Revison 0.8.6  Document No. 01424  Mobile Cardiac Monitor Operator Manual
Revison 0.8.6  Document No. 01424  Page 2 Table of Contents 1. Description……………………………………………………………..  32. Indications for Use……………………………………………………  43. Contraindications for Use……………………………………………  44. Precautions…………………………………………………………….  55. Parts……………………………………………………………………..  65.1  RhythmStar Monitor………………………………………………..  65.2  ECG Cable…………………………………………………………  85.3  Batteries…………………………………………………………….  85.4  Battery Charger…………………………………………………….  96. Electrode Application-Connecting the ECG……………………..  96.1  Connecting the ECG cable to the patient………………………..  96.1.1  2-Channel (3-lead) Electrode Placement……………………..  106.1.2  3-Channel (5-lead) Electrode Placement……………………..  116.2  Connecting the patient ECG lead cable to the monitor……….. 127. Using the Monitor……………………………………………………..  127.1  Powering on the monitor…………………………………………..  127.2  Checking ECG signal quality………………………………………  147.3  Recording a cardiac event…………………………………………  157.4  Entering symptoms and activity level associated withthe cardiac event………………………………………………………..  167.5  Recharging the battery…………………………………………….  178. Maintenance and Service……………………………………………  188.1  Cleaning…………………………………………………………….  188.2  Service………………………………………………………………  188.3  Troubleshooting……………………………………………………  198.4  Disposal of Battery…………………………………………………  209. Medical Device Symbols and Safety Signs………………………  2110. Specifications………………………………………………………….  2211  RhythMedix Limited Warranty………………………………………  23 12  Obtaining Warranty Service…………………………………………  23 13. Wireless Compliance…………………………………………………  2414. Arrhythmia Detection Performance……………………………….  2615. Accessories…………………………………………………………….  27Appendix A - Handling Instructions for Lithium Ion Battery….  28
Revison 0.8.6  Document No. 01424   Page 3 1. DescriponCaution:  US Federal law restricts this device to sale by or on the order of a physician. The  RhythmStar  device  is  a  portable,  battery  powered,  wireless  cardiac  monitor which  may  be  worn  by  a  patient  to record ECG and  activity level data  for  up  to 30 consecutive days.    The  device  can  capture  patient  activated  and  auto-triggered      cardiac events such as Bradycardia,  Tachycardia, and Atrial  Fibrillation  as  identified by  an  embedded  arrhythmia  detection  algorithm.  The  device  is  capable  to           automatically deliver the data to the server. The data can be delivered to the server wirelessly by using a built-in wireless data modem or via USB connection. A medical professional, using the server,  can adjust and program  the  device  configuration  and auto-triggering parameters. The  RhythmStar  device  consists  of  a  monitor,    a  patient  ECG  lead  cable,    an           externally rechargeable battery, and a wall battery charger. The RhythmStar device is intended to be used with 3rd party lead electrodes supplied to a patient by a physician or a monitoring  enter. High quality FDA approved lead electrodes should be used.  The  RhythmStar  system  supports  USB  connectivity  that  can  be  used  to  send  and    receive  data  from/to  RhythmStar  and  other  devices.  The  server  can  deliver           configuration  parameters  to  the  device,  such  as  monitoring  duration,  pre/post          activation  recording  duration,  auto-trigger  rate  and  duration  limits,  user  interface     preferences, and requests for additional data stored in the device memory. The device receives continuous ECG signal from lead electrode sensors attached to the  patient’s  body,  analyzes  the  data  and  stores  it  in  the  on-board  flash  memory.     Activity level data that represents the patient's activity related to the physical motion is measured by a built-in accelerometer and also stored in the device memory.
Revison 0.8.6  Document No. 01424  Page 4 2. Indications for UseThe RhythmStar System is intended for use by patients who either have or are at risk of  having  cardiac  disease  and  those  that  demonstrate  intermittent  symptoms           indicative  of  cardiac  disease  and  require  cardiac  monitoring  on  a  continuing           basis.    The  device  continuously  records  ECG  data  and  upon  detection  by  an  ECG analysis  algorithm  or  manually  initiated  by  the  patient,  automatically  delivers  the     recorded cardiac activity to the server where it is presented and can be reviewed by a medical professional. The  data  received  from  RhythmStar  device  can  be  used  by  another  device  for         arrhythmia analysis, reporting and signal measurements. The RhythmStar system is not intended to sound any alarms. The device does not deliver any therapy, administer any drugs, provide interpretive or  diagnostic  statements  or  provide  for  any  life  support.  RhythmStar  is  for  prescription use only (Part 21 CFR 801 Subpart D).  3. Contraindications for UseThe RhythmStar system is not intended for use under the following conditions:  Patients  with  potentially  life-threatening  arrhythmias  who  require  inpatientmonitoring.  Patents who the attending physician believes should be hospitalized. Infants weighing less than 10 kg. (22 lbs.).
