PHOENIX L300i Operating Manual GA10218_002_C0 (1609) Phoenixl300i
User Manual: Pdf phoenixl300i
Open the PDF directly: View PDF .
Page Count: 112
Download | |
Open PDF In Browser | View PDF |
PHOENIX L300i Leak detector Operating Manual GA10218_002_C0 Part numbers 250000V01 250001V01 250002V01 251000V01 251001V01 251100V01 251101V01 Content Important safety information 5 1 9 1.1 Design and Function .................................................................... 9 1.1.1 Vacuum Diagram PHOENIX L300i .............................................. 10 1.1.2 Vacuum Diagram PHOENIX L300i DRY ...................................... 11 1.1.3 Vacuum Diagram PHOENIX L300i MODUL ................................ 12 1.1.4 Vacuum Method ........................................................................ 13 1.1.5 Partial Flow Method ................................................................... 14 1.1.6 Sniffer Mode .............................................................................. 14 1.2 Supplied Equipment ................................................................... 14 1.3 Technical Data ........................................................................... 15 1.3.1 Technical Data PHOENIX L300i ................................................. 15 1.3.2 Technical Data PHOENIX L300i DRY ......................................... 17 1.3.3 Technical Data PHOENIX L300i MODUL .................................... 19 1.3.4 Dimensional Drawings ................................................................ 21 1.4 Ordering Information .................................................................. 24 1.5 Accessories ............................................................................... 24 1.6 Default Settings .......................................................................... 26 2 Installation 27 2.1 Placement ................................................................................. 27 2.2 Conforming Utilisation ................................................................ 27 2.3 Ambient Conditions ................................................................... 27 2.4 Electrical Connections ................................................................ 27 2.4.1 Mains Power .............................................................................. 27 2.4.2 Connection for the Controller Signal and Accessories ................ 29 2.4.3 Vacuum Connections ................................................................. 32 3 2 Description Operation 34 3.1 Media Compatibility/Purge Gas .................................................. 34 3.2 Start-Up ..................................................................................... 34 3.2.1 Display ....................................................................................... 34 3.2.2 The Display in Run-Up Mode ..................................................... 35 3.2.3 The Display in Standby Mode ..................................................... 35 3.2.4 Gas Ballast/Purge ..................................................................... 35 3.2.5 The Display in Measurement Mode ............................................ 35 3.2.6 Call for Calibration ...................................................................... 36 GA10218_002_C0 - 10/2016 - © Leybold GmbH Content 3.2.7 Speaker Volume ........................................................................ 36 3.2.8 Status Line in the Display ........................................................... 36 3.2.9 Measurement Mode with Bargraph ............................................ 37 3.2.10 Measurement Mode with Trend Information ............................... 37 3.3 First Operation Check ................................................................ 37 3.4 Control Panel ............................................................................. 39 3.5 Interfaces ................................................................................... 44 3.6 Operation .................................................................................. 51 3.6.1 Main Menu ................................................................................ 52 3.6.2 View .......................................................................................... 54 3.6.3 Scale Linear/Logarithmic ........................................................... 54 3.6.4 Display-Range Auto/Manual ...................................................... 55 3.6.5 Time Axis ................................................................................... 56 3.6.6 Contrast .................................................................................... 56 3.6.7 Background in standby .............................................................. 57 3.6.8 Lower Display Limit .................................................................... 57 3.7 Mode ......................................................................................... 58 3.8 Trigger & Alarms ........................................................................ 59 3.8.1 Trigger Level 1 ........................................................................... 59 3.8.2 Trigger Level 2 ........................................................................... 60 3.8.3 Trigger Level 3 ........................................................................... 60 3.8.4 Units .......................................................................................... 60 3.8.5 Volume ...................................................................................... 61 3.8.6 Alarm Delay ............................................................................... 61 3.8.7 Audio Alarm Type ...................................................................... 62 3.9 Calibration ................................................................................. 63 3.10 Settings ..................................................................................... 64 3.10.1 Vacuum Settings ....................................................................... 65 3.10.2 Filter & Background ................................................................... 73 3.10.3 Mass ......................................................................................... 75 3.10.4 Miscellaneous (Language, Calibration request, Service interval...) 75 3.10.5 Parameter Save/Load ................................................................ 78 3.10.6 Monitoring Functions ................................................................. 78 3.11 Information ................................................................................ 82 3.11.1 Service ...................................................................................... 83 GA10218_002_C0 - 10/2016 - © Leybold GmbH 3 Content 3.12 Access Control .......................................................................... 83 3.12.1 Zero ........................................................................................... 83 3.12.2 Access to CAL Function ............................................................ 84 3.12.3 Change Menu-PIN ..................................................................... 84 3.13 Calibration ................................................................................. 84 3.13.1 Introduction ............................................................................... 84 3.13.2 The calibration Routines ............................................................. 85 3.13.3 Internal Calibration ..................................................................... 85 3.13.4 External Calibration .................................................................... 86 3.14 4 Maintenance 89 4.1 Safety Information ...................................................................... 89 4.2 Maintenance Intervals ................................................................ 89 4.3 Service at Leybold ..................................................................... 89 4.4 Maintenance Work ..................................................................... 90 4.4.1 Opening of the PHOENIX L300i ................................................. 90 4.4.2 Exchanging the Filter Mats ......................................................... 91 4.4.3 Exchanging the Oil ..................................................................... 92 4.4.4 Cleaning .................................................................................... 92 4.4.5 Exchanging the Fuses ................................................................ 92 4.4.6 Exhaust Oil Filter ........................................................................ 94 4.4.7 Turbo Molecular Pump ............................................................... 94 4.5 Calibrated Leak TL7 ................................................................... 95 4.5.1 Technical Data ........................................................................... 95 4.5.2 Factory Inspection ..................................................................... 95 5 4 Switching Off/Shutting Down ..................................................... 88 Troubleshooting 96 5.1 Hints for Troubleshooting ........................................................... 96 5.2 List of Errors & Warnings ............................................................ 96 6 Waste Disposal 103 7 Certificates 105 GA10218_002_C0 - 10/2016 - © Leybold GmbH Safety Information Important Safety Information Indicates procedures that must be strictly observed to prevent hazards to persons. Warning Indicates procedures that must be strictly observed to prevent damage to, or destruction of the product. Caution Emphasizes additional application information and other useful information provided within these Operating Instructions. Note The Leybold PHOENIX L300i leak detector has been designed for safe and efficient operation when used properly and in accordance with these Operating Instructions. It is the responsibility of the user to carefully read and strictly observe all safety precautions described in this section and throughout the Operating Instructions. The PHOENIX L300i must only be operated in the proper condition and under the conditions described in the Operating Instructions. It must be operated and maintained by trained personnel only. Consult local, state, and national agencies regarding specific requirements and regulations. Address any further safety, operation and/or maintenance questions to our nearest office. Failure to observe the following precautions could result in serious personal injury! Electrical hazards During all maintenance and connection work, make sure that the mains cable have been reliable disconnected and do not carry a mains voltage.The leak detector must only be used in with the hoods closed. The electrical connections must only be provided by a trained electrician as specified, for example, by the regulations EN 50110-1. Warning Avoid exposing any part of the human body to the vacuum. Only handle the leak detector when the pump is vented. After a mains power failure the leak detector can run up automatically once more. Before changing oil or fuses, make sure that the mains cable have been disconnected. The leak detector is not suited for operation in explosion hazard areas. GA10218_002_C0 - 10/2016 - © Leybold GmbH 5 Safety Information During operation the pump can become so hot that there is a danger of burns (>70°C, 158°F).Provide protection against contact with the hot components. Contaminated parts can be detrimental to health and environment. Before beginning to work, find out whether any parts are contaminated. Adhere to the relevant regulations and take the necessary precautions when handling contaminated parts. Caution Failure to observe the following precautions could result in damage to the pump! Unauthorized opening or modifications of the mechanical or electrical components of the leak detector void the warranty. The leak detector must only be opened by such persons who have been authorised by Leybold to do so. The leak detector can be damaged when using the wrong voltage. The voltage must be in the range 230V (+/- 5%), 115V (+/- 5%) e.g. 100V (+/- 5%) depending on the leak detector version. Make sure that the mains voltage rating on the PHOENIX L300i coincides with the locally available mains voltage. When the PHOENIX L300i is running in closed rooms the exhaust has to be put out of doors so that the oil vapor can not be breathed in. Ensure a sufficient air cooling. The air inlet as well the air discharge openings must never be obstructed. By using the optional iPad it is recommended to pull the fixture of the iPad into a raised position to optimize the cooling. The PHOENIX L300i is designed for indoor use only. Operate the PHOENIX L300i only in the permitted temperature range between +10°C and +40°C. Only Leybonol-LVO-310 (cat. no. L31001) must be used in the TIVAC D2,5E. Pumping condensable gases and steams: When pumping test sample water vapour that is inside the test object can attain to the forepump. With the water vapor that is in the air - especially in humid areas or when using humid or wet test samples - the acceptable compatibility of water vapor or capacity of water vapor respectively can be exceeded. The steam in the oil of the pump condenses when the water vapor rises over the acceptable value. So the attribute of the oil changes and danger of corrosion occurs for the pump. While using the leak detector with condensable gases and steams the oil of the forepump has to be controlled regularly. So you can recognize a condensation of water vapor in the pump. Usually the oil is light and lucent. When water vapor is 6 GA10218_002_C0 - 10/2016 - © Leybold GmbH Safety Information inside it gets blear and milky at operating state temperature. When turning the pump off water vapor condensates and raises the part of water in the oil. The leak detector must not directly be switched off after the process, in which condensable gases or steams are pumped, is finished. It must be running (at least 20 minutes) with opened gas ballast valve until the oil of the pump is freed from detached steam. When not taking care of this instruction there can be a corrosion within the pump, which will not be covered by our warranty. The height of the oil of the pump has to be controlled regularly. The normal intervals of changing the oil from the producer have to be taken care of. See instructions of the rotary vane pump. The references to diagrams, e.g. (1/2) consist of the Fig. no. and the item No. in that order. We reserve the right to alter the design or any data given in these Operating Instructions. The illustrations are not binding. Retain the Operating instructions for further use. Figures Symbols of vacuum technology Given in the following are some important symbols which are used in this Operating Instructions. Vacuum pump in general Turbo molecular pump Measuring instrument Valve Definition of Terms The range of the preamplifier and the vacuum ranges are selected automatically. The auto ranging feature of the PHOENIX L300i covers the entire range or leak rates depending on the selected operating mode. Not only the leak rate signal, but also the pressure in the test sample (inlet pressure P1) and the fore vacuum pressure (P2) are used for control purposes. Range switching between the main ranges is performed via valves. Fine range switching within the main ranges is implemented by switching over the gain factor of the preamplifier. Auto ranging GA10218_002_C0 - 10/2016 - © Leybold GmbH 7 Safety Information Mass alignment Mass alignment This function automatically aligns the mass spectrometer so that a maximum leak rate is displayed. The control processor changes the voltage which accelerates the ions in the selected mass range until a maximum ion current is detected by the ion detector. During each calibration the mass alignment is run automatically. Auto Zero Determination and automatic adaptation of the internal background. Through this function, the internal zero level of the instrument is determined which is then subtracted from the current leak rate signal. This function is run during the calibration process or when operating the start push button, provided the PHOENIX L300i has been running previously for at least 20 seconds in the standby or vent mode. GROSS GROSS is a measurement mode which allows high inlet pressure (15 to 0,2 mbar). The smallest detectable leak rate is 1 · 10-7 mbar l / s. FINE FINE is a measurement mode with inlet pressure < 0.2 mbar. The minimum detectable leak rate is 5 · 10-12 mbar l / s. PRECISION Precision is a measurement mode for the PHOENIX L300i DRY only from an inlet pressure < 0,01 mbar. In this mode the PHOENIX L300i DRY has the highest sensitivity, the minimum detectable leak rate is 3 · 10-11 mbar l / s. Fore-Vacuum Pressure Pressure in the fore vacuum between turbo pump and rotary vane pump. Internal helium background The existing helium partial pressure in the measurement system. The level of the internal helium background is measured in the standby mode and subtracted from the measured signal. Minimum detectable leak rate The smallest leak rate the PHOENIX L300i is able to detect ( 5·10-12 mbar l / s) in vacuum mode. Menu The menu allows the user to program the PHOENIX L300i according to his requirements. The menu has a tree architecture. Measurement mode The PHOENIX L300i measures the leak rate of the test sample. Default Status of the PHOENIX L300i when supplied by the factory. 