Healix Ti Brochure

Healix Ti Brochure Healix_Ti_Brochure Healix_Ti_Brochure 3 2013 pdf 258413772373414384

2013-02-24

: Pdf Healix Ti Brochure Healix_Ti_Brochure 2 2013 pdf

Open the PDF directly: View PDF PDF.
Page Count: 4

TM
1-800-382-4682 | www.depuy.com
©DePuy Mitek, Inc. 2009. All rights reserved. Printed in the USA.
Dyneema®Purity is a registered trademark of Royal DSM N.V.
P/N 901072, Rev. A 10/09
TM
Dual Thread Suture Anchor with Titanium
Optimized Fixation in Rotator Cuff Repair
Cortical
Threads
Cancellous
Threads
4
0
8
12
16
20
24
12.71 13.88
2.95 3.38
4.5mm Corkscrew FT*4.5mm HEALIX Ti
Insertion Torque [lbf]
4.5mm HEALIX Ti Torque Capability1
Safe Driver/Handle Strip - 17.60
Driver/Anchor Hex Strip Failure - 18.70
Soft Foam
Hard Foam
Soft Foam
Hard Foam
Difference before the force to
insert reaches the force of failure
Driver/Anchor
Hex Strip -12.50
5
0
10
15
20
25
30
21.34 22.00
5.19 5.14
5.5mm Corkscrew FT*5.5mm HEALIX Ti
Insertion Torque [lbf]
5.5mm HEALIX Ti Torque Capability3
Driver/Anchor Hex Strip Failure - 25.00
Driver/Anchor Hex Strip - 22.80
Internally driven design provides increased
torque capabilities and insertion confidence2
DePuy Mitek’s evolutionary HEALIX Suture Anchor
is now offered with titanium for added strength
• Cannulation channels blood
to the surface
Smooth suture bridge provides
improved resistance to
suture abrasion
2
4
6
8
10
12
14
12
4
5.5mm Corkscrew FT*5.5mm HEALIX Ti
Threads per Anchor Comparison1
0
Threads per anchor
Strength Data Attributes
20
0
40
60
80
100
120
110.00 74.20 72.00
5.5mm
Corkscrew FT*
4.5mm
Corkscrew FT*
5.5mm
HEALIX Ti
Suture Abrasion in Number of Suture Cycles to Break
Pullout Force [lbf]
Pullout Comparison1
50
0
100
150
200
250
300
242 126 82
4.5mm HEALIX Ti
82.90
4.5mm
HEALIX Ti
FASTIN®RC5.5mm HEALIX Ti
Tip Separation Force [lbf]
Trocar Tip Strength1
0 1600140012001000800600400200
1346 Cycles
4.5mm Corkscrew FT
FASTIN®RC
1171 Cycles
972 Cycles
4.5mm HEALIX Ti
5.5mm HEALIX Ti
Suture Abrasion Resistance1
Dual thread pattern maximizes pull-out
strength2by independently engaging both
cortical & cancellous bone
• Tactile feedback during transition from
cancellous to cortical threads confirms
anchor engagement and security
HEALIX Ti is engineered to engage denser
bone than Arthrex’s Corkscrew FT1
• Increased number of threads provides
improved bone contact, purchase & bite
• ORTHOCORD®Suture now with the
NEW blue striped 3rd suture
- 55 lbs of tensile strength3
- 45% less stiff than Fiberwire**
• Efficient trocar tip designs allows quick &
easy anchor insertion without the need to
pre-awl, or request additional instruments
Suture Bridge
Optimized Fixation in
Rotator Cuff Repair
Engineered with HEALIX Dual
Thread Technology™ Insertion Torque
169 Cycles
25
50
75
100
125
150
175
128.3
5
5.5mm Corkscrew FT*5.5mm HEALIX Ti
Cadaveric Perfusion (Flow Rate)1
0
Perfusion (cm3/min)
The two simple stitches are then tied over
the mattress stitch. A microfracture is
then performed laterally on the tuberosity
to stimulate blood flow to the surface
(Crimson Duvet).
3
The two blue striped ORTHOCORD®suture
limbs are passed through the RC as a hori-
zontal mattress with the medial limbs of
the blue and purple passed more medially
and between the mattress. Simply, pass 1 is
posterior lateral and is the blue striped.
Next, pass 2 and 3(purple & blue) are made
as simple passes using more medial limbs
closer to the tissue. Pass 4 is blue striped
and that is anterior lateral.
1
Technique II
A wide lateral horizontal mattress is cre-
ated with passes 1 and 4 and the mattress