Revison 0.8.6  Document No. 01424   Page 5 4. Precautions Use only patient ECG lead cables provided by Rhythmedix with RhythmStar. Disconnect patient ECG lead cable before using a defibrillator. When viewing the ECG data, the presence of pacemaker signals in the ECG traceshould  not  be  considered  true  representations  of  the  actual  pacemaker  stimulus amplitude.  To  avoid  unintended  battery  discharge,  do  not  leave  the  battery  in  RhythmStarwhen it is not in use.  To  receive  the  best  recording  results,  instruct  patients  to  stay  away  from  heavyelectrical equipment or other sources of electromagnetic interference.  Equipment such as electric blankets and heating pads are included in this group.  Avoid exposing RhythmStar or RhythmStar battery to water or excessive moisture. Do  not  expose  RhythmStar  or  RhythmStar  battery  to  extreme  temperaturesbeyond the limits shown in the environmental specifications. PLEASE NOTE:  No Computerized information is completely reliable and physicians should review all ECG results.
Revison 0.8.6  Document No. 01424  Page 6 5. PartsRhythmStar is shipped with the following components: 5.1  RhythmStar monitorThe RhythmStar monitor is a small, lightweight, portable, battery powered device which is typically worn on a patient’s belt or waistband by placing the monitor in a “cell phone type pouch” (not supplied by RhythMedix).  Monitor Components Letter  Description A  LCD display B  Record Button C  Menu Selection Buttons ACBFront view of the RhythmStar Monitor
Revison 0.8.6  Document No. 01424   Page 7 Monitor Components continued Letter  Description D  Battery Compartment E  Serial Number F  Battery Door Back view of the RhythmStar Monitor  (battery door removed, no battery inserted) DEF
Revison 0.8.6  Document No. 01424  Page 8 5.2  ECG CableThe patient’s ECG signal is conducted using the patient ECG lead cable.  RhythmStar will not work with any other cables than those supplied by the manufacturer.  With RhythmStar, you can choose between a 3– or a 5-lead ECG cable. 5.3  BatteriesRhythmStar uses an external rechargeable Lithium-Ion (Li-Ion) battery.  This easily removable battery must be recharged with the supplied battery charger.  Only IEC 62133 certified Li-Ion batteries supplied by RhythMedix are to be used with RhythmStar. 3-lead ECG Cable:  5-lead ECG Cable:
Revison 0.8.6  Document No. 01424   Page 9 5.3  Battery ChargerRhythmStar batteries must be charged with the included battery charger. 6. Electrode Application—Connecting the ECGRhythmStar operates with the supplied patient ECG lead cables only!  RhythmStar can operate with either a 3 or a 5-lead cable.  Any attempt to connect another type of cable could damage the device or cause injury to the patient.  Every effort should be made to handle and store the patient ECG lead cables with care to avoid damaging the lead wires and lead wires snaps.   6.1  Connecting the ECG cable to the patientThe quality of the ECG signal greatly depends on the contact between the electrode and the patient’s skin.  RhythMedix recommends use of high-quality, Holter          electrodes that have been approved by the FDA.  Proper preparation of the patient’s skin is required for obtaining a quality ECG recording.  It is best to refer to your    electrode manufacturer for instructions on proper skin preparation techniques.     However, the following points can assist in obtaining quality ECG recordings:  If necessary, shave hair from the areas on the chest where the electrodes are tobe placed.  At each location where an electrode is to be placed, clean the skin with water orrubbing alcohol and let the skin dry.   To avoid applying excessive pressure to the patient’s body, attach the snaps of theECG cable to the electrodes prior to placing them on the patient’s skin.  Generally, electrodes should be placed over bone structures.  Artifact and noiseresult from placement of electrodes over large muscles or fatty tissue. NOTE:  RhythmStar device is intended to be used with 3rd party lead electrodes supplied to a patient by a physician or a monitoring center. High quality FDA ap-proved lead electrodes should be used.
