BCBCAT

2015-05-08

: Pdf Bcbcat BCBCAT 04 2015 x7mag uploads wp-content

Open the PDF directly: View PDF PDF.
Page Count: 77

DownloadBCBCAT
Open PDF In BrowserView PDF
A UNIT OF BC GROUP INTERNATIONAL, INC.

2014-2015 CATALOG

✔ Cardiac Output
✔ Fetal Maternal
✔ NIBP
✔ IBP
✔ ECG
✔ Arrhythmias
✔ Performance
✔ SPO2
✔ Temp

ESU-2400H
ESU Analyzer

The Only Multi-Pulse
Analyzer

See Page 25-26

Are You IAC
Section 2.2.3B
Compliant?

NIBP-1040Kit-E
Patient Simulator
See Page 21-22

Ultrasound Transducer
Leakage Tester
See Page 47-52
BC Group International, Inc.
3081 Elm Point Industrial Dr.
St. Charles, MO 63301 USA
1-314-638-3800 or 1-888-223-6763
1-314-638-3200 Fax
Shop Online at...
www.bcgroupintl.com

History

History of BC Biomedical

BC Group was founded in 1988. In 2000, we
began manufacturing our own product line under
the now familiar Green and Gold “BC Biomedical”
label.
In January of 2005, Lloyd Industries, the company
that engineered and manufactured most of the BC
Biomedical brand products, purchased BC Group.
This acquisition has allowed our products and
services to expand at an even faster pace. BC
Group, under the “BC Biomedical” brand, is now
the second largest manufacturer of Biomedical
Test and Measurement Equipment in the world.
Our philosophy with BC Biomedical products runs
counter to the current trend of “one-size-fits-all.”
We offer families of products that provide the users
with a choice of models so they can pick the
features they need at a price they can afford.
We just introduced our new ESU-2400H
Electrosurgery Analyzer with enhanced ability to
detect waveforms with a higher sampling rate.
(Upgrade path for previous models available.)
BC Group has a major commitment to Quality at all
levels of our operation. The company is certified to
ISO 9001:2008 and ISO 13485:2003.Our Service
Center is accredited to both ANSI Z540-1 and
ISO/IEC 17025:2005 with full traceability to NIST.
We are Registered with and Inspected by the FDA
and follow the Good Manufacturing Practices
(GMP) as well as comply with the QSR (Quality
Systems Regulation) 21CFR820.
Keep checking our website for the latest additions
to the BC Biomedical Line.

SF-SLS-180 (A) BC Biomedical Catalog

1-888-223-6763 - bcgroupintl.com

Table

of

Contents

Safety Analyzers	
Patient Simulators	

1-6
7-14
15-20

NIBP Simulators	

21-24

ESU Analyzers	

25-36

Digital Pressure Meters	

37-42

Pulse Oximetry Testing	

43-44

Infusion Pump Analyzers	

45-46

Ultrasound Leakage Tester	

47-52

Ultrasound Watt­meters	

53-54

Dew Point Meter	

55-56

Anesthetic Agent Analyzer	

57-58

Multi Gas Analyzer	

59-60

Benchtop Flow Analyzer	

61-62

Handheld Flow Analyzer	

63-64

TSI Certifier 	

63-64

Lung Simulators	

67-68

Tachometer Series	

69-70

Any of the trademarks, service marks or similar rights that are
mentioned, used or cited within are the property of their respective
owners. Their use here does not imply endorsement or affiliation
with any of the holders of any such rights.

Table of Contents

Defibrillator Analyzers	

Safety Analyzer Series
1

Safety Analyzer Series

Features - SA-2000 Series
±	

Small & Lightweight Portable Analyzer

±	

Line Voltage Measurement

±	

Device Under Test Current Measurement

±	

	

Earth/Ground Lead Resistance/ 		
Leakage Current

±	

Point-to-Point Testing

±	

Enclosure/Chassis Leakage Current

±	

External Resistance

±	

External Leakage Current

±	 Source Receptacle Wiring Integrity 	
	Monitor
±	

MAP (Isolation) Function

±	

Internal ECG Simulator (SA-2010S Only)

±	

True RMS Readings

±	

	

AAMI ES1-1993 or IEC 601 Selectable 	
Test Loads

±	

90 to 264 VAC Operation

±	

Touch Control Keys - No Knobs

±	

LED Status Indicators

±	

Audible Feedback

±	

Self-Test Points

±	

Externally Replaceable Ground Fuse

±	

	

Automatic Delay on Power Polarity 	
Reversal to DUT

±	

Soft-Sided Carrying Case

±	

International Versions Available

Optional
±	

External Point-to-Point (Red) Cable

±	 Kelvin Cable (For four wire resistance
	testing)
±	

RS232 add “-R” to part number

1-888-223-6763 - bcgroupintl.com

SA-2010S
	The SA-2000 Series is a Microprocessor
based Electrical Safety Analyzer Family.
There are 5 models that provide everything
from the most basic testing functions to
the most complete set of features of any
handheld unit on the market.
	The compact size makes the SA-2000
Family easy to hold and operate. All models
operate from 90 to 264 Volts @ 50/60 Hz,
with Load Currents up to 20 Amps, so there
is no need to purchase 2 different analyzers
to cover your testing needs. The SA-2010S
is the smallest safety analyzer available with
a built-in ECG Patient simulator.
	The SA-2000 Family of Safety Analyzers
allows the user to choose the model that
best fits the application. The combination of
the best features at the best price makes this
Family of products a natural choice.
	 All models come with a soft-sided case and
Chassis (Black) cable. An optional External
Point-to-Point (Red) cable or a Kelvin cable
is also available.

Specifications

22

SA-2000

SA-2001

SA-2005

SA-2010

SA-2010S

Voltage (Rating)

90 to 264 VAC

90 to 264 VAC

90 to 264 VAC

90 to 264 VAC

90 to 264 VAC

Current (Rating)

20 A

20 A

20 A

20 A

20 A

RS232

On “-R”
Measures:

Voltage (VAC)

No

90 to 264
(± 3% R)

90 to 264
(± 3% R)

90 to 264
(± 3% R)

90 to 264
(± 3% R)

Current (A)

No

0 to 19.99
(± 5% R)

0 to 19.99
(± 5% R)

0 to 19.99
(± 5% R)

0 to 19.99
(± 5% R)

Leakage Current
(μA)

0 to 1999
0 to 1999
0 to 1999
0 to 1999
0 to 1999
DC & 25 to 1 kHz DC & 25 to 1 kHz DC & 25 to 1 kHz DC & 25 to 1 kHz DC & 25 to 1 kHz
(± 1% R)
(± 1% R)
(± 1% R)
(± 1% R)
(± 1% R)
1 kHz to 100kHz 1 kHz to 100kHz 1 kHz to 100kHz 1 kHz to 100kHz 1 kHz to 100kHz
(± 2.5% R)
(± 2.5% R)
(± 2.5% R)
(± 2.5% R)
(± 2.5% R)
100 kHz to 1 MHz 100 kHz to 1 MHz 100 kHz to 1 MHz 100 kHz to 1 MHz 100 kHz to 1 MHz
(± 5% R)
(± 5% R)
(± 5% R)
(± 5% R)
(± 5% R)

Resistance (ohms)

0 to 1.99
(± 1% R)
2 to 19.99
(± 1% R)
10 mA

0 to 1.99
(± 1% R)
2 to 19.99
(± 1% R)
10 mA

0 to 1.99
(± 1% R)
2 to 19.99
(± 1% R)
10 mA

0 to 1.99
(± 1% R)
2 to 19.99
(± 1% R)
10 mA

0 to 1.99
(± 1% R)
2 to 19.99
(± 1% R)
10 mA

Patient Leads

0

0

5

10

10

MAP (Isolation)

No

No

Yes

Yes

Yes

No

No

No

No

Yes

Yes
Yes
Yes

Yes
Yes
Yes

Yes
Yes
Yes

Yes
Yes
Yes

Built-in ECG
Simuator
AAMI & IEC Loads
Point-to-Point
Test Points

AAMI Only
No
No

Open Ground
Reverse Polarity
Open Neutral

Yes
Yes
Yes

Yes
Yes
Yes
Yes
Yes
Yes
Test Capability:
Yes
Yes
Yes
Yes
Yes
Yes

Open Line

Yes

Yes

Yes

Yes

Yes

Size

8.65 x 5.73 x
1.92 Inches

8.65 x 5.73 x
1.92 Inches

8.65 x 5.73 x
2.40 Inches

8.65 x 5.73 x
2.40 Inches

8.65 x 5.73 x
2.40 Inches

Weight

≤ 2.5 Lbs
(1.14kg)

≤ 3.0 Lbs
(1.36kg)

≤ 3.0 Lbs
(1.36kg)

≤ 3.0 Lbs
(1.36kg)

≤ 3.5 Lbs
(1.59kg)

Flip for product pricing - Page A

Safety Analyzer Series

Models:

Safety Analyzer Series
3

Features - SA-2000-INTL Series
±	

	
	

±	

All of the Same Features Available as 	
the Domestic Counterpart

±	

Universal Power Receptacle Accepts 	
Almost Any International Power Plug 	
Configuration

±	

Detachable Power Cord Allows for use 	
in 	Any Country

±	

Multiple International Configuration 	
Power Cords Available

±	

	

AAMI ES1-1993 or IEC 601 Selectable 	
Test Loads

±	

90 to 264 VAC Operation

	

Safety Analyzer Series

Five Models Available: SA-2000-INTL, 	
SA-2001-INTL, SA-2005-INTL, 		
SA-2010-INTL & SA-2010S-INTL

	
	
	

	

±	Soft-sided

	

Carrying Case & Chassis
Cable Included

Features - CS-2000 Series
±	

	
	
	

Universal Power Receptacle
with BC20-20221 Installed for
Schuko use

SA-2001-INTL
	 The SA-2000-INTL Series is ideal for international customers or for those that simply
need a power cord configuration that is different from the standard NEMA 5-15P configuration that is standard on our SA-2000
Series. Interchangeable power cords make
power input switching extremely easy and
the universal configuration power receptacle is compatible with almost all countries
around the world.

Two Models Available: CS-2000-U (US 	
Version; has US Plug & Socket) and CS-	
2000-E (European Version; has		
European “Schuko” Plug & Socket)

±	

Facilitates Testing to IEC 601 		
	Requirements
±	

	
	

Interfaces Directly to BC Biomedical 	
SA-2000 & SA-2000-INTL Series of 	
Safety Analyzers

±	
Interfaces

with Dale Technology®
LT544D, 	 LT544D Lite, 601 & 601E Safety 	
	Analyzers

	

±	

CS-2000

Universal Power Supply (90 to 264 VAC, 	 	The CS-2000 Series is a 1 Amp current
	 50/60 Hz)
source for testing the ground continuity in
equipment. This device facilitates testing to
±	 1 Amp Current Source for Resistance 	
IEC 601 requirements, where a 1 Amp source
	Measurements
current is required. This device interfaces
±	 No Protruding Push Buttons
directly with the BC Biomedical SA-2000
Series of Safety Analyzers as well as the
±	 Light Touch Key Used to Operate
Dale Technology® LT544D, LT544D Lite &
±	 Small Size
601/601E. This Series comes in 2 models, US
and European plug & socket configurations.

1-888-223-6763 - bcgroupintl.com

SA-2000-INTL Series Information

Typical Power Cord Options Available for the SA-2000-INTL Series

BC20-20402 United
Kingdom

BC20-20401
Japan

BC20-20400 North
America Hospital
Grade

BC20-20412 Italy

Pick Your Country’s Connector
from Above

BC20-20403
SchukoContinental
Europe

BC20-20416
Australia

to Connect to the
SA-2000-INTL Series

Standard Power Cord

BC20-20200 Use
with
SA-2010S INTL

Standard Product Plug

BC20-20221 Schuko Plug
Ground Adapter

CS-2000 Specifications
Electrical

CURRENT SOURCE

1.00 to 1.02 A DC

OPERATING LINE VOLTAGE

90-264 VAC, 50/60 Hz

POWER CONSUMPTION

12 VA (12 W)

LOAD CURRENT

This device is an accessory to a Safety Analyzer, refer to applicable Safety Analyzer User
Manual for this Rating

LINE PLUG
DUT
RECEPTACLE

CONNECTIONS

CS-2000-U

NEMA 5-15P

CS-2000-E

European CEE 7/7 “Schuko”

CS-2000-U

Hospital Grade NEMA 5-15R

CS-2000-E

European CEE 7/7 “Schuko”
LINE PLUG 3 Pin Locking Plug for Safety Analyzer Chassis Cable Connection (See
Description section for Compatible Safety Analyzers)
Detachable Kelvin Cable (BC20-20114) for
DUT Ground Connection
Physical

SIZE
ENCLOSURE

7.09 x 3.94 x 1.56 Inches
(180 x 100 x 40 mm)
ABS Plastic
Back-printed Lexan Overlay

WEIGHT

< 1 Lbs (0.45 kg)

OPERATING RANGE

0 to 50 °C (32 to 122 °F)
10 to 90% RH, Non-Condensing

STORAGE RANGE

-40 to 60 °C (-40 to 140 °F)

Flip for product pricing - Page B

Safety Analyzer Series

BC20-20410
Switzerland

BC20-20409 India/
South Africa

44

Safety Analyzer Series
5

Safety Analyzer Series

Features - SA-2500
±	

	

Full PC Automation with 			
Procedures and Reports

±	

Small & Lightweight Portable Analyzer

±	

Line Voltage Measurement

±	

Device Under Test Current Measurement

±	

Earth/Ground Lead Resistance/Leakage 	
	Current
±	

Point-to-Point Testing

±	

Enclosure/Chassis Leakage Current

±	

External Resistance

±	

External Leakage Current

±	

Source Receptacle Wiring Integrity 	
	Monitor
±	

MAP (Isolation) Function

±	

True RMS Readings

±	

90 to 240 VAC Operation

±	

Touch Control Keys - No Knobs

±	

Audible Feedback

±	

	

Automatic Delay on Power Polarity 	
Reversal to DUT

±	

Soft-Sided Carrying Case

±	

International Versions Available

SA-2500
	The SA-2500 is an Electrical Safety
Analyzer with Full PC Automation.
	 There are 2 Modes of operation, Manual
and Remote. The Manual mode allows
most basic testing functions to be stepped
through one at a time. With the Remote
mode, measurements are controlled via
a PC. The user has the possibility to integrate all measuremetns into the individual
operator interface at the PC and to define
test sequences.

PC Interface
(Included)

Optional
±	

	
	

±	

	

BC20-20150 – Chassis Cable (Non-	
Kelvin), optional for external 		
measurement mode
BC20-20151 – Chassis Cable (Kelvin), 	
one comes with the unit

1-888-223-6763 - bcgroupintl.com

Sample
Report
(Included)

SA-2500

Specifications

6

Electrical
Voltage (Rating)

90-240 VAC

Current (Rating)

16A

Remote Control

USB

Voltage (VAC)

90 - 240 (+/-5%)

Current (A)

0 to 16.00 (+/- 5%)

Leakage Current

10 to 300 uA(+/- 5%)300 uA to 30 mA (+/- 5%)

Resistance (Ohms)

0.01 to 30.00 (+/- 10%)

Patient Leads

10

MAP (Isolation)

Yes

Built-In ECG Simulator

No

Test Load

IEC

Point to Point

Yes

Test Points

No
Test Capability

Open Ground

Yes

Reverse Polarity

Yes

Open Neutral

Yes

Open Line

Yes

Size

12.8 x 9.8 x 3.5 Inches

Weight

< 4.4 Lbs (2 kg)
Measuring Leakage Current

Frequency response is taken into
consideration in accordance with
the diagram to the right when leak
age current is measured.

Flip for product pricing - Page B

Safety Analyzer Series

Measurements

Multi-Parameter Simulators
7

Features - PS-2200 Series
±	

Simple to Operate

±	

	

Independent Lead Outputs Produce a 	
True 12 Lead ECG Signal

±	

1, 2 or 4 Invasive BP Channels

Patient Simulator Series

±	

All BP Waveforms Available on all BP 	
	Channels
±	

49 Arrhythmia Selections

±	

Respiration & Temperature Simulation

±	

ECG Performance Waveforms

±	

Dual Graphical Displays with Backlighting

±	

Drop Down Choice Screens List all 	
Options for Parameters

	

±	

Special Power Up Feature Allows User 	
	 to 	Choose to use Default, Last or Custom
	Settings
±	

Auto Sequences for BPM, 			
Static-Pressure Levels & Performance

±	

10 AHA & European Color Coded 		
Universal Patient Lead Connectors 	
(with Enough Space Between Posts for 	
the Large Diameter Connectors)

	
	

±	

Mini DIN Connectors Compatible with	
Standard BP Cables (Optional 		
Mini DIN-M to DIN-F Cable Adapter)

±	

% Battery Life & Low Battery Indicator

	

±	

Powered by Two 9V Batteries or 		
Optional Battery Eliminator

±	

Full Remote Operation via RS232

±	

Flash Programmable, Field Upgradeable

±	

Battery Eliminator

	

	
	
	

Optional
±	SpO2

Simulation

±	

Cardiac Output

±	

Fetal/Maternal Output

±	

Remote PSR-2200

1-888-223-6763 - bcgroupintl.com

PS-2240
	The PS-2200 Multi-Parameter Patient
Simulator Series is the most comprehensive,
easy to use simulator on the market. These
new portable simulators are intuitive to use
since there are NO MORE CODES to enter.
	 Each model provides ECG, Blood Pressure,
Respiration and Temperature Simulation.
Models are available with one, two or four
independent blood pressure
outputs. All
models will support the Fetal/Maternal, SpO2,
Cardiac Output and PSR-2200 Remote
options.
	The PS-2200 Series makes viewing and
selecting the desired waveforms and parameters quick and intuitive. All operational information is available at the same time on two
cursor-based graphic displays, allowing for
easy maneuvering through parameters and
scrolling through available options.

PS-2200

Specifications
ECG NSR
30,40,45,60,80,90,100,120,
140,160,180,200,220,240,2
60,280, 300 BPM

Rate
Accuracy

± 1%
.05,.10,.15,.20,.25,.30,.35,.4
0,.45,.50,1.0,1.5,2.0,2.5,3.0,
3.5,4.0,4.5,5.0 mV

Amplitude
Accuracy

Amplitude

Pacemaker Waveforms
1,2,3,4,5,6,7,8,9,10 mV

Accuracy

± 10%

Width

0.1,0.5,1.0,1.5,2.0 ms

Accuracy

± 5%

Rate

75 BPM

Accuracy

±1%
Respiration

± 2% @ Lead II
200 times Amplitude
± 5%
Adult (80ms), Pediatric (40ms)
± 0,.05,.1,.2,.3,.4,.5,.6,.7,.8 mV

ECG Performance
Sine Wave

0.1,0.5,5,10,40,50,60,100 Hz

Square Wave

0.125, 2.000 Hz

Triangle Wave

2.000, 2.500 Hz

Pulse Wave

30,60,120 BPM; 60 ms
.05,.10,.15,.20,.25,.30,.35,.4
0,.45,.50,1.0,1.5,2.0,2.5,3.0,
3.5,4.0,4.5,5.0 mV

Amplitude
R-Wave (Havertriangle)
Rate

30,60,80,120,200,250 BPM

Width

8,10,12,20,30,40,50,60,70,8
0,90,100,120,130,140,150,1
60,170180,190,200 ms
.05,.10,.15,.20,.25,.30,.35,.4
0,.45,.50,1.0,1.5,2.0,2.5,3.0,
3.5,4.0,4.5,5.0 mV

Amplitude
Rate Accuracy

± 1%

Amplitude Accuracy

± 2% @ Lead II

R-Wave  20 hours)

Programming Software

Assign names to each of the
18 programmable keys.

Select outputs from any simulator
waveform or output group.

Compatibility
Requires a PC running Microsoft® Windows 8, Windows 7,
Windows Vista, Windows XP Home Version, Windows XP
Professional, Windows 2000 (Service Pack 2 or higher),
Windows NT, Windows 98, with a USB Port (required optional
USB to Serial Adapter).
100% compatible with the BC Biomedical PS-2200 Family
of Patient Simulators (older simulators may require a Flash
firmware upgrade, which can be accomplished in the field
with a special software update utility). The PSR-2200 can be
configured by special request to work with competitive brand
patient simulators from leading manufacturers.

