WTW PH 3110 Operating Manual WTW_p H_3110 P H

User Manual: Pdf WTW_pH_3110

Open the PDF directly: View PDF PDF.
Page Count: 50

DownloadWTW PH 3110 Operating Manual WTW_p H_3110 P H
Open PDF In BrowserView PDF
Operating manual

pH 3110

69

pH

92

24

8

°C

TP

AR

pH meter
For Sales & Service Contact

2650 E. 40th Ave. • Denver, CO 80205
Phone 303-320-4764 • Fax 303-322-7242

1-800-833-7958
ba75791e05

05/2011

www.geotechenv.com

pH 3110

Accuracy when
going to press

Copyright

The use of advanced technology and the high quality standard of our
instruments are the result of a continuous development. This may result in
differences between this operating manual and your meter. Also, we cannot
guarantee that there are absolutely no errors in this manual. Therefore, we
are sure you will understand that we cannot accept any legal claims resulting
from the data, figures or descriptions.

© Weilheim 2008, WTW GmbH
Reproduction in whole - or even in part - is prohibited without the express
written permission of WTW GmbH, Weilheim.
Printed in Germany.

ba75791e05

05/2011

pH 3110

Contents

pH 3110 - Contents

1

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1
1.2
1.3

2

Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1
2.2

3

4.2

4.3

4.4
4.5

4.6

5.2
5.3

05/2011

13
13
13
14

General operating principles . . . . . . . . . . . . . . . . . . . . . .
4.1.1 Operating modes . . . . . . . . . . . . . . . . . . . . . . . .
4.1.2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.2.1 Measuring the pH value . . . . . . . . . . . . . . . . . . .
4.2.2 Measuring the ORP . . . . . . . . . . . . . . . . . . . . . .
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.3.1 Calibration interval (Int.C) . . . . . . . . . . . . . . . . .
4.3.2 Automatic calibration (AutoCal) . . . . . . . . . . . . .
4.3.3 Conventional calibration (ConCal) . . . . . . . . . . .
Downloading calibration data . . . . . . . . . . . . . . . . . . . . .
Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.5.1 System settings . . . . . . . . . . . . . . . . . . . . . . . . .
4.5.2 Measurement settings . . . . . . . . . . . . . . . . . . . .
Reset. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.6.1 Resetting calibration values . . . . . . . . . . . . . . . .
4.6.2 Resetting all meter settings . . . . . . . . . . . . . . . .

15
15
15
16
17
18
19
21
22
24
27
28
29
30
34
34
35

Maintenance, cleaning, disposal . . . . . . . . . . . . . . . . . 37
5.1

ba75791e05

Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Initial commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2.1 Inserting the batteries. . . . . . . . . . . . . . . . . . . . .
3.2.2 Switching on the meter. . . . . . . . . . . . . . . . . . . .

Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1

5

Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
General safety instructions . . . . . . . . . . . . . . . . . . . . . . . 10

Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1
3.2

4

Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Socket field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.1.1 Replacing the batteries . . . . . . . . . . . . . . . . . . .
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Packing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

37
37
38
38

3

Contents

pH 3110

5.4

6

What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

7

Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.1
7.2

8

4

Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Measuring ranges, resolution, accuracy . . . . . . . . . . . . . 44

Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

ba75791e05

05/2011

pH 3110

Overview

1

Overview

The compact pH 3110 precision pH meter enables you to perform pH
measurements rapidly and reliably. The pH 3110 provides the
maximum degree of operating comfort, reliability and measuring
certainty for all applications.
The proven calibration procedures and automatic stability control
function (AR) support your work with the pH meter.

3

69

pH

92

24

8

°C

2
TP

AR

1

ba75791e05

05/2011

1

Keypad

2

Display

3

Socket field

5

Overview

pH 3110

1.1

Keypad

pH 3110

In this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. ) generally indicates a short keystroke
(under 2 sec) in this operating manual. A long keystroke (approx.
2 sec) is indicated by the underscore behind the key symbol (e.g.
).

_

:
:

Switches the meter on/off
Resets calibration data

:
:

Selects the measured parameter
Opens the setting menu for
calibration and measurements

:
:

Calls up the calibration procedure
Displays the calibration data

<▲>:

Increments values, scrolls

<▼>:

Decrements values, scrolls

_

_

_

:

:

_

6

Confirms entries
Opens the setting menu for system
settings

ba75791e05

05/2011

pH 3110

Overview

1.2

Display
Status display
Sensor symbol
(calibration
evaluation)
Measured value
display
Function and
temperature display
indicators

Function display
indicators

ba75791e05

05/2011

AutoCal TEC

Calibration with automatic buffer recognition
(buffer set: Technical buffers )

AutoCal DIN

Calibration with automatic buffer recognition
(buffer set: DIN buffers )

ConCal

Calibration with any buffers

CalError

An error occurred during calibration

LoBat

With battery operation: Batteries almost empty

AR

Stability control (AutoRead) is active

TP

Temperature measurement active

TIME

Setting of calibration interval

7

Overview

pH 3110

1.3

Socket field

1

4

2

3

Connectors:
1

pH electrode

2

Reference electrode

3

Temperature sensor

4

Service interface

Caution
Only connect sensors to the meter that cannot return any voltages
or currents that are not allowed (> SELV and > current circuit with
current limiting).
Almost all customary sensors fulfill these conditions.

