Edwards Diffuision Pumps Diffstak

User Manual: Pdf diffstak

Open the PDF directly: View PDF PDF.
Page Count: 16

DownloadEdwards Diffuision Pumps Diffstak
Open PDF In BrowserView PDF
Scientific vapor
pumps
3
PAGE
1

Vapor pumps for scientific
instruments and R&D applications
Diffusion pumps – technology and applications
Diffstak
Diffstak installation
EO50/60
Diffstak combined outfits
Vapor pump accessories

Shop online at www.bocedwards.com

3-2
3-4
3-12
3-14
3-15
3-15

VAPOR PUMPS FOR SCIENTIFIC
INSTRUMENTS AND R&D APPLICATIONS
Diffusion
Vapor
pumps
pumps
for–scientific
technology
instruments
and applications
and R&D applications

Scientific and R & D applications require special vapor pumps and accessories.
It is important to minimize any backstreaming of the vapor pump fluid, and the
number of elastomer seals used in system design needs to be kept to a
minimum, to give clean pumping with minimal outgassing. For bench-top or
transportable instruments, compact air-cooled pumps are essential.
BOC Edwards offers a range of vapor pumps and accessories which are
designed to meet these needs.

Measurement methods

3
PAGE
2

The methods used to measure pressure and flow have become more
accurate in recent years allowing much tighter controls over system
conditions. The speeds and throughputs quoted in the catalog for the
diffusion pumps are based on actual pump data derived from measurements
made with the latest technology total pressure gauges and mass flow
transducers in accordance with ISO standards.
Some confusion could arise with previously published speed and throughput
figures for older designed pumps where, historically, pressure
measurements were made with partial pressure gauges like the McLeod
gauge. This older gauge can indicate speeds up to 30% higher than that
expected using state of the art total pressure gauges.
Further confusion could arise from the measurement standards chosen to
determine pumping performance. In the case of AVS (American Vacuum
Society), this can indicate speeds and throughput up to 15% higher than
ISO figures.
Historical industry practice meant that in considering the above, and
potential gauge accuracy of ± 15%, it was possible to have speeds quoted
60% higher than might be expected using modern total pressure
measurement equipment. This should be carefully considered when
comparing specifications for older derived data published for similarly sized
competitor pumps and BOC Edwards diffusion pumps.

Diffstak vapor diffusion pumps
The compact water-cooled Diffstak pumps with an integral cooled baffle
offer exceptionally clean pumping with very low backstreaming, reduced
outgassing, and a reduction in the number of elastomer seals required for
installation.
The Diffstak design has been proven over many years with thousands of
pumps installed. They are supplied in three types: standard, cryo-cooled
and unvalved. (C – collar model pumps).
The standard Diffstak pumps have integral high vacuum valves and watercooled baffles. The cryo-cooled versions (CR – model pumps) have a liquid
nitrogen cooled baffle supplied from a reservoir attached to the pump and
are intended for pumping heavy vapor loads. Valves on both versions are
supplied as either manually operated (M-model pumps) or pneumatically
operated (P-model pumps). When comparing pumping speeds, note that
the speeds quoted for valved Diffstaks are the speeds above the high
vacuum valve, taking full account of the valve's impedance.
The unvalved Diffstaks are for systems requiring the highest possible
ultimate vacuum or for those which do not need a high vacuum valve. All
sizes are available with ISO flanges while two sizes are also available with
CF flanges. (F – ConFlat® model pump).
The complete range is shown in the table below. Refer to the following
pages for full technical data for each of the pumps and also for full details
of installations, spares and accessories.
STANDARD
DIFFSTAK

63/150M or P
100/300M or P
160/700M or P
250/2000M or P

CRYO-COOLED
DIFFSTAK

CR63/150M or P
CR100/300M or P
CR160/700M or P
CR250/2000M or P

UNVALVED DIFFSTAK
ISO FLANGE CF FLANGE

63/150C
100/300C
160/700C
250/2000C

–
100/300F
160/700F
–

Fast cycling on unvalved systems
For fast cycling pumping systems we recommend our TV range of water
or air-cooled pumps. Primarily designed for the evacuation of cathode ray
tubes, they will also suit other fast cycling pumping applications.

Ultimate vacuum
The ultimate vacuum of a vapor pump is the lowest pressure achieved in
tests on the pump, measured above the inlet of the pump (or above the
high vacuum valve for the Diffstaks). The ultimate vacuum depends on: the
type of fluid used in the pump; the temperature of the inlet baffle; the
amount of outgassing from the vacuum system; and the amount of leakage
into the system.

Critical backing pressure
The critical backing pressure is the highest pressure that a pump can
tolerate in the backing line. If the pressure is higher than the critical backing
pressure, the pump may stall. The critical backing pressure depends on: the
pump design; the power of the heaters; and the fluid used in the vapor
pump.

Backstreaming
Backstreaming is the direct movement of molecules of pump fluid vapor
from the pump toward the vacuum system. All BOC Edwards pumps are
specially designed to minimize backstreaming; the pumps have a guard ring
(sometimes called a cool-cap) fitted above the top jet. The guard ring
condenses vapor molecules moving from the top jet toward the vacuum
system.
In some sensitive applications, backstreaming may be very undesirable; you
can minimize backstreaming if you fit a baffle or a trap to the top of the
pump.

Accessories
For vapor pump accessories see page 3-15.

APPLICATIONS
•
•
•
•
•
•
•
•
•

Mass spectrometry
Gas analysis
Leak detection
Thin film coating
High vacuum systems
Lamp evacuation
Surface physics
Small vacuum ovens
Vacuum isolation/cryo transfer

EO50/60 diffusion pump
The EO50/60 is a compact, fast warm-up, air-cooled diffusion pump and
makes a good choice for small, low-cost systems. The high helium pumping
speed from a pump only 176 mm high is a particular benefit for GCMS
applications.
For EO2, EO4, EO6 diffusion pumps contact BOC Edwards.

