Flash HT Book 31707010 (Rev.B) EA Iso Link CNSOH Operating Manual

User Manual: Pdf

Open the PDF directly: View PDF PDF.
Page Count: 186

DownloadFlash HT Book 31707010 (Rev.B) EA Iso Link CNSOH Operating Manual
Open PDF In BrowserView PDF
EA IsoLink IRMS System
for CNSOH
Including the Flash IRMS Elemental Analyzer

Operating Manual
31707010

Revision B

December 2016

EA IsoLink IRMS System
for CNSOH
Including the Flash IRMS Elemental Analyzer

Operating Manual
31707010

Revision B

December 2016

© 2016 Thermo Fisher Scientific Inc. All rights reserved.
Thermo Scientific™ EA IsoLink™ IRMS System for CNSOH, Thermo Scientific™ Isodat™ Software Suite,
EagerSmart™ Data Handling and Thermo Scientific™ AI 1310/AS 1310 Autosamplers are trademarks of
Thermo Fisher Scientific Inc., and its subsidiaries.
Other brand and product names may be trademarks or registered trademarks of their respective companies.

Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or errorfree and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that
might result from any use of this document, even if the information in the document is followed
properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.

Release history:
Revision A, released October 2016, “Original Instructions”
Revision B, released December 2016

For Research Use Only. Not for use in diagnostic procedures.

Reader’s Survey
EA IsoLink IRMS System for CNSOH, PN 31707010, Revision B

The manual is well organized.
The manual is clearly written.
The manual contains all the information I need.
The instructions are easy to follow.
The instructions are complete.
The technical information is easy to understand.
Examples of operation are clear and useful.
The figures are helpful.
I was able to operate the system using this manual.

fold

Agree

Neutral

Disagree

2
2
2
2
2
2
2
2
2

3
3
3
3
3
3
3
3
3

4
4
4
4
4
4
4
4
4

Strongly
Disagree
5
5
5
5
5
5
5
5
5

If not, please comment below. Attach additional sheets if necessary.
__________________________________________________________
__________________________________________________________
__________________________________________________________
__________________________________________________________
__________________________________________________________
__________________________________________________________
__________________________________________________________
__________________________________________________________
__________________________________________________________

Customer Registration Card

fold

Strongly
Agree
1
1
1
1
1
1
1
1
1

__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________
__________________________________________________________________

Register now…and receive all the privileges associated with being a Thermo Fisher Scientific product user including customer
support, application reports, and technical reports.
MY ORGANIZATION IS: (Check only one)
MY PRIMARY APPLICATION IS: (Check only one)
❏ Commercial (for profit) lab
❏ Analytical
❏ Government lab
❏ Biomedical
❏ Hospital/Clinic
❏ Clinical/Toxicology
❏ Industrial lab
❏ Energy
❏ Research Institute
❏ Environmental
❏ University/College
❏ Food/Agricultural
❏ Veterinary
❏ Forensic/Toxicology
❏ Other______________________
❏ Pharmaceutical
❏ Research/Education
MY PRIMARY JOB FUNCTION IS: (Check only one) ❏ Other______________________
❏ Administration
❏ Lab management
❏ Operator
❏ Other______________________
Name __________________________________________________Title__________________________________________________________________
Company __________________________________________________ __________________________________________________________________
Address ___________________________________________________ __________________________________________________________________
City/State _________________________________________ Postal Code__________________________________________________________________
Country ___________________________________________________ __________________________________________________________________
Telephone ______________________________________________ Ext. __________________________________________________________________
Serial Number __________________________________ Date purchased __________________________________________________________________

Fold and mail or email to:
Editors, Technical Documentation Thermo Fisher Scientific (Bremen) GmbH
Hanna-Kunath-Str. 11
28199 Bremen, Germany
Tel. +49 421 54930
Email: documentation.bremen@thermofisher.com

Declaration
Manufacturer: Thermo Fisher Scientific
Thermo Fisher Scientific is the manufacturer of the instrument described in this manual and, as such, is responsible
for the instrument safety, reliability and performance only if:
• installation
• re-calibration
• changes and repairs
have been carried out by authorized personnel and if:
• the local installation complies with local law regulations
• the instrument is used according to the instructions provided and if its operation is only entrusted to qualified
trained personnel
Thermo Fisher Scientific is not liable for any damages derived from the non-compliance with the aforementioned
recommendations.
Thermo Fisher Scientific (Bremen) GmbH
Hanna-Kunath-Str. 11; 28199 Bremen, Germany; Tel. +49 421 54930

Regulatory Compliance
Thermo Fisher Scientific performs complete testing and evaluation of its products to ensure full compliance with
applicable domestic and international regulations.
When the system is delivered to you, it meets all pertinent electromagnetic compatibility (EMC) and safety
standards.

Safety
This device complies with the following safety standards according to Low Voltage Directive 2014/35/EU.
•

EN 61010-1:2010 (3a ed.); IEC 61010-1:2010 (3rd ed.); CAN/CSA C22.2 No. 61010-1-12;
UL 61010-1:2012; IEC 61010-2-010:2014 (3rd ed.); IEC 61010-2-081:2015 (2nd ed.)

Electromagnetic Compatibility
This device complies with the following regulations on Electromagnetic Compatibility (EMC) and Radio
Frequency Interference (RFI) according to directive 2014/30/EU:
•

IEC 61326-1:2013 (2a ed.); IEC 51326-1:2012 (2nd ed.)

IMPORTANT: Class A equipment is intended for use in an industrial environment. In others
environments there may be potential difficulties in ensuring electromagnetic compatibility, due
to the conducted as well as radiated disturbances.

FCC Compliance Statement
THIS DEVICE COMPLIES WITH PART 15 OF THE FCC RULES. OPERATION IS SUBJECT TO
THE FOLLOWING TWO CONDITIONS: (1) THIS DEVICE MAY NOT CAUSE HARMFUL
INTERFERENCE, AND (2) THIS DEVICE MUST ACCEPT ANY INTERFERENCE RECEIVED,
INCLUDING INTERFERENCE THAT MAY CAUSE UNDESIRED OPERATION.

CAUTION Read and understand the various precautionary notes, signs, and symbols contained
inside this manual pertaining to the safe use and operation of this product before using the device.

Notice on Lifting and Handling of
Thermo Scientific Instruments
For your safety, and in compliance with international regulations, the physical handling of this Thermo Fisher
Scientific instrument requires a team effort to lift and/or move the instrument. This instrument is too heavy and/
or bulky for one person alone to handle safely.

Notice on the Proper Use of
Thermo Scientific Instruments
In compliance with international regulations: Use of this instrument in a manner not specified by Thermo Fisher
Scientific could impair any protection provided by the instrument.

Notice on the Susceptibility
to Electromagnetic Transmissions
Do not use radio frequency transmitters, such as mobile phones, in close proximity to the instrument.

WEEE Directive
2012/19/EU

Thermo Fisher Scientific is registered with B2B Compliance (B2Bcompliance.org.uk) in the UK and with the European
Recycling Platform (ERP-recycling.org) in all other countries of the European Union and in Norway.
If this product is located in Europe and you want to participate in the Thermo Fisher Scientific Business-to-Business
(B2B) Recycling Program, send an email request to weee.recycle@thermofisher.com with the following information:
• WEEE product class
• Name of the manufacturer or distributor (where you purchased the product)
• Number of product pieces, and the estimated total weight and volume
• Pick-up address and contact person (include contact information)
• Appropriate pick-up time
• Declaration of decontamination, stating that all hazardous fluids or material have been removed from the product
For additional information about the Restriction on Hazardous Substances (RoHS) Directive for the European Union,
search for RoHS on the Thermo Fisher Scientific European language websites.
IMPORTANT This recycling program is not for biological hazard products or for products that have been medically
contaminated. You must treat these types of products as biohazard waste and dispose of them in accordance with
your local regulations.

Directive DEEE
2012/19/EU

Thermo Fisher Scientific s'est associé avec une ou plusieurs sociétés de recyclage dans chaque état membre de l’Union
Européenne et ce produit devrait être collecté ou recyclé par celle(s)-ci. Pour davantage d'informations, rendez-vous sur
la page www.thermoscientific.fr/rohs.

WEEE Direktive
2012/19/EU

Thermo Fisher Scientific hat Vereinbarungen mit Verwertungs-/Entsorgungsfirmen in allen EU-Mitgliedsstaaten
getroffen, damit dieses Produkt durch diese Firmen wiederverwertet oder entsorgt werden kann. Weitere Informationen
finden Sie unter www.thermoscientific.de/rohs.

15-Years Warranty for Furnace and Thermal Conductivity Detector
Thermo Fisher Scientific provides a 15-year warranty on the combustion and reduction furnaces (P/N 354 06100)
of the Flash IRMS Elemental Analyzer. The combustion and reduction furnaces are assembled using the highest
quality materials and operationally tested. Each furnace is supplied with a unique serial number to identify it.
If the combustion or reduction furnaces have a manufacturing or material defect during the 15-year warranty
period, from the date of delivery of the system, they will be replaced free of charge by a service engineer.

What is excluded from the Furnaces Warranty?
Damage caused to the furnace as a result of improper use, which is defined as unwarranted maintenance and
changes by the user, are not covered.

Thermal Conductivity Detector Warranty
The Thermal Conductivity Detector (P/N 419 07510) is assembled using the highest quality materials and
operationally tested.
The operator must not touch the Thermal Conductivity Detector and in case of any suspected problems, should
contact a service engineer for diagnosis.
Only in the case that a qualified service engineer determines that the Thermal Conductivity Detector has a
manufacturing or material fault, there is a 15-year warranty covering replacement.

What is excluded from the Thermal Conductivity Detector Warranty?
Damage caused to the Thermal Conductivity Detector as a result of improper use, which is defined as unwarranted
maintenance and changes by the user, are not covered.

Who to contact in case of problems?
In the event of Furnace or Thermal Conductivity Detector issue, please contact the Organic Elemental Analyzer
Product Manager, Guido Giazzi at guido.giazzi@thermofisher.com.

C
Contents
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xix
About Your System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xix
Power Rating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xx
Contacting Us. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xx
Safety Alerts and Important Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xx
Safety Symbols and Signal Words . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xx
Instrument Markings and Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxii
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xxiii
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xxiii
Use of Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xxiii
Safety Cut Off Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xxiii
Gases Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxiv
Precaution for Helium . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxiv
Precaution for Oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxiv
Precaution for Hydrogen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxiv
Precaution for Nitrogen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxiv
Precaution for Sulfur Dioxide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxiv
Precaution for Carbon Dioxide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxiv
Precaution for Carbon Monoxide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxv
Hazardous Substances Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxv
Venting Toxic Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxv
Residual Risks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxv
High Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxv
Hot Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxv
Hazardous Chemical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxvi
Maintenance Precaution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxvi
Personal Protective Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxvi
Training . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xxvii
Chapter 1

Thermo Scientific

Getting Familiar with your EA IsoLink IRMS System for CNSOH . . . . . . . . . . . . . 1
Technical Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Instrument Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Software Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Instrument Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety Cut Off Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Labels Location on the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

EA IsoLink IRMS System for CNSOH Operating Manual

xiii

Contents

Use of Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Gas Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Gas Purity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Maximum Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Nominal Pressure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Precaution for Carbon Monoxide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Precaution for Sulfur Dioxide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Using Hydrogen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Leak Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Back Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Top Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Furnaces Compartment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Oven Compartment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Detection System Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Electrical Compartment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Low Voltage Compartment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Main Voltage Compartment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Connections Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Interface Area. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Power Supply Area. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Gases Supply Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Transformers Compartment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Devices for the Furnaces Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Devices Supplying the Furnaces. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Status Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Autosamplers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
MAS Plus Autosampler for Solid Samples . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
AI 1310 /AS 1310 Autosampler for Liquid Samples . . . . . . . . . . . . . . . . . . . 27
Ramped GC Oven. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Chapter 2

xiv

Analytical Principles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
About Your System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Pneumatic Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Pressure Regulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Electronic Flow Controller (EFC-t) Module . . . . . . . . . . . . . . . . . . . . . . . . . 36
Automatic Switching Valve Block (ASV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Helium Management (HeM) Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Preparation of Reactors and Adsorption Filter . . . . . . . . . . . . . . . . . . . . . . . . . . 44
NC Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
H and O Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Preparing the Reactors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Preparing the Adsorption Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
The Ramped GC Oven . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

Contents

Thermo Scientific

Chapter 3

Installing Flash IRMS Elemental Analyzer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Preliminary Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Who Performs the Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Standard Outfit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Verify Site Preparation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Unpacking the Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Placing the Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Making the Gas Supply Plumbing Connections . . . . . . . . . . . . . . . . . . . . . . . . 53
Building the Gas Lines. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Purging Gas Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Connecting the Gas Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Mounting Peripheral Devices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Performing Electrical Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Installing the Reactors into the Furnaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Setting Pressure and Flow Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
NCS Determination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
H-O Determinations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Installing Autosampler. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Isodat Software Suite . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64

Chapter 4

Installing MAS Plus Autosampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
MAS Plus Autosampler Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
MAS Plus Autosampler Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67

Chapter 5

Installing AI 1310/AS 1310 Autosampler. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Preliminary Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Who Performs the Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Electrical Requirement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Lift and Carry the Sampling Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Installing the Direct Injection Device for Flash IRMS Elemental Analyzer . . . . 72
Installing the Sampler Support on the Flash IRMS Elemental Analyzer . . . . . . . 73
Installing the AI 1310/AS 1310 on the Flash IRMS Elemental Analyzer . . . . . . 76
Installation of the Sampling Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Installing the Syringe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Starting-up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82

Chapter 6

Using EA IsoLink IRMS System for CNSOH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Dual Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Performing a Jump Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Creating an Isodat Method for N+C Measurement (Dual Measurement) . . . . . 88
Creating an EA Method for N+C Measurement . . . . . . . . . . . . . . . . . . . . . . . . 92
Creating an Isodat Method for Single Mode S Measurements . . . . . . . . . . . . . . 93
High-Temperature Conversion – Analysis of H and O Isotopes . . . . . . . . . . . . 94
Bottom Field Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94

EA IsoLink IRMS System for CNSOH Operating Manual

xv

Contents

Setting Up a HO Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Procedure Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Dual Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Creating a New Method for Dual Measurement . . . . . . . . . . . . . . . . . . . . . . 97
Creating an EA Method for H+O Measurement . . . . . . . . . . . . . . . . . . . . . . . 102
Creating a Sequence for Dual Measurement . . . . . . . . . . . . . . . . . . . . . . . . 104
Start Data Acquisition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Dual Measurement Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
H3-Factor Determination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Measuring Sulfur Isotopes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Sulfur Measurement Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Preparing the System for a Sulfur Measurement. . . . . . . . . . . . . . . . . . . . . . 109
Before Starting a Sulfur Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Setting for Flash Elemental Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Create a Gas Configuration for a Sulfur Measurement. . . . . . . . . . . . . . . . . . . 111
Starting a Sulfur Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
EA IsoLink Setting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Defining an Isodat Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Events During Acquisition. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116

xvi

Chapter 7

Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Post-Analysis Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Putting the Instrument in Standby Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Maintaining the Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
Current Maintenance Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
Replacing Reactors and Adsorption Filters . . . . . . . . . . . . . . . . . . . . . . . . . 122
Replacing the Filling Materials. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Ashes Removal and Crucible Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Maintaining the Reactor of the Pyrolysis Unit . . . . . . . . . . . . . . . . . . . . . . . 125
Replacing the Gas Chromatographic Column . . . . . . . . . . . . . . . . . . . . . . . 127
Maintaining the MAS Plus Autosampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Upgrading EA for 25 mm OD Macro Reactor. . . . . . . . . . . . . . . . . . . . . . . . . 133
Cleaning the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135

Chapter 8

Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Analytical Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Safety Cutoff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
EFC-t Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140

Chapter 9

Running the Flash IRMS as Stand-alone Instrument . . . . . . . . . . . . . . . . . . . . . . 141
EagerSmart Software Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
EagerSmart Main Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Configuring the Analyzer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Performing a Leak Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Adjusting the Detector Signal Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
View Sample Being Acquired. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Configuring EagerSmart Data Handling Software . . . . . . . . . . . . . . . . . . . . . . 150

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

Contents

Using Isodat Software Suite and EagerSmart Data Handling Software at the
Same Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

xvii

P
Preface
This manual contains detailed information for the use of the Thermo Scientific™ EA IsoLink™ IRMS
System for CNSOH.
Contents

• About Your System
• Power Rating
• Contacting Us
• Safety Alerts and Important Information
• Instrument Markings and Symbols
• Environmental Conditions
• Safety Information
• Gases Precautions
• Hazardous Substances Precautions
• Residual Risks
• Training

About Your System
Thermo Scientific systems provide high-caliber elemental analysis instrumentation.
IMPORTANT Thermo Scientific systems optimize the separation and detection capabilities of EA
IsoLink IRMS System for CNSOH providing high performance analytical capabilities for both
research, and routine applications. More information about the use of this system can be found in
related documentation sources, and by using the provided contact information.
WARNING Thermo Scientific systems operate safely and reliably under carefully controlled
environmental conditions. If the equipment is used in a manner not specified by the manufacturer,
the protections provided by the equipment might be impaired. If you maintain a system outside the
specifications listed in this guide, failures of many types, including personal injury or death, might
occur. The repair of instrument failures caused by operation in a manner not specified by the
manufacturer is specifically excluded from the standard warranty and service contract coverage.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

xix

Preface

WARNING Operation of this system requires the use of chemical substances with different hazard
specifications. Before using any chemicals, read the hazard indications and information reported in
the Safety Sheet supplied by the manufacturer, referring to the relevant CAS (Chemical Abstract
Service) number.

Power Rating
EA IsoLink IRMS System for CNSOH
• 230 Vac ±10%, 50/60 Hz ±2 Hz, 1400 VA
Detailed instrument specifications are in the Product Specifications Sheet.

Contacting Us
Thermo Fisher Scientific provides comprehensive technical assistance worldwide and is dedicated to
the quality of our customer relationships and services.
Use http://www.thermofisher.com address for products information.
Service contact details are available under: www.unitylabservices.com
For other information please contact your local Thermo Fisher Scientific office or affiliate.

Safety Alerts and Important Information
Make sure you follow the precautionary notices presented in this guide. The safety and other special
notices appear in boxes.
IMPORTANT Highlights information necessary to prevent damage to software, loss of data, or
invalid test results, or it might contain information that is critical for optimal performance of the
system.
Note Emphasizes important information about a task.
Tip Provides information that can make a task easier.

Safety Symbols and Signal Words
All safety symbols are followed by WARNING or CAUTION, which indicates the degree of risk of
personal injury, instrument damage, or both. Cautions and warnings are followed by a descriptor, such
as BURN HAZARD.
A WARNING is intended to prevent improper actions that could cause personal injury. A CAUTION is
intended to prevent improper actions that might cause personal injury and/or instrument damage. You
find the following safety symbols on your instrument or in this guide:

xx

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

Preface

Symbol

Description

BIOHAZARD: Indicates that a biohazard will, could, or might occur.
BURN HAZARD: Alerts you to the presence of a hot surface that could or might cause
burn injuries.
ELECTRICAL SHOCK HAZARD: Indicates that an electrical shock could or might occur.
FIRE HAZARD: Indicates a risk of fire or flammability could or might occur.
GLOVES REQUIRED: Indicates that you must wear gloves when performing a task or
physical injury could or might occur.
MATERIAL AND EYE HAZARD. Indicates you must wear eye protection when
performing a task.
MATERIAL AND EYE HAZARD: Indicates that eye damage could or might occur.
HAND AND CHEMICAL HAZARD: Indicates that chemical damage or physical injury
could or might occur.
HARMFUL. Indicates that the presence of harmful material will, could, or might occur.
INSTRUMENT DAMAGE: Indicates that damage to the instrument or component
might occur. This damage might not be covered under the standard warranty.
LIFTING HAZARD. Indicates that a physical injury could or might occur if two or more
people do not lift an object.
READ MANUAL: Alerts you to carefully read your instrument’s documentation to
ensure your safety and the instrument’s operational ability. Failing to carefully read the
documentation could or might put you at risk for a physical injury.
TOXIC SUBSTANCES HAZARD: Indicates that exposure to a toxic substance could
occur and that exposure could or might cause personal injury or death.
RADIOACTIVE HAZARD. Indicates that the presence of radioactive material could or
might occur.
For the prevention of personal injury, this general warning symbol precedes the
WARNING safety alert word and meets the ISO 3864-2 standard. In the vocabulary of
ANSI Z535 signs, this symbol indicates a possible personal injury hazard exists if the
instrument is improperly used or if unsafe actions occur. This symbol and another
appropriate safety symbol alerts you to an imminent or potential hazard that could
cause personal injury.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

xxi

Preface

Instrument Markings and Symbols
The following table explains the symbols used on Thermo Fisher Scientific instruments. Only a few of
them are used on the EA IsoLink IRMS System for CNSOH IRMS System, which are annotated with
an asterisk below.
Table 1. Instrument Marking and Symbols
Symbol

Description

Direct current
*

Alternating current
Both direct and alternating current
Three-phase alternating current

3

Earth (ground) terminal
Protective conductor terminal
Frame or chassis terminal
Equipotentiality
*

On (supply)

*

Off (supply)
Equipment protected throughout by DOUBLE INSULATION or REINFORCED
INSULATION (Equivalent to Class II of IEC 536)

*

Instruction manual symbol affixed to product. Indicates that the user must refer to
the manual for specific WARNING or CAUTION information to avoid personal
injury or damage to the product.
Caution, risk of electric shock

*

Caution, hot surface

*

Caution, biohazard
In-position of a bistable push control
Out-position of a bistable push control
+

*

xxii

-

Jack socket
Symbol in compliance to the Directive 2012/19/EU on Waste Electrical and
Electronic Equipment (WEEE) placed on the European market after August, 13,
2005.

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

Preface

Environmental Conditions
• For use and operation indoor only.
• Altitude up to 2000 meters.
• Operating temperature range from 15 to 35 °C.
• Maximum relative humidity between 30% and 85%.
• Voltage variations not exceeding ±10% of the nominal value.
• Transients according to installation categories II.
• Degree of pollution according to IEC 664 (3.7.3) 2.

Safety Information
WARNING The instrument must be used according to the specifications of this guide. Improper
use can adversely affect the instrument protection. If the equipment is connected to optional
instruments, such as computer, balance, and so on, the degree of insulation of peripheral devices
should be equivalent or higher (double or reinforced) than that of the EA IsoLink CNSOH IRMS
System. The analyzer operation requires the use of chemical substances having different hazard
specifications.
Before using chemicals, please read the hazard indications and information reported in the Material
Safety Data Sheet supplied by the manufacturer referring to the relevant CAS (Chemical Abstract
Service) number.

Use of Gases
The following gases are used with the instrument:
• Helium (He) as carrier gas.
• Oxygen (O2) as gas for sample oxidation.
• Gases Purity — The EA IsoLink IRMS System for CNSOH uses helium, argon, and oxygen with
99.995% minimum purity.
• Maximum Pressure — The maximum pressures of the gases to supply the Flash IRMS Elemental
Analyzer is 700 kPa (7 bar).
• Nominal Pressure of Gases — The nominal pressure of the gases to supply the Flash IRMS
Elemental Analyzer:
−

Maximum 250-300 kPa (2.5-3 bar) for helium (He).

−

Maximum 250-300 kPa (2.5-3 bar) for oxygen (O2).

Safety Cut Off Device
When an alarm condition is detected, this device cuts off the power to the heating resistors of the
oxidation, reduction furnaces and to the GC Oven. For more details see “Safety Cut Off Device” on
page 5.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

xxiii

Preface

Gases Precautions
WARNING Before using gases, carefully read the hazard indications and information reported in
the Material Safety Data Sheet (MSDS) supplied by the manufacturer referring to the CAS
(Chemical Abstract Service) number. It is the user’s responsibility to ensure compliance with all
local safety regulations for the use of gases.

Precaution for Helium
Helium is a nontoxic, odorless, colorless, nonflammable gas stored in cylinders at high pressure. It can
cause rapid suffocation when concentrations are sufficient to reduce oxygen levels below 19.5%. It is
lighter than air and may collect in high points or along ceilings.

Precaution for Oxygen
Oxygen is an odorless, colorless, nonflammable gas stored in cylinders at high pressure. It is an
oxidizing gas and vigorously accelerates combustion. Keep away from oils or grease. Rescue personnel
should be aware of the extreme fire hazards associated with oxygen-enriched (greater than 23%)
atmospheres.

Precaution for Hydrogen
Hydrogen is a colorless, odorless, highly flammable gas. The use of hydrogen requires the operator’s
strict attention and compliance with special precautions due to the hazards involved. Hydrogen is a
dangerous gas, particularly in an enclosed area when it reaches a concentration corresponding to its
lower explosion level (4% in volume). When mixed with air it can create an explosive mixture.

Precaution for Nitrogen
Liquid nitrogen is a colorless and odorless gas. It can cause rapid suffocation when concentrations are
sufficient to reduce oxygen levels below 19.5%. A Self Contained Breathing Apparatus (SCBA) may be
required. Oxygen concentrations must be monitored in the release area.

Precaution for Sulfur Dioxide
Sulfur dioxide is a colorless gas with an irritating odor. May be fatal if inhaled. Causes severe respiratory
tract, eye and skin burns. Use only with adequate ventilation.