 Revison 0.8.6  Document No. 01424  Page 10 STEP 1:  Connect the patient ECG lead cable snap to the electrode.  STEP 2:  Remove the protective backing from the electrode.  STEP 3:  Apply the electrode to the patient’s skin.  Apply each electrode to match     Figures 1 or 2 in this manual or as instructed by the physician. 6.1.1  3-lead configuration Three color-coded lead wires are used to create two channels of ECG re    cording.  This is a typical electrode placement:Figure 1—3-Lead Electrode Placement #  Color  Location RA  White  Upper-right portion of chest, 2 to 3 inches below collarbone LL  Red  Below left breast, over lower      ribcage LA  Black  Upper-left portion of chest, 2 to 3 inches below collarbone
Revison 0.8.6  Document No. 01424   Page 11 6.1.2  5-lead configurationFive color-coded lead wires are used to create three channels of ECG              recording.  This is a typical electrode placement:Figure 2—5-Lead Electrode Placement #  Color  Location RA  White  Upper-right portion of chest, 2 to 3 inches below collarbone LL  Red  Below left breast, over lower rib-cage LA  Black  Upper-left portion of chest, 2 to 3 inches below collarbone RL  Green  Lower right rib margin over bone V1  Brown  Fourth intercostal space at the right boarder of the sternum
Revison 0.8.6  Document No. 01424  Page 12 6.2  Connecting the patient ECG lead cable to the  monitorThe cable has a metal connector which can only be inserted one way into the cable receptacle opening on the top portion of the RhythmStar monitor.  Once aligned, firmly insert the  cable connector into the cable receptacle as pictured below:   7. Using the monitor7.1  Powering on the monitorRhythmStar does not contain a separate On/Off power button.  When you are ready to begin using RhythmStar, insert a charged battery into the monitor as pictured on the following page.
Revison 0.8.6  Document No. 01424   Page 13 Step 1:  Locate the battery cover on the back of the monitor.  Use your thumb to push down and toward the bottom of the monitor to slide the battery cover off. Step 2:  Insert a 3.7 volt battery, included with the RhythmStar,  into the battery       compartment following the polarity instructions (“+” on the battery to “+” on the wall of the   battery compartment and “-” on the battery to “-” on the wall of the battery compartment).  Step 3:  Replace battery cover on the back of monitor by gently aligning it as pictured and use your thumb to push up and toward the top of the monitor to slide the battery cover on. Step 4:  If the battery was successfully     installed, after a few seconds, the screen should show the RhythmStar name and logo as pictured.   RhythmStar is now powered and ready for use.
Revison 0.8.6  Document No. 01424  Page 14 Step 1:  Make sure the patient ECG lead cable is  connected to RhythmStar and insert the battery.   Immediately after inserting the battery (and before the RhythmStar name and logo screen appears, you will briefly see the screen as pictured (a circle with counterclockwise movement). Step 2:  Press both Menu Selection buttons simultaneously.  Step 3:  Provided the patient ECG lead wires are  properly connected to the patient and the patient ECG lead wire connector is properly connected to RhythmStar, the patient’s ECG signal will display as  pictured.  Use the Menu Selection buttons to select Ch1, Ch2 and Ch3. (Ch3 only available when using the 5-lead ECG cable) Step 5:  When you are satisfied with the ECG quality, press  the Record Button to exit the ECG signal display and begin Monitoring the patient.  RhythmStar will now display “MONITORING” as  pictured.  7.2  Checking ECG signal qualityTo visualize the patient’s ECG signal on the RhythmStar’s LCD screen to inspect the quality of the ECG signal, follow the steps below:
Revison 0.8.6  Document No. 01424   Page 15 7.3  Recording a cardiac eventThe medical technician or physician should instruct the patient with an explanation of what should be marked as a cardiac event.  The patient should be instructed to press the Record Button to start a recording when he or she experiences symptoms for which RhythmStar was prescribed.  It is recommended that this feature be used by patients who can comprehend the instructions provided by the medical technician or physician and are capable of pressing the necessary buttons.  RhythmStar will also be automatically recording and transmitting asymptomatic cardiac events as they    occur without the patient needing to press the Record Button or interface with RhythmStar in any way.  Follow the below instructions to record a cardiac event:  Step 1:  Instruct the patient to press the   Record Button.STEP 2:  Upon pressing the Record Button, the screen will change to “RECORDING”.  A red progress bar will begin and last until the           recording is complete.NOTE:  Instruct patients to remain as still as   possible for the duration of the recording.
