HMO3000 Series Datasheet En

Hmo3000 Datasheet En HMO3000_datasheet_en HMO3000_datasheet_en original 35 attachments listing_attachments images rentogproduction

2016-07-24

: Pdf Hmo3000 Datasheet En HMO3000_datasheet_en original 34 attachments listing_attachments images

Open the PDF directly: View PDF PDF.
Page Count: 16

DownloadHMO3000 Series  Datasheet En
Open PDF In BrowserView PDF
Product Brochure | 02.00

Test & Measurement

R&S®HMO3000 Series
Mixed Signal Oscilloscopes
300/400/500 MHz Bandwidth

Precise signal analysis

Intelligent user interface

FFT

Quick view

4 Gsamples/s sampling rate
8 Msamples memory

To optimize the screen
­display the instrument
shows and hides menus

Superb FFT
functionality

At the push of a button
the 16 most important
values of the measured
signal are permanently
updated and displayed

Product of the Year

2014

HMO3000
Winner in category
Test & Measurement

Setup
Intuitive, multi-lingual
user menu

Help
Context-sensitive help

Math
Wide range of programmable
math functions

Zoom
Memory zoom up to
250,000 : 1

Analog channels
Vertical sensitivity of up to
1 mV/div.

Serial Bus Analysis

Always with MSO functionality

Bus signal source

Fan

Hardware-based triggering and
decoding (optional)

Analyze analog channels plus up to
an additional 16 digital channels

To create SPI, I2C, UART
and counter signals

Maximum noise reduction by
temperature-controlled fan

2 Rohde & Schwarz R&S®HMO3000 Series

At a glance
Systems that are constantly becoming faster and more complex lead to ever
higher demands on the required measurement technology. The oscilloscope series R&S®HMO3000 offers the solution for current requirements in regards to
bandwidth, sampling rate and memory depth. Its bandwidth of up to 500 MHz
allows to set a new milestone in the development of high-performance mixedsignal oscilloscopes at an attractive price.

The 2- and 4-channel instruments provide bandwidths of 300, 400 and 500 MHz,
a sampling rate of 4 GSa/s and a memory depth of 8 MPts. The instruments
are rounded off with a standard inclusion of the MSO functionality and several
options for serial bus analysis to meet all requirements of modern development
designs.
Rohde & Schwarz is offering the new R&S®HMO3000 series exclusively as mixedsignal oscilloscopes. It is also unnecessary to initially activate the mixed-signal functions via software options, as is the case with other suppliers. The low
capacitance logic probe R&S®HO3508 (also available as double pack HO3516) is
optional. It allows the analysis of up to 16 logic channels with a sampling rate of
1 GSa/s. The logic probe is not linked to a specific instrument serial number. This
allows its use with all digital oscilloscope of the R&S®HMO series.
For communications between embedded systems and the environment the
R&S®HMO3000 includes hardware-based signal triggering and decoding for all
common protocols (I2C, SPI, UART, CAN and LIN). This option can be activated
with an upgrade voucher at any time.
The integrated three-digit digital voltmeter enables service technicians to
simultaneously perform voltage measurements on all analog channels with four
values totalling.

The segmented memory option R&S®HOO14 enables you to divide the available
memory of your R&S®HMO3000 into up to 1000 segments. This procedure allows sampling rates of 200 000 Wfm/s, which makes it possible to capture rare
anomalies occurring during many short events in quick succession. For the analysis of the recorded signals, all measurement functions of the R&S®HMO are
available, including the Pass/Fail function.
Thanks to the FFT analysis function with 64 k test points the R&S®HMO3000
series keeps pace with significantly larger oscilloscopes also in the frequency
domain. The time domain signal, measurement window, FFT analysis result are
displayed together on a single screen, which makes it easier to evaluate the input
waveform.
The R&S®HMO3000 series offers time domain, logic, protocol and frequency
analysis in a single instrument and is a member of the Rohde & Schwarz family of
scope-of-the-art oscilloscopes.
Rohde & Schwarz R&S®HMO3000 Series 3

