • Sector • Carrier • Channel When an ATP report for a UBS is generated, all saved test results will be included in the report. ATP Report Each time an ATP test is run, a report may be created for the UBS being tested. If a previous report exists for the UBS, it is updated with the new test results as noted above. The report includes the following for each test: • Test name • Channel number • Carrier number • Sector number • Test result • PASS or FAIL • Description information • Time stamp • Details/Warning information (if applicable) The report can be printed if the LMF computer is connected to a printer. Follow the procedure in Procedure 4-9 to view and/or print the ATP report for a UBS. 4-26 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Procedure 4-9 Generate ATP Report Generating an ATP Report Click Util. Save Test Report from the LMF menu. Click on a column heading to sort the report. Do one of the following: • If a printable file copy is not desired, click on the Dismiss button. • If a printable copy is required, select the desired file type in the pick list and click on the Save button. 68P09283A63-5 AUG 2007 4-27 FOA Generate ATP Report Chapter 4: Acceptance Test Procedures 4-28 68P09283A63-5 FOA AUG 2007 Chapter Leave the Site ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ 68P09283A63-5 AUG 2007 5-1 FOA Conguring Backhaul Chapter 5: Leave the Site Conguring Backhaul ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Before disconnecting the LMF and test equipment, the span backhaul must be configured to allow the UBS to communicate with the OMC-R. Introduction General This section allows the Cellular Field Engineers (CFEs) to pre-configure the UBS backhaul parameters prior to cutting over the UBS. It also contains information on how to make adjustments to individual backhaul parameters. Conguring the UBS Backhaul Prior to BTS cutover, some span configuration and connection parameters must be set to match the type of backhaul chosen for the particular BTS. These parameters can be checked and changed if required using the LMF Configure Backhaul Basic Screen. With these values set correctly, the OMC-R should be able to establish communications with the BTS. In the event that communication between the OMC-R and the BTS is lost and cannot be re-established, the LMF Configure Backhaul Advanced Screen can be used to restore the backhaul configuration to default values. See Figure 5-1. The default mode is Full Configuration Backhaul unless the Fractional Span box is checked. Backhaul Conguration Procedure Prior to executing Procedure 5-1, the information determined and gathered while working through the System Engineering Section of the UBS Set-up Procedure should be available. This information documents the intended backhaul configuration for the site and should match the configuration the OMC-R is expecting to use for the UBS. Failure to congure the UBS backhaul correctly with the LMF using the same DS0 conguration may result in the OMC-R being unable to establish communication with the UBS at cutover. 5-2 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Conguring Backhaul Figure 5-1 Congure Backhaul Basic Screen ti-cdma-06184.eps Procedure 5-1 Fractional Span Backhaul Conguration Procedure Go to the LMF Configure Backhaul Screen. See Figure 5-1 The default value for both the LcpNcpNegRetryInt and the PPPKeepAliveMsg parameters is 1000 ms. If the UBS will use satellite backhaul, the value for these parameters must be set to 2000 ms. If the UBS will use Fractional Span backhaul, click on the Fractional Span box. The Start DS0 and End DS0 boxes will now accept operator - entered values. Enter the starting DS0 and ending DS0 intended for Span 1 into the appropriate boxes. The LMF will configure Span 1, which is sufficient to allow the cutover to succeed. Span 2 will be configured from the OMC-R after cutover. Select the correct span type, frame type, coding type and span equalization from the pull-down menus in the Span Configuration section. Click on the OK button. 68P09283A63-5 5-3 FOA AUG 2007 Conguring Backhaul Chapter 5: Leave the Site Using the LMF Congure Backhaul Advanced Screen In the event that a backhaul configuration entered at the Operation and maintenance Center-Radio (OMC-R) results in the OMC-R being unable to communicate with the BTS, it may be necessary to reset backhaul parameters to their default values using the LMF. The LMF Configure Backhaul Advanced Screen is used to reset these values. Prior to executing Procedure 5-2, the information determined and gathered while working through the System Engineering Section of the BTS Set-up Procedure should be available. This information documents the intended backhaul configuration for the site and should match the configuration the OMC-R is expecting to use for the BTS. Default values for those parameters not covered above are listed below: • Muxing: enabled • Header Compression: enabled • Max Transport Unit Size (bytes): 512 • Hello Interval Timer (seconds): 30 • Periodic Transmission intervals for join/prune messages (seconds): 60 Connection Configuration (ubsCon) • Interval between LCP/IPCP retries during PPP link negotiation (milliseconds): 1000 For satellite backhaul: 2000 • Time between PPP keep-alive messages (milliseconds) 1000 For satellite backhaul: 2000 • Number of PPP keep-alive messages that can be lost before link is marked down: 5 • Number of retries for LCP/IPCP negotiations: 10 • Fractional Span backhaul: Per SE config 5-4 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Conguring Backhaul Span configuration (ubsSpan) • Span type: T1 • Frame type: ESF, Extended ... • Coding type: B8ZS, Bipolar 8–... • Termination type: TX131, 131 to 2... • Link alarm set threshold (bit errors / second): 30 • Link alarm set duration (seconds): 10 • Link alarm clear threshold (bit errors / second): 30 • Link alarm clear duration (seconds): 10 • Link removal threshold (bit errors / second): 75 • Link removal duration (seconds): 10 • Link recovery threshold (bit errors / second): 75 • Link recovery duration (seconds): 10 Procedure 5-2 Resetting Backhaul Parameters to their Default Values Go to the LMF Configure Backhaul Screen. Click on the Show Advanced box at the bottom of the screen. A full list of backhaul parameters will now be displayed. See Figure 5-2 Change the values in the entry boxes to align with the default values noted above. Click on the OK button. 68P09283A63-5 5-5 FOA AUG 2007 Conguring Backhaul Chapter 5: Leave the Site Figure 5-2 Congure Backhaul Advanced Screen en ter_filena me_her e_and_mo ve_per_st ep_8 5-6 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Prepare to Leave the Site Prepare to Leave the Site ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ External Test Equipment Removal Perform the procedure in Procedure 5-3 to disconnect the test equipment and configure the UBS for active service. Procedure 5-3 External Test Equipment Removal Procedure Disconnect all external test equipment from all TX and RX connectors on the UBS. Reconnect and visually inspect all TX and RX antenna feed lines on the UBS. Verify that all sector antenna feed lines are connected to the correct ports on the frame. Crossed antenna cables will cause system degradation of call processing. LMF Removal Before removing the LMF, verify that the UBS code load is synched with the code load in the OMC-R. DO NOT power down the LMF without performing the procedure below. Corrupted/lost data files may result, and in some cases, the LMF may lock up Follow the procedure in Procedure 5-4 to terminate the LMF session and remove the terminal. Continued 68P09283A63-5 5-7 FOA AUG 2007 Prepare to Leave the Site Chapter 5: Leave the Site Procedure 5-4 LMF Termination and Removal Procedure From the Local Terminal window select File > Exit. Anytime the LMF is exited from the UBS, the DMI controller(s) will automatically reboot within 1 minute in order to clear out any test configurations and boot up under the original configuration in the NECB and NECJ files. From the Windows Task Bar click Start > Shutdown. Click Yes when the Shut Down Windows message appears. Disconnect the LMF terminal Ethernet connector from the UBS cabinet. Disconnect the LMF serial port, the RS-232 to GPIB interface box, and the GPIB cables as required for equipment transport. Reset All Devices and Initialize Site Remotely Devices in the UBS should not be left with data and code loaded from the LMF. The configuration data and code loads used for normal operation could be different from those stored in the LMF files. Perform the procedure in Procedure 5-5 to reset all devices and initialize site remotely. Procedure 5-5 Reset UBS Devices and Remote Site Initialization Terminate the LMF session by following the procedure in Procedure 5-4 Verify that the OMC-R operator has re-activated the spans. Verify at the OMC-R that the UBS has been integrated and has synchronized its code and data. Account for all tools used and all parts removed. Visually inspect the UBS for any foreign objects and remove them. Visually inspect all cable connections, ensuring that they are connected as required for normal UBS operation. Verify no alarm conditions are being reported to the OMC-R. After all activities at the site have been completed, contact the OMC-R and confirm that the UBS is under OMC-R control. 