Urologix CW5000A RFID Card Transmitter User Manual 532897

Urologix Inc. RFID Card Transmitter 532897

Manual

Download: Urologix CW5000A RFID Card Transmitter User Manual 532897
Mirror Download [FCC.gov]Urologix CW5000A RFID Card Transmitter User Manual 532897
Document ID532897
Application IDmUSDswohlVjfyQ7ioQVSrw==
Document DescriptionManual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatMicrosoft Word - pdf
Filesize123.25kB (1540609 bits)
Date Submitted2005-04-15 00:00:00
Date Available2005-04-13 00:00:00
Creation Date2005-04-15 13:13:15
Producing SoftwareAcrobat Distiller 5.0.5 (Windows)
Document Lastmod2005-04-15 13:13:15
Document TitleMicrosoft Word - 532897.DOC
Document CreatorPScript5.dll Version 5.2
Document Author: Administrator

Section 1
System Description
1.1 Notice
The information contained in this manual is subject to change. This manual does not necessarily
address all safety concerns associated with the Urologix® Targis® System.
You must establish appropriate safety and health practices and determine the applicability of
regulatory limitations prior to use.
The Targis System is intended for use only by qualified medical personnel. Federal law in the
United States restricts this device for sale, distribution, use by or on the order of physicians.
Federal (USA) law restricts this device to sale by or on the order of a physician trained and/or
experienced in the use of this device as outlined in the required training program.
Medical equipment, however sophisticated, should never be a substitute for the human care,
attention, and critical judgment that only trained healthcare professionals can provide.
1.2 Safety symbols and definitions
The following safety symbols are used throughout this manual. Familiarize yourself with each
symbol and its meaning before using this equipment. You can find additional symbols
associated with the Targis System in Section 5.3, Description of symbols.
Safety Symbol
Note
Caution
Warning
Instruction Manual
Part number 250023-001 Rev 02
Definition
A note indicates important information that
helps you operate the Targis Control Unit or
use the disposable devices.
A caution contains instructions that must be
followed to avoid the Control Unit or
disposable devices from malfunctioning or
from damage. Do not proceed beyond a
caution sign until the indicated conditions are
fully understood and met.
A warning contains important information
about possible danger to you or the patient.
Do not proceed beyond a warning sign until
the indicated conditions are fully understood
and met.
The instruction manual symbol is displayed
on the product when it is necessary for you to
refer to the Targis System User Manual.
1 of 19
System Description
Section 1
1.3 User manual overview
This manual combines technical reference material as well as information on how to use the
Targis System.
Notes:
o For information regarding the contents of this manual, please call Urologix
Customer Service at 1-888-229-0772.
o Read this manual before operating equipment, and keep the manual in the holder on
the Control Unit for reference.
Section 1: System Description provides an overview of the Targis System equipment. This
section also provides important notes about using the Targis System including information on
Control Unit installation, use environment, equipment connections, equipment testing, and
safety instructions.
Section 2: Treatment Session Setup describes how to prepare the patient, prepare the Control
Unit, install the Targis Coolant Bag, and insert the MDS (treatment catheter) and the RTU
(rectal unit).
Section 3: Treatment Instructions provides instruction on how to use the Control Unit from
logging into the Control Unit to beginning and ending a Targis treatment. You will also find
information on how to change system settings, handle system errors, and work in demonstration
mode.
Section 4: Equipment Maintenance presents information on post-treatment cleaning
procedures and storage instructions. You will also find information on how to move and ship the
Control Unit. Finally, this section discusses how to maintain the equipment, though some
maintenance requires a Urologix trained service representative.
Section 5: Appendix includes a troubleshooting guide, a flowchart of the treatment screens, a
description of the symbols used in the manual and on the labels, an overview of the Targis
Patient Comfort Kit, and a glossary of terms used in Targis System literature.
1.4 Precautions
Only those physicians who have been thoroughly trained on the operation of the Targis System
and the Targis procedure should deliver the Targis procedure.
The Targis procedure must not be initiated without assurance that the treatment catheter is
properly positioned in the patient. The correct positioning of the catheter must always be
checked by ultrasound imaging prior to commencing treatment. Improper placement or
orientation of the treatment catheter may lead to procedure failures or heating damage of
nontarget tissues such as the bladder neck, external sphincter, or penile urethra.
2 of 19
Part number 250023-001 Rev 02
Section 1
System Description
All components of the Targis System must be used in a manner consistent with the instructions
set forth in their respective instructions for use insert and the Targis System User Manual.
