Toshiba Dfp 2000A Users Manual DICOM CONFORMANCE STATEMENT

728c1a6f-d17c-4c4e-9dd8-18121880d6f3 Toshiba Portable Media Storage XIDF-037A User Guide |

2014-12-13

: Toshiba Toshiba-Dfp-2000A-Users-Manual-131044 toshiba-dfp-2000a-users-manual-131044 toshiba pdf

Open the PDF directly: View PDF PDF.
Page Count: 23

DownloadToshiba Toshiba-Dfp-2000A-Users-Manual- DICOM CONFORMANCE STATEMENT  Toshiba-dfp-2000a-users-manual
Open PDF In BrowserView PDF
No. MIIXR0002EAC

DICOM CONFORMANCE STATEMENT
MEDIA STORAGE
FOR
TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM
MODEL DFP-2000A
with XIDF-037A and XIDF-038A
or
SRS-1000A
(MIIXR0002EAC)

2001

No. MIIXR0002EAC

IMPORTANT!
(1) No part of this manual may be copied or reprinted, in whole or in part,
without written permission.
(2) The contents of this manual are subject to change without prior notice
and without our legal obligation.

C-1

*

No. MIIXR0002EAC

Table of Contents

Page
1

INTRODUCTION................................................................................................................ 1

1.1
1.2
1.3
2

REFERENCES ..................................................................................................... 1
DEFINITIONS ....................................................................................................... 1
ACRONYMS, ABBREVIATIONS AND SYMBOLS ....................................... 2

IMPLEMENTATION MODEL ............................................................................................. 3
2.1
APPLICATION DATA FLOW DIAGRAM ............................................................................... 3
2.2
FUNCTIONAL DEFINITIONS OF AE'S ................................................................................. 3
2.2.1
DICOM Recording ............................................................................................... 3
2.2.2
DICOM Reader .................................................................................................... 4
2.3
SEQUENCING OF REAL WORLD ACTIVITIES ..................................................................... 4
2.3.1
Features............................................................................................................... 4
2.3.1.1
2.3.1.2

2.3.2
2.3.2.1
2.3.2.2

2.4
3

Transfer of X-ray Examination....................................................................................... 4
Review of X-ray Examination ........................................................................................ 4

Operation ............................................................................................................. 5
Transfer of X-ray Examination....................................................................................... 5
Review of X-ray Examination ........................................................................................ 5

FILE META INFORMATION FOR IMPLEMENTATION CLASS AND VERSION ............................. 5

AE SPECIFICATIONS ....................................................................................................... 6
3.1
DICOM RECORDING SPECIFICATION .............................................................................. 6
3.1.1
File Meta Information for the Application Entity ................................................... 6
3.1.2
Real-World Activity .............................................................................................. 6
3.1.2.1

Real-World Activity - Transfer of X-ray Examination ..................................................... 6

3.1.3
Media Storage Application Profile........................................................................ 6
3.1.4
Information Object Definition and DICOMDIR keys ............................................ 6
3.2
DICOM READER SPECIFICATION ................................................................................... 6
3.2.1
File Meta Information for the Application Entity ................................................... 7
3.2.2
Real-World Activity .............................................................................................. 7
3.2.2.1

3.2.3
3.2.4
4

Real-World Activity - Review of X-ray Examination ....................................................... 7

Media Storage Application Profile........................................................................ 7
Information Object Definition and DICOMDIR keys ............................................ 7

AUGMENTED AND PRIVATE APPLICATION PROFILES .............................................. 7
4.1
AUGMENTED APPLICATION PROFILES ............................................................................. 7
4.1.1
AUG-XABC-CD.................................................................................................... 7
4.1.1.1
4.1.1.2
4.1.1.3

4.2

SOP Class Augmentations............................................................................................ 7
Directory Augmentations ............................................................................................... 7
Other Augmentations .................................................................................................... 7

PRIVATE APPLICATION PROFILES.................................................................................... 7

5 EXTENSIONS, SPECIALIZATIONS, AND PRIVATIZATIONS OF SOP CLASSES AND
TRANSFER SYNTAXES........................................................................................................... 8
6

CONFIGURATION ............................................................................................................. 8

