Relion Prime Owner S Manual

2014-07-06

: Relion Relion-Relion-Prime-Owner-S-Manual relion-relion-prime-owner-s-manual relion pdf

Open the PDF directly: View PDF PDF.
Page Count: 70

Prime
Blood Glucose Monitoring System
User
Instruction
Manual
2
CAUTION:
Before using any product to test your blood sugar (blood
glucose), read all instructions and practice the test. Do all
quality control checks as directed and consult with a diabetes
healthcare professional. These recommendations apply to all
blood glucose monitoring systems and are supported by the
American Association of Diabetes Educators, the American
Diabetes Association, the U.S. Food and Drug Administration,
and the Health Industry Manufacturers Association.
3
Introduction ...........................................6
Intended Use .......................................7
Important Information .................................8
The ReliOn® Prime System ................................9
Understanding the ReliOn® Prime Blood Glucose Meter ........10
Explanation of the Full Display Screen ...................11
ReliOn® Prime Test Strips ................................12
Testing with ReliOn® Prime test strips .......................13
Important Test Strip Information ........................14
Changing the Battery ................................15
Meter Set Up .......................................18
Setting the Time and Date ............................18
Blood Glucose Testing ...................................22
ReliOn® Lancing Device ...............................22
Obtaining a Drop of Blood ............................25
Performing a Blood Glucose Test ..........................27
TABLE OF CONTENTS
4
Alternate Site Testing (AST) ............................... 31
Important Information About AST ....................... 31
Preparing to Test Your Blood Glucose From an Alternate Site .......33
Performing a Blood Glucose Test From an Alternate Site .....34
Using Control Solution ..................................39
Recommended Use of Control Solution ..................40
Performing a Control Solution Test ......................41
Comparing Control Solution Results .....................45
Downloading Test Results to a Computer ....................46
Understanding Test Results ............................... 47
Comparing Meter and Laboratory Results ...............47
Expected Blood Glucose Readings .....................50
High Blood Glucose Readings .........................50
Low Blood Glucose Readings .........................51
Diabetes Findings ..................................51
Meter Memory ........................................54
5
Marking Results in the Memory ........................54
Marking Control Solution Tests .........................54
Marking Other Results. ...............................55
Recalling Results ....................................55
Display Screen Images ..................................59
Caring For Your ReliOn® Prime Blood Glucose Meter ...........64
Cleaning the Meter ..................................64
Storage. ...........................................64
Specifications .........................................66
Warnings, Precautions, and Limitations .....................67
Warranty. .............................................69
6
INTRODUCTION
Thank you for choosing the ReliOn® Prime Blood Glucose Monitoring
System. Frequent blood glucose testing is important. It tells you how your
diabetes is affected by factors such as medication, diet, exercise, and stress
management. Blood glucose test results can also tell you if your diabetes
is changing in ways that affect your treatment plan. Always consult your
diabetes healthcare professional before making any changes. All of the
information needed to use the ReliOn® Prime System is included in this
manual. Please read it thoroughly before using the system. If you have any
questions, call Customer Service at 800.631.0076 (24 hours a day, 7 days a
week). Contact your healthcare professional with questions if you are unable
to reach Customer Service.
7
Intended Use
The ReliOn® Prime Blood Glucose Monitoring System is intended for the
quantitative measurement of glucose in fresh capillary whole blood samples
drawn from the fingertips or palm. Testing is done outside the body (
In Vitro
Diagnostic Use). It is indicated for use at home (over the counter [OTC]) by
persons with diabetes, or in clinical settings by healthcare professionals, as an
aid to monitor the effectiveness of diabetes control. It is not intended for the
diagnosis of or screening for Diabetes Mellitus, and is not intended for use on
neonates.
8
Important Information
 
 
glucose (hyperglycemia)
 
 
consult your diabetes healthcare professional immediately
 
have been followed carefully, contact your diabetes healthcare
professional
 
operating temperature
9
The ReliOn® Prime System Includes:
 