Failure to do so may result in insufficient treatment or increased risk of injury or infection to the
patient.
Use of the Targis System results in the deposition of microwave energy within the patient's
prostate and in adjacent regions of the body. Some animal studies in the literature suggest that
there may be as yet unknown health effects from exposure to microwave radiation, including an
increased incidence of tumors. Although it is not possible to extrapolate these studies to humans,
they suggest that unnecessary microwave radiation exposure should be avoided.
At least 20 cm of ventilation clearance must be provided around the base of the Control Unit.
Note: This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio frequency energy and, if not
installed and used in accordance with the instruction manual, may cause harmful interference to
radio communications. Operation of this equipment in a residential area is likely to cause
harmful interference in which case the user will be required to correct the interference at his
own expense.
The Urologix Targis System emits a small amount of electromagnetic energy during a
procedure. Urologix recommends that all electronic medical devices be kept at a minimum
distance of 1.0 meter from the Targis System when performing a treatment. However, a 1-meter
separation of electronic medical equipment from the Targis System does not guarantee that
operation of other devices will not be impacted. The effect of this electromagnetic energy on all
equipment can not be predicted due to age and quality of maintenance. The performance of each
piece of equipment operated near the Targis System, during a procedure, must be evaluated for
degradation.
Since microwave energy can travel through walls, ceilings, and floors to affect other devices, it
is important to understand that the 1-meter safety distance applies not only to the treatment
room, but also to all adjacent rooms in the building, including the rooms above and below the
treatment room.
Do not operate the Targis System near equipment that emits electromagnetic energy, unless the
effect on the Targis System has been evaluated and no degradation of performance was found.
The national standard ANSI/IEEE C95.1 - 1999 Edition (Safety Levels with Respect to Human
Exposure to Radio Frequency Electromagnetic Fields) recommends a maximum stray field
exposure level for whole body exposure of 3 mW/cm2, as averaged for any 6 minute period. The
maximum radiated field, at full power, from the Targis Control Unit patient cable and treatment
catheter, at 5 centimers, is 2.1 mW/cm2. Urologix recommends that the operator maintain a
minimum distance of 5 centimeters from the patient cable and exposed portions of the catheter
Part number 250023-001 Rev 02
3 of 19
System Description
Section 1
during the procedure.
Operate the Control Unit and connected devices only in clinical environments where the
electrical installation is in accordance with international standard, DIN VDE 0107, and the
national standard, ANSI/NFPA 70. The equipment must be connected to a fully tested, hospital
grade power outlet with adequate grounding.
The Control Unit must be plugged into the appropriate voltage outlet.
The electrical equipment inside the Targis System uses voltages capable of causing serious
injury or death from electric shock. To avoid this hazard, never open the housing of the Control
Unit.
1.5 Introduction to the Targis System
The Targis System treats Benign Prostatic Hyperplasia (BPH) by applying microwave power to
the prostate. This microwave power, when applied to the prostate, heats the diseased tissue via a
treatment catheter. This catheter also minimizes patient discomfort and risk to the urethra by
circulating cooling fluid. In addition, the treatment catheter protects the rectal wall from
damage: The system targets microwave power at the prostate while continuously monitoring
rectal wall temperature readings throughout treatment. However, if during a treatment, urethra
or rectal temperatures exceed protocol parameters, the system will adjust microwave power to
protect the urethra or rectal wall from overheating.
1.6 Targis System equipment overview
The Targis System includes the Targis CoolWave™ Control Unit, Model 5000A, the Targis
Procedure Kit (comprised of a treatment catheter, a rectal unit, and a Coolant Bag), and
accessories. For this user manual, the representative treatment catheter is the Targis Cooled
ThermoCath® treatment catheter.
1.6.1 Control Unit
The Control Unit (Figure 1) supplies microwave energy and coolant to the treatment catheter
and collects temperature data from the catheter and the rectal unit. The Control Unit also
provides a way of entering patient data, controlling treatment parameters (eg, ramp rate, coolant
temperature, and treatment time), and monitoring rectal and urethra temperatures.