8 GA10218_002_C0 - 10/2016 - © Leybold GmbH Description 1 Description The PHOENIX L300i is a leak detector for helium or hydrogen. This instrument may be used to detect the location and the size of leaks on objects under test in two different ways: When the test sample has been evacuated first and is sprayed with helium on the outside. It is required that a vacuum connection is provided between the PHOENIX L300i and the test sample (vacuum mode). or when a helium overpressure is provided in the test sample and the test sample is searched from the outside with a sniffer probe which is attached to the inlet port (sniffer mode). Fig. 1 View of the PHOENIX L300i 1 inlet flange 2 control panel 1.1 Design and Function The PHOENIX L300i basically is a helium leak detector for vacuum applications, i.e. the part under test is evacuated while the test is performed. The vacuum is achieved with a pumping system that is part of the PHOENIX L300i. In addition the vacuum can be generated by pumps which are set up in parallel to the PHOENIX L300i. The PHOENIX L300i MODUL needs a fore vacuum pump, dry or wet version, to be connected because this unit has no internal roughing pump. The connection (DN25 KF) is on the side or under the bottom of the PHOENIX L300i MODUL (Fig. 7). GA10218_002_C0 - 10/2016 - © Leybold GmbH 9 Description Another operating mode of the PHOENIX L300i is the Sniffer mode which can only be used when a sniffer line (See Chapter 1.1.6) is hooked up. 1.1.1 Vacuum Diagram PHOENIX L300i The vacuum diagram below shows the major components inside the PHOENIX L300i. 5 1 6 2 7 3 4 Fig. 2 Vacuum diagram PHOENIX L300i Pos. Description 1 MS: Mass Spectrometer, Helium sensor (180° magnetic field mass spectrometer) 2 Turbo molecular pump (TMP, provides high vacuum conditions in the MS) 3 Pirani gauge P2 (fore vacuum pressure) 4 Fore pressure pump (provides the fore vacuum pressure for the TMP and pumps down the parts under test) 5 Inlet Port 6 Pirani gauge P1 (inlet pressure) V1 … V7: Electromagnetic Valves to control the gas flows The mass spectrometer (MS) is mainly composed of the ion source with cathode, the magnetic separator and the ion collector. Gas molecules getting into the mass spectrometer are ionized by the ion source. These positively charged particles are accelerated into the magnetic field following a circular path, the radius of which depends on the mass-to-charge ratio of the ions. When mass 4 is selected (Default setting) only helium ions can pass this filter and reach the ion collector where the stream of the ions is measured as an electrical current. When selected another mass than 4, only the corresponding ions 10 GA10218_002_C0 - 10/2016 - © Leybold GmbH Description can pass the filter. For operation the mass spectrometer requires a vacuum level in the range of 1·104 mbar and lower. This pressure is provided by the turbo molecular pump which in turn is backed up by a fore vacuum pump. Besides maintaining the pressure in the mass spectrometer the pump system is used to evacuate the test parts. It is made sure that the pressure in the mass spectrometer is low enough under all circumstances. The valves V1, V2a, V2b, V4a control the gas flows when measuring. Valves V3 and V5 are used to vent the system and the Turbo pump, valve V6 controls the gas ballast function of the fore vacuum pump. Valve V7 opens and closes the internal test leak during calibration. With the pressure in the test part being lower than ambient pressure sprayed helium (or Hydrogen as forming gas) can penetrate into the part in case of a leakage. As soon as the pressure conditions allow it one of the valves to the TMP opens. Now Helium can penetrate into the mass spectrometer contrary to the pumping direction of the TMP. 1.1.2 Vacuum Diagram PHOENIX L300i DRY The PHOENIX L300i DRY has a diaphragm pump as fore vacuum pump, making it suitable for applications where oil sealed systems can not be used. Furthermore the PHOENIX L300i DRY contains one more valve, the valve 4b. This valve opens step by step to regulate the inlet pressure into the turbo pump. 5 1 6 MS P 1 V4 b 2 V 3 V7 V4 a V6 3 P 2 V V 2a V V 2b 5 1 4 Diaphragm pump Fig. 3 Vacuum diagram PHOENIX L300i DRY GA10218_002_C0 - 10/2016 - © Leybold GmbH 11 Description Pos. Description 1 MS: Mass Spectrometer, Helium sensor (180° magnetic field mass spectrometer) 2 Turbo molecular pump (TMP, provides high vacuum conditions in the MS) 3 Pirani gauge P2 (fore vacuum pressure) 4 Diaphragm pump (provides the fore vacuum pressure for the TMP and pumps down the parts under test) 5 Inlet Port 6 Pirani gauge P1 (inlet pressure) V1 … V7: Electromagnetic Valves to control the gas flows 1.1.3 Vacuum Diagram PHOENIX L300i MODUL The PHOENIX L300i MODUL has no roughing pump integrated as the other models. Therefore it can be used with an external pump only. This pump can be oil sealed or a dry version with a roughing capacity between 2.5 and 65 m3/h. This pump has to be connected to the DN25 KF at the side or under the bottom of the PHOENIX L300i MODUL. 6 1 5 MS P 1 V V7 V 4a 2 3 P 2 V 2a V6 external pump 4 Fig. 4 Vacuum diagram PHOENIX L300i MODUL 12 GA10218_002_C0 - 10/2016 - © Leybold GmbH V V V 2b 1 5 3 Description Pos. Description 1 MS: Mass Spectrometer, Helium sensor (180° magnetic field mass spectrometer) 2 Turbo molecular pump (TMP, provides high vacuum conditions in the MS) 3 Pirani gauge P2 (fore vacuum pressure) 4 Fore pressure pump (provides the fore vacuum pressure for the TMP and pumps down the parts under test) 5 Inlet Port 6 Pirani gauge P1 (inlet pressure) V1 … V7: Electromagnetic Valves to control the gas flows 1.1.4 Vacuum Method For the purpose of leak detection on a test sample (vacuum method), the sample has to be evacuated so that Helium or Hydrogen which is sprayed on to the outside, can enter through any leaks due to the pressure differential for detection by the PHOENIX L300i. The test sample is evacuated - START button (Fig 16./10) - by the backing pump or the external pump. In the case of larger test samples an additional external partial flow pump with a corresponding linking valve may be connected in parallel as required. Inlet valve V1 is opened so that the evacuation can take place. At the same time all other valves are closed in order to prevent an unacceptable pressure increase in the mass spectrometer. In this context (valve V2a closed) the turbo molecular pump is operated without being supported by the rotary vane pump. Since generally no gas is pumped out of the mass spectrometer, p2 remains constant or increases only slowly. The condition for the evacuation process described here is maintained until the inlet pressure p1 has dropped <15 mbar. Now the valves V2a and V2b are opened additionally. Possibly present Helium or Hydrogen may now flow upstream against the pumping direction of the turbo molecular pump into the mass spectrometer where it is detected. This measurement mode is called GROSS. In this mode, leak rates down to 10-8 mbar l/s can be detected. Since the rotary vane vacuum pump continues to evacuate the test sample via valves V2a, V2b and V1 the inlet pressure p1 will continue to drop. When the pressure drops below p1 < 0.2 mbar, the PHOENIX L300i will switch to the FINE mode, i.e. valve V1and V2b closes and valve V4a opens so that the gas flow enters the turbo molecular pump at the side. This offer two advantages: a) b) A part of the high pumping speed of the turbo molecular pump remains available for further evacuation of the test sample. The response time is inversely proportional to pumping speed.). The advantages offered by the counter flow principle can still be utilized In the FINE mode the full sensitivity of the PHOENIX L300i is reached. Because of the higher base pressure of the diaphragm pump the switching from GROSS to FINE mode of the PHOENIX L300i DRY is done by the valve V4b. When the pressure drops below 3,5 mbar the valves V1 and V2b will be close and V4b GA10218_002_C0 - 10/2016 - © Leybold GmbH 13 Description opens step by step. When valve V4b is open completely, pressure < 0,1 mbar, V4a will open also to get the maximum pumping speed. In PRECISION mode the PHOENIX L300i DRY opens the valve V4b only, with the disadvantage of low pumping speed but with the highest sensitivity. When the leak detection process is stopped – STOP-button – all valves except valve V2a are closed. Valve V3 is opened during venting of the inlet or test sample. 1.1.5 Partial Flow Method In the partial flow mode the test sample is additionally evacuated by an auxiliary pump. Using the optional partial flow pump set offers to the user the following advantages (PHOENIX L300i and PHOENIX L300i MODUL: - faster response time - entry into the measure mode already at an inlet pressure of 1000 mbar - faster venting of large test objects Alternatively to a partial flow pump set an external auxiliary pump may also be connected via a tee, this option is possible for the PHOENIX L300i dry and PHOENIX L300i MODUL also. However, in such a case the PHOENIX L300i will not be able to make measurements already at an inlet pressure of 1000 mbar. 1.1.6 Sniffer Mode The PHOENIX L300i may simply be converted into a sniffer leak detector via the rugged sniffer line (Cat. No. 252003) For this the KF flange of the sniffer line is connected to the inlet flange (Fig. 1/1) and the sniffer mode is selected through menu mode. After pressing START, the inlet valve V1 (Fig. 2) opens. The sniffer lines have been designed in such a way that the PHOENIX L300i is operated in the FINE mode. If the fore vacuum pressure P2 increases over 0,2 mbar respectively 0,1 mbar a warning sign and audio alarm comes up in the display In the measurement mode the helium present in the ambient air is now indicated as the leak rate (about 2 · 10-6 mbar l/s). Smaller leaks may be detected by pressing the ZERO-button. In sniffer mode the smallest detectable leak rate is < 1· 10-7 mbarl/s. 1.2 Supplied Equipment The PHOENIX L300i will be shipped in a special cardboard packed separately in a plastic foil as protection against dust. Supplied equipment includes: Leak Detector PHOENIX L300i Set of fuses Power cord Folder with documents (Operating instructions, Spare part list) 2 L-type screwed connections (hose connections) 1 hose nozzle Blank flange DN 25 KF Clamping ring DN 25 KF Centering ring DN 25 KF 14 GA10218_002_C0 - 10/2016 - © Leybold GmbH Description 1.3 Technical Data 1.3.1 Technical Data PHOENIX L300i Physical data Max. inlet pressure Physical data 15 mbar Minimum detectable Helium leak rates - in vacuum mode <5·10-12 mbar l / s - in sniffer mode <1·10-7 mbar l / s Minimum detectable Hydrogen leak rates - in vacuum mode <1·10-8 mbar l / s - in sniffer mode <1·10-7 mbar l / s Maximum Helium leak rate which can be displayed 0.1 mbar l / s Measurement range 12 decades Time constant of the leak rate signal (blanked off, <1 s 63% of the final value) Pumping speed (Helium) at the inlet - GROSS mode 0.4 l/s - FINE mode > 2.5 l/s Detectable masses 2, 3 and 4 Mass spectrometer 180° magnetic sector field Ion source 2 filaments; Iridium/Yttrium-oxide Inlet port DN 25 KF Run-up time (after starting) < 2 min To get down to the minimum detected leak rate range some conditions must be fulfilled: PHOENIX L300i has to run at least 20 minutes Ambient conditions must be stable (temperature, no vibration/accelerations.) The part under test has been evacuated long enough without using the zero function (background is no longer decreasing) ZERO must be active GA10218_002_C0 - 10/2016 - © Leybold GmbH 15 Description Electrical data Electrical data Power supply Other data Ambient conditions Cat.-No.250000V01 230 V, +/- 5%, 50/60 Hz Cat.-No.251000V01 115 V, +/- 5%, 60 Hz Cat.-No.251100V01 100 V, +/- 5%, 50/60 Hz Power consumption 420 VA Type of protection IP40 Other data Valves solenoid Dimensions (L × W × H) in mm 495 x 456 x 314 Weight in kg 40.0 Noise level dB (A) < 54 max. Audio alarm dB (A) 90 Contamination level (to IEC 60664-1) 2 Overvoltage category (to IEC 60664-1) II Ambient conditions For use within buildings Permissible ambient temperature (during operation) +10 °C … +40 °C 16 Permissible storage temperature -10 °C … +60 °C Maximum relative humidity 80% (up to 31°C) linear decreasing to 50% at 40°C Max. permissible height above sea level (during operation) 2000 m GA10218_002_C0 - 10/2016 - © Leybold GmbH Description 1.3.2 Technical Data PHOENIX L300i DRY Physical data Physical data Max. inlet pressure 15 mbar Minimum detectable Helium leak rates - in vacuum mode < 3·10-11 mbar l / s - in sniffer mode < 1·10-7 mbar l / s Minimum detectable Hydrogen leak rates - in vacuum mode < 1·10-8 mbar l / s - in sniffer mode < 1·10-7 mbar l / s Maximum Helium leak rate which can be displayed 0.1 mbar l / s Measurement range 11 decades Time constant of the leak rate signal (blanked off, <1 s 63% of the final value) Pumping speed (Helium) at the inlet GROSS mode 0.02 l / s PRECISION mode 0,4 l / s FINE mode > 2.5 l / s Detectable masses 2, 3 and 4 Mass spectrometer 180° magnetic sector field Ion source 2 filaments; Iridium/Yttrium-oxide Inlet port DN 25 KF Run-up time (after starting) < 2 min To get down to the minimum detected leak rate range some conditions must be fulfilled: PHOENIX L300i has to run at least 20 minutes Ambient conditions must be stable (temperature, no vibration/accelerations.) The part under test has been evacuated long enough without using the zero function (background is no longer decreasing) ZERO must be active GA10218_002_C0 - 10/2016 - © Leybold GmbH 17 Description Electrical data Electrical data Power supply Other data Ambient conditions Cat.-No.250001V01 230 V, 50 Hz Cat.-No.251001V01 115 V, 60 Hz Cat.-No.251101V01 100 V, 50/60 Hz Power consumption 350 VA Type of protection IP40 Other data Valves solenoid Dimensions (L × W × H) in mm 495 x 456 x 314 Weight in kg 35.5 Noise level dB (A) < 54 max. Audio alarm dB (A) 90 Contamination level (to IEC 60664-1) 2 Overvoltage category (to IEC 60664-1) II Ambient conditions For use within buildings Permissible ambient temperature (during operation) +10 °C … +40 °C 18 Permissible storage temperature -10 °C … +60 °C Maximum relative humidity 80% (up to 31°C) linear decreasing to 50% at 40°C Maximum permissible height above sea level (during operation) 2000 m GA10218_002_C0 - 10/2016 - © Leybold GmbH Description 1.3.3 Technical Data PHOENIX L300i MODUL Physical data Physical data Max. inlet pressure 15 mbar Minimum detectable Helium leak rates in vacuum mode - with Scroll pump < 8·10-12 mbar l / s - with oil sealed pump < 5·10-12 mbar l / s in sniffer mode <1·10-7 mbar l / s Minimum detectable Hydrogen leak rates - in vacuum mode < 1·10-8 mbar l / s - in sniffer mode < 1·10-7 mbar l / s Maximum Helium leak rate which can be displayed 0.1 mbar l / s Measurement range 12 decades Time constant of the leak rate signal (blanked off, < 1 s 63% of the final value) Pumping speed (Helium) at the inlet GROSS mode 1,0 l / s FINE mode > 2.5 l / s Detectable masses 2, 3 and 4 Mass spectrometer 180° magnetic sector field Ion source 2 filaments; Iridium/Yttrium-oxide Inlet port DN 25 KF Run-up time (after starting) < 2 min To get down to the minimum detected leak rate range some conditions must be fulfilled: PHOENIX L300i has to run at least 20 minutes Ambient conditions must be stable (temperature, no vibration/accelerations.) The part under test has been evacuated long enough without using the zero function (background is no longer decreasing) ZERO must be active GA10218_002_C0 - 10/2016 - © Leybold GmbH 19 Description Electrical data Electrical data Power supply Cat. No. 250002V01 100 - 240 V 50/60 Hz Power consumption 200 VA Type of protection IP40 Power cords (EU, USA, UK) 2.5 m Other data Other data Valves solenoid Dimensions (L × W × H) in mm 495 x 456 x 314 Weight in kg 35.5 Noise level dB (A) < 54 max. Audio alarm dB (A) 90 Contamination level (to IEC 60664-1) 2 Overvoltage category (to IEC 60664-1) II Ambient conditions For use within buildings Ambient conditions 20 Permissible ambient temperature (during operation) +10 °C … +40 °C Permissible storage temperature -10 °C … +60 °C Maximum relative humidity 80% (up to 31°C) linear decreasing to 50% at 40°C Maximum permissible height above sea level (during operation) 2000 m GA10218_002_C0 - 10/2016 - © Leybold GmbH Description 1.3.4 Dimensional Drawings Abmaße PHOENIX L300i Familie in mm und inch. (in Klammern) Dimensions PHOENIX L300i family in mm and inches (in brackets). Fig. 5 Dimensions PHOENIX L300i GA10218_002_C0 - 10/2016 - © Leybold GmbH 21 Description Abmessungen PHOENIX L300i Familie in mm und inch. (in Klammern) Dimensions PHOENIX L300i family in mm and inches (in brackets) Fig. 6 Dimensions PHOENIX L300i side view 22 GA10218_002_C0 - 10/2016 - © Leybold GmbH Description Abmessungen PHOENIX L300i in mm und inch (in Klammern). Dimensions PHOENIX L300i in mm and inches (in brackets). Fig. 7 Dimensions PHOENIX L300i MODUL GA10218_002_C0 - 10/2016 - © Leybold GmbH 23 Description 1.4 Ordering Information PHOENIX L300i EU-Version 250000V01 PHOENIX L300i DRY EU-Version 250001V01 PHOENIX L300i MODUL 250002V01 PHOENIX L300i US-Version 251000V01 PHOENIX L300i DRY US-Version 251001V01 PHOENIX L300i JP-Version 251100V01 PHOENIX L300i JP-Version 251101V01 1.5 Accessories The following parts can be ordered additionally: Sniffer line SL300 252003 Sniffer line SL301, 4 m 252025V01 Sniffer line SL301, 10 m 252026V01 Leak Ware (Software for data acquisition) 14090 Helium Sniffer QUICK-TEST Qt 100 15594 Remote control set RC310C consisting of: 252013V01 - Remote control - cable 4 m - mounting parts Extension cable for remote control,10m 14022 Remote control set RC310WL wireless consisting of: 252014V01 - Remote control - Radio transmitter - mounting parts Radio transmitter for remote control 252015V01 Spray gun with hose 16555 Set of connection plugs 20099024 Partial flow system (PHOENIX L300i and PHOENIX 14020 L300i MODUL) Adapter USB/RS232 24 GA10218_002_C0 - 10/2016 - © Leybold GmbH 800110V0103 Description Sniffer line With the use of a sniffer line the PHOENIX L300i can easily be converted to a sniffer leak detector. Sniffer line Available versions: SL300: 4 m SL301: 4 m and 10 m For further information on the sniffer lines SL300 and SL301 see the enclosed manuals. Remote Control RC310WL Wireless and RC310C Wired Remote control Fig. 8 Remote Control RC310WL wireless For further information on the remote controls RC310WL and RC310C see instruction manual with the document number 300306406_002_A1. Partial Flow System (PHOENIX L300i and PHOENIX L300i In the partial flow mode the test sample is additionally evacuated by an auxiliary pump. Using the optional partial flow pump set offers to the user the following advantages: Partial flow system - faster response time - entry into the measurement mode already at an inlet pressure of 1000 mbar - faster venting of large test objects The partial flow system consists of the components partial flow valve block, right angle valve DN 25 KF, control cable and vacuum hose with flange connections. The partial flow valve block with the right angle valve has to be connected to the inlet flange of the PHOENIX L300i. Connect the control cable to the Option port and the vacuum hose to the auxiliary pump. The PHOENIX L300i has to be configurated as described in chapter. For further detailed information please refer to operating instructions GA 10.277 partial flow system. GA10218_002_C0 - 10/2016 - © Leybold GmbH 25 Description 1.6 Default Settings The following parameters are set like shown when in the menu of the PHOENIX L300i under Settings > Parameters, Load / Save is chosen. 