Heuresis Pb200i User Guide
Heuesis_Pb200i_UserG.. Heuesis_Pb200i_UserGuide
User Manual: Pdf Heuresis Pb200i UserGuide
Open the PDF directly: View PDF
Page Count: 48
- User’s Guide USA
- Revision 10.0.1
- Before operating the Pb200i, be sure to read this entire User’s Guide. We strongly recommend that you store this User’s Guide with the instrument in its carrying case.
- Pb200i Features
- Charging the Batteries
- Installing the Batteries
- Powering On and Initializing the Pb Application Software
- Taking a Measurement
- Figure 13
- Testing the Accuracy and Precision of the Pb200i
- The Pb200i should be tested for Quality Control (QC) before each inspection, every 4 hours, and when the inspection is complete.
- 1. Start by placing the wooden block on top of the Pb200i case. Hold the device with the proximity button against the wooden reference block , with the snout of the instrument centered on the paint film nearest 1.0 mg/cm2. When the LED turns green, sq...
- 2. Compare the result to the value on the reference block and make a note of the reading.
- 3. Complete the test 3 times, then average the readings. The average (rounded to 1 decimal place) of the three readings must fall between 0.8 and 1.2 mg/cm2 (inclusive) for the Pb200i to pass its QC check.
- Figure 14
- Figure 15
- Figure 16
- Figure 17
- Figure 18
- Figure 22
- Figure 23
- Figure 24
- Figure 25
- Figure 26
- Using Wi-Fi Connectivity
- Figure 28
- Figure 29
- Figure 30
- Figure 31
- Figure 32
- Figure 33
- Figure 34
- Figure 35
- Figure 36
- Figure 37
- Enabling Bluetooth™ Connectivity
- Figure 38
- Figure 39
- a) Select the “Settings” app.
- b) Scroll down to “Location Access” and select.
- c) Enable “Access to my location.”
- d) This icon in Figure 45 below will appear, indicating that GPS is enabled.
- To check the status of you GPS, go to the following screens.
- Creating an Administrator
- To view the rights of the administrator, return to the main menu. Select setup, followed by users, then modify and finally settings. (Figure 53 to 56)
- Figure 57 Figure 58
- Changing passwords/user name
- Follow these steps to change your password or user name. From the main menu select setup, then users, then change password or change user name (Figure 59 to 63).
- Caution: User names must be at least four characters long, and passwords must be at least six numbers long.
- Creating a User/Supervisor
- Modifying Permissions
- Turn the “delete data” toggle to the “off” position and back out with the left arrow. (Figure 84 to 86)
- Changing the Action Level in Action Level Mode
- Extended Reading Mode
- Stop at Set Level
- Fixed Time
- Unlimited Time
- Stop at Statistics
- Additional Features
- Data Entry Smart Fields
- Camera
- Enlarging Text on the Pb200i
- Radiation Safety :
- Typical dose rates (5 milliCurie [185 MBq] source)
- Typical dose rates (5 milliCuries [185 MBq] source)
- Operating Conditions & Other Safety
- Radiation Dosimetry
- Radiation dosimetry is worn to monitor radiation levels. It should be worn when required by your regulatory jurisdiction, company safety policy, or RSO. Typically several rounds of dosimetry are used, along with exposure time and use estimates to dete...
- Leak Testing
- Emergency Procedures
- Lost or Stolen Instrument
- Damaged Instrument
- Emergency Numbers
- Regulatory authority_________________________________________________
- Your organization’s RSO _____________________________________________
- Additional company contact(s)____________________________________
- Storage location________________________________________________
- City police____________________________________________________
- State police________________________________________________________
- Fire department________________________________________
- Heuresis emergency contact: Ken Martin 1-617-751-8286 (call or message any time)
- Caution: these contacts should be kept on the operator’s person and with the instrument.
- Warranty