Precaution for Carbon Dioxide
Carbon dioxide is a colorless, cryogenic liquid. At low concentrations, is odorless. At higher
concentrations carbon dioxide will have a sharp, acidic odor. At concentrations between 2 and 10%,
Carbon dioxide can cause nausea, dizziness, headache, mental confusion, increased blood pressure, and
increased respiratory rate. If the gas concentration reaches 10% or more, suffocation and death can
occur within minutes.

xxiv

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

Preface

Precaution for Carbon Monoxide
Carbon monoxide is a colorless gas. May be fatal if inhaled. Causes severe respiratory tract, Use only
with adequate ventilation.

Hazardous Substances Precautions
WARNING Before using hazardous substances (toxic, harmful, and so on), read the hazard
indications and information reported in the applicable Material Safety Data Sheet (MSDS.)
Use personal protection according to the safety requirements.

Venting Toxic Gases
When analyzing toxic compounds be aware that during the normal operation of the EA some of the
sample might be vented outside the instrument through the inlet and detector exits; therefore, make
sure to vent the exhaust gases to a fume hood. Consult local Environmental and Safety Regulations for
instructions in exhausting fumes from your system.

Residual Risks
Users of Flash IRMS Elemental Analyzer must pay attention to the following residual hazards.

High Voltage
WARNING DO NOT OPEN the electrical compartment because there are no user serviceable
parts inside. Any operation inside the compartment must be carried out by authorized and trained
Thermo Fisher Scientific personnel.

Hot Parts
WARNING Do not open the furnace compartment during operation due to the very high
temperatures reached during operation. The protecting plate should only be removed when the
temperature of the furnace is near that of room temperature.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

xxv

Hazardous Chemical
WARNING Samples, consumables, reactors and filters filling materials might contain toxic,
carcinogenic, mutagenic, or corrosive/irritant chemicals. Avoid exposure to potentially harmful
materials. Always wear protective clothing, gloves, mask, and safety glasses when you handle
consumables, reactors and filters filling materials and during the cleaning of the MAS Plus’s piston
and the remotion of the ashes from the crucible when used.
Also contain waste streams and use proper ventilation. Refer to our supplier's Material Safety Data
Sheet (MSDS) for proper handling of a particular compound.
CAUTION Always use original Thermo Fisher Scientific materials and products. The use of
materials not meeting the technical specifications of our products does not ensure a good operation
of the instrument and may even damage it.

Maintenance Precaution
WARNING When, for technical reasons, it is necessary to work on instrument parts which might
involve an hazard (moving parts, components under voltage, and so on), the authorized Technical
Service must be contacted. This type of situations can be identified because access to these parts is
possible only by using a tool. The removable protective covers bear a warning symbol suggesting to
refer to the documentation accompanying the instrument. When a maintenance operation is
performed, the operator must have received proper training to carry out specific actions.
CAUTION When the instrument is switched off, consider that its does not cool down immediately,
but heat tends to concentration in the upper part of the furnaces area. It should be made clear that
it is better to cool down the furnaces first before switching off the instrument. Switching it off
means that the fan in the back does not remove the hot air concentrating at the top and the surface
thus becomes very hot. The openings provided for the chamber aeration will cause a slow cooling
of the same, which however, in the vicinity of the areas marked with the symbol “hot surfaces”,
might even reach temperatures higher than ambient temperature. Therefore in the minutes
immediately following the instrument switching off, the operator must consider this risk and pay
adequate attention during any maintenance operations following the use of the instrument.

Personal Protective Equipment
This manual can only give general suggestions for personal protective equipment (PPE), which protects
the wearer from hazardous substances. Refer to the Material Safety Data Sheets (MSDSs) of the
chemicals handled in your laboratory for advice on specific hazards or additional equipment.
• Eye Protection — The type of eye protection required depends on the hazard. For most situations,
safety glasses with side shields are adequate. Where there is a risk of splashing chemicals, goggles
are required.
• Protective Clothing — When the possibility of chemical contamination exists, protective clothing
that resists physical and chemical hazards should be worn over street clothes. Lab coats are
appropriate for minor chemical splashes and solids contamination, while plastic or rubber aprons
are best for protection from corrosive or irritating liquids.

Preface

• Gloves — For handling chemical compounds, Thermo Fisher Scientific recommends the
following gloves: white nitrile clean room gloves (Fisher Scientific P/N 19-120-2947B [size
medium] or P/N19-120-2947C [size large]; Thermo Scientific P/N23827-0008 [size medium] or
P/N 23827-0009 [size large]). For handling hot objects, gloves made of heat-resistant materials
(for leather) should be available.
• Mask — For handling chemical compounds and filling materials for preparing and cleaning
reactors and filters, and for removing the ashes from the crucible when used.

Training
Thermo Fisher Scientific offers worldwide training on instruments and software. Experience has shown
that maximum results can be obtained from a scientific instrument, if the instrument operator receives
an adequate training.
We recommend that the key operator undertakes training at Thermo Fisher Scientific Rodano - Milan
(Italy), Thermo Fisher Scientific Bremen (Germany), at your site, or at one of the local Thermo Fisher
Scientific offices. For information on training courses and enrollment, please contact your local
Thermo Fisher Scientific office.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

xxvii

1
Getting Familiar with your EA IsoLink IRMS System
for CNSOH
This chapter provides information to familiarize you with your Thermo Scientific™ EA IsoLink™
IRMS System for CNSOH. Here, a detailed description of the instrument’s components is provided.
The EA IsoLink IRMS System includes the Flash IRMS Elemental Analyzer, the Thermo Scientific™
ConFlo IV Universal Interface and a Thermo Scientific Isotope Ratio Mass Spectrometer.
Contents

• Technical Features
• Instrument Configurations
• Software Requirements
• Instrument Basics
• Labels Location on the Instrument
• Use of Gases
• Front Panel
• Back Panel
• Top Panel
• Furnaces Compartment
• Oven Compartment
• Detection System Description
• Electrical Compartment
• Connections Panel
• Transformers Compartment
• Status Panel
• Autosamplers
• Ramped GC Oven

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

1

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Technical Features

Technical Features
Table 1 summarizes the major technical features of the EA IsoLink IRMS System for CNSOH.
Table 1.

Technical Features of the Flash IRMS Elemental Analyzer

Features

Description

Detector

Thermal conductivity detector (TCD)

External interface

RS 232 serial line

Instrument control

Isodat Software Suite 3.0, EagerSmart Data Handling Software

Power supply

230 Vac ±10%, 50/60 Hz ±2 Hz, 1400 VA

Dimensions (cm)

Height 50 (54 with fittings); Width 59, Depth 58

Mass (kg)

65

Sound Pressure Level

Less than 70 db (A)

Instrument Configurations
The Flash IRMS Elemental Analyzer is able to be coupled to an Isotope Ratio Mass Spectrometer
(IRMS) for the following determinations. See Table 2.
Table 2. EA IsoLink IRMS System for CNSOH Analytical Capabilities
Configuration

Description

H

Quantitative (weight%) and stable isotope analysis of hydrogen on H2.

O

Quantitative (weight%) and stable isotope analysis of oxygen on CO.

N

Quantitative (weight%) and stable isotope analysis of nitrogen on N2.

C

Quantitative (weight%) and stable isotope analysis of carbon on CO2.

S

Quantitative (weight%) and stable isotope analysis of sulfur on SO2.

IMPORTANT Oxygen and hydrogen analyses are performed using the left furnace (high
temperature conversion). Nitrogen, carbon and sulfur analyses are performed using the right
furnace for combustion.

Software Requirements
Thermo Scientific Isotope Ratio Mass Spectrometer series require the use of the Isodat Software Suite.
In most cases the installation of Isodat Software Suite is sufficient to operate the EA IsoLink IRMS
System for CNSOH. See “Isodat Software Suite” on page 64.
Table 3 describes the minimum hardware required to run the software.
ATTENTION If you want to operate with the EA IsoLink CNSOH as a stand-alone instrument, you
need to install EagerSmart Data Handling Software on you computer. See Chapter 9, “Running the
Flash IRMS as Stand-alone Instrument.”

2

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

1

Table 3.

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Instrument Basics

Minimum Requirements for Personal Computer

Components

Description

Computer

Any PC can be used, including laptop computer
Operating System: Windows® 2000 / XP* / Vista / 7* / 8
Pentium Processor minimum 256 MHz
Hard drive with at least 1 GB free
One free COM port for instrument control
One free COM port for balance, if required
One free USB port

Monitor

Color 1024 x 768 or better

Printer

Any printer accepted by the operating system

* Required by Isodat 3.0

Instrument Basics
The Flash IRMS Elemental Analyzer is part of the EA IsoLink CNSOH IRMS System.
The Elemental Analyzer comprises:
• Two furnaces — The left furnace, for high temperature conversion analysis, contains a silicon
carbide heater element. The heater element is immersed into a refractory material contained in a
metallic compartment. This furnace is used only for the H and O isotope ratio determinations.
The right furnace contains a quartz candle on which an electric resistance is wound.
The candle is immersed into a refractory material contained in a metallic compartment.
This furnace is used only for the N, C and S isotope ratio determination.
• Furnace Temperature Regulation — The temperature is monitored by a thermocouple of
Pt-Pt/Rh type appropriately located in the furnace.
• Furnace Cooling — The cooling time varies according to the operating temperature setting and
room temperature.
• Thermal conductivity detector (TCD) — Located in a thermostatic chamber at controlled
programmable temperature. This chamber also accommodates the analytical column.
• Chromatographic columns — The chromatographic column performs the chromatographic
separation of the reaction products generated during the combustion or high temperature
conversion process. The chromatographic column for high temperature conversion is placed in the
thermostatic chamber of the TCD detector according to the instrument configuration.
The analytical column for the combustion side (N, C and S analysis) is placed in a separate ramped
GC oven which can be heated and cooled in a fast way.

Thermo Scientific

Description

Qty

Part number

H/O separation column (SS; 1 m; 1/8-in. OD)

1

260 08240

IRMS separation column (SS; 60 cm; 1/8-in. OD)

1

260 08241

IRMS NC separation column (SS; 60 cm; 1/8-in. OD)

1

260 08242

Sulfur separation column (Teflon®; 80 cm; 1/8-in. OD)

1

260 08243

EA IsoLink IRMS System for CNSOH Operating Manual

3

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Instrument Basics

• Adsorption filters — They can be made of glass or Plexiglas® according to the analytical
configuration.
• Reactors — The H and O configurations require the use of a special cylindrical ceramic reactor
filled with special catalyst.
There are two different sizes available:
−

reactor of 18 mm OD

−

reactor of 25 mm OD (Macro Reactor)
Note The EA IsoLink IRMS System for CNSOH comes standardized for a 18 mm OD
reactor. For 25 mm reactor tubes the MAS Plus and the top of the EA must be modified.
See the section “Upgrading EA for 25 mm OD Macro Reactor” on page 133.

 To set the temperature

The following procedure is recommended for heating up the furnace of the combustion reactor when
operating with chromium oxide (Cr2O3) at 1020°C:
1. Increase the temperature from room temperature to 400°C and check the background signals on
the Mass Spectrometer. Check the system for leaks.
2. Increase the temperature from room 400°C to 900°C in steps of 100°C and hold at 900°C for at
least 15 min. This avoids melting the copper within the reactor, which will cause poor
performance.
3. Increase it to the operating temperature of 1020°C..
WARNING Immediate temperature increase to 1020 °C can cause copper melting. Cooling down
in steps can avoid reactor breaking. Alternatively, you can use the HeatUp and CoolDown script in
Isodat. Right-click on the Flash IRMS visualization window in Isodat and choose the appropriate
context menu item. See page 82.
• Protective Cap
IMPORTANT Before the EA IsoLink IRMS System for CNSOH is shipped, the pneumatic circuit is
protected by a cap mounted on the coupling union located on the base of the left furnace
compartment as shown in Figure 1 on page 4. This is to avoid humidity in the chromatographic
column. The cap must be removed prior to the reactor installation.
Figure 1.

Protective Cap

Protection Cap
Coupling Union

4

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Instrument Basics

• Autosampler — Performs the automatic injection of samples into the reactor.
• Pneumatic compartment — Consists of two pressure reducers, two pressure gauges, and of several
lines fitted with an thermo-regulator electronic flow controller (EFC-t), which ensures the
switching between carrier gas and oxygen, and controls the flow values.
−

HeM device — Reduces the consumption of helium.

−

Automatic switching box — Switches automatically from combustion N,C, S configurations
to pyrolysis O, H configuration, and vice-versa.

• Electrical compartment — Comprises the electronic boards for the instrument power supply and
control.
• User interface — The instrument is not provided with independent keyboard and display.
A synoptic on the instrument front allows you to monitor the instrument status.

Safety Information
CAUTION The instrument must be used according to the specifications of this guide. Improper use
can adversely affect the instrument protection. If the equipment is connected to optional
instruments, such as computer, balance, and so on, the degree of insulation of peripheral devices
should be equivalent or higher (double or reinforced) than that of the EA IsoLink IRMS System for
CNSOH. The analyzer operation requires the use of chemical substances having different hazard
specifications. Before using chemicals, please read the hazard indications and information reported
in the Material Safety Data Sheet supplied by the manufacturer referring to the relevant CAS
(Chemical Abstract Service) number.

Safety Cut Off Device
When an alarm condition is detected, this device cuts off the power to the heating resistors of the
furnaces and to the EA Oven.
Note For more details, see the section “Safety Cutoff ” on page 138.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

5

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Labels Location on the Instrument

Labels Location on the Instrument
The following illustrations show the location of the safety labels attached on the instrument.
Figure 2.

EA IsoLink IRMS System for CNSOH: Analyzer Model, Serial Number and Electrical Data

Analyzer Model, Serial Number
and Electrical Data Label

Figure 3.

EA IsoLink IRMS System for CNSOH: Alert Labels

Alert Label

Hot Surface Labels

Alert Label

Alert Label

6

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

1

Figure 4.

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Use of Gases

EA IsoLink IRMS System for CNSOH: Hot Surface Labels

Hot Surface Label

Use of Gases
CAUTION Before using gases, carefully read the hazard indications and information reported in the
Material Safety Data Sheet (MSDS) supplied by the manufacturer referring to the CAS (Chemical
Abstract Service) number.
It is the your responsibility to see that all local safety regulations for the use of gases are obeyed.

Gas Supply
The Flash IRMS Elemental Analyzer uses the following gases:
• Helium (He) as carrier gas and purge gas [CAS number 7440-59-7]
• Oxygen (O2) as gas for sample oxidation [CAS number 7782-44-7]
Note ConFlo II/III/IV and IRMS Delta models (TF Bremen manufacturing) use the following
gases:
• Helium (He) as carrier gas and purge gas [CAS number 7440-59-7]
• Carbon Monoxide (CO) as standard gas [CAS number 630-08-0]
• Hydrogen (H2) as standard gas [CAS number 1333-74-0]
• Nitrogen (N2) as standard gas [CAS number 7727-37-9]
• Carbon Dioxide (CO2) as standard gas [CAS number 124-38-9]
• Sulfur Dioxide (SO2) as standard gas [CAS number 7446-09-5]

Gas Purity
The EA IsoLink IRMS System for CNSOH requires the use of ultra-high purity gas chromatography
grade (99.999%) gases.

Maximum Pressure
The maximum pressure of the Elemental Analyzer gas supplies is 700 kPa (7 bar).

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

7

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Use of Gases

Nominal Pressure
The nominal pressure of the Flash IRMS EA gas supplies are:
• 250-300 kPa (2.5-3 bar) for He
• 250-300 kPa (2.5-3) for O2
Note For the nominal pressure of the gases used by ConFlo II/III/IV and IRMS Delta models
(TF Bremen manufacturing), refer to the relevant manuals.
Note The pressure must be adjusted to about 4 kPa through the reducing valves mounted at the
cylinder.

Precaution for Carbon Monoxide
ATTENTION Carbon Monoxide is toxic!
When working with this gas good ventilation is essential. Otherwise, the gas can be hazardous to
your health! The use of a CO detector with an alarm is strongly recommended.
Install an exhaust tube on top of your ConFlo II/III/IV, as shown in of the example of Figure 5, to
remove the toxic carbon monoxide (CO) from inside the ConFlo II/III/IV out of your working area.
Figure 5.

Exhaust Tube Connection

Working Area

Outside of
Working Area
Fan
Exhaust tube

ConFlo II/III/IV

Precaution for Sulfur Dioxide
ATTENTION Sulfur Dioxide is toxic!
When working with this gas good ventilation is essential. Otherwise, the gas can be hazardous to
your health!

8

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Use of Gases

Sulfur dioxide is s colorless gas with a distinctive pungent odor that will alert the user of a leak in ample
time before irritation can occur. It is somewhat denser then air and will sink to low-lying spaces.
Because of the density and expansion factors, never place sulfur dioxide containers in direct sunlight or
exposed to other direct or indirect heating sources.
Sulfur dioxide is corrosive to many common metals and other substances.

Using Hydrogen
CAUTION The use of hydrogen requires the operator’s strict attention and compliance with special
precautions due to the hazards involved. The use of an H2 detector with an alarm is strongly
recommended.
Hydrogen is a dangerous gas, particularly in an enclosed area when it reaches a concentration
corresponding to its lower explosion level (4% in volume). When mixed with air it can create an
explosive mixture.
Use the following safety precautions when using hydrogen:
• Ensure that the hydrogen cylinder complies with the safety requirements for proper use and
storage. Hydrogen cylinders and delivery systems must comply with local regulations.
• Make sure the gas supply is turned completely off when connecting hydrogen lines.
• Perform a bubble test to ensure that the hydrogen lines are leak-tight before using the instrument as
described in the paragraph “Leak Test.” Avoid spraying any electrical components during the
bubble test.

Leak Test
Before starting the system, a leak check has to be performed as described in the following “Performing
a leak check,” operating sequence.
 Performing a leak check

Figure 6.

Gas Cylinder
Main valve
Manometer

On/Off valve
Reducing valve

Gas cylinder

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

9

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Use of Gases

Referring to Figure 6, proceed as follows:
1. After mounting the reducing valve to the gas cylinder both “on/off-valve” and “reducing-valve”
must be open.
2. Open the main valve for two or three seconds to let the gas purge the whole valve system.
3. Close the on/off-valve, then close the main-valve.
4. Mark manometer positions of the on/off-valve and main-valve, and wait for 10 - 15 min.
5. A leak may be present if the manometer positions have changed.
6. To detect a leak, use soap solution on all valves and connections. Check for bubble formation.
Remove soap solution quickly and carefully after test.

10

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Front Panel

Front Panel
The front panel of the Elemental Analyzer is shown in Figure 7.
Figure 7.

Instrument Front Panel

LED Status Panel
Access to HeM, Isothermal GC Oven and
Absorption Filters, He and O2 pressure regulators
Access to Furnaces

The front panel comprises:
• A furnaces compartment. Also see the section “Furnaces Compartment” on page 13.
• A LED status panel. Also see the section “Status Panel” on page 25.
• Oven compartment including the pressure regulators, pressure gauges, adsorption filters,
isothermal GC oven housing, the TCD detector, the gas chromatographic column, and the HeM
module. See the section “Oven Compartment” on page 16.
Note On the internal wall of the front door you will find holders designed for a paper copy of the
Consumables and Spare Parts Catalog, and tools for maintenance. See Figure 8.
Figure 8.

Right Door Internal View

Consumable and Spare Parts Catalog
Tools

Tool Storage

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

11

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Back Panel

Back Panel
The back panel of the Elemental Analyzer is shown in Figure 9.
Figure 9.

Instrument Back Panel

Autosamplers

Cooling Fan

Connections
Panel

Transformer
Compartment

On the back panel, you will find:
• The cooling fan.
• Interface, gas inlets, and electrical connections. See the section “Connections Panel” on page 21.
• Transformers compartment. Access to the transformer compartment is obtained by removing the
back panel cover. See the section “Transformers Compartment” on page 23.
• Place holders for the ramped GC Oven for EA IsoLink CNSOH.
See Figure 10 and the section “Ramped GC Oven” on page 28.
Figure 10. Instrument Back Panel with the Ramped GC Oven

Ramped
GC Oven

12

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Top Panel

Top Panel
The top panel of the Elemental Analyzer is shown in Figure 11.
Figure 11. Instrument Top Panel
Gas Connections

Fittings for Reactors
Autosamplers

The top panel comprises:
• Fittings for mounting and securing reactors and autosamplers.
• Fittings for Helium carrier gas and ConFlo™ Universial Interface™ connection.

Furnaces Compartment
The furnaces compartment can be accessed from the instrument front by removing (lifting) the cover.
See Figure 12.
Figure 12. Furnace Compartment with Protection Plate

Furnace Compartment Cover

Furnace compartment protection plate

WARNING Do not open the furnace compartment during operation due to the very high
temperatures reached during operation. The protecting plate should only be removed when the
temperature of the furnace is near that of room temperature.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

13

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Furnaces Compartment

The furnaces are accessible by removing the protecting plate. See Figure 13.
Figure 13. Furnaces Compartment

Protection Plate

Left Furnace for high temperature conversion analysis (OH)
Right Furnace for combustion analysis (NCS)

In the software, the furnaces are limited to the following maximum temperatures to avoid reduced
lifetime and damage:
• LEFT Furnace: 1450 °C
• RIGHT Furnace: 1100 °C
The furnace temperature is monitored by a thermocouple located inside the furnace. The furnaces are
cooled when required by the operator or if there is a safety cut-off. The cooling time depends on the
operating temperature and the room temperature.

14

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Furnaces Compartment

 Instruction for operation of the furnace

The furnace temperature is monitored by a thermocouple located inside the furnace. The furnaces are
cooled when required by the operator or if there is a safety cut-off. The cooling time depends on the
operating temperature and the room temperature.
1. The furnaces should not be operated above the maximum temperature stated in the operating
instructions, which is 1100 °C and is limited by the software.
2. When the Elemental Analyzer is not being using for sample analysis, either for periods of several
hours (e.g. overnight) or periods lasting several days or longer, the furnaces should be placed into
Stand-By Mode as defined by the Standy-By Mode which can be activated automatically at the end
of a sample sequence.
a. If the instrument is not used for a long period of time (e.g. several weeks) it is recommended
to switch off the furnaces. The furnaces should be cooled down using the furnace Cool Down
procedure in the Isodat Software Suite.
b. After a Stand-By period or period where the furnaces are switched off, the furnaces should be
brought back to operating temperature following the furnace heat-up procedure defined in the
manual.
3. The operator must not cause a short-circuit with the furnace heating element or it’s connections,
or the furnace thermocouple. Only a qualified service engineer should check or test this. Operator
interference on these parts is not covered by the warranty.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

15

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Oven Compartment

Oven Compartment
The oven compartment is located behind the right front door of the instrument and can be accessed by
opening the front door. Figure 14 shows the inside of the oven compartment.
Figure 14. Oven Compartment Internal View

He

Pressure
Regulators
O2

Pressure
Gauges

Fixing Screws
Needle Valves for
Split Adjustment
Securing Clips for
Adsorption Filters

Protection Plate

Fixing Screws
HeM Device

The oven compartment houses the helium (He) and oxygen (O2) pressure regulators and pressure
gauges, and the thermostatic chamber containing the thermal conductivity detector (TCD) and the gas
chromatographic column located behind the protecting plate. See “To access to the thermostatic
chamber” on page 17. The adsorption filters are housed in this compartment and are attached to the
protecting plate using securing clips.
Note Two adsorption filters are required as shown in Figure 15. One filter is required for the High
Temperature Conversion analytical channel and one filter for the combustion analytical channel.
Figure 15. Adsorption Filters Installed

Adsorption Filters

16

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Oven Compartment

 To access to the thermostatic chamber

1. Open the right side door, which can be moved 180°.
2. To access the TCD and GC Columns, first remove the adsorption filters from the fastening clips.
3. Then, remove the four fixing screws on the protecting plate. See Figure 16.
Figure 16. Access to Thermostatic Chamber

Fixing Screws

Protection Plate

Fixing Screws

Figure 17 shows the detector compartment, the heating block surrounding the thermal
conductivity detector (TCD), and the gas chromatographic column.
Note One chromatographic columns is installed for OH analysis with the high temperature
furnace. A second chromatographic column may be installed for combustion if no ramped GC
oven is used.
Figure 17. Thermostatic Chamber Internal View

Heating Block
TCD Detector

Chromatographic Columns

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

17

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Detection System Description

Detection System Description
It consists of a thermal conductivity detector (TCD) sensitive to any substance with thermal
conductivity other than that of the carrier gas used.
The detector consists of a stainless steel block provided with two pairs of filaments (generally of
tungsten/rhenium) having the same electrical resistance. The detector is housed in a thermally
insulated metal block (detector oven) and maintained at constant temperature.
The two pairs of filaments are electrically connected according to a Wheatstone bridge circuit powered
at constant voltage. The first pair of filaments is fed with pure carrier gas (reference channel), whereas
the second pair is fed with the gas flowing from the reactor (analytical channel). When the bridge is
powered, the filaments heat at a temperature (resistance) that is a function of the thermal conductivity
of the gas feeding the filaments. The reference channel is exposed only to pure carrier gas, whereas the
analytical channel is exposed to the reactor effluents (carrier gas + sample).
When pure carrier gas flows through both the reference and the analytical channels, a constant
temperature gradient is established between the elements and the detector walls, and the Wheatstone
bridge is balanced, namely there is no output signal. As a component is eluted, a change in heat transfer
occurs, with consequent variation of the filaments temperature. Since electrical resistance is a function
of temperature, the bridge unbalances and the detector generates a signal proportional to the difference
in thermal conductivity between the eluted component and the carrier gas. The output signal is then
sent to the data acquisition board.
Note The filaments are powered constantly at 5 V and are electrically protected if their
temperature exceeds 220 °C (Safety Cut Off ).