Revison 0.8.6  Document No. 01424  Page 16 7.4  Entering symptoms and activity level associated with the cardiac eventAfter a recording is completed, RhythmStar will then prompt the patient to select the  symptom or symptoms and the level of activity that he or she was feeling and doing at the time of pressing the Record Button.   Follow the below instructions to enter symptom(s) and activity level associated with the cardiac event: Step  1:    After  each  time  a  patient       activated  event  recording  is  completed, RhythmStar  will  then  display  the  “Select Symptoms” screen. The patient should be instructed  to  use  the  Menu  Selection    buttons to highlight the symptom selection(s) that best represent the reason for the event. By pressing the Record Button of a highlighted  symptom,  the  empty  circle next  to  the  selection  will  fill.    The patient can  repeat  this  step  to  select  more  than one  symptom  if  appropriate.    Once     symptom(s)  are  properly  selected,  the  patient  should  highlight  the  “NEXT”       selection and then press the Record But-ton to move on to the next screen. Step 2:  RhythmStar  will  next  display the  “Select  Activity”  screen.    The  patient should  be  instructed  to  use  the  Menu   Selection  buttons  to  highlight  the  activity level  that  best  represents  their  activity   level  at  the  time  that  the  Record  Button was  pressed.    Only  one  activity  level     selection  may  be  selected.    Once  the    activity  level  is  properly  selected,  the    patient  should  highlight  the  “SUBMIT”   selection  and  then  press  the  Record   Button.    RhythmStar  is  now  ready  to    record additional events.
Revison 0.8.6  Document No. 01424   Page 17 7.5  Recharging the battery NOTE:  Only IEC 62133 certified Lithium-Ion batteries supplied by RhythMedix are to be used with RhythmStar. RhythmStar  is  powered  by  an  externally  rechargeable  Lithium-Ion  battery.    Each RhythmStar  device  is  supplied  with  (2)  batteries  and  one  battery  charger.    It  is        recommended  that  while  one  battery  is  being  used  in  the  RhythmStar  monitor,  the  other battery is being charged.  We recommend  replacing the battery being used in the  RhythmStar  monitor  with  a  fully  charged  battery  every  24  hours  to  avoid  any    power outage on the monitor while RhythmStar is monitoring the patient.  To  charge  the  battery,  place  the  rechargeable  Lithium-Ion  battery  into  the  charger.  Next, plug the charger into the wall outlet (110 or 220V).  Once properly placed, the charger will show either a solid red or blue indication light as shown below: Charging—Indicated by solid Red light:  Charged—Indicated by solid Blue light:
Revison 0.8.6  Document No. 01424  Page 18 8. Maintenance and Service8.1  CleaningTo clean RhythmStar: 1. Remove the battery from the monitor (DO NOT ATTEMPT TO CLEAN BATTERY):2. Dampen  a  soft  cloth  with  a  mild  detergent  and  water  mixture.   An example of amild detergent is an alcohol-free  hand  soap  or  sodium  hypochlorite  (bleach)solution 10% in water.3. Clean the monitor and patient ECG lead cables.4. Remove  any    remaining  adhesives  from  the  patient  ECG  lead  cables  with  anadhesive tape remover solution or swab of mild detergent.NOTE:   Do not use alcohol or acetone to clean the lead wires as this can cause the wires to stiffen and the insulating plastic to crack.    Do not immerse the monitor in water or any other cleaning solution. Store the patient ECG lead cables suspended when possible. Avoid kinking the patient ECG lead cables.8.2  ServiceIf there is a problem with the monitor, review the Troubleshooting section for a listing of problems and solutions.  If additional assistance is required, contact RhythMedix customer support via phone at (856) 282-1080.  Be prepared to provide the following information:  Serial number of the monitor Description of the problemCall customer support before returning a recorder to make shipping arrangements.
Revison 0.8.6  Document No. 01424   Page 19 8.3  Troubleshooting
Revison 0.8.6  Document No. 01424  Page  20 8.4  Disposal of BatteryStorage and Disposing of Lithium-Ion (Li-Ion) batteries instructions: The RhythmStar monitor and battery should be stored at room temperature in a dry area.  Make sure that the battery is removed from the monitor while in storage.  In the event that a battery becomes damaged or no longer holds its charge, the       battery should no longer be used and should be recycled.  To locate a Rechargeable Battery Recycling Corporation (RBRC) collection point   within the US, visit the website listed below  and type in your postal or zip code.  The   closest drop off points in many stores such as Wal-Mar, Home Depot, Radio Shack, Best Buy will display. http://www.call2recycle.org/locator/ Place the batteries that you wish to recycle into a box or bag and take them to the   collection point.  Hand the batteries over to the RBRC representatives at the collection point for recycling.
