Berthold Technologies FCC01X01 Moisture Measuring System User Manual
Berthold Technologies Moisture Measuring System
User Manual
Process Control detect and identify Concentration Meter Micro-Polar Brix LB 565 Hardware Manual User’s Guide ID No. 39531 BA2 Rev. No.: 00 01/11/04 Soft. Version: 1.00 The units supplied should not be repaired by anyone other than Berthold Service engineers or technicians authorized by Berthold. In case of operation trouble, please address our central service department. The complete user’s guide consists of two manuals, the hardware description and the software description. The hardware manual comprises: mechanical components assembly electrical installation technical data electrical and mechanical drawings The software manual comprises: operation of the evaluation unit parameter description basic setting calibration error messages The present manual is the hardware description. Subject to change without prior notice. Table of Contents Table of Contents Page Chapter 1. Safety Summary 1.1 Symbol and Pictograms 1.2 Use and Function 1.3 General Safety Instructions Chapter 2. General Information 2.1 Intended Use 2.2 Definitions Chapter 3. System Description Principle of Measurement Calculation of Measured Values Temperature Compensation Mechanical Components 3.4.1 The Evaluation Unit 3.4.2 The Microwave Probes 3.4.3 High-frequency Cable 3.5 Measurement Configuration 10 13 13 14 15 3.1 3.2 3.3 3.4 15 16 17 19 20 22 25 26 Chapter 4. Getting Started 27 4.1 Transport to the Installation Site 4.2 Installation 4.2.1 Pan Probe Installation 4.2.2 Setting up the Evaluation Unit 4.3 Connecting the Evaluation Unit 4.3.1 Connecting the HF Cable 4.3.2 Pin Configuration of the Connector Strip 4.3.3 Digital Outputs, Relays Chapter 5. Service Instructions 5.1 5.2 5.3 5.4 General Information Instrument Cleaning Batteries Fuse Replacement 27 27 27 29 30 30 32 37 38 38 38 38 39 Chapter 6. Technical Data 40 6.1 Evaluation Unit 6.2 Probes 40 43 Chapter 7. Certificate 7.1 EC-Declaration of Conformity 7.2 Frequency approval Chapter 8. Engineering drawings 8.1 Dimensional drawing for housing Micro-Polar Brix LB 565 44 44 45 47 47 Table of Contents 8.2 Electrical Wiring Diagram 8.3 Dimensional drawings for batch pan probes 8.3.1 Type LB 5650-01 8.3.2 Type LB 5650-02 8.3.3 Type LB 5650-03 8.3.4 Type LB 5650-04 8.3.5 Type LB 5650-05 8.4 Dimensional drawings for continuous pan probes 8.4.1 Type LB 5651-01 8.4.2 Type LB 5651-02 8.4.3 Type LB 5651-03 8.4.4 Type LB 5651-04 8.4.5 Type LB 5651-05 8.5 Installation sheets for LB 5650 (Batch pan probes) 8.6 Installation sheets for LB 5651 (Continuous pan probes) 48 49 49 50 51 52 53 54 54 55 56 57 58 59 61 Micro-Polar Brix LB 565 Chapter 1 Safety Summary Chapter 1. Safety Summary 1.1 Symbol and Pictograms The following symbols identify safety instructions in this user’s guide: Danger! Possible danger of life and health hazard Caution! Possible hazard Minor personal injuries Warning! Possible hazard Property damages Note! Tips for application and useful information The safety instructions are supplemented by explanatory pictograms, for example: Micro-Polar Brix LB 565 Chapter 1 Safety Summary 1.2 Use and Function FrequencyApprovals The Micro-Polar fulfils the regulations of FCC1, Part 15 and the regulations of the Canadian Industry according to RSS-210. This measuring system complies with interference immunity and radiated interference and is permissible for operation. FCCApproval Sign “IC2:” before the equipment certification number signifies that the Industry Canada technical specifications were met. It does not guarantee that the certified product will operate to the user’s satisfaction. The Micro-Polar Brix complies with the R&TTE regulations 1999/5/EG and fulfils herein all requirements for such high frequency systems. The Micro-Polar Brix contains the identification of conformity according to CE symbol of No. 0682 of the certification office. The certificate can be found in chapter 7.2 Frequency approval. FCC … Federal Communications Commission IC ... Industry Canada Micro-Polar Brix LB 565 Chapter 1 Safety Summary Note! The Micro-Polar Brix is a system for concentration measurement using microwave technology. The microwave probe is immersed into the product being measured. The emitted microwaves have a very low power (< 0.1 mW) and are, therefore, not at all hazardous to human beings or the environment. Also, the product is not altered at all. Danger! The Micro-Polar Brix has been manufactured in compliance with the safety requirements for microwave devices. If special legal provisions exist regarding the use of microwaves, it will be the responsibility of the user to adhere to them. Any change in frequency or any other manipulation on the microwave device will result in a loss of the frequency license and will be prosecuted. The microwave modules do not include any replaceable components and must not be opened. Micro-Polar Brix LB 565 Chapter 1 Safety Summary 1.3 General Safety Instructions Note! The Micro-Polar Brix has been manufactured in accordance with the best available technology and in compliance with acknowledged safety rules to ensure the highest level of occupational safety. The instrument housing is protected according to protection type IP 65 and is suitable for outdoor application. The instrument has been tested by the manufacturer and is delivered in a condition that allows safe and reliable operation. Caution! The safety instructions and warnings in this user’s guide have to be observed without fail to ensure safe operation of the instrument! The system may only be used in technically good order and only according to regulations! Only authorized persons may work with the system who have been assigned to do this and who have the proper qualification and have received the necessary instructions! Installations and modifications on the system which could affect the operational safety are not permitted! Warning! Ambient conditions: All system components require non-corrosive ambient conditions during transport, storage and starting up. 10 Micro-Polar Brix LB 565 Chapter 1 Safety Summary Danger! Electrical shock hazard: Disconnect power to ensure that contact with live part is avoided during installation and when servicing. Disconnect power before opening the instrument. Work on open and live instruments is prohibited. Warning! Spare fuses must match the rating specified by the device manufacturer. Short-circuiting or manipulation is not permitted. Warning! The LB 565 and all ancillary units have to be connected to mains via grounded connection. Caution! If liquid gets inside the instrument, cut off the power supply. The instrument has to be checked and cleaned by an authorized service center. Warning! Never change the installation and the parameter settings without full knowledge of these operating instructions, as well as full knowledge of the behavior of the connected controller and the possible influence on the operating process to be controlled. Micro-Polar Brix LB 565 11 12 Micro-Polar Brix LB 565 Chapter 2 General Information Chapter 2. General Information 2.1 Intended Use The measuring system LB 565 can be used to determine the concentration of nearly all materials which can be dissolved or suspended in water using microwave technology. Pan probes with and without rinsing device are available. The pan probes have been designed for installation in pipelines with a nominal width of ≥ 200 mm and in containers, for example, evaporation crystallizers. The probe is installed so that both measuring rods (sender and receiver) are immersed into the product being measured. During operation, the concentration meter Micro-Polar Brix sends out electromagnetic radiation in a frequency range between 2.4 GHz and 2.5 GHz (range restrictions depend on local regulations in your Country). The microwaves emerge only between both measuring rods of the probe, and are not dangerous to human beings and the environment (power emission < 0.1 mW). The microwaves are emitted targeted from the microwave window; the product is not altered by the microwaves. To ensure proper function of the meter, please pay attention to the following: Note! The material being measured must not be electrically conduc- tive, i.e. the ohmic resistance is infinite. The product must not contain any gas bubbles, or gas bubbles have to be compressed with adequate pressure when carrying out measurements in pipelines. The ion concentration, e.g. salt content, has to be near constant. Micro-Polar Brix LB 565 13 Chapter 2 General Information 2.2 Definitions Attenuation Weakening of the microwave signals, microwaves measurement effect. Batch pan probe Pan probe without rinsing device. Application e.g. in evaporation crystallization process. Continuous probe 14 pan Pan probe with rinsing device. Application e.g. in continuous evaporation crystallization process. Factory setting All parameters have been set by the manufacturer using standard values. In most cases this simplifies calibration of the instrument significantly. Despite factory setting, calibration always has to be performed. HF cable High-frequency cable. Microwaves Term for electromagnetic waves in a certain frequency range. Phase Phase or phase shift. Microwave measurement effect. Quad cable Combination of four HF cables of equal length in a corrugated tube. Softkeys Buttons associated with the software. TC Temperature compensation. Micro-Polar Brix LB 565 Chapter 3 System Description Chapter 3. System Description 3.1 Principle of Measurement The microwaves that spread between the rods transmit the product being measured; their propagation speed is slowed down (= phase shift) and their intensity is damped (= attenuation). Figure 1 illustrates the principle of measurement: The propagation speed of microwaves passing through the product being measured is slowed down (phase shift) and their intensity (attenuation) is reduced, relative to a reference signal. Reference signal HF sources Reference path Sender Phase comparison --> Phase Receiver Product Amplitude comp. --> Attenuation Measured value: Concentration %TS Measurement signal Phase shift Figure 1: Schematic diagram: Change of microwave by product Prerequisite is that the product being measured shows some dielectric properties. In sugar strike, it is the water which presents a very distinct dielectric fluid. The water or dry mass concentration can therefore be determined by measuring the phase shift and/or attenuation. The concentration to be detected in the product depends in good approximation linear on phase shift and attenuation. For this reason we can measure the concentration or the dry mass content of the sugar solution using a linear calibration (see chapter 3.2 Calculation of Measured Values). Micro-Polar Brix LB 565 15 Chapter 3 System Description 3.2 Calculation of Measured Values The microwave parameters phase and attenuation are calibrated according to the plausibility analysis. Linear calibration: K = A ⋅ ϕ + B ⋅D + C Quadratic calibration: K = A1 ⋅ ϕ + A 2 ⋅ ϕ ⋅ arg(ϕ) + B1 ⋅ D + B2 ⋅ D ⋅ arg(D) + C A, B, C ϕ ϕ, D arg(ϕ) , arg(D) Concentration Coefficients of respective calibration function Phase Attenuation Rate of phase, attenuation algebraic sign of phase, attenuation The formula is simplified accordingly for a pure phase or attenuation measurement. In case of varying measurement parameters (product density, temperature, etc.), the microwave phase shift method yields a much better accuracy compared to the microwave attenuation method. The phase measurement is significantly less affected by material parameters such as temperature or salt contents (electrolytic conductivity), bubbles and grain size than an attenuation measurement. Note! In most cases we recommend using a pure phase measurement to determine the concentration of sugar strike. It is calculated in good approximation through linear calibration: K = A ⋅ϕ+ C Micro-Polar Brix allows you to calibrate, display and output two concentrations K1 and K2. You have to enter the calibration coefficients separately for concentration 1 and 2. For more information please refer to the Software Manual. 