4 of 19
Part number 250023-001 Rev 02
Section 1
System Description
Figure 1. Control Unit
Part number 250023-001 Rev 02
5 of 19
System Description
Section 1
The Control Unit includes these features:
•
•
•
•
•
•
•
•
•
•
Patient connection cable and patient connection cable housing
Touchscreen monitor
Keyboard
Printer
ON/OFF power switch
Lockable wheels
Coolant system
Main power indicator and microwave power indicator
Emergency Stop button
Other: Volume control, headphone jack, USB port, and service port
Patient connection cable and patient connection cable housing
Data from the treatment catheter and the rectal unit enters the Control Unit via the patient
connection cable and patient connection cable housing (Figure 2).
Figure 2. Patient connection cable housing
The patient connection cable housing contains connectors for the rectal unit, fiber optic
connector, and microwave connector.
6 of 19
Part number 250023-001 Rev 02
Section 1
System Description
Touchscreen monitor
View a treatment using the touchscreen monitor (Figure 3) and, when necessary, adjust
treatment parameters. The monitor can be tilted for improved viewing. To open the monitor,
operate the latch and lift.
Figure 3. Touchscreen monitor
Keyboard
Enter patient data using the keyboard (Figure 4), located under the touchscreen monitor. The
keys are sealed to prevent damage from spillage onto the keyboard. To access the keyboard, lift
the touchscreen monitor into an upright position.
Figure 4. Keyboard
Part number 250023-001 Rev 02
7 of 19
System Description
Section 1
Printer
Caution: The Control Unit can tip over if you press down on the open printer drawer with
too much weight. Do not press down on the open drawer with more than 20 kg (44 lbs).
Print data from any treatment using the Canon® color bubble jet printer (Figure 5). The printer
drawer pulls out and holds the printer.
Figure 5. Printer
ON/OFF power switch
Instruction Manual: Read the Targis System User Manual before turning ON and
operating the system.
Turn the Control Unit ON or OFF with this power switch (Figure 6).
Figure 6. ON/OFF power switch
8 of 19
Part number 250023-001 Rev 02
Section 1
System Description
Lockable wheels
Keep the Control Unit from moving by locking the front wheels (Figure 7).
Figure 7. Wheel lock tabs
To lock the wheels, use your foot to press down on the wheel lock tabs. To unlock the wheels,
press on the back part of the wheel lock tabs.
Coolant system
The coolant system consists of a chill plate, temperature and pressure sensors, a peristaltic pump
mechanism, and a coolant bag (Figure 8).
Figure 8. Coolant system features
Part number 250023-001 Rev 02
9 of 19
System Description
Section 1
Chill plate: The chill plate, located behind the coolant door (Figure 8), is equipped with
mounting pins to hold the coolant bag securely against the chill plate surface.
Temperature and pressure sensors: These sensors, located on the Control Unit, monitor the
coolant temperature and the coolant pressure (Figure 8).
Peristaltic pump mechanism: The peristaltic pump mechanism (Figure 8) circulates the
coolant.
Coolant bag: The coolant bag serves as the reservoir for the coolant.
Main power indicator and microwave power indicator
The main power indicator (Figure 9), located in front of the keyboard, is a green LED that
illuminates when the Control Unit is ON. The microwave power indicator (Figure 9), also
located in front of the keyboard, is an amber LED that illuminates when the Control Unit
delivers microwave power to the prostate.
Figure 9. Power indicators
10 of 19
Part number 250023-001 Rev 02
Section 1
System Description
Emergency Stop button
If there is an emergency, press the red Emergency Stop button (Figure 10), located to the left of
the amber LED, to immediately turn the microwave power OFF.
Figure 10. Emergency Stop button
Other: Volume control, headphone jack, USB port, and service port
The Control Unit includes a volume control, a headphone jack, a USB (Universal Serial Bus)
port, and a service port (Figure 11). The USB port can accommodate a USB flash drive.
Note: The service port is reserved for use by Urologix service personnel only.
Figure 11. Volume control, headphone jack, USB port, and service port
Part number 250023-001 Rev 02
11 of 19
System Description
Section 1
1.6.2 Treatment catheter
The single-use treatment catheter includes a fiber optic temperature sensor to measure urethral
temperature, a microwave antenna and cable, coolant channels and connectors, a urine drainage
port that connects to a standard urine drainage bag, and a location balloon to position the
catheter at the bladder neck (Figure 12). The microwave antenna and temperature sensor are
connected to the Control Unit via the patient connection cable housing and patient connection
cable. To ensure that the treatment catheter is positioned properly within the urethra, the
location balloon is inflated to hold the catheter in place during treatment.
Figure 12. Treatment catheter
The treatment catheter is used for:
•
•
•
•
Delivering microwave energy to the targeted prostatic tissue.
