Senior Technologies IDUT002 WanderGuard Universal Tester User Manual 0163 390 Universal Tester D

Senior Technologies Inc WanderGuard Universal Tester 0163 390 Universal Tester D

user manual

Download: Senior Technologies IDUT002 WanderGuard Universal Tester User Manual  0163 390  Universal Tester D
Mirror Download [FCC.gov]Senior Technologies IDUT002 WanderGuard Universal Tester User Manual  0163 390  Universal Tester D
Document ID703188
Application IDaYyNZ8OLDC059iHszW2qsA==
Document Descriptionuser manual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize12.14kB (151750 bits)
Date Submitted2006-09-11 00:00:00
Date Available2006-09-11 00:00:00
Creation Date2006-07-31 09:00:51
Producing SoftwareAcrobat Distiller 6.0.1 (Windows)
Document Lastmod2006-07-31 09:00:51
Document Title(0163-390) Universal Tester-D.qxd
Document CreatorPScript5.dll Version 5.2.2
Document Author: skw0621

WanderGuard® Universal Tester
Operating Instructions
Model 18029
WARNING
Read these instructions, particularly the Warnings and Cautions,
before using the WanderGuard® Universal Tester. Failure to do
so could result in injury or death to a person in your care.
0163-390
FCC ID:LA5IDUT002
This device complies with Part 15 of the FCC rules. Operation is subject to the
following two conditions: 1) this device may not cause harmful interference, and
2) this device must accept any interference received, including interference that
may cause undesired operation.
Changes or modifications not expressly approved by Stanley Senior
Technologies could void the user's authority to operate the equipment.
All content herein is the property of The Stanley Works, its Affiliates, or their
content suppliers and is protected by United States and international copyright
laws. The compilation of all content is likewise the exclusive property of The
Stanley Works (or the Affiliate identified in any copyright notice) and is protected
by United States and international copyright laws.
The trademarks, service marks and logos (the 'Trademarks') used and displayed
in this publication are registered and unregistered Trademarks of The Stanley
Works, its Affiliates and others. Nothing herein should be construed as granting,
by implication, estoppel or otherwise, any license or right to use any Trademark
displayed herein without the prior consent of the Trademark owner. ARIAL,
AUTISM ALLY, AUTISM ALLY design, the CATCH-ALL design, SENIOR
TECHNOLOGIES, TABS, TIM, WANDERGUARD, and the WANDERGUARD
design are some of the registered and unregistered trademarks owned by
Stanley Logistics, Inc. and/or its Affiliates ("Stanley Trademarks"). Stanley
Trademarks may not be used in connection with any product or service that is
not manufactured by or under license from The Stanley Works or its appropriate
Affiliate.
©2006 Printed in USA
WARNING
Test bracelets daily as detailed in these instructions, in the
bracelet instructions, and the WanderGuard® Door Monitor User
Manual. Record the results in the resident’s records. It is
important to test the bracelet before putting it into use and daily
thereafter. Failure to do so could result in injury to or death to a
person in your care.
The Universal Tester contains a long-life lithium battery that is
not user replaceable. Do not disassemble, heat above 100o C, or
submerge in water.
CAUTION
Dispose of the Universal Tester immediately when low battery is
indicated on the display.
When testing the bracelet, make regular inspections to check for
frayed or loose bands.
Stanley-Senior Technologies recommends the removal of the
bracelet during x-ray or MRI procedures.
Changes or modifications not expressly approved by StanleyStenior Technologies could void the user’s authority to operate
the equipment.
Features of the WanderGuard® Universal Tester
The Universal Tester is a hand-held, battery-powered device used to test
WanderGuard® door monitors and WanderGuard® bracelets.
When using the tester to test bracelets, the tester will detect if a bracelet
is working properly, display the individual ID number of the resident
wearing the bracelet, and indicate when the bracelet is reaching its
“replace bracelet” date.
The Universal Tester has a non-replaceable battery. Battery life will
vary depending on usage. The battery can be expected to last 2 years
under typical use.*
The tester will indicate when it’s battery life is low by displaying “LO”
on the LCD.
This message will flash three times when any button is pressed. When
this happens, the Universal Tester needs to be replaced immediately. A
dimming display is also an indication that the battery is low and the
tester needs to be replaced.
* Based on testing 40 bracelets per day and 10 doors per week.
The tester has an ambient light sensor built-in. When the tester is used
in low light conditions, it will automatically back-light the LCD display.
Using the RECALL button will display the result of the last bracelet
test.
AMBIENT LIGHT
SENSOR
BRACELET STATUS
LED
DOOR STATUS LED
BRACELET
TEST
DOOR
TEST
RECALL
Testing Bracelets
“End of Service” feature
The Model 36000 series has an “End of Service” feature. During routine
testing the Universal Tester will display when the bracelet is reaching
the end of its useful service. This lets you know it is time to replace the
bracelet BEFORE the bracelet becomes inoperable. Bracelets must be
tested daily. IMPORTANT: Failure to replace the bracelet “on time”
could result in injury to or the death of a person in your care.
To test a WanderGuard® bracelet with ID capability (36000 series):
•
•
Hold the Universal Tester within one foot of the bracelet.
Press the BRACELET button once. The LED above the button will
light briefly.
Press
button
The display will
animate until a
bracelet
is detected, or for
10
seconds.
•
