BTS BTSWEMG2 Wireless EMG sensor for EMG system User Manual EREMB 01091 06 FREEEMG 100 RT ENG v 2 0 3

BTS SpA Wireless EMG sensor for EMG system EREMB 01091 06 FREEEMG 100 RT ENG v 2 0 3

EREMB-01091-06 FREEEMG 100 RT User Manual ENG v.2.0.3

Download: BTS BTSWEMG2 Wireless EMG sensor for EMG system User Manual EREMB 01091 06 FREEEMG 100 RT  ENG v 2 0 3
Mirror Download [FCC.gov]BTS BTSWEMG2 Wireless EMG sensor for EMG system User Manual EREMB 01091 06 FREEEMG 100 RT  ENG v 2 0 3
Document ID2039279
Application IDTwX+Kn8Cv5HGkw1jec3WVw==
Document DescriptionEREMB-01091-06 FREEEMG 100 RT User Manual ENG v.2.0.3
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize246.99kB (3087436 bits)
Date Submitted2013-08-09 00:00:00
Date Available2013-08-09 00:00:00
Creation Date2013-06-14 12:41:53
Producing SoftwareAdobe PDF Library 9.0
Document Lastmod2013-06-24 14:52:25
Document TitleEREMB-01091-06 FREEEMG 100 RT User Manual ENG v.2.0.3
Document CreatorAdobe InDesign CS4 (6.0.4)

user manual
english
version 2.0.3
Document Number : EREMB-01091-06
Published: June 2013
Copyright © 2010 - 2013 BTS SpA. All Rights Reserved.
contents
contents							
icons, symbols and acronyms					
radio regulation							
disposal (WEEE)						
intended use 							
regulatory label							
receiving unit regulatory label				
EMG wireless probes regulatory label 			
AUX wireless probes regulatory label 			
BTS charger regulatory label 				
warnings
copyright
introduction 							
general description 					
case contents						
system components						
receiving unit						
wireless EMG probes					
wireless FSW/EGN probes (optional)			
on-off analysis (optional)					
electrogoniometers (optional)				
charger							
installation							
user PC minimum configuration 				
connections						
BTS Bioengineering
10
10
10
11
12
13
16
17
17
18
21
21
22
25
27
29
30
33
33
33
contents
hardware installation					
description of the software on the user PC			
BTS EMG-Analyzer configuration check			
acquisition protocol						
34
37
38
41
appendix A – technical specifications 				
wireless probes						
receiving unit						
BTS internal coding					
appendix B – environmental specifications			
appendix C – power supply and switch off		
appendix D – battery						
appendix E – troubleshooting guide				
warning – invalid trigger mode 			
appendix F – declaration of conformity 				
appendix G – regulatory notice 				
FDA medical device reporting system—reportable events
adverse event report (21 CFR 803.32)			
47
47
47
48
49
50
51
52
52
53
54
54
55
BTS Bioengineering
icons, symbols and acronyms
Symbol in the instructions for the function.
The icon represents the information which requires special attention.
Symbol in the instructions for the function.
This icon makes reference to a more detailed discussion of the subject in
hand.
Symbol on the equipment:
The data appearing next to the manufacturer’s symbol refer to the place of
manufacture of the equipment itself.
Symbol on the equipment:
The “FCC” simbol refers to the Federal Communication Commission of
the USA. The device complies with the relevant regulations put forth by
the FCC as long as it is operated according to the instructions contained
in this manual and to all national and local regulations.
BTS Bioengineering
icons, symbols and acronyms
Symbol on the equipment:
The figure in the square indicates the insulation class and the part types
used. In accordance with Standard ISO 60601-1, the equipment has an
internal power supply and the parts used are type BF.
Symbol on the equipment:
Attention, read the information in the users’ manual carefully before using
the equipment.
Symbol on the equipment:
The double square indicates that the product is a medical device of II
Class (In accordance with the law EN 60601-1).
Symbol on the equipment:
CE mark with the code of the Notified Body. The CE mark certifies that
the product conforms to the standards applicable in the member states of
BTS Bioengineering
icons, symbols and acronyms
the European Union (see Declaration of Conformity).
Symbol on the equipment:
CE mark with the code of the Notified Body. The CE mark certifies
that the product conforms to the Directive 99/05/EEC - R&TTE and
obtained the Expert Opinion by IMQ.
Symbol on the equipment and in the users’ instructions:
Symbol for the separate disposal of electrical and electronic equipment, in
accordance with Directive 2002/96/CE (WEEE).
The equipment belongs to Group 8 (medical equipment).
In force in the nations of the European Union, Norway and Switzerland.
Rx only
Symbol for prescription only. U.S. Federal law restricts this device to sale
by or on the order of a physician or properly licensed practitioner.
BTS Bioengineering
icons, symbols and acronyms
REF
Symbol on the equipment:
Symbol located next to the model number (ref.to catalogue).
SN
Symbol on the equipment:
Symbol located next to the series number on the equipment.
Acronyms used in this manual:
RU
Receiving Unit
EMG Electromyography
WS
Workstation
BTS Bioengineering
radio regulation
Radio equipment identification:
- EMG probes:
FCC ID: YQH-BTSWEMG2
IC: 9188A-BTSWEMG2
- FSWEGN probes:
FCC ID: YQH-BTSWAUX
IC: 9188A-BTSWAUX
- Receiving Unit contains:
FCC ID: TFB-MATRIXLP
IC: 5969A-MATRIXLP
This device complies with part 15 of the FCC Rules. Operation
is subject to the following two conditions: (1) This device may
not cause harmful interference, and (2) this device must accept
any interference received, including interference that may cause
undesired operation.
Modifications not expressly approved by BTS SpA could void the
user’s authority to operate the equipment under FCC rules.
BTS Bioengineering
disposal (WEEE)
In disposing of the equipment observe the legal prescriptions.
In accordance with Directive 2002/96/CE (WEEE) all equipment supplied
after 13/08/2005 may not be disposed of in general domestic waste. This
equipment belongs to Category 8 (medical equipment) and is classified in
the Business-to-Business sector.
The symbol of the crossed out rubbish bin
indicates that the equipment must not be
disposed of in normal domestic waste.
The regulations for disposal may differ between
individual countries in the EU. In cases of doubt,
refer to the respective sales outlet.
This is a battery-powered equipment.
See Appendix D for information about the batteries used. Operate
and dispose of this equipment according to the instructions set in
the “warnings” section.
BTS Bioengineering
intended use
This equipment is an instrument for the EMG surface analysis, classified
as medical equipment in accordance with European Directive 93/42/CE
(and its amendments).
BTS FREEEMG 100 RT must always be used only for this purpose, by
qualified persons, in an environment suitable for the execution of EMG
analyses and respecting the prevailing regulations in the countries in which
it is being utilized.
BTS Bioengineering
regulatory label
Receiving Unit Regulatory label
EMG Wireless Probes Regulatory label
Label on the probes:
Probe ID
10
BTS Bioengineering
regulatory label
Label not on actual probes due to size constraints:
AUX Wireless Probes Regulatory label
Label on the probes:
Probe ID
Label not on actual probes due to size constraints:
BTS Bioengineering
11
regulatory label
BTS Charger Regulatory label
9V
12
BTS Bioengineering
3A
9V
3A
3A
warnings
We recommend to carry out any kind of operation keeping
strictly to the security regulations contained in this manual. The
safety of the instrument cannot be guaranteed if these conditions
are not respected.
BTS FreeEMG 100 RT is a medical device (EU Directive 93/42/CE and
its amendments, including Directive 2007/47/CE) which use must be at
all times be supervised by qualified and authorized personnel, according to
the laws in force in the nation it is in use. The EMG probes are classified as
ETSI EN 300 440 “Receiver category 3” according to Directive R&TTE
99/5/EEC.
The results of the acquisitions must be assessed by people legally authorised
by national law, who possess the suitable necessary knowledge of anatomy
and muscular function.
The instrument must be used in a medical environment, since it has a high
level of sensitivity (measured voltage levels of between 1 microvolt and 6
millivolt).
The uses of the device for other purposes and with methodologies
different from of those indicated in this manual are not to be
considered congruent with the precise use of the device.
During the preparation of the patient, take particular care that the system’s
components do not impede in any way the normal movements of the
subject. Apply the probes only on undamaged skin.
BTS Bioengineering
13
warnings
• Only use CE branded probes and hypoallergenic double-sided tape,
compatible with the usage on undamaged skin for brief periods of time
• Periodically verify the integrity of the system and of its components.
• To not wet or dip in water the parts which make up the system.
No modification of this equipment is allowed.
• Only BTS S.p.A. authorized technicians may maintain and operate
servicing to the instrument. BTS S.p.A. cannot be held responsible
for system safety should the instrument be opened, repairs carried out,
third parties software be installed, or system components be replaced by
persons other than those authorized by BTS S.p.A.
• Users cannot change any software configuration (including O.S. and
CD writer software).
• In case the device accidentally falls, tear of the probes or other accidents
always address authorized technical support.
• Use only the provided power supply unit FW7363M/09 (FRIWO) or
the one provided by BTS S.p.A. for supplying the charger unit. If a
different power supply unit is used, the compliance to IEC 60601-1 is
not ensured.
• Only original cables must be used, otherwise BTS S.p.A. cannot assure
the safety of the instrument. Should it be necessary to replace any part
of the system, only original BTS S.p.A. parts may be used.
In addition to the users’ instructions, the prescriptions regarding
accident prevention and technical regulations regarding
occupational safety must also be complied with. The appertaining
national regulations and standards of the country of use, with
regards prevention of accidents and environment, are an extension
of the users’ instructions.