Key facts
Superior hardware-based acquisition for precise measurement results
❙❙ 4 Gsample/s sampling rate, 8 Msample memory depth
❙❙ High vertical sensitivity down to 1 mV/div
❙❙ Low-noise measurement due to state-of-the-art A/D converter
❙❙ High acquisition rate to identify signal faults
❙❙ Segmented memory and manually adjustable memory depth
Versatile measurement functions and fast results
❙❙ Wide selection of automatic measurement functions
❙❙ QuickView: key results at the push of a button
❙❙ Mask test: a new mask can be easily created with just a few keystrokes
❙❙ FFT: the easy way to analyze the signal spectrum
Logic analysis with the MSO option
❙❙ Mixed signal function as standard
❙❙ Precise triggering on signal events
❙❙ Straightforward display of digital signals
❙❙ Low test point loading due to active probe solution
Serial bus analysis: hardware-based triggering and decoding
❙❙ Versatile trigger options for isolating specific data packets
❙❙ Color-coded display of decoded bus signals
❙❙ Direct export of analyzed data to USB memory drive
❙❙ Simultaneous decoding of two buses in realtime

Voltmeter measurements using an oscilloscope
❙❙ Three-digit display for precise voltage measurements
❙❙ Simultaneous measurement on all analog channels of up to four voltage
values totalling
Future-ready investment and scalability
❙❙ Free firmware updates
❙❙ Bandwidth upgrades as required
❙❙ Serial bus analysis and segmented memory via optional software licenses

Application

How the R&S®HMO3000 meets your needs

Engineering lab

❙❙ Adjustable memory depth
❙❙ Advanced math functions available as standard, math on math possible
❙❙ Automeasurement for 28 user-defined parameters
❙❙ Segmented memory (R&S®HOO14, R&S®HV114)
❙❙ Low-noise amplifier and A/D converter
❙❙ 1 mV/div. sensitivity
❙❙ 50 Ω/1 MΩ input impedance, switchable
❙❙ Bandwidth upgrades via software options
❙❙ Simultaneous voltmeter measurements on all analog channels
❙❙ Mixed signal option (MSO) with 16 logic channels
❙❙ Serial bus trigger and hardware-accelerated decode (R&S®HOO10/11/12,
R&S®HV110/111/112)
❙❙ 7-digit hardware counter
❙❙ Superb FFT functionality
❙❙ Remote control for automated data acquisition
❙❙ Pass/fail tests based on user-defined masks with error signal output
❙❙ Automatic signal measurement at the push of a button
❙❙ USB/RS-232, Ethernet/USB or GPIB (IEEE 488) interfaces
❙❙ Fast boot time
❙❙ Low-noise, intelligent temperature management
❙❙ Extended display size through Virtual Screen technology
❙❙ DVI-D output for external display

Analog
circuit design

Embedded
debugging

Model overview
500 MHz

400 MHz

300 MHz

4 channel

R&S®HMO3054

R&S®HMO3044

R&S®HMO3034

2 channel

R&S®HMO3052

R&S®HMO3042

R&S®HMO3032

4 Rohde & Schwarz R&S®HMO3000 Series

Production
environment

General
purpose and
education

300 MHz, 400 MHz, 500 MHz
Should your requirements change, then so does the R&S®HMO3000, as the
300 MHz models can be extended to 400 MHz and 500 MHz bandwidth via
software
upgrades whenever required. This is done with option upgrade vouchers
available at your dealer.

300MHz

400MHz

500MHz

HV35x

❙❙ For 300 MHz to 400 MHz: HV342 (2 channel) and HV344 (4 channel)
❙❙ For 300 MHz to 500 MHz: HV352 (2 channel) and HV354 (4 channel)
❙❙ For 400 MHz to 500 MHz: HV452 (2 channel) and HV454 (4 channel)
Vouchers for bandwidth upgrades or serial bus analysis options are available at
your dealer. The individual voucher number and the serial number of the
instrument to be upgraded is entered at http://voucher.rohde-schwarz.com. The
customer immediately receives the respective licence key which can be loaded
via USB memory drive into the instrument.

Standard

HV34x

HV45x

Rohde & Schwarz R&S®HMO3000 Series 5

Always a MSO
The mixed signal functionality is always included in the R&S®HMO3000 series
with no software option being necessary to unlock it.

6 Rohde & Schwarz R&S®HMO3000 Series

Rohde & Schwarz is offering the R&S®HMO3000 series exclusively as a mixed-signal
oscilloscope. The great advantages of these instruments are best illustrated by taking a look
at how ADCs (Analog Digital Converter) or DACs (Digital Analog Converter) are integrated.

