Omitec OM4720 OmiScan Gas Handset User Manual

Omitec Limited OmiScan Gas Handset

User Manual

Part Number - I317719US - Issue 1© Omitec 2008OmiScan GasOperating InstructionsPart Number - I317719USIssue 1© Omitec  2008
ContentsiiiIntroductionIntroduction.............................................................................................  5Power Management ...............................................................................  7Safety Precautions .................................................................................  7Electrical Safety Precautions..................................................................  7Assembling the OmiScan Gas Analyzer.................................................  8Safe Removal and Disposal of Batteries ................................................  9OmiScan Gas Sample Probe ...............................................................  10Getting StartedEmissions Testing ................................................................................  11Handset Navigation ..............................................................................  11Gas Analyzer ........................................................................................  12Simulated Tests....................................................................................  23System Checks.....................................................................................  35PC/USB Mode ......................................................................................  37Settings.................................................................................................  37Low Battery Indication ..........................................................................  43Out of Range ........................................................................................  44PC Gas Analyzer SoftwareSystem requirements............................................................................  46Installation ............................................................................................  46Using the Software ...............................................................................  46Analyzer Software Functions ................................................................  47Communications...................................................................................  68Software Build ......................................................................................  68Gas AnalysisOmiScan Gas Analysis of Results........................................................  69
ContentsivRoutine Maintenance & CleaningRoutine Maintenance............................................................................  74NOx Cell Replacement .........................................................................  76Calibration and Accuracy Checks.........................................................  77Cleaning ...............................................................................................  77Technical SpecificationTechnical Specification.........................................................................  78General Data ........................................................................................  80Ancillary Equipment..............................................................................  81Pneumatic Specifications .....................................................................  81Service and calibration frequency ........................................................  82Country Specific Approvals ..................................................................  82Spare Parts List ....................................................................................  84
Introduction5IntroductionIntroductionOmiScan Gas is an exhaust gas analyzer for spark ignition vehicles and comprises a fully portable handset with a base station and optional software to install on a PC. 1. Handset2. Base unit3. Exhaust Probe attachment4. Base unit exhaust pipe extension5. Mains power converter supply6. CD ROM with PC software 7. Vehicle power supply cable (optional)8. USB cable handset to laptop9. FiltersSpark ignition vehicles includes both catalyst and non-catalyst, supercharged (including Turbo) and lean-burn engines, vehicles that are fuelled by Petrol, Liquid Petroleum Gas (LPG) and Compressed Natural Gas (CNG) and also injection or carburettor fuel control.NOTE: Diesel engined vehicles are specifically excluded from testing using the OmiScan Gas system.78123OM13754569
Introduction6There are two options available:1. OM4720/5G•Measures 5 gases, provides live readings of all gases and λ•PC software to graph and/or diagnose emissions faults from gas readings2. OM4720/4G•Measures 4 gases, provides live readings of all gases and λ•PC software to graph gas readingsThe OmiScan Gas system enables the measurement of four or five of the main exhaust gases emitted from a spark ignition engine. The gases measured are: Carbon Dioxide (CO2), Carbon Monoxide (CO), Hydrocarbons (HC), Oxygen (O2) and Nitric Oxide (NOx) (optional). Measurement accuracy meets or exceeds requirements for OIML Class 0 and BAR97:•CO:Range - 0% to 15%; accuracy typically 3% relative•CO2: Range - 0% to 20%; accuracy typically 3% relative•O2: Range 0% to 25%; accuracy typically 1% relative•HC n-hexane: 0 to 2000ppm; accuracy typically 3%•HC n-propane: 0 to 30000ppm; accuracy typically 5%•NOx :Range - 0 to 5000ppm, accuracy typically 3% relative (optional)The optional PC based system provides enhanced facilities for data capture and graphing as an additional feature.The OmiScan Gas handset communicates with the base station using Bluetooth® Wireless Technology and with the PC option, the handset communicates with a PC via a USB port.
Introduction7Power ManagementThe handset monitors both itself and the base station for the charge status of the battery packs. If either battery pack voltage falls below a predetermined level the handset will display a warning to cease testing and recharge the units. Approximately five minutes before the batteries reach the automatic shutdown point, a message is flagged to warn you of the low battery state and thus the imminent shutdown.1. Low Base Station Battery•Indicated by the handset battery symbol flashing ‘empty’ and the green LED on the base station flashing2. Low Handset Battery•Indicated only by the battery symbol on handset display flashing empty.NOTE: Once the units have been put on charge, both LED and battery symbol will stop flashing.Safety PrecautionsWARNING: Exhaust gases are poisonous and contain carbon monoxide, which can cause unconsciousness and can be fatal - NEVER start or leave the test vehicle engine running in an unventilated area.WARNING: DO NOT handle the OmiScan Gas exhaust probe tip, or place the exhaust probe on plastic or combustible surfaces after the test procedure. The probe may become extremely hot during the test and could cause burns and a risk of fire.Electrical Safety PrecautionsWARNING: Any power or extension cord must have a ground conductor.WARNING: When replacing fuses make sure the correct type and rating are used.WARNING: Any adjustment, maintenance or repair must be carried out by an approved service engineer.
Introduction8Assembling the OmiScan Gas AnalyzerUnpack the units carefully, checking off each item and component against the itemized packaging list included. If any component is missing or damaged, contact your supplier immediately.•The batteries in the base unit and the handset will need to be installed and charged before either unit can be used (see ‘Power Management’, page 7).•The base station battery compartment is on the underside of the unit, secured by two cross-headed bolts. Unscrew the cover, and secure the bigger battery pack into the cover tray with the velcro strap; connect the in-line connector to the base station. Refit the cover and tighten the two screws (do not over tighten).•Remove the handset rubber protection cover to access the underside of the case cover; remove the four cross-headed screws and place to one side. Lift off the bottom cover and place to one side. Connect the smaller of the battery packs to the in-line connector and place the battery pack in the recess through the PCB; taking care to keep the supply cables from catching the cover. Position the cover and replace the four screws (do not over tighten). Refit rubber protection cover.•Place the handset in the base station ensuring that it is pushed firmly into place and clips into the bottom securing clip.IMPORTANT: BATTERY PACKS SHOULD ALWAYS BE FITTED BEFORE THE CHARGER UNIT IS CONNECTED.•Connect the mains charging unit into the round connector at the rear of the base station and the mains plug into a wall socket. With the mains switched on, the middle red LED on the top surface of the base station should illuminate to indicate that the base station is charging. The handset display will switch ON automatically and the battery symbol in the top right hand corner will indicate progressive charging.•If applicable, while the batteries are charging, install the Analyzer software onto your PC (see ‘Installation’, page 46).•When fully charged the red base station LED will extinguish and the lower amber LED will illuminate to indicate the base station battery is on trickle charge until the charging unit is disconnected. Likewise, the battery symbol on the handset will remain static showing a fully charged battery, which will be on trickle charge until mains charging unit is disconnected.
Introduction9CAUTION: Before connecting the base station to the mains, ensure that you have read and fully understand the operating instructions given in this document.•Attach the exhaust sample probe to the water trap fitting (2) at the rear of the base station.•Attach the exhaust outlet pipe to the Exhaust Gas outlet (4). The exhaust outlet pipe is used to expel the sampled exhaust gases from the test vehicle well away from the active working area, to prevent technicians inadvertently inhaling the gases. Run the outlet pipe to a well ventilated area.Safe Removal and Disposal of BatteriesBoth the Base station and the handset use rechargeable Nickel Metal Hydride battery packs. The expected working life for each pack should be in excess of 18 months. Both packs are consumable spare parts and are designed to be replaced by the user. See ‘Assembling the OmiScan Gas Analyzer’, page 8 to see how to access the batteries.CAUTION: Ensure both Handset and Base station are powered OFF and disconnected from charger Power supply BEFORE replacing battery packs!  The Handset also contains a Lithium button cell for time and date back up. This battery’s life expectancy is in excess of 5 years. This battery should not be replaced by the user.  This should be carried out by an Omitec Inc. authorized Service Agent.CAUTION: DO NOT DISPOSE OF BATTERIES IN HOUSEHOLD REFUGE. SOME COMMUNITIES OFFER   RECYCLING OR THE COLLECTION OF USED BATTERIES. CONTACT YOUR LOCAL GOVERNMENT FOR DISPOSAL PRACTICES IN YOUR AREA.Ni-MH LI-ION
Introduction10OmiScan Gas Sample ProbeWARNING: Before attaching any OmiScan Gas probe to the test vehicle, ensure that you have read and understand the safety precautions listed earlier in this manual.Exhaust probeThe exhaust probe collects gas samples from the test vehicle exhaust, which are then analyzed by OmiScan Gas.The probe connects between the Sample Gas Input fitting (item 2 under ‘Assembling the OmiScan Gas Analyzer’, page 8) and the test vehicle exhaust pipe. Insert the probe as far into the vehicle exhaust as possible and secure by attaching the alligator clip to the end of the exhaust pipe.WARNING: Take care to avoid contact with the exhaust system when inserting or removing the exhaust probe. The exhaust system will be hot and severe burns could occur.The exhaust probe is supplied with a sealing cap (Part No. OM 4700/23) which must be fitted over the end of the probe when a leak test is performed. Keep the cap in a safe place when not in use (see ‘System Checks’, page 35, for further details of a leak test).