Electrical Compartment
It is located on the right part of the instrument, and it is accessible by removing the right side cover.
Behind the electrical compartment, there is the Connections Panel. For more details, see the section
“Connections Panel” on page 21.
WARNING DO NOT OPEN the electrical compartment because there are no user serviceable
parts inside. Any operation inside the compartment must be carried out by authorized and trained
Thermo Fisher Scientific personnel.

The electrical compartment, shown in Figure 18, comprises:
• Low voltage compartment
• High voltage compartment
• EFC electronic flow controller for gas regulation

18

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Electrical Compartment

Figure 18. Electrical Compartment Internal View

Low Voltage Compartment
It contains the electronic boards to operate and control the instrument. These boards are interlocked
through a mother board.
Table 4 reports the function of each electronic board present in the low voltage section:
Table 4.

Thermo Scientific

Description of the Function of the Electronic Boards (Sheet 1 of 2)

Board

Function

MB 1112

Mother board. It provides interlocking between low voltage boards and with
the rest of the instrument. This board can be connected to a NiCd 3,6 V; 280
mA/h rechargeable battery located nearby. The rechargeable battery
replacement must be performed by specialized technical personnel

CPU 1112

This board has full control of the instrument operation.
It controls the communication between operator and machine through
EagerSmart Data Handling Software. Actuates the Safety Cut Off device,
which puts the instrument in safe conditions, when an alarm condition occurs.

HWD 1112

Provides power supply to the TCD detector filaments. Allows the detector
oven thermo-regulation and also amplifies and converts the detector signal to
send it to the PC.

EA IsoLink IRMS System for CNSOH Operating Manual

19

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Electrical Compartment

Table 4.

Description of the Function of the Electronic Boards (Sheet 2 of 2)

Board

Function

TCR 1112

Operates and controls the furnaces thermoregulation.

PWR 1112

Receives voltage supplies from the TRF 1112 transformers board.
Generates voltage supply for the electronic control circuits.

FP 1112

Synoptic panel

Main Voltage Compartment
It contains the mains power circuits and the Safety Cut Off device.
Table 5 details the function of each component present in the mains voltage section.
Table 5.

Description of the components of the Main voltage compartment

Component

Description

TRF 1112 Transformers
Board

Receives the mains power and supplies it to the following devices:
• Cooling fan
• Furnaces transformers
• Heater of the detector thermostatic chamber
Six fuses are provided on the board. See Table 6.

AC 1112 Furnaces
Power Supply

Supplies 48 Vac power to the furnaces. It contains the SSR relays for the
furnaces control. Also see the section “Devices for the Furnaces Control”
on page 25. Two fuses are provided on the board. See Table 6.

Breaker

Instrument On/Off main switch.

Table 6.
Board

Fuse

Type

Protection

TRF 1112

F1

F1A; IEC 127/I (5 x 20 mm)

Power supply to LEFT and RIGHT
furnaces transformers

F2

F0.315A; IEC 127/I (5 x 20 mm)

Fan

F3

F1.6A; IEC 127/I (5 x 20 mm)

Main power (Breaker)

F4

F1A; IEC 127/I (5 x 20 mm)

LEFT and RIGHT furnaces
transformers

F5

F0.315A; IEC 127/I (5 x 20 mm)

Fan

F6

F1,6A; IEC 127/I (5 x 20 mm)

Mains power (Breaker)

F1

FF12 A; IEC 269 (1.3 x 38 mm)

LEFT Furnace power circuit

F2

FF12 A; IEC 269(1.3 x 38 mm)

RIGHT Furnace power circuit

AC 1112

20

Fuses of the High voltage compartment

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Connections Panel

Connections Panel
The connections panel is shown in Figure 19.
Figure 19. View of the Connections Panel

Interface

Power Supply

Gas Inlets

The connections panel is subdivided into the following connecting areas: interface, power supply, and
gases supply.

Interface Area
The interface area comprises the connectors for connecting the autosamplers and the helium
management device, and the automatic switching valve. See Figure 20.
Figure 20. Interface Area

• A 9-pin connector marked RS 232 to dialog with the computer via serial line.
• A 25-pin connector marked Aux Connector for the remote start of the instrument.
• A 2-pin connector marked Autosampler HO for the MAS Plus autosampler installed on the left
channel for HO determinations.
• A 2-pin connector marked Autosampler NCS for the MAS Plus autosampler installed on the
Right channel for NCS determinations.
• A 9-pin connector marked Trigger Input for the control of the automatic switching valve.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

21

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Connections Panel

• A 15-pin connector marked HeM for the control of helium management device.

Power Supply Area
The power supply area comprises the following components. See Figure 21.
Figure 21. Power Supply Are

• Breaker marked Mains to power the instrument ON/OFF.
−

Position I = instrument powered ON; — Position O = instrument powered OFF.

• 230 V; 50/60 Hz mains connector.

Gases Supply Area
The gas supply area comprises the gas inlet ports. See Figure 22.
Figure 22. Gases Supply Area

Helium Inlet Port
Oxygen Inlet Port

• The helium and oxygen gas inlet ports, marked He and O2 are directly connected to the pressure
regulators. Table 7 details the pressure value to be set for each gas inlet port.
Table 7.

Gas Inlet Ports and Pressure Setting

Port

Description

Set TO:

He

Inlet port for helium or argon

250 kPa (2.5 bar, 36 psig)

O2

Inlet port for oxygen

250-300 kPa (2.5-3 bar, 36-44 psig)

Gas pressures must be set and controlled through the pressure regulators and the pressure gauges of the
instrument. Table 8 provides indications on the most currently used units of pressure.

22

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

1

Table 8.

Pressure Units Conversion

To convert

Into

Multiply By:

kPa

bar

0.01

psi

0.145

kPa

100

psi

14.51

kPa

6.89476

bar

0.0689476

bar

psig

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Transformers Compartment

Transformers Compartment
Located in the right bottom part of the instrument, the compartment of the transformer is accessible
firstly by removing the back panel of the instrument, then by removing the relevant protection plate.
See Figure 23.
Figure 23. Accessing the Transformer Compartment

Back Panel

Thermo Scientific

Transformer Compartment Protection Plate

EA IsoLink IRMS System for CNSOH Operating Manual

23

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Transformers Compartment

The compartment contains the electrical devices for powering the furnaces and controlling their
temperature.
WARNING DO NOT OPEN the transformers compartment, there are no user serviceable parts
inside. Any operation inside the compartment must be carried out only by authorized and trained
Thermo Fisher Scientific personnel.
Figure 24 shows the devices contained in the compartment.
Figure 24. Transformer Compartment Internal View

PWR 1112 Board
Transformers

LTA-2

24

LTA-1 (HT)

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Status Panel

Devices for the Furnaces Control
Table 9 describes the function of each device:
Table 9.

Description of the Devices Controlling the Furnaces

Device

Function

LTA-1 LEFT
LTA-1 RIGHT

They read the values of the thermocouple present in the relevant furnace and
send the signals to the TCR 1112 board.

SSR LEFT
SSR RIGHT

Solid State Relays (SSR) contained in the AC 1112 board.
Each SSR is coupled with a proper safety sensor, which detects any malfunction.
The SSR control the power supply to the relevant furnace and cut off power to
the heating resistor when the thermocouple detects temperature values
exceeding the setpoint.

Devices Supplying the Furnaces
Table 10 details the function of each device:
Table 10. Description of the Devices Supplying the Furnaces
Device

Function

T1 Transformer

Supplies 48 V voltage to the right furnace resistor. It is provided with a safety
thermal protection, which cuts off power in case of overheating.

T2 Transformer

Supplies 48 V voltage to the left furnace resistor. It is provided with a safety
thermal protection, which cuts off power in case of overheating.

Status Panel
The LED status panel shows the instrument operating conditions, and it is located on the right side of
the instrument front panel. See Figure 25.
Figure 25. LED Status Panel

Each LED lights up when the relevant function is active. Table 11 illustrates the meaning of each
function:

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

25

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Autosamplers

Table 11. Status LED Description
LED

Meaning

Power On

When lit, the instrument is powered on.

Ready

When lit, the instrument is ready to run analyses.

Run

When lit, an analysis is in progress.

Standby

When lit, the instrument is in standby condition.
During this condition, helium gas flows are decreased to 10 mL/min, and
the furnace temperatures reduced to 50% of the set value.

Wake Up

When lit, the instrument has been programmed for a timed automatic
startup (Ready Condition).

Safety Cutoff

When lit, the instrument is in safety shut-off. Gas flows are stopped,
furnaces are switched off and TCD is switched off.

Autosamplers
The EA IsoLink IRMS System for CNSOH can be configured with the following autosamplers: MAS
Plus for solid samples, and AI 1310/AS 1310 for liquid samples.

MAS Plus Autosampler for Solid Samples
It is the standard autosampler for solid samples mounted directly on the connecting fitting of the
concerned channel provided with the proper reactor tube. See Figure 26.
Figure 26. MAS Plus Autosampler for Solid Samples

26

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Autosamplers

Its modular structure allows to run up to 125 unattended analyses. The base unit is provided with one
32-position sample tray. It can accommodate three additional 32-position trays to reach a capacity of
125 samples. Each sample tray is installed in a specific position defined by the numbering. Therefore,
they are not interchangeable. The sample numbering is detailed in Table 12.
Table 12. Sample Tray Numbering
Sample Tray

Locating Mark

Numbering

#1

Seat marked 1 (one)

from 1 to 32

#2

Seat marked 0 (zero) from 33 to 63

#3

Seat marked 0 (zero) from 64 to 94

#4

Seat marked 0 (zero) from 95 to 125

Note For installing the MAS Plus autosampler on your EA IsoLink IRMS System for CNSOH see
Chapter 4, “Installing MAS Plus Autosampler.”

AI 1310 /AS 1310 Autosampler for Liquid Samples
These are optional autosamplers for the analysis of liquid samples. It is mounted on the Flash IRMS
EA by means of the appropriate support. See Figure 27.
Figure 27. AI 1310 and AS 1310 Autosamplers for Liquid Samples

AI 1310

AS 1310

The autosampler consists of:
• A sampling unit
• An 8-position (AI 1310) or 105-position (AS 1310) sample tray
Note For installing the AI 1310/AS 1310 autosamplers for liquids on your EA IsoLink IRMS
System for CNSOH see Chapter 5, “Installing AI 1310/AS 1310 Autosampler.”
For operation with the AI 1310/AS 1310 autosampler refer to the AI 1310/AS 1310 for FLASH
Elemental Analyzers User Guide.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

27

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Ramped GC Oven

Ramped GC Oven
The Ramped GC Module for the temperature control of the chromatographic column.
Figure 28 shows the Ramped GC Oven installed on the back of the Flash IRMS EA.
Figure 28. Ramped GC Oven

MAS Plus on the
Left Channel for OH
Determinations

MAS Plus on the
Right Channel for NCS
Determinations

Ramped
GC Oven

Left Side View

Right Side View

28

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Ramped GC Oven

Figure 29 shows the components of the Ramped GC Oven.
Figure 29. Components of the Ramped GC Oven
4

6

5
3
2
1

7

The Ramped GC Oven includes the following components. See the numbering in the images in
Figure 29.
1. AC Input Connector
2. Power Switch to power the Ramped GC Oven On/Off.
−

Position I = instrument On

−

Position O = instrument Off

3. 9-pin connector for the control of the GC.
4. Compartment for the Thermo Scientific™ smartEA™ option.
5. Jumo controller for the manual temperature program setting.

ATTENTION The start and the stop of the temperature program is performed through Isodat
Software Suite by clicking GC Start/Stop function in the method.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

29

1

Getting Familiar with your EA IsoLink IRMS System for CNSOH
Ramped GC Oven

6. Gas inlet/outlet ports for the connection of the carrier gas.
7. Cooling fan.
CAUTION For installing, connecting and operating the Ramped GC Oven refer to the Ramped GC
Oven Operating Manual.

30

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

2
Analytical Principles
This chapter describes the analytical techniques with the correlated pneumatic circuits used for the
configurations of the EA IsoLink IRMS System for CNSOH.
Contents

• About Your System
• Pneumatic Circuit
• Preparation of Reactors and Adsorption Filter
• The Ramped GC Oven

About Your System
The Elemental Analyzer is equipped with an automatic switch valve (ASV) between the two modes of
high temperature conversion (H and O isotope analysis), and combustion (N, C, and/or S isotope
analysis).
The switch also involves the transfer of the electronic autosampler start signal to either side.
The EA IsoLink IRMS System for CNSOH is configured with an optimized reactor and separation
column for triple analysis of N, C and, S isotopes, and Sulfinert® capillary for minimized water
adsorption, and reduced nitrogen background.
The NCS separation column is operated in a ramped GC oven attached to the back panel of the Flash
IRMS EA. The molecular sieve for high temperature conversion analysis is installed in the isothermal
GC oven of the EA.
Two autosamplers on top and full software control automate the analysis of five isotopes with two
sample drops.
The additional bottom feed connector for the pyrolysis side extends the sample throughput to more
than 400 solid samples without maintenance.
This section describes the new features and software setup of the EA IsoLink IRMS System for
CNSOH, and shall be seen in addition to the existing:
• ConFlo IV Operating Manual Rev. B
It is recommended to read the above manuals in addition to this guide.
The following analytical setups are possible:

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

31

2

Analytical Principles
Pneumatic Circuit

• H and O isotope analysis in dual or single mode.
−

Using the high temperature conversion technique (also referred to as pyrolysis) with glassy
carbon reactor.
See “High-Temperature Conversion – Analysis of H and O Isotopes” on page 94.
See Setting Up a HO Method.

• N, C, and S isotope analysis (triple analysis):
−

Using a dedicated single reactor for triple NCS analysis, and an optimized separation column
and temperature program.
Please refer to the Ramped GC Oven Operating Manual.

• N and C isotope analysis in dual or single mode.
−

Using a single reactor setup with chemical trapping of SO2.
See “Creating an Isodat Method for N+C Measurement (Dual Measurement)” on page 88.

−

Installing an additional CO2 trap for N isotope analysis in single mode reduces the analytical
time.

• S isotope analysis in single mode.
−

If no separation of N2 and CO2 is needed, a dedicated single reactor for triple NCS analysis
and an optimized separation column and temperature program can be used. This reduces the
analytical time.
See “Creating an Isodat Method for Single Mode S Measurements” on page 93.

The EA IsoLink IRMS System for CNSOH configuration and its features like triple analysis are
available from Isodat Software Suite version 3.0 on with Service Pack 0.94 or higher.
For recent service packs please contact your local Thermo Fisher Scientific service office.

Pneumatic Circuit
This section describes the pneumatic circuits of the EA IsoLink IRMS System for CNSOH in the
combustion NCS configuration and high temperature mode OH configuration. See Figure 30 and
Figure 31.

IMPORTANT All the pneumatic diagrams are visualized in the Pre-analysis stage.

32

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

2

(2)

(1)

1

2

(4)

S1

PRV2

2

2

EV2

1

1

EV3

1

EV4

1

3

(7)

F

EV1

6.5 mL/min.
300 kPa

PI2

2

(6)

F

F

O2

EV4

(4)

(1)

(2)

Carrier Gas

EFC - T

(5)

1

S2

Reference Gas

(5)

2

EVP2

F

He

(9)

HO

Purge NCS

Purge HO

MS

NCS

2

1

3
2

1

EV2
3

EV1

2

1

3
2

1

AUTOMATIC SWITCHING BLOCK

EV3

3

(3)

(8)

(6)

(7)

COMBUSTION NCS

(4)

(1)

(2)

(5)

AUTOMATIC
SWITCHING BLOCK

Purge HO
Purge NCS
(3)

MS

NCS

HO

(9)

(8)

(6)

(7)

ConFlo

2

1

IN

Left Furnace 0 ..1450 °C

O2

BFC

AS

H2O/CO2TRAP

EA IsoLink CNSOH (NCS Mode)

R1
OUT

1
2

OUT

2

1

NV1
IN

3

2

1

HeM

EV1

Right Furnace 0 ..1100 °C

PI1

AS

H O TRAP
2

PRV1

2

R2

2

1

1

3

3

NV3

EV2

EV3

IN

OUT

Column

TCD

Ramped GC Oven

NV2

Column

EVP1

IN

Thermo Scientific

3

3

Purge Gas

Carrier Gas

OUT

He

4

2
Analytical Principles
Pneumatic Circuit

Figure 30. EA IsoLink IRMS System for CNSOH in Combustion NCS Analysis Pneumatic Diagram

EA IsoLink IRMS System for CNSOH Operating Manual

33

2

(2)

O2

(1)

1

2

(4)

S1

PRV2

2

2

EV2

1

1

EV3

1

EV4

1

3

(7)

F

EV1

6.5 mL/min.
300 kPa

PI2

2

(6)

F

F

O2

EV4

(4)

(1)

(2)

Carrier Gas

EFC - T

(5)

1

S2

Reference Gas

(5)

2

EVP2

F

He

HIGH TEMPERATURE MODE
PYROLYSIS, O, H

Purge NCS

Purge HO

2

1

3
2

1

EV2
3

EV1

2

1

3
2

1

AUTOMATIC SWITCHING BLOCK

EV3

3

(3)

(9)

MS

(4)

NCS

(6)
(8)

HO

(2)

(7)

(1)

(5)

AUTOMATIC
SWITCHING BLOCK

(3)

(9)

(8)

(6)

(7)

ConFlo

Purge NCS

Purge HO

MS

NCS

HO

EA IsoLink CNSOH (OH Mode)

Cap

2

1

IN

Left Furnace 0 ..1450 °C

PI1

BFC

AS

R1
OUT

1

H2O/CO2TRAP

PRV1

2

OUT

2

1

IN

NV1

3

2

1

Right Furnace 0 ..1100 °C

AS

HeM

EV1

H2O TRAP

EA IsoLink IRMS System for CNSOH Operating Manual
2

R2

2

1

1

3

3

NV3

EV2

EV3

IN

OUT

Column

TCD

Ramped GC Oven

NV2

Column

EVP1

IN

34

3

3

Purge Gas

Carrier Gas

OUT

He

4

2

Analytical Principles
Pneumatic Circuit

Figure 31. EA IsoLink IRMS System for CNSOH in Pyrolysis OH Analysis Pneumatic Diagram

Thermo Scientific

2

Analytical Principles
Pneumatic Circuit

The external pneumatic connection is shown in Figure 32.
Figure 32. External Pneumatic Connection for NCS and OH Analysis
OH

NCS
NCS Configuration uses the right
channel.

To ConFlo

REFERENCE

1
2

1

3 (Purge Autosampler NCS)
4 CARRIER

1

(Purge Autosampler H/O)

Note: The reference flow enters the
pyrolysis reactor from the bottom
before it ends as purge in the
Autosampler NCS.

(Carrier Gas NCS)

2

2
Cap

OH

NCS
OH Configuration uses the
left channel.

To ConFlo

1
2

1

1

(Purge Autosampler H/O)

3
4 REFERENCE
(Purge Autosampler NCS)

2

2
Cap

Note: The carrier gas flow is not
shown here. It enters the pyrolysis
reactor from the bottom.
The reference flow enters the system
at the combustion reactor and ends
as purge for the Autosampler H/O.

Pressure Regulators
The pressure regulators allow the manual adjustment of the carrier gas (helium or argon) and oxygen
inlet pressures. The regulators are located in the detector compartment and they are schematically
shown in Figure 33.
Figure 33. Pressure Regulators
He

PRV1
PI1

PRV2

O2

PI2

The pressure regulators are common in all the configurations of the EA IsoLink IRMS System for
CNSOH. The regulators consist of the following components. See Table 13.
Table 13. Pressure Regulators Pneumatic Diagram (Sheet 1 of 2)

Thermo Scientific

Component

Description and function

He

Helium inlet port

O2

Oxygen inlet port

PRV1

Helium pressure regulator

PI1

Helium pressure gauge

EA IsoLink IRMS System for CNSOH Operating Manual

35

2

Analytical Principles
Pneumatic Circuit

Table 13. Pressure Regulators Pneumatic Diagram (Sheet 2 of 2)
Component

Description and function

PRV2

Oxygen pressure regulator

PI2

Oxygen pressure gauge

Electronic Flow Controller (EFC-t) Module
The EFC-t module is common in all the configurations of the EA IsoLink IRMS System for CNSOH.
The module is schematically shown in Figure 34, and consists of the following components.
See Table 14.
Figure 34. Thermo-regulated EFC-t Module Pneumatic Diagram
S2
EVP2

He

1

F

2

(5)

O2

F

F
EV1

(6)

2

F
3

(7)

1

6.5 mL/min.
300 kPa

EV4
3

1

EV3

1

1
2

2

2

EV2

(4)

EFC-t

S1
EVP1

1

2

(1)

ASV

(2)

EFC - T

Table 14. Parts of the EFC-t Module (Sheet 1 of 2)

36

Component

Description and function

He

Helium inlet port

O2

Oxygen inlet port

F

Stainless steel filter

EV1

Two-way solenoid valve to control oxygen inlet.

EV2

Three-way solenoid valve to control helium inlet and to allow switching between helium and
oxygen.

EV3

Two-way solenoid valve, normally open, to control the inlet of helium flowing back from the TCD
detector analytical channel. The valve is closed during the leak test.

EV4

Two-way solenoid valve, normally open, to control the inlet of helium flowing back from the TCD
detector reference channel. The gas leads to port 1 of the MAS Plus autosampler and then to the
ConFlo..

S1

Electronic flow sensor for helium as carrier gas and oxygen during the sampling stage.
It cooperates with the EVP1 electronic controller (proportional valve).

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

2

Analytical Principles
Pneumatic Circuit

Table 14. Parts of the EFC-t Module (Sheet 2 of 2)
Component

Description and function

S2

Electronic flow sensor for helium as reference gas. It cooperates with the EVP2 electronic
controller (proportional valve).

EVP1

Electronic flow controller for helium as carrier gas and oxygen to control the flow rates of gases
according to the flow values set.

EVP2

Electronic flow controller for helium as reference gas to control the flow rate according to the
required flow value.

Automatic Switching Valve Block (ASV)
Allows the automatic switching from the combustion N,C, S configurations to the pyrolysis O, H
configuration, and vice-versa. The module is schematically shown in Figure 35, and consists of the
following components. See Table 15.
Figure 35. Automatic Switching Valve Bock Pneumatic Diagram
(5)
Reference Gas

1

EV1
(2)

2

Carrier Gas

(7)

HO

(6)

NCS

1

EV2
(1)

3

2

(8)

MS

(9)

Purge HO

(3)

Purge NCS

1

EV3
(4)

3

2

1

EV4
3

2

AUTOMATIC
SWITCHING BLOCK

AUTOMATIC
SWITCHING BLOCK
(5)

(7)

HO

(5)

(7)

HO

(2)

(6)

NCS

(2)

(6)

NCS

(1)

(8)

MS

(1)

(8)

MS

(4)

(9)

Purge HO

(4)

(9)

Purge HO

(3)

(3)

Purge NCS

Purge NCS

HIGH TEMPERATURE MODE
PYROLYSIS, O, H

COMBUSTION NCS

ASV

EFC-t

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

37

2

Analytical Principles
Pneumatic Circuit

S2
EVP2

He

PRV1

He

F

O2

F

1

2

PI1
(5)
PRV2

O2

F
EV1

PI2

(6)

2

F
3

(7)

1

6.5 mL/min.
300 kPa

EV4
3

1

EV3

1

1
2

2

2

EV2

(4)

S1

COMBUSTION NCS
(1)

1

(2)

2

Carrier Gas

EFC - T

Reference Gas

EVP1

AUTOMATIC
SWITCHING BLOCK
(5)

(7)

HO

(2)

(6)

NCS

(1)

(8)

MS

(4)

(9)
(3)

Purge HO
Purge NCS

S2
EVP2

He

PRV1

He

F

O2

F

1

2

PI1
(5)

EV1

PI2

(6)

2

F
3

6.5 mL/min.
300 kPa

EV4
3

1

EV3

2

S1
EVP1

1

1
2

EV2

(7)

1

1

EFC - T

2

2

HIGH TEMPERATURE MODE
PYROLYSIS, O, H

(4)
(1)

(2)

Reference Gas

O2

F

Carrier Gas

PRV2

AUTOMATIC
SWITCHING BLOCK
(5)

(7)

HO

(2)

(6)

NCS

(1)

(8)

MS

(4)

(9)
(3)

Purge HO
Purge NCS

Table 15. Parts of the Automatic Switching Block

38

Component

Description and function

EV1

In combustion NCS analysis controls the reference gas flow through the right channel.
In pyrolysis O,H analysis controls the carrier gas flow through the right channel.

EV2

In combustion NCS analysis controls the carrier gas flow through the right channel.
In pyrolysis O,H analysis controls the reference gas flow through the right channel.

EV3

In combustion NCS analysis controls the carrier gas flow to the ConFlow.
In pyrolysis O,H analysis controls the reference gas flow to the ConFlow.

