WTW PH 3110 Operating Manual WTW_p H_3110 P H
User Manual: Pdf WTW_pH_3110
Open the PDF directly: View PDF .
Page Count: 50
Download | |
Open PDF In Browser | View PDF |
Operating manual pH 3110 69 pH 92 24 8 °C TP AR pH meter For Sales & Service Contact 2650 E. 40th Ave. • Denver, CO 80205 Phone 303-320-4764 • Fax 303-322-7242 1-800-833-7958 ba75791e05 05/2011 www.geotechenv.com pH 3110 Accuracy when going to press Copyright The use of advanced technology and the high quality standard of our instruments are the result of a continuous development. This may result in differences between this operating manual and your meter. Also, we cannot guarantee that there are absolutely no errors in this manual. Therefore, we are sure you will understand that we cannot accept any legal claims resulting from the data, figures or descriptions. © Weilheim 2008, WTW GmbH Reproduction in whole - or even in part - is prohibited without the express written permission of WTW GmbH, Weilheim. Printed in Germany. ba75791e05 05/2011 pH 3110 Contents pH 3110 - Contents 1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 1.1 1.2 1.3 2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 2.1 2.2 3 4.2 4.3 4.4 4.5 4.6 5.2 5.3 05/2011 13 13 13 14 General operating principles . . . . . . . . . . . . . . . . . . . . . . 4.1.1 Operating modes . . . . . . . . . . . . . . . . . . . . . . . . 4.1.2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2.1 Measuring the pH value . . . . . . . . . . . . . . . . . . . 4.2.2 Measuring the ORP . . . . . . . . . . . . . . . . . . . . . . Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.3.1 Calibration interval (Int.C) . . . . . . . . . . . . . . . . . 4.3.2 Automatic calibration (AutoCal) . . . . . . . . . . . . . 4.3.3 Conventional calibration (ConCal) . . . . . . . . . . . Downloading calibration data . . . . . . . . . . . . . . . . . . . . . Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.5.1 System settings . . . . . . . . . . . . . . . . . . . . . . . . . 4.5.2 Measurement settings . . . . . . . . . . . . . . . . . . . . Reset. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6.1 Resetting calibration values . . . . . . . . . . . . . . . . 4.6.2 Resetting all meter settings . . . . . . . . . . . . . . . . 15 15 15 16 17 18 19 21 22 24 27 28 29 30 34 34 35 Maintenance, cleaning, disposal . . . . . . . . . . . . . . . . . 37 5.1 ba75791e05 Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Initial commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.2.1 Inserting the batteries. . . . . . . . . . . . . . . . . . . . . 3.2.2 Switching on the meter. . . . . . . . . . . . . . . . . . . . Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 4.1 5 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . 10 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 3.1 3.2 4 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Socket field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1.1 Replacing the batteries . . . . . . . . . . . . . . . . . . . Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Packing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 37 38 38 3 Contents pH 3110 5.4 6 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 7 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 7.1 7.2 8 4 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Measuring ranges, resolution, accuracy . . . . . . . . . . . . . 44 Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 ba75791e05 05/2011 pH 3110 Overview 1 Overview The compact pH 3110 precision pH meter enables you to perform pH measurements rapidly and reliably. The pH 3110 provides the maximum degree of operating comfort, reliability and measuring certainty for all applications. The proven calibration procedures and automatic stability control function (AR) support your work with the pH meter. 3 69 pH 92 24 8 °C 2 TP AR 1 ba75791e05 05/2011 1 Keypad 2 Display 3 Socket field 5 Overview pH 3110 1.1 Keypad pH 3110 In this operating manual, keys are indicated by brackets <..> . The key symbol (e.g.) generally indicates a short keystroke (under 2 sec) in this operating manual. A long keystroke (approx. 2 sec) is indicated by the underscore behind the key symbol (e.g. ). _ : : Switches the meter on/off Resets calibration data : : Selects the measured parameter Opens the setting menu for calibration and measurements : : Calls up the calibration procedure Displays the calibration data <▲>: Increments values, scrolls <▼>: Decrements values, scrolls _ _ _ : : _ 6 Confirms entries Opens the setting menu for system settings ba75791e05 05/2011 pH 3110 Overview 1.2 Display Status display Sensor symbol (calibration evaluation) Measured value display Function and temperature display indicators Function display indicators ba75791e05 05/2011 AutoCal TEC Calibration with automatic buffer recognition (buffer set: Technical buffers ) AutoCal DIN Calibration with automatic buffer recognition (buffer set: DIN buffers ) ConCal Calibration with any buffers CalError An error occurred during calibration LoBat With battery operation: Batteries almost empty AR Stability control (AutoRead) is active TP Temperature measurement active TIME Setting of calibration interval 7 Overview pH 3110 1.3 Socket field 1 4 2 3 Connectors: 1 pH electrode 2 Reference electrode 3 Temperature sensor 4 Service interface Caution Only connect sensors to the meter that cannot return any voltages or currents that are not allowed (> SELV and > current circuit with current limiting). Almost all customary sensors fulfill these conditions. 