Revison 0.8.6  Document No. 01424   Page  21 9. Medical Device Symbols and Safety Signs
Revison 0.8.6  Document No. 01424  Page 22 10. SpecificationsCharacteristics Test Conditions Min. Typical Max. Unit Physical   Length  101  mm  Width  66  mm  Thickness  13  mm    Weight  With battery  90  gm Functional ECG Channels  Cable selectable  2  2  3  n/a Accelerometer  ±2g/±4g/±8g Sampling Rate  Dynamically selectable  1.56  30  800  Hz Resolution  Variable  8  12  14  Bits Memory Recording Time  30  days Data Retention  microSDHC flash media  10  years Wireless Communication Technology  UMTS/HSDPA  850/1900 MHz GSM/GPRS/EDGE  850/900/1800/1900 Time-Slot Class  12  n/a Output Power  WCDMA Band II & V 0.302 GSM850  2.13 W PCS1900  1.74 Electrical CMRR  90  105  dB AC Range  ±6  mV DC Range  ±300  mV Input Impedance  >10  MOhm Frequency Response  0.05 to 40  Hz Recovery Time  1.33  sec ADC Resolution  12  12  16  Bits ADC Sample Rate  125  250  1000  Hz Battery Type  Replicable, externally re-chargeable Li-Ion 3.7  V Life  From full charge  72  hours    Storage Temperature  < 30 days  -20  50  C (shipped condition)  > 30 days  -20  35  C    Relative Humidity  45  85  % Environmental    Operating Temperature  0  40  C    Storage Temperature  -25  70  C    Relative Humidity  Non-condensing 23 C  10  93  %    Atmospheric Pressure  700  1060  hPa
 Revison 0.8.6  Document No. 01424   Page 23 11.  RhythMedix Limited Warranty  This RhythMedix product is warranted to be free from manufacturing and material   defects, excluding batteries, charging adapters, and patient cables, for a period of one year from the date of shipment from RhythMedix to the original purchaser (“Warranty Period”).  If a hardware defect arises and a valid claim is received within the Warranty Period, RhythMedix will repair or replace (at RhythMedix’s option) the defective product free of charge for parts and labor.   This warranty does not apply to any product which has been damaged by accident or which has been misused, abused, altered, or repaired by anyone other than       RhythMedix or its Representatives.  Except for the express warranties stated above, RhythMedix disclaims all warranties including implied warranties of merchantability and fitness.  The stated express     warranties are in lieu of all obligations of liabilities on the part of RhythMedix for    damages, including but not limited to, special indirect or consequential, arising out of or in connection with the use or performance of RhythMedix products.  Any warranty claims must be initiated within the one (1) year warranty period.  Any  repairs made to the product that are not covered by the warranty are billed to the  customer. 12.  Obtaining Warranty Service  To obtain repairs, first obtain a Returned Merchandise Authorization (RMA) number from your RhythMedix Representative.  Include the RMA number on the shipment and ship postage prepaid to:  RhythMedix, 5000 Atrium Way, Ste 1, Mt. Laurel, NJ  08054  Attention:  Repair, RMA # ___________  RhythMedix will return in-warranty units postage prepaid. 