stitch is then tied first.
2
Mason-Alex7with HEALIX Ti Triple Loaded Anchor
For the simple repair three suture passes
are made per anchor. The medial limb of
each suture closer to the RC is then passed
through the RC using either a shuttle
technique with DePuy Mitek IDEAL™
Suture Graspers or an EXPRESSEW® II
device. The three limbs are equidistantly
spaced apart. Once all sutures are passed
they are sequentially tied using a locking
sliding knot and then follow up with three
alternating half hitches.
1A lateral microfracture is performed with
a small awl (2-3mm diameter holes by
3-4mm in depth) to promote blood flow
to the healing site (Crimson Duvet).
2
Technique I Single-Row Rotator Cuff Repair with HEALIX Ti Triple Loaded Anchor
Purple
Purple
Blue
Blue
Blue Striped
Blue Striped
Passes are typically made from posterior to anterior with
the purple stitch as a laterally based wider mattress (pass
1,5) the blue ORTHOCORD®stitch as a narrower and more
medial mattress (pass 2,4) and the blue striped medial limb
passed at the apex (most medial) of the triangle as pass 3.
1The blue ORTHOCORD sutures are tied first as a medial
mattress (limbs 2 and 4).
2
The blue striped ORTHOCORD is then tied over the top of
the medial mattress as a simple stitch.
3
The second most lateral mattress (purple ORTHOCORD) is lastly
tied over the top of the simple stitch completing the repair. A
microfracture is then performed laterally on the tuberosity to
stimulate blood flow to the surface (Crimson Duvet).
4
Technique III Pyramid Stitch with HEALIX Ti Triple Loaded Anchor
Triple-Loaded HEALIX TiTechniques
Introduction - For rotator cuff repairs, there are cases where double row techniques can be beneficial, however, the majority of
cases can be managed well with single row repairs. Several prospective randomized studies have documented equivalent
outcomes with either single or double row repairs4,5,6. In the following, we present different suture options for single row repair.
Mark H. Getelman, M.D.
Co-Director Sports Medicine Fellowship
Southern California Orthopedic Institute
HEALIX Ti 4.5mm
HEALIX Ti 5.5mm
HEALIX Ti 6.5mm
Purple
Purple
Blue
Blue
Blue Striped
Blue Striped
CUFFLOKSurgical Technique3Pascal Boileau M.D.
Department of Orthopaedic Surgery, Hôpital de L’Archet,
University of Nice, Nice, France
Insert HEALIX Ti™ Dual Threaded
Suture Anchor adjacent to the articular
margin on the medial tuberosity.
1Group like colored suture limbs
together and pass either anteriorly
or posteriorly with EXPRESSEW® II
Flexible Suture Passer.
2Grasp and remove inner suture limbs
(one purple and one blue) together and
tie a secure knot outside the shoulder.
3
Pull opposite limbs to advance knot
into the subacromial space securing
the medial aspect in a mattress fashion.
4
Place limbs into a VERSALOK® Suture
Anchor. Insert VERSALOK Anchor into the
bone and load deployment gun. Once
sutures are tensioned deploy anchor.
5Rotate VERSALOK inserter in
a counterclockwise manner to
remove inserter. Cut sutures
and probe final repair.
6
RO TATOR C UFF R EPAI R S Y S TE M S
Engineered with HEALIX Dual Thread Technology
HEALIX TiAnchor
222253 4.5mm HEALIX Ti Anchor w/ORTHOCORD®
222245 5.5mm HEALIX Ti Anchor w/ORTHOCORD
222249 6.5mm HEALIX Ti Anchor w/ORTHOCORD
222242 5.5mm HEALIX Ti Anchor w/ORTHOCORD and Needles
222248 6.5mm HEALIX Ti Anchor w/ORTHOCORD and Needles
222271 4.5mm HEALIX Ti 3-Suture Anchor w/ORTHOCORD
222243 5.5mm HEALIX Ti 3-Suture Anchor w/ORTHOCORD
222241 6.5mm HEALIX Ti 3-Suture Anchor w/ORTHOCORD
HEALIX BRAnchor