The PSR-2200 comes with a Microsoft® Windows
Based Application Program that allows the user to
quickly and easily develop and upload almost any
configuration. Complex testing and trending 	routines
can easily be generated and saved.
The software is self-installing and designed to mimic
the simulator screens, so there is almost no learning
curve, just point and click. The software works with
both RS232 and USB ports and will  automatically find
the PSR-2200 and identify it. Single click uploading (writing to) and downloading (reading from) the PSR-2200
allow for full cloning  and file transfers.

Flip for product pricing - Page D

Patient Simulator Series

Key
Number

Storage Range	

Patient Simulator Series
11

Patient Simulator Series

Features - PS-2100 Series
±	

Simple to Operate

±	

	

Independent Lead Outputs Produce a 	
True 12 Lead ECG Signal

±	

One Invasive BP Channel (PS-2110)

±	

Arrhythmia Selections

±	

Resp & Temp Simulation (PS-2110)

±	

ECG Performance Waveforms

±	

Graphical Display with Backlighting

±	

Drop Down Choice Screen List All 	
Options for Parameters

±	

Special Power Up Feature Allows the 	
User to Choose to use Default, Last or 	
Custom Settings

±	

Auto Sequences for BPM, Static-		
Pressure Levels & Performance

±	

10 AHA & European Color Coded 		
Universal Patient Lead Connectors 	
(with Enough Space Between Posts for 	
the Large Diameter Connectors)

	
	
	
	

	
	
	

±	

	
	

Mini DIN Connectors Compatible with 	
Standard BP Cables (Optional Mini
DIN-M to DIN-F Cable Adapter)

±	

% Battery Life Display & Low Battery 	
	Indicator
±	

	

Powered by Two 9V Batteries or 		
Optional Battery Eliminator

±	

Full Remote Operation via RS232

±	

Flash Programmable, Field Upgradeable

±	

Battery Eliminator

Optional
±	SpO2

Simulation

1-888-223-6763 - bcgroupintl.com

PS-2110
	The PS-2100 Patient Simulator Series
offers cost effective simulation with many
features. If you only need ECG simulation,
the PS-2105 offers all of the features you
need. If you want basic Multi-Parameter
simulation, the PS-2110 is the right choice. It
provides ECG, Blood Pressure, Respiration
and Temperature Simulation, plus limited
Fetal/Maternal Simulation.
	The PS-2100 Series makes viewing and
selecting the desired waveforms and parameters quick and intuitive, with all operational
information being available at the same time
on the cursor-based graphic display, allowing for easy maneuvering through parameters and scrolling through available options.

PS-2105 Specifications

PS-2110 Specifications

ECG Functions

PS-2110

Rate

30 to 300 BPM

Amplitude

0.05 to 5mV

S-T Segment Elevation

ECG Functions
Rate
Accuracy

± 0.05 to 0.8 mV

Amplitude

Artifacts

50 Hz, 60 Hz, Muscle,
Baseline Wander, Respiration

High Level

QRS Interval

Adult (80 ms) or Pediatric (40
ms)
Arrhythmia Functions

ECG Performance
Sine Wave

2.000 Hz

Supraventricular, Premature,
Ventricular, Conduction

Rate Accuracy

± 1%

Atrial, Pacer, Asynchronous,
Non-Capture, Non-Function,
Demand - Occasional, Demand - Frequent, AV - Sequential

Pulse Height

1 to 10 mV

Pulse Width

0.1 to 2 ms
ECG-Performance Functions

Sine Waves

Amplitude Accuracy

0.125, 2 Hz

Triangle Waves

5, 2.5 Hz
30, 60, 120 BPM; 60 ms Width

Amplitude

0.05 to 5 mV (Lead II)

Automatic Mode
R Wave Rate

Yes
30, 60, 80, 120, 200, 250 BPM

R Wave Width

8 to 200 ms

R Wave Amplitude

0.05 to 5 mV (Lead II)
Training
Manual, 10, 15, 20, 25, 30 s

Randomizer

Off, On

Arrhythmias

All, Subset

SpO2 Simulation (MSP-2100 & Finger SIMTM Req’d.)

± 2% @ Lead II
Blood Pressure
0,20,40,80,100,200,250,300
mmHg

Static Pressure
Accuracy

± (2% of Reading + 2 mmHg)

Impedance

300 Ω

Accuracy

± 10%

Excitation Range

2 to 16 V RMS

Excitation Frequency

DC to 5 kHz

Sensitivity

5 or 40 µV/V/mmHg
Respiration

0.1 to 100 Hz

Square Waves

Pulse Waves

0.1,0.5,5,10,40,50,60,100 Hz
0.125, 2.000 Hz

Manual & Automatic

Waveforms

Rate

± 2% @ Lead II
200 times Amplitude

Triangle Wave

Pacemaker Functions

SpO2 Output

Accuracy

Square Wave

Triggering

Timer

0.5, 1.0, 1.5, 2.0 mV

Apnea,15,20,30,40,60,80,100,1
20 BrPM

Rate
Accuracy

± 1%

Impedance delta

0.1,0.2,0.5,1.0,2.0,3.0 Ω

Accuracy

± 10%

Baseline

500,1000 Ω

Accuracy

± 5%

Lead

LA or LL
Temperature

Temperature

30,37,40 °C (86.0,98.6,104.0 °F)

Accuracy

± 0.1 °C ( ± 0.1 °F)

Type

YSI Series 400 and 700

SpO2 Simulation (MSP-2100 & Finger SIMTM Req’d.)
Rate
SpO2 Output

30 to 300 BPM
80, 90 & 97%

30 to 300 BPM

PS-2105
80, 90 & 97%

PS-2110
with (optional)
MSP-2100
& FingerSim
TM

To C ompar e Pa tient Simulators,
s ee char t on Page 14
Flip for product pricing - Page D

Patient Simulator Series

General Groups

± 1%

49 different

Arrhythmias

12

30,60,80,120,180,240,300 BPM

ECG Simulators
13
Overview - PS-2010
The PS-2010 is the perfect tool for
providing a 10 lead wire hook up for those diagnostic 12 lead ECG machines. This small,
lightweight, feature packed simulator is a
must for your test equipment tool box.

Patient Simulator Series

	

10 AHA & European Color Coded 	
		Leads
	

±

PS-2010

	 ± Normal Sinus Rhythm & 			
		Performance Waveforms
	

±

Universal Connectors

	

±

Lead Test

Overview - PS-2006
The PS-2006 is unique in the market
place. Many of the new monitors now being
released have a six lead configuration.   By
popular demand, the PS-2006, six lead ECG/
PVC simulator was developed. Press a button and insert PVCs into the ECG trace.
	

	

±

6 AHA & European Color Coded Leads

	 ± Normal Sinus Rhythm & 			
		Performance Waveforms
	

±

Snap Connectors

	

±

PVC Insertion

PS-2006

Overview - PS-2005
	 The PS-2005 is a small unit for those who
need five lead ECG simulation only and are
looking for a cost effective simulator. Need
to check your lead wires? There is a built
in lead wire tester for your convenience.
	

±

5 AHA & European Color Coded Leads

Normal Sinus Rhythm & 			
		Performance Waveforms
	

±

	

±

Snap Connectors

	

±

Lead Test

1-888-223-6763 - bcgroupintl.com

PS-2005

Comparison Chart

Models

14

PS-2005

PS-2006

PS-2010

PS-2105

PS-2110

PS-2200

11 LED

11 LED

11 LED

1 LCD Graphical
128x64 pixels

1 LCD Graphical
128x64 pixels

2 LCD Graphical
128x64 pixels

General
Display

No

No

No

Yes

Yes

Yes

Yes

No

Yes

No

Yes

Yes

RS232

No

No

No

Yes

Yes

Yes

Power

1-9 Volt Battery

1-9 Volt Battery

1-9 Volt Battery

No

No

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Leads

5

6

10

10

10

10

Rates
Amplitudes
QRS Interval

4

4

4

17

7

17

1

1

1

4

4

19

Adult

Adult

Adult

Adult & Pediatric

Adult

Adult & Pediatric

ST Segment Elevation

No

No

No

19

No

19

ECG-Performance
Sine Waves
Square Waves
Triangle Waves
Pulse Waves

Yes

Yes

Yes

Yes

Yes

Yes

3
2

3
2

3
2

8
2

8
2

8
2

1

1

1

2

1

2

--

--

--

3

--

3

No

No

No

No

Yes

Yes

--

--

--

--

8

8

--

--

--

--

2

4

--

--

--

--

6

6

Pacemaker

No

No

No

Yes

No

No

Arrhythmias

No

PVC on Demand

No

49

12

49

Blood Pressure
Temperature

No

No

No

No

1 Channel

1, 2, or 4 Channel

No

No

No

No

YSI 400/700

YSI 400/700

No

No

No

w/MSP-2100

w/MSP-2100

w/MSP-2100

No
No

No
No

No
No

No
No

Yes
No

Optional
Optional

No

No

No

Yes

No

Yes

Yes

No

Yes

No

Yes

Yes

Battery Eliminator
ECG-NSR

Respiration
Rate
Baseline
Delta Impedance

Special Modes
SpO2 Simulation
Fetal/Maternal
Cardiac Output
Training Mode
Lead Test
Physical
Size
Weight

2-9 Volt Batteries 2-9 Volt Batteries 2-9 Volt Batteries

ABS Plastic Case ABS Plastic Case ABS Plastic Case ABS Plastic Case ABS Plastic Case ABS Plastic Case
5.12” x 2.56” x
.97”

5.15” x 2.89” x
1.06”

7.26” x 4.46” x
1.51”

8.63” x 5.5” x
1.63”

8.63” x 5.5” x
1.63”

8.63” x 5.5” x
1.63”

< 1 Lbs

≤ 0.5 Lbs
(0.23 kg)

< 1 Lbs

< 2 Lbs

< 2 Lbs

< 2 Lbs

Flip for product pricing - Page D

Patient Simulator Series

Backlight
Lead Test Terminals

Defibrillator Analyzers
15

Features - DA-2006 Series
±	

	

Pulsed Biphasic, Biphasic and 		
Monophasic Compatible

±	

Fully AED Compatible

Defibrillator Analyzer

±	Graphical Display with Backlighting
	 &	 Simultaneous Details of Parameters
	 with Scrolling Option Control
±	

On-Screen Viewing of Defib & Pace 	
	Waveforms
±	

5000 V, 1000 Joule Capacity

±	

High & Low Ranges

±	

Cardioversion Delay Measurement

±	

Charge Time Measurement

±	

Waveform Storage & Playback

±	

	

10 AHA & European Color Coded 		
Universal Patient Lead Connectors

±	

25 Pin Connector for Centronics Printer

	

±	

Power, two 9 V Batteries (Battery 		
Eliminator Furnished)

±	

Low Battery Indicator

±	

Flash Programmable, Field Upgradeable

±	

Auto Sequence Testing

	

±	

PC Utility Software for Auto Sequence 	
Development & Maintenance

±	

Storage for 50 Custom Sequences

Features - DA-2006P
±	

26 Selectable Internal Loads

±	

Full Pulse Analysis

±	

On-Screen Viewing of Pace Waveform

±	

Demand Sensitivity Test

±	

Refractory Period Tests

±	

50/60 Hz Interference Test Signals

±	

Pacer Input Defib Protection

1-888-223-6763 - bcgroupintl.com

DA-2006P
The DA-2006 and DA-2006P Defibrillator
Analyzer Series take advantage of the latest
electronic technology and deliver accurate,
consistent test results on all defibrillators,
regardless of manufacturer or model.
Whether you need to test output energy,
cardioversion delay time, maximum energy
charge time, or test your AED, the DA-2006
and DA-2006P will deliver. You can even
capture and view the actual output energy and pacer waveforms to check for any
abnormalities.
With 26 internal test loads, the DA-2006P
delivers a full range of capabilities for testing the Transcutaneous Pacemaker function of your advanced level defibrillators,
including tests like demand sensitivity, refractory
period, rate, pacer pulse width, pacer
pulse amplitude, etc. Test pacer functionality with the peace of mind that the Pacer
Input
terminals on the DA-2006P are
internally protected against accidental defibrillator
discharge.
The DA-2006 & DA-2006P offers automated
testing. Create and store up to 50 unique
testing “auto sequences”, including both defibrillator and pacer tests. You can easily edit
existing sequences and create new ones
with our unique PC-based utility software.
You can also “clone” a specific set of auto
sequences to multiple analyzers.

-2006_a

Specifications

ENERGY OUTPUT MEASUREMENT
METHOD
Monophasic, Biphasic or Pulsed Biphasic
LOAD RESISTANCE
50 Ω ± 1%, Non-Inductive (< 1 µH)
DISPLAY RESOLUTION
0.1 J
MEASUREMENT TIME WINDOW
100 ms
ABSOLUTE MAX PEAK VOLTAGE
6000 V
PULSE WIDTH
100 ms
CHARGE TIME MEASUREMENT
0.1 to 99.9 s
VOLTAGE
CURRENT
ENERGY
ACCURACY
TRIGGER LEVEL
PLAYBACK AMPLITUDE
TEST PULSE
OSCILLOSCOPE
OUTPUT ATTENUATION
CARDIOVERSION

SYNC TIME
MEASUREMENTS

LOW RANGE
≤ 1000 V
≤ 20 A
≤ 50 J
≤ 20 J ± 0.4 J
> 20 J ± 2% of reading
20 V
1 mV / 1000 V Lead 1
5 J ± 20%

1000:1

200:1

DELAY
RESOLUTION
ACCURACY
OUTPUT
SCREEN

0 to 6000 ms
0.1 ms
± 2 ms
LEAD I & PLATES
200:1TimeBaseExpansion
Starts at peak of each
R-wave
All waveform
simulations available
± 1 ms

TIMING WINDOW
TEST WAVEFORMS
DELAY TIME ACCURACY

PULSE RATE
PULSE WIDTH
VOLTAGE
OSCILLOSCOPE
OUTPUT

OUTPUT
AMPLITUDE

DEMAND
SENSITIVITY

PATIENT SIMULATOR
ECG NSR
ECG WAVEFORM
RATES

SINE
SQUARE
TRIANGLE

30, 40, 45, 60, 80, 90, 100, 120,
140, 160, 180, 200, 220, 240,
260, 280, 300 BPM
0.1, 0.2, 0.5, 5, 10, 40, 50, 60,
100 Hz
0.125, 2.000 Hz
2.000, 2.500 Hz

30, 60, 120 BPM; 60 ms
width
0.5, 1.0, 1.5, 2.0 mV (Lead II)
RATE
± 1%
AMPLITUDE
± 2% @ Lead II
OUTPUT
200 times Amplitude
ACCURACY
± 5%
80 ms

REFRACTORY
PERIOD

PULSE WAVE
AMPLITUDE
ACCURACY
HIGH LEVEL
QRS DURATION
LEAD TO LEAD
IMPEDANCE

1000 Ω

RANGE
30 to 800 ppm
ACCURACY
± 1% or 2 ppm (whichever is greater)
RANGE
0.6 to 80 ms
ACCURACY
± 1% or ± 0.3 ms (whichever is greater)
VARIABLELOADINPUT
200 V
FIXED LOAD INPUT
15 V
0 – 15 V
10.24:1
AMPLITUDE ATTEN15 – 60 V
41:1
UATION
> 60 V
164:1
MAX OUTPUT
200 V
WAVEFORMS
Square, Triangle, Haversine
WIDTH
10, 25, 40, 100, 200 ms
ECG OUTPUT
0 to 4 mV

ECG ARRHYTHMIA
SELECTIONS

ECG SHOCK ADVISORY
ALGORITHM TEST
SELECTIONS

DA-2006-VL

0 to 10
mV / 50 Ω

PACER INPUT (500
0 to 100 mV
TO 2300 OHMS &
OPEN)
DEFIBRILLATOR
0 to 10 mV
PLATES
ECG OUTPUT
40 µV
PACER INPUT (50
40 µV
TO 400 OHMS)
OUTPUT
PACER INPUT (500
RESOLUTION
1 mV
TO 2300 OHMS &
OPEN)
DEFIBRILLATOR
0.1 mV
PLATES
OUTPUT ACCURACY
± 2%
ECG OUTPUT
N/A
PACER INPUT
30 to 100 ppm
INPUT RATE
DEFIBRILLATOR
30 to 100 ppm
PLATES
PACING
20 to 500 ms
SENSING
20 to 500 ms
ACCURACY
± 2 ms
ECG OUTPUT
0,0.4,0.8,1.2,1.6,2.0,2.4,2.8,3.2,3.6,4.0mV
50 Ω

0,1,2,3,4,5,6,7,8,9,10mV
0,2,4,6,8,10,12,14,16,
18,20mV
0,3,6,9,12,15,18,21,
150 Ω
24,27,30mV
0,4,8,12,16,20,24,28,
200 Ω
32,36,40mV
0,6,12,18,24,30,36,
300 Ω
42,48,54,60mV
0,8,16,24,32,40,48,
400 Ω
56,64,72,80mV
0,10,20,30,40,50,60,
≥ 500 Ω
70,80,90,100mV
0, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10 mV
50, 100, 150, 200, 300, 400, 500, 600, 700, 800,
900, 1000, 1100, 1200, 1300, 1400, 1500, 1600,
1700, 1800, 1900, 2000, 2100, 2200, 2300 Ω
50 to 1300 Ω
± 1%
1400 to 2300 Ω
± 1.5%
4 to 300 mA (100 Ω load)
±5%or±0.5mA(whicheverisgreater)
50 – 600 Ω
300 mA
700 Ω
286 mA
800 Ω
250 mA
900 Ω
222 mA
1000 Ω
200 mA
1100 Ω
182 mA
1200 Ω
167 mA
1300 Ω
154 mA
1400 Ω
143mA
1500 Ω
133 mA
1600 Ω
125 mA
1700 Ω
118 mA
1800 Ω
111 mA
1900 Ω
105 mA
2000 Ω
100 mA
2100 Ω
95 mA
2200 Ω
91 mA
2300 Ω
87 mA
50 Ω

50/60 HZ INTER
FERENCE TEST

PACER INPUT

SIGNAL

Ventricular Fibrillation
Atrial Fibrillation
Second Degree A-V Block
Right Bundle Branch Block
Premature Atrial Contraction
PVC Early
PVC Standard
PVC R on T
Multifocal PVC
Bigeminy
Run of 5 PVCs
Ventricular Tachycardia
Asystole
Coarse Ventricular Fibrillation
Fine Ventricular Fibrillation
Multifocal Ventricular Tachycardia @ 140 BPM
Multifocal Ventricular Tachycardia @ 160 BPM
Polyfocal Ventricular Tachycardia @ 140 BPM
Polyfocal Ventricular Tachycardia @ 160 BPM
SupraVentricular Tachycardia @ 90 BPM

PACER INPUT (50
TO 400 OHMS)

DEFIB PLATES

TEST LOAD

LOAD VALUES
ACCURACY
RANGE
ACCURACY

PULSE
CURRENT

LIMIT

Our DA-2006-VL, works in conjunction with our DA-2006
Series Defibrillator Analyzers, providing Variable Loads
used when testing Defibrillators to assure the proper
electrical current is delivered to the heart, per IEC
60601-2-4 and AAMI DF80 standards. See page 19.

Flip for product pricing - Page E

16
16

Defibrillator Analyzer

WAVEFORM PLAYBACK

HIGH RANGE
≤ 5000 V
≤ 100 A
≤ 1000 J
≤ 100 J ± 2 J
> 100 J ± 2% of reading
100 V
1 mV / 1000 V Lead 1
125 J ± 20%

TRANSCUTANEOUS PACEMAKER ANALYZER

Defibrillator Analyzers
17

Defibrillator Analyzer

Programmable Autosequence
±	

PC Based Software

±	

Create up to 50 Autosequence

±	

Up to 20 Steps Per Autosequence

±	Autosequences

can be cloned

	

to Multiple Units

±	

Easily Share Autosequence Files
via Email

	

DA-2006
The BC Biomedical DA-2006 and DA-2006P defibrillator analyzers take autosequence
test development to new levels. Never before has it been so easy to create new automated testing sequences, or edit existing ones. Our unique PC-based autosequence
development software allows you to build and edit these automated testing sequences on
your PC under the power and utility of the Windows operating system, rather than having
to work on a cryptic small instrument display like other analyzers. You can create up to
50 autosequences to download to your DA-2006 or DA-2006P.
A starter set of 19 autosequences is provided. Autosequences can be defibrillator-only,
transcutaneous pacer-only, or a combination of both. Each autosequence can support
up to 20 output energy test steps, a maximum energy test (including charging time), a
cardioversion test (at up to three different power levels), and transcutaneous pacer tests.
You can take an existing autosequence and copy it to a new one, rename it, and make
any minor changes necessary to have a brand new sequence for a different make and
model defibrillator.
You can “clone” multiple DA-2006 or DA-2006P analyzers by downloading the same
autosequence file to multiple analyzers. You can e-mail autosequence file sets for use in
field or district offices away from the main office when changes are made. You can even
save your DA-2006 or DA-2006P autosequence setup and easily put it back exactly the
same way it was, prior to repair and calibration of your instrument.

