Zimmer®
Herbert
Cannulated
Bone Screw
An innovative option for managing the reduction and fixation
of difficult fractures
Zimmer®
Herbert
Cannulated
Bone Screw
Specifically Designed for Problem Internal Fixation
The headless design of the Herbert Screw means that the screw is completely
embedded in the bone, without any protrusions to cause tissue irritation even in
intra-articular placement. • Cannulation helps ensure precise placement of the
screw. • Herbert Cannulated Guide Pins hold the fragment and act as guides for
drilling, tapping, and screw placement. • The Tivanium® Ti-6A1-4V Alloy used in
fabricating these screws is highly biocompatible and corrosion resistant. • The
thread design of the Herbert Screw creates compression and provides fracture
stability. As the proximal threads engage the bone, the fracture is drawn together,
helping to create and maintain stability of the fracture site.
The Herbert Cannulated Bone Screw is
designed for the management of fractures
where minimal tissue coverage makes
standard screw use inappropriate or where
extreme precision in fragment alignment is
imperative. Such locations might include
fractures of the:
Olecranon and malleolar region
Distal femur
Proximal tibia
Distal tibia
4.5mm Herbert Cannulated
Bone Screw
6.5mm Herbert Cannulated
Bone Screw
INDICATIONS FOR USE
The Herbert Cannulated Bone Screw System
(4.5mm and 6.5mm diameter) are indicated
for fracture fixation, reconstruction, osteotomy,
and arthrodesis of various bones and bone
fragments including joint fusions (arthrodeses)
in the foot and fixation of intra-articular
fractures of the humerus, femur and tibia.
3
6
In determining screw length, the
surgeon may wish to choose an
implant 5mm to 10mm shorter than
the measured pilot length. Using the
Cannulated Screwdriver, insert the
Herbert Cannulated Screw. As the
trailing threads engage the bone,
reduction is achieved. These
trailing threads should be seated
approximately 1mm below the
cortex of the bone to ensure that
there is no intra-articular protrusion
or interference with the joint
function.
1Guide Pin
After initial reduction of the fracture
fragments is obtained, the
specifically designed Herbert
Cannulated Guide Pin is placed
through the fragments to act as a
Guide Wire for the rest of the
placement operation. The Guide Pin
should not perforate the opposite
cortex.
2
Use the Depth Gauge to measure the
length of the Guide Pin in the bone.
If the Guide perforates the opposite
cortex, corresponding compensation
must be made when selecting the
implant length.
Insert the Proximal Drill Bit over the
Guide Pin to drill the cortex. The Drill
Bit should be advanced until the
built-in stop contacts the cortex of
the bone.
4
Insert the Distal Drill Bit over the
Guide Pin and drill to the desired
depth. At this time the surgeon may
choose to drill 5mm less than the
pilot length.
Option: The Distal Drill Bit may be
inserted into a Jacob’s Chuck to
the appropriate calibration, thus
allowing the surgeon to utilize the
Chuck as an automatic stop.
5
Insert the Cannulated Tap over
the Guide Pin and tap the channel
to prepare for the leading screw
threads of the implant. Tap depth
should be equal to the depth
created by the Distal Drill Bit.
