Radiometer analytical TTL075 Sample Changer User Manual D21T085 UG SAC9xx gb

Radiometer analytical S.A.S. Sample Changer D21T085 UG SAC9xx gb

Manual

Download: Radiometer analytical TTL075 Sample Changer User Manual D21T085 UG SAC9xx gb
Mirror Download [FCC.gov]Radiometer analytical TTL075 Sample Changer User Manual D21T085 UG SAC9xx gb
Document ID881679
Application IDv4aipYlJT1E42Y2gbnSmaA==
Document DescriptionManual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize251.06kB (3138250 bits)
Date Submitted2007-12-19 00:00:00
Date Available2007-12-19 00:00:00
Creation Date2007-12-14 10:38:29
Producing SoftwareGPL Ghostscript 8.50
Document Lastmod2007-12-14 10:38:29
Document TitleD21T085_UG_SAC9xx_gb.pdf
Document CreatorPDFCreator Version 0.9.0
Document Author: lchapus

SAC850 & SAC950
Sample Changers
User’s Guide
D21T085
D21T085 • Printed by Radiometer Analytical SAS • 2007-09A
Contents
General information ............................................................... 5
1. Introduction ........................................................................ 7
2. Presentation ........................................................................ 9
SAC850 ............................................................................................. 9
SAC950 ............................................................................................. 11
RFID technology on turntables ........................................................... 14
3. Setting up ............................................................................ 17
Mounting the turntable ........................................................................
Placing beakers .................................................................................
Mounting the beaker presence detector ..............................................
Assembling the electrode head ..........................................................
Mounting the propeller stirrer ..............................................................
Mounting the electrodes and addition tips ..........................................
Connections .......................................................................................
Assembling option accessories ..........................................................
Dynamic rinse module ........................................................
Reagent addition module .....................................................
18
19
21
21
22
23
24
27
27
34
4. Use ....................................................................................... 39
Switching on ...................................................................................... 39
Running sequence with a workstation ................................................. 40
5. Troubleshooting ................................................................. 45
Autotest and manual mode errors ...................................................... 45
Errors displayed by the workstation ................................................... 46
Contents
6. Maintenance ........................................................................ 49
Cleaning .............................................................................................
Transporting the instrument ................................................................
Servicing ............................................................................................
Replacing the fuses ...........................................................................
49
49
49
50
7. Accessories ........................................................................ 51
8. Technical specifications ..................................................... 61
Contents
General infor
mation
information
Safety Information
Please read this entire manual before unpacking, setting up, or operating
this equipment. Pay attention to all danger and caution statements.
Failure to do so could result in serious injury to the operator or damage to
the equipment. To ensure that the protection provided by this equipment
is not impaired, do not use and do not install this equipment in any
manner other than that specified in this manual.
Precautionary Labels
Read all labels and tags attached to the instrument. Personal injury or
damage to the instrument could occur if not observed.
This symbol, if noted on the instrument, references the
instruction manual for operation and/or safety information.
Electrical and electronical equipment marked with this
symbol may not be disposed of in European public
disposal systems after 13 August of 2005. In conformity
with European local and national regulations (EU Directive
2002/96/EC), European electrical equipment users must
now return old or end-of life equipment to the Producer for
disposal at no charge to the user.
Note: For equipment supplied or produced by "Radiometer
Analytical", please contact www.hach-lange.com and
select your country for instructions on how to return your
equipment for proper disposal."
This symbol, when noted on the product, identifies the
location of a fuse or current limiting device.
General information
Warning !
The sample changer has been developed to meet the requirements of
volumetric titration applications. It is therefore aimed at experienced
users who have the knowledge required to operate the instrument and
implement the security instructions enclosed. Please remember that the
sample changer must not, under any circumstances, be used to perform
tests on living beings.
We accept no responsibility for using the sample changer and its
peripheral devices under conditions that are not specified in this
User's Guide.
Compliance with FCC rules, part 15
Information to the user
NOTE: This equipment has been tested and found to comply with the
limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instruction, may cause harmful
interference to radio communications. However, there is no guarantee
that interference will not occur in a particular installation. If this equipment
does cause harmful interference to radio or television reception which can
be determined by turning the equipment off and on, the user is
encouraged to try to correct interference by one or more of the following
measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on circuit different from that to
which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
Use a shielded RS232 serial cord and a shielded USB cord fitted
with a ferrite.
Warning!!
Changes and modifications that are not expressly approved by the party
responsible for compliance could void the user's authority to operate the
equipment.
General information
1. Intr
oduction
Introduction
The SAC850 and SAC950 Sample Changers are designed for automatic
analysis in various types of trays.
Each tray can hold up to 70 samples depending on the sample volume
and the application to be performed.
The SAC850 Basic model is dedicated for small sample series
automation. The SAC850 is fitted with a single radius tray and can hold
14 to 25 samples.
The SAC950 Extended model is dedicated for large sample series
automation. The SAC950 is fitted with a 1, 2 or 3-radii tray and can hold
14 to 70 samples.
The SAC850 and SAC950 Sample Changers are ideal for handling
samples in TitraLab® or MeterLab® compatible systems.
The aim of this User's Guide is to provide you with detailed information
about setting up and operating your sample changer.
For information on how to enter data into your workstation, please refer to
the instructions delivered with the workstation.
Introduction
SAC850
5a
5b
11
10
Introduction
2. Pr
esentation
Presentation
SAC850
4 locations for spare electrodes and antidiffusion delivery tips
(2 locations on each side).
Sample/Rinse/Park beakers (1 Park, 0 to 2 Rinses, remaining
positions dedicated for samples).
Mono radius turntable. Capacities depend on the beaker size and
diameter : see chapter "Accessories" - Table "Turntables and
beakers (SAC850 and SAC950)".
Keypad.
Electrodes/Stirrer/Beaker presence detector holder.
Beaker presence detector location (5a).
Electrodes/stirrer head holder location (5b).
Shower rinse holder represented here with the shower head
assembly.
Optional peristaltic pump (optional).
Turntable holder.
Optional shower rinse pump/emptying pump module. Comprises 2
diaphragm pumps (rinsing and emptying).