1-888-223-6763 - bcgroupintl.com

Programming Auto Sequences Overview

Program Menu Bar

File Control
This section is utilized to load/save
configuration files on the PC as
well as read/write the auto configuration in the DA-2006/P

18

ECG Sequence
Programming
Use this section to easily
configure each step of the  
ECG Auto Sequence

Status Message

Com Port
being used

Sequence Configuration
Use this section to configure
each Auto Sequence test.

Compatibility
Requires a PC running Microsoft® Windows 8, Windows 7,
Windows Vista, Windows XP Home Version, Windows XP
Professional, Windows 2000 (Service Pack 2 or higher),
Windows NT, Windows 98, with a USB Port (required optional
USB to Serial Adapter).

Task
Progress
Indicator

Today’s
Date

Current Time

BC Group makes defibrillator autosequence
development and editing easier than competitive analyzers that offer such capability. Why
struggle with limited on-instrument displays
and cryptic development tools? Our DA-2006
and DA-2006P autosequence capability is in a
class of its own.

Flip for product pricing - Page E

Defibrillator Analyzer

Sequence Selection
Use this list to select
which sequence to
view /edit.

Defibrillator Analyzers

Defibrillator Analyzer

19

Features - DA-2006-VL
	

±	

Pulsed Biphasic, Biphasic and 		
Monophasic Compatible

±	

Fully AED Compatible

±	

5000 V, 1000 Joule Capacity

DA-2006-VL

Our DA-2006-VL, works in conjunction with
our DA-2006 Series Defibrillator Analyzers,
±	 Smart Loads, no Settings to Change in 	 providing Variable Loads used when testing
	 DA-2006 or DA-2006P
Defibrillators to assure the proper electrical current is delivered to the heart, per IEC
±	 25-200 ohm Loads, 25 ohm Steps
60601-2-4 and AAMI DF80 standards.

1-888-223-6763 - bcgroupintl.com

Automatic detection of loads, no settings
to change or configure in the DA-2006/P.
Ensure that the correct load is selected in the
Defibrillator output pulse measurement.

Are you testing your defibrillator with the right loads?  

20

The DA-2006 series and DA-2006-VL are the right answer for you. Not all
patients are the same, use the DA-2006-VL to test your defibrillator with
test loads from 25 to 200 ohms, as required by IEC 60601-2-24 and AAMI
DF80 standards.

If you already have a DA-2006, there is no need to purchase a new analyzer in order to measure pulsed biphasic waveforms. Simply update the
software in your DA-2006/P and you will be ready for testing.

Need to calibrate any
Defibrillator?

Typical DA-CS-06 Setup
Defibrillator Analyzer Calibration
System

Flip for product pricing - Page E

Defibrillator Analyzer

Do you need to measure Pulsed Biphasic waveforms?

NIBP Simulators
21

NIBP Simulator Series

Features - NIBP-1040-BE
	

±	

NIBP, IBP, ECG, Temp, Arrhythmias,
Respiration, Leak Detect & Much More

±	

Small, Hand-Held, Lightweight

±	

Cardiac Output

±	

Fetal Maternal

±	

Simple One Key Operation

±	

± 500 mmHg Manometer

±	

Digital Pressure Envelope Offset

±	

Total Pressure & BP Waveform Displays

±	

Adult, Neonatal, Hypertensive &
Hypotensive Modes

	

±	SpO2

Shown with
NIBP-1040-BE
Cardiac Output

Ready–Compatible with MSP-2100

	

Module & FingerSimsTM

±	

Multiple Display Screens & Digit Sizes

±	

Flash Programmable, Field Upgradeable

±	

Peak Pressure Detection with Easy Reset

±	

ECG Output with Full NSR Waveform

±	

Sinusoidal Respiration Simulation

±	

ECG Performance Test Waveforms

±	

Pace Waveform

±	

ECG Alarm Test

±	

Synchronized Invasive Blood

	

Pressure Output

±	

Selectable IBP Sensitivity - 5 or 40 μV/V/
mmHg

±	

Leak Rate Test

±	

ECG Arrhythmia Waveforms

±	

ECG Arrhythmia Sequence

±	

YSI 400 & 700 Simulation Temperature

1-888-223-6763 - bcgroupintl.com

NIBP-1040-BE
		
The NIBP-1040-BE is the newest member
of the NIBP-1000 family of high quality, low
cost, high function microprocessor-based
NIBP simulators. The NIBP-1040-BE comes
with the Cardiac Output and Fetal Maternal
feature built-in. It is designed to meet the
demand for a small, easy to use unit with
multiple features to fit your exact needs.
Though small, this complete Multiparameter
Simulator makes no compromise on providing true outputs.There is no other NIBP
Simulator on the market that offers all the
features of the NIBP-1040-BE in this price
range. The ability to provide offsets for
various manufacturers NIBP’S can only be
found in units costing twice as much.
The graphic display not only allows you to
see what is going on digitally, it also allows
you to view a plot of the overall pressure or
a close-up of the BP waveform.
The Flash Memory allows for field upgrades
with downloads from our website.

Screen Views
Main Display

Specifications
Blood Pressure
Pressure Range: 	
Accuracy:
	
Simulation Rate: 	
Accuracy:
	

Output Waveform Screen
(NOTE: The output Waveform is NOT
Physiologically Correct)

± 500 mmHg @ 20 °C
± (1% of Reading + 0.5 mmHg)
80, 94 BPM (synchronized to ECG)
±1%

ECG NSR
Rate:	
30, 60, 120, 240 BPM
	(± 1% Accuracy)
Amplitude:	
2.75 mV (± 2% @ Lead II
	Accuracy)

Pressure Graph Screen

ECG Performance
Sine Wave:	
Square Wave:	
Triangle Wave:	
Rate Accuracy:	
Amplitude:	
	

FETAL/MATERNAL

10, 60, 100 Hz
0.125, 2.000 Hz
2.000 Hz
± 1%
2.75 mV (± 2% @ Lead II
Accuracy)

Pacemaker Waveforms
Amplitude:	
3 mV (± 10% Accuracy)
Width:	
6 ms (± 5% Accuracy)
CARDIAC OUTPUT

Model Summary
NIBP-1040Kit-E 	
	
	

- Includes MCO-2100, built-in 		
rechargeable battery and Universal 		
Patient Lead Connectors

Main Test Modes:
Adult 120/80
Adult 120/80 w/Pace
Adult High 190/120
Adult Low 80/40
Neonatal 70/40
Alarm Test
Arrhythmia Sequence
Leak Test
Manometer
Cardiac Output
Fetal Maternal

Other Modes/Features:
Peak Detect
ECG Performance Waves
Individual Arrhythmias
Synchronized Dynamic
IBP
Static IBP
RS232

Respiration
Rate Accuracy:	
Impedance:	
Baseline:	

± 1%
Delta 3.0 Ω (± 10% Accuracy)
1000 Ω (± 5% Accuracy)

Temperature
Settings:	
	
Type:	

0, 24, 30, 37, 40 °C (± 0.1 °C Accuracy)	
(32.0, 75.2, 86.0, 98.6, 104.0 °F)
YSI Series 400 and 700

Fetal/Maternal
Fetal heart rate: 	
60, 90, 120, 140, 150, 210, 240 BPM
IUP Period:	
Single Contraction (Manual) 2, 3, 5, Min
	
(AUTO)
FHR
(During Contraction):	Uniform Deceleration
	
Early Deceleration
	
Late Deceleration
	
Uniform Acceleration
Cardiac Output
Injectate
Temperature: 	
0, 24 °C
Injectate Flow:	
2.5, 5.0, 10.0 L/MIN
Other Outputs:	
Faulty – Injectate
	
Left-to-Right Shunt
	
Calibrated Temperature Pulse

Flip for product pricing - Page H

NIBP Simulator Series

Invasive Blood Pressure
Static Pressure:	
-10, -5, 0, 20, 40, 50, 60, 80, 100, 		
	
120, 150, 160, 200, 240, 250, 300,
	
320, 400 mmHg
Accuracy: 	
± (1% Full Range + 1 mmHg) or
	
± (2% Setting + 2 mmHg)
Impedance:	
300 Ω (± 10% Accuracy)
Excitation Range:
2 to 16 Vrms
Excitation
Frequency: 	
DC to 5 kHz
Sensitivity:	
5 or 40 μV/V/mmHg

22
22

NIBP Simulators
23

NIBP-1030-BE

Overview - NIBP-1030-BE
The NIBP-1000 Series is a Family
of high quality, low cost, high function
Microprocessor based NIBP Simulators.
They are designed to meet the demand for
a small, easy to use unit with multiple features ti fit your exact needs.

NIBP Simulator Series

	

	 You have the option of operating by
power cord or you can add the optional
internal rechargeable NiMH Battery. You
can perform hundreds of test on a fully
charged battery.
	 Though small, this Family makes no
compromise on providing true outputs. The
optional ECG waveforms are full QRS and
the respiration waveforms look real.

NIBP-1020

	 The graphic display not only allows you
to see what is going on digitally, it also
allows you to view a plot of the overall pressure or a close-up of the BP waveform.
	 All models have the ability to adjust
the output depending on the manufactures
algorithm. (operators choice)
	 The Flash Memory allows for field
upgrades with downloads from our website.
	 NIBP-1030-BE
NIBP, Peak Detect,
Full Range Manometer, IBP, ECG, Pace,
Temperature, Arrhythmias, Respiration,
Leak Detect and Performance waveforms.
(Extended ECG post optional)

	NIBP-1020

NIBP, Peak Detect, Full
Range Manometer, ECG, Pace Respiration
and Performance waveforms.
	 NIBP-1010 NIBP, Peak Detect and Full
Range Manometer

1-888-223-6763 - bcgroupintl.com

NIBP-1010

Specifications

Screen Views

Blood Pressure
Pressure Range: ± 500 mmHg @ 20 °C
Accuracy:
	
± (1% of Reading + 0.5 mmHg)
Simulation Rate: 	80, 94 BPM (synchronized to ECG)
Accuracy:
	
±1%

Main Display

ECG NSR
Rate:	
Amplitude:	
Composite Pressure Display

30, 60, 120, 240 BPM; ± 1%
2.75 mV; ± 2% @ Lead II

Pacemaker Waveforms
Amplitude:
	 3 mV; (± 10% Accuracy)
Width:
	 6 ms; (± 5% Accuracy)
Respiration
Impedance:
Baseline:

	Delta 3.0 Ω (± 10% Accuracy)
	 1000 Ω (± 5% Accuracy)

Model Summary
AVAILABLE MODELS:
NIBP-1010 	
Basic unit
NIBP-1010-P 	
+ Peak
NIBP-1010-BP 	
+ Battery & peak
NIBP-1010-KIT 	
NIBP-1010-BP, 5 adapters & soft carrying case
NIBP-1020 	
Basic unit with ECG
NIBP-1020-PA 	
+ Peak & alarm
NIBP-1020-BPA 	 + Battery, peak & alarm
NIBP-1020-KIT 	
NIBP-1020-BPA, 5 adapters & soft carrying case
NIBP-1030
Base unit
NIBP-1030-B
+ built-in rechargeable battery
NIBP-1030-BE
+ Universal Patient Lead Connectors
NIBP-1030Kit
NIBP-1030-B, soft carrying case and 5 adapters
NIBP-1030Kit-E
Same as above with Universal Patient Lead
Connectors
NIBP-1040Kit-E
NIBP-1040-BE with Cardiac Output & Fetal
Maternal, Universal Patient Lead Connectors
soft carrying case and 5 adapters

NIBP-1010 with (optional)
MSP-2100 & FingerSim
TM

Flip for product pricing - Page I

NIBP Simulator Series

ECG Performance
Sine Wave:	
10, 60, 100 Hz; ± 1%
Square Wave:	 0.125, 2.000 Hz; ± 1%
Triangle Wave:	 2.000 Hz; ± 1%

Output Pressure Display
(NOTE: The output Waveform is NOT
Physiologically Correct.)

24

ESU Analyzer Series
25
25

Features - ESU-2400H
±	 The Industry’s Most Comprehensive and 		
	 Most Accurate Full-Featured Analyzer
±	 Measure Advanced Generator Outputs with 	
	 Pulsed RF at Up to 3 Different Amplitudes
±	 Up to 16 million samples

ESU-2400H

±	 1% Measurement Accuracy
±	 Based Upon Years of Collaborative Work 	
	 with Leading Electrosurgery Industry
	
	Manufacturers

ESU Analyzer Series

±	 Industry Standard RF Current Measurement
±	DFATM Technology Ultra High Speed Digitization 	
	 of The Complex RF Waveform
±	 Compatible with The Latest ESU Generator 	
	 Platforms by The Major ESU Manufacturers
±	 Continuous & Pulsed Output Waveform 		
	Compatible
±	 Embedded Real-Time Operating System with 	
	 ¼ VGA Color Touch Screen Display
±	 Displays Up to 15 Different Measurement 	
	 Parameters with User Selectable and 		
	
Definable Screens
±	 Internal Precision Test Loads From 0 Ω to 	
	
6400 Ω in 1 Ω Increments
±	 External Test Load Compatibility
±	 Automated Power Load Curves with Multiple 	
	 Power Settings Per Load Setting
±	 Automated User-Definable Testing 	
	Sequences

	

±	 Print test Reports to PDF format or USB Printer
±	 USB (3), RS232, and Ethernet Communications 	
	Ports
±	 External Keyboard and Mouse Compatible 	
	 Via Dedicated Ports
±	 Automatic or Manual Activation of ESU 		
	 Generator During Power Load Curve Tests
±	 Remote Communications Capability with 		
	 ESU Generators
±	
	
	
	

100% Compatible with Covidien/Valleylab 	
Force Triad & Ligasure Generators, Conmed 	
System 5000 Generators, and All Legacy 		
Generators by Other Manufacturers

±	 REM/ARM/CQM Testing Via 500 Ω Adjustable 	
	
Load in 1 Ω Increments
±	 RF Leakage Current Measurement
±	 Capture, Store, Print RF Waveform

1-888-223-6763 - bcgroupintl.com

	 The ESU-2400H offers an unprecedented buffer
depth of over 16 million samples. This allows for
ultra-stable measurements of ESU Generator
output waveforms and provides future-proofing
for new generator outputs modes such as multiple pulsed waveforms.
	 The ESU-2400H uses the same base platform
as the ESU-2400. This means that you can
upgrade your existing analyzer without having to
purchase a whole new unit. You have the freedom to get started with the ESU-2400 and then
upgrade as the need arises.
	 The low impedance internal load bank has a
range of 0 to 6400 ohms in 1 ohm increments. It
is microprocessor based and utilizes a combination of unique hardware and software to provide
accurate and reliable test results, even from
“noisy” Electrosurgical Generator waveforms
such as “Spray”. The DFATM Technology utilized in the ESU-2400H allows the system to
aggressively digitize the complex RF waveforms
produced by Electrosurgical Generators. Each
data point is analyzed to provide highly accurate
measurement results.
	The ESU-2400H,unlike most conventional ESU Analyzers, has internal high voltage
setup relays to control the measurement path.
This allows the user to switch between Power
measurements,Leakage measurements, REM/
ARM/CQM testing, or even run an autosequence
that could include any or all of these tests – without evenmoving wires around.
	 The current transformer internal to the ESU2400H senses the RF current flowing through the
internal test load and produces a ratiometric
voltage, which is digitized and analyzed by the
microprocessor. Combining the standard and low
ranges of the ESU-2400H with the use of the current transformer, the user has full control over the
ability to get high accuracy and high resolution
readings from all types of Electrosurgical
Generators

Specifications

ESU-2400H
ESU-2400
page26

MEASUREMENTS

Advanced Pulsed Mode Measurements

A/D Resolution

14 Bits

A/D Speed

64 MSPS

Bandwidth

50 kHz – 10 MHz

Measurement Accuracy

± 1% Reading
2.0 to 700.0 mA RMS (Low Range)
20 to 7000 mA RMS (High Range)

Current Range
Pulse Mode Config

0.1 mA RMS (Low Range)
1 mA RMS (High Range)

Current Resolution
Power Range (Watts)

500 Watts

Power Resolution (Watts)
Crest Factor Range

0.1 Watt
1.4 to 500

Crest Factor Resolution

0.1

0.01 mV (Low Range)
0.1 mV (High Range)

Voltage Resolution
mV Peak/Peak-to-Peak Range

0.0 to 1.0

mV Peak/Peak-to-Peak Resolution

0.1

Load Bank Specifications
Internal Setup/Load Selection
Relays

10kV, 5A rated Reed Relays

Internal Load Selection
Internal Load Range

0 to 6400 Ω

Internal Load Accuracy
Pulse Mode 2

Internal Load Power Ratings

± 1% Non-inductive
1 Ω: 25 W
2 Ω: 50 W
4 Ω: 100 W
Remaining Loads: 225 W

Load Bank Duty Cycle

10 seconds on, 30 seconds off

Load Cooling

Dual 120mm Variable Speed DC
Fans

External Load Selection
External Load Range
External Load Resolution

Pulse Mode 3

SCREEN SIZE
SETUP MEMORY
MEMORY RETENTION
OPERATING RANGE
STORAGE RANGE

Pulse Mode 3
Scope Capture

CONSTRUCTION
SIZE
WEIGHT

0 to 6400 Ω
1Ω
5.7" QVGA 18 bit color touch
screen
EEPROM, All Parameters
10 Years w/o Power
15 to 30 °C (59 to 86 °F)
-20 to 60 °C (-4 to 140 °F)
Enclosure – Aluminum
Face – Lexan, Back Printed
7.8 x 15.0 x 22.5 inches
198.1 x 381 x 571.5 mm
31 lbs. (14 kg)

CONNECTIONS

Input: I/O 4mm Safety Jacks
3xUSB, 1xSerial, 1xEthernet
1xPS/2 Keyboard/Mouse
Output: 1xBNC Scope
Hypertronics 25-pin Footswitch
connector

POWER SUPPLY ADAPTER

Input: Universal 100-240 VAC, 5060 Hz
Output: 12 VDC
(Specify power cord see page Z)

Flip for product pricing - Page G

ESU Analyzer Series

0.20 to 70.00 mV RMS (Low
Range)
2.0 to 700.0 mV RMS (High Range)

Input Voltage Range

Pulse Mode

26

ESU Analyzer Series
27
27

Features - ESU-2400
±	 Automated One Button Force Triad 		
	 Autosequence included
±	 Industry Standard RF Current Measurement
±	DFATM Technology Ultra High Speed 		
	 Digitization of The Complex RF Waveform
±	 1% Measurement Accuracy

ESU Analyzer Series

±	 Compatible with The Latest ESU Generator 	
	 Platforms by The Major ESU Manufacturers
±	 Continuous & Pulsed Output Waveform 	
	Compatible
±	 Embedded Real-Time Operating System with 	
	 ¼ VGA Color Touch Screen Display
±	 Displays Up to 15 Different Measurement 	
	 Parameters with User Selectable and 		
	 Definable Screens
±	 Internal Precision Test Loads From 0 Ω to 	
	 6400 Ω in 1 Ω Increments
±	 External Test Load Compatibility
±	 Automated Power Load Curves with Multiple 	
	 Power Settings Per Load Setting
±	 Automated User-Definable Testing 	
	Sequences

	

±	 Print test Reports to PDF format or USB 	
	Printer
±	 USB (3), RS232, and Ethernet
	 Communications Ports

	

±	 External Keyboard and Mouse Compatible 	
	 Via Dedicated Ports
±	 Automatic or Manual Activation of ESU 	
	 Generator During Power Load Curve Tests
±	 Remote Communications Capability with 	
	 ESU Generators
±	
	
	
	

100% Compatible with Covidien/Valleylab 	
Force Triad & Ligasure Generators, Conmed 	
System 5000 Generators, and All Legacy 	
Generators by Other Manufacturers

±	 REM/ARM/CQM Testing Via 500 Ω Adjustable 	
	 Load in 1 Ω Increments
±	 RF Leakage Current Measurement
±	 Capture, Store, Print RF Waveform

1-888-223-6763 - bcgroupintl.com

ESU-2400
The ESU-2400 Electrosurgical Unit
Analyzer is a high-accuracy, True RMS RF
Measurement system designed to be used in
the calibration and routine performance verification of Electrosurgical Generators. It offers
a higher degree of accuracy than previously
attainable with conventional Electrosurgical
Unit Analyzer designs. The ESU-2400 provides an advanced low inductance internal
load bank with a range of 0 to 6400 ohms
in 1 ohm increments. It is microprocessor
based and utilizes a combination of unique
hardware and software to provide accurate
and reliable test results, even from “noisy”
Electrosurgical Generator waveforms such as
“Spray”. The DFATM Technology utilized in the
ESU-2400 allows the system to aggressively
digitize the complex RF waveforms produced
by Electrosurgical Generators, analyze each
individual digital data point, and provide
highly accurate measurement results.
The ESU-2400, unlike most conventional
ESU Analyzers, has internal high voltage
setup relays to control the measurement
path, allowing the user to switch between
Power measurements, Leakage measurements, REM/ARM/CQM testing, or even run
an autosequence that could include any or all
of these tests – without even moving wires
around.
The current transformer internal to the
ESU-2400 senses the RF current flowing
through the internal test load and produces
a ratiometric voltage, which is digitized and
analyzed by the microprocessor. Combining
the standard and low ranges of the ESU2400 with the use of the current transformer,
the user has full control over the ability to get
high accuracy and high resolution readings
from all types of Electrosurgical Generators.

