Zimmer® Herbert™ Cannulated Bone Screw Zimmer Herbert Surgical Technique

2016-04-01

: Pdf Zimmer Herbert Cannulated Screw Surgical Technique Zimmer_Herbert_Cannulated_Screw_Surgical_Technique 3 2016 pdf

Open the PDF directly: View PDF PDF.
Page Count: 4

DownloadZimmer® Herbert™ Cannulated Bone Screw  Zimmer Herbert Surgical Technique
Open PDF In BrowserView PDF
Zimmer
Zimmer®®
Herbert
Herbert™™
Cannulated
Cannulated
Bone
BoneScrew
Screw

An innovative option for managing the reduction and fixation
of difficult fractures

Specifically Designed for Problem Internal Fixation
The headless design of the Herbert Screw means that the screw is completely
embedded in the bone, without any protrusions to cause tissue irritation even in
intra-articular placement. • Cannulation helps ensure precise placement of the
screw. • Herbert Cannulated Guide Pins hold the fragment and act as guides for
drilling, tapping, and screw placement. • The Tivanium® Ti-6A1-4V Alloy used in
fabricating these screws is highly biocompatible and corrosion resistant. • The
thread design of the Herbert Screw creates compression and provides fracture
stability. As the proximal threads engage the bone, the fracture is drawn together,
helping to create and maintain stability of the fracture site.
INDICATIONS FOR USE
The Herbert Cannulated Bone Screw System
(4.5mm and 6.5mm diameter) are indicated
for fracture fixation, reconstruction, osteotomy,
and arthrodesis of various bones and bone
fragments including joint fusions (arthrodeses)
in the foot and fixation of intra-articular
fractures of the humerus, femur and tibia.

4.5mm Herbert Cannulated
Bone Screw

6.5mm Herbert Cannulated
Bone Screw

The Herbert Cannulated Bone Screw is
designed for the management of fractures
where minimal tissue coverage makes
standard screw use inappropriate or where
extreme precision in fragment alignment is
imperative. Such locations might include
fractures of the:
• Olecranon and malleolar region
• Distal femur
• Proximal tibia
• Distal tibia

1

2

4

5

Guide Pin

After initial reduction of the fracture
fragments is obtained, the
specifically designed Herbert
Cannulated Guide Pin is placed
through the fragments to act as a
Guide Wire for the rest of the
placement operation. The Guide Pin
should not perforate the opposite
cortex.

Distal Drill Bit

Insert the Distal Drill Bit over the
Guide Pin and drill to the desired
depth. At this time the surgeon may
choose to drill 5mm less than the
pilot length.
Option: The Distal Drill Bit may be
inserted into a Jacob’s Chuck to
the appropriate calibration, thus
allowing the surgeon to utilize the
Chuck as an automatic stop.

Depth Gauge

Use the Depth Gauge to measure the
length of the Guide Pin in the bone.
If the Guide perforates the opposite
cortex, corresponding compensation
must be made when selecting the
implant length.

Cannulated
Tap

Insert the Cannulated Tap over
the Guide Pin and tap the channel
to prepare for the leading screw
threads of the implant. Tap depth
should be equal to the depth
created by the Distal Drill Bit.

3

Proximal
Drill Bit

Insert the Proximal Drill Bit over the
Guide Pin to drill the cortex. The Drill
Bit should be advanced until the
built-in stop contacts the cortex of
the bone.

6

Insert Screw

In determining screw length, the
surgeon may wish to choose an
implant 5mm to 10mm shorter than
the measured pilot length. Using the
Cannulated Screwdriver, insert the
Herbert Cannulated Screw. As the
trailing threads engage the bone,
reduction is achieved. These
trailing threads should be seated
approximately 1mm below the
cortex of the bone to ensure that
there is no intra-articular protrusion
or interference with the joint
function.

