RAE Systems TOXIRAEPRO QRAE3 User Manual ToxiRAE Pro PID User s Guide

RAE Systems, Inc QRAE3 ToxiRAE Pro PID User s Guide

manual

Download: RAE Systems TOXIRAEPRO QRAE3 User Manual ToxiRAE Pro PID User s Guide
Mirror Download [FCC.gov]RAE Systems TOXIRAEPRO QRAE3 User Manual ToxiRAE Pro PID User s Guide
Document ID3228900
Application IDajkLBrHXuqNHE0QXzbR7sA==
Document Descriptionmanual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize218.24kB (2728049 bits)
Date Submitted2016-12-15 00:00:00
Date Available2016-12-19 00:00:00
Creation Date2016-09-30 11:49:27
Producing SoftwareAdobe PDF Library 15.0
Document Lastmod2016-09-30 11:50:18
Document TitleToxiRAE Pro PID User's Guide
Document CreatorAcrobat PDFMaker 15 for Word
Document Author: tmulhern

ToxiRAE Pro PID
User’s Guide
P/N G02-4008-000 Rev. F
September 2016
© Copyright 2016 RAE Systems by Honeywell.
ToxiRAE Pro PID User’s Guide
Contents
1. General Information ........................................................................................................ 7
2. Specifications .................................................................................................................. 8
Radio Specifications ........................................................................................................... 9
3. Overview ....................................................................................................................... 10
4. User Interface ................................................................................................................ 10
4.1.
Display Overview ........................................................................................ 11
4.2.
Status Indicator Icons................................................................................... 11
5. Charging The ToxiRAE Pro PID .................................................................................. 12
5.1.
Removing The ToxiRAE Pro PID From The Charging Cradle.................. 13
5.2.
Low Voltage Shutoff.................................................................................... 13
6. Turning The ToxiRAE Pro PID On .............................................................................. 14
7. Turning The ToxiRAE Pro PID Off ............................................................................. 15
8. Testing The Alarms (Manually).................................................................................... 15
9. Low Battery Warnings .................................................................................................. 16
10. Dead Battery Warning & Shutoff ............................................................................... 16
11. Operating The ToxiRAE Pro PID ............................................................................... 17
11.1.
Overview ...................................................................................................... 17
11.1.1. Normal Mode ................................................................................................ 17
11.1.2. Basic User Mode ........................................................................................... 18
11.1.3. Advanced User Mode.................................................................................... 19
11.2.
Navigation In Normal Mode ........................................................................ 19
11.2.1. TWA, STEL, Peak ........................................................................................ 20
11.2.2. Date, Time, And Temperature ...................................................................... 20
11.2.3. Wireless Settings........................................................................................... 20
11.2.4. Calibration Gas, Measurement Gas, Correction Factor ................................ 20
11.2.5. Enter Communications and stop measurement? ........................................... 21
11.3.
Programming................................................................................................ 21
11.4.
Entering Programming Mode ...................................................................... 21
11.4.1. Navigating Programming Menus .................................................................. 23
11.5.
Programming Menus.................................................................................... 23
11.5.1. Calibration..................................................................................................... 24
11.5.2. Measurement ................................................................................................. 25
11.5.3. Alarm Setting ................................................................................................ 26
11.5.4. Datalog .......................................................................................................... 30
11.5.5. Monitor Setup ............................................................................................... 32
11.6.
Exit (Exit Programming Mode) ................................................................... 38
12. Policy Enforcement..................................................................................................... 39
12.1.
Setting Policy Enforcement ......................................................................... 39
12.1.1. AutoRAE 2 Cradle ........................................................................................ 39
12.1.2. ToxiRAE Pro PID Desktop Cradle Or Travel Charger ................................ 43
12.2.
Deactivating Policy Enforcement ................................................................ 45
12.2.1. AutoRAE 2 Cradle ........................................................................................ 45
12.2.2. ToxiRAE Pro PID Desktop Cradle Or Travel Charger ................................ 45
13. Bump Testing & Calibrating The ToxiRAE Pro PID ................................................. 47
13.1.
Bump Testing & Calibrating With An AutoRAE 2™ ................................. 47
13.1.1. Bump (Functional) Testing ........................................................................... 47
ToxiRAE Pro PID User’s Guide
13.2.
Calibration Adapter...................................................................................... 47
13.2.1. Connecting The Calibration Adapter ............................................................ 47
13.2.2. Disconnecting The Calibration Adapter ....................................................... 48
13.3.
Zero Calibration ........................................................................................... 48
13.4.
Span Calibration........................................................................................... 49
13.4.1. Changing Span Gas Concentration ............................................................... 50
14. Diagnostic Mode ......................................................................................................... 51
14.1.
Entering Diagnostic Mode ........................................................................... 51
14.2.
Navigating Diagnostic Mode ....................................................................... 51
14.3.
Exiting Diagnostic Mode ............................................................................. 51
15. Alarm Signal Summary............................................................................................... 52
16. Downloading Datalogs And Performing PC-Based
Instrument Configuration and Firmware Upgrades .................................................... 53
17. Maintenance ................................................................................................................ 54
17.1.
Replacing the ToxiRAE Pro PID Li-Ion Battery ......................................... 54
17.2.
Proper Battery Disposal ............................................................................... 57
17.3.
Removing And Installing The Rubber Boot ................................................ 58
17.4.
Filter Replacement ....................................................................................... 59
17.5.
Cleaning Or Replacing The PID .................................................................. 61
17.6.
Replacing The Fan ....................................................................................... 63
17.7.
Cleaning The Exterior .................................................................................. 65
17.7.1. Ordering Replacement Parts ......................................................................... 65
18. Troubleshooting .......................................................................................................... 66
19. Technical Support ....................................................................................................... 67
20. RAE Systems Contacts ............................................................................................... 67
21. Appendix A: Regulatory Information ......................................................................... 68
ToxiRAE Pro PID User’s Guide
WARNINGS
Read Before Operating
This manual must be carefully read by all individuals who have or will have the
responsibility of using, maintaining, or servicing this product. The product will perform
as designed only if it is used, maintained, and serviced in accordance with the
manufacturer’s instructions.
CAUTION!
Never operate the monitor when the cover is removed. Remove the monitor top cap, rear
cover, fan, or battery only in an area known to be non-hazardous.
Product Registration
Register your product online by visiting:
http://www.raesystems.com/support/product-registration
By registering your product, you can:
•
•
•
Receive notification of product upgrades or enhancements
Be alerted to Training classes in your area
Take advantage of RAE Systems special offers and promotions
ToxiRAE Pro PID User’s Guide
WARNINGS
Use only RAE Systems lithium- ion battery part number G02-3004-000 (3.7V,
1,800mAh). This instrument has not been tested in an explosive gas/air atmosphere
having an oxygen concentration greater than 21%. Substitution of components may
impair suitability for intrinsic safety. Replace batteries only in non-hazardous locations.
STATIC HAZARD: Clean only with a damp cloth to avoid damage from electrostatic
discharge.
For safety reasons this equipment must be operated and serviced by qualified personnel
only. Read and understand this instruction manual completely before operating or
servicing.
The monitor should be calibrated every time it does not pass a bump test, but no less
frequently than every six months, depending on use and exposure to gas and
contamination, and its operational mode.
• Calibration intervals and bump test procedures may vary due to national legislation.
• RAE Systems recommends using RAE calibration gas cylinders with one of the
following:
•
•
•
•
Isobutylene, 34L, 100 ppm (balance air), PN 600-0002-000
Isobutylene, 100 ppm 103L steel cylinder, PN 600-0002-001
Isobutylene Calibration Gas, 100 ppm, 34L cylinder; constant-flow CGA-600
regulator with adapter, PN 002-3025-000
Isobutylene, 100 ppm (balance air), 34L cylinder, Dry Air, 0 ppm, 34L cylinder;
Constant-flow DGA-600 regulator with adapter, PN 007-3001-000
ToxiRAE Pro PID User’s Guide
AVERTISSEMENT
Utiliser seulement la batterie Lithium- ion RAE Systems référence G02-3004-000 (3.7V,
1,800mAh). Cet instrument n’a pas été testé dans une atmosphère de gaz/air explosive
ayant une concentration d’oxygène plus élevée que 21%. La substitution de composants
peut compromettre la sécurité intrinsèque. Ne changer les batteries que dans un
emplacement désigné non dangereux.
RISQUE D'ORIGINE ELECTROSTATIQUE: Nettoyer uniquement avec un chiffon
humide.
Pour des raisons de sécurité, cet équipement doit être utilisé et entretenu uniquement par
un personnel qualifié. Étudier le manuel d’instructions en entier avant d’utiliser, ou
d’entretenir l’équipement.
• Une vérification d’étalonnage est définie par une exposition du détecteur au gaz
d’étalonnage qui doit déclencher le seuil d’alarmes bas.
• Les périodes de calibration peuvent varier suivant les législations locales.
RAE Systems préconise d’utiliser les gaz de calibration suivants:
•
•
•
•
Isobutylene, 34L, 100 ppm (balance air), PN 600-0002-000
Isobutylene, 100 ppm 103L Cylindre acier, PN 600-0002-001
Isobutylene Calibration, 100 ppm, 34L cylindre; débit constant CGA-600
regulateur avec adaptateur, PN 002-3025-000
Isobutylene, 100 ppm (balance air), 34L cylindre, Air sec, 0 ppm, 34L cylindre;
débit constant DGA-600 regulateur avec adaptateur, PN 007-3001-000
ToxiRAE Pro PID User’s Guide
CAUTION
This device complies with Part 15 of the FCC Rules / Industry Canada license-exempt
RSS standard(s). Operation is subject to the following two conditions: (1) this device may
not cause harmful interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux
appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions
suivantes : (1) l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur de
l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est
susceptible d'en compromettre le fonctionnement.
Changes or modifications not expressly approved by the party responsible for compliance
could void the user's authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a residential installation. This
equipment generates uses and can radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the following
measures:
•
•
•
•
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
Under Industry Canada regulations, this radio transmitter may only operate using an
antenna of a type and maximum (or lesser) gain approved for the transmitter by Industry
Canada. To reduce potential radio interference to other users, the antenna type and its
gain should be so chosen that the equivalent isotropically radiated power
(e.i.r.p.) is not more than that necessary for successful communication.
Conformément à la réglementation d'Industrie Canada, le présent émetteur radio peut
fonctionner avec une antenne d'un type et d'un gain maximal (ou inférieur) approuvé
pour l'émetteur par Industrie Canada. Dans le but de réduire les risques de
brouillage radioélectrique à l'intention des autres utilisateurs, il faut choisir le type
d'antenne et son gain de sorte que la puissance isotrope rayonnée équivalente
(p.i.r.e.) ne dépasse pas l'intensité nécessaire à l'établissement d'une communication
satisfaisante.
ToxiRAE Pro PID User’s Guide
1. General Information
ToxiRAE Pro PID is a personal wireless monitor for volatile organic compounds
(VOCs), the world’s smallest. Featuring RAE Systems’ next-generation PID sensor, the
ToxiRAE Pro PID can promptly detect and accurately monitor over 300 VOCs. With an
onboard library of 190 correction factors, the ToxiRAE Pro PID can be programmed to
automatically read in concentrations of the specified compound.