5-8 68P09283A63-5 FOA AUG 2007 Appendix Data Sheets 68P09283A63-5 AUG 2007 A-1 FOA Optimization/ATP Checklist Appendix A: Data Sheets Optimization/ATP Checklist ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Verication of Test Equipment Used Table A-1 Test Equipment Used Checklist Manufacturer Model Serial Number A-2 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Optimization/ATP Checklist Checklist Table A-2 Procedure Checklist Parameter Specication Comments Power Up Procedures Pre-Power Up Test Per Procedure 2-1 Initial Power Up Test Per Procedure 2-2 andProcedure 2-3 Optimization/ATP Procedures LMF-to-UBS Connection Per Procedure 3-2 UBS GUI Login Per Procedure 3-3 UBS GUI Logout Per Procedure 3-4 Generate ATP Report Per Procedure 4-9 Leave the Site Test Equipment Removal Per Procedure 5-3, Procedure 5-5 Reset Devices and Initialize Site Remotely Per Procedure 5-5 68P09283A63-5 AUG 2007 A-3 FOA A-4 68P09283A63-5 FOA AUG 2007 Appendix FRU Optimization/ATP Matrix 68P09283A63-5 AUG 2007 B-1 FOA FRU Optimization/ATP Test Matrix Appendix B: FRU Optimization/ATP Matrix FRU Optimization/ATP Test Matrix ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Usage & Background Re-optimization of specific portions of the site may be required. An outline of some basic guidelines is included in the following tables. Re-optimization steps listed for any assembly detailed in the tables below must be performed anytime an RF cable associated with it is replaced. UBS Optimization Required Table B-1 When Optimization is Required Item Replaced XMI, IDRF, DMI, SSI, CRMS Optimize All sector / carrier TX / RX paths Detailed Reduced ATP There Reduced ATP consists of the TX Audit and RSSI tests. Detailed Optional Full ATP Test Matrix Table B-2 outlines the optional ATP tests that would need to be performed if the user wants to fully test the UBS components or if the reduced ATP fails. It is also assumed that all modules are placed in a Locked State via the LMF. The following guidelines should also be noted when using this table. B-2 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP FRU Optimization/ATP Test Matrix Not every procedure required to bring the site back on line is indicated in Table B-2. It is meant to be used as a guideline ONLY. The table assumes that the user is familiar enough with the UBS Optimization/ATP procedure to understand which test equipment set ups, test set calibrations, and UBS site preparation will be required before performing the Table # procedures referenced. Various passive UBS components (such as the TX and RX directional couplers, etc.) only call for a TX or RX calibration audit to be performed in lieu of a full path calibration. If the RX or TX path calibration audit fails, the entire RF path calibration will need to be repeated. If the RF path calibration fails, further troubleshooting is warranted. Whenever any unit is replaced, it is assumed that only power to the unit being replaced is turned off via the breaker supplying that unit. Table B-2 Full ATP Test Matrix Test Procedure Description Reference RX Cables Run the appropriate LMF applications RSSI Test FER Test TX Cables Run the appropriate LMF application TX Audit GPS Run the appropriate LMF applications GPS Initialization/Verification TX Audit RSSI Test Waveform Quality ATP Pilot Time Offset ATP Code Domain Power/Noise Floor FER Test • OPTIMIZATION AND TEST LEGEND •Required * Perform if necessary for additional fault isolation, repair assurance, or site certification 1 Perform on all carrier and sector TX paths to the UBS. 2 Perform on all carrier and sector RX paths to the UBS. 3 Perform on all primary and redundant TX paths of the affected carrier. 4 Perform on the affected carrier and sector TX paths. 5 Perform on the affected carrier and sector RX paths. 6 Verify performance by performing on one sector of on carrier only. 68P09283A63-5 AUG 2007 B-3 FOA B-4 68P09283A63-5 FOA AUG 2007 Appendix CDMA Operating Frequency 68P09283A63-5 AUG 2007 C-1 FOA 800 MHz CDMA Operating Frequency Programming Information Appendix C: CDMA Operating Frequency 800 MHz CDMA Operating Frequency Programming Information ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Introduction Programming of each of the BTS XMIs is performed by the BTS DMI modules over the Concentration Highway Interface (CHI) bus. This programming data determines the transmit and receive operating frequencies (channels) for each XMI. 800 MHz Channels Figure C-1 shows the total channels for the 800 MHz frequency spectrum and the channels allocated to CDMA. There are 10 CDMA channels used in a CDMA system. The channels used are unique to each customer system. Table C-1 shows the channel allocations and frequencies covered by the UBS Macro. Table C-1 800 MHz Channel Allocations System Designator CDMA Channel Validity Channel No. RX (MHz) TX (MHz) (see NOTE below) Not Valid Conditionally Valid Valid 991–1012 1013–1020 1021–1023 824.04–824.67 824.70–824.91 824.94–825.00 869.04–869.67 869.70–869.91 869.94–870.00 Valid Conditionally Valid Not Valid 1–303 304–311 312–333 825.03–834.09 834.12–834.33 834.36–834.99 870.03–879.09 879.12–879.33 879.36–879.99 Not Valid Conditionally Valid Valid Conditionally Valid Not Valid 334–355 356–363 364–636 637–644 645–666 835.02–835.65 835.68–835.89 835.92–844.32 844.11–844.32 844.35–844.98 880.02–880.65 880.68–880.89 880.92–889.32 889.11–889.32 889.35–889.98 A’ Not Valid Conditionally Valid Not Valid 667–688 689–694 695–716 845.01–845.64 845.67–845.82 845.85–846.48 890.01–890.64 890.67–890.82 890.85–891.48 B’ Not Valid Conditionally Valid Valid Conditionally Valid Not Valid 717–738 739–746 747–769 770–777 778–799 846.51–847.14 847.17–847.38 847.41–848.07 848.10–848.31 848.34–848.97 891.51–892.14 892.17–892.38 892.41–893.07 893.10–893.31 893.34–893.97 India Conditionally Valid Valid 1013–1020 1021–594 824.70–824.91 824.94–842.82 869.70–869.91 869.94–887.82 C-2 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP 800 MHz CDMA Operating Frequency Programming Information Table C-1 800 MHz Channel Allocations (Continued) System Designator China CDMA Channel Validity Valid Channel No. 37–283 RX (MHz) 826.11–833.49 TX (MHz) 871.11–878.49 The UBS supports all valid channels within the 800 MHz band and two Conditionally Valid channels: 1019 and 1020. On any other conditionally valid channels within the 800 MHz band, full TX Output power cannot be supported. 68P09283A63-5 AUG 2007 C-3 FOA 800 MHz CDMA Operating Frequency Programming Information Appendix C: CDMA Operating Frequency Figure C-1 800 MHz Frequency Spectrum (CDMA Allocation) F R E Q U E N C Y IN C R E AS E RX (MHz) CHANNELS TX 824.700 824.880 824.910 825.000 825.030 869.700 869.880 869.910 870.000 870.030 1013 1019 1020 1023 825.110 870.110 37 833.490 878.490 283 834.090 834.120 834.330 879.090 879.120 897.330 303 304 835.680 880.680 356 835.860 835.890 880.860 880.890 362 363 844.110 844.140 889.110 889.140 637 638 844.320 889.320 644 845.670 890.670 689 845.820 890.820 694 847.170 892.170 739 847.380 847.410 892.380 892.410 746 747 848.100 848.130 893.100 893.130 770 771 848.330 893.330 777 311 ti-cdma-05810.eps C-4 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP 800 MHz CDMA Operating Frequency Programming Information 800 MHz Channel Center Frequencies Table C-2 shows the valid 800 MHz CDMA channels, listed in both decimal and hexadecimal, and the corresponding transmit and receive frequency for each. Actual frequencies used depend on customer CDMA system frequency plan. Other channels and frequencies can be calculated using the following formula: For channels 1–799: • TX Freq = 870.00 + (0.03 x Channel) • RX Freq = (0.03 x Channel) + 825.00 For channels 991–1023: • TX Freq = 870.00 + [0.03 x (Channel - 1023)] • RX Freq = [0.030 x (Channel - 1023)] + 825.00 For channels 1024–1323: • TX Freq = 860.04 + [0.03 x (Channel- 1024)] • RX Freq = [0.030 x (Channel - 1024)] + 815.04 For channels 1324–1490: • TX Freq = 860.04 + [0.03 x (Channel- 1024)] • RX Freq = [0.030 x (Channel - 1024)] + 815.04 Table C-2 800 MHz TX and RX Channel Center Frequencies Decimal Hex Transmit Center Frequency (MHz) Receive Center Frequency (MHz) 1021 03FD 869.94 824.94 1023 03FF 870.00 825.00 0001 870.03 825.03 303 012F 879.09 834.09 364 016C 880.92 835.92 636 027C 889.32 844.32 666 029A 889.99 844.35 716 02CC 891.49 844.98 747 02EB 893.10 847.71 769 0301 893.31 848.07 68P09283A63-5 AUG 2007 C-5 FOA 1900 MHz CDMA Operating Frequency Programming Information Appendix C: CDMA Operating Frequency 1900 MHz CDMA Operating Frequency Programming Information ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Introduction Programming of each of the BTS XMIs is performed by the BTS DMI modules over the Concentration Highway Interface (CHI) bus. This programming data determines the transmit and receive operating frequencies (channels) for each XMI. 1900 MHz Channels Figure C-2 shows the total channels for the 1900 MHz frequency spectrum and the channels allocated to CDMA. The channels used are unique to each customer system. Table C-3 shows the channel allocations and frequencies covered by the UBS Macro. Table C-3 1900 MHz Band Class 1 Channel Allocations CDMA Channel Validity Channel Number TX (MHz) RX (MHz) Not Valid 0–24 1930.00–1931.20 1850.00–1851.20 Valid 25–1175 1931.25–1988.75 1851.25–1908.75 Not Valid 1176–1199 1988.80–1989.95 1908.80–1909.95 C-6 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP 1900 MHz CDMA Operating Frequency Programming Information Figure C-2 1900 MHz Frequency Spectrum (CDMA Allocation) F R E Q U E N C Y IN C R E AS E RX (MHz) CHANNELS TX 1851.25 1931.25 25 1855.00 1935.00 100 1860.00 1940.00 200 1867.50 1947.50 350 1872.50 1952.50 450 1876.25 1956.25 525 1882.50 1962.50 650 1887.50 1967.50 750 1891.25 1971.25 825 1895.00 1975.00 900 1902.50 1982.50 1050 1905.00 1985.00 1100 1908.75 1988.75 1175 ti-cdma-05810-B.eps 68P09283A63-5 AUG 2007 