Getting Started11Getting StartedEmissions TestingBefore attempting to test vehicle emissions levels, ensure that the base station and handset are fully charged (see ‘Power Management’, page 7) for remote operation. For on-road testing if the base station batteries are low, connect the base station to the vehicle supply cable (optional). With this configuration the batteries will not charge but the supply to the base station is maintained.OmiScan Gas emissions testing can be undertaken by using the handset separated from, or in position on, the base station. Handset NavigationPress the ‘Power up’ buttons   on the base station and then on the handset, the handset screen displays the ‘Main Menu’ and lists the functions available:•Gas Analyzer•PC/USB Mode•Settings•LanguageThe handset will automatically start searching for the Bluetooth® Wireless Technology address specific to the base station. The handset and the base station are ‘paired’ together to prevent any interference or acquisition from other Bluetooth® Wireless Technology devices such as laptop PCs or mobile telephones. After a few seconds an antenna icon will appear in the top LH corner of the screen indicating that the Bluetooth® Wireless Technology link has been established. This icon will always be displayed whilst the link is made.The battery icon in the top RH corner reflects the charge status of the handset battery pack.NOTE: The battery icon will flash empty to indicate a low battery charge for either the handset or the base station - see ‘Power Management’, page 7.The Main Menu has four options:•Gas Analyzer - Emission testing and analyzer system checks•PC/USB Mode - Used for down loading test result and updating handsetMAIN MENUGas Analyzer  PC/USB Mode  Settings  Lang, Idioma                     
Getting Started12•Settings - Results record, clock and general system settings•Language - Allows the user to configure handset in English, French or SpanishGas AnalyzerTo access Emission Menu press the   button with the   pointing at Gas Analyzer.Follow the instructions and press the   button.The handset will display ‘Analyzer SetUp - Please Wait’ and a progress bar. On completion the progress bar will be replaced briefly by ‘Passed’. During this time the analyzer is initializing settings ready to conduct emissions testing. Once completed the display screen will change back to the ‘Main Menu’.The handset has an automatic power saving facility, if no action is taken after one minute the display backlight powers down. Pressing any button will turn the backlight back on. If the handset does not detect a button press for more than 15 minutes it will assume that the OmiScan Gas is not in use and the Base Station and the Handset will automatically power down.NOTE: The automatic shutdown feature is disabled during ‘Live Reading’ test.MAIN MENUGas Analyzer  PC/USB Mode  Settings  Lang, Idioma                     REMOVE THE PROBE         Ensure the exhaust           probe is in clean                 air-OK                                      ANALYZER SETUP         Please wait. . . . .
Getting Started13Manual Power DownTo power down both Handset and Base station at the same time, simply press and hold power button on handset for >1 second.Live ReadingsPress the   again to enter the ‘Gas Analyzer’ menu and again to start ‘Live Readings’.The screen momentarily displays ‘Live Readings’, then ‘Please wait’ then asks the technician to ‘Repeat’  or ‘Restart’   a test, displaying the vehicle data from the last test: Manufacturer, Licence plate, year and fuel type.NOTE: When using for the first time, there will not be any previous tests stored, therefore the handset will proceed directly to the series of screen reminders before starting a test.For ‘Repeat’ see page 18, ‘Restart’ sequence follows.Other fuels offered are: •Ethanol E85•Methanol M85GAS ANALYZERLive Readings  Simulated Tests  System Checks                                       SAME VEHICLE?            FORD          OU54XYZ***         2006   Gasoline-Repeat X-Restart                                       SELECT FUEL TYPEGasoline  USA Reformulated  LPG  CNG                                       
Getting Started14•Gasohol E10•Gasohol E20Pressing Restart   will display a screen asking for fuel type.The remaining memory will be shown. On pressing  -OK the series of screens and procedures will be displayed.A series of screen reminders are displayed: ‘Ensure oil and coolant levels OK’, press  to continue, or press   to abort’.‘Ensure brake on and in neutral/park, press   to continue, or press   to abort’.LIVE READINGS        Remaining Memory:          56Mins 40secs          -OK X-Cancel                                       LIVE READINGS        Ensure oil and       coolant levels OK          -Continue X-Abort                                       LIVE READINGS       Ensure brake on and          in neutral/park          -Continue X-Abort                                       LIVE READINGS       Ensure engine is       running and up to               temp.          -Continue X-Abort                                       
Getting Started15‘Ensure engine is running and up to temp., press   to continue, or press   to abort’.‘HC HANGUP Ensure the exhaust probe is in clean air, press   to continue.HC HANGUP’ ‘HC residue check’, ‘Please wait’, Lim: 20ppm HC: XX’, ‘XXsecs’, press  - to abort. On completion of the countdown the screen will briefly display ‘HC residue check passed.’The base station will automatically carry out a ZERO CALIBRATION, on completion the progress bar will be replaced briefly by ‘Passed’.On completion there is a prompt to ‘INSERT PROBE’.HC HANGUP       Ensure the exhaust         probe is in clean                 air          -OK                                       HC HANGUP       HC residue check        Please wait . . . .      Limit: 20ppm HC: 00           9 secs     X-Abort                                       XZERO CALIBRATION               Please wait . . .   INSERT PROBE       Exhaust probe check.        Please wait . . . .             -Continue X-Abort                                       X
Getting Started16As soon as gas is sensed there is a request for the number of gases to display, all, two or one and also an ‘Exit Test’.If ‘1 Gas Display’ or ‘2 Gas Display’ is selected, an additional menu will be displayed listing the gases or outputs that can be displayed, which are CO, HC, O2, CO2, Lambda, NOx, Air Fuel Ratio (AFR) and corrected CO (COK).To make a selection, scroll up or down and press  . In ‘2 Gas Display’ option, the selection will be tagged with an arrow. To deselect press   again. Once the second gas has been selected, the handset will exit the menu and display the selected gases.On selection the screen will display the gas and the gas level in % or ppm, as appropriate. There is no fixed time for this test, however, there is a prompt to ‘Start’ and ‘Stop’ using the   button. When sampling, a ‘Mark’ can be made at a specific point in the test by pressing the   button. This will set a marker in the data that is displayed in the analyzer data graph on the PC.LIVE READINGSDisplay All.  2 Gas Display  1 Gas Display  Exit Test                                                 XLIVE READINGSHC:     0               λ: 0.00CO: 0.001%         CO2: 0.00%O2: 21.02%         NOx: 0ppm-Start X-Go back                XLIVE READINGSHC:     0               λ: 0.00CO: 0.001%         CO2: 0.00%O2: 21.02%         NOx: 0ppm-Stop     -Mark  X-Go back   X
Getting Started17When stop is pressed the screen displays ‘Test Result XX’ ‘Live Readings’ ‘Time: XX:XX’ ‘Date: XX/XX/XX’ press   to save, or press   to erase.If saving, the next screen displays ‘SELECT MAKE’ and makes of vehicles can be scrolled using the   and   buttons. Press   to select the make. The next screen displays ‘ENTER VEHICLE YEAR’, 200X (model year) and the   and   buttons will scroll through the years, press   to select the year, the bottom line, i.e.  -OK  -Cancel  -Move will flash until   is pressed to confirm. TEST RESULTS 1           Live Readings              Time: 09:11          Date: Feb 20 08-Save  X-Erase                  XSELECT MAKENot recorded  ACURA  ALFA ROMERO  ASTON MARTIN                                         LIVE READINGS        Enter Vehicle Year                           200X          -OK X-Cancel  - Move                                       
Getting Started18The next screen displays ‘LIVE READINGS’ ‘Vehicle License No:’ ‘XXXXXXXXXX’. The   and   buttons will set each character from A to Z and 0 to 9, when the first character is correct press the camera   button. Move to the next character and repeat the procedure until the registration number has been added press  . The bottom line, i.e.  -OK  -Cancel  -Move will flash until   is pressed again to confirm.Next screen displays ‘SELECT TYPE’ and a list of vehicle type, eg. Hatchback, Sedan, Station wagon, Convertible, Coupe and 4x4.The next screen displays ‘TESTER NAME’ and a list of stored testers, a default list shows ‘Tester 1’, ‘Tester 2’ etc.On selection of the tester the remaining memory is displayed.LIVE READINGS        Vehicle License No:                       **********             ----------          -OK X-Cancel  - Move                                   SELECT TYPENot recorded  Sedan  Coupe  Convertible                                         TESTER NAMENot recorded  TESTER 1  TESTER 2  TESTER 3  TESTER 4                                       LIVE READINGS        Remaining Memory:          61Mins 43secs          -OK X-Cancel                                       
Getting Started19Pressing the  -OK will start a ‘System Purge’ procedure to remove any residual exhaust gases in the probe and pipework.On completion of the ‘System Purge’ the display will return to the Gas Analyzer menu.If pressing ‘Repeat’ the following sequence will occur:Pressing   will display ‘Remaining memory: XXMins XXsecs’  -OK,  -Cancel. If there is insufficient time available to complete another test the handset will need to have the existing files deleted (see ‘Settings’, page 37). For ‘Restart’ see page 13.NOTE: If you have the PC option, make sure that any tests you wish to keep/analyze are down loaded to the PC before deleting the files.REMOVE THE PROBE          Please Wait. . .                     -OK X-Cancel                                       SYSTEM PURGE            Please wait . . . X-Cancel                                  XLIVE READINGS        Remaining Memory:          61Mins 43secs          -OK X-Cancel                                       