10
Magnetic stirrer assembly comprising also a RFID reader device
for turntable automatic recognition.
11
Status lamp:
Blue : Free for operation. Manual operations possible.
Purple: Sample changer controlled by the workstation (analysis in
progress). Only the keyboard Stop key is available.
Red:
An error has been detected.
Presentation
6a
SAC950
6b
5c
5b
5a
12
11
10
10
Presentation
SAC950
4 locations for spare electrodes and antidiffusion delivery tips
(2 locations on each side).
Sample beakers.
1, 2 or 3 radii turntable. Capacities depend on the beaker size and
diameter : see chapter "Accessories" - Tables "Turntables and
beakers".
Keypad.
Beaker extension with Park (5a), Rinse 1 (5b) and Rinse 2 (5c)
beaker locations.
Electrodes/Stirrer/Beaker presence detector holder.
Beaker presence detector location (6a).
Electrodes/stirrer head holder location (6b).
Optional peristaltic pump.
Turntable holder.
Optional shower rinse pump/emptying pump module. Comprises 2
diaphragm pumps (rinsing and emptying).
10
Magnetic stirrer assembly comprising also a RFID reader device for
turntable automatic recognition.
11
Pilot lamp indicating the operating status of the instrument:
Blue :
Free for operation.
Manual operations on keyboard are possible
Purple:
The sample changer is controlled by the workstation
unit (analysis in progress). Only the keyboard Stop key
is available.
Red:
An error has been detected.
12
Shower rinse holder represented here with the shower head
assembly.
Presentation
11
Keypad (SAC950 only)
Starts adding reagent with peristaltic pump (*). Stops addition
when the key is latched down.
Starts shower rinsing using a membrane pump of the Dynamic
rinsing module (**). Stops action when the key is latched down.
Starts emptying a beaker using the second membrane pump of
the Dynamic rinsing module (**). Stops action when the key is
latched down.
(*)
(**)
12
If a Reagent addition module is installed (see chapter "Setting up")
and if the electrode holder is at its downmost position.
If a Dynamic rinsing module is installed (see chapter "Setting up")
and if the electrode holder is at its downmost position.
Presentation
4
Moves to the previous beaker. If the electrodes dip into the
beaker (or are above the beaker) before pressing the key, at the
end of the operation, they will dip into the previous beaker (or will
be above the previous beaker). The beakers are ranked as
follows:
Rinse beaker 2 (R2)
Rinse beaker 1 (R1)
Park beaker
Beaker no.1 of the turntable
...........................
Beaker no.n-1 of the turntable
Beaker no.n (last beaker) of the turntable
Rinse beaker 2 (R2)
...........................
Moves the arm (electrode head holder) downwards.
Moves the arm (electrode head holder) upwards or moves the
electrodes above the Park beaker if the electrodes are already
on service position (upmost position).
Moves to the next beaker. Same observations as for the "Moves
to the previous beaker" key (4).
Blue pilot lamp. Lights on when a key is pressed and lights off
when the key is latched down. This lamp blinks when the key is
not available (example: during an analysis, when a key other
than Stop is pressed).
Stop key. Stops the action in progress while a sequence is
running. Removes an error status (red status pilot lamp).
The Stop key is the only active key while a sequence is running.
Presentation
13
RFID technology on turntables
What’s RFID means and how does it work?
Radio Frequency IDentification (RFID) is a generic term that is used to
describe a system that transmits the identity (in the form of a unique
serial number) of an object or person wirelessly, using radio waves. It's
grouped under the broad category of automatic identification
technologies.
It’s the combination of a “tag” and a “reader”. A typical RFID tag consists
of a microchip attached to a radio antenna mounted on a substrate. The
chip can store data. The reader is used to retrieve the data stored on an
RFID tag. A typical reader is a device that has one or more antennas that
emit radio waves and receive signals back from the tag. The reader then
passes the information in digital form to a computer system.
In our assembly an RFID tag is attached inside the turntable at the first
beaker level position. This tag is clearly visible.
RFID tag placed on the turntable
14
Presentation
RFID reader mounted inside the Magnetic stirrer assembly
SAC850 on the left and SAC950 on the right
The RFID reader is mounted inside the Magnetic stirrer assembly (see
figure above). When an analysis is started and a turntable is in place, the
sample changer rotates the tray (and the arm in the case of a SAC950)
so that the tag and the reader are face to face for a moment.
Which data contains a turntable RFID tag?
The turntable identification data (type of turntable, number of beakers,
beaker location and beaker size/height).
Benefits of using RFID technology on turntables
The Sample Changer knows at any moment if a turntable is installed and
which turntable is used.
Possibility of working with magnetic bars without risks of interferences.
Presentation
15
16
Presentation
3. Setting up
Unless otherwise indicated, the following setting up instructions are
common to both SAC850 and SAC950 Sample Changers.
Set up the instrument on a flat surface and in a properly ventilated place.
The power supply connectors on the rear panel must remain accessible
so that you can quickly disconnect the cables in case of emergency. The
room temperature should not exceed 40°C.
The relative humidity must be between 20 and 80%.
Minimum space requirements: 80 cm x 60 cm (32" x 24")
60 cm
80 cm
Setting up
17
Mounting the turntable
Several turntables are available for
each model of sample changer:
see chapter "Accessories",
table "Turntables and beakers".
The turntable is identified on
starting an analysis. The RFID
reader reads the turntable RFID tag
18
Setting up
Placing beakers
Park beaker
At the end of a sample series, the electrodes are dipped in the park beaker
or left above the park beaker. The park beaker can also be used to rinse
dynamically the electrodes.
Positioning of the park beaker (see figures below):
SAC850:
Last position of the turntable
SAC950:
On the electrode reconditioning extension
Static rinse beakers
Between 2 sample runs, the electrodes are dipped once or twice into a rinse
solution under stirring. 1 or 2 static rinse beakers can be used.