Key Features
Wireless1 . Personal. Proven
• Wireless access to real-time instrument readings and alarm status from any location
• Unmistakable five-way local and remote wireless notification of alarm conditions
• Cutting-edge PID with 190 built- in correction factors
• Largest display in its class
• Reliable, rugged, and intrinsically safe
• Continuous datalogging2 capability
• Fully automated bump testing and calibration with AutoRAE 2 3
• Easy to maintain with replaceable sensor, fan, and Lithium-ion battery
• Available in Industrial Hygiene (advanced) or Safety (basic) 3 configurations
• Wireless or non-wireless versions4 are available
Wireless functionality is currently offered in 2011 in North America and Europe. Contact RAE Systems for
availability in other regions.
Available with Industrial Hygiene configurations only.
Contact RAE Systems for availability.
Wireless versions have a functioning wireless modem and are ready for wireless deployments. “Nonwireless” versions cannot be upgraded to wireless in the future. Additional equipment and/or software
licenses may be required to enable remote wireless monitoring and alarm transmission.
ToxiRAE Pro PID User’s Guide
2. Specifications
ToxiRAE Pro PID Specifications
Dimensions
Weight
Battery
Charging Time
Operating Time
Charger
118 mm x 60 mm x 30 mm (4.6" x 2.4" x 1.2")
235g (8.3 oz)
Internal rechargeable Li-ion
Less than 4 hours
More than 12 hours on full charge (normal operation)
Charging and PC communication cradle: AC / DC adapter with international
prong kit (100 to 240V, 50/60 Hz AC to 12V DC)
Multi-unit charging station for up to 5 units (data download from one):
AC/DC adapter for the multi-unit charging station
Alarms Audio alarm – 95 dB @ 30cm
Visual alarm – 4 bright red LED lights
Vibration alarm
Activated display backlighting and on-screen indication of alarm conditions
Wireless alarm notification sent to remote observers (wireless versions only)
Display Monochrome dot-matrix graphical LCD display
Display Backlight Manually and on alarm
Keypad 2 programming/operation keys
Direct Readout Real-time reading of VOC concentrations in parts per million or mg/m3 ;
battery status; data logging on/off status; wireless reception quality; STEL,
TWA, and peak values; calibration and measurement gases, correction factors
Sampling Method Fan-assisted diffusion
Datalogging Capacity: 3 months at 1-minute interval
Datalog interval: user-configurable from 1 to 3,600 seconds
Calibration Two-point field calibration of zero and standard reference gas
Adjustable span value
Maintains calibration date
IP Rating IP-54
Docking Station AutoRAE 2 Automatic Test and Calibration Station
Protection:
Hazardous
Location
Approvals
CE Compliance
(European
Conformity)
Operating
Temperature
Humidity
Warranty
Password-protected calibration settings, alarm limits, and data
Class I, Div. 1; Gr. A,B,C,D T4
Class I, Zone 0 A/Exia IIC T4
II 1G,
ATEX:
Ex ia IIC Ga T4
IECEx: Ex ia IIC Ga T4
China: Ex ia IIC T4
EMC directive: 2004/108/EC
R&TTE directive: 1999/5/EC
ATEX directive: 94/9/EC
-4° to 131° F (-20° to 55° C)
0 to 95% relative humidity (non-condensing)
Two years on non-consumable components (per RAE Systems Standard
Warranty); one year on fan, sensor, battery, and other consumable components
ToxiRAE Pro PID User’s Guide
Radio Specifications
Radio model: RM900
Frequency range: Within 902 to 907.5 MHz and 915 to 928 MHz, use IEEE 802.15.4
channel 1, 6, 7, 8, 9 and 10
Modulation: 802.15.4 DSSS BPSK
RF power (Tx):10dBm
Data rate: 40kbps
Wireless Approval For UAE In Middle East
TRA REGISTERED No: ER36635/15
DEALER No: HONEYWELL INTERNATIONAL MIDDLE EAST - LTD - DUBAI
BR
Wireless Approval For QATAR In Middle East
ictQATAR
Type Approval Reg. No.: R-4635
ToxiRAE Pro PID User’s Guide
3. Overview
The ToxiRAE Pro PID is a wireless, personal volatile organic compounds (VOC)
monitor, the world’s smallest. The ToxiRAE Pro PID takes worker protection to the next
level by providing safety professionals wireless access to real-time instrument readings
and alarm status from any location for better visibility and faster response. The ToxiRAE
Pro PID features RAE Systems’ next-generation PID sensor that delivers fast, reliable
detection and accurate monitoring of a wide variety of VOCs in a broad range of
temperature and humidity conditions.
4. User Interface
The ToxiRAE Pro PID’s user interface consists of the display, four LEDs, an alarm
buzzer, and two keys, [MODE] and [Y/+]. The LCD displays real-time reading in
specified measurement units, sensor type, alarm type (when in alarm, including cal.
overdue), battery status, datalog (if on), and radio and connection quality (if available).
Gas inlet
Top cap
Belt clip (on back)
LED
LEDs
Alarm
Buzzer
Display
[MODE] key
[Y/+] key
10
ToxiRAE Pro PID User’s Guide
4.1. Display Overview
The LCD display provides visual feedback that includes the sensor types, readings, time,
battery condition, and other functions.
Calibration and bump
test up to date
Sensor name and “Alarm” are
displayed alternately when in alarm
Wireless on/off status
Wireless signal strength
Datalog on
Bump test due
Battery status
Sensor
calibration due
Man Down alarm on
Reading
Unit of measure
Alarm type
(when in alarm)
Soft keys
(functions change by activity)
4.2. Status Indicator Icons
Along the top of most screens are status indicators that tell you whether a function is
operating and/or its strength or level.
Icon
Function
Wireless status: the radio is on (replaced by “R” when Roaming is on)
Wireless status: the radio is off (replaced by “R” when Roaming is on)
Roaming status: “R” blinks when trying to find a network (replaced by antenna
when Roaming is off)
Roaming status: “R” solid when network communication established (replaced
by antenna when Roaming is off)
The instrument is not equipped with a radio (no icon)
Wireless strength (0 to 5 bars)
1 bar = Network joined, signal
2 bars = Network joined, signal
3 bars = Network joined, signal
4 bars = Network joined, signal
5 bars = Network joined, signal
very low RSSI (less than 20%)
low RSSI (21% to 40%)
medium RSSI (41% to 60%)
good RSSI (61% to 80%)
very good RSSI (81% to 100%)
Datalogging status (shown when datalogging is on, blank when off)
Battery status (three segments show battery charge level)
Man Down alarm enabled
Sensor calibration due
11
ToxiRAE Pro PID User’s Guide
Icon
Function
Bump test due
All sensors tested and calibrated tick mark (all sensors have been bump tested
and calibrated; no sensor is overdue for a bump test or calibration according to
the intervals configured on the instrument
5. Charging The ToxiRAE Pro PID
Always fully charge the battery before using the ToxiRAE Pro PID. The ToxiRAE Pro PID’s
Li-ion battery is charged by placing the ToxiRAE Pro PID in its cradle. Contacts on the
bottom of the ToxiRAE Pro PID meet the cradle’s contact pins, transferring power.
Rear view of
ToxiRAE Pro PID
Raised lip on
rear of ToxiRAE Pro PID
locks into cradle’s latch
Charging status
LED
ToxiRAE Pro
Charging and PC
Communications
Cradle
Latch release
Rear view of
charging cradle
Connect AC
adapter to the
charging cradle
Note: Before setting the ToxiRAE Pro PID into its charging cradle, visually inspect the
contacts to make sure they are clean. If they are not, wipe them with a soft, dry cloth. Do
not use solvents or cleaners.
12
ToxiRAE Pro PID User’s Guide
Follow this procedure to charge the ToxiRAE Pro PID (the procedure is applicable to the
instrument with or without the rubber boot installed):
1. Plug the AC/DC adapter into the ToxiRAE Pro PID’s cradle.
2. Plug the AC/DC adapter into the wall outlet.
3. Place the ToxiRAE Pro PID into the cradle and press down until it is locked in
place.
The ToxiRAE Pro PID begins charging automatically. The LED in the cradle should
glow red to indicate charging. When charging is complete, the LED in the cradle glows
green.
WARNING!
To reduce the risk of ignition of hazardous atmospheres, recharge the instrument
only in areas known to be non-hazardous. Remove and replace batteries only in
areas known to be non-hazardous.
5.1. Removing The ToxiRAE Pro PID From The Charging
Cradle
To remove the ToxiRAE Pro PID from its charging cradle, press down on the latch
release on the back of the charging cradle while slightly tilting the ToxiRAE Pro PID
toward the front of the cradle, and lift it out.
Lift ToxiRAE Pro PID out of
charging cradle
Press on latch
5.2. Low Voltage Shutoff
When the battery’s charge falls below the shutoff voltage, the ToxiRAE Pro PID then
turns itself off. Recharge the battery by placing the ToxiRAE Pro PID in its cradle (the
cradle should be connected to the AC adapter).
13
ToxiRAE Pro PID User’s Guide
6. Turning The ToxiRAE Pro PID On
To turn the ToxiRAE Pro PID on, press and hold [MODE] for 3 seconds.
Caution: The alarm buzzer is very loud. During startup, you can mute most of the sound
by holding a finger over the alarm buzzer port.
Note: Never put tape over the alarm buzzer port to permanently mute it or otherwise limit
the loudness of the buzzer.
1. When the ToxiRAE Pro PID turns on, the buzzer, vibration alarm, and LEDs are
tested. The firmware version is shown in the display (for example, “V1.00” indicates
firmware version 1.00).
2. A battery test is performed. If the battery has sufficient charge, then the next step is a
self test of the system.
Note: If the battery is almost fully discharged, the message “Battery too low! Needs
charging. Powering off!” shows in the display and the ToxiRAE Pro PID shuts off.
You must charge the battery before you can use the instrument.
3. Self-testing is performed by the ToxiRAE Pro PID.
Note: If one of the following error messages is displayed, shut off the ToxiRAE Pro
PID by pressing [MODE]:
•
•
Sensor error!
Wrong sensor!
If any of the following error messages occurs, you can either shut off the instrument
by pressing [MODE], or continue to the next startup test by pressing [Y/+]:
•
•
•
•
•
No fan or fan error!
Real time clock error!
Memory error!
Sensor changed!
Failed to turn on lamp!
For any of these error types, try restarting the instrument. If the error message appears
again after five more attempts to restart the instrument, consult the distributor from
whom the instrument was purchased for technical support.
14
ToxiRAE Pro PID User’s Guide
4. After self-testing, two start-up options are available for the ToxiRAE Pro PID:
Normal and Fast Start-up. The following information is displayed in sequence under
Normal Start-up and skipped under Fast Start-up:
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Sensor type
Sensor serial number
Date, time, and temperature
Calibration gas
Last calibration date
Measurement gas
Correction factor
User mode
Battery type, voltage, and shutoff voltage
Alarm mode
Buzzer enabled/disabled status
Light enabled/disabled status
Vibration alarm enabled/disabled status
Alarm settings for High, Low, STEL, and TWA
Datalog type and interval
5. Once the startup process is complete, a sensor warm-up screen is shown (with
countdown, if necessary). Then the main measurement screen is shown. The type of
monitored hazard is displayed, with current reading. The ToxiRAE Pro PID is now
ready for use.
7. Turning The ToxiRAE Pro PID Off
Press and hold [MODE]. In 2 seconds, a 5-second countdown to shutoff begins,
accompanied by one buzzer beep per second. You must hold your finger on the key for
the entire shutoff process. If you remove your finger from the key during the countdown,
the shutoff operation is canceled and the ToxiRAE Pro PID continues normal operation.