26 Scale linear Display range: 4 decades Time axis: 32 seconds LCD inverted off Background in standby mode: off Calibration request: off Mass: 4 (helium) Recorder: leak rate Volume: 2 Leak rate unit: mbar l/s Mode: Vacuum Trigger level 1: 1E-9 mbar l/s Trigger level 2: 1E-8 mbar l/s Trigger level 3: 1E-7 mbar l/s Leak rate external test leak (vacuum): 1E-7 mbar l/s Leak rate external test leak (sniffer): 1E-5 mbar l/s Vent delay: 2 seconds Automatic purge (PHOENIX L300i DRY and PHOENIX L300i MODUL only) on Pressure: mbar Minimum volume: 0 Beep: on Maximum evacuation time: 30 minutes Audio Alarm Type: Trigger Alarm Max. pressure limit for sniff mode 0.15 mbar Min. pressure limit for sniff mode 0.05 mbar Control location local Alarm delay: 30 seconds Leak rate filter: auto Zero: enable Vacuum ranges normal Upper display limit 1E-5 mbar l/s Service message exhaust oil filter (PHOENIX L300i only) on Contamination protection Off Switch-off limit for contamination protection: (Limits: 1E-6 mbar*l/s … 1E+3 mbar*l/s) 1E-3 mbar*l/s GA10218_002_C0 - 10/2016 - © Leybold GmbH Installation 2 Installation 2.1 Placement Unpack the PHOENIX L300i immediately after delivery, even if it will be installed later on. Examine the shipping container for any external damage. Completely remove the packaging materials. Check the PHOENIX L300i is complete (see Chapter Supplied Equipment) and carefully examine the PHOENIX L300i visually. If any damage is discovered, report it immediately to the forwarding agent and insurer. If the damaged part has to be replaced, please contact the orders department. Retain the packaging materials in the case of complaints about damage. 2.2 Conforming Utilisation The PHOENIX L300i is a leak detector for Helium or Hydrogen. This instrument may be used to detect the location and the size of leaks on objects under test in two different ways: when the test sample has been evacuated first and is sprayed with helium on the outside. It is required that a vacuum connection is provided between the PHOENIX L300i and the test sample (vacuum mode). or when a helium overpressure is provided in the test sample and the test sample is searched from the outside with a sniffer probe which is attached to the inlet port (sniffer mode). The PHOENIX L300i is to be used for leak detection only. It must not be used as a pumping system (esp. pumping aggressive or humid gases.) The leak detector is not suitable for - pumping liquids or gases containing dust or particles - pumping corrosive or reactive gases 2.3 Ambient Conditions The permissible ambient temperature is between +10°C (50°F) and +40°C (104°F). Ambient temperatures The PHOENIX L300i must not be operated in explosive gas atmospheres. Make sure to avoid dripping water. Ensure a sufficient air cooling. 2.4 Electrical Connections 2.4.1 Mains Power Generally the local regulations for electrical connections must be observed. Warning Before connecting the PHOENIX L300i to the mains you must make sure that the mains voltage rating of the PHOENIX L300i coincides with the locally available mains voltage. The instrument must exclusively be connected to a single phase supply with fuses for installation (Circuit breaker 10A max. according to IEC/EN 60898 with tripping characteristic B). GA10218_002_C0 - 10/2016 - © Leybold GmbH 27 Installation Only 3-core mains cables having a protection ground conductor must be used. Operation of the PHOENIX L300i where the ground conductor has been left unconnected is not permissible. The PHOENIX L300i can be damaged when using the wrong voltage. The voltage must be in the range 230V (+/- 5%), 115V (+/- 5%) or 100V (+/- 5%) depending on the version. The mains voltage rating for the PHOENIX L300i can be read off from the name plate beneath the mains socket Fig. 9/4 at the side. This voltage is fixed and can not be changed. A separate fuse for each of the mains conductors has been integrated into the mains switch. The mains voltage is applied to the instrument via the detachable mains cable which is supplied with the instrument. A mains socket Fig. 9/4 is available for this purpose at the side of the instrument. Left side: 1. VENT 2. EXHAUST 3. GASBALLAST 1 5 6 4. On / Off switch, mains socket 7 2 Right side: 8 3 5. SERIAL 4 6. CONTROL2 7. RECORD 9 8. CONTROL 9. OPTION 10. REMOTE1 11 11. REMOTE2 12 12. CHARGE (iPad) Fig. 9 Side views of the PHOENIX L300i 28 GA10218_002_C0 - 10/2016 - © Leybold GmbH 10 Installation 2.4.2 Connection for the Controller Signal and Accessories Option (Accessories) The sniffer line SL300 or the partial flow system may be connected to the option port (Fig. 9/9): OPTION Contact pins 1 and 3 are fused together with a 0.8 A slow-blow fuse. The amount of power which can be drawn is limited to 10 W. The contacts are numbered from bottom to top. Pin Assignment 1 +24 V, constantly applied, power supply for the Leybold partial flow valve or sniffer line. 2 GND 3, +24V switched by the PHOENIX L300i for an external venting valve 4, 5, 6, 7, 8 These pins are used in connection with accessories. Digital Out (CONTROL) The following relay outputs Fig. 9/8 are available for further signal processing. The maximum rating for the relay contacts is 60V AC/1A. All pins of digital I/O, digital out and recorder must not be connected with voltages higher than 60V DC/25V AC (to grounding equipment conductor) or reach this threshold. Digital out (CONTROL) Caution The contacts are numbered from bottom to top. Pin Assignment 1 PLC in free selectable 2 PLC in free selectable 3 PLC in free selectable 4 GND 5 to 7 Digital out free selectable, 5 center contact, 6 normally open contact, 7 normally closed contact 8 to 10 Digital out free selectable 11 to 13 Digital out free selectable 14 to 16 Digital out free selectable The pin assignment for contacts 8 to 16 follows the same order as for pins 5 to 7. For further information see chapter Interfaces. GA10218_002_C0 - 10/2016 - © Leybold GmbH 29 Installation Digital In (CONTROL2) Digital In (Control 2) These inputs can be used to control the PHOENIX L300i with a programmable logic control (PLC). The contacts are numbered from bottom to top. Caution Maximum input voltage 35V. Pin Assignment 1 PLC in free selectable 2 PLC in free selectable 3 PLC in free selectable 4 PLC GND These inputs Fig. 9/6 are working only, if the correct location of control is chosen. See chapter Interfaces. To avoid a mistake between the connection Control 2 and Record, pin 1 and 4 are blocked. When using the connectors the guiding nose for pin 1 and 4 must be removed. Recorder/RECORD Recorder/RECORD The recorder output Fig. 9/7 may be used to chart the leak rate, the inlet pressure and the fore vacuum pressure. Both recorder activities can be adjusted individually for showing leak rates and pressures. The measured values are provided by way of an analogue signal in the range of 0 V … 10 V. The resolution is limited to 10 mV. The instrument which is connected to the recorder output (e. g. X(t) chart recorder) should have an input resistance of no less than 2.5 kW. The measured values are available through pins 1 and 4. The reference potential (GND) is available at pins 2 and 3. The contacts are numbered from bottom to top. The chart recorder outputs are electrically isolated from other plugs. If, in spite of this, hum interference is apparent it is recommended to operate the PHOENIX L300i and the chart recorder from the same mains phase. If this is not possible, you must make sure that the frame ground of both instruments is kept at the same potential. Pin Assignment 1 Analog 1, leak rate, inlet pressure P1 or fore vacuum pressure P2 2 GND 3 GND 4 Analog 2, leak rate, inlet pressure P1 or fore vacuum pressure P2 For further information see chapter Interfaces. 30 GA10218_002_C0 - 10/2016 - © Leybold GmbH Installation SERIAL/RS232 This RS232 interface Fig. 9/5 is wired as data communication equipment (DCE) and permits the connection of a personal computer (PC) for monitoring and data logging. The connection is made through a 9 pin sub-D socket. For more information refer to chapter Interfaces and the Interface Description. Pin Assignment 1 24 V switchable, default setting 0 2 TXD 3 RXD 4 GND 24V 5 GND 6 DSR 7 CTS 8 RTS 9 free Remote Control (REMOTE1) The remote control interface Fig. 9/10 is a serial interface to control the PHOENIX L300i by the remote control. The remote control can be connected via an extension cable with a RJ45 plug. Refer to the Interface Description for more information. The remote control does not belong to the standard equipment. If the remote control is connected via a cable, wireless communication over REMOTE2 is excluded. Pin Assignment 2 +24V (fuse 0.8 A time lag) 3 0V 4 RXD (intern. RS232) 5 TXD (intern. RS232) REMOTE2 Through this interface the PHOENIX L300i is controlled wirelessly via Bluetooth or WLAN. The Bluetooth transmitter connects to the remote control RC310 WL. The WLAN module provides the connection to Apple I-Pad or other handheld devices. Is the remote control RC310 WL connected via a cable to REMOTE1 REMOTE2 can not be used. The Bluetooth transmitter and the WLAN module are not included PHOENIX L300i delivery. SERIAL/RS 232 Remote control REMOTE1 Remote control REMOTE2 GA10218_002_C0 - 10/2016 - © Leybold GmbH 31 Installation CHARGE Pin Assignment 1 free 2 RxD 3 TxD 4 connected to pins 6 and 7 5 GND 6 connected to pins 4 and 7 7 connected to pins 4 and 6 8 free 9 7.5V CHARGE The USB interface can be charged an Apple iPad only. The USB port does not allow data exchange. 2.4.3 Inlet port Vacuum Connections Inlet Port The inlet port is located on the top of the PHOENIX L300i Fig. 1/1. The size of the flange is DN 25 KF. A test object or a test chamber has to be connected to the inlet port if the vacuum mode is chosen (See Chapter Mode). The inlet port is also used for the connection of the sniffer line. Exhaust Exhaust The exhaust Fig. 9/2 flange is located on the side of the PHOENIX L300i. There is a filter mounted in the exhaust that absorbs the oil steams occur i ng during the use of the rotary vane pump. The exhaust filter has to be cleaned when doing the maintenance (see Chapter.Exhaust Oil Filter). When the PHOENIX L300i is running in closed rooms the exhaust has to be put out-of-doors using the provided adapter. So the oil steams that are harmful to health are lead off. With the provided connection a hose line can be connected to the exhaust of the PHOENIX L300i and lead off. Vent Vent Usually the parts under test are vented with ambient air when the test is finished. If it is required the parts can be vented with a different gas (i. e. fresh air, dry air, nitrogen, …) at atmospheric pressure. In this case a vent hose has to be connected to the hose coupling Fig. 9/1. The pressure in the venting line must nor exceed 1050 mbar. Gas ballast connection Gas ballast connection For the mode gas ballast it is recommended to use helium-free gases at atmospheric pressure. Ambient air can be contaminated with helium due to spraying or 32 GA10218_002_C0 - 10/2016 - © Leybold GmbH Installation charging. In this case a gas supply line (i. e. nitrogen, fresh air, …) should be connected to the hose coupling Fig. 9/3. The pressure of these gas line must not exceed 1050 mbar. The connectors 1,2 and 3 in Fig. 9 are quick connectors for hose diameters of 8/ 6 mm. Connection of an external pump (only PHOENIX L300i MODUL) The PHOENIX L300i MODUL offers two possibilities to connect the external fore vacuum pump to the DN 25 KF flange. One on the side of the PHOENIX L300i or one in the bottom (measurements see Fig. 5 - Fig. 7). As default setting the flange on the side is chosen. To change the connection proceed as follows: Connection of external pumps 1. Take of the mechanical hood, see chapter Opening of the PHOENIX L300i. 2. Loose the flange with the connection piece on the side of the PHOENIX L300i. 3. Disconnect the blind flange on the bottom, therefore lay the PHOENIX L300i carefully on the electronic hood. 4. Screw in the connection piece into the flange in the bottom. 5. Connect the hose for the fore vacuum pump. 6. Connect the blind flange to the sidewise flange. 7. Put on the mechanical hood. GA10218_002_C0 - 10/2016 - © Leybold GmbH 33 Operation 3 Purge gas/Gas ballast Operation 3.1 Media Compatibility/Purge Gas The PHOENIX L300i is a leak detector for helium and hydrogen. Only air and clean gases must be used with the PHOENIX L300i. The leak detector is not suitable for - pumping liquids or gases containing dust - pumping reactive or corrosive gases As purge gas all gases can be used that - does not contain helium - are dry, clean and dust free - generate no corrosion. For venting or gas ballast a helium free gas at atmospheric pressure should be used. Ambient air can be contaminated with helium due to spraying or charging, so it is recommended to connect a hose to the vent- and fore vacuum port. The pressure in this hose must not exceed 1050 mbar. 3.2 Start-Up The PHOENIX L300i is switched on by pushing the mains switch (Please refer to Chapter 2.4.1). After about 2 minutes the run-up procedure is finished; the unit is in standby-mode and ready to measure. When using the PHOENIX L300i MODUL an additional fore vacuum pump (dry or wet version) has to be connected to the fore vacuum connection (DN25 KF) on the side or the bottom. Please connect the part to be tested to the inlet port and press START. The PHOENIX L300i starts to evacuate the part. The evacuation time depends on the volume of the test part. During evacuation the screen shows the inlet pressure online. Once the pressure of 15 mbar (11 Torr or 1500 Pa) is reached the unit switches to measurement mode. The corresponding leak rate is displayed. For further explanations of the screen please refer to Chapter 3.2.5. The displayed leak rate corresponds to the helium background concentration in the part under test. Since the PHOENIX L300i continues to pump down the part this background leak rate will further reduce. As soon as the leak rate is low enough in respect to your requirements you may start spraying Helium or Hydrogen to search for possible leaks. When the measurement is finished please press STOP and hold the button a few seconds to vent the part under test. 3.2.1 Display The display is used to either show leak rates or program specific set-ups and get information by means of the software menu (Please refer to Chapter 3.6.1). In addition messages and maintenance instructions are displayed on the screen (Please refer to Chapter 5). 34 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation 3.2.2 The Display in Run-Up Mode In run-up mode the display shows: Speed of the turbomolecular pump Fore-vacuum pressure State of emission Active filament A bar graph which shows the run-up progress If the display is too bright or too dark you can change the contrast. Please see Chapter 3.4.6. During run-up phase the menu button can be pushed to get to the selection menu. 3.2.3 The Display in Standby Mode In standby mode the conditions are shown in the upper edge of the display (Fig. 12). Furthermore calibration (Please refer to Chapter 3.13) can also be started in standby mode and purging, too (Fig. 1/1) 3.2.4 Gas Ballast/Purge In standby mode the gas ballast of the forepump can be switched on/off manually or via soft key 7. The gas ballast is for abolishing a too huge helium background. Additionally a condensation of water vapor in the pump will be avoided. After 20 minutes the machine closes the gas ballast valve automatically to limit the loss of oil. This function can be chosen automatically for the PHOENIX L300i MODUL. Every time the unit changes into standby mode the purge starts automatically for 20 seconds. During this time the scroll pump will be purged by the valve V6. In case there was a large quantum of water vapor pumped with the machine please activate the gas ballast for about 20 minutes before running the machine down. 3.2.5 The Display in Measurement Mode In measurement mode the leak rates can be displayed in two different modes: - Numerically, combined with a bargraph 10. Fig. 10. - As trend: numerically, combined with a diagram (leak rate versus time) Fig. 11. In the lower right corner of the display (next to the Soft Key no. 8, Fig. 10 and Fig. 11) you will find a symbol that allows to switch between the display modes by pressing Soft Key no. 8. See chapter 3.2.9 and 3.2.10 for explanations of the different display modes. Access to calibration (Soft Key no. 5, Fig. 10 and Fig. 11) and access to the speaker volume (Soft Keys no. 2 and no. 3, Fig. 10 and Fig. 11) is the same in all modes. Also the status icons in the upper line are in common in both display modes. GA10218_002_C0 - 10/2016 - © Leybold GmbH 35 Operation Fig. 10 Display: Measurement mode with bargraph 3.2.6 Call for Calibration In all modes the soft key no.5 is used to get to the calibration routine. Refer to Chapter. 3.13for further information regarding calibration. 3.2.7 Speaker Volume On the left hand side two loud speaker symbols are shown, combined with the signs + and - (Fig. 10 and Fig. 11). By pressing the corresponding soft keys (Soft Keys no. 2 and no. 3) the volume can be adjusted for convenient loudness. In the bottom line of the display another loud speaker symbol is shown, combined with a number. This number indicates the level of the current loudness (ranges from 0 to 15). Refer to Chapter 3.8 Volume for information on loudness, alarms, and sound tracks. 3.2.8 Status Line in the Display The status line at the top of the display (Fig. 10 and Fig. 11) informs about (reading from left to right): Symbol of display Meaning 36 Explanation Volume level Please refer to Chapter 3.8 Speaker volume. S1 Trigger 1 If the trigger values are exceeded these signs are inverted. (White on black background.) S2 Trigger 2 see: Trigger 1 S3 Trigger 3 see: Trigger 1 ! Warning triangle Please refer to Chapter 5 VAC Working mode VAC or SNIFF indicate which working mode was selected GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation Symbol of display Meaning Explanation FINE Vacuum area Depending on the inlet pressure the PHOENIX L300i may be in GROSS, PRECISION (PHOENIX L300i DRY only) or FINE mode, which is indicated here (Chapter 3.10.1) ZERO ZERO Indicates if ZERO-function is active. 