SECTION FOUR
CARE AND MAINTENANCE
INTRODUCTION
This section of the manual is designed to inform the operator of basic care and
maintenance of the instrument. It is recommended to verify the calibration annually.
In the event that recalibration is required, it is also recommended that the procedure
to recalibrate the system be performed by Alcon Technical Services personnel. If
a problem occurs on the instrument, call the Alcon Technical Services department
and give details of the circumstances and effects. From these elements, a specialized
technician will evaluate the problem and determine the maintenance requirements.
WARNINGS!
Maintenance on any part of the laser system must be performed with the laser
off and the main power plug disconnected.
When keyswitch power is on, all individuals in the laser room must wear laser
protective eyewear, OD 4 or above at 532 nm.
CAUTION
There are no operator replaceable parts other than the fuse. Contact Alcon Technical
Services for all servicing issues.
Care and Cleaning
WARNING!
A qualified technician must perform a visual inspection of the following components
every twelve months:
• Warning labels (see Section One)
• Power Cord
• Fuses
In case of a deficiency, do not use the system; call Alcon Technical Services.
A qualified technician must check ground continuity and both polarities for leakage
current every twelve months to ensure they are within the applicable standards
(for example: EN60601-1/IEC601-1). Values must be recorded. If they are above the
applicable standards, or 50% above initial measurement, do not use the system;
call Alcon Technical Services.
The following tips are recommended for proper care of the PurePoint™ Laser
system:
• Turn off the system correctly after each use with the rear panel switch.
• Cover the slit lamp with the plastic cover.
• Cover the fiber optic connector with the dust cover.
• Cover the fiber port with the dust cover.
• Clean the exterior portion of the equipment with a dry, lint-free cloth or tissue. No other
products can be used.
• Use care not to damage or scratch the laser apertures or fiber optic connector.
• Place the system into its traveling case when moving to another location.
8065751131
4.1
• Inspect fibers to ensure that they have not been compromised, i.e., chips, cracks, or loose
connectors.
The condition of the following system hardware components must be checked
periodically to identify any fault that may affect system operation:
• Chassis appearance.
• Operation of controls and indicators.
• State of the fibers and connecting cables.
Damaged hardware must be replaced to ensure safe operation. Call Alcon Technical
Services for assistance.
Mirror and Lens Cleaning
The mirrors and lenses of the LIO headpiece and Slit Lamp adaptation must be
kept clean and unscratched. Cleaning them requires special care and the following
materials:
• Standard lens cleaning paper
• Methanol of spectrographic quality.
The following tips will aid you in cleaning the optics:
• Use each piece of cleaning paper only once.
• Move the cleaning paper across the optic surface from one end to the other in one
continuous motion. Discard the cleaning paper and use a new piece for the next cleaning
pass.
• Do not use a back and forth rubbing motion on the optic surface.
CAUTION
Care and cleaning operations must be performed with the instrument turned off and
power cord disconnected. Use only optical quality paper and spectroscopic quality
methanol when cleaning the mirrors and lenses, otherwise the optics could be
scratched and their coatings destroyed.
FUSE REPLACEMENT PROCEDURE
NOTE: Use only the recommended fuses for the PurePoint ™ Laser as listed on
the fuse label.
4.2
Turn system power off and disconnect power cord
from the PurePoint ™ Laser before changing fuses.
Remove the fuse clip from the fuse holder using a
small screwdriver.
Inspect fuses in holder for damage or a burnt
connection.
Place new fuses in each side of holder in fuse clip
(replace with fuses rated T5A/250V or contact your
local Alcon representative).
Replace fuse clip into the fuse holder. Close fuse
holder.
Fuse
8065751131
CALIBRATION VERIFICATION
Calibration verification must be performed at least every twelve months to verify that
the laser output is within tolerance and calibration is not required. It is recommended
to call Alcon Technical Services before conducting the calibration verification
procedure.
CAUTION
Serious damage to the instrument may occur if these procedures are not
performed by qualified personnel.
SPECIAL TOOLS
• Computer, with browser software; MS Internet Explorer or equivalent
• Custom service ethernet cable (Alcon p/n 023-100)
• Power Meter, Thermopile type (Coherent FieldMaster w/ LM-10 head or
equivalent)
• Laser Safety Goggles (OD4 or above, at 532 nm wavelength)
• Optics cleaning kit, including spectroscopic grade methanol, lens paper and
air blower
• Light Meter (Labsphere HLMS 200P or equivalent) - A power meter may be
used instead, using the following conversion factors:
• Ophir Nova with PD300-SH head
(use conversion factor 2.5mW=1 Lumen; divide meter reading by 2.5)
• Newport 840-C with 818SL
(use conversion factor 1.9mW=1 Lumen; divide meter reading by 1.9)
• Coherent Field Master w/LM-10 head
(use conversion factor 2.9mW=1 Lumen; divide meter reading by 2.9)
WARNING!
Laser light emitted from the fiber and laser head is powerful enough to cause
serious eye or skin damage. Maintenance should be performed only by properly
trained personnel, following established guidelines for laser safety. The use of
protective eye wear is mandatory.