16 Micro-Polar Brix LB 565 Chapter 3 System Description 3.3 Temperature Compensation Temperature compensation (TC) has to be carried out if varying operating states are likely to occur, for example vapour pressure and product temperature. TC has to be carried out whenever you are working with cooling crystallizers. The following formulas are used for temperature compensation. Linear compensation ϕ comp = ϕmeas + TC ϕ ⋅ ∆θ Additive: D comp = Dmeas + TCD ⋅ ∆θ Multiplicative: ϕ comp = ϕmeas (1 + TC ϕ ⋅ ∆θ) D comp = Dmeas (1 + TCD ⋅ ∆θ) Quadratic compensation ϕ comp = ϕmeas + TC ϕ1 ⋅ ∆θ + TC ϕ2 ⋅ ∆θ ⋅ arg(∆θ ) D comp = Dmeas + TCD1 ⋅ ∆θ + TCD2 ⋅ ∆θ ⋅ arg(∆θ ) Additive: Multiplicative: ϕ comp = ϕmeas (1 + TC ϕ1 ⋅ ∆θ + TC ϕ 2 ⋅ ∆θ ⋅ arg(∆θ)) D comp = Dmeas (1 + TCD1 ⋅ ∆θ + TCD2 ⋅ ∆θ ⋅ arg(∆θ )) Where ϕmeas ϕcomp Dmeas Dmeas TCϕx TCDx ∆θ = measured temperature (Tmeas ) – reference temp. (TRef) ∆θ measured phase [°] compensated phase [°] measured attenuation [dB] compensated attenuation [dB] temperature coefficient [° / °C] temperature coefficient [dB/ °C] = difference temperature arg(∆θ) = algebraic sign of difference temperature Depending on the selected function (additive, multiplicative, quadratic, cubic), the required temperature coefficients appear on the Calibration menu. Temperature coefficients that are not used are set to zero. Micro-Polar Brix LB 565 17 Chapter 3 System Description If you select two-range calibration (split concentration), separate TC’s have to be entered for both concentration ranges. The coefficients are entered in the course of calibration. TC can be carried out via Pt100 or via current input. This has to be defined on the Calibration menu. The Pt100 temperature range is between –50 °C and +200 °C. How to work with the temperature compensation is described in detail in the Software Manual. Note! In the most cases, linear multiplicative temperature compensation in good approximation is the correct mode of calculation, e.g. for cooling crystallizers. 18 Micro-Polar Brix LB 565 Chapter 3 System Description 3.4 Mechanical Components The measuring system comprises an evaluation unit (see Figure 2), one probe and one set of special high-frequency (HF) cables. The probes are available in different versions, as pan probe with and without rinsing device (see Figure 3 and Figure 4). Figure 2: Evaluation unit Micro-Polar Brix LB 565 Figure 3: Batch pan probe LB 5650 Figure 4: Continuous pan probe LB 5651 Micro-Polar Brix LB 565 19 Chapter 3 System Description 3.4.1 The Evaluation Unit The evaluation unit comprises the evaluation computer with microwave unit. The microwaves are generated, received and analyzed in the microwave unit. Signal processing and communication take place in the evaluation computer. For simple operation, the measuring system includes a display, 4 softkeys and an alphanumeric keypad. Different functions are assigned to the softkeys on the display. An RS232 interface is included on the underside of the instrument. Micro-Polar LCD display Lock Softkey buttons Numerical keypad LED’s RS232 connection Figure 5: Front view of evaluation unit HF connections for signal cable reference cable and cable feed-through Cable feed-through M 20 and M 16 M-Rx Figure 6: Evaluation unit Bottom view 20 M-Tx R-Tx R-Rx RS232 9-pole Sub-D-connector High-frequency connections Micro-Polar Brix LB 565 Chapter 3 System Description LED’s on the Front Panel Five LED’s on the instrument front panel indicate the instrument status. Run Error Signal 1 Signal 2 Figure 7: LED’s on the front panel Comm LED Function Run Instrument in measurement mode A flashing LED indicates an error Error Error Cancelled after reset or if error has been eliminated Signal 1 Display depending on the selected function of relay 1, possible functions: error, no product, alarm min., alarm max., measurement stopped Signal 2 Display depending on the selected function of relay 2, possible functions: error, no product, alarm min., alarm max., measurement stopped. Comm Communication active, e.g. via RS 232 or RS 485 Terminal Block The electrical connections of the LB 565 are located on a connector strip in the wall housing. The terminal block is accessible from the front after you have opened the cover. There, you also find the power cut-off switch and the fuse. The high-frequency connections are located on the outside of the housing. All other elements, especially the live elements (on the motherboard) are provided with a protection cap. Micro-Polar Brix LB 565 21 Chapter 3 System Description 3.4.2 The Microwave Probes The pan probe is available in two different versions: with and without rinsing device (see Figure 8). For technical data please refer to chapter 8. Technical Data. LB 5651 with rinsing device LB 5650 without rinsing device Figure 8: Pan probe A: High-frequency connections B: Process connection, flanges of different sizes Pan Probe Type LB 5650 and Type LB 5651 The pan probe has been specially designed for concentration measurements in containers. Both measuring rods are immersed into the product. The microwaves are emitted from one end of the rod and received at the other end of the rod; they are emitted only in the direction of the opposite end of the rod. This direction characteristic of the probe minimizes the interfering influence of metal parts in the vicinity of the probe and allows installation if only little space is available. For example, the concentration of sugar strike can be measured continuously to find the suitable inoculation time. The plastic rods meet the special requirements for application in food. Two different probe types are available: the "batch pan probe" is used for discontinuous crystallization processes 22 Micro-Polar Brix LB 565 Chapter 3 System Description the "continuous pan probe" with rinsing device is used for continuous crystallization processes; this type of probe prevents crystallization on the microwave exit windows and therefore provides exact measurement results. The flow direction of the product being measured should be vertical, as shown in Figure 9. This ensures a representative product between the measuring rods, provided it is mixed thoroughly. Product Microwave measuring field ~ 190 mm Plastic rods Pt100 100 mm Container wall Figure 9: Batch pan probe LB 5650 HF connections PT-100 Micro-Polar Brix LB 565 Only the batch pan probe is provided with a PT-100 and is connected to the evaluation unit via a 4-wire cable. The wiring diagram for the PT-100 is described in chapter 4.3.2 Pin Configuration of the Connector Strip. Usually, no temperature compensation is required for measurements with the continuous pan probe; therefore, the continuous pan probe is not equipped with a PT 100. 23 Chapter 3 System Description Pan Probe Type LB 5651 with Rinsing Device The probe LB 5651 with rinsing device has been designed for measurements in continuous crystallization processes. The continuous pan probe can be used, for example, for sugar manufacturing, in the afterproduct, raw sugar and white sugar. The continuous pan probe has two rinsing channels which keep the plastic rods free from incrustations; this ensures that the microwaves come into direct contact with the product being measured. All parts coming into contact with the product meet the specific requirements for application in foodstuffs. Figure 10 shows the design of the probe. Microwave measuring field Plastic rods ~ 320 mm Rinsing medium 100 mm Container wall Figure 10: Continuous pan probe LB 5651 24 4 x HF connections 2 x 3/8’’ Rinsing connections, internal thread Micro-Polar Brix LB 565 Chapter 3 System Description 3.4.3 High-frequency Cable High-frequency cables (HF cable) are used to transmit microwaves between probe and evaluation electronics. The measuring system includes a HF quad cable, see Figure 11. The quad cable comprises four individual HF cables of equal length whose ends are terminated by an HF connector (N-type). However, these high-quality special cables are rugged and temperature-resistant only to a certain degree. Warning! Never bend HF cables! The bending radius should not be less than 100 mm. After installation, fix cables with cable binders. HF cables change the conductivity (of microwaves) depending on the temperature. Therefore, they would create measurement errors if the ambient temperature varies. This error is compensated for by enabling the cable compensation. The influences of the ambient temperature on the signal cable are compensated for by means of the reference cable. The reference cable has the same length as the signal cable; during operation, it should be exposed to the same ambient temperature. We, therefore, recommend installing both cable types together in a corrugated tube; this also simplifies installation. 4 x N-connector ø18.5 mm 4 x N-connector Corrugated tube Figure 11: Quad cable Micro-Polar Brix LB 565 0.35 m 0.35 m Lengths 2, 6, 10 m 25 Chapter 3 System Description 3.5 Measurement Configuration The evaluation unit is connected to the microwave probe via HF quad cable. Cable lengths of up to 10 m allow you to place the evaluation unit in a sufficient distance to the evaporation apparatus to rule out any influence by the radiation heat. The pan probe is connected using a HF quad cable, with two individual cables making up the signal and the reference channel. Our example below (Figure 12) shows the measurement configuration on a discontinuous evaporation crystallizer. Micro-Polar Distance up to 10 m (no radiation heat) PT-100 HF quad cable Product Microwave measuring field Pt100 Container wall Figure 12: Typical system configuration of an evaporation crystallizer 26 HF connections Micro-Polar Brix LB 565 Chapter 4 Getting Started Chapter 4. Getting Started 4.1 Transport to the Installation Site Warning! Risk of damage! System parts may get damaged during transportation! Transport probe and the evaluation unit in their original packaging. Protect parts against shocks. Especially the plastic rods of the pan probes have to be protected against mechanical impact! After unpacking, make sure all parts listed on the packing list have been delivered and show no sign of damage; if necessary, clean these parts. If you detect any damage, please notify the forwarder and the manufacturer immediately. 