When you see “Unit Off,” release your finger from the [MODE] key. The ToxiRAE Pro
PID is now off.
8. Testing The Alarms (Manually)
Under normal non-alarm conditions, the buzzer, vibration alarm, and LEDs can be tested
at any time by pressing and holding [Y/+].
Important! If any of the alarms does not respond to this test, check the Alarm Settings. It is
possible that any or all of the alarms have been turned off. If all the alarms are turned on, but
one or more of them (buzzer, LED lights, or vibration alarm) does not respond to this test, do
not use the instrument; contact your RAE Systems distributor for technical support.
15
ToxiRAE Pro PID User’s Guide
9. Low Battery Warnings
If, during operation, the battery’s charge is low, the battery icon in the display screen
turns empty and blinks once every second. At the same time, the following notifications
occur every minute: the buzzer beeps and LEDs blink once, followed by a vibration and a
screen message “Battery low! Shut off in X minutes.” The countdown to shutoff starts 10
minutes prior to automatic shutoff.
10. Dead Battery Warning & Shutoff
If you turn on the ToxiRAE Pro PID and the battery’s charge is below its lowest
threshold, the display shows this message:
Battery too low!
Needs charging.
Powering off!
It then automatically shuts off. You must charge the ToxiRAE Pro PID’s battery before
you can continue to use this instrument.
Important! Do not operate the ToxiRAE Pro PID again until its battery has been
charged.
WARNING!
To reduce the risk of ignition of hazardous atmospheres, charge the battery only in
areas known to be non-hazardous.
16
ToxiRAE Pro PID User’s Guide
11. Operating The ToxiRAE Pro PID
11.1. Overview
The ToxiRAE Pro PID has two operational modes:
•
•
Normal — see page 17 for detailed instructions.
Diagnostic — see page 51 for detailed instructions.
The following is an overview of the two modes:
Normal Mode is the default mode. It is accessed when you turn on the ToxiRAE Pro
PID. Two sub-modes are available: Basic User and Advanced User. In Basic User, some
restrictions are applied, including password protection that guards against entering
Programming Mode by unauthorized personnel. In Advanced, there are no such access
restrictions (you do not need a password), and it provides the indications and data you
need most for typical monitoring applications.
Diagnostic Mode is primarily intended for technicians to use during troubleshooting, although it
also offers access to a few changeable parameters that you may rarely (if ever) need to change. A
password is required to enter Diagnostic Mode (the password is the same as (and
automatically kept in sync with) the one used to enter Programming Mode). The default
password is 0000. To enter Diagnostic Mode, power the instrument off and then turn it
back on by pressing and holding both the buttons. In Diagnostic Mode, ToxiRAE Pro PID
displays readings in raw counts instead of units such as parts per million (ppm).
It is not possible to switch to Programming from Diagnostic Mode without turning off the
instrument first and then restarting it in the desired mode.
11.1.1. Normal Mode
Normal Mode is the default mode of the ToxiRAE Pro PID when it is turned on. The
selected sub-mode (Basic or Advanced) determines the number of screens that are
accessible.
17
ToxiRAE Pro PID User’s Guide
11.1.2. Basic User Mode
When the ToxiRAE Pro PID is in Basic User Mode, pressing [MODE] allows you to step
through the screens in the following fashion:
Note: The gray boxes indicate the datalog functions when datalogging is in Manual
mode. When datalogging is in Automatic mode, these screens do not appear. Refer to
page 30 for information on selecting Automatic or Manual datalogging.
18
ToxiRAE Pro PID User’s Guide
11.1.3. Advanced User Mode
Note: To enter Advanced User mode from Basic User mode, you must enter
Programming mode on an instrument that is running in Basic User mode by
simultaneously pressing and holding both the buttons for 3 seconds and supply a
password.
To exit Programming Menu and return to the measurement screen (in Advanced User
mode), press the [MODE] button repeatedly to highlight the Exit icon in Programming
Menu and press Y to exit to the measurement screen.
When the ToxiRAE Pro PID is in Advanced User mode, pressing [MODE] steps through
the same screens, but with the addition of a screen that allows for selection of different
measurement gases from the internal library. Also, once in Advanced User mode, you
may enter Programming without using a password.
Note: In Advanced User Mode, you may select different measurement gases from the
internal library.
Note: The gray box indicates the datalog functions when datalogging is in Manual mode.
When datalogging is in Automatic mode, these screens do not appear. Refer to page 30
for information on selecting Automatic or Manual datalogging.
11.2. Navigation In Normal Mode
Step through the screens in Normal Mode by pressing [MODE]. Some screens allow
actions on your part, such as clearing the data. In these instances, pressing [Y/+] performs
the action.
19
ToxiRAE Pro PID User’s Guide
11.2.1. TWA, STEL, Peak
The first screen includes a readout for TWA, STEL, and Peak readings.
TWA
The TWA (time-weighted average) reading is the average reading of the gas
concentration times that the monitor has been on, prorated to an 8-hour period.
STEL
Short Term Exposure Limit (STEL) data based on 15-minute STEL values,
expressed in ppm or mg/m3 . A STEL value is only displayed after the
ToxiRAE Pro PID has been on for 15 minutes or longer. After exiting PC
Communications mode, the STEL is reset.
Peak
The Peak value tells you the highest reading for the sensor since the ToxiRAE
Pro PID was turned on. Press [Y/+] twice to clear the Peak value.
Note: If you do not press a key within 1 minute, the display reverts to the
normal reading.
Press [MODE] to advance from screen to screen.
11.2.2. Date, Time, And Temperature
The current Date is expressed as Month/Day/Year, with two digits for month and day,
and four digits for the year. Below the date is the time, which is expressed as
Hours/Minutes/Seconds, with two digits for each. The time is in 24-hour (military)
format. Below the time is the internal temperature, which can be in degrees Fahrenheit or
Celsius (settable in Programming Mode).
11.2.3. Wireless Settings
The current wireless settings are shown here. These include the Roaming, PAN ID,
Channel, and Unit ID, which are helpful when creating or troubleshooting networks.
Parent ID data does not apply.
11.2.4. Calibration Gas, Measurement Gas, Correction Factor
This screen shows calibration gas (Cal. Gas), measurement gas (Meas. Gas), and
correction factor (CF).
20
ToxiRAE Pro PID User’s Guide
11.2.5. Enter Communications and stop measurement?
Before communicating with a PC running ProRAE Studio II Instrument Configuration
and Data Management software (for downloading datalogs, configuring the instrument,
or uploading new firmware), the ToxiRAE Pro PID must be set to communicate with the
computer. Press [MODE] repeatedly until you reach the “Enter Communications and stop
measurement?” screen. Then press [Y/+]. Measurement and datalogging are stopped, and
the instrument is now ready to communicate with the PC.
11.3. Programming
You must provide a 4-digit password to enter Programming Mode, unless the ToxiRAE
Pro PID is already running in Advanced User mode. The default password is 0000.
11.4. Entering Programming Mode
1. Press and hold [MODE] and [Y/+] simultaneously until the Password screen appears.
2. Input the 4-digit password:
• Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to 9.
Once 9 is reached, pressing [Y/+] again "wraps" around back to 0.
• Step from digit to digit using [MODE].
• When you are done, continue pressing [MODE] until OK or Cancel is highlighted.
• Press [Y/+] to register the password (if OK is highlighted) or to exit without entering
the password (if Cancel is selected).
If you make a mistake, you can cycle through the digits by pressing [MODE] repeatedly
and then using [Y/+] to change the number in each position.
Note: The default password is 0000.
21
ToxiRAE Pro PID User’s Guide
When you have successfully entered Programming Mode, the following screen appears:
Note: The password can only be changed by connecting the instrument to a PC running
ProRAE Studio II software. Follow the instructions in ProRAE Studio II to change it.
The Calibration icon is highlighted first by default, but you can press [MODE] to step
from one programming menu item to the next, with the name of the menu shown at the
top of the display and the corresponding icon highlighted. As you repeatedly press
[MODE], the selection moves from left to right, and the following screens appear in
sequence:
Note: When you reach Exit and press [MODE], the menu cycles back to Calibration.
To exit the Programming Menu, highlight Exit by pressing the [MODE] button, and press
the [Y/+] key.
22
ToxiRAE Pro PID User’s Guide
11.4.1. Navigating Programming Menus
Navigating through the Programming menus is easy and consistent, using a single
interface format of a “forward arrow” and “Select,” controlled by the [MODE] key and
[Y/+] key, respectively.
In the submenus, the functions of a “down arrow” and “Select” are controlled by the
[MODE] key and [Y/+] key, respectively.
Also in submenus, the functions of the keys may correspond to the instruction directly
above each key. Therefore, if it says “OK” above the [MODE] key, pressing [MODE]
selects “OK.” When setting numerical data, the [MODE] key typically selects which digit
is highlighted, while each press of [Y/+] increases the digit by 1.
11.5. Programming Menus
Programming allows anyone with the password to change the instrument’s settings, calibrate
the instrument, enter user information, etc. Programming has five menus. Each menu
includes several submenus to perform additional programming functions.
This table shows the menus (and their icons) and submenus:
Calibration
Zero Calibration
Measurement
Meas. Gas
Alarm Setting
Datalog
High Alarm
Clear Datalog
Monitor Setup
Wireless
Span Calibration
Meas. Unit
Low Alarm
STEL Alarm
TWA Alarm
Site ID
User ID
User Mode
Interval
Data Selection
Datalog Type
Alarm Mode
Buzzer & Light
Date
Time
Man Down
Alarm
Temp. Unit
Language
Zero At Start
LCD Contrast
Note: There is also an “Exit” icon. It has only one function. When you select it, the
instrument exits Programming Mode, updates the settings, and returns to monitoring.
23
ToxiRAE Pro PID User’s Guide
Once you enter Programming, the LCD displays the first menu, Calibration. Each
subsequent menu is accessed by pressing [MODE] repeatedly until the desired menu is
displayed. To enter a submenu of a menu, press [Y/+].
11.5.1. Calibration
Two types of calibration are available: Zero (fresh air) and Span.
In the Calibration submenu, select Zero or Span Calibration by pressing [MODE]. Once
your choice is highlighted, press [Y/+].
Zero Calib
The procedure for performing a zero calibration is covered on page 39.
24
ToxiRAE Pro PID User’s Guide
Span Calib
The procedure for performing a basic span calibration is covered on page 49.
Exit
Press [Y/+] to exit. Calibration is exited, and Measurement is highlighted in the
Programming Menu screen.
11.5.2. Measurement
The submenus for Measurement are Measurement Gas and Measurement Unit.
Meas Gas
Measurement gases are organized in a list called Gas Library:
The Gas Library consists of 190 compounds found in RAE Systems’ Technical Note TN106 (available online at www.raesystems.com).
1. Scroll through the Gas Library by pressing [Y/+] until you find the gas you want
as your measurement gas.
2. When you are on the gas of your choice, press [MODE] to highlight “Save.”
3. To save your choice, press [Y/+]. Otherwise, press [MODE] to highlight “Quit,”
followed by [Y/+]; the newly selected gas will not be saved.
Meas Unit
Standard available measurement units include:
Abbreviation
ppm
mg/m3
Unit
parts per million
milligrams per cubic meter
1. Choose between the two measurement units by pressing [MODE].
2. If you switch to a different measurement unit, press [Y/+] to select the new
choice. Press [MODE] until “Exit” is highlighted.