 
 
 



THE ReliOn® Prime SYSTEM
10
Back
Battery Compartment
Cover
UNDERSTANDING THE ReliOn® Prime BLOOD GLUCOSE METER
Front
Display Screen
Back Button

Forward Button
Test Strip Port
Data
Connection
Port
11
Explanation of the Full Display Screen
Time
Date
Average Test
Reading
Control
Solution Test
Low
Battery
Unit of
Measurement
Temperature
Warning
Test Strip
Apply Blood

Test Strip
Memory Mode
Error Code
Result Warning

Test Countdown
Test Results
Error Codes
If any part of the display screen is missing when the meter is turned on call
Customer Service at 800.631.0076.
12
These capillary action test strips require
only 0.5 micro-liter (µl) of blood and provide
an end-fill sample application tip for easy
blood application. The ReliOn® Prime Blood
Glucose Monitoring System measures the
amount of glucose (sugar) in fresh capillary
whole blood. When a drop of blood is added
to the sample application tip, the sample is
drawn into the test strip reaction site. The
glucose in the blood reacts with the enzyme
on the test strip. This reaction produces
a current, proportional to the amount of
glucose in the blood. The meter detects the
current and converts it into a blood glucose
reading.
TEST STRIP
Reaction
Site
Test Strip
Contact Bars
Sample Application Tip
(apply blood to end of test strip
and capillary action will pull blood
to reaction site quickly and easily)
ReliOn® Prime TEST STRIPS
13
Important Test Strip Information
  

 
 
 
 
bottle
 
expiration date ( )
The ReliOn® Prime Blood Glucose Monitoring System works with
the ReliOn® Prime Test Strips.
TESTING WITH ReliOn® Prime TEST STRIPS
14
Important Test Strip Information
 
strips after 90 days (3 months)
 
 
 
 
15
Note:
Changing the battery should only take a few seconds. If the battery is
changed within 30 seconds, the meter memory is not affected. You will not
lose any past test results and the settings are not affected.
However, if it takes longer than 30 seconds to change the battery, or if the
battery is dead before changing it, the time and date settings may need to
be reset.
The first time the meter is turned on after changing the battery, you may see
a display screen message (E00). This means you should check that the time
and date are correct. (See pages 18-21 for detailed instructions on changing
the time and date).
Changing the Battery
The ReliOn® Prime Blood Glucose Meter has a CR2032 battery installed.
16
When the battery power is getting low, your meter shows . You have
enough power for about 100 tests before you need to replace the battery.
If the is flashing, there is only enough power for about 20 tests. Replace the
battery as soon as possible.
When the battery has run out, the meter will not work.
CR 2032
Li-Mn 3V
+
WARNING:

doctor immediately.
17
How to Change the Battery
Step 1
Turn the meter off. The battery cover is on the back of the meter.
Step 2
Pull the battery cover up with your finger or fingernail. Remove battery cover.
Step 3
Remove the old battery. If necessary, turn the meter over and tap it gently
against the palm of your hand until the battery falls out. Avoid the use of
sharp objects that may damage the meter.
Step 4
Insert the new battery. The “+” on the battery should be facing up.
Step 5
Replace the battery cover.
18
Meter Set Up
Before using the ReliOn® Prime Meter for the first time, you must set the time

or pm) and the Military 24-hour format (0-23). The date format can either be

Do not perform a blood glucose test until you have set the meter correctly.
Setting the Time and Date
Step 1: Turn on meter

any part of the display screen is missing (see page 11), call Customer Service.
The meter then displays the time, date, unit of measurement, and the insert
test strip symbol.
Step 2: Enter set-up mode

set-up mode.
19
Step 3: Set the time format
The time can be displayed in either 24-hour format or
12-hour format with AM and PM. The meter displays
the current time format. To change format, press

button to select the time format. The date format then
appears on the display.