4.2 Installation 4.2.1 Pan Probe Installation Figure 12 shows the position of the batch pan probe on the container. This position is equally valid for the continuous pan probe. The installation sheet in chapter 8 includes all information required for installation. For installation, please keep in mind: Select the installation site such that good mixing and a homogeneous product is ensured and the formation of bubbles on the probe is prevented. A tap should be provided in the direct vicinity to allow representative sampling. Micro-Polar Brix LB 565 27 Chapter 4 Getting Started The probe has to be flange-mounted on the container such that the product being measured flows between both measuring rods. In other words, the fork (both measuring rods) has to be arranged vertically to the material flow (e.g. sugar strike). The distance between the measuring rod tips and any metallized walls (heating element, stirrer, container wall) should be at least a = 60 mm. For installation of the probe you need a hole diameter in the connection flange of at least ∅ 102 mm. For further installation dimensions please refer to chapter 8 (see installation sheets). Rinse Parameters (only with continuous pan probe) The sugar quality or purity is essential for the rinse parameters, i.e. rinse frequency and duration. The rinse parameters may change with the purity of the solution and have to be adapted, in individual cases, to the product and the crystallization process. The following typical rinse parameters are used: Rinse parameters Frequency Duration Rinse solution Temperature of rinse solution Pressure Connections Supply pipe Every 2 hours 12 seconds Water, condensate Average product temperature, typically 65 ±5°C ≥ 3 bar, max. 8 bar 2 x 3/8 inch (female screw thread) ≥ 1/2 inch The rinse connections can be rinsed at the same time or one after the other. The above rinse parameters have to be observed for each connection. Note! The required rinse duration has to take into account a possible inertness of the system, e.g. valve openings. The rinse supply pipes have to be insulated well against heat, in order to prevent an initially colder rinse solution. 28 Micro-Polar Brix LB 565 Chapter 4 Getting Started 4.2.2 Setting up the Evaluation Unit For installation of the evaluation unit please keep in mind: Position the evaluation unit depending on the length of the HF cable in the vicinity of the microwave probe. The HF cable is available in a length of 2, 6 and 10 m; the standard cable length is 2 m. The instrument has to be protected against vibrations. For instrument installation you should foresee a cutoff device to allow you to disconnect the meter easily and quickly from the power supply. When installing the evaluation unit on an evaporation crystallizer, you should use a distance rail to minimize the thermal radiation and conduction. See Figure 13 Insulation of pan wall Spacer Figure 13: Top view: Installation of evaluation unit on the evaporation crystallizer Micro-Polar Brix LB 565 Evaluation unit 29 Chapter 4 Getting Started 4.3 Connecting the Evaluation Unit 4.3.1 Connecting the HF Cable You need a HF quad cable to connect the pan probe to the evaluation unit. Note! Correct cable installation: Make sure the cable does not get into contact with hot pipes over the entire length (corrugated tube and single cable section after splitting), e.g. direct contact with the device wall (not insulated). This alone guarantees that all single cables are subject to the same ambient conditions and that the compensation of the cable drift works properly. M-Tx Microwave measuring field ø100 Figure 14: Connection of the discontinuous probe on the evaluation unit M-Rx M-Tx R-Tx R-Rx M-Rx M-Tx R-Tx R-Rx Evaluation unit, underside PT-100 R-Tx R-Rx M-Rx HF quad cable The HF cables and the HF connections on the evaluation unit and on the probe are labeled. Connect the probe to the evaluation unit as shown in Figure 14. 30 Micro-Polar Brix LB 565 Chapter 4 Getting Started Note! When tightening the 21 mm screw nut, make sure that the connector is not twisted on the cable. If the connector is twisted relative to the cable, the shielding may get damaged and this could result in mismatching and bad sealing. Hand tighten all screwed connections of the HF cable (2 Nm = 0.2 kg/m)! Before tightening, carefully screw on the cable by hand. Caution! Threaded joint jams easily. Occasionally you should check if the screwed connection is still properly tightened. If the installation is exposed to vibrations, the screwed connection may come loose and this may result in inaccurate measurements or corrosion of the connections. As long as the cables are not connected, the coaxial sockets have to be covered immediately with plastic caps and the cable connectors have to be protected by suitable provisions against moisture and dirt. Note! Never bend HF cables! The bending radius should not be less than 100 mm. After installation, fix cables with cable binders or other suitable means, so that the cable cannot slip any more! Micro-Polar Brix LB 565 31 Chapter 4 Getting Started 4.3.2 Pin Configuration of the Connector Strip Danger! Electrical hazards: Turn off power supply to ensure that contact with voltage-carrying parts is avoided during installation and when servicing. Turn off power supply before opening the instrument. Work on open and live instruments is prohibited. Temperature Signal Connection A PT100 or a temperature current signal has to be connected to current input 1 or 2 if significant temperature fluctuations occur in the product and if a temperature dependence of the phase or attenuation measurement is likely to occur. The temperature sensor has to measure the material temperature in the vicinity of the microwave probe. When taking the batch pan probe into operation, connect the 4wire cable of the PT-100 to the connector strip of the evaluation unit as follows: blue (23) Evaluation unit Connector strip Figure 15: PT-100 Connection batch pan probe 32 (11) blue white PT-100 sensor in the batch pan probe white ( ) Terminal no. Micro-Polar Brix LB 565 Chapter 4 Getting Started Other Connections Connect all desired input and output signals to the terminal strip as shown on the following pages. Use the M feed-through to keep the degree of protection. Check if the voltage indicated on the type plate matches your local supply voltage. Connect the line cable to the terminals 3(L1), 2(N) and 1(PE). Check if the test switch (mains interruption) is in position „on“ (see Figure 17). Close the instrument housing and turn on the power supply. The line cross-section for power supply must be at least 1.0 mm2. Micro-Polar Brix LB 565 33 Chapter 4 Getting Started On the connector strip of the evaluation unit you find the following connections: PE N L1 Figure 16: Wiring diagram LB 565 Power supply: Terminals 3 (L1), 2 (N) and 1 (PE) 90 V - 260 V AC, 45 - 65 Hz or 24 V AC/DC, DC: 18 V … 36 V ; AC: 24 V +5%, -20%; 45…65 Hz Depending on the type of system, see type plate on the outer wall of the housing. Current input no. 1 (terminals 20+ and 8-), insulated Input as 0/4 - 20 mA signal. For example, for temperature compensation or reference signal recording. Current input no. 2 (terminals 22+ and 10-), not insulated Input as 0/4 - 20 mA signal. For example, for temperature compensation or reference signal recording. Current output no. 1 (terminals 27+ and 15-), insulated Output as 4 - 20 mA signal. Output options: concentrations (1 / 2), current inputs signals (1 / 2) and PT100 signal 34 Micro-Polar Brix LB 565 Chapter 4 Getting Started Current output no. 2 (terminals 19+ and 7-), insulated Output as 0/4 - 20 mA signal. Output options: concentrations 1 and 2, current input signals 1 and 2 and PT100 signal PT100 (terminals 23+ and 11-) Connection for temperature measurement. Digital input 1: DI1 (terminals 24+ and 12-) Configuration options: no function measurement: start (closed) and stop (open) Digital input 2: DI2 (terminals 25+ and 13-) Configuration options: no function average value: hold (closed) and continue averaging (open) product selection: product 1 (open) and product 2 (closed) Digital input 3: DI3 (terminals 26+ and 14-) Configuration options: no function start sampling, open: no action, closed: unique measurement starts product selection Relay 1: (terminals 4, 5 and 6) Changeover contacts, insulated, configuration option: no function error message measurement stopped limit value min. and max. no product Micro-Polar Brix LB 565 35 Chapter 4 Getting Started Relay 2: (terminals 16, 17 and 18) Changeover contacts, insulated, configuration option: no function error message measurement stopped limit value min. and max. no product RS485 interface (terminals 21 (RS1) and 9 (RS2)) Transfer rate 38400 baud, 8 data bits, 1 stop bit, no parity, no handshake. Bidirectional, serial data interface for communication and protocol output RS232 interface (on instrument underside) 9-pole D-connectors. Data transfer rate 38400 baud, 8 data bits, 1 stop bit, no parity, no handshake. Bidirectional, serial data interface for communication and protocol output. 36 Micro-Polar Brix LB 565 Chapter 4 Getting Started 4.3.3 Digital Outputs, Relays The status of the measurement is output via two relays: Error Alarm (alarm min. and max.) No product Under menu item Plausibility, you may enter a min. attenuation for pause detection (e.g. for discontinuous evaporation crystallizer); if this value is not reached, „no product“ is signalled via a relay and the current output drops to 0 or 4 mA. Measurement stopped The respective switching status is also signaled via LED’s on the front panel (LED’s: signal 1 and 2). Relay no. Error, alarm, no product, measurement stopped, current-free status Normal com 16 16 17 18 com 17 com com 18 The relays with changeover contacts can either be operated as make contact, terminals 4 & 5 (open at error, alarm ...) or as break contact, terminals 5 & 6 (closed at error, alarm ...). Micro-Polar Brix LB 565 37 Chapter 5 Service Instructions Chapter 5. Service Instructions 5.1 General Information The evaluation unit has no wearing parts or components requiring any special maintenance. A malfunction of the measuring system is not always due to a defect in the instrument. Often the error is caused by incorrect operation, wrong installation, or irregularities in the product being measured. If a malfunction occurs, anyway, the measuring system helps you to identify and eliminate errors by displaying error messages on the LCD, indicating operator errors and defects of the electronics. Moreover, the modular design of the measuring system simplifies the replacement of individual module components. Usually, faulty modules of the evaluation unit cannot be repaired but have to be replaced. The microwave module is fixed with screws to a shielding cover and must not be opened. 