C-7 FOA 1900 MHz CDMA Operating Frequency Programming Information Appendix C: CDMA Operating Frequency 1900 MHz Channel Center Frequencies Table C-4 shows some of the valid 1900 MHz CDMA channels, listed in both decimal and hexadecimal, and the corresponding transmit and receive frequency for each. Actual frequencies used depend on customer CDMA system frequency plan. Other channels and frequencies can be calculated using the following formula: For channels 0–1199: • TX Freq = 1930.00 + (0.05 x Channel) • RX Freq = 1850.00 + (0.05 x Channel) Table C-4 1900 MHz TX and RX Channel Center Frequencies Decimal Hex Transmit Center Frequency (MHz) Receive Center Frequency (MHz) 25 0019 1931.25 1851.25 100 0064 1935.00 1855.00 200 00C8 1940.00 1860.00 350 015E 1947.50 1867.50 450 01C2 1952.50 1872.50 525 020D 1956.25 1876.25 650 028A 1962.50 1882.50 750 02EE 1967.50 1887.50 825 0339 1971.25 1891.25 900 0384 1975.00 1895.00 1050 041A 1982.50 1902.50 1100 044C 1985.00 1905.00 1175 0497 1988.75 1908.75 C-8 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP CDMA Operating Frequency Programming Information CDMA Operating Frequency Programming Information ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Introduction Programming of each of the BTS XMIs is performed by the BTS GLI cards over the Concentration Highway Interface (CHI) bus. This programming data determines the transmit and receive operating frequencies (channels) for each XMI. 68P09283A63-5 AUG 2007 C-9 FOA CDMA Operating Frequency Programming Information Appendix C: CDMA Operating Frequency 2100 MHz Channels Figure C-1Figure C-3 shows the total channels for the 2100 MHz frequency spectrum and the channels allocated to CDMA. There are 10 CDMA channels used in a CDMA system. The channels used are unique to each customer system. Figure C-3 2100 MHz Frequency Spectrum (CDMA Allocation) FREQ (MHz) RX TX CHANNEL 125 1926.25 2116.25 1978.75 2168.75 1175 1199 CDMA2100MHZ-1 ti-cdma-00115.eps C-10 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP CDMA Operating Frequency Programming Information Calculating 2100 MHz Channel Center Frequencies Table C-5 shows the valid 2100 MHz CDMA channels, listed in both decimal and hexadecimal, and the corresponding transmit and receive frequencies for each. Center frequencies (in MHz) for channels not shown in the table may be calculated as follows: • TX = 2110 + 0.05 * Channel# Example: Channel 262 TX = 2110 + 0.05 * 262 = 2123.10 MHz • RX = 1920 + 0.05 * Channel# Example: Channel 262 RX = 1920 + 0.05 * 262 = 1933.10 MHz Actual frequencies used depend on customer CDMA system frequency plan. Each CDMA channel requires a 1.77 MHz frequency segment. The actual CDMA carrier is 1.23 MHz wide, with a 0.27 MHz guard band on both sides of the carrier. Minimum frequency separation required between any CDMA carrier and the nearest NAMPS/AMPS carrier is 900 kHz (center-to-center). Table C-5 2100 MHz TX and RX Channel Center Frequencies Channel Number Receive Center Frequency (MHz) Decimal Hex 125 007D 2116.25 1926.25 150 0096 2117.50 1927.50 175 00AF 2118.75 1928.75 200 00C8 2120.00 1930.00 225 00E1 2121.25 1931.25 250 00FA 2122.50 1932.50 275 0113 2123.75 1933.75 300 012C 2125.00 1935.00 325 0145 2126.25 1936.25 350 015E 2127.50 1937.50 375 0177 2128.75 1938.75 400 0190 2130.00 1940.00 425 01A9 2131.25 1941.25 450 01C2 2132.50 1942.50 475 01DB 2133.75 1943.75 500 01F4 2135.00 1945.00 525 020D 2136.25 1946.25 550 0226 2137.50 1947.50 575 023F 2138.75 1948.75 600 0258 2140.00 1950.00 68P09283A63-5 AUG 2007 Transmit Center Frequency (MHz) C-11 FOA CDMA Operating Frequency Programming Information Appendix C: CDMA Operating Frequency Table C-5 2100 MHz TX and RX Channel Center Frequencies (Continued) Channel Number Transmit Center Frequency (MHz) Receive Center Frequency (MHz) Decimal Hex 625 0271 2141.25 1951.25 650 028A 2142.50 1952.50 675 02A3 2143.75 1953.75 700 02BC 2145.00 1955.00 725 02D5 2146.25 1956.25 750 02EE 2147.50 1957.50 775 0307 2148.75 1958.75 800 0320 2150.00 1960.00 825 0339 2151.25 1961.25 850 0352 2152.50 1962.50 875 036B 2153.75 1963.75 900 0384 2155.00 1965.00 925 039D 2156.25 1966.25 950 03B6 2157.50 1967.50 975 03CF 2158.75 1968.75 1000 03E8 2160.00 1970.00 1025 0401 2161.25 1971.25 1050 041A 2162.50 1972.50 1075 0433 2163.75 1973.75 1100 044C 2165.00 1975.00 1125 0465 2166.25 1976.25 1150 047E 2167.50 1977.50 1175 0497 2168.75 1978.75 C-12 68P09283A63-5 FOA AUG 2007 Appendix Test Equipment Preparation 68P09283A63-5 AUG 2007 D-1 FOA Test Equipment Preparation Appendix D: Test Equipment Preparation Test Equipment Preparation ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ This section provides pre–testing set–up information for test equipment supported by the LMF. Purpose Pretesting set–up information covered includes verification and setting GPIB addresses, inter–unit cabling, connectivity testing, pre–test control settings, and equipment calibration for items which are not calibrated with the Calibrate Test Equipment function of the LMF. GPIB addresses This appendix provides information on verification and changing GPIB addresses for the following test equipment items: • Agilent E7495A/B test equipment set–up • Agilent E4406A transmitter test set • Agilent E4432B signal generator • Advantest R3267 spectrum analyzer • Advantest R3562 signal generator • Agilent 8935 analyzer (formerly HP 8935) • Advantest R3465 analyzer • Gigatronics 8541C power meter • Agilent E4418 power meter • GPIB adapter Calibration actions The following calibration–related subjects are also covered for the test equipment items indicated: Cable calibration set-up – Calibrating Test Cable Set–up using Advantest R3465 D-2 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Test Equipment Preparation Calibrating test sets – • Agilent E4406A Transmitter Tester Self-alignment (Calibration) • Gigatronics 8541C power meter (Calibration) 68P09283A63-5 AUG 2007 D-3 FOA Agilent E7495A/B Test Equipment Setup Appendix D: Test Equipment Preparation Agilent E7495A/B Test Equipment Setup ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Initial Requirement This test equipment requires a warm-up period of at least 30 minutes before BTS testing or calibration begins. Using the Agilent E7495A/B with the LMF The Agilent E7495A/B does not require the use of the 19MHz frequency reference; if connected, it will be ignored. The Even Sec SYNC connection is required. The Agilent E7495A/B signal generator is only calibrated down to -80dB. In order to achieve accurate FER testing, be sure the RX setup includes at least 40dB of attenuation. This will ensure the signal generator will output sufficient power to operate in the calibrated range. Set the IP Address in the test set as described in . Procedure D-1 Set IP Address on Agilent E7495A/B test set Use the System Button > Controls > IPAdmin to set an IP address on the E7495A/B as 128.0.0.11, and Netmask to 255.255.255.128. If the IP addressed was changed, reboot the E7495 For compatibility, change the LMF computer Network Interface Card (NIC) IP address to 128.0.0.48, subnetmask 255.255.255.128. Connection Connect the E7495A/B test set as described in the Figure 3-29Test Equipment Connection to the LMF. Power Sensor Calibration Procedure D-2 describes the E7495A/B Power Sensor Calibration. Procedure D-2 E7495A/B Power Sensor Calibration Display the power meter screen. Make sure equipment is connected as shown in Figure D-1. D-4 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Agilent E7495A/B Test Equipment Setup Procedure D-2 E7495A/B Power Sensor Calibration (Continued) Zero the power meter by doing the following: 1. Press the Zero softkey. 2. Press the Continue softkey. Calibrate the power meter by doing the following: 1. Press Ref CF. 2. Enter the reference cal factor, reading it off the label on the power sensor head. 3. Press Calibrate. 4. Connect the power sensor (see Figure D-2). 5. Press Continue. 6. Press Cal Factor. 7. Enter the cal factor from the label on the power sensor head. Select a cal factor that is within the operating frequency of the base station being calibrated. Figure D-1 Agilent E7495A/B Pre-Power Sensor Calibration Connection Power REF 50 MHz GPIO Even Second Sync In Serial 1 Port 2 RF In Use only Agilent supplied power adapter Ext Ref In Sensor Port 1 RF Out / SWR GPS Antenna Serial 2 POWER SENSOR NOT CONNECTED ti-cdma-00116.eps Figure D-2 Agilent E7495A/B Power Sensor Calibration Connection POWER SENSOR CONNECTED GPIO Port 2 RF In Power REF 50 MHz Ext Ref In Even Second Sync In Serial 1 Sensor Serial 2 Use only Agilent supplied power adapter Port 1 RF Out / SWR GPS Antenna ti-cdma-00117.eps 68P09283A63-5 AUG 2007 D-5 FOA Agilent E7495A/B Test Equipment Setup Appendix D: Test Equipment Preparation Cable Calibration Follow the directions in the LMF application program to calibrate cables. Additional cable calibration details - Calibrate the short cable (see Figure 3-25) and two 10 dB attenuators to establish a baseline and then calibrate the TX and RX set-ups. Because at least 40 dB of attenuation is needed when testing the FER, the set-up for RX is the same as for TX. D-6 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Verifying and Setting GPIB Addresses – Agilent E4406A Transmitter Tester Verifying and Setting GPIB Addresses – Agilent E4406A Transmitter Tester ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Figure D-3 Setting Agilent E4406A GPIB Address Active Function Area Softkey Label Display Area System Key Bk Sp Key Enter Key Softkey Buttons Data Entry Keypad ti-CDMA-WP-00085-v01-ildoc-ftw ti-cdma-00118.eps Refer to Figure D-3 when performing the following: Procedure D-3 Verify and Set/Change Agilent E4406A Transmitter Tester GPIB Address In the SYSTEM section of the instrument front panel, press the System key. Result: The softkey labels displayed on the right side of the instrument screen will change. Press the Config I/O softkey button to the right of the instrument screen. Result: • The softkey labels will change. • The current instrument GPIB address will be displayed below the GPIB Address softkey label. If the current GPIB address is not set to 18, perform the following to change it: 1. Press the GPIB Address softkey button. Result: In the on-screen Active Function Area, GPIB Address will be displayed followed by the current GPIB address. 