Positioning of the static rinse beaker(s)
SAC850: Last but one (R1) and previous (R2) positions of the turntable
SAC950: On the electrode reconditioning extension, positions (R1) and (R2)
R1: Static rinse beaker no. 1
R2: Static rinse beaker no. 2
Other positions: Sample
beakers (15 free position if 2
static rinses programmed)
Park
R2
R1
15
14
13
Rotation
direction
12
10
11
Position of the beakers on a SAC850
Example of a 18-position turntable
Setting up
19
R2
R1
Park
1 30
29
3 2
28
32 31 53
52 27
54 70
51 26
6 34 56 55
69
50 25
7 35 57
68
49 24
8 36 58
67
48 23
9 37 59
66
47 22
10 38 60
65
11 39 61
64 46 21
62
63
45 20
12 40 41
44
42 43
13
19
14 15
18
16 17
Rotation
direction
33
R1: Static rinse beaker no. 1
R2: Static rinse beaker no. 2
Positions 1 to 70:
Sample beakers
Position of the beakers on a SAC950
Example of a 70-position turntable
SAC850 users:
Park, static rinse and sample beaker sizes depend on the type of
turntable used. These beakers are supplied with the turntable kit (see
chapter "Accessories").
SAC950 users:
Park and static rinse beakers are 40-100 ml PP titration vessel, supplied
with the SAC950, part no.: X91T055 (10 pcs).
Sample beaker sizes depend on the type of turntable used. These
beakers are supplied with the turntable kit (see chapter "Accessories").
20
Setting up
Mounting the beaker presence detector
The beaker presence detector is an ultrasonic transducer that detects
empty positions on the turntable.
Assemble the 2
parts until awaiting
a sound "Click!".
Click!
Assembling the electrode head
Assemble the 2
parts until awaiting
a sound "Click!".
X31T151
B = 12 mm
Click!
X31T151
X31T102
Setting up
21
Mounting the propeller stirrer
Screw the stirrer in
its housing (screw
completely without
tightening).
∅ = 18 mm
∅ = 10 mm
The propeller stirrer is supplied with 2
propellers 10 & 18 mm.
Fit the right propeller according to the
size of the beaker
(see chapter "Accessories",
table "Turntables and beakers").
22
Setting up
Mounting the electrodes and addition tips
842-152
842-302
842-306
842-303
X51T011
ø = 12; L = 103mm
ø = 9.5; L = 103 mm
ø = 12; L = 120 mm
X51T012
ø = 12; L = 103 mm
A25B800
ø = 7.5 mm; L = 103 mm
956-309
X51T010
X31T017
X31T018
X31T018
X31T013
X31T018
X31T015
X31T082
X31T016
X31T014
X31T031
B B
X31T151
A set of adapters for bayonet electrode heads, part
no. X91T002 is supplied with the sample changer
turntable.
Setting up
23
4 locations for spare electrodes
and antidiffusion delivery tips (2
locations on both side)
Connections
Press the electrode
cable through the slot
of the stirrer head.
Repeat the same
operation for the
beaker detector cable.
24
Setting up
Engage the stirrer cable in the chain from the
3rd chain link. Repeat the same operation for
the beaker detector and electrode cables.
A1
A9
A9
A1
Connect the electrode the
corresponding socket of
the workstation unit.
Insert the stirrer and beaker detector cables in
the opening of the arm then slip these cables
under the sample changer in order to present the
plugs at the connection panel of the instrument.
Slightly level up the instrument to facilitate this
operation.
Setting up
25
To a mains socket.
Mains cable, Part no.
A95S001 (230 V)
A95S002 (115 V)
To workstation SAC socket
Serial Cable, SubD9/SubD9
Part no. A95X501
26
Setting up
Assembling option accessories
Dynamic rinse module
The Dynamic rinse module is used for shower rinsing and emptying
operations. It consists of a pump assembly (two membrane pumps, one
for rinsing and the other for emptying), a shower head, a suction tip and
silicon tubings.
Switch on the sample changer
and move the arm at its upmost
position:
- SAC850: p ush up under part
(b) (see arrow opposite).
- SAC950: as for the SAC850 or
press key
(see chapter "Presentation Keypad", key (7)).
Switch off the sample changer.
Remove the electrode head
holder (a) (2 screws).
Setting up
27
a
b2
b1
28
Setting up
Separate part (b) in 2
pieces (b1) and (b2) (2
screws).
Assemble the shower
head (c) and shower
head skirt (d) or (e) (or
clip (f)) on part (b1). The
use of a (d) or (e) shower
head skirt depends on
the diameter of the
beaker used (see chapter
"Accessories", table
"Turntables and
beakers"). Use a clip (f)
if you do not use a
shower head skirt.
Fit the electrode head (a)
onto part (b2) (2 screws).
If not already done,
remove the turntable.
Locate the housing of the
pump assembly.
Remove the protection sticker.
Unscrew the 3 screws then
remove the cover plate.
Connect the 2 pump cable plugs to
the corresponding sockets.
Setting up
29
Insert the pump
assembly and fix it
in place with the 3
screws.
48 mm
X31T155
30
Bevelled edge
Setting up
Setting up of the
suction tip,
part no. X31T155.
2m
X31T154
Cut the silicon tubing, d=4/6 mm, L= 8 m, part no. X31T154,
in 4 pieces of 2 meters each.
Mount one end of tubing 1 on the metal
rod of the shower head. Form a loop of 5 cm
diameter at the level of the 2nd chain link.
Engage the tubing in the chain by simple
pressure on each link.
Important
It is important to check at this stage that the tubing slides well and when
moving up part A , this particularly after having set up all tubings and
wires. To move up part A , push up under this part or press key
SAC950 keypad.
Setting up
of
31
2
Mount one end of tubing 2 on the head of the
suction tip X31T155. Engage the tubing in the
chain from the 3rd chain link.
Insert tubings 1 and 2 in the opening of the arm then slip these
tubings under the sample changer in order to present them at the rinsing/
draining pumps. The tubings must pass on the side of the black cable
guide part. Slightly level up the instrument to facilitate this operation.
32
Setting up
RINSE WASTE
Rinsing
solvent
Waste
Mount tubing 1 on the outlet of the rinsing pump and tubing 2 on the
inlet draining pump.
Mount one end of tubing 3 on the inlet of rinsing pump and dip the other
end in a can filled with rinsing solvent (demineralised water for example).
Mount one end of tubing 4 on the oulet of draining pump and dip the
other end into a waste.
