OSANG Healthcare IGM-1001D Lipid profile and glucose measuring system for self testing User Manual 1

Infopia Co.,Ltd. Lipid profile and glucose measuring system for self testing 1

User Manual

Download: OSANG Healthcare IGM-1001D Lipid profile and glucose measuring system for self testing User Manual 1
Mirror Download [FCC.gov]OSANG Healthcare IGM-1001D Lipid profile and glucose measuring system for self testing User Manual 1
Document ID2419965
Application IDkscz11l4fPP6a1qRagg2kA==
Document DescriptionUser Manual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize119.59kB (1494864 bits)
Date Submitted2014-10-16 00:00:00
Date Available2014-10-16 00:00:00
Creation Date2014-09-29 17:16:50
Producing SoftwareMicrosoft® Word 2010
Document Lastmod2014-09-29 17:16:50
Document Title1
Document CreatorMicrosoft® Word 2010
Document Author: 남증희

Operation Manual
Dear IGM-1001D Blood Glucose Monitoring System
Owner,
Thank you for choosing the IGM-1001D Blood Glucose
Monitoring System. This manual contains everything you
need to know about your new glucometer and how it works.
Please take a moment to read the instructions carefully.
We understand that self-testing of blood glucose level
provides a way to control your diabetes and may give you
peace of mind by testing regularly. As a result, IGM-1001D
has been developed to provide you with a fast and accurate
reading in a convenient and simple process. Our goal for the
IGM-1001D Blood Glucose Monitoring System is to provide
you with the best quality healthcare products coupled with
superior customer service. Always consult with your
healthcare professional, before making any changes to your
diabetes management. The IGM-1001D Blood Glucose
Monitoring System is for in-vitro diagnostic use only.
Customer Service is available 24 hours a day, 7 days a week,
and 365 days a year. Please call customer support center in
your local area.
Important Information
The IGM-1001D Blood Glucose Monitoring System is
intended for use outside the body (in-vitro diagnostic use
only). It should be used only for testing blood glucose and
only with fresh capillary whole blood samples. It should not
be used for the diagnosis of diabetes.
Consult with your physician or diabetic healthcare
professional about daily management of your diabetes and
proper use of the glucometer.
Please pay close attention when handling blood. Improper
procedures may cause serious hazards to your health.
The IGM-1001D Blood Glucose Monitoring System contains
small parts, which could be a choking hazard for children if
swallowed.
WARNING:
1. All parts of the kit are considered bio-hazardous and can
potentially transmit infectious diseases, even after you
have taken cleaning and disinfection measures.
2. Always use a new, sterile lancet. Lancets are for single
use only.
3. Avoid getting hands lotions, oil, dirt or debris in and/or on
the lancets and reusable lancing device.
4. Please refer to page 12 for cleaning and disinfecting IGM1001D Blood Glucose Monitoring System.
5. Do not change your medication based on the IGM-1001D
meter test results without contacting your physician or
healthcare professional first.
6. Do not assemble or disassemble the blood glucose
test meter
7. Keep out of direct sunlight.
CAUTION:
Following list of IMPORTANT SAFETY INSTRUCTIONS provided:
1. Close supervision is necessary when equipment is used
by, on, or near children, handicapped persons or invalids.
2. Do not place the equipment in liquid, nor put it where it
could fall into liquid. If the equipment becomes wet,
remove the batteries and dry the device.
3. Use the equipment only for the purpose described in the
instructions for use.
4. Do not use accessories which are not supplied or
recommended by the manufacturer.
5. Do not place anything on top of the equipment.
Test Principle
IGM-1001D blood glucose test strip is the bio-sensor which
has been made by the FAD-dependent glucose
dehydrogenase(enzyme). FAD-GDH(enzyme) quantitatively
reacts with the glucose in the whole blood and the mediator.
34At this moment, mediator Fe(CN)6 is reduced to Fe(CN)6 .
4This Fe(CN)6 generated by the reaction is re-oxidized on
carbon electrode, for which the method is called
amperometry. The measured oxidation current exists in
proportion to the glucose concentration in the whole blood.
INFOPIA CO., LTD
Page 1
UM-M (Rev.0)
Learning the System
Table of Contents
Intended use of Blood Glucose Test Meter
The IGM-1001D Blood Glucose Monitoring System is
intended for use outside the body only (in-vitro
diagnostic use) and self-testing. Recorded results
make it easier for your doctor to access your
metabolic control. However, the self-testing should not
take the place of overall diabetes management
assessment from your doctor. It should be used for
testing blood glucose (sugar) only with fresh capillary
whole blood samples. IGM-1001D Blood Glucose
Monitoring System is intended for use in places like
home and in professional setting to monitor blood
glucose levels.
Learning the System
Intended use of blood glucose test
meter
IGM-1001D kit components
IGM-1001D meter
IGM-1001D meter Display
IGM-1001D test strip
--- 2
--- 3
--- 3
--- 3
--- 4
Before Testing:
Setting your meter
Auto-coding function
Performing a control solution test
--- 4
--- 5
--- 5
CAUTION
- Do not use system for any other purpose than blood
glucose test.
- Cholesterol concentrations > 500 mg/dL or triglyceride
concentrations > 3,000 mg/dL may produce elevated
reading.
- Icodextrin does not interfere with this version of
IGM-1001D test strips.
- Do not use during or soon after xylose absorption
testing. Xylose in the blood will cause interference.