8 ba75791e05 05/2011 pH 3110 Safety 2 Safety This operating manual contains basic instructions that you must follow during the commissioning, operation and maintenance of the meter. Consequently, all responsible personnel must read this operating manual before working with the meter. The operating manual must always be available within the vicinity of the meter. Target group Safety instructions The meter was developed for work in the field and in the laboratory. Thus, we assume that, as a result of their professional training and experience, the operators will know the necessary safety precautions to take when handling chemicals. Safety instructions in this operating manual are indicated by the warning symbol (triangle) in the left column. The signal word (e.g. "Caution") indicates the level of danger: Warning indicates instructions that must be followed precisely in order to avoid possibly great dangers to personnel. Caution indicates instructions that must be followed precisely in order to avoid the possibility of slight injuries or damage to the meter or the environment. Further notes Note indicates notes that draw your attention to special features. Note indicates cross-references to other documents, e.g. operating manuals. ba75791e05 05/2011 9 Safety pH 3110 2.1 Authorized use This meter is authorized exclusively for pH and ORP measurements in a laboratory or field environment. The technical specifications as given in chapter 7 TECHNICAL DATA must be observed. Only the operation and running of the meter according to the instructions given in this operating manual is authorized. Any other use is considered unauthorized. 2.2 General safety instructions This meter is constructed and tested in compliance with the IEC 1010 safety regulations for electronic measuring instruments. It left the factory in a safe and secure technical condition. Function and operational safety The smooth functioning and operational safety of the meter can only be guaranteed if the generally applicable safety measures and the specific safety instructions in this operating manual are followed during operation. The smooth functioning and operational safety of the meter can only be guaranteed under the environmental conditions that are specified in chapter 7 TECHNICAL DATA. If the meter was transported from a cold environment to a warm environment, the formation of condensate can lead to the faulty functioning of the meter. In this event, wait until the temperature of the meter reaches room temperature before putting the meter back into operation. Caution The meter is only allowed to be opened by authorized personnel. 10 ba75791e05 05/2011 pH 3110 Safety Safe operation If safe operation is no longer possible, the meter must be taken out of service and secured against inadvertent operation! Safe operation is no longer possible if the meter: has been damaged in transport has been stored under adverse conditions for a lengthy period of time is visibly damaged no longer operates as described in this manual. If you are in any doubt, please contact the supplier of the meter. Obligations of the purchaser The purchaser of this meter must ensure that the following laws and guidelines are observed when using dangerous substances: EEC directives for protective labor legislation National protective labor legislation Safety regulations Safety datasheets of the chemical manufacturers. Caution In addition to the safety instructions mentioned here, also follow the safety instructions of the sensors used. The operating manuals of the sensors are available on the supplied CD and on the Internet under www.WTW.com. ba75791e05 05/2011 11 Safety 12 pH 3110 ba75791e05 05/2011 pH 3110 Commissioning 3 Commissioning 3.1 Scope of delivery Handheld meter, pH 3110 4 batteries 1.5 V Mignon type AA Short instructions CD-ROM with detailed operating manual 3.2 Initial commissioning Perform the following activities: Insert the supplied batteries Switch on the meter 3.2.1 Inserting the batteries 1 Unscrew the two screws (1) on the underside of the meter. 2 Open the battery compartment (2) on the underside of the meter. 1 2 3 ba75791e05 05/2011 Place four batteries (type Mignon AA) in the battery compartment. 13 Commissioning pH 3110 Note Alternatively, you can also use Ni-MH rechargeable batteries (type Mignon AA). In order to charge the batteries, an external charging device is required. Caution Make sure that the poles of the batteries are positioned correctly. The ± signs on the batteries must correspond to the ± signs in the battery compartment. 4 3.2.2 1 Close the battery compartment (2) and tighten the screws (1). Switching on the meter Press the key. A display test is briefly displayed. Subsequently, the meter switches to the measuring mode (measured value display). Note The meter has an energy saving feature to avoid unnecessary battery depletion during battery operation. The energy saving feature switches off the meter if no key was pressed during the specified interval (setting the switch-off interval see section 4.5.1). 