222229 4.5mm HEALIX BR Anchor w/ ORTHOCORD
222233 5.5mm HEALIX BR Anchor w/ ORTHOCORD
222238 6.5mm HEALIX BR Anchor w/ ORTHOCORD
222232 5.5mm HEALIX BR Anchor w/ ORTHOCORD w/ Needles
222239 6.5mm HEALIX BR Anchor w/ ORTHOCORD w/ Needles
HEALIX PEEKAnchor
222205 4.5mm HEALIX PEEK Anchor w/ ORTHOCORD
222209 5.5mm HEALIX PEEK Anchor w/ ORTHOCRD
222221 6.5mm HEALIX PEEK Anchor w/ ORTHOCORD
222210 5.5mm HEALIX PEEK Anchor w/ ORTHOCORD and Needles
222215 6.5mm HEALIX PEEK Anchor w/ ORTHOCORD and Needles
222207 5.5mm HEALIX PEEK 3-Suture w/ ORTHOCORD
222219 6.5mm HEALIX PEEK 3-Suture w/ ORTHOCORD
HEALIX Instruments (For use with HEALIX BR & HEALIX PEEK)
222223 HEALIX Awl
222226 4.5mm HEALIX Cortical Awl/Tap
222251 5.5mm HEALIX Cortical Awl/Tap
222272 4.5mm HEALIX Full Awl/Tap
222224 5.5mm HEALIX Full Awl/Tap
222225 6.5mm HEALIX Full Awl/Tap
PRODUCT# DESCRIPTION
VERSALOK®
210808 VERSALOK Anchor w/ ORTHOCORD
214710 Deployment Gun
214711 2.9mm Awl
EXPRESSEW®II Flexible Suture Passer
214004 EXPRESSEW II Device
214005 EXPRESSEW II Needles 5/box
270120 Grasper-Grabber Suture/Tendon Grabber
NEW Clear Cannula System
214106 5.5mm x 75mm Smooth
214114 7.0mm x 75mm Smooth
214107 5.5mm x 75mm Smooth with Distal Rib
214115 7.0mm x 75mm Smooth with Distal Rib
214104 5.5mm x 55mm Threaded
214118 8.5mm x 55mm Threaded
214108 5.5mm x 75mm Threaded
214116 7.0mm x 75mm Threaded
214120 8.5mm x 75mm Threaded
214110 5.5mm x 90mm Threaded
214122 8.5mm x 90mm Threaded
IDEALSuture Shuttle with CHIA
251001 IDEAL Suture Shuttle, 25º Left
251002 IDEAL Suture Shuttle, 25º Right
251003 IDEAL Suture Shuttle, 45º Left
251004 IDEAL Suture Shuttle, 45º Right
251005 IDEAL Suture Shuttle, 90º Up
251006 IDEAL Suture Shuttle, Straight Hook
251007 IDEAL Suture Shuttle, Straight Crescent
CHIA PERCPASSER™
214101 CHIA PERCPASSER Suture Passer
PRODUCT# DESCRIPTION
*Corkscrew FT is a trademark of Arthrex Inc. Naples FL
**Fiberwire is a registered trademark of Arthrex, Inc. Naples FL
1. Data on File at DePuy Mitek, Inc.
2. Compared with Arthrex Corkscrew FT Anchor.
3. Professor Pascal Boileau M.D. et al, The Mattress-Tension-Band (MTB)
Technique: A Knotless Double-Row Arthroscopic Rotator Cuff Repair, Shoulder
Concepts 2008 Arthroscopy & Arthroplasty, pp 245, 2008 May.
4. Burks RT., Crim J., Brown N., Fink B. Greis PE. A prospective randomized
clinical trial comparing arthroscopic single- and double-row rotator cuff repair:
magnetic resonance imaging and early clinical evaluation. American Journal
of Sports Medicine. 37(4):674-82, 2009 Apr.
Sizes (4.5mm, 5.5mm, 6.5mm)
Arthroscopic and Needle versions
Comes preloaded with 2 or 3 strands of ORTHOCORD
®
Suture
HEALIX Ti product breadth provides versatility
to address the demands from a wide variety of
techniques & tear configurations
Ordering Information
TM
HEALIX Ti 5.5mm
EXPRESSEW II
VERSALOK
5. Grasso A., Milano G., Salvatore M., Falcone G., Deriu L.,
Fabbriciani C., Single-row versus double-row arthroscopic
rotator cuff repair: a prospective randomized clinical study.
Arthroscopy. 25(1):4-12, 2009 Jan.
6. Franceschi F., Ruzzini L., Longo UG., Martina FM., Zobel BB.,
Maffulli N., Denaro V. Equivalent clinical results of arthroscopic
single-row and double-row suture anchor repair for rotator
cuff tears: a randomized controlled trial. American Journal of
Sports Medicine. 35(8):1254-60, 2007 Aug.
7. Castagna A., Garofalo R., Conti M., Borroni M., Snyder SJ.
Arthroscopic rotator cuff repair using a triple-loaded suture
anchor and a modified Mason-Allen technique (Alex stitch).
Arthroscopy. 23(4):440.e1-4, 2007 Apr.

Navigation menu