14
BTS Bioengineering
warnings
• Make sure that the cables have been connected correctly. When
disconnecting cables, use the connectors and not the cables themselves
to unplug the connectors.
• Mains plug of external power supply unit is considered as disconnecting
device. Avoid connecting the probes to the charger with inverted
polarity with respect to that shown on the cover of the recharger – This
could cause irreparable damage to them.
• For a safe use and adequate maintenance of rechargeable batteries strictly
follow the instructions given in this manual. If rechargeable batteries
are used in such a way that is not the one specified by BTS S.p.A. the
shelf life, functionality and the integrity of the batteries is not ensured.
• ESD application to EGN probes, causes a loss of link to the device.
Temporary loss of function or performance which is recoverable with
operator action.
BTS FREEEMG 100 RT is a device that is able to function
COUNTINUOUSLY, this is of course limited by the battery
duration and by the memory available for the acquisition data
storing.
The device uses lithium ion battery. For the battery replacement
and disposal please contact the technical support. At any rate,
ensure that device component (i.e.) probes, receiving unit, …)
integrity is never compromised.
The information contained in this manual is subject to change
without notice and does not constitute product specifications or
any obligation on the part of BTS S.p.A.
BTS Bioengineering
15
copyright
The software of the system described in this manual is supplied with the
“licence to use” contract. The software may be used or copied only as
stipulated under the terms of this contract.
No part of this manual may be copied or transmitted in any form or
means, electronic or mechanical, including photocopying, without prior
written permission from BTS S.p.A.
Unless otherwise specified, any reference to companies, names, data and
addresses used in the reproduction of the screens and the examples are
purely incidental, and has the sole purpose of illustrating the use of the
BTS product. All trademarks are registered by the respective owners.
This publication contains reserved information which is the
property of BTS S.p.A.
The recipient acknowledges that the illustrations and information
supplied in this manual shall not be made available to third parties
without explicit written agreement by BTS S.p.A.
16
BTS Bioengineering
introduction
General description
BTS FREEEMG 100 RT allows for a functional evaluation and a realtime visualization of the acquired signals for biofeedback and monitoring
applications.
That makes it an indispensable instrument in the fields of sport,
rehabilitation, ergonomics, neurology and orthopaedics.
With BTS FREEEMG 100 RT the patient set up in very quick: the
lightweight probes are attached directly to the pre-gelled electrodes and do
not require any additional fixing such as adhesive tape. Thanks to the total
absence of cables the patient can move around freely.
BTS FREEEMG 100 RT is supplied with EMG-Analyzer the most
complete software solution for analyzing electromyographic signals and
with BTS MIOFEED, an easy-to-use therapeutic solution, studied also
for home applications.
BTS EMG-Analyzer software includes predefined templates for evaluations
in the clinical, sports, and research field: Jump, plyometrics, walking,
fatigue analysis, isokinetic, etc...
BTS EMG-Analyzer also has an editor for creating elaboration protocols:
thanks to an innovative object interface, that translates the biomechanical
analysis language into graphical form, the user can develop quickly and
effectively customized analysis protocols.
BTS Bioengineering
17
introduction
BTS MIOFEED software allows a real-time monitoring of muscular
activations during rehabilitative or sports tests execution.
The EMG signal acquired with BTS FREEEMG 100 RT are transmitted
in real-time directly at the receiving unit connected to the PC and
transformed in graphic form and sounds to provide a prompt audio-visual
feedback to the subject. This increase of audio and visual afferences is
fundamental in order to improve the contractile capacities of the muscles.,
also of the paretic ones.
BTS FREEEMG 100 RT seamlessly integrates with BTS motion analysis
systems, through the SMART (and ELITE) dedicated software.
Case contents
Standard components:
• Receiving Unit
• Up to 6 wireless EMG probes (identificative labels available in 4 different
colors)
• Magnet for EMG probes activations
18
BTS Bioengineering
introduction
• Probes Charger (AC adapter included)
• Set of disposable electrodes
• USB extension cable
• User manual
Optional components:
• Footswitch Kit
• 2 FSW/EGN Wireless Probes
• 2 Connectors of 4 Switches
• 10 single Switches
BTS Bioengineering
19
introduction
• Electrogoniometer Kit
• 1 FSW/EGN Wireless Probe
• 1 Connector			
• 1 Electrogoniometer
You will receive the instructions for use for other possible optional
components not mentioned in this manual.
20
BTS Bioengineering
system components
BTS FREEEMG 100 RT system consists of two parts: the receiving unit
and the wireless probes.
Receiving unit
The receiving unit, connected to the PC trough USB, allows the WiFi
reception of the signal acquired by the wireless probes.
The receiving unit is able to handle simultaneously 6 wireless probes with
the following limitations:
• up to 6 EMG probes, used for acquisition of the same number
of electromyographic signals;
• up to 2 probes for the connection of 4 switches each, used for
the basographic acquisition;
• up to 6 probes for electrogoniometers connection, used for
acquisition of up to 2 angle components each;
• it is not possible to acquire simultaneously FSW and EGN
probes.
BTS Bioengineering
21
system components
All the probe combinations that respect what indicated above are possible.
WiFi transmission
Wireless EMG Probes
BTS FREEEMG 100 RT utilizes miniaturized probes with active electrodes
weighing less than 13 grams.
The special design ensures maximum space-saving and comfort for the
patient who is free to move around without obstacles.
The probes can be hooked on directly to the pre-gelled electrodes without
requiring additional fixing with plasters or double-sided tape.
This together with the total absence of cables enables a much faster patient
preparation, drastically reducing the time of each session.
Each probe consists of a mother electrode and a satellite electrode, each
fitted with a clip. The two parts, connected via a flexible cable, may be
positioned as needed by the user at adjustable distance (electrodes with
variable geometry).
22
BTS Bioengineering
system components
All probes are equipped with a solid state memory buffer, to prevent data
loss for problems due to the WiFi network or due to exceeding the useful
operating range.
mother electrode
LED
satellite electrode
mother electrode clip
satellite electrode clip
flexible cable
probe ID
Each probe is fitted with an LED indicating its state.
The probes can be in one of a number of different states:
• Charge: steady blue LED.
During the recharging phase the steady blue LED is on.
This phase occurs when the probes are connected to the charger
turned on, and the charge level is less than 90%.
When the charge level reaches 90% the led turns OFF.
Since, by connecting a probe to the charger on, it enters in “Deep
Sleep” mode, even while charging the probe will be completely
passive and does not respond to any commands.
• Active-Scanning: white LED which cyclically lights for a few
seconds.
In this mode the probe is searching for the receiving unit.
At intervals of about 1 minutes it carries out a scan of the
frequencies of few seconds. During the scan the white LED
flashes quickly.
BTS Bioengineering
23
system components
• Active-Connected: white LED which flashes slowly.
When the probe and the receiving unit establish a connection,
the white LED begins to pulse slowly: the probe is waiting for
commands.
If the connection is interrupted, the probe returns to “ActiveScanning” mode and attempts to re-establish the connection
with the receiving unit.
• Active-Capturing: white LED which lights and goes out at
regular intervals.
During acquisition the white LED flashes at regular intervals of
approximately one second. At the conclusion of the acquisition,
the probe returns to the “Active-Connected” condition.
If during the acquisition, connection to the receiver unit is lost,
the probe continues to acquire, storing the data locally for one
minute and at the same time scans the assigned channel trying
to reconnect to the receiving unit. If after one minute the scan
is unsuccessful, the probe returns to the “Active-Scanning”
condition interrupting the storage of data.
• Completely discharged or in “Deep Sleep” mode: LED is off.
If the probe is completely discharged the LED does not display
any flashing cycle and is off.
The same happens when it is in “Deep Sleep” mode (except
during the recharging phase in witch the led is steady blue).
The probes in “Deep Sleep” mode do not perform any
scan cycle, but are turned off. Is therefore guaranteed
energy savings.
24
BTS Bioengineering
system components
To put the probe in “Deep Sleep” mode it is necessary to connect
them to the Charger switched on, or to put them in contact
with a magnet for half a second.
Before the next use is necessary to reactivate the probes,
putting them in contact with a magnet.
The probes are charged by a dedicated charger to which the probes are
connected via their respective clips.
For more info about the probes charge sse the paragraph “Charger”
of this chapter.
Wireless FSW/EGN Probes (optional)
For collecting the on-off analysis signals coming from the Footswitch or for
collecting data from the Electrogoniometers (optional system components)
BTS FREEEMG 100 RT uses wireless probes which must be connected to
the FSW or EGN probes using a special connector.
The probe will work differently if used with one or the other probe
and will receive from the same receiving unit information on its
work modality during activation.
BTS Bioengineering
25
system components
The probe consists of a single parallelepiped-shaped block.
The upper face has an ID tag characterized by a color (Green, Red, Yellow,
Blue) and a letter (A, B, C, D, E or F) and a status LED.
The probes can be in one of a number of different states:
• Charge: steady red LED.
During recharging the red LED is on, the probe is completely
passive and does not respond to any command.
When the battery is fully recharged or when the probe is removed
from the charger, if sufficiently recharged it passes to the “ActiveScanning” mode.
• Active-Scanning: green LED which cyclically lights for a few
seconds.
In this mode the probe is searching for a receiving unit.
At intervals of about 3 minutes for 3 seconds its carries out a
scan of the frequencies.
During the scan the green LED flashes quickly.
• Active-Connected: Green LED which flashes slowly.
When the probe and the receiving unit establish a connection,
the green LED begins to pulse slowly: the probe is waiting for
commands.
If the connection is interrupted, the probe returns to “ActiveScanning” mode and attempts to re-establish the connection.
• Active-Capturing: Green LED which lights and goes out at
regular intervals.