8

ba75791e05

05/2011

pH 3110

Safety

2

Safety

This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the meter.
Consequently, all responsible personnel must read this operating
manual before working with the meter.
The operating manual must always be available within the
vicinity of the meter.
Target group

Safety instructions

The meter was developed for work in the field and in the laboratory.
Thus, we assume that, as a result of their professional training and
experience, the operators will know the necessary safety precautions
to take when handling chemicals.
Safety instructions in this operating manual are indicated by the
warning symbol (triangle) in the left column. The signal word (e.g.
"Caution") indicates the level of danger:

Warning
indicates instructions that must be followed precisely in order to
avoid possibly great dangers to personnel.

Caution
indicates instructions that must be followed precisely in order to
avoid the possibility of slight injuries or damage to the meter or
the environment.
Further notes
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating
manuals.

ba75791e05

05/2011

9

Safety

pH 3110

2.1

Authorized use

This meter is authorized exclusively for pH and ORP measurements in
a laboratory or field environment.
The technical specifications as given in chapter 7 TECHNICAL DATA must
be observed. Only the operation and running of the meter according to
the instructions given in this operating manual is authorized.
Any other use is considered unauthorized.

2.2

General safety instructions

This meter is constructed and tested in compliance with the IEC 1010
safety regulations for electronic measuring instruments.
It left the factory in a safe and secure technical condition.
Function and
operational safety

The smooth functioning and operational safety of the meter can only be
guaranteed if the generally applicable safety measures and the specific
safety instructions in this operating manual are followed during
operation.
The smooth functioning and operational safety of the meter can only be
guaranteed under the environmental conditions that are specified in
chapter 7 TECHNICAL DATA.
If the meter was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty
functioning of the meter. In this event, wait until the temperature of the
meter reaches room temperature before putting the meter back into
operation.

Caution
The meter is only allowed to be opened by authorized personnel.

10

ba75791e05

05/2011

pH 3110

Safety

Safe operation

If safe operation is no longer possible, the meter must be taken out of
service and secured against inadvertent operation!
Safe operation is no longer possible if the meter:
 has been damaged in transport
 has been stored under adverse conditions for a lengthy period of
time
 is visibly damaged
 no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the meter.

Obligations of the
purchaser

The purchaser of this meter must ensure that the following laws and
guidelines are observed when using dangerous substances:
 EEC directives for protective labor legislation
 National protective labor legislation
 Safety regulations
 Safety datasheets of the chemical manufacturers.

Caution
In addition to the safety instructions mentioned here, also follow
the safety instructions of the sensors used.
The operating manuals of the sensors are available on the
supplied CD and on the Internet under www.WTW.com.

ba75791e05

05/2011

11

Safety

12

pH 3110

ba75791e05

05/2011

pH 3110

Commissioning

3

Commissioning

3.1

Scope of delivery

 Handheld meter, pH 3110
 4 batteries 1.5 V Mignon type AA
 Short instructions
 CD-ROM with detailed operating manual

3.2

Initial commissioning

Perform the following activities:
 Insert the supplied batteries
 Switch on the meter

3.2.1

Inserting the batteries

1

Unscrew the two screws (1) on the underside of the meter.

2

Open the battery compartment (2) on the underside of the
meter.

1

2

3

ba75791e05

05/2011

Place four batteries (type Mignon AA) in the battery
compartment.

13

Commissioning

pH 3110

Note
Alternatively, you can also use Ni-MH rechargeable batteries (type
Mignon AA). In order to charge the batteries, an external charging
device is required.

Caution
Make sure that the poles of the batteries are positioned correctly.
The ± signs on the batteries must correspond to the ± signs in the
battery compartment.

4

3.2.2

1

Close the battery compartment (2) and tighten the screws (1).

Switching on the meter

Press the  key.
A display test is briefly displayed.
Subsequently, the meter switches to the measuring mode
(measured value display).

Note
The meter has an energy saving feature to avoid unnecessary battery
depletion during battery operation.
The energy saving feature switches off the meter if no key was pressed
during the specified interval (setting the switch-off interval see section
4.5.1).

14

ba75791e05

05/2011

pH 3110

Operation

4

Operation

4.1

General operating principles

This section contains basic information on the operation of the
pH 3110.

4.1.1

Operating modes

The meter has the following operating modes:
 Measurement
The display indicates the measurement data in the measured value
display
 Calibration
The display guides you through a calibration procedure with
calibration information
 Configuration
The system menu or a sensor menu with submenus, settings and
functions is displayed

4.1.2
Keys
Functions

Operation

The meter is operated via keys. The keys can have different functions
with long or short keystrokes.
Generally, with a short keystroke a function is carried out. A long
keystroke opens a setting menu.
In a setting menu, settings are selected with the <▲><▼> keys.
A setting is confirmed with . With confirming, the setting is
finished and the next setting is displayed.