Specifications

Screen
Views
ESU-2400
page26

(NOTE: User can select measurement values & locations to
be displayed on each screen choosing from RMS Current,
RMS Voltage, RMS Power, Peak Voltage & Crest Factor)

MEASUREMENTS
A/D Resolution

14 Bits

A/D Speed

64 MSPS

Bandwidth

50 kHz – 10 MHz

Measurement Accuracy

± 1% Reading
2.0 to 700.0 mA RMS (Low Range)
20 to 7000 mA RMS (High Range)

Current Range
Main Screen

0.1 mA RMS (Low Range)
1 mA RMS (High Range)

Current Resolution
Power Range (Watts)

500 Watts

Power Resolution (Watts)
Crest Factor Range

0.1 Watt
1.4 to 500

Crest Factor Resolution

0.1

0.01 mV (Low Range)
0.1 mV (High Range)

Voltage Resolution
mV Peak/Peak-to-Peak Range

0.0 to 1.0

mV Peak/Peak-to-Peak Resolution

0.1

Load Bank Specifications
Internal Setup/Load Selection
Relays

10kV, 5A rated Reed Relays

Internal Load Selection
Internal Load Range

0 to 6400 Ω

Internal Load Accuracy
CQM Screen

Internal Load Power Ratings

RF Leakage Mode

Load Bank Duty Cycle

10 seconds on, 30 seconds off

Load Cooling

Dual 120mm Variable Speed DC
Fans

External Load Range
External Load Resolution

SCREEN SIZE
SETUP MEMORY
MEMORY RETENTION
OPERATING RANGE
On - Screen
Graphing

1 Ω: 25 W
2 Ω: 50 W
4 Ω: 100 W
Remaining Loads: 225 W

External Load Selection

ESU-2400 Graph Mode Example
Load Curve

± 1% Non-inductive

STORAGE RANGE
CONSTRUCTION
SIZE
WEIGHT

0 to 6400 Ω
1Ω
5.7" QVGA 18 bit color touch
screen
EEPROM, All Parameters
10 Years w/o Power
15 to 30 °C (59 to 86 °F)
-20 to 60 °C (-4 to 140 °F)
Enclosure – Aluminum
Face – Lexan, Back Printed
7.8 x 15.0 x 22.5 inches
198.1 x 381 x 571.5 mm
31 lbs. (14 kg)

CONNECTIONS

Input: I/O 4mm Safety Jacks
3xUSB, 1xSerial, 1xEthernet
1xPS/2 Keyboard/Mouse
Output: 1xBNC Scope
Hypertronics 25-pin Footswitch
connector

POWER SUPPLY ADAPTER

Input: Universal 100-240 VAC, 5060 Hz
Output: 12 VDC
(Specify power cord see page Z)

Flip for product pricing - Page G

ESU Analyzer Series

0.20 to 70.00 mV RMS (Low
Range)
2.0 to 700.0 mV RMS (High Range)

Input Voltage Range

RF Measurement Mode

28

ESU Analyzer Series
29

Features - ESU-2300
±	

	

Measurements via Industry Standard 	
Current Sensing Method

ESU Analyzer Series

±	 Built-In RF Current Transformer 		
	(Pearson® Coil)
±	

True RMS Readings Using DFATM 		
	Technology
±	 Performs Output, RF Leakage & CQM 	
	Tests

	

±	

Main Test Loads from 50 to 750 Ohms 	
In 50 Ohm Steps

±	

External Test Loads Supported

±	

Auxiliary Test Load – 200 Ohms

	

±	

Independent Variable CQM Test Load 	
1 to 500 Ohms In 1 Ohm Steps

±	

Non-Inductive Internal Load Resistors

±	

	
	

GraphicalDisplay With Simultaneous 	
Details of Parameters & Scrolling 		
Option Control

±	

Bright-White Display Backlight

±	

Rechargeable Battery or Line Powered 	
	Operation
±	

Isolated Oscilloscope Output

±	

Full Remote Operation

±	

USB Port

±	

Digital Battery Monitor

±	

Flash Programmable, Field Upgradeable

±	

Soft Touch Keys With Audible Feedback

1-888-223-6763 - bcgroupintl.com

ESU-2300
	The ESU-2300 Analyzer is for users
who prefer a conventional instrument with
internal, selectable test loads. Utilizing the
same Patent-Pending DFATM Technology as
our ESU-2050, the ESU-2300 uses industry
standard current sensing technology rather
than relying on less accurate voltage measurement techniques employed by some
competitive products.
	 The ESU-2300 uses advanced high-speed
waveform sampling techniques to accurately
analyze even the most complex electrosurgery
generator waveforms. You can easily analyze
Coag waveforms like Desiccate, Fulgurate or
even Spray with the same accuracy as pure
sinusoidal Cut waveforms. RMS current (mA)
and power (watts) can be easily read from
the large LCD graphical display, as well as
other parameters such as mV, mV peak, and
CF. A whisper quiet fan keeps the internal
non-inductive load resistors running cool.
		

Added features like CQM Testing, RF
Leakage Current measurement, a Rechargeable
Battery, USB and RS232 com ports, oscilloscope output, universal power supply and the
ability to easily update the instrument’s firmware in the field via our unique Flash Update
Utility Software put the ESU-2300 in a class
of its own.

Specifications

ESU-2300
Measurement

Industry Standard Current
Sensing, using RF Current
Transformer (Pearson Coil)

Method

Power
Range

1.0 to 400.0 Watts RMS

Resolution

0.1 Watts

High Impact Plastic, UL 94 V-0
Face-Lexan, Back Printed

Enclosure

6.30 x 13.50 x 13.40 inches
160.0 x 342.9 x 340.4 mm

Size
Weight

≤ 17.5 Lbs (7.95 kg)

± 5% Reading or ± 3 Watts
(whichever is greater)

Accuracy

Current
Range

Electrical
Power Supply

20 to 2500 mA RMS

Resolution

± 1 mA
± 2.5% Reading or ± 15 mA

Sealed Lead Acid
6 VDC, 7.2 AH

Battery

General

Limits
Bandwidth

10 kHz to 10 MHz

Crest Factor

1.4 to 500

Voltage

10,000 V Peak

Main Test Load
50 to 750 Ω

Resolution

50 Ω

Accuracy

± 1% (DC)
(10 SEC ON, 30 SEC OFF)

Auxiliary Test Load
Fixed

200 Ω

Accuracy

LCD Graphical 240 x 64
Pixels, Backlight

Display
Ventilation

Variable speed internal fan

Oscilloscope Output

Range

Duty Cycle

Input: Universal 100-240 VAC, 5060 hz
Output: 9 VDC

± 1% (DC)

Setup Memory
Memory Retention
Operating Range

Isolated (uncalibrated), BNC
Connector
EEPROM, All Parameters
10 years w/o Power
15 to 30 °C (59 to 86 °F)

Storage Range

-20 to 60 °C (-4 to 140 °F)

Humidity Limit

20 to 80% RH, Non-Condensing
Communications USB, DB9
Loads: 4mm safety jacks

Connections

CQM Test Load
Range
Resolution
Accuracy

1 to 500 Ω
1Ω
± 2% or ± 2 Ω

Internal Load Values NOT
Exactly what YOU need?
The ESU-2300 allows for the use of
External or a combination of Internal &
External Load Resistors to cover ranges
outside the provided Internal Load Range.
Setup Parameters allow the user to tell the
unit about External Loads. This ensures
the continued accurate measurement &
display of power.

Optional: External Load
Precision Power Resistor
(See Page F for a
listing of available values)

Flip for product pricing - Page F

ESU Analyzer Series

Accuracy

30
30

Physical

ESU Analyzer Series
31
Features - ESU-2050 & ESU-2050P
±	

True RMS Readings Using DFATM 		
	Technology
±	

mV, mV Peak, mA, Crest Factor and 	
Wattage Displays

±	

	
	

Large Graphical Display with 		
Backlighting & Cursor Selection of 	
Options and Setup of Parameters

±	

1% Measurement Accuracy

±	

Digital Data Output Via USB and RS232

±	

Digital Calibration - No Pots to Turn

±	

Display Contrast is Software Adjustable

±	

On-Screen Graphical Representation of
Generator Waveform with Scroll & 	
Zoom Capabilities

ESU Analyzer Series

	

	
	

±	

Standard Range Uses 0.1:1 RF Current 	
	Transformer

ESU-2050
	 The ESU-2050 ESU Analyzer is the first
instrument of its kind on the market. Designed
through extensive collaboration with several
of the world’s leading electrosurgery generator manufacturers (including the worldwide
market leader), the ESU-2050 is a highly
accurate calibration-quality instrument intended for use by OEM factory technicians and
field service engineers, as well as customers
who desire to test their ESU generators in the
exact same way the medical device manufacturers do. It is the only ESU Analyzer on
today’s market with a 1% of reading level of
accuracy.

	DFATM Technology achieves the high level of
±	 Special Low Range Uses 1:1 RF Current 	 accuracy and functionality demanded by ESU
	Transformer
manufacturers worldwide, and replaces the
Fluke® Model 8920A instrument previously
±	 Internally Protected Input Circuitry
utilized in such applications. The ESU-2050
±	 Capture, Store & Print ESU Generator 	
uses external precision load resistors and an
	 Output Waveforms with up to 32,768 	
external wide band toroidal current transform	 Discrete Data Points through the 		
er to take industry standard RF current mea	 Specialized PC Software
surements. It can function as a stand-alone
meter or in conjunction with our optional PC
±	 Internal Data Storage for Three Full ESU 	
Utility Software.
	 Waveform Data Sets
Create Customized Load Resistor Table 	 	 Measurement data, including RMS current,
RMS voltage, Peak voltage, Power (watts),
	 within the Instrument Based Upon the 	
and Crest Factor are displayed on a series
	 Load Resistors Commonly Used
of user configurable screens. You can even
±	 User Selectable Data Display Screens
display all of these values on a single screen,
including the load resistance. Digitized ESU
±	 Smallest & Lightest Weight ESU
waveforms with up to 32,768 data points can
±	 Pulsed Output Waveform Compatible 	
be stored, displayed on screen, or exported
	 (ESU-2050P Only)
to a PC for analysis. Our companion PC
Utility Software will create an Excel graph of
even the most complex output waveforms for
viewing.
±	

®

1-888-223-6763 - bcgroupintl.com

ESU-2050
page24
Screen
Views

Specifications

(NOTE: User can select measurement values & locations to
be displayed on each screen choosing from RMS Current,
RMS Voltage, RMS Power, Peak Voltage & Crest Factor)

INPUT RANGE
Voltage (RMS)

2.0 – 700.0 mV RMS

Input Resolution

0.1 mV RMS

Voltage (Peak)
One Display Zone Screen
with mV Parameter Selected

1000.0 mV

Resolution

0.1 mV

Frequency

10 KHz – 10 MHz

Accuracy

1% reading
3.3 V p-p
Internally Protected

Maximum Input Voltage
Two Display Zone Screen
with mV & Watts
Parameters Selected

CALCULATED RANGES
Current (with 0.1:1 CT)

7000 mA RMS

Resolution

1 mA
700.0 mA RMS

Resolution

0.1 mA

Wattage

10 KHz – 10 MHz
Resolution

0.1 Watt

Crest Factor

1.4 to 500

Resolution

0.1

INPUT IMPEDANCE
Four Display Zone Screen
with mV, Watts, mA & mV Peak
Parameters Selected

50 Ω

INPUT COMPATIBILITY
RF Current Transformer (50ohm)
RF Current Transformer
Attenuation

Five Display Zone Screen
with mV, Watts, mA, mV
Peak & CF Parameters
Selected

DISPLAY
MEMORY RETENTION
OPERATING RANGE

ESU-2050 Graph Mode Example

STORAGE RANGE

Fully Zoomed In

CONSTRUCTION

WEIGHT

POWER SUPPLY ADAPTER

Valleylab Force FX Blend Cut Waveform

POWER CONSUMPTION
DATA STORAGE (Internal)

Voltage (mV)

400

10 Years w/o Power
15 to 30 °C (59 to 86 °F)
-20 to 60 °C (-4 to 140 °F)
Enclosure – ABS Plastic
Face – Lexan, Back Printed

≤ 3 lbs. (1.36 kg)

CONNECTIONS

600

EEPROM, All Parameters

3.4” x 9.1” x 8.0”
86.4 x 231.4 x 203.2 mm

SIZE

PC Software Waveform Example created from the
Excel® graphing function using the over 32,000
discrete datapoints exported from the ESU-2050

0.1: 1
1: 1
User Selectable
LCD Graphical 128 X 64 Pixels

SETUP MEMORY

Fully Zoomed Out

Pearson Electronics 411 OR 4100
(Typical)

Input: BNC
Output: Serial DB-9 or USB
Input: Universal 100-240 VAC, 5060 Hz
Output: 6 VDC
ON: less than 150 mA
OFF: less than 40 µA
3 Sets of 32768 Data Points

200

	

0
-200
-400
-600
0

500

1000

1500
Sample

2000

2500

The ESU-2050 & ESU-2400 are
used and recommended
by the world’s leading
electrosurgery generator
manufacturer

3000

Flip for product pricing - Page F

ESU Analyzer Series

Current (with 1:1 CT)

Three Display Zone Screen
with mV, Watts & mA
Parameters Selected

32

ESU Analyzer Series
33

ESU Analyzer Series

Features - ESU-2000A
±	

Small & Simple to Use

±	Lightweight
±	

6 Selectable Loads from 50 to 500 ohms

±	

No Load Switching Relays

±	

Return Contact Monitor Test with 1 ohm

	Steps
±	

RF Power, Current & Leakage 		
	Measurement
±	

No Batteries or Power Cords

±	

Easy-to-Read Color-Coded Meter

±	

High Speed Fuse Protects Meter

Accessories Included
±	

ESM-1 Switching Module

±	

Red, Black & Green Patch Cords

±	 Insulated Alligator Clips & Ground 	
	adapter
±	

Spare Fuse

±	

Operator Manual

1-888-223-6763 - bcgroupintl.com

ESU-2000A
	 The ESU-2000A ESU Tester is full-featured
& versatile.Test modes include RF output
power and leakage, measured with a large
analog meter. There are different color codes
for each range at the input connector and
on the meter, making measurements quick
& easy. For Return Contact Monitor (CQM,
REM, ARM, PSS) testing, we’ve built in a 0
to 1000 ohm variable resistance with 1 ohm
digital resolution. Even more expensive ESU
testers generally don’t offer this much range
or resolution.
To improve reliability and accuracy, the
ESU-2000A has no front-panel switches.
Instead, the user simply changes the position
of our unique ESM-1 Switching Module to
select either RF Power or RF Leakage mode.
For safety, the module also blocks access to
unused input connectors, virtually eliminating
worry about improper connections & resultant
equipment damage. Plus you won’t lose the
switching module because it is conveniently
stored on the front panel at all times.

SU-2000A

Specifications
RF MEASUREMENT
CURRENT
POWER

RANGE
ACCURACY
FREQUENCY
FUSE
OSCILLOSCOPE OUTPUT

CURRENT

± 2% FS

POWER

± 5% FS

0 to 10 MHz
500 mA Fast-Blow FWH-1/2

OUTPUT
CONNECTION
VALUES

VARIABLE LOAD
POWER RATING
(MAX)

CONNECTIONS

DUTY CYCLE
VALUE
TYPE
ACCURACY
POWER RATING

Isolated, Uncalibrated
BNC (50 Ω)
50 Ω, 100 Ω, 200 Ω, 300 Ω, 400 Ω
500 Ω
Non-Inductive
± 2% ± 1 Ω
50 Ω
50 W
100 Ω
100 W
200 Ω
200 W
300 Ω
300 W
400 Ω
400 W
500 Ω
500 W
10 seconds on, 20 seconds off
200 Ω
Non-Inductive
± 2% ± 1 Ω
50 W
4mm Banana

RETURN ELECTRODE MONITOR TEST LOAD
RANGE
RESOLUTION
ACCURACY
CONNECTIONS

0 to 999 Ω
1Ω
± 3% + 3 Ω
4mm Banana
PHYSICAL & ENVIRONMENTAL

DISPLAY
CONSTRUCTION
SIZE
WEIGHT
OPERATING RANGE
STORAGE RANGE

4-1/2” Analog Meter, Color-coded display (W and mA)
ENCLOSURE
ABS Plastic
FACE PLATE
Back-printed Lexan
5.36 x 8.68 x 9.10 Inches(136.1 x 220.5 x 231.1 mm)
≤4 Lbs (1.82 kg)
15 to 30 °C (59 to 86 °F) 20 to 80% RH, Non-Condensing
-40 to 60 °C (-40 to 140 °F)

Flip for product pricing - Page E

ESU Analyzer Series

TYPE
ACCURACY

LEAKAGE LOAD

0 to 1000 mA RMS
0 to 500 W

34
34

Comparison Chart
Comparison Chart
35
ESU-2400/H

ESU-2300

ESU-2050

ESU-2050P

ESU-2000A

Yes

Yes

Yes

Yes

Yes

Test Loads

1 to 6400 Ohms, 1
ohm increments,
internal load bank

50 to 750 Ohms,
50 ohm increments, internal
load bank

Requires external
precision non-inductive test loads,
1-9999 Ohms
compatible

Requires external
precision non-inductive test loads,
1-9999 Ohms
compatible

50, 100, 200, 300,
400, 500 Ohms

RF Current Range

<2 ma to 7,000 mA 20 ma to 2,500 mA

<2 ma to 7,000 mA

<2 ma to 7000 mA

0 - 1000 mA

Models:
Industry-standard
RF current measurement

ESU Analyzer Series

RF Power (Watts RMS) Range

To 999.9 Watts

To 400 Watts

To 999.9 Watts

To 999.9 Watts

To 500 Watts

Bandwidth

50 kHz to 10 MHz

10 kHz to 10 MHz

10 kHz to 10 MHz

10 kHz to 10 MHz

0 - 10 MHz

Display

QVGA Color Touch
Screen

260 x 64 pixel LCD
Graphical Display
with Backlight

128 X 64 Pixels
LCD Graphical
with Backlight

128 X 64 Pixels
LCD Graphical
with Backlight

4.5 inch Analog
Meter

External Test Load
Compatible

Yes, can use internal + external or
external only, or
internal only

Yes, can use interYes, requires exter- Yes, requires external + external or
nal test loads
nal test loads
external only, or internal only