Getting Started20A series of screen reminders are displayed: ‘Ensure oil and coolant levels OK’, press  to continue, or press   to abort’.‘Ensure brake on and in neutral/park, press   to continue, or press   to abort’.‘Ensure engine is running and up to temp., press   to continue, or press   to abort’.‘HC HANGUP Ensure the exhaust probe is in clean air, press   to continue.LIVE READINGS        Ensure oil and       coolant levels OK          -Continue X-Abort                                       LIVE READINGS       Ensure brake on and          in neutral/park          -Continue X-Abort                                       LIVE READINGS       Ensure engine is       running and up to               temp.          -Continue X-Abort                                       HC HANGUP       Ensure the exhaust         probe is in clean                 air          -OK                                       
Getting Started21HC HANGUP’ ‘HC residue check’, ‘Please wait’, Limit: 20ppm HC: XX’, ‘XXsecs’, press   - to abort. On completion the screen will briefly display ‘HC residue check passed.’The base station will automatically carry out a ZERO CALIBRATION, on completion the progress bar will be replaced briefly by ‘Passed’.On completion there is a prompt to ‘INSERT PROBE’.As soon as a gas is sensed there is a request for the number of gases to display, all, two or one and also an ‘Exit Test’.HC HANGUP       HC residue check        Please wait . . . .      Limit: 20ppm HC: 00           9 secs     X-Abort                                       XZERO CALIBRATION               Please wait . . .   INSERT PROBE       Exhaust probe check.        Please wait . . . .             -Continue X-Abort                                       XLIVE READINGSDisplay All.  2 Gas Display  1 Gas Display  Exit Test                                                 X
Getting Started22If ‘1 Gas Display’ or ‘2 Gas Display’ is selected, an additional menu will be displayed listing the gases and outputs that can be displayed, which are CO, HC, O2, CO2, Lambda, NOx, Air Fuel Ratio (AFR) and corrected CO (COK).To make a selection, scroll up or down and press  . In ‘2 Gas Display’ option, the selection will be tagged with an arrow. To deselect press   again. Once the second gas has been selected, the handset will exit the menu and display the selected gases.On selection the screen will display the gas and the gas level in % or ppm, as appropriate. There is no fixed time for this test, however, there is a prompt to ‘Start’ and ‘Stop’ using the   button. When sampling, a ‘Mark’ can be made at a specific point in the test by pressing the   button. This will set a marker in the data that is displayed in the analyzer data graph on the PC.When stop is pressed the screen displays ‘Test Result XX’ ‘Live Readings’ ‘Time: XX:XX’ ‘Date: XX/XX/XX’ press   to save, or press   to erase.LIVE READINGSHC:     0               λ: 0.00CO: 0.001%         CO2: 0.00%O2: 21.02%         NOx: 0ppm-Start X-Go back                XLIVE READINGSHC:     0               λ: 0.00CO: 0.001%         CO2: 0.00%O2: 21.02%         NOx: 0ppm-Stop     -Mark  X-Go back   XTEST RESULTS 1           Live Readings              Time: 09:11          Date: Feb 20 08-Save  X-Erase                  X
Getting Started23Pressing   -Save will store the results and the screen will change to live readings available memory. Pressing  -OK will display the Gas Analyzer menu, after the ‘remove probe’ and system purge’ screens.Simulated TestsSimulated tests has three test procedures:•CAT•TSI•ASMNote: The 4 Gas version of OmiScan Gas only offers TSI as NOx is not fitted. CAT and ASM require NOx readings.LIVE READINGS        Remaining Memory:          56Mins 40secs          -OK X-Cancel                                       GAS ANALYZER  Live ReadingsSimulated Tests  System Checks                                      SIMULATED TESTSCAT  TSI  ASM                                        
Getting Started24CATIn this test all five gases HC, CO, CO2, NOx and O2 are analyzed to gauge the performance of a Catalytic convertor. The test will report that the Catalytic convertor is good, suspect or if the gas readings suggest there may be another  fault which may not be related to the Catalytic convertor (only available for 5-gas option).Selecting petrol will start a sequence of screens, remaining memory, oil and coolant levels, hand brake and engine running and up to temperature. HC Hangup sequence will be started ending with a Zero Calibration and a prompt to insert probe into the exhaust pipe. The process will continue with fast idle.Open the throttle to increase engine speed to within the indicated RPM range.CAT           Please Wait. . . .                                                 SELECT FUEL TYPE   Gasoline     USA Reformulated     LPG     CNG                                                    CAT       Run engine between          2500 - 3000 RPM                 -OK X-Abort                                       
Getting Started25A short preconditioning before continuing the fast idle test.If the gas level is within the limits the test will be complete and a report of ‘Within Limits’ will be displayed.If the test fails an extended test can be started.The additioning conditioning comprises an extended period at fast idle to ensure that the catalyzer is at the correct temperature to start and maintain the catalytic reaction.CAT        Preconditioning              5 secs                       X-Abort                                       XCAT        Fast Idle Test              30 secs                       X-Abort                                       XCAT        Outside Limits           X = Quit            = Extended test                                                       or XCAT TEST             Additional           conditioning           Fast Idle test?                       -Continue X-Abort                                or X
Getting Started26The vehicle engine speed must be set to within the range indicated.If the cat fails again a results page is displayed.The test results can be saved or erased as required.CAT        Run engine betweeen            2500-3000 RPM                 -OK X-Abort                                       CAT           Fast Idle Test              180 secs                       X-Abort                                       XNON-CAT FAULT    CO = -0.00%,  HC = 0ppm    O2 = 20.97%,  CO2 = 0.00%           NOx = 0ppm-OK                                       
Getting Started27TSI TestThis test simulates the ‘first chance’ IDLE and HIGH SPEED modes outlined in the EPA91 Two Speed Idle emissions test. The test is designed to allow the user to evaluate how a vehicle may perform when tested officially.Note: This is not an official test.A warning screen displays the fact that this test routine is not an official test and can be used only as a guide to see if gas levels pass/fail the preset gas limits.The option to carry out the test on the last vehicle tested or a new vehicle will be offered. If a new vehicle is selected, you will be prompted for fuel type at the start of the test, so that the correct parameters can be set. At the end of the test, before saving the results you will be prompted for vehicle make, year etc. If the test is being carried out on the last vehicle tested then fuel type, vehicle make and year etc. will have already been saved by the handset and will be automatically added to the test result if it is chosen to be saved.WARNING! This is NOT an officialTSI test           -OK, X-Cancel                                  X or SAME VEHICLE?CITROEN OU54XYZ*** 2006   Gasoline-Repeat, X-Restart                                        
Getting Started28The test limits to be applied can be selected from the next menu screen. Simply scroll down and press to select. Note TSI (Custom 1) and TSI (Custom 2) limits can be edited to suit testing requirements using PC Gas Analyzer Software – see ‘Tests’, page 61‘ detailed in the PC Analyzer Software section of this manual.Test limits are displayed. Press   to proceed or X to go back to Limit Menu screen to select other limits.Remaining memory screen will be displayed. Note the TSI can require up to approximately 3 minutes to complete.The handset will then detail the same preconditioning, HC hang, Zero Calibration and Insert Probe Screens as in other test routines before displaying:TEST LIMITSTSI (Default)TSI (Custom 1)TSI (Custom 2)TEST LIMITSHC = 200ppmCO = 1.00%           -OK, X-Cancel                                  TSIRemaining Memory:59 mins 44 secs           -OK, X-Cancel                                   X or 
Getting Started29Press   to Start test.The test readings will be displayed along with a test countdown. Maximum duration of test mode is 90 seconds.At the end of the IDLE Mode test the result will be displayed. Press   to proceed to the HIGH Speed Mode part of the test.Screen prompts for the vehicle idle speed to be raised and maintained between an upper and lower limit for the duration of the HIGH Speed Mode of the test. Press   to proceed.TSI TESTALLOW VEHICLE SPEED TO IDLE BETWEEN 350 - 1100 RPM           -OK,    X-ABORT                                   IDLE MODEHC (200) = 54ppmCO (1.00) = 0.05%           77 Secs      X-ABORT                                   WITHIN LIMITSHC (200) = 54ppmCO (1.00) = 0.05%           -OK,    X-ABORT                                  TSI TESTRUN ENGINE BETWEEN2200-2800 RPM           -OK,    X-ABORT
Getting Started30The test readings will be displayed along with a test countdown. Maximum duration of test mode is 90 seconds.At the end of the HIGH SPEED Mode test the result will be displayed. Press   to proceed.The overall result of the TSI test will be displayed. Press the ‘Camera’ button to see a summary of both modes of the test. Press   to proceed.HIGH SPEED MODEHC (200) = 54ppmCO (1.00) = 0.05%           77 Secs    X-ABORT                                  WITHIN LIMITSHC (200) = 54ppmCO (1.00) = 0.05%           -OK,    X-ABORT                                  TSI TEST RESULTWITHIN LIMITS           -OK,    X-ABORT