ConFlat® is a registered trademark of Varian, Inc

Shop online at www.bocedwards.com

DATA SUMMARY VAPOR PUMPS
PUMP
Catalog page
Pumping speed
Nitrogen
Hydrogen
Hydrogen (baffled)
Minimum backing pump displacement
Inlet connection
Backing connection

1

63/150M
63/150P
3-4

63/150C

l s-1

135

150

ls

-1

200

225

ls

-1

–

–

3 -1

5

5

ISO63
NW10

ISO63
NW10

100/300M
100/300P
3-6

100/300C
100/300F
3-7

m h

PUMP

l s-1

280

300

Hydrogen

l s-1

500

535

Hydrogen (baffled)

l s-1

–

–

m3h-1

5

5

ISO100
NW25

ISO100 / 6 in
NW25

160/700M
160/700P
3-8

160/700C
160/700F
3-9

PUMP
Catalog page
Pumping speed
Nitrogen

l s-1

700

760

Hydrogen

l s-1

1300

1410

Hydrogen (baffled)
Minimum backing pump displacement
Inlet connection

1

l s-1

–

–

m3h-1

12

12

ISO160

ISO160 / 8 in

250/2000M
250/2000P
3-10

250/2000C

PUMP
Catalog page
Pumping speed
Nitrogen

l s-1

2000

2130

Hydrogen

l s-1

3000

3200

Hydrogen (baffled)

l s-1

–

–

m3h-1

40

40

ISO250
NW40

ISO250
NW40

Minimum backing pump displacement1
Inlet connection
Backing connection

PUMP
Catalog page
Pumping speed
Air

l s-1

60

Air (baffled)

l s-1

27

Nitrogen

l s-1

–

Hydrogen

l s-1

70

l s-1

50

EO50/60
3-14

Hydrogen (baffled)
Minimum backing pump displacement
Inlet connection
Backing connection

3-11

1

3 -1

m h

1
NW50 flange
NW10 flange

For maximum throughput. (63/150M and 63/150P – For applications where maximum throughput is not required, use an RV3.)

Shop online at www.bocedwards.com

3
PAGE
3

SCIENTIFIC VAPOR PUMPS

Catalog page
Pumping speed
Nitrogen

Minimum backing pump displacement1
Inlet connection
Backing connection

1

3-5

(3.70)

70

50

(2.75)

85

(1.97)

65

360 (14.17)

4 x Ø9 (0.35)
x Ø110 (4.33)

(2.56) (3.35)
131
(5.16)
216 (8.50)

17
105

94

(0.67)

213 (8.38)
Ø130 (5.12)

(4.13)

STANDARD DIFFSTAK 63/150M AND
63/150P

Ø76
(2.99)
125 (4.92)

Diffstak 63/150M

TECHNICAL DATA

85

(2.56) (3.35)
220 (8.66)

Standard Diffstak 63/150P

(2.75)

65

Standard Diffstak 63/150M
Diffstak

70

50
(1.97)
4 x Ø9 (0.35)
x Ø110 (4.33)

360 (14.17)

105
(4.13)

PAGE
4

94
(3.70)

135 (5.31)

213 (8.38)
Ø130 (5.12)

17
(0.67)

3

Ø76

M – model pumps have a manually operated high vacuum isolation valve.
P – model pumps have a pneumatically actuated high vacuum isolation valve.
Pumping speed
nitrogen
135 l s-1
hydrogen
200 l s-1
Minimum backing pump displacement1
5 m3h-1
Recommended backing pump
RV5
Recommended fluid
Santovac® 5
Fluid charge (dry)
60 ml
Inlet connection compatible with
ISO63
Backing connection
NW10
Cooling water connection
6 mm compression fittings
Heater power
0.45 kW
Minimum cooling-water flow at 20 °C
42 l h-1
2
Pneumatic connections
6 mm coupling x !/4 BSP male stud
Pneumatic actuating pressure2
Minimum
2.4 bar / 35 psi
Maximum
6.9 bar / 100 psi
Weight
63/150M
8 kg
63/150P
9 kg
1

For maximum throughput. For applications where maximum throughput is not required, use an RV3.

2

63/150P only

Diffstak 63/150P
The BRV and PVK valves shown in these dimension diagrams are optional accessories.

ORDERING INFORMATION
PRODUCT DESCRIPTION

Santovac® 5
Silicone DC702
Silicone DC704EU
Silicone DC705

Heater (0.45 kW)
110-125 V
210-225 V
230-250 V
Valve plate "O" ring VIT0335
Valve shaft seal "O" ring VIT00121
Valve blanking plate "O" ring VIT01121
Backing "O" ring assembly (fluoroelastomer)
Drain and filler plug "O" rings1

5 × 10-9
7 × 10-6
7 × 10-8
3 × 10-8

0.6
1.2
0.8
0.6

3

ORDERING NUMBER

Diffstak 63/150M
110-125 V 1-ph 50/60 Hz
B34431976
210-225 V 1-ph 50/60 Hz
B34431977
230-250 V 1-ph 50/60 Hz
B34431978
Diffstak 63/150P
110-125 V 1-ph 50/60 Hz
B34432976
210-225 V 1-ph 50/60 Hz
B34432977
230-250 V 1-ph 50/60 Hz
B34432978
Supplied with: NW10 elbow, NW10 centering-ring, NW10 clamp, water pipe couplings and
ferrules, inlet ISO 63 Co-Seal
SPARES

CRITICAL BACKING
PRESSURE
ULTIMATE PRESSURE
(mbar)
(mbar)

FLUID

10

(2.99)
125 (4.92)

ORDERING NUMBER

H01700182
H01700186
H01700191
H02106050
H02106010
H02106012
C10511395
H02123027

l s -1

1 Pack of 5
10

For information about installation of Diffstak pumps, see page 3-12. For
details of suitable backing/ roughing valves, see page 5-14; for
electropneumatic control valves, see page 5-20; for thermal snap-switch
accessories, see page 3-15; for hardware components, see page 6-1; and
for vapor pump fluids, see page 6-46.

2

Diffstak 63/150
10

1

10

10

-8

-6

10

10

-7

-5

10

10

-6

-4

10

10

-5

-3

10

10

-4

-2

10

10

-3

-1

10

10

-2

0

mbar

Pa

Santovac® is a registered trademark of Santovac Fluids Inc, USA

Shop online at www.bocedwards.com

UNVALVED DIFFSTAK 63/150C

70

Ø95 (3.74)

(2.76)

85 (3.35)

155 (6.10)

(2.99)

334 (13.15)

148 (5.83)

179 (7.05)

69
(2.72)

50
(1.97)

Ø76

94 (3.70)
122 (4.80)
192 (7.56)

ORDERING INFORMATION
PRODUCT DESCRIPTION

ORDERING NUMBER

Unvalved Diffstak 63/150C

SPARES

Heater (0.45 kW)
110-125 V
210-225 V
230-250 V

TECHNICAL DATA
C – model pumps have an unvalved ISO collar inlet flange.
Pumping speed
nitrogen
150 l s-1
hydrogen
225 l s-1
Minimum backing pump displacement1
5 m3h-1
Recommended backing pump
RV5
Recommended fluid
Santovac® 5
Fluid charge (dry)
60 ml
Inlet connection
ISO63
Backing connection
NW10
Cooling water connection
6 mm compression fittings
Minimum cooling-water flow at 20 °C
42 l h-1
Heater power
0.45 kW
Weight
5 kg
1

For information on the Cryo-cooled version, contact BOC Edwards.