Optional: Logic probe R&S®HO3508

These transformer modules include an analog signal on the one side and a digital signal on
the other side. As shown in the image below the latency time of a DAC can be determined
with one simple cursor measurement. Therefore a MSO allows developers to devote their full
attention to the circuit without having to waste energy on the measurement setup.
The active logic probe R&S®HO3508 (also available as double pack HO3516) is available
separately and is not linked to a specific serial number of an instrument. It can be used with
any R&S®HMO oscilloscope.

❙❙ Logic probe R&S®HO3508 fits to all
R&S®HMO series oscilloscopes
(also available as double pack HO3516)
❙❙ No hardware lock to a specific device
❙❙ 8 logic channels available on each logic probe
❙❙ Signal threshold adjustable for each logic pod
Specifications R&S®HO3508
Channels

8

Memory depth per channel

4 MPts (R&S®HMO3000 series)

Input impedance

100  kΩ || <4  pF

Max. input frequency

350 MHz

Max. input voltage

40  V (DC + AC)

Thresholds
Measuring category

TTL, CMOS, ECL, user-defined
(-2 V to +8 V)
CAT I

Cable length

approx. 1 m

14 bit DAC signal change
Rohde & Schwarz R&S®HMO3000 Series 7

Serial Bus Analysis
I2C, SPI, CAN or LIN – in terms of interaction with the outside world for embedded systems, it is safe to
say that these are the most commonly used communication protocols. The R&S®HMO3000 series by
Rohde & Schwarz offers you hardware-accelerated signal triggering and decoding for all of these
protocols. You can upgrade your instrument via software licence keys with those functions required to
develop your application:
❙❙ R&S®HOO10 / R&S®HV110: Analysis of I2C, SPI and UART/RS-232 signals on analog and logic channels
❙❙ R&S®HOO11 / R&S®HV111: Analysis of I2C, SPI and UART/RS-232 signals on all analog channels
❙❙ R&S®HOO12 / R&S®HV112: Analysis of CAN and LIN signals on analog and logic channels

SPI bus signal, MISO / MOSI decoded

Serial bus trigger types:
❙❙ I2C: Start, Stop, ACK, nACK,
Address/Data
❙❙ SPI: Start, End, Serial Pattern (32Bit)
❙❙ UART/RS-232: Startbit, Frame Start,
Symbol, Pattern

HEX decoded CAN bus signal

❙❙ LIN: Frame Start, Wake Up,
Identifier, Data, Error
❙❙ CAN: Frame Start, Frame End,
Identifier, Data, Error

I2C bus signal in zoom view
8 Rohde & Schwarz R&S®HMO3000 Series

Segmented Memory
The segmented memory option R&S®HOO14 enables you to divide
the available acquisition memory of the R&S®HMO3000 into up to
1000 segments.
This procedure allows sampling rates of 200 000 Wfm/s, which
makes it possible to capture rare anomalies occurring during many
short events in quick succession.
Segmentation can be applied on the acquisition of analog and
digital channels as well as onto the decoding of serial busses.
Additionally, all measurement functions for analyzing the recorded
signals are available, including the Pass/Fail function.
You can upgrade to option R&S®HOO14 at any time with voucher
R&S®HV114. The individual voucher number and the serial number
of the instrument is entered at http://voucher.rohde-schwarz.com.

Segmented Memory (R&S®HOO14, R&S®HV114)
Acquisition memory divided into segments
Maximum segments

1000

Minimum segment size

5 kPts

Maximum segment size

1 MPts

Re-arming time

<3 µs

Maximum acquisition rate

200 000Wfm/s

Segment player

Displays all recorded segments manually or
automatically, all measurement functions including
pass/fail can be used with recorded segments

Sources

all analog and digital channels, busses

Rohde & Schwarz R&S®HMO3000 Series 9

Digital Voltmeter (DVM)
The three-digit digital voltmeter is also a standard feature which makes the work
of service technicians in particular easier. Voltage measurements can be
performed simultaneously for all analog channels. Integrated into a single
compact device it allows you to keep your workplace tidy.
❙❙ Perform measurements simultaneously on all analog channels, with up to
four freely definable parameters totalling
❙❙ These options are available: DC, DCrms, ACrms, Crest Factor, Vpp, Vp+, Vp❙❙ You decide about the position of the values on the screen