7

SUPPORT OF EXTENDED CHARACTER SETS............................................................. 8

8

X-RAY ANGIOGRAPHIC INFORMATION OBJECT DEFINITION................................... 9
8.1

ENTITY MODULE DEFINITIONS ........................................................................................ 9

-a-

No. MIIXR0002EAC
8.1.1
Media Storage Directory IOD Modules................................................................ 9
8.1.2
Media Storage XA IOD Modules.......................................................................... 9
8.1.3
XA IOD Modules ................................................................................................ 10
8.2
DICOM FILE META INFORMATION ................................................................................ 11
8.2.1
DICOM File Meta Information of Directory IOD ................................................. 11
8.2.2
DICOM File Meta Information of XA IOD........................................................... 11
8.3
BASIC DIRECTORY INFORMATION OBJECT DEFINITIONS ................................................. 11
8.3.1
File-set Identification Module ............................................................................. 11
8.3.2
Directory Information Module............................................................................. 12
8.4
DEFINITION OF SPECIFIC DIRECTORY RECORDS ............................................................ 12
8.4.1
PATIENT keys ................................................................................................... 12
8.4.2
STUDY keys ...................................................................................................... 13
8.4.3
SERIES keys ..................................................................................................... 13
8.4.4
IMAGE keys ....................................................................................................... 14
8.5
XA INFORMATION OBJECT DEFINISIONS........................................................................ 15
8.5.1
Patient Module ................................................................................................... 15
8.5.2
General Study Module ....................................................................................... 15
8.5.3
Patient Study Module......................................................................................... 15
8.5.4
General Series Module ...................................................................................... 16
8.5.5
General Equipment Module ............................................................................... 16
8.5.6
General Image Module ...................................................................................... 16
8.5.7
Image Pixel Module ........................................................................................... 17
8.5.8
Contrast/Bolus Module ...................................................................................... 17
8.5.9
Cine Module....................................................................................................... 17
8.5.10 Multi-frame Module ............................................................................................ 17
8.5.11 X-ray Image Module .......................................................................................... 18
8.5.12 X-ray Acquisition Module ................................................................................... 18
8.5.13 XA Positioner Module ........................................................................................ 19
8.5.14 Curve Module..................................................................................................... 19
8.5.15 VOI LUT Module ................................................................................................ 19
8.5.16 SOP Common Module ....................................................................................... 19

-b-

*

No. MIIXR0002EAC

1

Introduction

This document is a DICOM Conformance Statement for Toshiba's Digital Fluorography
Systems. It is intended to provide the reader with the knowledge of how to integrate this product
within DICOM compliant equipments. It details the DICOM Service Classes, Information
Objects, and Application Profiles which are supported by this product.
If the reader is unfamiliar with DICOM, it is recommended that they read the DICOM
Specification (referenced below) prior to reading this conformance statement. Also note that
this document is formatted according to the DICOM Specification, Part 2: Conformance.

1.1
•

References

ACR-NEMA Digital Imaging and Communications in Medicine, DICOM V3.0.

1.2

Definitions

•

Information Object Definition (IOD) - An IOD is a data model which is an abstraction of
real-world information. This data model defines the nature and attributes relevant to the
class of real-world objects represented.

•

Service/Object Pair (SOP) Class - A SOP Class is defined by the union of an Information
Object Definition and a set of DIMSE Services. A DICOM Application Entity may support
one or more SOP Classes. Each SOP Class is uniquely identified by a SOP Class UID.

•

SOP Instance - A specific occurrence of a Information Object.

•

Transfer Syntax - The Transfer Syntax is a set of encoding rules that allow DICOM
Application Entities to negotiate the encoding techniques (e.g. data element structure, byte
ordering, compression) they are able to support.

•

Unique Identifier (UID) - A Unique Identifier is a globally unique, ISO compliant,
ASCII-numeric string. It guarantees uniqueness across multiple countries, sites, vendors
and equipment.