Representation

In this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. ) generally indicates a short keystroke
(under 2 sec) in this operating manual. A long keystroke (approx.
2 sec) is indicated by the underscore behind the key symbol (e.g.
).

_

ba75791e05

05/2011

15

Operation

pH 3110

4.2
Preparatory activities

Measuring

Perform the following preparatory activities when you want to measure:
1

Connect the pH combination electrode to the meter.

2

Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a
temperature sensor.

3

Calibrate or check the meter with the combination electrode.

4

Select the measured parameter with .

Note
Incorrect calibration of pH combination electrodes leads to incorrect
measured values. Calibrate regularly before measuring.
Stability control
AutoRead

During the measuring procedure, the stability control function is
automatically activated. The stability control function (drift control)
checks the stability of the measured pH signal and the stability of the
measured temperature signal. The stability has a considerable effect
on the reproducibility of the measured value.
For identical measurement conditions, the following applies:

Temperature sensor

Measured
parameter

Reproducibility

Response time

pH value

< 0.02 pH units

> 15 seconds

Temperature

< 0.3 °C of temperature value

> 15 seconds

You can measure with or without a temperature sensor. If a
temperature sensor is connected, it is indicated on the display by TP.
Note
The pH meter automatically recognizes the type of the temperature
sensor used. Therefore, you can connect combination electrodes with
an NTC30 or Pt1000.
The temperature measurement is absolutely essential for a
reproducible pH measurement. If the measurement is made without a
temperature sensor, proceed as follows:

16

1

Measure the current temperature using a thermometer.

2

Set the temperature value with <▲><▼>.

ba75791e05

05/2011

pH 3110

Operation

Note
When calibrating without temperature sensor, also set the current
temperature of each buffer solution manually with the <▲><▼> keys.

4.2.1

Measuring the pH value

1

Perform the preparatory activities according to section 4.2.

2

Immerse the pH combination electrode in the test sample.

3

Using , scroll as necessary until the measured parameter
pH is displayed.

4

Wait for a stable measured value.
The AR display indicator flashes as long as the measured
value is not yet stable.

pH

6992
248

TP

°C
AR

ba75791e05

05/2011

17

Operation

pH 3110

4.2.2

Measuring the ORP

The meter can, in conjunction with an ORP electrode, measure the
ORP (mV) of a solution.
1

Perform the preparatory activities according to section 4.2.

2

Immerse the ORP electrode in the test sample.

3

If necessary, call up the measured parameter U with .

4

Wait for a stable measured value.
The AR display indicator flashes as long as the measured
value is not yet stable.

U

4 8 63
213

mV

TP

°C
AR

Note
ORP electrodes are not calibrated. However, you can check ORP
electrodes using a test solution.

18

ba75791e05

05/2011

pH 3110

Operation

4.3
Why calibrate?

When do you have to
calibrate?

Calibration

pH combination electrodes age. This changes the zero point
(asymmetry) and slope of the pH combination electrode. As a result, an
inexact measured value is displayed. Calibration determines the
current values of the zero point and slope of the combination electrode
and stores them in the measuring instrument. Thus, you should
calibrate at regular intervals.
 After connecting another combination electrode
 When the sensor symbol flashes on the display:
– e.g. after the calibration interval has expired

AutoCal

is adapted to the permanently programmed buffer solutions as a fully
automatic single-point, two-point or three-point calibration. The buffer
solutions are automatically recognized by the meter.
The following buffer sets are suitable:
 Technical buffers (AutoCal TEC)
 DIN buffers (AutoCal DIN)

Calibration points

Calibration can be performed using one, two or three buffer solutions in
any order (single-point, two-point or three-point calibration). The meter
determines the following values and calculates the calibration line as
follows:

1-point

Determined values

Displayed calibration data

ASY

 Zero point = ASY
 Slope = Nernst slope
(-59.16 mV/pH at 25 °C)

2-point
3-point

ASY
SLO

 Zero point = ASY
 Slope = SLO

Note
You can display the slope in the units, mV/pH or %.
You can display the zero point in the units, mV or pH.
ConCal

Stability control
AutoRead

ba75791e05

05/2011

This function is a conventional two-point calibration using two buffer
solutions (pH 7.0 ± 0.5 and any other buffer solution) or a single-point
calibration using any buffer solution and is used as a high-speed
method.
The calibration procedure automatically activates the stability control
function. The current measurement with stability control can be
terminated at any time (accepting the current value) by pressing
.
19

Operation

Calibration record
Displaying the
calibration data
Calibration evaluation

pH 3110

When finishing a calibration, the new calibration values are first
displayed as an informative message and stored.
You can display the data of the last calibration (see section 4.4).
After calibrating, the meter automatically evaluates the calibration. The
zero point and slope are evaluated separately. The worse evaluation of
both is taken into account. The evaluation appears on the display as
the sensor symbol and in the calibration record.
Sensor symbol

Zero point
[mV]

Slope
[mV/pH]

-15 ... +15

-60.5 ... -58

-20 ... +20

-58 ... -57

-25 ... +25

-61 ... -60.5
or
-57 ... -56

-30 ... +30

-62 ... -61
or
-56 ... -50

< -30 or
> 30

... -62 or
... -50

Clean the combination electrode
according to the electrode operating
manual
CalError
Eliminate the error according to
chapter 6 WHAT TO DO IF...