14 ba75791e05 05/2011 pH 3110 Operation 4 Operation 4.1 General operating principles This section contains basic information on the operation of the pH 3110. 4.1.1 Operating modes The meter has the following operating modes: Measurement The display indicates the measurement data in the measured value display Calibration The display guides you through a calibration procedure with calibration information Configuration The system menu or a sensor menu with submenus, settings and functions is displayed 4.1.2 Keys Functions Operation The meter is operated via keys. The keys can have different functions with long or short keystrokes. Generally, with a short keystroke a function is carried out. A long keystroke opens a setting menu. In a setting menu, settings are selected with the <▲><▼> keys. A setting is confirmed with . With confirming, the setting is finished and the next setting is displayed. Representation In this operating manual, keys are indicated by brackets <..> . The key symbol (e.g. ) generally indicates a short keystroke (under 2 sec) in this operating manual. A long keystroke (approx. 2 sec) is indicated by the underscore behind the key symbol (e.g. ). _ ba75791e05 05/2011 15 Operation pH 3110 4.2 Preparatory activities Measuring Perform the following preparatory activities when you want to measure: 1 Connect the pH combination electrode to the meter. 2 Adjust the temperature of the buffer solutions or test solutions, or measure the current temperature, if you measure without a temperature sensor. 3 Calibrate or check the meter with the combination electrode. 4 Select the measured parameter with . Note Incorrect calibration of pH combination electrodes leads to incorrect measured values. Calibrate regularly before measuring. Stability control AutoRead During the measuring procedure, the stability control function is automatically activated. The stability control function (drift control) checks the stability of the measured pH signal and the stability of the measured temperature signal. The stability has a considerable effect on the reproducibility of the measured value. For identical measurement conditions, the following applies: Temperature sensor Measured parameter Reproducibility Response time pH value < 0.02 pH units > 15 seconds Temperature < 0.3 °C of temperature value > 15 seconds You can measure with or without a temperature sensor. If a temperature sensor is connected, it is indicated on the display by TP. Note The pH meter automatically recognizes the type of the temperature sensor used. Therefore, you can connect combination electrodes with an NTC30 or Pt1000. The temperature measurement is absolutely essential for a reproducible pH measurement. If the measurement is made without a temperature sensor, proceed as follows: 16 1 Measure the current temperature using a thermometer. 2 Set the temperature value with <▲><▼>. ba75791e05 05/2011 pH 3110 Operation Note When calibrating without temperature sensor, also set the current temperature of each buffer solution manually with the <▲><▼> keys. 4.2.1 Measuring the pH value 1 Perform the preparatory activities according to section 4.2. 2 Immerse the pH combination electrode in the test sample. 3 Using , scroll as necessary until the measured parameter pH is displayed. 4 Wait for a stable measured value. The AR display indicator flashes as long as the measured value is not yet stable. pH 6992 248 TP °C AR ba75791e05 05/2011 17 Operation pH 3110 4.2.2 Measuring the ORP The meter can, in conjunction with an ORP electrode, measure the ORP (mV) of a solution. 1 Perform the preparatory activities according to section 4.2. 2 Immerse the ORP electrode in the test sample. 3 If necessary, call up the measured parameter U with . 4 Wait for a stable measured value. The AR display indicator flashes as long as the measured value is not yet stable. U 4 8 63 213 mV TP °C AR Note ORP electrodes are not calibrated. However, you can check ORP electrodes using a test solution. 18 ba75791e05 05/2011 pH 3110 Operation 4.3 Why calibrate? When do you have to calibrate? Calibration pH combination electrodes age. This changes the zero point (asymmetry) and slope of the pH combination electrode. As a result, an inexact measured value is displayed. Calibration determines the current values of the zero point and slope of the combination electrode and stores them in the measuring instrument. Thus, you should calibrate at regular intervals. After connecting another combination electrode When the sensor symbol flashes on the display: – e.g. after the calibration interval has expired AutoCal is adapted to the permanently programmed buffer solutions as a fully automatic single-point, two-point or three-point calibration. The buffer solutions are automatically recognized by the meter. The following buffer sets are suitable: Technical buffers (AutoCal TEC) DIN buffers (AutoCal DIN) Calibration points Calibration can be performed using one, two or three buffer