Step 4: Set the date format


the current format. To change format, press either the

select the date format. The hour setting then flashes
on the display.
20
Step 5: Set the hour
Press and release the Back or Forward button to change the hour.
Hold the Back or Forward button down to scroll through quickly.

minute setting then flashes on the display.
Step 6: Set the minute
Press and release the Back or Forward button until the correct
minute appears. Hold the Back or Forward button down to
scroll through quickly. When the correct minute is displayed,

display.
Step 7: Set the month (If day/month format is selected then
Step 7 will set day)
Press and release the Back or Forward button until the correct
month appears. Hold the Back or Forward button down to scroll
through quickly. When the correct month is displayed, press

Hour flashing
Minute flashing
Month flashing
21
Note:

date and exits the set-up mode. The meter automatically shuts off after 2 minutes if no
buttons are pressed. Settings stored up to that point are saved.
Step 8: Set the day
(If day/month format is selected then Step 8 will set
month) Press and release the Back or Forward button until
the correct day appears. Hold the Back or Forward button
down to scroll through quickly. When the correct day is

flashes on the display.
Step 9: Set the year
Press and release the Back or Forward button until the
correct year appears. Hold the Back or Forward button down
to scroll through quickly. When the correct year is displayed,

mode.
Day flashing
Year flashing
22
ReliOn® Lancing Device
BLOOD GLUCOSE TESTING
Step 1
Pry the depth
adjustment cap
sideways to remove it.
NOTE: Do not use the lancet if the cap is missing or damaged.
Release Button
Charger
Lancet
Safety Tab
Depth
Indicator
Deep Depth Setting
(Longest Marker)
Shallow Depth Setting
(Shortest Marker)
Medium Depth Setting
Depth Adjustment Cap
Clear Cap
for Alternate
Site Testing
NOTE: Lancing device is for single-patient use only and should never be shared with
another person, even a family member. This lancing device should not be used for
multiple patient testing by healthcare providers or at healthcare provider sites.
23
Step 3
Twist off the safety tab.
Step 4
Attach the depth adjustment cap.
Step 2
Insert a new, unused lancet into the Holder
until it stops.
Holder
24
Step 5
Adjust the depth of puncture setting if necessary. If the obtained amount of
blood is excessive, turn the dial to a lower setting. If the obtained amount of
blood is insufficient, turn the dial to a larger setting.
1 2 3 4 5
Deeper
Lancing Device.
Step 6
Pull the charger to reset the device.
You may hear a click.
25
Obtaining a Drop of Blood
Step 1
Wash hands with soap and warm water. Dry hands
thoroughly. If you use alcohol wipes to cleanse fingers,
make sure the area is dry before lancing finger.
Residual alcohol may lead to inaccurate readings.
Step 2
Let the arm hang down for 10 to 15 seconds. (Holding the
arm below the heart and massaging the wrist, palm and
finger makes it easier to obtain a blood drop).
Step 3
Choose a site on the side of a fingertip to minimize pain.
Step 4
Press the depth adjustment cap to the side of the
fingertip. Push the release button.
26
Step 5
Set aside the Lancing Device and wait a few seconds for a blood drop to

massaging your wrist, palm or the base of the finger helps the flow of
blood. Use a different site each time you test to help prevent soreness and
calluses.
27
Note: For Instructions on Alternate
Site Testing, refer to pages 31-38.
Step 1
Insert a ReliOn® Prime Test Strip
into the meter. Insert the contact bars
into the test strip port, with the reaction
site facing up.
Sample Application Tip
Reaction Site
Contact Bars
Test Strip Port
When the test strip is inserted, the meter turns on
automatically. If any part of the display screen is missing
(see page 11), call Customer Service.
PERFORMING A BLOOD GLUCOSE TEST
28
The test strip and flashing blood drop tells you the meter
is ready for you to apply a drop of blood to the sample
application tip at the end of the test strip.

flashing
Step 2
Lance your finger using the Lancing Device. Touch the
sample application tip at the end of the strip to the blood
drop on your finger. The capillary action will pull the blood
to the reaction site.