Cover the pump assembly with the protection lid provided.
Setting up
33
Reagent addition module
A reagent addition module can be mounted on the SAC850 or SAC950
sample changer. This peristaltic pump is used to add reagent as be part of
a sample preparation. Volumes are added with an accuracy around 1%.
A reagent addition module consists of a peristaltic pump assembly, a
reagent addition tip, silicon and Maprene  tubings.
If not already done, remove the
turntable.
Locate the
housing of
peristaltic
pump.
Remove the
protection sticker,
unscrew the 3
screws then
remove the cover
plate.
34
Setting up
Setup the
Maprene  tubing,
part no. X31T156,
in the peristaltic
pump.
Connect the 2 pump cable plugs
to the corresponding sockets.
Insert the
peristaltic pump
assembly and fix it
in place with the 3
screws.
Setting up
35
X31T157
X31T018
Set up the addition tip,
part no. X31T157.
X31T154
2m
Cut a piece from 2 m long in the silicon tubing, d=4/6 mm,
L= 8 m, part no. X31T154.
Mount one end of tubing 1 on the head of the
addition tip X31T157. Engage the tubing in the
chain from the 3rd chain link (by simple
pressure on each link).
36
Setting up
Insert tubing 1 in the opening of the arm
then slip this tubing under the sample
changer in order to present it at the
peristaltic pump. The tubing must pass on
the side of the black cable guide part.
Slightly level up the instrument to facilitate
this operation.
X31T154
2-6m
Cut a piece from 2 to 6 m long in the remaining part of silicon
tubing, part no. X31T154.
Setting up
37
1
Reagent
Mount tubing 1 on the outlet of the peristaltic pump.
Mount one end of tubing 2 on the inlet of peristaltic pump and dip the
other end in the reagent flask.
Cover the pump with the protection lid provided.
38
Setting up
4. Use
Switching on
1.
Connect the mains socket of the SAC 850/SAC950 sample changer
and the workstation (TIMxxx or IONxxx) to the mains supply using
the 3-lead power cord provided.
In the USA or Canada, use a UL listed power cord only.
Important:
The sample changer and the workstation must be connected to an
earthed mains socket for safety reasons. Efficient grounding is
essential for reliable measurements and security.
The power supply connector on the rear panel must remain
accessible and close to the user (2 m maximum) so that you can
quickly disconnect the cables in case of emergency.
2.
Switch on the sample changer (O/I power switch on the rear panel
set to "I", see below) then the workstation.
Mains supply socket
The sample changer runs its autotest routine:
Lifts/lowers the arm to search for the arm upmost position,
Rotates the tray and arm to search for reference positions of
both arm and tray.
Dips the electrodes into the Park beaker. If no Park beaker
detected, leaves the electrodes above the Park beaker position.
Use
39
Running sequence with a workstation
After having started a sequence from the workstation or TitraMaster 85,
the sample changer performs the following operations if 1 to 9 dynamic
rinses and if 2 static rinses R1 and R2 have been programmed (*):
1.
Moves to Park beaker (**) and starts rinsing dynamically the
electrodes. At the end, the Park beaker is left empty.
See paragraph "Dynamic rinsing".
2.
Moves to R1 beaker (***), dips the electrodes into that beaker and
rinses under stirring during a preset delay. See paragraph "Stirring".
3.
Moves to R2 beaker (***), dips the electrodes into that beaker and
rinses under stirring during a preset delay. See paragraph "Stirring".
4.
Moves to beaker no.1, prepares sample/standard running the
sample preparation number attached to the method. During this
time, the workstation displays the message "Sample changer
processing in beaker 1" message.
See paragraph "Sample preparation".
5.
At the end of analysis, starts rinsing dynamically the electrodes. At
the end, the beaker no.1 is left nearly empty with a small amount of
solvent that was used to flush the electrodes and delivery tips.
(*)
The number of static rinses (0, 1 (R1) or 2 (R1 and R2)), the duration of
static rinses and the number of dynamic rinses (0 to 9) are programmed
in the Setup > Configuration parameters of the workstation (refer to the
Reference Manual of the workstation used).
(**)
A dynamic rinse before first beaker analysis can occur in the Park or in
the R1 beaker if 0 or 1 static rinse has been programmed. Selection
between Park and R1 is made in the Setup > Configuration parameters
of the workstation (refer to the Reference Manual of the workstation
used).
(***) R1: Static rinse beaker no.1. R2: Static rinse beaker no.2. The position of
these beakers is given in chapter "Setting up - Placing beakers".
40
Use
6.
Repeats steps 2 and 3.
8.
Dips the electrodes in beaker no. 2, prepares sample/standard
running the sample preparation number attached to the method.
7.
At end of the sequence (end of analysis in the last beaker of the
sequence):
- starts rinsing dynamically the electrodes,
- repeats steps 2 and 3,
- moves the electrodes above the Park beaker,
- left the electrodes as is or dip the electrodes into the Park beaker
depending on the selection made in the Setup > Configuration
parameters of the workstation (refer to the Reference Manual of the
workstation used).
At the position indicated, the vertical axis moving is likely to
pinch the skin of the user!
Use
41
Dynamic rinsing
1 to 9 Dynamic rinses (or shower rinse/emptying operations) can be
performed between 2 sample runs. A Dynamic rinse is carried out in the
just analysed sample beaker or in the Park beaker (*). At the start of a
sequence, the dynamic rinse can be performed either in the Park beaker or
in R2 (static rinse no.2) beaker (*). Dynamic rinses never take place in
calibration beakers. At the end of a calibration beaker analysis, the
dynamic rinse is performed in the Park or R2 (static rinse no.2) beaker (*).
A Dynamic rinse consists of the following operations:
1.
The electrode are positioned above the rinse beaker.
2.
The electrodes move down in the beaker. The beaker is emptied to a
waste in the same time. At the end, the electrodes are located in the
emptied beaker at their downmost position.
3.
The electrodes move up slowly. In the same time, the electrodes are
washed with rinse solution (usually demineralised water) and the
resulting solution is emptied very slowly to a waste. At the end, the
electrodes are located above a beaker filled with rinse solution and
some remaining impurities.
4.