 Revison 0.8.6  Document No. 01424  Page 24 13.  Wireless Compliance FCC ID: 2ACA9-10002  FCC Regulations Compliance Statement  The Federal Communications Commission (FCC) has adopted a safety standard for human ex-posure to Radio Frequency (RF) electromagnetic energy emitted by FCC-certified equipment. This wireless electro- cardiovascular monitor has been evaluated under FCC Bulletin OET 65C (01-01) and found to be compliant to the requirements of uncontrolled environmental limits as set forth in CFR 47 Sections 2.1091, 2.1093 addressing RF Exposure from radio frequency devices when operated in accordance with the operation guidelines described in this manual.  Proper op-eration of this radio device according to the instruction in this publication will result in user expo-sure substantially below the FCC recommended limits.  The RhythmStar contains a radio transmitter and receiver.  It is designed and manufactured not to exceed the emission limits for exposure to radio frequency (RF) energy set by the Federal Com-munications Commission of the U.S. Government.  These limits are part of comprehensive guide-lines and establish permitted levels of RF energy for the general population.  The guidelines are based on standards that were developed by independent scientific organizations through periodic and thorough evaluation of scientific studies.  The standards include a substantial safety margin designed to assure the safety of all persons, regardless of age and health.  The exposure standard for wireless mobile devices employs a unit of measurement known as the Specific Absorption Rate, or SAR.  The SAR limit set by the FCC is 1.6W/kg.  Tests for SAR are conducted using standard operating positions accepted by the FCC with the device transmitting at its highest certified power level in all tested frequency bands.  Although the SAR is determined at the highest certified power level, the actual SAR level of the device while op-erating can be well below the maximum value.  This is because the device is designed to oper-ate at multiple power levels so as to use only the power required to reach the network.  In gen-eral, the closer you are to a wireless base station antenna, the lower the power output.  Before a device model is available for sale to the public, it must be tested and certified to the FCC that it does not exceed the limit established by the government-adopted requirement for safe exposure.  The tests are performed in positions and locations (for example, worn on the body) as required by the FCC for each model.  This device has been tested and meets FCC RF exposure guidelines when worn in a holster or belt clip over belt or waistband of an outer garment with 15mm minimum separation distance from the body.   The highest reported SAR value for this body-worn device is: 1.44 W/kg.  The FCC has granted an Equipment Authorization for this wireless electro-cardiovascular monitor with all reported SAR levels evaluated as in compliance with the FCC RF exposure guidelines.  SAR information on this device is on file with the FCC and can be found under the “Display Grant” section of: http://transition.fcc.gov/oet/ea/fccid/ after searching on FCC ID: 2ACA9-10002.
Third-party belt-clips, holsters, and similar accessories used by this device should not contain any metallic components. Body-worn accessories that do not meet these requirements may not comply with RF exposure requirements and should be avoided.  FCC statement ·This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/TV technician for help.  NOTE: THE GRANTEE IS NOT RESPONSIBLE FOR ANY CHANGES OR MODIFICATIONS NOT EXPRESSLY APPROVED BY THE PARTY RESPONSIBLE FOR COMPLIANCE. SUCH MODIFICATIONS COULD VOID THE USER’S AUTHORITY TO OPERATE THE EQUIPMENT.
 Revison 0.8.6  Document No. 01424   Page 25  IC ID: 11948A-10002  Canadian Regulations Compliance Statement  RhythmStar Complies with the Canadian RSS - 102 Issue 4: 2010 and IEC 62209-2: 2010 re-quirements. RhythmStar Conforme à la norme RSS - 102 Issue 4: 2010 et IEC 62209-2: 2010 reguirements. This device complies with RSS 210 of Industry Canada.  This Class B device meets all the requirements of the Canadian interference-causing equipment regulations. Cet ap-pareil numérique de la Classe B respecte toutes les exigences du Réglement sur le matériel brouilleur du Canada.    This device complies with Industry Canada license exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.   Cet appareil est conforme aux normes CNR exemptes de licence d’Industrie Canada. Le fonc-tionnement est soumis aux deux conditions suivantes : (1) cet appareil ne doit pas provoquer d’interférences et (2) cet appareil doit accepter toute interférence, y compris celles susceptibles de provoquer un fonctionnement non souhaité de l’appareil.