8065751131
4.3
1
Exposure Time Verification
1.1
Setup the system as shown in Figure 4-1. (Where slit lamp is not used, connect a
test fiber or endoprobe and direct the distal output into the photo cell.)
Slit lamp
adaptation
Laser Beam
Photo Cell
Slit lamp
Oscilloscope
Figure 4-1
4.4
Exposure Time Test Configuration (shown for slit lamp)
1.2
Set spot size to 250 microns on the zoom. Adjust the distance between the slit
lamp and photo cell to obtain a beam size of 2mm or more on the photo cell.
Use aiming beam to determine spot size on the photo cell.
1.3
Set the exposure time to 0.1s and treatment beam power to minimum then select
READY mode.
1.4
Fire the laser and record the exposure time as determined from the oscilloscope.
1.5
Repeat steps 1.3 and 1.4 for each time value listed in Table 4-1.
8065751131
2
Slit Lamp Power Verification
2.1
Setup the system as shown in Figure 4-2.
Slit lamp
adaptation
Laser Beam
Wattmeter Cell
Slit lamp
Wattmeter
Figure 4-2
8065751131
Power Test Configuration (shown for Slit Lamp)
2.2
Set the exposure time to CW.
2.3
Set spot size to 250 microns on the zoom. Adjust the distance between the slit
lamp and wattmeter cell to obtain a beam size of 2mm or more on the wattmeter
cell. Use aiming beam to determine spot size on the wattmeter cell.
2.4
Set the treatment power to 0.10 W then press the Standby/Ready key.
2.5
Fire the laser and record the wattmeter power reading into Table 4-1.
2.6
Repeat steps 2.4 and 2.5 for each value listed in the Slit Lamp section of Table
4-1.
Endoprobe Power Verification
3.1
Setup the system in a similar configuration as shown in Figure 4-2, to direct the
endoprobe distal output beam into the wattmeter cell.
3.2
Set the exposure time to CW.
3.3
Set the treatment power to 0.10 W then press the Standby/Ready key.
3.4
Fire the laser and record the power reading as determined from the Wattmeter.
3.5
Repeat steps 3.3 and 3.4 for each value listed in the Endoprobe section of Table
4-1.
4.5
4
LIO Power Verification
4.1
Setup the system in a similar configuration as shown in Figure 4-2, to direct the
LIO distal output beam into the wattmeter cell.
4.2
Set the exposure time to CW.
4.3
Set the treatment power to 0.10 W then press the Standby/Ready key.
4.4
Fire the laser and record the power reading as determined from the Wattmeter.
4.5
Repeat steps 4.3 and 4.4 for each value listed in the LIO section of Table 4-1.
Energy Matrix Completion
5.1
Complete the matrix by multiplying actual power by actual exposure time and
recording the result, as shown in the example below.
��������������
������������
������
����
������
����
������
�����������������
5.2
4.6
���������
���������
�������������
����
�����������
������x�����
���������������
���������
�������������
������������
Ensure that all calculated results are within the values listed in each matrix cell.
The listed values are ±15% of the set energy.
•
If all calculated energy values are within the specified limits, the system
calibration is OK.
•
If any of the calculated energy results are not within the specified limits,
the Terminal Efficiencies will need to be adjusted. Perform the Setting the
Terminal Efficiencies procedure following Table 4-1 or call Alcon Technical
Services.
8065751131
���������������
������
������
������
�����������������
�����������������
�����������������
�����������������
�����������������
�����������������
�����������������
����������������
�����������������
�����������������
�����������������
�����������������
�����������������
�����������������
�����������������
�����������������
�����������������
�����������������
�����������������
�����������������
�����������������
�����������������
�����������������
�����������������
�����������������
�����������������
�����������������
�
������
����������
������
������
������
���������
������
������
������
����������������
������
������
������
Table 4-1
8065751131
Energy Matrix
4.7
6
Setting the Terminal Efficiencies
If unable to sucessfully complete the Energy Matrix table, use the following
procedure to adjust the Terminal Efficiencies, and retest.
6.1
With the unit off, connect the service ethernet cable between the console
and service computer. and turn unit ON. Turn the computer ON and start the
browser program.
6.2
Type the IP address into the address box: 161.61.112.69, and hit return. When
the page loads, enter the password: ngl1
6.3
For slit lamp, endoprobe, and/or LIO, use a power test configuration similar to
as shown in Figure 4-1, to direct the distal output beam into the wattmeter cell.
6.5
Set power to 0.50 watts on the console, exposure time to CW, and select
READY mode.
6.6
Fire the laser and record the power reading as determined from the wattmeter.
6.7
Calculate the new Terminal Efficiency cooefficient using the following formula:
new coefficient = (old coefficient) X (measured power in Step 6.6)
0.5
4.8
6.8
Enter the new value in the Terminal Efficiency window for the respective
device, and click the SAVE button.
6.9
Repeat as needed to bring all values within compliance to complete the Energy
Matrix Table 4-1, for each delivery device. If unable to sucessfully complete
the matrix, the unit will need a System Calibration, as outlined in the following
procedure.
8065751131
SYSTEM CALIBRATION
WARNING!
