Technology Guide Blood Gas Analyzers
ALERE INC INSTRUMENTATION LABORATORY NOVA BIOMEDICAL
Product epoc®GEM® Premier™ 4000 critical care analyzer Stat Prole® pHOx®
Web site www.alere.com www.ilus.com www.novabiomedical.com
Calibration/Measurement
Sample volume 92 μL 65 μL micro mode, 100 μL CO-Ox/Total
Bilirubin, 135 μL full menu 45 µL to 125 µL
Calibration time ~170 seconds Automated continuous with iQM 3 to 5 minutes
Measurement time ~35 seconds 70 seconds for electrochemical and 25
additional seconds for CO-Ox 45 to 52 seconds
Analytes
Measured: pH, Pc o 2, Po2, Hct, Na, K, iCa,
Lactate, Glucose; Calculated: Hgb, O2SAT,
BEb, BEecf, tc o 2, Hc o 3
Measured: pH, Pc o
2
, Po
2
, Na
+
, K
+
, Cl
-
, iCa
++
,
Glu, Lact, Hct, tHb, COHb, HHb, MetHb, O2Hb,
Total Bilirubin; Calculated: Hct, tc o
2
, BEecf (in
vivo), BE(B) (in vivo), tHb(c), Ca
++
(7.4), anion
gap, P/F ratio, pa o
2
, Cao
2
, Cvo
2
, P50, O2cap,
sO
2
(c), Hc o
3
-
std, Hc o 3
3
-
(c), AaDO
2
, Pao
2
/Pa o
2
,
RI, CcO
2
, a-vDO
2
, Qsp/Qt(est), Qsp/Qt
pH, Po2, Pc o 2, SO2%, Hct, Hb, Na+, K+, Cl,
Ca, Glu, Lac
Typical cycle time ~30 seconds 90 seconds Throughput of 40 to 50 samples per hour
Operating temperature 15°C to 30°C (59°F to 86°F) 12°C to 32°C (53.6°F to 89.6°F) Ambient operating temperature is 15°C
to 30°C (59°F to 86°F)
Sample type Heparinized or un-anticoagulated arterial,
venous, or capillary whole blood
Whole blood, capillary, mixed venous,
arterial, venous
Whole blood, arterial, mixed venous,
venous, capillary, serum/plasma
Measurement principle
pH, iCa, Pc o 2, Na, K: potentiometry;
Po2, lactate, glucose: amperometry;
Hct: conductometry
Electrochemical, conductivity, CO-Ox Direct ISE, Severinghaus, Optical,
Conductivity, Enzyme/Amperometric
Sample application Injection Syringe, capillary, tube, ampoule Syringe, tube, capillary tube (no adapter
required)
Data storage Data Management System, Database and
Web Server Security 100 MB (unlimited patient samples)
All quality control data is automatically
stored. Daily and cumulative statistical
reports and Levey-Jennings graphs can
be printed at any time.
Security
Password Yes Yes Multilevel password with unique user
ID No.
QC Lockout Yes
User denable QC and new lot lockout options
Yes
Dimensions/
Weight
Height x width x depth
(inches) 3 x 3.4 x 8.5 18 x 12 x 15 15 × 12 × 15
Weight ~1.5 pounds 44 pounds 18 pounds
Miscellaneous
Voltage/frequency
Reader: AC input: 100-240 Vac, .5 amps,
50-60 Hz; DC output: 5 volts, 3 amps;
Host: Input: 100-240 Vac, 50/60 Hz, 500
mA; Output: 5 Vdc, 3000 mA
100 VAC/3 Amps Voltage - /Freq - 50/60
Hz 90-264 VAC, 50/60 Hz
Power consumption (max) DC output: 5 volts, 3 amps 300 VA 200 W
Interface HL7 ASTM1394/HL7
RS232 serial communication and HL7
ASTM Standard data protocols, HL7 on
TCP/IP, USB ports
Approvals FDA 501(k) clearance on epoc System
and all analytes represented
CE, CSA, EMC Emissions and Immunity,
European Union Directive 2002/96/EC on
WEEE, CEI/IEC 61010.1, FDA
ISO 13485; 2003, ISO 9001; 2000
Quality System, Registration, CSA,
TüV, IVDD CE Self Declared, CSA, TüV.
Complies to EN 61010, EN 50081,82
Other Features
Room-temperature card storage (up to 6
months); bar-coded test cards for quality
and inventory management; fully wire-
less data transfer to data manager, real
time (no need to dock for download).
iQM detects, corrects, and documents instrument
errors, reducing error detection time to minutes;
single component, multi-use PAK includes
all testing components, is changed every 30
days, requires no refrigeration or maintenance;
GEMweb Plus Custom Connectivity software
allows access and control from any networked
PC or GEM Premier 4000 analyzer.