Monitoring the urethral temperature.
Cooling the urethra during treatment.
Draining urine during treatment.
1.6.3 Rectal unit
The Targis System requires the use of either the single-use standard Targis RTU or the Targis
RTU Plus® with reusable handle and single-use balloon. The standard RTU and RTU Plus
(Figure 13 and Figure 14) both consist of an inflatable balloon with 5 temperature sensors.
These sensors monitor rectal temperature along the anterior rectal wall and sends this
information to the Control Unit during a treatment. After inserting one of the rectal units into the
rectum, inflating the rectal balloon with air holds the thermosensors in place against the anterior
rectal wall nearest the prostate.
12 of 19
Part number 250023-001 Rev 02
Section 1
System Description
Figure 13. Single-use standard RTU
Figure 14. Assembled RTU Plus (with the reusable handle and single use balloon)
1.6.4 Coolant Bag
The single-use Coolant Bag includes a coolant bag, inlet and outlet tubing, connectors, and a
sensor module (Figure 15). The sensor module allows the Control Unit to monitor coolant
temperature and pressure in order to maintain coolant temperatures within acceptable limits and
to ensure that the coolant circulates properly. The coolant bag also includes 2 small holes on the
top and 1 hole on the bottom for mounting the bag on the chill plate and a hydrophobic vent to
release air (but not coolant) from inside the bag.
Part number 250023-001 Rev 02
13 of 19
System Description
Section 1
Figure 15. Coolant Bag
The function of the coolant bag is to provide a reservoir for the coolant that circulates through
the treatment catheter during a treatment. The coolant circulates through the coolant bag via the
peristaltic pump mechanism: The coolant inlet tubing runs across the pump mechanism, which
pushes the coolant through the inlet tubing, the sensor module, the coolant bag, and the outlet
tubing. Coolant continuously circulates through the treatment catheter, connected to the coolant
outlet tubing, and returns to the coolant bag via the coolant inlet tubing. The coolant bag resides
against the chill plate, which chills the circulating coolant.
1.7 Required equipment
The following equipment, including an ultrasound system, is needed to successfully treat
patients with the Targis System.
1.7.1 Equipment provided by Urologix
Urologix provides the following equipment.
Quantity
Equipment/Material
Control Unit
Targis System User Manual
Targis Procedure Kit containing:
• 1 treatment catheter
• 1 rectal unit (standard RTU or RTU Plus)
• 1 Coolant Bag
Patient cable holder
Patient Comfort Kit (2 knee cushions and an MDS
Holder)
Targis Transport Kit, optional (trolley and electrical
safety tester)
1.7.2 Equipment provided by the clinic
The clinic typically provides the following equipment.
14 of 19
Part number 250023-001 Rev 02
Section 1
Quantity
As needed
As needed
50 cc
As needed
200 cc
As needed
As needed
As needed
System Description
Equipment/Material
Foley catheter, 16-18 French
Straight catheter, 14-16 French
Urine drainage bag
Sterile gloves
Anesthetic lubricating jelly (Urojet or lidocaine jelly)
Local bladder anesthetic of choice (eg, 50 cc of 1% or 2%
lidocaine without epinephrine)
Water soluble lubricating gel (eg, K-Y® Jelly)
60 cc luer-lok syringe
60 cc catheter-tip syringe (ie, Toomey™ syringe)
10 cc luer-lok syringe
Sterile water for coolant bag and catheter balloons
Ultrasound system
Catheter plug
Permanent marker or tape
Nonsterile gloves
Penile clamp
Specimen cup
Urinal or graduate
Ice or ice pack
1.8 Targis System installation and use environment
1.8.1 Installation
Warning: DO NOT USE components that have evidence of a compromised package or
damage.
Before unpacking the Control Unit, inspect the shipping crate for signs of damage. Remove the
Control Unit from the shipping crate, and retain the shipping crate to return the Control Unit for
service, if needed. Then, prior to using the Targis System, visually inspect the following system
components for damage:
• Control Unit for obvious damage
• Pump latch and coolant door to see that they are operating correctly
• Patient connection cable, patient connection cable housing, and connectors for kinks,
cuts, dirt, contamination, or obvious damage
• Treatment catheter for kinks, cuts, or obvious damage
• Rectal unit for kinks, cuts, or obvious damage
• Coolant Bag for kinks, cuts, or obvious damage
Operate the Control Unit and its connected devices only in clinical environments where the
electrical installation is in accordance with the international standard, DIN VDE 0107; and the
national standard, ANSI/NFPA 70-1993. (Refer to section 517 of the National Electric Code®.)