solutions in any order (single-point, two-point or three-point calibration). The meter determines the following values and calculates the calibration line as follows: 1-point Determined values Displayed calibration data ASY Zero point = ASY Slope = Nernst slope (-59.16 mV/pH at 25 °C) 2-point 3-point ASY SLO Zero point = ASY Slope = SLO Note You can display the slope in the units, mV/pH or %. You can display the zero point in the units, mV or pH. ConCal Stability control AutoRead ba75791e05 05/2011 This function is a conventional two-point calibration using two buffer solutions (pH 7.0 ± 0.5 and any other buffer solution) or a single-point calibration using any buffer solution and is used as a high-speed method. The calibration procedure automatically activates the stability control function. The current measurement with stability control can be terminated at any time (accepting the current value) by pressing . 19 Operation Calibration record Displaying the calibration data Calibration evaluation pH 3110 When finishing a calibration, the new calibration values are first displayed as an informative message and stored. You can display the data of the last calibration (see section 4.4). After calibrating, the meter automatically evaluates the calibration. The zero point and slope are evaluated separately. The worse evaluation of both is taken into account. The evaluation appears on the display as the sensor symbol and in the calibration record. Sensor symbol Zero point [mV] Slope [mV/pH] -15 ... +15 -60.5 ... -58 -20 ... +20 -58 ... -57 -25 ... +25 -61 ... -60.5 or -57 ... -56 -30 ... +30 -62 ... -61 or -56 ... -50 < -30 or > 30 ... -62 or ... -50 Clean the combination electrode according to the electrode operating manual CalError Eliminate the error according to chapter 6 WHAT TO DO IF... Preparatory activities 20 1 Switch on the meter with . 2 Connect the pH combination electrode to the meter. 3 Keep the buffer solutions ready. 4 Adjust the temperature of the solutions and measure the current temperature if the measurement is made without temperature sensor. ba75791e05 05/2011 pH 3110 Operation 5 4.3.1 Set the buffer set to be used for calibration as necessary. Calibration interval (Int.C) The calibration interval reminds you to calibrate regularly. When the adjusted calibration interval (Int.C) has expired, the sensor symbol flashes. It is still possible to measure. Note To ensure the high measuring accuracy of the measuring system, calibrate after the calibration interval has expired. Setting the calibration interval The calibration interval (Int.C) is set to 7 days (d7) in the factory. You can change the interval (1 ... 999 days): _>. 1 Open the menu for measurement settings with until Int.C is displayed. pH Time ba75791e05 05/2011 1NT C D 8 3 Set the calibration interval with <▲><▼>. 4 Confirm the setting with . 21 Operation pH 3110 4.3.2 Automatic calibration (AutoCal) For the AutoCal TEC and AutoCal DIN procedures, use one to three buffer solutions of the relevant buffer set (Technical buffers or DIN buffers) in any order. Note The steps 2 and 6 are not necessary if you use a temperature sensor. 1 Press repeatedly until the AutoCal TEC or AutoCal DIN function display indicator appears. pH pH C DI CT1 2 4 8 °C TP AutoCal DIN 2 4 8 °C TP AutoCal TEC 2 If necessary, set the temperature of the buffer solution with <▲><▼>. 3 Immerse the pH combination electrode in the first buffer solution. 4 Start the measurement with . The AR display indicator flashes. The electrode voltage (mV) or the nominal value of the buffer (setting: see section 4.5.2) is displayed. As soon as a stable value is recognized, Ct2 or Cd2 appears. pH pH CT2 2 4 8 °C C D2 TP AutoCal DIN 2 4 8 °C TP ba75791e05 05/2011 AutoCal TEC 22 pH 3110 Operation Note Here you can cancel the calibration procedure with . This corresponds to a single-point calibration. The value of the zero point (ASY) is displayed. Pressing displays the value of the slope (SLO). Continuing with twopoint calibration 5 Thoroughly rinse the combination electrode with distilled water. 6 If necessary, set the temperature of the second buffer solution with <▲><▼>. 7 Immerse the combination electrode in the second buffer solution. 8 Press the key. The AR display indicator flashes. The electrode voltage (mV) or the nominal value of the buffer (setting: see section 4.5.2) is displayed. As soon as a stable value is recognized, Ct3 or Cd3 appears. pH pH CT3 2 4 8 °C C D3 TP AutoCal DIN 2 4 8 °C TP AutoCal TEC Note Here you can cancel the calibration procedure with . This corresponds to a two-point calibration. The value of the zero point (ASY) is displayed. Pressing displays the value of the slope (SLO). Continuing with threepoint calibration ba75791e05 05/2011 9 Thoroughly rinse the combination electrode with distilled water. 10 If necessary, set the temperature of the third buffer solution with <▲><▼>. 11 Immerse the combination electrode in the third buffer solution. 23 Operation pH 3110 12 Press the key. The AR display indicator flashes. The electrode voltage (mV) or the nominal value of the buffer (setting: see section 4.5.2) is displayed. The asymmetry is displayed as soon as a stable value is recognized. 