Yes

External Current Transformer
Required

No

No

Yes

Yes

No

Continuous Output Waveform
Compatibility

Yes

Yes

Yes

Yes

Yes

Pulse Output Waveforms
Compatibility

Yes

No

No

Yes

Yes

2400H Only

No

No

No

No

1 % of reading on
RMS Current

2.5 % of reading
on RMS Current, 5
% for reading RMS
Power

1% reading, > 50
mV, up to 1 MHz,
3% reading, > 50
mV, 1 to 10 MHz

1% reading, > 50
mV, up to 1 MHz,
3% reading, > 50
mV, 1 to 10 MHz

± 5% of Full Scale

Yes

No

Yes

Yes

No

1.4 to 500

1.4 to 500

1.4 to 500

1.4 to 500

N/A

Covidien/Valleylab Force Triad
Calibration Procedure
Compatibility

Yes

No

Yes

Yes

No

Covidien/Valleylab Ligasure
Generator Calibration
Compatibility

Yes

No

Yes

Yes

No

Number of parameters
measured/displayed

15

5

15

15

x2

Yes, 500 ohm load
adjustable in 1
ohm increments

Yes, 500 ohm load
adjustable in 1
ohm increments

No, requires external CQM Test Box
(CQM-2000)

No, requires external CQM Test Box
(CQM-2000)

Yes

No

No

No

No

Multiple Pulsed Waveforms
Compatibility
Measurement accuracy
Calibration level accuracy
Crest Factor Range

Internal REM/ARM/CQM
Testing

Yes, "x" to "y" Ohms
in "z" ohm increments OR by userAutomated power-load curves defined load table
values, multiple
power levels @
each load possible

1-888-223-6763 - bcgroupintl.com

Comparison Chart
Comparison Chart
36
36
ESU-2300

ESU-2050

ESU-2050P

ESU-2000A

Yes

No

No

No

No

User definable testing
autosequences

Yes, unlimited steps
with unlimited ascii
text description per
step, embedded
measurement
capability within
sequence

No

No

No

No

Communication ports

USB, RS232,
Ethernet, PS/2

USB, RS232

USB, RS232

USB, RS232

No

Capture & Store RF
Waveform

Yes

No

Yes

Yes

No

RF Leakage Current
Measurement

Yes

Yes

Yes, with external
200 ohm test load
and CT

Yes, with external
200 ohm test load
and CT

Yes

Remote control compatibility
of ESU generator

Yes

No

No

No

No

User definable data screens

Yes

Yes

Yes

Yes

No

FLASH field upgrade of
firmware

Yes

Yes

Yes

Yes

No

Test load temperature
monitor

Yes

Yes

No

No

No

Oscilloscope Output Jack

Yes

Yes

No

No

Yes

User-adjustable signal
averaging

Yes

Yes

Yes

Yes

No

Test form printout

Yes

No

No

No

No

Browse Internet

Yes

No

No

No

No

Battery Operation

No

Yes

No

No

N/A

External keyboard & mouse
compatible

Yes

No

No

No

No

Fan Cooling of Internal Load
Resistor Bank

Yes

Yes

No

No

No

7.8 x 15.0 x 22.5
inches
198.1 x 381 x
571.5 mm

6.30 x 13.50 x 13.40
inches
160.0 x 342.9 x
340.4 mm

3.4 x 9.1 x 8.0
inches
86.4 x 231.4 x
203.2 mm

3.4 x 9.1 x 8.0
inches
86.4 x 231.4 x
203.2 mm

5.36 x 8.68 x 9.10
inches
136.1 x 220.5 x
231.1 mm

31 lbs. (14 kg)

≤ 17.5 lbs (7.95 kg)

≤ 3 lbs. (1.36 kg)

≤ 3 lbs. (1.36 kg)

≤ 4 lbs (1.82 kg)

Auto-activation of ESU
generator for power load
curve testing

Size

Weight

Flip for product pricing - Page E-F

ESU Analyzer Series

ESU-2400/H

Models:

Pressure Meters
37
37

Features - DPM 2300 Series

Digital Pressure Meter Series

±	 Large Graphics Display With Cursor
	 Selection Of Options And Setup
±	 High Resolution 24 Bit Measurement
±	 ± 0.05% FS Pressure Accuracy
±	 Standard Pressure Scales Include 13
	 Different Engineering Unit Ranges
	 (PSI, mmHg@0 °C, mmHg@20 °C, inHg@0 °C,

	
inHg@20 °C, cmH2O@20 °C, inH2O@4 °C,
	inH2O@20 °C, inH2O@60 °F, kg/cm2, kPa, mBar, Bar)

±	 MAX And MIN Pressure Value Capture
	 And Storage
±	 Digital Calibration & Zero Adjust
±	 Selectable Display Options & Digit Sizes
±	 Battery Life Display (0 To 100%)
±	 Programmable Digital Averaging Filter
±	 Display Contrast Is Software Adjustable
±	 Digital Zero Adjustment

DPM-2302 Adds:
±	 RS232 Serial Communications

Second Pressure Sensor Adds:
±	 Independent Pressure Channel
±	 Separate And Combined Display Option

Temperature Option Adds:
±	

YSI 700 or 100 Ω RTD Temperature Probe
	Interface

±	

-20.0 to 100.0 °C (-4.0 to 212.0 °F) Range

±	

± 0.5% FS Accuracy

Analog Output Option Adds:
	

±	

DC Output Proportional to the 		
Displayed Pressure or Temperature

±	

± 0.1% FS Output Accuracy

1-888-223-6763 - bcgroupintl.com

DPM-2300
	The DPM-2300 Series is a microprocessor based, High Precision, High Resolution
Digital Pressure Meter Family. These meters
measure both gas and liquid pressures and
provide multiple engineering unit displays
for the results. The unit may have one or
two pressure sensors and an optional temperature sensor input to measure pressure and
temperature all in one meter.
	 There are five standard pressure ranges.
Models are available with any combination of
ranges. The DPM-2350 Subset provides a
differential pressure option using a single
sensor in any of the five ranges. Optional
features allow the addition of RS232
communication, Temperature measurement
and DC Analog Outputs.

Model Summary

Specifications

38
38

PRESSURE MEASUREMENT (GAUGE)

RANGE

MAXIMUM
OVERPRESSURE

-13.5 TO 100.0 PSI
-13.5 TO 75.0 PSI

10 PSI SENSOR

-10.0 TO 10.0 PSI

5 PSI SENSOR

-5.0 TO 5.0 PSI

0.3 PSI SENSOR

-0.3 TO 0.3 PSI

100 PSI SENSOR

0.001 PSI

75 PSI SENSOR

0.001 PSI

10 PSI SENSOR

0.0001 PSI

5 PSI SENSOR

0.0001 PSI

0.3 PSI SENSOR

0.00001 PSI

100 PSI SENSOR

200 PSI

75 PSI SENSOR

200 PSI

10 PSI SENSOR

45 PSI

5 PSI SENSOR

15 PSI

0.3 PSI SENSOR

5 PSI

ACCURACY
DIGITAL FILTER
COMPATIBLE
MEDIA
CONNECTIONS
DISPLAY
CONSTRUCTION
SIZE
WEIGHT
OPERATING RANGE
STORAGE RANGE
BATTERY

± 0.05% FS
0 to 10 seconds, Selectable
Only non-corrosive, non-ionic, or otherwise pure
fluids and/or gases that are compatible with sensor
materials including glass, silicon, ceramic, epoxy,
RTV, gold, aluminum and nickel.
See Model Number Breakdown
PHYSICAL & ENVIRONMENTAL
128 X 64 Pixels Non-Backlit Graphical LCD
ENCLOSURE
ABS Plastic
OVERLAY

Back-printed Lexan

7.69 x 3.97 x 1.80 Inches
(195.3 x 100.8 x 45.7 mm)
< 1 Lbs (0.45 kg)
15 to 30 °C (59 to 86 °F)
-40 to 60 °C (-40 to 140 °F)
ELECTRICAL & MISC.
9V Alkaline Battery (ANSI/NEDA 1604A or equivalent)
9 VDC, 200 mA

BATTERYELIMINATOR
BC20-21110 BATTERY ELIMINATOR ( UNIVERSAL)
ON
< 35 mA
OFF
< 40 µA
CONTINUOUS
80 hours
BATTERY LIFE
OFF
1 year
BAUD
115200
DATA BITS
8
START BITS
1
STOP BITS
1
PARITY
None
HANDSHAKING
None
RS-232
Seven (7) pin Mini-DIN
COMMUNICATIONS
Receptacle
Pinout:
POWER
CONSUMPTION

Special Model:

Recommended model for use with TSI
Certifier-FA Respiratory Vent Test System:
DPM-235175NNFC
or DPM-23017510NFC

CONNECTIONS

DPM-2350
w/Female CPC

Flip for product pricing - Page O

Digital Pressure Meter Series

RESOLUTION

100 PSI SENSOR
75 PSI SENSOR

Pressure Meters
39
39

Digital Pressure Meter Series

Features - DPM-2200 Series
±	

Single or Dual Pressure Sensors

±	

	

Large Graphics Display with Cursor 	
Selection of Options & Parameters Setup

±	

± 0.05% FS Pressure Accuracy

±	

16 Bit Measurement

±	

Standard Pressure Scales Include 13
Different Engineering Unit Ranges

	
	

(PSI, mmHg@0 °C, mmHg@20 °C, inHg@0 °C,
	
inHg@20 °C, cmH2O@20 °C, inH2O@4 °C,
	inH2O@20 °C, inH2O@60 °F, kg/cm2, kPa, mBar, Bar)

±	

Digital Calibration & Zero Adjust

±	

Selectable Display Options & Digit Sizes

±	

Battery Life Display (0 to 100%)

±	

Programmable Digital Filter

±	

Display Contrast is Software Adjustable

±	

Flash Programmable, Field Upgradeable

	

±	

Optional DC or High Frequency (up to 	
100 Hz Cycle Rate) Outputs

±	

Pressure & Vacuum

±	

Liquids & Gases

DPM-2202 Adds:
±	 Min/Max Pressure Value Capture and 	
	Storage
±	

RS232 Serial Communications

Second Pressure Sensor Adds:
±	

Independent Pressure Channel

±	

Separate & Combined Display Option

Temperature Option Adds:
±	

YSI 700 or 100 Ω RTD Temperature Probe
	Interface

±	

-20.0 to 100.0 °C (-4.0 to 212.0 °F) Range

±	

± 0.5% FS Accuracy

Analog Output Option Adds:
	

±	

DC Output Proportional to the 		
Displayed Pressure or Temperature

±	

± 0.1% FS Output Accuracy

1-888-223-6763 - bcgroupintl.com

DPM-2200
	 The DPM-2200 Series is a Microprocessorbased, High Precision, Digital Pressure Meter
Family. They measure both gas and liquid
pressures and provide multiple
engineering unit displays for the results. The
unit may have one or two pressure sensors
and an optional temperature sensor input
to
measure pressure and temperature all
in one meter. An analog output option is also
available.
	 There are five standard pressure ranges.
Models are available with any combination of
ranges. The DPM-2250 Subset provides a
differential pressure option using a single
sensor in any of the five ranges. Optional
features allow the addition of RS232
communication, Temperature measurement
and DC Analog Outputs.

Specifications
PRESSURE MEASUREMENT (GAUGE)

RANGE

RESOLUTION

100 PSI SENSOR
75 PSI SENSOR

-13.5 TO 100.0 PSI
-13.5 TO 75.0 PSI

10 PSI SENSOR

-10.0 TO 10.0 PSI

5 PSI SENSOR

-5.0 TO 5.0 PSI

0.3 PSI SENSOR

-0.3 TO 0.3 PSI

100 PSI SENSOR

0.01 PSI

75 PSI SENSOR

0.01 PSI

10 PSI SENSOR

0.001 PSI
0.001 PSI
0.0001 PSI

ACCURACY
DIGITAL FILTER
COMPATIBLE
MEDIA
CONNECTIONS
DISPLAY
CONSTRUCTION
SIZE
WEIGHT
OPERATING RANGE
STORAGE RANGE
BATTERY

± 0.05% FS
0 to 10 seconds, Selectable
Only non-corrosive, non-ionic, or otherwise pure
fluids and/or gases that are compatible with sensor
materials including glass, silicon, ceramic, epoxy,
RTV, gold, aluminum and nickel.
Male Luer Coupler
PHYSICAL & ENVIRONMENTAL
128 X 64 Pixels Non-Backlit Graphical LCD
ENCLOSURE
ABS Plastic
OVERLAY

MODEL INFORMATION

Model Summary

Use the following model configuration guide to construct or decode a DPM220X series
differential Digital Pressure Meter model number:
DPM

220X

XX

XX

XX

XX
Options
Blank = Not Applicable
DC = DC (Direct Current) Analog Output
HF = High Frequency DC Analog Output

Back-printed Lexan

7.69 x 3.97 x 1.80 Inches
(195.3 x 100.8 x 45.7 mm)

Temperature Sensor Option
N
= Not Applicable
Y7
= YSI 700
R1 = 100 Ω RTD

< 1 lbs (0.45 kg)
15 to 30 °C (59 to 86 °F)

Pressure Range (Sensor 2)
N = Not Applicable
100 = Max 100 PSI
75 = Max 75 PSI
10 = Max 10 PSI
5 = Max 5 PSI
.3 = Max 0.3 PSI

-40 to 60 °C (-40 to 140 °F)
ELECTRICAL & MISC.
9V Alkaline Battery (ANSI/NEDA 1604A or equivalent)
9 VDC, 200 mA

Pressure Range (Sensor 1)
100 = Max 100 PSI
75 = Max 75 PSI
10 = Max 10 PSI
5 = Max 5 PSI
.3 = Max 0.3 PSI

BATTERYELIMINATOR
BC20-21110 BATTERY ELIMINATOR ( UNIVERSAL)
ON
< 35 mA
OFF
< 40 µA
CONTINUOUS
80 hours
BATTERY LIFE
OFF
1 year
BAUD
115200
DATA BITS
8
START BITS
1
STOP BITS
1
PARITY
None
HANDSHAKING
None
RS-232
Seven (7) pin Mini-DIN
COMMUNICATIONS
Receptacle
Pinout:
CONNECTIONS
POWER
CONSUMPTION

Model
2201 = Basic Model
2202 = Adds Min/Max and RS-232

DPM

225X

XX

N

XX

XX
Options

Blank = Not Applicable
DC = DC (Direct Current) Analog Output
HF = High Frequency DC Analog Output

Temperature Sensor
N = Not Applicable
Y7 = YSI 700
R1 = 100 Ω RTD

3

Pressure Sensor 2
N = Not Applicable
Pressure Sensor 1
N = Not Applicable
100 = Max 100 PSI (CPC Connector)
75 = Max 75 PSI (CPC Connector)
10 = Max 10 PSI (Luer Connector)
5 = Max 5 PSI (Luer Connector)
.3 = Max .3 PSI (Luer Connector)
2250 Series
2251 = Basic Model (Includes Min/Max)
2252 = RS232 Add
Digital Pressure Meter

OPTIONAL ACCESSORIES FOR DPM SERIES
BC20
BC20
BC20
BC20
BC20
BC20

-

21110	
41361 	
30106	
01005	
01006	
01008	

Digital Pressure Meter Series

5 PSI SENSOR
0.3 PSI SENSOR

40

DPM-2250

BATTERY ELIMINATOR (UNIVERSAL)
RS232 COMMUNICATION CABLE, USB(A)M TO MINI DIN MALE
BC BIOMEDICAL SMALL SOFT SIDED CARRYING CASE
UNIVERSAL PRESSURE ADAPTER KIT
YSI TEMPERATURE PROBE
RTD (100 Ω) TEMPERATURE PROBE

Flip for product pricing - Page P

Pressure Meters

Digital Pressure Meter Series

41
41

Features - DPM-2000 Series
±	

Pressure Ranges:

	

- 13.50 to 100.00 PSI

	

- 701 to 5190 mmHg @ 20 °C

	

- 951 to 7043 cmH2O @ 20 °C

	

- 374 to 2773 inH2O @ 20 °C

±	±

0.1% FS Pressure Accuracy

±	

5 Digit LCD Display

±	

Programmable Digital Filter

±	

Digital Calibration & Zero Adjust

±	

Male Luer Connector

±	

Battery Life Display (0 to 100%)

±	

Liquids & Gases

DPM-2001PLUS Adds:
±	

Min/Max Value Capture

±	

RS232 Serial Communications

DPM-2100 Adds:
±	

YSI-700 Temperature Probe Interface

±	

0.0 to 100.0 °C (32.0 to 212.0 °F) Range

±	

± 0.5% FS Accuracy

1-888-223-6763 - bcgroupintl.com

DPM-2001PLUS
	 The DPM-2000 Series is a Microprocessor
based Family of Meters that offers not only
positive and negative pressure sensing of
compatible liquids and gases, but also has
models with Temperature and RS232.
	The DPM-2001 is the base model. The
DPM-2001PLUS adds a “Min/Max” feature
plus a RS232 port. The top-of-the-line DPM2100 is a
pressure/temperature meter. It
has all the features of the PLUS model and
also includes a YSI 700 temperature probe
interface.
	 Each model has a battery life display (0 to
100%), Software adjustable LCD contrast,
digital calibration (no pots to turn), .1% of full
scale accuracy, -13.50 to 100.00 PSI range,
digital zero adjust and four pressure scales:
PSI, inH2O, cmH2O and mmHg.

Specifications

42
42

PRESSURE
-13.50 to 100.00 PSI
-701 to 5190 mmHg @ 20 °C
-374 to 2773 inH2O @ 20 °C
-951 to 7043 cmH2O @ 20 °C

RANGE

MAXIMUM
OVERPRESSURE

200 PSI

ACCURACY

± 0.1% FS

DIGITAL FILTER
COMPATIBLE
MEDIA

1 – 255 samples
Only non-corrosive, non-ionic, or otherwise pure
fluids and/or gases that are compatible with sensor
materials including glass, silicon, ceramic, epoxy,
RTV, gold, aluminum and nickel.

CONNECTIONS
DISPLAY
CONSTRUCTION

Male Luer Coupler
PHYSICAL & ENVIRONMENTAL
5 Digit, 7-Segment Non-Backlit LCD
ENCLOSURE
ABS Plastic

MEMORY
SIZE
WEIGHT
OPERATING RANGE
STORAGE RANGE
BATTERY
BATTERY
ELIMINATOR

OVERLAY

Lexan, Back Printed

SETUP

EEPROM, All Parameters

RETENTION

10 Years w/o Power

7.69 x 3.97 x 1.82 Inches (195.3 x 100.8 x 46.2 mm)
≤ 1 Lbs (0.45 kg)
0 to 50 °C (32 to 122 °F)
-40 to 60 °C (-40 to 140 °F)

DPM-2001

ELECTRICAL & MISC.
9V Alkaline Battery (ANSI/NEDA 1604A or equivalent)
9 VDC, 200 mA

BC20-21110 BATTERY ELIMINATOR (UNIVERSAL)
ON
< 15 mA
OFF
< 40 µA
CONTINUOUS
80 hours
BATTERY LIFE
OFF
1 year
BAUD
115200
DATA BITS
8
START BITS
1
STOP BITS
1
RS-232
PARITY
None
COMMUNICATIONS
HANDSHAKING
None
1/8 Stereo Receptacle
for use with
CONNECTIONS
BC20-21102 RS-232
Cable
POWER
CONSUMPTION

Model Summary
BC Model	
DPM-2001	
DPM-2001PLUS	
DPM-2100	

Description
Basic Unit
Basic Unit + Capture and RS232
Basic Unit + Capture, RS232 and Temp

Optional Accessories
BC20 - 21110	
Battery Eliminator (Universal)
BC20 - 21102	
RS232 Communication Cable
BC20 - 30106	
Soft Carrying Case
BC20 - 01005	
Universal Adapter Kit

DPM-2100

Flip for product pricing - Page P

Digital Pressure Meter Series

0.01 PSI
1 mmHg @ 20 °C
1 inH2O @ 20 °C
1 cmH2O @ 20 °C

RESOLUTION

Pulse Oximetry Testing
43
FingerSims
Replacement Set
FS-1000RS
TM

Pulse Oximetry Testing

Features - FingerSimsTM
±	

Easy to Use SpO2 Simulator

	

±	

Quick Testing of Pulse Oximetry
Systems including Sensors

±	

80, 90 & 97% Saturation Levels

±	

Inexpensive & Accurate

Features - MSP-2100
±	

Accepts the FingerSimsTM to Provide 	
SpO2 Outputs

	
	

±	

Pulse Synchronized with PS-2100, 	
PS-2200 & NIBP-1000 Series Products
via Aux Port

±	

Powered via Aux Port

	

Features - SPO-2000
±	

	

Accepts the FingerSimsTM to Provide 	
SpO2 Outputs

The FingerSimsTM tests your SpO2 monitor
and finger sensor as a system and works very
similar to a human finger, providing for variable
absorption and reflectance of visible red and
infrared light. The dye-based technology of the
FingerSimsTM is a proven simulation technique for
generating accurate SpO2 saturation levels. Plus
FingerSimsTM work with any style Pulse Oximetry
Finger Sensor. The FingerSimsTM provide SpO2
outputs of 80, 90 & 97%. Simply push on the
FingerSimsTM to generate a “pulse” or include a
calibrated pulse rate by adding the SPO-2000 or
the MSP-2100 Pulse Oximetry Module.