3. Press [Y/+] to save your choice and exit.
Exit
Press [Y/+] to exit. Measurement is exited, and Alarm Setting is highlighted in the
Programming Menu screen.
25
ToxiRAE Pro PID User’s Guide
11.5.3. Alarm Setting
High Alarm
Set the High Alarm as follows:
1. In the High Alarm submenu, press [Y/+] repeatedly to select a desired number.
Numbers increase from 0 to 9. Once 9 is reached, pressing [Y/+] again “wraps”
around back to 0.
2. Press [MODE] to move to the next digit.
3. When you are satisfied with the number, press [MODE] until “OK” is
highlighted.
4. To save the new value, press [Y/+]. The value is saved, and the menu advances to
Low Alarm.
To exit without saving the new value:
1. Press [MODE] to advance to Cancel.
2. Press [Y/+] to exit the High Alarm submenu and advance to the Low Alarm
submenu.
Low Alarm
Set the Low Alarm as follows:
1. In the Low Alarm submenu, press [Y/+] repeatedly to select a desired number.
Numbers increase from 0 to 9. Once 9 is reached, pressing [Y/+] again “wraps”
around back to 0.
2. Press [MODE] to move to the next digit.
3. When you are satisfied with the number, press [MODE] until “OK” is highlighted.
4. To save the new value, press [Y/+]. The value is saved, and the menu advances to
STEL Alarm.
To exit without saving the new value:
1. Press [MODE] to advance to Cancel.
2. Press [Y/+] to exit the Low Alarm submenu and advance to STEL Alarm
submenu.
STEL Alarm
Set the STEL Alarm as follows:
1. In the STEL Alarm submenu, press [Y/+] repeatedly to select a desired number.
Numbers increase from 0 to 9. Once 9 is reached, pressing [Y/+] again “wraps”
around back to 0.
2. Press [MODE] to move to the next digit.
3. When you are satisfied with the number, press [MODE] several times until “OK”
is highlighted.
26
ToxiRAE Pro PID User’s Guide
4. To save the new value, press [Y/+]. The value is saved, and the menu advances to
TWA Alarm.
To exit without saving the new value:
1. Press [MODE] to advance to Cancel.
2. Press [Y/+] to exit the STEL Alarm submenu and advance to the TWA Alarm
submenu.
TWA
Set the TWA Alarm as follows:
1. In the TWA Alarm submenu, press [Y/+] repeatedly to select a desired number.
Numbers increase from 0 to 9. Once 9 is reached, pressing [Y/+] again “wraps”
around back to 0.
2. Press [MODE] to move to the next digit.
3. When you are satisfied with the number, press [MODE] until “OK” is
highlighted.
4. To save the new value, press [Y/+]. The value is saved, and the menu advances to
Alarm Mode.
To exit without saving the new value:
1. Press [MODE] to advance to Cancel.
2. Press [Y/+] to exit the TWA Alarm submenu and advance to the Alarm Mode
submenu.
Alarm Mode
There are two selectable alarm types:
Auto Reset
When an alarm is triggered and after the alarm condition is no
longer present, the alarm stops and resets by itself.
Latch
When the alarm is triggered, it will not cease until you manually
turn it off.
The latched setting only controls alarms for High Alarm and Low
Alarm, not the STEL or TWA alarms.
Note: To clear an alarm when the instrument is set to “Latched,”
press [Y/+] when the main (measurement) display is shown.
1. Press [MODE] to step from one alarm type to the other.
2. Press [Y/+] to select an alarm type.
3. When you have completed your selection, press [MODE] until “Exit” is
highlighted. Then press [Y/+] to save the change and exit to “Buzzer and Light.”
27
ToxiRAE Pro PID User’s Guide
Buzzer & Light
The buzzer, light, and vibration alarms can be programmed so that any combination, all,
or none of them are activated when the ToxiRAE Pro PID goes into alarm.
Important! If you program the ToxiRAE Pro PID so that no alarms are activated in an
alarm condition, you may put yourself or your workers at risk of being exposed to a
hazardous atmosphere and being unaware of it, as the indication of a dangerous condition
will be limited to the concentration shown in the display and the name of the type of
alarm flashing on the display.
1. Choose an alarm combination setting by pressing [MODE] until your choice is
highlighted.
2. Press [Y/+] to select it.
3. Press [MODE] repeatedly until “Exit” is highlighted.
4. Press [Y/+] to register your selection and exit.
To exit without saving the new value:
1. Make sure your previous setting is selected.
2. Press [MODE] repeatedly to advance to Exit.
3. Press [Y/+] to exit.
Man Down Alarm
The Man Down Alarm is a critical and potentially lifesaving safety feature of every
ToxiRAE Pro PID. The Man Down Alarm is based on the premise that if the instrument
is motionless when it is not supposed to be, something wrong may be happening to its
user. If that is the case, the ToxiRAE Pro PID not only goes into alarm locally on the
instrument, but also remotely, over the RAE Systems Dedicated Wireless Network, to
notify people in the vicinity, as well as remote safety officers at a command center, that a
person is down, so that help can be dispatched quickly.
Whenever the Man Down feature is on, the main screen displays a Man Down icon along the
top to indicate it is active:
28
ToxiRAE Pro PID User’s Guide
When the Man Down feature is on and there is no gas alarm, the ToxiRAE Pro PID senses that
it is motionless for the amount of time set in the “Motionless Time” parameter. If the
instrument is not moved during that time, then a pre-alarm is activated to alert the user, and
shows the “Are You OK?” screen. Pressing [Y/+] clears the alarm and returns the ToxiRAE
Pro PID to its normal operation. Pressing [MODE] sets it into Man Down Alarm (and if
wireless connectivity is enabled, a Man Down message is sent in real time to remote
observers). If neither key is pressed, then after the countdown, it goes into Man Down Alarm
(again sending a message to remote observers if wirelessly enabled).
Settings are available for:
• Off/On
• Motionless Time (time the instrument is motionless before initiating a pre-alarm)
• Sensitivity (set to low, medium, or high to compensate for ambient vibration or motion)
• Warning Time (countdown, in seconds, from pre-alarm to Man Down alarm)
When the Man Down alarm is activated, the buzzer sounds and LEDs flash continuously, and a
countdown begins.
•
•
•
If the ToxiRAE Pro PID’s user presses [Y/+] for “Yes” in response to the “Are You
OK?” question on the screen before the countdown reaches zero, the Man Down alarm
stops and the main reading screen is displayed.
If the person does not press [Y/+] for “Yes” in response to the “Are You OK?”
question on the screen before the countdown reaches zero, the Man Down alarm is
sounded and LEDs flash continuously.
If the person presses [MODE] during the countdown, answering the “Are You OK?”
question with “No,” the Man Down alarm starts.
If the ToxiRAE Pro PID’s wireless connectivity is enabled, a Man Down message is also sent to
remote observers.
Exit (the Alarm Settings Submenu)
Press [Y/+] to exit the Alarm Settings submenu. Alarm Settings is exited, and Datalog is
highlighted.
If you want to cycle through the Alarm Setting submenus, press [MODE] instead of [Y/+].
29
ToxiRAE Pro PID User’s Guide
Exit (the Alarm Settings Submenu)
Press [Y/+] to exit the Alarm Settings submenu. Alarm Settings is exited, and Datalog is
highlighted.
If you want to cycle through the Alarm Setting submenus, press [MODE] instead of
[Y/+].
11.5.4. Datalog
Clear Datalog
This operation erases all data stored in the datalog.
Note: Once the datalog is cleared, the data cannot be recovered.
Press [Y/+] to clear the datalog. The display asks, “Are you sure?”
•
•
Press [Y/+] if you want to clear the datalog. When it has been cleared, the display
shows “Datalog Cleared!”
Press [MODE] if you do not want to clear the datalog.
You are taken to the next submenu, Interval.
Interval
Intervals are shown in seconds. The default value is 60 seconds. The maximum interval is
3600 seconds, and the minimum is 1 second.
1. Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to 9.
Once 9 is reached, pressing [Y/+] again “wraps” around back to 0.
2. Press [MODE] to advance to the next digit.
3. Repeat this process until the desired value is entered.
When you have completed your selections, press [MODE] repeatedly until “OK” is
highlighted.
•
•
To save the new value, press [Y/+].
To cancel the changes and keep the settings as they were, press [MODE] to
highlight “Cancel.” Then press [Y/+].
You are taken to the next submenu, Data Selection.
30
ToxiRAE Pro PID User’s Guide
Data Selection
Data Selection allows you to select which types of data are recorded and made available
when you download your datalog to a computer via ProRAE Studio II software.
You can choose any or all of four types of data (you must choose at least one):
•
•
•
•
Minimum
Average
Maximum
Real Time
1. Press [MODE] to step from one option to the next. The highlighter indicates your
choice.
2. Press [Y/+] to toggle your selection on or off (the check box indicates “on” with
an “X”).
3. When you have completed your selections, press [MODE] repeatedly until “Exit”
is highlighted.
4. Press [Y/+] to save the changes and exit.
Datalog Type
The instrument has two datalog types:
Auto
Manual
Collects datalog information when the instrument is sampling until the
datalog memory is full.
Datalogging occurs only when you manually initiate datalogging (see
below for details).
1. Press [MODE] to step from one option to the next.
2. Press [Y/+] to make your selection (the dark circle in the “radio button” indicates
that an option has been selected).
3. When you have completed your selection, press [MODE] until “Exit” is
highlighted.
4. Press [Y/+] to save your selection and exit.
About Manual Datalog
When the instrument is set to Manual Datalog, you can turn datalogging on and off from
the main measurement display in the Normal Mode by pressing the [MODE] key three
times.
•
•
When you reach the screen that says “Start Datalog?” press [Y/+] to start it. You
see “Datalog Started,” confirming that datalogging is now on.
When you reach the screen that says “Stop Datalog?” press [Y/+] to stop it. You
see “Datalog Stopped,” confirming that datalogging is now off.
Exit (the Datalog Submenu)
Press [Y/+] to exit. Datalog is exited, and Monitor Setup is highlighted.
31
ToxiRAE Pro PID User’s Guide
11.5.5. Monitor Setup
Wireless
There are seven Wireless parameters:
•
•
•
•
•
•
•
•
Radio On/Off
Roaming
PAN ID
Channel
Join Network
Interval
Off Network Alarm
Factory Reset
Radio On/Off
This turns the wireless modem on or off.
Note: Turning the radio off does not change any of the settings. They are
stored and ready for use the next time the wireless modem is turned on.
1. Choose an “On” or “Off” by pressing [MODE] until your choice is highlighted. (the dark
circle in the “radio button” indicates that an option has been selected).
2. Press [Y/+] to select it.
3. Press [MODE] repeatedly until “Exit” is highlighted.
4. Press [Y/+] to register your selection and exit.
To exit without saving the new value:
1. Make sure your previous setting is selected.
2. Press [MODE] repeatedly to advance to Exit.
3. Press [Y/+] to exit.
Roaming
The Roaming function provides continuous wireless connectivity
between zones enabling users of wireless monitors to travel from one
zone/work area to another without losing communication between
their monitor and ProRAE Guardian. You can turn Roaming on and
off in your monitor, or you can set it via ProRAE Studio II.
Note: When Roaming is on, you cannot change the instrument’s
PAN ID. When Roaming is on, the “PAN ID” is not shown in the
Wireless menu. To turn “PAN ID” back on:
1.