13 Press the key. The value of the slope (mV/pH) appears on the display. 14 To return to the measuring mode: Press the key. Note While the zero point (ASY) is being displayed, you can change the unit of the zero point with <▲><▼>. While the slope (SLO) is being displayed, you can change the unit of the slope with <▲><▼>. The % display refers to the Nernst slope of 59.2 mV/pH at 25° C (100 x determined slope/Nernst slope). The unit of zero point and slope can also be changed permanently in the measurement settings (see section 4.5.2). 4.3.3 Single-point calibration Two-point calibration Conventional calibration (ConCal) Use any buffer solution for this rapid method. The calibration will be the more exact the nearer the pH value of the buffer solution is to that of the test sample. Use two buffer solutions for this procedure: pH 7.0 ± 0.5 any other buffer solution Note The steps 2 and 8 are not necessary if you use a pH combination electrode with a temperature sensor. 1 24 Press the key repeatedly until the ASY display indicator and the ConCal function indicator appear. ba75791e05 05/2011 pH 3110 Operation ASY 248 TP °C ConCal 2 If necessary, set the temperature of the first buffer solution with <▲><▼>. 3 Immerse the pH combination electrode in the first buffer solution (pH 7.0 ± 0.5 for two-point calibration). 4 Press the key. The measured pH value appears on the display. 5 When the measured value is stable, using <▲><▼>, set the measured value to the nominal pH value of the buffer solution (at the current temperature). 6 Press the key. SLO appears on the display. Note Here you can cancel the calibration procedure with . This corresponds to a single-point calibration. The value of the zero point (ASY) is displayed. The sensor symbol shows the evaluation of the single-point calibration. Pressing displays the value of the slope (SLO). sL O 248 TP °C ConCal ba75791e05 05/2011 25 Operation pH 3110 7 To continue the two-point calibration, thoroughly rinse the combination electrode with deionized water. 8 If necessary, set the temperature of the second buffer solution with <▲><▼>. 9 Immerse the combination electrode in the second buffer solution. 10 Press the key. The second pH value appears on the display. 11 When the measured value is stable, using <▲><▼>, set the measured value to the nominal pH value of the buffer solution (at the current temperature). 12 Press the key. The value of the zero point (ASY) is displayed. 13 Press the key. The value of the slope (SLO) is displayed. 14 Press the key. The measuring mode is active. Note While the zero point (ASY) is being displayed, you can change the unit of the zero point with <▲><▼>. While the slope (SLO) is being displayed, you can change the unit of the slope with <▲><▼>. The % display refers to the Nernst slope of 59.2 mV/pH at 25° C (100 x determined slope/Nernst slope). The units of the zero point and slope can also be changed permanently in the measurement settings (see section 4.5.2). 26 ba75791e05 05/2011 pH 3110 Operation 4.4 Downloading calibration data You can download calibration data to the display as follows: 1 Pressing displays the calibration data (asymmetry). 02 mV ASy ConCal Note While the calibration data is displayed you can press <▲><▼> to switch over the unit of the zero point (ASY). 2 With display the slope. - 592 mV/pH SL O ConCal Note While the calibration data is displayed you can press <▲><▼> to switch over the unit of the slope (SLO). ba75791e05 05/2011 27 Operation pH 3110 4.5 Settings You can adapt the meter to your individual requirements. The settings are done in the following menus: System settings ( ) – Switch-off interval (t.Off) Measurement settings ( ) – Display of the buffer during calibration (pH nominal value or measured voltage value in mV) – Unit of the value for the slope (mV/pH or %) – Unit of the value for the zero point (mV, pH) – Measured value resolution, pH (0.000 / 0.00 / 0.0) – Measured value resolution, U (0.0 / 0) – Temperature unit (°C / °F) – Calibration interval (Int.C [0 ... 999]) Note You can exit the setting menu at any time by pressing . Settings already modified and confirmed with are stored. 28 ba75791e05 05/2011 pH 3110 Operation 4.5.1 System settings The default setting is printed in bold. Switch-off interval (t.Off) 1 10, 20, 30, 40, 50 min, 1, 2, 3, 4, 5, 10, 15, 20, 24 h Open the menu for system settings with . Switch-off interval (t.Off) Time ba75791e05 05/2011 t Of f 010 0 2 Set the switch-off interval with <▲><▼>. 3 Confirm with . The system settings are completed. The meter switches to the measuring mode. 29 Operation pH 3110 4.5.2 Measurement settings These settings apply to calibration and measurement (the default setting is printed in bold). Display during calibration (BUFFER) pH (buffer nominal value), U (combination electrode voltage) Unit of the value for the slope (SLO) mV/pH, % Unit of the value for the zero point (ASY) mV, pH Measured value resolution, pH 0.000 , 0.00 , 0.0 Measured value resolution, U 0.0 , 0 Temperature unit (UnIt) °C, °F Calibration interval (Int.C) 0 ... 7 ... 