68P09283A63-5 AUG 2007 D-7 FOA Verifying and Setting GPIB Addresses – Agilent E4406A Transmitter Tester Appendix D: Test Equipment Preparation Procedure D-3 Verify and Set/Change Agilent E4406A Transmitter Tester GPIB Address (Continued) 2. On the front panel Data Entry keypad, enter the communications system analyzer GPIB address of 18 Result: The GPIB Address label will change to Enter. Digits entered with the keypad will replace the current GPIB address in the display. To correct an entry, press the Bk Sp key at the upper right of the keypad to delete one character at a time. 3. Press the Enter softkey button or the keypad Enter key to set the new GPIB address. Result: The Cong I/O softkey labels will reappear. The new GPIB address will be displayed under the GPIB Address softkey label. D-8 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Verifying and Setting GPIB Addresses – Agilent E4432B Signal Generator Verifying and Setting GPIB Addresses – Agilent E4432B Signal Generator ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Figure D-4 Setting Agilent E4432B GPIB Address Active Entry Area Softkey Label Display Area Utility Key Softkey Buttons Numeric Keypad Backspace Key ti-CDMA-WP-00086-v01-ildoc-ftw ti-cdma-00119.eps Refer to Figure D-4 when performing the following: Procedure D-4 Verify and Change Agilent E4432B Signal Generator GPIB Address In the MENUS section of the instrument front panel, press the Utility key. Result: • The softkey labels displayed on the right side of the instrument screen will change. Press the GPIB/RS232 softkey button to the right of the instrument screen. Result: • The softkey labels will change. • The current instrument GPIB address will be displayed below the GPIB Address softkey label. If the current GPIB address is not set to 1, perform the following to change it: 1. Press the GPIB Address softkey button. Result: 68P09283A63-5 AUG 2007 D-9 FOA Verifying and Setting GPIB Addresses – Agilent E4432B Signal Generator Appendix D: Test Equipment Preparation Procedure D-4 Verify and Change Agilent E4432B Signal Generator GPIB Address (Continued) The GPIB Address label and current GPIB address will change to boldface. In the on-screen Active Entry Area, Address: will be displayed followed by the current GPIB address. 2. On the front panel Numeric keypad, enter the signal generator GPIB address of 1. Result: The GPIB Address label will change to Enter. Digits entered with the keypad will replace the current GPIB address in the Active Entry display. To correct an entry, press the backspace key at the lower right of the keypad to delete one character at a time. 3. Press the Enter softkey button to set the new GPIB address. Result: The new GPIB address will be displayed under the GPIB Address softkey label. D-10 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Verifying and Setting GPIB Addresses – Advantest R3267 Spectrum Analyzer Verifying and Setting GPIB Addresses – Advantest R3267 Spectrum Analyzer ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Figure D-5 Setting Advantest R3267 GPIB Address Softkey Lable Display Area Softkey Buttons on REMOTE LED LCL Key CONFIG Key Keypad BS Key ENTR Key ti-CDMA-WP-00083-v01-ildoc-ftw ti-cdma-00120.eps Refer to Figure D-5 when performing the following. Procedure D-5 Verify and Set/Change Advantest R3267 GPIB Address If the REMOTE LED is lighted, press the LCL key. Result: • The LED extinguishes. Press the CONFIG key. Result: • The CONFIG softkey labels will appear in the softkey label display area of the instrument display. • The current GPIB address will be displayed below the GPIB Address softkey label. If the current GPIB address is not set to 18, perform the following to change it: 1. Press the GPIB Address softkey. Result: A GPIB Address entry window will open in the instrument display showing the current GPIB address. 68P09283A63-5 AUG 2007 D-11 FOA Verifying and Setting GPIB Addresses – Advantest R3267 Spectrum Analyzer Procedure D-5 2. Appendix D: Test Equipment Preparation Verify and Set/Change Advantest R3267 GPIB Address (Continued) Enter 18 on the keypad in the ENTRY section of the instrument front panel. To correct an entry, press the BS (backspace) key at the lower right of the keypad to delete one character at a time. Result: Characters typed on the keypad will replace the address displayed in the GPIB Address entry window. 3. Press the ENTR key to the lower right of the keypad to enter the address. Result: The GPIB Address entry window closes. The new address is displayed in the bottom portion of the GPIB Address softkey label. D-12 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Verifying and Setting GPIB Addresses – Advantest R3562 Signal Generator Verifying and Setting GPIB Addresses – Advantest R3562 Signal Generator ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Set the GP-IB ADDRESS switch on the rear of the Advantest R3562 signal generator to address 1 as shown in Figure D-6. Figure D-6 Advantest R3562 GPIB Address Switch Setting SERIALI/O ON GP-IBADDRESSES 54321 AIN POW ER OFF GP-IP ADDRESS GPIB SYNTHEREF 10MHZREF CLOCK OUT OUT IN GPIB Address set to ?1" LOCAL OUT IN/OUT IN/OUT 5 4 3 2 1 ti-cdma-00121.eps 68P09283A63-5 AUG 2007 D-13 FOA Verifying and Setting GPIB Addresses – Agilent 8935 Series E6380 Test Set Appendix D: Test Equipment Preparation Verifying and Setting GPIB Addresses – Agilent 8935 Series E6380 Test Set ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Figure D-7 Setting Agilent 8935 Test Set GPIB Address Preset Local Inst Config FW00885 Shift Cursor Control ti-cdma-00122.eps This procedure assumes that the test equipment is set up and ready for testing. Refer to Figure D-7 when performing the following. Procedure D-6 Verify and Set/Change Agilent 8935 Series E6380 Test Set GPIB Address The HP I/O configuration MUST be set to Talk & Listen, or NO device on the GPIB bus will be accessible. (Consult test equipment manufacturer’s documentation for additional information as required.) D-14 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Verifying and Setting GPIB Addresses – Agilent 8935 Series E6380 Test Set Procedure D-6 Verify and Set/Change Agilent 8935 Series E6380 Test Set GPIB Address (Continued) To verify that the GPIB addresses are set correctly, press Shift and LOCAL on the 8935. The current HP-IB address is displayed at the top of the screen. HP-IB is the same as GPIB. If the current GPIB address is not set to 18, perform the following to change it: 1. Press Shift and Inst Cong. 2. Turn the Cursor Control knob to move the cursor to the HP-IB Adrs field. 3. Press the Cursor Control knob to select the field. 4. Turn the Cursor Control knob as required to change the address to 18. 5. Press the Cursor Control knob to set the address. Press Preset to return to normal operation. 68P09283A63-5 AUG 2007 D-15 FOA Verifying and Setting GPIB Addresses – Gigatronics 8541C Power Meter Appendix D: Test Equipment Preparation Verifying and Setting GPIB Addresses – Gigatronics 8541C Power Meter ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Figure D-8 Gigatronics 8541C Power Meter Detail CONNECT POWER SENSOR TO CALIBRATOR POWER REFERENCE WHEN CALIBRATING/ZEROING UNIT CONNECT POWER SENSOR WITH POWER METER TURNED OFF AC POWER LIN ELV LT AN GE SE EO CT IO Giga-tronics8542CUniversalPowerMeter - 100 CALIBRATOR dBm GPIB CONNECTION MODE GPIB 120VAC ZERO FREQ REL CAL MENUdB/MWRECALL ENTER ESCAPE LOCAL GPIB POWER FRONT View REAR View ti-cdma-00128.eps Refer to Figure D-8 when performing the following: This procedure assumes that the test equipment is set up and ready for testing. Procedure D-7 Verify and Set/Change Gigatronics 8541C Power Meter GPIB Address Do not connect/disconnect the power meter sensor cable with AC power applied to the meter. Disconnection could result in destruction of the sensing element or miscalibration. Press MENU. Use the [Symbol_arrowblackdown] arrow key to select CONFIG MENU and press ENTER. Use the [Symbol_arrowblackdown] arrow key to select GPIB and press ENTER. Result: The current Mode and GPIB Address are displayed. D-16 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Procedure D-7 (Continued) Verifying and Setting GPIB Addresses – Gigatronics 8541C Power Meter Verify and Set/Change Gigatronics 8541C Power Meter GPIB Address If the Mode is not set to 8541C, perform the following to change it: 1. Use the ◄ ► arrow keys as required to select MODE. 2. Use the arrow keys as required to set MODE to 8541C If the GPIB address is not set to 13, perform the following to change it: 1. Use the ►arrow key to select ADDRESS. 2. Use the arrow keys as required to set the GPIB address to 13. Press ENTER to return to normal operation. 