Datasheet

FSM – Field Strength Meter (IEC60118-4)
The FSM is a cost effective and simple solution for measuring, setting
up and commissioning an induction loop system to the requirements
of IEC60118-4:2006. The meter is an ergonomically designed hand
held instrument for measuring loop system performance. There are
three calibrated operational modes for the assessment of Background
Noise, Field Strength and Frequency Response as required to ensure
correct function of the loop system. The meter also doubles as a loop
listener, with a headphone output to listen to the signal in the loop.
The meter is supplied in a soft case with an audio CD with the
required test signals, and full operating instructions including a guide
to commissioning. The Ampetronic FSM can be used to monitor,
set up or commission any induction loop system regardless of the
manufacturer or type.

Features

Operational modes

• Simple assessment of any system to
IEC60118-4:2006

				

• Three modes of operation for three test types
• A weighted background noise
• Broad band mode (50Hz - 8kHz)
• Frequency response (100Hz, 1kHz, 5kHz)
• True RMS detection calibrated to
400mA/m = 0dB

Background Noise

To determine the level of the background magnetic field present in
the intended location for the loop system. Also used to measure
low level signals to assess overspill outside a loop system.
• A weighted filter
• True RMS detection referenced to 400mA/m
• Scale -42 to -12dB
• Audio (headphone) output is post filter for monitoring

• Wide viewing angle LED display
Field Strength

• Colour coded LEDs for simple readout
• Resolution to 1dB
• Head phone output with volume control
• Ergonomic, rugged, light weight construction
• Test signals supplied on CD
• Soft carry case
• 5 year warranty

Applications include:
•
•
•
•
•
•
•

Accurate set-up and commissioning
System monitoring and maintenance
Site surveys
Certification to IEC60118-4:2006
Assessment of frequency losses due to metal
Assessment of loop coverage and overspill
Assessment of background noise

A broad band measurement to measure field strength delivered by
the system.
• Broad band measurement 50Hz to 8kHz
• True RMS detection referenced to 400 mA/m
• Scale -22 to +8dB with 1dB intervals from -3 to +6dB
• Suitable for use with sine wave, pink noise, or
combination signals (provided), or any other real signals

  

Frequency Response

Third octave filters for measuring performance across the required
frequency spectrum as required by IEC60118-4.
Used to confirm adequate power at high frequencies as required
for good intelligibility. Used to assess frequency dependent loss
due to metal structures, and to optimise frequency compensation.
• Third octave bands at 100Hz, 1kHz and 5kHz
• True RMS detection
• Scale resolution to 1dB
• For use with pink noise only (signal provided)
• Audio (headphone) output is post filter

Commissioning Procedure
The FSM allows a complete test of an induction loop system to be performed with a simple six or seven step procedure as
shown in the table below. Using the FSM with the CD of test signals (included), all aspects of an installation can be examined
and adjusted to meet the requirements. A Certificate of Conformity (template included) is provided to record test results from this
procedure, and to certify that the installation meets the requirements of IEC 60118-4:2006.

Step

Audio Input

1

Volume

SYSTEM

of use

OFF

2

Background

SYSTEM

Noise

OFF

3

Field

Track: 1

Strength (1)

COMBINATION*

4

Frequency

Track: 2

Response

PINK NOISE

5

Field

Track: 1

Strength (2)

COMBINATION*

6

Overspill

Track: 1

(if required)

COMBINATION

7

System use

ACTUAL SIGNALS

FSM settings

Adjustments

METER off

n/a
Sources of

5kHz
1kHz
100hz

Performance
requirements
Determine volume of
use Sketch Layout
<22dB essential

magnetic noise

<32dB acceptable

Loop current

-3 to +3dB peaks

MLC / tone

-3 to +3dB peaks

control
Loop current
n/a
Input gain

compared to 1kHz
-3 to +3dB peaks
<42dB
(OFF SCALE)
-9 to 0dB peaks
Subjective - >OK

* Other signals may be used with revised performance requirements: PINK NOISE -9 to -3dB, 1kHz SINE -3 to +3dB

Accessories

Magnetic field measurement

• User guide/handbook

Coil orientation: Vertical when unit held upright
Reference level: 400mA/m (In Field strength mode)
Frequency response
50Hz to 8kHz ±0.25dB
30Hz to 10kHz -3dB
Gain stability: Better than 0.5dB over all conditions

• Commissioning Certificate (template)

Outputs

• Soft carry pouch

• Designing Induction Loops handbook
• Batteries
• CD of Test Signals
3 different test signals are provided on CD to enable
the set up of any induction loop system.
Tracklist
• Combination: Pink noise with 1kHz sine
		 wave bursts (30 mins)
• Pink noise (30 mins)
• Sine wave (1 min)

