AFFIXUS Hip Fracture Nail Surgical Technique

2016-04-01

: Pdf Affixus Hip Fracture Nail Surgical Technique AFFIXUS_Hip_Fracture_Nail_Surgical_Technique 3 2016 pdf

Open the PDF directly: View PDF PDF.
Page Count: 36

DownloadAFFIXUS Hip Fracture Nail Surgical Technique
Open PDF In BrowserView PDF
Surgical Technique

One Surgeon. One Patient.

Over 1 million times per year, Biomet helps one surgeon
provide personalized care to one patient.
The science and art of medical care is to provide the right
solution for each individual patient. This requires clinical
mastery, a human connection between the surgeon and the
patient, and the right tools for each situation.
At Biomet, we strive to view our work through the eyes of one
surgeon and one patient. We treat every solution we provide
as if it’s meant for a family member.
Our approach to innovation creates real solutions that assist
each surgeon in the delivery of durable personalized care
to each patient, whether that solution requires a minimally
invasive surgical technique, advanced biomaterials or a
patient-matched implant.
When one surgeon connects with one patient to provide
personalized care, the promise of medicine is fulfilled.

AFFIXUS Hip Fracture Nail
Contents

Features and Benefits..........................................................................................................................................................................................................3
Indications and Pre-op Planning.....................................................................................................................................................................................7
Patient Positioning and Reduction.................................................................................................................................................................................8
Entry and Canal Preparation.............................................................................................................................................................................................9
Nail Insertion........................................................................................................................................................................................................................ 12
Proximal Locking................................................................................................................................................................................................................ 15
Distal Locking...................................................................................................................................................................................................................... 23
End Cap Placement........................................................................................................................................................................................................... 25
Implant Removal................................................................................................................................................................................................................ 26
Implant Diagrams.............................................................................................................................................................................................................. 27
Product Ordering Information...................................................................................................................................................................................... 28

US Surgeon Design Team
George J. Haidukewych, MD
Orlando, FL
Daniel S. Horwitz, MD
Salt Lake City, UT
Frank A. Liporace, MD
Newark, NJ
S. Andrew Sems, MD
Rochester, MN

International Surgeon
Designer
Peter Giannoudis, MD
Leeds, UK

1

2

AFFIXUS Hip Fracture Nail

Strength and Stability in the Proximal Femur
• 	 Optimal lag screw design for resistance to cut-out
• 	 Easy-to-use instrumentation and targeting jig,
which includes Goal Post technology,
aids in lag screw placement
• 	Extensive range of neck/shaft angles, distal diameters,
and nail lengths – combined with a small proximal nail
diameter – allows the surgeon to achieve a close match
for each patient’s anatomy
• 	Unique distal bend facilitates entry through
the proximal 1/3 of the femur and reduces potential
for anterior cortex penetration

3

AFFIXUS Hip Fracture Nail
A system of choices for effective treatment
of proximal femoral fractures

•	Short (180 mm) and long (260 - 460 mm) nail options
treat a wide range of proximal fracture indications using
a single set of user-friendly instruments
•	15.6 mm proximal nail diameter
•	Proximal 4˚ lateral bend allows for greater
trochanteric entry site
•	125˚ and 130˚ neck angles provide a range
of anatomical options

•	 3° distal bend facilitates ease of insertion through
the proximal intertrochanteric/subtrochanteric region
•	Pre-loaded set screw for ease of use
•	5.0 mm anti-rotation (AR) screw for rotational
control (optional)
•	Shouldered lag screw and AR screw help prevent medial
screw disengagement

•	10˚ of proximal anteversion built into the nails

•	Long nail maintains a 1.8 M radius of curvature
to closely match the femoral anatomy

•	10.5 mm diameter cannulated lag screw for bone
preservation

•	5.0 mm diameter distal interlocking screws have
a large core diameter for strong fixation

•	Unique thread spacing and design of the lag screw
helps to resist displacement and cut-out

•	Static or dynamic distal locking options
with a 6 mm dynamization range

•	 Chamfer on the front distal tip helps facilitate
insertion and minimizes risk of stress on
the anterior cortex in the distal femur

4

Multiple locking options for optimal implant stability

The AFFIXUS Hip Fracture Nail System, comprised of short and long nails, provides surgeons with
an intramedullary hip screw to stabilize fractures of the proximal femur. The AFFIXUS Hip Fracture
Nail combines the principles of a compression hip screw with the biomechanical advantages of an
intramedullary nail.

5

AFFIXUS Hip Fracture Nail

Figure 1

6

6

AFFIXUS Hip Fracture Nail

Figure 1

AFFIXUS Hip Fracture Nail
Short – (180 mm)

AFFIXUS Hip Fracture Nail
Long – (260 - 460 mm)

Indications and Pre-op Planning
The AFFIXUS Hip Fracture Nail System* is designed for
antegrade trochanteric insertion to treat the following
fractures (Figure 1):
The AFFIXUS Hip Fracture Nail System is intended to
treat stable and unstable proximal fractures of the femur
including pertrochanteric fractures, intertrochanteric
fractures, high subtrochanteric fractures and combinations
of these fractures, including non-union, malunion and
tumor resections. The Long Nail system is additionally
indicated to treat pertrochanteric fractures associated
with shaft fractures, pathologic fractures in osteoporotic
bone (including prophylactic use) of the trochanteric and
diaphyseal areas, impending pathological fractures, long
subtrochanteric fractures, ipsilateral femoral fractures,
proximal or distal non-unions, malunions, revision
procedures and tumor resections.

Note: Bone screws referenced in this material are not
intended for screw attachment or fixation to the posterior
elements (pedicles) of the cervical thoracic or lumbar
spine.
* System includes short (180 mm) and long (260-460 mm) nails,
in 20 mm increments.

7

AFFIXUS Hip Fracture Nail

Figure 2

Patient Positioning and Reduction
Place the patient in the supine or lateral position on a
fracture table or radiolucent imaging table. Lateral access
to the proximal femur is required. Intraoperative image
intensification with a C-arm is required to obtain AP and
lateral imaging of the operative area during preoperative
preparation (reduction) and throughout the procedure
for nail insertion, nail locking, and anteversion alignment.
Avoid excessive abduction of the hip during reduction
as the access to the starting point and nail insertion may
be impeded. The trunk may be laterally flexed away from
the operative side to improve access to the starting point.
The contralateral leg may be flexed at the hip or scissored
below the affected leg in the supine position (Figure 2).