3.2.9 Measurement Mode with Bargraph The display shows the leak rate in big digital figures, see Fig. 10. The unit of the leak rate is shown, too. Underneath the leak rate the inlet pressure is displayed in smaller digits. The units of leak rate and pressure can be defined in the menu (See Chapter 3.8.4). Below this the same leak rate is shown graphically as a bar. The scale of this bar, i.e. the number of decades included in this bar can be defined in the menu (Please refer to Chapter 3.6.4). The programmed trigger levels (Please refer to Chapters 3.8.1 and 3.8.2) are indicated at the bar by short vertical lines: a straight line for trigger 1 and a dotted line for trigger 2. 3.2.10 Measurement Mode with Trend Information In trend mode the leak rates are displayed over time Fig.11. In addition the actual leak rate and inlet pressure also are displayed digitally. The time axis can be defined in the menu (Please refer to Chapter 3.6.5). The intensity axis (y-axis) is defined the same way as the bargraph (Please refer to Chapter 3.6.4). Fig. 11 Display: Measurement mode with bargraph 3.3 First Operation Check The steps for an initial operation are described here. It is explained how to switch on the PHOENIX L300i, how to measure and how to carry out an internal calibration. First operation check If anything unexpected happens during the initial operation or the leak detector acts in a strange way the PHOENIX L300i can be switched off by the mains switch at any time. GA10218_002_C0 - 10/2016 - © Leybold GmbH 37 Operation Needed equipment Needed Equipment The following parts will be needed: A blind flange 25 KF (if not assembled at the inlet port). A helium test leak with a DN 25 KF connection (optional). A fore vacuum pump connected to the DN25 KF flange on the side or under the bottom (dry or wet version) for use with the PHOENIX L300i MODUL Startup and measure Startup and Measure 1. Unpack the PHOENIX L300i and inspect it for any external damage (refer to Chapter 2.1). 2. Connect the instrument to the mains power (refer to Chapter 2.4). For the PHOENIX L300i MODUL connect the fore vacuum pump and switch it on. 3. Switch on the PHOENIX L300i by using the mains switch. Caution Don’t switch the PHOENIX L300i on when ambient temperature is below 10°C or above 40°C After power on a welcoming picture appears on the screen of the control panel Fig. 12, the status information on the speed of the turbo pump, the fore vacuum pressure, the emission and the active filament are given. The start up procedure takes less than 2 minutes and the end is indicated by a signal. The PHOENIX L300i is in standby mode now.(Fig. 12) 4. Check if the inlet port (Fig. 1/1) is blanked off. If not, please mount a blind flange with o-ring on the inlet port. 5. Press the START button. The inlet will be evacuated and if the inlet pressure drops below 15 mbar a measured leak rate will be displayed. 6. Press the STOP button, the PHOENIX L300i will go to standby. If you press STOP a few seconds the inlet of the PHOENIX L300i will be vented. 7. To finish the startup procedure please proceed with step 21. For calibration proceed with step 8. Internal calibration Internal Calibration 8. Proceed the internal calibration (Please refer to Chapter). For better quantitative measurements please let the unit warm up (15 … 20 minutes). Press Calibration (Soft Key no. 5 Fig. 12 /5) to get into the calibration menu. Select internal (Soft Key no. 4, Fig. 12 /4) to choose the internal calibration. The internal calibration starts automatically and takes about 30 seconds. After a successful calibration a visual and audible signal comes up. 9. Press the STOP button Fig. 12 /12 until the message STANDBY/VENTED appears on the display. The inlet is vented now. Verification with external test leak Verification with an external test leak To verify the accuracy please proceed through the following steps. A test leak is required. If a test leak is not available please continue with step 21. 10. Remove the blind flange from the inlet port and connect the open helium test leak to the inlet port. 38 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation 11. Press the START button Fig. 12 /10. The inlet will be evacuated and the leak rate of the test leak will be measured and displayed. 12. Press the STOP button Fig. 12 /12 to stop the measurement. The PHOENIX L300i goes into standby mode. 13. Press the STOP button Fig. 12 /12 again until the message STANDBY/VENTED appears on the display. The inlet is vented now. 14. Remove the helium test leak from the inlet port and put a blind flange onto the inlet port again. Measure with a test object 15. Remove the blind flange from the inlet port and connect the test object to the inlet port Measure with a test object 16. Press the START button Fig. 12 /10. The test object will be evacuated. 17. Start spraying Helium onto the outside of the test object. The leak rate of the test object will be shown in the display. 18. Press the STOP button Fig. 12 /12 to stop the measurement. The PHOENIX L300i goes into standby mode. 19. Press the STOP button Fig. 12 /12 again until the message STANDBY/Vented appears on the display. The inlet is vented now. 20. Remove the test object and put on a blind flange on the inlet port. Switch off 21. Switch off the PHOENIX L300i if the unit is in STANDBY or VENTED mode by using the mains switch Fig. 9/4. Switch off 3.4 Control Panel The Control panel Fig. 12 contains a liquid crystal display (LC Display), the START, STOP, ZERO and MENU buttons and eight soft Keys for the different menus and inputs selections. 1 2 9 5 10 6 3 7 4 8 11 12 13 Fig. 12 Control panel GA10218_002_C0 - 10/2016 - © Leybold GmbH 39 Operation Pos. Description Pos. Description 1 2 3 4 5 6 7 8 9 10 11 12 13 Soft Key no. 1 Soft Key no. 2 Soft Key no. 3 Soft Key no. 4 Soft Key no. 5 Soft Key no. 6 Soft Key no. 7 Soft Key no. 8 LC Display START ZERO STOP MENU LC-display LC Display The LC Display Fig. 12 /9 is the communication interface to the operator. It displays the leak rates, the status report of the PHOENIX L300i, messages, warnings and errors. With the soft keys no.1 to no. 8 various functions which are shown in the display can be selected Start button START Button Pushing the START Button Fig. 12/10 enables the PHOENIX L300i to start the measure procedure. The measured leak rate is shown in the display. If the START button is pushed again in measurement mode, the maximum leak rate indicator („hold“ function) is activated. This indicator shows the maximum leak rate since „START“. By pressing the START-button again the „hold“ function will be started again. Stop button STOP Button Pushing the STOP Button Fig. 12/12 interrupts the measure procedure. If the button is pressed longer the inlet is vented according to the conditions defined in the menu Vent delay. See Chapter 3.10.1 to select the time parameters of the venting. ZERO button ZERO Button Pushing the ZERO button Fig. 12/11 enables the zero mode. (see also Chapter 3.10.2) When pressing ZERO the currently measured leak rate is taken as a background signal and is subtracted from all further measurements. As a result the displayed leak rate then is 1·10-8 mbarl/s 1·10-12 mbarl/s in GROSS in FINE After pressing ZERO (Fig. 13, t=1) the decreasing background is fitted to the course (Fig. 13, t=2) automatically. When the measurement signal declines below the saved background the underground value will automatically be equated with the measurement signal. As soon as the measurement signal is increasing again the saved decreasing value remains constant. Increasing of the signal are displayed clearly as a leak So it is possible to recognize leaks even when the signal is decreasing rapidly. 40 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation Fig. 13 Zero activation Fig. 14 Undo zero When you want to see the measurement signal (including underground) please press the ZERO button again. The saved value will be reset to zero. The underground signal will not be suppressed anymore (Fig. 14). MENU Button When pressing the MENU button (Fig. 12/13) the selecting menu is shown at the display. This function is not depending on the operating mode when calibrating. Menu button If the menu is opened during the current session the operator will lead to the last screen before the menu was left. Pushing the MENU button again leads back to the screen of the previous working mode. The software shows the last screen that was used before. GA10218_002_C0 - 10/2016 - © Leybold GmbH 41 Operation Softkeys Soft Keys The function of the eight Soft Keys Fig.12/1…8 depends on the current menu. Special functions Special Functions When inputs are allowed or when settings can be selected in a submenu two of the Soft Keys always have the same function: Soft Key no. 1 Fig. 12/1 is Cancle. It allows to escape from the submenu without any changes of the present settings and return to the previous menu page. Soft Key no. 8 Fig. 12/8 is OK. The selected settings or edited values will be stored and the previous menu page will be displayed again. Numerical entries Numerical Entries If you have opened a menu page where a digit can be changed please proceed in the following way: If you don’t want to change anything, press Soft Key no. 1 Cancle. If you want to change the digit please proceed as follows: 1. The digit that can be changed is displayed inverted. With the arrows (Soft Key no. 8) and (Soft Key no. 4) you can choose which digit you want to change. 2. To change a digit to a specific number press the corresponding pair of numbers. A submenu opens and the desired number can be selected. The submenu closes automatically and the next digit of the total number now is inverted. 3. Having reached the last digit all changes have to be confirmed by OK (softkey no. 8). To correct a wrong entry press Cancel (softkey 1) or soft key 4 and enter the desired value again. Example change trigger Example To change the trigger level 1.0 · 10-7 mbar l/s to 3 · 10-7 mbar l/s please press 2/3 (Soft Key no. 3, Fig. 16) . Please consider that the first digit is displayed inverted. If not please change the digit with (softkey no. 8) or (softkey no. 4). With the soft key no. 4 (Fig. 12) the chosen value can be selected. 42 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation Fig. 15 Numerical entry of the trigger level, sample of the digit In the submenu press 3 (soft key no.4) Fig. 16 Fig. 16 Trigger level, change of the first digit GA10218_002_C0 - 10/2016 - © Leybold GmbH 43 Operation 3.5 Interfaces Under Main menu > Settings > Interfaces the parameters for the interface can be set. Softkey 2: Location of control The location of control for the leak detector can be defined Softkey 3: Define recorder output Customer defined selection for the recorder output Softkey 4: RS232 Selection for the RS232 Softkey 5: Define PLC outputs (Control, digital out) Customer defined selection for PLC outputs Softkey 6: Define PLC inputs (Control 2, Digital in) Customer defined selection for PLC inputs Softkey 7: Scaling recorder outputs Selection for the scaling of the recorder output Softkey 8: PLC sample rate Selection of the PLC sample rate 44 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation Location of Control Main menu > Settings > Interfaces > Location of control Softkey 2: Location of control PLC The PHOENIX L300i is controlled via the Digital In connector. The START, STOP and ZERO buttons at the control panel and remote control are locked. Softkey 3: RS232 The PHOENIX L300i is controlled via RS232 interface by an external computer. In this mode the PHOENIX L300i can not be controlled via keyboard. The START, STOP and ZERO button at the machine are deactivated. Softkey 4: All The PHOENIX L300i is controlled via all possible controls, e.g. PLC, RS232, Local. Softkey 5: Local & PLC The PHOENIX L300i is controlled via the Digital In connector or/and the START, STOP and ZERO buttons at the control panel and remote control. Softkey 6: Local & RS232 The PHOENIX L300i is controlled via the Digital In connector or/and the START, STOP and ZERO buttons at the control panel and remote control. Softkey 7: Local The PHOENIX L300i is controlled via the START, STOP and ZERO buttons at the control panel or remote control. This is the default setting. Define recorder output Main menu > Settings > Interfaces > Define recorder output Define recorder output The signals to be recorded can be selected in this submenu. With the left keys the pin can be selected, with the right keys a function is assigned to the selected pin. The recorder output has 2 channels (Fig. 18) 0 0,45 0,52 0,97 2,6 5,97 10 -3 10 -2 10 -1 10 0 8,2 8,65 10 1 10 2 10 3 10 -4 10 -3 10 -2 10 -1 10 0 10 1 10 2 -4 7,5E -3 7,5E -2 7,5E 0,75 7,5 75 750 V Recorder output pirani P1/P2 mbar,atm Pa Torr Fig. 17 Recorder output: Pirani PHOENIX L300i, P1 and P2. The complete characteristic Pirani line is shown in the appendix. GA10218_002_C0 - 10/2016 - © Leybold GmbH 45 Operation Selection recorder outputs The following functions can be selected: Off The recorder output is switched off P1 Pirani PHOENIX L300i The inlet pressure P1 of the PHOENIX L300i will be shown logarithmic. Fig. 17 P2 Pirani PHOENIX L300i The fore vacuum pressure P2 of the PHOENIX L300i will be shown logarithmic. Fig. 17 P1 (L200) The setting for the inlet pressure P1 is identical with those of the L200, i.e. 1000 mbar comply 4V with 0.5V/decade logarithmic scale. Users can exchange their L200 for a PHOENIX L300i without changing pin assignment. Fig. 18 P2 (L200) The setting for the fore vacuum pressure P2 is identical with those of the L200,i.e. 1000 mbar comply 4V with 0.5V/decade logarithmic scale. Users can exchange their L200 for a PHOENIX L300i without changing pin assignment. Fig. 18 LR mantissa The leak rate mantissa is recorded linearly from 1...10V. Fig. 20 LR exponent The exponent is recorded as step function: U = 1...10V with steps of 0,5V per decade, starting with 1V = 1·10-12 mbarl/s. Fig. 19 LR linear There will be two decades, the lower decade with voltages from 0.1 to 1.0V, the upper from 1.0 to 10.0V output. The scale is linear. The "Zoom recorder output (leak rate) upper limit ( = 10V)" specified exponent determines the upper decade. LR log Recorder output P1/P2 L200 46 The fundamental output voltage is scaled logarithmic and can be freely chosen. The voltage output ranges from 1 to 10 V with adjustable steps of 0,5/1/2/2,5/5 to 10 V per decade (Fig. 21 shows the default setting) GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation 0 1 2 3 4 5 10 -3 10 -1 10 +1 10 +3 10 10 -6 10 -4 10 -2 10 6 7 8 9 +5 10 V Recorder output leak rate exponent mbar,Pa,Torr -0 atm Fig. 18 Recorder output: P1 and P2 L200 0 0,5 1 2 -13 -12 10 10 10 -10 3 4 5 6 7 8 9 10 10 -8 10 -6 10 -4 10 -2 10 0 10 +2 10 +4 V mbarl/s, Pam•/s Torrl/s, sft•/yr oz/yr, g/a Recorder output leak rate mantissa Fig. 19 Recorder output: Leak rate exponent 0 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 10 V Recorder output leak rate logarithmic Mantissa Fig. 20 Recorder output: Leak rate mantissa 0 1 2 3 4 5 6 7 8 9 10 V 10 -14 10 -13 10 -12 10 -11 10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 mbarl/s, Pam•/s Torrl/s, sft•/yr oz/yr, g/a Fig. 21 Recorder output: Leak rate logarithmic, default setting GA10218_002_C0 - 10/2016 - © Leybold GmbH 47 Operation RS232 RS232 Main Menu > Settings > Interfaces > RS232 Softkey 2: Baudrate and end sign Settings for the baudrate selectable between 1200 and 19200, as well as endsign between CR+LF, CR or LF. Softkey 3: Data, Parity, Stop bits The settings for Data (7 or 8), Parity (Even, Odd, None) and Stop bits (1 or 2) can be selected Softkey 7: RS 232 Protocol The protocol from the RS 232 can be chosen between: L200 size, Diagnostic, L200 Leak Ware and ASCII code. The calibration function of Leak Ware is not suitable with the use of a Phoenix. Define PLC outputs Define PLC outputs Main Menu > Settings > Interfaces > Define PLC outputs The following relay outputs are available for further signal processing. The maximum rating for the relay contacts is 60V AC/1A. The contacts are numbered from bottom to top. Caution All pins of digital I/O, digital out and recorder must not be connected with voltages higher than 60V DC/25V AC (to grounding equipment conductor) or reach this threshold. Pin Assignment 1 PLC in free selectable 2 PLC in free selectable 3 PLC in free selectable 4 GND 5 to 7 Digital out free selectable, 5 center contact, 6 normally open contact, 7 normally closed contact 8 to 10 Digital out free selectable 11 to 13 Digital out free selectable 14 to 16 Digital out free selectable Description of the operation mode of the Digital Out. The pin assignment for contacts 8 to 16 follows the same order as for pins 5 to 7. The actual pin setting can be seen under Info / View internal data / page 7. The following digital out signals are selectable. Trigger 1; Trigger 2 and 3 analog Trigger 1 Is open in case Trigger Level 1 is exceeded or the machine is not in condition of 48 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation measuring. Zero active: Is closed in case Zero function is running. Ready: Is closed in case machine is in measurement mode (Emission on, no error). CAL active Closed when machine is in calibrating routine. CAL Request Is opened in case of calibration request. During external calibration a open output indicates that the external calibrated leak has to be closed. Fail Open when a error is shown. Warning Open when a warning is shown. Gas ballast Closed when gas ballast is active. Open Open all time. Close Closed all time. Recorder Strobe Closed in case recorder output is invalid. Only used when record output is set on „leak rate“. Pump down Open when machine is evacuating the test object. Standby Open when machine is in Standby or Vent mode. Vented Open when machine is in Vent mode. Emission on Open when emission is on. Define PLC inputs Main menu > Settings > Interfaces > Define PLC inputs These inputs can be used to control the PHOENIX L300i with a programmable logic control (PLC). Maximum input voltage 35V. Pin Assignment 1 PLC in free selectable 2 PLC in free selectable Define PLC inputs Caution GA10218_002_C0 - 10/2016 - © Leybold GmbH 49 Operation Pin Assignment 3 PLC in free selectable 4 PLC GND Description of operation mode of the Digital In. The contacts are numbered from bottom to top. The PLC inputs are working only if the correct location of control has been set. The actual pin setting can be seen under Info / View internal data / page8. Zero: Change from low to high: activate zero Change from high to low: deactivate zero Start: Change from low to high: activate START Stop: Change from low to high: activate STOP When this inlet is longer high than chosen then ventilate it additionally. Purge/gas ballast: Change from low to high: activate purge/gas ballast Change from high to low: deactivate purge/gas ballast Clear: Change from low to high: confirm error message CAL: Change from low to high: When machine is in standby mode: start internal calibration. In case machine is measurement mode: start external calibration. (Premise: external calibration test leak has to be open and leak rate signal is stable) Change from high to low: External calibration: approve that external test leak is closed and leak rate signal is stable. High means: U > 13V(approximately 7mA) Low means: U < 7V The level of the logic signals must not exceed 35V. CAL intern: Change from low to high: The Machine starts an internal calibration independent from the mode the machine is running in. CAL extern: Change from low to high: The Machine starts an external calibration independent from the mode the machine is running in. Change from high to low: approve that external test leak is closed and leak rate signal is stable. Signals at these inputs are only accepted if the location of control is set to „PLC“, “All“ or „Local and PLC“. 