Standards Compliance
The FSM field strength meter and loop receiver is
CE marked to all relevant safety and EMC standards

Display:
Flying spot LED bar graph with wide viewing angle
Colour coded (green for -3dB to +3dB)
Audio (headphone):
16Ω min (32Ω per side)
3.5mm stereo jack connector

Power
Battery: 2 x AA alkaline LED (supplied)
Battery monitor: Battery OK when LED illuminated
Battery life: Up to 100hrs use, dependent on use pattern

Physical
Dimension: 84 x 27 x 140mm (meter only)
Weight: 150g excluding batteries
Operating environment:
-10 to +45˚C
10 to 85% RH

www.ampetronic.com
sales@ampetronic.com
support@ampetronic.com
phone +44 (0)1636 610062
fax +44 (0)1636 610063
Issue no UP34302-1

Northern Road, Newark NG24 2ET. United Kingdom

DISTRIBUTOR



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : Yes
XMP Toolkit                     : Adobe XMP Core 4.0-c006 1.236519, Wed Jun 14 2006 08:31:24
Create Date                     : 2007:05:23 10:37:24+01:00
Metadata Date                   : 2007:05:23 10:37:31+01:00
Modify Date                     : 2007:05:23 10:37:31+01:00
Creator Tool                    : Adobe InDesign CS3 (5.0)
Thumbnail Format                : JPEG
Thumbnail Width                 : 256
Thumbnail Height                : 256
Thumbnail Image                 : (Binary data 12713 bytes, use -b option to extract)
Instance ID                     : uuid:69a368c0-e613-4ab1-bbd1-fa17fe40a131
Document ID                     : adobe:docid:indd:3f57cd3d-03e9-11dc-bb30-c63adec0d664
Rendition Class                 : proof:pdf
Derived From Instance ID        : 3f57cd3c-03e9-11dc-bb30-c63adec0d664
Derived From Document ID        : adobe:docid:indd:c0fa6e51-c350-11db-84e1-81ea5501162e
Manifest Link Form              : ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream
Manifest Placed X Resolution    : 72.00, 314.00, 299.99, 299.99, 299.99, 72.00, 72.00, 299.99, 299.99, 299.99, 299.99, 299.99, 299.99
Manifest Placed Y Resolution    : 72.00, 314.00, 299.99, 299.99, 299.99, 72.00, 72.00, 299.99, 299.99, 299.99, 299.99, 299.99, 299.99
Manifest Placed Resolution Unit : Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches
Manifest Reference Instance ID  : uuid:780f3d17-9c6e-11da-811d-000a95b4fbbc, uuid:850CEE44C33711DBA981B5CD9B0EADD6, uuid:7BDE5CBAF57E11DBA031A9720FC26C15, uuid:7BDE5CB7F57E11DBA031A9720FC26C15, uuid:69F65DF9F57E11DBA031A9720FC26C15, uuid:7142aba4-9a5d-11da-8afc-000d935d820e, uuid:780f3d17-9c6e-11da-811d-000a95b4fbbc, uuid:7BDE5CB7F57E11DBA031A9720FC26C15, uuid:7BDE5CBAF57E11DBA031A9720FC26C15, uuid:69F65DF9F57E11DBA031A9720FC26C15, uuid:7BDE5CB7F57E11DBA031A9720FC26C15, uuid:7BDE5CBAF57E11DBA031A9720FC26C15, uuid:7BDE5CBAF57E11DBA031A9720FC26C15
Manifest Reference Document ID  : adobe:docid:photoshop:56aad482-31c3-11da-a505-a5d1208279a2, uuid:850CEE43C33711DBA981B5CD9B0EADD6, uuid:69F65DF6F57E11DBA031A9720FC26C15, uuid:69F65DF7F57E11DBA031A9720FC26C15, uuid:69F65DF8F57E11DBA031A9720FC26C15, adobe:docid:photoshop:56aad482-31c3-11da-a505-a5d1208279a2, uuid:69F65DF7F57E11DBA031A9720FC26C15, uuid:69F65DF6F57E11DBA031A9720FC26C15, uuid:69F65DF8F57E11DBA031A9720FC26C15, uuid:69F65DF7F57E11DBA031A9720FC26C15, uuid:69F65DF6F57E11DBA031A9720FC26C15, uuid:69F65DF6F57E11DBA031A9720FC26C15
Format                          : application/pdf
Producer                        : Adobe PDF Library 8.0
Trapped                         : False
Page Count                      : 2
Creator                         : Adobe InDesign CS3 (5.0)
EXIF Metadata provided by EXIF.tools

Navigation menu