5.2 Instrument Cleaning Clean all system components using only a moistened cloth without chemical cleaning agent. Parts coming into contact with the product (during regular operation) can be cleaned with hot water, taking into account the temperature limits (see chapter 6.2). 5.3 Batteries If the measuring system LB 565 is without power supply (power failure or disconnected from mains), the system clock is supplied with power via the Lithium battery on the CPU. The instrument works correctly even with empty battery, only measured data which are output via one of the serial interfaces may become useless as a result of the faulty date and time information. The service life of the battery, even under continuous load, is at least 8 years. To replace the battery, you have to disconnect the instrument from mains. Battery data see chapter 8.2 Electrical Wiring Diagram. Battery type: 3 Volt Lithium cell (button cell), type CR2032. 38 Micro-Polar Brix LB 565 Chapter 5 Service Instructions 5.4 Fuse Replacement The mains fuse of the LB 565 is located in the wall housing. Replace the fuse only if the instrument is disconnected from mains. Be sure that the new fuse matches the rating specified. Use only fuse with correct rating: Instrument version with 90 V ... 260 V AC: 2.0 A slow-blow Instrument version with 24 V AC/DC: 2.0 A slow-blow Warning! Spare fuses must match the rating specified by the device manufacturer. Short-circuiting or manipulation is not permitted. Protective cover Netzteil Test switch Motherboard Fuse EEprom (can be pulled off) Line connector Figure 17: View with open wall housing Feed-through for line connector Micro-Polar Brix LB 565 Terminal strip (can be pulled off) Battery 39 Chapter 6 Technical Data Chapter 6. Technical Data General Specifications Method Microwave transmission measurement Working frequency 2.4 – 2.5 GHz (ISM band), depending on local regulations Transmission power < 0.1 mW (< - 10 dBm), coaxial line power Applications Concentration measurement in sugar solutions in containers and pipes 6.1 Evaluation Unit Evaluation Unit Housing Wall housing made of stainless steel, material 1.4571 (~316+Ti), HxWxD: 300 x 323 x 140 mm, see dimensional drawing in chapter 8 Protection type IP 65 Weight approx. 6.5 kg Operating temperature 0 °C ... + 60 °C ( 273 K ...333 K ), no condensation Storage temperature - 20 °C ... + 80 °C ( 253 K ... 353 K ), no condensation Display Dot matrix LC display, 114 mm x 64 mm, 240 x 128 pixels, with back-lighting, automatic contrast setting Keyboard Freely accessible foil keypad, light-stable and weatherproof: alphanumeric keyboard and 4 softkeys (software-assigned buttons) Power supply Depending on instrument version: 1. 90 V - 260 V AC, 45 Hz – 65 Hz 2. 24 V AC/DC; DC: 18 V – 36 V; AC: 24 V +5%, -20%, 45 Hz – 65Hz Power consumption 40 max. 30 VA (AC/DC), depending on configuration Micro-Polar Brix LB 565 Chapter 6 Technical Data Fuses 2.0 A / slow-blow Battery type 3 V Lithium button cell, type CR2032 Measured value Concentration, dry mass content, Brix Communication module Preparation for HART Inputs and Outputs Cable cross-section min. 1.0 mm² (mains supply) Cable feed-through 2 x M20x1.5 for cable 5...14mm (depending on application), 4 x M16x1.5 for cable 5 ...8 mm (depending on application) Probe connection Inputs and outputs for signal and reference channel, 50 Ω N-socket Probes Cable lengths: 2 m, 6 m and 10 m; 50 Ω ; both sides with 4 N connectors Current input 2 x current input 0/4 ...20 mA, ohmic resistance 50 Ω, 1x insulated, 1x instrument ground e.g. for temperature compensation Current output Current output 1: 4...20 mA, ohmic resistance max. 800 Ω , insulated current output 2: 0/4...20 mA, ohmic resistance max. 800 Ω , insulated e.g. for measured value or temperature output PT-100 connection Measuring range: - 50 °C ... + 200 °C (223 K ... 473 K); measurement tolerance: < 0,4 °C Digital input 3 x digital inputs (DI1..3), for floating contacts, TTL-level connected to instrument ground Configuration options: DI1: none, measurement start/stop DI2: none, measurement hold, product DI3: none, sample measurement, product Function description: 1. Measurement (Start/Stop), open: measurement stopped, closed: measurement started or measurement running 2. Hold measurement, open: measurement running, closed: measurement stopped, i.e. average values and current output are held Micro-Polar Brix LB 565 41 Chapter 6 Technical Data 3. Product selection, open: product 1 (P1), closed: P2; with two DI’s: DI2 open & DI3 open: P1, DI2 closed & DI3 open: P2, DI2 open & DI3 closed: P3, DI2 closed &DI3 closed: P4 4. Start sample measurement, open: no actions, closed: single measurement starts Relay outputs 2 x relay (SPDT), insulated Configuration options: collective error message measurement hold threshold (min. and max.) no product Loading capacity: AC: max. 400VA DC: AC / DC: max. 90W max. 250V or max. 2A, not inductive ≥ 150V: voltage must be grounded Restrictions for 24 V AC/DC (DC: 18 V...36 V; AC: 24 V +5 %, -20 %) mains supply, if the ground conductor is not connected to terminal 1 (PE): AC: max. 50 V DC: max. 70 V Serial interface (bidirectional) RS 232 on the underside of the instrument, 38400 baud, no handshake, data format 8 data bits, 1 stop bit RS 485 via terminal block, 38400 baud, no handshake, data format 8 data bits, 1 stop bit 42 Micro-Polar Brix LB 565 Chapter 6 Technical Data 6.2 Probes Pan Probes Micro-Polar Brix LB 565 Application Pan probes with and without rinsing device for concentration measurement in process containers and in pipelines with nominal width ≥ 200 mm. Material Plastic rod, stainless steel 1.4301 Process coupling Flange according to DIN 2527 DN65 / PN6, DN 80, 100, 150 / PN16; ANSI Flange 2.5’’ / 150 PSI (others on request) Process pressure 0 ... + 3 bar (relative) Temperature range Product temperature: + 10 °C ...+100 °C, short-term up to + 120 °C; storage temperature: + 10 °C to + 80 °C Connections 4 x HF cable: 50 Ω , N-connectors on both sides (max. 10 m HF cable) Dimensions See dimensional drawings in chapter 8 43 Chapter 7 Certificate Chapter 7. Certificate 7.1 EC-Declaration of Conformity 44 Micro-Polar Brix LB 565 Chapter 7 Certificate 7.2 Frequency approval Micro-Polar Brix LB 565 45 Chapter 7 Certificate 46 Micro-Polar Brix LB 565 Chapter 8 Engineering drawings Chapter 8. Engineering drawings 8.1 Dimensional drawing for housing Micro-Polar Brix LB 565 47 Chapter 8 Engineering drawings 8.2 Electrical Wiring Diagram PE N L1 48 Micro-Polar Brix LB 565 Chapter 8 Engineering drawings 8.3 Dimensional drawings for batch pan probes 8.3.1 Type LB 5650-01 Micro-Polar Brix LB 565 49 Chapter 8 Engineering drawings 8.3.2 50 Type LB 5650-02 Micro-Polar Brix LB 565 Chapter 8 Engineering drawings 8.3.3 Type LB 5650-03 Micro-Polar Brix LB 565 51 Chapter 8 Engineering drawings 8.3.4 52 Type LB 5650-04 Micro-Polar Brix LB 565 Chapter 8 Engineering drawings 8.3.5 Type LB 5650-05 Micro-Polar Brix LB 565 53 Chapter 8 Engineering drawings 8.4 Dimensional drawings for continuous pan probes 8.4.1 54 Type LB 5651-01 Micro-Polar Brix LB 565 Chapter 8 Engineering drawings 8.4.2 Type LB 5651-02 Micro-Polar Brix LB 565 55 Chapter 8 Engineering drawings 8.4.3 56 Type LB 5651-03 Micro-Polar Brix LB 565 Chapter 8 Engineering drawings 8.4.4 Type LB 5651-04 Micro-Polar Brix LB 565 57 Chapter 8 Engineering drawings 8.4.5 58 Type LB 5651-05 Micro-Polar Brix LB 565 Chapter 8 Engineering drawings 8.5 Installation sheets for LB 5650 (Batch pan probes) Micro-Polar Brix With Pan Probe - Installation In Evaporation Crystallizer - 90 V - 260 V AC or 24 V AC/DC output 0/4 - 20 mA top view into the evaporation crystallizer R-Tx R-Rx R-Tx R-Rx M-Tx M-Tx 190 M-Rx M-Rx evaluation unit botten side M-Tx PT100 R-Tx microwave measuring field HF cable quad ø100 PT100 R-Rx M-Rx ø160 60 evaluation unit 300 sealing pan wall 220 Micro-Polar probe flange DN65 / PN6 pan probe Oben/Top do not remove, only in case of service by Berthold Technologies ø1 installation position insulation R-Tx M-Rx 30 M-Tx ø1 heating PT100 R-Rx 60 ø1 60 current stream fitting flange ø130 / 4 x M12 area without metalic installations ø102 Id.-No. 41975TI2 Rev.01 ± 0,5 BERTHOLD TECHNOLOGIES GmbH & Co. KG . P.O. Box 100 163 . D-75312 Bad Wildbad, Germany Phone +49 7081 177-0 . Fax +49 7081 177-100 . industry@BertholdTech.com www.BertholdTech.com Micro-Polar Brix LB 565 59 Chapter 8 Engineering drawings Micro-Polar Brix With Pan Probe - Installation In Evaporation Crystallizer - 90 V - 260 V AC or 24 V AC/DC output 0/4 - 20 mA top view into the evaporation crystallizer M-Tx R-Tx R-Rx R-Rx M-Tx R-Tx M-Rx M-Rx evaluation unit botten side pan wall 190 M-Tx PT100 R-Tx HF cable quad ø100 PT100 microwave measuring field R-Rx M-Rx evaluation unit øB 60 300 probe flange DN80, 100, 150 / PN16 pan probe DN A B C S sealing Micro-Polar installation position 220 80 160 200 18 M16 100 180 220 18 M16 150 240 285 22 M20 øC insulation do not remove, only in case of service by Berthold Technologies Oben/Top R-Tx M-Rx M-Tx øB øA heating R-Rx current stream øA / 8 x S area without metalic installations ø102± 0,5 BERTHOLD TECHNOLOGIES GmbH & Co. KG . P.O. Box 100 163 . D-75312 Bad Wildbad, Germany Phone +49 7081 177-0 . Fax +49 7081 177-100 . industry@BertholdTech.com Id.-No. 41975TI21 Rev.01 fitting flange 60 PT100 www.BertholdTech.com 60 Micro-Polar Brix LB 565 Chapter 8 Engineering drawings 8.6 Installation sheets for LB 5651 (Continuous pan probes) Micro-Polar Brix With Continuous Pan Probe - Installation In Evaporation Crystallizer - 90 V - 260 V AC or 24 V AC/DC output 0/4 - 20 mA top view into the evaporation crystallizer R-Tx R-Rx R-Tx R-Rx M-Tx M-Tx 320 M-Rx M-Rx evaluation unit botten side M-Tx R-Tx HF cable quad ø100 microwave measuring field R-Rx M-Rx ø160 160 evaluation unit 300 sealing pan wall 220 Micro-Polar probe flange DN65 / PN6 cont. pan probe do not remove, only in case of service by Berthold Technologies Oben/Top ø1 installation position insulation R-Tx M-Rx M-Tx 55 ø1 30 2 x 3/8’’ flushing connection (internal thread) R-Rx heating 60 ø1 60 current stream ø130 / 4 x M12 area without metalic installations ø102± 0,5 BERTHOLD TECHNOLOGIES GmbH & Co. KG . P.O. Box 100 163 . D-75312 Bad Wildbad, Germany Phone +49 7081 177-0 . Fax +49 7081 177-100 . industry@BertholdTech.com Id.