SVC Anatomy & Bridge™ Deployment
Bridge™ Occlusion Balloon has proven to be a life-saving technology.1
Procedural planning that considers superior vena cava (SVC) anatomy in
relation to the pericardium can help ensure successful deployment and
utilization.

BC

BC

SVC Tears Can Be Intra- and Extra-Pericardial
•	 The pericardium extends above the right atrium and covers
between 30-45% of the SVC.

SVC

•	 TEE may identify an intra-pericardial tear, but not recognize
the tear also extends above the pericardium in the SVC.
•	 Tears that extend into both areas can bleed into
pericardium and the right chest.
•	 Bleeding in the pericardium should not rule out a tear that
extends higher in the SVC.
Perica
rdium

Diagram of the pericardium covering about 1/3 of the SVC. The right atrium (RA)
and brachiocephalic vein (BC, also known as the innominate vein) are also shown for
reference.

RA

“When in doubt, inflate Bridge.”
- Dr. Roger Carrillo, MD. University of Miami

SVC lengths of patient population at University of Miami
93% of patients would be completely covered by the Bridge Balloon. The
remaining 7% of patients would be mostly covered by the 8cm Bridge Balloon.
Courtesy of Roger Carrillo, MD.
25

23

Number of Patients

20

16

15

93%

22

16
14

13

119 out of 128 patients had
SVC lengths that would be
completely covered by Bridge.

11

10

5

4

4
0 to 40

40 to 45

45 to 50

50 to 55

55 to 60

60 to 65

65 to 70

70 to 75

75 to 80

80 to 85

SVC Length in mm

Bridge is designed to cover the entire length of the SVC,
including tears in the intra- and extra-pericardial space.

3
85 to 90

2
90 to 95

95 to 100

STOPS ON
AVERAGE 90%
OF BLOOD
LOSS3

DEPLOYS IN
UNDER
2 MINUTES2

30 MINUTES
OF ACCEPTABLE
HEMOSTASIS4

“I cannot overstate that this is a tool that saves lives.1”
- Dr. Roger Carrillo, MD., University of Miami

Important Safety Information
INDICATIONS
The Bridge Occlusion Balloon Catheter is indicated for use for temporary
vessel occlusion of the superior vena cava in applications including
perioperative occlusion and emergency control of hemorrhage. Any use
for procedures other than those indicated in these instructions is not
recommended

exiting the vasculature at the tear site, result in an inability to place the Bridge
Occlusion Balloon Catheter or delay or prevent the ability to achieve occlusion.
Lead extraction should be performed at institutions with cardiothoracic
surgical capabilities by physicians knowledgeable in the techniques
and devices for lead or catheter removal. Complication prevention and
management protocols should be in place and routinely practiced. It is strongly
suggested that the recommendations for lead management of the Heart
Rhythm Society (HRS) and European Heart Rhythm Association (EHRA) be
followed for best results.

CONTRAINDICATIONS
None known.
WARNINGS
Do not position the Bridge Occlusion Balloon Catheter in a manner that
would obstruct the right atrium. Obstruction of the atrium could lead to
arrhythmias and/or hemodynamic compromise.
Prior to initiating the lead extraction procedure, a Bridge Occlusion Balloon
Catheter compatible guidewire should be placed through a venous access
site and across the length of the superior vena cava. Attempting to place a
compatible guidewire after a venous tear occurs may result in an inability to
traverse the superior vena cava with the guidewire, result in the guidewire

Failure to observe recommended inflation techniques may result in formation
of contrast crystals which could prevent deflation. Do not over-inflate the
Bridge Occlusion Balloon Catheter after fully occluding the vessel. Over
inflation may result in damage to the vessel Do not exceed the Maximum
Inflation Volume. Over inflation may result in damage to the vessel or rupture
of the balloon. Occlusion of the superior vena cava beyond 30 minutes is
not recommended as this may increase the risk of adverse physiologic or
neurologic complications.
Refer to the IFU for additional information.

Always Reaching Farther

Corporate Headquarters
The Spectranetics Corporation
9965 Federal Dr., Colorado Springs, CO 80921
Tel: 719-447-2000 • Fax: 719-447-2022
Customer Service: 800-231-0978
German Office
Spectranetics Deutschland GmbH
Schweinfurter Str. 7
97080 Würzburg, Germany
Phone: +49 931/4520080 • Fax: +49 931/45200811
Spectranetics International B.V.
Plesmanstraat 6, 3833 LA Leusden
The Netherlands
Tel: +31 33 4347 050 • Fax: +31 33 4347 051

References
1.	 Elrod, Jodia. Use of Bridge™ Occlusion Balloon in Lead Extraction: Interview with Dr. Roger Carrillo. EP Lab Digest.
November, 2016.
2.	 Document on file D027562. Bridge can be fully deployed in under one minute (53 seconds) in an animal model
when pre-positioned on a guidewire, or in under two minutes (1 minute, 46 seconds) when not pre-positioned.
3.	 Document on file D027561. When deployed, the Bridge occlusion balloon reduces blood loss by up to 90%, on
average, in an animal model of an SVC tear. Testing was conducted in a heparinzed porcine model which has
shorter SVC length than is typical in humans. A balloon design scaled for use specifically in the porcine model was
used in generating this data.
4.	 Document on file, D026197. In an animal model with SVC tears up to 3.5 cm, with 2 pacing leads and 1 ICD lead.
©2016 Spectranetics. All Rights Reserved. Approved for External Distribution. D032863-00 112016



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : Yes
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 5.6-c015 84.158975, 2016/02/13-02:40:29
Create Date                     : 2016:12:05 12:53:31-07:00
Metadata Date                   : 2016:12:05 13:49:54-07:00
Modify Date                     : 2016:12:05 13:49:54-07:00
Creator Tool                    : Adobe InDesign CC 2015 (Macintosh)
Instance ID                     : uuid:7d47d3d0-0ea0-4dad-b71b-7c150402b4a5
Original Document ID            : xmp.did:bb22bfda-5de4-443e-845a-53847d69ad29
Document ID                     : xmp.id:cc566e02-127d-48f4-bca6-6555bd438098
Rendition Class                 : proof:pdf
Derived From Instance ID        : xmp.iid:8ea651c8-a670-4d85-aeee-b2a5f656dd2d
Derived From Document ID        : xmp.did:8ea651c8-a670-4d85-aeee-b2a5f656dd2d
Derived From Original Document ID: xmp.did:bb22bfda-5de4-443e-845a-53847d69ad29
Derived From Rendition Class    : default
History Action                  : converted
History Parameters              : from application/x-indesign to application/pdf
History Software Agent          : Adobe InDesign CC 2015 (Macintosh)
History Changed                 : /
History When                    : 2016:12:05 12:53:31-07:00
Format                          : application/pdf
Producer                        : Adobe PDF Library 15.0
Trapped                         : False
Page Layout                     : SinglePage
Page Count                      : 2
Creator                         : Adobe InDesign CC 2015 (Macintosh)
EXIF Metadata provided by EXIF.tools

Navigation menu