EV4

In combustion NCS analysis controls the purge gas flow to the right autosampler.
In pyrolysis O,H analysis controls the purge gas flow to the left autosampler.

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

2

Analytical Principles
Pneumatic Circuit

Helium Management (HeM) Module
This module allows the reduction of helium consumption for each sample. The module is shown in
Figure 36, and consists of the following components. See Table 16.
Figure 36. Helium Management Valve Block (Left) and Needle Valve Holder for Split Adjustment (Right)

Table 16. Parts of the HeM Module
Component

Description and Function

EV1 - NV1

Pyrolysis — opens and closes the split and controls the split flow

EV2 - NV2

Combustion — opens and closes the split and controls the split flow

EV3 - NV3

Vent — opens and closes the split and controls the split flow for large sample analysis

Helium Management (HeM) Valve Block Description
The Flash IRMS Elemental Analyzer is equipped as standard with the Helium Management (HeM)
module. The HeM module reduces the He consumption in an EA significantly, e.g. up to >60% for a
CNS triple analysis from a single sample. Large volumes like the reactors require high flows of
>80mL/min. Since peak broadening starts in the reactors higher flows are favorable to enhance peak
shape. HeM module allows high reactor flows of 180 mL/min. This increase in carrier flow is
compensated by recirculating up to 2/3 of the He flow back into the system for purging. HeM module
works through a valve block which is installed underneath the isothermal GC oven of the EA). See
Figure 37.
Figure 37. HeM Valve Block Installed

Needle Valves

HeM Valve Block

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

39

2

Analytical Principles
Pneumatic Circuit

The valves are controlled by Isodat Software Suite. The split ratios are set by needle valves which are
located above the isothermal GC oven. The needle valves are preset.
Note Any changes may have influence on the system performance and should be done only when
necessary and after reading this manual.

HeM Working Principle
When working with HeM module the analysis is separated into two phases:
1. Analysis mode
2. He save mode
HeM is switched on as default. It can be switched off if no He savings are desired.
• Combustion — Before an analysis Isodat uploads an EA method to the EA. This method includes
the settings of carrier and reference flow (=purge flow). It is thus obligatory that an EA method is
set in the sequence of Isodat to ensure proper functionality of the HeM module. With the upload
of the EA method the flows are set.

During initialization of the system (magnet settings, peak center, reference gas flow equilibration
in the ConFlo) the reactor flow (carrier gas) is stabilized.
When the sample drops (or is injected by liquids autosampler) the carrier gas flow is maintained
for a fixed period until all analyte gases produced upon conversion of the sample in the reactor are
certainly on the chromatographic column (analysis mode). Split valve 2 (EV2) is open at that time.
In general, the flow through the chromatographic column is 50 mL/min if the carrier flow is set to
180 ml/min. The split flow is thus 130 mL/min and merges into the reference gas (purge) line to
contribute to a sufficient purge flow of the sample in the autosampler (100 – 200 mL/min).
The minimum flow for the reference should be 10 mL/min.

As soon as all analyte gases are on the chromatographic column EV2 is closed and the carrier gas
flow is automatically set to 50 mL/min (He save mode). The reference flow is kept at its set value
(10 - 70 mL/min) but the contribution from the split is stopped. This status is maintained until
the next analysis is started.

40

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

2

Analytical Principles
Pneumatic Circuit

• Pyrolysis — The carrier gas flow through the pyrolysis reactor is lower than in the combustion
mode. It is commonly set to 100 mL/min. The flow through the separation column is the same as
in combustion mode: 50 ml/min if the carrier gas flow is 100 mL/min. Unlike in combustion
mode the carrier flow is not reduced during the analysis. V1 is open until acquisition is completed.
He saving is achieved by using the split flow for purge just as in combustion mode.
To avoid high pressure on the chromatographic column in the combustion system the reference
flow (50 – 150 mL/min) is split using EV3 and reunited after the chromatographic column.
Thus V3 is also open all the time.

Setting Different Flows and Switching Times
It is possible to change the default settings for switching the He Save Mode. Click on the button and
choose the tab "HeM".

Here you can define when the He Save Mode is switched on and how high the carrier flow shall be
during the He Save Mode.
Note Changing the values will have influence on your analytical results. All instructions in this
manual are based on the default settings (100s, 50 mL/min).
Note The flow during He Save Mode must be identical to the split flow through the GC column
during Analysis Mode. This is mandatory to maintain a constant TCD signal.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

41

2

Analytical Principles
Pneumatic Circuit

Adjusting the Split Flow
The split ratio is managed by the valve block and an adjustable restriction, the needle valves on top of
the GC oven in the EA. See Figure 38.
Figure 38. Needle Valves

Needle Valve Holder

HeM Valve Block

In general, there is no need to adjust the splits. It may become necessary if the restrictions after the
valve block (i.e. water trap, GC column) change or if the flow conditions (i.e. carrier gas flows) are
modified. This chapter is about the procedure to change the settings.
Note Any changes may have influence on the system performance and should be done only when
necessary and after reading this manual.

42

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

2

Analytical Principles
Pneumatic Circuit

Pneumatic Circuit and Measuring Points
Figure 39 is a simplified pneumatic circuit of the Flash IRMS. It shows two measuring points “TP”.
Here you must disconnect the lines and connect to a flow meter. The split ratio of combustion side
should be 130/50 mL/min where 130 mL/min are fed back into the purge line and 50 mL/min go
through the GC column.
Note The flow through the GC column should not exceed 60 mL/min in analytical conditions.
Figure 39. Simplified Pneumatic Circuit

1. Make sure you are in combustion mode (arrow points to NCS).
2. Make sure V2 Comb is on.

3. Set carrier flow to 180 mL/min and the reference flow to 70 mL/min.
4. Disconnect the purge line of your autosampler (see measuring point TP1 in Figure 39) and
measure the flow. It should read 200 mL/min (±2 mL/min).
−

If the flow is higher, use a screwdriver to gently close needle valve V2 turning it
counter-clockwise.

−

If the flow is lower, use a screwdriver to open the needle valve turning it clockwise.

5. Now remove the line to the ConFlo (or SmartEA, measuring point TP2 in Figure 39) and measure
at the exit of the EA. It should read 50 mL/min (±2 mL/min).
CAUTION The needle valves are very sensitive. It is best if the fixing nut is already fixed before you
use the screwdriver. This allows a finer adjustment.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

43

2

Analytical Principles
Preparation of Reactors and Adsorption Filter

Preparation of Reactors and Adsorption Filter
NC Configuration
Table 17 report the characteristics of the components required for NC determination, and the type and
size of the filling materials to be used for a proper preparation of the reactor.
Table 17. Components required for NC Determination
Reference

Component

Characteristic

Filling Materials

Right Channel

Reactor

Material: Quartz

Quartz Wool
Chromium Oxide
Reduced Copper
Silvered Cobaltous-Cobaltic Oxide

F2

Absorption Filter

Material: Plexiglas

Quartz Wool
Magnesium Perchlorate (Anhydrone)

CC2

Gas chromatographic
Column

Material: Stainless
Steel

---

Figure 40 shows the size of the filling material.
Figure 40. Size of Filling Material

NC Configuration with EA IsoLink CN/OH
F2 Adsorption Filter: PlexiglasTube

Right Channel Reactor : Quartz Tube

Right
Channel
10 mm

Quartz Wool

50 mm

Chromium Oxide

10 mm

Quartz Wool

100 mm

F2
Copper Oxide

10 mm

Quartz Wool

50 mm

Silvered Cobaltous-Cobaltic Oxide

20 mm

Quartz Wool

ConFlow IV
or “CF IV”

Quartz Wool

Magnesium Perchlorate
(Anhydrone)

Copper Reduced

10 mm

IRMS

10 mm

Quartz Wool

TCD

CC2

Gas Chromatography Column

Note In some cases, it can be advantageous to use copper oxide instead of chromium oxide.
Reactors filled with copper oxide must be operated at 900°C - 920°C. Check your calibration with
appropriate standards to maintain analytical precision and quality after changing the reactor filling
from chromium oxide to copper oxide.
• Reactors — The reactor is a quartz tube having a conical bottom end. There are two different sizes
available:
−

44

quartz glass reactor of 18 mm OD

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

2 Analytical Principles
Preparation of Reactors and Adsorption Filter
−

quartz glass reactor of 25 mm OD (Macro Reactor)
Note The EA IsoLink IRMS System for CN/OH comes standardized for a 18 mm OD
quartz glass reactor. For 25 mm reactor tubes the MAS Plus and the top of the EA must be
modified. See the section “Upgrading EA for 25 mm OD Macro Reactor” on page 131.

• Filter — It is a Plexiglas filter. The filling materials used are according to the NC analytical
determination.
 To set the temperature

The following procedure is recommended for heating up the furnace of the combustion reactor when
operating with chromium oxide (Cr2O3) at 1020°C:
1. Increase the temperature from room temperature to 400°C and check the background signals on
the Mass Spectrometer. Check the system for leaks.
2. Increase the temperature from room 400°C to 900°C in steps of 100°C and hold at 900°C for at
least 15 min. This avoids melting the copper within the reactor, which will cause poor
performance.
3. Increase it to the operating temperature of 1020 °C.
WARNING Immediate temperature increase to 1020 °C can cause copper melting. Cooling down
in steps can avoid reactor breaking. Alternatively, you can use the HeatUp and CoolDown script in
Isodat. Right-click on the Flash IRMS visualization window in Isodat and choose the appropriate
context menu item. See page 82.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

45

2

Analytical Principles
Preparation of Reactors and Adsorption Filter

H and O Configurations
These configurations require the use of a special ceramic reactor, as shown in Figure 41, and the use of
a trap filled with Carbosorb (Ascarite® P/N 33835236) and Magnesium Perchlorate
(P/N 338 21900). after the reactor and before the HeM valve block. This additional trap is not
included in delivery.
Figure 41. High Temperature Conversion Reactor for H and O Analyses of Solid (A) and Liquid (B) samples

Reactor for H and O Configurations with EA IsoLink CN/OH
17 mm

A

Reactor for solid sample

17 mm

B

Ceramic tube

Ceramic tube

60 mm

60 mm

Special insert

270 mm

270 mm

Insert

Glassy carbon reactor

Glassy carbon reactor

355 mm

355 mm

30 mm

Reactor for liquid sample

Graphite crucible

100 mm

130 mm

Glassy carbon granulate

5 mm
15 mm

Silver wool
Glassy carbon granulate
Quartz wool
Silver wool

Glassy carbon granulate

5 mm
15 mm

Silver wool

15 mm
5 mm

Quartz wool
Silver wool

Glassy carbon granulate

15 mm
5 mm

IMPORTANT The filling of the reactor must be performed rigorously respecting the exact
proportions of the packing materials. The distance of 270 mm from the head of the ceramic tube to
the graphite crucible is of extreme importance, because the crucible must be placed in the hottest
zone of the furnace. Additionally avoid that any organic material enters the reactor.
The filling of the trap is given in Figure 42 using quartz wool at top and end and to separate the filling
material.
Figure 42. CO2/H2O Trap Between High Temperature Conversion Reactor and HeM Valve Block

Preparing the Reactors
According to the instrument configurations, the filling materials are introduced into the reactor in a
way to form a series of layers of defined dimensions. For a proper preparation of the filling layers, refer
to the filling diagram of the concerned instrument configuration, as described in the section
“Preparation of Reactors and Adsorption Filter” on page 44. When using macro reactors (25 mm D)
then use the same packing instruction.

46

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

2 Analytical Principles
Preparation of Reactors and Adsorption Filter

Note The conical end of the quartz glass tube must be at the bottom.
WARNING Before using the filling materials required for this operation, please read the hazard
warnings and information reported in the Material Safety Data Sheets (MSDS) provided, referring
to the relevant CAS (Chemical Abstract Service) number.
The filling of reactors requires the use of quartz wool. Before handling quartz
wool, we recommend to wear gloves and face protection.
Always use original Thermo Fisher Scientific materials and products. The use of materials not
meeting the technical specifications of our products does not ensure a good operation of the
instrument and may even damage it.
The filling procedure should be carried out on a wide and clean workbench. See the following
procedures:
• “To fill the quartz reactor” on page 47
• “To fill the adsorption filter” on page 49
 To fill the quartz reactor

The following procedure provides instructions for filling a quartz reactor.
Material Required

Quartz reactor
Compression rod
Filling material
1. Starting from the reactor conical bottom end, introduce a sufficient amount of quartz wool to
form the required layer, as shown in Figure 43.
Figure 43. Introduction of Quartz Wool into the Conical End of the Reactor

Quartz Wool
Conical Bottom End

2. Plug with your finger the mouth of the reactor conical end. Gently press the quartz wool using the
rod provided, as shown in Figure 44.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

47

2

Analytical Principles
Preparation of Reactors and Adsorption Filter

Figure 44. Compression of Quartz Wool into the Quartz Reactor

Figure 45. Filling of the Quartz Reactor

3. Turn the reactor conical end downward and rest it delicately onto the workbench.
4. Pour sequentially the required filling materials into the reactor, as shown in Figure 45, ensuring
that each layer has the indicated size. At each step gently press the quartz wool using the rod
provided.
5. The last step of the sequence consists in introducing a sufficient quantity of quartz wool to form
the last required layer, as shown in Figure 46.
Figure 46. Introduction of Quartz Wool as Last Layer of the Sequence

Quartz Wool

6. Delicately press the quartz wool using the rod provided.

48

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

2 Analytical Principles
Preparation of Reactors and Adsorption Filter

Preparing the Adsorption Filter
The filling materials are introduced into the empty filter to form a series of layers of defined
dimensions. For a proper preparation of the layers, refer to the filling diagram of the concerned
instrument configuration, as described in the section “Preparation of Reactors and Adsorption Filter”
on page 44.
WARNING Before using the filling materials required for this operation, please read the
hazard warnings and information reported in the Material Safety Data Sheets (MSDS)
provided, referring to the relevant CAS (Chemical Abstract Service) number.
The filling of reactors requires the use of quartz wool. Before handling
quartz wool, we recommend to wear gloves and face protection.
Always use original Thermo Fisher Scientific materials and products. The use of materials
not meeting the technical specifications of our products does not ensure a good operation
of the instrument and may even damage it.
The filling procedure should be carried out on a wide and clean workbench. See “To fill the adsorption
filter” on page 49.
 To fill the adsorption filter

The following procedure provides instructions for filling an adsorption filter.
Material required
Plexiglas filter
Compression rod
Filling materials

1. Introduce into either of the tube ends a sufficient amount of quartz wool to form the required layer
as shown in Figure 47.
Figure 47. Introduction of Quartz Wool into the Tube
Threaded End

Quartz Wool

2. While plugging the tube mouth with your hand, press gently the quartz wool using the rod
provided.
3. Screw the nut complete with its seal onto the threaded mouth, as shown in Figure 48.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

49

2

Analytical Principles
The Ramped GC Oven

Figure 48. Nuts and Seals for Adsorption Filter
Seal
Nut

4. Pour sequentially the required filling materials into the adsorption filter, ensuring that each layer
has the indicated size. At each step gently press the quartz wool using the rod provided.
5. Do the last layer using a sufficient quantity of quartz wool to form the required layer.
6. Complete the procedure by screwing on the second nut complete with its seal, as shown in
Figure 49.
Figure 49. Preparation of the Adsorption Filter

The Ramped GC Oven
The Ramped GC Oven is controlled by a JUMO controller. The controller is
pre-programmed and locked. It allows to switch between two setpoints, SP1
and SP2.
• SP1 is the start temperature of 70 °C
• SP2 is the end temperature of 240 °C
When receiving the GC start signal (GC On) from Isodat, the Ramped GC Oven increases
automatically the temperature from SP1 to SP2. When stopping the GC (GC Off ) the oven cools
down to SP1. The starting temperature of 70 °C is defined to separate the N2 and CO2 peaks.
After the CO2 peak a rapid increase of the temperature accelerates the elution of the SO2.
Detailed information can be found in the Ramped GC Oven Operating Manual.

50

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

3
Installing Flash IRMS Elemental Analyzer
This chapter provides the instruction for installing the Flash IRMS.
Contents

• Preliminary Information
• Making the Gas Supply Plumbing Connections
• Electrical Connections
• Installing the Reactors into the Furnaces
• Setting Pressure and Flow Parameters
• Installing Autosampler
• Isodat Software Suite

Preliminary Information
This chapter contains the preliminary information for installing the EA IsoLink IRMS System for
CNSOH, and for the electrical requirements.

Who Performs the Installation
Your EA IsoLink IRMS System for CNSOH will be installed by an authorized Thermo Fisher
Scientific engineer (FSE), who will verify the instrument operation. If, for any reason, your system is
not installed by a Thermo Fisher Scientific FSE, you should ensure that the following operations are
performed.

Standard Outfit
Use the standard outfit checklist accompanying the instrument to verify that all items have been
received.

Verify Site Preparation
Before installing the EA IsoLink IRMS System for CNSOH, your laboratory must be in compliance
with the guidelines and requirements described in the EA IsoLink IRMS System for CNSOH
Preinstallation Requirements Guide.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

51

3

Installing Flash IRMS Elemental Analyzer
Preliminary Information

Unpacking the Instrument
ATTENTION This operation must be carried out by a Thermo Scientific Field Service Engineer.

After inspecting the exterior of the shipping container for damage carefully unpack the instrument and
do the following:
1. Check the contents of each box against the packing list to verify the shipment is complete.
2. Inspect each item for damage.
a. If equipment is damaged, keep boxes and their equipment in their existing condition and
immediately notify the carrier.
b. Submit a damage claim directly to the carrier, and send a copy (including any shortage claim)
to your authorized Thermo Fisher Scientific sales representative.
c. Do not return any equipment to the dealer or the factory without prior Thermo Fisher
Scientific authorization.

Placing the Instrument
Place the EA IsoLink IRMS System for CNSOH on the workbench, allowing free access to electrical
connections and gas lines.
LIFTING HAZARD The Flash IRMS Elemental Analyzer weighs approximately 65 kg (145 lb)
when unpacked. Pay attention when lifting the instrument onto the workbench. At least TWO
people should perform this operation, each standing on left/right side of the instrument and
putting their hands near its supporting feet.

You should already have prepared your laboratory according to the space requirements specified in the
Preinstallation Requirements Guide. The gas and power supplies should have been made accessible.
Optional equipment should be placed near the analyzer to be easily connected.

52

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

3

Installing Flash IRMS Elemental Analyzer
Making the Gas Supply Plumbing Connections

Making the Gas Supply Plumbing Connections
This section provides instructions for making the gas supply plumbing connections.

Building the Gas Lines
Building the gas supply lines from the supply cylinders to the elemental analyzer includes connecting
the gas lines to the supply tanks and installing any traps or filters on the line.
To properly connect the gas lines to the gas tanks, you will need the following materials:
• 1/8-in. in diameter (gas lines longer than 3 m [10 ft]
• 1/8-in. stainless steel tubing, properly cleaned
• a 1/16-in. stainless steel tubing, properly cleaned (for the last couple of meters to the instrument)
• a tubing cutter
• connecting nuts and relevant ferrules, reducing pieces 1/8-in. to 1/16-in.
• two wrenches
WARNING Secure gas cylinders to an immovable structure or wall. Handle all gases according to
local safety regulations.
Use the following procedure to connect regulators and tubing to the gas supply tanks:
 To regulate and connect tubing

1. Make sure the initial supply valves are turned off.
2. Connect the regulator to the gas supply tank. Use an open-ended wrench or adjustable wrench to
tighten the regulator connection.
3. Determine the length of tubing you need. Use only enough tubing to connect the instrument to
the gas cylinders, but allow enough slack in case the instrument should be moved at least 40 cm
(16 in.) from other equipment. This allows enough room to perform system maintenance.
4. Use a tubing cutter to cut the tubing.

Purging Gas Lines
We recommend to purge the lines any time you make a cut in the tubing during the gas line assembly
process. This will clear them of any debris from the cut. You should also purge the completely
assembled gas lines before you connect the gas supply to the EA IsoLink IRMS System for CNSOH.
Use the following procedure to purge the gas lines:
 To purge the gas lines

1. Turn the gas supply on, and set the pressure to 35 kPa (0.35 bar).
2. Allow the line to purge for 10 minutes.
3. Turn off the gas supply.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

53

3

Installing Flash IRMS Elemental Analyzer
Electrical Connections

Connecting the Gas Lines
The gas supply lines must be connected to the instrument back panel using the proper inlets and
fittings.
CAUTION The maximum pressures of the gases to supply the Flash IRMS Elemental Analyzer is
700 kPa (7 bar).

 To connect gas lines

1. Connect the Helium gas line to the inlet marked He on the instrument rear panel.
Gas inlet pressure must be set 400-500 kPa (4-5 bar, 58-73 psig).
2. Connect the Oxygen gas line to the inlet marked O2 on the instrument rear panel.
Gas inlet pressure must be set 400-500 kPa (4-5 bar, 58-73 psig) according to the instrument
configuration.
3. By using the pressure regulators and the pressure gauges located in the detector compartment of
the instrument, set the pressure of the gases as follows:
• 250-300 kPa (2.5-3 bar) for helium (He)
• 250-300 kPa (2.5-3 bar) for oxygen (O2)

Electrical Connections
This paragraph explains the electrical connections of the EA IsoLink IRMS System for CNSOH, and
helps you install and configure peripheral devices and EagerSmart Data Handling Software.
CAUTION The power line and the connections among the instruments must maintain good
electrical grounding. Poor grounding represents a danger for the operator and may seriously affect
the instrument performance.
Do not connect the EA IsoLink IRMS System for CNSOH to lines feeding devices of a heavy duty
nature, such as motors, UV lamps, refrigerators and other devices that can generate disturbances. If
other instruments, such as computer, balance, printer, ans so forth, have to be connected to the
same electrical line as the EA IsoLink IRMS System for CNSOH, ensure that such electrical line is
capable of withstanding such electrical consumptions by calculating the total absorption.

Mounting Peripheral Devices
Unpack them and follow instructions included with them. Follow the instructions in the paragraphs
below to connect your peripheral devices to the EA IsoLink IRMS System for CNSOH.

54

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

3

Installing Flash IRMS Elemental Analyzer
Electrical Connections

Power Connections
WARNING This instrument is electrically powered, and therefore all electrical connections must
be provided with good grounding.
Poor grounding can represent a danger to the operator and adversely affect the instrument
efficiency.

Performing Electrical Connections
 To connect autosampler cable

1. Connect the signal cable of the MAS Plus autosampler to the connector marked Autosampler, on
the back panel of the Elemental Analyzer.
 To connect RS232 cable

Perform this operation if the RS 232 serial line connection is required.
2. Connect the RS 232 cable supplied in the standard outfit between the COM1 or COM2 ports of
your computer and the 9-pin connector marked RS 232 on the instrument connection panel.
If your computer is equipped with USB ports, a Serial-to-USB adapter is required for properly
connecting the cable.
3. Plug in the instrument and computer power cables.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

55

3

Installing Flash IRMS Elemental Analyzer
Installing the Reactors into the Furnaces

Installing the Reactors into the Furnaces
Note In the following figures, the connections of the tubings are not visible for convenience.
Before installing the reactors for pyrolysis and combustion the following preliminary operation must be
carried out.
1. Make sure that the furnaces are at room temperature.
2. Open the furnaces compartment by lifting the cover and removing the protection plate.
See Figure 50.
Figure 50. Access to the Furnace Compartment

Furnace Compartment Cover

Furnace Compartment Protection Plate

Protection Plate
Left Furnace
Right Furnace

3. If present, remove the protection cap from the coupling union of each channel.
a. By using a 32-mm wrench, unscrew the reactor retaining nut. Remove the protection cap, the
washer, and the conical O-ring as shown in Figure 51.

56

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

3

Installing Flash IRMS Elemental Analyzer
Installing the Reactors into the Furnaces

Figure 51. Protection Cap Removal

Protection Cap

Reactor Retaining
Nut
Left Channel - Pyrolysis

Washer
Conical O-ring
Coupling Union

Bottom Feed
Connector

Right Channel - Combustion)

Protection Cap

Reactor Retaining
Nut
Washer
Conical O-ring
Coupling Union

ATTENTION Do not dispose of the protection cap but keep it in your lab for future use.
The protection cap should be reinstalled on the coupling union in case the instrument is not used
for a long time. The washer and the conical O-ring removed are used for the installation of the
reactor.
4. Remove the MAS Plus autosampler, if installed, by manually undoing its fixing nut. See Figure 52.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

57

3

Installing Flash IRMS Elemental Analyzer
Installing the Reactors into the Furnaces

Figure 52. Removing MAS Plus Autosampler
Nut

5. Install the reactors following the instruction reported in the To insert the reactor with a bottom
Feed Connector (BFC) and To Install the combustion reactor into the right furnace operating
sequences.

CAUTION The reactors must be installed with the furnace at room temperature.

 To insert the reactor with a bottom Feed Connector (BFC)
Material required
Reactor retaining nut
Washer
Conical O-ring
Coupling union fixing tool
32-mm wrench

1. Loosely screw the nut with washer and o-ring on the BFC. Insert the ceramic tube from top. Fix it
in this position using the upper o-ring. See Figure 53.
Figure 53. Introduction of the Reactor with a Bottom Feed Connector (1)

Conical O-ring
Reactor

2. Insert the bottle brush in the glassy carbon tube and introduce it in the ceramic tube from top.
Use your other hand to carefully slip it over the o-rings of the BFC while keeping the ceramic tube
up. Hold it tight while removing the bottle brush. See Figure and Figure 55.