999 d 1 Display during calibration (bUFF) _ With open the menu for measurement and calibration settings. The first setting is displayed. pH b UF F AutoCal DIN pH AutoCal TEC 30 2 Using <▲><▼>, select the pH or U display during calibration. 3 Confirm with . SLO, the unit of the value for the slope (mV/pH or %) is displayed. ba75791e05 05/2011 pH 3110 Operation Unit of the value for the slope (SLO) pH SL O 4 Using <▲><▼>, select the unit for the slope. 5 Confirm with . ASY, the unit of the value for the zero point (mV or pH) is displayed. Unit of the value for the zero point (ASY) ba75791e05 05/2011 mV/pH pH ASY mV 6 Using <▲><▼>, select the unit for the zero point. 7 Confirm with . res, the resolution of the pH display (0.0, 0.00 or 0.000) appears on the display. 31 Operation pH 3110 pH resolution (res) pH res o o oo 8 Select the resolution for the pH display with <▲><▼>. 9 Confirm with . res, the resolution of the voltage display (0.0, 0.00 or 0.000) appears on the display. U resolution (res) U res mV oo 32 10 Using <▲><▼>, select the unit for the voltage display. 11 Confirm with . Unit, the setting of the unit of the temperature value is displayed. ba75791e05 05/2011 pH 3110 Operation Temperature unit (Unit) Unit oC TP °C 12 Using <▲><▼>, toggle between °C and °F. 13 Confirm with . Int.C, the setting of the calibration interval is displayed. Calibration interval (Int.C) Time INT C d AutoCal DIN 7 AutoCal TEC ba75791e05 05/2011 14 Set the interval with <▲><▼>. 15 Confirm with . The measurement settings are completed. The meter switches to the measuring mode. 33 Operation pH 3110 4.6 Reset 4.6.1 Resetting calibration values This function resets the calibration values to the default condition. All other meter settings are retained. Calibration values in the default condition Zero point 0 mV (pH 7.000) Slope -59.16 mV/pH (100 %) Note The measuring system is not calibrated after a reset. Before measuring, recalibrate the meter. Resetting calibration values 1 _ Press to open the menu for the reset of the calibration data. Init.C is displayed. pH I ni t C no 34 2 Press <▲><▼> to display no or YES. YES: Reset the calibration values. no: Retain the calibration values. 3 Confirm with . The menu is finished. The meter switches to the measuring mode. ba75791e05 05/2011 pH 3110 Operation 4.6.2 Resetting all meter settings This function resets all meter settings to the default condition. The relevant values are given in the following sections: Resetting the meter settings System settings section 4.5.1 Measurement settings section 4.5.2 1 Switch on the meter with . The display test appears briefly on the display. 2 During the display test, press to open the menu for the reset of the meter settings. Init is displayed. Init no 3 Press <▲><▼> to display no or YES. YES: Reset the meter settings. no: Retain the meter settings. 4 Confirm with . The menu is finished. The meter switches to the measuring mode. Note The measuring system is not calibrated after a reset. Before measuring, recalibrate the meter. ba75791e05 05/2011 35 Operation 36 pH 3110 ba75791e05 05/2011 pH 3110 Maintenance, cleaning, disposal 5 Maintenance, cleaning, disposal 5.1 Maintenance The only maintenance activity required is replacing the batteries. Note See the relevant operating manuals of the combination electrodes for instructions on maintenance. 5.1.1 Replacing the batteries 1 Unscrew the two screws (1) on the underside of the meter. 2 Open the battery compartment (2) on the underside of the meter. 1 2 3 Remove the four batteries from the battery compartment. 4 Place four new batteries (type Mignon AA) in the battery compartment. Note Alternatively, you can also use Ni-MH rechargeable batteries (type Mignon AA). In order to charge the batteries, an external charging device is required. Caution Make sure that the poles of the batteries are positioned correctly. The ± signs on the batteries must correspond to the ± signs in the battery compartment. ba75791e05 05/2011 37 Maintenance, cleaning, disposal pH 3110 5 5.2 Close the battery compartment (2) and tighten the screws (1). Cleaning Occasionally wipe the outside of the meter with a damp, lint-free cloth. Disinfect the housing with isopropanol as required. Caution The housing is made of synthetic material (ABS). Thus, avoid contact with acetone or similar detergents that contain solvents. Remove any splashes immediately. 5.3 Packing This meter is sent out in a protective transport packing. We recommend: Keep the packing material. The original packing protects the meter against damage during transport. 5.4 Disposal Note This meter contains batteries. Batteries that have been removed must only be disposed of at a recycling facility set up for this purpose or via the retail outlet. It is illegal to dispose of them in household refuse. 