Part number 250023-001 Rev 02
15 of 19
System Description
Section 1
The equipment must be connected to a fully tested, hospital-grade power outlet with adequate
grounding.
Power requirements
Control Unit, Model 5000E (Europe):
Control Unit, Model 5000A (US):
Control Unit, Model 5000J (Japan):
220/240 V [+/- 10%] (8 A) Single phase 50 or 60 Hz
110/120 V [+/- 10%] (15 A) Single phase 50 or 60 Hz
100 V [+/- 10%] (15 A) Single phase 50 or 60 Hz
If required, an equal potential ground cable should be connected to the Control Unit (Figure 16)
and the appropriate ground.
Figure 16. Control Unit rear cover
16 of 19
Part number 250023-001 Rev 02
Section 1
System Description
Control Unit power cords
Urologix provides the following power cords for use with the Control Unit.
End View
Power Cord Catalog Number
Catalog Number: AC1011
France
Austria
Germany
Norway
Belgium
Sweden
Netherlands
Finland
Catalog Number: AC1012
Australia
New Zealand
Catalog Number: AC1013
United Kingdom
Ireland
Catalog Number: AC1014
Denmark
Catalog Number: AC1015
Italy
Catalog Number: AC1016
Japan
Catalog Number: AC1017
Canada
United States
Mexico
Catalog Number: AC1018
Switzerland
Figure 17. Control Unit power cords
Part number 250023-001 Rev 02
17 of 19
System Description
Section 1
Equipment connections
The Control Unit must not be connected to any device other than the treatment catheter, rectal
unit, or Coolant Bag. In addition, the treatment catheter, rectal unit, and Coolant Bag must not
be connected to any other device or outlet.
Equipment testing
Turn the Control Unit ON, and verify that the Login screen display appears. Do not use the
Control Unit if there are any irregular sounds or vibrations present.
1.8.2 Use environment
Cautions:
o Do not stack any objects on top of Control Unit, treatment catheter, rectal unit,
or Coolant Bag.
o Do not place the Control Unit near any electronic device or other equipment
emitting electromagnetic waves. The interference may compromise the
operation of the equipment.
¾ Provide ventilation space of at least 20-cm clearance around the base of the Control Unit
for operation.
¾ Do not turn ON the Control Unit with the touchscreen monitor lid closed. The touchscreen
will turn OFF, and the lid may become warm to the touch.
¾ Operate the Control Unit on a level surface.
¾ Operate the Control Unit under these operating conditions:
An ambient temperature range of +10°C to +30°C
A relative humidity range of 30% to 75%
An atmospheric pressure range of 700 hPa to 1,060 hPa
1.9 Safety instructions
Warning: Do not open the housing of the Control Unit; doing so risks receiving an electric
shock.
¾ Refer all Control Unit servicing to qualified Urologix personnel.
Warning: This equipment is not intended for use in areas where there is a danger of
explosion.
¾ Do not use the Control Unit in the presence of flammable substances.
Caution: The Targis System must be operated by trained and authorized personnel. You
should read and understand the instructions in this manual before operating the system.
18 of 19
Part number 250023-001 Rev 02
Section 1
System Description
This manual does not claim to address all of the safety concerns associated with the use of this
equipment. You must establish appropriate safety and health practices and determine the
applicability of regulatory limitations prior to use.
Perform the following Control Unit safety checks at least every 12 months:
• Cables and connectors for damage
• Equipment for physical damage
• Safety labels are readable
Maintain a written record of these safety checks, and service any equipment that does not meet
these standards.
Part number 250023-001 Rev 02
19 of 19

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.3
Linearized                      : Yes
Create Date                     : 2005:04:15 13:13:15-04:00
Modify Date                     : 2005:04:15 13:13:15-04:00
Page Count                      : 19
Creation Date                   : 2005:04:15 17:13:15Z
Mod Date                        : 2005:04:15 17:13:15Z
Producer                        : Acrobat Distiller 5.0.5 (Windows)
Author                          : Administrator
Metadata Date                   : 2005:04:15 17:13:15Z
Creator                         : Administrator
Title                           : Microsoft Word - 532897.DOC
EXIF Metadata provided by EXIF.tools
FCC ID Filing: S7M-CW5000A

Navigation menu