CRITICAL BACKING
PRESSURE
ULTIMATE PRESSURE
(mbar)
(mbar)

FLUID

l s -1

10

10

5 × 10-9
7 × 10-6
7 × 10-8
3 × 10-8

0.6
1.2
0.8
0.6

3

2

Diffstak 63/150C
10

1

10

10

-8

-6

10

10

-7

-5

10

10

-6

-4

10

10

-5

-3

10

10

-4

-2

10

10

-3

-1

10

10

H01700182
H01700186
H01700191

For information about spares, refer to standard Diffstak 63/150
(see page 3-4). For information about installation of Diffstak pumps,
see page 3-12. For details of suitable backing/ roughing valves,
see page 5-14; for electropneumatic control valves, see page 5-20; for
thermal snap-switch accessories, see page 3-15; for hardware components,
see page 6-1; and for vapor pump fluids, see page 6-46.

For maximum throughput.

Santovac® 5
Silicone DC702
Silicone DC704EU
Silicone DC705

ORDERING NUMBER

-2

0

mbar

Pa

Shop online at www.bocedwards.com

SCIENTIFIC VAPOR PUMPS

Diffstak 63/150C
110-125 V 1-ph 50/60 Hz
B34433976
210-225 V 1-ph 50/60 Hz
B34433977
230-250 V 1-ph 50/60 Hz
B34433978
Supplied with: NW10 elbow, NW10 centering-ring, NW10 clamp, water pipe couplings and
ferrules, inlet ISO 63 Co-Seal.

3
PAGE
5

(4.8)

271 (10.67)

106

86

108
(4.25)

(4.17)

40 (1.58)

(3.39)
122

276 (10.87)
Ø165 (6.5)

35
(1.38)

STANDARD DIFFSTAK 100/300M AND
100/300P

437 (17.2)

8 x Ø9 (0.35)
x Ø145 (5.71)
Ø111
(4.37)
127
(5.0)
183
(7.2)

Diffstak 100/300M

TECHNICAL DATA

For maximum throughput

2

100/300P only

Santovac® 5
Silicone DC702
Silicone DC704EU
Silicone DC705

l s -1

10

10

275 (10.83)

122
(4.8)

183 (7.2)

Diffstak 100/300P
The BRV and PVK valves shown in these dimension diagrams are optional accessories.

ORDERING INFORMATION
PRODUCT DESCRIPTION

ULTIMATE PRESSURE
(mbar)

5 × 10-9
7 × 10-6
7 × 10-8
3 × 10-8

0.6
1.2
0.8
0.6

3

ORDERING NUMBER

Diffstak 100/300M
110-125 V 1-ph 50/60 Hz
B34631976
210-225 V 1-ph 50/60 Hz
B34631977
230-250 V 1-ph 50/60 Hz
B34631978
Diffstak 100/300P
110-125 V 1-ph 50/60 Hz
B34632976
210-225 V 1-ph 50/60 Hz
B34632977
230-250 V 1-ph 50/60 Hz
B34632978
Supplied with: NW25 elbow, NW25 centering-ring, NW25 clamp, water pipe couplings and
ferrules, inlet ISO 100 Co-Seal
Heater (0.65 kW)
110-125 V
210-225 V
230-250 V
Valve plate "O" ring VIT0346
Valve shaft seal "O" ring VIT00121
Valve blanking plate "O" ring VIT01121
Backing "O" ring assembly (fluoroelastomer)
Drain and filler plug "O" rings1

ORDERING NUMBER

H01700199
H01700097
H01700190
H02106061
H02106010
H02106012
C10514395
H02123027

1 Pack of 5

For information about installation of Diffstak pumps, see page 3-12. For
details of suitable backing/ roughing valves, see page 5-14; for
electropneumatic control valves, see page 5-20; for thermal snap-switch
accessories, see page 3-15; for hardware components, see page 6-1; and
for vapor pump fluids, see page 6-46.

2

Diffstak 100/300
10

86
(3.39)
127 (5.0)

SPARES

CRITICAL BACKING
PRESSURE
(mbar)

FLUID

Ø110
(4.33)

M – model pumps have a manually operated high vacuum isolation valve.
P – model pumps have a pneumatically actuated high vacuum isolation valve.
Pumping speed
nitrogen
280 l s-1
hydrogen
500 l s-1
Minimum backing pump displacement1
5 m3h-1
Recommended backing pump
RV5 or RV8
Recommended fluid
Santovac® 5
Fluid charge (dry)
125 ml
Inlet connection compatible with
ISO100
Backing connection
NW25
Cooling water connection
6 mm compression fittings
Heater power
0.65 kW
Minimum cooling-water flow at 20 °C
60 l h-1
2
Pneumatic connections
6 mm coupling x ¼ BSP male
stud
Pneumatic actuating pressure2
Minimum
2.4 bar / 35 psi
Maximum
6.9 bar / 100 psi
Weight
100/300M
12 kg
100/300P
13 kg
1

106
(4.17)

Standard Diffstak 100/300P

437 (17.2)

Standard Diffstak 100/300M

40 (1.57)
8 x Ø9 (0.35)
x Ø145 (5.71)

153
(6.02)

35
108

PAGE
6

(1.38)

127 (5.0)

276 (10.87)
Ø165 (6.5)

(4.25)

3

1

10

10

-8

-6

10

10

-7

-5

10

10

-6

-4

10

10

-5

-3

10

10

-4

-2

10

10

-3

-1

10

10

-2

0

mbar

Pa

Shop online at www.bocedwards.com

127 (5.0)
40
(1.57)

122 (4.8)
86
(3.39)

414 (16.30)

177 (6.97)

Ø111
(4.37)

206 (8.11)

84
(3.31)

Ø130 (5.12)

216 (8.5)

UNVALVED DIFFSTAK 100/300C AND
100/300F

127 (5.0)
159 (6.26)
252 (9.92)

Diffstak 100/300C

3

1

For maximum throughput

CRITICAL BACKING
PRESSURE
ULTIMATE PRESSURE
(mbar)
(mbar)

FLUID

Santovac® 5
Silicone DC702
Silicone DC704EU
Silicone DC705

l s -1

10

10

122 (4.8)
86
(3.39)
127 (5.0)
159 (6.26)
252 (9.92)

Diffstak 100/300F

ORDERING INFORMATION
PRODUCT DESCRIPTION

ORDERING NUMBER

Diffstak 100/300C
110-125 V 1-ph 50/60 Hz
B34633976
210-225 V 1-ph 50/60 Hz
B34633977
230-250 V 1-ph 50/60 Hz
B34633978
Diffstak 100/300F
110-125 V 1-ph 50/60 Hz
B34640976
210-225 V 1-ph 50/60 Hz
B34640977
230-250 V 1-ph 50/60 Hz
B34640978
Supplied with: NW25 elbow, NW25 centering-ring, NW25 clamp, water pipe couplings and
ferrules inlet ISO Co-Seal (C version only).
SPARES

5 × 10-9
7 × 10-6
7 × 10-8
3 × 10-8

0.6
1.2
0.8
0.6

435 (17.13)