58

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

3

Installing Flash IRMS Elemental Analyzer
Installing the Reactors into the Furnaces

Figure 54. Introduction of the Reactor with a Bottom Feed Connector (2)
Retaining Nut

Conical
O-ring

Bottom Feed Connector

Washer

Figure 55. Introduction of the Reactor with a Bottom Feed Connector (3)

3. Now unscrew the nut and slip it with washer and o-ring over the ceramic tube at the bottom. Fix
it using the steel rod and a 32 mm wrench. See Figure 56.
Figure 56. Introduction of the Reactor with a Bottom Feed Connector (4)
32-mm Wrench

Coupling Union Fixing Tool

4. Slowly insert the granules into the glassy carbon tube from top. Make sure no granule blocks the
tube. Check the height with the mark on the graphite crucible remover tool by placing the crucible

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

59

3

Installing Flash IRMS Elemental Analyzer
Installing the Reactors into the Furnaces

on the crucible remover and inserting it into the glassy carbon tube. The distance from the top of
the crucible to the rim of the ceramic must be 270 mm. Here is the hottest zone of the furnace.
5. Drop the crucible into the reactor and introduce the insert on top of the glassy carbon tube.
6. To complete the operation manually screw the fixing nut of the autosampler. See Figure 63.
Figure 57. Reinstall MAS Plus Autosampler
Nut

7. Close the furnace compartment with the protecting plate and the front cover.
 To Install the combustion reactor into the right furnace

Note The figures in this operating sequence show the installation of a reactor into the right
furnace.
Material required
Reactor retaining nut
Washer
Conical O-ring
Coupling union fixing tool
32-mm wrench

1. Carefully introduce and guide the reactor into the furnace until the conical O-ring on the top of
the reactor touches the furnace fitting. See Figure 58.

60

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

3

Installing Flash IRMS Elemental Analyzer
Installing the Reactors into the Furnaces

Figure 58. Introduction of the Combustion Reactor Into the Right Furnace (1)
Reactor
Conical O-ring

2. Onto the bottom end of the reactor, which protrudes from the bottom of the furnaces, slide first
the reactor retaining nut, then the washer, and finally the conical O-ring, paying attention that the
conical section of the O-ring must be turned downwards. See Figure 59 and Figure 60.
Figure 59. Conical O-ring

Conical Section

Figure 60. Introduction of the Combustion Reactor Into the Right Furnace (2)
Retaining Nut

Conical O-ring

Washer

3. Insert the reactor end into the coupling union located on the base of the furnace compartment.
See Figure 61.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

61

3

Installing Flash IRMS Elemental Analyzer
Installing the Reactors into the Furnaces

Figure 61. Introduction of the Combustion Reactor into the Right Furnace (3)
Retaining Nut

Coupling Union

4. Finger-tighten the reactor retaining nut until it starts to grip the coupling union.
5. Use the coupling union fixing tool and the 32-mm wrench to tighten the retaining nut.
See Figure 62.
Figure 62. Introduction of the Combustion Reactor Into the Right Furnace (4)
32-mm Wrench

Coupling Union Fixing Tool

CAUTION Use appropriate pressure to obtain a good sealing (1/4 to 1/2 turn). Do not overtighten
the retaining nut to avoid reactor damaging.
6. To complete the operation manually screw the fixing nut of the autosampler. See Figure 63.
Figure 63. Reinstall MAS Plus Autosampler
Nut

7. Close the furnace compartment with the protecting plate and the front cover.

62

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

3

Installing Flash IRMS Elemental Analyzer
Setting Pressure and Flow Parameters

Setting Pressure and Flow Parameters
The following tables report the pressure and flow parameters to set if the helium and oxygen flows are:
• Preset at the helium, and oxygen pressure regulators.
• Adjusted and controlled through the software by means of the EFC-t control.

NCS Determination
Table 18 details the pressure and flow parameters for NCS determinations.
Table 18. NCS Determination Pressure and Flow Parameters Setting
Parameters

Flow adjustment/control through
software by EFC-t control

Flow to set through Isodat Software Suite/EagerSmart Data Handling Software
Helium as carrier gas

180 mL/min (for pressure 280-300 kPa)

Helium as reference gas

70 mL/min (for pressure 280-300 kPa)

Oxygen

250 mL/min (for pressure 250 kPa)

H-O Determinations
See Table 19 details the pressure and flow parameters for H-O determinations.
Table 19. O-H Determination Pressure and Flow Parameters Setting
Parameters

Flow adjustment/control through
software by EFC-t control

Flow to set through EagerSmart Data Handling Software/sodat Software Suite
Helium as carrier gas

100 mL/min (for pressure 280-300 kPa)

Helium as reference gas

100 mL/min (for pressure 280-300 kPa)

Oxygen

0 mL/min (for pressure 250 kPa)

IMPORTANT For H-O Analysis only: From Isodat Software Suite open the Elemental Analyzer
Status page. Select Special Function then check Disable Oxygen Injection to ensure no oxygen
introduction into the carrier circuit. This is usually taken care of by Isodat Software Suite
automatically.
The color of the button turns red. Always remember to maintain a flow of minimum 10 mL/min
through the NCS Separation Column during H-O determination in order to avoid damage to the
separation column in heated condition. This is the case in particular for liquid injections when no
purge of the autosampler is required.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

63

3

Installing Flash IRMS Elemental Analyzer
Installing Autosampler

Installing Autosampler
As standard you must install the two MAS Plus autosamplers for solid samples furnished with the
instrument. See Chapter 4, “Installing MAS Plus Autosampler.”
Alternatively, if you need to analyze liquid samples, you must install the AI/AS 1310 autosampler.
See Chapter 5, “Installing AI 1310/AS 1310 Autosampler.”

Isodat Software Suite
The instrument comes with installed Isodat Software Suite and appropriate configuration.
The software includes all necessary tools to operate the Flash IRMS and the mass spectrometer
simultaneously. If for any reason Isodat Software Suite is not installed, install Isodat Software Suite as
described in the Isodat Software Suite manual.
IMPORTANT A Flash IRMS can also be run as a stand-alone instrument for elemental analysis. In
this case the instrument must be modified and the full EagerSmart Data Handling Software version
must be installed. Please see the section Chapter 9, “Running the Flash IRMS as Stand-alone
Instrument,” and read the Technical Note.

64

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

4
Installing MAS Plus Autosampler
This chapter provides the instruction for installing the MAS Plus autosampler on the Flash IRMS.
Contents

• MAS Plus Autosampler Overview
• MAS Plus Autosampler Installation

MAS Plus Autosampler Overview
The MAS Plus autosampler for solid samples is provided with the Elemental Analyzer. See Figure 64.
Figure 64. MAS Plus Autosampler

Sample Tray
Alignment Pin

Viewer
Sampler Body
Carrier Gas Inlet
Reference Gas Inlet
Autosampler Nut
Tray Motor Box

It consists of:
• An anodized aluminum block (sampler body) provided on the left side with fittings for carrier gas
and reference gas lines connection.
• A 32-position sample-holding tray numbered 1 to 32. It is provided with a reference pin to be
introduced into the seat marked 1(one) which has a locating mark. See Figure 65.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

65

4

Installing MAS Plus Autosampler
MAS Plus Autosampler Overview

Purge/reference gas inlet

Figure 65. Sample Tray
Locating Mark
Seat “one”

The modular design of the MAS Plus allows up to three additional 32-position sample trays can be
added to reach a capacity of 125 samples. Each sample tray is installed in a specific position
defined by the numbering, and therefore they are not interchangeable. See Figure 66
Figure 66. Additional Sample Trays

Additional
Sample Trays
(Drums)

The sample numbering is detailed in Table 20.
Table 20. Sample Tray Numbering
Sample Tray

Locating Mark

Numbering

#1

Seat marked 1 (one)

from 1 to 32

#2

Seat marked 0 (zero) from 33 to 63

#3

Seat marked 0 (zero) from 64 to 94

#4

Seat marked 0 (zero) from 95 to 125

Any sample has to place in the locating mark position. The correct alignment of the locating
mark is important for the installation of the sample tray on the MAS Plus autosampler.
See Figure 67.

66

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

4

Installing MAS Plus Autosampler
MAS Plus Autosampler Installation

Figure 67. Sample Trays

Sample Tray (Drum) #4
Sample Tray (Drum) #3
Locating Mark
Sample Tray (Drum) #2
Sample Tray (Drum) #1

• A motor for the tray.
• A viewer on the sampler body.
Tip The viewer on the sampler body allows observation of the Flash combustion.

Before Sampling

Flash Combustion

MAS Plus Autosampler Installation
WARNING Before starting, make sure that the EA IsoLink IRMS System for CNSOH is powered
off and that the reactors required for your analyses are installed in their corresponding furnaces.

 To install a MAS Plus autosampler
Material Required
8 mm wrench

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

67

4

Installing MAS Plus Autosampler
MAS Plus Autosampler Installation

1. Place the autosampler on the connecting nut of the concerned channel.
2. Manually screw the autosampler nut on the concerned channel.
3. Connect the tubings coming from the gas connections, located on the analyzer, to the relevant
connections of the autosampler.
4. Connect the signal cable of the MAS Plus autosampler to the 2-pin connector, marked
Autosampler, on the back panel of the analyzer.
Take care that the autosampler on top of the left furnace is connected to the connector marked
HO and the autosampler on top of the combustion furnace is connected to the connector marked
NCS. See Figure 20 on page 21.
5. Install the samples tray.
a. Manually rotate the toothed wheel clockwise until the guide located on its rim is perfectly
aligned with the metal pin of the autosampler body.

Toothed Wheel

b. Check that the sample tray (drum) reference pin is in correspondence with the seat marked
“1”.
c. Place the sample tray, with the reference pin in correspondence with the “1” seat, onto the
toothed wheel, paying attention to have the base match with the guides.

d. Place the protection cover over the sample tray with the surface marked “Side up” turned
towards you.

68

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

4

Installing MAS Plus Autosampler
MAS Plus Autosampler Installation

6. If additional sample trays are required, install them in the correct order one over the other, paying
attention that the relevant locating marks are in correspondence with the relevant seats marked 0
(zero), and placing the samples properly.

IMPORTANT Before installing an additional sample tray, make sure that the samples to analyze are
placed in all the seats of the previous tray installed.

CAUTION Before starting samples analyses, make sure that the protection cover is positioned over
the sample tray. A complete deaeration of the area where samples are housed is only possible if the
cover is in place.
Pay attention not to invert the cover; the surface marked “Side-up” must be turned towards you.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

69

5
Installing AI 1310/AS 1310 Autosampler
This chapter provides the instruction for installing the AI 1310/AS 1310 autosampler for liquid
samples on the Flash IRMS Elemental Analyzer.
Contents

• Preliminary Information
• Installing the Direct Injection Device for Flash IRMS Elemental Analyzer
• Installing the Sampler Support on the Flash IRMS Elemental Analyzer
• Installing the AI 1310/AS 1310 on the Flash IRMS Elemental Analyzer

Preliminary Information
This section contains the preliminary information for the installation and the connection of the
AI 1310/AS 1310 sampling system to the Flash IRMS Elemental Analyzer.

Who Performs the Installation
The AI 1310/AS 1310 is installed by authorized Thermo Fisher Scientific technical engineers, who will
check its correct operation. For more details, please contact Thermo Fisher Scientific local
representatives. Should the instrument not be installed by Thermo Fisher Scientific personnel, strictly
adhere to the instructions reported in this section.

Electrical Requirement
The instrument has the following power supply rating:
• 24 Vdc through a portable external power supply, level VI efficiency
−

input 100-240 Vac; 50-60 Hz — output 24 Vdc; 3 A - 3.75 A

WARNING YOU MUST ONLY USE THE PORTABLE EXTERNAL POWER SUPPLY
FURNISHED WITH THE INSTRUMENT BY THERMO FISHER SCIENTIFIC.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

71

5

Installing AI 1310/AS 1310 Autosampler
Installing the Direct Injection Device for Flash IRMS Elemental Analyzer

CAUTION The power line and the connections between the instruments must maintain good
electrical grounding. Poor grounding represents a danger for you and might seriously affect the
instrument performance. Do not connect the AI 1310/AS 1310 sampling system to lines feeding
devices of a heavy duty nature, such as motors, UV lamps, refrigerators, and other devices that can
generate disturbances.

Lift and Carry the Sampling Unit
Lift and carry the sampling unit by hand. See Figure 68.
Figure 68. How to Lift and Carry the Sampling Unit

Installing the Direct Injection Device for Flash IRMS Elemental
Analyzer
This device is installed in replacement of the MAS Plus autosampler, when present.
See Figure 69.
Figure 69. Direct Injection Device

72

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

5 Installing AI 1310/AS 1310 Autosampler
Installing the Sampler Support on the Flash IRMS Elemental Analyzer

 To install the direct injection device

1. If present, disconnect the MAS Plus autosampler from the reactor, and place the autosampler nut
on the stainless steel plate. If the MAS Plus autosampler is not present, remove the reactor fitting
unscrewing the relevant fixing nut.
2. Disconnect the gas connection.
3. Install the direct injection device over the reactor. See Figure 70.
Figure 70. Direct Injection Device Installation
Septum Holder
Septum
Tube 2 mm OD
6MB Nut
Ferrule
Nut
Injector Body

Combustion Reactor
O-ring

Note When installing the injection device on the pyrolysis side the insert in the reactor must
be replaced by the stainless steel insert which is provided in the water injection kit.
For detailed information please refer to the corresponding manual.
a. Mount the septum by using the septum holder provided in the standard outfit.
b. Connect the gas line to the direct injection device.

Installing the Sampler Support on the Flash IRMS Elemental Analyzer
The AI 1310/AS 1310 sampling system is installed on the Flash IRMS Elemental Analyzer by using the
appropriate support provided. See Figure 71.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

73

5

Installing AI 1310/AS 1310 Autosampler
Installing the Sampler Support on the Flash IRMS Elemental Analyzer

Figure 71. Sampler Support

Fixing Screws

Slots

Guide Pivot

Spacers

Semi-Circular Plate

Support Bracket

The support consists of a semi-circular plate resting on three spacers non adjustable in height.
The top surface of the plate is provided with two slots for the introduction of the corresponding fixing
screws. Use the guide pivot for the accommodation and the centering of the sampling unit.
Before mounting the support, you must place and fix the support bracket on the top panel of the Flash
IRMS Elemental Analyzer.
 To install the sampler support on the Flash IRMS Elemental Analyzer

1. Mount the support bracket.
Note The support bracket can be installed on the left side as well as the right side of the Flash
IRMS Elemental Analyzer. Install the support bracket on the side of interest according to the
instrument configuration.
a. From the top panel of the Flash IRMS Elemental Analyzer remove the four plastic caps
covering the corresponding fixing holes. See Figure 72.
Figure 72. Plastic Caps Removal

b. Mount and fix the support bracket on the top panel of the Flash IRMS Elemental Analyzer by
using the provided fixing screws. See Figure 73.

74

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

5 Installing AI 1310/AS 1310 Autosampler
Installing the Sampler Support on the Flash IRMS Elemental Analyzer

Figure 73. Mounting Support Bracket

2. Mount the sampler support. See Figure 74.
Figure 74. Mounting Sampler Support

a. Insert the provided fixing screw into each slot present on the support.
b. Insert each screw into the relevant spacer paying attention to keep its largest surface turned
toward the support base.
c. Hold the spacers in position with their flat side toward the inside, then place the sampler
support on the support bracket.
d. Guide the two fixing screws located on the external spacers into the corresponding fixing
holes.
e.

Thermo Scientific

Loosely tighten the screws.

EA IsoLink IRMS System for CNSOH Operating Manual

75

5

Installing AI 1310/AS 1310 Autosampler
Installing the AI 1310/AS 1310 on the Flash IRMS Elemental Analyzer

f.

If you must install an AS 1310, screw the support pin into the dedicated hole on the top plate.
See Figure 75.
Note The support pin is NOT required if you must install an AI 1310.
Figure 75. Mounting Support Pin
Support Pin

Installing the AI 1310/AS 1310 on the Flash IRMS Elemental Analyzer
This section provides the instruction for installing the AI 1310/AS 1310 on the Flash IRMS Elemental
Analyzer. See Figure 76.
Figure 76. AI 1310/AS 1310 Installation

AI 1310

76

EA IsoLink IRMS System for CNSOH Operating Manual

AS 1310

Thermo Scientific

5 Installing AI 1310/AS 1310 Autosampler
Installing the AI 1310/AS 1310 on the Flash IRMS Elemental Analyzer

The installation procedure includes the following steps:
• Installation of the Sampling Unit
• Installing the Syringe
• Electrical Connections
• Starting-up

Installation of the Sampling Unit
 To install the sampling unit

1. Lift the sampling unit and insert it into the guide pivot located on the sampling system support.
Introduce the guide pivot into the hole provided on the bottom of the base. See Figure 77.
Figure 77. Installation of the Sampling Unit
AI 1310

AS 1310

Hole for the introduction
of the guide pivot

2. Open the safety door and remove the protection of the injection assembly. See Figure 78.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

77

5

Installing AI 1310/AS 1310 Autosampler
Installing the AI 1310/AS 1310 on the Flash IRMS Elemental Analyzer

Figure 78. Remove Protection

Protection

3. Insert the centering plate into its seat located in the right side section of the sampling unit base
paying attention that the guide hole, present on the arm of the centering plate, correctly fits the
injector nut. See Figure 79.
Figure 79. Centering Plate
Injection Nut

Centering Plate

.

78

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

5 Installing AI 1310/AS 1310 Autosampler
Installing the AI 1310/AS 1310 on the Flash IRMS Elemental Analyzer

4. Check the correct alignment of the sampling system support, then fix it by tightening the proper
fixing screws.
5. Insert the sample tray into the sampling unit base.
• AI 1310 — Insert the 8-position sample tray into the appropriate housing of the sampling
unit base. See Figure 80.
Figure 80. 8-position Sample Tray

8-position sample tray

• AS 1310 — Insert the dedicated support plate into its appropriate housing of the sampling
unit base. Place the 105-position sample tray on the hub located on the support. The system
will automatically recognize the sample tray at the instrument power on. See Figure 81.
Figure 81. 105-position Sample Tray

Hub

Support Plate

105-position Sample Tray

Installing the Syringe
The installation of the syringe is a simple operation. However, it must be performed with caution to
avoid damages to the syringe needle and ensure an optimal performance of the injection device.
The standard syringes have 10 μL and 250 μL capacity with a 50 mm needle. It is also possible to
install 50 μL and 100 μL syringes with needles of 50 mm.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

79

5

Installing AI 1310/AS 1310 Autosampler
Installing the AI 1310/AS 1310 on the Flash IRMS Elemental Analyzer

 To install the syringe

1. Open the safety door of the turret.
2. Insert the syringe needle into the vial capture device. See Figure 82.
Figure 82. Syringe Installation (1)

Plunger Head Guide
Flange Guide
Plunger Head
Plunger
Flange

Syringe Body

Syringe Needle

Vial Capture Device
Syringe Body Seat

80

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

5 Installing AI 1310/AS 1310 Autosampler
Installing the AI 1310/AS 1310 on the Flash IRMS Elemental Analyzer

3. Accommodate the syringe body into its seat paying attention to insert the flange and the head of
the syringe plunger simultaneously into their relevant guides.
4. Turn the lock knob by approximately 180° clockwise to lock the syringe. See Figure 83.
Figure 83. Syringe Installation (2)

Lock Knob

5. Close the safety door of the rotating turret.

Electrical Connections
 To perform electrical connections

1. By using the cable provided, connect the 9-pin female connector marked RS232 located on the
sampling unit back side to a 9-pin serial port connector (COM) of the PC.
2. Plug in the tampered connector provided into the 6-pin female connector marked GC located on
the sampling unit back side.
3. Only for AS 1310, connect the 15-pin female connector of the cable, coming from the support
plate of the 105-position sample tray, to the connector marked TRAY located on the back side of
the sampling unit.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

81

5

Installing AI 1310/AS 1310 Autosampler
Installing the AI 1310/AS 1310 on the Flash IRMS Elemental Analyzer

Starting-up
 To start-up your AI 1310/AS 1310 sampling system

1. Plug in the Vdc power cable of the external portable power supply level VI efficiency into the jack
marked 24 Vdc located on the sampling unit back side.
2. Connect the power cord of the external power supply to the mains outlet.
The AI 1310/AS 1310 sampling system will automatically run the self-testing routine during
which the following automatic checks and settings are carried out:
• Alignment between AI 1310/AS 1310 sampling system and injector
• Check of the turret travel
• Acknowledgment of the installed sample tray
• Calculation of the syringe zero
Note The self-test routine is automatically carried out every time the safety door of the
turret is closed.

82

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6
Using EA IsoLink IRMS System for CNSOH
This chapter provides instruction for using the EA IsoLink IRMS System for CNSOH.

Introduction
The EA IsoLink IRMS System for CNSOH is an enhanced elemental analyzer for O and H as well as
N, C, and S isotope analyses and element analysis. It is equipped with an automatic switching valve
(ASV) that allows fast switching from NCS analysis (combustion) to the OH analysis (high
temperature conversion/pyrolysis). Isodat Software Suite is able to detect the required analysis mode by
the gas configuration defined in the method, and automatically switches the carrier gas from one mode
to the other. The system can be set up to run elements in single, dual or even triple mode, i.e. one
element per sample, two elements per sample or three elements per sample. The ASV allows to switch
unattended between pyrolysis and combustion so that five elements of a sample can be analyzed from
two sample drops. This chapter covers the setup of dual measurements. Single measurement can be
derived from it. Triple measurements are only recommended with the ramped GC Oven. These
measurements are covered in the Ramped GC Oven Operating Manual.
Contents

• Introduction
• Dual Measurement
• Performing a Jump Calibration
• Creating an Isodat Method for N+C Measurement (Dual Measurement)
• Creating an EA Method for N+C Measurement
• Creating an Isodat Method for Single Mode S Measurements
• High-Temperature Conversion – Analysis of H and O Isotopes
• Setting Up a HO Method
• Creating an Isodat Method for Single Mode S Measurements
• H3-Factor Determination
• Measuring Sulfur Isotopes
• Before Starting a Sulfur Measurement
• Create a Gas Configuration for a Sulfur Measurement
• Starting a Sulfur Measurement

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

83

6

Using EA IsoLink IRMS System for CNSOH
Dual Measurement

Dual Measurement
Analyses of N and C in dual or single mode have to be performed with a reactor, which traps SO2.
The packing instructions for such a reactor are given in Figure 84.
Figure 84. Reactor Setup NC

After installation of a new NC reactor heat up the furnace in the following stepwise manner:
room temperature > 900 °C > 1020 °C. Upon reaching 900 °C step, allow at least 15 minutes of
equilibration time before commencing the next heating step.
A stepwise heating up of the reactor avoids temperature overshooting, and hence melting of the
reduced copper contained in the reactor. Alternatively, use HeatUpFlash from the Context Menu with
a right-click on the Flash IRMS peripheral visualization. See
Figure 85. HeatUpFlash

It is possible to perform dual measurements of hydrogen and oxygen as well as carbon and nitrogen
from a single sample with the system. This analysis mode is suitable for both solid and liquid samples.
If no S analysis is required, it is recommended to use a reactor setup that traps SO2. If S analysis is
required, it is recommended to use a triple analysis.

Procedure
For the analysis of two isotopic species (hydrogen and oxygen, or nitrogen and carbon) from a single
sample, an Isodat Method, which comprises both of them has to be defined.

84

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6

Using EA IsoLink IRMS System for CNSOH
Performing a Jump Calibration

Due to the separation of the gas chromatographic column the acquisition is divided in two parts, each
dedicated to the isotopic species (i.e. gas configuration).
As soon as the first species has been identified (i.e. H2 or N2), Isodat Software Suite automatically
changes the magnetic field so that the mass traces of the second isotopic species (i.e. CO or CO2) can
be recorded (see: Switch Gas column in the Time Events list).
If no hydrogen or nitrogen peak can be found, Isodat waits a defined time, for example 25 s, before
switching to the second gas configuration.
Figure 86. Dual Measurement for NC - Schematic Chromatogram

Performing a Jump Calibration
In order to determine the isotope ratios of different elements during the same run, switching to
another Gas Configuration is necessary. In contrast to a single element measurement, in which the
magnetic field runs the gamut from high to low and after that to the pre-calculated magnetic field,
there is not sufficient time in dual measurement to perform this procedure for the next gas
configuration. For this reason, a so-called Jump Calibration from the first gas configuration to the
next gas configuration is necessary. After the Jump Calibration has been performed, the computer
finds exactly the peak center even without performing any peak center procedure. A Jump Calibration
should be performed daily to stabilize the performance of the magnet jump.
 How to perform a jump calibration

The following procedure describes how to perform a jump calibration for H2 --> CO.
For an N2 --> CO2 jump calibration proceed similarly.
1. Open Instrument Control.
2. Choose your Configuration containing ConFlo IV and EA IsoLink IRMS System for CNSOH.
3. Allow CO reference gas to enter the source. In case of an N2 --> CO2 jump calibration open CO2
reference gas.
4. Select Scan | Jump Calibration.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

85

6

Using EA IsoLink IRMS System for CNSOH
Performing a Jump Calibration

a. In the list of available Jump Calibrations, mark the one for H2 --> CO by clicking its No.
(for example 001).
–

Slow: compensate hysteresis by Max/Min settings of the magnet.