The electrodes move down in the beaker. The beaker is emptied to a
waste in the same time. At the end, the electrodes are located in the
emptied beaker at their downmost position. Then, a last rinsing of the
electrodes and delivery tips is carried out.
5.
The electrodes move up. At the end, the electrodes are located above
a nearly emptied beaker containing a little solvent that was used to
flush the electrodes and delivery tips.
Steps 3 and 4 can be repeated up to 8 times as up to 9 dynamic rinses can
be programmed. Steps 2 to 4 are performed under stirring.
(*)
42
Depending on the selection made in the Setup > Configuration
parameters of the workstation (refer to the Reference Manual of the
workstation used).
Use
Stirring
A magnetic and a propeller stirrer can be used to stir the solutions. When
the propeller stirrer is connected (see chapter "Setting up"), it will be
automatically used in place of the built-in magnetic stirrer. The stirrer
speed is set in the Cell menu of the workstation (value adjustable
between 100 and 1100 rpm by steps of 50 rpm).
Dynamic and static rinses are performed under stirring.
Magnetic stirring with a SAC950
When using a 2 or 3-radii turntable with a SAC950, you must use a
propeller stirrer to stir the solutions that are placed on the smallest row (2radii turntable) or on the 2 smallest rows (3-radii turntable).
Green positions : magnetic and propeller stirring possible
Red positions : propeller stirring only
Use
43
Sample preparation
A SAC850 or SAC950 can run up to 10 different sample preparations in a
sequence. Sample preparations are created and edited from
TitraMaster 85 PC PC software.
Each element of a sample changer sequence is associated to a sample
preparation. A sample preparation is a succession of up to 10 elementary
actions required by the sample changer to prepare each sample beakers
to be run thereafter.
On running a sequence, the workstation sends to the sample changer all
sample preparations to be runned for the sequence. The compatibility
between the sample preparation and the sample changer used is tested
and an error message is displayed in case of failure (example: use of a
peristaltic pump that is not installed yet).
See TitraMaster 85 on line help, topics "Handling sample preparations".
Important:
When starting a sample preparation, if you have a solid or powder sample
in the beaker or if you have, for some beakers, less liquid than a
minimum height limit (see chapter "Accessories - Beakers"), then you
must disable the "beaker presence detection" option before running the
sequence. If you keep that option activated, then the sample changer will
not prepare the sample in that beaker and will jump to the next beaker
considering that the current position is empty.
To disable the "beaker presence detection" option in TitraMaster 85 , see
TitraMaster 85 on line help, topics "Editing an application - Configuration".
Actions that can be done during a sample preparation:
Control peristaltic pump.
Start/stop electrode rinsing.
Start/stop beaker emptying.
Control stirring.
Set delay.
Control stirrer.
TTL command (wait for a TTL IN for a preset time).
44
Use
5. T
Trroubleshooting
A number of tests are automatically performed when the sample changer
is switched on and during calibration and sample batch measurements.
The errors reported can be divided in two groups:
•
Autotest errors occuring on switching on the sample changer and
errors occuring while operating the sample changer alone (manual
mode without workstation).
•
Errors displayed by the workstation.
Autotest and manual mode errors
If an error is detected, the sample changer stops the current action and
the status lamp turns to red.
Switch off the sample changer, identify the causes of the error (generally,
arm or turntable blocked or arm or turntable movement obstructed),
correct the error then switch on the instrument.
Troubleshooting
45
Errors displayed by the workstation
Messages
Error/Action
ARM or TURNTABLE
OBSTRUCTED
The arm or turntable has been blocked and
cannot function properly. Remove the
obstruction and press Resume analysis
(key 1) to continue the sequence from the
point it stopped.
On starting a sequence,the sample changer
cannot read the RFID tag of the turntable.
Check that a turntable is correctly mounted
on the sample changer. Check the model of
turntable used. See chapter "Accessories".
Restart the sequence.
A rinse beaker is missing or has not been
positioned properly on the sample changer.
Either position a beaker or terminate the
operation and enter the correct number of
rinses.
If you want to continue with the next beaker
in the analysis, press 1 Resume analysis.
Either the data transmission between
sample changer and the workstation cannot
be performed properly, in which case you
should check the cable connections, or the
measurement stopped due to a movement
error. Terminate the cycle by pressing key
Stop on the SAC950 sample changer
keypad or key Stop of the workstation, and
check that nothing is obstructing sample
changer's movements.
MISSING TRAY
NO BEAKER
SAMPLE CHANGER
46
Troubleshooting
Messages
Error/Action
SAMPLE PREPARATION
IMPOSSIBLE
The sample changer cannot run the sample
preparation the workstation is asking for.
Example:
A reagent addition is programmed in the
sample preparation and no peristaltic
pump is installed on the sample changer.
RINSE PROCEDURE
IMPOSSIBLE
Review the Edit method parameters,
parameter Sample preparation no and/or
install missing pumps
For TitraMaster 85 users only, check and,
if necessary, edit the sample preparation
routine.
The sample changer cannot run the rinse
programmed.
Example:
You have programmed 2 static rinses and a
first dymamic rinse in the second static
rinse beaker (R2 beaker). Only 1 beaker is
available for static rinses (beaker R1) if you
intend to rinse dynamically the electrodes in
a static rinse beaker.
Review the Setup > Configuration
parameters.
Troubleshooting
47
48
Troubleshooting
6. Maintenance
Cleaning
The sample changer requires minimum maintenance. The exterior surface
shoudd be cleaned regularly with tepid water or a mild detergent and
wiped dry with a soft damp cloth.
Transpor
ting the instrument
ransporting
Always use the packaging supplied by the manufacturer.
IMPORTANT!
Remove the tray before transporting the instrument.
Never pick-up or carry the instrument by the electrode head holder or by
the arm.