- Before using any product, please read all
instructions.
- Consult with your physician or diabetic healthcare
professional about daily management of your
diabetes and proper use of the glucometer. If you
have any questions about the IGM-1001D product,
please contact your local representative.
Performing Your Test:
Testing Blood - preparation
Performing a blood test with your
finger tip
Test strip ejector function
--- 6
--- 7
--- 9
Reviewing Your Results:
Reviewing your blood sugar
Reviewing your averages
---9
--10
Alternate Site Test:
Lancing and sampling from an
alternate site area
Lancing device clear cap is used for
alternate site testing
--10
--11
WARNING
- Before using IGM-1001D system, read all
instructions in this manual, and practice for accurate
and safe test. You should have commentary
recommendation
from
your
diabetes
care
professional for the proper use of the meter and daily
management of your diabetes.
- If you think your blood glucose results are too low or
too high, or if the results are doubtful, please contact
your doctor.
- If your blood glucose result is unusually low or high,
or you do not feel the result is correct, repeat the test
again with a new test strip.
- If the results are still inconsistent, please consult
with your physician before making any decision in
controlling your diabetes.
Caring for Your IGM-1001D System:
Cleaning your meter and
maintenance
Storage of your IGM-1001D system
Transferring Your Result Data
Download Application
Battery Installation
Dispose of meter and batteries
Trouble-shooting
Warranty
Specifications
Symbol Reference
INFOPIA CO., LTD
--12
--12
--12
--13
--13
--13
--14
--15
--15
--15
Page 2
UM-M (Rev.0)
IGM-1001D Kit Components
1. IGM-1001D Meter
2. IGM-1001D Test Strip
3. Reusable Lancing Device
4. Lancets
5. Operation Manual
6. Warranty Registration
Card
7. Carrying Case
8. Patient Logbook
9. 2EA of 3V Coin
Battery (CR2032)
Your IGM-1001D Blood Glucose Monitoring System has
been sealed to protect the contents.
If you find your seal has been broken, please return it to the
place of purchase.
CAUTION
1. Do not use IGM-1001D glucose test meter in a dry
environment, especially if synthetic materials are present.
Synthetic clothes, carpets, etc., may cause damaging
static discharge in a dry environment.
2. Do not use IGM-1001D glucose test meter near cellular
or cordless telephones, walkie-talkies, garage door
openers, radio transmitters, or other electrical equipment
that are sources of electromagnetic radiation, as these
may interfere with the proper operation of the glucose
test meter.
IGM-1001D meter display
Low Battery Warning Symbol
IGM-1001D meter
Alarm Symbol
Test Strip Port:
Insert the IGM-1001D test strip
Troubleshooting Symbol
Delete Results
Display Screen:
Displays your test results,
symbols and messages.
Average Test Result System
Before Having a Meal
After Having a Meal
Power/Enter Button:
Power the meter on/off, prompts
memory data and acts as an
“Enter” button during setup and
after tests.
After Taking Medication
After Sport Activity
Control Solution Symbol
Blood Drop Symbol for Test
Strip Code Symbol
Up/Down Arrow Buttons:
Scroll up or down to adjust
settings or retrieve memory.
Test Result Unit Symbol

Temperature / Date
Time / Test Result Amount
IGM-1001D test strip
Test Strip Ejector:
Releases and discards test strip
Serial Number:
Battery Cover:

INFOPIA CO., LTD
Page 3
Number
Name
Function
Entrance
Blood goes into this area.
Air exit
As blood enters, air exits thru this area
Confirmation
Able to see through this window whether
Window
blood is fully filled.
Connector
Connect between the electrodes and meter
Auto-coding label
Automatic coding recognition
UM-M (Rev.0)
Setting Your Meter
CAUTION
1. Store the IGM-1001D test strip vials in a cool, dry place.
Keep out of direct sunlight. Do not freeze.
2. Store test strips in their original vial only. Do not mix the
test strips in new vials or in any other container.
3. Immediately replace the vial cap and close tightly after
removing any test strips from the vial.
4. Make a note of the expiry date, which is three months
from the date you first open a new vial of strips. Throw
IGM-1001D test strips and vial away after the expiry date.
5. Do not use the test strips after the expiration date printed
on the package or vial, since it may cause inaccurate
results.
6. IGM-1001D test strips should be single use only. DO
NOT RE-USE.
7. Do not test at temperatures below 10℃(33.8°F) or above
40℃(104°F).
8. Do not test in the condition that humidity is below 10% or
above 90%.
9. Do not bend, cut, or alter the test strip.
10. Avoid getting dirt, food, and water on the test strip with
wet hands.
11. Avoid getting dirt, food, and water on the color-coding
label (backside of test strip)
12. Refer to additional information in the IGM-1001D test
strip package.
The IGM-1001D meter has a wide variety of functions to
choose from. From the setup mode you have the ability to
turn activity/meal flags, set the date/time, designate three
unique averages and set up to five daily alarms.