2.
3.
4.
5.
6.
While in the Wireless menu, press [MODE] to scroll to “Roaming.”
Press [Y/+] to select “Roaming.”
Press [MODE] to scroll to “Off.”
Press [Y/+] to select.
Press [MODE] to scroll to “Exit.”
Press [Y/+] to exit.
32
ToxiRAE Pro PID User’s Guide
The message “Applying Settings” is shown and then the menu advances
automatically to “PAN ID.”
PAN ID
All units on a network must be programmed with the same Pan ID
(Personal Network Identifier) to ensure communication compatibility.
Important! The allowed range for Pan ID numbers is 001 through
999.
Note: When Roaming is on, you cannot change the instrument’s PAN
ID. When Roaming is on, the “PAN ID” is not shown in the Wireless
menu.
Choose and enter a 3-digit PAN ID:
1. Press [Y/+] and the display shows the current PAN ID. Example: “866.” The leftmost digit is highlighted.
2. Press [Y/+] repeatedly to select a desired number. Once 9 is reached, pressing
[Y/+] again “wraps” around back to 0.
3. Press [MODE] to advance to the next digit.
Repeat this process until all eight characters of the new PAN ID are entered.
To save your new PAN ID and exit, press [MODE] repeatedly until “OK” is
highlighted. Then press [Y/+]. It says, “Applying settings…” and then “Settings
applied successfully!” to indicate they have been registered. The menu now advances
automatically to “Channel.”
Channel
Every network operates on its own channel, between 01 and 10 (do not
use 00). If you need to know the network’s channel, check the
EchoView Host’s setting. Note: In Europe, only Channel 00 can be
used, and it is preset (it cannot be changed).
Choose and enter a 2-digit Channel:
1. Press [Y/+] and the display shows the current Channel. Example: “10.” The leftmost digit is highlighted.
2. Press [Y/+] repeatedly to select a desired number. Once 9 is reached, pressing
[Y/+] again “wraps” around back to 0.
3. Press [MODE] to advance to the next digit.
Repeat this process until all eight characters of the new Channel are entered.
To save your new Channel number and exit, press [MODE] repeatedly until “OK” is
highlighted. Then press [Y/+]. It says, “Applying settings…” and then “Settings
applied successfully!” to indicate they have been registered. The menu now advances
automatically to “Join Network.”
33
ToxiRAE Pro PID User’s Guide
Join Network
This screen is for joining an existing network. When you see this
screen, press [Y/+] to join, or [MODE] to exit without joining.
While searching for the network, the message “Joining network…” is displayed.
If the network is found and successfully joined, then “Join network successful!” is
displayed. Press [MODE] to acknowledge the connection and return to the Wireless
menu.
If the network is not joined, the display shows: “Join network failed!” You have two
options: Exit and Retry. If you want to retry, check that your network setting is
correct (matches the EchoView Host) and that the EchoView Host is turned on and
operating. Then press [Y/+] to retry.
Interval
This sets the amount of time between data transmissions to the
EchoView Host.
To change the Interval:
1. Press [MODE] to step from one option to the other. The highlighted option
changes each time you press [MODE].
2. Press [Y/+] to make your selection (the dark circle in the “radio button” indicates
the option has been selected).
3. When you have completed your selection, press [MODE] repeatedly until “Exit”
is highlighted.
4. Press [Y/+] to save your selection, exit the Interval menu, and advance to “Off
Netwk Alarm.”
Off Netwk Alarm When this is enabled, the unit goes into alarm whenever an active
network connection is lost. This informs you that connectivity has
been broken between the ToxiRAE Pro and the EchoView Host.
1. Choose “On” or “Off” by pressing [MODE] until your choice is highlighted. (the
dark circle in the “radio button” indicates that an option has been selected).
2. Press [Y/+] to select it.
3. Press [MODE] repeatedly until “Exit” is highlighted.
4. Press [Y/+] to register your selection and exit.
To exit without saving the new value:
1. Make sure your previous setting is selected.
2. Press [MODE] repeatedly to advance to Exit.
3. Press [Y/+] to exit.
34
ToxiRAE Pro PID User’s Guide
Factory Reset
CAUTION!
Selecting this resets the wireless settings to their original factory
configurations.
A factory reset of the wireless settings deletes any customized
settings and cannot be undone.
1. Press [Y/+] to select “Factory Reset.” You will see the message “Reset wireless
settings?”
2. Press [Y/+] for “Yes” or [MODE] for “Exit.” If you select “Exit,” the ToxiRAE
Pro exits without performing a factory reset.
Exit (the Wireless Submenu)
Press [Y/+] to exit. Wireless is exited, and Monitor Setup is highlighted.
Site ID
Choose and enter an 8-digit alphanumeric Site ID to uniquely identify the particular site where
the instrument is to be used. This Site ID is included in the datalog report.
1. Press [Y/+] and the display shows the current Site ID. Example: “SITE0001.” The leftmost digit is highlighted.
2. Press [Y/+] repeatedly to select a desired character. Letters increase from current value to
Z followed by digits from 0 to 9. Once 9 is reached, pressing [Y/+] again “wraps” around
to the letter A. For the last four characters that must be numerical, numbers increase from
0 to 9. Once 9 is reached, pressing [Y/+] again “wraps” around back to 0.
Note: The last four characters must be numerals.
3. Press [MODE] to advance to the next character.
Repeat this process until all eight characters of the new Site ID are entered.
To save your new Site ID and exit, press [MODE] repeatedly until “OK” is highlighted. Then
press [Y/+].
User ID
Enter an 8-digit alphanumeric User ID to uniquely identify a user. This User ID is
included in the datalog report.
1. Press [Y/+] and the display shows the current User ID. Example: “USER0000.”
The left-most character is highlighted.
2. Press [Y/+] repeatedly to select a desired character. Letters increase from current
value to Z followed by digits from 0 to 9. Once 9 is reached, pressing [Y/+] again
“wraps” around to the letter A.
3. Press [MODE] to advance to the next character.
Repeat this process until all eight characters of the new User ID are entered.
To save your new User ID and exit, press [MODE] repeatedly until “OK” is highlighted.
Then press [Y/+].
35
ToxiRAE Pro PID User’s Guide
User Mode
The ToxiRAE Pro PID has two user modes:
Basic
Basic users can only see and use a basic set of functions.
Advanced
Advanced users can see all screens and take advantage of all available
functions.
Note: The default value for User Mode is Basic.
To change the User Mode:
1. Press [MODE] to step from one option to the other. The highlighting changes
each time you press [MODE].
2. Press [Y/+] to make your selection (the dark circle in the “radio button” indicates
the option has been selected).
3. When you have completed your selection, press [MODE] repeatedly until “Exit”
is highlighted.
4. Press [Y/+] to register your selection, exit the User Mode submenu and advance
to Date.
Date
The Date is expressed as Month/Day/Year, with two digits for each (MM/DD/YY).
The left-most digit is highlighted to indicate it is selected when you enter the Date
submenu.
1. Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to 9.
Once 9 is reached, pressing [Y/+] again “wraps” around back to 0.
2. Press [MODE] to advance to the next digit.
Repeat this process until all six digits of the new date are entered.
3. Press [MODE] repeatedly until “OK” is highlighted.
4. Press [Y/+] to save your selection and advance to Time.
Note: If you do not want to save the change, press [MODE] repeatedly until
“Cancel” is highlighted. Press [Y/+] to exit to Time without saving the changes.
Time
The Time is expressed as Hours/Minutes/Seconds, with two digits for each. The time is in
24-hour (military) format of HH:MM:SS.
The left-most digit is highlighted to indicate it is selected.
1. Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to 9.
Once 9 is reached, pressing [Y/+] again “wraps” around back to 0.
36
ToxiRAE Pro PID User’s Guide
2. Press [MODE] to advance to the next digit.
Repeat this process until all six digits of the new time are entered.
3. Press [MODE] repeatedly until “OK” is highlighted.
4. Press [Y/+] to save your selection, exit the Time submenu and advance to Temp.
Unit.
Note: If you do not want to save the change, press [MODE] until “Cancel” is
highlighted. Press [Y/+] to exit to Temp. Unit without saving the changes.
Temp. Unit
The display unit of the internal temperature sensor can be switched between Fahrenheit
and Celsius.
To change the Temp. Unit:
1. Press [MODE] to step from one option to the other. The highlighted option
changes each time you press [MODE].
2. Press [Y/+] to make your selection (the dark circle in the “radio button” indicates
the option has been selected).
3. When you have completed your selection, press [MODE] repeatedly until “Exit”
is highlighted.
4. Press [Y/+] to save your selection, exit the Temp. Unit menu, and advance to
Language.
Language
English is the default language, but other languages can also be selected for the
instrument.
1. Press [MODE] to step from one option to the next. The highlighted option
changes each time you press [MODE].
2. Press [Y/+] to make your selection (the dark circle in the “radio button” indicates
this option has been selected).
3. When you have completed your selection, press [MODE] repeatedly until “Exit”
is highlighted.
4. Press [Y/+] to save your selection, exit the Language submenu, and advance to
Zero At Start.
Zero At Start
When Zero At Start is turned on, the instrument starts with a zero calibration every time
it is turned on.
1. Press [MODE] to choose between “Off” and “On.”
2. Press [Y/+] to make your selection (the dark circle in the “radio button” indicates
your selection).
3. When you have completed your selection, press [MODE] repeatedly until “Exit”
is highlighted.
37
ToxiRAE Pro PID User’s Guide
4. Press [Y/+] to save your selection, exit the Zero At Start submenu, and advance to
LCD Contrast.
LCD Contrast
The display’s contrast can be increased or decreased from its default setting. You may
not need to ever change the default setting, but sometimes you can optimize the display to
suit extreme temperature and ambient brightness/darkness conditions.
1. Press and hold [Y/+] to select a desired level of contrast. The range is 1 to 100.
Once 100 is reached, pressing [Y/+] again "wraps" around back to 1.
2. When you have completed your selection, press [MODE] repeatedly until “OK” is
highlighted.
3. Press [Y/+] to save your selection, and exit the LCD Contrast.
Note: If you do not want to save the change, press [MODE] repeatedly until
“Cancel” is highlighted. Press [Y/+] to exit without saving the changes.
Exit (Monitor Setup Submenu)
Press [Y/+] to exit. Monitor Setup is exited. Exit (to exit Programming Mode) is
highlighted.
11.6. Exit (Exit Programming Mode)
To exit Programming Mode and return to normal operation, press [MODE] until the Exit
icon is highlighted. Press [Y/+] to exit. You will see “Updating Settings…” as changes
are registered and the mode changes to normal operation.
38
ToxiRAE Pro PID User’s Guide
12. Policy Enforcement
The ToxiRAE Pro PID can be configured to enforce a facility/company’s requirements
that calibration and/or bump testing be performed at specified intervals, and to warn the
user that calibration/bump testing is required. In addition, it can require calibration or
bump testing and not allow normal operation of the instrument unless calibration or bump
testing is performed.
If Policy Enforcement is enabled and if the instrument has been bump tested and
calibrated in compliance with the policy settings, a check-mark icon is included along the
top of the ToxiRAE Pro PID screen:
If Policy Enforcement is enabled, then after startup the ToxiRAE Pro PID displays a
screen that informs the user that the instrument requires either a bump test or a
calibration. If both are required, then they are shown in sequence.