Preparatory activities

20

1

Switch on the meter with .

2

Connect the pH combination electrode to the meter.

3

Keep the buffer solutions ready.

4

Adjust the temperature of the solutions and measure the
current temperature if the measurement is made without
temperature sensor.

ba75791e05

05/2011

pH 3110

Operation

5
4.3.1

Set the buffer set to be used for calibration as necessary.
Calibration interval (Int.C)

The calibration interval reminds you to calibrate regularly. When the
adjusted calibration interval (Int.C) has expired, the sensor symbol
flashes. It is still possible to measure.
Note
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.
Setting the calibration
interval

The calibration interval (Int.C) is set to 7 days (d7) in the factory.
You can change the interval (1 ... 999 days):

_>.

1

Open the menu for measurement settings with  until Int.C is displayed.

pH

Time

ba75791e05

05/2011

1NT C
D

8

3

Set the calibration interval with <▲><▼>.

4

Confirm the setting with .

21

Operation

pH 3110

4.3.2

Automatic calibration (AutoCal)

For the AutoCal TEC and AutoCal DIN procedures, use one to three
buffer solutions of the relevant buffer set (Technical buffers or DIN
buffers) in any order.
Note
The steps 2 and 6 are not necessary if you use a temperature sensor.
1

Press  repeatedly until the AutoCal TEC or AutoCal DIN
function display indicator appears.

pH

pH

C DI

CT1

2 4 8 °C

TP
AutoCal DIN

2 4 8 °C

TP

AutoCal TEC

2

If necessary, set the temperature of the buffer solution with
<▲><▼>.

3

Immerse the pH combination electrode in the first buffer
solution.

4

Start the measurement with .
The AR display indicator flashes.
The electrode voltage (mV) or the nominal value of the buffer
(setting: see section 4.5.2) is displayed. As soon as a stable
value is recognized, Ct2 or Cd2 appears.

pH

pH

CT2

2 4 8 °C

C D2
TP
AutoCal DIN

2 4 8 °C

TP

ba75791e05

05/2011

AutoCal TEC

22

pH 3110

Operation

Note
Here you can cancel the calibration procedure with . This
corresponds to a single-point calibration. The value of the zero point
(ASY) is displayed. Pressing  displays the value of the slope
(SLO).
Continuing with twopoint calibration

5

Thoroughly rinse the combination electrode with distilled water.

6

If necessary, set the temperature of the second buffer solution
with <▲><▼>.

7

Immerse the combination electrode in the second buffer
solution.

8

Press the  key.
The AR display indicator flashes.
The electrode voltage (mV) or the nominal value of the buffer
(setting: see section 4.5.2) is displayed. As soon as a stable
value is recognized, Ct3 or Cd3 appears.

pH

pH

CT3

2 4 8 °C

C D3
TP
AutoCal DIN

2 4 8 °C

TP

AutoCal TEC

Note
Here you can cancel the calibration procedure with . This
corresponds to a two-point calibration. The value of the zero point
(ASY) is displayed. Pressing  displays the value of the slope
(SLO).
Continuing with threepoint calibration

ba75791e05

05/2011

9

Thoroughly rinse the combination electrode with distilled water.

10

If necessary, set the temperature of the third buffer solution
with <▲><▼>.

11

Immerse the combination electrode in the third buffer solution.

23

Operation

pH 3110

12

Press the  key.
The AR display indicator flashes.
The electrode voltage (mV) or the nominal value of the buffer
(setting: see section 4.5.2) is displayed. The asymmetry is
displayed as soon as a stable value is recognized.

13

Press the  key.
The value of the slope (mV/pH) appears on the display.

14

To return to the measuring mode: Press the  key.

Note
While the zero point (ASY) is being displayed, you can change the unit
of the zero point with <▲><▼>.
While the slope (SLO) is being displayed, you can change the unit of
the slope with <▲><▼>.
The % display refers to the Nernst slope of 59.2 mV/pH at 25° C
(100 x determined slope/Nernst slope).
The unit of zero point and slope can also be changed permanently in
the measurement settings (see section 4.5.2).

4.3.3
Single-point calibration

Two-point calibration

Conventional calibration (ConCal)

Use any buffer solution for this rapid method.
The calibration will be the more exact the nearer the pH value of the
buffer solution is to that of the test sample.
Use two buffer solutions for this procedure:
 pH 7.0 ± 0.5
 any other buffer solution
Note
The steps 2 and 8 are not necessary if you use a pH combination
electrode with a temperature sensor.