C – model pumps have an unvalved ISO collar inlet flange.
F – model pumps are unvalved and have a copper compression gasket flange
(ConFlat®).
Pumping speed
nitrogen
300 l s-1
hydrogen
535 l s-1
Minimum backing pump displacement1
5 m3h-1
Recommended backing pump
RV5 or RV8
Recommended fluid
Santovac® 5
Fluid charge (dry)
125 ml
Inlet connection
100/300C
ISO100
100/300F
6 inch
Backing connection
NW25
Cooling water connection
6 mm compression fittings
Minimum cooling-water flow at 20 °C
60 l h-1
Heater power
0.65 kW
Weight
100/300C
9 kg
100/300F
10 kg

(1.57)

Heater (0.65 kW)
110-125 V
210-225 V
230-250 V

ORDERING NUMBER

H01700199
H01700097
H01700190

3

For information about spares, refer to standard Diffstak 100/300
(see page 3-6). For information about installation of Diffstak pumps,
see page 3-12. For details of suitable backing/ roughing valves,
see page 5-14; for electropneumatic control valves, see page 5-20; for
thermal snap-switch accessories, see page 3-15; for hardware components,
see page 6-1; and for vapor pump fluids, see page 6-46.

2

For information on the Cryo-cooled version, contact BOC Edwards.
Diffstak 100/300C, F
10

1

10

10

-8

-6

10

10

-7

-5

10

10

-6

-4

10

10

-5

-3

10

10

-4

-2

10

10

-3

-1

10

10

-2

0

mbar

Pa

Shop online at www.bocedwards.com

SCIENTIFIC VAPOR PUMPS

TECHNICAL DATA

198 (7.8)

Unvalved Diffstak 100/300F

40
Ø109
(4.29)

227 (8.94)

105
(4.13)

Unvalved Diffstak 100/300C

PAGE
7

127 (5.0)

16 x Ø8.3 (0.33)
x Ø130.2 (5.13)

216 (8.5)

Ø152 (5.98)

192 (7.56)

(4.88) (5.91)

124

150

123
(4.84)
610 (24.02)

8 x Ø11 (0.43)
x Ø200 (7.87)

354 (13.94)

150
(5.91)

90
(3.54)

364 (14.33)
Ø225 (8.86)

58
(2.28)

STANDARD DIFFSTAK 160/700M AND
160/700P

Ø157
(6.18)
170
(6.69)
218 (8.58)

Diffstak 160/700M

M – model pumps have a manually operated high vacuum isolation valve.
P – model pumps have a pneumatically actuated high vacuum isolation valve.
Pumping speed
nitrogen
700 l s-1
hydrogen
1300 l s-1
Minimum backing pump displacement1
12 m3h-1
Recommended backing pump
RV12 or E2M18
Recommended fluid
Santovac® 5
Fluid charge (dry)
255 ml
Inlet connection compatible with
ISO160
Backing connection
NW25
Cooling water connection
10 mm compression fittings
Heater power
1.35 kW
Minimum cooling-water flow at 20 °C
115 l h-1
2
Pneumatic connections
6 mm coupling x ¼ BSP male
stud
2
Pneumatic actuating pressure
Minimum
2.4 bar / 35 psi
Maximum
6.9 bar / 100 psi
Weight
160/700M
26 kg
160/700P
27 kg
For maximum throughput

2

160/700P only

CRITICAL BACKING
PRESSURE
(mbar)

Santovac® 5
Silicone DC702
Silicone DC704EU
Silicone DC705

l s -1

10

ULTIMATE PRESSURE
(mbar)

5 × 10-9
7 × 10-6
7 × 10-8
3 × 10-8

0.6
1.2
0.8
0.6

4

3

330 (12.99)

(3.50)
150
124

191

(4.88) (5.91)

Ø157
(6.18)
170
(6.69)

Diffstak 160/700P
The BRV and PVK valves shown in these dimension diagrams are optional accessories.

ORDERING INFORMATION
PRODUCT DESCRIPTION

ORDERING NUMBER

Diffstak 160/700M
110-125 V 1-ph 50/60 Hz
B34831976
210-225 V 1-ph 50/60 Hz
B34831977
230-250 V 1-ph 50/60 Hz
B34831978
Diffstak 160/700P
110-125 V 1-ph 50/60 Hz
B34832976
210-225 V 1-ph 50/60 Hz
B34832977
230-250 V 1-ph 50/60 Hz
B34832978
Supplied with: NW25 elbow, NW25 centering-ring, NW25 clamp, water pipe couplings and
ferrules, inlet ISO 160 Co-Seal
Heater (one heater of each power is required)
110-125 V 0.35 kW
110-125 V 1 kW
210-225 V 0.35 kW
210-225 V 1 kW
230-250 V 0.35 kW
230-250 V 1 kW
Valve plate "O" ring VIT1221
Valve shaft seal "O" ring VIT01111
Valve blanking plate "O" ring VIT01111
Backing "O" ring assembly (fluoroelastomer)
Drain and filler plug "O" rings1

ORDERING NUMBER

H01700102
H01700059
H01700107
H01700063
H01700113
H01700054
H02106221
H02106011
H02106011
C10514395
H02123027

1 Pack of 5

Diffstak 160/700
10

150
(5.91)

218 (8.58)

SPARES

FLUID

10

8 x Ø11 (0.43)
x Ø200 (7.87)

(7.52)

TECHNICAL DATA

610 (24.02)

Standard Diffstak 160/700P

1

192 (7.56)

123
(4.84)

Ø225 (8.86)

89

490 (19.29)

58
(2.28)

3
PAGE
8

2

10

10

-8

-6

10

10

-7

-5

10

10

-6

-4

10

10

-5

-3

10

10

-4

-2

10

10

-3

-1

10

10

-2

0

mbar

Pa

For information about installation of Diffstak pumps, see page 3-12. For
details of suitable backing/ roughing valves, see page 5-14; for
electropneumatic control valves, see page 5-20; for thermal snap-switch
accessories, see page 3-15; for hardware components, see page 6-1; and
for vapor pump fluids, see page 6-46.