receive an abnormal (high) result.
CORRECT
INCORRECT
29

not start the meter or can start the meter count down and provide an “E13”
error message. Ensure blood completely fills reaction site.

when it detects that blood has been applied. Discard test strip if meter does
not start to count down after blood drop is applied.
Displaying the Result
The meter will count down
showing a flashing dash. The
result is displayed in 7 seconds.
Enough
Blood

Enough
Blood
Too
Much
Blood
30
Step 3
Always record your results in your self-testing logbook along with other
information such as insulin dosage, diet, and exercise. The result is
automatically stored in memory with the time and date.
Step 4
A lancet should be used once. After use, put protective cap back on and
remove lancet from lancing device. Dispose lancet in a safe manner as not to
cause accidental injury. Remove the used test strip and dispose of it carefully.
The meter shuts off automatically when the test strip is removed.
31
Important Information About AST
Sites other than your fingertip may have fewer nerve endings so obtaining
a blood sample from these sites may be less painful. The technique for
alternate site testing is different from fingertip testing. Blood glucose results
from sites other than your fingertip could be significantly different due to
blood glucose levels changing rapidly after a meal, insulin, or exercise.
Consult with your diabetes healthcare professional prior to testing from
a site other than your fingertips.
Results obtained from sites other than the fingertip may differ significantly
as is the case with all blood glucose monitoring systems. Rapidly changing
results are likely to occur after eating. Insulin dosing and physical activity
are shown in the fingertip more quickly than in alternate sites.
On average, blood taken from the palm may give results as much as 5%
higher than blood taken from the fingertip.
ALTERNATE SITE TESTING (AST)
32
Use Fingertip When Testing:
 
 
 
  
sugar, or suffer from hypoglycemic unawareness (you cannot tell when
you have low blood glucose)
 
how you feel
 
Consider Alternate Site Testing When:





33
Ask your diabetes healthcare professional about recommended testing
procedures before operating machinery or driving a car, as fingertip testing is
usually the preferred method of testing under these circumstances.
If bruising occurs, you may choose to lance a fingertip instead.
Preparing to Test Your Blood Glucose From an Alternate
Site
Choose a Site: Select a soft, fleshy area on the
palm that is free of visible veins and moles and away from
bone. Wash the area with warm, soapy water. Rinse and
dry completely. If you use alcohol wipes to cleanse the site,
make sure the area is dry before lancing the site.
34
Step 2
a) Insert a test strip into the meter.
Insert contact bars into test strip port
with reaction site facing up. When the
strip is inserted, the meter turns on
automatically. Make sure all symbols
appear on the screen (see page 11).
Performing a Blood Glucose Test From an Alternate Site
Step 1
Attach clear cap for Alternate Site Testing
to lancing device. Insert lancet and cock
lancing device.
Sample
Application Tip
Reaction Site
Contact Bars
Test Strip Port
IMPORTANT!
Insert into
meter this
way
35
b) Wait until strip with flashing blood drop appears.
(You now have 2 minutes to apply blood drop before
meter shuts off).
Step 3
a) Vigorously rub the selected area for 5 to 10 seconds
until it starts to feel warm to the touch.
b) Firmly hold the cocked lancing device against the
clean skin for 5 to 10 seconds.
c) Press the release button on the lancing device to
lance the skin. Continue to hold the lancing device
firmly against the skin until a 0.5 µL blood drop forms
d) Once a large enough drop of blood has formed,
remove the lancing device.
36
IMPORTANT

µ

IF NOT, REPEAT TEST.
Step 4
Pick up the meter. Touch the end of the test strip
to the blood sample until the reaction site is full.
Immediately remove the meter and test strip from the
blood drop.
37
Step 5
The meter will count down, showing a flashing dash and
display result in 7 seconds.
CONSIDER YOUR RESULT.
REPEAT ALTERNATE SITE TEST:





Any of the above situations can lead to an inaccurate test result.
Enough
Blood

Enough
Blood
Too
Much
Blood
38
If the repeated alternate site result is still not consistent with how you feel,
confirm your blood glucose level with fingertip testing.
Step 6
Always record your results in your self-testing log book along with other
information such as insulin dosage, diet, and exercise. The result is
automatically stored in memory with the time and date.
Step 7
Remove the test strip and dispose of it carefully. A lancet should be used
once. After use, put protective cap back on and remove lancet from lancing
device. Dispose lancet in a safe manner as not to cause injury. The meter
shuts off automatically when the test strip is removed.
39
Use Control Solution to check that the meter and test strips are working
correctly. It is important that you carry out this simple check regularly.
Control Solution contains a known amount of glucose that reacts with the

and High.
Compare the control solution test results with the range printed on the test
strip bottle label.
To order control solution, call ReliOn® at 800.631.0076.
USING CONTROL SOLUTION
40
It is recommended to use Control Solution:
 
 
properly.
 
with clinical symptoms.
 

levels above 80%.
 
 

 
41
Performing a Control Solution Test
Before performing a control solution test, always check the expiration date
( ) of the control solution. Do not use control solution if it is past the
expiration date or if the solution has not been stored correctly. When you open
a new bottle of control solution, always write the date of opening on the control
solution label. As soon as you have used the control solution, put the cap back
on the bottle. Always use the control solution within 90 days (3 months) of first
opening.
42
Step 1
Insert a ReliOn® Prime Test Strip into the
test strip port of the meter with contact bars
facing up. The meter turns on automatically.
The full screen appears briefly. If any part of
the display screen is missing (see page 11), call
Customer Service. The screen then displays the
test strip symbol and flashing blood drop.
Contact
Bars
Test Strip
Port
Reaction Site Sample
Application Tip
43
Step 2
You must press the Back or Forward button
BEFORE performing control solution test.
Hold the Back or Forward button until
appears in the upper right corner of the
screen. Verify that remains displayed on
the screen. If you do not do this, the control

will prompt you to add control solution.
Step 3
Mix control solution by gently inverting control solution
bottle several times. Remove the cap from the control
solution bottle. Place cap on flat surface. Squeeze the
bottle and discard the first drop. Apply the second drop
to the top of the cap.
44
Note: Dye added to control solution may
stain surfaces or fabric. Use with care.
Step 5
Compare the result with the range printed on
the test strip bottle you are using. Values may
vary from bottle to bottle. Verify that the result is
within range of the control solution level that you
are using. If the result falls within the range on the
bottle, the meter is working correctly.
Step 6
Remove the test strip. The meter shuts off automatically when the test
strip is removed.
Dispose of the test strip immediately.
06146A
2010-10
80-120
200-300
Control Solution Range (mg/dL)
(Rango de la solución de control)
Normal:
High:
Step 4
Bring meter and test strip to the drop. Test strip will draw up the
solution. The meter will show result in 7 seconds.
45
Comparing Control Solution Results
If the result is outside the range printed on the test strip bottle, repeat the

 
solution
 
 

 
 
 
 
 
 
test strip
46
DOWNLOADING TEST RESULTS TO A COMPUTER
We offer diabetes management software to help you download your test
results to a computer. For more information on Diabetes Management
Software, call Customer Service at 800.631.0076 (24 hours a day, 7 days
a week).
Repeat the test, carefully following instructions, using control solution mode
with new control solution or a new test strip, as appropriate. Compare the
result with the range printed on the test strip bottle you are using.
Note:
If your control solution result continues to fall outside the required range, the

If you are unable to resolve the problem, call Customer Service at 800.631.0076 (24
hours a day, 7 days a week).
47
NOTE: The ReliOn® Prime Meter is designed to give results at temperatures


Comparing Meter and Laboratory Results
Your ReliOn® Prime Blood Glucose Meter is referenced to fresh plasma blood
measurement. This was performed in a clinical laboratory with glucose