26
BTS Bioengineering
system components
During acquisition the green LED flashes at regular intervals of
approximately one second.
At the conclusion of the acquisition, the probe returns to the
“Active-Connected” condition.
If during the acquisition, connection to the receiver unit is lost,
the probe continues to acquire, storing the data locally for one
minute and at the same time scans the assigned channel trying
to reconnect to the receiving unit.
If after one minute the scan is unsuccessful, the probe returns
to the “Active-Scanning” condition interrupting the storage of
data.
• Probe discharged: LED is off.
If the probe is completely discharged the LED does not display
any flashing cycle and is off.
The two lateral faces have two connectors; the one on the ID tag side
serves to charge the probe, and a cable will be connected to this to enable
connection to the Charger.
The one on the other side is for the probe connections (FSW or EGN).
Finally, the ID identifier of the probe is on the bottom left corner of the
back side.
On-off analysis (optional)
The Footswitches are useful in defining the contact points during the
contact phases of deambulation.
BTS Bioengineering
27
system components
The footswitches consist of a resistive membrane, (FSR technology), of
diameter 18 mm and thickness less than 0.5 mm, expressly designed for
applications in the analysis of movement.
The compact size of the instrument permits a maximum of flexibility in
positioning on the patient’s foot.
For applications other than gait analysis, there are available on request
smaller diameter (8 mm) switch probes (applicable, for example, to the
finger), and square (useful for tapping tests), 44 mm x 44 mm.
BTS FREEEMG 100 RT permits up to 8 basographic zones to be
measured, through 2 connectors from 4 single switches (usually right and
left side) that are connected to the two FSW/EGN wireless probes.
The footswitch channels are not supplementary to the 6
electromyographic channels. Moreover it is not possible to acquire
simultaneously FSW and EGN.
Refer to § “receiving unit” for the possible probes combination.
28
BTS Bioengineering
system components
Electrogoniometers (optional)
The Electrogoniometer is an easy to use device that allows the measurement
of joint angle progress over time.
There are primarily two types of electrogoniometers: the potentiometer
and the strain gauge. BTS FREEEMG 100 RT uses the strain gauge
electrogoniometer of Biometrics LTD.
There are single-axle models for the neck (axial rotation) and the forearm
(prone-supination) and biaxial models for other main joints: wrist, elbow,
knee, ankle, hip and back.
The strain gauge electrogoniometers are made up of two sensors, connected
to each other, that are fixed to the bone segments involved in the joint to
value.
The measure of the angle is provided by the relative angle between the
axes of the two sensors and, unlike the potentiometric electrogoniometers,
it doesn’t depend on the linear slidings in which the two extremities can
incur.
Each electrogoniometer is connected to a FSW/EGN wireless probe
BTS Bioengineering
29
system components
using the appropriate connector. Each receiving unit can handle up to 6
electrogoniometers.
The EGN channels are not supplementary to the electromyographic
ones, each EGN probe replace a EMG probe. Each EGN probe
allows acquiring up to two angular components. Moreover it is
not possible to acquire simultaneously FSW and EGN.
Refer to § “receiving unit” for the possible probes combinations.
Charger
The Charger, included with the product, charges the BTS FREEEMG 100
RT probes.
The Charger can simultaneously charge up to 8 EMG probes and 2 FSW/
EGN probes.
More units can be connected in series for simultaneous power
supply through the same AC adapter, using the cable included.
30
BTS Bioengineering
system components
The EMG probes are connected to the charger using the same clips that
normally collect the EMG signal.
Refer to the cover of the charger to identify the correct polarity.
The probes cannot be recharged if the polarities are inverted.
The Charger comes with an output short circuit protection
system also in case of reversed recharging poles. At any rate, poles
connected incorrectly will not recharge.
While the FSW/EGN probes connect to the charger using the special
connector as shown in the figure below:
To recharge connect all the probes that you would like to charge to the
Charger (follow the instructions described above) and connect the AC/
BTS Bioengineering
31
system components
DC adaptor to the mains and turn on the switch located on the rear panel.
When the Charger is properly connected to the mains and has been turned
on, the status LED “Power” will show a steady GREEN light.
The charging status of the EMG probes and of the FSW/EGN
probes is indicated by the respective status LED (see § “Wireless
EMG Probes” and “Wireless FSW/EGN Probes”).
Note that inserting the EMG probes into the Charger when it is
on, these come in “Deep Sleep” mode.
It is necessary to reactivate the probes, prior to use them, using a
magnet.
32
BTS Bioengineering
installation
User PC minimum configuration
Operating system		
Windows 7
Processor			
Intel Dual Core
RAM				
2 GB
Video resolution			
1280x800
Disk space			
100 MB for the application,
not including storage for acquired data
USB				
2.0
Connections
The wireless probes transmit in real-time the acquired data to the receiver
connected via USB to the Workstation.
Connect the receiver to the WS using the USB connector.
If the morphology of the WS does not allow direct connection of
the receiver, use the USB extension cable.
Also verify that the probes are fully charged and ready for use:
Note that to be recognized and activated by the system it is
necessary that all the probes have been disconnected from the
Charger (if switched-on) and that the EMG probes have been
also reactivated using a magnet.
BTS Bioengineering
33
installation
Hardware installation
The first step of the hardware installation procedure consist in the USB
receiver driver installation.
To do that enter in the CD player of the workstation, the BTS CD
software containing the installation files that contains the “EmbEMGX.X.XX.X.zip” file.
Extract the .zip file in the local disk C and then perform a double click on
the file “CDM20814_Setup.exe” and wait for the window that has just
opened, to close.
After have performed all the steps above, it will be necessary to configure
the port for the USB receiver.
To set the characteristics of the receiver the system requires the
receiver to be connected to the PC via a USB port.
Then, clicking the mouse right button on “My Computer”, select
“Properties”, choose “Hardware” and then enter the “Device Manager”:
34
BTS Bioengineering
installation
Once inside the Device Manager menu, select “Ports (COM & LPT)” and
click the right button on the item below “USB Serial Port (COMx)” and
select “Properties”.
Entering the port properties window it will be displayed the following:
BTS Bioengineering
35
installation
In which you must select the TAB “Port Settings”.
From this TAB set in the drop box menu related to the “Bits per second”,
the value 115200 and then click “Advanced”.
In the window that opens, you must set the value of “Latency Timer
(msec)” equal to 1msec, and then click “OK”, returning to the previous
screen.
36
BTS Bioengineering
installation
Also in the “Port Settings” window click on “OK” then close the windows
remained still open.
During the setting operations of the COM port, take note of the
port number that the PC automatically associates to the device,
because you will need it at the software first start.
Description of the software on the user PC
BTS EMG-Analyzer is the complete and highly flexible solution for
making advanced elaborations of electromyographic signals and angular
measurements of body segments.
Includes predefined templates for evaluations in clinics, sports, and
research and an editor to develop customized elaboration protocols.
BTS Bioengineering
37
installation
In the following paragraph we will refer to this software, however BTS
FREEEMG 100 RT is manageable also by BTS MIOFEED software and
all BTS applications of SMART family dedicated to the motion analysis.
For more details concerning the use of BTS FREEEMG 100 RT
with other software, please refer to its specific manual.
Before to proceed verify that the software BTS EMG-Analyzer has
been properly installed on the user PC (please refer to its manual
for the installation procedure).
BTS EMG-Analyzer configuration check
Keeping the receiver still inserted in the USB port, launch BTS EMGAnalyzer double-clicking on the relative icon.
If the following windows appears:
Click on “OK” to reach the software main screen.
On the menu bar at the voice “Laboratory” select “Set Emg Device”.
The following windows will open:
38
BTS Bioengineering
installation
Verify that the check is on “EMG embedded”.
Select the correct Port number acting on the drop box menu, according
with the port number that the PC has automatically associated to the device
during the COM port setting operations (see §”Hardware installation”).
If the port number selected is the correct one the software will activate the
Sensors area of the windows and all the info about the sensors associated
with the USB receiver connected to the workstation will be visualized.
BTS Bioengineering
39
installation
Verify that all the parameters values ​​coincide with those reported on the
supplied sensors that you want to use (see § “Wireless EMG probes” and
“Wireless EGN/FSW probes”).
During the acquisition the system will acquire always all the
sensors enable in this windows. Is possible to disable the sensors
that you don’t need for the acquisition. Click on “Update” before
closing the window to save the new configuration.
To enable a sensor click on the white square correspondent, the check
mark will appear and then fill the fields “Serial”, “Label Code” and “Label
Color”,​ according with the value reported on the probes (see § “Wireless
EMG probes” and “Wireless EGN/FSW probes”), and select a “Type”
between “EMG” and “AUX”.
Click on “Update” before closing the window to save the new configuration.
40
BTS Bioengineering
acquisition protocol
Now we will describe a basic procedure about how to create a new protocol
using BTS EMG-Analyzer software.
Refer to the EMG-Analyzer manual for further details about the
“Protocol builder” function.
The procedure may be different if performed with other
applications. In this case, refer to the specific manual of the
software you are using.
Please note that with BTS FREEEMG 100 RT it is possible to acquire up
to 6 probes.
Refer to § “Receiving unit” for the details about possible probe
combinations.
Launch the software EMG-Analyzer.
Before to proceed we suggest you to make the procedure described
in the § “Software configuration check” to be sure to have selected
the EMG embedded device and to have enabled all the probes
required for the acquisition.
To create a new protocol, from the menu voice “Laboratory” of BTS
EMG-Analyzer, select “Create Emg Protocol” .
BTS Bioengineering
41
acquisition protocol
The window for the creation or modification of protocols will open.