MSP-2100
with FingerSims

TM

	 The MSP-2100 can plug into the Auxiliary port of
BC Biomedical’s PS-2100, PS-2200 & the NIBP1000 Series providing a calibrated pulse rate.
(FingerSimsTM Required, but Not Included.)

±	 30, 60, 90, 120, 150 & 180 BPM Pulse 	
	Rates
±	

External AC Adapter

±	

Compact & Cost-Effective

±	

Maintenance Free

SPO-2000
with FingerSims

TM

1-888-223-6763 - bcgroupintl.com

	 The SPO-2000 is a stand-alone device that provides a calibrated pulse rate for the FingerSimsTM.
This allows the FingerSimsTM to be used as a
hands-free test system. (FingerSimsTM Required,
but Not Included.)

TM

FingerSims Specifications

Saturation Nominal Values	
80,90 & 97%
These are the nominal Red-to-Infrared Ratios (AC)
@ 72.5 F & 660 nM/910 nM
Light Absorption	
	

Infrared: 10 to 40 dB
Red: 15 to 40 dB

Typical Infrared Percent	
Modulation when Squeezed

0 to 5%

44
44

Long Term Storage Temp Range	 32 to 104 °F
65 to 90 °F

MSP-2100 Specifications

FingerSims Starter Kit
FS-1000-SK
TM

Accuracy	

±1 BPM

Power	

Aux Port

Weight	

≤ 1 Lbs (0.45 kg)

Size	

2.3 x 3.6 x 1.5 Inches

SPO-2000 Specifications

NIBP-1010 with MSP-2100 & FingerSims

Accuracy(Rate)	

± 1%

Power	

AC Adapter

Weight	

≤ 1 Lbs (0.45 kg)

Size	
	

5.2 x 5.2 x 2.5 Inches
(132.1 x 132.1 x 63.5 mm)

TM

Wouldn’t it be nice to have
calibrated fingers?
?%

Pulse Oximetry Testing

Operating Temp.	

?%
97%
90%
80%

PS-2110 with MSP-2100 & FingerSims

TM

Flip for product pricing - Page I

BC Biomedical
FingerSims TM

Infusion Pump Analyzer
45
45

35 mL
Chamber

Infusion Pump Analyzer

Features - IPA-1000 Series
±	

Easy to Use

±	

Easy to Clean

±	 Simultaneous Operation of Two 		
	Chambers

3.5 mL
Chamber

IPA-1000

±	 Large Graphics Display with Cursor 	
	 Selection of Options & Setup of
	
	Parameters
±	

Auto Chamber Recognition

±	

2 Chamber Sizes (3.5 & 35 mL)

±	

Out of Sequence Sensing

±	

Battery Display (0 to 100%)

±	

	

Similar Function to Dynatech Nevada 	
Model 404A

±	

Ranges 0 to 999.9 mL/hr & 0 to 9999 mL/hr

±	±

1% of Reading Flow Accuracy

±	

Digital Calibration – No Pots to Turn

±	

Software Adjustable Contrast

±	

Replaceable Tubes

±	

Sealed Level Sensors

±	

Programmable End of Test Audio

±	 Programmable Auto or Manual Test 	
	Start

Optional
±	

Battery Eliminator

1-888-223-6763 - bcgroupintl.com

	The IPA-1000 Infusion Pump Analyzer
Family is designed to test the flow rates of
intravenous infusion pumps to ensure their
correct operation. It is able to test up to
two pumps simultaneously by the volumetric method in milliliters per hour. All tests
are controlled by a Microprocessor, which
calculates and displays the results.
	 The fluid path is unobstructed, making
them easy to clean. The tubes are plastic,
not glass so they are rugged. The cables
and chambers are interchangeable. The
unit auto-identifies each chamber when it
is plugged in. The chamber calibration is
stored electronically in each chamber, so
all units are interchangeable. This unit is
designed to be tough, simple, portable & very
user friendly.

Specifications

Model Summary

Flow Rate Tests

BC Model	

Description

BC10-03103	
		

IPA-1000, (1) 3.5 mL Chamber,
(1) Interconnect cable

BC10-03104	
		

IPA-1000, (1) 35 mL Chamber,
(1) Interconnect cable

BC10-03105	
		

IPA-1000, (2) 3.5 mL Chambers,
(2) Interconnect cables
IPA-1000, (2) 35 mL Chambers,
(2) Interconnect cables

BC10-03107	
		
		

IPA-1000, (1) 3.5 mL Chamber,
(1) 35 mL Chamber,
(2) Interconnect cables

Individual Components
IPA-1000 	
MC3.5-1000 	
MC35-1000 	

Infusion Pump Analyzer Base Unit
3.5 mL Chamber
35 mL Chamber

Performs two independent tests simultaneously

±

Controls tests automatically, starting, stopping & 	
resetting according to the position of fluid in chamber

	
±

	
	

Ranges: Automatically selected when chamber is 	
plugged in 0-999.9 mL/hr with 3.5 mL chamber 0-	
9999 mL/hr with 35 mL chamber

±	 Accuracy:

		

Volumetric chamber: 3.5 or 35 mL(nominal)
Overall: ± 1% of reading ± 1 digit
Display

±

Graphic LCD, 128 X 64 pixel
Power Requirements

±

9V Alkaline

±

Optional AC adapter

Optional Accessories
BC20-30109 	
Soft Carrying Case
BC20-40607	
IPA-1000 Interconnect Cable
BC20-21110	
Battery Eliminator (Universal)

Optional Soft Carrying Case
shown with Chambers mounted in
shock-resistant inserts. Analyzer
& Accessories may be stored in
upper compartment.

Flip for product pricing - Page J

Infusion Pump Analyzer

BC10-03106	
		

±

46
46

Ultrasound Leakage Tester
47

Ultrasound Leakage Tester Series

Features - ULT-2000 Series
±	

Tests the Upper & Lower Leakage 	
Current Limits per Ultrasound Mfr. 	
Specifications

±	

Range 0.5 to 500 µA - Designed to Meet 	
Mfr. Transducer Specifications

±	

Large Graphic Display with Backlighting 	
Allows Easy Selection of Options & 	
Setup of Parameters

	
	
	

	
	

±	

	
	

User-Selectable Challenge (Test) 		
Voltage (90 to 275 VAC) & Frequency 	
(50 or 60 Hz)

±	

User-Selectable Test Limits

±	

Programmable Test Limits Based on 	
Ultrasound Transducer Mfr. & Type

	

±	±1%

F.S. Range Accuracy

±	

	

Auto Ranging for Enhanced Accuracy 	
Over Entire Range

±	

Single Button Press for Full System Test

±	 Selectable Pass/Fail or Numerical Test 	
	Results
±	

User-Selectable Display Options

±	

Programmable Backlight Timer

±	

Flash Programmable, Field Upgradeable

±	

	

Compatible with Dale Technology® & 	
Fluke Biomedical® Adapters & Probes

±	

RS232 Interface

±	

	

PC Utility Software for Configuration 	
Setup & Remote Control

±	

Test Results Printable w/Optional Printer

±	

On-Board Clock & Calendar Function 	
for Date/Time Stamp of Test Records

	

±	On-Board

	Records

Storage for up to 99 Test 	

±	

Meter Mode

±	

Audio-Visual Test Status Indication

1-888-223-6763 - bcgroupintl.com

Are You IAC
Section 2.2.3B
Compliant?

ULT-2000
The ULT-2000 Series is specifically designed
to test the electrical safety of all types of
diagnostic ultrasound transducers, totally
independent of the ultrasound machines
on which they are typically used. Although
the ULT-2000 can be used on virtually any
type of ultrasound transducer, it is especially recommended in the testing of TEE
(Transesophageal Echocardiography) transducers prior to each use; as recommended
by many TEE ultrasound device manufacturers. The ULT-2000 tests the integrity of the
outer insulation barrier of the transducer as
well as the capacitive leakage currents that
exist.
The ULT-2000 Series is the most advanced
instrument of its kind on the market today
and adds a totally new dimension to diagnostic ultrasound transducer electrical safety
testing. With features and functionality that
far surpasses competitive products from
other manufacturers, the ULT-2000 is easy
to set up and use. Operating modes include
a simple PASS/FAIL mode as well as a
QUANTITATIVE mode that offers actual
readings. You can print test results to an
optional printer.
Compatibility with Dale Technology®
DALE800, Fluke Biomedical® ULT800 adapters and Dual Conductivity Probes allow
you to upgrade to the very latest
technology platform, while safeguarding your
prior investment in adapters and probes. A
wide variety of BC Biomedical adapters and
accessories are available.

Specifications

ULT-2000
Screen
Views
Typical Passing Results Display

Source Voltage
Leakage Current
DUT Information
Manufacturer: Model

Conductivity Current
Display

Progress Indicator
Test Measurements

Setup Memory
Memory Retention

Typical Passing Results Display Choices

Operating Range

Pass/Fail Mode

EEPROM, All Parameters
10 years w/o Power
15 to 30 °C (59 to 86 °F)

Construction

Enclosure - ABS Plastic
Face - Lexan, Back Printed

Size

Power Consumption
Battery

Battery Life

Quantitative Mode & Pass/Fail Mode
Test your TEE Transducers While
They are being Disinfected in a Commercial Cleaner

LCD Graphical w/Backlighting
128 x 64 Pixels

-40 to 60 °C (-40 to 140 °F)

Weight

Typical Failed Results Display

0.5 to 500 µA, ± 1% FS

Storage Range

or

Quantitative Mode

0.50 to 10.00 µA, ± 0.5 µA
10 to 250 µA, ± 1% FS
250 to 500 µA, ± 1% FS

Battery Eliminator (Optional)

7.27 x 3.97 x 1.80 Inches
(184.7 x 100.8 x 45.7 mm)
< 1.1 lbs (0.68 kg)
On: < 300 mA
Off: < 250 µA
9V Lithium
(ANSI/NEDA 1604LC or equivalent)
Continuous: > 100 Full Tests
(Note: Backlight set to OFF)
OFF: > 1 Year
BC20-21110 Battery Eliminator
(Universal)

Typical Test Setup for Testing Your Ultrasound Transducers with a
Transducer-Specific Adapter & a Soak Tray.

BC20-42300
Optional
ULT-2000 Printer
ULT-PA-29 Adapter
Flip for product pricing - Page C

48
40

Ultrasound Leakage Tester Series

Test Procedure:
SOURCE TEST to verify voltage
used during test.
CIRCUIT TEST to verify
measurement circuitry.
BATH CONDUCTIVITY TEST to
validate cleaning solution.
PROBE LEAKAGE TEST to
measure the leakage current of
the DUT.

90 to 275 VAC, ± 1% FS

Ultrasound Transducer Adapter Cross Reference

Ultrasound Leakage Tester Series

49

BC Adapter
Part #

Manufacturer

Transducer Compatibility

Ultrasound
Platform

Connector
Key
Micro-pinless (PC board)

ULT-PA-10

Siemens/ Acuson

V5M (TEE), V7M (TEE), EV8-C4, etc.

Sequoia

Micro-pinless (PC board)
ULT-PA-11

SonoSite

ICT7-4, ITC8-5, C60, etc.

180/180+, Titan

260-pin ITT Cannon ZIF
ULT-PA-12

ULT-PA-13

Siemens

6.5EV13, EC9-4, etc.

Acuson/Siemens 156-pin and V510B
Siemens/ Acuson Transducers, and ATL UM4, UM9 and ATL
5MHz Bi Plane

Call for Details

128XP

156-pin ITT Cannon ZIF

260-pin ITT Cannon ZIF
ULT-PA-14

Philips/ATL

HP/Agilent/Philips 21311A, 21369A,
21378A, 21381A Transducers

Philips/HP 4500,
Sonos 5500/7500

260-pin ITT Cannon ZIF

ULT-PA-16

GE

GE LogiQ (for use with GE LogiQ Transducers 3, 5, 7, 9 and GE Vivid Transducers 3, 5, 7, 6T, 9T and GE P9603AU
Transducers), 6T (TEE), 9T (TEE), E8C)

LOGIQ, Vivid &
GE P9603AU

260-pin ITT Cannon ZIF

ULT-PA-17

Philips/ATL

Philips iE33 and iU22 diagnostic ultrasound TEE transducers - S7-2 (TEE), S73t (TEE), S3-1, C8-4v, C9-5, etc. with
belled housing

ULT-PA-18

Philips/HP

HP/Agilent/Philips (no center post on
transducer connector) 21202A, 21364A,
21365A, 21366A, 21367A Transducers

To download the current cross reference sheet in PDF format, go to:
http://testequipmentandtools.com/acatalog/ULT2000SeriesReference.pdf

1-888-223-6763 - bcgroupintl.com

iE33/iU22

Call for Details

Micro-pinless (PC board)

Ultrasound Transducer Adapter Cross Reference

ULT-PA-19

ULT-PA-20

Manufacturer

Philips

Transducer Compatibility

All Philips CX50 Laptop System based
transducers

Ultrasound
Platform

CX50 Platform

Acuson/Toshiba (for use with Acuson/
Siemens XP, Aspen, Capasee, 3-Needle
Toshiba PVF Series,
Acuson/ Toshiba
Guide C3 Transducers; ATL 3.5 DFT
2B701-753E, etc.
Transducers; Toshiba PSF-37HT and F
Series Transducers)

Connector
Key
260-pin Tyco Electronics Mini
Connector

156-pin ZIF

260-pin ITT Cannon ZIF
EUP-ES52M, EUPSeries Probes in
general

Hitachi

HI VISION 900, 5500, 6500, 8500, EUB2000, EUB-525, EUB-405 Plus

ULT-PA-22

Ultraschallkopf Aloka

UST-934N/9395, UST-945BP/945BP,
ASU-32-3-M, ASU-32-WSJ, UST556/5512, UST-5514DTU

SSD-620, SSD-650

ULT-PA-23

ALL

ALL Transducers

All Platforms

ULT-PA-21

260-pin ITT Cannon ZIF

ALL Configurations
260-pin ITT Cannon ZIF

ULT-PA-24

Philips/ ATL

L7-4 and similar 260-pin transducers with
belled housing

Call for Details

260-pin Tyco Electronics Mini
Connector
ULT-PA-25

GE

GE LogiqBook Probes - GE VIVID I 6T,
9T, etc. Probes

Vivid

360-pin ITT Canon ZIF
ULT-PA-26

Acuson

For use with Acuson/Siemens 360-pin
Transducers

Antares

ULT-PA-27

GE

GE YMS/RT (for use with GE YMS/RT
Transducers)

Call for Details

Call for Details
168-pin Proprietary Custom
Connector

ULT-PA-29

ULT-PA-30

Zonare

	 GE Vivid E9

Zonare Medical Systems (all transducAll Zonare Platforms
ers)

	 GE Vivid E9

Call for Details

To download the current cross reference sheet in PDF format, go to:
http://testequipmentandtools.com/acatalog/ULT2000SeriesReference.pdf

Flip for product pricing - Page C

50

Ultrasound Leakage Tester Series

BC Adapter
Part #

“Soft Touch” ULT Adapter Family Continues to Grow

Ultrasound Leakage Tester Series

51

Did you know that all ultrasound transducer
adapters are not created equally? There are
basically two types of adapters available in
the market today. The traditional adapters
available from other manufacturers utilize
a conductive rubber pad within a rigid body
that captures the center mounting post of
the transducer and pushes the pins of the
transducer electrical connector into the surface of this relatively rigid conductive rubber pad. This action creates stress on the
individual pins of the transducer connector
and can actually bend or break pins under
certain conditions.

Old Rubber Pad

New “Soft Touch” Connector

1-888-223-6763 - bcgroupintl.com

Ultrasound manufacturers are aware of the
existence of this type of adapter, and they
will typically not provide warranty repairs to
transducers with bent or broken pins caused
by the use of these adapters. In fact, most
ultrasound manufacturers frown upon the
use of this type of adapter because of the
physical stress placed upon the pins of
the electrical connector. Replacement of a
transducer electrical connector due to bent
or broken pins can be a very costly experience.
The ULT line of second generation “Soft
Touch” transducer adapters, available from
BC Group International, completely eliminates the possibility of costly damage.
Our second generation adapters utilize the
actual multi-pin mating connector for the
ultrasound transducer to be tested. The
transducer electrical connector is subjected
only to the same level of insertion and locking force that is normally seen when the
transducer is connected to the ultrasound
machine. There are no other undesired
mechanical forces on the transducer electrical connector, or the individual pins within it.
So why not take steps to avoid unnecessary
and costly transducer electrical connector
damage as a result of using older generation adapters? Second generation improved
design adapters are now available from BC
Group International.

Ultrasound Transducer Adapter Reference

ULT-PA-11

ULT-PA-12

ULT-PA-13

ULT-PA-14

ULT-PA-16

ULT-PA-17

ULT-PA-18

ULT-PA-19

ULT-PA-20

ULT-PA-21

ULT-PA-22

ULT-PA-23

ULT-PA-24

ULT-PA-25

ULT-PA-27

Conductivity Probes

ULT-PA-29

ULT-PC-20

ULT-PA-30

ULT-PC-10

ULT-PC-25
Flip for product pricing - Page C

ULT-PC-15

ULT-PC-30

Ultrasound Leakage Tester Series

ULT-PA-10

52

Ultrasound Wattmeters
53
53
Features - USP-100A
±	 Digital Reading
±	 Continuous or Pulsed Power
±	 Integrated Carrying Case

Ultrasound Wattmeters

±	 0.2 to 30.0 Watt Range
±	 Accuracy = ± 3%+One Count Reading
±	 0.5 to 10 MHz Frequency Response
±	 3” Max Transducer Diameter
±	 Built-in Transducer Holder
±	 Degassed Water is the Testing Medium
±	 Line Power or Battery Operation

Features - USP-50SP
±	

Digital Reading

±	

0 to 30 Watts Range

±	

± 3% + One Count Accuracy

±	

2.5 Second Integration

±	

0.05 Watt Display Sensitivity

Features - USP-30
±	

Analog Reading

±	

0.15 to 30.0 Watt Range

±	

Continuous or Pulsed Power

±	

1 to 10 MHz Frequency Response

±	

Integrated Carrying Case

±	

Built-in Transducer Holder

1-888-223-6763 - bcgroupintl.com

USP-100A
	The USP-100A Ultrasound Power Meter
was developed to test therapeutic
transducers according to AIUM, JCAHO, FDA
and NIST guidelines. The compact and light
weight USP-100A meets all the requirements of a portable test instrument with the
reliability and proven performance of the
Series.
	The USP-50SP is designed to measure
the ultrasound power output of diagnostic or
therapeutic transducers up to 30
watts. The water tank requires only one pint
of
de-gassed distilled water. If degassed water is not available, use distilled
water. The principle of measurement is the
radiant force method.
	
	The USP-30 is designed to measure the
ultrasonic power output of diagnostic or
therapeutic transducers up to 30 Watts using
the radiation force balance method. The
USP-30 consists of a precision mechanical
balance, which measures from 0.01 to 10.00
grams, a conical target
suspended on a
triangular hanger, a water tank, a transducer
support rod with clamp and a carrying case.
The transducer to be tested is center mounted over the conical target with the transducer’s radiating area immersed facedown
in the water medium. This system provides
a sturdy instrument with good measurement
accuracy and economical cost.