If the bracelet is still operational, the LED will flash green four times
and the ID number of the bracelet will be displayed.
LED blinks
GREEN 4 times
•
If the bracelet has reached the End-of-Service phase, the LED will be a
steady red.
LED is steady
RED
•
If the bracelet is not detected, the LED will be a steady red and display
“- - -”.
LED is steady
RED
If the Universal Tester does not detect the bracelet:
•
Move the tester closer to the bracelet.
•
Hold the tester at a different orientation in relation to the bracelet.
•
‘Sweep’ the tester by the bracelet.
NOTE: Test one bracelet at a time. Surrounding bracelets should be at
least 5 feet away.
To test a WanderGuard® bracelet WITHOUT ID capability
(32000 Series):
• Hold the Universal Tester within one foot of the bracelet.
• Press the BRACELET button once. The LED above the button will
light briefly.
Press
button
The display will
animate until a
bracelet
is detected, or for
10
seconds.
•
•
If the bracelet is detected, “000” will be displayed.
If the bracelet is still operational, the LED will flash green 4 times.
LED blinks
GREEN 4 times
•
If the bracelet has reached the End-of-Service phase, “000” will be
displayed and the LED will be a steady red.
LED is steady
RED
•
If the bracelet is not detected, the LED will be a steady red and
display “- - -”.
LED is steady
RED
If the Universal Tester does not detect the bracelet:
• Move the tester closer to the bracelet.
• Hold the tester at a different orientation in relation to the bracelet.
• ‘Sweep’ the tester by the bracelet.
NOTE: Test one bracelet at a time. Surrounding bracelets should be at
least 5 feet away. Other electronic devices, such as computers and
monitors, should also be at least 5 feet away when testing.
Testing Door Monitors
Using the Door Test Function
The Universal Tester will test all WanderGuard® door monitors except
the FirstQTM series. Test door monitors at least weekly on each shift with
all surrounding power devices turned on to maximize possible
electromagnetic interference.
Press
button
The display
will animate
while the
button is
pressed.
The Amber LED
above the button will light.
Repeat the button press or hold the button down for continous
transmission.
To test a WanderGuard® door monitor for bracelet detection:
1. Open the door(s).
2. Move to X' from the antenna of the door, and hold the tester at
waist level.
3. Press and hold the DOOR button. “Door Test” will be displayed on
the System Status Display.
4. The antenna should respond with a double “chirp” noise every
second.
If the Universal Tester is not detected, insure the door test LED is on
and the Universal Tester is X inches away from the antenna. To
verify your Universal Tester is working properly, try testing another
door. If your Universal Tester does not work at several different
doors, contact Senior Technologies’ Technical Service at 800-8242996.
5. Repeat steps 1 through 4 for each antenna if there is more than one
located at the doorway.
The Universal Tester will automatically stop after button is
depressed for 30 seconds.
To test a WanderGuard® door alarm or motion detection alarm:
1. Open the door(s) or activate the motion sensor with the universal
tester off.
2. Move to X inches in front of the middle of the door, or middle of
doorway if testing a double door system, and hold the tester at
waist level.
3. Press the DOOR button. The alarm on the WanderGuard® door
monitor will sound as long as the door button is held. Reset the
alarm.
If the alarm does not sound, make sure the door is open and the
tester is within X inches of the monitor. To verify your universal
tester is working properly, try testing another door. If the alarm still
does not sound at several different doors, contact Senior
Technologies’ Technical Service at 800-824-2996.
The Universal Tester will automatically stop after button is
depressed for 30 seconds.
To test a WanderGuard® door coverage:
1. While holding the tester in a vertical position (screen facing you),
place the tester in one of the positions illustrated below - you may
have to kneel to keep the tester in the vertical position.
2. Press the DOOR button.
3. The antenna should chirp and “Door Test” will be displayed on the
System Status Display (if installed).
If the alarm does not sound, make sure the door is open and the tester is
within X inches of the monitor. To verify your universal tester is
working properly, try testing another door. If the alarm still does
not sound at several different doors, contact Senior Technologies’
Technical Service at 800-824-2996.
Single Door Coverage Test
Test Point
Antenna
Double Door Coverage Test
Test Point
Antenna
Antenna
For assistance, please call Stanley-Senior Technologies at
800-824-2996 or 402-475-4002
Mailing address: Stanley-Senior Technologies • P. O. Box 80238
Lincoln, NE 68501-0238
Shipping address: Stanley-Senior Technologies• 1620 N. 20th Circle
Lincoln, NE 68503-1103
10

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : No
Page Count                      : 12
XMP Toolkit                     : XMP toolkit 2.9.1-13, framework 1.6
About                           : uuid:a3e69533-b1b0-4a43-adc5-fefc166f3672
Producer                        : Acrobat Distiller 6.0.1 (Windows)
Creator Tool                    : PScript5.dll Version 5.2.2
Modify Date                     : 2006:07:31 09:00:51-05:00
Create Date                     : 2006:07:31 09:00:51-05:00
Document ID                     : uuid:e09f784a-d345-4b9d-9cd4-bdc90b1a3e83
Format                          : application/pdf
Title                           : (0163-390) Universal Tester-D.qxd
Creator                         : skw0621
Author                          : skw0621
EXIF Metadata provided by EXIF.tools
FCC ID Filing: LA5IDUT002

Navigation menu