If the receiver is connected to the PC and correctly set, in the first column of
the protocol table there will be reported the labels of the probes connected
to the receiver, and the active protocol will be displayed.
If the active protocol is not consistent with the set of probes
connected to the receiver the following warning message will be
displayed on the bottom part of the windows:
42
BTS Bioengineering
acquisition protocol
To proceed with the acquisitions it is necessary to have the probes
set and the selected protocol compatible.
To reach this condition it is possible:
- to select a line of the protocol and modify any fields to make
it consistent with the probes set,
- to select an other protocol compatible with the probes set
among the ones saved in the database and displayed in the
“Protocol folder” area of this windows, and set it as “Active
protocol” clicking on “Set Active”,
- coming back to the “Set Emg Device” windows and modify
the probes set.
To create a new protocol follow this procedure:
- click on “New”. The fields to be filled for the protocol creation will
be enabled.
- select the anatomical map containing the muscle you want to add to
the protocol and click on the little yellow square that identified it;
immediately the first free row available in the protocol table will be
filled with the info related to the selected muscle.
- repeat this operation for all the muscles to be acquired.
- if you want to acquire Footswitches, first of all verify that there were
no EGN already selected in the protocols, otherwise the protocol
won’t be valid and the software won’t allow saving it.
Select the “District” anatomical map and proceed click on the right
and then on the left foot by selecting the corresponding square in the
BTS Bioengineering
43
acquisition protocol
anatomical map:
- If you want to acquire Electrogoniometers, first of all verify that there
were no FSW already selected in the protocols, otherwise the protocol
won’t be valid and the software won’t allow saving it.
Check the first free line in the protocol table, click on the “Type” field
and select EGN among the voices available in the drop box menu.
In the same way, click on the “Description” field and select among the
drop box menu voices the angles you want to acquire with the EGN,
then on the “Side” field to select the body side to which the data refers.
- indicate a name for the protocol specifying it on the “Name” box.
-it is possible to insert also a short description of the protocol using the
appropriate box “Description”.
When every signal of the protocol has been inserted, proceed saving the
44
BTS Bioengineering
acquisition protocol
protocol, clicking on “Save”.
If the protocol is compatible with the probes combination allowed by the
system the protocol could be saved.
If the Probes setup is compatible with the protocol the following windows
will be displayed:
Clicking on “Yes” it will be set as “Active protocol” and will be used in the
next acquisition session.
Instead if the Probes setup is not compatible with the protocol the message
displayed will be:
To proceed with the acquisition it is necessary make the probes set
and “Active protocol” consistent.
In both cases clicking on “NO” the protocol will be saved in the “Protocol
Folder” and will be available in the future, but the “Active protocol” won’t
be modified.
BTS Bioengineering
45
appendix A
technical specifications
Wireless Probes
Geometry: 			
Electrodes: 			
Separation: 			
Autonomy: 			
Battery: 			
Dimensions: 			
Weight: 			
Frequency used: 			
Input impedance: 		
CMMR: 			
Resolution: 			
Acquisition frequency:		
Sensitivity: 			
Accuratezza di misura*:		
variable
standard with clip connection
min: 16mm - max: 66mm
8h of use
some days stand-by
some months deep sleep
rechargeable, lithium ion
14x41,5x24,8mm mother electrode
diameter 16x12mm satellite electrode
13g battery included
ISM band
2.4GHz (standard IEEE802.15.4)
>10 GOhm
>110 dB @ 50-60Hz
16bit
1KHz
1μV
± 2%
Receiving Unit
Connection: 			
Dimensions: 			
Weight: 			
USB
82x44x22,5mm
80 gr
BTS Bioengineering
47
appendix
Frequency used: 			
ISM band
2.4GHz (standard IEEE802.15.4)
* The system is calibrated at the factory. No further calibration is required
BTS internal coding
Name of device, components, parts and/or
accessories as per product label
FREEEMG 100 RT
USB receiving unit
EMG wireless probes
Identifier for
device (bar code,
catalogue, model or
part number)
FREEEMG 100 RT
FSW/EGN wireless probe
Foot switches: Insole individual foot switches for the
automatic identification of the gait events.
Electrogoniometers: Sensors for the measurement of
FRESPU12
joint angles progress over the time.
EMG-Analyzer: software application for EMG signal
analysis
Charger
48
BTS Bioengineering
SMAN0901
FRESPU06
appendix B
environmental specifications
Min
Max
Operating Temperature
-20°
+45°
Operating Humidity
50%
80%
0°
+40°
50%
80%
0m
2000m
Storage and Transport
Temperature
Storage and Transport
Humidity
Altitude
Note
Relative,
non-condensing
Relative,
non-condensing
Degrees of protection provided by the dangerous enclosures of water and
dust (IEC 60529): IPX0.
BTS Bioengineering
49
appendix C
power supply and switch off
The receiving unit is powered by the USB port.
To switch off the system the following operation must be done:
- Exit from the application software
- Unplug the receiving unit from the USB port.
- Put the probes in “Deep Sleep Mode” placing them on the “Charger”.
50
BTS Bioengineering
appendix D
battery
BTS FREEEMG 100 RT probes are internally powered.
The Probe battery replacement can be done only by BTS qualified
personnel.
The probes are sealed to avoid the access to the internal circuit
components.
Batteries are equipped with battery protection circuit to:
-over-voltage, threshold 4.3V
-under-voltage, threshold 2.8V
- short-circuits
The specific characteristic of the Wireless probes equipped with the battery
are:
Quantity: 		
1 per each EMG probe
Technology: 		
lithium polymer (Li-Poly)
Removable: 		
NO, BTS technical service is required
BTS Bioengineering
51
appendix E
troubleshooting guide
Warning – Invalid trigger mode
During the hardware installation, the USB port dedicated to the receiver,
is configured. May happen that these port settings went lost.
This event may occur for example when another USB device is connected
to the PC and the PC assigns to it the same port previously assigned to the
receiver, returning the port setting to
​​ the default configuration.
In these cases it may occur that by launching an acquisition the following
warning message may appear:
In this case to proceed immediately with the acquisition, it is necessary
to restart the application, however the same problem will recur in later
acquisitions.
To resolve this problem you must repeat the USB port configuration
procedure described in the “Hardware Installation” paragraph of the
“Installation” chapter.				
52
BTS Bioengineering
appendix E
declartion of conformity
DECLARATION OF CONFORMITY
BTS SpA
Via della Croce Rossa 11, 35129 Padova (PD)
– Italy
Tel. +39 049 981 5500
Fax +39 049 792 9260
declare under our sole responsibility that the product(s):
name / description:
model:
S/N:
Electromyographic system / device for recording myoelectrical activity.
FreeEMG 100 RT
SN
satisfies the essential requirements of the Medical Devices Directive 93/42/EC (and its amendments inluding
2007/47/CE), and therefore carries the CE marking of the European Union. The conformity assessment procedure is
according to the article 11 of the directive (Annex II.3 full quality assurance) and the article 12 it is not applicable.
In accordance with Annex IX of the 93/42/EC directive it is classified as follow:
CLASS “IIa” (rule 10)
In accordance with IEC 60601-1 is also classified as follow:
Class: internally powered device
Applied part type: BF
The product conforms to the following standards:
EN ISO 14971
Medical Devices - Application of risk management to medical devices.
IEC 60601-1
Medical Electrical Equipment - Part 1: General Requirements for basic safety and
essential performance.
IEC 60601-1-2
Medical Electrical Equipment - Part 1-2: General requirements for basic safety and
essential performance - Collateral standard: electromagnetic compatibility Requirements and tests.
IEC 60601-1-6
Medical Electrical Equipment - Part 1-6: General Requirements for basic safety and
essential performance - collateral Standard: Usability
EN 62304
Medical device software - Software life-cycle processes
ETSI EN 301 489-3 Electromagnetic compatibility and Radio spectrum Matters (ERM) – Electromagnetic
Compatibility (EMC) – standard for radio equipment and services – Part 3: Specific
conditions for Short-Range Devices (SRD) operating on frequencies between 9 KHz and
40 GHz.
ETSI EN 301 440-2 Electromagnetic compatibility and Radio spectrum Matters (ERM) – Short Range Devices
(SRD) – Radio equipment to be used in the 1 GHz to 40 GHz frequency range – Part 2:
Harmonized EN covering essential requirements of Article 3(2) of the R&TTE Directive.
This compliance is valid ONLY for the equipment identified when used in a manner consistent
with the intent of the referenced documents and according to the product’s usage manual.
Notified Body : TÜV Product Service GmbH, Zertifizierstelle, Ridierstrasse 65, 80339 München – Germany, Identification N. 0123.
EC certificate N. G1 12 10 65301 003 valid until January, 16 2018.
Padova, Date
Bruno Ros
CEO
BTS S.p.A.
BTS Bioengineering
53
appendix G
regulatory notice
FDA Medical Device Reporting System—Reportable Events
Notice to Agents: for inclusion in all BTS systems supplied to the
United States of America, the master Medical Device Reporting
(MDR) file is located at BTS S.p.A. Should an adverse event occur,
the following form is to be completed and forwarded within one
working day to BTS S.p.A.
Department of Health & Human Services,
US Food and Drug Administration
Medical Device Reporting System—Reportable Events
Code of Federal Regulations
Title 21, Volume 8
Revised as of April 1, 2006
Cite: 21CFG803.32
Under 803.1(a) device user facilities and manufacturers must report
deaths and serious injuries to which a device has or may have caused or
contributed. Should such an event occur, please complete the following
details and forward the document in accordance with the applicable
regulations and time limits to one of the following addresses:
BTS S.p.A.
via della Croce Rossa 11
35129 Padova PD - Italy
tel +39 049 981 5500
fax +39 049 792 9260
54
BTS Bioengineering
appendix
Adverse Event Report (21 CFR 803.32)
Use blank pages if required.