Gaithersburg, MD, USA). The results you get from your meter may differ from
the tightly controlled clinical laboratory result. This is due to normal variation.
To make an accurate comparison between the meter and laboratory results,
follow the guidelines below.
UNDERSTANDING TEST RESULTS
48
Before You Compare
Perform a control solution test to make sure your meter and test strips are
working correctly. Wait at least 2 hours after a meal or drink (other than water)
before doing comparison blood tests. Fasting is essential if venous blood is
required for the reference analyzer.
When at the Laboratory
The capillary samples for both tests must be taken and tested within 15
minutes of each other. The sample must be collected with anticoagulant
appropriate to the laboratory analyzer. The recommended technique should

guidelines.
ALWAYS wash hands in warm, soapy water. Rinse and dry them thoroughly
before testing.
ALWAYS use fresh capillary whole blood on meter tests.
49
You may still experience a difference from the laboratory result because
blood glucose levels can change over short periods. This can be for a variety
of reasons, such as food, exercise, medication, hematocrit, stress, or loss of
body fluids.
Data analysis should be performed using a clinically acceptable method.
Circulating blood glucose fluctuates naturally over time. Test results can vary
over time and can be affected by several factors. Some of these include, but


your result does not seem correct after testing your blood, repeat the test. If
abnormally high or low readings persist, contact your healthcare professional.
50
Expected Blood Glucose Readings
Expected blood glucose levels for people without diabetes:1, 2
High Blood Glucose Readings

the test with a new test strip. If this message shows again, contact your
healthcare professional immediately!
Contact your physician for advice if test results are very high3 (above 240

include dry mouth, thirst, frequent urination, nausea, vomiting, blurred vision,
sleepiness, or abdominal pain. Symptoms will vary from person to person.
You may have one or all of these symptoms.
mg/dL
  
  
51
Low Blood Glucose Readings

test with a new test strip. If this message shows again, contact your healthcare
professional immediately!
Contact your physician for advice if test results are very low4 

glucose include sweating, shakiness, trembling, blurred vision, hunger, headache,
confusion, rapid heartbeat, or tingling or numbness around the mouth or
fingertips. Symptoms will vary from person to person. You may have one or all of
these symptoms.
Diabetes Findings

of people with Type 1 diabetes. This study, called the Diabetes Control and
Complications Trial (DCCT), found that by keeping your blood glucose close to the
levels of people without diabetes you can reduce the risk of complications involving
the eyes, kidneys, and nervous system by approximately 60%5.
52
What This Means for You
Frequent blood glucose testing is the best means you have for keeping track of
how well you are doing with the factors that affect your diabetes-medication, diet,
exercise, and stress management. Blood glucose test results can also tell you
whether your diabetes is changing in ways that might require an adjustment to
your treatment plan. Always consult your healthcare professional before making
any changes in managing your diabetes.
Frequency of Testing
How often you need to test your blood glucose will vary according to your
age, the type of diabetes you have, the medications you are taking, and your
physical and emotional health. Your healthcare professional will guide you.
After deciding when and how often you should test, it is important that you
make testing part of your routine.
53
References
1 “Denitionanddiagnosisofdiabetesmellitusandintermediatehyperglycaemia”WorldHealthOrganization,2006
2 DiabetesCare,vol.33Jan;1S82-86,2010
3 Krall,L.P.andBeaserR.S.:JoslinDiabetesManual.Philadelphia:LeaandFibiger(1989),261–263.
4 Kahn,R.andWeir,G.:Joslin’sDiabetesMellitus.Philadelphia:LeaandFibiger(1994),489.
5 AmericanDiabetesAssociationpositionstatementontheDiabetesControlandComplicationsTrial(1993).
Recommendations for better diabetes control
 
 
professional.
 