Shop online at www.bocedwards.com

UNVALVED DIFFSTAK 160/700C AND
160/700F

192 (7.56)
Ø180 (7.09)

90
(3.54)

287 (11.3)

150
(5.91)

(6.14)

578 (22.76)

242 (9.53)

290 (11.42)

108

(4.25)

150
(5.91)
Ø156

170 (6.69)

PHOTO REQUIRED

218 (8.58)
355 (13.98)

Diffstak 160/700C

3

192 (7.56)

90

TECHNICAL DATA

170 (6.69)

C – model pumps have an unvalved ISO collar inlet flange.
F – model pumps are unvalved and have a copper compression gasket flange
(ConFlat®).
Pumping speed
nitrogen
760 l s-1
hydrogen
1410 l s-1
Minimum backing pump displacement1
12 m3h-1
Recommended backing pump
RV12 or E2M18
Recommended fluid
Santovac® 5
Fluid charge (dry)
250 ml
Inlet connection
160/700C
ISO160
160/700F
8 inch
Backing connection
NW25
Cooling water connection
10 mm compression fittings
Minimum cooling-water flow at 20 °C
115 l h-1
Heater power
1.35 kW
Weight
160/700C
18 kg
160/700F
20 kg
1

For maximum throughput

FLUID

l s -1

10

10

355 (13.98)

Diffstak 160/700F

ORDERING INFORMATION
PRODUCT DESCRIPTION

3

ORDERING NUMBER

Diffstak 160/700C
110-125 V 1-ph 50/60 Hz
B34833976
210-225 V 1-ph 50/60 Hz
B34833977
230-250 V 1-ph 50/60 Hz
B34833978
Diffstak 160/700F
110-125 V 1-ph 50/60 Hz
B34840976
210-225 V 1-ph 50/60 Hz
B34840977
230-250 V 1-ph 50/60 Hz
B34840978
Supplied with: NW25 elbow, NW25 centering-ring, NW25 clamp, water pipe couplings and
ferrules, inlet ISO 160 Co-Seal (C version only).
Heater (one heater of each power is required)
110-125 V 0.35 kW
110-125 V 1 kW
210-225 V 0.35 kW
210-225 V 1 kW
230-250 V 0.35 kW
230-250 V 1 kW

5 × 10-9
7 × 10-6
7 × 10-8
3 × 10-8

0.6
1.2
0.8
0.6

218 (8.58)

SPARES

CRITICAL BACKING
PRESSURE
ULTIMATE PRESSURE
(mbar)
(mbar)

Santovac® 5
Silicone DC702
Silicone DC704EU
Silicone DC705

(5.91)
287 (11.3)

150

(6.14)

588 (23.15)

252 (9.92)

300 (11.81)

118

Unvalved Diffstak 160/700F

PAGE
9

150
(5.91)
Ø156

ORDERING NUMBER

H01700102
H01700059
H01700107
H01700063
H01700113
H01700154

For information about spares, refer to standard Diffstak 160/700
(see page 3-8). For information about installation of Diffstak pumps,
see page 3-12. For details of suitable backing/ roughing valves,
see page 5-14; for electropneumatic control valves, see page 5-14; for
thermal snap-switch accessories, see page 3-15; for hardware components,
see page 6-1; and for vapor pump fluids, see page 6-46.

2

For information on the Cryo-cooled version, contact BOC Edwards.
Diffstak 160/700C, F
10

1

10

10

-8

-6

10

10

-7

-5

10

10

-6

-4

10

10

-5

-3

10

10

-4

-2

10

10

-3

-1

10

10

-2

0

mbar

Pa

Shop online at www.bocedwards.com

SCIENTIFIC VAPOR PUMPS

Unvalved Diffstak 160/700C

(4.65)

20 x Ø8.3 (0.33)
x Ø181 (7.12)

(3.54)

Ø203 (7.99)

625 (24.61)

343 (13.5)

633 (24.92)

180 (7.09)

12 x Ø11 (0.43)
x Ø310 (12.2)

408 (18.31)

769 (30.28)

212
(8.35)

(13.19)

290 (11.42)

94
(3.7)

Ø335

100
(3.94)
134
310
(5.28) (12.2)

STANDARD DIFFSTAK 250/2000M AND
250/2000P

Ø261
(10.28)
242
(9.53)
412 (16.22)

Diffstak 250/2000M

290 (11.42)

TECHNICAL DATA
M – model pumps have a manually operated high vacuum isolation valve.
P – model pumps have a pneumatically actuated high vacuum isolation valve.
Pumping speed
nitrogen
2000 l s-1
hydrogen
3000 l s-1
Minimum backing pump displacement1
40 m3h-1
Recommended backing pump
E2M40
Recommended fluid
Santovac® 5
Fluid charge (dry)
500 ml
Inlet connection compatible with
ISO250
Backing connection
NW40
Cooling water connection
10 mm compression fittings
Heater power
2.25 kW
Minimum cooling-water flow at 20 °C
180 l h-1
2
Pneumatic connections
6 mm coupling x ¼ BSP male
stud
Pneumatic actuating pressure2
Minimum
2.4 bar / 35 psi
Maximum
6.9 bar / 100 psi
Weight
250/2000M
59 kg
250/2000P
60 kg
1

For maximum throughput.

2

250/2000P only

CRITICAL BACKING
PRESSURE
(mbar)

FLUID

Santovac® 5
Silicone DC702
Silicone DC704EU
Silicone DC705

l s -1

10

10

ULTIMATE PRESSURE
(mbar)

5 × 10-9
7 × 10-6
7 × 10-8
3 × 10-8

0.6
1.2
0.8
0.6

4

590 (23.23)

Ø261
(10.28)
242
(9.53)
412 (16.22)

Diffstak 250/2000P
The BRV and PVK valves shown in these dimension diagrams are optional accessories.

ORDERING INFORMATION
PRODUCT DESCRIPTION

ORDERING NUMBER

Diffstak 250/2000M
110-125 V 1-ph 50/60 Hz
B35031976
210-225 V 1-ph 50/60 Hz
B35031977
230-250 V 1-ph 50/60 Hz
B35031978
Diffstak 250/2000P
110-125 V 1-ph 50/60 Hz
B35032976
210-225 V 1-ph 50/60 Hz
B35032977
230-250 V 1-ph 50/60 Hz
B35032978
Supplied with: NW40 elbow, NW40 centering-ring, NW40 clamp, water pipe couplings and
ferrules, inlet ISO 250 trapped "O" ring
SPARES

Heater (one heater of each power is required)
110-125 V 0.85 kW
110-125 V 1.4 kW
210-225 V 0.85 kW
210-225 V 1.4 kW
230-250 V 0.85 kW
230-250 V 1.4 kW
Valve plate "O" ring VIT0449
Valve shaft seal "O" ring VIT1121 (pack of 2)
Valve blanking plate "O" ring VIT1128 (pack of 2)
Backing "O" ring assembly (fluoroelastomer)
Drain and filler plug "O" rings (pack of 5)

ORDERING NUMBER

H01700140
H01700161
H01700134
H01700155
H01700137
H01700158
H02106093
H02106121
H02106128
C10516395
H02123027

For information about installation of Diffstak pumps, see page 3-12. For
details of suitable backing/ roughing valves, see page 5-14; for
electropneumatic control valves, see page 5-20; for thermal snap-switch
accessories, see page 3-15; for hardware components, see page 6-1; and
for vapor pump fluids, see page 6-46.