Section A. Patient Information
Patient confidentiality to be maintained unless authorized
otherwise in writing by User Facility.
Patient name or other identifier
Age at the time of the event (Years, Months),
or Date of birth (MM/DD/YYYY)
Female
Gender
Male
lb
Weight
kg
Section B. Adverse Event or Product Problem
Identification of adverse event or product problem
(check all that apply)
Adverse Event
Product Use Error
Product Problem (e.g. defects/malfunctions)
Problem with Different Manufacturer of Same System
BTS Bioengineering
55
appendix
Outcomes attributed to the adverse event
(check all that apply)
Death:
(MM/DD/YYYY)
Life-threatening injury or illness
Hospitalization—initial or prolonged
Required intervention to prevent permanent
impairment/ damage (Devices)
Disability or permanent damage:
Congenital Anomaly/Birth Defect
Other Serious (Important Medical Events)
Date of Event
(MM/DD/YYYY)
Date of this report
(MM/DD/YYYY)
Describe event, problem or product use error
(include a discussion of how the device was involved, nature of the
problem, patient follow-up or required treatment, and any environmental
conditions that may have influenced the event)
56
BTS Bioengineering
appendix
Relevant tests/laboratory data, including dates
Other relevant history, including preexisting medical conditions
(e.g. allergies, race, pregnancy, smoking and alcohol use, liver/kidney
problems, etc.)
Section C. Device Information
Brand Name
Type of Device
Manufacturer name
and address
Model #
Lot #
Catalog #
Expiration Date
(MM/DD/YYYY)
Serial #
Other #
BTS Bioengineering
57
appendix
Operator of the device
(delete not applicable)
health professional
patient
lay user
other:
(specify)
Date of system installation
(MM/DD/YYYY)
Device available for
evaluation?
(Do not send to FDA)
yes
no
Returned to BTS S.p.A.
or its agents on:
(MM/DD/YYYY)
Concomitant medical products and therapy dates
(do not report products that were used to treat the event)
58
BTS Bioengineering
appendix
Section D. Initial Reporter Information
For the reporter who initially provided information to you, or to the
manufacturer or distributor:
Name
Address
Telephone Number
E-mail Address
Health Professional?
Yes
No
Occupation (include
speciality if appropriate)
Initial reporter also sent a
copy of report to FDA?
Yes
No
BTS Bioengineering
59
appendix
Section E. User Facility Information
Health Professional?
User Facility
Importer
User Facility Number
User Facility Name
User Facility Address
Contact Person
Phone Number
Date User Facility became
aware of event
Type of Report
(MM/DD/YYYY)
Initial
Follow-up #
Date of this Report
(MM/DD/YYYY)
Approximate age of system
Event problem Codes (refer to “MEDWATCH Medical Device Reporting
Code Instructions”)
Report sent to FDA?
Yes
(MM/DD/YYYY)
No
Report sent to
manufacturer?
Yes
(MM/DD/YYYY)
No
60
BTS Bioengineering
appendix
Location where event
occurred
Hospital
Home
Nursing Home
Outpatient Treatment Facility
Outpatient Diagnostic Facility
Ambulatory Surgical Facility
other:
(specify)
Manufacturer Name/Address
BTS Bioengineering
61
BTS Bioengineering Corp.
147 Prince Street - Suite 11
11201 Brooklyn NY
USA
info: +1 347 204 7027
helpdesk: +1 646 575 0426
www.btsbioengineering.com
info@btsbioengineering.com
BTS S.p.A.
viale Forlanini 40
20024 Garbagnate M.se MI
Italy
tel +39 02 366 490 00
fax +39 02 366 490 24

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : Yes
Create Date                     : 2013:06:14 12:41:53+02:00
Creator                         : Adobe InDesign CS4 (6.0.4)
Modify Date                     : 2013:06:24 14:52:25+02:00
XMP Toolkit                     : Adobe XMP Core 4.2.1-c041 52.342996, 2008/05/07-21:37:19
Metadata Date                   : 2013:06:24 14:52:25+02:00
Creator Tool                    : Adobe InDesign CS4 (6.0.4)
Thumbnail Format                : JPEG
Thumbnail Width                 : 256
Thumbnail Height                : 256
Thumbnail Image                 : (Binary data 7749 bytes, use -b option to extract)
Instance ID                     : uuid:6a1d36c3-456e-f84e-9065-d0fffe87763f
Document ID                     : xmp.did:4BAA366B1B20681188C6BE2E08047DDD
Original Document ID            : adobe:docid:indd:e3c3ba9d-865b-11dd-b6fc-cce9e642b6aa
Rendition Class                 : proof:pdf
Derived From Instance ID        : xmp.iid:4AAA366B1B20681188C6BE2E08047DDD
Derived From Document ID        : xmp.did:2CFA93711E2068118C149D94BBB6A924
Derived From Original Document ID: adobe:docid:indd:e3c3ba9d-865b-11dd-b6fc-cce9e642b6aa
Derived From Rendition Class    : default
History Action                  : saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved, saved
History Instance ID             : xmp.iid:78C503AB68206811A722B72A0BC84C0F, xmp.iid:79C503AB68206811A722B72A0BC84C0F, xmp.iid:7AC503AB68206811A722B72A0BC84C0F, xmp.iid:7BC503AB68206811A722B72A0BC84C0F, xmp.iid:7CC503AB68206811A722B72A0BC84C0F, xmp.iid:7DC503AB68206811A722B72A0BC84C0F, xmp.iid:7EC503AB68206811A722B72A0BC84C0F, xmp.iid:7FC503AB68206811A722B72A0BC84C0F, xmp.iid:391456CB502068119D8CFB4AF82B2296, xmp.iid:3B1456CB502068119D8CFB4AF82B2296, xmp.iid:3F1456CB502068119D8CFB4AF82B2296, xmp.iid:401456CB502068119D8CFB4AF82B2296, xmp.iid:7E4AD82CBC2068119D8CFB4AF82B2296, xmp.iid:824AD82CBC2068119D8CFB4AF82B2296, xmp.iid:834AD82CBC2068119D8CFB4AF82B2296, xmp.iid:01801174072068118209C6184ACF5441, xmp.iid:02801174072068118209C6184ACF5441, xmp.iid:0680117407206811BA67DDABDCF5F87A, xmp.iid:0780117407206811BA67DDABDCF5F87A, xmp.iid:0A80117407206811BA67DDABDCF5F87A, xmp.iid:60130AA710206811BA67DDABDCF5F87A, xmp.iid:63130AA710206811BA67DDABDCF5F87A, xmp.iid:F77F11740720681199EA8C8F681EAA11, xmp.iid:F87F11740720681199EA8C8F681EAA11, xmp.iid:F97F11740720681199EA8C8F681EAA11, xmp.iid:FA7F11740720681199EA8C8F681EAA11, xmp.iid:FB7F11740720681199EA8C8F681EAA11, xmp.iid:FC7F11740720681199EA8C8F681EAA11, xmp.iid:FD7F11740720681199EA8C8F681EAA11, xmp.iid:FE7F11740720681199EA8C8F681EAA11, xmp.iid:FF7F11740720681199EA8C8F681EAA11, xmp.iid:008011740720681199EA8C8F681EAA11, xmp.iid:B673BE5D7B20681199EA8C8F681EAA11, xmp.iid:B773BE5D7B20681199EA8C8F681EAA11, xmp.iid:08F3A5E1F420681199EA8C8F681EAA11, xmp.iid:0EF3A5E1F420681199EA8C8F681EAA11, xmp.iid:0FF3A5E1F420681199EA8C8F681EAA11, xmp.iid:F77F117407206811934E9A6FBD29FA76, xmp.iid:F87F117407206811934E9A6FBD29FA76, xmp.iid:F97F117407206811934E9A6FBD29FA76, xmp.iid:FA7F117407206811934E9A6FBD29FA76, xmp.iid:FB7F117407206811934E9A6FBD29FA76, xmp.iid:018011740720681192B0C34E08416F7A, xmp.iid:028011740720681192B0C34E08416F7A, xmp.iid:038011740720681192B0C34E08416F7A, xmp.iid:048011740720681192B0C34E08416F7A, xmp.iid:01801174072068119109F3C016E378A6, xmp.iid:02801174072068119109F3C016E378A6, xmp.iid:04801174072068119109F3C016E378A6, xmp.iid:05801174072068119109F3C016E378A6, xmp.iid:67567BDA682068119109F3C016E378A6, xmp.iid:0180117407206811B8B9837A83461EAD, xmp.iid:0280117407206811B8B9