-1-

No. MIIXR0002EAC

1.3

Acronyms, Abbreviations and Symbols

•

ACC

American College of Cardiology

•

ACR

American College of Radiology

•

ASCII

American Standard Code for Information Interchange

•

AE

Application Entity

•

ANSI

American National Standards Institute

•

AP

Application Profile

•

CEN TC251 Comite Europeen de Normalisation - Technical Committee 251 - Medical
Informatics

•

DICOM

Digital Imaging and Communications in Medicine

•

FSC

File-set Creator

•

FSR

File-set Reader

•

FSU

File-set Updater

•

IOD

Information Object Definition

•

ISO

International Standards Organization

•

JIRA

Japan Industries Association of Radiological Systems

•

NEMA

National Electrical Manufacturers Association

•

RWA

Real-World Activity

•

SOP

Service-Object Pair

•

UID

Unique Identifier

-2-

No. MIIXR0002EAC

2

Implementation Model

2.1

Application Data Flow Diagram

Transfer of
X-ray
Examination

DICOM
Recording

CD-R
Medium
Review of
X-ray
Examination

DICOM
Reader
Legend
Application Entity
Local Real-World Activity
File-set Creator (FSC)
File-set Reader (FSR)
Local Relationship
Figure 1

2.2

Functional Definitions of AE's

2.2.1

DICOM Recording

The DICOM Recording supports the following tasks:
•

Builds DICOM XA Information Objects.

•

Creates a DICOMDIR file that represents the contents of DICOM XA Information
Objects to be recorded.

•

Records DICOM XA Information Objects and DICOMDIR from the local storage to
the CD-R media.

•

Duplicates the File-set of the CD-R media or the local storage onto the blank CD-R
media.

-3-

No. MIIXR0002EAC

2.2.2

DICOM Reader

The DICOM Reader supports the following tasks:

2.3

•

Reads the DICOMDIR file that represents the contents of the data as recorded.

•

Displays the ordered list of patient and image identifying information or icon
images.

•

Reads the selected image SOP instance from the CD-R media and displays it on
the monitor of the DICOM Reader.

•

Reads the File-set of the CD-R media and writes it onto the local storage of DICOM
Reader.

Sequencing of Real World Activities

2.3.1
2.3.1.1

Features
Transfer of X-ray Examination

Operator requests to transfer images to the local storage after selecting the transferred
images from the Patient List or from the Image List.
Operator requests to record the images of the local storage onto the CD-R media after
transferring all data that are to be recorded.
When the size of images exceeds the capacity of a single CD-R, the request of
transferring images is canceled.
Each CD-R media has a DICOMDIR file that represents the contents of the data as
recorded.
Operator can request to duplicate the CD-R media from the recorded CD-R media to the
blank CD-R media

2.3.1.2

Review of X-ray Examination

Operator inserts the CD-R media to be read.
Operator requests to display the list of images or the catalog of icon images that are
recorded on the CD-R media.
Operator selects one image from the list or from the thumbnail images and requests to
review the image from the CD-R media.
Operator can request to read images from the CD-R media onto the local storage and then
operator can request to review the image from the local storage.
Operator can request to delete the images on the local storage.
The images on the local storage remain after being recorded onto the CD-R media, so
operators can display these images from the local storage also.

-4-

No. MIIXR0002EAC

2.3.2

Operation

2.3.2.1

Transfer of X-ray Examination

The operation for manual transfer of X-ray examination is described below:
STEP-1: Select the images or the patient to be transferred.
STEP-2: Request transfer.
STEP-3: Insert the blank CD-R media.
STEP-4: Request recording the transferred images onto the CD-R media.

2.3.2.2

Review of X-ray Examination

The operation for manual review of X-ray examination is described below:
STEP-1: Insert the CD-R media to be review.
STEP-2: Select one image from the list or from the thumbnail images.
STEP-3: Request forward play, backward play, step by step, or other image processings.