–

Fast: magnet setting of Jump Calibration.

–

HV: high voltage setting of Jump Calibration.

b. Press Recalibrate Icon

.

Note If no Jump Calibration is available in the list, create a new one as shown below:
How to Create a new Jump Calibration:
• Click New.
• From Gasconfiguration, select H2.
• To Gasconfiguration, choose CO.
• Accept the default values.
• Click OK.
Note Magnet jumps are always from a low mass to a high mass (e.g. 2 to 28 at
H2 -> CO, or 28 to 44 at N2 ->CO2).
5. Confirm with Yes, because you have already opened the Reference Port.
6. Start Jump Calibration Procedure.
a. Jump to CO (along hysteresis curve).
b. Perform a peak center for CO in order to get the signal height.
c. Jump to H2 (along hysteresis curve; H is origin).
d. Jump to CO (not along hysteresis curve).

86

e.

Perform a peak center for CO in order to catch the peak.

f.

Repeat until magnet field setting is within the magnet window defined.

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6

Thermo Scientific

Using EA IsoLink IRMS System for CNSOH
Performing a Jump Calibration

EA IsoLink IRMS System for CNSOH Operating Manual

87

6

Using EA IsoLink IRMS System for CNSOH
Creating an Isodat Method for N+C Measurement (Dual Measurement)

Creating an Isodat Method for N+C Measurement (Dual
Measurement)
For setting up an Isodat method for N+C analysis only, please refer to the figures from Figure 87 to
Figure 96.
Figure 87. NC Method: Instrument Tab

Note
• The gas configuration denotes the starting gas configuration for an analysis. When analyzing
only one element, e.g. CO2, the starting gas configuration must be changed accordingly.
• The dilution for the second gas configuration appears only after defining a Switch Gas in the
Time Event List (see below). The settings are examples and must be adjusted to your sample
type. Dynamic dilution is only available with the smartEA Option.
• HeM settings — Ticking the box Disable He Save Mode will maintain the starting flow over
the complete acquisition time (Analysis Mode). This mode does not switch into He Save
Mode.

88

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6 Using EA IsoLink IRMS System for CNSOH
Creating an Isodat Method for N+C Measurement (Dual Measurement)

Figure 88. NC Method: Time Events List

Note By double-clicking on the relevant gas name in the time events list in column Switch Gas,
you access to Isodat Object Method Switcher window. See Figure 90.
Figure 89. NC Method: Acquisition and Time

Figure 90. NC Method: Method Switcher

Note Waiting Time specifies the maximum period of time, for how long the system/software will
wait for the previous peak (for example N2 in a dual N-C measurement), to reach background
again, and delay the subsequent peak jump.
For example: if you enter a waiting time of 10 seconds, the software will delay the peak jump as
specified in the time event list by a maximum of 10 seconds.
• If the N2 peak has already reached background until the time specified in the time events list,
then the peak jump will occur straight away at that specified time.
• If the N2 peak hasn't reached background until the specified time in the time events list and the
entry for waiting time is for example 10 seconds (you can also enter a different waiting time or
0 seconds here), the peak jump will be delayed by a maximum of that number of seconds, so
that the N2 peak still has that number of seconds to reach background.
• If the N2 peak reaches background within these seconds of waiting time, the peak jump will
then occur. If the N2 peak does not reach background within the waiting time, the peak jump
will definitely occur after the waiting time period.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

89

6

Using EA IsoLink IRMS System for CNSOH
Creating an Isodat Method for N+C Measurement (Dual Measurement)

Figure 91. NC Method: Evaluation@N2

Figure 92. NC Method: Peak Detection@N2

Figure 93. NC Method: Printout@N2Tab

90

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6 Using EA IsoLink IRMS System for CNSOH
Creating an Isodat Method for N+C Measurement (Dual Measurement)

Figure 94. NC Method: Evaluation@CO2 Tab

Figure 95. NC Method: Peak Detection@CO2 Tab

Figure 96. NC Method: Printout@CO2 Tab

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

91

6

Using EA IsoLink IRMS System for CNSOH
Creating an EA Method for N+C Measurement

Creating an EA Method for N+C Measurement
The following figures (Figure 97 to Figure 101) show how an EA-method for a dual analysis N+C may
be set up.
Figure 97. NC EA Method: Instrument Temperature

Note The temperature of the high-temperature conversion furnace (left furnace) may be different
depending on the intended use of it.
Figure 98. NC EA Method: Instrument Flow-Timing

Note Sampling delay and oxygen injection end vary depending on the sample nature. O2 injection
can be crucial in a single reactor if it exceeds the required amount.
The reference flow is used for the autosampler purge. This flow can be set to values between
20 mL/min and 70 mL/min (150 - 200 mL/min in total when using HeM), with the goal to
minimize or eliminate the blank contribution of the autosampler.
For single mode N analyses, a second trap can be installed filled with Carbosorb® or Ascarite® to
remove CO2. The analysis time has to be reduced even further. In this case the entry Switch Gas CO2
can be deleted in the Time Event List of the Isodat Method.

92

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6 Using EA IsoLink IRMS System for CNSOH
Creating an Isodat Method for Single Mode S Measurements

Figure 99. TCD Settings

Figure 100. Peak Integration and Evaluation Parameters

Figure 101. TCD Control Parameters

Creating an Isodat Method for Single Mode S Measurements
This Isodat method must be used with reactor packing as for the triple NCS measurements.
The elution time of the SO2 peak can be decreased if the GC oven temperature is ramped earlier.
For setting up an Isodat method for single mode S analysis only please refer to the Ramped GC Oven
Operating Manual.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

93

6

Using EA IsoLink IRMS System for CNSOH
High-Temperature Conversion – Analysis of H and O Isotopes

High-Temperature Conversion – Analysis of H and O Isotopes
The separation column is installed in the GC Oven of the EA together with the TCD. During analysis
the temperature should be kept in a range between 70 - 90 °C. Nitrogen containing samples may be
run at low temperature using dilution at the elution of the N2 prior to the CO
Note Bake-out at maximum temperature for at least 24 hours (plus carrier flow for both analytical
sides) must be considered when separation of the peaks is not sufficient any more or if δ18O
precision, accuracy, or both are deteriorating.
To achieve longer maintenance periods, the pyrolysis side has a trap filled with Carbosorb® (e.g.
Ascarite® P/N 338 35236) and Magnesium Perchlorate (P/N 338 21900). These chemicals remove
CO2 and water from the carrier gas at standby temperatures of the reactor minimizing the bake-out
intervals. The filling of the trap is given in Figure 102 using quartz wool at top and end, and to
separate the filling material.
Figure 102. Trap Pyrolysis

Bottom Field Connector
The EA IsoLink IRMS System for CNSOH is equipped with a bottom feed connector for high
temperature conversion analysis, also referred to as pyrolysis. It allows high sample throughput.
This connector supplies the carrier gas helium from the bottom into the reactor. The helium flows up
and is redirected downwards through the glassy carbon tube where it exits at the bottom connector
again. See Figure 103.
Figure 103. Flow Path with Conventional and Bottom Feed Connector

94

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6 Using EA IsoLink IRMS System for CNSOH
High-Temperature Conversion – Analysis of H and O Isotopes

Inserting the Reactor with a Bottom Feed Connector (BFC)
 To insert the reactor with a bottom feed connector

1. Loosely screw the nut with washer and o-ring on the BFC. Insert the ceramic tube from top.
Fix it in this position using the upper o-ring.
2. Insert the bottle brush in the glassy carbon tube and introduce it in the ceramic tube from top.
Use your other hand to carefully slip it over the o-rings of the BFC while keeping the ceramic tube
up. Hold it tight while removing the bottle brush.
3. Now unscrew the nut and slip it with washer and o-ring over the ceramic tube at the bottom.
Fix it using the steel rod and a 32 mm wrench.
4. Slowly insert the granules into the glassy carbon tube from top. Make sure no granule blocks the
tube. Check the height with the mark on the graphite crucible remover tool by placing the crucible
on the crucible remover and inserting it into the glassy carbon tube. The distance from the top of
the crucible to the rim of the ceramic must be 270 mm. Here is the hottest zone of the furnace.
5. Drop the crucible into the reactor and introduce the insert on top of the glassy carbon tube.

Removing the Reactor from the Bottom Feed Connector (BFC)
 To remove the reactor from the bottom feed connector

1. Wait to cool down.
2. Put in sample dilution in the ConFlo, or close the needle valve at the mass spectrometer. Stop the
carrier gas flow through the pyrolysis system. Wait for the system to de-pressurize (3-5 minutes).
3. Remove the autosampler for liquids.
4. Unscrew the BFC using the steel rod and a 32 mm wrench.
5. Put on lint free gloves. Remove upper graphite tube insert from inside the ceramic tube.
6. Insert the bottle brush into the glassy carbon reactor.
7. Lift the ceramic tube with one hand and get a finger underneath the glassy carbon tube keeping
the ceramic tube up with the same hand.
8. Slip the glassy carbon tube from the o-rings of the BFC while carefully pulling the tube with the
bottle brush from top.
9. Remove o-ring, washer and nut at the bottom of the ceramic tube and take it out.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

95

6

Using EA IsoLink IRMS System for CNSOH
Setting Up a HO Method

Setting Up a HO Method
This section describes the dual measurement to determine hydrogen and oxygen from a single sample.
It contains the following topics:
• “Procedure Overview” on page 96
• “Dual Measurement” on page 97
• “Creating a New Method for Dual Measurement” on page 97
• “Creating a Sequence for Dual Measurement” on page 104
• “Start Data Acquisition” on page 105
• “Dual Measurement Results” on page 106

Procedure Overview
It is possible to perform dual measurements of hydrogen and oxygen from a single sample with the
system (Thermo Chemical Elemental Analyzer, ConFloIV and IRMS) in less than 7min offering fast
sample throughput and high productivity. The technique is suitable for both solid and liquid samples.
Different autosamplers are used to measure different sample types, see Table 21.
Table 21. Dual Measurement and Sample Type
Sample Type

Autosampler

Solids and viscous liquids (for example, honey,
serum, olive oil)

MAS 200, MAS Plus or others

Liquids, for example, water, alcohol, urine

AS3000, AS 1310, GC Pal, TriPlus 100 LS.

For the analysis of two isotopic species (hydrogen and oxygen) from a single sample, a method that
comprises both must be defined. As soon as the hydrogen peak (3 in Figure 104) has been identified,
Isodat Software Suite stops the HD acquisition. The magnet jumps to the CO configuration (defined
in the Switch Gas column in the time events list, see Figure 106 on page 98).
HO

Figure 104. Schematic Chromatogram of a HO Dual Measurement

96

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6

Using EA IsoLink IRMS System for CNSOH
Setting Up a HO Method

If no hydrogen peak can be found, Isodat Software Suite waits, for example, 25s. Data acquisition
follows immediately after injection. See Figure 104.

Dual Measurement
Thermo Fisher Scientific assumes that the user already has working experience with the ConFlo IV
interface and the IRMS. Thermo Fisher Scientific recommends to perform a simple check in order to
test the analytical condition of the complete system before measuring any samples. Before the dual
measurement can be started, some preparation are necessary.
Before running a dual measurement, the Peak Jump from H2 to CO must be created. See “Performing
a Jump Calibration” on page 85. If a jump is already defined, make sure the jump is re-calibrated
regularly. It is recommended to perform a re-calibration on a daily basis.

Preparing the Hardware
 To prepare the hardware components

1. Set the reactor temperature to 1450 °C.
2. Set the GC column temperature to 90 °C. Set it to 70 °C if you are running N-containing samples.
3. Set the helium flow to >100mL/min.
4. Set the reference flow to 100 mL/min.
5. Make sure that the standard gas (H2 and CO) on ConFloIV are available.

Determining the H3-Factor
The H3-factor is determined as described in the section “H3-Factor Determination” on page 107.

Creating a New Method for Dual Measurement
As a guideline for creating a new dual measurement method, the method HD_CO.met should be
used.
 To create a new method for an dual measurement

1. Open Acquisition.
2. Click New to create a new method.
3. In the File New dialog box, double-click Method.
4. Click the tab Instrument of the new dual measurement method, see Figure 105.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

97

6

Using EA IsoLink IRMS System for CNSOH
Setting Up a HO Method

Figure 105. Instrument Page of Dual Measurement Method

5. Set the Instrument parameters for your new dual measurement method as shown in Figure 105.
6. Click the tab Time Events, see Figure 106.
Figure 106. Time Events Page of the Dual Measurement Method.

7. Set the Time Events parameters for your new dual measurement method as shown in Figure 106.
8. Double-click on CO in the column Switch Gas. The Event Time is active since for example, 130s.

98

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6

Using EA IsoLink IRMS System for CNSOH
Setting Up a HO Method

The MethodSwicher window opens, see Figure 107.
Figure 107. Switching Between Methods

9. For Gasconfiguration, select the gas configuration to be switched to, for example, CO.
10. Select the Event when the switch is to be performed, for example, Isodat@Eval@PeakFound.The
switch is performed when the peak is found.
11. For Waiting Time, enter for example, after 30 s.
If a peak is found, the jump takes place. If no peak is found, the waiting time elapses before the
jump takes place immediately.
12. Click OK.
13. Click the tab Evaluation@H2. See Figure 108.
Figure 108. Evaluation@H2 Page of the Dual Measurement Method

Note There are Evaluation pages, Peak Detection pages and Printout pages are for each, H2
and CO, for example, Evaluation@H2 and Evaluation@CO2.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

99

6

Using EA IsoLink IRMS System for CNSOH
Setting Up a HO Method

Note The values for reference gas and Amount% are examples and must be adjusted according
to the isotope value of the reference gas and amount% of H in your sample. Please refer also to
the ConFlo IV manual, Chapter 4, where the settings for reference/blank are described.
14. Set the Evaluation@H2 parameters for your new dual measurement method as shown in
Figure 108.
15. Click the tab Peak Detection@H2, see Figure 109.
Figure 109. Peak Detection@H2 Page of the Dual Measurement Method

16. Set the Peak Detection@H2 parameters for your new dual measurement method as shown in
Figure 109.
17. Click Advanced Parameter to open the advanced parameter settings, see Figure 110.
Figure 110. Peak Detection@H2 Page – Advanced Parameters of the Dual Measurement Method

18. Set the Advanced Parameters parameters for your dual measurement method as shown in
Figure 110.
19. Select the Printout@H2 tab of your new dual measurement method, see Figure 111.
Figure 111. Printout@H2 of the Dual Measurement Method

100

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6

Using EA IsoLink IRMS System for CNSOH
Setting Up a HO Method

Note Usually, the printout of results is not performed until the complete measurement has
been finished. Therefore, it is recommended to choose No Printout.irw on the Printout@H2
page. Printout options are best defined at the end of the Printouts page.
20. Click the tab Evaluation@CO. See Figure 112.
Figure 112. Evaluation@CO Page of the Dual Measurement Method

21. Set the Evaluation@CO parameters for your new dual measurement method as shown in
Figure 112.
22. 3.Click the tab Peak Detection@CO, see Figure 113.
Figure 113. Peak Detection@CO Page of the Dual Measurement Method

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

101

6

Using EA IsoLink IRMS System for CNSOH
Creating an EA Method for H+O Measurement

Note The Start Detection time must be adjusted so that the spikes on the mass traces from the
magnet jump are excluded. This may vary depending on the Time Event List setup.
23. Set the Peak Detection@CO parameters for your new dual measurement method as shown in
Figure 113.
24. Select the Printout@CO tab of your new dual measurement method, see Figure 114.
Figure 114. Printout@CO Page

25. Select your Printout templates.
26. Click Save | Save as to save this method with the file ending *.met.30.
27. Enter a name for your method and click Save.

Creating an EA Method for H+O Measurement
The following figures show how an EA-method for a dual analysis H+O may be set up.
Figure 115. HO EA Method: Instrument Temperature

Note The temperature of the high-temperature conversion furnace (left furnace) may be different
depending on the intended use of it.

102

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6

Using EA IsoLink IRMS System for CNSOH
Creating an EA Method for H+O Measurement

Figure 116. NC EA Method: Instrument Flow-Timing

Note Sampling delay and oxygen injection end vary depending on the sample nature. O2 injection
can be crucial in a single reactor if it exceeds the required amount.
The reference flow is used for the autosampler purge. This flow can be set to values between
20 mL/min and 70 mL/min (150 - 200 mL/min in total when using HeM), with the goal to
minimize or eliminate the blank contribution of the autosampler.
For single mode N analyses, a second trap can be installed filled with Carbosorb® or Ascarite® to
remove CO2. The analysis time has to be reduced even further. In this case the entry Switch Gas CO2
can be deleted in the Time Event List of the Isodat Method.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

103

6

Using EA IsoLink IRMS System for CNSOH
Creating an EA Method for H+O Measurement

Creating a Sequence for Dual Measurement
 To create a new sequence for dual measurement

1. Place a sample (for example, 0.281 mg of benzoic acid) in the solids autosampler.
2. Open Acquisition.
3. Click the New button to create a new sequence.The File New dialog box opens, see Figure 117.
Figure 117. Creating a New Sequence

104

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6

Using EA IsoLink IRMS System for CNSOH
Creating an EA Method for H+O Measurement

4. In the File New dialog box, double-click Sequence. The Sequence Properties window opens, see
Figure 118.
Figure 118. Defining Number of Samples

5. In the Sequence Properties window, enter the Number of Samples, for example, 3.6.
6. Click OK. The sequence grid opens, see Figure 119.
Figure 119. Creating a Sequence for Dual Measurement

7. For each row, enter the Amount of the individual sample, for example 0.260.
8. Enable

to perform a Peak Center

prior to measurement, see Figure 119.

9. Select an EA Method, for example Py_iHeM.eam.
10. Edit text in Identifier 1 to identify the sample, for example benzoic acid.
11. Select Method, for example Py_iHeM.
12. Save the sequence with a name of your choice.
13. Continue with “Start Data Acquisition” on page 105.

Start Data Acquisition
 To start the data acquisition

1. In the Sequence window, click Start to start the sequence.The Start Sequence window opens, see
Figure 120.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

105

6

Using EA IsoLink IRMS System for CNSOH
Creating an EA Method for H+O Measurement

Figure 120. Defining Results Export and Printout Parameters

2. Select your destination Folder Name and File Name for the Results.
3. Select the Format for the export file.
4. Select Printout, for example, Yes.
5. Click OK.

Dual Measurement Results
The following events occur during the acquisition:
• Peak center procedure is performed.
• The H2 reference gas pulse is activated after 20 seconds for 20 seconds.
• The H2 reference gas pulse is activated after 60 seconds for 20 seconds.
• The sample is dropped into the reactor.
• The H2 sample peak appears approximately 15 seconds after reaction start. See Figure 121.
• The CO sample peak appears approximately 100 seconds after reaction start. See Figure 121.
• The CO reference gas pulse is activated after 340 seconds for20 seconds.
• The CO reference gas pulse is activated after 400 seconds for 20 seconds.
• The acquisition stops after 450 seconds.
• After finishing data acquisition, the printer creates a data output sheet as defined by the Result
Workshop template (*.irw) selected. The results are exported to a spreadsheet file if defined in
Figure 120.

106

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6

Using EA IsoLink IRMS System for CNSOH
H3-Factor Determination

• The chromatogram and result grid are shown in Figure 121, Figure 122 and Figure 123:
Figure 121. Dual Measurement Chromatogram

The H2 page displays information about the hydrogen-related peaks, see Peaks Nr. 1, 2 and 3 in
Figure 122
Figure 122. Dual Measurement H2 Result Grid

The CO page displays information about the CO-related peaks. See Peaks Nr. 4, 5 and 6 in
Figure 123.
Figure 123. Dual Measurement CO Result Grid

H3-Factor Determination
Protonation reactions in the ion source result in H3+-ion production. The H3+ portion of the m/z ion
beam is determined as the H3-factor. The H3-factor is used to correct the H3+ contribution to the
m/z3 signal. A low and stable H3-factor is needed for a good DH/H2 determination.
H3-factor determination is based on the peak area information of a “standard gas on/off
chromatogram” (*.dxf ) with standard gas pulses of different amplitudes. How to perform a H3-Factor
Determination is explained in Chapter 3 of the ConFlo IV Operating Manual.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

107

6

Using EA IsoLink IRMS System for CNSOH
Measuring Sulfur Isotopes

Measuring Sulfur Isotopes
This section provides the instructions for measuring sulfur isotopes.

Introduction
In comparison to isotope ratio measurements of nitrogen (15N) and carbon (13C) in organic and
inorganic matter, the analysis of sulfur (34S) has always been more challenging.
Analyzing biological sulfur makes experimental difficulties caused by the low abundance of sulfur in
organisms (for example 0.2 wt% [mg/mg] in plants), as well as by the fact that sulfur is present as a
mixture of organic and inorganic compounds.
Difficulties in using the Elemental Analyzer arise due to a large amount of carbon in the same samples,
which quickly exceeds the capacity of combustion reactors. This causes incomplete combustion.
Thermo Fisher Scientific has developed a technique for precise, accurate and fast sulfur measurement,
which puts it on a par with carbon and nitrogen in terms of ease of use and sample size.
Due to the high natural abundance of the heavier isotope 34S, less amplification is required and may
become necessary for all Delta IRMS before Deltaplus XP (since 2002).
It is recommended to use a smaller resistor (1 * 1010 Ω) on the cup for mass 66 (usually 3 * 1010 Ω).

Procedure
Sulfur measurements are performed using a specially equipped Elemental Analyzer. Combustion and
reduction are carried out in a single reactor filled with tungsten oxide (WO3) and copper (Cu) as
reducing agent.
The technique used for sulfur determination is based on the quantitative Dynamic Flash Combustion
method. The samples - sometimes together with vanadium pentoxide (V2O5) - are wrapped in tin
capsules and placed into the autosampler. Then they are continuously purged with helium to remove
any traces of water and nitrogen. When a sample is dropped into the reactor, the helium stream is
temporarily enriched with pure oxygen.
The sample and its container melt as the tin promotes a violent reaction flash combustion). Under
these favorable conditions, even thermally resistant substances are completely oxidized.
In the reactor, for example barium sulfate (BaSO4) is thermally decomposed within a tin capsule.
The following reactions can then take place. See Bailey, S.A. and Smith, J.W., 1972):
BaSO 4 → BaO + SO 2 + 1--- O 2
2
BaSO 4 → BaO + SO 3
1
SO 3 → SO 2 + --- O 2
2
Note Bailey, S.A. and Smith, J.W. (1972): An improved method for the preparation of sulfur

dioxide from barium sulfate BaSO4) for isotope ratio studies. Anal. Chem. 44, 1542-1543.
Although the process does not require oxygen (O2), better combustion has been experienced when
oxygen (O2) is injected and vanadium pentoxide (V2O5) is added to the sample. If either the oxygen
(O2) pressure is low or a bad catalyst is selected or the reactor has too much ash, combustion will
proceed slowly.

108

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6

Using EA IsoLink IRMS System for CNSOH
Measuring Sulfur Isotopes

A slow stream of sulfur dioxide (SO2) through the system causes adsorption at the tubing wall.
Note It is important that enough reduced copper (Cu) is present in the combustion tube.
If this is not assured, sulfur trioxide (SO3) will only be reduced partially and isotope fractionation will
occur.
Light isotopes (32SO3) are reduced more easily than heavy isotopes (34SO3). Therefore, sulfur
dioxide (SO2) gas is depleted in 34S compared to the original sample, and the δ value becomes more
negative.

Sulfur Measurement Kits
To perform a sulfur measurement your Elemental Analyzer needs to be modified. The required
hardware is available in a kit. This kit contains amongst some helpful chemicals and spare parts:
• a dedicated reactor
• a dedicated separation column for S analysis
• a special tubing to minimize water adsorption
Table 22. Parts of Sulfur Measurement Kit for ConFlo (P/N 115 7100)
Quantity

Description

Part Number

1

Attachment for exhaust tube

112 1390

1

Self-adhesive heating foil for ConFlo

114 1180

1

Power supply for self-adhesive heating foil for ConFlo

204 8580

WARNING When working with sulfur dioxide (SO2), good ventilation is essential. Otherwise, the
gas can be hazardous to your health.
WARNING TOXIC SUBSTANCES HAZARD: To ensure operational safety, a CO, SO2 and H2
detector with an alarm must be installed. The exhaust tube should be installed on top of the ConFlo
interface to remove the toxic sulfur dioxide (SO2) from inside of ConFlo out of your working area.
See Figure 5 on page 8.

Preparing the System for a Sulfur Measurement
 To prepare the system for a sulfur measurement

1. Ensure proper ventilation by connecting the exhaust to an outlet.
2. Ensure heating of the pressure regulator for SO2.
3. Install the SO2 GC column.
4. Installing the Self-Adhesive Heating Foil when using a ConFlo III.
Note SO2 is liquid at higher pressures. Therefore, the manometer should be heated to avoid
condensation of SO2. The temperature must be between 60 °C and 70 °C. It may vary with
ventilation.
a. Remove the backing paper from the self-adhesive heating foil.
b. Paste the heating foil in the middle of the manometer’s rear side (of Ref 2).