Power On
Press
for 3
seconds at least
User Activity Option
After pressing and
releasing either ▲ or
▼ buttons to turn the
User Activity Option
on/off, confirm with
Year
After pressing and
releasing either▲ or
▼ buttons to set the
year, confirm with
Date / Time
After pressing and
releasing either▲ or
▼ buttons to set the
date and time,
confirm each with
Number of days for
average
After pressing and
releasing either▲ or
▼ buttons to set the
number of days for
average calculation (3
different possible),
confirm with
Alarm on/off
After pressing and
releasing either▲ or
▼ buttons to turn the
alarm on/off, confirm
with
Alarm time
After pressing and
releasing either ▲ or
▼ buttons to set the
alarm hour and
minute, confirm with
Alarm
After programming
the number of desired
alarms (5 alarms
possible), confirm
with
Before Testing
WARNING
To reduce the chance of infection:
1. Lancing device and lancets should NOT be shared with
others.
2. Always use a new lancet and a new blood glucose test
strip.
3. Practice using the lancing device and become
accustomed with its use.
4. Lancets and blood glucose test strips should be single
use only.
5. Wash your hands in warm clean running water using
soap before testing.
6. Avoid getting hand lotion, oils, dirt or debris on the
lancets or on the lancing device.
7. Dry your hands completely before testing.
INFOPIA CO., LTD
Page 4
CAUTION
1. VERY IMPORTANT: Set the right test unit.
Your IGM-1001D Meter is pre-set to mg/dL as test unit. A
wrong test unit will read to wrong test result display.
Please consult your healthcare professional if you’re not
sure which test unit to use. Without setting the date
properly, the average glucose level and the results in
memory will not show proper values. It is recommended
to set the glucose test meter before use and also when
new batteries are installed.
2. You cannot test your blood glucose in the setting mode.
3. To turn the meter off during the setup, press
for at
least 3 seconds.
UM-M (Rev.0)
Auto-coding Function
Performing a Control Solution Test
1. IGM-1001D has automatic
code recognition function(e.g.
C26)
IGM-1001D control solutions contain a known amount of
glucose that reacts with IGM-1001D test strip. By comparing
your control solution test results with the expected range
printed on the test strip vial, it is able to check the meter and
test strips are working properly and that you are performing
the test correctly. It is very important that you perform the
check routinely to make sure you get accurate results.
2. It recognizes the code number
automatically. This is a very
convenient function preventing
the inconvenience of setting the
code number on the glucose test
meter every time. Check the code
on the vial to see if the codes
match each other
The Glucose control solution should be used
 Before using a new box of strips
 Whenever you suspect the blood glucose test meter or
blood glucose test strip is not functioning properly.
 If your blood glucose test results are not consistent with
your symptoms or if you think they are not accurate.
 If you have dropped the blood glucose test meter.
 For quality control in the point of care usage.
 For teaching or learning the system.
 When using the meter for blood glucose tests after any
disinfection procedure.
CAUTION
1. Do not bend the glucose test strip to prevent the
automatic code recognition failure.
2. If the code recognition label is damaged, code
recognition may fail. Please check the code number on
the LCD window together with the code number on the
glucose test strip vial to see if code numbers match each
other.
Firmly insert the test strip into the meter
test port. Insert down firmly in direction of
arrow on test strip (arrow up). Please do
not insert the glucose test strip upside
down.
When you insert the glucose test strip
into the glucose test meter, the power
automatically adjusts the meter
accordingly.
IGM-1001D
26
Backlight Function
When inserting IGM-1001D
glucose test strip (starting
measuring process), backlight
will turn on for 10 seconds.
Before inserting blood, Scroll through the
icons with the ▲ or ▼ button until the
control solution bottle is blinking, press
After the test result displays,
backlight will last for 10 seconds.
By pressing the power button
(moving into memory mode),
backlight will turn on for 10
seconds.
1. Check the expiration date before
performing a control solution test.
Do not use if expired. Please note
the expiration date on the control
solution bottle.
When pressing the power button
for 3 seconds (entering the
setting mode), backlight will turn
on for 10 seconds.
2. Gently shake the control solution
before use.
When inserting the PC cable into
the USB port, backlight will turn
on for 10 seconds.
3. Discard the 1st drop of control
solution, this will eliminate any
residue.
Place a drop of control solution on
a clean dry surface (e.g. the lid of
the test strip vial)
INFOPIA CO., LTD
Page 5
UM-M (Rev.0)
Please call customer support toll: 82-31-460-0300
4. Dip the test strip into the control
solution.
Results will appear in 5 seconds.
Compare the result to the range
printed on the test strip vial, the
result should fall within that range
CAUTION
1. Control solutions is for in-vitro diagnostic use only. This
means that it is only used for testing outside the body.
2. Control solution is exclusively made for the IGM-1001D
Blood Glucose Monitoring System. Do not use other brand
control solutions made by other manufacturers.
3. Check the expiration date on the vial. Do not use, if
expired.
4. Once opened, use it for the next three months only.
Record the discard date on the control solution bottle.
Discard after three months.
5. Do not swallow. This is NOT for human or animal
consumption.
6. The control solution bottle icon contains letter ‘C,’ so as
not to be confused with the medication bottle.
This will allow you to understand differences between a
control solution test and an actual blood test.
For example only :
If you are using the normal control solution, according to this
particular vial of strips, your meter should show a number
between 80 and 120.
IGM-1001D
IGM-1001D
H. 255-345
N. 80-120
L. 35-65 IGM-1006
IGM-1001D
IGM-1001D
Follow-up Action
1. Check your meter and strip with the glucose control
solution.
2. Perform your test again.
CAUTION
1. If your glucose control solution falls out of range, do not
perform a glucose test and please contact your local
distributor.