68P09283A63-5 AUG 2007 D-17 FOA Verifying and Setting GPIB Addresses – Agilent E4418 Power Meter Appendix D: Test Equipment Preparation Verifying and Setting GPIB Addresses – Agilent E4418 Power Meter ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Figure D-9 Agilent E4418B Power Meter Verify and Set/Change GPIB Address ARROW KEYS PRESET/ LOCAL STATUS REPORTING LINE SOFTKEY LABEL AREA SYSTEM/ INPUTS SOFTKEY MENU TITLE LEFT DOWN UP SOFTKEY MENU PAGE NUMBER RIGHT ti-cdma-05962.eps Refer to Figure D-9 when performing the following: Procedure D-8 Verify and Set/Change Agilent E4418 Power Meter GPIB Address If RMT (remote operation) is displayed on the left end of the instrument screen status reporting line, press the Preset/Local front panel button. Result: The status line entry will change to LCL (local or front panel operation). On the instrument front panel, press the System/Inputs key. Result: The softkey labels displayed on the right side of the instrument screen will change. Press the Remote Interface > Congure Interface > GPIB softkeys to the right of the instrument screen. Result: • The softkey labels will change. • The current instrument GPIB address will be displayed below the GP-IB Addr softkey label. D-18 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Verifying and Setting GPIB Addresses – Agilent E4418 Power Meter Procedure D-8 Verify and Set/Change Agilent E4418 Power Meter GPIB Address (Continued) If the current GPIB address is set to 13, skip to step step 8. If the current GPIB address is not set to 13, press the GP-IB Addr softkey. Result: A pop-up window will appear and display the current GPIB address. Modify the GPIB address in the pop-up as necessary using the following front panel keys to perform the indicated functions: • Up and down arrow keys to change the digit on which the cursor is currently located • Left and right arrow keys to move the cursor to other digits When the GPIB address is correct, press the Enter softkey. Return the instrument to the required setup configuration for testing. 68P09283A63-5 AUG 2007 D-19 FOA Verifying and Setting GPIB Addresses – RS232 GPIB Interface Box Appendix D: Test Equipment Preparation Verifying and Setting GPIB Addresses – RS232 GPIB Interface Box ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Be sure that the RS232 GPIB interface box DIP switches are set as shown in Figure D-10. Figure D-10 RS232 GPIB Interface Box DIP SWITCH SETTINGS S MODE DATA FORMAT BAUD RATE ON GPIB ADRS G MODE RS232-GPIB Interface Box ti-cdma-00125.eps D-20 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Test Equipment Calibration – Agilent 4406 Self-alignment Test Equipment Calibration – Agilent 4406 Self-alignment ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Refer to Figure D-11 when performing the following. Figure D-11 Agilent E4406A Self-alignment Softkey Label Display Area Softkey Buttons System Key ti-cdma-00126.eps Procedure D-9 Agilent E4406A Self-alignment (Calibration) In the SYSTEM section of the instrument front panel, press the System key. Result: The softkey labels displayed on the right side of the instrument screen will change. Press the Alignments softkey button to the right of the instrument screen. Result: The softkey labels will change. Press the Align All Now softkey button. Result: • All other instrument functions will be suspended during the alignment. • The display will change to show progress and results of the alignments performed. • The alignment will take less than one minute. 68P09283A63-5 AUG 2007 D-21 FOA Test Equipment Calibration – Gigatronics 8542 Power Meter Appendix D: Test Equipment Preparation Test Equipment Calibration – Gigatronics 8542 Power Meter ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Precise transmit output power calibration measurements are made using a bolometer-type broadband power meter with a sensitive power sensor. Follow the steps in Procedure D-10 to enter information unique to the power sensor. Refer to Figure D-12 as necessary. Figure D-12 Gigatronics 8541C Power Meter Calibration CONNECT POWER SENSOR TO CALIBRATOR POWER REFERENCE WHEN CALIBRATING/ZEROING UNIT CONNECT POWER SENSOR WITH POWER METER TURNED OFF AC POWER LIN ELV LT AN GE SE EO CT IO Giga-tronics8542CUniversalPowerMeter - 100 CALIBRATOR dBm GPIB CONNECTION MODE GPIB 120VAC ZERO FREQ REL CAL MENUdB/MWRECALL ENTER ESCAPE LOCAL GPIB POWER FRONT View REAR View ti-cdma-00128.eps Procedure D-10 Calibrate Gigatronics 8542 Power Meter Do not connect/disconnect the power meter sensor cable with AC power applied to the meter. Disconnection could result in destruction of the sensing element or miscalibration. Make sure the power meter POWER pushbutton is OFF. Connect the power sensor cable to the SENSOR input. Set the POWER pushbutton to ON. Allow the power meter and sensor to warm up and stabilize for a minimum of 60 minutes before performing the calibration procedure. D-22 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Test Equipment Calibration – Gigatronics 8542 Power Meter Procedure D-10 Calibrate Gigatronics 8542 Power Meter (Continued) Connect the power sensor to the CALIBRATOR output connector. Press ZERO, and wait for the process to complete. Result: Sensor factory calibration data is read to power meter during this process. When the zeroing process is complete, disconnect the power sensor from the CALIBRATOR output. 68P09283A63-5 AUG 2007 D-23 FOA D-24 68P09283A63-5 FOA AUG 2007 Appendix Optimization and Calibration Procedures 68P09283A63-5 AUG 2007 E-1 FOA Introduction to Calibration Appendix E: Optimization and Calibration Procedures Introduction to Calibration ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Overview This appendix is not normally applicable to the UBS, but is provided for reference. This section describes procedures for isolating the UBS from the span lines, preparing and using the LMF, downloading system operating software, set up and calibration of the supported test equipment, and transmit/receive path verification. Before using the LMF, use a browser to view the CAVEATS section in the readme.html file in the LMF home directory (e.g., c:\wlmf) for any applicable information. Optimization Process Summary After a UBS is physically installed and the preliminary operations, such as power up, have been completed, the LMF is used to optimize the UBS. The basic optimization process consists of the following: E-2 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Introduction to Calibration 1. Use the status function and verify that all of the installed devices of the following types respond with status information: HDModem and DMI. 2. Verify the code load of all the devices. 3. Verify the operation of the GPS and QHSO or MSO signals. 4. Unlock the following devices: UBS 5. Connect the required test equipment for a full optimization. 6. Select the test equipment. 7. Perform the TX Audit. 8. If the TX Audit fails, there is a hardware problem, correct the problem that caused the failure and repeat the optimization for the failed path. 9. If the TX Audit portion of the optimization passes for the RF path, but some of the TX or RX tests fail, correct the problem that caused the failure and run the individual tests as required until all TX and RX tests have passed for all the paths. 68P09283A63-5 AUG 2007 E-3 FOA Take Control of UBS Resources Appendix E: Optimization and Calibration Procedures Take Control of UBS Resources ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ UBS Control In a UBS, the DMI controls all UBS resources, such as the Modem and QHSO , which were previously managed by the Central Base Station Controller (CBSC) and Mobility Manager (MM) in BTSes. As a result, the LMF cannot perform any of the normal code/data download, optimization, or test functions with the UBS until the DMI gives up control of these items. Taking Control of UBS Resources After logging into a UBS for optimization or acceptance testing, the first action which must be performed is transferring control of the UBS from the Controller to the LMF Prerequisites LMF must be logged into the UBS. after logging into the UBS, the LMF will be operating in monitoring mode. (It allows the user to view status/monitor all UBS cards and Site Controller (SC)) To take control of all devices from the UBS, perform UBS-> Enter Invasive Mode (DMI comes under the control of the LMF and disconnects it from the UBS span lines). E-4 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Code Syncing to the UBS Code Syncing to the UBS ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ General Information Before a UBS can operate, each equipped device must contain device application (RAM) code and must be code sync’d to each equipped device by the user before the UBS can be made fully functional for the site where it is installed 68P09283A63-5 AUG 2007 E-5 FOA Bay Level Offset Calibration Appendix E: Optimization and Calibration Procedures Bay Level Offset Calibration ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ To avoid over-driving the XMIs, the BLO for UBS must be -2 dB or greater. Introduction Bay Level Offset (BLO) calibration is used to optimize non-standard UBS installations. Factory calibration of the UBS components assures proper TX performance of the UBS system in all installations that use Motorola supplied hardware in the TX path. In the event of a non-standard installation, BLO calibration can be used to compensate for TX path loss variations due to non-standard equipment. RF Path Bay Level Offset Calibration Calibration identifies the accumulated gain in every transmit path (sector–carrier) at the UBS site and stores that value in a BLO data base calibration table in the LMF. For single frames, each receive path starts at an RX antenna port and terminates at the UBS. Calibration identifies the accumulated