3

Diffstak 250/2000
10

100
(3.94)
310
(12.2)

343 (13.5)

94

(3.7)

134
(5.28)
229

(9.02)

Standard Diffstak 250/2000M

12 x Ø11 (0.43)
x Ø310 (12.2)

180 (7.09)

465 (18.31)

769 (30.28)

PAGE
10

212

3

(8.35)

724 (28.5)
Ø335
(13.19)

2

10

10

-8

-6

10

10

-7

-5

10

10

-6

-4

10

10

-5

-3

10

10

-4

-2

10

10

-3

-1

10

10

-2

0

mbar

Pa

Shop online at www.bocedwards.com

ORDERING INFORMATION

UNVALVED DIFFSTAK 250/2000C

PRODUCT DESCRIPTION

C – model pumps have an unvalved ISO collar inlet flange.
Pumping speed
nitrogen
2130 l s-1
hydrogen
3200 l s-1
Minimum backing pump displacement1
40 m3h-1
Recommended backing pump
E2M40
Recommended fluid
Santovac® 5
Fluid charge (dry)
500 ml
Inlet connection
ISO250
Backing connection
NW40
Cooling water connection
10 mm compression fittings
Minimum cooling-water flow at 20 °C
180 l h-1
Heater power
2.25 kW
Weight
46 kg

SPARES

Heater (one heater of each power is required)
110-125 V 0.85 kW
110-125 V 1.4 kW
210-225 V 0.85 kW
210-225 V 1.4 kW
230-250 V 0.85 kW
230-250 V 1.4 kW

For maximum throughput

10

For information on the Cryo-cooled version, contact BOC Edwards.

4

3

Diffstak 250/2000C
10

2

10

10

-8

-6

10

10

-7

10

-5

10

-6

10

-4

10

-5

-3

10

10

-4

10

-2

10

-3

10

-1

10

-2

0

mbar

Pa

290 (11.42)

Ø290 (11.42)

180
(7.09)

310 (12.2)

522 (20.55)

100

(10.28)
(3.94)

357 (14.06)

Ø261

381 (15.0)

142
(5.59)

l s -1

10

5 × 10-9
7 × 10-6
7 × 10-8
3 × 10-8

0.6
1.2
0.8
0.6

H01700140
H01700161
H01700134
H01700155
H01700137
H01700158

242 (9.53)
290 (11.42)
503 (19.8)

Shop online at www.bocedwards.com

3
PAGE
11

SCIENTIFIC VAPOR PUMPS

FLUID

Santovac® 5
Silicone DC702
Silicone DC704EU
Silicone DC705

ORDERING NUMBER

For information about spares, refer to standard Diffstak 250/2000
(see page 3-10). For information about installation of Diffstak pumps,
see page 3-12. For details of suitable backing/ roughing valves,
see page 5-14; for electropneumatic control valves, see page 5-20; for
thermal snap-switch accessories, see page 3-15; for hardware components,
see page 6-1; and for vapor pump fluids, see page 6-46.

CRITICAL BACKING
PRESSURE
ULTIMATE PRESSURE
(mbar)
(mbar)

706 (27.8)

1

ORDERING NUMBER

Diffstak 250/2000C
110-125 V 1-ph 50/60 Hz
B35033976
210-225 V 1-ph 50/60 Hz
B35033977
230-250 V 1-ph 50/60 Hz
B35033978
Supplied with: NW40 elbow, NW40 centering-ring, NW40 clamp, water pipe couplings and
ferrules, inlet ISO 250 trapped "O" ring

TECHNICAL DATA

DIFFSTAK INSTALLATION
• M-model pumps have a manually operated high vacuum valve. P-model
pumps have a pneumatically operated high vacuum valve. Both M-model
and P-model pumps have inlet flanges which are compatible with ISO
flanges: the internal diameter of the inlet flange is narrower and the
flange is deeper than a standard ISO flange, to accommodate the high
vacuum valve.
Diffstak installation

• C-model pumps do not have a high vacuum valve and have an ISO flange
on the inlet (see pages 3-5 – 3-11).
• F-model pumps do not have a high vacuum valve and have a CF flange
on the inlet (see pages 3-7 and 3-9).
Refer to the diagrams and the tables on these pages to identify the pipeline
components and valves required to complete the typical Diffstak
installation shown. Items supplied with the pump are shown as a dotted
line. Read the footnotes below the diagrams and the tables for more
information and for details of the installation requirements for the different
models of Diffstaks.

63/150, 100/300, 160/700 installation

3 **

3

1

PAGE
12

10 ***

9

5P

2

7

4

6

*

5P
5

4P

4P

*
8

DIAGRAM
KEY

12
2 23
34

45
5
6
7
86
96
10 ***

COMPONENT
DESCRIPTION

ISO tube/collar assembly
Rotable flange pack
Bolts (size × minimum length, mm)
M- and P-model
F-model
Inlet seal
BRV valve, manual
BRV valve, pneumatic
Clamp
"O' ring assembly
4-port light-weight electropneumatic
control valve
5-port electropneumatic control valve
5-port electropneumatic control valve

63/150 COMPONENT
SIZE
QUANTITY

100/300 COMPONENT
SIZE
QUANTITY

160/700 COMPONENT
SIZE
QUANTITY

ISO63
ISO63

1
1

ISO100
ISO100

1
1

ISO160
ISO160

1
1

M8 × 75
–
ISO63
BRV10M
BRV10P
NW10
NW10
–

4
–
1
1
1
1
1
2

M8 × 75
M8 × 55
ISO100
BRV25M
BRV25P
NW25
NW25
–

8
16
1
1
1
1
1
2

M10 × 90
M8 × 60
ISO160
BRV25M
BRV25P
NW25
NW25
–

8
20
1
1
1
1
1
2

–
–

1
1

–
–

1
1

–
–

1
1

1
2

These items are supplied with the pumps, except that the inlet seal for the F-model pumps is not supplied.
Not required for F-model pumps; use CF fittings (which must be obtained from another supplier)

3
4

Not required for C-model pumps; use claw clamps to bolt the pump ISO inlet flange directly to the ISO tube/collar assembly.
Bolts are not available from BOC Edwards. Bolts are not required for C-model pumps; use claw clamps to bolt the pump ISO inlet flange directly to the ISO tube/collar assembly; use 4 claw clamps for ISO63,
ISO100 and ISO160 flanges.

5 These inlet seals are suitable for standard, cryo-cooled and C-model pumps only; use CF fittings (which must be obtained from another supplier) for F-model pumps.
6 Required only for pneumatic operation BRV valves; use either 1 5-port control valve or 2 4-port control valves. If you use 2 4-port control valves, you can use the isolation position of the BRV valve.
7 Required only for P-model pumps, to control the operation of the high vacuum valve.