38 ba75791e05 05/2011 pH 3110 What to do if... 6 Error message OFL, UFL Error message CalError What to do if... Cause Remedy pH combination electrode: – Measured value outside the measuring range – Use suitable combination electrode – Air bubble in front of the diaphragm – Remove air bubble – Air in the diaphragm – Extract air or moisten diaphragm – Cable broken – Replace combination electrode – Gel electrolyte dried out – Replace combination electrode Cause Remedy pH combination electrode: – The values determined for zero point and slope of the combination electrode are outside the allowed limits. – Recalibrate – Diaphragm contaminated – Clean diaphragm – Combination electrode broken – Replace combination electrode Buffer solutions ba75791e05 05/2011 – Incorrect buffer solutions – Change calibration procedure – Buffer solutions too old – Use only once. Note the shelf life – Buffer solutions depleted – Change solutions 39 What to do if... pH 3110 No stable measured value Cause Remedy pH combination electrode: – Diaphragm contaminated – Clean diaphragm – Membrane contaminated – Clean membrane Test sample: – pH value not stable – Measure with air excluded if necessary – Temperature not stable – Adjust temperature if necessary Combination electrode + test sample: Sensor symbol flashes – Conductivity too low – Use suitable combination electrode – Temperature too high – Use suitable combination electrode – Organic liquids – Use suitable combination electrode Cause – Calibration interval expired Display, LoBat 40 Cause – Batteries almost empty Remedy – Recalibrate the measuring system Remedy – Replace the batteries (see section 5.1 MAINTENANCE) ba75791e05 05/2011 pH 3110 What to do if... Obviously incorrect measured values Cause pH combination electrode: – pH combination electrode unsuitable – Use suitable combination electrode – Temperature difference between buffer and test sample too high – Adjust temperature of buffer or sample solutions – Measurement procedure not suitable – Follow special procedure Meter does not react to keystroke Cause You want to know which software version is in the meter Cause ba75791e05 05/2011 Remedy – Operating condition undefined or EMC load unallowed – E. g., a question by the service department Remedy – Processor reset: Press the and key simultaneously Remedy – Switch on the meter. During the display test, display the software version with . 41 What to do if... 42 pH 3110 ba75791e05 05/2011 pH 3110 Technical data Dimensions Weight Mechanical structure 7 Technical data 7.1 General data approx. 180 x 80 x 55 mm approx. 0.4 kg Type of protection IP 67 Electrical safety Protective class III Test certificates CE, cETLus Ambient conditions Power supply Sensor input Guidelines and norms used ba75791e05 05/2011 Storage - 25 °C ... + 65 °C Operation -10 °C ... + 55 °C Allowable relative humidity Annual mean: < 75 % 30 days/year: 95 % Other days: 85 % Batteries 4 x 1.5 V alkali-manganese batteries, type AA Rechargeable batteries 4 x 1,2 V NiMH rechargeable batteries, type AA (no charging function) Operational life Approx. 2500 h operating hours (batteries) Input resistance > 5 * 1012 ohm Input current < 1 * 10-12 A EMC EC directive 2004/108/EC EN 61326-1 EN 61000-3-2 EN 61000-3-3 FCC Class A Meter safety EC directive 2006/95/EC EN 61010-1 IP protection EN 60529 43 Technical data pH 3110 7.2 Measuring ranges, resolution Measuring ranges, resolution, accuracy Variable Measuring range Resolution pH - 2.0 ... + 20.0 0.1 - 2.00 ... + 20.00 0.01 - 2.000 ... + 19.999 0.001 - 1200.0 ... + 1200.0 0.1 - 2000 ... + 2000 1 T [°C] - 5.0 ... + 105.0 0.1 T [°F] 23.0 ... + 221.0 0.1 Variable Range Increment Tmanual [°C] - 25 ... + 130 1 Tmanual [°F] -13 ... + 266 1 Variable Accuracy U [mV] Manual temperature input Accuracy (± 1 digit) Temperature of the test sample pH / range * - 2.0 ... + 20.0 ± 0.1 + 15 °C ... + 35 °C - 2.00 ... + 20.00 ± 0.01 + 15 °C ... + 35 °C - 2.000 ... + 19.999 ± 0.005 + 15 °C ... + 35 °C - 2000 ... + 2000 ±1 + 15 °C ... + 35 °C -1200.0 ... +1200.0 ± 0.3 + 15 °C ... + 35 °C U [mV] / range T [°C] / temperature sensor NTC 30 ± 0.1 PT 1000 ± 0.1 * when measuring in a range of ± 2 pH around a calibration point Note The accuracy values specified here apply exclusively to the meter. The accuracy of the combination electrodes and buffer solutions has to be taken into account additionally. 44 ba75791e05 05/2011 pH 3110 Lists 8 Lists This chapter provides additional information and orientation aids. Specialist terms Index The glossary briefly explains the meaning of the specialist terms. However, terms that should already be familiar to the target group are not described here. The index helps you to find the topics that you are looking for. Glossary Adjusting To manipulate a measuring system so that the relevant value (e. g. the displayed value) differs as little as possible from the correct value or a value that is regarded as correct, or that the difference remains within the tolerance. Asymmetry see zero point AutoRange Name of the automatic selection of the measuring range. AutoRead Calibration Function to control the measured value stability. Comparing the value from a measuring system (e. g. the displayed value) to the correct value or a value that is regarded as correct. Often, this expression is also used when the measuring system is adjusted at the same time (see adjusting). Electromotive force of a combination electrode The electromotive force U of the combination electrode is the measurable electromotive force of an electrode in a solution. It equals the sum of all the galvanic voltages of the electrode. Its dependency on the pH results in the electrode function, which is characterized by the parameters, slope and zero point. Junction The junction is a porous body in the housing wall of reference electrodes or electrolyte bridges. It arranges the electrical contact between two solutions and makes the electrolyte exchange more difficult. The expression, junction, is also used for ground or junctionless transitions. Measured parameter ba75791e05 05/2011 The measured parameter is the physical dimension determined by measuring, e. g. pH, conductivity or D.O. concentration. 