DVM on two analog channels with four measurement parameters

HZ15 probe
(sold separately)

10 Rohde & Schwarz R&S®HMO3000 Series

Ramp waveform measured by DVM

Frequency Analysis
Due to the outstanding FFT functionality of the R&S®HMO series
oscilloscopes signals can also be analysed in the frequency domain with
up to 65,536 points. Additional practical tools such as cursor
measurement as well as peak-detect functions are also available. They
allow engineers to complete their analysis significantly faster, also in the
frequency domain.

Sinusoid signal in the time domain

Frequency spectrum exposes the distortion

Easy analysis in frequency domain
In the time domain quite often the distortion of input s­ ignals cannot be
detected with the naked eye. For instance, an acquired sine wave signal
­appears to be undistorted. Only the frequency spectrum – available with
just one push of a button – clearly displays additional harmonics that
occur as harmonic oscillations for multiples of the basic frequency.
Since FFT is also active for previously stored signals, it is possible to
subsequently analyze any sections of those signals captured in single
shot mode or stop mode with an adjustable window width.

Rohde & Schwarz R&S®HMO3000 Series 11

R&S®HMO3002 series 2-channel mixed signal oscilloscope
R&S®HMO3004 series 4-channel mixed signal oscilloscope
R&S®HMO3032, R&S®HMO3034: 300 MHz
R&S®HMO3042, R&S®HMO3044: 400 MHz
R&S®HMO3052, R&S®HMO3054: 500 MHz

Rise time (calculated, 10% to 90%)
< 0.875 ns

Trigger level setting

R&S®HMO305x

< 0.700 ns

with auto level

Linking peak value and trigger level, adjustable
between peak values of a signal

without auto level

±8 Div (from center of screen)

external

±5 V

DC gain accuracy

2 % of full scale

Input sensitivity
1 mV/Div to 5 V/Div (1 MΩ and 50 Ω)

Display

coarse stepping

12 calibrated steps, 1-2-5

variable stepping

freely between calibrated steps

Screen size / type

16.5 cm (6.5 “) VGA Color Display

Resolution

640 (H) x 480 (V) pixel

Impedance

1 MΩ II 13 pF ±2 pF (50 Ω switchable)

Backlight

500 cd/m2 (LED)

Coupling

DC, AC, GND

Max. input voltage

(derates at 20 db/decade to 5 Vrms above
100 kHz)

12 Div (600 pixel)

with menu bar

10 Div (500 pixel)

0.5 Vpp to 10 Vpp

R&S®HMO304x

all analog channels

without menu bar

0.8 Div

external

< 1.166 ns

from firmware version 5.520

Display range in horizontal direction

from 5 mV/Div

R&S®HMO303x

1 MΩ

200 Vp

50 Ω

5 Vrms, max. 30 Vp

Trigger coupling
Auto level

5 Hz to 300/400/500 MHz

AC

5 Hz to 300/400/500 MHz

DC

DC to 300/400/500 MHz

HF

30 kHz to 300/400/500 MHz

selectable filters
LF

DC to 5 kHz, selectable in DC and auto level
mode

low-pass
(noise rejection)

200 MHz, selectable in AC, DC, HF and auto
level mode

Display range in vertical
direction

8 Div (400 pixel)

Position range

with VirtualScreen usage

20 Div

Offset control

Color depth

256 colors

1 mV/Div, 2 mV/Div

±0.2 V - 8 Div x sensitivity

External Trigger Input (BNC)

Trace display

pseudo-color, inverse intensity

5 mV/Div to 20 mV/Div

±1.0 V - 8 Div x sensitivity

Impedance

1 MΩ || 14 pF ±2 pF

Levels of trace brightness

32

50 mV/Div

±2.5 V - 8 Div x sensitivity

Sensitivity

0.5 Vpp to 10 Vpp

Button brightness

light, dark

100 mV/Div, 200 mV/Div

±20 V - 8 Div x sensitivity

Trigger level

±5 V

500 mV/Div to 5 V/Div

±50 V - 8 Div x sensitivity

Max. input voltage

100 Vp

XY/XYZ mode

selectively all analog channels

Coupling

DC, AC

selectively all analog channels

Trigger/Auxiliary Output (BNC)