Shop online at www.bocedwards.com

250/2000 installation

3
1
7

11

6

10 **

3P

5P

2
4 *

5
8

7
6
9
7

6

3

6

PAGE
13

10 **

SCIENTIFIC VAPOR PUMPS

3P

*

7

5

DIAGRAM
KEY

1
22
33
4
5
6
7
8
9
10 4
11 5

COMPONENT
DESCRIPTION

ISO tube/collar assembly
Rotatable flange pack
Bolts (size, minimum length,
mm) M- & P-model
Inlet seal, trapped "O" ring
PV40 valve, manual
PV40 valve, pneumatic
Clamp
"O' ring assembly
"T' piece
Elbow
3-port electropneumatic
control valve
5-port electropneumatic
control valve

COMPONENT
SIZE
QUANTITY

ISO250
ISO250
M10 × 110

1
1
12

ISO250
PV40MK
PV40PK
NW40
NW40
NW40
NW40

1
2
2
5
5
1
1
2

–
1

1 These items are supplied with the pump.
2 Not required for C-model pumps; use 6 claw clamps to bolt the pump ISO inlet flange directly to
the ISO tube/collar assembly.
3 Bolts are not available from BOC Edwards. Bolts are not required for Model-C pumps; use 6 claw
clamps to clamp the pump ISO inlet flange directly to the ISO tube/collar assembly.
4 Required only for pneumatic operation PVPK valves; use 1 3-port control valve for each of the two
PVPK valves.
5 Required only for P-model pumps, to control the high vacuum valve.

Shop online at www.bocedwards.com

16
(0.63)
76 (2.99)

176 (6.93)

41
(1.61)

54

(2.13)

EO50/60 AIR COOLED DIFFUSION PUMP

82 (2.23)
540 (21.26)
91 (3.58)

3
62 (2.44)

PAGE
14

80 (3.15)

NW50

NW10
EO50/60

The model EO50/60 is a compact, fast warm-up air-cooled diffusion pump
supplied complete with high integrity inlet Co-Seal, flange clamping ring and
optional baffle. The jet assembly is removable for easy cleaning. It meets
the need for a low-cost, high vacuum pump for use in desk top or mobile
scientific instruments, leak detectors, and general laboratory applications.
In many applications the rapid warm-up capability will enable the EO50/60
to be used without an isolation valve. However, the pump is not intended
for use on valveless systems where rapid cycling from atmospheric
pressure through the hot pump is required. BOC Edwards offers the TV
range of diffusion pumps for these applications.

TECHNICAL DATA
Pumping speed
Air
Hydrogen
Fluid charge
Recommended fluid
Minimum backing pump displacement
Recommended backing pump
Inlet connection
Backing connection
Heater power
Warm-up time
Weight
Minimum cooling air flow

60 l s-1 (27 l s-1 baffled)
70 l s-1 (50 l s-1 baffled)
20 ml
Santovac® 5, Silicone DC704EU
1 m3h-1
E2M1.5
NW50 flange
NW10 flange
200 W or 250 W
4 min
2 kg
4.5 m3h-1

CRITICAL BACKING
PRESSURE
ULTIMATE PRESSURE
(mbar)
(mbar)

FLUID

Santovac® 5
Silicone DC702
Silicone DC704EU
Silicone DC705
70

PRODUCT DESCRIPTION

1
50

SPARES

ACCESSORIES

-1

ls
30
20
10

1

-5

10

-3

DC704EU

10

-4

10

-2

2

10

-3

10

-1

H01707060
H01707062
H01707063
H01707075
H01707073
B27158466
C10511393
ORDERING NUMBER

For EO2, EO4, EO6 diffusion pumps contact BOC Edwards.

2

10

ORDERING NUMBER

Heater (0.2 kW)
110-125 V
210-225 V
230-250 V
Heater (0.25 kW)
90 V
180 V
Inlet Co-Seal fluoroelastomer
Backing NW10 "O" ring assembly, nitrile

40

0

ORDERING NUMBER

EO50/60 pump, 50/60 Hz, single phase
Pump voltage
Fan voltage
90 V (0.25 kW)
115 V
B30207090
110-125 V (0.2 kW)
115 V
B30207110
180 V (0.25 kW)
230 V
B30207180
210-225 V (0.2 kW)
230 V
B30207210
230-250 V (0.2 kW)
230 V
B30207240
Supplied with: NW50 combined baffle and centering-ring, NW50 centering-ring and "O"
ring, NW50 clamp and NW10 Co-Seal and clamp.

10

-2

10

0

B02304000
B27903006
H11401001
H11201040

For valves, see page 5-4. For NW pipeline fittings, see page 6-7. For vapor
pump fluids, see page 6-46.

EO50/60

60

ORDERING INFORMATION

Pump ready thermal snap-switch
Cooling fail thermal snap-switch
Santovac® 5 (100 ml)
Silicone DC704EU (500 ml)

5 × 10-8
5 × 10-6
5 × 10-7
5 × 10-7

0.6
1.2
0.8
0.6

142 (5.59)

10

-1

mbar

10

1

Pa

Santovac® 5

Shop online at www.bocedwards.com

615 (24.21)

DIFFSTAK COMBINED OUTFITS

540 (21.26)

340 (13.39)

ORDERING INFORMATION
PRODUCT DESCRIPTION

These are compact, high performance units featuring either the 63 mm or
100 mm Diffstak both backed by the RV5 two-stage rotary pump. Inlet
pumping speeds above the valve are 135 l s-1 and 280 l s-1 respectively.
The Diffstak has an integral water cooled baffle and is fitted with manual
quarterswing and backing/roughing valves. A manual air admittance valve
and a Pirani gauge are fitted in the backing line to the rotary pump. The
unit is supplied complete with rotary pump oil but vapor pump fluid should
be selected from the fluids and sealants section. Additional gauges are
optional extras.
1 A separate display or controller is required to show the pressure measured by the Pirani gauge head.

ACCESSORIES

B62426990
B62426952
B62426960
B62613990
B62613952
B62613960
ORDERING NUMBER

Line cords for combined outfit – 2m electrical supply cable with IEC socket
UK, three pin plug
North European plug
No plug
USA plug

A50505000
A50506000
A50508000
A50507000

The Diffstak combination includes the inlet flange Co-Seal, rotary pump oil
and instruction manual.
The Diffusion pump oil is not included, but can be ordered through
BOC Edwards, see page 6-46.
The power cable (line cord) is not included and needs to be ordered
separately, through BOC Edwards, see above.
To meet your special requirements we also offer custom built Diffstak
combined outfits or combined outfits featuring turbomolecular pumps,
see page 2-27. Please contact us for further details.