2.4

File Meta Information for Implementation Class and Version
•

File Meta Information Version

0x0001

•

Implementation Class UID

1.2.392.200036.9116.31

•

Implementation Version Name

TM_XA_DCM_V12

-5-

No. MIIXR0002EAC

3

AE Specifications

3.1

DICOM Recording Specification

The DICOM Recording AE provides Standard Conformance to the DICOM Media Storage
Service, File Format Class and the Media Storage Application Profile.
Table 1
Application Profile

Identifier

Real World Activity

Role

SC Option

Basic Cardiac X-ray
Angiographic Studies on
CD-R Media

STD-XABC-CD

Transfer of X-ray
Examination

FSC

Interchange

1024 X-ray Angiographic
Studies on CD-R Media

STD-XA1K-CD

Transfer of X-ray
Examination

FSC

Interchange

3.1.1

File Meta Information for the Application Entity

The Application Entity Title is registered into the DICOM File Meta Information header.
•

3.1.2

Source Application Entity Title

TM_XA_DCM_V12

Real-World Activity

3.1.2.1

Real-World Activity - Transfer of X-ray Examination

DICOM Recording AE acts as a FSC when chosen by the operator. This AE writes the SOP
instances onto the CD-R media, creates a DICOMDIR and writes it onto the CD-R media.

3.1.3

Media Storage Application Profile

DICOM Recording AE uses the Application Profile defined in Table 1.

3.1.4

Information Object Definition and DICOMDIR keys

Information Object Definition and DICOMDIR keys are described in Chapter 8.

3.2

DICOM Reader Specification

The DICOM Reader AE provides Conformance to the DICOM Media Storage Service,
File Format Class and the Media Storage Application Profile.
Table 2
Application Profile

Identifier

Real World Activity

Role

SC Option

Basic Cardiac X-ray
Angiographic Studies on
CD-R Media

STD-XABC-CD

Review of X-ray
Examination

FSR

Interchange

Extended Cardiac X-ray
Angiographic Studies on
CD-R Media

AUG-XABC-CD

Review of X-ray
Examination

FSR

Interchange

-6-

No. MIIXR0002EAC
3.2.1

File Meta Information for the Application Entity

The Application Entity Title is not relevant because this DICOM Reader AE only reads CD-R
media.

3.2.2

Real-World Activity

3.2.2.1

Real-World Activity - Review of X-ray Examination

DICOM Reader AE acts as a FSR when chosen by the operator. This AE reads the DICOMDIR
from the CD-R media and reads the SOP instances from the CD-R media.

3.2.3

Media Storage Application Profile

DICOM Reader AE uses the Application Profile defined in Table 2.

3.2.4

Information Object Definition and DICOMDIR keys

Information Object Definition and DICOMDIR keys are described in Chapter 8.

4

Augmented and Private Application Profiles

4.1

Augmented Application Profiles

DICOM Reader AE supports Augmented Application Profile: AUG-XABC-CD.
This is a extension of the standard Application Profile: STD-XABC-CD.

4.1.1

AUG-XABC-CD

4.1.1.1

SOP Class Augmentations

None

4.1.1.2

Directory Augmentations

None

4.1.1.3

Other Augmentations

The following attributes used within the X-ray Angiographic Image files have the specified
values.
Table 3
Attribute Name

4.2

Tag

Value

Modality

(0008, 0060)

XA

Rows

(0028, 0010)

1024

Columns

(0028, 0011)

1024

Bits Allocated

(0028, 0100)

8

Bits Stored

(0028, 0101)

8

Private Application Profiles

Not applicable to this product.

-7-

No. MIIXR0002EAC

5 Extensions, Specializations, and Privatizations of SOP
Classes and Transfer Syntaxes
Not applicable to this product.

6

Configuration

For the Digital Fluorography System, the configuration can be set.
Note: Settings and changes are performed by Toshiba Service Personnel at the time of
installation of the Digital Fluorography System.

7

Support of Extended Character Sets

This product supports the following character sets:
Table 4
•

ISO-IR 6 (default)

Basic G0 Set

-8-

No. MIIXR0002EAC

8

X-Ray Angiographic Information Object Definition

8.1

Entity Module Definitions

The information modules for the X-Ray Angiographic devices are defined below.