USP-100A Specifications

USP-50SP Specifications
Power Range 				
0 to 30 Watts
Resolution 				
+ 50 mW, ± .01 gram
Minimum Detectable Power 		
0.05 Watts
Display Sensitivity 			
0.05 Watts
Accuracy 				
± 3% @10 Watts
Stabilization 				
2.5 Second Integration
Maximum Transducer Diameter 	
4 1/2” Inches
Transducer Operating Frequency 	
0.5 to 10 MHz
Test Media 				
Degassed Water
Display					
7 segment 6 Digit 0.6”
					(15mm) high LCD
Battery		
			
4-AA: battery life over 40 hrs.; “Lo-Bat” Indicator
Power Supply 				
120 VAC to 12 VAC 500 mA adapter with 6 Ft. cord
Size 					
7.0” x 10.5” x 8.5” (H x L x W)
Weight 				10 Lbs

USP-30 Specifications
Power Range 				
150 mW to 30 Watts
Resolution 				
± 150 mW
Minimum Detectable Power Level	
± 150 mW
Minimum Accuracy			
± 5%
Expected Accuracy 			
± 4%
Display					Vernier Dial
Display Sensitivity 			
± 0.01g or ± 150mW
Zeroing Method 			
Manually Adjusted
Test Tank Lining	
		
0.5” neoprene rubber
Test			 		Medium Degassed, 					
					
distilled water at normal room temp
Test Tank Liquid Capacity	
	
1,350 mL
Test Tank Dimensions 		
Outside 5.75” Dia. x 5.55”
Inside					
4.5” Dia. x 5.25” D
Cone Target	 			
3.5” Dia. x 2” H
Dimensions(w/Case)			
19” H x 17” L x 7” W
Power					Not required
Weight(w/Case)			18 Lbs

Flip for product pricing - Page L

Ultrasound Wattmeters

Measure Range 			
0.2 to 30 Watts
Resolution 				
± 200 mW
Accuracy 				
± 3% +One Count Reading
Transducer Frequency			
0.5 to 10 MHz
Maximum Transducer Diameter 	
3”
Test Medium				
Degassed, distilled water
					at normal room temp
Stabilization Time			
2.5 sec
Display					3-Digit LCD, 1.3” High
Line Power				
120 or 240 VAC Plug-in AC Adapter
Battery 				
4-AA, 20 hrs. continuous battery life
Dimensions				
8.5” H x 12” L x 8.75” W
Weight 				11 Lbs

54

Dew Point Meter
55
55

Features - AMM-15
±	 Microprocessor Based Portable Battery 	
	 Operated Instrument

Dew Point Meter

±	 Digital Display - 40 to +80 °F Dew Point
±	 ± 2 °F From + 20 TO +80 °F
±	 ± 4 °F From - 40 TO +20 °F
±	 Samples Compressed Air Up to
	 120 PSIG
±	 Quick Connect Fittings, Connection 	
	 Tubing and Other Adapters Supplied
±	 Utilizes Connect Fittings, Filter, 		
	 Connection Tubing and Other Adapters 	
	Supplied
±	 Utilizes HC-610 Thin Film Polymer 	
	 Integrated Water Vapor Sensor
±	 RS-232 Computer Interface, Data Logger
±	 Logs 2,860 Readings, Selectable 		
	 Logging Rate

1-888-223-6763 - bcgroupintl.com

AMM-15
	 AMM-15 portable Dew Point Meter(Air-Line
Moisture Monitor) is designed to measure
the performance of drying systems for air,
nitrogen and other compressed gases.
	The model AMM-15 measures dew point
by utilizing microprocessor circuitry to convert the voltage output of OHMIC’s HC-610
precision water vapor sensor and a temperature sensor to dew/frost point temperature by the use of psychrometric equations.
Assembled in a rugged hand-held ABS
plastic enclosure and equipped with quick
disconnect tubing couplings and an internal
sensor manifold. An orifice maintains line
pressure for true pressure dew point readings. Air or industrial gases are sampled
from receivers, pipes, and hoses up to 120
PSIG and 185ºF.
	
	 In addition, heat sink tubing is available for
sampling from drying systems and ovens up
to 550ºF ( depending on ambient temperature). The AMM-15 can also be connected to
a vacuum pump for sampling from piping,
ovens, tanks or manifolds operating at ambient or low pressure.

AMM-15

56
AMM-15 Specifications
Water Vapor Sensor

HC-610 (Integrated Thin Film Polymer)

Temperature Sensor

10K NTC Thermistor

Dew Point Range

-40 to +80 °F (-40 to +27°C)

Dry Bulb Temperature Range

0-200 °F (-18 to 93°C)
±4 °F Dew point from -40 to +20ºF
±2 °F Dew point from+20 to +80ºF

Dew Point Accuracy
Dry Bulb Accuracy

±1 °F
185°F. Heat Sink Tubing Permits Maximum Inlet
Temperature to 550 ºF (depending on ambient temp.)

Sensor Response Time
Display Type

10 Seconds for 63% Step Change
3-1/2 Digit LCD; 0.5"

Analog Outputs

0 to 1 Vdc

Power

6 VDC, (4) AA Batteries & 120 Vac Adapter

Battery Life

35 Hours Nominal

Inlet Sample Port

Quick Disconnect for 1/4” OD Tubing

Inlet Pressure

120 PSIG Max

Computer Interface

Bi-Directional RS-232. ASCII Delimited Format,
9600 Baud, 8 Bits, 1 Stop, 1 Start, No Parity

Software Required

HyperTerminal or PROCOMM

Data Logger

32 KByte, Non-Volatile Memory

Internal Clock

MM:DD:YY HH:MM:SS Military Time Format

Dimensions

8.25" x 4.0" x 1.6", ABS Plastic Enclosure

Weight

2 Lbs Net

ORDERING INFORMATION
AMM-15: Portable Dew-Point Meter/Logger with carrying case, connection
parts, filter, 120 Vac adapter.
In addition, OHMIC Instruments has dew/frost point station monitoring units
with alarm features, portable dew point monitors with built-in sampling pumps,
and kits for economically testing and installation in OEM dryer panels.
Flip for product pricing - Page T

Dew Point Meter

Maximum Inlet Temperature

Anesthetic Agent
Analyzer

Anesthetic Agent Analyzer

57
57

Features - AA-2005
±	

State-of-the-art non-dispersive 			
	 infrared 	 (NDIR) technology identifies 	 	
	 and measures five anesthetic agent 	
	
	gases:
	

Halothane, Enflurane,

	

Isoflurane, Desflurane, and

	

Sevoflurane

±	

Also measures O2, CO2, and N2O

±	

Automatic or manual modes

±	

Gas flow sampled at 100, 150 and

	

200 ml/min

AA-2005
	A powerful monitoring tool – proprietary
state-of-the-art digital 5-agent gas analysis
technology. Versatile, reliable, and incredibly
compact. Designed for testing Anesthetic
vaporizer output concentration in hospitals
and outpatient surgical centers.

±	 Measurement

accuracy is not affected
by 	 alcohol or ketones

	The system’s reliable performance, ease
Fast warm-up time ensures full accuracy 	 of use, flexible design, and affordable cost
within minutes
makes it the ideal service tool for anesthesia
±	Auto-calibration
service applications in hospitals and surgical
±	 Bright, real-time display of waveforms 		
centers.
	 and numerical values provided
±	

±	 Lightweight, portable design provides
flexible workspace options

	

• Automatically identifies and quantifies five

±	

	

Service tool for anesthesia service 		 	 anesthetic agents.
application (NOT FOR CLINICAL USE)
• Detects and quantifies mixed agents.

±	

VSI-1466 Now Available

• Compatible with color VGA displays and
	 has large screen output capability.

1-888-223-6763 - bcgroupintl.com

Specifications

58
58

Display

Gas Monitoring

Measurement Range: 		

Halothane: 0 to 10%

					

Enflurane: 0 to 10%

					

Isoflurane: 0 to 10%

					

Desflurane: 0 to 20%

					

Sevoflurane: 0 to 10%

					CO2: 0 to 12.5%

Active color TFT
5.5” Display Area
Internal Display; 320 x 240 pixels
640 x 480 pixels
3 maximum
7; membrane-activated
Push and rotate; 24 steps/turn

Com Ports: 		
				

Com 1; RS232-compatible,
Serial DB-9

				
				

Com 2; Serial/Analog, mini-		
DIN8

Video Port:		

Color VGA compatible,

				

High Density D-Sub 15

	

Alarms
Alarm Characteristics: 	
				
				

EN 475, Adjustable; with 		
audible and visual
indications.
High, Medium, Low, Informational

					

Agents: ± (0.1% abs. + 4% reading Alarm Levels: 		
CO2: ± 0.2% abs. or 4% of reading Alarm Modes:		
				
N2O: ± (1.5% abs. + 4% of reading 				

					

Oxygen: ± 3 vol% (for 0-90%),

					

± 4 vol% (for 91-99%)

					N2O: 0 to 99%
					
Measurement Accuracy: 		
					

Oxygen: 0 to 100%

Time to Detect Agent: 		

< 15 seconds @ 200ml/min

Agent Detection Resolution: 	

0.1 Volume Percent

Mixed Gas Threshold: 		

0.2 vol. % +10% of total

					concentrations
Rise Time: 			

Agents: 		

450 msec

					CO2: 		

350 msec

					N2O: 		

400 msec

					Oxygen: 		600 msec
Respiration Rate
Range: 				
Accuracy: 			

1 to 60 breaths/minute

Trends
Memory: 		
				
				
Display: 			
Power Requirements
Voltage: 		
				

24 hours of stored data, 		
History available with
external VGA display
Tabular, Graphical
100, 120, 220, 240 VAC;
50/60 Hz

Power Consumption: 	

40 W, typical

Battery: 			

NiMH; Life: 1 hour, typical;

				

Recharge time: 3 hours

Mechanical
Weight: 			

Size: 			
± 2 breaths/minute or 2% of reading 				

System Features

Adult/Pediatric/Neonate High 	
and low limit settings for each
mode.

				

≤13 Lbs (5.9 kg )
16.5cm (H) x 27.9cm (W) x 		
30.5cm (D);
6.5” (H) x 11.0”(W) x 12.0” (D)

Occlusion Clearing: 		

Automatic

Environmental

Auto Zeroing: 			

Occurs 30 to 60 minutes

Operating Temperature: 	

15 to 35 ºC; 59 to 95 ºF

Duration: 3.0 to 7.0 seconds

Storage Temperature: 	

-5 to 50 ºC; 23 to 122 ºF

					

					
Manual user calibration not 		
					required.

Operating/Storage Humidity: 	 15% to 95%, noncondensing

Altitude: 		
					
Temperature stabilized optical 	 				
					assembly
Type of Protection:	
					
Auto-calibration; verification 		
Degree of Protection: 	
					recommended
					every 6-months.
Warm-up Time: 			

1 minute to first waveforms;

					

< 20 minutes to full accuracy

-300m to 3000m
(-1,000 to 10,000ft.)
Class I Equipment
Type CF, Defibrillator-proof

Protection against ingress:	 Ordinary

Flip for product pricing - Page Z

Anesthetic Agent Analyzer

Main Display: 		
Method: 			
Sidestream; 			
					Non-dispersive 			
Size: 			
					infrared (NDIR)
Resolution: 		
Identified Gases: 			
Halothane,Enflurane, Isoflurane, 	
External Video; 		
					
Desflurane, 			
					
Sevoflurane, CO2, N2O, 		
Number of Traces: 	
					Oxygen
Controls
Concentration Units:		
Vol%, Torr, kPa, mmHg
Keys: 			
Flow Rates: 			
100 ml/min, 150 ml/min, 200 	
Rotary knob: 		
					ml/min
System Outputs
Agent Detection

Multi-Gas Analyzer

Multi-Gas Analyzer

59
59

Features - MGA-3050
Displays anesthetic agent output 	
	 in 	percentage values
±	

±	

Easy to use

±	

Fast response time

±	

Very compact

Connects directly to common gas
	outlet
±	

1-888-223-6763 - bcgroupintl.com

MGA-3050 Multi Gas Analyzer
	
The
MGA-3050
Multi-gas
Analyzer connects to any PFC3000 series Flow Analyzer and
is used to measure anesthetic
agent concentrations for testing purposes. The MGA-3050
measures Halothane, Enflurane,
Isoflurane, Sevoflurane and
Desflurane inpercentage values.
The MGA-3050 also measures
percentage values of CO2 and
N2O. The MGA-3050 is an excellent instrument for performing vaporizer efficacy tests, but a PFC-3000
series unit is required, since it
uses its display to display measurement values.

60

Specifications
Range
CO2
N2O
Measurements

HAL, ISO, ENF

DES

0 - 10%

± (0.2% ABS + 2% REL)

10 - 20%

± (0.3% ABS + 4% REL)

0 - 100%

± (2% ABS + 2% REL)

0 - 8%

± (0.15% ABS + 5% REL)

8 - 12%

± (0.2% ABS + 10% REL)

0 - 10%

± (0.15% ABS + 5% REL)

10 - 15%

± (0.2% ABS + 10% REL)

0 - 22%

± (0.15% ABS + 5% REL)

22 - 25%

± (0.2% ABS + 10% REL)

Response Time CO2 < 90 ms, N2O, AA < 300 ms, O2 < 300 ms
Numerical data available with the FlowAnalyzer
Monitoring

Physical Data

Environmental
Data

Numerical data and real-time curves available with FlowLab software
Interface

through RS-232 port

Weight

< 25 g (excluding cable)

Size

1.49 x 1.14 x 1.22 inches (38 x 29 x 31 mm)

Operating

10 - 40 °C (50 - 104 °F)

Storage

-20 - 50 °C (-4 - 122 °F)

Humidity

10 - 95%, non-condensing

Atm. pressure

700 - 1200 mbar (3048 m)

CE marked according to the 93/42/EEC MDD
Compliance
and Approvals

ISO 11196:1997, EN 864:1996, EN 12598:1999
ISO/DIN 21647:2003,
ISO 7767, ASTM-F 1452-92, ASTM-F 1456-92 and ASTM-F 1462-93

Flip for product pricing - Page W

Multi-Gas Analyzer

SEV

Accuracy

61
61

Benchtop Flow Analyzer
Respiratory Parameters
16 respiratory parameters can be calculated
including PEEP, VTI and Compliance

Benchtop Flow Analyzer

Bidirectional Flow Measurement
Two measuring ports evaluate flow, pressure, temperature, humidity and O2
Pressure Measurements
All pressure information included with up to 6
different pressure sensors
Gas Standards
13 gas standards and 10 gas types adapt the
unit to the tested device
RT-200 Emulation Mode
The RT-200 emulation mode simulates the
RT-200 operating mode while offering a contemporary replacement
Data Storage
Internally stores all measured and respiratory
parameters in order to simplify the testing
procedure
Battery Operation
Convenient and independent work when you
are on the go
USB, RS-232 and External Trigger
Communicates with your test software and
ventilator

Optional Multi-Gas Analyzer

The optional MGA-3050 Sensor offers instant
gas concentration measurements of CO2, N2O,
Halothane, Enflurane, Isoflurane, Sevoflurane
and Desflurane

1-888-223-6763 - bcgroupintl.com

PFC-3000
	The PFC-3000 series of instruments
measures flow, pressure, temperature,
humidity and O2concentrations bidirectionally. The one-of-a-kind Adult, Pediatric
and High Frequency ventilation measuring
modes make the PFC-3000 the ideal calibration tool for all ventilators, anesthesia
machines and spirometers.
	The PFC-3000 distinguishes itself from
other calibration tools by combining a
simple, intuitive multilingual user interface
with the highest precision. With the push
of a button, all measured values can be
stored directly on the PFC-3000 and later
retrieved for documentation purposes.

62

Specifications
Flow

Pressure

Oxygen
Temperature
Dew point
Air humidity

Pressure
Additional Measuring Values
Concentration
Pressure compensated
High Flow Port
High Flow Port
High Flow Port

CO2

Concentration

N2O

Concentration

HAL, ISO, ENF

Concentration

SEV

Concentration

DES

Concentration

Gas types

Respiratory Parameters
TI,TE

Vti, Vte(@Flow Low)
Vti, Vte(@Flow High)

Minute volumes

Vi, Ve

Pressure
Peakflow
Compliance

Ppeak, Pmean, PEEP, Pplateau
Peakflow Insp./Exp.
Cstat

Trigger

Adult, Pediatric, HFO

Graphic display

Communication Interfaces
Calibration
Conditions
Approvals

Ambient temperature
Humidity

bidirectional
yes
yes
yes
yes
± 300 L/min
± 1.75%* or ± 0.1 L/min**
± 20 L/min
± 1.75%* or ± 0.04 L/min**
0 – 145 PSI
± 1%* or ± 0.15 PSI**
± 112.5 mmHg
± 0.75%* or ± 0.08 mmHg**
0 – 3.75 mmHg
± 1%* or ± 0.01 mmHg**
0 – 112.5 mmHg
± 0.75%* or ± 0.08 mmHg**
0 – 862.5 mmHg (abs)
± 1%* or ± 3.75 mmHg**
± 750 mmHg
± 0.5%* or ± 1.5 mmHg**
L/min, L/s, cfm, mL/min, mL/s
bar, mbar, cmH 2O, inH2O, Torr, inHg,
hPa, kPa, mmHg, PSI
Range
Accuracy
0 - 100%
± 1% O2**
yes
0 - 50°C
± 1.75%* or ± 0.5°C**
-10 - 50°C
± 2%* or ± 1°C**
0 - 100%RH
± 3%**
0 - 10%
± (0.2% ABS + 2% REL)
10 - 20%
± (0.3% ABS + 4% REL)
0 - 100%
± (2% ABS + 2% REL)
0 - 8%
± (0.15% ABS + 5% REL)
8 - 12%
± (0.2% ABS + 10% REL)
0 - 10%
± (0.15% ABS + 5% REL)
10 - 15%
± (0.2% ABS + 10% REL)
0 - 22%
± (0.15% ABS + 5% REL)
22 - 25%
± (0.2% ABS + 10% REL)
Air, Air/O2, N2O/O2, Heliox (21% O2),
He/O2, N2, CO2, customized gas types

0 - 300 L/min

General Information
AC input
Battery
(lead rechargeable battery)
Electrical & Physical Data
Power consumption
Weight
Dimensions (w x d x h)
Data Storage
Display

Accuracy

ATP, ATPD, ATPS, AP21, STP, STPH, BTPS, BTPD,
0/1013, 20/981, 15/1013, 25/991, 20/1013
Range
Accuracy
1 - 1000 bpm
± 2.5%** or ±1 bpm
0.05 - 60 s
± 0.02 s
1:300 - 300:1
± 2.5%*
0 - 100%
± 5%*
± 1.75%* or ± 0.10 mL(>2.4 L/min)
± 10 L
± 1.75%* or ± 0.20 mL(>6.0 L/min)
± 10 L

Gas Conditions
Rate
Time
I:E ratio
Ti/Ttotal
Breath volumes

Range

± 2.5%* or ± 20 mL/min(High)
± 10 mL/min(low)

0 – 152.96 cmH2O
± 0.75%* or ± 0.1 cmH2O**
± 300 L/min
± 1.75%* or ± 0.1 L/min**
0 - 1000 mL/mbar
± 3%* or ± 1 mL/mbar**
Adjustable on flow or pressure curves
with user-defined limits.