-No. 41976TI2 Rev.02 fitting flange www.BertholdTech.com Micro-Polar Brix LB 565 61 Chapter 8 Engineering drawings Micro-Polar Brix With Continuous Pan Probe - Installation In Evaporation Crystallizer - 90 V - 260 V AC or 24 V AC/DC output 0/4 - 20 mA R-Tx R-Rx R-Rx M-Tx R-Tx 320 M-Rx M-Tx pan wall M-Rx evaluation unit botten side top view into the evaporation crystallizer M-Tx R-Tx HF cable quad ø100 microwave measuring field R-Rx M-Rx øB 160 evaluation unit 300 sealing probe flange DN80, 100, 150 / PN16 cont. pan probe DN A B C S øC installation position Micro-Polar 220 80 160 200 18 M16 100 180 220 18 M16 150 240 285 22 M20 insulation do not remove, only in case of service by Berthold Technologies Oben/Top R-Tx M-Rx M-Tx 55 øB øA heating R-Rx 60 2 x 3/8’’ flushing connection (internal thread) current stream fitting flange øA / 8 x S area without metalic installations Id.-No. 41976TI21 Rev.02 ø102± 0,5 BERTHOLD TECHNOLOGIES GmbH & Co. KG . P.O. Box 100 163 . D-75312 Bad Wildbad, Germany Phone +49 7081 177-0 . Fax +49 7081 177-100 . industry@BertholdTech.com www.BertholdTech.com 62 Micro-Polar Brix LB 565 Notes Micro-Polar Brix LB 565 63 Notes 64 Micro-Polar Brix LB 565 Micro-Polar Brix LB 565 Micro-Polar Brix LB 565 Process Control detect and identify Concentration Meter Micro-Polar Brix LB 565 Software Manual User’s Guide ID No. 39531BA2 Rev. No.: 00 01/07/04 Soft. Version 1.00 The units supplied should not be repaired by anyone other than Berthold Service engineers or technicians authorized by Berthold. In case of operation trouble, please address to our central service department. The complete user’s guide consists of two manuals, the hardware description and the software description. The hardware manual comprises: mechanical components assembly electrical installation radiation protection guidelines technical data electrical and mechanical drawings The software manual comprises: operation of the evaluation unit parameter description basic setting calibration error messages The present manual is the software description. Subject to change without prior notice. Micro-Polar Brix LB 565 Table of Contents Table of Contents Page Chapter 1. Communication with Micro-Polar Brix ......................7 1.1 Brief Instructions........................................................8 1.2 System Configuration..................................................8 1.3 System Calibration .....................................................8 Chapter 2. Software Functions ..................................................9 2.1 Information on Menu Structure ....................................9 2.2 Menu Structure ........................................................ 10 2.2.1 Start Menu...............................................................12 2.2.2 Diagnostic................................................................12 2.2.3 Setup ......................................................................13 2.2.4 Access Level.............................................................13 2.2.5 Language.................................................................14 2.2.6 Configuration ...........................................................14 2.2.7 General ...................................................................14 2.2.8 Measurement ...........................................................15 2.2.9 Plausibility ...............................................................15 2.2.10 Microwave..............................................................16 2.2.11 Calibration .............................................................16 2.2.12 System Adjust ........................................................16 2.2.13 Calibrate Concentration ...........................................17 2.2.14 Sample No. ............................................................17 2.2.15 Sample Data (expanded) .........................................18 2.2.16 Advanced Settings ..................................................18 2.2.17 Calibr. manual ........................................................19 2.2.18 Input / Output ........................................................19 2.2.19 Current Output .......................................................20 2.2.20 Current Out 1 .........................................................20 2.2.21 Current Out 2 .........................................................21 2.2.22 Current Input .........................................................21 2.2.23 Current In 1 ...........................................................21 2.2.24 Current In 2 ...........................................................21 2.2.25 Pt 100 ...................................................................22 2.2.26 Digital Output.........................................................22 2.2.27 Digital Input ...........................................................23 Chapter 3. Configuration .........................................................24 3.1 Configuration Setup .................................................. 24 3.1.1 3.1.2 3.1.3 3.1.4 General Data............................................................24 Measurement ...........................................................25 Plausibility ...............................................................26 Microwave ...............................................................26 Chapter 4. Calibration .............................................................27 4.1 System Calibration ................................................... 27 4.2 Start Calibration ....................................................... 29 4.2.1 Tuning.....................................................................29 4.2.2 Sampling .................................................................32 Micro-Polar Brix LB 565 Table of Contents 4.2.3 Entering the Laboratory Values ...................................33 4.2.4 Automatic Calibration ................................................35 4.3 Manual Calibration.................................................... 36 4.3.1 4.3.2 4.3.3 4.3.4 Manual Calibration with One Concentration ..................36 Calibration with Two Concentrations............................39 Calibration with Split Value ........................................43 Calibration with Temperature Compensation.................45 Chapter 5. Password .............................................................. 48 5.1 Password................................................................. 48 Chapter 6. Inputs / Outputs ................................................... 49 6.1 Current Outputs ....................................................... 49 6.1.1 Current Output Setup ................................................50 6.1.2 Test and Adjustment .................................................51 6.1.3 Error Current ............................................................53 6.1.4 Current Output 2 ......................................................54 6.2 Current Inputs ......................................................... 55 6.2.1 Enabling the Current Input.........................................55 6.2.2 Range Setting and Adjustment ...................................56 6.3 Pt 100..................................................................... 58 6.3.1 Pt 100 Enabling ........................................................58 6.3.2 Pt 100 Calibration .....................................................59 6.4 Digital Output .......................................................... 60 6.4.1 Digital Output Assignment .........................................60 6.5 Digital Input ............................................................ 61 6.5.1 External Product Switchover.......................................62 Chapter 7. Factory Settings .................................................... 63 Chapter 8. Error Lists ............................................................. 64 8.1 Error Lists ............................................................... 64 8.1.1 Hardware Error .........................................................64 8.1.2 Input Error...............................................................64 8.1.3 Measurement Error ...................................................65 Chapter 9. Calibration Data Sheet .......................................... 66 9.1 Configuration ........................................................... 66 9.1.1 9.1.2 9.1.3 9.1.4 General ...................................................................66 Measurement ...........................................................66 Plausibility ...............................................................66 Microwave................................................................66 9.3.1 9.3.2 9.3.3 9.3.4 9.3.5 Current Output .........................................................67 Current Input ...........................................................67 Pt 100 Input.............................................................68 Digital Output...........................................................68 Digital Input.............................................................68 9.2 Product ................................................................... 67 9.3 Input / Output ......................................................... 67 9.4 Calibration Data ....................................................... 68 9.4.1 Calibration Coefficients ..............................................68 9.4.2 Sampling Table.........................................................69 Micro-Polar Brix LB 565 Safety Summary GENERAL WARNINGS Parameter settings Never change the parameter settings without a full knowledge of these operating instructions, as well as a full knowledge of the behavior of the connected controller and the possible influence on the operating process to be controlled. Micro-Polar Brix LB 565 Chapter 1. Communication with Micro-Polar Brix Chapter 1. Communication with Micro-Polar Brix Communication with the Micro-Polar Brix takes place via 4 softkey buttons. The function of the individual buttons changes relative to the position in the menu. Values and texts are entered via an alphanumeric keyboard. The instrument status is indicated by 5 LED’s. Micro-Polar LCD display Lock Softkey buttons Numerical keypad LED’s RS232 connection HF connections for signal cable reference cable and cable feed-through Micro-Polar Brix LB 565 Chapter 1. Communication with Micro-Polar Brix 1.1 Brief Instructions The main menu is displayed automatically after power on, provided the Micro-Polar Brix has been connected properly. To get correct measurement values, the instrument has to be configured and calibrated before running the first measurement. 