Servicing
DO NOT ATTEMPT TO SERVICE THIS PRODUCT YOURSELF, except
as noted in this User's Guide. For servicing, please contact your
Radiometer Analytical service representative at the address given below:
RADIOMETER ANALYTICAL SAS
72, rue d'Alsace
69627 Villeurbanne CEDEX - France
Tel.: +33 (0) 4 78 03 38 38
Fax: +33 (0) 4 78 03 38 27
E-mail: radiometer@nalytical.com
or your local service representative:
_________________________________
_________________________________
_________________________________
_________________________________
_________________________________
Maintenance
49
Replacing the fuses
For continued protection replace the fuse with one of a high interrupting
capacity, same type and rating:
2 x fuses, slow blow, T2.5A L250V (5 x 20 mm), part no. X16T017.
To replace the fuses:
Locate the fuse housing on the rear panel of the sample changer.
1. Disconnect line cord
2. Remove the fuse holder
(use tweezers)
hook
3. Replace the used fuses with
ones of the same type and rating
4.
50
Maintenance
Put the cap back in place
7. Accessories
Dynamic rinse module for SACxxx
Dynamic rinse/waste module for SACxxx, part no. X91T095
comprising:
1 X31T153
Shower head for SACxxx
1 X16T019
Pump membrane assembly for SACxxx
1 X31T154
Silicon crystal tube, d=4/6 mm, L= 8 m
1 X31T155
Suction tip for SACxxx
1 X31T158
Shower head skirt for 22-45 & 8-25 ml PP beakers
1 X31T159
Shower head skirt for 100 ml glass beakers
Reagent Addition Module for SACxxx
Reagent Addition Module for SACxxx, part no. X91T096 comprising:
1 X16T020
Peristaltic pump assembly for SACxxx
1 X31T156
Maprene  tube, d=4/7,2 mm, L= 2 m
1 X31T154
Silicon crystal tube, d=4/6 mm, L= 8 m
1 X31T157
Addition Tip for SACxxx
Kit of cables for SACxxx
Kit of cables for SACxxx, part no. X91T097 comprising:
1 A95S001
Line Cord 230V Euro
1 A95S002
Line Cord 115V
1 A95X501
Serial Cable, SubD9/SubD9
2 X16T017
Fuse, slow blow, 2.5A (5x20mm)
1 X16T018
Beaker detection module - Ultrasonic transducer
2 A95A126
Cable, BNC/1m/BNC with Adapter BNC f/f
1 A95A127
Cable, Banana-m/1m/Banana-m with Adapter Banana f/f
Accessories
51
Turntables (SAC850 and SAC950)
Kit 25-position turntable, SAC850, 50 ml glass low form, 22-45 ml & 825 ml PP beakers, X91T080 comprising:
1 X31T136
25-position turntable, beakers d= 44 mm
1 X91T002
Set of bayonet adapters
2 842-302
Delivery Tubing Long, 325 cm, for burette stand >=5 ml
1 904-488
Titration Vessel PP, 8-25 ml (50 pcs)
1 904-489
Titration Vessel PP, 22-45 ml (50 pcs)
1 X31T151
SACxxx Electrode Head, standard
Kit 22-positions, turntable, SAC850, glass 100ml tall form beakers,
X91T082 comprising:
1 X31T138
22-position turntable, beakers d= 48 mm
1 X91T002
Set of bayonet adapters
2 842-302
Delivery Tubing Long, 325 cm, for burette stand >=5 ml
1 X31T151
SACxxx Electrode Head, standard
Kit 20-position turntable, SAC850, glass 150ml tall form beakers,
40-100 ml PP beakers, X91T084 comprising:
1 X31T140
20-position turntable, beakers d= 60 mm
1 X91T002
Set of bayonet adapters
2 842-302
Delivery Tubing Long, 325 cm, for burette stand >=5 ml
1 904-490
Titration Vessel PP, 40-100 ml (50 pcs)
1 X31T151
SACxxx Electrode Head, standard
Kit 20-position turntable, SAC850, GOSSELIN 125-180 ml PP beakers,
X91T086 comprising:
1 X31T142
20-position turntable, beakers d= 54 mm
1 X91T002
Set of bayonet adapters
2 842-302
Delivery Tubing Long, 325 cm, for burette stand >=5 ml
1 X31T087
Beakers with cover, PP, 125 ml, 50 pcs
1 X31V005
Beakers with cover, PP, 180 ml, 50 pcs
1 X31T151
SACxxx Electrode Head, standard
Kit 18-position turntable, SAC850, glass 250ml tall form beakers,
X91T088 comprising:
1 X31T144
18-position turntable, beakers d= 60 mm
1 X91T002
Set of bayonet adapters
2 842-302
Delivery Tubing Long, 325 cm, for burette stand >=5 ml
1 X31T151
SACxxx Electrode Head, standard
52
Accessories
Kit 15-position turntable, SAC850, glass 250ml low form beakers,
X91T090 comprising:
1 X31T146
15-position turntable, beakers d= 70 mm
1 X91T002
Set of bayonet adapters
2 842-302
Delivery Tubing Long, 325 cm, for burette stand >=5 ml
1 X31T151
SACxxx Electrode Head, standard
Kit 15-position turntable, SAC850, glass 400ml tall form beakers,
X91T092 comprising:
1 X31T148
15-position turntable, beakers d= 70 mm
1 X91T002
Set of bayonet adapters
2 842-302
Delivery Tubing Long, 325 cm, for burette stand >=5 ml
1 X31T151
SACxxx Electrode Head, standard
Accessories
53
Turntables (SAC950 only)
Kit 70-position turntable, 50 ml glass low form, 22-45 ml & 8-25 ml PP
beakers, X91T081 comprising:
1 X31T137
70-position turntable, beakers d= 44 mm
1 X91T002
Set of bayonet adapters
2 842-302
Delivery Tubing Long, 325 cm, for burette stand >=5 ml
2 904-488
Titration Vessel PP, 8-25 ml (50 pcs)
2 904-489
Titration Vessel PP, 22-45 ml (50 pcs)
1 X31T151
SACxxx Electrode Head, standard
Kit 42-position turntable, 100 ml glass tall form beakers, X91T083
comprising:
1 X31T139
42-position turntable, beakers d= 48 mm