1

24

Press the  key repeatedly until the ASY display indicator
and the ConCal function indicator appear.

ba75791e05

05/2011

pH 3110

Operation

ASY
248

TP

°C

ConCal

2

If necessary, set the temperature of the first buffer solution with
<▲><▼>.

3

Immerse the pH combination electrode in the first buffer
solution (pH 7.0 ± 0.5 for two-point calibration).

4

Press the  key.
The measured pH value appears on the display.

5

When the measured value is stable,
using <▲><▼>, set the measured value to the nominal pH
value of the buffer solution (at the current temperature).

6

Press the  key. SLO appears on the display.

Note
Here you can cancel the calibration procedure with . This
corresponds to a single-point calibration. The value of the zero point
(ASY) is displayed. The sensor symbol shows the evaluation of the
single-point calibration. Pressing  displays the value of the
slope (SLO).

sL O
248

TP

°C

ConCal

ba75791e05

05/2011

25

Operation

pH 3110

7

To continue the two-point calibration, thoroughly rinse the
combination electrode with deionized water.

8

If necessary, set the temperature of the second buffer solution
with <▲><▼>.

9

Immerse the combination electrode in the second buffer
solution.

10

Press the  key. The second pH value appears on the
display.

11

When the measured value is stable,
using <▲><▼>, set the measured value to the nominal pH
value of the buffer solution (at the current temperature).

12

Press the  key.
The value of the zero point (ASY) is displayed.

13

Press the  key.
The value of the slope (SLO) is displayed.

14

Press the  key.
The measuring mode is active.

Note
While the zero point (ASY) is being displayed, you can change the unit
of the zero point with <▲><▼>.
While the slope (SLO) is being displayed, you can change the unit of
the slope with <▲><▼>.
The % display refers to the Nernst slope of 59.2 mV/pH at 25° C
(100 x determined slope/Nernst slope).
The units of the zero point and slope can also be changed permanently
in the measurement settings (see section 4.5.2).

26

ba75791e05

05/2011

pH 3110

Operation

4.4

Downloading calibration data

You can download calibration data to the display as follows:
1

Pressing  displays the calibration data (asymmetry).

02

mV

ASy
ConCal

Note
While the calibration data is displayed you can press <▲><▼> to
switch over the unit of the zero point (ASY).

2

With  display the slope.

- 592

mV/pH

SL O
ConCal

Note
While the calibration data is displayed you can press <▲><▼> to
switch over the unit of the slope (SLO).

ba75791e05

05/2011

27

Operation

pH 3110

4.5

Settings

You can adapt the meter to your individual requirements. The settings
are done in the following menus:
 System settings ()

– Switch-off interval (t.Off)
 Measurement settings ()

– Display of the buffer during calibration
(pH nominal value or measured voltage value in mV)
– Unit of the value for the slope (mV/pH or %)
– Unit of the value for the zero point (mV, pH)
– Measured value resolution, pH (0.000 / 0.00 / 0.0)
– Measured value resolution, U (0.0 / 0)
– Temperature unit (°C / °F)
– Calibration interval (Int.C [0 ... 999])

Note
You can exit the setting menu at any time by pressing . Settings
already modified and confirmed with  are stored.

28

ba75791e05

05/2011

pH 3110

Operation

4.5.1

System settings

The default setting is printed in bold.
Switch-off interval (t.Off)

1

10, 20, 30, 40, 50 min,
1, 2, 3, 4, 5, 10, 15, 20, 24 h

Open the menu for system settings with .

Switch-off interval (t.Off)

Time

ba75791e05

05/2011

t Of f
010 0

2

Set the switch-off interval with <▲><▼>.

3

Confirm with .
The system settings are completed.
The meter switches to the measuring mode.

29

Operation

pH 3110

4.5.2

Measurement settings

These settings apply to calibration and measurement (the default
setting is printed in bold).
Display during calibration (BUFFER)

pH (buffer nominal value),
U (combination electrode
voltage)

Unit of the value for the slope (SLO)

mV/pH, %

Unit of the value for the zero point
(ASY)

mV, pH

Measured value resolution, pH

0.000 , 0.00 , 0.0

Measured value resolution, U

0.0 , 0

Temperature unit (UnIt)

°C, °F

Calibration interval (Int.C)

0 ... 7 ... 999 d

1

Display during
calibration
(bUFF)

_

With  open the menu for measurement and calibration
settings.
The first setting is displayed.

pH

b UF F
AutoCal DIN

pH

AutoCal TEC

30

2

Using <▲><▼>, select the pH or U display during calibration.

3

Confirm with .
SLO, the unit of the value for the slope (mV/pH or %) is
displayed.

ba75791e05

05/2011

pH 3110

Operation

Unit of the value for the
slope
(SLO)

pH

SL O
4

Using <▲><▼>, select the unit for the slope.

5

Confirm with .
ASY, the unit of the value for the zero point (mV or pH) is
displayed.

Unit of the value for the
zero point
(ASY)

ba75791e05

05/2011

mV/pH

pH

ASY

mV

6

Using <▲><▼>, select the unit for the zero point.