PFC-3000A

PFC-3000V

PFC-3000L

•
•
•
•
•
•
•
•
Differential

•
•
•
•
•
•
•
•
Relative

•
•
•

•
•
•
•

•
•
•
•
•
•
•
•
Relative
•
•
•

•

•

•

•
•
•
•
•
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050

•
•
•
•
•
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050

•
•
•
•
•
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050
w/ MGA-3050

•

•

•

•

•

•

•
•
•
•
•

•
•
•
•
•

•
•
•
•
•

•

•

•

•

•
•
•

•
•
•

•
•
•

•

•

•

100 - 240 VAC, 50/60 Hz

•

•

•

3 hrs (with MGA-3050 2 hrs)

•

•

•

25 VA(W)
<8.5 Lbs (3.8 kg)
8.67 x 9.84 x 4.72 inches (220 x 250 x 120 mm)
all parameters (measured as well as respiratory values)
Intuitive user interface with numerical measuring values,
statistics, volume trigger configuration,
gas type selection and calibration menus.
USB for Windows Software FlowLab,
Interfaces RS-232 for individual communication,
TTL for external trigger.
annually
15 - 40 °C (59 - 104 °F)
10 - 90% RH, Non-condensing
CE, CSA

•
•
•
•

•
•
•
•

•
•
•
•

•

•

•

•

•

•

•
•
•
•

•
•
•
•

•
•
•
•

Legend
* Tolerance related to the measured value
** Absolute tolerance (Whichever is greater)

Flip for product pricing - Page W

Benchtop Flow Analyzer

Measuring unit

Flow & Pressure Measurements
Measuring direction
Temperature compensated
Pressure compensated
Humidity compensated
O2 compensated
High
Low
High
Average
Low
High Flow Port
Barometer
Vacuum pressure
Flow

TSI Certifier ® FA Plus
Ventilator Test System
®

63

Certifier® FA Plus is the full-feature system capable of testing virtually all models
of ventilators: adult, pediatric, anesthesia,
neonatal and high-frequency

Highlights

Handheld Flow Analyzer

±	Color

touch screen graphical user
	interface
±	

Real-time graphing mode

±	

Bi-directional flow measurement

±	

Data storage using SD Flash card and
internal memory

	

±	Access

	interface

stored

data

through

USB

±	

Report printing capability

±	

Rechargeable battery plus AC operation

Test Parameters
±	Flow
±	

Peak & Minimum Flow

±	

Volume (Inhaled and Exhaled)

±	

Minute Volume

±	

Low Pressure (Differential)

±	

Peak & PEEP Pressure

±	

Mean Airway Pressure

±	

High Pressure

±	

Barometric Pressure

±	

Inspiratory Time

±	

Expiratory Time

±	

I:E Ratio

±	

Respiratory Rate

±	

Flow & Volume modes - STP, ATP, BTPS,
BTPD, plus user-defined

	

±	Oxygen

	

4073 Kit)

Concentration (with optional

1-888-223-6763 - bcgroupintl.com

TSI Certifier® FA Plus
	 Certifier® FA Plus Ventilator Test System
4080 can also test a variety of other medical
equipment such as anesthesia gas delivery
machines, insufflators and oxygen concentrators. Its compact size makes this ventilator
tester ideal for use in field service, biomedical shops,and manufacturing.
	 The fast response and bi-directional sensing sensors makes the Certifier® FA Plus
Ventilator Test Systems 4080 capable of
testing all types of ventilators including
neo-natal and high frequency. This ventilator tester is designed to measure air, oxygen, and nitrous oxide flow and pressure in
institutional, home care, field service,
laboratory, and production applications

Optional LS-2000A
Lung Simulator

Certifier FA Plus

Specifications
Certifier FA PLUS

Range

Accuracy**

Air, O2

-200 to +300 slpm*

±2% or ±0.075 slpm

Air/O2 Mixtures

0 to 300 slpm

±4% or ±0.1 slpm

N2

-200 to +300 slpm

±3% or ±0.1 slpm

CO2

-40 to +40 slpm

±3% or ±0.1 slpm

Air, O2

0.01 to 20 slpm

±2% or ±0.01 slpm

N2, CO2

0.01 to 20 slpm

±3% or ±0.01 slpm

N 2O

0.01 to 20 slpm

±4% or ±0.025 slpm

Air, O2

0.01 to 10 liters STP

±2% Plus 0.02 liters

Air/O2 Mixtures

0.01 to 10 liters STP

±4% Plus 0.02 liters

Air, O2

0.01 to 10 liters STP

±3% Plus 0.03 liters

Air/O2 Mixtures

0.01 to 10 liters STP

±4% Plus 0.04 liters

Air, O2

0.01 to 10 liters STP

±2% or ±0.01 liters

N 2O

0.01 to 10 liters STP

±4% or ±0.01 liters

Minute Volume-High Flow

0.01 to 100 liters STP

±3%

Minute Volume-Low Flow

0 to 10 liters STP

±3%

Flow—High Flow Module

Flow—Low Flow Module

Volume-High Flow-Inhaled

Volume-High Flow-Exhaled

Volume-Low Flow-Inhaled

Respiratory Times

Ti, Tip, Te

0.04 to 30 seconds

±2% or ±0.01 seconds

I:E Ratios

I:E, I:Eip

1:100 to 100:1

±4%

Respiratory Rate

f

1 to 1500 breaths per minute

±2% or 0.1 bpm

Low Pressure

All

-25 to +150 cmH2O

±0.5% or ±0.15 cmH2O

High Pressure

-10 to +150 PSI (-0.7 to 10 bar)

±1% or 0.1 PSI (7 mbar)

Barometric Pressure

7 to 23 PSI (500 to 1600 mbar)

±0.16 PSI (11 mbar)

Oxygen Concentration

21% to 100%

2% of concentration

*slpm = Standard Liters per Minute **Accuracy stated as a percent of reading at TSI standard gas conditions.

To Order Certifier FA Plus
Model
4080

Description
High-Flow Standard Kit

Optional Modules and Accessories
Model
Description
4073
Oxygen Sensor Kit
PSR-11-917-J Replacement Oxygen Sensor
4082
Low-Flow Module
Part #
1208061
1303860
1602342
BC20-40702

Description
Extra battery pack and charger kit
Printer cable
Low Flow Filter
High Flow Filter

4080 High-flow test system
4082 Low-flow kit (sold separately)

Flip for product pricing - Page V

64
64

Handheld Flow Analyzer

Gas/Mode

TSI Certifier® FA
Ventilator Test System

65
65

Certifier® FA is a low-cost test system
capable of testing multiple parameters of
ventilator performance.

®

TSI Certifier FA

Highlights
±	

Simple, easy to read, user interface

±	

Backlit LCD display

±	

Entire kit weighs less than 3 Lbs (1.4 kg)

±	

Volumes and flow in BTPS, ATP, or STP

±	

Powered by 4-AA Alkaline Batteries

Test Parameters
±	Flow
±	

Peak Flow

±	Volume
±	

Stacked Volume

±	

Minute Volume

±	

Low Pressure

±	

Peak & PEEP Pressure

±	

Barometric Pressure

±	

Inspiratory Time

±	

I:E Ratio

±	

Respiratory Rate

±	

Oxygen Concentration (with optional
4073 Kit)

	

Respiratory Ventilator Test System
	 The Certifier® FA Ventilator Test System
4070 is designed to measure air, oxygen,
and nitrous oxide flow and pressure in
institutional, home care, field service,
laboratory, and production applications. The
Certifier® FA ventilator test system is a
battery-operated flow analyzer that can also
test a variety of other medical equipment,
such as anesthesia gas delivery machines
and oxygen concentrators.
	Their compact size makes these ventilator testers ideal for use in field service,
biomedical shops, and manufacturing.

Displays 2 test parameters

1-888-223-6763 - bcgroupintl.com

Certifier FA

Specifications

Certifier FA

66
66

Range

Accuracy**

Air, O2

0 to 300 slpm*

±2% or ±0.075 slpm

Air/O2 Mixtures

0 to 300 slpm

±4% or ±0.1 slpm

Air, O2

0.01 to 15 slpm

±2% or ±0.01 slpm

N 2O

0.01 to 15 slpm

±4% or ±0.025 slpm

Air, O2

0.01 to 10 liters STP

±2% Plus 0.02 liters

Air/O2 Mixtures

0.01 to 10 liters STP

±4% Plus 0.02 liters

Air, O2

0.01 to 9.999 liters STP

±2% or ±0.01 liters

N 2O

0.01 to 10 liters STP

±4% or ±0.01 liters

Minute Volume-High Flow

0.01 to 99 liters STP

±7%

Minute Volume-Low Flow

0 to 9.999 liters STP

±7%

Inspiratory Time

0.25 to 60 seconds

±0.01 seconds

I:E Ratio– High Flow

1:100 to 100:1

±5%

I:E Ratio– Low Flow

1:15 to 15:1

±5%

Respiratory Rate

0.5 to 120 breaths per minute

±5%

Low Pressure

-25 to +150 cmH2O

±0.75% or ±0.2 cmH2O

Barometric Pressure

7 to 29 PSI (500 to 2000 mbar)

±0.16 PSI (11 mbar)

Oxygen Concentration

21% to 100%

2% of concentration

Flow—High Flow Module

Flow—Low Flow Module

Volume-High Flow-Inhaled

Volume-Low Flow-Inhaled

*slpm = Standard Liters per Minute **Accuracy stated as a percent of reading at TSI standard gas conditions.

To Order Certifier FA
Model
4070

Description
High-Flow Standard Kit

Optional Modules and Accessories
Model
Description
4073
Oxygen Sensor Kit
PSR-11-917-J Replacement Oxygen Sensor
BC20-40701
Replacement Low Flow Filter
BC20-40702
Replacement High Flow Filter
4072
Low-Flow Module
Part #
1319288

Description
Hard shell carrying case

DPM-235175NNFC Digital Pressure Meter

Flip for product pricing - Page V

TSI Certifier FA

Gas

Lung Simulators
67
67

Lung Simulators

General Features
±	

Easy to Use

±

Ultra-Portable (< 1 Lbs)

±	

Quick Disassembly

±	

Simple Design

±	

Replacement Parts Available

±	

All Components can be Sterilized

±	

Connects Directly to your Ventilator’s
Existing Tubing

LS-2000 Series Features
±	

Adjustable Compliance Settings

±

Adjustable Resistance Settings

±	

Variable Leakage Valve

±	

Simple Design

±	

Fraction of the Price and Size of any
Comparable Lung Sim

LS-2000A, LS-2000I, LS-1000E

LS-1000E Features
±	

Unique Double-conus Connector
(OD 22 conical, OD 15 conical)

±

Unbeatable Simplicity & Price

LS-1000N Features
±	

Two Compliances

±

Two Separately Adjustable Resistances

±

Leak Simulator on Each Lung

LS-1000N

LS-2000A, LS-2000I, LS-1000E, LS-1000N
	BC Biomedical’s Lung Simulator series
is an affordable, ultra-portable and easy to
maintain alternative to older models. The
LS-2000(A/I) SmartLungs come in Adult
and Infant models,
offering all the performance and features of large and expensive
test lungs in an easy to use, compact package. The LS-1000E EasyLung is a simple,
low-cost, general purpose test lung with no
variable controls. LS-1000N is the ideal tool
to test neonatal and infant ventilators.

1-888-223-6763 - bcgroupintl.com

Specifications

68

Lung Sim

Models

LS-2000I

LS-2000A

General Purpose

Infant

Adult

Resistance

20 mbar/L/s

5, 20, 50, 200 mbar/L/s

5, 20, 50, 200 mbar/L/s

Compliance

25 mL/mbar
(Vt = 500mL,
PEEP = 0 mbar)

1, 2, 3, 5 mL/mbar

10, 15, 20, 30 mL/mbar

0-1000 mL (with 1L bag)

0-200 mL (with 0.5L bag)

0-600 mL (with 1L bag)

N/A

0-10 L/min

0-10 L/min

< 0.5 Lbs (0.23 kg)

< 0.7 Lbs (0.32 kg)

< 0.8 Lbs (0.36 kg)

Dimensions
(LxWxH)

11.9" x 4.6" x 1.7"
(302.3 x 116.8 x 43.7 mm)

10.7" x 4.6" x 1.6"
(271.8 x 116.8 x 40.6 mm)

11.9" x 4.6" x 1.6"
(302.3 x 116.8 x 40.6 mm)

Replaceable
Components

Yes

Yes

Yes

Sterilizable

Yes

Yes

Yes

Use

Volume
Leak
Weight

The BC Biomedical LS-1000E EasyLung is
an affordable, versatile, general purpose test
lung that provides a broad spectrum of benefits to the respiratory care field. It is ideal for
ventilator manufacturers validating the safety
of their products, for ventilator training and
for biomedical engineers
performing
general service procedures. The EasyLung
combines singular design with high-quality,
replaceable parts. Its unique double-conus
multi-connector also ensures a direct connection to all tubing systems.

The SmartLung is unbeatable in terms of
price/performance ratio. Resistance (airway
resistance), lung compliance and leakage
are all adjustable. Its size eliminates the
cumbersome side-tables needed by large
test lungs. Different bag sizes ensure that
virtually all patient lungs can be simulated;
from babies to adults. The SmartLung does
not require any additional adapters and the
leak simulation is infinitely adjustable.The
SmartLung enables verification of premature
baby ventilators as well as mask ventilation.
Even the sensitive function of patient flow
triggering can be tested with the SmartLung.

These test lungs are manufactured in accordance with CE requirements.
Flip for product pricing - Page X

Lung Simulators

LS-1000E

Tachometer Series
69
Applications
Ideal for use in production, 			
	 engineering, inspection, quality 		
	 control and maintenance.

±	

Tachometer Series

Highlights
Contact or non-contact
	measurement
±	

±	

	

Eight character display with
floating decimal point

±	

Sixteen units of measure

±	

Unit measure always displayed

Four memories (last, maximum,
	 minimum and average)
±	

±	

	

Most significant display digit 	
	
flashes to indicate need for battery

Two models available:
±	

TC-1726 without calibration

±	

TC-1726C with NIST calibration 		
certificate

	

TC-1726
	The BC Biomedical Model 1726 handheld Digital Tachometer is a dual function
instrument providing contact and non-contact measurement of rotational and linear
motions, with an accuracy of ±0.025% of
indicated reading ±1 LSD.
	 The Model TC-1726 comes complete with
a convex tip, concave tip, 10 cm linear measuring wheel, 9V Alkaline battery, and carrying
case.
	 The compact size and light weight of the
TC-1726 makes for easy, one-hand operation while the rugged, high impact housing
provides for years of dependable service.

1-888-223-6763 - bcgroupintl.com

TC-1726

70
70

Specifications
Angular Velocity

< 100 Hz: ±(0.025% reading + 1 digit)
100 to 1000 Hz: ±(0.05% reading + 1 digit)
>1000 Hz: ±(0.1% reading + 1 digit)

Linear Measurement

± 0.2% ± 1 LSD
Built-in photoelectric transducer, 12" operating
distance

Transducer

8 character dot matrix LCD, contrast is
adjustable for different viewing angles, 3/8"
character height

Display

Battery

9V Alkaline battery (ANSI/NEDA 1604A or
equivalent)

Low Battery Indicator
Memory

Most significant display character flashes to
indicate need for battery replacement
Last reading, maximum, minimum and average

Dimensions Unit

7.65 x 3.83 x 1.84 Inches
(194.3 x 97.3 x 46.7 mm)

Case

10.53 x 7.18 x 2.39 Inches
(267.5 x 182.4 x 60.7 mm)

Weight

< 2 Lbs. (0.91 kg)

Operating Temperature

0 to 50 °C (32 to 122 °F)

Storage Temperature

-20 to 70 °C (-4 to 158 °F)

Humidity

< 95% R.H. , Non-condensing

Contact measurement
using standard
accessories

Flip for product pricing BB

Standard
Accessories

Tachometer Series

Contact Adapter

Utilizes interchangeable accessories.
Convex tip, concave tip and 10 cm linear
measuring wheel included

BC Structure

BC Group consists of three interlinking operations: BC Biomedical,
BC Marketplace and BC Service. In January 2005, Lloyd Industries
added Engineering, Design and Manufacturing capability when they
purchased BC Group. With these new tools at our disposal, the list of
BC Biomedical products has grown substantially, as is evident in the BC
Biomedical portion of this catalog. In January 2009, BC Group moved
all departments into one new facility, allowing us to better serve our
customers. Additionally, we also have a sales office in Atlanta, GA. BC
Group’s growth has come from listening to customers and then providing what they want. BC Group is now the second largest manufacturer
of Biomedical Test Equipment in the world and growing.
We are truly your ONE-STOP BIOMED SHOP.

BC Group designs and manufactures
a complete line of test equipment
under the BC Biomedical label. The
familiar Green and Gold overlays
tell you that you are getting the
latest technology available in the
biomedical field today.
BC Biomedical products have been
designed with the customer in mind.
Most of our instruments are offered
as families of products. We do not
subscribe to the theory that ones
size fits all. You choose the features
you want and only pay for what you
need. The BC Biomedical product
offering is extensive but we have
many more in the works that will be
introduced after this catalog is published. Please check our website for
the latest products being introduced
or contact us about being added to
our BC Group email newsletter list.

1-888-223-6763 - bcgroupintl.com

BC Group distributes over 75
product lines under BC Marketplace.
This large group of high quality
products, coupled with our own BC
Biomedical Line, truly makes us
your ONE STOP BIOMED SHOP.
Why should you buy items from
BC Group that can be purchased
directly from the OEM? (1) Price
- We buy in volume. (2) We have
stock. (3) It is expensive for an
organization to cut a purchase
order. If you need to order 5 items
from 5 different vendors, you will be
paying much more than if you cut
one PO to BC Group for all items.
(4) You may only buy one or two
items from a vendor in a year where
we purchase hundreds of items
from that same vendor. If you have
any problems with a manufacturer’s
product, you’ll have BC Group as
your advocate.

BC Group offers support for the Biomedical Industry’s test equipment. We calibrate, repair, manage and store test equipment for our
customers. There are three components to BC Service: BC OnSite, BC
Depot and BC eTEAMS. We have a dedicated technical staff that has
been specializing in biomedical test equipment repair and calibration for
many years.

BC Group offers a nation wide,
onsite calibration service program
that provides some unique benefits. Allow us to come to your
facility to minimize downtime and
shipping costs, plus there are no
travel charges. We offer multi-year
discounts on your calibration cost,
plus you will enjoy special discounting
when purchasing test equipment if
you use BC Group as your onsite
provider.

BC eTEAMS (Test Equipment Asset Management System) is a webbased program, custom tailored to customer needs. Secure access to
the on-line database of equipment, complete with location, calibration
certificates and customized management reports make this program an
instant favorite of anyone attempting to manage a fleet of test equipment
in more than one location. This program can also add the tracking of
tools, consumables & other specialized needs as they arise.

1-888-223-6763 - bcgroupintl.com

BC Structure

BC Group offers test equipment
calibration and repair service for all
BC Biomedical and most other manufacturers of test equipment at
our St. Louis Depot facility. We are
ISO 9001:2008 Certified, ISO/
IEC
17025:2005
Accredited,
NIST Traceable and ANSI Z540
compliant. Quick turnaround time,
years of experience calibrating and
attractive prices are just some of
the reasons our customers keep
sending their equipment to BC
Group.

Quality

BC Quality Program

Is at the heart of everything we do at BC Group

BC Group is committed to exceeding customer expectations and enhancing customer
satisfaction through the continual improvement of our products, services and quality
management system.
BC Group has been ISO certified for over a Decade

Our Quality Programs
ISO 9001:2008 Certified
ISO 13485:2003 Certified
ISO/IEC 17025:2005 Accredited
ANSI Z540-1 Accredited
FDA Registered & Inspected
FDA - QSR 21 CFR 820 Compliant
FDA - CGMP Compliant
NIST Traceable

BC Group Manufactures and Sells the familiar Green Labeled BC Biomedical product
line with all products falling under our overall Quality Program

1-888-223-6763 - bcgroupintl.com

Our Production Department Meets the Highest
Standards and Operates Under Our Quality Program

For BC Biomedical products and most other Brands of Biomedical Equipment

BC Group offers Onsite Calibration Services

We bring our Technicians and Expertise to your facility to Minimize Downtime

1-888-223-6763 - bcgroupintl.com

BC Quality Program

BC Group offers Depot Calibration and Repair Services



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : Yes
XMP Toolkit                     : Adobe XMP Core 5.5-c021 79.154911, 2013/10/29-11:47:16
Create Date                     : 2015:03:24 09:55:36-05:00
Metadata Date                   : 2015:03:24 09:57:35-05:00
Modify Date                     : 2015:03:24 09:57:35-05:00
Creator Tool                    : Adobe InDesign CC (Windows)
Instance ID                     : uuid:4b171881-3328-46cf-ba76-045b22bcfb8a
Original Document ID            : xmp.did:EF34D4C9F6C0E4118A4CA5E5B92856FC
Document ID                     : xmp.id:6be2e907-8991-e944-a4ec-155b626e5e29
Rendition Class                 : proof:pdf
Derived From Instance ID        : xmp.iid:919c931d-7798-dc46-97fb-6c5b08d41424
Derived From Document ID        : xmp.did:019076BAF7C0E4118A4CA5E5B92856FC
Derived From Original Document ID: xmp.did:EF34D4C9F6C0E4118A4CA5E5B92856FC
Derived From Rendition Class    : default
History Action                  : converted
History Parameters              : from application/x-indesign to application/pdf
History Software Agent          : Adobe InDesign CC (Windows)
History Changed                 : /
History When                    : 2015:03:24 09:55:36-05:00
Format                          : application/pdf
Producer                        : Adobe PDF Library 11.0
Trapped                         : False
Page Count                      : 77
Creator                         : Adobe InDesign CC (Windows)
EXIF Metadata provided by EXIF.tools

Navigation menu