1 X91T002
Set of bayonet adapters
2 842-302
Delivery Tubing Long, 325 cm, for burette stand >=5 ml
1 X31T151
SACxxx Electrode Head, standard
Kit 36-position turntable, 150 ml glass tall form,40-100 ml PP beakers,
X91T085 comprising:
1 X31T141
36-position turntable, beakers d= 60 mm
1 X91T002
Set of bayonet adapters
2 842-302
Delivery Tubing Long, 325 cm, for burette stand >=5 ml
1 904-490
Titration Vessel PP, 40-100 ml (50 pcs)
1 X31T151
SACxxx Electrode Head, standard
Kit 36-position turntable, GOSSELIN 125/180 ml PP beakers, X91T087
comprising:
1 X31T143
36-position turntable, beakers d= 54 mm
1 X91T002
Set of bayonet adapters
2 842-302
Delivery Tubing Long, 325 cm, for burette stand >=5 ml
1 X31T087
Beakers with cover, PP, 125 ml, 50 pcs
1 X31V005
Beakers with cover, PP, 180 ml, 50 pcs
1 X31T151
SACxxx Electrode Head, standard
Kit 30-position turntable, 250 ml glass tall form beakers, X91T089
comprising:
1 X31T145
30-position turntable, beakers d= 60 mm
1 X91T002
Set of bayonet adapters
2 842-302
Delivery Tubing Long, 325 cm, for burette stand >=5 ml
1 X31T151
SACxxx Electrode Head, standard
54
Accessories
Kit 25-position turntable, 250 ml glass low form beakers, X91T091
comprising:
1 X31T147
25-position turntable, beakers d= 70 mm
1 X91T002
Set of bayonet adapters
2 842-302
Delivery Tubing Long, 325 cm, for burette stand >=5 ml
1 X31T151
SACxxx Electrode Head, standard
Kit 25-position turntable, 400 ml glass tall form beakers, X91T093
comprising:
1 X31T149
25-position turntable, beakers d= 70 mm
1 X91T002
Set of bayonet adapters
2 842-302
Delivery Tubing Long, 325 cm, for burette stand >=5 ml
1 X31T151
SACxxx Electrode Head, standard
Accessories
55
Turntables and beakers (SAC850 and SAC950)
Beaker
part no.
Volume
min./max.
400 ml tall form *
d = 70 mm
300 - 360 ml
Turntable
to be used
Dynamic rinsing Shower head skirt
Propeller size
Not used
18 mm
Not used
18 mm
Not used
18 mm
Not used
18 mm
Not used
18 mm
Not used
18 mm
Not used
18 mm
Required, X31T159
18 mm
X31T148
15 positions
250 ml low form *
d = 70 mm
130 - 225 ml
X31T146
15 positions
250 ml tall form *
d = 60 mm
180 - 225 ml
X31T144
18 positions
125 ml Gosselin,
d = 54 mm, PP,**
X31T087 (50pcs)
70 - 110 ml
180 ml Gosselin,
d = 54 mm, PP,**
X31V005 (50pcs)
100 - 160 ml
150 ml tall form *
d = 60 mm
95 - 135 ml
40 - 100 ml PP
904-490 (50 pcs)
d = 60 mm
55 - 90 ml
100 ml tall form *
d = 48 mm
50 - 90 ml
X31T142
20 positions
X31T140
20 positions
X31T138
22 positions
50 ml low form *
d = 44 mm
30 - 45 ml
Required, X31T158
18 mm
904-489 (50 pcs)
d = 44 mm
22 - 45 ml
Required, X31T158
18 mm
904-488 (50 pcs)
d = 44 mm
10 - 22 ml
Required, X31T158
10 mm
X31T136
25 positions
* Standard glass beakers specified in accordance with DIN12331 ISO3819.
** Metal covers, X31T129 (for 180 ml beakers) and X31T130 (for 125 ml beakers)
Beakers can be covered/uncovered using the Beaker cover module for SACxxx, X16T016.
56
Accessories
Turntables and beakers (SAC950 only)
Beaker
part no.
Volume
min./max.
400 ml tall form *
d = 70 mm
300 - 360 ml
Turntable
to be used
Dynamic rinsing Shower head skirt
Propeller size
Not used
18 mm
Not used
18 mm
Not used
18 mm
Not used
18 mm
Not used
18 mm
Not used
18 mm
Not used
18 mm
Required, X31T159
18 mm
X31T149
25 positions
250 ml low form *
d = 70 mm
130 - 225 ml
X31T147
25 positions
250 ml tall form *
d = 60 mm
180 - 225 ml
X31T145
30 positions
125 ml Gosselin,
d = 54 mm, PP,**
X31T087 (50pcs)
70 - 110 ml
180 ml Gosselin,
d = 54 mm, PP,**
X31V005 (50pcs)
100 - 160 ml
150 ml tall form *
d = 60 mm
95 - 135 ml
40 - 100 ml PP
904-490 (50 pcs)
d = 60 mm
55 - 90 ml
100 ml tall form *
d = 48 mm
50 - 90 ml
X31T143
36 positions
X31T141
36 positions
X31T139
42 positions
50 ml low form *
d = 44 mm
30 - 45 ml
Required, X31T158
18 mm
904-489 (50 pcs)
d = 44 mm
22 - 45 ml
Required, X31T158
18 mm
904-488 (50 pcs)
d = 44 mm
10 - 22 ml
Required, X31T158
10 mm
X31T137
70 positions
* Standard glass beakers specified in accordance with DIN12331 ISO3819.
** Metal covers, X31T129 (for 180 ml beakers) and X31T130 (for 125 ml beakers)
Beakers can be covered/uncovered using the Beaker cover module for SACxxx, X16T016.
Accessories
57
Beakers
Beaker type
Ø x H (mm)
Volume
min./max.
Detection min.
height **
Part no.
400 ml tall form *
70 x 130
300 - 360 ml
10 mm
(40 ml)
250 ml low form *
70 x 95
130 - 225 ml
0 mm
250 ml tall form *
60 x 120
180 - 225 ml
10 mm
(30 ml)
125 ml, Gosselin,
PP, 54 x 72
70 - 110 ml
0 mm
X31T087
(pack of 50)
180 ml, Gosselin,
PP, 54 x 102
100 - 160 ml
10 mm
(25 ml)
X31V005
(pack of 50)
150 ml tall form *
60 x 80
95 - 135 ml
0 mm
40-100 ml, PP
60 x 80
55 - 90 ml
0 mm
100 ml tall form *
48 x 80
50 - 90 ml
0 mm
50 ml low form *
42 x 60
30 - 45 ml
0 mm
22-45 ml, PP
44 x 70
22 - 45 ml
0 mm
904-489
(pack of 50)
8-25 ml, PP
44 x 70
10 - 22 ml
0 mm
904-488
(pack of 50)
904-490
(pack of 50)
* Standard glass beakers specified in accordance with DIN12331 ISO3819.