Laser light emitted from the fiber and laser head is powerful enough to cause
serious eye or skin damage. Maintenance should be performed only by properly
trained personnel, following established guidelines for laser safety. The use of
protective eye wear is mandatory.
With the unit off, connect the service ethernet cable between the console
and service computer. and turn unit ON. Turn the computer ON and start the
browser program.
Type the IP address into the address box: 161.61.112.69, and hit return. When
the page loads, enter the password: ngl1
Enter and save the following values:
3.1
10 ºC for Minimum Diode Temperature value.
3.2
50 ºC for Maximum Diode Temperature value.
3.3
5 ºC for Minimum LBO Temperature value.
3.4
60 ºC for Maximum LBO Temperature value.
Enter and save the following values as denoted on the laser engine:
4.1 Diode Temperature
4.2. LBO Temperature
4.3 Maximum Current
Wait at least 2 minutes for the engine to come to proper working temperature for
the calibration. Select Port 1 on the console.
On the computer, set Simmer Value to 11 amps. Be aware the laser engine is
now lasing and producing visible output.
LBO Temperature Optimization 1.7.1 On the computer, vary the LBO Temperature Setpoint to maximize the
Pmon1 reading. Use the browser refresh button after each change, and allow
5 seconds for adjustment before reading the new Pmon value. Begin with the
default value noted in Step 1.4.2, and sequentially change in +/- 1, 0.5, and 0.25
-degree steps, to fine-tune the setpoint for maximum Pmon reading.
8065751131
On the computer, set the Simmer Value to 5 amps.
Select Output Calibration on the computer
4.9
10
11
4.10
Pmon 1 Low-Power Calibration 10.1
Select Port 1 on the console.
10.2
Press “Start Pmon Calibration” on computer.
10.3
Set laser in CONTINUOUS mode on the console.
10.4
Set POWER to 100mw on the console.
10.5
Select READY mode on the console.
10.6
Fire the laser and measure output power directly from Port 1.
10.7
Press “Start Calibration” on the computer
10.8
Input 100mW into the Low Power Display field.
10.9
Input the power, as previously measured, into the Actual Power Field,
and press Save
10.10
Return to STANDBY mode on the console.
10.11
Repeat the Pmon 1 Low-Power Calibration as needed (2 or 3 times) to
bring Displayed/Actual tracking as close as possible.
Pmon 1 High-Power Calibration
11.1
Press “Start Pmon Calibration” on computer
11.2
Set laser in CONTINUOUS mode on the console.
11.3
Set POWER to 1 Watt on the console.
11.4
Select READY mode on the console.
11.5
Fire the laser and measure the actual output power directly from Port
1.
11.6
Press “Start Calibration” on the computer
11.7
Input 1 Watt into the Low Power Display field.
11.8
Input the power, as previously measured, into the Actual Power Field,
and press Save
11.9
Return to STANDBY mode on the console.
11.10
Repeat the Pmon 1 High-Power Calibration as needed (2 or 3
times) to bring Displayed/Actual tracking as close as possible.
12
Repeat steps 1.10 and 1.11 for Low/High Power Calibration for Pmon 2.
13
Repeat the Delivered Power Calibration, adjusting the Terminal Efficiencies as
required, so to sucessfully complete the Energy Matrix for each delivery device.
8065751131
AIMING BEAM / LIO ILLUMINATION CALIBRATION
With the unit off, connect the service ethernet cable between the console
and service computer. and turn unit ON. Turn the computer ON and start the
browser program.
Type the IP address into the address box: 161.61.112.69, and hit return. When
the page loads, enter the password: ngl1
Aiming Beam Calibration -
4.
8065751131
3.1
Adjust aiming beam power output for Port 1 to 0.9 - 0.99mW.
3.2
Click “Set Max Value” on the computer.
3.3
Repeat for Port 2.
LIO Illumination Calibration 4.1
Adjust light output to 90 foot-candle.
4.2
Click “Set Max Value”
4.11
THIS PAGE INTENTIONALLY BLANK
4.12
LAST PAGE OF THIS SECTION
8065751131

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : No
Create Date                     : 2006:03:29 07:31:58Z
Creator                         : Adobe InDesign CS (3.0.1)
Modify Date                     : 2008:01:15 15:51:44-08:00
About                           : uuid:3f94c82c-a1e4-11dc-aac6-0003933e69a0
Producer                        : Adobe PDF Library 6.0
Trapped                         : False
Creation Date                   : 2006:03:29 07:31:58Z
Mod Date                        : 2008:01:15 15:51:44-08:00
Creator Tool                    : Adobe InDesign CS (3.0.1)
Metadata Date                   : 2008:01:15 15:51:44-08:00
Rendition Class                 : proof:pdf
Document ID                     : adobe:docid:indd:6f862a62-c088-11da-b0cf-836282a3a012
Instance ID                     : uuid:5519e28a-a1e5-11dc-88d7-0003933e69a0
Format                          : application/pdf
Title                           : NGL Book 1.indb
Page Count                      : 12
Has XFA                         : No
EXIF Metadata provided by EXIF.tools
FCC ID Filing: VMCPUREPT1

Navigation menu