Vertical System
DSO mode

±8 Div (from center of screen)

Trigger hold-off

2-channel models

CH1, CH2

Inversion

4-channel models

CH1, CH2, CH3, CH4

Logic Channels (with logic probe R&S®HO3508/HO3516)

MSO mode

Thresholds

TTL, CMOS, ECL, user-defined (-2 V to +8 V)
100 kΩ || 4 pF

2-channel models

CH1, CH2, POD1, POD2

Impedance

4-channel models

CH1, CH2, CH3|POD1, CH4|POD2

Coupling

Analog Channels

Max. input voltage

Y-bandwidth (-3dB)

Trigger System

(1mV, 2mV)/Div

R&S®R&S®HMO303x: 180 MHz
R&S®HMO304x, HMO305x: 200 MHz

(5mV bis 5V)/Div

R&S®HMO303x: 300 MHz
R&S®HMO304x: 400 MHz
R&S®HMO305x: 500 MHz

Lower AC bandwidth
Bandwidth limitation
(switchable)

2 Hz
about 20 MHz

40 Vp

Auto

Functions

Pulse output for every acquisition trigger event,
error output on mask violation

Output level

3.8 V

Pulse polarity

positive

Pulse width

>150 ns (trigger event), >0.5 µs (mask violation)

Trigger Types
Edge

Trigger Mode
Triggers automatically also without any specific
trigger event

Direction

increasing, decreasing, both

Trigger coupling

auto level AC, DC, HF

Normal

Triggers only on specific trigger events

Switchable filters

LF, noise rejection

Single

Triggers once on a trigger event

Sources

Trigger indicator

Screen and panel (LED)

all analog and digital channels, mains, external
(AC, DC)

Trigger sensitivity
up to 2 mV/Div
2 mV/Div to 5 mV/Div

12 Rohde & Schwarz R&S®HMO3000 Series

DC

50 ns to 17 s

1.5 Div
1.0 Div

Edge A/B
Direction

increasing, decreasing, both

Source A, B

all analog channels, external (AC, DC)

Frequency range

DC to 300/400/500 MHz

Min. signal amplitude

0.8 Div

Trigger level range (seperately ±8 Div (from center of screen)
adjustable with different
sources)
external

±5.0 V

Trigger coupling
State A

Runt
positive, negative, both

2-channel models

2 x 4 MPts or 1 x 8 MPts

Sources

all analog channels

4-channel models

4 x 4 MPts or 2 x 8 MPts

Serial Busses (optional)

Resolution

8 bit, (HiRes up to 16 bit)

Bus representation

Waveform arithmetics

refresh, roll (loose/triggered), average (up to 1024),
envelope, peak detect (500 ps), filter (low-pass,
adjustable), high resolution (up to 16 bit)

Record modes

automatic, max. sampling rate, max. waveform
update rate, specific record length (10 kPts to
2 MPts)

auto level, AC, DC, HF, LF, low-pass

State B
same sources

as state A

different sources

DC, HF, low-pass

Trigger setting
time based

16 ns to 8.589 s, resolution min. 4 ns

event based

1 to 216 events

Pulse Width

Memory Depth

Polarity

Up to two busses can be analyzed at the same
time. Color-coded display of decoded data in
ASCII, binary, decimal and hexadecimal format.

Option / Voucher codes
R&S®HOO10 / R&S®HV110

Analysis of I2C, SPI, UART/RS-232 signals on
analog and logic channels

R&S®HOO11 / R&S®HV111

Analysis of I2C, SPI, UART/RS-232 signals on
all analog channels

R&S®HOO12 / R&S®HV112

Analysis of CAN and LIN signals on analog and
logic channels

Trigger types by protocols

Polarity

positive, negative

I2C

Start, Stop, ACK, NACK, Address/Data

Functions

equal, not equal, lower, higher, within/without
a range

SPI

Start, End, Serial Pattern (32 Bit)