Closed Fracture Reduction
Fluoroscopy must be used to verify proper fracture
reduction.
8

• Acceptable fracture alignment must be obtained prior to
implant insertion
• Surgeon must avoid varus malreductions
• Use a combination of traction, rotation, adduction, and
flexion/extension of the leg to obtain an acceptable
reduction
• Open reductions may be required for more complicated
fracture patterns and should be used when an acceptable
closed reduction cannot be obtained (see page 10)

Initial Incision
Make an incision proximal to the tip of the greater
trochanter in line with the femoral axis. Divide the fascia
lata in line with its fibers and access the tip of the greater
trochanter.

Trochanteric entry point

Figure 3
Entry reamer shape
matches proximal nail
shape

Awl

Figure 4
Figure 5

Figure 6

Entry and Canal Preparation
Femoral Entry Preparation
Attach the standard 3.2 mm guide pin to the pistol
guidewire gripper or power source and pass it through the
tip of the greater trochanter into the center of the femoral
canal. Position the entry on the tip of the greater trochanter
(Figure 3). Confirm on AP and lateral fluoroscopy views that
the entry pin is centered on the trochanter.

Option 1: Cannulated Entry Reamer
(One-step 16.6 mm)
Attach the cannulated entry reamer to the power source
and pass it over the guide pin through the entry portal
(Figure 4).
It is essential to ream until the reamer’s proximal shaft
passes with the greater trochanter’s cortical bone as the
shape of the entry reamer matches the nail shape and the
top of the cylindrical segment of the reamer corresponds
to the top of the nail (Figure 4). Reaming should continue
until the tip of the entry reamer is at the level of the lesser
trochanter and not beyond.

Option 2: Cannulated Awl
Pass the cannulated awl over the guide pin and introduce
with a rotation motion until the awl is buried to at least half
its blade length (Figure 5 & 6).
9

AFFIXUS Hip Fracture Nail

Figure 7

Figure 8

Open Fracture Reduction
Once access to the femoral canal has been gained, place
the ball nose guide wire into the entry site utilizing the
pistol guide wire gripper (Figure 7).
Obtain appropriate anatomic reduction in order to
restore length, anatomic axis alignment, and rotation
of the injured limb. Reduction can be achieved through
the surgeon’s preferred method such as traction,
external fixator, external aids, or joysticks. To aid in
manipulating the fracture fragments and passing the
ball nose guide wire, long (7.5 mm diameter) and short
(6.5 mm diameter) reduction tools are available.

10

Insert the reduction tool into the medullary canal, past the
fracture site. Once the fracture is in alignment, pass the
ball nose guide wire, available in both 80 cm and 100 cm
lengths, across the fracture site. Remove the reduction tool
(Figure 8).

Figure 9

Figure 10

Canal Preparation
Short Nail

Nail Length Selection

Confirm that the femoral diaphysis is wide enough and
long enough to allow the selected nail diameter to pass.
Ream as necessary to enlarge the diaphysis to accept the
selected nail.

With the tip of the ball nose guide wire at the level of the
desired depth of nail insertion, slide or snap the nail depth
gauge onto the ball nose guide wire until it contacts the
bone, ensuring that the tip does not fall into the existing
trochanteric entry canal, thus providing an inaccurate
measurement. To obtain the appropriate nail length, read
the measurement mark on the nail depth gauge that is
closest to the beginning of the black transition area on
the guide wire (Figure 10). If a nail of the exact measured
length is not available, choose a shorter nail of the next
closest available length. A direct measurement can also be
taken of the uninjured extremity using either radiographs
with magnification markers, or directly on the uninjured
limb.

Long Nail
Achieve proper alignment of the injured limb prior to reaming.
Maintain alignment throughout the reaming process to
avoid eccentric reaming. Commence reaming by placing
the flexible reamer over the ball nose guide wire (Figure 9).
Ream the medullary canal in millimeter increments
until cortical bone is reached and in half-millimeter
increments thereafter. Surgeon preference should dictate
the actual extent of intramedullary reaming. Monitor
the reaming procedure using image intensification
to avoid eccentric or excessive cortical reaming.
Note: It is recommended to over-ream the diaphysis by 2 mm.

11

AFFIXUS Hip Fracture Nail

Al
ig
nm

en
to
f
rg
et pro
xim
in
g

ta

al

Figure 11

Figure 12

Nail Insertion
Jig Assembly
Select the appropriate targeting jig that corresponds to
the neck shaft angle of the implant selected. Insert the
jig bolt through the targeting jig using the jig bolt driver
(Figure 11).

12

Note: 130º neck angle is most commonly used (Figure 12).

Jig Tab
Nail Slot

Alignment of distal
targeting

Figure 13A

When assembling the nail to the insertion jig, ensure that
the jig tabs align with the slots on the nail so that the nail
fully seats in the targeting jig (Figure 13A). Once the nail is
fully seated, securely tighten the jig bolt using the jig bolt
driver

Note: If it is difficult to attach the nail to the jig, doublecheck that the nail and jig are identified with the same
angle. The nail will only align with the jig if they have the
same neck-shaft angle.

Figure 13B

Check the assembly prior to nail introduction. Pass the
lag screw sheath through the targeting jig. A properly
assebled nail and jig will allow the lag screw drill to be
directed through the sleeve and through the center of the
lag screw hole in the nail.
When using a short (180 mm) nail, confirm the targeting
alignment of the distal interlocking screws using the green
sheaths and drill bits in the same manner (Figure 13B).

13

AFFIXUS Hip Fracture Nail

Figure 14

Figure 15

Insert the nail by hand over the 3 mm ball nose guide wire
into the medullary canal. Take care not to strike the jig or
targeting arm with the mallet. A curved impaction tool is
included in the set and is meant to be used for gentle taps
of the mallet to fine tune the final seating of the nail.

As the nail passes the medial cortex of the proximal
fragment, slowly derotate the jig handle into the usual
lateral position, so that the anterior bow of the nail now
corresponds with the anterior bow of the femur (Figure 14).
If the nail requires substantial force to advance, remove it
and ream an additional millimeter. Avoid excessive force
when inserting the nail. Advance the nail until the lag screw
aligns to the desired position into the femoral head and
neck to allow ideal placement of the lag screw (Figure 15).