8.1.1

Media Storage Directory IOD Modules
Table 5
Object

Module / key

DICOM File Meta
Information

Reference

Usage1

8.2.1

M

Basic Directory
Information Object
Definition

File-set Identification Module

8.3.1

M

Directory Information Module

8.3.2

U

Definition of Specific
Directory Records

PATIENT keys

8.4.1

M

STUDY keys

8.4.2

M

SERIES keys

8.4.3

M

IMAGE keys

8.4.4

M

Reference

Usage1

8.2.2

M

8.1.3

M

1

8.1.2

M=Mandatory, C=Conditional, U=User option

Media Storage XA IOD Modules
Table 6
Object

Module / key

DICOM File Meta
Information
XA Information
Object Definition
1

XA IOD Modules

M=Mandatory, C=Conditional, U=User option

-9-

No. MIIXR0002EAC

8.1.3

XA IOD Modules
Table 7
Information Entity

Module

Reference

Usage1

Patient

Patient Module

8.5.1

M

Study

General Study Module

8.5.2

M

Study

Patient Study Module

8.5.3

U

Series

General Series Module

8.5.4

M

Equipment

General Equipment Module

8.5.5

M

Image

General Image Module

8.5.6

M

Image

Image Pixel Module

8.5.7

M

Image

Contrast/bolus Module

8.5.8

C

Image

Cine Module

8.5.9

C

Image

Multi-frame Module

8.5.10

C

Image

Frame Pointers Module

Not Used

U

Image

Mask Module

Not Used

C

Image

Display Shutter Module

Not Used

U

Image

Device Module

Not Used

U

Image

Therapy Module

Not Used

U

Image

X-ray Image Module

8.5.11

M

Image

X-ray Acquisition Module

8.5.12

M

Image

X-ray Collimator Module

Not Used

U

Image

X-ray Table Module

Not Used

C

Image

XA Positioner Module

8.5.13

M

Image

Overlay Plane Module

Not Used

U

Image

Multi-Frame Overlay Module

Not Used

C

Image

Curve Module

8.5.14

C (Note)

Image

Modality LUT Module

Not Used

C/U

Image

VOI LUT Module

8.5.15

U

Image

SOP Common Module

8.5.16

M

1

M=Mandatory, C=Conditional, U=User option
Note : Always set when ECG data is available

- 10 -

No. MIIXR0002EAC

8.2

DICOM File Meta Information

8.2.1

DICOM File Meta Information of Directory IOD
Table 8
Attribute Name

Tag

Type

File Preamble

No Tag

1

Always set

DICOM Prefix

No Tag

1

Always set ("DICM")

Group Length

(0002, 0000)

1

Always set

File Meta Information Version

(0002, 0001)

1

Always set (0x0001)

Media Storage SOP Class UID

(0002, 0002)

1

Always set ("1.2.840.10008.1.3.10")

Media Storage SOP Instance UID

(0002, 0003)

1

Always set

Transfer Syntax UID

(0002, 0010)

1

Always set ("1.2.840.10008.1.2.1")

Implementation Class UID

(0002, 0012)

1

Always set

8.2.2

Attribute Description

DICOM File Meta Information of XA IOD
Table 9
Attribute Name

Tag

Type

File Preamble

No Tag

1

Always set

DICOM Prefix

No Tag

1

Always set ("DICM")

Group Length

(0002, 0000)

1

Always set

File Meta Information Version

(0002, 0001)

1

Always set (0x0001)

Media Storage SOP Class UID

(0002, 0002)

1

Always set
("1.2.840.10008.5.1.4.1.1.12.1")

Media Storage SOP Instance UID

(0002, 0003)

1

Always set

Transfer Syntax UID

(0002, 0010)

1

Always set ("1.2.840.10008.1.2.4.70")

Implementation Class UID

(0002, 0012)

1

Always set

Implementation Version Name

(0002, 0013)

3

Always set

8.3
8.3.1

Attribute Description

Basic Directory Information Object Definitions
File-set Identification Module
Table 10
Attribute Name

File-set ID

Tag

Type

(0004, 1130)

2

- 11 -

Attribute Description
Always set

No. MIIXR0002EAC

8.3.2

Directory Information Module
Table 11
Attribute Name

Tag

Type

Offset of the First Directory Record of
the Root Directory Entity

(0004, 1200)

1

Always set

Offset of the Last Directory Record of
the Root Directory Entity

(0004, 1202)

1

Always set

File-set Consistency Flag

(0004, 1212)

1

Always set

Directory Record Sequence

(0004, 1220)