7

Confirm with .
res, the resolution of the pH display
(0.0, 0.00 or 0.000) appears on the display.

31

Operation

pH 3110

pH resolution
(res)

pH

res

o o oo

8

Select the resolution for the pH display with <▲><▼>.

9

Confirm with .
res, the resolution of the voltage display (0.0, 0.00 or 0.000)
appears on the display.

U resolution
(res)

U

res

mV

oo

32

10

Using <▲><▼>, select the unit for the voltage display.

11

Confirm with .
Unit, the setting of the unit of the temperature value is
displayed.

ba75791e05

05/2011

pH 3110

Operation

Temperature unit (Unit)

Unit
oC

TP

°C

12

Using <▲><▼>, toggle between °C and °F.

13

Confirm with .
Int.C, the setting of the calibration interval is displayed.

Calibration
interval
(Int.C)

Time

INT C
d

AutoCal DIN

7

AutoCal TEC

ba75791e05

05/2011

14

Set the interval with <▲><▼>.

15

Confirm with .
The measurement settings are completed.
The meter switches to the measuring mode.

33

Operation

pH 3110

4.6

Reset

4.6.1

Resetting calibration values

This function resets the calibration values to the default condition. All
other meter settings are retained.
Calibration values in the
default condition

Zero point

0 mV (pH 7.000)

Slope

-59.16 mV/pH (100 %)

Note
The measuring system is not calibrated after a reset. Before
measuring, recalibrate the meter.
Resetting calibration
values

1

_

Press  to open the menu for the reset of the
calibration data.
Init.C is displayed.

pH

I ni t C
no

34

2

Press <▲><▼> to display no or YES.
YES: Reset the calibration values.
no: Retain the calibration values.

3

Confirm with .
The menu is finished.
The meter switches to the measuring mode.

ba75791e05

05/2011

pH 3110

Operation

4.6.2

Resetting all meter settings

This function resets all meter settings to the default condition. The
relevant values are given in the following sections:

Resetting the meter
settings

System settings

section 4.5.1

Measurement settings

section 4.5.2

1

Switch on the meter with .
The display test appears briefly on the display.

2

During the display test, press  to open the menu for the
reset of the meter settings.
Init is displayed.

Init
no

3

Press <▲><▼> to display no or YES.
YES: Reset the meter settings.
no: Retain the meter settings.

4

Confirm with .
The menu is finished.
The meter switches to the measuring mode.

Note
The measuring system is not calibrated after a reset. Before
measuring, recalibrate the meter.

ba75791e05

05/2011

35

Operation

36

pH 3110

ba75791e05

05/2011

pH 3110

Maintenance, cleaning, disposal

5

Maintenance, cleaning, disposal

5.1

Maintenance

The only maintenance activity required is replacing the batteries.
Note
See the relevant operating manuals of the combination electrodes for
instructions on maintenance.
5.1.1

Replacing the batteries

1

Unscrew the two screws (1) on the underside of the meter.

2

Open the battery compartment (2) on the underside of the
meter.

1

2

3

Remove the four batteries from the battery compartment.

4

Place four new batteries (type Mignon AA) in the battery
compartment.

Note
Alternatively, you can also use Ni-MH rechargeable batteries (type
Mignon AA). In order to charge the batteries, an external charging
device is required.

Caution
Make sure that the poles of the batteries are positioned correctly.
The ± signs on the batteries must correspond to the ± signs in the
battery compartment.

ba75791e05

05/2011

37

Maintenance, cleaning, disposal

pH 3110

5

5.2

Close the battery compartment (2) and tighten the screws (1).

Cleaning

Occasionally wipe the outside of the meter with a damp, lint-free cloth.
Disinfect the housing with isopropanol as required.

Caution
The housing is made of synthetic material (ABS). Thus, avoid
contact with acetone or similar detergents that contain solvents.
Remove any splashes immediately.

5.3

Packing

This meter is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing
protects the meter against damage during transport.

5.4

Disposal

Note
This meter contains batteries. Batteries that have been removed must
only be disposed of at a recycling facility set up for this purpose or via
the retail outlet.
It is illegal to dispose of them in household refuse.

38

ba75791e05

05/2011

pH 3110

What to do if...

6
Error message
OFL, UFL

Error message
CalError

What to do if...

Cause

Remedy

pH combination electrode:
– Measured value outside the
measuring range

– Use suitable combination
electrode

– Air bubble in front of the
diaphragm

– Remove air bubble

– Air in the diaphragm

– Extract air or moisten
diaphragm

– Cable broken

– Replace combination
electrode

– Gel electrolyte dried out

– Replace combination
electrode

Cause

Remedy

pH combination electrode:
– The values determined for zero
point and slope of the
combination electrode are
outside the allowed limits.