 
 
 
hyperglycemia (high blood glucose).
 
keeping a record of your blood glucose results, along with information on
your food intake, exercise, and medication.
 
regularly.
54
Marking Results in the Memory
You can mark results in the meter memory in two ways.
Marking Control Solution Tests
You must mark results with the control bottle symbol . Insert test strip into
the meter. Hold the Back or Forward button until appears in the upper right
corner of the screen. Do this BEFORE performing a control solution test. If
you do not mark control solution tests before applying control solution to the
test strip, the results may not be accurate and control solution test results will
be included in your test result averages.
The meter does not use results marked when it calculates the 14- and
30-day average. This ensures that your 14- and 30-day average includes only
results from testing your own blood.
To mark a result , press the Back or Forward button BEFORE testing with
control solution. (See page 41 for detailed instructions on testing with control
solution).
METER MEMORY
55
Marking Other Results
You can also mark results with an when you do NOT want to include the result
in your 14- and 30-day average. For example, if someone else used your meter, or
if you believe the test result is incorrect because you applied blood to the test strip
incorrectly, use the exclamation mark to ensure that your 14- and 30-day average
includes only those results from testing your own blood that you want included in
your average. To mark a result with , press the Back or Forward button AFTER
the meter displays the test result.
Recalling Results
Step 1

meter. Then press the Back or Forward
button. Do not insert a test strip to turn on
the meter.
is displayed when the memory mode is
entered. The meter displays the most
recent test result stored in the
memory.
mg/dL
56
Step 2
Press the Back or Forward button to display the
average of tests over the last 14 days. This can include
1 to 250 results, depending on how frequently you
test. The number at the top (20n) shows the number of
tests that have produced the average.
is displayed when the average is shown.
Step 3
Press the Back or Forward button again to display the
30-day average. The greater the frequency of testing
the higher this number will be.
is displayed when the average is shown.
If no results are stored, or you are using the meter for
the first time, the meter will display 3 bars.
Displaying the
average (14 day)
Displaying the aver-
age (30 day)

57
If a test has not been performed in the last 14 days the meter
will display the 14-day average screen with 3 bars. This
means there are no results stored to produce the average. If
a test has not been performed in the last 30 days the meter
will display the 30-day average screen with 3 bars.
Step 4
Press the Back and Forward buttons to recall other results
stored in the memory. Pressing the Forward button after
the average displays the oldest test in memory. Pressing
the Back button after the average displays the second most
recent test in memory.
The meter displays the time each test was recorded in the
upper left corner. The date of the test appears below the
time.
58
Results marked with when the test was performed are not
included in the average. The marked results can be seen when
you scroll through the results saved in the memory.
The memory mode also displays if the result was marked
before using control solution. These results are not included in
the 14- and 30-day average.
If the thermometer symbol is flashing when you perform a
test, the thermometer symbol will be displayed in the memory
with the result. These results will not be included in the meter
averages.
NOTE:


Step 5


59
A list of all display screen messages is included below.
If you have any questions after reading these messages, or if the meter
displays a message not listed here, call Customer Service at 800.631.0076
(24 hours a day, 7 days a week). If the display screen shows , you should be
able to clear the error message yourself. Do not send your meter to your local
dealer without authorization from a company representative.

test using a new test strip. If this message appears again,
contact your healthcare professional immediately.

the test using a new test strip. If this message appears
again, contact your healthcare professional immediately.
DISPLAY SCREEN MESSAGES
60
The battery has been replaced.
Check that the time and date are set correctly.
The battery power is getting low.
There is enough power for approximately 100 tests
before the battery needs to be replaced.
A used strip has been inserted or the meter is having
difficulty reading the strip.
Repeat the test with a new test strip.
61
The battery power is low.
There is only enough power for about 20 more tests.
Replace with a new CR2032 battery as soon as
possible.
The battery power is too low for testing.
Replace with a new CR2032 battery.
62
Thermometer symbol - E6.


meter and strips to warm up or cool down slowly to


Thermometer flashing.
The meter and test strips are outside the optimum

sure your meter and strips are in an environment with a

test result obtained where the thermometer is flashing
with caution. The flashing thermometer will be displayed
on all screens during glucose testing until the meter has
reached acceptable temperature. Readings obtained with
the flashing thermometer will show a thermometer when
results are recalled in memory.
63
Corrupted memory.
Call Customer Service at 800.631.0076.