Getting Started31Press   to save or X to discard. If the test was carried out on a new vehicle the screen will prompt for make, year etc. before returning to the Simulated Test Menu. If the test was carried out on the last vehicle tested then the screen will revert straight back to the Simulated Test Menu.ASM TestThis test simulates measurement and analysis carried out during a ‘Acceleration Simulation Mode’ Test as outlined in specification EPA420-B-04-011. The gases measured are CO, HC and NOx. The test is designed to allow the user to evaluate how a vehicle may perform in an official test.Note: This is not an official test.A warning screen displays the fact that this test routine is not an official test and can be used only as a guide to see if gas levels pass/fail the preset gas limits.The option to carry out the test on the last vehicle tested or a new vehicle will be offered. If a new vehicle is selected, you will be prompted for fuel type at the start of TEST RESULTDefault TSITIME:   11:45DATE: Oct 03 08           -OK,    X-Erase                                   WARNING!This is NOTAn official ASM test           -OK,    X-Cancel                                   SAME VEHICLE?CITROEN OU54XYZ*** 2006   Gasoline-Repeat, X-Restart                                        
Getting Started32the test, so that the correct parameters can be set. At the end of the test, before saving the results you will be prompted for vehicle make, year etc. If the test is being carried out on the last vehicle tested then fuel type, vehicle make and year etc. will have already been saved by the handset and will be automatically added to the test result if it is chosen to be saved.The test limits to be applied can be selected from the next menu screen. Simply scroll down and press  . Note ASM (Custom 1) and ASM (Custom 2) limits can be edited to suit testing requirements using PC Analyzer Software – see ‘Tests’, page 61‘ detailed in the PC Analyzer Software section of this manual. Test limits are displayed. Press   to proceed or X to go back to Limit Menu screen to select other limits.Remaining memory screen will be displayed. Note the ASM can require up to a minute and a half to complete.The handset will then detail the same preconditioning , HC hang , Zero Calibration as in other test routines before displaying:TEST LIMITS ASM (Default) ASM  (Custom 1) ASM  (Custom 2)TEST LIMITSHC = 150ppmCO = 1.50%NOx= 1000ppm-OK,    X-Go backDEFAULT ASM       Remaining Memory:         54mins 56 secs-OK,    X-Cancel
Getting Started33The test requires values for both Ambient Temperature and Relative Humidity to be entered. These parameters are required for a Humidity correction factor which is applied to the NOx readings. Use the Up and Down keys to alter each reading then press   enter. When both values have been entered the bottom line of the display will flash. Either press    confirm or ‘Camera’ to change value.The Handset will then display the Insert Probe Screen before displaying screens advising on precautions that should be taken if the ASM test is being conducted during a road test.WARNING – SAFETY PRE CAUTION: ANY ASM TEST CARRIED OUT DURING A ROAD TEST SHOULD BE CONTROLLED BY A PASSENGER AND NOT THE DRIVER.The handset will prompt for the vehicle to be bought to the load or speed for the test.PLEASE ENTER Amb. Temp.   68 F Rel. Humidity 45%-OK,  X-Cancel  ‘Camera’ - RedoASM TESTIf Road Test ensure Probe is fitted  securely-OK,    X-AbortASM TESTIf Road Test ensureAnalyser exhaust is vented out of vehicle-OK,    X-Abort
Getting Started34Press     to Start test.The test readings will be displayed along with a test countdown. Maximum duration of test mode is 90 seconds.At the end of the test the handset will display the result either ‘Inside Limits’ or ‘Outside Limits’. Press     to enter Save Result screen.Press     to save or X to discard. If the test was carried out on a new vehicle the screen will prompt for make, year etc. If the test was carried out on the last vehicle tested then the result will be saved with the previous details. Once the test result has been saved or discarded the handset will display a revised ‘Remaining Memory’ screen before offering the option to ‘Repeat Test’.ASM TEST           Set vehicle to       Desired speed/load-OK,    X-AbortASM TESTHC (150) = 75ppmC0 (1.00) = 0.30%NOx (1000) = 120ppm           77 Secs     X: Abort                                  WITHIN LIMITSHC (150) = 75ppmC0 (1.00) = 0.30%NOx (1000) = 120ppm           -Continue                                TEST RESULT      Default  ASM       TIME:  11:45      DATE: Oct 03 08           -OK,    X-Erase
Getting Started35If the option to finish test is taken the handset will display a prompt to remove the probe when the vehicle is stationary. The Analyzer will automatically detect the probe being removed before conducting a ‘System purge’ then returning to ‘Gas Analyzer’ Menu.System ChecksThere are eight system checks:•Leak Test•Manual HC HangUp•Display PEF•Environmental Data•O2 Cell Functions•Zero Calibration•Base Stn Battery•Bench Data•4/5-gas setupWITHIN LIMITS      Repeat Test?                  -OK,    X-Finish                                REMOVE PROBE      Ensure vehicle is       Stationary and      Parking brake on                  - O K                                   GAS ANALYZER  Live Readings  Simulated TestsSystem Checks                                       
Getting Started36Leak Test - Follow on screen instruction to check integrity of water trap and probe.Manual HC Hangup - Follow the on screen instructions to clear the system of residual HC gas.Display PEF - Displays the current Petrol Equivalent Factor.Environmental Data - This function is for use by a calibration engineer and has no user functionality.O2 Cell Functions - There are two items:•Oxygen Cell Value•Displays the output sensor voltage in mV•Renew Oxygen Cell•This resets the cell error code and recalibrates the analyzer with the new cell (see ‘Oxygen cell replacement procedure’, page 75)Zero Calibration - Follow the on screen instructions to zero the measurement gases by reading fresh air.Base Stn Battery - Displays current base station battery level and whether or not it is being charged 0 for off charge, >200 if on charge.Bench Data - Provides data about the gas bench, has no user functionality.4/5-gas setup - This indicates whether or not a NOx sensor has been fitted, which will determine if the CAT simulated test is activated. BENCH PEF VALUE               0.533      -OK                                BENCH DATAID: 01            S/W Rev: X4        SerNum: 0000196        Date: Oct 12 2007-OK                                BUILD STATE          Bench is 5-Gas         (NOx cell fitted)-OK                                
Getting Started37PC/USB ModeSimply follow on screen instructions.SettingsUnder ‘Settings’ there are seven items:•Manage Files•Display Clock•Change Contrast•View Thresholds•Wireless Setup•S/W Version•Factory DefaultPC/USB MODE      Ensure USB cable is     connected to the PC        and the HandsetX-Cancel                            XSETTINGSManage Files  Display Clock  Change Contrast  View Thresholds                                       X
Getting Started38Manage FilesThere four items under Manage Files:•Summary•List by Date•List by Plate•ActionsThe ‘Summary’ screen displays the number of results stored and the amount of time left in memory, to exit from the screen press  .The ‘List by Date’ screen displays the list of stored tests by date in MM/DD/YY and time of the test. Each result is also listed by ‘Last Test First’. Where relevant, a test result will be listed, e.g. CAT test may list ‘CAT OK’ or ‘Suspect CAT’. The integrity of the data saved will also be listed, e.g. if the test is cut short. To select a test use   or   keys, to view the test details press   - VIEW or to return to ‘Manage Files’ menu press   - GOBACK. When viewing files the following data is displayed: Date and time (of test), Plate (License No.), Type: (live Readings, Simulated tests), Durn: (length of test), Manuf: (vehicle manufacturer), Year (vehicle MY), Fuel: (fuel type), Class: (sedan, station wagon, 4x4 etc.), Opr: (name of the Technician carrying out the test).MANAGE FILESSummary  List by Date  List by Plate  Actions                                       XSTORAGE SUMMARY  2 results  Mem Left:   61m 43s  -OK -VIEW X- GOBACK  02/05/08   10:50    01/31/08    14:20                        
Getting Started39The ‘List by Plate’ screen displays the list of stored by license plate (Registration N°). Each result is also listed by ‘Last Test First’. Where relevant a test result will be listed, e.g. CAT test may list ‘CAT OK’ or ‘Suspect CAT’. The integrity of the data saved will also be listed, e.g. if the test is cut short. To select a test use   or   keys, to view the test details press   - VIEW or to return to ‘Manage Files’ menu press   - GOBACK. When viewing files the following data is displayed: Date and time (of test), Plate (Registration No.), Type: (live Readings, Simulated tests), Durn: (length of test), Manuf: (vehicle manufacturer), Year: (vehicle MY), Fuel: (fuel type), Class: (sedan, station wagon etc.) and Opr: (name of tester).The ‘Actions’ screen displays ‘TEST RESULTS’ and ‘Delete all’ the bottom right of the screen will alternately display   and  . Press   and the screen will display ‘DELETE FILE’ ‘Do you really want to delete ALL results?’. Press   to delete and   to return to ‘Manage Files’ menu.Display ClockWhen you select ‘Display Clock’ the screen displays ‘DATE AND TIME’ and the day, date, and time in hours/mins/secs. If date and time are correct, press   - ‘OK’ returns to the ‘Settings’ menu, press   - ‘Set Clock‘ will allow you to change the year, month, day and time.-VIEW X- GOBACK  OU54XYZ***    --                        TEST RESULTS  Delete all                            DATE AND TIME  Mon 11/02/08   15:05:26  -OK  X-Set Clock                       
Getting Started40The screen will display ‘SETUP DATE AND TIME’, ‘YY 200X’, ‘X’ will flash to indicate that it can be changed by using the   and   will increase or decrease the number. When the next number needs to be changed, e.g. if the year is 2009 and needs to be changed to 2010, the 9 will flash, when changed to 0, press the   to move to the preceding number to change that to 1; this procedure is used for days, hours and minutes. When the correct year is set, press   and cursor moves to ‘MM’ and ‘XXX’ will flash, month can be set by using the   and   buttons until the correct month is set, press  . ‘DD’, ‘Day’, ‘hh’, ‘mm’ are changed and set in the same manner. On completion of setting ‘mm’, the bottom line containing   - OK,   - Cancel will flash, press   to accept the settings and return to ‘DATE AND TIME’ screen, press   to abort settings and return to ‘DATE AND TIME ‘ screen.Change ContrastWhen you select ‘Change Contrast’, the screen displays ‘CHANGE CONTRAST’ and there are four options press   to increase,   to decrease,   - OK to retain changes and return to ‘Settings’ menu, or   - Cancel to exit without saving changes and return to ‘Settings’ menu.SET DATE AND TIME  YY 200X           DAY Wed     MM Feb           hh 11     DD 27            mm 07  -OK,   X-Cancel                         CHANGE CONTRASTÏ-IncreaseÐ-Decrease-OKX-Cancel                       