50 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation Scaling Recorder Output Main menu > Settings > Interfaces > Scaling recorder output Scaling recorder output Here the scaling of the recorder output can be adjusted. This adjustment is possible only when the signal LR lin or LR log is chosen. Softkey 2: The decade of the upper leak rate can be decreased Softkey 3: Decrease scaling of the previously adjusted value in steps of 0.5, 1, 2, 2.5, 5, 10 Volt/decade. The complete array covers 10 V. Softkey 5: ? Help text. Softkey 6: The decade of the upper leak rate can be increased Softkey 7: Increase scaling of the previously adjusted value in steps of 0.5, 1, 2, 2.5, 5, 10 Volt/decade. The complete array covers 10 V. Example: Upper limit value is adjusted to 10-5 (= 10V) Scaled to 5 V /decade Lower limit value consequently is 10-7 (= 0 V) PLC Sample Rate Main menu > Settings > Interfaces > PLC Sample rate Softkey 2: Decreasing the PLC sample rate down to the minimum of 10 ms. This might be necessary if exchanging an L200 to the PHOENIX L300i to stay compatible. Softkey 3: Increasing the PLC sample rate to the maximum of 100 ms. 3.6 Operation By pressing the MENU push button Fig. 12/13 the main menu will be displayed regardless of the current working mode or status of the PHOENIX L300i. The main menu Fig. 23 leads the operator to several submenus described in the following chapters. The main menu is identically for all PHOENIX L300i models. The next page gives an overview of the entire menu architecture Fig. 23. The overview of the menu architecture corresponds to the PHOENIX L300i, differing menu points or setting possibilities for the PHOENIX L300i DRY and PHOENIX L300i MODUL are described in the respective menu point. GA10218_002_C0 - 10/2016 - © Leybold GmbH 51 Operation 3.6.1 Main Menu The main menu Fig.23 shows 7 sub-menus. In these sub-menus groups of technical features are put together logically. From here the next levels of the menu tree can be reached. All following chapters show the path to get to the described menu line right underneath the headline. Fig. 22 Display: Main menu Explanation to Fig.22: 52 Key No. Name Description 1 Back Return to the previous screen. 2 View Display settings like scaling, contrast, system background. Please refer to Chapter 3.6.2. 3 Mode Selection of the working modes Vacuum or Sniff Please refer to Chapter 3.7. 4 Trigger & Alarms Settings of units, trigger levels and alarms. Please refer to Chapter 3.7. 5 Calibration Calibration of the PHOENIX L300i. Please refer to Chapter 3.13. 6 Settings Settings of internal machine parameters. Please refer to Chapter 3.10. 7 Information Information on the PHOENIX L300i (electrical and vacuum data) and service menu. Please refer to Chapter 3.11. 8 Access Control Access restrictions. Please refer to Chapter 3.12. GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation 1. Level 2. Level View Scale linear / logarithmic 3. Level Display range Time axis Contrast Background in standby Lower display limit Mode Trigger & Alarms Sniff/Vacuum Trigger Level 1 Trigger Level 2 Trigger Level 3 Units Volume Alarm delay Audio alarm type Calibration internal external Settings Vacuum settings Vent delay Vacuum ranges Partial flow setup/pump setup Sniffer factor Machine factor Leak rate internal test leak Purge in measurement Filter & Background Background suppression Calculate inlet area background Main Menui Leak rate filter Mass Interfaces Location of control Define recorder output RS232 Define PLC outputs Define PLC inputs Scaling recorder output PLC sample rate Miscellaneous Time & Date Language Calibration request Service internal forepump Service internal exhaust oil filter Service message exhaust oil filter Parameter save / load Load parameter set Monitoring functions Pressure limits for sniff mode Maximum evacuation time Pressure limits for vacuum mode Information View settings View internal data Vacuum diagram View error list Calibration history Calibration factors Service Access Control Access to CAL function Change Device-PIN Change Menu-Pin Zero Fig. 23 Overview menu structure GA10218_002_C0 - 10/2016 - © Leybold GmbH 53 Operation 3.6.2 View Main menu > View In this menu Fig.24 all features that influence the data display are put together. Fig. 24 Display: View Menu Explanation to Fig.24: Key No. Name Description 1 Back Return to the main menu. 2 Scale linear/logarithmic Settings for bargraph and trend mode. Please refer to Chapter 3.6.3 3 Display range auto/ manual Manual or automatic scaling. Please refer to Chapter 3.6.4 4 Time axis Time axis in trend. Please refer to Chapter 3.6.5 5 Contrast Display contrast. Please refer to Chapter 3.6.6 6 Background in standby Background displayed or not. Please refer to Chapter 3.6.7 8 Lower display limit 3.6.3 Setting of the display limit. Please refer to Chapter 3.6.8 Scale Linear/Logarithmic Main menu > View > Scale linear/logarithmic These settings apply to the bargraph (= bar underneath the digital figures in the measurement mode) and Y-axis in the trend diagram. The scale of the bargraph can either be linear or logarithmic. With the arrows ( and ) it can be determined how many decades the bargraph and Y-axis are covered. Usually a logarithmic scale is recommended because leak rates may change easily over several decades. Default setting is logarithmic with 4 decades. 54 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation Softkey 2: Linear Pressing this key switches the display to a linear scale, starting at zero. Softkey 3: (Number of decades) Pressing this key reduces the number of displayed decades. The minimum value is 2 decades. Only available if log (softkey 6) was chosen. Softkey 6: Logarithmic The scaling will be displayed logarithmically. Softkey 7: (Number of decades) Increase the number of displayed decades. Maximum value is 9 decades. Only available if log (softkey 6) was chosen. 3.6.4 Display-Range Auto/Manual Main menu > View > Display range auto/manual The upper limit of the displayed leak rate range can be set manually or automatically. These settings apply to the bargraph (= bar underneath the digital figures in the measurement mode and y-axis in the trend mode). With the upper limit defined here the lower limit is set to a value based on the number of decades. Softkey 2: Manual The upper limit of the displayed range can be set manually. Softkey 3: Decrease the upper limit if manual is chosen. The minimum value is 10-11 mbar l/s Softkey 5: ? Help text Softkey 6: Automatic The limit of the displayed range will be chosen automatically. Softkey 7: Increase the upper limit if manual is chosen. The maximum value is 10+3 mbar l/s Softkey 8: Save the settings and return to the previous menu. If linear scale is selected, the lower limit is always zero. The upper limit is only a default value. You can change this on the measurement screen with the Soft Key 6 and 7 if you have chosen manual display ranging. GA10218_002_C0 - 10/2016 - © Leybold GmbH 55 Operation 3.6.5 Time Axis Main menu > View > Time axis The length of the time axis in trend mode can be changed in given steps between 16 … 960 seconds. Softkey 3: Decrease the length of the time axis. The minimum value is 16 seconds. Softkey 5: ? Help text Softkey 7: Increase the length of the time axis. The maximum adjustable value is 960 seconds. 3.6.6 Contrast Main menu > View > Contrast The contrast of the display can be changed. The recommended value under regular conditions is about 50 (Default setting). Softkey 3: Fade the contrast to dark. The minimum values is 0. Softkey 4: Invert display Invert the contrast of the screen, that means background dark and font bright. Softkey 5: ? Help text Softkey 7: Fade the contrast to light. The maximum value is 99. If by accident the display has been set too bright or too dark so that it can not be read off, this may be changed as follows: Switch off the PHOENIX L300i and turn it on again. During the run-up phase press the key no. 3 or 7 so long until the display can be read properly again. This setting is saved to the EPROM only after confirming this through the contrast menu. If this setting is not confirmed, the former setting will be applied after switching on the instrument on again. 56 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation 3.6.7 Background in standby Main menu > View > Background in standby The internal background leak rate can be displayed in standby mode or not.The default setting is OFF. Softkey 3: Off The background leak rate will not be shown. Softkey 5: ? Help text. Softkey 7: ON The background leak rate will be shown. The internal background is generated by residual gas (e. g. helium) that has not been pumped away yet. Sources for residual gas are air or absorbed gases from the inner surfaces of the PHOENIX L300i. This internal background will never disappear totally. Very clean systems which have been pumped for a long time will show a background in the 10-11 mbar l/s range. Under normal conditions the background level is in the 10-10 mbar l/s or low 10-9 mbar l/s range. When pressing START the current internal background is subtracted from all further measured signals automatically. Thus it is made sure that only the net leak rate from the part under test is measured. When switched to standby / Vent again a new internal background is calculated after 25 s. The updated value is underlined. This means that if you press START when the value is underlined, the actual background signal will be subtracted. If you press START when the value is not underlined, the old background signal from the last standby will be subtracted. 3.6.8 Lower Display Limit Main menu > View > Lower display limit This mode limits the lower detection limit of the measured leak rate. This is valid for vacuum mode only. Softkey 3, 7: Changing of the lower detection limit between 1 · 10-9 and 1 · 10-12 mbarl/s. The lower limit for the PHOENIX L300i DRY ranges between 1 · 10-9 and 1 · 10-11 mbarl/s. Softkey 5: ? Help text GA10218_002_C0 - 10/2016 - © Leybold GmbH 57 Operation 3.7 Mode Main menu > Mode The mode menu Fig.25 enables the submenu to select the different working modes. Fig. 25 Display: Mode menu Explanation to Fig.25: 58 Key No. Name Description 1 Cancel Return to the main menu without any changes of the present settings. 3 Sniff The sniffer mode will be used. 7 Vacuum The vacuum mode will be used. 8 OK Save the settings and return to the previous menu. GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation 3.8 Trigger & Alarms Main menu > Trigger & Alarms The trigger levels, the volume of the loudspeaker and the units of leak rates and pressures can be set in this menu Fig.26. Fig. 26 Display: Trigger & Alarms Menu Explanation for Fig.26: Key No. Name Description 1 Back Return to the main menu. 2 Trigger level 1 Definition of Trigger level 1. Chapter 3.8.1 3 Trigger level 2 Definition of Trigger level 2. Chapter 3.8.2 4 Trigger level 3 Definition of Trigger level 2. Chapter 3.8.3 5 Units Selection of leak rate and pressure units. Refer to Chapter 3.8.4 6 Volume Refer to Chapter 3.8.5 7 Alarm delay Refer to Chapter 3.8.6 8 Audio alarm type Choice of different alarm types. Refer to Chapter 3.8.7 3.8.1 Trigger Level 1 Main menu > Trigger & Alarms > Trigger level 1 The value of the first trigger level can be set. See Numerical entries for the description of the entry. Trigger 1, 2 and Trigger 3 are programmable switching thresholds. When these thresholds will be exceeded the PHOENIX L300i reacts as follows: Display In the status line of the display the signs for Trigger 1, 2 and Trigger 3 are displayed inverted if the leak rate exceeds (becomes higher than) the programmed value. (see Fig. 10) GA10218_002_C0 - 10/2016 - © Leybold GmbH 59 Operation Relay Output The trigger-relais of the digital out switches. Please refer to Chapter 3.5 Digital out, for further details. Alarm/Loudspeaker Additionally Trigger level 1 defines at which level the various alarm types react (See Chapter 3.8.7) 3.8.2 Trigger Level 2 Main menu > Trigger & Alarms > Trigger level 2 The value of the second trigger level can be set. Please refer to Numerical entries for the description of the entry. If Trigger 2 is exceeded the corresponding relay will switch. This is also indicated at the display (see Chapter 3.8.1). 3.8.3 Trigger Level 3 Main menu > Trigger & Alarms > Trigger level 3 The value of the third trigger level can be set. Please refer to Numerical entries for the description of the entry. If Trigger 3 is exceeded the corresponding relay will switch. This is also indicated at the display (see Chapter.3.8.1). 3.8.4 Units Main menu > Trigger & Alarms > Units The preferred leak rate unit can be selected. There is the choice of 4 (mbar, Pa, Torr, atm) pressure units and 5 leak rate units (mbar l/s, Pa m3/s, Torr l/s, atm cc/ s, s ft3/yr). In Sniff mode the following measuring units are selectable: ppm, g/a eq (helium leak rate is equivalent with leak rate R134a), oz/yr eq (helium leak rate is equivalent with leak rate R134a). Softkey 2: Scroll up to select a pressure unit. Softkey 3: Scroll down to select a pressure unit. Softkey 6: Scroll up to select a leak rate unit. Softkey 7: Scroll down to select a leak rate unit. 60 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation 3.8.5 Volume Main menu > Trigger & Alarms > Volume The minimum loudness and the actual volume of the loudspeaker can be adjusted. The minimum loudness is the minimum speaker volume that cannot be exceeded to even lower values. Thus it is avoided that the actual volume is accidentally adjusted to a value that is below the noise level of the environment. The actual volume can be adjusted between 15 (maximum) and the value defined as minimum loudness. Softkey 2: Decrease the minimum loudness. The minimum value is 0. Softkey 3: Decrease the actual volume. The minimum value is limited by the minimum volume. Softkey 4: Beep off / Beep on Softkey 5: ? Help text. Softkey 6: Increase the minimum volume. The maximum value is 15. Softkey 7: Increase the regular volume. The maximum value is 15. 3.8.6 Alarm Delay Main menu > Trigger & Alarms > Alarm delay In some applications (for instance during pump down in a „chamber test system“) it might be necessary to block an alarm for some time after pressing START. This delay time of the alarm can be changed. Softkey 3: Decrease the delay time. The minimum value is 0 seconds. Softkey 5: ? Help text. Softkey 7: Increase the delay time. The maximum value is 10 minutes up to infinity. After pressing START the loudspeaker is activated as soon as the leak rate drops below trigger level 1 or after the entered alarm delay time has elapsed. This setting is only active for the audio alarm types SETPOINT and TRIGGER ALARM (See chapter 3.8.7). GA10218_002_C0 - 10/2016 - © Leybold GmbH 61 Operation 3.8.7 Audio Alarm Type Main menu > Trigger & Alarms > Audio alarm type The audio alarm type can be chosen. Softkey 2: Pinpoint This function is for localization of a known leak rate value. Softkey 3: Leak rate prop. The sound will be proportional to the leak rate signal. Softkey 5: ? Help text Softkey 6: Setpoint The sound will be proportional to the leak rate signal only if trigger 1 is exceeded. Softkey 7: Trigger alarm An alarm sounds when the trigger 1 is exceeded. Pinpoint Pinpoint The tone of the acoustical signal changes its frequency only in a leak rate-window Fig. 27 which ranges from one decade below the Trigger level 1 up to one decade above the Trigger level 1. Below the window the tone is constantly low, above the window it is constantly high. Example: The Trigger level 1 is 4·10-7 mbar l/s. So the window where the tone changes reaches from 4·10-8 mbar l/s up to 4·10-6 mbar l/s. f 2570 Hz - 491 Hz - 0.1 • Tr. 1 Tr. 1 10 • TR. 1 LR Fig. 27 Pinpoint: Change of the frequency in the leak rate window (TR1=Triggerlevel 1) Leak rate prop. 62 Leak rate prop. The frequency of the acoustic output is proportional to the reading on the bargraph display. The frequency ranges from 300 Hz to 3300 Hz. Please refer to Chapter 3.6.3 for the definition of the number of decades. GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation Setpoint The tone is off as long as the leak rate is below the Trigger level 1. Above Trigger 1 the tone varies proportional to the leak rate Fig. 28. Setpoint f 3000 Hz - 375 Hz - no tone Tr. 1 1000 • TR.1 LR Fig. 28 Setpoint: Change of the frequency above the trigger level 1 (TR1= Trigger level 1) Trigger alarm As soon as the leak rate increases above trigger level 1, a multi-tone signal is generated. The tone does not vary with the leak rate. 3.9 Trigger Alarm Calibration Main menu > Calibration In the menu Calibration Fig. 29 the selection between internal and external calibration can be chosen. Please refer to Chapter 3.13 Calibration for a detailed description of the calibration. Fig. 29 Display: Calibration menu GA10218_002_C0 - 10/2016 - © Leybold GmbH 63 Operation 3.10 Settings Main menu > Settings This menu Fig. 30 allows to observe and to change the adjustment of the internal machine settings. Fig. 30 Display: Settings menu Explanation for Fig.30: 64 Key No. Name Description 1 Back Return to the main menu. 2 Vacuum settings Settings of vacuum system related functions. See chapter 3.10.1 3 Filter & Background See Chapter 3.10.2 4 Mass Switching between Helium and Hydrogen. See Chapter 3.10.3 5 Interfaces Settings for electrical communication and selection for control location See Chapter 3.5 for detailed information. 