Retest with new test strip. Apply at least 0.5µL of
blood to sample application tip.
64
Cleaning the Meter
The ReliOn® Prime Blood Glucose Meter is a precise instrument. Please handle
it with care. Clean the outside of the meter with a damp cloth only. Dirt, dust,
blood, control solution, or water entering the meter could cause damage.
Your ReliOn® Prime Blood Glucose Meter should not need decontamination
as no blood or control solution should come into contact with the meter if the
instructions are followed correctly.
Storage
The ReliOn® Prime Blood Glucose Meter is designed to be stored within a

 
source (radiator) or in a car in very hot or cold weather
 
to high humidity, such as in a bathroom or kitchen
CARING FOR YOUR ReliOn® Prime BLOOD GLUCOSE METER
65
 
 
such as a microwave oven, stereo amplifiers, or a mobile phone
 
bleach
66
TEST STRIPS: ReliOn® Prime
RESULT RANGE:  
CALIBRATION: 
SAMPLE SIZE: micro-liter (µL)
BLOOD SOURCE: CAPILLARY WHOLE BLOOD
HEMATOCRIT RANGE: 33%-52%
TEST TIME: 
GLUCOSE UNITS: 
ASSAY METHOD: 

SOURCED)
POWER SOURCE: 
BATTERY LIFE: 
DISPLAY TYPE: LIQUID CRYSTAL DISPLAY (LCD)
MEMORY: 
AUTOMATIC SHUT-OFF: 
SIZE: ""0.6"
WEIGHT: .48 OZ. (42 G)
OPTIMUM OPERATING RANGES: 
RELATIVE HUMIDITY: 20% - 80%
ALTITUDE: 10,000 FT. (3,048 METERS)
SPECIFICATIONS
67
 
substitute for pathology laboratory equipment and should not be used for
the diagnosis of diabetes.

you believe you are suffering from dehydration, consult your healthcare
professional immediately.

abnormally low results. A very low hematocrit level (below 33%) can give
abnormally high results.

in shock.

hyperglycemic-hyperosmolar state, with or without ketosis.


neonatal use.
WARNINGS, PRECAUTIONS, AND LIMITATIONS
68
IMPORTANT: Loworhighbloodglucosereadingscanindicateapotentially
seriousmedicalcondition.Ifyourbloodglucosereadingisunusuallylowor
high,oryoudonotfeelthewayyourreadingsindicate,repeatthetestusing
anewstrip.Ifyourresultisstillnotconsistentwithyoursymptomsorifyour
bloodglucoseresultislessthan60mg/dLorhigherthan240mg/dL,contact
yourdiabeteshealthcareprofessionalforadvice.

(3,048 m).

However, glucose values in specimens beyond this triglycerides level should
be interpreted with caution.

affect test results.

whole blood.
69
You are a valued ReliOn® customer of. It is important to us that you are completely
satisfied with your Blood Glucose Meter.
ReliOn® warrants that your Meter will be free from defects in materials and
workmanship for a period of five years from the date of the original purchase.
If during this time the meter does not work properly because of a defect in
materials or workmanship, ReliO agrees to replace or repair, free of charge
any and all parts proven to be defective and subject to warranty.
This warranty is in lieu of all other warranties, expressed or implied, including
any implied warranty of merchantability or fitness for any purpose, other than
stated herein.
This warranty does not apply to the performance of the ReliOn® Prime meter
that has been damaged by accident or has been altered, misused, tampered
with, or abused in any way. This warranty only applies to the original

WARRANTY
 

RETURNS
You must contact Customer Service at 800.631.0076 before returning your

meters without this authorization will not be accepted.

Navigation menu