2

Always set

>Offset of the Next Directory Record

(0004, 1400)

1C

Always set when condition is met

>Record In-use Flag

(0004, 1410)

1C

Always set when condition is met

>Offset of Referenced Lower-Level
Directory Entity

(0004, 1420)

1C

Always set when condition is met

>Directory Record Type

(0004, 1430)

1C

Always set when condition is met

>Referenced File ID

(0004, 1500)

1C

Always set when condition is met

>Referenced SOP Class UID in File

(0004, 1510)

1C

Always set when condition is met

>Referenced SOP Instance UID in File

(0004, 1511)

1C

Always set when condition is met

>Referenced Transfer Syntax UID in
File

(0004, 1512)

1C

Always set when condition is met

8.4

Attribute Description

Definition of Specific Directory Records

8.4.1

PATIENT keys
Table 12
Attribute Name

Tag

Type

Patient's Name

(0010, 0010)

2

Always set except for urgent patient

Patient ID

(0010, 0020)

1

Always set

Patient's Birth Date

(0010, 0030)

2

Length=0 when no entry is made

Patient's Sex

(0010, 0040)

2

Always set

- 12 -

Attribute Description

No. MIIXR0002EAC

8.4.2

STUDY keys
Table 13
Attribute Name

Tag

Type

Study Date

(0008, 0020)

1

Always set

Study Time

(0008, 0030)

1

Always set

Study Description

(0008, 1030)

2

Length=0 when no entry is made

Study Instance UID

(0020, 000D)

1C

Always set

Study ID

(0020, 0010)

1

Always set

Accession Number

(0008, 0050)

2

Length=0 when no entry is made

8.4.3

Attribute Description

SERIES keys
Table 14
Attribute Name

Tag

Type

Modality

(0008, 0060)

1

Always set ("XA")

Series Instance UID

(0020, 000E)

1

Always set

Series Number

(0020, 0011)

1

Always set

Institution Name

(0008, 0080)

2

Always set

Institution Address

(0008, 0081)

2

Always set

Performing Physician's Name

(0008, 1050)

2

Length=0 when no entry is made

- 13 -

Attribute Description

No. MIIXR0002EAC

8.4.4

IMAGE keys
Table 15
Attribute Name

Tag

Type

Instance Number

(0020, 0013)

1

Always set

Image Type

(0008, 0008)

1

Always set

Image Date

(0008, 0023)

3

Always set

Image Time

(0008, 0033)

3

Always set

Positioner Primary Angle

(0018, 1510)

3

Always set [degree]

Positioner Secondary Angle

(0018, 1511)

3

Always set [degree]

Image Comments

(0020, 4000)

3

Length=0 when no entry is made

Number of Frames

(0028, 0008)

3

Always set

Calibration Image

(0050, 0004)

2

Always set

Icon Image Sequence

(0088, 0200)

1

Always set

>Samples per Pixel

(0028, 0002)

1

Always set (0x0001)

>Photometric Interpretation

(0028, 0004)

1

Always set ("MONOCHROME2")

>Rows

(0028, 0010)

1

Always set (0x0100)

>Columns

(0028, 0011)

1

Always set (0x0100)

>Bits Allocated

(0028, 0100)

1

Always set (0x0008)

>Bits Stored

(0028, 0101)

1

Always set (0x0008)

>High Bit

(0028, 0102)

1

Always set (0x0007)

>Pixel Representation

(0028, 0103)

1

Always set (0x0000)

>Pixel Data

(7FE0, 0010)

1

Always set

- 14 -

Attribute Description

No. MIIXR0002EAC

8.5

XA Information Object Definisions

8.5.1

Patient Module
Table 16
Attribute Name

Tag

Type

Patient's Name

(0010, 0010)

2

Always set except for urgent patient

Patient ID

(0010, 0020)

2

Always set

Patient's Birth Date

(0010, 0030)

2

Length=0 when no entry is made

Patient's Sex

(0010, 0040)

2

Always set

Patient Comments

(0010, 4000)

3

Length=0 when no entry is made

8.5.2

Attribute Description

General Study Module
Table 17
Attribute Name

Tag

Type

Study Instance UID

(0020, 000D)