6 Miscellaneous Settings like date or time. See Chapter 3.10.4 7 Parameter save / load Store and load sets of parameters. See Chapter 3.10.5 8 Monitoring functions GA10218_002_C0 - 10/2016 - © Leybold GmbH Choose functions of protection of the PHOENIX L300i in this mode. See Chapter 3.10.6 Operation 3.10.1 Vacuum Settings Main menu > Settings > Vacuum settings This menu allows to observe and to change the settings belonging to the vacuum system. Softkey 3: Vacuum settings PHOENIX L300i Vent delay Definition of delay time until the PHOENIX L300i is vented Softkey 4: Vacuum ranges Selection of the vacuum working modes Softkey 5: Partial flow setup/Forepump setup Selection of partial flow mode and forepump type Softkey 6: Sniffer factor Setting for the sniffer factor Softkey 7: Machine factor Setting for the machine factor Softkey 8: Leak rate internal test leak Setting for the internal test leak The menu for the PHOENIX L300i DRY version allows the following vacuum settings which varies to the PHOENIX L300i: Softkey 3: Vacuum settings PHOENIX L300i DRY Vent delay Definition of delay time until the PHOENIX L300i is vented Softkey 4: Vacuum ranges Selection of the vacuum working modes Softkey 5: Purge in measurement The purging during measurement mode can be switched on Softkey 6: Sniffer factor Setting for the sniffer factor Softkey 7: Machine factor Setting for the machine factor Softkey 8: Leak rate internal test leak Setting for the internal test leak GA10218_002_C0 - 10/2016 - © Leybold GmbH 65 Operation Vacuum settings PHOENIX L300i DRY MODUL The menu for the PHOENIX L300i MODUL version allows the following vacuum settings which varies to the PHOENIX L300i: Softkey 2: Automatic purge Definition of automatic purge function in standby Softkey 3: Vent delay Definition of delay time until the PHOENIX L300i is vented Softkey 4: Vacuum ranges Selection of the vacuum working modes Softkey 5: Partial flow setup/Forepump setup Selection of partial flow mode and forepump type Softkey 6: Sniffer factor Setting for the sniffer factor Softkey 7: Machine factor Setting for the machine factor Softkey 8: Leak rate internal test leak Setting for the internal test leak Automatic purge Automatic purge Main menu > Settings > Vacuum settings > Automatic purge Through this menu is it possible to start automatic purge for 20 seconds automatically. This setting is only possible for the PHOENIX L300i MODUL (refer to vacuum settings PHOENIX L300i MODUL) Softkey 3: Off The function automatic purge is off. Softkey 6: ON The function automatic purge is on. When changing from measure to standby the forevacuum pump will be purged automatically for 20 seconds. Vent delay Vent delay Main menu > Settings > Vacuum settings > Vent delay Through this menu item it is possible to define the delay time until the inlet port is vented when operating the STOP button. When the STOP button is pressed for a period of time which is shorter than the delay time specified here, the PHOENIX L300i will just change to standby mode. When the STOP button is pressed for a period of time which is longer than the delay time specified here, the PHOENIX L300i will vent the inlet port. 66 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation Softkey 2: Immediately The inlet port will be vented immediately after pressing the STOP button. Softkey 3: After 1 second The inlet port will be vented with a time delay of 1 second. Softkey 4: After 1.5 seconds The inlet port will be vented with a time delay of 1.5 second. Softkey 5: ? Help Softkey 6: after 2 seconds The inlet port will be vented with a time delay of 2 second. Softkey 7: No venting The inlet port cannot be vented with the STOP button. Vacuum ranges Main menu > Settings > Vacuum settings > Vacuum ranges With this menu you can adjust different modes concerning the activity of leak detection. This setting is only active in vacuum mode (see Chapter 3.5). Vacuum ranges PHOENIX L300i DRY/PHOENIX L300i MODUL Softkey No. 2: GROSS only In this mode the PHOENIX L300i remains at the inlet flange after falling below 15 mbar. When the pressure is increasing over 15 mbar the PHOENIX L300i switches automatically into evacuation mode. The smallest detectable leak rate is 1 ·10-8 mbarl/s. Softkey No. 3: FINE only In this mode the PHOENIX L300i remains after falling below 0,2 mbar at the inlet flange. Valve V1 will be closed. When the pressure at the inlet flange is increasing > 0.2 mbar the PHOENIX L300i switches immediately into evacuation mode. The advantage of FINE only is that while this mode is running no valve will switch and the PHOENIX L300i has a high pumping speed. Softkey No. 4: GROSS only 920 Hz In this mode the turbo pump of the PHOENIX L300i works at reduced speed. Therefore the smallest detectable leak rate is 3 ·10-9 mbarl/s. This function can be used to replace the L200 in Gross only mode equivalent Taste No. 5: ? Help text Softkey No.6 Partial flow enable GA10218_002_C0 - 10/2016 - © Leybold GmbH 67 Operation If the PHOENIX L300i or the PHOENIX L300i MODUL is used with a partial flow system this vacuum mode must be enabled before. The setup for the partial flow mode is described in partial flow setup Softkey No. 7: Normal (default settings) This is the default setting. The activity runs as explained in Chapter 1.1.1. Vacuum ranges PHOENIX L300i DRY The PHOENIX L300i DRY allows the vacuum ranges as follows: Softkey No. 2: GROSS only In this mode the PHOENIX L300i DRY remains at the inlet flange after falling below 15 mbar. When the pressure is increasing over 15 mbar the PHOENIX L300i DRY switches automatically into evacuation mode. The smallest detactable leak rate is 1 ·10-8 mbarl/s. Softkey No. 3: FINE only In this mode the PHOENIX L300i DRY remains after falling below 0,1 mbar at the inlet flange. Valve V1 will be closed. When the pressure at the inlet flange is increasing > 0.1 mbar the PHOENIX L300i DRY switches immediately into evacuation mode. The advantage of FINE only is that while this mode is running no valve will switch and the leak detector has a high pumping speed. Softkey No. 4: GROSS only 920 Hz In this mode the turbo pump of the PHOENIX L300i works at reduced speed. Therefore the smallest detectable leak rate is 3 ·10-9 mbarl/s. This function can be used to replace the L200 in Gross only mode equivalent Taste No. 5: ? Help text Softkey No. 6: Precision In this mode the PHOENIX L300i DRY achieves the maximum sensitivity. Softkey No. 7: Normal (default settings) This is the default setting. The activity runs through the vacuum ranges from GROSS to FINE. Partial flow setup/Pump setup Partial flow setup/pump setup Main menu > Settings > Vacuum settings > Partial flow setup/pump setup Through this menu item the settings for a use of a partial flow system can be set. In the partial flow mode the test sample is additionally evacuated by an auxiliary pump, which offers the advantage of measuring from 1000 mbar on. Enable partial flow mode Before setting up the parameters the partial flow mode has to be enabled in the main menu under Settings > Vacuum settings > Vacuum ranges with softkey no. 6 partial flow enable and confirm with the soft key OK. This setting is not possible for the PHOENIX L300i DRY version. 68 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation The PHOENIX L300i MODUL allows additionally the settings for the forepump (oil sealed or dry) and selectable pumping speed for the forepump. Partial flow setup for the PHOENIX L300i. Softkey 2: Partial flow setup PHOENIX L300i The entry of the nominal pumping speed of the partial flow pump can be decreased. The minimum pumping speed is 4m3/h. Softkey 3: Decrease Quick-pump time. The quick-pump time defines whether and how long valve V10 of the partial flow block is opened. (For detailed descriptions please refer to the operating instructions “GA 10.277” of the partial flow system.) At TQ = 0 seconds valve V10 will not be open for the time being. This selection is recommended for large volumes and dirty objects. At TQ = endless valve V10 will open when pressing start. At an inlet pressure p1 < 15 mbar the PHOENIX L300i DRY switches to measurement mode and display leak rates. This is recommended if it is acceptable to wait for a while until measurement mode is reached and leak rate reading at high inlet pressures are not needed. With times between 0 and endless V10 is opened and the leak detector tries to reach a inlet pressure of less than 15 mbar within this time TQ. When TQ has gone by V10 is closed and the PHOENIX L300i switches to measurement mode (Helium/Hydrogen comes through the orifice of the partial flow valve block). Softkey 4: Changing behavior of the valve V8 of the partial flow system Closed: In partial flow mode valve V8 (see GA.10.277 partial flow system) switches dependent on the inlet pressure Open: Valve V8 stays open, even when the inlet pressure is low enough Softkey 5: ? Help text. Softkey 6: Increase the pumping speed of the partial flow pump. The entry of the nominal pumping speed of the partial flow pump can be increased. The maximum pumping speed is 80 m3/h. Default setting is 25 m3/h. Softkey 7: Increase of the quick-pump time up to the maximum. GA10218_002_C0 - 10/2016 - © Leybold GmbH 69 Operation Partial flow setup PHOENIX L300i MODUL Partial flow setup/pump configuration for the PHOENIX L300i MODUL: Before setting up the parameters the partial flow mode has to be enabled in the main menu under Settings > Vacuum settings > Vacuum ranges with softkey no. 6 partial flow enable and confirm with the soft key OK. Softkey 2: Pump setup Setting for the forepump if the PHOENIX L300i MODUL is operated with a partial flow system. Softkey 7: Partial flow setup Options for setting up the partial flow system. Vacuum settings PHOENIX L300i DRY MODUL Forepump setup for PHOENIX L300i MODUL Softkey 2: The entry of the nominal pumping speed of the partial flow pump can be decreased. The minimum pumping speed is 4 m3/h. Softkey 3: Forepump type The PHOENIX L300i MODUL with partial flow system can be operated with a dry forepump (for ex. Scroll pump) or a wet forepump (oil sealed). This key is for choosing a dry (Scroll, piston) fore vacuum pump. Softkey 5: ? Help text Softkey 6: The entry of the nominal pumping speed of the partial flow pump can be increased. The maximum pumping speed is 80m3/h. Softkey 7: Forepump type The PHOENIX L300i MODUL with partial flow system can be operated with a dry (for ex. Scroll pump) forepump or a wet pump (oil sealed). This key is for choosing a wet (oil sealed) fore vacuum pump. 70 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation Partial flow setup for the PHOENIX L300i MODUL Softkey 2: Partial flow setup PHOENIX L300i MODUL The entry of the nominal pumping speed of the partial flow pump can be decreased. The minimum pumping speed is 4m3/h. Softkey 3: Decrease Quick-pump time. The quick-pump time defines whether and how long valve V10 of the partial flow block is opened. (For detailed descriptions please refer to the operating instructions “GA 10.277” of the partial flow system.) At TQ = 0 seconds valve V10 will not be open for the time being. This selection is recommended for large volumes and dirty objects. At TQ = endless valve V10 will open when pressing start. At an inlet pressure p1 < 15 mbar the PHOENIX L300i switches to measurement mode and display leak rates. This is recommended if it is acceptable to wait for a while until measurement mode is reached and leak rate reading at high inlet pressures are not needed. With times between 0 and endless V10 is opened and the leak detector tries to reach a inlet pressure of less than 15 mbar within this time TQ. When TQ has gone by V10 is closed and the PHOENIX L300i switches to measurement mode (Helium/Hydrogen comes through the orifice of the partial flow valve block). Softkey 4: Changing behavior of the valve V8 of the partial flow system Closed: In partial flow mode valve V8 (see GA.10.277 partial flow system) switches dependent on the inlet pressure Open: Valve V8 stays open, even when the inlet pressure is low enough Softkey 5: ? Help text. Softkey 6: Increase the pumping speed of the partial flow pump. The entry of the nominal pumping speed of the partial flow pump can be increased. The maximum pumping speed is 80 m3/h. Default setting is 25 m3/h. Softkey 7: Increase of the quick-pump time up to the maximum. Sniffer factor Main menu > Settings > Vacuum settings > Sniffer factor Sniffer factor The sniffer factor takes into account, after an internal calibration, an external partial flow ratio, for example the Quicktest or a auxiliary pump with sniffer line connected via T-peace to the leak detector. During an internal calibration the internal sensitivity of the PHOENIX L300i is calibrated. The calculated number is multiplied with the sniffer factor and the result is the sniffer factor for this application. GA10218_002_C0 - 10/2016 - © Leybold GmbH 71 Operation Softkey 4: Set default value Setting between the default value 1 for the sniffer line SL300 or the correction factor (1000) for the use of the Quicktest. Quicktest setting Machine factor For the use of a PHOENIX L300i with a Quicktest the setting for the Quicktest under Main menu > Settings > Monitoring functions > Pressure limits for sniff mode has to be chosen. Machine factor Main menu > Settings > Vacuum settings > Machine factor The machine factor takes into account, after an internal calibration, the ratio between the effective helium pumping rate of the PHOENIX L300i and the pumps in the pump system in measurement mode as well as the measurement mode used (GROSS/FINE).Based on an internal calibration only, all measured leak rate would be measured too small. The measured leak rate is multiplied with the machine factor and the result is displayed. This factor is only used for vacuum measurement modes (not for sniff mode). See Numerical entries for the description of the entry. Since the effective pumping rates are usually not known due to the conductances of the vacuum connections, we recommend the following indirect measurement: 1. Set up the PHOENIX L300i for operation 2. First an internal calibration must be performed with machine factor = 1 (Refer to Chapter 7) 3. Connect an external calibrated leak (for example 2.0·10-6 mbar l/s) to the test chamber 4. Measure leak rate of the external test leak, for example 5.0·10-8 mbar l/ s 5. The machine factor is the quotient of the desired value and the actual value. Desired value: 2.0·10-6 mbar l/ s / 5.0·10-8 mbar l/ s = machine factor 40 6. Set the acquired value in the menu point 7. Calibrate again internally so that the machine factor is taken over 8. All signals that are measured in further measurements are multiplied by factor 40 and then shown in the display Leak rate internal test leak Leak rate internal test leak Main menu > Settings > Vacuum settings > Leak rate internal test leak The value of the internal test leak can be set. See Numerical entries for the description of the entry. Normally there is no reason to edit the leak rate of the internal test leak besides after a change or a recertification of the internal test leak. A wrong leak rate of the internal test leak will lead to wrong leak rate readings! Purge in measurement 72 Purge in measurement Main menu > Settings > Vacuum settings > Purge in measurement This function is only possible for the PHOENIX L300i DRY. In vacuum mode the GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation fore vacuum pump is purged constantly to avoid helium accumulation. If the use of the PHOENIX L300i DRY needs the possibility to shut off this function, fore ex. because of a high helium background in the ambient and no option to connect a hose line with fresh air to the gas ballast port, this function can be shut of here. If you disable the purge mode in measurement some specifications as minimum detectable leak rate, internal background or time constant may degrade. Softkey 7: Enable (default) The fore vacuum pump is purged constantly during the measurement mode to avoid helium accumulation. This setting is the preferred setting and should be used. Softkey 3: Disable This setting enables the purging during measuring mode, all disadvantages listed above may occur. It is recommended to use this setting only in special applications. 3.10.2 Filter & Background Main menu > Settings > Filter & Background The type of leak rate filters and background condition can be chosen. The default setting for the leak rate filter is auto. Softkey 2: Calculate inlet area background This function is for assignation of the background in the inlet area. Softkey 3: Background suppression Setting of the internal condition for the background. Softkey 7: Leak rate filter The type of leak rate filter can be chosen. The PHOENIX L300i DRY allows the following settings Softkey 3: Background suppression Setting of the internal condition for the background. Softkey 7: Leak rate filter The type of leak rate filter can be chosen. Calculate inlet area background Main menu > Settings > Filter & Background > Calculate inlet area background This function calculates the background of the inlet area. The PHOENIX L300i has to be in the following conditions: Calculate inlet area background 1. Mode vacuum 2. Mode VENTED (min. 25 seconds) GA10218_002_C0 - 10/2016 - © Leybold GmbH 73 Operation 3. Inlet port blanked off 4. Minimum 20 minutes since power on After starting this function the leak detector starts with evacuating the inlet area. Earliest two minutes after start the measured value can be accepted as „Background inlet area“. This value will be saved. Background suppression Background suppression Main menu > Settings > Filter & Background > Background suppression Softkey 3: Off Deactivation of the offset function. Under certain circumstances a positive leak rate can be displayed. This setting should be used by experienced users only because of the high possibilities of measuring wrong leak rates. Softkey 6: inlet area Additionally to the internal offset (background) the offset of the inlet area will be subtracted. This function for the inlet area is only possible in standby mode, therefore this value has to be determined with the menu point „Calculate inlet area background“. Softkey 7: internal only (default) With start the PHOENIX L300i defines the internal offset (background) and subtracts this value from the leak rate signal, so that just the leak rate is shown in the display. This setting should be used as standard setting for the PHOENIX L300i. Leak rate filter Leak rate filter Main menu > Settings > Filter & Background > Leak rate filter Softkey 3: Fixed A leak rate filter with a fixed time constant is used Softkey 6: Auto Auto makes sure, that the signals are averaged in optimized time intervals, based on the leak rate intensity. Auto also eliminates noise peaks that are not related to leak rate signals and provides extraordinary short response times for low leak rate signals. This setting should be used for the PHOENIX L300i. 74 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation 3.10.3 Mass Main menu > Settings > Mass The requested mass of the measured gas can be selected. The PHOENIX L300i must be in standby mode for changing to another mass. Softkey 2: H2 (2 amu) Hydrogen with the mass of 2 amu will be measured. Softkey 3: 3He (3 amu) Isotope of helium with the mass of 3 amu will be measured. Softkey 7: 4He (4 amu) Helium with the mass of 4 amu will be measured. Default setting After changing the mass a calibration for the selected mass should be done. See chapter 3.13. 