Order Information
Description 	
Herbert Cannulated Bone Screw Set
(Includes the following implants,
instruments and case components)

Qty. in Set

Sets
00-1155-093-00 	
		

Herbert Cannulated Bone Screw Instrumentation Set
(Includes all instrumentation listed)

00-1155-094-00 	
		

Herbert Cannulated 4.5mm Bone Screw instrumentation used for the 	
4.5mm bone screw (Denoted with a • )
Herbert Cannulated Bone Screw Implant Set
(Includes two each of every size bone screw)

Prod. No. 	
00-1155-001-00	
	
	
4.5mm Implants
47-1155-025-05 	
47-1155-030-05 	
47-1155-035-05 	
47-1155-040-05 	
47-1155-045-05 	
47-1155-050-05 	
47-1155-055-05 	
47-1155-060-05 	
47-1155-065-05 	
47-1155-070-05 	
47-1155-075-05 	
47-1155-080-05 	
47-1155-085-05 	
47-1155-090-05 	
47-1155-095-05 	
47-1155-100-05 	

25mm 	
30mm 	
35mm 	
40mm 	
45mm 	
50mm 	
55mm 	
60mm 	
65mm 	
70mm 	
75mm 	
80mm 	
85mm 	
90mm 	
95mm 	
100mm 	

2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2

00-1155-099-00 	
		

6.5mm Implants
47-1155-025-07 	
47-1155-030-07 	
47-1155-035-07 	
47-1155-040-07 	
47-1155-045-07 	
47-1155-050-07 	
47-1155-055-07 	
47-1155-060-07 	
47-1155-065-07 	
47-1155-070-07 	
47-1155-075-07 	
47-1155-080-07 	
47-1155-085-07	
47-1155-090-07 	
47-1155-095-07 	
47-1155-100-07 	

Length 	
	
	

Leading 	
Thread 	
Length 	

Trailing 	
Thread 	
Length 	

Pitch 	
Flush 	
Flush 1 Turn
Difference 	 Reduction 	 Reduction
Potential 	 Potential

25mm 	
30mm 	
35mm 	
40mm 	
45mm 	
50mm 	
55mm 	
60mm 	
65mm 	
70mm 	
75mm 	
80mm 	
85mm 	
90mm 	
95mm 	
100mm	

2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2

25-35mm 	

9.0mm 	

5.8mm 	

.41mm 	

1.23mm 	

1.64mm

40-70mm 	

11.9mm 	

6.4mm 	

.41mm 	

1.42mm 	

1.85mm

75-100mm 	

16.0mm 	

6.4mm 	

.41mm 	

1.42mm 	

1.85mm

Instrumentation
• 00-1155-011-00 	
• 00-1155-012-00 	
• 00-1155-013-00 	
• 00-1155-014-00 	
• 00-1155-015-00 	
• 00-1155-016-00 	
00-1155-021-00 	
00-1155-022-00 	
00-1155-023-00	
00-1155-024-00 	
00-1155-026-00	
00-1155-027-00	
• 00-1180-040-00	
• 00-4808-020-00 	
• 00-1155-085-00	

Cannulated Tap, 4.5mm 	
Cannulated Screwdriver, 4.5mm 	
Trephine, 4.5mm 	
Pilot Drill Bit, 3.8mm 	
Main Drill Bit, 3.1mm 	
Guide Pin, Partially Threaded, 1.6mm 	
Cannulated Tap, 6.5mm 	
Cannulated Screwdriver, 6.5mm 	
Trephine 	
Pilot Drill Bit, 5.5mm 	
Guide Pin, Partially Threaded, 2.0mm 	
Main Drill Bit, 4.6mm 	
Depth Guide 	
Parallel Drill Guide, 2.0mm 	
Sterilization Tray 	

1
1
1
1
1
1
1
1
1
1
1
1
1
1
1

Herbert Cannulated Bone Screws
4.5mm Diameter
Leading Thread Major Diameter 4.5mm
Trailing Thread Major Diameter 5.8mm
Length	
Leading 	
Trailing 	
	
Thread 	
Thread 	
	
Length 	
Length 	

Pitch 	
Flush 	
Flush 1 Turn
Difference 	 Reduction 	 Reduction
Potential† 	 Potential††

25-35mm 	

8.2mm 	

5.8mm 	

.41mm 	

1.23mm 	

1.64mm

40-70mm 	

12.0mm 	

6.4mm	

.41mm 	

1.42mm 	

1.85mm

75-100mm 	

16.0mm 	

6.4mm 	

.41mm 	

1.42mm 	

1.85mm

6.5mm Diameter
Leading Thread Major Diameter 6.5mm
Trailing Thread Major Diameter 8.1mm

† Amount of fracture reduction potential with screw tightened flush with bone surface
†† Amount of fracture reduction potential with screw tightened 1mm below bone surface

Warning: This device is not approved by the U.S. FDA for screw attachment or
fixation to the posterior elements (pedicles) of clavical, thoracic, or lumbar spine.
	