2. The glucose control solution’s measurement range has
nothing to do with individual’s blood glucose level, as its
own purpose is to check the status of strip and meter.
Discard the used control solution and test strips carefully,
according to the local regulation.
Performing Your Test
Testing Blood - Preparation
Lancing device is a convenient medical tool for collecting
capillary blood samples for glucose monitoring or other
tests which require one or two drops of blood.
Being a more advanced lancing device, it provides
ultimate safety and comfort for obtaining blood samples.
An adjustable tip offers 5 different strength levels of skin
penetration for individual user’s comfort, while a lancet
ejector enables the safe disposal of the used lancet. As
an added bonus, this also comes with a transparent tip
for AST (Alternative Site Testing).
CAUTION
1.
Please note that the control values shown in the picture are not
control ranges.
2.
NOTE:
- It is recommended that glucose control solution is stored at room
temperature 20~25℃ (68~77℉) before testing.
- Check the expiration date before performing a glucose control
solution test.
- Do not use if it expired. Once opened, the control solution expires
after 3 months.
3.
4.
If your control solution test falls out of range, please follow
the next steps before contacting customer support:
 Check the expiration dates on all the products you are
using. Be sure you are using IGM-1001D control solution.
 Try another control solution test.
If this test falls out of range, try another control solution
test with a new unopened bottle of strips.
 After following the appropriate steps and the control test
still falls out of range, do not perform a glucose test.
7.
INFOPIA CO., LTD
Page 6
5.
6.
Never use a lancet that has been used by someone else. This
could lead to contamination.
If the lancing device is to be used by another person, the unit
must use a new tip or properly disinfected
Do not leave the lancet in the device after use.
A new lancet must only be placed into the lancing device
directly before testing.
This device has many small parts and could be a choking
hazard for children if swallowed.
In the case of hospital use, hospitals need to observe their
own infection control protocols in order to avoid any
contamination.
Always dispose of the used lancet in a biohazard garbage
container.
UM-M (Rev.0)
Performing a blood test with your fingertip
Components
Step 1
1. Firmly insert the test strip into the
meter test port. Insert down firmly in
direction of arrow on test strip (arrow up).
Please do not insert the glucose test strip
upside down.
When you insert the glucose test strip
into the glucose test meter, the power
automatically turns on with the code and
temperature.
Using the Clear Tip(AST)
1. Place the lancing device and lancet on the skin in the
chosen area, press and hold it continuously for a few
seconds, then push the release button to take a blood
sample.
2. Watch through the clear tip until a sufficient blood sample
is taken. If there is not enough blood, gently massage the
area until a sufficient sample is collected.
STORAGE
Products must be stored at room temperature protected
from sunlight or moisture.
CLEANING AND DISINFECTION
-Clean and wipe the outer part of lancing device once a
week at the minimum with a soft cloth dampened with mild
soap and water. Wash the adjustable tip and clear tip, once
a week with mild soap.
-DO NOT immerse the lancing device in water or liquid.
WARRANTY
Lancing device has a 2-year warranty from the date of
purchase.
A malfunctioning device should be returned to the following
address:
Manufacturer :
GMMC
Room No. 1112, Ace Tower 9th Bldg., 345-30, Gasandong, Geumcheon-gu, Seoul, South Korea
Tel : +82 2 2104 0470
Fax : +82 2 2104 0472
EU Authorized Representative :
GMMC, S.L.
C/Vinalopo' 4 - 1° - Pta 4 Valencia, Spain
Tel : +34 963 691 484
Fax: +34 961 125 949
INFOPIA CO., LTD
2. Make sure the code number matches
on the strip bottle its code (e.g. C26).
The IGM-1001D automatically
recognizes the test strip code number
and adjusts the meter accordingly.
A blinking test strip will appear at the top
of the screen indication the meter is
ready for testing
CAUTION
1. If the code is not displayed after for a while, pull the
glucose test strip out of the port, and re-start the
procedure from the beginning.
2. If the code on the LCD display window as well as on the
glucose test strip vial does not match, try another new
glucose test strip. If the mismatch persists, please
contact your local representative for assistance.
3. Avoid testing under direct sunlight, for a more accurate
test result.
4. If you apply your blood sample too early, Er5 message
will appear on LCD. Please refer to Page 12.
Step 2
To obtain a suitable blood samples
 Before obtains a blood sample, wash your hands with
warm, clean water and soap. Dry hands completely
before testing.
WARNING
1. All parts of the kit are considered bio hazardous and can
potentially transmit infectious disease, even after you
have conducted cleaning and disinfection measures.
2. To reduce the chances of getting infection: The test
strips are for single use only. Do not reuse.
3. Avoid getting hands lotion, oil, dirt or debris in or on the
lancets (single use only) and the lancing device.
4. Wash your hands thoroughly with soap and clean water
after handling the meter, lancing device and/or test strip.
Page 7
UM-M (Rev.0)
Step 3
1. Unscrew the lancing
device cap.
2. Insert a sterile lancet into
the lancing device.
3. Twist the protective
cover off. Do not
discard.
IGM-1001D test meter only requires 0.5 ul sample volume.
 Place the edge of the test strip to the drop of blood.
 The blood will automatically be ‘drawn’ into the test strip
channel.
 When your sample blood is enough, your meter will
automatically count down
4. Put the lancing cap
back and twist it
tightly.
5. To adjust the depth
setting:
Use strength level 1-2 for
soft skin, 3 for average
skin, 4-5 for thick or
calloused skin.