gain in every transmit path at the UBS site and stores that value in a BLO database. Only those slots (sectors) actually equipped in the current NEC file are tested. When to TX Audit TX Audit will be performed: • After initial UBS installation • After UBS replacement TX Path Calibration The TX Path Calibration assures correct site installation, cabling, and the first order functionality of all installed equipment. The proper function of each RF path is verified during calibration. The external test equipment is used to validate/calibrate the TX paths of the UBS. E-6 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Bay Level Offset Calibration Before connecting any test equipment directly to any TX OUT connector you must rst verify that there are no CDMA channels keyed. Have the OMC–R place all sectors under test OOS. Failure to do so can result in serious personal injury and/or equipment damage. Always wear an approved anti–static wrist strap while handling any circuit card/module. If this is not done, there is a high probability that the card/module could be damaged by ESD. BLO Calibration Data File During the calibration process, the LMF creates a UBS–#.cal calibration (BLO) offset data file (CAL file) in the UBS–# folder. After calibration has been completed, Cal file has to be downloaded to the UBS using UBS–>Upload CAL File function. An explanation of the file is shown below. Test Equipment Set-up for RF Path Calibration Follow the procedure in Procedure E-1 to set up test equipment. Procedure E-1 Test Equipment Setup (RF Path Calibration) Verify the GPIB controller is properly connected and turned on (does not apply to the Agilent E7495A/B). To prevent damage to the test equipment, all transmit (TX) test connections must be via the 30 dB directional coupler. Total cable loss should be no less than 30 dB. If it has not already been done, connect the LMF computer to the UBS LMT connector on the UBS. 68P09283A63-5 AUG 2007 E-7 FOA Bay Level Offset Calibration Appendix E: Optimization and Calibration Procedures Transmit (TX) Path Calibration Description The Line Maintenance Facility (LMF) will automatically use the channel numbers and power levels assigned to the UBS if the UBS has been connected to an Operation Maintenance Center — Radio (OMC-R) and brought into service. If the OMC-R connection has not yet been established, the LMF operator will have to manually select each calibration point using a specific channel number and power level for that channel. Bay Level Offset (BLO) calibration is only applied to the specific channel numbers, thus calibration must be done on the exact channel number to be used when the site is operational. TX Calibration and the LMF The LMF Tests > TX > TX Calibration... and Tests > All Cal/Audit... selections perform TX BLO calibration testing for the installed RDF. The All Cal/Audit... selection initiates a series of actions to perform TX calibration, and, if calibration is successful, download BLO and perform TX audit. The TX Calibration... selection performs only TX calibration. When TX Calibration... is used, BLO download and TX Audit must be performed as separate activities. The CDMA Test Parameters window which opens when TX Calibration... or All Cal/Audit... is selected contains several user–selectable features which are described in the following subsections. Pilot Generation Channels / Carrier pick–list containing the list of carriers available for the selected Sector/Carriers. The list is derived from NECB file. DMI pick–list containing the list of DMIs being able to generate Pilot signal of selected Sector/Carrier. The list is derived from NECB file. The list is derived from NECB file. PN non–modifiable text box(es). The value is the PilotPn derived from NECB files Carrier # Channels: text box(es). Carrier numbers and default channels are derived from NECB file. Frequency Band non–modifiable text box(es). The value is the frequency band derived from NECB files. Verify BLO In both the TX Calibration and All Cal/Audit dialog boxes, a Verify BLO checkbox is provided and checked by default. After the actual TX calibration is completed during either the TX Calibration or AllCal/Audit process, the BLO derived from the calibration is compared to a standard, acceptable BLO tolerance for the UBS. In some installations, additional items may be installed in the transmit path. The additional change in gain from these items could cause BLO verification failure and, therefore, failure of the entire calibration. In these cases, either the Verify BLO checkbox should be unchecked or the additional path losses should be added into each applicable sector using the Util > Edit > TX Coupler Loss... function. E-8 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Bay Level Offset Calibration Single–sided BLO Checkbox Another option that appears in the pull–down menu is Single–sided BLO. Normally valid BLO values are some value plus–or–minus some offset. Single–sided BLO spec is >–3.5 dB. Double–sided BLO spec is –1.5 +/– 2.0 dB. To get the more stringent conditions, the operator checks Single–sided BLO when calibrating non-redundant transceivers. Single–sided BLO carries the likelihood of more failures. This option should only be used by experienced CFEs. The Tests > TX > TX Calibration... menu window has a Test Pattern pull–down menu. This menu has the following choices. • Pilot (default) – performs tests using a pilot signal only. This pattern should be used when running in–service tests. It requires the DMIs to do the test. • Standard – performs the tests using pilot, synch, paging and six traffic channels. • CDFPilot – performs the tests using the pilot signal, however, the gain is specified in the NEC file. Advanced users may use CDFPilot to generate a Pilot pattern using the value specified by the Pilot Gain parameter in the NEC file instead of a predetermined value. • CDF – performs the tests using pilot, synch, paging and six traffic channels, however, the gain for the channel elements is specified in the NEC file. Advanced users may use NEC to generate a standard pattern. Instead of using the values specified by IS–97, the settings for the following NEC parameters are used: PilotGain PchGain SchGain NomGain1Way 68P09283A63-5 AUG 2007 E-9 FOA UBS Redundancy/Alarm Testing Appendix E: Optimization and Calibration Procedures UBS Redundancy/Alarm Testing ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Test Equipment Setup Follow the procedure in Procedure E-2 to set up test equipment: All alarm tests are performed using the sector 1 IDRF. Procedure E-2 Test Equipment Setup for Redundancy/Alarm Tests Connect the LMF computer to the CRMS/LMT CUSTOMER ENET connector on the SSI. Login to the UBS. Set up test equipment for TX Audit at the sector 1 IDRF TX/RX M connector (see Figure 3-23, Figure 3-24, Figure 3-25, or Figure 3-26). Display the alarm monitor by selecting Util > Alarm Monitor. Unequip all customer–defined alarms reported through the SSI Customer I/O connectors (IN 1–12 OUT 1–4 and IN 13–24 OUT 5–8) then selecting Device > Set Alarm Relays > Unequipped. During alarm reporting, spurious alarms may report. Allow the UBS to stabilize for 10 seconds. If any alarms are actively being reported after the UBS has stabilized, determine the cause before proceeding further. GPS and QHSO Redundancy/Alarm Tests Follow the procedure in Procedure E-3 to verify the manual redundancy of the GPS and QHSO boards. Verification of alarms reported is also covered. E-10 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP UBS Redundancy/Alarm Testing DO NOT perform Procedure E-3, unless the site is configured with an QHSO or MSO time base as a backup for the GPS. Procedure E-3 GPS and QHSO/MSO Redundancy/Alarm Tests Before enabling any XMI, always verify that the TX output is terminated into a 200 W non-radiating RF load! Failure to do so could result in serious personal injury and/or damage to the equipment. If the XMI is not INS (green), enable it by selecting it and then selecting Device > Unlock. Click OK to close the status report window. Key the XMI by selecting it and then selecting Device > XCVR > Key. Disconnect the cable from the SSI RGPS connector. This forces the QHSO board time base to become the CDMA timing source. Observe that a CDMA timing reference alarm and source change are reported in the Alarm Monitor. Allow the QHSO/MSO to become the active timing source, and verify the following: • The XMI remains keyed and INS. • No other modules went OOS due to the transfer to QHSO/MSO reference. Reconnect the cable to the SSI RGPS connector. Allow the GPS to become the active timing source, and verify the following: • The XMI remains keyed and INS. • No other modules went OOS due to the transfer to GPS reference. De-key the UBS by selecting Device > XCVR > Dekey. 10 Disable the XMI by selecting it and then selectingDevice > Lock. Result: The XMI color changes to yellow (OOS). 