Nail Insertion

Note: The insertion jig should not be hammered on.

It may be helpful to preliminarily insert the trochanteric
nail utilizing its bow to facilitate clearance of the medial
femoral cortex of the proximal fragment. To do this, rotate
the insertion jig anteriorly (toward the ceiling). In this
position the distal bend in the nail will be angled laterally
to aid in passing the nail through the greater trochanteric
entry site, and avoid medial cortical penetration.

14

Maintenance of reduction must be confirmed prior to lag
screw insertion. If the reduction has shifted to a suboptimal
position, further hip adduction, traction, and rotational
adjustments can be made prior to lag screw placement.
Remove the ball nose guide wire.

Lag Screw Sheath
Lag Screw
3.2 mm sleeve

Lag Screw Trochar

Figure 16A

Figure 16B

Proximal Locking
Lag Screw Guide Pin Introduction
Note: At the distal end of the jig assembly, the jig knob can
be tightened to secure the position of the lag screw sheath
to maintain contact against the lateral femoral cortex.

Insert the lag screw sheath assembly (lag screw sheath, lag
screw trochar, lag screw 3.2 mm sleeve) through the lag
screw hole in the jig. Pass the trochar through the sheath
and make an appropriate skin incision where the trochar
contacts the skin. Advance the trochar through the tissue
until the tip is seated against the lateral femoral cortex and
confirm with fluoroscopy. The trochar may be impacted
into the lateral cortex with a mallet to create a starting
point for the guide pin and minimize migration during
insertion (Figure 16A).

Introduce the 3.2 mm guide pin into the 3.2 mm sleeve and
drill into position under fluoroscopic guidance. Check the
guide pin position within the center of the femoral head
and neck in both AP and lateral planes. Advance the guide
pin to a distance within 5 mm from the subchondral bone
(Figure 16B).

Remove the trochar and maintain the lag screw sheath
position against the lateral femoral cortex.

Note: If at any time a guide pin is bent, replace it
immediately.

15

AFFIXUS Hip Fracture Nail

Figure 18A

Figure 18B

Figure 17
Fluoroscopic true lateral of the
proximal femur with insertion jig

Goal Post Technology

Lag Screw Length Selection

The Goal Post Technology is designed to facilitate
visualization of the femoral neck on the lateral view in order
to more accurately place the guide pin for the lag screw.
The anterior and posterior metal posts on the proximal
aspect of the insertion jig allow for an unobstructed
fluoroscopic view down to the base of the femoral neck
(Figure 17) and assist with fine tuning of the guide pin
before it is fully seated in the femoral head.

Before selecting a lag screw length, verify that the lag
screw sheath and 3.2 mm sleeve are in place and fully
seated against the lateral femoral cortex.
• 	The depth gauge seats against the lag screw sheath, not
the 3.2 mm sleeve
•	 The system measures to the tip of the guide pin
• 	The measurement represents the length of a lag screw
that begins at the end of the lag screw sheath and
terminates at the tip of the guide pin (Figure 18A and 18B)

16

Notch

Measurements taken
here, i.e., 100 mm
Lag Screw Drill

Depth Stop

Figure 19A
Push button control

Figure 19B

Depth Stop Adjustment

Lag Screw Drilling and Tapping

Adjust the depth stop on the lag screw drill to the desired
depth. The measurement on the depth stop should be
set to the depth measured by the lag screw depth gauge
(Figure 19A).

Advance the lag screw drill over the guide pin and drill to
the desired depth. Use fluoroscopy to confirm the position
of the lag screw drill and that the guide pin is not advanced
into the hip joint or acetabulum by the drill.

Adjust the depth stop by pushing in the button and sliding
the stop forward or backward until desired depth is seen
on the end of the depth stop closest to the gold drill bit tip
(Figure 19B).

If the bone is particularly dense, use the cannulated tap to
cut a thread for the lag screw.

Note: There is a “notch” on the lag screw drill that is
visible under fluoroscopy; this “notch” references 100 mm
(Figure 19A).

Note: There is a guide pin repositioning tool to aid
inreinserting the guide pin if it backs out with removal of
the lag screw drill.

17

AFFIXUS Hip Fracture Nail

Figure 20

Figure 21

Lag Screw Insertion

Fracture Compression

Insert the lag screw coupling rod through the lag screw
driver and position the selected lag screw on the end of
the lag screw driver. Tighten the coupling rod to secure the
lag screw to the driver.

Compression of the intertrochanteric component of
the fracture, if desired, can be achieved by utilizing the
compression wheel. Once the lag screw has been fully
seated, release traction from the leg and firmly seat the
lag screw sheath against the lateral cortex. Confirm that
the sheath is tightly secured in the jig by tightening the jig
knob, and place the compression wheel on the lag screw
driver and advance against the lateral side of the sheath. In
osteoporotic bone, care should be taken to avoid pulling
the lag screw out of the femoral head with this technique
(Figure 21).

Advance the lag screw manually into the femoral neck and
head over the guide pin. Confirm the terminal position of
the lag screw with fluoroscopy, with a goal of seating the
screw between 5 and 10 mm from the subchondral bone.
The handle of the lag screw driver must be positioned
either parallel or perpendicular to the targeting jig when
the lag screw has been advanced to the desired depth
(Figure 20). This will ensure that the set screw will engage
one of the grooves of the lag screw.

18

Note: Hash marks on lag screw driver represent 5 mm
intervals. It is recommended that no more than 4-6 mm
of compression is applied and should be applied prior to
placing the Anti-Rotation (AR) screw.

Figure 22A

Figure 22B
Cross section of set screw engaging lag screw

Lag Screw Fixation
The set screw is pre-loaded in the nail. Using the 5 mm set
screw hex driver, engage the set screw and advance in a
clockwise direction 2 to 3 full rotations until the set screw
contacts the lag screw in one of the four lag screw grooves
(Figure 22A & 22B).
To confirm proper position of the set screw, gently
attempt to rotate the lag screw both clockwise and
counterclockwise. If there is firm resistance and the lag
screw will not rotate, the set screw has properly engaged
the lag screw grooves. However, if you are able to rotate
the lag screw, the set screw has not engaged a groove and
the lag screw handle should be realigned and the set screw
tightened again.