Getting Started41View ThresholdsWhen you select ‘View Thresholds’ the screen will display the threshold levels of the various gases for all of the simulated test procedures, TSI Default, Custom TSI 1, Custom TSI 2, ASM Default, Custom ASM 1, Custom ASM 2 and CAT. Press   to toggle through thresholds one at a time.Wireless SetupWhen you select ‘Wireless Setup’ the screen will display ‘WIRELESS SETUP’ with item:•Base StationIn ‘Base Station’ the screen displays ‘Set base station address’, e.g. ‘008098E5CCBC’ (this address is preset on build and would not normally require changing). The address is an alphanumerical address and ensures that the handset talks only to that Bluetooth® Wireless Technology address. The Bluetooth® Wireless Technology address should be written in the Quick Guide in the area provided. Use the   and/or   on the handset to set each address character; when correct, press the   button to move to the next character. On completion of entering all of the characters, press   to save, the bottom line (-OK X-Cancel -Move) will flash until  is pressed again to confirm the address and exit. Press   to cancel change and exit.CAT DEFAULT      CO<0.3%     HC<200      O2<1.00%    CO2>14.0%      NOx<100  -OK, X-CancelWIRELESS SETUP Base Station                        WIRELESS SETUP      Set base station            address         008098E5CCBC__________________________  -OK X-Cancel -Move                                       
Getting Started42S/W VersionWith selection of ‘S/W Version’ the screen will display the current down loaded handset software version; for example 00:01:31. This is not a user changeable item, press   to exit.Factory DefaultsWith selection of ‘Factory Defaults’ pressing   will over write any lists that have been downloaded from the PC. Lists that are affected are: Tester Name, Vehicle Type and Vehicle Make.Note: Stored test results will not be affected by this function.Language SelectionWhen selected the handset reverts to the Language Select Menu.S/W VERSION                   00:01:31                    -OK                                       FACTORY DEFAULTS                   WARNING:    Lists downloaded from        PC will be lost                -OK, X_Cancel                                       MAIN MENU  Gas Analyzer  PC/USB Mode  SettingsLang, Idioma                     
Getting Started43To select language simply scroll down menu then press  . The handset will automatically change language and then return to the Main Menu.Low Battery IndicationBase station Low Battery - The battery icon   on the handset flashes empty and the ‘POWER/LO BATT’ LED on the base station flashes.Handset Low Battery - The battery icon   flashes empty and the ‘POWER/LO BATT’ LED on the base station is not affected.The low battery warning is five minutes, if the analyzer is not either placed on charge or switched OFF within that time, the base station will power down automatically, in this event the handset will display the following screen.LANG, IDIOMAUS English  Français  Español                       COMMS LOST           Test terminated.                          S o m e  d a t a  l o s t .                               Low Battery           -OK                                       
Getting Started44Press   to proceed and the handset will prompt you to save the results and to remove the probe from the exhaust.Note: Any saved data will state ‘Some Data Lost’ due to loss in Communications when the base station is powered down. Likewise, if the data is analyzed on the PC software, any data lost will be clearly indicated and all the gas readings for that part of the test will be recorded as zero. After data has been either saved or discarded, the handset will automatically shut down.Out of RangeThe OmiScan Gas should function normally with the handset up to at least 10 metres (30 feet) away from the base station. However, if the handset is taken out of range the display communication icon will disappear and the following text will be displayed:If the handset is taken back within range within a few seconds, the handset will revert to the previous screen shown before it was taken out of range.At the end of the test the handset will give a warning ‘Some Data Lost’ giving the option to save the test.Note: Any saved data will state ‘Some Data Lost’ due to loss in communications when the base station is powered down. Likewise, if the data is analyzed on the PC software, any data lost will be clearly indicated and all the gas readings for that part of the test will be recorded as zero.WARNING            Base station out of             range                    X-Abort                                       XLIVE READINGS              Some data was lost         during the test        due to signal loss.  -OK                                       
Getting Started45If, when out of range, the handset stays out of range the message above will be displayed.You will be given the option of saving the data that was captured up to the point of lost communications.TEST RESULT 3            Live Readings              Time: 14:07          Date: Feb 21 08  -Save   X-Erase                                       LIVE READINGS        Test terminated        due to signal loss              -OK                                       
PC Gas Analyzer Software46PC Gas Analyzer SoftwareSystem requirementsThe minimum PC system requirements to load the OmiScan Gas Analysis Utility software are:•Windows Server 2003, Windows Vista Business, Windows Vista Business 64-bit edition, Windows Vista Enterprise, Windows Vista Enterprise 64-bit edition, Windows Vista Home basic, Windows Vista Home Basic 64-bit edition, Windows Vista Home Premium, Windows Vista Home Premium 64-bit edition, Windows Vista Starter, Windows Vista Ultimate, Windows Vista Ultimate 64-bit edition, Windows XP Service pack 2 or later.Required software: Windows Installer 3.1 or later is recommended.IE 6 or later (required for loading .NET Framework).•Intel or AMD 1200MHz processor or higher•512 MB RAM (This is the recommended requirement by MS for the above operating systems)•330 MB (x86), 660 (x64) of free disk space1024 X 768 minimum screen resolutionCD ROM or DVD Combo driveUSB 2.0The requirements above are the minimum, the higher the PC specification the better the operation of the analysis software.InstallationThe PC software, including the necessary drivers for the handset USB connection, are on the CD that is supplied as part of the OmiScan Gas PC kit.Insert the CD into the CD drive of the PC and the installation program should AutoStart, if there is no response, start Windows Explorer, click on the CD drive and identify the setup.exe file. Double click on this file and the installation should now proceed.Using the SoftwareDouble click on the OmiScan Gas icon on the desktop. The OmiScan Gas application will start to load. You can also use the ‘All Programs’ menu to access the application; press the start button at the bottom left hand side of the screen, press ‘All Programs’. From the displayed list select Omitec\OmiScan Gas\OmiScan Gas Analysis Utility.
PC Gas Analyzer Software47Analyzer Software FunctionsThe OmiScan Gas Analyzer provides the facility to view the data collected by the OmiScan Gas handset, which is transferred to the PC. The data is analyzed in bar or line graph format. The title bar displays ‘OmiScan Gas Analysis Utility’ and five ‘buttons’: •CSV File   - Export the current test as a Comma Separated Values (CSV) file•Bitmap file   - Export the currently visible graph section as a bitmap•Save   - Saves the test currently being quick viewed•Search   - Searches the tests saved to PC using the selected search criteria•Print   - Prints the test details and the visible graph sectionThe buttons will be ‘disabled’ if they are not applicable to the function currently on screen.There are six tabs to select screens and the default screen is ‘Search’, which is the second of the tabs:•Handset Results•Search•Results Details•Graph•Setup•Help
PC Gas Analyzer Software48Handset ResultsWhen the handset is connected to the PC with the USB cable, communication with the handset is established; the left hand pane displays all of the stored vehicle test results in the handset identified by test type, date, license plate details and manufacturer. The right hand side displays three icons representing the functions of‘Save data’  , ‘Display data’   and ‘Delete data’  . To activate any of these functions, click and hold the left mouse button over the data item, drag the item over the desired function, release the left mouse button.Note: Handset Results tab will be disabled if the handset is not connected.‘Save Data’ will display a dialog box prompting you to select either ‘Create New’ or ‘Add To’ one of the list of existing customers. When the data is saved the screen will automatically change to ‘Results Detail’. To use the ‘Add To’ save option the license number, or part of it, must be entered.
PC Gas Analyzer Software49‘Display Data’ will automatically load the data into the ‘Graph’ screen. The ‘Save’ button in the title bar is now activated and when pressed displays the same dialog box as in ‘Save Data’ above.‘Delete Data’ displays a dialog box asking you to confirm deletion.
PC Gas Analyzer Software50Search‘Search’ allows stored data to be retrieved for later analysis, reference, historical recall etc. A general search for all stored data files can be started by pressing the ‘Search’ button (binoculars).
PC Gas Analyzer Software51A specific search can be initiated from one or all of the following input cells:•Customer Name•Vehicle License•State•Make•Model•Type•Use Date Range with ‘From’ and ‘To’ date selectionAny text item used as a search parameter can have a ‘wildcard’ character ‘*’ within it if the whole text is not known. For example, if searching for a vehicle but the whole registration is not known, entering ‘OU5*’ will match with any registration staring with those characters.With data input, pressing either ‘Enter’ on the keyboard or the ‘Binocular’ button in the title bar will invoke a search with the result/s displayed on the right hand side of the screen.Select and double click on a file and the screen will change to ‘Result Detail’. The left pane will list all of the tests completed on the vehicle; the test list is expanded by double clicking on the ‘+’ sign. Double clicking on a specific test will change the screen to ‘Graph’ displaying the data from that test. Selecting a test and then clicking on the ‘Graph’ tab will also display that data in graphical detail.