6. Pull the end of the
lancet device back.
CAUTION
1. If the countdown does not start, do not add more blood
to the glucose test strip. Discard the glucose test strip
and re-start testing.
2. If you do not conduct the test within 3 minutes, the
glucose test meter will automatically power off to save
battery life. In this case, test procedure should start
again from the beginning.
3. You may get an inaccurate result if the blood sample is
not completely filled.
4. You must inject the blood sample until beep sounds.
Step 4
7. To prick your finger,
push
the
center
button.
CAUTION
1. If the blood smears or runs, do not use such blood
sample. Dry the area and gently squeeze another drop
or puncture a new area of the finger.
Do not share reusable lancing device with anyone
 After the beeping sound, the test will begin automatically
and your results will appear in 5 seconds. It should begin
counting down from 5 to 1 second on the LCD display
window. The LCD window will display the result of your
blood glucose level. Temperature and time.
 After results displayed, you choose to use the
activity/meal flags (
), press and
release ▲ or ▼ button to select the activity, then
press and release the power button
INFOPIA CO., LTD
Page 8
UM-M (Rev.0)
 Record the result value in your logbook. When glucose
test strip is removed, the glucose test meter turns off
automatically.
CAUTION
1. If the test result is out of the test range, “Hi/Lo” message
will be shown on the LCD window.
2. Safely discard used glucose test strips and lancets in the
proper place, according to the local regulation.
CAUTION
1. If you push the ejector button forward too much, it may be
part of the failure.
2. Do not give strong impact to blood glucose test meter.
3. Please follow the waste disposal regulations on the used
lancets and strips imposed by your local authorities.
Reviewing Your Results
The IGM-1001D meter stores up to 365 test results in its
built-in memory, along with the average sugar level for the
number of days, which you had preset the glucose test
meter to calculate.
Test Strip Ejector Function
Reviewing Your blood sugar
Disposing of your test strip
Once the test is complete,
push the strip forward to
remove the test strip from
the meter
Press and release
button. The most recent result appears first.
Note the time and day of your blood sugar result. Press and release
the ▼ button and your previous result will appear on the display
Reviewing Your averages
Used test strips may be
considered bio hazardous
waste in your area.
Be sure to contact your
local representative for
proper disposal
Press and release the ▲ button and your average result will appear
on the display.
Average Test Result
Disposing of your lancet
Test Unit
Test Result Amount
Average Test Result Days
Push the needle into the
protective cover.
Deleting Test Results
Pull the lancet out and
discard accordingly.
INFOPIA CO., LTD
Page 9
Deleting individual test result
To delete any individual test result in the
memory press ▲ or ▼ button for 3 seconds,
while the test result is to delete, displayed
blinking. After the third beep sound you
may hear, the test result is deleted.
UM-M (Rev.0)
Deleting all test results
To delete all test results in the memory press
▲ and ▼ button for 3 seconds, while the test
result is to delete, displayed
blinking. After the third beep sound you may
hear, the test results are deleted.
alternate site is less painful than using a fingertip. Getting a
blood sample from your forearm or palm is different than
getting a sample from your fingertip.
Ventral palm/Dorsal hand
Choose a fleshy area on the
palm, below your thumb or
pinky finger. Select a spot
without any visible veins and
away from any deep lines. This
may cause your blood sample
to smear.
CAUTION
The deleted test results cannot be recovered. Please be
careful in deleting the test results.
Alternate Site Test
The IGM-1001D Blood Glucose Monitoring System gives
you the ability to obtain a blood sample for testing your
sugar from different areas of your body.
The figure shown above displays the areas where you can
test using your IGM-1001D meter.
You may test you blood sugar from your forearm, upper arm,
palm (ventral palm or dorsal hand), fingertips, thigh, or calf.
IMPORTANT
It is recommended that alternate site testing be used when
sugars are stable: before meals and before bedtime.
However, when sugars are changing, blood from the
fingertip may show these changes sooner than blood from
other sites.
LIMITATION
1. Alternative site testing results should never be used to
calibrate continuous glucose monitoring systems.
2. Alternative sampling results should never be used in
insulin dosing calculations.
Forearm
Upper arm
Calf
Thigh
Choose a fleshy area of the forearm, upper arm, thigh, or
calf away from bone, visible veins and/or hair. Sometimes
there is less blood flow to these areas than to the fingertip.
To help you get a large enough drop of blood, you may
gently message or apply a heating pad to the site to
increase blood flow.
IMPORTANT
We recommend that you test on your fingertips if you are
testing for hypoglycemia (low blood glucose) or if you are
suffering from hypoglycemia unawareness
To ensure accurate results when lancing your arm (forearm
or upper arm), leg (calf or thigh), or palm (ventral palm or
dorsal hand), clean the test site with soap and water.
Make sure there is no cream or lotion on the test site.
Thoroughly dry your hands and test site.
WARNING
To reduce the chances of getting infection: never share a
lancet or reusable lancing device with anyone.
Lancing and sampling form an alternate site area
Sampling from your upper arm, forearm, dorsal hand, ventral
palm, thigh, or calf allows you to use your fingertips less
often.
You may find that obtaining a blood sample from an
INFOPIA CO., LTD
Page 10
UM-M (Rev.0)
The lancing device clear cap is used for alternate site
testing
Lance the test site with the lancing device
1. Remove the lancing cap by
twisting off.