3.10.4 Miscellaneous (Language, Calibration request, Service interval...) Main menu > Settings > Miscellaneous The actual date and time, the preferred language and the mains frequency can be set in this submenu. Softkey 2: Time & Date Setting of time and date Softkey 3: Language Selection of the language Softkey 4: Calibration request Setting if the PHOENIX L300i should remind for a calibration Softkey 5 Service interval forepump Setting service time interval forepump Softkey 7: Service interval exhaust oil filter Setting service time interval exhaust oil filter Softkey 8: Service message exhaust oil filter Setting service message exhaust oil filter Time & Date Main menu > Settings > Miscellaneous > Time & Date Time & Date Time and date can be changed on two subsequent pages. Please refer to Numerical entries for the description of the entry. GA10218_002_C0 - 10/2016 - © Leybold GmbH 75 Operation Language Language Main menu > Settings > Miscellaneous > Language The preferred language can be selected. The default setting is english. The following languages can be chosen: English, German, French, Italian, Spanish, Chinese, Japanese, Polish, Russian Softkey 3: Scrolling down to select the language. Press OK to confirm the selected language. Softkey 7: Scrolling up to select the language. Press OK to confirm the selected language. Calibration request Calibration request Main menu > Settings > Miscellaneous > Calibration request It can be selected whether the operator is reminded of the fact that a calibration may has become necessary or not. The default value is off. Softkey 3: Off The calibration request will be switched off. Softkey 5: ? Helptext Softkey 7: ON The calibration request will be switched on. If the calibration request is switched on, a corresponding message will appear when 30 minutes have elapsed after power on or if the temperature of the PHOENIX L300i has changed by more than 5 °C (9 °F) since the last calibration. Service interval forepump Service interval forepump Main menu > Settings > Miscellaneous > Service interval forepump Setting for the service interval of the forepump. This setting depends on the use of the PHOENIX L300i but latest after 4000 running hours or one year the oil in the pump should be controlled. See also manual for the Trivac D2,5E which is included in the document folder. See chapter 4.4.3 also. This setting is possible for the PHOENIX L300i DRY only. Softkey 3: Ø The time for the service interval can be decreased in steps of 500 hours. Softkey 7: ¦ The time for the service interval can be increased in steps of 500 hours to the upper limit of 4000 hours. Service interval exhaust oil filter 76 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation Main menu > Settings > Miscellaneous > Service interval exhaust oil filter Here you can enter the service interval for the exhaust oil filter. This setting is only possible for the PHOENIX L300i. This setting depends on the use and application of your PHOENIX L300i and therefore no recommendations can be given (see chap. maintenance). Softkey 3: Service interval exhaust oil filter Decrease of the service interval steps of within 500 hours. The limit is 1000 hours Softkey 5: ? Help text Softkey 7: Increase of the service interval within steps of 500 hours. The limit is 4000 hours. Service message exhaust oil filter Main menu > Settings > Miscellaneous > Service message exhaust oil filter The exhaust oil filter must be maintained at regular intervals to ensure the correct function of the PHOENIX L300i. If the service message is activated, the PHOENIX L300i reminds you of the required maintenance. Service message exhaust oil filter This setting is only possible for the PHOENIX L300i. If the service messages are ignored and the exhaust is not replaced a risk for overheating the pump motor exists. Softkey 3: The service message for the oil filter can be reduced to the min. 1000 hours Softkey 7: The service message for the oil filter can be increased up to the max. 4000 hours. GA10218_002_C0 - 10/2016 - © Leybold GmbH 77 Operation 3.10.5 Parameter Save/Load Main menu > Settings > Parameter save load Enables to save and load individual settings or reload the default settings. Softkey 2 to 4: The names of the current values can be saved under a free selectable name. The saving of 3 different sets is possible. Softkey 5: load default values The factory settings will be loaded again. Softkey 6 to 8: One of three saved parameter sets can be loaded. Load parameter set Load parameter set Main menu > Settings > Parameter save load > Load parameter set Save the current parameter settings. Softkey 4: Edit a file name Rename the parameter set. Softkey 8: Save Save the edited parameter set. Save parameter set Save parameter set Main menu > Settings > Parameter save load > Save parameter set The settings of the selected saved parameter set will be displayed and can be reloaded. Softkey 6: Upward to the previous screen. Softkey 7: Downward to the next screen. 3.10.6 Monitoring Functions Main menu > Settings > Monitoring functions This Submenu explains the monitoring functions. Softkey No. 3: Maximum evacuation time Settings for the gross leak measurement Softkey No. 4: Contamination protection Set the switch off limit and enable the function Softkey No. 6: Pressure limits for vacuum mode Setting for the pressure limits between evacuation, Gross and Fine mode. 78 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation Softkey No. 7: Pressure limits for sniff mode Definition of the upper and lower limit of the sniffer pressure Maximum evacuation time Main menu > Settings > Monitoring functions > Maximum evacuation time Maximum evacuation time This menu item is used to define when the gross leak message is to occur. The gross leak detection process operates in two steps and the limits can be adapted as required. This menu item is particularly useful in series testing under the same conditions at all times. After pressing the start button the test sample is evacuated. If the pressure conditions (p1 < 100 mbar) are not attained, or if the pressure does not drop low enough within the periods of time specified here, the pump down process is terminated and the display will indicate a message (see Ch.Trouble shooting, W75 and W76). The periods which are selected in each case depend firstly on the desired reaction time for the gross leak message, and secondly on the volume of the test sample and the effective pumping speed. If the evacuation time was set to endless, the oil level of the mechanical pump should be checked more often. Softkey No. 2: Decreasing maximum evacuation time until p1 < 100 mbar. Within this period of time the inlet pressure at the test flange must have dropped below 100 mbar. The duration may be selected freely between 1 second and 9 minutes or can be set to endless. The default is 30 seconds. Softkey No. 3: Decreasing maximum time until measurement. Within the period of this time the status of measurement readiness must have been attained, i.e. the inlet pressure must have dropped below 15 mbar. The duration may be freely selected between 5 seconds and 30 minutes or can be set to endless. Softkey No. 5: ? Help text Softkey No. 6: Increasing maximum evacuation time until p1 < 100 mbar Softkey No. 7 Increasing maximum time until measurement. Contamination Protection Main menu > Settings > Surveillance > Contamination Protection Contamination Protection If enabled, this mode will cause the PHOENIX to close all inlet valves as soon as the measured leak rate exceeds the programmed limit. Once the START key is GA10218_002_C0 - 10/2016 - © Leybold GmbH 79 Operation pressed, the value must drop below the limit once as the function will otherwise be triggered when the alarm delay time has elapsed. This will prevent an excess amount of helium from entering the mass spectrometer. Consequently, the leak detection unit is prevented from becoming contaminated by helium. The helium entering the specimen can be pumped out using an external pump. If no pump is available, we recommend that you ventilate the specimen before continuing with the measurements. Softkey No. 1: Back Softkey No. 3: Off Softkey No. 4 Enter limit value Selection between air or Argon Softkey No. 5: ? Help text Pressure limits for vacuum mode Softkey No. 6: On Softkey No. 7: Off Pressure limits for vacuum mode. Main menu > Settings > Monitoring functions > Pressure limits for vacuum mode With this function the default settings for the pressure limits EVAC - GROSS and FINE can be changed. This might be necessary if other gases than air will be pumped by the PHOENIX L300i. The pressure signal from the gas dependent inlet pressure (P1) will dump false signals. With changing the pressure limits this performance will be adjusted. Softkey No. 2: Decrease change over threshold EVAC-GROSS Selectable between15-3 mbar (Default value 15 mbar) Softkey No. 3: Decrease change over threshold GROSS-FINE Selectable between 0,2-0,05 mbar (Default value 0,2 mbar). For the PHOENIX L300i DRY and the PHOENIX L300i MODUL the changeover is between 0,1-0,05 mbar (Default value 0,2 mbar). Softkey No. 4 Adjustment for ARGON Selection between air or Argon Softkey No. 5: Help text 80 GA10218_002_C0 - 10/2016 - © Leybold GmbH ? Operation Softkey No. 6: Increase change over threshold EVAC-GROSS Selectable between 3 - 15 mbar Softkey No. 7: Increase change over threshold GROSS-FINE Selectable between 0,05 - 0,2 mbar, respectively 0,05 - 0,1 mbar for PHOENIX L300i DRY and PHOENIX L300i MODUL Pressure limits for sniff mode Main menu > Settings > Monitoring functions > Pressure limits for sniff mode Pressure limits for sniff mode This function is automatically activated in sniff mode. The pressure limits define an upper and lower limit of the inlet pressure P1. If the pressure is not in this range error messages are generated: P > upper limit: Capillary broken P < lower limit: Flow through capillary too low (Capillary blocked) Softkey No. 2: Decreasing the maximum pressure, Upper limit is 0,15 mbar (default) Softkey No. 3: Decreasing the minimum pressure Softkey No. 4: Setting for Quicktest Setting for use with the Quicktest, upper limit is 0,05 mbar, lower limit 0 mbar. Pushing the button again for default setting. Softkey No. 6 Increasing the maximum pressure Softkey No. 7: Increase the minimum pressure, Lower limit is 0,05 mbar (default) GA10218_002_C0 - 10/2016 - © Leybold GmbH 81 Operation 3.11 Information Main menu > Information The Information Menu Fig.31 enables submenus to select different kinds of information belonging to the PHOENIX L300i. Fig. 31 Display: Information Menu Softkey 2: View settings The current settings will be displayed on 5 pages, e.g. trigger levels, test leak mass, date and time. Softkey 3: View internal data Information on measured internal data is provided on 10 screens. Softkey 4: Vacuum diagram The vacuum diagram of the PHOENIX L300i is shown. Here you can see which valves are opened or closed momentarily. See chapter 4.1.1 Softkey 5: View error list The list of occurred errors and warnings will be displayed. See chapter 8.2. Softkey 6: Calibration history The carried out calibrations will be listed. Softkey 7: Calibration factors The calibration factors for the different masses and the machine factor will be displayed. Softkey 8: Service 82 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation 3.11.1 Service Main menu > Information > Service With the main menu special functions can be accomplished (e. g. manual switching of the valves). The access to the service menu is protected by a PIN. This PIN is not communicated with the delivery of the leak detector but after an adequate service training. 3.12 Access Control Main menu > Access control With this menu you can deny or allow access to specific functions of the PHOENIX L300i. Fig. 32 Display: Access Control Menu Softkey 3 Zero Settings of the zero function Softkey 4: Access to CAL function Settings for restriction of the CAL function Softkey 8: Change Menu-Pin Access to menu pin can be restricted 3.12.1 Zero Main menu > Access control > Zero This setting enables (respectively disables) the ZERO button at the control panel. With „Zero at FINE“, the ZERO functions executes automatically as soon as the measuring range FINE is reached for the first time after START. In this mode the ZERO function also can be executed manually via the ZERO button. GA10218_002_C0 - 10/2016 - © Leybold GmbH 83 Operation Softkey 3: Disable ZERO button disabled Softkey 5: ? Help text Softkey 6: Zero at FINE When reaching the FINE mode and leak rate falls below trigger level 3 the ZERO function is started automatically. Softkey 7: Enable Softkey ZERO is selectable. Default setting. 3.12.2 Access to CAL Function Main menu > Access control > Access to CAl function It can be selected whether the access to the calibration menu is restricted or not. Softkey 3: Off The calibration function is only available at the main menu. If the MenuPIN (See Chapter 3.12.3) is activated you need this PIN to start a calibration. Default setting. Softkey 5: ? Help text Softkey 7: ON The calibration function is available at the main menu and in standby and the measure mode. 3.12.3 Change Menu-PIN Main menu > Access control > Change Menu-Pin The access to the menu can be restricted by entering or changing the personal identification number (PIN). No PIN will be checked if 0000 is entered. The default setting for the Pin is 0013. Please refer to Numerical entries for the description of the entry. 3.13 Calibration 3.13.1 Introduction The PHOENIX L300i can be calibrated in two different ways: Internal calibration by means of a built-in test leak or external calibration by means of an additional test leak which then is attached to the inlet port or the component under test. During the calibration procedure the mass spectrometer is tuned to the maximum 84 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation helium or hydrogen signal and this signal is referred to the known leak rate of the internal or external test leak. Although the PHOENIX L300i is a very stable instrument a calibration is recommended every day with heavy use, or before using the PHOENIX L300i from time to time, to make sure that ambient temperature changes or dirt or other impacts don’t adulterate the measurements. To get an optimized calibration the machine has to warm up at least 20 minutes before use, otherwise a warning will come up which might be ignored. 3.13.2 The calibration Routines The calibration routines can be started by pressing button CAL (Softkey 5) via 3 different locations: main menu (Fig. 23) standby mode (Fig. 12) measurement mode (Fig. 10) The access via standby mode or measurement mode can possibly be blocked (see Chapter 3.12.2). In this case the softkey is not labeled. Default: Access on. Once the calibration mode is activated the user must choose between an internal and an external calibration. Please press the corresponding Soft Key (Fig. 29). A calibration may be terminated at any time by pressing the Stop button or using the Soft Key no. 1 (Cancel) Fig. 34. 3.13.3 Internal Calibration Mass 4 must be selected (Default setting) Press Softkey no 4 Fig. 29 to start the calibration. Once this procedure is started the entire procedure is performed automatically. At the end (after about 25 s) a visual and audio signal is released. Thereafter the unit is ready for further use. GA10218_002_C0 - 10/2016 - © Leybold GmbH 85 Operation 3.13.4 External Calibration For an external calibration a test leak has to be attached to the part under test or the inlet port directly depending on the application. After External calibration (Fig. 29, Soft Key no. 8) has been chosen the following messages are displayed and the described actions are required: Make sure that the correct mass is selected. Make sure that the test leak is connected and opened. Check the leak rate printed on the test leak and compare it with the leak rate at the display. If the leak rates are not identical press Edit leak rate (Soft Key no. 4) and correct the value. If the leak rates are okay press START (Soft Key no. 8). Fig. 33 Display: External Calibration, Step 1 No action required. Fig. 34 Display: External Calibration, Step 2 The bargraph display shows a signal which must not vary much. If so please press OK (Soft Key no. 8). Fig. 35 Display: External Calibration, Step 3 86 GA10218_002_C0 - 10/2016 - © Leybold GmbH Operation No action required. Fig. 36 Display: External Calibration, Step 4 The bargraph display shows a signal which must not decrease any more. There might be a small fluctuation which is okay. If so please press OK (Soft Key no. 8). Fig. 37 Display: External Calibration, Step 5 When the signal is stable confirm with o.k. Fig. 38 Display: External Calibration, Step 6 The PHOENIX L300i shows the old and the new calculated calibration factor. Fig. 39 Display: External Calibration, Step 7 GA10218_002_C0 - 10/2016 - © Leybold GmbH 87 Operation Factor of calibration/Range of values Factor of calibration - Range of values To avoid a faulty calibration the factor of calibration is tested for plausibility at the end of the calibration routine. When the new factor of calibration is not considerable higher or lower (5mV). W32 Preamplifier temperature is too high (> 65°C). Ambient temperature is too high. W33 Preamplifier temperature is too low (< 2°C). Ambient temperature is too low. E34 Signal MVPZN on the MSV board is active. 