PC Gas Analyzer Software52Results DetailThe ‘Results Detail’ saves the test data under the vehicle rather than customer, therefore any new data from an already saved vehicle will be added to the vehicle records. The vehicle records are listed in the left pane in three areas.Car Symbol (License) •This lists details of make, type, year, fuel type, State and an area for you to add notes. (The data is down loaded from the handset but can be edited.)
PC Gas Analyzer Software53Man Symbol (Customer) •Cells are available to enter customer’s details.Clipboard Symbol (Test results) •Each test is identified by type, test name, date, tester’s name and notes area.Note: Selecting any test will enable the ‘Graph’ tab.
PC Gas Analyzer Software54GraphTest Result can be presented in raw data line graph format or by test specific bar and line graphs. The available graph views depend upon the type of test carried out but the raw data line graph is always available.The test specific graphs (identified by ‘R’ on the graph icons) present a view of the data, but not the raw data, from only a portion of the entire gas recording. This may be, for example, the last 10 seconds of readings as an average. The bar graph option displays the highest levels within this data portion whilst the line graph will plot this data portion against time.
PC Gas Analyzer Software55The raw data line graph is available for all tests and is a plot of non-adjusted values against time.On the graph tab you can export the test results into a CSV file by pressing the file button on the top tool bar. You can also export the graph image as a bitmap by pressing the picture button on the top tool bar.
PC Gas Analyzer Software56Bar graph: Readings that can be selected are: CO, CO2, O2, HC, NOx, MPH, AFR, and LAMBDA. On opening results the displayed readings will default to relevant gases. For example, if the test was a CAT then CO, CO2, HC, NOx and O2 will be displayed. Because all gases are recorded during a test they can be added to the graph by clicking on the gases in the DISPLAY bar down the left hand side of the screen.The scale of the Y axis can be changed to read any gas by selecting from the key down the right hand side of the screen, double click on the gas required.If preset cut off points are set in test then the gas colour will change to red over the cut off point, therefore any area in red is in excess of the predefined acceptance level.A horizontal guide line, moveable with the mouse, helps the user to read gas values.For a print out of result, click on ‘Printer’ symbol in the top right hand corner of the screen, when pressed you are prompted for language of print out – English, Spanish or French. A prompt will be displayed for which graph is to be printed. The Garage Details, vehicle details and time and date stamp will be added to the print out.
PC Gas Analyzer Software57Line graph: CO, CO2, O2, HC, NOx, MPH, RPM, AFR and LAMBDA are plotted against time.•Gases displayed depend on type of test, but like bar graph mode other gases can be added to plot.•Y axis scale can be selected to a chosen gas and its graph will become bolder.•Cut off limits will be shown on plot for the active series.•A Landscape graph (only available on the raw data graph view), showing a complete record of the gas readings, is under the main display. The Main display seen on the screen will be presented as a window on the landscape graph. To view readings anywhere through out the test sequence simply click and drag landscape window. To zoom in or out further on a reading click on arrows in the bottom right hand corner of the main graph.•To interrogate gas readings further, hold mouse anywhere in the main graph. A vertical ‘Wand’ will appear and when held over a reading a ‘INFO’ box (only available in raw data graph view) will appear. This box will:1. Display gas values on plot.2. Give Cut off Status – A green tick indicates all gases are below cut off points, a red exclamation mark will list all gases above limits whether they are on the plot or not.3. Information speech bubble – based on combination of gas readings a suggested diagnostic hint on what could be wrong with the vehicle will be displayed. Simply click on bubble to revert to a diagnostic guide.
PC Gas Analyzer Software584. Add ‘Snap Shot/View Snap Shot’ Camera symbol – Click on camera symbol to add user comment and save. A green circle will be placed top and bottom of both graph displays to tag Snap shot.Note: Any Snap shot made during a test by pressing the Camera button on the handset will be automatically tagged on graph. To view a Snapshot, click near the diamond on the landscape graph to set the window over snapshot, place mouse over reading on main graph to bring on the INFO box then simply click on ‘VIEW SNAP SHOT’. To delete Snapshot click Delete when viewing Snap shot information.5. Print out function is the same as Bar graph mode.When the graph is in line view you can select the analyze button   (below the line graph button), left hand side of screen.
PC Gas Analyzer Software59This will activate diagnostics for the current visible frame of graph. Any possible diagnosis made within this section will be displayed as a blue triangle as a prompt to move the cursor to that graph point. Once the mouse is over that point, the ‘INFO’ box will be displayed as normal. The purpose of this is to be able to quickly check a section of graph for possible diagnosable problems.
PC Gas Analyzer Software60TSI GraphThe left hand bar graph indicates CO and HC values for the idle mode section of the test. The right hand bar graph indicates the CO and HC values for the High Speed Mode of the test. A window in the top left hand of each graph will state ‘Inside Limits’ or ‘Outside Limits’ as the case may be.Note: ASM graph - Test result will be presented as a single bar graph indicating the CO, HC and NOx values from the test. Again ‘Inside or Outside Limits’ will be shown as appropriate.SetupThe setup section has four sub-sections selected by tabs on the right of the screen:1. Tests2. Lists3. Shop details4. Setup
PC Gas Analyzer Software61TestsBoth the ASM and TSI simulated tests offered by the OmiScan Gas can be carried out using different limits (Cut points). Each Test offers two sets of Custom limits as well as a set of default limits. The PC analyzer software allows the custom limits to be changed on the handset. Click on Tests Tab then on the drop down menu arrow to list tests.Click on either ASM or TSI to list limits. If the limits on the handset match those on the
PC Gas Analyzer Software62PC Analyzer software the limits will be displayed in green (In Sync). Click on either of the Custom limit boxes to display the individual gases and their limits. To change limits click on up or down arrows for the relevant gas to set the desired value. Click on limit box or press return on the PC keyboard to enter value. The test limit box will change to Red to show the test limits on the handset no longer match those set on the PC Analyzer software (Out of Sync).To update the handset with the new edited test held in the PC, click and hold over the specific test to transfer, and drag the test over the handset image and release the mouse button.On completion of downloading, the list box will change from red to green.Note: Drag and drop facility will be disabled whilst the handset is disconnected from the PC.
PC Gas Analyzer Software63ListsVehicle make, vehicle types and testers can be edited and saved to the handset by using the drag and drop facility. Double click on any of the lists, ‘Vehicle Makes’, ‘Vehicle Types’ or ‘Testers’ and the current data stored on the handset is shown in the right pane.The + button can be used to add to a list, likewise, the - button can be used to delete from a list.  Any list will be in alphabetical order. To position an item, for instance, a vehicle make towards the top of the list, click on the appropriate make then click on the Up button to move it up the list. This useful feature allows the list to customized to individual requirements.Like Tests a list will be either green if the list is In Sync with the handset and Red if list is Out of Sync with the handset.
PC Gas Analyzer Software64If the Handset is not connected to the PC there will be no picture of the Handset and a warning ‘Handset Not Present’ will appear in the left pane and the status of any lists or tests will be red.
PC Gas Analyzer Software65The lists for Vehicle type and Tester can be edited in the same way as Vehicle Make.Again like Tests to update the handset to the list held in the PC, click and hold over the specific list to transfer, and drag the list over the handset image and release the mouse button. On completion of downloading, the list box will change from red to green.Note: Drag and drop facility will be disabled whilst the handset is disconnected from the PC.
PC Gas Analyzer Software66Shop: Contained in the PC and is added to the print out.If no printer is installed on the PC a print preview is not available and warning will be shown if the ‘Preview’ button is pressed.
PC Gas Analyzer Software67Setup: Sets language and enables/disables tool tips.HelpThe help tab will link to a document providing a guide to the functionality tabs. The Navigation pane on the left also allows access to the diagnostic documents that are linked from the dialog box in the ‘Graph’ section if any of the gases are out of limits.
PC Gas Analyzer Software68CommunicationsThe Bluetooth® Wireless Technology communications provides connection between the base station and the handset. The base station has a unique addresses and it is advisable to note this address in the data section of this manual, especially if either unit is replaced.Software BuildDetails of the software version and build can be accessed from the ‘Help’ function in the top bar left hand side. Click on this function and then on ‘About’ will display a dialog box containing version and build details of the OmiScan Gas PC Analyzer Utility program.