DISCLAIMER:
This documentation is intended exclusively for physicians and
is not intended for laypersons. Information on the products and
procedures contained in this document is of a general nature
and does not represent and does not constitute medical advice
or recommendations. Because this information does not purport
to constitute any diagnostic or therapeutic statement with regard
to any individual medical case, each patient must be examined
and advised individually, and this document does not replace the
need for such examination and/or advise in whole or in part.
Please refer to the package inserts for important product
information, including, but not limited to, indications,
contraindications, warnings, precautions, and adverse effects.

Contact your Zimmer representative or visit us at www.zimmer.com

97-1155-401-00 Rev. 3 7-27-15 ©2015 Zimmer, Inc.



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : Yes
Author                          : Zimmer, Inc.
Create Date                     : 2015:08:06 10:06:47-05:00
Date Released                   : 7-27-15
I D                             : 97-1155-401-00 Rev. 3
ID                              : 97-1155-401-00
Keywords                        : Zimmer® Herbert™ Cannulated Bone Screw, completely embedded, intra-articular placement, Cannulation helps ensure precise placement of the screw, Herbert Cannulated Guide Pins, drilling, tapping, screw placement, Tivanium® Ti-6A1-4V Alloy, highly biocompatible, corrosion resistant, compression and fracture stability, proximal threads engage the bone, maintain stability of the fracture site, Cannulated tap
Modify Date                     : 2015:12:02 09:34:46-05:00
Has XFA                         : No
XMP Toolkit                     : Adobe XMP Core 5.4-c005 78.147326, 2012/08/23-13:03:03
Format                          : application/pdf
Modified                        : 2015:08:18 09:36:12.585-05:00
Subject                         : Zimmer® Herbert™ Cannulated Bone Screw, completely embedded, intra-articular placement, Cannulation helps ensure precise placement of the screw, Herbert Cannulated Guide Pins, drilling, tapping, screw placement, Tivanium® Ti-6A1-4V Alloy, highly biocompatible, corrosion resistant, compression and fracture stability, proximal threads engage the bone, maintain stability of the fracture site, Cannulated tap
Description                     : Zimmer® Herbert™ Cannulated Bone Screw
Creator                         : Zimmer, Inc.
Title                           : Zimmer® Herbert™ Cannulated Bone Screw
Size                            : 657029
Extracted                       : 2015:08:18 09:35:02.194-05:00
Sha 1                           : 30e7a3971b7f81c20c41bc307cffc7c52ff87f5c
Metadata Date                   : 2015:12:02 09:34:46-05:00
Creator Tool                    : Adobe InDesign CC 2015 (Macintosh)
Rendition Class                 : proof:pdf
Document ID                     : xmp.id:d1baf3d6-f2c2-4681-8ab6-f82ec0c53d4c
Original Document ID            : adobe:docid:indd:7a6a3f0d-bcb1-11dd-a59d-eed3222edd58
Instance ID                     : uuid:66e63043-549b-764f-8d50-44f2bb89156e
Derived From Rendition Class    : default
Derived From Instance ID        : xmp.iid:bdd99ee5-469a-4683-8dd9-01c47bbf4999
Derived From Document ID        : adobe:docid:indd:7a6a3f0d-bcb1-11dd-a59d-eed3222edd58
Derived From Original Document ID: adobe:docid:indd:7a6a3f0d-bcb1-11dd-a59d-eed3222edd58
History Action                  : converted
History Software Agent          : Adobe InDesign CC 2015 (Macintosh)
History Parameters              : from application/x-indesign to application/pdf
History When                    : 2015:08:06 10:06:47.000-05:00
History Changed                 : /
Id                              : 97-1155-401-00
Date 0020 Released              : 7-27-15
Producer                        : Adobe PDF Library 15.0
Trapped                         : False
Page Layout                     : TwoPageRight
Page Count                      : 4
EXIF Metadata provided by EXIF.tools

Navigation menu