The set screw may be backed off one-quarter turn to allow
dynamic compression of the lag screw in the nail, while still
providing rotational control of the lag screw.
Note: The set screw can be engaged before or after
inserting the AR screw (if the AR screw is to be used). The
AR screw will align through an oblong hole within the set
screw.

19

AFFIXUS Hip Fracture Nail

Figure 23

Anti-rotation (AR) Guide Pin and Screw
Placement (optional)
This system allows multiple techniques for placement of an
anti-rotation (AR) screw if desired.
• 	The AR screw may be inserted either before or after the
lag screw is placed, based upon surgeon preference and
the fracture pattern.
•	 The surgeon has the option to place a guide pin through
the AR hole to provisionally stabilize the fracture during
lag screw placement, or he/she may choose to use an AR
screw. The guide pin used through the AR hole is also
useful to assist in stabilizing the femoral neck and head
segment during lag screw placement to resist rotation
around the axis of the femoral neck. Once the lag screw
has been placed and secured, the surgeon may choose to
remove the guide pin from the AR hole and place a screw
in this position to provide further rotational control.

20

Place the AR screw sheath and trochar through the AR
hole in the insertion jig. Make a small incision where the
trochar meets the skin and advance the trochar to the
lateral aspect of the femoral cortex. Alternatively, in cases
where the lag screw has already been inserted, extend
the incision for the lag screw proximally to allow the AR
screw sheath and trochar to be seated against the femur
(Figure 23).
Note: When the anti-rotation and lag screw sheaths are
seated at the same time, they must be rotated so the
groove on the lag screw sheath faces the anti-rotation
screw sheath (so the colored handles are 180 degree to
each other) in order to allow both sheaths to fully seat
(Figure 23).

5 – 10 mm

15 – 20 mm

Figure 24

Figure 25

Remove the trochar and insert the AR 3.2 mm sleeve. Insert
the 3.2 mm guide pin and advance into desired position. It
is recommended to leave the AR guide pin 15-20 mm from
the subchondral bone (Figure 24).
Note: In cases where very dense cortical bone is
encountered, the cortex may be opened up with the antirotation screw drill prior to advancing the 3.2 mm guide
pin to prevent the guide pin from “walking” up the lateral
cortex.

Remove the guide pin and 3.2 mm sleeve. Confirm that the
screw sheath is advanced against the lateral femoral cortex
and use the AR drill to drill to the desired depth. Measure
the length of the desired screw by reading the depth of the
AR drill against the screw sheath.
Note: It is recommended that the tip of the AR screw be
15-20 mm shorter than the lag screw to avoid perforation
of the femoral head (Figure 25).

21

AFFIXUS Hip Fracture Nail

Figure 26

Select an AR screw of the desired length. Place the AR
screw on the 3.5 mm hex driver and manually insert the
screw into the femur through the AR screw sheath.
Advance until the tip of the screw reaches the desired
depth and confirm with fluoroscopy. The screwdriver and
sheath may now be removed.

Securing the AR Screw (optional)
The AR screw may be secured with an impinging end cap
that is inserted through the end of the nail.
Note: The impinging end cap will make the AR screw a
static construct and is recommended to only be used when
the lag screw is also fixed in a static position (this can be
achieved by not backing off the pre-loaded set

22

screw a quarter turn). Otherwise there is risk of creating
the Z-effect.
It is recommended to only lock the AR screw in instances
in which the set screw has been left fully engaged into the
lag screw, thus preventing any collapse of the 10.5 mm
compression screw (Figure 26).
The impinging end cap may be utilized at the end of
the case, after the set screw for the lag screw has been
tightened, and the insertion jig has been removed.

Figure 27

Figure 28

Distal Locking
Distal Locking (short nails)
The short nail may be locked either statically, dynamically,
or left unlocked based on the particular fracture pattern
and stability (Figure 27).
Pass the distal screw sheath and trochar through the hole
labeled “static” on the insertion jig and advance to the
lateral femoral cortex. Remove the trochar and use the
distal screw drill sleeve and 4.3 mm graduated drill bit. Drill
until the far cortex is either reached or penetrated. The drill
is calibrated and may be used to determine screw length
by reading the depth off the end of the distal screw drill
sleeve (Figure 28).
An optional distal screw depth gauge is available to
confirm screw length. This gauge measures off of the
lateral side of the 4.3 mm distal screw drill sleeve.

Select a 5.0 mm diameter screw of the desired depth and
use the 3.5 mm hex driver long to introduce the screw
through the screw sheath and advance until it is fully
seated against the lateral cortex.
Repeat the above steps for dynamic locking, except pass
the distal screw sheath and trochar through the hole
labeled “dynamic” on the insertion jig.
Note: Maintain contact of the drill sheath on the lateral
femoral cortex to ensure accurate measurement of the
distal locking screw. Verify screw position using AP and
lateral fluoroscopy imaging.
Note: There are two 4.3 mm drill bits available. Use the
long bit when drilling through the jig assembly and use the
short bit when performing the freehand approach.
23

AFFIXUS Hip Fracture Nail

Figure 30

Figure 29

Distal Locking (long nails)

Option 2

Prior to locking the distal screw(s), check femoral length
and rotation under fluoroscopy. Distal locking of long
nails should be conducted using the standard image
intensification freehand technique.

Using the short 4.3 mm graduated drill, drill until the far
cortex is either reached or penetrated. Remove the 4.3 mm
graduated drill and measure using the distal screw depth
gauge. Ensure that the sheath of the distal screw depth
gauge is fully seated on the bone (Figure 30).

Option 1
Using the short 4.3 mm graduated drill and the 4.3 mm drill
measuring sleeve, drill until the far cortex is either reached
or penetrated. Verify the drill bit position fluoroscopically
prior to taking any measurements. Read the calibration
directly off of the 4.3 mm graduated drill by using the
drill measuring sleeve. The measurement should be taken
from the end of the measuring sleeve, closest to the power
source (Figure 29).

24

Remove the drill bit and advance the 5.0 mm screw using
the solidlok screwdriver or 3.5 mm hex driver. Repeat the
above steps for additional screw placement.

Figure 31

End Cap Placement
End Cap Placement (optional)
Unscrew the jig bolt that connects the insertion jig to
the end of the nail using the jig bolt driver. Remove the
insertion jig and use fluoroscopy to determine the length
of the end cap desired, with a goal of leaving the proximal
aspect of the end cap flush with the tip of the greater
trochanter.