Pulse duration

4 ns to 8.5 s, resolution min. 0.5 ns

UART/RS-232

Startbit, Frame Start, Symbol, Pattern

Sources

all analog and digital channels

Logic

LIN
CAN

Frame Start, Wake Up, Identifier, Data, Error
Frame Start, Frame End, Identifier, Data, Error

boolean operators

AND, OR, TRUE, FALSE

time based operators

equal, not equal, lower, higher, within/without
a time range, timeout

Duration

4 ns to 8.5 s, resolution min. 0.5 ns

States

H, L, X

Sources

all logic channels

Video

Time domain (Yt)

main screen, time domain and zoom window

Frequency domain (FFT)

time domain and frequency domain window
(FFT)

Risetime
Functions

rise/fall time, both

Time range

4 ns to 8.5 s, resolution min. 0.5 ns

Time based operators

equal, not equal, lower, higher

Variance

±2 ns to ±33.5 ms, resolution 2 ns

Sources

all analog channels

0 to 8,59 x 109 Sa x (1/sample rate)

Waveform update rate

up to 5000 Wfm/s

Waveform display

dots, vectors, persistence afterglow

Persistence afterglow

min. 50 ms

Segment size

5 kPts to 1 MPts

max. number of segments

up to 1,000

Re-arming time

less than 3 µs
200.000 Wfm/s

Segment player

Reference signals

up to 4 references

Channel deskew

-62.5 ns to +61.5 ns, step size 500 ps
up to 250,000:1

Displays all recorded segments manually or
automatically. All measurement functions
including pass/fail testing can be applied on
the recorded segments.
all analog and digital channels, busses

Time basis

all analog channels, external (AC, DC)

post-trigger

virtual display of 20 Div for all math, logic, bus,
reference signals

NTSC, SECAM, PAL, PAL-M, SDTV 576i, HDTV
720p, HDTV 1080i, HDTV 1080p

Sources

0 to 4x106 Sa x (1/sample rate), x2 in interlaced
mode

VirtualScreen

Supported standards

line number selectable, all

pre-trigger

Sampling rate

Memory zoom

even/odd, either

pulse

Delay

voltage (XY), intensity (Z)

positive, negative

Line

sin(x)/x, linear, sample-hold

logic channels

XY/XYZ mode

Sync. pulse polarity

Field

all analog channels

Segmented Memory (optional, R&S®HOO14 / R&S®HV114)

Horizontal System

Functions

Interpolation

Sources

Waveform Measurements and Operation
Operation

accuracy

±15.0 x 10-6

aging

±5.0 x 10-6 per year

Automatic measurements

Operation modes
REFRESH

1 ns/Div to 50 s/Div

ROLL

50 ms/Div to 50 s/Div

Acquisition System
Automatic search functions

Realtime Sampling Rate
2-channel models

2 x 2 GSa/s or 1 x 4 GSa/s

4-channel models

4 x 2 GSa/s or 2 x 4 GSa/s

logic channels

16 x 1 GSa/s

menu-driven (multilingual), auto-set, help
functions (multilingual)
voltage (Vpp, Vp+, Vp-, Vrms, Vavg, Vmin, Vmax),
amplitude, phase, frequency, period, rise/fall
time (80%, 90%), overshoot (pos/neg), pulse
width (pos/neg), burst width, duty cycle (pos/
neg), standard deviation, delay, crest factor,
edge/pulse count (pos/neg), trigger period,
trigger frequency
edge, pulse, peak, rise/fall time, runt

Rohde & Schwarz R&S®HMO3000 Series 13

Cursor measurements

Quick measurements
(QUICKVIEW)

Marker

voltage (V1, V2, ∆V), time (t1, t2, ∆t, 1/∆t), ratio
X, ratio Y, pulse and edge count (pos/neg),
peak values (Vpp, Vp+, Vp-), mean/RMS/standard
deviation, duty cycle (pos/neg), burst width,
rise/fall time (80%, 90%), ratio marker, crest
factor
voltage (Vpp, Vp+, Vp-, Vrms), frequency, period
(predefined), 6 additional measure-ment
functions (see automatic measurement
functions) freely selectable plus statistics
up to 8 freely positionable markers for easy
navigation, automatic marker positioning
based on search specification

Frequency Analysis (FFT)
Parameters
FFT length

Math. equation sets
frequency span, center frequency, vertical
scale, vertical position

Realtime clock (RTC)