837A83461EAD, xmp.iid:ED7F1174072068119BADDD8C681FFF8E, xmp.iid:EE7F1174072068119BADDD8C681FFF8E, xmp.iid:EF7F1174072068119BADDD8C681FFF8E, xmp.iid:F07F1174072068119BADDD8C681FFF8E, xmp.iid:D336DBF7782068119BADDD8C681FFF8E, xmp.iid:D436DBF7782068119BADDD8C681FFF8E, xmp.iid:04801174072068119D99A5E9812948D7, xmp.iid:0A801174072068119D99A5E9812948D7, xmp.iid:4AB03C445F2068119D99A5E9812948D7, xmp.iid:4BB03C445F2068119D99A5E9812948D7, xmp.iid:A4CA2D9C7E2068119D99A5E9812948D7, xmp.iid:A9CA2D9C7E2068119D99A5E9812948D7, xmp.iid:ADCA2D9C7E2068119D99A5E9812948D7, xmp.iid:FA7F1174072068118906C836A5E92104, xmp.iid:FB7F1174072068118906C836A5E92104, xmp.iid:FE7F1174072068118906C836A5E92104, xmp.iid:BC7731853B2068118906C836A5E92104, xmp.iid:BD7731853B2068118906C836A5E92104, xmp.iid:BE7731853B2068118906C836A5E92104, xmp.iid:BF7731853B2068118906C836A5E92104, xmp.iid:C07731853B2068118906C836A5E92104, xmp.iid:C17731853B2068118906C836A5E92104, xmp.iid:C27731853B2068118906C836A5E92104, xmp.iid:C37731853B2068118906C836A5E92104, xmp.iid:C47731853B2068118906C836A5E92104, xmp.iid:C57731853B2068118906C836A5E92104, xmp.iid:C67731853B2068118906C836A5E92104, xmp.iid:62E1F1068A2068118906C836A5E92104, xmp.iid:63E1F1068A2068118906C836A5E92104, xmp.iid:64E1F1068A2068118906C836A5E92104, xmp.iid:116B2A5CA1206811B6ED8CEE1D28ADD3, xmp.iid:126B2A5CA1206811B6ED8CEE1D28ADD3, xmp.iid:F77F117407206811853399C08CEFBB07, xmp.iid:F87F117407206811B0DFA2A0CD293944, xmp.iid:FB7F117407206811B0DFA2A0CD293944, xmp.iid:F77F117407206811B8CFF70E6DDB4D01, xmp.iid:F87F117407206811B8CFF70E6DDB4D01, xmp.iid:FA7F117407206811B8CFF70E6DDB4D01, xmp.iid:FB7F117407206811B8CFF70E6DDB4D01, xmp.iid:FC7F117407206811B8CFF70E6DDB4D01, xmp.iid:FD7F117407206811B8CFF70E6DDB4D01, xmp.iid:FE7F117407206811B8CFF70E6DDB4D01, xmp.iid:FF7F117407206811B8CFF70E6DDB4D01, xmp.iid:0080117407206811B8CFF70E6DDB4D01, xmp.iid:E01AD04C92206811B8CFF70E6DDB4D01, xmp.iid:E11AD04C92206811B8CFF70E6DDB4D01, xmp.iid:F97F117407206811B1D0E6BFF8D841BD, xmp.iid:FA7F117407206811B1D0E6BFF8D841BD, xmp.iid:FB7F117407206811B1D0E6BFF8D841BD, xmp.iid:FC7F117407206811B1D0E6BFF8D841BD, xmp.iid:FD7F117407206811B1D0E6BFF8D841BD, xmp.iid:FE7F117407206811B1D0E6BFF8D841BD, xmp.iid:FF7F117407206811B1D0E6BFF8D841BD, xmp.iid:0080117407206811B1D0E6BFF8D841BD, xmp.iid:7CE6512F79206811B1D0E6BFF8D841BD, xmp.iid:7DE6512F79206811B1D0E6BFF8D841BD, xmp.iid:7EE6512F79206811B1D0E6BFF8D841BD, xmp.iid:7FE6512F79206811B1D0E6BFF8D841BD, xmp.iid:CC2E34BB3320681192B0944CC1C3CA1E, xmp.iid:CD2E34BB3320681192B0944CC1C3CA1E, xmp.iid:CE2E34BB3320681192B0944CC1C3CA1E, xmp.iid:CF2E34BB3320681192B0944CC1C3CA1E, xmp.iid:D02E34BB3320681192B0944CC1C3CA1E, xmp.iid:D12E34BB3320681192B0944CC1C3CA1E, xmp.iid:D22E34BB3320681192B0944CC1C3CA1E, xmp.iid:D32E34BB3320681192B0944CC1C3CA1E, xmp.iid:D42E34BB3320681192B0944CC1C3CA1E, xmp.iid:01801174072068118DE7A62DA7D5D004, xmp.iid:02801174072068118DE7A62DA7D5D004, xmp.iid:03801174072068118DE7A62DA7D5D004, xmp.iid:04801174072068118DE7A62DA7D5D004, xmp.iid:05801174072068118DE7A62DA7D5D004, xmp.iid:06801174072068118DE7A62DA7D5D004, xmp.iid:F77F1174072068118BE3A2EE2B4B9A94, xmp.iid:F87F1174072068118BE3A2EE2B4B9A94, xmp.iid:F97F1174072068118BE3A2EE2B4B9A94, xmp.iid:FA7F1174072068118BE3A2EE2B4B9A94, xmp.iid:F77F1174072068119FEEED2328480636, xmp.iid:F87F1174072068119FEEED2328480636, xmp.iid:F97F1174072068119FEEED2328480636, xmp.iid:FA7F1174072068119FEEED2328480636, xmp.iid:FB7F1174072068119FEEED2328480636, xmp.iid:FC7F1174072068119FEEED2328480636, xmp.iid:FD7F1174072068119FEEED2328480636, xmp.iid:FE7F1174072068119FEEED2328480636, xmp.iid:FF7F1174072068119FEEED2328480636, xmp.iid:00801174072068119FEEED2328480636, xmp.iid:06F7EC26512068119FEEED2328480636, xmp.iid:07F7EC26512068119FEEED2328480636, xmp.iid:08F7EC26512068119FEEED2328480636, xmp.iid:09F7EC26512068119FEEED2328480636, xmp.iid:0AF7EC26512068119FEEED2328480636, xmp.iid:0BF7EC26512068119FEEED2328480636, xmp.iid:0CF7EC26512068119FEEED2328480636, xmp.iid:0DF7EC26512068119FEEED2328480636, xmp.iid:018011740720681191099BC29A54DBD3, xmp.iid:028011740720681191099BC29A54DBD3, xmp.iid:038011740720681191099BC29A54DBD3, xmp.iid:048011740720681191099BC29A54DBD3, xmp.iid:058011740720681191099BC29A54DBD3, xmp.iid:068011740720681191099BC29A54DBD3, xmp.iid:078011740720681191099BC29A54DBD3, xmp.iid:088011740720681191099BC29A54DBD3, xmp.iid:098011740720681191099BC29A54DBD3, xmp.iid:0A8011740720681191099BC29A54DBD3, xmp.iid:0280117407206811B1A495791501A103, xmp.iid:0380117407206811B1A495791501A103, xmp.iid:0580117407206811B1A495791501A103, xmp.iid:0780117407206811B1A495791501A103, xmp.iid:0980117407206811B1A495791501A103, xmp.iid:1AD3DE5A83206811B1A495791501A103, xmp.iid:1CD3DE5A83206811B1A495791501A103, xmp.iid:F77F11740720681191FAC4188F855E27, xmp.iid:FB7F11740720681191FAC4188F855E27, xmp.iid:FC7F11740720681191FAC4188F855E27, xmp.iid:FD7F11740720681191FAC4188F855E27, xmp.iid:FE7F11740720681191FAC4188F855E27, xmp.iid:FF7F11740720681191FAC4188F855E27, xmp.iid:008011740720681191FAC4188F855E27, xmp.iid:F97F1174072068119BF28D6F147B3D8E, xmp.iid:FB7F1174072068119BF28D6F147B3D8E, xmp.iid:F87F11740720681192B0D83898469DDA, xmp.iid:F97F11740720681192B0D83898469DDA, xmp.iid:FA7F11740720681192B0D83898469DDA, xmp.iid:FC7F11740720681192B0D83898469DDA, xmp.iid:FC7F117407206811A791C2E80EE53E78, xmp.iid:058011740720681188C6C0C2A71FD2AD, xmp.iid:068011740720681188C6C0C2A71FD2AD, xmp.iid:078011740720681188C6C0C2A71FD2AD, xmp.iid:018011740720681192B0B67D4F38C3DC, xmp.iid:0A8011740720681192B0B67D4F38C3DC, xmp.iid:FA7F117407206811A63ADEAFEB523233, xmp.iid:FB7F117407206811A63ADEAFEB523233, xmp.iid:FC7F117407206811A63ADEAFEB523233, xmp.iid:FD7F117407206811A63ADEAFEB523233, xmp.iid:FE7F117407206811A63ADEAFEB523233, xmp.iid:FF7F117407206811A63ADEAFEB523233, xmp.iid:0080117407206811A63ADEAFEB523233, xmp.iid:AF0670A9B8206811A63ADEAFEB523233, xmp.iid:B00670A9B8206811A63ADEAFEB523233, xmp.iid:F77F117407206811B1C1BFD89F4AAA4C, xmp.iid:F87F117407206811B1C1BFD89F4AAA4C, xmp.iid:F97F117407206811B1C1BFD89F4AAA4C, xmp.iid:0580117407206811910999112586E600, xmp.iid:0780117407206811910999112586E600, xmp.iid:84E243FD41206811910999112586E600, xmp.iid:86E243FD41206811910999112586E600, xmp.iid:88E243FD41206811910999112586E600, xmp.iid:F87F11740720681188C6C2BC02575663, xmp.iid:F97F11740720681188C6C2BC02575663, xmp.iid:0180117407206811871F9414368FFF8A, xmp.iid:0280117407206811871F9414368FFF8A, xmp.iid:0380117407206811871F9414368FFF8A, xmp.iid:0480117407206811871F9414368FFF8A, xmp.iid:0580117407206811871F9414368FFF8A, xmp.iid:0680117407206811871F9414368FFF8A, xmp.iid:0780117407206811871F9414368FFF8A, xmp.iid:0880117407206811871F9414368FFF8A, xmp.iid:0980117407206811871F9414368FFF8A, xmp.iid:0A80117407206811871F9414368FFF8A, xmp.iid:B8E365D52B206811871F9414368FFF8A, xmp.iid:B9E365D52B206811871F9414368FFF8A, xmp.iid:BAE365D52B206811871F9414368FFF8A, xmp.iid:038011740720681192B097AD857B9D5E, xmp.iid:048011