 Revison 0.8.6  Document No. 01424  Page 26 14.  Arrhythmia Detection Performance  The RhythmStar incorporates a real-time embedded arrhythmia detection algorithm. The processing steps include signal bandpass and morphological filtering, estimation of the motion artifact, analysis of the slope, width and amplitude of the signal,          decision making logic to determine location and length of the QRS complexes, atrial activity analysis and template matching for AF detection.  Rhythmedix conducted algorithm performance testing according to the ANSI/AAMI EC57 standard. The MIT-BIH, AHA, and NST databases were used to verify           performance of the algorithm. The algorithm performance testing summary is         provided in the tables below:  QRS Detection Performance    AF Detection Performance    Database QRS Sensitivity QRS Positive Predictivity MIT-BIH  99.77%  99.81% AHA  99.72%  99.82% NST  96.59%  79.86% Database AF Sensitivity AF Specificity MIT-BIH AF  93.43%  96.88%
 Revison 0.8.6  Document No. 01424   Page 27 15.  Accessories  The following accessories are available for use with the RhythmStar: The following materials are provided with the RhythmStar: Part Number:   Description:  PC-10003  RhythmStar two-channel patient ECG lead cable PC-10005  RhythmStar three-channel patient ECG lead cable BP-10006  RhythmStar Rechargeable Li-Ion battery BC-10007  RhythmStar Li-Ion wall charger for battery PA-10008  RhythmStar replaceable battery door Part Number:   Description:  PC-10003  RhythmStar two-channel patient ECG lead cable BP-10006  RhythmStar Rechargeable Li-Ion battery BC-10007  RhythmStar Li-Ion wall charger for battery PA-10008  RhythmStar replaceable battery door
 Revison 0.8.6  Document No. 01424  Page 28 Appendix A—Handling Instructions for Lithium Ion    Battery Please read and follow the handling instructions for the battery before use. Improp-er use of the battery  may cause heat, fire, explosion, damage or capacity deterio-ration of the battery. However, the  manufacturer will not guarantee against any ac-cident caused by the usage which is not written here. (When using the battery) DANGER  Do  not  dip  or  wet  the  battery  in  water,  seawater,  or  other  liquid.  If  the  protecting  device  assembled in the battery is damaged, the battery may be charged with an abnormal current and  voltage, which may result in the cause of heat generation, explosion, or fire of the battery.  The battery has a predetermined polarity. If the battery will not connect well to the charger or  equipment, do not  try  to  connect  the  battery  forcefully.  Check  the  polarity  first.  In  the  case  the  battery  is  connected  in       reverse,  it  is  charged  reversely  and  may  cause  acid  leakage,  heat  generation, explosion,  or  fire  due  to  an   abnormal chemical reaction.  Do not put the battery into a fire or heat it.  In such a case, the insulator in the battery may be  melted, the gas release vent and protection mechanism may be damaged,  all of which may cause  heat generation, explosion, or fire.  Do  not  connect  the  battery  reversed  in  positive  (+)  and  negative  (-)  terminals  in  the  charger  or  equipment. In the case the battery is connected in reverse, it is charged reversely during charge,  and causes an  excessive current  during  discharge,  and  may  cause  heat  generation,  explosion,  or  fire  due to  an  abnormal  chemical  reaction.  Do not let the battery terminals ( + and - ) contact a wire or any metal ( like a metal necklace or a  hairpin  )  with which  it  carried  or  stored  together.  In  such  a  case,  the  battery  is  shorted  and  causes an excessive current, which may result in heat generation, explosion, or fire.  Do not apply heavy impact to the battery, or throw or drop it.    Strong impact may damage the  protecting device, which may result in heat generation, explosion, or fire of the battery.  Do  not  drive  a  nail  in,  hit  with  a  hammer,  or  stamp  on  the  battery.  In  such  a  case,  the  battery  may  be       deformed  and  shorted,  and  the  protecting  device  may  be  damaged,  which  may  cause  heat  generation,       explosion, or fire of the battery.  Do  not  solder  the  battery  directly.  Heat  applied  during  soldering  may  damage  the  insulator  or  the  gas     release vent and  protection mechanism, which  may result in acid leakage, heat  generation, explosion, or fire of the battery.  Do  not  disassemble  or  alter  the  battery.  The  battery  contains  the  protection  mechanism  and  protection device in order to avoid any danger. If these are damaged, heat, explosion or fire may  be caused.  Charge  the  battery  every  6  months  to  the  amount  specified  by  the  manufacturer,  even  if  the  battery is not used. An excessive over-discharge may cause an abnormal chemical reaction, which  may result in the cause of acid leakage, or fire of the battery.