1

Always set

Study Date

(0008, 0020)

2

Always set

Study Time

(0008, 0030)

2

Always set

Referring Physician's Name

(0008, 0090)

2

Length=0 when no entry is made

Study ID

(0020, 0010)

2

Always set

Accession Number

(0008, 0050)

2

Length=0 when no entry is made

Study Description

(0008, 1030)

3

Length=0 when no entry is made

8.5.3

Attribute Description

Patient Study Module
Table 18
Attribute Name

Tag

Type

Patient's Size

(0010, 1020)

3

Length=0 when no entry is made

Patient's Weight

(0010, 1030)

3

Length=0 when no entry is made

- 15 -

Attribute Description

No. MIIXR0002EAC

8.5.4

General Series Module
Table 19
Attribute Name

Tag

Type

Modality

(0008, 0060)

1

Always set ("XA")

Series Instance UID

(0020, 000E)

1

Always set

Series Number

(0020, 0011)

2

Always set

Series Date

(0008, 0021)

3

Always set

Series Time

(0008, 0031)

3

Always set

Performing Physician's Name

(0008, 1050)

3

Length=0 when no entry is made

Series Description

(0008, 103E)

3

Always set

8.5.5

Attribute Description

General Equipment Module
Table 20
Attribute Name

Tag

Type

Manufacturer

(0008, 0070)

2

Always set

Institution Name

(0008, 0080)

3

Always set

Institution Address

(0008, 0081)

3

Always set

Station Name

(0008, 1010)

3

Always set

Manufacturer's Model Name

(0008, 1090)

3

Always set

Device Serial Number

(0018, 1000)

3

Always set

Software Versions

(0018, 1020)

3

Always set

8.5.6

Attribute Description

General Image Module
Table 21

Attribute Name

Tag

Type

Image Number

(0020, 0013)

2

Patient Orientation

(0020, 0020)

2C

Always Length=0

Image Date

(0008, 0023)

2C

Always set

Image Time

(0008, 0033)

2C

Always set

Image Type

(0008, 0008)

3

Always set

Acquisition Date

(0008, 0022)

3

Always set

Acquisition Time

(0008, 0032)

3

Always set

Image Comments

(0020, 4000)

3

Length=0 when no entry is made

- 16 -

Attribute Description
Always set

No. MIIXR0002EAC

8.5.7

Image Pixel Module
Table 22
Attribute Name

Tag

Type

Samples per Pixel

(0028, 0002)

1

Always set (0x0001)

Photometric Interpretation

(0028, 0004)

1

Always set ("MONOCHROME2")

Rows

(0028, 0010)

1

Always set (0x0400 or 0x0200)

Columns

(0028, 0011)

1

Always set (0x0400 or 0x0200)

Bits Allocated

(0028, 0100)

1

Always set (0x0008)

Bits Stored

(0028, 0101)

1

Always set (0x0008)

High Bit

(0028, 0102)

1

Always set (0x0007)

Pixel Representation

(0028, 0103)

1

Always set

Pixel Data

(7FE0, 0010)

1

Always set

8.5.8

Attribute Description

Contrast/Bolus Module
Table 23
Attribute Name

Tag

Type

Contrast/Bolus Agent

(0018, 0010)

2

Length=0 when no entry is made

Contrast/Bolus Route

(0018, 1040)

3

Length=0 when no entry is made

Contrast/Bolus Volume

(0018, 1041)

3

Length=0 when no entry is made

Contrast Flow Rate(s)

(0018, 1046)

3

Length=0 when no entry is made

8.5.9

Attribute Description

Cine Module
Table 24
Attribute Name

Frame Time

8.5.10

Tag

Type

(0018, 1063)

1C

Attribute Description
Always set

Multi-frame Module
Table 25

Attribute Name

Tag

Type

Number of Frames

(0028, 0008)

1

Always set

Frame Increment Pointer

(0028, 0009)

1

Always set (0x00181063)

- 17 -

Attribute Description

No. MIIXR0002EAC

8.5.11

X-ray Image Module
Table 26

Attribute Name

Tag

Type

Frame Increment Pointer

(0028, 0009)

1C

Image Type

(0008, 0008)