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

109

6

Using EA IsoLink IRMS System for CNSOH
Before Starting a Sulfur Measurement

5. Ensure the inlet heater and the ion source are heated.
6. Install the properly packed reactor within the Elemental Analyzer.
7. Replace all stainless steel tubing with Teflon®/Sulfinert® tubing to avoid water condensation in the
system.
Note Take care that the wires of the heating foil point downwards when pasting it on the
manometer’s rear side.
8. Simply insert the wires into the plug socket where they fix themselves.

Reactor Filling
We recommend to use a Thermo Fisher Scientific ready for use reactor (P/N 468 02021). The reactor
is 470 mm long. If the reactor must be packed by the user, the packing should be as shown in
Figure 124.
Figure 124. Packing of the Reactor

10 mm

Quartz Wool

60 mm

Tungsten Oxide

20 mm

Quartz Wool

130 mm

30 mm

Electrolytic Copper

Quartz Wool

About 150 to 200 analyses can be performed using the reactor type described above.

Before Starting a Sulfur Measurement
We assume that the user already has working experience with the isotope ratio mass spectrometer.
Before starting a sulfur measurement make sure that:
1. The Elemental Analyzer is set up, that is:
• The SO2 reactor is packed and installed.
• The water trap is installed.
2. The special SO2 stainless steel (or Teflon®) GC column is installed.
3. The connection between reactor, water trap, GC column and ConFlo is made up of Teflon®
tubing.

110

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6 Using EA IsoLink IRMS System for CNSOH
Create a Gas Configuration for a Sulfur Measurement

4. The gases needed (He, O2) are available and have been connected to the corresponding positions.
5. Leak-check all connections outside of ConFlo by brushing all fittings carefully with soapsuds.
6. The exhaust tube is connected to the ConFlo and ventilation is working.
7. SO2 is connected to the ConFlo.
8. The pressure regulator for SO2 is heated.
9. The IRMS has been calibrated for SO2 measurements.
10. The Elemental Analyzer is switched on.
11. Adjust the gas pressure of Elemental Analyzer and ConFlo. Take the following settings as a
guideline.
• “Setting for Flash Elemental Analyzer” on page 111

Setting for Flash Elemental Analyzer
Table 23. Setting for Flash Elemental Analyzer in EagerSmart software
Carrier

180 mL/min

Oxygen

25 mL/min

Reference

20 mL/min

Cycle run time

1000 s

Sampling delay

10 s

Oxygen Injection End

3 s (or higher, depending on sample size and sample type)

12. Increase reactor temperature to 900 °C. After 10 minutes at this temperature, increase to 1020 °C.
Alternatively, use the HeatUp function in the context menu of the Flash IRMS peripheral
visualization
13. After 10 minutes at this temperature, increase to 1020 °C. Increase column temperature to 100 °C.
14. It is recommended to keep the reactor over night at end temperature to stabilize the conditions.
Note Teflon® tubings are not absolutely tight against atmosphere. Therefore, the backgrounds of
argon and nitrogen are higher than those of carbon and nitrogen measurements. Nevertheless, take
care of leaks.
Note When using reduced amplification (i.e. resistor of 1×1010 Ohm on cup for mass 66), the
background values will be three times higher.
Note Background values may vary depending on sensitivity and focus settings.

Create a Gas Configuration for a Sulfur Measurement
A Gas Configuration determines a combination of masses, which are collected in the cups, for
evaluation of ratios and eventually δ values. The Gas Configuration is specific for the particular gas and
is combined with a magnet field value taken from the mass calibration of your IRMS. The ratio groups
determine the reported ratios of predefined masses.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

111

6

Using EA IsoLink IRMS System for CNSOH
Create a Gas Configuration for a Sulfur Measurement

Prior to defining this Gas Configuration ensure that the connected IRMS has the cups for the
simultaneous detection of masses 64 and 66 and mass calibration for these cups has already been
performed.
For a 34S measurement, a Gas Configuration must be available for the masses 64 (i.e. 34S16O16O) and
66 (i.e. 32S16O16O). Otherwise, it must be created as follows.
 To create a gas configuration for sulfur measurement

15. Open the Acquisition module.
16. Open theGas Configuration Editor. It is only available if no acquisition is running.

• Per default, the Gas Configuration CO2 is being created as the first one.
• If the Gas Configuration SO2 has already beencreated, it occurs in the list above.
• If the Gas Configuration SO2 has not been created yet, it does not occur in the list above.
Then, proceed as follows.
17. Add a new Gas Configuration.

a. Type SO2 for the Name.
b. Select a Gas Configuration as Template, for example CO2. In the pull-down menu, only the
already existing Gas Configurations are displayed. When creating the first Gas Configuration,
CO2 is displayed.
c. Confirm by OK.

112

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6 Using EA IsoLink IRMS System for CNSOH
Create a Gas Configuration for a Sulfur Measurement

18. Type No. If you would type Yes, this would automatically mark the template (i.e. CO2) instead of
SO2 in the Ratio Groups window below.

a. Mark SO2.
b. If Ratio Groups other than SO2are marked, clear them all.
19. Confirm by OK.
20. The new Gas Configuration SO2 appears in the list as a row of its own.
Figure 125. Creating a New Gas Configuration

21. In the Calibration column select your current calibration file.
Note Figure 125 shows a common cup configuration as used in most Delta mass spectrometers,
i.e. universal triple collector. If you have a special cup configuration, the respective masses will be
collected in other cups.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

113

6

Using EA IsoLink IRMS System for CNSOH
Starting a Sulfur Measurement

22. Fill in the correct masses (64 and 66 replace for example 44, 45 and 46) to the appropriate cups
specific for your IRMS.

When highlighting the specific gas configuration by a click on its row, the number of cups required
for measurement is displayed together with the assigned masses.

23. Select a Calibration, which is valid for the selected cups.
24. Press the Save & Close button

.

Note A value of -1 denotes unlimited

Starting a Sulfur Measurement
EA IsoLink Setting
See Table 24.
Table 24. Setting for EA IsoLink

114

Carrier

180 mL/min

Oxygen

25 mL/min

Reference

20 mL/min

Cycle run time

1000 s

Sampling delay

10 s

Oxygen Injection End

3 s (or higher, depending on sample size and sample type)

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6

Using EA IsoLink IRMS System for CNSOH
Starting a Sulfur Measurement

Defining an Isodat Method
Instrument
Figure 126. Starting a Sulfur Measurement - Instrument Tab

Time Events
Figure 127. Starting a Sulfur Measurement - Time Events Tab

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

115

6

Using EA IsoLink IRMS System for CNSOH
Starting a Sulfur Measurement

Evaluation
Figure 128. Starting a Sulfur Measurement - Evaluation Tab

Peak Detection
Figure 129. Starting a Sulfur Measurement - Peak Detection Tab

Printout
Figure 130. Starting a Sulfur Measurement - Printout Tab

Events During Acquisition
See Figure 127 on page 115 and Figure 130 on page 116.
1. Peak center procedure.

116

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

6

Using EA IsoLink IRMS System for CNSOH
Starting a Sulfur Measurement

2. First SO2 Reference Gas pulse activated at 20 s (duration: 20 s).
3. Second SO2 Reference Gas pulse activated at 90 s (duration: 20 s). It is assigned as Standard pulse
for δ value calculation. See Time column in Figure 127 on page 115.
4. EA starts at 105 s. At this time the oxygen is injected into the reactor.
5. Autosampler activated by EA after a sampling delay defined in the EA method, e.g. 10 s after
oxygen injection.
6. Sample peak appears approximately 100 s after Autosampler activation.
7. Acquisition stops at 400 s.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

117

7
Maintenance
This chapter provides information on the post-analysis operation, the operation sequences for the
current and periodic maintenance of the instrument, and for maintaining the MAS Plus autosampler.
Contents

• Post-Analysis Operations
• Maintaining the Instrument
• Current Maintenance Program
• Maintaining the MAS Plus Autosampler
• Upgrading EA for 25 mm OD Macro Reactor
• Cleaning the Instrument
WARNING When, for technical reasons, it is necessary to work on instrument parts which might
involve an hazard (moving parts, components under voltage, and so on), the authorized Technical
Service must be contacted. This type of situations can be identified because access to these parts is
possible only by using a tool. The removable protective covers bear a warning symbol suggesting to
refer to the documentation accompanying the instrument. When a maintenance operation is
performed, the operator must have received proper training to carry out specific actions.
WARNING When the instrument is switched off, consider that its does not cool down
immediately, but heat tends to concentration in the upper part of the furnaces area. It should be
made clear that it is better to cool down the furnaces first before switching off the instrument.
Switching it off means that the fan in the back does not remove the hot air concentrating at the top
and the surface thus becomes very hot. The openings provided for the chamber aeration will cause a
slow cooling of the same, which however, in the vicinity of the areas marked with the symbol “hot
surfaces”, might even reach temperatures higher than ambient temperature. Therefore in the
minutes immediately following the instrument switching off, the operator must consider this risk
and pay adequate attention during any maintenance operations following the use of the instrument.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

119

7

Maintenance
Post-Analysis Operations

Post-Analysis Operations
For maintaining high analytical performance and to reduce operating costs, we recommend you follow
these practical suggestions.
• “Putting the Instrument in Standby Mode” on page 120
• “Maintaining the Instrument” on page 122
• “Maintaining the Instrument” on page 122

Putting the Instrument in Standby Mode
When the work session is over, the instrument can be placed into Standby mode. This mode is
intended to reduce the gas consumption significantly and will reduce operating costs. The standby
conditions can be defined by the operator.
There are these options:
• Reduce the flow only — This option reduces the carrier and reference flow to 10 mL/min and the
oxygen to 0 mL/min.
• Define an EA method — This option allows to define settings in an EA method which are
uploaded to the EA.
The Standby function can be activated immediately or automatically at the end of the analytical
sequence.
 To set the EA in standby immediately

1. Right-click on the EA window in Isodat Software Suite and open EA Standby options. When
clicking this button it opens the following page. See Figure 131.
Figure 131. Standby option for EA

Here you decide which standby properties you want to use when setting the EA in standby.
• Reduced gas flows (default)

120

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

7 Maintenance
Post-Analysis Operations

• Upload user defined method — This method can be found in the following folder:
C:\Thermo\Isodat NT\Global\User\ConFlo IV Interface\EA IsoLink Device\Method\Flash Default

The method can be modified to your purposes. Alternatively, it can be created a new by
renaming the given EA file name (EA Stdby Setting.eam)
Note If the method does not exist in the given folder Isodat Software Suite uses the default
settings for the EA (reduced gas flow options)
 To automatically set the standby function after sequence end

In the Sequence Options of an Isodat Software Suite Software Suite sequence there are two boxes.
See
Figure 132. Sequence Standby Options

• ConFlo IV Standby after Sequence — Reduces the helium flow in the ConFlo IV to a
minimum required to protect the ion source from ingress of air through the Open Split. The
following action are taken:

Thermo Scientific

–

The ConFlo is set into LF Mode.

–

The sample dilution is set to V1.

–

All reference gases are closed.

–

Reference dilution is set to zero.

EA IsoLink IRMS System for CNSOH Operating Manual

121

7

Maintenance
Maintaining the Instrument

• Flash Standby after Sequence — When selected, Isodat reduces the flows of the EA (default)
or uploads the EA Stdby Settings.eam to the EA if the box is checked in the EA options. For
this you must click on the options button in the Flash IRMS visualization and check the box
under (2). See Figure 133.
Figure 133. DynExternals

Maintaining the Instrument
Note The instrument will be generally serviced by Thermo Fisher Scientific authorized technical
personnel for all the warranty period or, after warranty, possibly according to a Programmed
Service Contract. For more information contact your local Thermo Fisher Scientific Office.
• Current Maintenance — Replacement of reactors and adsorption filters and their filling materials.
For instrument configurations using special steel reactors for combustion it may be necessary also
to clean and remove the ashes from the crucible done with same material. See “Current
Maintenance Program” on page 122.
• Periodic Maintenance — Replacement of the gas chromatographic column. The column lifetime
in EA IsoLink IRMS System for CNSOH instrument is evaluated in years. Replacement of the
seals of the reactors coupling unions placed on the furnace compartment base.
Note For some maintenance operations, furnaces and oven need to be at room temperature.

Current Maintenance Program
Each reactor, each filter and relevant fillings, need to be replaced according to the analytical
configuration used. Please refer to the CookBook for the best analytical setup for your samples.

Replacing Reactors and Adsorption Filters
The replacement of reactors and adsorption filters is performed after a preset number of analyses
according to the setting entered in the section “Replacing the Filling Materials” on page 123. Replace
and install reactors and filters.

122

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

7 Maintenance
Current Maintenance Program

Replacing the Filling Materials
The replacement of reactors and adsorption filters requires the replacement of their filling materials.
This operation comprises two steps. Alternatively, prepacked reactors can be obtained via your local
Thermo Fisher Scientific Office.
• Removing the exhausted filling material from the reactor.
• Restoring the sequence of the layers of filling materials using new reagents.
Perform these operations according to the instructions given in the following operating sequences.
• “To replace the filling material in quartz reactors” on page 123
• “To replace the filling material in adsorption filters” on page 124
 To replace the filling material in quartz reactors
Material required

Tool for cleaning quartz reactors P/N 27606010 (included in the Standard Outfit of all
EA IsoLink IRMS System for CNSOH Configurations)
Filling materials depending on analytical setup. Consult the CookBook for different reactor
setups.

CAUTION Before starting the operation, check that the furnaces are at room temperature.

Remove the quartz reactor from the furnace, then do the following:
1. Introduce the cleaning tool into the reactor as shown in Figure 134.
Figure 134. Removing the Filling Material from a Quartz Reactor

2. Rotate the tool exerting a slight pressure to scrape off the filling material.
3. Collect the removed filling material as shown in Figure 135.
4. Repeat steps 1 and 2 until complete elimination of the exhausted filling materials.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

123

7

Maintenance
Current Maintenance Program

Figure 135. Collection of the Material Removed from a Quartz Reactor

5. At the end of the operation restore the layers of filling materials introducing into the reactor the
new ones.
 To replace the filling material in adsorption filters
Material Required
Filling materials

Remove the adsorption filter from the detector compartment, then do the following:
1. Unscrew the filter nut and remove the filling material.
2. Restore the sequence of the layers of filling materials introducing the new ones into the filter.

Ashes Removal and Crucible Cleaning
 To remove the ashes and clean the crucible

CAUTION Perform the operation with the furnaces at room temperature.

Materials Required
Tool for cleaning quartz reactors P/N 27606010
Quartz wool

Remove the crucible from the reactor.
If the crucible is made of quartz, remove the ashes by using a spatula.
If the crucible is made of steel then do the following:
1. Introduce the cleaning tool into the crucible as shown in Figure 136.

124

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

7 Maintenance
Current Maintenance Program

Figure 136. Removal of Ashes and Quartz Wool from the Crucible

2. Rotate the cleaning tool exerting a slight pressure in a way to scrape off the ashes and quartz wool,
then collect the material removed. At the end of the operation, introduce new quartz wool into the
crucible.

Maintaining the Reactor of the Pyrolysis Unit
Note Please refer also to the section “Installing the Reactors into the Furnaces” on page 56.
 To clean the reactor

1. Take out the glassy carbon tube.
2. Remove the crucible, silver wool and quartz wool. Empty the granules in a separate bowl.
3. Clean the inside of the glassy carbon tube with the bottle brush.
4. Insert new quartz wool and silver wool.
5. Clean the granules with a tissue or a sieve until they are shiny again.
6. Remove all granules that are not shiny anymore or show a porous surface.
7. Clean the outside of the crucible with a tissue, and remove the molten silver inside or use a new
one.
 To clean/replace the graphite crucible

Clean or replace the graphite crucible after approximately 400 analysis, according to sample nature and
samples weight.
ATTENTION The use of gloves is strongly suggested. The graphite crucible is very hot after removal
from the reactor.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

125

7

Maintenance
Current Maintenance Program

Figure 137. Graphite Crucible Removal

Crucible Remover Tool

Graphite Tube

Graphite Crucible

A

B

C

1. Cool the furnace to a temperature below 600 °C.
2. Put the ConFlo in standby or close needle valve to IRMS.
3. After reaching the temperature switch Off the flow of helium. Wait approximately five minutes to
depressurize the system.
4. Remove the autosampler.
5. Carefully remove the graphite tube (for example with a pair of tweezers), see A of Figure 137, and
place it on a clean cloth.
6. Insert the crucible remover inside the graphite crucible. See B in Figure 137.
7. Pull the graphite crucible out of the reactor. See C in Figure 137.
8. Drop a new graphite crucible into the reactor, or remove the deposits outside and inside of the
graphite crucible and use it again.
9. Remount the parts proceeding in the reverse order.
 Maintenance measures for pyrolysis unit

To facilitate proper work, adhere to the described maintenance recommendations and procedures.
This will also increases the life time of the pyrolysis reactor and the furnace heater.

126

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

7 Maintenance
Current Maintenance Program

CAUTION As the lifetime of the furnace heater is limited, do not heat it to higher temperatures
than necessary. Put it to a lower standby temperature, when not in use for several days, or turn it off
for longer idle times.
• To measure:
Type of analysis

Suggested operating temperature

Oxygen of organic samples

1325 °C - 1350 °C

Oxygen of inorganic samples

1450 °C

Hydrogen of organic samples

1450 °C

Hydrogen of inorganic samples

1450 °C

Water (sample amount > 1 μL)

1400 °C

• To cool down the furnace:
If you do not want to perform any analysis for about one day (for example, over night), put the
instrument in stand-by condition, and set the oven temperature to 150 °C.
• If you do not want to perform any analysis for more than one week, put the instrument in
stand-by condition, cool down both furnaces and oven to ambient temperatures.
• If the pyrolysis unit has not been in use for a long time or if it was off, heat the column to a
temperature of 190 oC for at least 24 hours. Make sure to maintain a flow through both
separation columns during baking out
• Always watch the background of the masses 28 and 40. If the values exceed those you were used to,
perform a leak test. A small leak will crack the ceramic tube, and after a while the glassy carbon
reactor, too.
• After around 400 measurements (depending on size and type of the samples) clean or replace the
graphite crucible as described in the “To clean/replace the graphite crucible,” operating sequence.
After cleaning the crucible may be used again. See also reactor maintenance.

Replacing the Gas Chromatographic Column
The instrument rarely requires the gas chromatographic column replacement, however, in case, operate
according to the following operating sequence.
Note To replace the GC column in the ramped GC oven, please refer to the Ramped GC Oven
Operating Manual.
 To replace the gas chromatographic column in the GC oven of the EA.
Material Required
Open end wrenches for the column fittings

1. Open the instrument right door and remove adsorption filters from their securing clips.
Note According to the instrument configuration, there can be one or two adsorption filters in
the detector compartment.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

127

7

Maintenance
Current Maintenance Program

2. Undo the 4 fixing screws securing the protection panel. See Figure 138.
Figure 138. Access to the Detector

Fixing Screws

Securing Clips for
Adsorption Filters

Protection
Plate

Figure 139 shows the detector compartment, the heating block where the detector is housed, and
the gas chromatographic column.
Note According to the instrument configuration there can be one or two analytical columns.
Figure 139. Compartment Housing the Chromatographic Column

Heating Block
TCD Detector

Chromatographic Column

3. Unscrew the fittings from the column ends and remove the column from the compartment.
4. Introduce the new column and connect its ends to the fittings.
5. Re-mount the protection panel and the adsorption filters.

128

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

7 Maintenance
Maintaining the MAS Plus Autosampler

Maintaining the MAS Plus Autosampler
The MAS Plus autosampler does not normally require maintenance. However, when the instrument is
extensively used, it is a good practice to clean from time to time the shaft housed in the autosampler
operating as described in the following operating sequence.
 To clean the shaft of the MAS Plus autosampler
Materials Required
Phillips screwdriver
Cloth
Silicon grease (For use at pressures down to 10-6 mm Hg)

1. In Isodat Software Suite, right click in the Flash IRMS dialogue box and then select settings for
EA.
2. In EagerSmart Data Handling Software Main Menu Edit | Edit Elemental Analyzer parameters,
or just click the

icon. The Analyzer Parameter window is visualized.

3. Click Send and wait 2-3 minutes to discharge the gas out the circuit.
4. Remove the frontal cover of the MAS Plusautosampler as shown in Figure 140.
Figure 140. Removal of the MAS Plus Autosampler Cover

Cover

5. From Main Menu select the menu View | View Elemental Analyzer status, or click the icon

.

6. In the Status window select the Special function tag, then click Step sampler tray position.
The autosampler mechanism pushes the shaft forward and then ejects it.
7. Take out the shaft from the autosampler, as shown in Figure 141.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

129

Figure 141. Removal of the Shaft

8. Eliminate possible traces of dirt from the shaft seals using a dry clean cloth.
9. Place and smear a slight layer of silicon grease on the o-ring. Do not use solvents.
Figure 142 shows the shaft of the MAS Plus autosampler.
Figure 142. Shaft of the Autosampler
O-rings

Rack

10. Re-introduce the shaft into the autosampler until in place keeping its rack turned downward, as
shown in Figure 143.
Figure 143. Reinstalling the Shaft (1)

11. Slightly pushing the shaft with one finger of your hand, as shown in Figure 144, click Step
Sampler tray position again. The autosampler mechanism first tries to eject the shaft, then, on the
motor reversal, the mechanism will hook the shaft and draw it inside the autosampler (2).
Figure 144. Reinstalling the Shaft (2)

7 Maintenance
Maintaining the MAS Plus Autosampler

12. Re-install the autosampler front cover.
13. Return to Analyzer Parameter window. Restore the operating conditions setting the carrier gas
flow to the initial value.
 Replacing the o-rings of the MAS shaft
Materials Required
Tweezers or spatula
Green O-rings package (set of 3) for MAS Plus shaft (P/N 290 30343)
MAS O-ring Removal Tool (P/N 205 02613)

The shaft of the MAS plus has three conical O-rings (lip seal) that must be replaced when worn-out.
The MAS O-ring Removal tool allows an easier replacement of the internal and middle O-rings into
their relevant seats.

Rack

O-rings (Lip Seals)
Internal O-ring (Lip Turned Inwards)
Middle O-ring (Lip Turned Inwards)
External O-ring (Lip Turned Outwards)

Internal Seat
Lip Turned Inwards
Middle Seat
External Seat

Lip Turned Outwards

MAS O-ring Removal Tool

1. Remove the three from the shaft body by using tweezers or a spatula paying attention to not
scratch the shaft.
2. Take the new green O-rings package and the removal tool.
3. Insert the first O-ring into the internal seat of the shaft.
a. Place one of the three O-ring, with the lip turned inwards, on the conical section of the tool.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

131

b. Move the O-ring along the tool up to reach the border.
c. Move the tool on the shaft body up to reach the internal seat.
d. Move and inset the O-ring into the internal seat, then remove the tool.

4. Insert the second O-ring into the middle seat of the shaft.
a. Place the second O-ring, with the lip turned inwards, on the conical section of the tool.

b. Move the O-ring along the tool up to reach the border.
c. Move the tool on the shaft body up to reach the middle seat.
d. Move and inset the O-ring into the middle seat, then remove the tool.

5. Insert the third O-ring into the external cavity of the shaft. The use of the tool is not necessary.
a. Manually place the last O-ring, with the lip turned outwards, into the external seat.

7 Maintenance
Upgrading EA for 25 mm OD Macro Reactor

Upgrading EA for 25 mm OD Macro Reactor
This procedure gives the instructions for replacing the fittings for the 18 mm OD standard reactor
with the fittings for the 25 mm macro reactor.

WARNING This operation must be carried out with the furnaces at room temperature.

1. If already installed, remove the MAS Plus and the reactor from the left/right furnace of interest.
2. Using the 23-33 mm open-ended wrench, unscrew and remove the introduction sample fitting
with the nut for the 18 mm OD reactor (P/N 35008417) from the MAS Plus paying attention to
do not damage the internal gasket.

Introduction Sample Fitting
Nut for the 18 mm OD Reactor Tube

3. Screw manually the introduction sample fitting with the nut for the 25 mm OD macro reactor
(P/N 35008418, provided in the MAS Plus standard outfit) into the MAS Plus. Tighten the fitting
by using the 23-33 mm open-ended wrench.
4. Access the furnace compartment lifting the cover and removing the protecting plate.

Furnace Compartment Protection Plate

Thermo Scientific

Furnace Compartment Cover

EA IsoLink IRMS System for CNSOH Operating Manual

133

7

Maintenance
Upgrading EA for 25 mm OD Macro Reactor

5. On the top panel of the instrument remove the left/right upper connector for the 18 mm OD
reactor unscrewing the two fixing Allen screws with the proper Allen wrench. Place and fix the
upper connector for 25 mm OD macro reactor.

Upper Connector for
25 mm OD Macro Reactor

Upper Connector for
18 mm OD Reactor

6. On the top of left/right furnace remove the adaptor for 18 mm OD reactor and its sealing adapter.
Adaptor for 18 mm
OD Reactor

Sealing Adapter

7. Reinstall the protecting plate and the cover of the furnace compartment.
8. Install the macro reactor tube into the furnace and reinstall the MAS Plus.