2. Insert lancet and replace with
the clear cap.
3. To bring fresh blood to the
surface of the test site, rub the
test site vigorously for a few
seconds until you feel it
getting warm. Applying heat
may be helpful.
REMEMBER
1. Consult with your healthcare professional before using
alternate site testing.
2. Choose a different puncture site each time you test. Repeated
punctures in the same spot may cause soreness and calluses.
3. If bruising occurs at an alternate site or you have difficulty
getting a sample, consider sampling from a fingertip instead.
You may want to review the choice of sites with your
healthcare professional.
4. Do not share reusable lancing device with anyone including
other family members.
CAUTION
Do not test on you forearm or palm when
1. You think you blood glucose is rapidly falling. For example
within two hours of exercise, rapid-action insulin injection, or
an insulin pump bolus.
2. Testing with a fingertip sample may identify hypoglycemia or
an insulin reaction sooner than testing with a forearm or palm
sample.
4. Adjust clear cap to the highest
setting. Hold the clear cap
down against a fleshy area on
the alternative site. Press the
release button. Do not lift up.
5. Continue to hold the lancing
device and gradually increase
pressure for several seconds.
6. While holding the lancing
device on your test site, look
through the clear cap; until a
round drop of blood appears
7. Lift the lancing device straight
up; be careful not to smear the
blood on your testing site.
8. Place the edge of the test strip
to the drop of blood. The
blood will automatically be
drawn up.
INFOPIA CO., LTD
Page 11
UM-M (Rev.0)
Caring for your IGM-1001D System
Cleaning your meter and maintenance
Blood glucose test meter :
Your blood glucose test meter does not require special
maintenance or cleaning. Avoid getting dirt, dust, blood,
glucose control solution, or liquids on the blood glucose test
meter, the test port, or data port. Your blood glucose test
meter’s operation temperature is 10~40°C(50~104°F). It is
recommended that you store the blood glucose test meter in
the portable pouch after each use.
A cloth dampened with water and mild detergent can be
used to wipe down the outer part of the blood glucose test
meter. Your IGM-1001D blood glucose test meter is a
precision instrument. Please handle it with care.
Lancing device :
Clean the lancing device and caps with soap and tepid
water. To disinfect the lancing device, prepare a disinfectant
solution household bleach and water in 1(bleach) : 9(water)
ratio. Dampen a cloth with this solution and wipe the lancing
device thoroughly. Soak only the cap for at least 30 minutes
in the disinfectant solution. Do not soak the lancing device in
liquid. Rinse the lancing device and cap with water, and dry
thoroughly.
1. For detailed instructions concerning IGM-1001D blood
glucose test strips, refer to the blood glucose test strip
package inserted and found in the blood glucose test strip
box.
2. Please store the IGM-1001D blood glucose measuring
system in a cool and dry area, out of reach of children. Do
not freeze. For a more accurate test result, avoid testing
under direct sunlight.
3. Do not soak the blood glucose test meter or blood
glucose test strips in water or liquid.
4. Do not subject the meter or blood glucose test strips to
excessive heat.
5. Use your meter according to the instructions in the
manual.
6. If you need to purchase the lancing device, lancets or
blood glucose test strips, contact your local representative.
7. Discard the used lancets carefully, to prevent any
infection.
away from direct sunlight and heat.
 Do not freeze.
 After removing a test strip from the vial, immediately
close it tightly.
 Record the discarded date (date vial opened, plus 3
months) on the strip label.
 Do not use if it is expired(2 years) after manufacturing.
 Touch the test strip with clean and dry hands.
 Do not use if 3 months have been passed, since the
opening of the vial cap for strips.
Control Solution Storage
 Storage condition: Store in a cool, dry place at between
8~30 ℃ or 46~86 ℉ (Temperature)
 After use, close cap tightly.
 Make sure the control solution bottle is tightly closed
 Record the discarded date (date vial opened, plus 3
months) on the control solution on the bottle.
 Do not use if expiry date has passed.
 It expires 24 months after manufacturing date.
 It expires 3 months after opening the vial cap.
CAUTION
1. Store the IGM-1001D test strip vials in a cool, dry place.
Keep out of direct sunlight. Do not freeze.
2. Store test strips in their original vial only. Do not mix the
test strips in new vials or in any other container.
3. Immediately replace the vial cap and close tightly after
removing any test strips from the vial.
4. Make a note of the discard date, which is three months
from the date you first open a new vial of strips. Throw
IGM-1001D test strips and vial away after the discard date.
5. Do not use the test strips after the expiration date printed
on the package or vial, since it may cause inaccurate
results.
6. Do not test at temperatures below 10℃(50°F) or above 40℃
(104°F).
7. Do not test with humidity below 10% or above 90%.
8. Do not bend, cut, or alter the test strip.
9. Avoid getting dirt, food, and water on the test strip with
wet hands.
10. Avoid getting dirt, food, and water on the color-coding
label (backside of test strip)
Storage of your IGM-1001D System
To prevent the meter and test strips from getting dirt, dust or
other contaminants, please wash and dry your hands
thoroughly before use
Meter Storage
 Storage condition: 10~40 ℃ or 50~104 ℉
(Temperature) / 10 ~ 90% (Humidity)
 Always store and transport the meter in its original
storage case.
 Avoid dropping and strong impact.
 Avoid direct sunlight and humidity.