1

Always set

Pixel Intensity Relationship

(0028, 1040)

1

Always set ("DISP" or "LIN")

Samples per Pixel

(0028, 0002)

1

Always set (0x0001)

Photometric Interpretation

(0028, 0004)

1

Always set ("MONOCHROME2")

Bits Allocated

(0028, 0100)

1

Always set (0x0008)

Bits Stored

(0028, 0101)

1

Always set (0x0008)

High Bit

(0028, 0102)

1

Always set (0x0007)

Pixel Representation

(0028, 0103)

1

Always set

Reference Image Sequence

(0008, 1140)

1C

Not set when Image type(0008,0008)
Value 3 is "SINGLE PLANE"

>Reference SOP Class UID

(0008, 1150)

1C

Always set when Reference Image
Sequence is present

>Reference SOP Instance UID

(0008, 1155)

1C

Always set when Reference Image
Sequence is present

8.5.12

Attribute Description
Always set (0x00181063)

X-ray Acquisition Module
Table 27

Attribute Name

Tag

Type

KVP

(0018, 0060)

2

Always set [kV]

Radiation Setting

(0018, 1155)

1

Always set

X-ray Tube Current

(0018, 1151)

2C

Always set [mA]

Exposure Time

(0018, 1150)

2C

Always set [msec]

Average Pulse Width

(0018, 1154)

3

Always set [msec]

Intensifier Size

(0018, 1162)

3

Always set [mm]

Field of View Shape

(0018, 1147)

3

Always set("ROUND ")

Field of View Dimension(s)

(0018, 1149)

3

Always set [mm]

Imager Pixel Spacing

(0018, 1164)

3

Always set [mm]

- 18 -

Attribute Description

No. MIIXR0002EAC

8.5.13

XA Positioner Module
Table 28

Attribute Name

Tag

Type

Attribute Description

Distance Source to Detector

(0018, 1110)

3

Always set [mm]

Estimated Radiographic Magnification

(0018,1114)

3

Always set

Positioner Motion

(0018, 1500)

2C

Positioner Primary Angle

(0018, 1510)

2

Always set [degree]

Positioner Secondary Angle

(0018, 1511)

2

Always set [degree]

Factor

8.5.14

Always set ("STATIC")

Curve Module
Table 29

Attribute Name

Tag

Type

Attribute Description

Curve Dimensions

(5000, 0005)

1

Always set

Number of Points

(5000, 0010)

1

Always set

Type of Data

(5000, 0020)

1

Always set

Axis Units

(5000, 0030)

3

Always set

Data Value Representation

(5000, 0103)

1

Always set

Curve Data Descriptor

(5000, 0110)

1C

Always set

Coordinate Start Value

(5000, 0112)

1C

Always set

Coordinate Step Value

(5000, 0114)

1C

Always set

Curve Data

(5000, 3000)

1

Always set

Note : This module is set only if ECG data is available.

8.5.15

VOI LUT Module
Table 30

Attribute Name

Tag

Type

Window Center

(0028, 1050)

3

Always set

Window Width

(0028, 1051)

1C

Always set

8.5.16

Attribute Description

SOP Common Module
Table 31

Attribute Name

Tag

Type

SOP Class UID

(0008, 0016)

1

Always set

SOP Instance UID

(0008, 0018)

1

Always set

- 19 -

Attribute Description

E



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : Yes
Encryption                      : Standard V1.2 (40-bit)
User Access                     : Print, Fill forms, Extract, Assemble, Print high-res
Modify Date                     : 2001:12:20 13:11:46+09:00
Create Date                     : 2001:12:20 12:01:00+09:00
Page Count                      : 23
Creation Date                   : 2001:12:20 12:01:00+09:00
Mod Date                        : 2001:12:20 13:11:46+09:00
Producer                        : Acrobat Distiller 5.0 (Windows)
Author                          : Naoki Yamada
Metadata Date                   : 2001:12:20 13:11:46+09:00
Creator                         : Naoki Yamada
Title                           : DICOM CONFORMANCE STATEMENT
Page Mode                       : UseNone
Tagged PDF                      : Yes
EXIF Metadata provided by EXIF.tools

Navigation menu