Gas Analysis69Gas AnalysisOmiScan Gas Analysis of ResultsCO -Carbon Monoxide - Carbon Monoxide is the product of incorrectly burned fuel, so the obvious aim is to ensure that the CO level is as low as possible, the better the combustion the lower the CO. However, due to unavoidable inefficiencies with  Internal Combustion Engine (ICE) there will always be an output of CO.CO2 - Carbon Dioxide - Carbon Dioxide is the result of correctly burned fuel so the obvious aim is to ensure that the CO2 level is as high as possible, typically 13% - 15% for a catalytic converter vehicle. Non-cat would be typically 11% - 12.5%.HC - Hydrocarbons - Hydrocarbons are basically unburnt fuel and so the obvious aim is to have HC as low as possible but inefficiencies mean there is always an output of HC.O2 Oxygen - Oxygen has to be present for combustion to take place so obviously the aim is to have as little surplus oxygen as possible, typically up to 2%.NOx - Nitrous Oxide is a gas generated by high combustion temperature and pressure and this can be reduced by design of the exhaust and inlet valve overlap and exhaust gas recirculation where the inert exhaust gas introduced into the inlet will reduce the combustion temperature. NOx can only be measured when the engine is under load.AFR/Lambda - Air Fuel Ratio (AFR) or Lambda is the ratio of air to fuel and is calculated from the amount of oxygen sensed in the exhaust gases.From the exhaust gases the following relationships are formed:•O2 combines with CO to form CO2•CO is an indicator of air fuel mixture richness•HC is an indicator of fuel mixture leanness or richness and misfires•CO and O2 are equal at the stoichiometric air/fuel ratio
Gas Analysis70•CO2 and O2 are indicators of the integrity of the exhaust system and/or the sample hose and probe•CO2is an indicator of combustion efficiency that peaks at or near the stoichiometric air/fuel ratio and decreases with rich or lean air/fuel ratios•If CO2 increases, O2 decreases (inversely related)If O2 increase, CO2 decreases (inversely related)During a cold start, and no secondary air injection, CO will be above 1% and the catalytic converter will be O2 starved and will not ‘fire’ and the emissions will be the same as pre-cat levelsCorrected COThe Live Readings test offers the selection of Corrected CO (COK).Corrected CO takes into account any dilution of the sampled gas, for example leaks in the exhaust pipes and silencers. The formula to calculate COK is:COK = CO x 15/CO2When stoichiometric combustion is achieved and there is no dilution, the CO2 value drops to approximately 15%, therefore, the correction factor is unity. If the sample is diluted with air, the CO2 value falls below 15% and the correction factor is <1.Non-Catalyst SystemsPerfect combustion in an ICE has an AFR of 14.7:1. This is measured by weight, any reading higher than 14.7 indicates a lean mixture and likewise any reading lower than 14.7 indicates a rich mixture. For diagnostics purposes the AFR must be settled at 14.7:1 before analysing any other gas reading. With AFR correct the following table shows how the levels of other gases can indicate a problem with the engine with relation to the time of combustion.Non-Catalyst System Gases with AFRThese readings are typical and reference should be made to the manufacturers’ specifications, which usually only refer to CO and HC. Setting the CO level correctly should bring the other gases to high CO2, low HC and low O2.AFR 14.7Gas Perfect Minimum Maximum ProblemCO In SpecificationCO2Highest possible 11% Before combustionHC Lowest possible 400ppm During combustionO2Lowest possible 2% After combustion
Gas Analysis71Catalyst SystemsWhere AFR is used for Non-Cat systems, Lambda is used for Cat systems and the engine management systems control Lambda to 1. Any value >1 is a lean mixture and anything <1 is a rich mixture.Catalyst System Gases with LambdaThe catalytic converter converts CO and HC into water (H2O) and extra CO2. Cats increase the CO2 level, therefore they are useful for reducing acid rain rather than the ozone layer. With a good operating system it should be possible to achieve 0 for both CO and HC under normal operating conditions.Fault DiagnosisThe following tables show various relationships between gases and suggest a possible cause for these gas levels. H = High, VH = Very High, L = Low, VL = Very Low, N = Normal.5 Gas RelationshipLambda 1Gas Perfect Minimum Maximum ProblemCO In Specification 0.5CO2Highest possible 12.5% 15 Before combustionHC Lowest possible 50ppm During combustionO2Lowest possible 2% After combustionHC O2CO CO2NOx Possible causeH VH Vacuum leakH H VL Overactive EGR valveH L H L L Running richH H H L H Running leanH Increased combustion pressure or VH O2 after combustionVH N VL VL Zero Blown head gasketHHHL H Faulty CAT
Gas Analysis72High Gas Levels4 Gas RelationshipGas Possible causeHigh CO EVAP system problemPCV system problemFuel system problemControl system problemHigh HC/CO Fuel system problemRunning richBlown head gasketHigh HC Mechanical system problemElectrical system problemFuel system problemRunning leanWarm air intake problemCat problemEVAP system leakHigh NOx High combustion pressureHigh combustion temperatureHigh O2 contentHigh HC/NOx Cat problemCO CO2HC O2ProblemH L H H Rich mixture with ignition misfireH L H L Faulty thermostat or ECT sensor - rich mixture faultL L L H Exhaust leak after CatL H L H Injector misfire with Cat at operating temperatureH L N H Rich mixtureH H H H Injector misfire, Cat not working - Rich mixture & vacuum leakL L H H Ignition misfire; lean mixture; unmetered air leakL H L L Good combustion efficiency with Cat at operating temperature
Gas Analysis73Lambda status5 Gas - Catalytic Vehicle Typical ReadingsThese are typical limit values of the 5 gasses for fuel injected vehicles and are not specific to any vehicle:•CO < 0.3%CO2 > 14% HC < 200 ppmO2 <1%NOx < 100 ppmCondition ResultToo lean Poor engine powerMisfiring at cruise speedBurnt valvesBurnt pistonsScored cylindersSpark knockSlightly lean High gas mileageLow exhaust emissionsReduced engine powerSlight tendency to knockStoichiometric Best all round performanceSlightly richMaximum engine powerHigher emissionsHigher fuel consumptionLower tendency to knockToo rich Poor gas mileageMisfiringIncreased air pollutionOil contaminationBlack exhaust
Routine Maintenance & Cleaning74Routine M aintenance  & CleaningRoutine MaintenanceTo ensure that the analyzer gives long and reliable service, regular cleaning and changing of the filters is necessary. Failure to change the disposable coalescing filter elements regularly may void warranty and service agreements.Check the cleanliness of the disposable coalescing filter (Part No. FL5720A or OM 4700/11/10 (Pack of 10)) at the start of each day. If the filter shows signs of severe contamination or a build up of deposits has occurred, remove and clean out the housing using a lint-free cloth. Ensure that no deposits are pushed into the exit pipes of the filter housing. Renew the disposable filter if heavily contaminated.REPLACING COALESCING FILTERTo access Coalescing filter (FL5720A), firstly remove clear catch pot from the water trap by unscrewing it clockwise. Then unscrew black retaining column, again clockwise to release filter.Once replaced, screw filter back in and refit catch pot. NOTE: Both filter and catch pot only need to be finger tight. This will ease future filter replacement.NOTE: Any dismantling of probe tip, sample hose or filters MUST be followed by a manual leak test.CAUTION: DO NOT run the analyzer without the filters in place!At the beginning of each day, drain any water collected in the sample pipe and the water trap. OM1398
Routine Maintenance & Cleaning75DRAINING WATER FROM WATER TRAPTo drain water from the water trap simply screw the brass drain fully clockwise.When the water has drained, screw the brass drain fully counter clockwise to seal water trap.NOTE: The drain only needs to be finger tight. Do not over tighten as this may deform the rubber seal. Always conduct a Leak Test after draining the water trap to ensure an air leak has not been introduced into the gas sampling chain which will cause false readings. See ‘System Checks’ in ‘Gas Analyzer’ Menu.The sample probe tip can be replaced as a separate item, or may be removed for cleaning. Assembly will require a small quantity of “Loctite 222”, or similar thread sealant, to be used on the threads to ensure a good gas seal.At the start of each day, clean any accumulated deposits from the Sample Probe tip, making sure that the two cross-drillings in the tip are clear and check that the analyzer outlet pipes are not kinked or blocked in any manner.Ensure that the analyzer is not operated in a dusty environment, especially areas where paint spray and dust spray are present.CAUTION: Contamination of the analyzer by paint spray or excessive dust may void the warranty and/or service agreements!Oxygen cell replacement procedureWhen an exhausted oxygen cell is identified during a zero calibration, the handset automatically links to ‘System Checks/O2 Cell Functions’. Select ‘Oxygen Cell Value’ to ensure that the O2 cell output is outside of the expected level of between 9mV and 11mV. If the cell has failed, change the cell as detailed:1. Firstly ensure there are no leaks or blockages in the gas handling system. If a blockage is found, clear and retest ‘Zero Calibration’, else proceed to 2.2. Ensure power to the unit is switched off, then go to the rear of the unit and locate the O2/NOx cells cover (see ‘Assembling the OmiScan Gas Analyzer’, page 8).3. Remove the two retaining screws and remove the cover.OM1399
Routine Maintenance & Cleaning764. Disconnect the cable from the red oxygen cell. It may be advantageous to gently prize the locking tab of the connector away from the mating half, to facilitate disconnection.5. Unscrew the cell by hand (unscrew counter-clockwise). If discarding the cell, place to one side for later safe disposal.6. With the new cell, ensure the “O” ring is properly seated against the thread abutment. Screw in the cell (clockwise) to hand-tightness but do not over-tighten. Ensure a gas tight seal. Refit the connector and screw the ground lead to the centre post, making sure the NOx cell ground lead is also secured.7. Ensure the Oxygen Cell is properly seated in the housing and provides a gas tight seal.8. Refit the cover, make sure the gas tubing is not crushed or trapped and secure cover to the back of the unit.9. Turn on the power to the unit and allow it to warm up. Ensure that the analyzer is fully warmed up (i.e. that it has had at least 5 minutes since power on).10. Use handset to configure the OmiScan Gas with the new O2 cell:•Select ‘Renew O2 Cell’.•This will reset the error code and recalibrate the Analyzer with the new cell.11. If the unit still fails, contact your service engineer.The O2 cell can be monitored to see if it is approaching the end of its life. This can be done by checking its cell voltage. This should be done with the probe in clean air and accessing the following menu on the handset:•Select ‘Gas Analyzer’ – ‘System Checks’ – ‘O2 Cell functions’ – ‘Oxygen Cell Value’A good cell should read between 9mV and 11mV. The Analyzer will automatically flag up an error if the cell voltage falls below 7mV. It is recommended that the cell is changed if voltage is nearing the 7mV threshold.NOx Cell ReplacementIn theory the NOx cell should have a life expectancy of 2 years. Unlike the O2 cell, the OmiScan Gas will not flag an error in the NOx cell.It is recommended, therefore, that the NOx channel, like the rest of the OmiScan Gas, is regularly tested with calibration gases of known concentration. This can be carried out by an Omitec authorized Service agent.NOTE: Due to the operating characteristics of the NOX sensor, any necessary replacement will need to be calibrated by an authorized Omitec authorized Service agent using a NOx calibration gas.