1.2 System Configuration Select | Setup | Configuration | General Data Enter the general data (date, time, tag) □< Push Home button to return to the Configuration menu and select | Measurement | Enter the system parameters (measurement mode, start mode, averaging, units, ...) □< Push Home button to return to the Configuration menu Select | Plausibility | Enter the limit values (limits conc. 1, variance of phase measurement, pause detection ) □< Push Home button to return to the Configuration menu Select | Microwave | Enter the cable length (reference cable length, signal cable length ) □< Push Home button to return to the Setup menu Select | Input/Output | Enter the values for current output, current input, Pt100, digital out-/input □< Push Home button to return to the Setup menu. 1.3 System Calibration You have to power on the instrument at least 30 minutes prior to system calibration. On the main menu, select | Setup |Calibration | System Adjust | Adjust | Start the adjustment only if it is sure that the transducer is sufficiently covered by the product. The typical standard coefficients have been set up by the manufacturer for your application. A sample has to be taken during system adjustment. The lab value of the sample is needed for calculation of the offset. Calculation: analysis value – reading = offset. The offset can only be entered in the Profi mode. Upon completion of the system adjustment, push the □< Home button three times to return to the main menu. Push the „RUN“ softkey to start the measurement. The live display appears. Push the „ESC“ button to return to the main menu and select | Setup | Calibration | Calibrate Conc. | Tuning | Offset | Enter offset value. Push the □< Home button 4 times to return to the main menu. Select | Live display | If the product has not changed, the reading value corresponds to the laboratory value. Micro-Polar Brix LB 565 Chapter 2. Software Functions Chapter 2. Software Functions 2.1 Information on Menu Structure The menu structure on the following pages provides an overview of all functions of the Micro-Polar Brix. Using the page numbers indicated you can look up the function of the depicted window. The menu scope is reduced, depending on the access level. You have to enter an editable password to change from the level „Read only“ to „Basic“ or to „Profi“. For „Service“ the password is fixed. Micro-Polar Brix LB 565 10 Page 12 Live Display Diagnostic Setup Access Level Language 2.2 Menu Structure Chapter 2. Software Functions English German Language Read only Basic Profi Service Access Level Page 14 Page 13 See next page Setup Datalog Page 12 Errorlog Info System Monitor Diagnostic ZOOM Micro-Polar Brix LB 565 Conc act ..▼▲.. 65.50 % 64.35% 07.05 – 13:25 Concentration av. Live Display PROBE Conc av. ESC 1| Page 22 Page 21 Page 22 Page 21 Status DI 1 function DI 2 function DI 3 function Digital Input Relay1 Relay2 Test Digital Output Enabled Trim Pt 100 Pt 100 live Pt100 Current In 1 Current In 2 Current Out 1 Current Out 2 Page 20 Current Input Current Output Page 13 Configuration Calibration Input / Output Service Product Change password Micro-Polar Brix LB 565 Page 19 Current Output Current input Pt100 Digital Output Digital Input Input / Output Live Display Diagnostic Setup Access Level Language Setup Page 16 Enabled DAC trim Live current Current In 2 Enabled DAC trim Live current Current In 1 Assignment Upper value Lower value Range Test/Adjust Error current Page 21 Page 21 Page 21 Page 20 Current Out 2 Assignment Upper value Lower value Test/Adjust Error current Current Out 1 System Adjust Calibrate Conc Advanced Calibration Cal. Order Cal. Base Coefficients Compensation Start Calibr. Page 14 Page 19 Page 18 Page 17 Page 16 Calibr. manual Num Cal. Sweeps Process Type Split Value Advanced Sampling Calibr. autom. Calibr. manual Tuning View Calibrate Conc Adjust Num. Sweeps Ref. values Chart Phi Chart Attn. System Adjust General Data Measurement Plausibility Microwave Configuration Page 14 Page 16 Page 15 Current Input 1 Current Input 2 Pt 100 PHI (m) Attenuation Advanced 11 Page 18 Next sample Actice Meas Value Lab Value Advanced Page 17 Sample No Cables Microwave Process limits Phase Measure Pause Detection Plausibility Meas. Mode Start Mode Averaging Reset Averaging Markers Unit Page 15 Measurement Date Time Tag General Data Software Functions Chapter 2. Software Functions 2.2.1 Start Menu LIVE DISPLAY: Shows the live display. 1| - | LB 565 | 07.05 – 13:25 Live Display Diagnostic Setup Access Level Language RUN ▲ ▼ ► DIAGNOSTIC: This menu item contains the submenu items data logger, error log and further instrument information. SETUP: All necessary inputs for operation of the measuring system can be entered here. ACCESS LEVEL: Areas protected by passwords can be cleared. LANGUAGE: Select the dialog language. 2.2.2 Diagnostic 1 | - | Diagnostic | 07.05 – 13:25 Datalog Errorlog Info System Monitor ⌂◄ ▲ ▼ ► Datalog: • Log type • Log time • • Averaging time • Print log enable/disable logging period 15 minutes to 3 days Input in seconds Printout of tables Error log: • Shows the logged error Info: • • • • • • • • • • System • • • 12 Device type Supplier Manufacturer Device no. Serial no. Tag Date Production no. Software ver. Hardware ver. Monitor: Internal temperature Max int. temperature HF voltage LB565 Berthold Berthold 4294967295 4294967295 Input possible 05.04.2004 V1.00 P01 : 40.79°C : 42.27°C : 9.72 V Micro-Polar Brix LB 565 Chapter 2. Software Functions 2.2.3 Setup 1 | - |Setup | 07.05 – 13:25 Configuration Calibration Input / Output Service Product Change password ⌂◄ ▲ ▼ Configuration: • Setup of • General data • Measurement-specific data • Plausibility data • Microwave data Calibration: • System calibration • Concentration calibration • Advanced setup ► Input / • • • • • Output: Current outputs Current inputs Pt 100 Digital outputs Digital inputs Service: • Factory setting (default values) Product: • Product selection (1 – 4); if you select another product, the product-specific data will be loaded; outputs, inputs and calibration Change password: • The password for the access levels Basic / Advanced can be changed here. 2.2.4 Access Level 1 | - |Change Access| 07.05 –13:25 Basic: • Read only Basic Profi Service ESC Read only: • This mode can be selected on all levels without password. • ▲▼ ► No password required on higher levels. Password has to be entered for „Read only“. Password can be changed. Profi: • As described above. Should be used only if you have sufficient knowledge of the measuring system. Service: • This level is reserved to the service personnel and cannot be changed. Micro-Polar Brix LB 565 13 Chapter 2. Software Functions 2.2.5 Language 1| | LB 565 | 07.05 – 13:25 LANGUAGE Language: • Select the dialog language English German ESC ▲▼ .√. 2.2.6 Configuration 1 | - |Configuration | 07.05 – 13:25 General Data Measurement Plausibility Microwave ⌂◄ ▲ ▼ ► General Data: • Enter date, time and tag Measurement: • Measurement mode (batch/continuous) • Start mode (keyboard/external) • Averaging (number of measured values used for averaging) • Reset average value (yes/no) • Marker (enter: value, name and correlation) • Units for concentration and temperature Plausibility: • Variance • Ratio phase/dB Microwave: • Cable (enter the reference and signal cable length) 2.2.7 General Date: 1 | - |General Data| 07.05 – 13:25 Date Time Tag • Enter the current date • Enter the current time • Enter the name of the tag. Time: Tag: ⌂◄ 14 ▲ ▼ ► Micro-Polar Brix LB 565 Chapter 2. Software Functions 2.2.8 Measurement 1 | - |Measurement| 07.05 – 13:25 Meas. Mode Start Mode Averaging Reset Averaging Markers Units ⌂◄ ▲ ▼ ► Meas. Mode: • Continuous or batch (in the batch mode, an average value is calculated between start and stop; continuous means averaging depending on the defined number of averaging cycles) Start Mode: • The measurement device can be started or stopped via external terminals (digital input) or via keyboard. Averaging: • Enter the number of single measurement values over which a moving average is to be calculated. Reset Averaging: • Reset averaging (yes/no) Markers: • Depending on the configuration, up to two markers can be entered by name and by value. Units: • Depending on the configuration, different units can be selected for concentrations and temperature. 2.2.9 Plausibility 1 | - | Plausibility | 07.05 – 13:25 Process limits Phase Measure Pause Detection ⌂◄ ▲ ▼ ► Process Limits: • A minimum and maximum concentration can be set where averaging stops and a message is displayed. Phase Measure: • Here you set the maximum variance of the regression Phase vs. Frequency. • The correlation between Phase and Attenuation is another plausibility criterion. Pause Detection: • Can be enabled or disabled and utilizes the measured attenuation as switching variable; the current output, e.g. with empty container, will be set to min. display. Micro-Polar Brix LB 565 15 Chapter 2. Software Functions 2.2.10 Microwave 1 | - | Microwave | 07.05 – 13:25 Cables: • Enter the reference and signal cable length. Cables ⌂◄ ▲ ▼ ► 2.2.11 Calibration 1 | - | Calibration | 07.05 – 13:25 System Adjust Calibrate Conc Calibrate Conc2 Advanced ⌂◄ ▲ ▼ System Adjust: • The system calibration is started on this page. Calibrate Conc: • Opens the calibration menu of concentration 1. ► Calibrate Conc2: • Opens the calibration menu of concentration 2. Advanced: • Here you set are the number of sweeps, the process type and the split value. 2.2.12 System Adjust 1 | - |System Adjust| 07.05 –13:25 Adjust Num. Sweeps Ref. values Chart Phi Chart Attn. ⌂◄ ▲ Adjust: • System calibration is started. Num. Sweeps: • Here the number of measurement cycles is set. ▼ ► Ref. values: • Upon completion of the reference measurement, the reference values for phase, attenuation, slope and variance can be output. Chart Phi: • Shows the characteristic curve of Phase versus Frequency of the regression. Chart Atten.: • Shows the characteristic curve of Attenuation versus Frequency of the regression. 16 Micro-Polar Brix LB 565 Chapter 2. Software Functions 2.2.13 Calibrate Concentration 1 | - |Calibrate Conc| 07.05 –13:25 Sampling Calibr. autom. Calibr. manual Tuning View ⌂◄ ▲ ▼ ► Sampling: • Shows all measured samples. Calibr. autom.: • Calibration can be started after measurement of two samples and input of the respective laboratory values. Calibr. manual: • Here you can choose the calibration order [linear/quadratic], the basis [phase/attenuation or both] and compensation [temperature]. Tuning: • Subsequent correction of the reading is possible by entering a factor and an offset. View: • Presentation of calibration curve, display of correlation and coefficients. 2.2.14 Sample No. 1 | 1/1 | Sample # 1 | 07.05 –13:25 Next sample Active Meas value Lab value Advanced ◄ DEL Yes 65.50 % 0.00 % ▲▼ ..√.. Sample number, date and time of the sample measurement are displayed in the header. Next sample: • Continues with the next sample. Active: • Here you can choose if this sample should be taken into account in the calibration. Measured value: • Display of the measured values, calculated with the actual coefficient. Lab value: • Entry position for the laboratory value. Advanced: • Switches to the next data page. Micro-Polar Brix LB 565 17 Chapter 2. Software Functions 2.2.15 Sample Data (expanded) 1 | 1/1 | Sample # 1 | 07.05 –13:25 Current In 1 Current In 2 Pt 100 PHI (m) Atten ◄ DEL Current In 1: • Editable display of the first compensation input. Current In 2: • Editable display of the second compensation input. ▲▼ ..