740720681192B097AD857B9D5E, xmp.iid:058011740720681192B097AD857B9D5E, xmp.iid:0A8011740720681192B097AD857B9D5E, xmp.iid:883A1C883C20681192B097AD857B9D5E, xmp.iid:893A1C883C20681192B097AD857B9D5E, xmp.iid:8A3A1C883C20681192B097AD857B9D5E, xmp.iid:8B3A1C883C20681192B097AD857B9D5E, xmp.iid:8C3A1C883C20681192B097AD857B9D5E, xmp.iid:8D3A1C883C20681192B097AD857B9D5E, xmp.iid:8E3A1C883C20681192B097AD857B9D5E, xmp.iid:FA7F1174072068119109B3229D4371B4, xmp.iid:FB7F1174072068119109B3229D4371B4, xmp.iid:F97F1174072068118C149D94BBB6A924, xmp.iid:FA7F1174072068118C149D94BBB6A924, xmp.iid:FB7F1174072068118C149D94BBB6A924, xmp.iid:27FA93711E2068118C149D94BBB6A924, xmp.iid:28FA93711E2068118C149D94BBB6A924, xmp.iid:29FA93711E2068118C149D94BBB6A924, xmp.iid:2AFA93711E2068118C149D94BBB6A924, xmp.iid:2BFA93711E2068118C149D94BBB6A924, xmp.iid:2CFA93711E2068118C149D94BBB6A924, xmp.iid:E81F4030492068118C149D94BBB6A924, xmp.iid:E91F4030492068118C149D94BBB6A924, xmp.iid:EA1F4030492068118C149D94BBB6A924, xmp.iid:EB1F4030492068118C149D94BBB6A924, xmp.iid:0180117407206811871FA86E140ABC40, xmp.iid:FF7F11740720681188C6BE2E08047DDD, xmp.iid:008011740720681188C6BE2E08047DDD, xmp.iid:46AA366B1B20681188C6BE2E08047DDD, xmp.iid:47AA366B1B20681188C6BE2E08047DDD, xmp.iid:48AA366B1B20681188C6BE2E08047DDD, xmp.iid:49AA366B1B20681188C6BE2E08047DDD, xmp.iid:4AAA366B1B20681188C6BE2E08047DDD, xmp.iid:4BAA366B1B20681188C6BE2E08047DDD, xmp.iid:62C8DAD73720681188C6BE2E08047DDD, xmp.iid:FF7F11740720681195FECCAA251A9B1A, xmp.iid:008011740720681195FECCAA251A9B1A, xmp.iid:B452DE721420681195FECCAA251A9B1A, xmp.iid:B552DE721420681195FECCAA251A9B1A, xmp.iid:F77F117407206811A00587A5AD016718, xmp.iid:F87F117407206811A00587A5AD016718, xmp.iid:FB7F117407206811A00587A5AD016718
History When                    : 2010:01:18 09:42:44+01:00, 2010:01:18 09:42:44+01:00, 2010:01:18 09:45:55+01:00, 2010:01:18 09:55:34+01:00, 2010:01:18 10:00:35+01:00, 2010:01:18 11:32:54+01:00, 2010:01:18 12:24:24+01:00, 2010:01:18 12:29:28+01:00, 2010:01:20 16:16:52+01:00, 2010:01:20 18:36:46+01:00, 2010:01:21 16:05:28+01:00, 2010:01:21 16:22:06+01:00, 2010:01:21 17:20:17+01:00, 2010:01:21 18:20+01:00, 2010:01:21 18:24:32+01:00, 2010:01:22 14:57:35+01:00, 2010:01:22 15:10:19+01:00, 2010:02:11 14:18:16+01:00, 2010:02:11 14:19:46+01:00, 2010:02:11 14:35:19+01:00, 2010:02:11 15:02:45+01:00, 2010:02:11 15:05:42+01:00, 2010:04:19 18:23:52+02:00, 2010:04:19 18:32:53+02:00, 2010:04:20 10:03:22+02:00, 2010:04:20 10:04:09+02:00, 2010:04:20 10:10:18+02:00, 2010:04:20 10:24:48+02:00, 2010:04:20 10:34:57+02:00, 2010:04:20 10:39:35+02:00, 2010:04:20 10:46:15+02:00, 2010:04:20 10:51:58+02:00, 2010:04:21 10:49:56+02:00, 2010:04:21 15:12:17+02:00, 2010:04:22 15:51:38+02:00, 2010:04:23 17:19:25+02:00, 2010:04:23 17:36:03+02:00, 2010:05:03 15:33:50+02:00, 2010:05:03 15:41:35+02:00, 2010:05:03 16:17:56+02:00, 2010:05:03 16:25:29+02:00, 2010:05:03 16:39+02:00, 2010:05:04 10:13:47+02:00, 2010:05:04 10:28:27+02:00, 2010:05:04 11:05:50+02:00, 2010:05:04 11:11:33+02:00, 2010:05:05 12:51:28+02:00, 2010:05:05 12:54:42+02:00, 2010:05:05 14:15:57+02:00, 2010:05:05 14:15:57+02:00, 2010:05:06 16:43:03+02:00, 2010:05:13 14:10:32+02:00, 2010:05:13 14:19:05+02:00, 2010:05:13 18:20:27+02:00, 2010:05:13 18:27:43+02:00, 2010:05:13 18:36:15+02:00, 2010:05:14 09:49:13+02:00, 2010:05:17 16:49:05+02:00, 2010:05:17 17:08:22+02:00, 2011:09:26 10:41:33+02:00, 2011:09:26 15:13:57+02:00, 2011:09:26 17:23:49+02:00, 2011:09:26 18:13:28+02:00, 2011:09:27 12:56:31+02:00, 2011:09:27 18:21:28+02:00, 2011:09:30 09:35:54+02:00, 2011:10:25 12:15:42+02:00, 2011:10:25 13:50:16+02:00, 2011:10:25 14:36:12+02:00, 2011:10:25 17:46:06+02:00, 2011:10:25 17:53:17+02:00, 2011:10:25 18:52:02+02:00, 2011:10:25 19:04:22+02:00, 2011:10:25 19:07:23+02:00, 2011:10:26 13:47:28+02:00, 2011:10:26 14:03:31+02:00, 2011:10:26 14:26:13+02:00, 2011:10:26 15:18:15+02:00, 2011:10:26 15:29:31+02:00, 2011:10:26 16:51:27+02:00, 2011:10:26 17:12:49+02:00, 2011:10:26 17:37:33+02:00, 2011:10:26 17:45:35+02:00, 2011:11:16 10:24:29+01:00, 2011:11:16 11:02:57+01:00, 2011:11:16 16:41:14+01:00, 2011:11:17 18:13:42+01:00, 2011:11:17 18:35:35+01:00, 2011:11:18 18:52:07+01:00, 2011:11:21 13:04:58+01:00, 2011:11:21 15:03:29+01:00, 2011:11:22 10:27:51+01:00, 2011:11:22 10:36:47+01:00, 2011:11:22 10:54:50+01:00, 2011:11:22 14:44:32+01:00, 2011:11:22 16:51:16+01:00, 2011:11:22 16:53:41+01:00, 2011:11:22 17:12:18+01:00, 2011:11:22 17:28:14+01:00, 2011:11:23 18:19:46+01:00, 2011:11:24 11:04:35+01:00, 2011:11:24 15:28:05+01:00, 2011:11:24 18:41:37+01:00, 2011:11:25 12:09:53+01:00, 2011:11:25 12:12:03+01:00, 2011:11:25 12:50:48+01:00, 2011:11:25 13:08:36+01:00, 2011:11:25 13:08:51+01:00, 2011:11:25 14:34:03+01:00, 2011:11:25 15:35:08+01:00, 2011:11:25 15:37:51+01:00, 2011:11:29 10:07:56+01:00, 2011:11:29 11:42:33+01:00, 2011:11:29 11:43:37+01:00, 2011:11:29 11:43:37+01:00, 2011:11:29 12:56:15+01:00, 2011:11:29 13:15:20+01:00, 2011:11:29 14:23:47+01:00, 2011:11:29 14:25:03+01:00, 2011:11:29 14:25:04+01:00, 2011:12:15 11:17:35+01:00, 2011:12:15 11:39:04+01:00, 2011:12:15 12:01:23+01:00, 2011:12:15 12:02:45+01:00, 2011:12:15 12:34:04+01:00, 2011:12:15 12:44:01+01:00, 2011:12:16 11:01:02+01:00, 2011:12:16 11:35:34+01:00, 2011:12:16 12:22:12+01:00, 2011:12:16 18:07:48+01:00, 2011:12:19 15:08:10+01:00, 2011:12:19 17:38:09+01:00, 2011:12:19 18:07:07+01:00, 2011:12:20 10:57:04+01:00, 2011:12:20 11:23:21+01:00, 2011:12:20 12:13:04+01:00, 2011:12:20 12:50:10+01:00, 2011:12:20 12:52:53+01:00, 2011:12:20 12:52:53+01:00, 2011:12:20 13:26:44+01:00, 2011:12:20 14:07:34+01:00, 2011:12:20 14:07:58+01:00, 2011:12:20 15:25:51+01:00, 2011:12:20 17:24:31+01:00, 2011:12:20 17:25:29+01:00, 2011:12:20 17:27:49+01:00, 2011:12:20 17:42:05+01:00, 2011:12:20 17:42:59+01:00, 2011:12:21 11:50:42+01:00, 2011:12:21 12:00:58+01:00, 2011:12:21 12:24:18+01:00, 2011:12:21 12:37:09+01:00, 2011:12:21 13:14:02+01:00, 2011:12:21 17:12:51+01:00, 2011:12:21 18:02:09+01:00, 2011:12:21 18:05:32+01:00, 2011:12:21 18:06:20+01:00, 2011:12:21 18:06:20+01:00, 2012:04:05 10:55:59+02:00, 2012:04:05 10:59:42+02:00, 2012:04:05 12:15:09+02:00, 2012:04:05 16:18:13+02:00, 2012:04:05 18:13:42+02:00, 2012:04:06 16:47:44+02:00, 2012:04:06 17:12:44+02:00, 2012:04:11 14:56:08+02:00, 2012:04:11 15:48:50+02:00, 2012:04:11 16:39:31+02:00, 2012:04:11 17:03:21+02:00, 2012:04:11 17:04:47+02:00, 2012:04:11 17:05:47+02:00, 2012:04:11 17:05:47+02:00, 2012:04:12 13:29:42+02:00, 2012:04:12 13:36:14+02:00, 2012:10:22 10:13:43+02:00, 2012:10:22 11:55:32+02:00, 2012:10:22 12:24:39+02:00, 2012:10:22 16:10:18+02:00, 2012:10:26 17:47:57+02:00, 2012:11:13 13:07:41+01:00, 2012:11:13 14:06+01:00, 2012:11:13 15:09:56+01:00, 2012:11:14 16:04:53+01:00, 2012:11:14 19:00:08+01:00, 2012:11:15 10:39:55+01:00, 2012:11:15 10:40:54+01:00, 2012:11:15 10:40:54+01:00, 2012:11:15 11:21:38+01:00, 2012:11:15 11:22:52+01:00, 2012:11:15 11:23:36+01:00, 2012:11:15 