2 kpts, 4 kpts, 8 kpts, 16 kpts, 32 kpts, to 64 kpts

Power supply

Window

Hanning, Hamming, Rectangular, Blackman

Scale

dBm, dBV, Vrms

Waveform arithmetics

refresh, envelope, average (up to 512)

Cursor measurement
Sources

2 horizontal cursors, previous/next peak search

Pattern Generator

Resolution

7 digit

Functions

Frequency range

0.5 Hz to 300/400/500 MHz

probe adjust, bus signal source, counter,
random pattern

Accuracy

±15.0 x 10

Probe ADJ output

1 kHz, 1 MHz square wave: 1.0 Vpp (tr <4 ns)

Aging

±5.0 x 10-6 per year

Bus signal source (4 bit)

I2C (100 kBit/s, 400 kBit/s, 1 MBit/s), SPI
(100 kBit/s, 250 kBit/s, 1 MBit/s), UART
(9600 Bit/s, 115,2 kBit/s, 1 MBit/s)

Mask Testing
Functions

Pass/Fail comparison with an user-definied
mask performed on waveforms

Counter (4 bit)

frequency: 1 kHz, 1 MHz
direction: incrementing

Sources

all analog channels

Random pattern (4 bit)

frequency: 1 kHz, 1 MHz

Mask definition

mask enclosing acquired waveform with userdefined tolerance

Interfaces

Actions
on mask violations
during acquisiton

for mass storage

2 x USB-Host, Typ A (FAT16/32)

for remote control

HO730 dual interface:
Ethernet (RJ-45) / USB-Device (Typ B)

beep, acquisition stop, screenshot, trigger
pulse, automatically saving trace data

Optional interfaces

statistics: number of completed tests, number
of passes / failed acquisition (absolute and in
percent), test duration

HO720 dual interface:
USB-Device (Typ B) / RS-232
HO740 interface: IEEE-488 (GPIB)

External monitor interface

DVI-D (480 p, 60 Hz), HDMI compatible

Waveform Maths

Application memory

Functions

addition, substraction, multiplication, division

Sources

2 analog channels

Mathematics
Functions

addition, substraction, multiplication, division,
minimum / maximum, square, square root,
absolute value, pos/neg wave, reciprocal,
inverse, log10/ln, derivation, integration, filter
(lowpass/highpass)

Editing

formula editor, menu-driven

Sources

all analog channels, user-defined constants

Storage location

math. memory

Number of formula sets

5 formula sets

Number of equations

5 equations per formula set

Simultaneous display of math. 1 formula set with max. 4 equations
functions

14 Rohde & Schwarz R&S®HMO3000 Series

100 V to 240 V, 50 Hz to 60 Hz, CAT-II

2-channel models

max. 70 W

4-channel models

max. 90 W

Safety

in line with IEC 61010-1 (ed. 3), IEC 61010-230 (ed. 1), EN 61010-1, EN 61010-2-030 , CAN/
CSA-C22.2 No. 61010-1-12 , CAN/CSA-C22.2
No. 61010-2-030-12 ,UL Std. No. 61010-1 (3rd
Edition) , UL61010-2-030

Temperature
operating temperature range
storage temperature range
Rel. humidity

+5 °C to +40 °C
-20 °C to +70 °C
5 % to 80 % (without condensation)

Mechanical data
dimensions (W x H x D)

285 x 175 x 220 mm

weight

3.6 kg

All specifications at 23°C after 30 minute warm-up.
Accessories included:
R&S®HO732 Ethernet/USB dual interface card, line cord, printed operating
­manual, software-CD, 2/4 probes: R&S®RT-ZP05 (up to 500 MHz, 10:1 with
attenuation ID)

Bandwidth Upgrade Vouchers

General Data

Quickmath

date and time

Power consumption

all analog channels

Frequency Counter (hardware based)

-6

AC supply

on internal file system or external USB memory

8 MB for references, formulas, device settings,
languages and help functions

Description

Voucher-Codes

Bandwidth upgrade 300 MHz to 400 MHz

HV342 (2-channel models)
HV344 (4-channel models)

Device settings

on internal file system or external USB
memory, available file formats: SCP, HDS

Bandwidth upgrade 300 MHz to 500 MHz

HV352 (2-channel models)
HV354 (4-channel models)

Reference waveforms

on internal file system or external USB
memory, available file formats: BIN (MSB/LSB),
FLT (MSB/LSB), CSV, TXT, HRT