68P09283A63-5 AUG 2007 E-11 FOA Alarms Testing Appendix E: Optimization and Calibration Procedures Alarms Testing ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Alarm Verication ALARM connectors provide Customer Alarm Inputs and Outputs. The customer can connect UBS site alarm input sensors and output devices to the UBS, thus providing alarm reporting of active sensors as well controlling output devices. Alarm Reporting Display The Alarm Monitor window can be displayed to list alarms that occur after the window is displayed. To access the Alarm Monitor window, select Util>Alarm Monitor. The following buttons are included: • The Options button allows for a severity level (Warning, Minor, and Major) selection. The default is all levels. To change the level of alarms reported click on the Options button and highlight the desired alarm level(s). To select multiple levels press the Ctrl key (for individual selections) or Shift key (for a range of selections) while clicking on the desired levels. • The Pause button pauses/stops the display of alarms. When the Pause button is clicked the name of the button changes to Continue. When the Continue button is clicked, the display of alarms continues. Alarms that occur between the time the Pause button is clicked and the Continue button is clicked are not displayed. • The Clear button clears the Alarm Monitor display. New alarms that occur after the Clear button is clicked are displayed. • The Dismiss button dismisses/closes the Alarm Monitor display. E-12 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Alarms Testing Figure E-1 Alarm Connector Location CUSTOMER EQUIPMENT ti-cdma-05698-A.eps 68P09283A63-5 AUG 2007 E-13 FOA Alarms Testing Appendix E: Optimization and Calibration Procedures Purpose The following procedures verify the customer defined alarms and relay contacts are functioning properly. These tests are performed on all alarms/relays in a sequential manner until all have been verified. Perform these procedures periodically to ensure the external alarms are reported properly. Following these procedures ensures continued peak system performance. Study the site engineering documents and perform the following tests only after first verifying that the alarm cabling configuration required to interconnect the UBS frame with external alarm sensors and/or relays meet requirements called out in the 1X UBS Macro BTS Hardware Installation . Motorola highly recommends that you read and understand this procedure in its entirety before starting this procedure. Alarm input and output information Alarm connectors There are two ALARM connectors on the UBS Macro SSI: • CUSTOMER IP 1-12 OP 1-4 connector • CUSTOMER IP 13-24 OP 5-8 connector Each ALARM connector provides 12 inputs and 4 outputs. A total of 24 inputs and 8 outputs are available. Function ALARM connectors provide for Customer Alarm Inputs and Outputs. The customer can connect UBS site alarm input sensors and output devices to the UBS, thus providing alarm reporting of active sensors as well as controlling output devices. The SSI detects signals from customer input sensors and reports the detected signals to the DMI controller, which in turn reports the detected alarm to the OMC-R, where it is displayed as defined by the customer. The SSI also provides switched relay contacts to customer output devices. The SSI controls relay contacts according to output control signals from the DMI controller, in response to customer defined commands entered at the OMC-R. E-14 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Alarms Testing Input connectors Each input consists of a wire/pin pair (that is, input/Gnd). To ensure proper operation, each pair to be used must be connected to an external sensor that provides a dry-contact closure. The customer sensor output connects between an optically isolated 5 VDC signal and an isolated return. For an OPEN circuit (logic 0) between pins: • There is 10 K Ohms or greater across the input pair. • The signal to ground voltage is +5 VDC. For a CLOSED circuit (logic 1) between pins: • There is 1 K Ohms or less across the input pair. • The signal to ground current is a maximum of 9 mA. Either of the above states can be defined by the customer in system software as an alarm condition. Output connectors Each output consists of 3-wires/pins: • COM • NC • NO The customer output device control inputs connect between the common (COM) and either the normally closed (NC) or normally open (NO) contacts of a relay. The NC and NO state for relay contacts occurs when the relay coil is not energized. Relay contacts are load rated for a maximum of 1 A at 24 VDC and 0.5 A at 50 VDC. The toggling of relay contacts to the opposite state is controlled by system software. Figure E-2 Alarm Connector Pin Numbering 19 LOOSE WIRES, NO CONNECTOR 37 37-Pin-Type (Male) 20 ti-cdma-05692.eps 68P09283A63-5 AUG 2007 E-15 FOA Alarms Testing Appendix E: Optimization and Calibration Procedures Customer Alarm Input Verication Procedure E-4 describes how to test the Customer alarm input verification. Follow the steps as instructed and compare results with the LMF display. It may take a few seconds for alarms to be reported. The default delay is 5 seconds. When shorting alarm pins wait for the alarm report before removing the short. Procedure E-4 Customer Alarm Input Verication Connect the LMF to the UBS and log into the UBS. Select the Controller. Click on the Device menu. Click on the Set Alarm Relays menu item. Click on Normally Open. A status report window displays the results of the action. Click on OK to close the status report window. Refer to Figure E-2 and sequentially short the ALARM A connector CA IN 1 through CA IN 12 pins together. Sequentially short the ALARM B connector CA IN 13 through CA IN 24 pins together. An alarm should be reported for each pair of pins that are shorted. A clear alarm should be reported for each pair of pins when the short is removed. Select the Controller. Click on the Device menu. 10 Click on the Set Alarm Relays menu item. 11 Click on Normally Closed. A status report window displays the results of the action. 12 Click on OK to close the status report window. Alarms should be reported for alarm inputs 1 through 24. 13 Sequentially short the ALARM B connector CA IN 13 through CA IN 24 pins together. A clear alarm should be reported for each pair of pins that are shorted. An alarm should be reported for each pair of pins when the short is removed. 14 Select the Controller. 15 Click on the Device menu. E-16 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Procedure E-4 Alarms Testing Customer Alarm Input Verication (Continued) 16 Click on the Set Alarm Relays menu item. 17 Click on Unequipped. A status report window displays the results of the action. 18 Click on OK to close the status report window. 19 Refer to Figure E-2 and sequentially short the ALARM A connector CA IN 1 through CA IN 12 pins together. Sequentially short the ALARM B connector CA IN 13 through CA IN 24 pins together. No alarms should be displayed. 20 Load data to the Controller to reset the alarm relay conditions according to the NEC file. Pin and Signal Information for Alarm Connectors Table E-1 and Table E-2 lists the pins, wire color codes, and signal names. Table E-1 Connector Pinout for Cable T IP 1–12 OP 1–4 Pin Name Recommended Wire Color for Cable CA IN 1 White/Blue CA IN 2 White/Orange CA IN 3 White/Green CA IN 4 White/Brown CA IN 5 White/Gray CA IN 6 Red/Blue CA IN 7 Red/Orange CA IN 8 Red/Green CA IN 9 Red/Brown 10 NC NA 11 CA NO 1 Black/Blue 12 CA NO 2 Black/Orange 13 CA NC 2 Black/Green 14 CA NO 3 Black/Brown 15 CA NO 4 Black/Gray 16 CA NC 4 Yellow/Blue 17 CA IN 10 Yellow/Orange 18 CA IN 11 Yellow/Green 19 CA IN 12 Yellow/Brown 68P09283A63-5 AUG 2007 E-17 FOA Alarms Testing Table E-1 Appendix E: Optimization and Calibration Procedures Connector Pinout for Cable T IP 1–12 OP 1–4 (Continued) Pin Name Recommended Wire Color for Cable 20 GND 1 Blue/White 21 GND 2 Orange/White 22 GND 3 Green/White 23 GND 4 Brown/White 24 GND 5 Gray/White 25 GND 6 Blue/Red 26 GND 7 Orange/Red 27 GND 8 Green/Red 28 GND 9 Brown/Red 29 CA NC 1 Blue/Black 30 CA COM 1 Orange/Black 31 CA COM 2 Green/Black 32 CA NC 3 Brown/Black 33 CA COM 3 Gray/Black 34 CA COM 4 Blue/Yellow 35 GND 10 Orange/Yellow 36 GND 11 Green/Yellow 37 GND 12 Brown/Yellow CA IN— Customer Alarm signal into the M810 BTS, NO — Normally Open, NC — Normally Closed, COM — Common Table E-2 Connector Pinout for Cable T IP13–24 OP5–8 Pin Name Recommended Wire Color for Cable CA IN 13 White/Blue CA IN 14 White/Orange CA IN 15 White/Green CA IN 16 White/Brown CA IN 17 White/Gray CA IN 18 Red/Blue CA IN 19 Red/Orange CA IN 20 Red/Green CA IN 21 Red/Brown 10 NC NA E-18 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Table E-2 Alarms Testing Connector Pinout for Cable T IP13–24 OP5–8 (Continued) Pin Name Recommended Wire Color for Cable 11 CA NO 5 Black/Blue 12 CA NO 6 Black/Orange 13 CA NC 6 Black/Green 14 CA NO 7 Black/Brown 15 CA NO 8 Black/Gray 16 CA NC 8 Yellow/Blue 17 CA IN 22 Yellow/Orange 18 CA IN 23 Yellow/Green 19 CA IN 24 Yellow/Brown 20 GND 13 Blue/White 21 GND 14 Orange/White 22 GND 15 Green/White 23 GND 16 Brown/White 24 GND 17 Gray/White 25 GND 18 Blue/Red 26 GND 19 Orange/Red 27 GND 28 Green/Red 28 GND 21 Brown/Red 29 CA NC 5 Blue/Black 30 CA COM 5 Orange/Black 31 CA COM 6 Green/Black 32 CA NC 7 Brown/Black 33 CA COM 7 Gray/Black 34 CA COM 8 Blue/Yellow 35 GND 22 Orange/Yellow 36 GND 23 Green/Yellow 37 GND 24 Brown/Yellow CA IN— Customer Alarm signal into the M810 BTS, NO — Normally Open, NC — Normally Closed, COM — Common 68P09283A63-5 AUG 2007 E-19 FOA E-20 68P09283A63-5 FOA AUG 2007 Appendix MSO Calibration 68P09283A63-5 AUG 2007 F-1 FOA MSO Calibration Appendix F: MSO Calibration MSO Calibration ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ MSO Calibration Status The check the status of the Motorola Standard (Medium Stability) Oscillator (MSO) perform the following: • Log into the UBS with the LMF. • Click on BTS > MSO Calibration Status menu item • MSO status is displayed. MSO Calibration Procedure Perform the following procedure only if the BTS is in Island Mode configuration, the system has no configured references, and HSO, GPS, or external reference is connected. Procedure F-1 Procedure to Calibrate the