Onboard auto-cartridge QC; all-liquid cal-
ibration cartridge eliminates gas tanks;
single reagent cartridge has all supplies
for calibration and waste collection.
RT11_32-33_TechGuideBloodGas.indd 32 10/26/11 11:13:44 AM
November 2011 www.rtmagazine.com
3332
RT | For Decision Makers in Respiratory Care November 2011
Technology Guide Blood Gas Analyzers
NOVA BIOMEDICAL OPTI MEDICAL
SIEMENS HEALTHCARE DIAGNOSTICS INC
Product Stat Prole® pHOx® ULTRA OPTI® CCA-TS analyzer RAPIDPoint® 500 system
Web site www.novabiomedical.com www.optimedical.com www.siemens.com/diagnostics
Calibration/Measurement
Sample volume 60 µL to 210 µL 125 µL for most cassettes, 60 µL cassette
available
Max no. of patient samples per hour/max
no. measured; results per hour 25/up to
325; sample size 100 μL minimum
Calibration time 3 to 5 minutes ~1 minute 1 and 2 point (manual and automatic); 1
point: 30 minutes; 2 point: 2 hours
Measurement time 1 to 2 minutes ~1.5 minutes ~60 seconds to result
Analytes
pH, Po2, Pc o 2, SO2%, Hct, Hb, Na, K, Cl,
Ca, iMg, Glu, BUN, Creat, Lac, CO-Ox
(HHb, O2Hb, COHb, MetHb, tBil)
pH, Pc o 2, Po2, tHb, So2, Na+, K+, iCa, Cl-,
glucose, lactate, and BUN (in various test
congurations)
pH, Pc o 2, Po2, Hb, Na+, K+, Cl-, iCa, glu-
cose, neonatal total bilirubin, COOXimeter
fractions (fO2Hb, fCOHb, fMetHb, fHHb)
Typical cycle time Throughput up to 38 samples per hour ~2.5 minutes ~60 seconds
Operating temperature Ambient operating temperature is 15°C to
30°C (59°F to 86°F) 10°C to 32°C (50°F to 90°F) 15°C to 32°C (59°F to 90°F)
Sample type Whole blood, arterial, mixed venous,
venous, capillary, serum/plasma Whole blood, serum, plasma Whole blood
Measurement principle
Direct ISE, Severinghaus, Amperometric,
Optical, Conductivity, Enzyme/
Amperometric
Optical uorescence and reectance
pH, iCa, Na, Cl, K: potentiometry using
ISE; Pc o 2: potentiometry based on
Severinghaus; Po2: amperometric; glu-
cose: amperometric, glucose oxidase; Hb:
spectrophotometric
Sample application Syringe, tube, capillary tube (no adapter
required)
Syringe or capillary by automatic aspira-
tion
Specimen types suitable for device: whole
blood, capillary, mixed venous, arterial,
venous, heparin/aspiration
Data storage Onboard data management Holds 200 patient records and 1 month of
QC on board. RAPIDComm Data Management system
Security
Password Multilevel password with unique user
ID No.
Password protected setup and operation,
holds up to 300 secure operator IDs/PINs. Yes
QC Lockout Yes
User denable QC and new lot lockout options
Yes
Dimensions/
Weight
Height x width x depth
(inches) 17.2 × 22.4 × 17.3 5 x 14 x 9 21.5 × 11.5 × 16
Weight 53 pounds 12 pounds with rechargeable battery 36.5 pounds
Miscellaneous
Voltage/frequency 90-264 VAC, 50/60 Hz 120 or 240 AC/50-60 Hz n/a
Power consumption (max) 350 W 110 VA n/a
Interface
RS232 serial communication and HL7
ASTM Standard data protocols, HL7 on
TCP/IP, USB ports
Ethernet, serial (ASTM or ASCII)
Ethernet and RS232 via the Siemens
communication protocol LIS3 and serial
RS232 option
Approvals
ISO 13485; 2003, ISO 9001; 2000 Quality
System, Registration, CSA, TüV, IVDD CE
Self Declared, CSA, TüV, Complies to EN
61010, EN 50081,82
CE, UL, IVDD FDA approved
Other features
Large whole blood critical care menu (20
tests), BUN, ionized Mg available exclu-
sively from Nova; onboard co-oximeter;
onboard QC cartridge provides sufcient
QC materials for 30-day auto QC analysis;
allows user to program frequency and
select report protocol.
Single use measurement cassette with
automatic sample aspiration; color touch
screen interface; barcode scanner; heated
measurement for cold environments.
No maintenance, multi-use cartridge;
fast time to patient results and sample
to sample throughput; 28-day, onboard,
automatic quality control cartridge.