12.1. Setting Policy Enforcement
You must use ProRAE Studio II to make changes to Policy Enforcement settings. The
procedure differs, depending on whether you are using an AutoRAE 2 Cradle, a
ToxiRAE Pro PID Travel Charger, or a ToxiRAE Pro PID Desktop Cradle.
12.1.1. AutoRAE 2 Cradle
Note: This procedure only works with a single AutoRAE 2 Cradle. It will not work with
an AutoRAE 2 Controller.
To program a ToxiRAE Pro PID via an AutoRAE 2 Cradle, you need ProRAE Studio II
Instrument Configuration and Data Management Software, the AutoRAE 2 Cradle
connected to a power source, and a USB PC communications cable.
1. Connect a USB cable between a PC with ProRAE Studio II and the AutoRAE 2
Cradle.
2. Apply power to the AutoRAE 2 Cradle.
3. Turn off the ToxiRAE Pro PID (or put the ToxiRAE Pro PID into PC
Communications Mode) and set it in the cradle.
4. Start ProRAE Studio II software on the PC.
5. Select “Administrator” and input the password (the default is “rae”).
6. Click “Detect the instruments automatically” (the magnifying glass icon with the
letter “A” in it). After a few seconds, the AutoRAE 2 Cradle is found and it is shown,
along with its serial number:
39
ToxiRAE Pro PID User’s Guide
7. Click on the icon to highlight it, and then click “Select.”
40
ToxiRAE Pro PID User’s Guide
8. In ProRAE Studio II, the AutoRAE 2 Cradle is shown, including its Serial Number,
under “Online”:
9. Expand the view to show the ToxiRAE Pro PID in the AutoRAE 2 Cradle by clicking
the “+” to the left of the image of the AutoRAE 2 Cradle:
10. Double-click on the icon representing the ToxiRAE Pro PID.
11. Click “Setup.”
41
ToxiRAE Pro PID User’s Guide
12. In the menu that now appears on the left side, click “Policy Enforcement.” It is
highlighted, and the Policy Enforcement pane is shown:
For “Must Calibrate” and “Must Bump,” you have the options of no enforcement or
enforcement (including “Can’t Bypass,” and “Can Bypass”).
Must Calibrate. The user is prompted to calibrate the instrument when calibration is due
(as set by the calibration interval). There are two programmable options:
•
•
Can’t Bypass. Unless calibration is performed, the instrument cannot be used,
and the only option is to turn off the instrument.
Can Bypass. If calibration is due but the user does not want to perform a
calibration, the instrument can still be used.
Must Bump
•
•
Can’t Bypass. Unless a bump test is performed, the instrument cannot be used,
and the only option is to turn off the instrument.
Can Bypass. If a bump test is due but the user does not want to perform one, the
instrument can still be used.
42
ToxiRAE Pro PID User’s Guide
These are the screens that are shown on a ToxiRAE Pro PID after startup if “Can Bypass”
is selected:
If “Can’t Bypass” is selected, the display looks like this, and only allows the options of
performing the test or shutting down:
12.1.2. ToxiRAE Pro PID Desktop Cradle Or Travel Charger
Make sure the AC adapter is connected and that a USB cable is connected to a computer
running ProRAE Studio II.
1.
2.
3.
4.
5.
Turn on the ToxiRAE Pro PID.
Hold down [MODE] and [Y/+] to enter Programming Mode.
Provide the password.
Press [MODE] until “Enter Communications Mode?” is displayed.
Press [Y/+]. The screen shows that it is ready by saying “Communications Mode
ready…” It is now ready to communicate with the computer running ProRAE
Studio II. If you want to exit, press [MODE].
6. Start ProRAE Studio II.
7. Select “Administrator.”
8. Input the password (the default is “rae”).
9. Click “OK.”
10. Click “A” (detects instruments automatically).
11. Click on the instrument’s icon when it appears to highlight it.
12. Click “Select.”
13. Click “Setup.”
14. Click “Policy Enforcement.”
43
ToxiRAE Pro PID User’s Guide
The Policy Enforcement pane is shown:
You can select “Must Calibrate” and/or “Must Bump” and then set whether the user
must perform the selected operation in order to use the instrument.
15. Once you have made your selections in ProRAE Studio II, you must upload the
changes to the instrument. Click the icon labeled “Upload all settings to the
instrument.”
16. A confirmation screen is shown. Click “Yes” to perform the upload, or “No” to abort.
Uploading takes a few seconds, and this progress bar is shown. You can abort the
upload by clicking “Cancel.”
44
ToxiRAE Pro PID User’s Guide
17. Exit ProRAE Studio II.
18. Press [Y/+] on the ToxiRAE Pro PID to exit Communication Mode.
12.2. Deactivating Policy Enforcement
12.2.1. AutoRAE 2 Cradle
To deactivate Policy Enforcement when using an AutoRAE 2 Cradle, follow the
procedure for changing settings. See page 39 for details.
12.2.2. ToxiRAE Pro PID Desktop Cradle Or Travel Charger
If the ToxiRAE Pro PID screen displays the message that it must be bump tested or
calibrated, and if the option to bypass bump testing or calibration is not available, you
should shut off the instrument and follow the procedure outlined here if you want to
change the Policy Enforcement settings:
1. Use a USB cable to connect the ToxiRAE Pro PID in its Travel Charger or Desktop
Cradle to a computer running ProRAE Studio II.
2. Enter Diagnostic Mode on the ToxiRAE Pro PID (with the instrument turned off,
press and hold [Y/+] and [MODE] until it starts up.
3. After startup, enter the password when prompted (default is “0000”) and press
[MODE].
4. Press [MODE] repeatedly until you see the “Enter Communications Mode?” screen.
5. Press [Y/+] to enter Communications Mode.
6. Start ProRAE Studio II.
7. Select “Administrator.”
8. Input the password (the default is “rae”).
9. Click “OK.”
10. Click “A” (detect instruments automatically).
11. Click on the instrument’s icon when it appears.
12. Click “Select.”
13. Click “Setup.”
14. Click “Policy Enforcement.” The Policy Enforcement pane is shown.
15. Deselect Policy Enforcement features you do not wish to use.
16. Click “Upload all settings to the instrument.”
45
ToxiRAE Pro PID User’s Guide
17. When you see this confirmation. Click “Yes.”
Uploading will take a few seconds, and this progress bar is shown:
18. When the upload is done, exit ProRAE Studio II.
19. Press [Y/+] on the ToxiRAE Pro PID to exit Communication Mode.
46
ToxiRAE Pro PID User’s Guide
13. Bump Testing & Calibrating The ToxiRAE Pro PID
Note: Calibration intervals and bump test procedures may vary due to national
legislation.
13.1. Bump Testing & Calibrating With An AutoRAE 2™
Bump testing and calibrating a ToxiRAE Pro PID can be done with an AutoRAE 2
Automatic Test And Calibration Station. Follow all instructions in the AutoRAE 2 User’s
Guide for successful bump testing or calibration.
13.1.1. Bump (Functional) Testing
With the ToxiRAE Pro PID in Normal Mode:
1. Connect the calibration gas cylinder, flow regulator, and calibration adapter to the
ToxiRAE Pro PID and start the gas flow.
2. Make sure the unit goes into alarm and that the buzzer produces at least two beeps per
second, the LED lights flash on and off, and the vibration alarm functions. The
display backlight should illuminate and an alarm message should be shown in the
display.
3. Turn off the gas flow.
4. Remove the calibration adapter.
Note: Bump testing and calibration can also be performed by using the RAE Systems
AutoRAE 2 Automatic Test And Calibration Station. Refer to the AutoRAE 2 User’s
Guide for instructions.
13.2. Calibration Adapter
The ToxiRAE Pro PID must be calibrated using a fixed-flow regulator with a flow rate
between 0.5 and 1.0 liters per minute. The ToxiRAE Pro PID is supplied with a special
calibration adapter that covers the gas fan inlet. If the local air has impurities that might
interfere with zero (fresh air) calibration, the calibration adapter and a cylinder of clean
zero air should be used.
13.2.1. Connecting The Calibration Adapter
1. Place the Calibration Adapter on the top cap of the ToxiRAE Pro PID by pressing
it into place.
2. Attach the hose from the regulator/calibration gas cylinder to the Calibration
Adapter’s inlet.
47
ToxiRAE Pro PID User’s Guide
Caution! After calibration is complete, remove the Calibration Adapter. When
monitoring, never operate the ToxiRAE Pro PID with the Calibration Adapter attached. If
the Calibration Adapter is attached during normal operation, inconsistent and lower-thannormal readings will occur because of decreased concentration of the gas being
monitored.
13.2.2. Disconnecting The Calibration Adapter
After calibrating the ToxiRAE Pro PID, you must remove the Calibration Adapter by
lifting it off the ToxiRAE Pro PID.
13.3. Zero Calibration
This operation sets the zero point of the sensor calibration curve for clean air. Expose the
ToxiRAE Pro PID to a clean air source with 20.9% oxygen and without any organic,
toxic, or combustible gas impurities. This “Zero Air” can come from a cylinder or clean
ambient air.
Note: If you use a zero air cylinder, you must use the ToxiRAE Pro PID Calibration
Adapter. Using a calibration adapter is not necessary for calibration in fresh air.
Turn on the ToxiRAE Pro PID. Once it has been through its startup, enter Programming
Mode:
1. Press and hold [MODE] and [Y/+] until you see the Password screen.
2. Input the 4-digit password:
•
•
•
•
Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to
9. Once 9 is reached, pressing [Y/+] again “wraps” around back to 0.
Step from digit to digit using [MODE].
When you are done, press [MODE] until OK or Cancel is highlighted.
Press [Y/+] to register the password (if OK is highlighted) or to exit without
entering the password (if Cancel is selected).
48
ToxiRAE Pro PID User’s Guide
If you make a mistake, you can cycle through the digits by pressing [MODE] and
then using [Y/+] to change the number in each position.
Note: The default password is 0000.
If you do not know the password, do not enter anything. Just press [MODE]
repeatedly to highlight OK, and then press [Y/+].
Note: If the ToxiRAE Pro PID is in Programming Mode and you do not press a key
within 60 seconds, it exits Programming Mode and re-enters Normal Mode.
1. At the Calibration screen, press [Y/+].
2. “Zero Calib” is highlighted. Make sure the ToxiRAE Pro PID is in clean air.
Otherwise attach a cylinder of zero gas to the Calibration Adapter installed on the
instrument.
3. Press [Y/+]. The message “Apply zero gas…” is displayed.
4. Start the flow of the zero gas (if used) and press [Y/+] to start the zero calibration.
5. There is a 60-second countdown, during which time the screen shows the word
“Zeroing…” and the countdown.
6. Upon completion, it says “Zeroing is done! Reading = 0.0ppm” along with the
reading (it should be 0.0 ppm or very close to it).
7. “Span Calib” is highlighted.
Note: You can abort the zero calibration during its countdown by pressing the [MODE] key.
Important! If you used the Calibration Adapter and a zero air cylinder, you must remove
the Calibration Adapter from the ToxiRAE Pro PID. If the Calibration Adapter is
attached during normal operation, inconsistent and lower-than-normal readings will occur
because of decreased concentration of the gas being monitored.
13.4. Span Calibration
This procedure determines the second point of the sensor calibration curve for the sensor.
To begin calibration, connect the calibration gas cylinder, flow regulator, and calibration
adapter to the ToxiRAE Pro PID.