Proximal
Drill Bit
Depth Gauge
Cannulated
Tap
Distal Drill Bit Insert Screw
Contact your Zimmer representative or visit us at www.zimmer.com
Order Information
Prod. No. Description Qty. in Set
00-1155-001-00 Herbert Cannulated Bone Screw Set
(Includes the following implants,
instruments and case components)
4.5mm Implants
47-1155-025-05 25mm 2
47-1155-030-05 30mm 2
47-1155-035-05 35mm 2
47-1155-040-05 40mm 2
47-1155-045-05 45mm 2
47-1155-050-05 50mm 2
47-1155-055-05 55mm 2
47-1155-060-05 60mm 2
47-1155-065-05 65mm 2
47-1155-070-05 70mm 2
47-1155-075-05 75mm 2
47-1155-080-05 80mm 2
47-1155-085-05 85mm 2
47-1155-090-05 90mm 2
47-1155-095-05 95mm 2
47-1155-100-05 100mm 2
6.5mm Implants
47-1155-025-07 25mm 2
47-1155-030-07 30mm 2
47-1155-035-07 35mm 2
47-1155-040-07 40mm 2
47-1155-045-07 45mm 2
47-1155-050-07 50mm 2
47-1155-055-07 55mm 2
47-1155-060-07 60mm 2
47-1155-065-07 65mm 2
47-1155-070-07 70mm 2
47-1155-075-07 75mm 2
47-1155-080-07 80mm 2
47-1155-085-07 85mm 2
47-1155-090-07 90mm 2
47-1155-095-07 95mm 2
47-1155-100-07 100mm 2
Instrumentation
00-1155-011-00 Cannulated Tap, 4.5mm 1
00-1155-012-00 Cannulated Screwdriver, 4.5mm 1
00-1155-013-00 Trephine, 4.5mm 1
00-1155-014-00 Pilot Drill Bit, 3.8mm 1
00-1155-015-00 Main Drill Bit, 3.1mm 1
00-1155-016-00 Guide Pin, Partially Threaded, 1.6mm 1
00-1155-021-00 Cannulated Tap, 6.5mm 1
00-1155-022-00 Cannulated Screwdriver, 6.5mm 1
00-1155-023-00 Trephine 1
00-1155-024-00 Pilot Drill Bit, 5.5mm 1
00-1155-026-00 Guide Pin, Partially Threaded, 2.0mm 1
00-1155-027-00 Main Drill Bit, 4.6mm 1
00-1180-040-00 Depth Guide 1
00-4808-020-00 Parallel Drill Guide, 2.0mm 1
00-1155-085-00 Sterilization Tray 1
97-1155-401-00 Rev. 3 7-27-15 ©2015 Zimmer, Inc.
Sets
00-1155-093-00 Herbert Cannulated Bone Screw Instrumentation Set
(Includes all instrumentation listed)
00-1155-094-00 Herbert Cannulated 4.5mm Bone Screw instrumentation used for the
4.5mm bone screw (Denoted with a )
00-1155-099-00 Herbert Cannulated Bone Screw Implant Set
(Includes two each of every size bone screw)
Herbert Cannulated Bone Screws
4.5mm Diameter
Leading Thread Major Diameter 4.5mm
Trailing Thread Major Diameter 5.8mm
Length Leading Trailing Pitch Flush Flush 1 Turn
Thread Thread Difference Reduction Reduction
Length Length Potential† Potential††
25-35mm 8.2mm 5.8mm .41mm 1.23mm 1.64mm
40-70mm 12.0mm 6.4mm .41mm 1.42mm 1.85mm
75-100mm 16.0mm 6.4mm .41mm 1.42mm 1.85mm
6.5mm Diameter
Leading Thread Major Diameter 6.5mm
Trailing Thread Major Diameter 8.1mm
Length Leading Trailing Pitch Flush Flush 1 Turn
Thread Thread Difference Reduction Reduction
Length Length Potential Potential
25-35mm 9.0mm 5.8mm .41mm 1.23mm 1.64mm
40-70mm 11.9mm 6.4mm .41mm 1.42mm 1.85mm
75-100mm 16.0mm 6.4mm .41mm 1.42mm 1.85mm
† Amount of fracture reduction potential with screw tightened flush with bone surface
†† Amount of fracture reduction potential with screw tightened 1mm below bone surface
Warning: This device is not approved by the U.S. FDA for screw attachment or
fixation to the posterior elements (pedicles) of clavical, thoracic, or lumbar spine.
DISCLAIMER:
This documentation is intended exclusively for physicians and
is not intended for laypersons. Information on the products and
procedures contained in this document is of a general nature
and does not represent and does not constitute medical advice
or recommendations. Because this information does not purport
to constitute any diagnostic or therapeutic statement with regard
to any individual medical case, each patient must be examined
and advised individually, and this document does not replace the
need for such examination and/or advise in whole or in part.
Please refer to the package inserts for important product
information, including, but not limited to, indications,
contraindications, warnings, precautions, and adverse effects.

Navigation menu