134

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

7 Maintenance
Cleaning the Instrument

Cleaning the Instrument
 To clean the instrument

WARNING Cleaning must be performed with the instrument off, the furnaces at room
temperature and the power cord disconnected.
1. Externally clean the instrument with a soap and water solution, or with a household non-abrasive
product. Avoid that any liquid seeps into the instrument.
2. If you just suspect that a substance used for cleaning or a product submitted to analysis has
infiltrated inside the instrument, immediately shut down the instrument and call an authorized
customer support engineer for proper actions.The service engineer must be fully informed on the
nature of the concerned substance.
WARNING It is your responsibility to avoid that dangerous liquids and/or materials seeping inside
the EA IsoLink IRMS System for CNSOH during operation and maintenance.
If you have any doubt about compatibility of decontamination or cleaning agents with parts of the
equipment or with material contained in, please contact your Thermo Fisher Scientific
Representative.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

135

8
Troubleshooting
This chapter provides information necessary to find out instrument troubles and to solve them.
Contents

• Analytical Troubleshooting
• Safety Cutoff
• EFC-t Module

Analytical Troubleshooting
If the instrument has been correctly installed, the gas characteristics are as required and maintenance
has been regularly carried out, EA IsoLink IRMS System for CNSOH will provide correct results.
The lack of the above conditions will be indicated by anomalies in the chromatograms and the relevant
analytical reports. Table 25 reports the most common anomalies with the relevant diagnosis and
remedy.
Table 25. Analytical Troubleshooting Guide (Sheet 1 of 2)
Problem

Diagnosis

Remedy

High nitrogen blank.

Presence of leak.

Check that helium and oxygen lines
are sealed and in case eliminate
possible leak.

Oxygen line or cylinder contaminated. Purge for 10-20 minutes.
Replace the contaminated cylinder.
Autosampler not purged.

Check that the helium flow is correct.

High constant nitrogen blank in
several sequential analyses.

Oxygen cylinder contaminated.

Replace the oxygen cylinder.

Presence of leak in the autosampler
system.

Identify leaks and remove them.

Carbon peak tailing or split.

Too much ashes inside the reactor.

Check ashes and remove them.

The sample analyzed was too large.

Weigh a lower amounts of sample.

The tube connecting reactor and
column is clogged.

Cut off the clogged tube portion.

Hydrogen peak is a split peak.

Bad separation between nitrogen and High nitrogen blank value.
carbon peaks.
Copper exhausted.

Thermo Scientific

Check the nitrogen blank value.
Eventually repeat the analysis.
Replace the reactor.

EA IsoLink IRMS System for CNSOH Operating Manual

137

8

Troubleshooting
Safety Cutoff

Table 25. Analytical Troubleshooting Guide (Sheet 2 of 2)
Problem

Diagnosis

Remedy

Peak between nitrogen and carbon
peaks.

Oxygen line contaminated.

Exclude autosampler and check the
oxygen blank.

Inadequate oxygen purity.

Use Oxygen with adequate purity.
Exclude autosampler and check the
oxygen blank.

Inadequate oxygen dosage.

Increase oxygen injection time.

Tin containers contaminated.

Check the tin container box, tweezers,
work bench are clean.

High carbon blank.

Memory effect due to bad combustion Remove ash and analyze lower
of previously run analyses.
amounts of sample.
Decreasing nitrogen blank values.

Oxygen line contaminated.

Wait 10-20 minutes for complete
purging of the oxygen line.
Repeat blank analysis.

Increasing nitrogen blank values.

Copper exhausted

Replace the reactor.

Retention times very delayed respect
the normal chromatogram.

Presence of leaks in the pneumatic
circuit.

Perform Leak Test.

Presence of obstructions in the
pneumatic circuit.

Reach and remove the obstruction
dissecting the pneumatic circuit.

Safety Cutoff
Instrument malfunctioning, due to a component failure or to abnormal operating conditions, is
identified by the red lighting of the Safety Cutoff LED indicator. See Figure 145
Figure 145. Safety Cutoff LED

Safety Cutoff LED

When lit, this LED indicates that the furnaces and detector oven power has been cut off for safety
reasons.
The Safety Cutoff status is followed by an error message about the possible cause of error.

138

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

8

Troubleshooting
Safety Cutoff

 To display the error message

Proceed as follows:
Proceed as follows:
1. In Isodat Software Suite, right click in the Flash IRMS dialogue box and then select settings for
EA.
2. In the Main Menu for IRMS click the icon

.

3. In the visualized dialog window, select the Special Functions tabbed page. The dialog window of
Figure 146 is visualized.
4. Read the error message in the reading box located on the lower right side of the window, below the
buttons Help and OK.
5. Refer to Table 26 to find out the error status and have mode information.
Figure 146. Special function window

The following Table 26 reports the error messages and the explanation of the relevant correlated
problem.
Table 26. Error Messages (Sheet 1 of 2)
Message

What to do

Under voltage protection Voltages supplied to the electrical
circuits are too low or out of
(Safety Cut Off message)
tolerance.

Check all the voltages and main
power.

Oven over limit

The oven temperature exceed
190 °C or does not reach the
setting temperature.

Check the functionality of the PT
100 probe. Replace the HWD
1112 board

The left thermocouple is damaged
or interrupted.

Verify the continuity of the
thermocouple and replace it.

(Safety Cut Off message)

Left furnace over limit
(Safety Cut Off message)

Thermo Scientific

Description

EA IsoLink IRMS System for CNSOH Operating Manual

139

8

Troubleshooting
EFC-t Module

Table 26. Error Messages (Sheet 2 of 2)
Message

Description

What to do

Right furnace over limit

The right thermocouple is
damaged or interrupted.

Verify the continuity of the
thermocouple and replace it.

Anomalous temperature inside the
transformers compartment.

Check the functionality of the fans.
Replace the AS 1112 board.

(Safety Cut Off message)

Thermal pre-protection
(Safety Cut Off message)

Oxygen pressure too low If the oxygen pressure is too low
Verify that the pressure at the
(<150 kPa) the flow of oxygen does manometer is at second stage
(NO Safety Cut Off message) not reach the set point.
higher or equal to 350 kPa.
CAUTION The error message is generally due to the specific cause indicated. Sometimes, it may
generated by different electric factors or caused by failures not depending on the system. In this case
contact the Technical Service.

EFC-t Module
The failures that may be generated on the EFCt Module are connected to the breakage, or to the
malfunctioning of solenoid valves and flow sensors.
Refer to the Table 27 to find the component responsible of the EFC-t module malfunctioning and to
solve the relevant problem.
Table 27. EFC-t module troubleshooting
Failure

Defective Component

Oxygen does not flow to the point 2 EV1
of the autosampler

Check voltage supply
Replace the solenoid valve

EV2

Check voltage supply
Replace the solenoid valve

The Helium flow measured on point Flow sensor 1 or 2.
1 or 2 cannot be adjusted
EVP1 or EVP2
The pneumatic circuit is perfectly
close but the flow value do not
decrease up to zero performing the
Leak Test.

140

Remedy

EA IsoLink IRMS System for CNSOH Operating Manual

EV3 and/or EV4

Replace flow sensor
Check voltage supply
Replace the solenoid valve
Check voltage supply
Replace the solenoid valve

Thermo Scientific

9
Running the Flash IRMS as Stand-alone Instrument
This chapter provides instruction for installing and configuring the EagerSmart Data Handling
Software.
Contents

• EagerSmart Software Installation
• Configuring the Analyzer
• Performing a Leak Test
• Adjusting the Detector Signal Level
• View Sample Being Acquired
• Configuring EagerSmart Data Handling Software
• Using Isodat Software Suite and EagerSmart Data Handling Software at
the Same Time

ATTENTION If you want to operate with the EA IsoLink CNSOH as a stand-alone instrument, you
need to install EagerSmart software on you computer.
ATTENTION EagerSmart does not support the HeM. It is thus required that the HeM valve block is
bypassed. Please refer to the proper Technical Note.

EagerSmart Software Installation
EagerSmart is compatible with commercially available computers and Windows® 2000, XP, Vista, 7, or
8 operating system
The required software package includes the following items:
• USB stick containing EagerSmart Data Handling Software
• EagerSmart Data Handling Software Manual
 To install EagerSmart Data Handling Software
Material required

EagerSmart Data Handling Software package

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

141

9

Running the Flash IRMS as Stand-alone Instrument
EagerSmart Software Installation

EagerSmart Data Handling Software is the dedicated software that fully controls all the operations of
the EA IsoLink IRMS System for CNSOH, Flash 2000, Flash 4000, and EA 1110 Elemental
Analyzers. It is designed to be compatible with commercially available computers, and requiring the
use of Windows™ 2000 / XP / Vista / 7 / 8 operating system. The free space on the PC hard disk must
be at least 1 GB.
EagerSmart Data Handling Software is installed by using the USB stick provided in the standard outfit
kit and operating as follows:
IMPORTANT Prior installation of EagerSmart Data Handling Software make sure any precedent
version of Eager software is removed from the disk.

 To install EagerSmart Data Handling Software

1. Remove a precedent version of Eager software installed on your computer.
a. Select Control Panel | Add/Remove Programs.
b. In the dialog window visualized, select the precedent Eager version to remove.
c. Click Add/Remove.
2. Install the new version of EagerSmart Data Handling Software. When the USB stick is introduced
in a free USB port of the computer, the following installation window is visualized. See Figure 147.
Figure 147. EagerSmart Data Handling Software Installation Menu

Note If when the USB stick is introduced in a free USB port of the computer, the installation
menu does not automatically appear, through the Windows command Start-Run, start the
program Autorun on the USB stick.
3. Start installation by clicking the push-button

Install EagerSmart for Flash.

4. Follow the instructions prompted step by step.
5. At the end of installation, in the page Start-Program EagerSmart, double-click the EagerSmart
for Flash icon. The window of Figure 148 is visualized.

142

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

9

Running the Flash IRMS as Stand-alone Instrument
Configuring the Analyzer

Figure 148. Selection of the instrument

6. Click the icon of the instrument selected.The program is designed to work with four instruments.
Each icon corresponds to one instrument. The instrument name shown below the icon can be
changed. To do this, click on the existing name and overwrite the new one.
7. EagerSmart Data Handling Software proceeds with the registration and the activation of some
drivers needed for the correct functioning of the software.
a. Click Ok to the answers prompted step by step.
b. At the end of the operation, reboot the computer. Start EagerSmart Data Handling Software
again selecting Start | Programs | EagerSmart.
8. Follow the prompted indications. At the end of the installation, the Main Menu is visualized.

EagerSmart Main Menu
The Main Menu of EagerSmart Data Handling Software, shown in Figure 149, is the starting point to
enter all menus and relevant functions.
Figure 149. EagerSmart Data Handling Software Main Menu

Configuring the Analyzer
The analytical conditions are set in our laboratories during the final test of the analyzer. To put the
analyzer in operating conditions, you only have to follow the instructions reported in the next
operating sequence “To configure the analyzer” on page 143.
 To configure the analyzer

1. In Main Menu of Figure 149 on page 143, choose File | Instrument Name and Configuration.
The following window is visualized. See Figure 150.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

143

9

Running the Flash IRMS as Stand-alone Instrument
Configuring the Analyzer

Figure 150. Instrument Name and Configuration

a. In the field Instrument name, type the instrument serial number (6 digits; for example
991234). See the label located on the instrument rear panel.
b. In the field Analytical Instrument Configuration select Undefined.
c. In the field IRMS & Argon select Instrument control for Flash EA IRMS (NC) and HT.
d. Click OK. The system ask to exit and reboot EagerSmart.
e.

At the reboot of EagerSmart, the simplified Main Menu for IRMS is visualized. See
Figure 151.
Figure 151. Main Menu for IRMS

Menus and buttons of the Main Menu for IRMS are described in the following Table 28 and
Table 29 respectively.
Table 28. Main Menu for IRMS: Description of Menus

144

Menu

Description

File

Monitors the analysis in real time.

• Instrument name and configuration
• View sample being acquired
• Exit EagerSmart

Tools

Used when the ashes removal, the
reactor replacement, or both, are
required as maintenance.

• Ashes removal
• Reactor replacement

Help

Accesses to the help system of
EagerSmart.

• About EagerSmart

EA IsoLink IRMS System for CNSOH Operating Manual

Sub-menus and Options

Thermo Scientific

9

Running the Flash IRMS as Stand-alone Instrument
Configuring the Analyzer

Table 29. Main Menu for IRMS: Description of Icons (Graphic Buttons)
Icon

f.

Function

Description

Edit elemental analyzer
parameters

Gives the access to the three pages containing the commands
for setting temperatures, flows, times, detector, and the
analyzer control functions.

Elemental analyzer status

Comprises four pages displaying the analyzer conditions.

In Main Menu for IRMS select File | Instrument name and configuration and click
Elemental analyzer setup for visualizing the dialog window of Figure 152, where the
configuration parameters must be set.
Figure 152. Configuration Dialog Window

g. In the field Elemental Analyzer Connection select the computer serial port (COM1, COM2,
and so on), to which the instrument is connected.
h. In the field Instrument Settings choose the following settings:
• Line Frequency = 50 Hz
• TCD Settings Source = Internal
• TCD Settings Polarity = Positive
i.

In the field Sampler Setting select the type of autosampler installed on the instrument.
• In the case of autosampler for liquid samples, also specify the serial port of the computer
to which the autosampler is connected, and the number of vials.
• Click Ok to go back to the window of Figure 152, then click Ok to return to Main Menu
for IRMS.

2. In Main Menu for IRMS click the icon
. The following window appears where the analyzer
operating parameters are visualized. See Figure 153.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

145

9

Running the Flash IRMS as Stand-alone Instrument
Configuring the Analyzer

Figure 153. Example of Analyzer Parameters

a. Click Send to transfer the operating parameters to the instrument. From now on the analyzer
is working.
b. The furnaces begin to heat, and the helium flows in the circuit. After about 50 minutes the
furnaces reach the temperature settings, and the LED Ready on the status panel lights up.
The instrument is now ready to run analyses. However, before starting an analytical cycle, a
leak test must be carried out for checking the Carrier and Reference pneumatic circuits are
free of leaks.
Note The leak test must be performed also any time a component of the pneumatic
circuit is replaced for checking that reactors, filters, if any, and gas chromatographic
columns have been properly installed. See the section “Performing a Leak Test” on
page 147 for details.

146

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

9

Running the Flash IRMS as Stand-alone Instrument
Performing a Leak Test

Performing a Leak Test
For performing the leak test follow the operating sequence “To check the leaks” on page 147.
 To check the leaks

1. In Main Menu for IRMS click the icon
Figure 154.

. The following dialog window is visualized. See

Figure 154. Analyzer Status Page

2. Select Special Functions tabbed page. See Figure 155.
Figure 155. Special Function Tab

3. In the field Command click Leak Test. The window of Figure 156 is visualized indicating the
status of the leak test.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

147

9

Running the Flash IRMS as Stand-alone Instrument
Adjusting the Detector Signal Level

Figure 156. Leak Test Status Window

a. Click Start to begin the leak test. The system automatically selects Carrier and Reference gas
outlet closed check boxes. The system will ask if the zero of flow controllers has to be
calibrated; reply Yes to this question. This operation will take place in less than one minute.
b. Gas outlets are closed by the solenoid valves. Wait for some time (see Leak test time),
according to the instrument configuration, to let the gas circuit reach the equilibrium pressure.
From the values of Carrier Flow and Reference Flow must be within 0 and 3 mL/min.
Higher values indicate that the system is not leak-free.
Note Leaks in the system are generally due to incorrect closure of the reactors and filters
locking nuts. Rarely, leaks may be due to the autosampler.
c. Click Stop and Done for ending the leak test and restoring the flow operating values.

Adjusting the Detector Signal Level
To adjust the level of the TCD detector’s signal, follow the instructions reported in the operating
sequence “To adjust the level of the detector signal” on page 148.
 To adjust the level of the detector signal

1. In Main Menu select View | View Elemental Analyzer Status, or just click the icon
select the Detector tabbed page. See Figure 157.

, then

Figure 157. Detector Status Tab

2. In the field TCD, click Auto-Adjust Level at 1000 μV. At the end of the operation, the value
1000 is set representing the analysis starting point.

148

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

9

Running the Flash IRMS as Stand-alone Instrument
View Sample Being Acquired

View Sample Being Acquired
This section provides instruction for monitoring the sample currently being acquired.
View Sample Being Acquired is a tool to monitor the sample currently being acquired. See
Figure 158.
Figure 158. View Sample Being Acquired Page

Note This page shows the real chromatogram being acquired.
Icon

Description

See on:

Fit to Higher Peak

Icon

Description

See on:

page 149

Step Up

page 150

Expand (scale/2)

page 150

Step Down

page 150

Shrink (Scale*2)

page 150

Run Menu — gives access to the following functions:
• Start Single Sample Data Acquisition — Starts the acquisition of single sample of the instrument
currently in use. All parameters set in Detection Parameters and Integration Parameters are used
for starting the acquisition.
• Stop Data Acquisition — Stops the acquisition of the instrument currently in use.
The system, after a confirmation, stops the acquisition saving the chromatogram data up to the
moment of the stop command, and a report is generated if required.
• Exit Monitor Detector — Exits from detector monitoring. The program returns to the Main
Menu, but the acquisition continues in the background.
View Menu — Gives access to the following functions:
• Fit to Highest Peak — Expands or reduces the chromatogram in order to fit all peaks being part of
the partially acquired chromatogram centering the signal on the window display.
This modification has effect only on the real time display, but it does not affect the integration and
the chromatogram printout.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

149

9

Running the Flash IRMS as Stand-alone Instrument
Configuring EagerSmart Data Handling Software

• Step Up — Lifts the chromatogram lowering the beginning of the scale by about 1/10 of the full
scale. The result is a lift of the chromatogram baseline. This modification has effect only on the
real time display, but it does not affect the integration and the chromatogram printout.
• Step Down — Lowers the chromatogram increasing the beginning of the scale by about 1/10 of
the full scale. This modification has effect only on the real time display, but it does not affect the
integration and the chromatogram printout.
• Expand (scale/2) — Expands the chromatogram by dividing the full scale in two. The system
automatically re-draws the entire chromatogram with the new required scale. This modification
has effect only on the real time display, but it does not influence the integration and the
chromatogram printout.
• Shrink (scale*2) — Reduces the chromatogram by doubling the full scale. The system
automatically re-draws the entire chromatogram with the new required scale. This modification
has effect only on the real time display, but it does not affect the integration and the chromatogram
printout.
• Set Manual Scale — Sets manually the values for the beginning and the full scale. You can enter
the signal on the window display or you can display the signal with a desired attenuation.
This modification has effect only on the real time display, but it does not affect the integration and
the chromatogram printout.
• Set Focus to Eager Main Menu — Sets the focus on Main Menu.
The function is only used when you wants to use the keyboard instead of the mouse. All functions,
like in all other Windows™ applications, can be accessed with Alt key. After the use of View
chromatogram being acquired, the function returns the focus on the Main Menu functions.

Configuring EagerSmart Data Handling Software
To control the EA IsoLink IRMS System for CNSOH version, it is necessary to properly configure the
Eager Smart Data Handling Software:
 To Configure EagerSmart Data Handling Software for HT

1. Exit EagerSmart Data Handling Software if active.
2. On the desktop select the directory where EagerSmart Data Handling Software is installed.

3. Double-click twice to open the relevant contents.

150

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

9

Running the Flash IRMS as Stand-alone Instrument
Configuring EagerSmart Data Handling Software

4. Note the four sub-folders named Temp 1, Temp 2, Temp 3, and Temp 4. Each of them
corresponds to the relevant instrument (Analyzer #1, Analyzer #2, and so on), that may be
configured during the EagerSmart Data Handling Software installation procedure.

5. Double-click the sub-folder name that corresponds to your EA IsoLink IRMS System for CNSOH
for opening the relevant contents. For example: Temp 1 = Analyzer #1. Double-click on the
Eager.ini file for opening the file. See Figure 159.
Figure 159. Eager.ini File (1)

6. Search Pyrolysis=0 line. Indicates the configuration of EagerSmart Data Handling Software to
control a standard EA IsoLink IRMS System for CNSOH. For controlling a EA IsoLink IRMS
System for CNSOH, modify the Pyrolysis=0 line to Pyrolysis=1. See Figure 160.
Figure 160. Eager.ini File (2)

7. Save the modification and close the file. Run EagerSmart Data Handling Software. A message
appears indicating that the instrument is now configured to operate under pyrolysis condition.

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

151

9

Running the Flash IRMS as Stand-alone Instrument
Using Isodat Software Suite and EagerSmart Data Handling Software at the Same Time

Using Isodat Software Suite and EagerSmart Data Handling Software
at the Same Time
 To install the software version

1. Make a backup of Isodat Software Suite with the Version Handler if Isodat Software Suite is
installed.
2. Make an additional backup of the User folder under Isodat Software Suite.
3. Close down Isodat Software Suite completely by right-click task bar.

4. Install EagerSmart Data Handling Software.
 To switch fromEagerSmart Data Handling Software to Isodat Software Suite

1. Before using either software make sure the other software is completely closed.
2. Open Acquisition. The OCX is automatically registered for Isodat Software Suite if EagerSmart
Data Handling Software is completely closed down first. If problems occur a Safety Cut-Off
message is visualized. Go to the Isodat Software Suite Configurator and Re-Register OCX, then
wait 10 seconds.

3. Close Configurator and start Acquisition.
 To switch from Isodat Software Suite to EagerSmart Data Handling Software

1. Before using either software make sure the other software is completely closed.
2. Start EagerSmart Data Handling Software. The OCX is automatically registered for EagerSmart
Data Handling Software if Isodat Software Suite is completely closed down. If problems occur you
may register and unregister the OCX using the two bat –files in the following folder of EagerSmart
Data Handling Software:

152

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific

9 Running the Flash IRMS as Stand-alone Instrument
Using Isodat Software Suite and EagerSmart Data Handling Software at the Same Time

EAGER For FLASH | Active X | Vers_1_x

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

153

G
Glossary
This section lists and defines terms used in this guide.
It also includes acronyms, metric prefixes, symbols and
abbreviations.

A B

C

D

E

F

G H

I

J

K

L M N O

P

Q

R

S

T

U

V W X

Y

Z

A ampere

ºF Fahrenheit

ac alternating current

FSE Field Service Engineer

ADC analog-to-digital converter

ft foot

Ar Argon

g gram

b bit

GND electrical ground

B byte (8 b)

h height

baud rate data transmission speed in events per second

h hour

C Carbon

H hydrogen

°C Celsius
cm centimeter

harmonic distortion A high-frequency disturbance
that appears as distortion of the fundamental sine
wave

CPU central processing unit (of a computer)

He Helium

CSE Customer Service Engineer

HPAR High Performance Alloy Reactor

 control key of the keyboard

HV high voltage

d depth

Hz hertz (cycles per second)

DAC digital-to-analog converter

IEC International Electrotechnical Commission

dc direct current

in. inch

DS data system

I/O input/output

EMC electromagnetic compatibility

k kilo (103 or 1024)

ESD electrostatic discharge

K Kelvin

Thermo Scientific

EA IsoLink IRMS System for CNSOH Operating Manual

155

Glossary:

kg kilogram
kPa kilopascal
l length

slow average A gradual long-term change in average
RMS voltage level, with typical duration greater than
2 s.
surge A sudden change in average RMS voltage level,
with typical duration between 50 μs and 2 s.

l liter
LAN Local Area Network

S sulfur
TCD Thermal Conductivity Detector

lb pound
LED light-emitting diode
m meter (or milli [10-3])

transient A brief voltage surge of up to several
thousand volts, with a duration of less than 50 μs.
V volt

6

M mega (10 )

V ac volts, alternating current

-6

μ micro (10 )
min minute
mL o ml milliliter
mm millimeter
m/z mass-to-charge ratio
N nitrogen
n nano (10-9)
O oxygen

V dc volts, direct current
VGA Video Graphics Array
w width
W Watt
When a unit of measure has a quotient (e.g. Celsius degrees
per minute or grams per liter) this can be written as
negative exponent instead of the denominator:
For example:
°C min-1 instead of °C/min
g L-1 instead of g/L

Ω ohm
p pico (10-12)
Pa pascal
PCB printed circuit board
PN part number
psi pounds per square inch
RAM random access memory
  key on the keyboard
RF radio frequency
ROM read-only memory
RS-232 industry standard for serial communication
s second

156

EA IsoLink IRMS System for CNSOH Operating Manual

Thermo Scientific



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : No
Author                          : giuliano.galimberti
Create Date                     : 2016:11:24 09:25:31Z
Modify Date                     : 2016:11:24 15:36:53+01:00
XMP Toolkit                     : Adobe XMP Core 4.2.1-c041 52.342996, 2008/05/07-20:48:00
Creator Tool                    : FrameMaker 2015
Metadata Date                   : 2016:11:24 15:36:53+01:00
Format                          : application/pdf
Title                           : Flash HT Book.book
Creator                         : giuliano.galimberti
Producer                        : Acrobat Distiller 9.0.0 (Windows)
Document ID                     : uuid:0dc7fc89-17c7-4251-82dd-ed3965cfc140
Instance ID                     : uuid:e3a9bb18-39d0-4576-83eb-816207bd230c
Page Mode                       : UseOutlines
Page Count                      : 186
EXIF Metadata provided by EXIF.tools

Navigation menu