With the ToxiRAE Pro PID in Programming Mode, and with Span Calib highlighted,
follow this procedure:
1. Press [Y/+]. The screen displays the current (or default) calibration gas and its
concentration. For example:
Cal. Gas: Isobutylene
Span=100ppm
Change?
•
Press [MODE] if you do not want to change it.
49
ToxiRAE Pro PID User’s Guide
•
Press [Y/+] if you want to change it.
If you entered the Calibration Menu without supplying the correct password, or
supplied the correct password but chose not to change the Cal.Gas, you will see a
message like this:
Cal.Gas: Isobutylene
Span=100ppm
Apply gas…
2. Press [Y/+] to start calibration.
3. Press [MODE] if you want to quit without calibrating.
During calibration, the calibration gas name, its concentration, and the countdown are
displayed. Depending upon the device alarm settings, the monitor may go into low and
then high alarm once span calibration is complete, which is expected in response to the
test gas.
Note: You can abort span calibration during its countdown by pressing [MODE].
Upon completion, the screen “Span is done!” appears along with the reading (it should be
within 10% of the span gas concentration).
Important! After calibration is complete, you must remove the Calibration Adapter from
the ToxiRAE Pro PID. If the Calibration Adapter is attached during normal operation,
inconsistent and lower-than-normal readings will occur because of decreased
concentration of the gas being monitored.
13.4.1. Changing Span Gas Concentration
If you selected “Change” for span gas concentration, the following screen appears:
1. Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to 9.
Once 9 is reached, pressing [Y/+] again “wraps” around back to 0.
2. Step from one digit to the next by pressing [MODE].
3. When you are done, press [MODE] until either “OK” or “Cancel” is selected.
Press [Y/+] to register your selection (if “OK” is highlighted) or to exit without
registering it (if “Cancel” is selected).
50
ToxiRAE Pro PID User’s Guide
14. Diagnostic Mode
In Diagnostic Mode, the ToxiRAE Pro PID provides raw counts for sensor, battery, and
other readings. Most of these screens are useful only to service technicians. Therefore, if
you require information on Diagnostic Mode, refer to the ToxiRAE Pro PID Service
Guide.
The ToxiRAE Pro PID’s Diagnostic Mode can only be accessed at startup time. In
Diagnostic Mode, ToxiRAE Pro PID displays readings in raw counts instead of units such as
parts per million (ppm).
14.1. Entering Diagnostic Mode
1. With the ToxiRAE Pro PID turned off, press and hold both [MODE] and [Y/+].
2. When the display turns on and the password screen appears, release the keys.
3. Enter the 4-digit password (the password is the same as the one for the
Programming Mode):
•
•
•
Step from one position in the four-character string to the other by pressing
[MODE].
Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to
9.
Once 9 is reached, pressing [Y/+] again “wraps” around back to 0.
4. When you are done, press [MODE] until either OK or Cancel is selected. Press
[Y/+] to register the password (if “OK” is highlighted) or to exit without entering
the password (if “Cancel” is selected).
14.2. Navigating Diagnostic Mode
Step through Diagnostic Mode by pressing [MODE].
14.3. Exiting Diagnostic Mode
1. Turn off the ToxiRAE Pro PID by pressing and holding [MODE]. There will be a
standard shutoff countdown.
2. When the instrument shuts off, you will be alerted. Release your finger.
Note: The next time you start ToxiRAE Pro PID, hold only [MODE], and it will
automatically start in Normal Mode.
51
ToxiRAE Pro PID User’s Guide
15. Alarm Signal Summary
Alarm types include:
Alarm
Message/
Priority
Condition
Alarm Signal
Super Alarm
Man Down Alarm
4 beeps/flashes per second
3 beeps/flashes per second
Man Down Warning
2 beeps/flashes per second
Lamp failure
Gas exceeds electronics’
maximum range
Gas exceeds measurement
range
3 beeps/flashes per second
3 beeps/flashes per second
Gas exceeds “High Alarm”
limit
Gas exceeds “Low Alarm”
limit
3 beeps/flashes per second
1 beep/flash per second
STEL*
Gas reading measures less
than zero
Gas exceeds “STEL” limit
TWA*
Gas exceeds “TWA” limit
1 Beep/flash per second
CAL
Calibration failed, or needs
calibration
Offline from network
communcations
1 beep/flash per second
Low battery
1 flash, 1 beep per minute plus
battery icon flashes on display
(Man
Down)
(Man Down
Warning)
LAMP
MAX
OVER
HIGH
LOW
NEG
Offline
Bat
3 beeps/flashes per second
2 beeps/flashes per second
1 Beep/flash per second
1 beep/flash per second
The alarm priority is as follows (in the event of multiple simultaneous alarms):
Super Alarm > Man Down Alarm > Man Down Warning > Lamp > Fan > Max > Over >
High > Low > Neg > STEL > TWA > Cal Failed > Bump Failed > Cal Required > Bump
Required > Offline > Bat
52
ToxiRAE Pro PID User’s Guide
16. Downloading Datalogs And Performing PC-Based
Instrument Configuration and Firmware Upgrades
The ToxiRAE Pro PID communicates with a PC running ProRAE Studio II Instrument
Configuration and Data Management software (version 1.03 or higher) to download
datalogs, configure the instrument, or upload new firmware.
The ToxiRAE Pro PID must be connected to a PC through the supplied Charging / PC
Communications Cradle and must be in the PC communications mode.
1. Make sure the Charging/PC Communications Cradle has power from its AC
adapter (the LED on the front of the cradle should be illuminated).
2. Use the supplied PC Communications Cable (USB to mini-USB cable) to connect
the Cradle to a PC.
3. Turn on the ToxiRAE Pro PID. Make sure it is running in Normal mode (with the
main measurement screen with VOC reading showing).
4. Insert the ToxiRAE Pro PID in the cradle.
5. Activate the PC communications mode on the ToxiRAE Pro PID by pressing
[MODE] repeatedly, starting from the main measurement screen until you reach
the “Enter Communications Mode and stop measurement?” screen.
6. Press [Y/+]. Measurement and datalogging stop, and the instrument is now ready
to communicate with the PC. The display now says “PC comm ready…”
7. Start up the ProRAE Studio II software, enter a password, and detect the
instrument following the directions provided in the ProRAE Studio II User’s
Manual.
8. Follow the instructions in the ProRAE Studio II User’s Manual to download the
datalog, configure the instrument settings, or update the ToxiRAE Pro PID’s
firmware.
9. When you are done, press [MODE] to exit the PC communications mode on the
ToxiRAE Pro PID. The screen displays “Sensor initializing,” after which the
instrument returns to operating in Normal mode.
53
ToxiRAE Pro PID User’s Guide
17. Maintenance
WARNING!
Maintenance should be performed only by a qualified person who has proper
training and fully understands the contents of this manual.
The following guidelines should be followed when changing the battery:
1. Turn off the unit before changing the battery.
2. When replacing the battery, take note of its orientation. The connectors on the circuit
board should go into the matching slots in the battery.
WARNING
To reduce the risk of ignition of hazardous atmospheres, replace the battery only in
areas known to be non-hazardous.
17.1. Replacing the ToxiRAE Pro PID Li-Ion Battery
Caution: Turn off the ToxiRAE Pro PID before removing or replacing the battery.
1. If the rubber boot is on the ToxiRAE Pro PID, remove it (refer to page 58 for details).
2. Place the ToxiRAE Pro PID face-down on a soft surface.
3. Use a Philips screwdriver to loosen the screw holding the top cap in place, and then
carefully use the provided tool to pry up the top cap from the middle in the front and
back and use your fingers to remove the top cap.
54
ToxiRAE Pro PID User’s Guide
4. Remove the four screws on the back of the instrument by turning them counterclockwise.
5. Hold the ToxiRAE Pro PID down and separate the rear cover from the front by
pulling from the top and tilting it open without disconnecting the ribbon cable
connecting the two halves.
Ribbon cable
6. Carefully disconnect the battery from the battery connectors on the circuit board
by pulling it toward the bottom of the instrument. Be careful not to damage the
ribbon cable.
55
ToxiRAE Pro PID User’s Guide
7. Before installing a new battery, visually inspect the contact pins on the ToxiRAE
Pro PID circuit board to make sure they are not bent. If they are, carefully
straighten them before installing the battery. Install the new battery (part number
G02-3004-000), making sure it is inserted fully. Make sure the battery connectors
on the circuit board go all the way into the connector slots in the battery.
8. Inspect to make sure the ribbon connector is still connected, and plug it back in if
it has come unplugged.
9. Place the cover over the compartment by tilting it up and then fitting it in place.
10. Tighten all four screws by turning them clockwise with a torque-limiting
screwdriver (preferably an electric one). Tighten them in the numerical order
shown here, to a torque rating of 3.54 to 4.43 lbf-in (4.08 to 5.10 kgf-cm).
Note: Incorrectly tightening the screws may compromise the instrument’s
IP rating.
56
ToxiRAE Pro PID User’s Guide
11. Replace the top cap and tighten the Philips screw.
Important! Charge the battery completely before using the ToxiRAE Pro PID, and
check the calibration before placing the instrument in service.
17.2. Proper Battery Disposal
The Waste Electrical and Electronic Equipment (WEEE) directive
(2002/96/EC) is intended to promote recycling of electrical and electronic
equipment and their components at end of life. This symbol (crossed-out
wheeled bin) indicates separate collection of waste electrical and electronic
equipment in the EU countries. This product may contain one or more
Nickel-metal hydride (NiMH), Lithium-ion, or Alkaline batteries. Specific
battery information is given in this user guide. Batteries must be recycled or
disposed of properly.
At the end of its life, this product must undergo separate collection and recycling from
general or household waste. Please use the return and collection system available in your
country for the disposal of this product.
Sensor Specifications, Cross-Sensitivities, And Calibration
Information
For information on sensor specifications, cross-sensitivities, and calibration information,
refer to RAE Systems Technical Note TN-114: Sensor Specifications And CrossSensitivities (available for free download from www.raesystems.com). All specifications
presented in this Technical Note reflect the performance of standalone sensors. Actual
sensor characteristics may differ when the sensor is installed in different instruments. As
sensor performance may change over time, specifications provided are for brand-new
sensors.
57
ToxiRAE Pro PID User’s Guide
17.3. Removing And Installing The Rubber Boot
The protective rubber boot on the ToxiRAE Pro PID must be removed before the sensor
top cap or back of the instrument can be removed. When removing the boot, follow these
steps outlined here.
1. Flip up the tab on the belt clip on the rear of the instrument.
2. Remove the Philips screw that holds the clip in place.
3. Stretch the bottom corners of the boot and gently slide the bottom of the boot over
the rear of the ToxiRAE Pro PID.
4. Slide the boot up until it comes off of the instrument. Make sure that you do not
force it to move if it sticks against the top cap. Gently nudge it.
5. Inspect the rubber boot for damage, and replace it if it is cracked or torn.
When replacing the boot, simply reverse the procedure, sliding the boot onto the
instrument and then carefully stretching it over the bottom portion of the instrument. Do
not forget to replace the belt clip.
58
ToxiRAE Pro PID User’s Guide
17.4. Filter Replacement
A filter is permanently attached inside the ToxiRAE Pro’s top cap, which helps to keep
debris and moisture out. This filter should be inspected periodically, and the cap and
rubber insert should be inspected for damage. If the filter is dirty or the cap or rubber
insert shows signs of damage, replace the entire instrument cap assembly, including the
cap, screw, filter, and the rubber insert (part number G02-0305-003, package of three).