24 V signal voltage is too low, U < 18.3 V. 24V voltage at MSV board is too low! Air filter dirty. Temperature sensor is faulty. – Fuse F1 on the MSV board has blown. – Reference voltage UREF on the MSV board XT7/1 is too high, U > 5 V. – DC/DC converter on the MSV board is defective. – 24V power supply voltage of the main power supply is defective or stressed to much. GA10218_002_C0 - 10/2016 - © Leybold GmbH 97 Troubleshooting No. Displayed Message Description and possible solutions E35 Anode-cathode voltage is too high! Anode-cathode voltage is > 130 V Anode-cathode voltage is too low. Anode-cathode voltage is < 30 V. E36 MSV board is faulty. MSV board is faulty. Fuse F4 on MSV board has blown. E37 Suppressor voltage reference value too high! Signal MFSZH on MSV board is active. Suppressor signal command variable is too high. – Suppressor voltage has a short circuit. – MSV is faulty. E38 Suppressor potential too high! Suppressor potential is higher than 363V. MSV board is faulty. E39 Suppressor potential is too low. Suppressor potential is lower than 297V. MSV board is faulty. E40 The anode potential exceeds its nominal value by over 10%! The actual anode potential exceeds its nominal value by 10%. The nominal value can be displayed in the service menu. MSV is faulty. MC 68 is faulty E41 The anode potential has dropped The actual anode potential has dropped below its nominal value by 10%. The below its nominal value by over nominal value can be displayed in the service menu. 10%! MSV is faulty. MC 68 is faulty. E42 Nominal value of the anode potential is too high! Signal MFAZH on MSV board is active. Anode voltage has been short circuited. Nominal value of the anode voltage is too high. Anode voltage is limited to about 1,200 V. E43 Cathode current is too high! MSV Cat-Heater >> Signal MPKZH on MSV board is active. Cathode current is too high, I > 3.6 A. MSV is faulty. E44 Cathode current is too low! Signal MPKZN on MSV board is active. Cathode current is too low, I < 0.2 A. MSV is faulty. Faulty ion source connector or cable. W45 Emission for cathode 1 can not be switched on! Signal MSIBE on MSV board is not active. Emission for cathode 1 can not be switched on. PHOENIX L300i switches to cathode 2. Please order a new ion source. Cathode 1 is defective MSV board is defective 98 GA10218_002_C0 - 10/2016 - © Leybold GmbH Troubleshooting No. Displayed Message W46 Emission for cathode 2 can can not be switched. Description and possible solutions Signal MSIBE on MSV board is not active. Emission for cathode 2 can not be switched on. PHOENIX L300i switches to cathode 1. Order a new ion source. Cathode 2 is defective. MSV board is defective. E47 Emission for both cathodes can not be switched on! Signal MSIBE on MSV board is not active. Emission can not be switched on. Exchange the cathode by changing the ion source. After having exchanged the ion source it must be possible to switch on both cathodes manually via the service menu. Replace ion source. MSV board is defective. E48 Anode heater is faulty! Signal MSAFD on MSV board is active. Anode heater fuse has blown. Replace fuse F2 on the MSV board. E50 No communication with turbo pump converter. Clock from the frequency converter has failed. No communication to the frequency converter. Fuse F4 on the wiring backplane has blown. Drive electronics Turbo Drive S is defective. E51 Unknown TMP error The frequency converter Turbo Drive S indicates an unknown error code. Inform Leybold service E52 TMP frequency is too low! TMP frequency is too low! Frequency converter is faulty. Turbo molecular pump is faulty. W53 Temperature at electronic unit is too high (>59°C) Ambient temperature too high. Ventilation failure. Air filter dirty and have to be changed. E54 Temperature at electronic unit is too high (>60°C). Ambient temperature is too high. Internal ventilation has failed. Air filters are dirty and must be exchanged. W55 Temperature at electronic unit is too low (< 2°C) The temperature sensor on the wiring plane indicates T < 2 °C. Run-up time for the fore vacuum pump will be longer. Temperature sensor is faulty. Ambient temperature is too low. E56 Inlet pressure p1 too low! Output voltage Pirani P1 U < 0,27 V. Pirani sensor P1 is defective. Pirani electronics on the I/O board is defective. E58 fore vacuum pressure p2 too low! Output voltage Pirani P2 U < 0,27 V. Pirani sensor P2 is defective. Pirani electronics on the I/O board is defective. GA10218_002_C0 - 10/2016 - © Leybold GmbH 99 Troubleshooting No. Displayed Message Description and possible solutions E60 p2>10mbar after 5 minutes since Run-up time of the fore vacuum pump is too long. power on Fore pump is faulty. Valve V2 does not open. Leak in the high vacuum system. E61 Emission fail. Emission could not be switched on. MSV subassembly indicates a fault. MENB emission current not within range. MSV board is defective. Both cathodes are defective, replace ion source. W62 Flow through capillary to low. In the sniffer mode the intake pressure of the sniffer line is controlled. If the pressure falls below the minimum limit, the flow through the capillary is too low (contamination) or the capillary is blocked (foreign objects, particles). The minimum limit can be set by the menu. Default value is 0.05 mbar. Filter in the tip is clogged. W63 Capillary broken In the sniffer mode the intake pressure of the sniffer line is controlled. If the pressure exceeds the maximum limit, the flow through the capillary is too high (no leak tightness, broken capillary). The maximum limit can be set by the menu. Default value is 0,15 mbar. The capillary is broken or has been torn off. E64 TMP error: Nominal speed has been exceeded by over 10% Nominal speed of the pump has been exceeded by over 10% EMC problems: Check connecting cable, insert it properly. Switch the power supply voltage off and then on again. Turbo.Drive S faulty: Inform Leybold Service E65 TMP error: Pass through time exceeded Max. time for passing through the critical frequencies has been exceeded. Fore vacuum or high-vacuum pressure to high: Reduce the inlet pressure of the PHOENIX L300i Bearing defective: Inform Leybold service for repair E66 TMP error: Bearing temperature too high (>67°C) Max. bearing temperature has been exceeded Fore vacuum or high-vacuum pressure to high: Reduce the inlet pressure of the PHOENIX L300i Fan defective: Replace the fan Ambient temperature too high: Feed cooler air to the PHOENIX L300i Bearing defective: Inform Leybold service for repair E67 TMP error: Short circuit in TMPmotor or connecting cable Short circuit in the pump’s motor or the connecting cable Check to see if the connecting cable is undamaged, exchange if required. Inform Leybold service in case of short circuit in TMP motor 100 GA10218_002_C0 - 10/2016 - © Leybold GmbH Troubleshooting No. Displayed Message Description and possible solutions E68 TMP error: temperature converter too high (>75°C) Maximum temperature for the converter has been exceeded. Ambient temperature too high: Feed cooler air to the PHOENIX L300i Fan defective: Replace the fan Fore vacuum or high-vacuum pressure to high: reduce the inlet pressure of the PHOENIX L300i E69 TMP error: Run-up time out Max. time after which the pump must enter its normal operation mode has been exceeded. Fore vacuum or high-vacuum pressure to high: Reduce the inlet pressure of the PHOENIX L300i Bearing defective: Inform Leybold service for repair E70 TMP error: TMP motor temperature too high (>90°C) Max. motor temperature has been exceeded. Fore vacuum or high-vacuum pressure to high: Reduce the inlet pressure of the PHOENIX L300i Fan defective: Replace the fan Ambient temperature too high: Feed cooler air to the PHOENIX L300i. Bearing defective: Inform Leybold service for repair E71 TMP error: TMP could not be identified Pump could not be identified or no pump is connected. E72 Emission off (P1 too high) Air inrush E73 Emission off (p2 too high) The emission is switched off as soon as the pressure P2 > 0.2 mbar or 1,5 mbar in the measurement mode. If after closing the inlet valve the pressure drops again, the leak detector will revert to the standby mode. Pump not connected to Turbo.Drive S: Check connecting cable Air inrush in the measurement mode. W74 Error proportional valve or control Proportional valve V4b or control electronics for proportional valve V4b electronics mounted on top-hat rail next to forepump non-existent or faulty. (PHOENIX L300i DRY only) W75 Maximum evacuation time "Test sample has got a GROSS leak." and "Wrong setting (too short) of max. exceeded time of evacuation." Within the pre set evacuation time the pressure of 100 mbar was not reached. W76 Maximum evacuation time until "Test sample has got a GROSS leak." and "Wrong setting (too short) of max. measurement exceededWithin time of evacuation." the pre set evacuation time measurement mode was not reached. W77 Peak not in Range The signal maximum has shifted to mass range alignment limits. Signal of leak rate was instable during mass adjustment. Calibrate again. Check the basic setting for the anode voltage through the service menu. Check calibrated leak. GA10218_002_C0 - 10/2016 - © Leybold GmbH 101 Troubleshooting No. Displayed Message Description and possible solutions W78 Differences of signal between test The amplifier voltage difference between opened and closed calibrated leak leak open and closed is too low. is less than 10 mV. Calibrated leak has not been closed properly. W79 Signal of test leak is too small Calibrated leak is too small or has not been opened. Preamplifier voltage < 10 mV. W80 Please calibrate machine newly The automatic request of calibration is activated and has fulfilled at least one of the conditions: 30 minutes are passed since power on. Temperature of the pre-amplifier has changed more than 5°C since the last calibration. Mass adjustments were changed. Change of TMP speed to 920 Hz, Gross only mode W81 CAL Factor too low The calculated factor falls out of the valid range (<0,1). The old factor is retained. Possible fault cause: The conditions for calibration have not been maintained. The leak rate of the internal calibrated leak which was entered is much too small. The internal test leak is defect. W82 CAL Factor too high The calculated factor is out of the valid range (>10). The old factor is retained. Possible fault cause: The conditions for calibration have not been maintained. The leak rate of the internal calibrated leak which was entered is much too high or much too small. W83 All EEPROM parameter lost. Please check your settings. EEPROM on back plane is empty and was initialized with default valves. Enter all parameters again. The EEPROM might be faulty when warning comes up again after power up. W84 EEPROM parameter initialized. Please check your settings Software update has been installed W85 Lost EEPROM parameter! Please Writing access was interrupted. Please check all adjustments. check your settings! An software update was done. In this case the notice can be ignored. When warning comes up again after powering up the EEPROM might be faulty. 102 GA10218_002_C0 - 10/2016 - © Leybold GmbH Waste Disposal 6 Waste Disposal The equipment may have been contaminated by the process or by environmental influences. In this case the equipment must be decontaminated in accordance with the relevant regulations. We offer this service at fixed prices. Further details are available on request. Contaminated parts can be detrimental to health and environment. Before beginning any work, first find out whether any parts are contaminated. Adhere to the relevant regulations and take the necessary precautions when handling contaminated parts. Separate clean components according to their materials, and dispose of these accordingly. We offer this service. Further details are available on request. When sending us any equipment, observer the regulations given in Section 4.3 Service at Leybold. GA10218_002_C0 - 10/2016 - © Leybold GmbH 103 Waste Disposal 104 GA10218_002_C0 - 10/2016 - © Leybold GmbH Certificates 7 Certificates Fig. 43 GA10218_002_C0 - 10/2016 - © Leybold GmbH 10 Certificates 106 GA10218_002_C0 - 10/2016 - © Leybold GmbH Appendix A Menu Apple I-Pad 31 Autoranging 7 57 MENU Button 41 Numerical Entries 42 On / Off switch 28 N B Background in Stand-by 8 O C Calibrated leak TL7 95 OPTION 28 Calibration request 76 Option 29 Partial flow setup 68 CHARGE 28, 32 CONTROL 28–29 Control 2 30 CONTROL2 28 P precision 8 RECORD 28, 30 REMOTE1 28, 31 REMOTE2 28, 31 RS232 31, 48 SERIAL 28, 31 R D Define PLC inputs 49 Define PLC outputs 48 Define recorder output 45 E EXHAUST 28, 32 S F fine 8 START Button 40 STOP Button 40 T G Gas ballast connection 32 Trigger & Alarms 59 GASBALLAST 28 Troubleshooting 96 Turbo molecular pump 94 USB interface 32 Vacuum ranges 67 Vacuum settings 65 gross 8 U I Inlet Port 32 Internal Calibration 85 V L List of Errors & Warnings 96 VENT M mains socket 28 28, 32 Z ZERO Button 40 GA10218_002_C0 - 10/2016 - © Leybold Vacuum106 Appendix 107 GA10218_002_C0 - 10/2016 - © Leybold Vacuum Appendix GA10218_002_C0 - 10/2016 - © Leybold Vacuum 108 Appendix 109 GA10218_002_C0 - 10/2016 - © Leybold Vacuum Appendix GA10218_002_C0 - 10/2016 - © Leybold Vacuum110 Sales and Service Germany Great Britain America Leybold GmbH Sales, Service, Support Center (3SC) Bonner Strasse 498 D-50968 Cologne T: +49-(0)221-347 1234 F: +49-(0)221-347 31234 sales@leybold.com www.leybold.com Leybold UK LTD. Unit 9 Silverglade Business Park Leatherhead Road Chessington Surrey (London) KT9 2QL Sales: T: +44-13-7273 7300 F: +44-13-7273 7301 sales.ln@leybold.com Service: T: +44-13-7273 7320 F: +44-13-7273 7303 service.ln@leybold.com USA Leybold GmbH Sales Area North Branch Office Berlin Industriestrasse 10b D-12099 Berlin T: +49-(0)30-435 609 0 F: +49-(0)30-435 609 10 sales.bn@leybold.com Leybold GmbH Sales Office South Branch Office Munich Karl-Hammerschmidt-Strasse 34 D-85609 Aschheim-Dornach T: +49-(0)89-357 33 9-10 F: +49-(0)89-357 33 9-33 sales.mn@leybold.com service.mn@leybold.com Leybold Dresden GmbH Service Competence Center Zur Wetterwarte 50, Haus 304 D-01109 Dresden Service: T: +49-(0)351-88 55 00 F: +49-(0)351-88 55 041 info.dr@leybold.com Europe Belgium Leybold Nederland B.V. Belgisch bijkantoor Leuvensesteenweg 542-9A B-1930 Zaventem Sales: T: +32-2-711 00 83 F: +32-2-720 83 38 sales.zv@leybold.com Service: T: +32-2-711 00 82 F: +32-2-720 83 38 service.zv@leybold.com France Leybold France S.A.S. Parc du Technopolis, Bâtiment Beta 3, Avenue du Canada F-91940 Les Ulis cedex Sales and Service: T: +33-1-69 82 48 00 F: +33-1-69 07 57 38 info.ctb@leybold.com sales.ctb@leybold.com Leybold France S.A.S. Valence Factory 640, Rue A. Bergès B.P. 107 F-26501 Bourg-lès-Valence Cedex T: +33-4-75 82 33 00 F: +33-4-75 82 92 69 marketing.vc@leybold.com Italy Leybold Italia S.r.l. Via Trasimeno 8 I-20128 Mailand Sales: T: +39-02-27 22 31 F: +39-02-27 20 96 41 sales.mi@leybold.com Service: T: +39-02-27 22 31 F: +39-02-27 22 32 17 service.mi@leybold.com Netherlands Leybold Nederland B.V. Floridadreef 102 NL-3565 AM Utrecht Sales and Service: T: +31-(30) 242 63 30 F: +31-(30) 242 63 31 sales.ut@leybold.com service.ut@leybold.com Switzerland Leybold Schweiz AG, Pfäffikon Churerstrasse 120 CH-8808 Pfäffikon Warehouse and shipping address: Riedthofstrasse 214 CH-8105 Regensdorf Sales: T: +41-44-308 40 50 F: +41-44-302 43 73 sales.zh@leybold.com Service: T: +41-44-308 40 62 F: +41-44-308 40 60 service.zh@leybold.com Spain Leybold Spain, S.A. C/. Huelva, 7 E-08940 Cornellà de Llobregat (Barcelona) Sales: T: +34-93-666 43 11 F: +34-93-666 43 70 sales.ba@leybold.com Service: T: +34-93-666 46 11 F: +34-93-685 43 70 service.ba@leybold.com Headquarter Leybold GmbH Bonner Strasse 498 D-50968 Cologne T: +49-(0)221-347-0 F: +49-(0)221-347-1250 info@leybold.com Leybold USA Inc. 5700 Mellon Road USA-Export, PA 15632 T: +1-724-327-5700 F: +1-724-325-3577 info.ex@leybold.com Sales: T: +1-724-327-5700 F: +1-724-333-1217 Service: T: +1-724-327-5700 F: +1-724-325-3577 Brazil Leybold do Brasil Rod. Vice-Prefeito Hermenegildo Tonolli, nº. 4413 - 6B Distrito Industrial Jundiaí - SP CEP 13.213-086 Sales and Service: T: +55 11 3395 3180 F: +55 11 99467 5934 sales.ju@leybold.com service.ju@leybold.com Asia P. R. China Leybold (Tianjin) International Trade Co. Ltd. Beichen Economic Development Area (BEDA), No. 8 Western Shuangchen Road Tianjin 300400 China Sales and Service: T: +86-22-2697 0808 F: +86-22-2697 4061 F: +86-22-2697 2017 sales.tj@leybold.com service.tj@leybold.com India Leybold India Pvt Ltd. No. 82(P), 4th Phase K.I.A.D.B. Plot Bommasandra Industrial Area Bangalore - 560 099 Indien Sales and Service: T: +91-80-2783 9925 F: +91-80-2783 9926 sales.bgl@leybold.com service.bgl@leybold.com Japan Leybold Japan Co., Ltd. Headquarters Shin-Yokohama A.K.Bldg., 4th floor 3-23-3, Shin-Yokohama Kohoku-ku, Yokohama-shi Kanawaga 222-0033 Japan Sales: T: +81-45-471-3330 F: +81-45-471-3323 sales.yh@leybold.com Leybold Japan Co., Ltd. Tsukuba Technical Service Center 1959, Kami-yokoba Tsukuba-shi, Ibaraki-shi 305-0854 Japan Service: T: +81-29 839 5480 F: +81-29 839 5485 service.iik@leybold.com Malaysia Leybold Malaysia Leybold Singapore Pte Ltd. No. 1 Jalan Hi-Tech 2/6 Kulim Hi-Tech Park Kulim, Kedah Darul Aman 09000 Malaysia Sales and Service: T: +604 4020 222 F: +604 4020 221 sales.ku@leybold.com service.ku@leybold.com South Korea Leybold Korea Ltd. 3F. Jellzone 2 Tower Jeongja-dong 159-4 Bundang-gu Sungnam-si Gyeonggi-do Bundang 463-384, Korea Sales: T: +82-31 785 1367 F: +82-31 785 1359 sales.bd@leybold.com Service: 623-7, Upsung-Dong Cheonan-Si Chungcheongnam-Do Korea 330-290 T: +82-41 589 3035 F: +82-41 588 0166 service.cn@leybold.com Singapore Leybold Singapore Pte Ltd. 8 Commonwealth Lane #01-01 Singapore 149555 Singapore Sales and Service: T: +65-6303 7030 F: +65-6773 0039 sales.sg@leybold.com service.sg@leybold.com Taiwan Leybold Taiwan Ltd. No 416-1, Sec. 3 Chunghsin Rd., Chutung Hsinchu County 310 Taiwan, R.O.C. Sales and Service: T: +886-3-500 1688 F: +886-3-583 3999 sales.hc@leybold.com service.hc@leybold.com
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.5 Linearized : Yes Author : l0108719 Create Date : 2016:10:21 07:56:01Z Modify Date : 2017:01:12 09:16:42-05:00 Tagged PDF : Yes XMP Toolkit : Adobe XMP Core 5.6-c015 84.158975, 2016/02/13-02:40:29 Creator Tool : FrameMaker 9.0 Metadata Date : 2017:01:12 09:16:42-05:00 Producer : Acrobat Distiller 11.0 (Windows) Format : application/pdf Title : PHOENIX-L300i-Operating-manual-GA10218_002_C0-(1609).book Creator : l0108719 Document ID : uuid:caac37e9-5dbc-4f8d-ab2f-fdbb0b4233a9 Instance ID : uuid:c7e5db32-d9c5-45fb-b84e-258eecf57ac8 Page Mode : UseOutlines Page Count : 112EXIF Metadata provided by EXIF.tools