12:05:15+01:00, 2012:11:19 15:35:47+01:00, 2012:11:19 15:50:52+01:00, 2012:11:19 17:03:37+01:00, 2012:11:19 17:15:24+01:00, 2012:11:19 17:27:26+01:00, 2012:11:20 13:05:48+01:00, 2012:11:20 13:10+01:00, 2012:11:20 17:14:31+01:00, 2012:11:20 17:49:02+01:00, 2012:11:20 18:57:57+01:00, 2012:11:21 17:19:19+01:00, 2012:11:21 17:21:33+01:00, 2012:11:22 11:08:33+01:00, 2012:11:22 11:35:27+01:00, 2012:11:22 11:39:33+01:00, 2012:11:22 11:42:01+01:00, 2012:11:22 11:49:59+01:00, 2012:11:22 12:12:42+01:00, 2012:11:22 12:14:48+01:00, 2012:11:22 14:48:43+01:00, 2012:11:22 15:25:50+01:00, 2012:11:22 15:27:59+01:00, 2012:11:22 15:28:58+01:00, 2012:11:22 15:29:17+01:00, 2012:11:22 15:29:17+01:00, 2012:11:23 09:28:25+01:00, 2012:11:23 09:38:53+01:00, 2012:11:23 09:43:21+01:00, 2012:11:23 12:59:46+01:00, 2012:11:23 15:28:54+01:00, 2012:11:23 16:09:22+01:00, 2012:11:23 16:30:44+01:00, 2012:11:23 16:31:13+01:00, 2012:11:23 16:38:30+01:00, 2012:11:23 16:38:39+01:00, 2012:11:23 16:40:45+01:00, 2012:11:23 17:37:52+01:00, 2012:11:23 17:40:01+01:00, 2012:11:26 11:49:46+01:00, 2012:11:26 11:52:30+01:00, 2012:11:26 11:59:46+01:00, 2012:11:26 14:22:40+01:00, 2012:11:26 14:22:40+01:00, 2012:11:26 14:22:58+01:00, 2012:11:26 14:24:17+01:00, 2012:11:26 14:24:36+01:00, 2012:11:26 14:24:36+01:00, 2012:11:26 18:09:39+01:00, 2012:11:26 18:13:14+01:00, 2012:11:26 18:16:38+01:00, 2012:11:26 18:18:26+01:00, 2012:12:14 10:10:19+01:00, 2013:01:29 12:55:53+01:00, 2013:01:29 12:57:46+01:00, 2013:01:29 13:02:28+01:00, 2013:01:29 13:07:41+01:00, 2013:01:29 13:51:25+01:00, 2013:01:29 13:54:07+01:00, 2013:01:29 13:54:52+01:00, 2013:01:29 13:54:52+01:00, 2013:01:29 17:42:19+01:00, 2013:06:10 16:35:50+02:00, 2013:06:10 16:43:33+02:00, 2013:06:10 16:44:58+02:00, 2013:06:10 16:46:20+02:00, 2013:06:14 12:08:06+02:00, 2013:06:14 12:09:30+02:00, 2013:06:14 12:41:23+02:00
History Software Agent          : Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0, Adobe InDesign 6.0
History Changed                 : /, /metadata, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /metadata, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /metadata, /;/metadata, /, /, /, /metadata, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /metadata, /;/metadata, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /metadata, /, /, /, /, /, /, /, /, /, /, /, /, /, /metadata, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /metadata, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /metadata, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /, /metadata, /;/metadata, /, /, /metadata, /, /, /, /, /, /, /, /, /, /, /, /, /metadata, /, /, /, /, /, /, /, /, /
Manifest Link Form              : ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream
Manifest Placed X Resolution    : 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 300.00, 72.00, 180.00, 72.00, 180.00, 300.00, 180.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 300.00, 300.00, 300.00, 72.00, 180.00, 72.00, 180.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00
Manifest Placed Y Resolution    : 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 300.00, 72.00, 180.00, 72.00, 180.00, 300.00, 180.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 300.00, 300.00, 300.00, 72.00, 180.00, 72.00, 180.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00
Manifest Placed Resolution Unit : Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches
Manifest Reference Instance ID  : uuid:f5d8e17a-6363-11df-824c-00112437f484, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:0874192e-3ec4-224c-8722-f0dfdec48631, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:C47219A248EFDC11A4A1E704FA923036, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:C47219A248EFDC11A4A1E704FA923036, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, xmp.iid:C656BA9580206811B6E2F5A7912FDF45, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, xmp.iid:090B01C76E226811B4A7E490DFA0546F, uuid:E3BE151DB1D6DE1196008D7E915F331F, uuid:8BD0846AB2D6DE1196008D7E915F331F, uuid:d4dd54c9-ca4b-11da-a027-d979a7b9ea11, uuid:FAB625A7B9D6DE1196008D7E915F331F, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:6d10fb0d-3773-2f43-a172-8a2b9919f0e9, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:9ebe73d5-3a40-8841-a696-afd60b6cc4ef, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, xmp.iid:4465603B4F20681188C6BB677E4A850E, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:958b3718-423b-11da-b86e-bf20ec40c5f1, uuid:eaa8f5b7-02a8-11db-84b2-eb5e03a573d1, uuid:c1ebd737-2375-11da-8b8d-cf95e0f15102, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, xmp.iid:090B01C76E226811B4A7E490DFA0546F, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:CAE2489B11E8DE11A28C9104D7B2C53A, xmp.iid:D2A7D7443A20681197A5DE51F63AF961, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:e7090fba-ebb8-a845-adaf-3f4cf72d9f09, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:077f7546-c8d0-3644-be63-5f34055e2b06, uuid:f5d8e17a-6363-11df-824c-00112437f484
Manifest Reference Document ID  : uuid:F9AD8E2264BA11DF86ABAD673B37D118, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:F77F117407206811BD4DD844042CD875, xmp.did:0280117407206811871FA5A4E93130E1, uuid:C37219A248EFDC11A4A1E704FA923036, xmp.did:0280117407206811871FA5A4E93130E1, uuid:C37219A248EFDC11A4A1E704FA923036, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:C556BA9580206811B6E2F5A7912FDF45, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:ED7ACC3B65226811B4A7E490DFA0546F, uuid:E2BE151DB1D6DE1196008D7E915F331F, uuid:729F56A418FFDC119FF7EA092DE38A63, adobe:docid:photoshop:d4dd54c1-ca4b-11da-a027-d979a7b9ea11, uuid:E19DC935B8D6DE1196008D7E915F331F, xmp.did:0280117407206811871FA5A4E93130E1, uuid:FABE4238268D11DD8E3B866E1C63CE7B, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:038011740720681192B0FAD1636F3DAA, xmp.did:0280117407206811871FA5A4E93130E1, uuid:3C128BC866E0DE11800E8BB4AB8F2CBD, xmp.did:0280117407206811871FA5A4E93130E1, adobe:docid:photoshop:c1ebd731-2375-11da-8b8d-cf95e0f15102, adobe:docid:photoshop:eaa8f5b3-02a8-11db-84b2-eb5e03a573d1, adobe:docid:photoshop:ef987683-2373-11da-8b8d-cf95e0f15102, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:ED7ACC3B65226811B4A7E490DFA0546F, xmp.did:0280117407206811871FA5A4E93130E1, uuid:C9E2489B11E8DE11A28C9104D7B2C53A, uuid:B8E9BD720FD7DE1196008D7E915F331F, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, uuid:35bf5dfc-6b99-11e2-0000-fc64eff0f6b9, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, xmp.did:0280117407206811871FA5A4E93130E1, uuid:F9AD8E2264BA11DF86ABAD673B37D118
Doc Change Count                : 92091
Format                          : application/pdf
Producer                        : Adobe PDF Library 9.0
Trapped                         : False
Page Count                      : 64
EXIF Metadata provided by EXIF.tools
FCC ID Filing: YQH-BTSWEMG2

Navigation menu