45 Lists pH 3110 Measured value Molality Molality is the quantity (in Mol) of a dissolved substance in 1000 g solvent. ORP voltage The ORP is caused by oxidizing or reducing substances dissolved in water, if these substances become effective at an electrode surface (e. g. a gold or platinum surface). pH value The pH is a measure of the acidic or basic effect of an aqueous solution. It corresponds to the negative decadic logarithm of the molal hydrogen ions activity divided by the unit of the molality. The practical pH value is the value of a pH measurement. Potentiometry Name of a measuring technique. The signal (depending on the measured parameter) of the electrode is the electrical potential. The electrical current remains constant. Reset Resolution Slope Standard solution Test sample Zero point 46 The measured value is the special value of a measured parameter to be determined. It is given as a combination of the numerical value and unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K). Restoring the original condition of all settings of a measuring system. Smallest difference between two measured values that can be displayed by a measuring instrument. The slope of a linear calibration function. The standard solution is a solution where the measured value is known by definition. It is used to calibrate a measuring system. Designation of the test sample ready to be measured. Normally, a test sample is made by processing the original sample. The test sample and original sample are identical if the test sample was not processed. The zero point of a pH electrode is the pH value at which the electromotive force of the pH electrode at a specified temperature is zero. Normally, this is at 25 °C. ba75791e05 05/2011 pH 3110 Lists Index A P Authorized use . . . . . . . . . . . . . . . . . . . . . . 10 AutoCal . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 AutoRead . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Precautions . . . . . . . . . . . . . . . . . . . . . . . . . 9 B Battery compartment . . . . . . . . . . . . . . 13, 37 R Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 All meter settings . . . . . . . . . . . . . . . . . 35 Calibration values . . . . . . . . . . . . . . . . 34 Resolution setting . . . . . . . . . . . . . 30, 31, 32 C Calibration . . . . . . . . . . . . . . . . . . . . . . . . . 19 Calibration evaluation pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Calibration interval . . . . . . . . . . . . . . . . . . . 21 Calibration points pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 ConCal . . . . . . . . . . . . . . . . . . . . . . . . . 19, 24 S Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Scope of delivery . . . . . . . . . . . . . . . . . . . . 13 Single-point calibration . . . . . . . . . 19, 23, 25 Slope pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Socket field . . . . . . . . . . . . . . . . . . . . . . . . . 8 Stability control . . . . . . . . . . . . . . . . . . . . . 16 D Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 T E Two-point calibration . . . . . . . . . . . . . . . . . 19 ConCal . . . . . . . . . . . . . . . . . . . . . . . . . 24 Energy saving feature . . . . . . . . . . . . . . . . 14 I Z Zero point of pH combination electrode . . . 19 Initial commissioning . . . . . . . . . . . . . . . . . 14 Interval Calibration . . . . . . . . . . . . . . . . . . . . . . 21 K Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 L LoBat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 M Measurement accuracy . . . . . . . . . . . . . . . 21 O Operational safety . . . . . . . . . . . . . . . . . . . 10 ORP electrode . . . . . . . . . . . . . . . . . . . . . . 18 ORP voltage . . . . . . . . . . . . . . . . . . . . . . . . 18 ba75791e05 05/2011 47 Lists 48 pH 3110 ba75791e05 05/2011 Wissenschaftlich-Technische Werkstätten GmbH Dr.-Karl-Slevogt-Straße 1 D-82362 Weilheim Germany Tel: Fax: E-Mail: Internet: +49 (0) 881 183-0 +49 (0) 881 183-100 +49 (0) 881 183-420 Info@WTW.com http://www.WTW.com
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.6 Linearized : Yes Author : direim Create Date : 2011:05:20 09:07:47Z Modify Date : 2015:09:17 13:48:22-06:00 Has XFA : No XMP Toolkit : Adobe XMP Core 5.6-c015 81.157285, 2014/12/12-00:43:15 Format : application/pdf Title : WTW pH 3110 Operating Manual Creator : direim Creator Tool : FrameMaker 10.0 Metadata Date : 2015:09:17 13:48:22-06:00 Producer : Acrobat Elements 9.0.0 (Windows) Document ID : uuid:8d6d5442-926b-4db1-b644-aae6851137f3 Instance ID : uuid:0c954173-4f50-6844-88e8-959da1fdb808 Page Mode : UseOutlines Page Count : 50EXIF Metadata provided by EXIF.tools