Attach the end cap to the 5 mm end cap hex driver and
insert into the end of the nail. Tighten the end cap by
turning clockwise until the end cap fully seats against the
top of the nail. If fixation of the AR screw is desired, select
the impinging in cap instead of the standard end cap
(Figure 31).

25

AFFIXUS Hip Fracture Nail
Slotted Hammer

Figure 32

Sliding Hammer

Figure 33

Implant Removal
Identify the proximal end of the nail by opening the same
incision used for insertion of the implant. Clear bone from
the proximal end of nail if necessary or remove the end cap
(if present) with the 5 mm end cap hex driver.
•	Remove the distal screw using the 3.5 mm hex driver
after making an incision through the scar site.
•	 If an AR screw is present, use the AR screw removal tool
to extract the AR screw prior to loosening the set screw.
	 Reminder: The set screw should NOT be loosened prior to
removing the AR screw.
•	 Use the 5 mm set screw hex driver to loosen the set screw.
This will allow the lag screw to rotate counterclockwise.
Typically 2 to 3 full rotations is all that is necessary.
•	 Attach the lag screw driver and coupling rod to the lateral
end of the lag screw and confirm that it will freely rotate
in a counterclockwise direction.

•	Insert the cannulated extraction bolt into the proximal
end of the nail (Figure 32).
Note: If the extraction bolt is not threading into the
proximal end of the nails, the set screw may have been
backed out too far and should be advanced clockwise.
•	 Attach the extraction rod to the extraction bolt.
•	Remove the lag screw by turning counterclockwise and
then remove the distal interlocking screws.
•	Use the sliding hammer or slotted mallet over the
extraction rod and back slap to remove the nail (Figure
33).
Note: It is recommended that the extraction rod and
bolt be attached to the nail prior to removing the final
screw to prevent the nail from being forced down the
intramedullary canal.
Note: The conical extractor is designed to cross thread
onto the nail, and it is recommended that it is tightly
secured to the nail before the lag screw is removed to
prevent the nail from rotating in the femoral canal.

26

15.6 mm Proximal
Diameter

4˚

Implant Diagrams

125˚
&
130˚

12.76 mm

10° Proximal
Anteversion

Distal Screw, 20-80 mm
Sterile: 8145-50-0XX
• Diameter 5.0 mm
• 3.5 mm Hex Driver Socket
1.8 M Radius
of Curvature

Anti-rotation Screw, 50-110 mm
Sterile: 8145-01-XXX
• Diameter 5.0 mm
• 3.6 mm Hex Driver Socket
• 3 mm Inner Thread for Removal
• Self Tapping Threads

Lag Screw, 70-130 mm
Sterile: 8145-10-XXX
• Diameter 10.5 mm
• Reverse Buttress Thread
• 6.5 mm Square Driver Socket
• Self Tapping Threads