√.. Pt 100: • Editable display of the Pt 100 input. PHI (m): • Not editable display of the measured phase. Atten: • Not editable display of the measured attenuation. 2.2.16 Advanced Settings 1| - |Advanced | 07.05 –13:25 Num Cal. Sweeps Process Type Split Value ⌂◄ ▲ ▼ ► Number of Calibration Sweeps: • Freely selectable number of individual measurements [Sweeps] used for determination of a calibration point. Process Type: • Select the operation mode: o one concentration [1 measuring range] o two concentrations [2 measuring ranges] o split concentration [1 measuring range with switching point (split value) for coefficient switchover]. Split Value: • Setting of the switching point on a value basis. 18 Micro-Polar Brix LB 565 Chapter 2. Software Functions 2.2.17 Calibr. manual 1 | - | Calibr. manual | 07.05 –13:25 Cal. Order Cal. Base Coefficients Start Calibr. ⌂◄ ▲ ▼ ► Cal. Order: • Here you define ear/quadratic] the calibration order [lin- Cal. Base: • The following parameters can be set: o Phase o Attenuation o Phase and attenuation Coefficients: • Here you can edit all coefficients for phase and attenuation. Start Calibr.: • Starts the calibration using the parameters you have set earlier. 2.2.18 Input / Output 1 | - |Input / Output | 07.05 –13:25 Current Output Current Input Pt 100 Digital Output Digital Input ⌂◄ ▲ ▼ ► Current Output: • Both outputs can be adjusted, assigned and set up on the selected level. Current Input: • Activation level of current input, calibration and display of the live current signal. Pt 100: • Here you can enable and adjust a connected Pt 100. Display of the actual temperature signal. Digital Output: • Allocation of relays 1 and 2 and test function. Digital Input: • Status control and assignment of the digital inputs. Micro-Polar Brix LB 565 19 Chapter 2. Software Functions 2.2.19 Current Output 1 | - |Current Output | 07.05 –13:25 Current Out 1 Current Out 2 ⌂◄ ▲ Current Out 1: • Select the setup display for output 1. Current Out 2: • Select the setup display for output 2. ▼ ► 2.2.20 Current Out 1 1 | - |Current Out 1| 07.05 –13:25 Assignment Upper value Lower value Test / Adjust Error current ⌂◄ ▲ Assignment: • The current output can be assigned to a concentration or one of the activated inputs. Upper value: • Display value assigned to the 20mA value. ▼ ► Lower value: • Display value assigned to the 4mA value. (Current output 1 only 4 – 20mA possible) Test/Adjust: • Current test, calibration and display of live current. (In case of test function, the measurement should be stopped.) Error current: • Status of current output in case of error o 22 mA o 3.5 mA o Hold o Value 20 Micro-Polar Brix LB 565 Chapter 2. Software Functions 2.2.21 Current Out 2 All functions same as current output 1 1 | - |Current Out 2| 07.05 –13:25 Assingnment Upper value Lower value Range Test / Adjust Error current ⌂◄ ▲ ▼ ► ! Current output 2 can be set from 0 – 20mA to 4 – 20mA. Range: • Change of the current output o 0 – 20mA o 4 – 20mA 2.2.22 Current Input 1 | - |Current Input |07.05 –13:25 Current In 1 Current In 2 ⌂◄ ▲ Current In 1: • When selected, change to activation and calibration menu. Current In 2: • As described above. ▼ ► 2.2.23 Current In 1 1 | - | Current In 1 |07.05 –13:25 Enabled DAC Trim Live current ⌂◄ ▲ Enabled: • If you select yes/no, the current input is enabled or disabled. DAC Trim: • Follow the instructions on the display. ▼ ► Live current: • Display of the live current signal. 2.2.24 Current In 2 Set and enabled same as input 1. Micro-Polar Brix LB 565 21 Chapter 2. Software Functions 2.2.25 Pt 100 1 | - | Pt 100 |07.05 –13:25 Enabled Trim Pt 100 Pt 100 live ⌂◄ ▲ ▼ ► Enabled: • If a Pt 100 is connected, the input has to be enabled first. Trim Pt 100: • You need a 100 Ohm and a 138.5 Ohm resistance. Follow the instructions on the display. Pt 100 live: o Display of the live temperature. 2.2.26 Digital Output 1 | - |Digital Output | 07.05 –13:25 Relay 1 Relay 2 Test ⌂◄ ▲ ▼ ► Relay 1: • Relay 1 can be assigned to different functions: o None o Error o Halt o No product o Alarm min o Alarm max Relay 2: • Same assignments possible as above. Test: • 22 The switching status of the relays can be set here and checked at the respective terminals. Micro-Polar Brix LB 565 Chapter 2. Software Functions 2.2.27 Digital Input 1 | - |Digital Input | 07.05 –13:25 Status DI 1 function DI 2 function DI 3 function ⌂◄ ▲ ▼ ► Status: • Shows the status of the input circuit o open/closed DI 1 Function: • The following functions can be assigned to DI 1: o None o Start (external start) DI 2 Function: • The following functions can be assigned to DI 2: o None o Hold (averaging is stopped) o Product (external product selection) DI 3 Function: • Assignments for DI 3 o None o Sample (external control of sampling) o Product (external product selection) Micro-Polar Brix LB 565 23 Chapter 3. Configuration Chapter 3. Configuration Before carrying out any calibration work, you should check the configuration setup of the measuring system and, if necessary, correct it. 3.1 Configuration Setup 1 | - |Setup | 07.05 – 13:25 Configuration Calibration Input / Output Service Product Change password ⌂◄ ▲ ▼ CONFIGURATION ► 3.1.1 General Data 1 | - |Configuration | 07.05 – 13:25 General Data Measurement Plausibility Microwave ⌂◄ ▲ GENERAL DATA ▼ ► 1 | - | General Data| 07.05 – 13:25 Date Time Tag ⌂◄ 07.05.2004 13:25 App.1 ▲ ▼ Example: Select the respective entry, edit and store it. DATE ► 1 | - | General Data | 07.05 –13:25 Date Push DEL to delete the entry and then enter the new date. Push..√.. to confirm and store the changed date. 07.05.2004 | ESC 24 ..?.. DEL ..√.. Micro-Polar Brix LB 565 Chapter 3. Configuration 3.1.2 Measurement 1 | - |Measurement| 07.05 – 13:25 Meas. Mode Continuous Start Mode Keypad Averaging 20 Reset Averaging no Markers Units ⌂◄ ▲ ▼ ► You have to check the settings on this display and adapt them to the measurement conditions. For example, you have to adapt the measurement mode, the start mode and the averaging to the actual operating conditions. With continuous measurements, the averaged values do not have to be reset. You can set a marker comprising max. 5 characters which identify the value set on in the live display. Set the units as desired. UNITS 1 | - | Units | 07.05 – 13:25 Conc. Conc. 2 Temp. ⌂◄ ▲ % TS °C ▼ CONC. ► You can select three different units or no unit (none). 1| - | Units | 07.05 – 13:25 Conc. none % DS ° Bx ESC ▲▼ ..√.. Different units can be set for both concentrations. 1 | - | Units | 07.05 – 13:25 Conc. Conc. 2 Temp ⌂◄ ▲ % TS °C ▼ TEMP ► Micro-Polar Brix LB 565 25 Chapter 3. Configuration 1| - | Units | Temp. 07.05 – 13:25 none °C °F °C ESC ▲▼ ..√.. The temperature input can be set to °C, °F or none. 3.1.3 Plausibility 1 | - | Plausibility | 07.05 – 13:25 Process limits Phase Measure Pause Detection ⌂◄ ▲ ▼ ► To rule out any unnecessary disturbances during calibration, the process limits should be set, as far as possible, below or above the measuring range. You should keep the factory-set default values for the phase measurement and disable Pause detection. If a default value has to be changed, you have to check all entries that are relevant for the calibration and, if necessary, renew them. Upon completion of the calibration work, you can enable Pause detection again. 3.1.4 Microwave 1 | - | Microwave | 07.05 – 13:25 Cables CABLES ⌂◄ ▲ 1| - | Cables ▼ | 07.05 – 13:25 Refer. cable length Signal cable length ⌂◄ 26 ▲ ► ▼ 4m 4m ► If the factory-set cable lengths does not match the actual geometry conditions, you have to correct the values. Example: For a 2 m long HF quad cable, enter 4 m for the reference and signal cable length. The input value corresponds to twice the quad cable length. Micro-Polar Brix LB 565 Chapter 4. Calibration Chapter 4. Calibration Prerequisite: The measuring system has been connected properly and the normal operating temperature has been reached (approx. 30 to 45 min.). If the instrument is still in the delivery status, you may start calibration with section 4.1 System Calibration. If the manufacture’s default settings have been lost, you have to start with section 7.1 Factory Settings. 4.1 System Calibration 1| - | LB 565 | 07.05 – 13:25 Live Display Diagnostic Setup Access Level Language RUN Basic English ▲ ▼ 1 | - | Setup | ▲ SETUP ► 07.05 – 13:25 Configuartion Calibration Input / Output Product Change password ⌂◄ If you turn on the measuring system, the following display appears: ▼ CALIBRATION ► 1 | - | Calibration | 07.05 – 13:25 System Adjust Calibrate Conc ⌂◄ ▲ ▼ SYSTEM ADJUST ► Micro-Polar Brix LB 565 27 Chapter 4. Calibration 1 | - |System Adjust |07.05 – 13:25 Adjust Num. Sweeps ⌂◄ ▲ 10 ▼ The manufacturer has set the number of measurement cycles (sweeps) to 10. ADJUST ► 1 | - |System Adjust|07.05 – 13:25 System adjustment now ? ..X.. Confirm ..√.. During system calibration, a sample should be taken and analyzed. This analysis value is needed to calculate the offset value. 1 | - |System Adjust|07.05 – 13:25 Adjustment in procress... System adjustment is in process. …............................................. 1 | - |System Adjust|07.05 – 13:25 Adjusted ! Push OK to confirm and push ⌂◄ three times to return to the main menu. ..OK.. Fine tuning (calibration) is required to optimize the reading. 28 Micro-Polar Brix LB 565 Chapter 4. Calibration 4.2 Start Calibration 1| - | LB 565 | 07.05 – 13:25 Live Diplay Diagnostic Setup Access Level Language RUN ▲ Basic English ▼ The standard coefficients required for your application have already been set up by the manufacturer. Push RUN to start the measuring system. ► 1 | - |Toggle Run Mode|07.05 –13:25 Push ..√.. to confirm this prompt and the instrument switches to the run mode. Switch Run Mode ? ..X.. ..√.. 4.2.1 Tuning The display to the left appears if you push RUN. 1 | - | Live Display | 07.05 – 13:25 Concentration 1 av 65.50 % Conc av. Conc act 64.35% ESC PROBE ..▲▼.. ZOOM The live display is now compared with the analysis value of the system calibration sample. The difference has to be entered with the correct sign character as offset Calculation: Analysis value – display = offset Push ESC to return to the main menu. 1| - | LB 565 | 07.05 – 13:25 Live Display Diagnostic Setup Access Level Language STOP ▲ SETUP Basic English ▼ Micro-Polar Brix LB 565 ► 29 Chapter 4. Calibration 1 | - |Setup | 07.05 – 13:25 Configuration Calibration Input / Output Service Product ⌂◄ ▲ CALIBRATION ▼ ► 1 | - | Calibration | 07.05 – 13:25 System Adjust Calibrate Conc Advanced ⌂◄ ▲ CALIBRATE CONC ▼ ► 1 | - |Calibrate Conc| 07.05 –13:25 Sampling Calibr. autom. Tuning View ⌂◄ ▲ TUNING ▼ 1 | - | Tuning | 07.05 –13:25 Factor Offset ⌂◄ ► 1.00000 0.000 ▲ 1 | - | Tuning | ▼ ► 07.05 –13:25 Enter the calculated offset value, confirm with ..