Strip Storage
 Storage condition: 2~30 ℃ or 36~86 ℉ (Temperature)
/ 10 ~ 90% (Humidity)
 Store your test strips in their original vial only. Do not
store the strips in other containers.
 Store test strip packages in a cool and dry place. Keep
INFOPIA CO., LTD
Page 12
UM-M (Rev.0)
Transferring your results
lithium batteries (CR2032), which are included. When
replacing the batteries, only CR2032 or equivalent lithium
battery should be used.
USE to Data Transfer CABLE(Option)
You can transfer test results from the IGM-1001D meter to a
computer.
You will need to download to software from the following
web site (http://www.infopia21.com) and purchase the
computer cable from your local representative. This is sold
separately.
Follow-up Action
1. Check your meter with the glucose control solution
2. Perform your test again.
3. If you are experiencing symptoms that are not consistent
with your blood glucose test results, call you healthcare
professional.
NOTE:
1. Make sure your date and time are correct after changing
your battery.
2. Remove both batteries from the battery compartment
and dispose of them, according to your institution's
guidelines..
CAUTION
1. Risk of explosion if battery is replaced by an incorrect
type.
2. Dispose of used batteries according to the instructions.
USE to Bluetooth Unit
1. Press the ‘Up’ button of the IGM-1001 meter, running
pairing mode. LCD Screen displayed ‘P’ and start count
from 0.
2. run to application on mobile phone.
3. Operating search mode on mobile phone.
4. select the operating meter in search list.
5. if pairing is succeeded, meter LCD screen displayed ‘P’ ->
‘C’ and re-start count from 0.
6. ‘Data Update’ command from the mobile phone to meter.
7. Glucose data sent from meter to mobile phone. LCD
screen displayed re-start count to 0.
8. connection termination : Press ‘Power’ button or
application exit on mobile phone.
8. if connection is fail, LCD screen displayed ‘C’ -> ‘P’. After
some seconds meter is power off automatically.
9. if pairing is incompleted, re-start to list 1.
Download application
Dispose of meter :
The meter must be disposed of according to the local
regulations concerning the disposal of electrical and
electronic equipment. The Waste Electrical and Electronic
Equipment(WEEE) regulation implement provisions of the
European Parliament and Council Directive 2012/19/EU
aimed to reducing the amount of EEE waste going for final
disposal.
The manufacturer,has specific instructions for the recovery
of the meter.
Please contact your distributor.
Dispose of batteries :
Please recycle or dispose of used batteries using your local
battery collection systems and in compliance with your local
environmental laws and regulations. Batteries contain
chemicals that, if released, may affect the environment and
human health. The crossed-out wheeled-bin symbol
indicates the need for the
separate collection for batteries.
Use apple app store or google play.
Battery Installation
The low battery icon
will appear in the upper left corner
of the LCD display to alert you when the battery power is
running low, indication a new battery are needed.
Your IGM-1001D blood glucose test meter uses only two 3V
INFOPIA CO., LTD
Page 13
UM-M (Rev.0)
Troubleshooting
Message
What it means
The following chart may help you identify certain problems,
but may not solve all problems that can occur. Contact your
authorized representative or Infopia Co.,Ltd. customer
support if the problem persists.
Message
What it means
Action required
Problem with the
blood glucose test
meter
Place the battery
again, and set the
blood glucose test
meter. If the problem
persists, Please
contact your
authorized
representative or
customer support.
Caused by either used
or wet blood glucose
test strip
Please, insert a new
glucose test strip and
perform your test
again
Problem with the test
strip
Please insert a new
glucose test strip and
apply the blood
sample until
confirming the
window is completely
filled.
Blood glucose test
strip is damaged
User applied the
blood sample before
the LCD display
was blinking
Color bar of the strip
is dirty or not good or
there is too much
light. Action required.
The ambient
temperature is less
than 10℃ (50℉)
The ambient
temperature is over
40℃ (104℉)
The test result is
lower than 10mg/dl
INFOPIA CO., LTD
The test result is
higher than 700mg/dl
Action required
performing
a glucose control
solution
test. If the test results
in a
normal reading, retest your
blood sample two or
three times.
If "Lo" persists,
consult your doctor
immediately.
Check the accuracy of
your
blood glucose test
strip by performing
a glucose control
solution
test. If the test results
in a
normal reading, retest your
blood sample two or
three times.
If "Hi" persists,
consult your doctor
immediately.
There are no reading
stored in the meter
Not enough reading
in the memory to
display designated
average
Please insert a new
glucose test strip and
perform your test
again
Wait for the meter to
display the blinking
icon before applying
your blood sample.
Please insert a new
glucose test strip and
perform your test
again. If blink with
“Sun”, avoid to the
direct sunlight and
retest. Contact your
local representative or
customer support if
the problem persists.
Place the meter at a
temperature between
10 ~ 40℃ (50 ~ 104
℉) for more than 10
minutes or more and
test again.
The glucose test
meter does not power
its ON
Battery is dead or
there is a problem
with the meter
Change the battery
and if the problem
persists, contact your
local representative.
The glucose test
meter does not start
after applied blood
sample.
Insufficient amount of
blood.
Please insert a new
test strip and perform
your test again.
The results are
inconsistent
There may be a
problem with the
glucose test strip
Please insert a new
test strip and perform
your test again.
Place the meter at a
temperature between
10 ~ 40℃ (50 ~ 104
℉) for more than 10
minutes or more and
test again.