25 mm

6 mm
Dynamization
Range

30 mm
Diameter 9.0 mm
Diameter 11.0 mm
Diameter 13.0 mm
Diameter 15.0 mm

3˚

27

AFFIXUS Hip Fracture Nail
Product Ordering Information
Rights

Lefts

Rights

Long Nails, 9 mm, 125°

Lefts

Long Nails, 9 mm, 130°

8143-09-260

8144-09-260

125° 9 mm X 260 mm

8145-09-260

8146-09-260

130° 9 mm X 260 mm

8143-09-280

8144-09-280

125° 9 mm X 280 mm

8145-09-280

8146-09-280

130° 9 mm X 280 mm

8143-09-300

8144-09-300

125° 9 mm X 300 mm

8145-09-300

8146-09-300

130° 9 mm X 300 mm

8143-09-320

8144-09-320

125° 9 mm X 320 mm

8145-09-320

8146-09-320

130° 9 mm X 320 mm

8143-09-340

8144-09-340

125° 9 mm X 340 mm

8145-09-340

8146-09-340

130° 9 mm X 340 mm

8143-09-360

8144-09-360

125° 9 mm X 360 mm

8145-09-360

8146-09-360

130° 9 mm X 360 mm

8143-09-380

8144-09-380

125° 9 mm X 380 mm

8145-09-380

8146-09-380

130° 9 mm X 380 mm

8143-09-400

8144-09-400

125° 9 mm X 400 mm

8145-09-400

8146-09-400

130° 9 mm X 400 mm

8143-09-420

8144-09-420

125° 9 mm X 420 mm

8145-09-420

8146-09-420

130° 9 mm X 420 mm

8143-09-440

8144-09-440

125° 9 mm X 440 mm

8145-09-440

8146-09-440

130° 9 mm X 440 mm

8143-09-460

8144-09-460

125° 9 mm X 460 mm

8145-09-460

8146-09-460

130° 9 mm X 460 mm

Long Nails, 11 mm, 125°

Long Nails, 11 mm, 130°

8143-11-260

8144-11-260

125° 11 mm X 260 mm

8145-11-260

8146-11-260

130° 11 mm X 260 mm

8143-11-280

8144-11-280

125° 11 mm X 280 mm

8145-11-280

8146-11-280

130° 11 mm X 280 mm

8143-11-300

8144-11-300

125° 11 mm X 300 mm

8145-11-300

8146-11-300

130° 11 mm X 300 mm

8143-11-320

8144-11-320

125° 11 mm X 320 mm

8145-11-320

8146-11-320

130° 11 mm X 320 mm

8143-11-340

8144-11-340

125° 11 mm X 340 mm

8145-11-340

8146-11-340

130° 11 mm X 340 mm

8143-11-360

8144-11-360

125° 11 mm X 360 mm

8145-11-360

8146-11-360

130° 11 mm X 360 mm

8143-11-380

8144-11-380

125° 11 mm X 380 mm

8145-11-380

8146-11-380

130° 11 mm X 380 mm

8143-11-400

8144-11-400

125° 11 mm X 400 mm

8145-11-400

8146-11-400

130° 11 mm X 400 mm

8143-11-420

8144-11-420

125° 11 mm X 420 mm

8145-11-420

8146-11-420

130° 11 mm X 420 mm

8143-11-440

8144-11-440

125° 11 mm X 440 mm

8145-11-440

8146-11-440

130° 11 mm X 440 mm

8143-11-460

8144-11-460

125° 11 mm X 460 mm

8145-11-460

8146-11-460

130° 11 mm X 460 mm

Long Nails, 13 mm, 125°

Long Nails, 13 mm, 130°

8143-13-260

8144-13-260

125° 13 mm X 260 mm

8145-13-260

8146-13-260

130° 13 mm X 260 mm

8143-13-280

8144-13-280

125° 13 mm X 280 mm

8145-13-280

8146-13-280

130° 13 mm X 280 mm

8143-13-300

8144-13-300

125° 13 mm X 300 mm

8145-13-300

8146-13-300

130° 13 mm X 300 mm

8143-13-320

8144-13-320

125° 13 mm X 320 mm

8145-13-320

8146-13-320

130° 13 mm X 320 mm

8143-13-340

8144-13-340

125° 13 mm X 340 mm

8145-13-340

8146-13-340

130° 13 mm X 340 mm

8143-13-360

8144-13-360

125° 13 mm X 360 mm

8145-13-360

8146-13-360

130° 13 mm X 360 mm

8143-13-380

8144-13-380

125° 13 mm X 380 mm

8145-13-380

8146-13-380

130° 13 mm X 380 mm

8143-13-400

8144-13-400

125° 13 mm X 400 mm

8145-13-400

8146-13-400

130° 13 mm X 400 mm

8143-13-420

8144-13-420

125° 13 mm X 420 mm

8145-13-420

8146-13-420

130° 13 mm X 420 mm

8143-13-440

8144-13-440

125° 13 mm X 440 mm

8145-13-440

8146-13-440

130° 13 mm X 440 mm

8143-13-460

8144-13-460

125° 13 mm X 460 mm

8145-13-460

8146-13-460

130° 13 mm X 460 mm

Long Nails, 15 mm, 130°

28

8145-15-320

8146-15-320

130° 15 mm X 320 mm

8145-15-360

8146-15-360

130° 15 mm X 360 mm

8145-15-400

8146-15-400

130° 15 mm X 400 mm

8145-15-440

8146-15-440

130° 15 mm X 440 mm

Short Nails, 125°

Distal Screws

8143-09-180

125° 9 mm X 180 mm

8145-50-020

CORTICAL BONE SCR 5.0 mm X 20 mm

8143-11-180

125° 11 mm X 180 mm

8145-50-022

CORTICAL BONE SCR 5.0 mm X 22 mm

8143-13-180

125° 13 mm X 180 mm

8145-50-024

CORTICAL BONE SCR 5.0 mm X 24 mm

8145-50-026

CORTICAL BONE SCR 5.0 mm X 26 mm

Short Nails, 130°

8145-50-028

CORTICAL BONE SCR 5.0 mm X 28 mm

8145-09-180

130° 9 mm X 180 mm

8145-50-030

CORTICAL BONE SCR 5.0 mm X 30 mm

8145-11-180

130° 11 mm X 180 mm

8145-50-032

CORTICAL BONE SCR 5.0 mm X 32 mm

8145-13-180

130° 13 mm X 180 mm

8145-50-034

CORTICAL BONE SCR 5.0 mm X 34 mm

8145-50-036

CORTICAL BONE SCR 5.0 mm X 36 mm

8145-50-038

CORTICAL BONE SCR 5.0 mm X 38 mm

Lag Screws
8145-10-070

LAG SCREW 10.5 mm X 70 mm

8145-50-040

CORTICAL BONE SCR 5.0 mm X 40 mm

8145-10-075

LAG SCREW 10.5 mm X 75 mm

8145-50-042

CORTICAL BONE SCR 5.0 mm X 42 mm

8145-10-080

LAG SCREW 10.5 mm X 80 mm

8145-50-044

CORTICAL BONE SCR 5.0 mm X 44 mm

8145-10-085

LAG SCREW 10.5 mm X 85 mm

8145-50-046

CORTICAL BONE SCR 5.0 mm X 46 mm

8145-10-090

LAG SCREW 10.5 mm X 90 mm

8145-50-048

CORTICAL BONE SCR 5.0 mm X 48 mm

8145-10-095

LAG SCREW 10.5 mm X 95 mm

8145-50-050

CORTICAL BONE SCR 5.0 mm X 50 mm

8145-10-100

LAG SCREW 10.5 mm X 100 mm

8145-50-052

CORTICAL BONE SCR 5.0 mm X 52 mm

8145-10-105

LAG SCREW 10.5 mm X 105 mm

8145-50-054

CORTICAL BONE SCR 5.0 mm X 54 mm

8145-10-110

LAG SCREW 10.5 mm X 110 mm

8145-50-056

CORTICAL BONE SCR 5.0 mm X 56 mm

8145-10-115

LAG SCREW 10.5 mm X 115 mm

8145-50-058

CORTICAL BONE SCR 5.0 mm X 58 mm

8145-10-120

LAG SCREW 10.5 mm X 120 mm

8145-50-060

CORTICAL BONE SCR 5.0 mm X 60 mm

8145-10-125

LAG SCREW 10.5 mm X 125 mm

8145-50-065

CORTICAL BONE SCR 5.0 mm X 65 mm

8145-10-130

LAG SCREW 10.5 mm X 130 mm

8145-50-070

CORTICAL BONE SCR 5.0 mm X 70 mm

8145-50-075

CORTICAL BONE SCR 5.0 mm X 75 mm

8145-50-080

CORTICAL BONE SCR 5.0 mm X 80 mm

Anti-Rotation Screws
8145-01-050

A/R SCREW 50 mm

8145-01-055

A/R SCREW 55 mm

End Caps

8145-01-060

A/R SCREW 60 mm

8145-03-000

END CAP FLUSH

8145-01-065

A/R SCREW 65 mm

8145-03-005

END CAP 5 mm

8145-01-070

A/R SCREW 70 mm

8145-03-101

IN CAP FLUSH IMPINGING

8145-01-075

A/R SCREW 75 mm

8145-01-080

A/R SCREW 80 mm

8145-01-085

A/R SCREW 85 mm

8145-01-090

A/R SCREW 90 mm

8145-01-095

A/R SCREW 95 mm

8145-01-100

A/R SCREW 100 mm

8145-01-105

A/R SCREW 105 mm

8145-01-110

A/R SCREW 110 mm

29

AFFIXUS Hip Fracture Nail
Product Ordering Information

1

2

3

4

8

AFFIXUS Hip Fracture Nail System
2112-01-000	
Instrument Case 2
2112-01-001	
Instrument Case 1
2112-01-800	
Full Anatomy (FA) Instrument Case
General
2810-01-004	
8261-66-000	

T-Handle Hudson
Ratchet Screwdriver Handle Small

Entry
2112-01-100	
2112-01-102	
2112-01-103	
2112-01-104	
2810-13-004	

1 - AWL
2 - Entry Reamer Solid Shaft
3 - Entry Reamer Flexible Shaft
4 - Entry Portal
5 - Entry Portal Trochar

* Products are disposable.