 Revison 0.8.6  Document No. 01424   Page 29 WARNING  Do not put the battery in a microwave oven or a pressure cooker.    Sudden heat may damage the  seal of the battery and may cause heat generation, explosion, or fire of the battery.  Do not use the battery together with a dry battery or other primary battery or other battery of a  different  capac-ity,  types  and  /  or  brand.  In  such  a  case,  over-discharge  during  use,  or  over-charge  during  charge  may   occur and      abnormal  chemical  reactions  may  cause  heat  generation, explosion, or fire of the battery.  If  you  notice  any  bad  odor,  heating,  discoloration,  deformation,  or  any  other  change  from  what  your  are used  to  while  using,  charging,  storing  the  battery,  take  it  out  of  equipment  or  charger,  and  avoid using  it.   Using it in such state may result in heat generation, explosion, or fire.  If the battery leaks or emits a bad odor, take it away from any fire immediately.    The electrolyte  may catch fire, which may cause heat generation, explosion, or fire.  Do  not  let  leaked  electrolyte  come  into  contact  with  eyes.  In  such  a  case,  immediately  wash  the  area  of     contact  with  clean  water  and  seek  help  from  a  doctor.  If  not  treated  soon,  prolonged  contact may cause se-rious injury.  Do  not  place  or  leave  the  battery  and  equipment  in the  reach  of  infants.    Improper  use  of  the  battery may cause danger. CAUTION  Do not use or leave the battery in a place exposed to strong direct sunlight, or in a car under the  blazing sun, or high temperature sources.    Such a high temperature may cause acid leakage.  If you find the battery rusty, bad odor, heating, or any other defective before using the battery for  the first time after  purchase, do not use it. Take it back to the dealer instead.  In  case  young  children  use  the  battery,  instruct  them  on  the  contents  of  the  instructions  and  ensure the battery is correctly used by them at all times.  If the battery leaks and its electrolyte contact with skin or clothes, wash it well with tap water or  other clean   water right away. Leaving it, it is may cause a rash on skin.  If you  have  any  question  regarding  the  battery, contact RhythMedix.   Read the instructions of your equipment regarding the battery installation and removal from the  equipment so as not to mishandle and waste the battery.  The  battery  was  charged  a  little  before  shipment for  temporary  use  by an end  user.  In  case  your  equipment does not operate with the battery or in the case of a long use, charge the battery with a  specified charger once.  Careffully read the instructions of your equipment before use.  When the battery is expected not to be used for a long time, take the battery out of the equipment  and store it in a less humid area  In  the  case  the  battery  terminals  are  dirty,  clean  the  terminals  with  a  dry  cloth  before  use,  otherwise,  the contact  with  equipment  might  cause  insufficiency,  and  power  failure  or  charge  failure.  Despite  being  rechargeable,  the  battery  has  a  limited  life  span.  Replace  it,  when  usage  time  becomes short.  As  for  a  used  battery,  please  recycle,  after  covering  the  battery  terminals  (  +  and  -  )  with  a  insulation tape or    inserting it to individual poly-bag.
 Revison 0.8.6  Document No. 01424  Page 30 (When charging the battery)  (When discharging the battery)   DANGER  Do  not  use  any  battery  charger  not  specified  by  (  manufacturer's  name  ),  also,  follow  the  charge  conditions specified by ( manufacturer's name )    If the battery is charged under other conditions  (   a   high   temperature,  a   high   voltage   or  current,   or   an   altered   charger   )   not   specified   by  ( manufacturer's name ), the battery may cause heat generation, explosion, or fire with abnormal  chemical reactions.  Do  not  connect  the  battery  directly  to  an  electric  outlet  or  cigarette  heater  socket  in  a  car.  Applying a high     voltage may generate an excessive current, and get an electric shock. In such a  case, the battery may leak electrolyte, overheat, explode, or cause fire.  Do  not  charge the  battery near  fire  or  in  a  car  under  the  blazing sun.  Such  a  high  temperature  may  cause damage of the protecting device in the battery, which may result in heat generation,  explosion, or fire. DANGER  Do not use or leave the battery in a place near fire, heaters, or high temperature sources.    Such a  high           temperature may cause heat generation, explosion, or fire.  Do not use the battery with any equipment other than specified.  Any such practice may expose  some types of equipment to an abnormal current, which may result in heat generation, explosion,  or fire. WARNING  Discontinue   charging   after   specified   charging   time   even   if   the   charge   is   not   complete.  Otherwise, the battery might cause heat generation, explosion, or fire. CAUTION  Do  not  use  the  battery  in  other  than  the  following  conditions.    Otherwise,  the  battery  might  cause heat generation, damage.                     Charge:    0° C  ~  +45° C  Carefully read the instructions for the specified charger to learn how to charge the battery.  Do not charge the battery over the specified time described in the instructions. CAUTION  Do  not  use  the  battery  in  the  place  where  the  static  electricity  (  more  than  the  limit  of  the  manufacturer's guarantee  )  occurs.  Otherwise,  the  protecting  device  in  the  battery  might  be  damaged and cause heat  generation, explosion, or fire. Do not use the battery in other than the following conditions                                  Discharge  :  -20° C ~ +60° C Store (less than a month)  :  -20° C ~ +50° C (on the charge of 50 %)         Store (more than a month)  :  -20° C  ~  +35° C  (on the charge of 50 %)