ALERE INC INSTRUMENTATION LABORATORY NOVA BIOMEDICAL
Product epoc®GEM® Premier™ 4000 critical care analyzer Stat Prole® pHOx®
Web site www.alere.com www.ilus.com www.novabiomedical.com
Calibration/Measurement
Sample volume 92 μL 65 μL micro mode, 100 μL CO-Ox/Total
Bilirubin, 135 μL full menu 45 µL to 125 µL
Calibration time ~170 seconds Automated continuous with iQM 3 to 5 minutes
Measurement time ~35 seconds 70 seconds for electrochemical and 25
additional seconds for CO-Ox 45 to 52 seconds
Analytes
Measured: pH, Pc o 2, Po2, Hct, Na, K, iCa,
Lactate, Glucose; Calculated: Hgb, O2SAT,
BEb, BEecf, tc o 2, Hc o 3
Measured: pH, Pc o
2
, Po
2
, Na
+
, K
+
, Cl
-
, iCa
++
,
Glu, Lact, Hct, tHb, COHb, HHb, MetHb, O2Hb,
Total Bilirubin; Calculated: Hct, tc o
2
, BEecf (in
vivo), BE(B) (in vivo), tHb(c), Ca
++
(7.4), anion
gap, P/F ratio, pa o
2
, Cao
2
, Cvo
2
, P50, O2cap,
sO
2
(c), Hc o
3
-
std, Hc o 3
3
-
(c), AaDO
2
, Pao
2
/Pa o
2
,
RI, CcO
2
, a-vDO
2
, Qsp/Qt(est), Qsp/Qt
pH, Po2, Pc o 2, SO2%, Hct, Hb, Na+, K+, Cl,
Ca, Glu, Lac
Typical cycle time ~30 seconds 90 seconds Throughput of 40 to 50 samples per hour
Operating temperature 15°C to 30°C (59°F to 86°F) 12°C to 32°C (53.6°F to 89.6°F) Ambient operating temperature is 15°C
to 30°C (59°F to 86°F)
Sample type Heparinized or un-anticoagulated arterial,
venous, or capillary whole blood
Whole blood, capillary, mixed venous,
arterial, venous
Whole blood, arterial, mixed venous,
venous, capillary, serum/plasma
Measurement principle
pH, iCa, Pc o 2, Na, K: potentiometry;
Po2, lactate, glucose: amperometry;
Hct: conductometry
Electrochemical, conductivity, CO-Ox Direct ISE, Severinghaus, Optical,
Conductivity, Enzyme/Amperometric
Sample application Injection Syringe, capillary, tube, ampoule Syringe, tube, capillary tube (no adapter
required)
Data storage Data Management System, Database and
Web Server Security 100 MB (unlimited patient samples)
All quality control data is automatically
stored. Daily and cumulative statistical
reports and Levey-Jennings graphs can
be printed at any time.
Security
Password Yes Yes Multilevel password with unique user
ID No.
QC Lockout Yes
User denable QC and new lot lockout options
Yes
Dimensions/
Weight
Height x width x depth
(inches) 3 x 3.4 x 8.5 18 x 12 x 15 15 × 12 × 15
Weight ~1.5 pounds 44 pounds 18 pounds
Miscellaneous
Voltage/frequency
Reader: AC input: 100-240 Vac, .5 amps,
50-60 Hz; DC output: 5 volts, 3 amps;
Host: Input: 100-240 Vac, 50/60 Hz, 500
mA; Output: 5 Vdc, 3000 mA
100 VAC/3 Amps Voltage - /Freq - 50/60
Hz 90-264 VAC, 50/60 Hz
Power consumption (max) DC output: 5 volts, 3 amps 300 VA 200 W
Interface HL7 ASTM1394/HL7
RS232 serial communication and HL7
ASTM Standard data protocols, HL7 on
TCP/IP, USB ports
Approvals FDA 501(k) clearance on epoc System
and all analytes represented
CE, CSA, EMC Emissions and Immunity,
European Union Directive 2002/96/EC on
WEEE, CEI/IEC 61010.1, FDA
ISO 13485; 2003, ISO 9001; 2000
Quality System, Registration, CSA,
TüV, IVDD CE Self Declared, CSA, TüV.
Complies to EN 61010, EN 50081,82
Other Features
Room-temperature card storage (up to 6
months); bar-coded test cards for quality
and inventory management; fully wire-
less data transfer to data manager, real
time (no need to dock for download).
iQM detects, corrects, and documents instrument
errors, reducing error detection time to minutes;
single component, multi-use PAK includes
all testing components, is changed every 30
days, requires no refrigeration or maintenance;
GEMweb Plus Custom Connectivity software
allows access and control from any networked
PC or GEM Premier 4000 analyzer.
Onboard auto-cartridge QC; all-liquid cal-
ibration cartridge eliminates gas tanks;
single reagent cartridge has all supplies
for calibration and waste collection.