– Recalibrate

– Diaphragm contaminated

– Clean diaphragm

– Combination electrode broken

– Replace combination
electrode

Buffer solutions

ba75791e05

05/2011

– Incorrect buffer solutions

– Change calibration
procedure

– Buffer solutions too old

– Use only once.
Note the shelf life

– Buffer solutions depleted

– Change solutions

39

What to do if...

pH 3110

No stable measured
value

Cause

Remedy

pH combination electrode:
– Diaphragm contaminated

– Clean diaphragm

– Membrane contaminated

– Clean membrane

Test sample:
– pH value not stable

– Measure with air excluded if
necessary

– Temperature not stable

– Adjust temperature if
necessary

Combination electrode + test
sample:

Sensor symbol flashes

– Conductivity too low

– Use suitable combination
electrode

– Temperature too high

– Use suitable combination
electrode

– Organic liquids

– Use suitable combination
electrode

Cause
– Calibration interval expired

Display,
LoBat

40

Cause
– Batteries almost empty

Remedy
– Recalibrate the measuring
system

Remedy
– Replace the batteries (see
section 5.1 MAINTENANCE)

ba75791e05

05/2011

pH 3110

What to do if...

Obviously incorrect
measured values

Cause
pH combination electrode:
– pH combination electrode
unsuitable

– Use suitable combination
electrode

– Temperature difference
between buffer and test sample
too high

– Adjust temperature of buffer
or sample solutions

– Measurement procedure not
suitable

– Follow special procedure

Meter does not react to
keystroke

Cause

You want to know which
software
version is in the meter

Cause

ba75791e05

05/2011

Remedy

– Operating condition undefined
or EMC load unallowed

– E. g., a question by the service
department

Remedy
– Processor reset:
Press the  and
 key
simultaneously

Remedy
– Switch on the meter.
During the display test,
display the software version
with .

41

What to do if...

42

pH 3110

ba75791e05

05/2011

pH 3110

Technical data

Dimensions
Weight
Mechanical structure

7

Technical data

7.1

General data

approx. 180 x 80 x 55 mm
approx. 0.4 kg
Type of protection

IP 67

Electrical safety

Protective class

III

Test certificates

CE, cETLus

Ambient
conditions

Power
supply

Sensor input

Guidelines
and norms used

ba75791e05

05/2011

Storage

- 25 °C ... + 65 °C

Operation

-10 °C ... + 55 °C

Allowable relative
humidity

Annual mean: < 75 %
30 days/year: 95 %
Other days: 85 %

Batteries

4 x 1.5 V alkali-manganese batteries,
type AA

Rechargeable
batteries

4 x 1,2 V NiMH rechargeable batteries,
type AA
(no charging function)

Operational life

Approx. 2500 h operating hours (batteries)

Input resistance

> 5 * 1012 ohm

Input current

< 1 * 10-12 A

EMC

EC directive 2004/108/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A

Meter safety

EC directive 2006/95/EC
EN 61010-1

IP protection

EN 60529

43

Technical data

pH 3110

7.2
Measuring ranges,
resolution

Measuring ranges, resolution, accuracy

Variable

Measuring range

Resolution

pH

- 2.0 ... + 20.0

0.1

- 2.00 ... + 20.00

0.01

- 2.000 ... + 19.999

0.001

- 1200.0 ... + 1200.0

0.1

- 2000 ... + 2000

1

T [°C]

- 5.0 ... + 105.0

0.1

T [°F]

23.0 ... + 221.0

0.1

Variable

Range

Increment

Tmanual [°C]

- 25 ... + 130

1

Tmanual [°F]

-13 ... + 266

1

Variable

Accuracy

U [mV]

Manual
temperature input

Accuracy (± 1 digit)

Temperature of the test
sample

pH / range *
- 2.0 ... + 20.0

± 0.1

+ 15 °C ... + 35 °C

- 2.00 ... + 20.00

± 0.01

+ 15 °C ... + 35 °C

- 2.000 ... + 19.999

± 0.005

+ 15 °C ... + 35 °C

- 2000 ... + 2000

±1

+ 15 °C ... + 35 °C

-1200.0 ... +1200.0

± 0.3

+ 15 °C ... + 35 °C

U [mV] / range

T [°C] / temperature sensor
NTC 30

± 0.1

PT 1000

± 0.1

* when measuring in a range of ± 2 pH around a calibration point

Note
The accuracy values specified here apply exclusively to the meter. The
accuracy of the combination electrodes and buffer solutions has to be
taken into account additionally.

44

ba75791e05

05/2011

pH 3110

Lists

8

Lists

This chapter provides additional information and orientation aids.

Specialist terms

Index

The glossary briefly explains the meaning of the specialist terms.
However, terms that should already be familiar to the target group are
not described here.

The index helps you to find the topics that you are looking for.

Glossary
Adjusting

To manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or
a value that is regarded as correct, or that the difference remains
within the tolerance.

Asymmetry

see zero point

AutoRange

Name of the automatic selection of the measuring range.

AutoRead
Calibration

Function to control the measured value stability.
Comparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct.
Often, this expression is also used when the measuring system is
adjusted at the same time (see adjusting).