** When starting an analysis, if you have less liquid sample than this
specified value, then the beaker automatic detection can not work.
When working with a workstation, you must clear the option
"Beaker detection" in TitraMaster 85 (refer to the TitraMaster 85,
on-line help, topics "Editing an application - Configuration").
58
Accessories
Adapters for Bayonet Electrode Head
Adapters for Bayonet Electrode Head, X91T002 comprising:
5 X31T031
Bayonet Socket ø 12 mm with washer
2 X31T013
Bayonet Adapter for PAF-X3 and D4362 delivery tips
2 X31T014
Bayonet Adapter for 103 mm screw cap electrode
2 X31T015
Bayonet Adapter for 120 mm screw cap electrode
2 X31T016
Bayonet Adapter for ø 9.5 mm electrode, type GK2401
2 X31T018
Bayonet Adapter ø 12 mm to ø 7.5 mm
1 X31T127
Stirring Shaft, PP, L = 96 mm with propellers
10&18 mm
2 X31T032
Bayonet Stopper ø12 mm
2 X31T033
Bayonet Stopper ø7.5 mm
Accessories
59
60
Accessories
8. T
echnical specifications
Technical
Functions
Easy removable turntables, automatic identification of turntable model by
RFID:
- SAC850: 1 radius
- SAC950: 1, 2 or 3 radii
Dip rinse and/or reconditioning: 0 to 2 positions.
Service position (electrodes above the park beaker).
Automatic detection of empty position by ultrasonic transducer.
Stirring: magnetic or propeller stirring. Magnetic stirring if propeller stirrer
not connected. 0 to 1100 rpm by steps of 50 rpm.
Built-in keypad: lift electrode head up/down, move to next/previous
beaker, start/stop dynamic shower rinse, start/stop emptying to waste,
start/stop adding a reagent, stop action.
Embedded programmable sample preparation sequences.
Unused electrodes and delivery tips storage: 4 positions Ø16 mm, on the
top of the sample changer arm.
Optional modules
Dynamic shower rinse & Beaker emptying: Shower rinse and emptying
operations controlled by 2 PTFE membrane pumps, available in any
beaker size. Flow rate: up to 600 ml/min. Pumps can be controlled
manually and individually through keypad.
Addition of a reagent with a peristaltic pump to prepare sample. The
pump can be controlled manually through keypad. The pump can be
programmed individually as part of sample preparation sequence.
Beaker cover option (SAC950 only): Remove beaker covers using an
electro-magnet. Covers are disposed beside the sample changer in a
dedicated location once removed.
Technical specifications
61
Inputs/outputs
USB-B (Slave) connection to computer for sample preparation programming
sequence.
RS232C serial connection to workstation unit.
TTL I/O. Start external peripherals (RJ11 socket). Wait for external signal
to continue the sample preparation sequence.
Electro-magnet input for beaker cover option (RJ11 socket).
Propeller stirrer output (RJ11 socket).
Temperature input (CINCH socket) for use as input in a sample preparation
sequence.
Ultrasonic transducer input for beaker detection module
(Jack socket 3.5 mm).
Casing, keyboard
PETP – PVC and/or PE
Protected against projections, IP54 – For use in a laboratory environment
Dimensions (H x W x D)
70 x 58 x 68 cm.
Weight
27 kg.
Power requirements
47.5 - 63 Hz. 115/230 Vac +15 -18%. 76VA
Fuses: 2 mains fuses : slow blow T2.5A L250V (230 V)
Environmental operating conditions
Interior use only
10 to 40°C temperature.
20 to 80% relative humidity.
Altitude ≤ 2000 m.
Variations of the mains voltage do not exceed ± 10% of the nominal value.
Level of pollution: 2.
Transitory overvoltage: level II.
62
Technical specifications
Rear panel
On/Off switch
Fuse holder:
2 mains fuses
slow blow, T2.5A L250V
Line cord:
230 V (A95S001)
115 V (A95S002)
PC
USB cable
RJ11 socket
Jack 3.5 mm socket
Ultrasonic sensor
(beaker detection
module)
CINCH socket
Temperature sensor
TTL I/O
Start external
peripherals
RJ11 socket
Electromagnet
(Beaker cover module)
RJ11 socket
Propeller stirrer
24VDC
RS232 serial socket
Workstation control
unit cable A95X501
Technical specifications
63
International Standards
EMC Directive (89/336/CEE)
The sample changer complies with following standards:
Class A equipment for laboratory use, according to the standard
EN 61326-1.
Emission:
EN 61000-4-2 level 2
EN 61000-4-3 level 1
EN 61000-4-4 level 2
EN 61000-4-5 level 2
EN 61000-4-6 level 2
EN 61000-4-11
Low Voltage Directive (73/83/CEE)
The sample changer complies with the following standard:
Reference standard: EN 61010-1.
The Titration Manager complies with the following standards:
UL 61010A - 1
CAN / CSA C22 2 N° 1010.1 - 92.
64
Technical specifications
The sample changer complies with part 15 of the FCC Rules. Operation
is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including
interference that may cause undesired operation.
See "Compliance with FCC rules, part 15 - Information to the user" on
page 6.
Technical specifications
65
66
Technical specifications

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.3
Linearized                      : No
Page Count                      : 66
Producer                        : GPL Ghostscript 8.50
Create Date                     : 2007:12:14 10:38:29+01:00
Modify Date                     : 2007:12:14 10:38:29+01:00
Title                           : D21T085_UG_SAC9xx_gb.pdf
Creator                         : PDFCreator Version 0.9.0
Author                          : lchapus
Subject                         : 
EXIF Metadata provided by EXIF.tools
FCC ID Filing: UHC-TTL075

Navigation menu