MSO Log into the UBS. Click on BTS Menu. SelectEnter Invasive Mode (Figure F-1). Wait for Message window to appear (Figure F-2). From BTS Menu, select MSO Calibration Status (Figure F-3). MSO status is displayed (Figure F-4) From BTS Menu, select MSO Calibration Figure F-5) Select Clock reference HSO-1 to start MSO calibration (Figure F-6). Wait for start (Figure F-7) The calibration takes about 15 minutes to complete (it can take up to an 60 minutes). Read the calibration status (Figure F-8) Logout of UBS. Click on File->Exit (Figure F-9). Login to UBS (do not enter invasive mode). From BTS Menu, select MSO Calibration Status (Figure F-10). F-2 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP MSO Calibration Figure F-1 BTS Menu - Enter Invasive Mode Figure F-2 Message Window Figure F-3 BTS Menu - MSO CAL Status Figure F-4 Read MSO Cal Status Figure F-5 Select MSO Calibration Figure F-6 Select Clock Reference Figure F-7 MSO Calibration (Invasive Mode) Figure F-8 Read MSO Status Figure F-9 Exit Local Terminal Figure F-10 MSO Status (Non-Invasive Mode) 68P09283A63-5 AUG 2007 F-3 FOA F-4 68P09283A63-5 FOA AUG 2007 Appendix SSI Loopback Connector Information 68P09283A63-5 AUG 2007 G-1 FOA SSI Loopback Connector Appendix G: SSI Loopback Connector Information SSI Loopback Connector ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ There are two loopback connectors required, one for the Span and one for the Customer Input / Output. This appendix section provides the information necessary for making a 37 pin D-sub for the span and customer input / output connectors. Purpose The purpose of the SSI loopback connector is to enable the user to determine if the receive and transmit lines of the span cable circuitry are operating correctly. Required Parts Should a 37 pin D-sub loopback connector not be available, one can be made by the user to interface with the 37 pin D-sub on the SSI board. Table G-1 Parts Required to Fabricate Loopback Connector Item Part Number Description Qty Customer Supplied 37 pin, D-sub plug Figure G-1 Loopback Connector 37-Pin-Type (Male) 19 20 37 ti-cdma-05692-A.eps SSI Span Loopback Connector The pinouts for the SSI Span Loopback Connector are listed in Table G-2. Recommended wire is 22 AWG, solid or 24 AWG, stranded. G-2 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP Table G-2 SSI Loopback Connector SSI Span Loopback Connector Pinouts Pin # Signal Description Loopback Connection Pin Pin # Signal Description Loopback Connection Pin TX1 TIP 12 20 TX1 RING 30 TX2 TIP 13 21 TX2 RING 31 TX3 TIP 14 22 TX3 RING 32 TX4 TIP 15 23 TX4 RING 33 TX5 TIP 16 24 TX5 RING 34 TX6 TIP 17 25 TX6 RING 35 TX7 TIP 18 26 TX7 RING 36 TX8 TIP 19 27 TX8 RING 37 GND 28 GND 10 GND 29 GND 11 GND 30 RX1 RING 20 12 RX1 TIP 31 RX2 RING 21 13 RX2 TIP 32 RX3 RING 22 14 RX3 TIP 33 RX4 RING 23 15 RX4 TIP 34 RX5 RING 24 16 RX5 TIP 35 RX6 RING 25 17 RX6 TIP 36 RX7 RING 26 18 RX7 TIP 37 RX8 RING 27 19 RX8 TIP 68P09283A63-5 AUG 2007 G-3 FOA SSI Loopback Connector Appendix G: SSI Loopback Connector Information Figure G-2 SSI Span Loopback Connector Wiring Diagram 20 12 30 19 37 ti-cdma-05692-C.eps SSI Customer Input / Output Loopback Connector The pinouts for the SSI CIO Loopback Connector are listed in Table G-3. Recommended wire is 22 AWG, solid or 24 AWG, stranded. G-4 68P09283A63-5 FOA AUG 2007 1X UBS Macro BTS Optimization/ATP SSI Loopback Connector Table G-3 SSI CIO Loopback Connector Pinouts Pin # Signal Description Loopback Connection Pin Pin # Signal Description Loopback Connection Pin CA IN 1 30 20 GND 13 CA IN 2 30 21 GND CA IN 3 30 22 GND CA IN 4 31 23 GND CA IN 5 31 24 GND CA IN 6 31 25 GND CA IN 7 33 26 GND CA IN 8 33 27 GND CA IN 9 33 28 GND 29 10 NC 29 Relay NC 1 28 11 Relay NO 1 NC 30 Relay COM 1 1,2,3 12 Relay NO 2 NC 31 Relay COM 2 4,5,6 13 Relay NC 2 20 32 Relay NC 3 26 14 Relay NO 3 NC 33 Relay COM 3 7,8,9 15 Relay NO 4 NC 34 Relay COM 4 17,18,19 16 Relay NC 4 23 35 GND 17 CA IN 10 34 36 GND 18 CA IN 11 34 37 GND 19 CA IN 12 34 68P09283A63-5 AUG 2007 16 32 G-5 FOA SSI Loopback Connector Figure G-3 Appendix G: SSI Loopback Connector Information SSI CIO Loopback Connector Wiring Diagram 20 23 19 13 26 34 ti-cdma-05692-B.eps G-6 68P09283A63-5 FOA AUG 2007 Index Index ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ Abbreviations and Acronyms Computer Requirements, Minimum . Acceptance Testing . . . . . . . . . . Advantest R3267 GPIB Address . . . . . . . . . . . . Advantest R3465 Calibrating Test Cable . . . . . . . Advantest R3562 GPIB Address . . . . . . . . . . . . Agilent 8935 Series E6380 (formerly HP 8935) Test Set GPIB Address . . . . . . . . . . D-14 Agilent E4406A calibration . . . . . . . . . D-3, D-21 to D-22 GPIB Address . . . . . . . . . . . . . . . D-7 Agilent E4432B Signal Generator GPIB Address . . . . . . . . . . . . . . . D-9 ATP Test Matrix FRU Optimization . . . . . . . . . . . . . B-2 . . 1-13 . . . 1-5 . . . . . . . D-11 D-2 D-13 Backhaul Configuration Procedure . . . . . Bay Level Offset Calibration 5-2 Bay Level Offset Calibration (contd.) Calibration, BLO . . . . . . . . . . . . . E-6 Cable Calibrating . . . . . . . . . . . . Setting Loss Values . . . . . . . . Cable Calibration Manual . . . . . . . . . . . . . . Cable Calibration Set-up Automatic . . . . . . . . . . . . . Calibrate Test Cabling Communications System Analyzer Calibrating Cables . . . . . . . . . . . . . . Test Equipment . . . . . . . . . . Calibrating Test Cable Advantest R3465 . . . . . . . . . . . . . . . 3-64 3-72 . . . 3-35 . . . 3-35 . . . 3-64 . . . . . . 3-64 3-56 . . . . Calibrating Test Cabling Calibrating Test Cabling using Communications System Analyzer . . . . . . . . . . . . 3-69 calibration Gigatronics 8542B . . . . . . . . . . . D-22 Calibration . . . . . . . . . . . . . . . . . 1-4 Calibration Procedures Included Automatic . . . . . . . . . . . . . . . . 3-56 Calibration, Test Cable . . . . . . . . . . . 1-7 Calibration, Test Equipment . . . . . . . . 1-7 Code Domain ATP Noise Floor ATP . . . . . . . . . . . . . 4-21 Continuous Waveform, Lock. . . . . . . . 4-25 Continuous Waveform, Unlock . . . . . . 4-24 D-2 Documents, Required . . . . . . . . . . . 1-13 E4406A 68P09283A63-5 AUG 2007 IX-1 FOA Index E4406A (contd.) calibration . . . . . . . . . D-3, D-21 to D-22 FER ATP . . . . . . . . . . . . . . . . . 4-23 Generate ATP Report . . . . . . Gigatronics 8541C power meter Gigatronics 8542 power meter calibration . . . . . . . . . . Gigatronics 8542B power meter illustration . . . . . . . . . . GPIB power meter Gigatronics 8542B . . . . . . . . . . 4-26 . . . . . . D-3 . . . . . D-22 . . . . . D-16 . . . . . D-16 GPIB Address Advantest R3267 Advantest R3562 Agilent (formerly Agilent E4406A . Agilent E4432B . . . . . HP) . . . . . . . . . . 8935 . . . . . . . D-11 . D-13 . D-14 . . D-7 . . D-9 IP Addresses Agilent E7495A . . . . . . . . . . . . . 3-47 LMF Hardware Requirements Computer Requirements, Minimum . . . . LMF Operation . . . . . . . . . . . . . . . 3-1 1-9 MSO Calibration Calibration MSO . . . . . . . . . . . . . MSO Calibration Procedure Calibration Procedure, MSO . . . . . . . F-2 F-2 Optimization and Calibration Procedures . . Optimization, Periodic . . . . . . . . . . . Optimization/ATP Checklist CHecklist, Optimization/ATP . . . . . . . E-1 1-6 Optimization/ATP Setup . . . . . . . . . . 3-40 Optional Full ATP Test Matrix Full ATP Test Matrix . . . . . . . . . . . B-2 A-2 Pilot Time Offset ATP IX-2 68P09283A63-5 FOA AUG 2007 Index Pilot Time Offset ATP (contd.) PTO ATP . . . . . . . . . . . . . . . . Power meter GPIB Gigatronics 8542B . . . . . . . . . . Power meter (contd.) 4-19 illustration Gigatronics 8542B . . . . . . . . . . D-16 D-16 Reduced ATP . . . . . . . . . . . . . . . . B-2 Reference Documents . . . . . . . . . . . 1-13 Rho ATP Rho ATP (contd.) TX Waveform Quality ATP . . . . . . . . RS232 GPIB Interface Box . . . . . . . . 4-18 D-20 Selecting Test Equipment . . . . . . . . . Setting Cable Loss Values . . . . . . . . . Setup for Optimization/ATP . . . . . . . . 3-47 3-72 3-40 Signal Generator . . . . . . . . . . . 3-71, 3-4 Spectrum Analyzer . . . . . . . . . . 3-71, 3-4 Spectrum Analyzer . . . . . . . . . . . 3-33 Test Equipment Automatically Selecting . . . . 3-50, 3-52, Calibrating . . . . . . . . . . . . . . . Connecting test equipment to the BTS. . Selecting . . . . . . . . . . . . . . . . Test Equipment GPIB Address Settings GPIB Address Settings, Test Equip ment . . . . . . . . . . . . . . . . . . Test Equipment Preparation . . . . . . . . Test Equipment Setup, Agilent E7495 Agilent E7495 Setup . . . . . . . . . Test Equipment, Supported . . . . . . . Test Set Calibration Background . . . . . . . . . . . . . . TX and RX ATP . . . . . . . . . . . . . TX Mask ATP TX Spectral Purity Transmit Mask ATP. 3-54 3-56 3-32 3-48 3-32 D-2 . . D-4 3-33 3-55 . . 4-4 4-17 UBS Component Identification . . . . . . 1-21 Warm-up Equipment . . . . . . . . . . . 3-34 68P09283A63-5 AUG 2007 IX-3 FOA
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.6 Linearized : No Encryption : Standard V2.3 (128-bit) User Access : Print, Print high-res XMP Toolkit : Adobe XMP Core 4.0-c316 44.253921, Sun Oct 01 2006 17:14:39 Modify Date : 2007:10:11 10:04:20-06:00 Create Date : 2007:10:11 09:54:27-06:00 Metadata Date : 2007:10:11 10:04:20-06:00 Creator Tool : Adobe Acrobat 8.1 Combine Files Format : application/pdf Title : Creator : Motorola Gold Disk User Document ID : uuid:b659c099-3ec3-4f72-8a84-a0ab2a2d0a4f Instance ID : uuid:c51a817c-a347-4fd4-b30f-52c3fd608a4b Producer : Adobe Acrobat 8.1 Page Count : 101 Author : Motorola Gold Disk UserEXIF Metadata provided by EXIF.tools