30

5

9

10

6

7

11

Reduction	
9030-03-004	
2810-01-080	
2810-01-100	
2810-01-001	
2810-01-026	
2810-01-007	
2142-02-012	
2112-01-003	
2141-19-000	
2810-01-175	

Threaded Guide Pin 3.2 mm*
Ball Nose Guidewire 80 cm*
Ball Nose Guidewire 100 cm*
6 - Pistol Guidewire Gripper
7 - Guidewire Pusher
8 - Long Reduction Tool
9 - Ball Spike Pusher
10 - Bone Hook
11 - Femoral Bone Clamp
3.2 mm x 444 mm Threaded Guide Pin Sterile*

15

16

12

13

18

19

Nail Insertion
2112-01-106	
2112-01-200	
2112-01-207	
2112-01-201	
2112-01-208	
2112-01-202	
2112-01-209	
2112-01-205	
2112-01-206	
2810-13-037	
2810-13-006	
2112-01-204	

20

21

12 - Nail Depth Gauge
13 - Insertion jig 125°
FA Insertion jig 125°
14 - Insertion jig 130°
FA Insertion jig 130°
15 - Insertion jig bolt
FA Insertion jig bolt
Jig knob
Jig knob retainer
16 - Flexible jig bolt driver 8 mm
Jig bolt driver 8 mm
17 - Impaction tool

17

14

22

23

24

25

26

27

28

Lag Screw Placement
2112-01-300	
18 - Lag Screw Sheath
2112-01-301	
19 - Lag Screw Trochar
2112-01-302	
20 - Lag Screw 3.2 mm Sleeve
2112-01-304	
21 - Lag Screw Depth Gauge
2112-01-303	
22 - Lag Screw Drill
2112-01-310	
23 - Lag Screw Tap
2112-01-307	
24 - Lag Screw Driver
2112-01-306	
25 - Lag Screw Coupling Rod
2112-01-308	
26 - Compression Wheel
2112-01-309	
27 - 5 mm Hex Driver - Set Screw
2112-01-320	
FA 5 mm Hex Driver - Set Screw
2112-01-312	
28 - Guide Pin Positioning Tool

31

AFFIXUS Hip Fracture Nail
Product Ordering Information

29

30

31

32

37

38

39

40

33

41

34

42

35

43

44

45

36

46

47

AR Screw Placement

Distal Screw Insertion

2112-01-501	

29 - A/R Screw Sheath

2112-01-401	

37 - Distal Screw Sheath

2112-01-502	

30 - A/R Screw Trochar

2112-01-402	

38 - Distal Screw Trochar

2112-01-503	

31 - A/R Screw 3.2 mm sleeve

2112-01-403	

39 - Distal Screw Drill Sleeve

2112-01-505	

32 - A/R Screw Drill*

2112-01-404	

40 - Distal Screw Depth Gauge

2112-01-504	

33 - 3.5 mm Hex Driver Long-A/R Distal Screw

2112-01-406 	

41 - 4.3 mm Distal Graduated Drill Short*

2112-01-506	

34 - A/R Screw Removal Tool

2112-01-405	

42 - 4.3 mm Distal Graduated Drill Long*

2112-01-410	

43 - 4.3 mm Drill Measuring Sleeve*

End Cap Placement

2112-01-409	

44 - 3.5 mm Hex Driver Short - Distal Screw

2112-01-600	

2810-01-020	

45 - SolidLok Screwdriver Handle

2810-01-021	

46 - SolidLok Driver Inner Shaft

2112-01-601	
2112-01-602	

35 - 5 mm Hex Driver End Cap
5 mm Hex Can Driver End Cap
36 - End Cap Removal Tool

2810-01-019	
2112-01-504	

* Products are disposable.

32

SolidLok Hex Tip 3.5 mm*
33 - 3.5 mm Hex Driver Long - AR/Distal Screw

50
48

49

52

53

51

Extraction 	
2112-01-666	
1095	
1796	
1096	
2112-01-606	
2112-01-605	
2810-01-027	

48 - Cannulated Extraction Bolt
49 - Extraction Rod
50 - Sliding Hammer Small
Sliding Hammer Large
51 - Slotted Mallet
52 - Conical Extractor
53 - 3/4 in Hex Driver

Flexible Reamers
2810-02-400	
400 mm Nitinol Modular Reamer Hudson
2810-02-470	
470 mm Nitinol Modular Reamer Hudson
2810-02-015	
150 mm Reamer Extension
2810-02-081	
8 mm MNBLC Endcut Reamer Hudson
2810-02-091	
9 mm MNBLC Endcut Reamer Hudson
2810-04-090	
9.0 mm Modular Reamer Head
2810-04-095	
9.5 mm Modular Reamer Head

2810-04-100	
2810-04-105	
2810-04-110	
2810-04-115	
2810-04-120	
2810-04-125	
2810-04-130	
2810-04-135	
2810-04-140	
2810-04-145	
2810-04-150	
2810-04-155	
2810-04-160	
2810-04-165	
2810-04-170	

10.0 mm Modular Reamer Head
10.5 mm Modular Reamer Head
11.0 mm Modular Reamer Head
11.5 mm Modular Reamer Head
12.0 mm Modular Reamer Head
12.5 mm Modular Reamer Head
13.0 mm Modular Reamer Head
13.5 mm Modular Reamer Head
14.0 mm Modular Reamer Head
14.5 mm Modular Reamer Head
15.0 mm Modular Reamer Head
15.5 mm Modular Reamer Head
16.0 mm Modular Reamer Head
16.5 mm Modular Reamer Head
17.0 mm Modular Reamer Head

33

Important:

Additional Contraindication for Orthopaedic Screws and Plates only:

This Essential Product Information does not include all of the information
necessary for selection and use of a device. Please see full labeling for all
necessary information.