Check the accuracy of
your blood glucose
test strip by
Page 14
UM-M (Rev.0)
Inconsistent or Unexpected Test Results
Declaration of Conformity
If you continue to get unexpected results, check your
system with control solution. If you experience symptoms
that are not consistent with your glucose results, review and
follow all instruction in this manual. Never ignore symptoms
or make significant changes to you diabetes control program.
Inform health care professional of symptoms and/or
concerns.
FCC
CAUTION
1. Low Glucose Results: if your result is lower than
70mg/dL, it may mean hypoglycemia (low blood sugar).
This may require immediate treatment according to your
healthcare profession’s recommendations. Although this
result could be due to a test error, it is safer to treat first
and you may test another.
2. High Glucose Results: if your test result is higher than
180 mg/dL, it may mean hyperglycemia (high blood
sugar). If you are uncertain about your test results,
consider with re-test. Your health care profession will
help you to decide how it action. If the meter displays HI
all the time, please re-check your blood sugar but still it’s
higher, please find your healthcare profession
immediately.
Warranty
Manufacturer’s Warranty
Infopia Co.,Ltd. warrants to the original purchaser that this
instrument will be free from defects in workmanship for 3 years
from the date of original purchase.
Limitations of Warranty
This warranty is subject to the following exceptions and
limitations.
1. Infopia Co.,Ltd. shall not be required to replace any units
which are damaged or malfunction due to abuse, accidents,
alteration, neglect, misuse, maintenance by someone other
than Infopia Co.,Ltd. or failure to operate in accordance with
the instructions.
2. Infopia Co.,Ltd. reserves the right to make changes in design
without obligation to incorporate such changes into previously
manufactured instruments.
3. Infopia Co.,Ltd. has no knowledge of the performance of the
instrument when the test strip is altered or modified in any
manner.
THIS DEVICE COMPLIES WITH PART 15 OF THE FCC
RULES.
Operation is subject to the following two conditions; (1) this
device may not cause harmful interference, and (2) this
device must accept any interference received, including
interference that may cause undesired operation.
CAUTION
Changes or modifications not expressly approved by the
party responsible for compliance could void the user's
authority to operate the equipment.
NOTE:
This equipment has been tested and found to comply with
the limits for a Class B digital device, pursuant to Part 15 of
the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no
guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful
interference to radio or television reception, which can be
determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more
of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and
receiver.
• Connect the equipment into an outlet on a circuit different
from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician
for help.
For Warranty Service
Purchaser must contact the customer service department of Infopia
Co.,Ltd. by calling toll 82-31-460-0300, for assistance and/or
instructions for obtaining service of this instrument.
Service Information
Infopia Co.,Ltd. has trained specialists to help you 24 hours a day,
7 days a week, and 365 days a year.
IMPORTANT
Confirming to Infopia Co.,Ltd. representative before returning
your meter for any reason. You will be given the information
needed to get your problem handled correctly and efficiently.
Remaining your meter, test strip and control solution nearby when
you call.
INFOPIA CO., LTD
Page 15
UM-M (Rev.0)
Specifications
Separate collection of batteries
Sample Type
Sample Volume
Test Range
Reading Time
Hematocrit
Altitude
Operating Temperature
Operating Humidity
Strip Storage Temperature
Display Type
Dimension(HxDxT)
Weight
Power Source
Batter Life
Capillary, venous, arterial whole
blood
0.5 ㎕
10 ~ 600 mg/dL
5 seconds
20 ~ 60%
3048 meter Up to (10,000 feet)
10 ~ 40℃ (50 ~ 104℉)
10 ~ 90%
Store 2~30 ℃ or 36~86 ℉ and
no direct sunlight.
Do not Freeze
LCD (Including backlight model)
96 X 56 X 23.5 (mm)
66±1g (Including batteries)
3V (Coin Battery, CR2032 X 2EA)
IGM-1001D: Running 1,000 Tests
Infopia Co., Ltd.
132, Anyangcheondong-ro, Dongan-gu, Anyang-si,
Gyeonggi-do, Korea
www.infopia21.com
Obelis S.A
Bd. Genaral Wahis 53, 1030 Brussels, BELGIUM
Tel : +(32)2.732.59.54
Fax : +(32)2.732.60.03
E-Mail : mail@obelis.net
Rev. 2012-03
Symbol Reference
Symbol
MEMO
Description
Consult Instructions for use
Use By date
This product fulfills the
requirements of Directive
98/79/EC on in-vitro diagnostic
medical devices
Attention, See Instructions for
use
For in-vitro diagnostic use
Batch code
Do not reuse
Serial number
Date of Manufacture
Manufacturer
Keep away from sunlight
Waste electrical and electronic
equipment
symbol
Temperature limitation
Catalogue number
Self-testing used
Authorized representative in the
European Community
Biohazard
INFOPIA CO., LTD
Page 16
UM-M (Rev.0)

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : No
Page Count                      : 16
Language                        : ko-KR
Tagged PDF                      : Yes
Title                           : 1
Author                          : 남증희
Creator                         : Microsoft® Word 2010
Create Date                     : 2014:09:29 17:16:50+09:00
Modify Date                     : 2014:09:29 17:16:50+09:00
Producer                        : Microsoft® Word 2010
EXIF Metadata provided by EXIF.tools
FCC ID Filing: WSX-IGM-1001D

Navigation menu