TECHNICAL DATA

Minimum cooling water
Maximum power
Weight (approx.)
Inlet connection
Outlet connection
Vapor pump
Rotary vacuum pump
Backing/roughing valve
Vacuum gauge – pirani

135 l s
3 x 10-8 mbar
3 x 10-6 Pa
42 l h-1
0.8 kW
51 kg
Diffstak 63 / 150M
RV5
BRV10K
APG-M-NW16
280 l s-1
3 x 10-8 mbar
3 x 10-6 Pa
60 l h-1
1.0 kW
61 kg
ISO 100
NW25
Diffstak 100 / 300M
RV5
BRV25K
APG-M-NW25

VAPOR PUMP – THERMAL
SNAP-SWITCHES
Vapor pump accessories

COOLING-FAIL THERMAL SNAP-SWITCH
This switch opens at a preset temperature to indicate failure of the cooling
water flow. The switch resets (closes) when the pump cools down to a
preset temperature.
35 (1.38)

14
(0.55)

Minimum cooling water flow
Maximum power
Weight (approx.)
Vapor pump
Rotary vacuum pump
Backing/roughing valve
Vacuum gauge – pirani
100 mm Diffstak outfit
Pumping speed above Diffstak valve
Ultimate vacuum (with Santovac® 5 fluid)

-1

65 (2.56)

63 mm Diffstak outfit
Pumping speed above Diffstak valve
Ultimate vacuum (with Santovac® 5 fluid)

Use with these pumps
Electrical data
Preset operation temperatures
Open
Close (automatic reset)

Diffstak, EO2, EO4, EO6, EO9
18B4A and 30B5A
25 A at 120 V a.c. or 240 V a.c.
46 ± 5 ºC
38 ± 5 ºC

ORDERING INFORMATION
PRODUCT DESCRIPTION

Cooling-fail thermal snap-switch
Shop online at www.bocedwards.com

ORDERING NUMBER

B02302000

3
PAGE
15

SCIENTIFIC VAPOR PUMPS

• Compact unit, delivering ~3 x 10-8 mbar ultimate pressure (depending
on vapor pump fluid used)
• Integrated manual backing/roughing valve and water cooled baffle
• Active Pirani gauge head* in backing line automatically interlocks the
Diffstak
• Enhanced safety shields for the highest temperature surfaces
• Castors are fitted as standard
• Fully factory tested
Diffstak combined outfits

ORDERING NUMBER

63 mm Diffstak Combined Outfit (requires electrical supply cable)
115 V 1-ph 50/60 Hz
220 V 1-ph 50/60 Hz
245 V 1-ph 50/60 Hz
100 mm Diffstak Combined Outfit (requires electrical supply cable)
115 V 1-ph 50/60 Hz
220 V 1-ph 50/60 Hz
245 V 1-ph 50/60 Hz

COOLING-FAIL THERMAL SNAP-SWITCH
(EO50/60 AND SI100)
This switch opens at a preset temperature to indicate failure of the cooling
water flow. The switch resets (closes) when the pump cools down to a
preset temperature.
0.51
(0.02)

This switch opens at a preset temperature to indicate that the pump is at
operating temperature. It can be used to control the opening of a high
vacuum valve. The switch resets (closes) when the pump cools down to a
preset temperature.
Ø16.00

18.54
(0.73)

(0.63)

30.15
(1.19)

23.8

EO50/60, SI100, EO100/90
15 A at 120 V a.c. resistive or
10 A at 240 V a.c. resistive

Preset operation temperatures
Open
Close (automatic reset)

50 ± 3 ºC
35 ± 3 ºC

Use with these pumps
90

PRODUCT DESCRIPTION

ORDERING NUMBER

Cooling-fail thermal snap-switch

15.50
(0.61)

All Diffstak, EO50/60, SI100, EO100/
EO4, EO6, EO9
6 A at 120 V a.c. resistive or
3 A at 240 V a.c. resistive or
7 A ar 30 V a.c. or dc, resistive

Electrical data
Preset operation temperatures
Open
Close (automatic reset)

ORDERING INFORMATION
B27903006

80 ± 2.8 ºC
60 ± 2.8 ºC

ORDERING INFORMATION
PRODUCT DESCRIPTION

COOLING-FAIL THERMAL SNAP-SWITCH
(TVA50/60 AND TVW50/60)
This switch opens at a preset temperature to indicate failure of the cooling
water flow. The switch resets (closes) when the pump cools down to a
preset temperature.
0.51

18.54

(0.02)

(0.73)

Ø3.68
(0.14)

B02304000

BOILER PROTECTION THERMAL
SNAP-SWITCH
This switch opens at a preset temperature when the pump boiler base
exceeds the maximum safe operating temperature. The switch resets
(closes) when the pump cools down to a preset temperature.
Ø18.03

0.51
(0.02)

Use with these pumps
Electrical data
Preset operation temperatures
Open
Close (automatic reset)

TVA50/60, TVW50/60
15 A at 120 V a.c. resistive or
10 A at 240 V a.c. resistive
60 ± 3 ºC
50 ± 3 ºC

Cooling-fail thermal snap-switch

Use with these pumps
Electrical data
Pre-set operation temperatures
Open
Close (automatic reset)

ORDERING INFORMATION
PRODUCT DESCRIPTION

15.49
(0.61)

Ø3.62
(0.14)
30.15
(1.19)

23.8

(0.94)

(0.71)

ORDERING NUMBER

B02304002

23.8

(0.63)

ORDERING NUMBER

Pump-ready thermal snap-switch

(0.94)

Ø15.88

30.15
(1.19)

PAGE
16

Use with these pumps
Electrical data

(0.94)

30.15
(1.19)

Ø3.68
(0.14)

3

0.51
(0.02)
Ø3.62
(0.14)
23.8
(0.94)

Ø15.88
(0.63)

PUMP-READY THERMAL SNAP-SWITCH

EO50/60, EO100/90, SI100
15 A at 220 V a.c. 50/60 Hz
10 A at 240 V a.c. 50/60 Hz
200 ± 4.4 ºC
184 ± 4.4 ºC

ORDERING INFORMATION
PRODUCT DESCRIPTION

Boiler protection thermal snap-switch

Shop online at www.bocedwards.com

ORDERING NUMBER

B02304001



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : No
Create Date                     : 2006:10:13 14:08:47Z
Keywords                        : Edwards, Diffuision, Pumps
Modify Date                     : 2018:02:27 11:15:18-08:00
XMP Toolkit                     : Adobe XMP Core 5.6-c015 84.159810, 2016/09/10-02:41:30
Producer                        : Acrobat Distiller 7.0.5 (Windows)
Creator Tool                    : FrameMaker 7.1
Metadata Date                   : 2018:02:27 11:15:18-08:00
Format                          : application/pdf
Title                           : Edwards Diffuision Pumps
Description                     : Edwards Diffuision Pumps
Creator                         : 
Subject                         : Edwards Diffuision Pumps
Document ID                     : uuid:2a3da380-26a0-4fd5-9838-6758dbbb2a70
Instance ID                     : uuid:45943960-548d-6042-a6af-a38fbd3bcb66
Page Mode                       : UseOutlines
Page Count                      : 16
EXIF Metadata provided by EXIF.tools

Navigation menu