Bandwidth upgrade 400 MHz to 500 MHz

HV452 (2-channel models)
HV454 (4-channel models)

Traces

on external USB memory, available file formats:
BIN (MSB/LSB), FLT (MSB/LSB), CSV, TXT,
HRT

Save/Recall

data

display or acquisition data

sources

single or all analog channels

Screenshots

on external USB memory, available file formats:
BMP, GIF, PNG
(color, inverted, grey-scale)

Bus Analysis and Segmented Memory
Description

Option-Code

Voucher-Code

I2C, SPI, UART/RS-232
on analog and digital channels

R&S®HOO10

R&S®HV110

I2C, SPI, UART/RS-232
on all analog channels

R&S®HOO11

R&S®HV111

CAN und LIN
on analog and digital channels

R&S®HOO12

R&S®HV112

Segmented memory

R&S®HOO14

R&S®HV114

Recommended Accessories
R&S®HO720

R&S®HO740

HZO20

HZO30

HZ115

USB-device/RS-232 ­dualinterface card

IEEE-488 (GPIB) interface
card, galvanically isolated

High voltage probe 1000:1
(400 MHz, 1000 Vrms)

1 GHz active probe
(0.9 pF, 1 MΩ)

Differential Probe
100:1/1000:1

R&S®HO3508

HZO40

HZO41

8 channel logic probe (350 MHz, 4 pF)

Active differential probe
200 MHz (10:1, 3.5 pF, 1 MΩ)

Active differential probe
800 MHz (10:1, 1 pF, 200 kΩ)

HZO50

HZO51

HZ99

HZ46

AC/DC current probe
30 A, DC to 100 kHz

AC/DC current probe
100/1000 A, DC to 20 kHz

Carrying case for
protection and transport

19” rackmount kit, 4 RU

Rohde & Schwarz R&S®HMO3000 Series 15

© 2015 Rohde & Schwarz GmbH & Co. KG
Mühldorfstr. 15, 81671 München, Germany
Phone: +49 89 41 29 - 0
Fax: +49 89 41 29 12 164
E-mail: info@rohde-schwarz.com
Internet: www.rohde-schwarz.com
Subject to change – Data without tolerance limits is not binding.
R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
Trade names are trademarks of the owners.
01/16

PD 3607.0146.32 V 02.00 PDP1 en

3607014632



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.7
Linearized                      : No
XMP Toolkit                     : Adobe XMP Core 5.6-c015 81.157285, 2014/12/12-00:43:15
Create Date                     : 2016:02:29 11:15:04+01:00
Metadata Date                   : 2016:02:29 12:27:34+01:00
Modify Date                     : 2016:02:29 12:27:34+01:00
Creator Tool                    : Adobe InDesign CC 2015 (Windows)
Instance ID                     : uuid:78c55d4a-032e-4493-9286-08f51a0f0530
Original Document ID            : xmp.did:17D59702A02968118A6D840947D357D3
Document ID                     : xmp.id:f7a888a8-2c72-5143-88bf-45a2b5ddc2ff
Rendition Class                 : proof:pdf
Derived From Instance ID        : xmp.iid:836f4a57-6d4e-bc4f-afea-07e6d48e2ff8
Derived From Document ID        : xmp.did:FA3F3E7407206811822AFA79D327CB3D
Derived From Original Document ID: xmp.did:17D59702A02968118A6D840947D357D3
Derived From Rendition Class    : default
History Action                  : converted
History Parameters              : from application/x-indesign to application/pdf
History Software Agent          : Adobe InDesign CC 2015 (Windows)
History Changed                 : /
History When                    : 2016:02:29 11:15:04+01:00
Format                          : application/pdf
Title                           : HMO3000 Series
Description                     : Oscilloscope
Creator                         : Rohde & Schwarz
Subject                         : Mixed Signal Oscilloscopes 300/400/500 MHz Bandwidth
Producer                        : Adobe PDF Library 15.0
Trapped                         : False
Keywords                        : Mixed Signal Oscilloscopes 300/400/500 MHz Bandwidth
Page Count                      : 16
Author                          : Rohde & Schwarz
Warning                         : [Minor] Ignored duplicate Info dictionary
EXIF Metadata provided by EXIF.tools