Cases with malignant primary or metastatic tumors which preclude adequate bone
support or screw fixations, unless supplemental fixation or stabilization methods
are utilized.

The use of metallic surgical appliances (screws, plates, intramedullary nails,
compression hip screws, pins and wires) provides the orthopaedic surgeon a
means of bone fixation and helps generally in the management of fractures and
reconstructive surgeries. These implants are intended as a guide to normal healing,
and are NOT intended to replace normal body structure or bear the weight of the
body in the presence of incomplete bone healing. Delayed unions or nonunions in
the presence of load bearing or weight bearing might eventually cause the implant
to break due to metal fatigue. All metal surgical implants are subjected to repeated
stress in use, which can result in metal fatigue.

Additional Contraindication for Retrograde Femoral Nailing:
A history of septic arthritis of the knee and knee extension contracture with inability
to attain at least 45º of flexion.
Additional Contraindications for Compression Hip Screws only:
Inadequate implant support due to the lack of medial buttress.
Warnings and Precautions:

Indications
The AFFIXUS Hip Fracture Nail is intended to treat stable and unstable proximal
fractures of the femur including pertrochanteric fractures, intertrochanteric fractures,
high subtrochanteric fractures and combinations of these fractures, including nonunion, malunion and tumor resections. The Long Nail system is additionally indicated
to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures
in osteoporotic bone (including prophylactic use) of the trochanteric and diaphyseal
areas, impending pathological fractures, long subtrochanteric fractures, ipsilateral
femoral fractures, proximal or distal non-unions, malunions, revision procedures and
tumor resections
Contraindications:
Screws, plates, intramedullary nails, compression hip screws, pins and wires are
contraindicated in: active infection, conditions which tend to retard healing such
as blood supply limitations, previous infections, insufficient quantity or quality of
bone to permit stabilization of the fracture complex, conditions that restrict the
patient’s ability or willingness to follow postoperative instructions during the healing
process, foreign body sensitivity, and cases where the implant(s) would cross open
epiphyseal plates in skeletally immature patients.

Bone screws and pins are intended for partial weight bearing and non-weight bearing
applications. These components cannot be expected to withstand the unsupported
stresses of full weight bearing.
Adverse Events:
The following are the most frequent adverse events after fixation with orthopaedic
screws, plates, intramedullary nails, compression hip screws, pins and wires:
loosening, bending, cracking or fracture of the components or loss of fixation in bone
attributable to nonunion, osteoporosis, markedly unstable comminuted fractures;
loss of anatomic position with nonunion or malunion with rotation or angulation;
infection and allergies and adverse reactions to the device material. Surgeons
should take care when targeting and drilling for the proximal screws in any tibial nail
with oblique proximal screws. Care should be taken as the drill bit is advanced to
penetrate the far cortex. Advancing the drill bit too far in this area may cause injury
to the deep peroneal nerve. Fluoroscopy should be used to verify correct positioning
of the drill bit.
Additional Adverse Events for Compression Hip Screw only:
Screw cutout of the femoral head (usually associated with osteoporotic bone).

This material is intended for health care professionals and the Biomet sales
force only. Distribution to any other recipient is prohibited. All content herein is
protected by copyright, trademarks and other intellectual property rights owned
by or licensed to Biomet Inc. or its affiliates unless otherwise indicated. This material
must not be redistributed, duplicated or disclosed, in whole or in part, without the
express written consent of Biomet.
Check for country product clearances and reference product specific instructions
for use. For complete product information, including indications, contraindications,
warnings, precautions, and potential adverse effects, see the package insert and
Biomet’s website.
This technique was prepared in conjunction with a licensed health care professional.
Biomet does not practice medicine and does not recommend any particular
orthopedic implant or surgical technique for use on a specific patient. The surgeon
is responsible for determining the appropriate device(s) and technique(s) for each
individual patient.
Not for distribution in France.

Legal Manufacturer
Biomet Trauma
56 East Bell Drive
P.O. Box 587
Warsaw, Indiana 46581
USA
©2014 Biomet Trauma • Form No. BMET0022.0-GBL • REV0614

www.biomet.com

Authorised Representative
Biomet UK Ltd.
Waterton Industrial Estate
Bridgend, South Wales
CF31 3XA
UK

0086



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : No
XMP Toolkit                     : Adobe XMP Core 5.5-c021 79.154911, 2013/10/29-11:47:16
Instance ID                     : uuid:56fa0d96-a995-9a4c-9c20-6f22ae43f304
Original Document ID            : adobe:docid:indd:fd6c8236-e1ae-11de-964f-bc2110bdf6d5
Document ID                     : xmp.id:a81da9fe-e208-4aba-9707-13a6f196f0f9
Rendition Class                 : proof:pdf
Derived From Instance ID        : xmp.iid:7293f660-d3ca-4617-a946-3bce9509c365
Derived From Document ID        : xmp.did:60e88dfd-381d-43f9-829c-2729d5108f77
Derived From Original Document ID: adobe:docid:indd:fd6c8236-e1ae-11de-964f-bc2110bdf6d5
Derived From Rendition Class    : default
History Action                  : converted
History Parameters              : from application/x-indesign to application/pdf
History Software Agent          : Adobe InDesign CC (Macintosh)
History Changed                 : /
History When                    : 2014:07:23 10:38:12-04:00
Create Date                     : 2014:07:23 10:38:12-04:00
Modify Date                     : 2014:07:23 10:38:43-04:00
Metadata Date                   : 2014:07:23 10:38:43-04:00
Creator Tool                    : Adobe InDesign CC (Macintosh)
Format                          : application/pdf
Producer                        : Adobe PDF Library 11.0
Trapped                         : False
Slug Family                     : Myriad Pro
Slug Version                    : 2.102
Slug Outline File Size          : 0
Slug Kerning Checksum           : 0
Slug Foundry                    : Adobe Systems
Slug Font Kind                  : OpenType - PS
Slug Checksum                   : 55235293
Slug Post Script Name           : MyriadPro-Regular
Slug Font Sense 12 Checksum     : 55235293
Page Count                      : 36
Creator                         : Adobe InDesign CC (Macintosh)
EXIF Metadata provided by EXIF.tools

Navigation menu