Electromotive force of
a combination
electrode

The electromotive force U of the combination electrode is the
measurable electromotive force of an electrode in a solution. It equals
the sum of all the galvanic voltages of the electrode. Its dependency
on the pH results in the electrode function, which is characterized by
the parameters, slope and zero point.

Junction

The junction is a porous body in the housing wall of reference
electrodes or electrolyte bridges. It arranges the electrical contact
between two solutions and makes the electrolyte exchange more
difficult. The expression, junction, is also used for ground or junctionless transitions.

Measured parameter

ba75791e05

05/2011

The measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or D.O. concentration.

45

Lists

pH 3110

Measured value

Molality

Molality is the quantity (in Mol) of a dissolved substance in 1000 g
solvent.

ORP voltage

The ORP is caused by oxidizing or reducing substances dissolved in
water, if these substances become effective at an electrode surface
(e. g. a gold or platinum surface).

pH value

The pH is a measure of the acidic or basic effect of an aqueous
solution. It corresponds to the negative decadic logarithm of the molal
hydrogen ions activity divided by the unit of the molality. The practical
pH value is the value of a pH measurement.

Potentiometry

Name of a measuring technique. The signal (depending on the
measured parameter) of the electrode is the electrical potential. The
electrical current remains constant.

Reset
Resolution
Slope
Standard solution
Test sample

Zero point

46

The measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and
unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).

Restoring the original condition of all settings of a measuring system.
Smallest difference between two measured values that can be
displayed by a measuring instrument.
The slope of a linear calibration function.
The standard solution is a solution where the measured value is
known by definition. It is used to calibrate a measuring system.
Designation of the test sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample
and original sample are identical if the test sample was not processed.
The zero point of a pH electrode is the pH value at which the
electromotive force of the pH electrode at a specified temperature is
zero. Normally, this is at 25 °C.

ba75791e05

05/2011

pH 3110

Lists

Index

A

P

Authorized use . . . . . . . . . . . . . . . . . . . . . . 10
AutoCal . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
AutoRead . . . . . . . . . . . . . . . . . . . . . . . . . . 16

Precautions . . . . . . . . . . . . . . . . . . . . . . . . . 9

B
Battery compartment . . . . . . . . . . . . . . 13, 37

R
Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
All meter settings . . . . . . . . . . . . . . . . . 35
Calibration values . . . . . . . . . . . . . . . . 34
Resolution setting . . . . . . . . . . . . . 30, 31, 32

C
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . 19
Calibration evaluation
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Calibration interval . . . . . . . . . . . . . . . . . . . 21
Calibration points
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
ConCal . . . . . . . . . . . . . . . . . . . . . . . . . 19, 24

S
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Scope of delivery . . . . . . . . . . . . . . . . . . . . 13
Single-point calibration . . . . . . . . . 19, 23, 25
Slope
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Socket field . . . . . . . . . . . . . . . . . . . . . . . . . 8
Stability control . . . . . . . . . . . . . . . . . . . . . 16

D
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

T

E

Two-point calibration . . . . . . . . . . . . . . . . . 19
ConCal . . . . . . . . . . . . . . . . . . . . . . . . . 24

Energy saving feature . . . . . . . . . . . . . . . . 14

I

Z
Zero point of pH combination electrode . . . 19

Initial commissioning . . . . . . . . . . . . . . . . . 14
Interval
Calibration . . . . . . . . . . . . . . . . . . . . . . 21

K
Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

L
LoBat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

M
Measurement accuracy . . . . . . . . . . . . . . . 21

O
Operational safety . . . . . . . . . . . . . . . . . . . 10
ORP electrode . . . . . . . . . . . . . . . . . . . . . . 18
ORP voltage . . . . . . . . . . . . . . . . . . . . . . . . 18

ba75791e05

05/2011

47

Lists

48

pH 3110

ba75791e05

05/2011

Wissenschaftlich-Technische Werkstätten GmbH
Dr.-Karl-Slevogt-Straße 1
D-82362 Weilheim
Germany
Tel:
Fax:
E-Mail:
Internet:

+49 (0) 881 183-0
+49 (0) 881 183-100
+49 (0) 881 183-420
Info@WTW.com
http://www.WTW.com



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : Yes
Author                          : direim
Create Date                     : 2011:05:20 09:07:47Z
Modify Date                     : 2015:09:17 13:48:22-06:00
Has XFA                         : No
XMP Toolkit                     : Adobe XMP Core 5.6-c015 81.157285, 2014/12/12-00:43:15
Format                          : application/pdf
Title                           : WTW pH 3110 Operating Manual
Creator                         : direim
Creator Tool                    : FrameMaker 10.0
Metadata Date                   : 2015:09:17 13:48:22-06:00
Producer                        : Acrobat Elements 9.0.0 (Windows)
Document ID                     : uuid:8d6d5442-926b-4db1-b644-aae6851137f3
Instance ID                     : uuid:0c954173-4f50-6844-88e8-959da1fdb808
Page Mode                       : UseOutlines
Page Count                      : 50
EXIF Metadata provided by EXIF.tools

Navigation menu