To replace the cap assembly:
1. Remove the top cap. First remove the small Phillips screw on the back side:
2. Then use the special tool (part number G02-0306-003, package of three) and
apply leverage at the middle of the instrument to push the top cap loose, and then
remove it with your fingers.
59
ToxiRAE Pro PID User’s Guide
3. Once the top cap is removed, install the new top cap assembly.
4. Place the new rubber insert onto the ToxiRAE Pro PID, being careful to align the
insert correctly and that it is not distorted. Also, make sure that you do not press
on the fan.
5. Press the new top cap onto the ToxiRAE Pro until it is fully seated. Check to see
that the pin on the front side of the cap went into the corresponding groove on the
front side of the instrument housing. Then tighten the Phillips screw.
60
ToxiRAE Pro PID User’s Guide
17.5. Cleaning Or Replacing The PID
Follow the instructions for removing the top cap in the previous section.
1. Gently lift out the PID module with your fingers, being careful not to damage the
fan:
Fan
Sensor module
2. If the module requires replacement (because the lamp does not illuminate, or the
module has been used past its expiration date), place a new module into the slot,
being careful to match the indexing keys. The sensor can only go into its slot one
way. Refer to step 11 for alignment information.
3. If you want to open the sensor module to inspect and clean the lamp and sensor
electrode panel, you must use the special tool. The special tool’s “C”-shaped end
has small “teeth” inside. Slide the tool so that the teeth slip into the notch between
the module’s cap and body:
4. Gently pry up the cap using a rocking motion:
61
ToxiRAE Pro PID User’s Guide
5. Once the cap is removed, set it aside.
6. Now lift the sensor electrode panel from the module:
7. Clean the sensor electrode panel in a solution of methanol lamp cleaner (included,
along with cleaning swabs, in a PID Lamp Cleaning Kit, part number 500-0014010, package of 10), and allow it to dry.
8. Clean the lamp’s window with a cleaning swab dipped in methanol lamp cleaner,
and allow it to dry. Do not touch the lamp window with your fingers, as the
residual oils will shorten its life.
9. Inspect the electrical contacts. Clean them with a swab dipped in lamp cleaner if
they appear to need cleaning:
62
ToxiRAE Pro PID User’s Guide
10. Reassemble the sensor module by placing the sensor electrode panel back in place
and firmly pressing the cap back onto the top.
11. Place the sensor module back into the ToxiRAE Pro PID. Make sure the index
points are aligned (it can only go in one way).
Match index
points on sensor
module with
ToxiRAE Pro PID
12. Carefully place the cap back onto the ToxiRAE Pro PID and tighten the screw.
Note: Always calibrate the ToxiRAE Pro PID after replacing the sensor module.
17.6. Replacing The Fan
If the fan stops working or is damaged, it can be replaced (part number G02-3006-000).
Before replacing the fan, make sure the instrument is turned off.
1. Remove the top cap as shown in Section 15.2 (page 59).
2. Grip the fan with two fingers and pull up.
63
ToxiRAE Pro PID User’s Guide
3. Inspect the new fan and make sure its two electrical pins are straight.
4. Align the fan so that the indexing key (the raised rectangular portion) on its body
matches with the keyway in the slot and the electrical pins and carefully slide into
the contact points at the bottom of the slot.
Important! If you encounter resistance while pressing the new fan into the slot,
remove it and check that the electrical pins are not bent and that all alignment is
correct before inserting the fan again.
Indexing key
Electrical pins
5. Gently press the fan into place to make sure it is completely seated.
6. Replace the top cap.
7. Turn the instrument on and test it.
64
ToxiRAE Pro PID User’s Guide
17.7. Cleaning The Exterior
Occasional external cleaning of the instrument with a soft cloth is recommended. Do not
use detergents or chemicals. If necessary, you can use a damp cloth (water only). It is a
good idea to install the Calibration Adapter before cleaning the ToxiRAE Pro PID’s
housing, to keep dirt, dust, or moisture away from the sensor openings and to keep the
filter clean.
Visually inspect the contacts at the base of the ToxiRAE Pro PID and on the charging
cradle to make sure they are clean. If they are not, wipe them with a soft, dry cloth. Never
use solvents or cleaners.
17.7.1. Ordering Replacement Parts
If you need replacement parts, contact your local RAE Systems distributor. A list of
authorized distributors is available online:
http://www.raesystems.com
65
ToxiRAE Pro PID User’s Guide
18. Troubleshooting
Problem
Cannot turn on power
Lost password
Reading abnormally
Low
Possible Reasons & Solutions
Reasons:
Depleted battery.
Defective battery.
Solutions:
Solutions:
Reasons:
Solutions:
Buzzer, LEDs, or
vibration alarm
Inoperative
Reasons:
Solutions:
66
Replace battery.
Use ProRAE Studio II
software to change the
password.
Call Technical Support
at +1 408-752-0723 or
toll-free at
+1 888-723-4800
Calibration Adapter is
attached.
Incorrect calibration.
Remove Calibration
Adapter.
Calibrate the ToxiRAE
Pro PID.
Alarms may be turned
off.
Blocked alarm port.
Bad buzzer, LEDs, or
vibration alarm.
Enable alarms in the
Buzzer & Light section
of the Alarm Settings
sub-menu in the
Programming menu.
Unblock alarm port.
Call authorized service
center.
ToxiRAE Pro PID User’s Guide
19. Technical Support
To contact RAE Systems Technical Support:
Monday through Friday, 7:00AM to 5:00PM Pacific (US) Time
Phone (toll-free): +1 877-723-2878
Phone: +1 408-952-8200
Fax: +1 408-952-8480
Email: RAE-tech@honeywell.com
Outside the Americas:
E-Mail: HAexpert@honeywell.com
Honeywell Analytics Ltd.
4 Stinsford Road
Nuffield Industrial Estate
Poole, Dorset, BH17 0RZ
United Kingdom
Tel: +44 (0) 1202 645 544
Fax: +44 (0) 1202 645 555
Honeywell Analytics
ZAC Athélia 4 – 375 avenue du Mistral
Bât B, Expace Mistral
13600 La Ciotat
France
Tel: +33 (0) 4 42 98 17 75
Fax: +33 (0) 4 42 71 97 05
Honeywell Analytics
Elsenheimerstrasse 43
80687 München
Germany
Tel: +49 89 791 92 20
Fax: +49 89 791 92 43
Honeywell Analytics
P.O. Box-45595
6th Street
Musaffah Industrial Area
Abu Dhabi
UAE
Tel: +971 2 554 6672
Fax: +971 2 554 6672
20. RAE Systems Contacts
RAE Systems by Honeywell World Headquarters
3775 N. First St.
San Jose, CA 95134-1708 USA
Phone: +1 888-723-4800
E-mail: RAE-tech@honeywell.com
Web Site: www.raesystems.com
Training
Phone: +1 408-952-8260
Email: training@raesystems.com
67
ToxiRAE Pro PID User’s Guide
21. Appendix A: Regulatory Information
SAFETY INSTRUCTIONS
Read Before Operating
This manual must be carefully read by all individuals who have or will have the
responsibility of using, maintaining, or servicing this product. The product will perform
as designed only if it is used, maintained, and serviced in accordance with the
manufacturer’s instructions. The user should understand how to set the correct parameters
and interpret the obtained results.
CAUTION!
Only use RAE Systems battery pack P/N: G02-3004-000
Do only charge the battery pack using RAE Systems charger outside hazardous area.
Use of non-RAE Systems components will void the warranty and can compromise the
safe performance of this product.
Warning: Substitution of components may impair intrinsic safety.
ToxiRAE Pro Marking
The ToxiRAE Pro is certified according to the IECEx scheme, ATEX and CSA for US
and Canada as protected by intrinsic safety.
The ToxiRAE Pro is marked with the following information:
RAE SYSTEMS
3775 N. 1st . St., San Jose
CA 95134, USA
ToxiRAE Pro
Type PGM-1800.
Serial No/barcode: XXXX-XXXX-XX
IECEx SIR
10.0096X
Ex ia IIC T4 Ga
0575 II 1G
SIRA 10 ATEX 2189X
Ex ia IIC T4 Ga
2010 2360054
I.S. for Cl.I Dv 1, Grps A,B,C,D TCode T4.
Class I, Zone 0 A/Exia IIC T4
Only as to intrinsic safety for use in
hazardous loc. Intrinsically safe/
Securite Intrinseque/Exia
-20º C < Tamb < +55º C;
Um: 20V
Battery pack: G02-3004-000
Warning: Read User’s Manual for Intrinsic Safety Precautions
Warning: Substitution of components may impact intrinsically safety
68
ToxiRAE Pro PID User’s Guide
Operation Area and Conditions
Hazardous Areas classified by Zones
ToxiRAE Pro PID (PGM1800) are intended to be used in hazardous areas classified
zone 0, zone 1 or zone 2, where gases of explosion groups IIA, IIB or IIC and T4 may be
present.
For North America and Canada the equipment can further be used in Class I, Zone 0.
Hazardous Areas classified by Divisions
ToxiRAE Pro is intended to be used in hazardous areas classified for Class I Div. 1 or 2,
within the temperature range of -20º C to +55º C, where gases of explosion groups A, B,
C or D and temperature clode T4 may be present.
Instruction For Safe Use
Strictly follow the instructions for safe use. Application of the ToxiRAE Pro requires full
understanding and strict observation of the instructions.
Only use RAE Systems battery pack P/N: G02-3004-000
Do only charge the battery pack using RAE Systems charger outside hazardous area
(Um: 20V).
Use In Hazardous Areas
Equipment which is intended for use in explosive atmospheres and which has been
assessed and certified according to international regulations may be used only under
specified conditions. The components may not be modified in any way.
The appropriate regulations for service and repair must be properly observed during such
activities.
69
ToxiRAE Pro PID User’s Guide
Year of manufacture
To identify the year of manufacture, refer to the serial number of the instrument.
The second to last digit in the serial number indicates the year of manufacture. For
example, “M” indicates the manufacturing year is 2010.
First digit
Year
2008
2009
2010
2011
2012
2013
2014
2015
2016
2017
2018
2019
70
RAE Systems by Honeywell
World Headquarters
3775 N. First St.
San Jose, CA 95134-1708 USA
Phone: 408.952.8200
Fax: 408.952.8480
E-mail: customerserv@raesystems.com
Web Site: www.raesystems.coms
P/N G02-4008-000 Rev. F
September 2016

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : Yes
Author                          : tmulhern
Create Date                     : 2016:09:30 11:49:27-07:00
Modify Date                     : 2016:09:30 11:50:18-07:00
Subject                         : 
Language                        : EN-US
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 5.6-c015 84.158975, 2016/02/13-02:40:29
Metadata Date                   : 2016:09:30 11:50:18-07:00
Creator Tool                    : Acrobat PDFMaker 15 for Word
Document ID                     : uuid:fce0a047-7a15-49e1-8d58-f3f23269ae89
Instance ID                     : uuid:072a223b-9e09-4947-8b29-b7ac919bd86f
Format                          : application/pdf
Title                           : ToxiRAE Pro PID User's Guide
Description                     : 
Creator                         : tmulhern
Producer                        : Adobe PDF Library 15.0
Keywords                        : 
Source Modified                 : D:20160916192538
Company                         : .
Page Layout                     : OneColumn
Page Count                      : 73
EXIF Metadata provided by EXIF.tools
FCC ID Filing: SU3TOXIRAEPRO

Navigation menu