Routine Maintenance & Cleaning77Calibration and Accuracy ChecksIt is recommended that periodic Calibration and Accuracy testing should be carried out on the OmiScan Gas Analyzer to ensure reliable and accurate testing.It is also recommended that this be done once every 12 months on average. This can be carried out by an Omitec authorized Service agentCleaningOmiScan Gas unitThe surfaces of the analyzer cabinet can be cleaned with a soft sponge, dampened with a mild soap and water solution. CAUTION: ENSURE that the unit is disconnected from the mains before cleaning. Be careful not to wipe electrical connections or connectors while cleaning.CAUTION: On no account use solvents or strong detergents to clean the unit.
Technical Specification78Technical Spe cificationTechnical SpecificationThe OmiScan Gas unit (Part No. OM 4720/1)EnvironmentalOperational temperature range 0°C (32°F) to 45°C (113°F)Operational temperature (nominal) 20°C (68°F)Storage temperature range -4°C (24.8°F) to 70°C(158°F)Relative Humidity (operational) 95% RH (non-condensing)Relative Humidity (storage) 99% RH (non-condensing)Atmospheric pressure range 813 to 1060 mBarAtmospheric pressure (nominal) 1013 mBarAltitude (operational) -300 m (-984 ft) to +3000 m (+9836 ft)Sunlight No direct exposure to sunlightOperating Constraints Designed for covered use onlyNot for external useMechanicalOperating Position HorizontalVentilation Still AirFree from strong draftsWeight Total 2.21 kg (4.9 lbs)Handset 400g (0.88 lbs)Base 1.81 kg (3.97 lbs)Dimensions of main unit 192 mm wide283 mm deep180 mm high (including 10mm feet)Dimensions of handset 108 mm wide160 mm deep38 mm highCharger Power SupplyMains voltage range 100 – 240V ACMains frequency range 50-60 HzPower input 60 WBase station battery pack 9.6V Nom. 8 cell rechargeable NiMHHandset battery pack 2.4V Nom. 2 cell rechargeable NiMH
Technical Specification79Measurement range (OIML Class 0)Carbon monoxide (CO) range 0 to 15% vol.CO accuracy ±0.03% abs or ±5% rel.Carbon dioxide (CO2) range 0 to 20% vol.CO2 accuracy ±0.5% abs. or ±5% rel.Hydrocarbons as n-hexane (HC) range 0 to 2000 ppm vol.HC accuracy - 0 to 2000ppm ±10 ppm or ±5% rel.PEF Factor 0.490 to 0.540Transducer measurement rangeCarbon monoxide (CO) range 0.00 – 15.00%CO accuracy:0.00% to 10.00%10.01% to 15.00%±0.02% abs. or ±3% rel.±5% rel.Carbon dioxide (CO2) range 0.00 to 20.00%CO2 accuracy:0.00% to 16.00%16.01% to 20.00%±0.3% abs. or ±3% rel.±5% rel.Hydrocarbons as n-hexane (HC) range 0 to 2000 ppmHC accuracy:0 to 2000ppm ±3% rel.Hydrocarbons as propane (HC) range 0 to 30000 ppmHC accuracy0 to 4000 ppm2001 to 15000 ppm15001 to 3000 ppm±3% rel.±5% rel.±8% rel.Measurement of OxygenOxygen range 0.00% to 25.00%Accuracy ±0.01% abs.Response time < 30s to 95% of final value< 60s for change from 20.9% to < 0.10% vol.Measurement of Nitric Oxide (NOx)NOx range 0 to 5000 ppmAccuracy ±3% rel.
Technical Specification80Response time of analyzer: <15 secs to display 95% of final values (CO, CO2 & HC)Calculation of Lambda value according to Brettschneider (Source: Bosch Technische Berichte, VOL. 6 (1979) No. 4, p177-186) – As listed in OIMLR99 specification.Simplified Lambda calculation:General DataReal time clockThe handset has:•Day and Date, time in hours, minutes and seconds with automatic leap year compensation.•Time is user adjustable.Handset•Bluetooth® Wireless Technology CommsRange typically 10m (30 feet)PC Interface•USB 1.1 or 2.0Lambda Calculation (by Brettschneider formula)Lambda range 0.0 to 2.0Resolution 0.01[ ] Concentration in % vol (ppm vol for HC)K1 Conversion factor for HC if expressed in ppm vol n-hexane (C6H14) equivalent. It’s value in this formula is 6x10-4HCV Atomic ratio hydrogen to carbon (HCV = 1.7261)OCV Atomic ratio oxygen to carbon (OCV = 0.0176)
Technical Specification81Ancillary EquipmentPneumatic SpecificationsVehicle Supply LeadCable length 6 footFor vehicles with auxiliary socket in the luggage compartmentFlow RateAnalyzer Outlet (nominal) 1.0 l/minSample Inlet 1.0 l/minFiltrationCoalescing element-PVDF Fluorocarbon 95%+ efficiency at particles > 0.1 microns57 mm lengthOxygen CellCity Technology CiTicel AO2Calibration GasMinimum pressure 700 mBar above atmospheric pressureFlow rate 1.5 l/minComposition (4 Gas):Carbon dioxide 14% vol.Carbon Monoxide 3.5% vol.Propane 2000 ppm vol.Nitrogen BalanceComposition NOx Calibration Gas:NOx 1000 ppmNitrogen Balance
Technical Specification82Service and calibration frequencyCalibration - At 12 monthly intervalsCountry Specific ApprovalsFCCOmiScan Gas Handset OM4720/2:FCC CFR47, Part 15 Subpart BThis device complies with part 15 of the FCC Rules. Operation is subject to the following 2 conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation.Grantee and Equipment Product Code:FCC ID: SV4-OM4720‘We hereby declare that product OmiScan Gas Handset OM4720/2 is a class 2 BT device (max 2.7 mW EIRP). Due to the low level of transmit output power the device is not subject to the routine RF exposure valuation as per Section 2.1093 of the FCC rules’.OmiScan Gas Base Station OM4720/1:FCC CFR47, Part 15 Subpart BThis device complies with part 15 of the FCC Rules. Operation is subject to the following 2 conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation.Note: This Product contains FCC ID: PI403BServiceReplace disposable filter element (0.1 µm) – Part No. OM 4700/11/10 (Pack of 10)Weekly – if in continuous use, else as necessarySingle disposable coalescing filter (0.1 µm) – Part No. FL5720AWeekly – if in continuous use, else as necessaryIn Line hydrophobic pancake filter – Part No. FL5718AAs necessaryCleaning of sample probe and hose Weekly – if in continuous use, else as necessaryReplace oxygen cell - OM4000/12  Two years, or as neededReplace NOx sensor - OM4700/12 Two years, or as needed
Technical Specification83INDUSTRY CANADA:OmiScan Gas Handset OM4720/2:RSS210, RSSGENOperation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation.Company Number and UPN Number:IC: 6220A-OM4720‘We hereby declare that product OmiScan Gas Handset OM4720/2 is a class 2 BT device (max 2.7 mW EIRP). Due to the low level of transmit output power the device is not subject to the routine RF exposure valuation.’OmiScan Gas Base Station OM4720/1:RSS210, RSSGENOperation is subject to the following 2 conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation.Note: This Product contains IC:1931B-EUSB.
Technical Specification84Spare Parts ListPlease quote the part number when ordering or enquiring about the availability or price of spares for you OmiScan Gas analyzer.Description Part numberTop level kit OM4720Base station OM 4720/1OmiScan Gas Handset OM 4720/2PC Analyzer Application Software CD OM 4720/3Instruction Manual OM 4720/4Exhaust Sample Probe complete with hose OM 4700/5Battery Charger PSU OM 4700/6Handset 2 battery pack OM 4700/7Base Station 4 battery pack OM 4700/8Sample Probe Hose extension OM 4700/10Coalescing Filter (box of 10) OM 4700/11/10NOx sensor OM 4700/12Vehicle power lead OM 4700/13Oxygen Sensor Cell OM 4000/12Sample Probe Cap RB4001AOmiScan Gas USB Lead CB9166AIn line hydrophobic pancake filter FL5718ACoalescing Filter FL5720A