√.. and push the Home button four times to return to the main menu. Offset 0.000 | ESC 30 ..?.. OFFSET DEL ..√.. Micro-Polar Brix LB 565 Chapter 4. Calibration 1| - | LB 565 | 07.05 – 13:25 Select Live Display Diagnostic Setup Access Level Language STOP ▲ Basic English ▼ ► LIVE DISPLAY to return to the display. 1 | - | Live Display | 07.05 – 13:25 Concentration 1 av 75.50 % Conc av. The display value should now correspond to the actual value. Conc act 64.35% ESC PROBE ..▲▼.. ZOOM Micro-Polar Brix LB 565 31 Chapter 4. Calibration 4.2.2 Sampling The display to the left appears if you push RUN. 1 | - | Live Display | 07.05 – 13:25 Concentration 1 av 65.50 % Conc av. Conc act 64.35% ESC PROBE ..▲▼.. ZOOM Note: Push the PROBE button to start measurement of the raw data. At the same time, the laboratory sample has to be taken and marked. The analysis may be performed later, provided the product is not changed by this. 1 | - | Live Display | 07.05 – 13:25 Taking Sample # 1 Sampling is in process...... Push the ..X.. button to stop the sample taking any time. Conc av. ..X.. Conc act 64.35% ..▲▼.. 1 | - | Live Display | 07.05 – 13:25 Save sample no. 1? Conc av. Conc act 64.35% ..X.. ..▲▼.. If the sample taking took place without any problem, push the ..√.. button to store the sample in the table and the measurement continues. ..√.. You have to repeat the process described above for each further sample. The second sample taking should be started only when the display shows a significant difference to the first sample taking. 32 Micro-Polar Brix LB 565 Chapter 4. Calibration 4.2.3 Entering the Laboratory Values 1 | - | Live Display | 07.05 – 13:25 Concentration 1 av 65.50 % Conc av. Conc act 64.35% ESC PROBE ..▲▼.. 1| Push the ESC button to go to the main menu. A measurement can be stopped only on the main menu. ZOOM - | LB 565 | 07.05 – 13:25 Live Display Diagnostic Setup Access Level Language Stop ▲ Push STOP to stop the measuring system. Basic English ▼ ► 1 | - |Toggle Run Mode|07.05 –13:25 Push ..√.. to confirm the prompt and the measurement switches to the STOP mode. Switch Run Mode ? ..X.. 1| ..√.. - | LB 565 | 07.05 – 13:25 Live Display Diagnostic Setup Access Level Language RUN ▲ 1 | - |Setup | Basic English ▼ ► 07.05 – 13:25 Comfiguration Calibration Input / Output Service Product Change password ⌂◄ ▲ SETUP ▼ CALIBRATION ► Micro-Polar Brix LB 565 33 Chapter 4. Calibration 1 | - | Calibration | 07.05 – 13:25 System Adjust Calibrate Conc Advanced ⌂◄ ▲ CALIBRATE CONC ▼ ► 1 | - |Calibrate Conc| 07.05 –13:25 Sampling Calibr. autom. Calibr. manual Tuning View ⌂◄ ▲ SAMPLING ▼ ► 1 | 1/1 | Sample # 1 | 07.05 –13:25 Next sample Actice Meas value Lab value Advanced ◄ DEL Yes 65.50 % 0.00 % ▲▼ LAB VALUE ..√.. 1 | 1/1 | Sample # 1 | 07.05 –13:25 Lab value 0.00| ◄ DEL Push DEL to delete default value, enter new value and push ..√.. to confirm. ▲▼ ..√.. 1 | 1/1 | Sample # 1 | 07.05 –13:25 Next sample Active Meas value Lab value Advanced ◄ DEL Yes 65.50 % 72.40 % ▲▼ ..√.. NEXT SAMPLE and repeat the step described above with the next sample. After you have entered the last sample, push the ◄ button to return to the calibration menu. (short push – one page, longer push of the button – you get back to the Calibration menu immediately) 34 Micro-Polar Brix LB 565 Chapter 4. Calibration 4.2.4 Automatic Calibration 1 | - |Calibrate Conc| 07.05 –13:25 Phase calibration including the enabled compensation is carried out during automatic calibration. Sampling Calibr. autom. Tuning View ⌂◄ ▲ ▼ ► 1 | - |Calibr. autom.| 07.05 –13:25 Start Calibr. ⌂◄ ▲ ▼ 1 | - | Calibration | START CALIBR. ► 07.05 –13:25 Calibrate now ? ..X.. CALIBR. AUTOM. Push the ..√.. button to start the calibration. ..√.. 1 | - | Calibrate Conc | 07.05 –13:25 Calibrated ! Push the ..OK.. button to confirm calibration. ..OK.. When calculating the new coefficient set, the factor and offset will be reset (factor 1.00000 and offset 0.000). 1 | - |Calibrate Conc| 07.05 –13:25 Sampling Calibr. autom. Tuning View ⌂◄ ▲ Push the ..⌂◄.. button four times to return to the main menu. ▼ ► Micro-Polar Brix LB 565 35 Chapter 4. Calibration 4.3 Manual Calibration Manual calibration is possible only on the Advanced level. Prerequisites for manual calibration are chapters 4.1 System Calibration 4.2 Start Calibration 4.2.1 Tuning 4.2.2 Sampling and 4.2.3 Entering the Laboratory Values 4.3.1 Manual Calibration with One Concentration 1 | - |Calibrate Conc| 07.05 –13:25 Sampling Calibr. autom. Calibr. manual Tuning View ⌂◄ ▲ ▼ If the display depicted to the left is not visible, do the following on the live display: ESC|SETUP|CALIBRATION|CALIBRATE CONC| CALIBR. MANUAL ► 1 | - | Calibr. manual | 07.05 –13:25 Cal. Order Cal. Base Coefficients Start Calibr. ⌂◄ ▲ Linear Phase ▼ CAL. ORDER ► 1 | - | Calibr. manual | 07.05 –13:25 Cal. Order Linear Quadratic ⌂◄ 36 ▲ LINEAR ▼ ► Quadratic calibration is possible only for a calibration with three and more samples. Micro-Polar Brix LB 565 Chapter 4. Calibration 1 | - | Calibr. manual | 07.05 –13:25 Cal. Order Cal. Basic Coefficients Start Calibr. ⌂◄ ▲ Linear Phase ▼ CAL. BASE ► 1 | - | Calibr. manual | 07.05 –13:25 Cal. Base PHI Phase Attenuation Both ⌂◄ ▲ ▼ ► (phase measurement) The calibration base is selected depending on the number of samples and their raw data. Initial calibration should be as simple as possible, since calibration can be optimized any time. 1 | - | Calibr. manual | 07.05 –13:25 Cal. Order Cal. Base Coefficients Start Calibr. ⌂◄ ▲ Linear Phase ▼ 1 | - | Calibration | ► 07.05 –13:25 Push the ..√.. button to start the calibration, push ..X.. to go back one page without calibration. Calibrate now ? ..X.. 1 | - | Calibration | START CALIBRATION ..√.. 07.05 –13:25 ..OK.. accepts the calibration and changes to the next display. Calibrated ! ..OK.. Micro-Polar Brix LB 565 37 Chapter 4. Calibration 1 | - | Calibration | 07.05 –13:25 The curve to the left shows the characteristic curve lab / measured value. ► 1 | - | Calibration | 07.05 –13:25 Correlation Lab/Meas. value 0.998726 ..►.. The correlation shows the average deviation of the curve from the sample series. ..OK.. ..OK.. 1 | - | Calibration | 07.05 –13:25 Calibration OK? ..X.. 38 ..√.. As soon as you confirm this prompt, the calibration display appears again; from there you get back to the main menu by pushing .⌂◄. four times and you can start the measurement again. Micro-Polar Brix LB 565 Chapter 4. Calibration 4.3.2 Calibration with Two Concentrations Calibration for two concentrations starts with changing the process type as described below. Prerequisites for manual calibration are chapters 4.1 System Calibration 4.2.1 Tuning and 4.2.2 Sampling 1 | - | Calibration | 07.05 – 13:25 System Adjust Calibrate Conc Advanced ⌂◄ ▲ ADVANCED ▼ 1| - | Advanced | ► 07.05 –13:25 Num. Cal. Sweeps Process Type ⌂◄ ▲ ▼ 1| - |Advanced | Process Type 1 Conc. 2 Conc. Split Conc. .ESC. 10 PROCESS TYPE ► 07.05 –13:25 2 CONC ..?.. ▲▼ ..√.. Push the ..√.. button to accept the selected process type and push the .⌂◄. button once to get to the following display. 1 | - | Calibration | 07.05 – 13:25 System Adjust Calibrate Conc Calibrate Conc2 Advanced ⌂◄ ▲ ▼ CALIBRATE CONCENTRATION 1 ► Micro-Polar Brix LB 565 39 Chapter 4. Calibration 1 | - |Calibrate Conc | 07.05 –13:25 Sampling Calibr. autom. Calibr. manual Tuning View ⌂◄ ▲ SAMPLING ▼ ► There is only one sample table for both calibrations. The lab values have to be entered for all samples used for calibration of concentration 1. All other samples have to be disabled (active.... yes/no). 1 | 1/4 | Sample # 1 | 07.05 –13:25 Next sample Active Meas value Lab value Advanced ◄ DEL Yes 65.50 % 0.00 % ▲▼ LAB VALUE ..√.. 1 | 1/4 | Sample # 1 | 07.05 – 13:25 Lab value 60.40 | ESC ..?.. DEL Delete default value with DEL, enter new value and confirm with ..√... ..√.. 1 |1 /4 | Sample # 1 | 07.05 –13:25 Next sample Active Meas value Lab value Advanced ◄ DEL Yes 65.50 % 60.40 % ▲▼ NEXT SAMPLE with next sample. ..√.. 1 |2 /4 | Sample # 2 | 07.05 –13:25 Next sample Active Meas value Lab value Advanced ◄ 40 DEL Yes 74.35 % 67.80 % ACTIVE Disable sample. ▲▼ ..√.. Micro-Polar Brix LB 565 Chapter 4. Calibration 1 | 2/4 | Sample # 2 | 07.05 – 13:25 Active No Yes ESC NO ..?.. DEL ..√.. 1 |2 /4 | Sample # 2 | 07.05 –13:25 Next sample Active Meas value Lab value Advanced ◄ DEL No 74.35 % 67.80 % ▲▼ Sampling Calibr. autom. Calibr. manual Tuning View ▲ Push ◄ to return to the calibration page. ..√.. 1 | - |Calibrate Conc | 07.05 –13:25 ⌂◄ Make sure that all samples have been processed and only those samples are active which are relevant for this calibration. ▼ Automatic as well as manual calibration is possible in this calibration mode. For non-professional users we recommend the automatic mode. ► 1 | - |Calibr. autom. |07.05 –13:25 START CALIBRATION Start Calibr. ⌂◄ ▲ ▼ ► 1 | - | Calibration | 07.05 –13:25 Push the ..√.. button to start the calibration; push ..X.. to return one page without calibration. Calibrate now ? ..X.. ..√.. Micro-Polar Brix LB 565 41 Chapter 4. Calibration 1 | - | Calibration | 07.05 –13:25 Push ..OK.. to accept the calibration and to change to the next display. Calibrated ! ..OK.. 1 | - |Calibr. autom. |07.05 –13:25 Start Calibr. ⌂◄ ▲ Push .⌂◄. twice to return two pages. ▼ ► 1 | - | Calibration | 07.05 – 13:25 System Adjust Calibrate Conc Calibrate Conc2 Advanced ⌂◄ ▲ ▼ CALIBRATE CONCENTRATION 2 ► Repeat the steps as described above for concentration 2; all samples have to be enabled again in the sample table. Now you have to disable all samples which are not used for concentration 2. 1 | - |Calibrate Conc2|07.05 –13:25 Sampling Calibr. autom. Calibr. manual Tuning View ⌂◄ 42 ▲ ▼ SAMPLING ► Micro-Polar Brix LB 565 Chapter 4. Calibration 4.3.3 Calibration with Split Value With this type of calibration, two characteristic curves (concentrations) are combined in one measuring range; their point of intersection defines the split value. 1 | - | Calibration | 07.05 – 13:25 System Adjust Calibrate Conc Advanced ⌂◄ ▲ ADVANCED ▼ 1| - | Advanced | ► 07.05 –13:25 Num Cal. Sweeps Process Type ⌂◄ ▲ ▼ 1| - | Advanced | Process Type 1 Conc. 2 Conc. Split Conc. .ESC. 10 ► 07.05 –13:25 SPLIT CONC ..?.. ▲▼ 1| - | Advanced ..√.. |07.05 –13:25 Num. Cal. Sweeps 10 Process Type Split Conc. Split Value 75.00 % ⌂◄ ▲ PROCESS TYPE ▼ ► Micro-Polar Brix LB 565 Push the ..√.. button to accept the selected process type and push the .⌂◄. button once to go to the display depicted below. The displayed split value has been set by the manufacturer, but has to be adapted to the respective application. The sample measurement should be selected such that the last sample of the lower concentration is fairly close to the first sample measurement of the upper concentration. Ideally, the last sample of the initial concentration is the first sample of the final concentration. 43 Chapter 4. Calibration 1 | - | Live Display | 07.05 – 13:25 The sample measurement is carried out continuously over the entire measuring range with the display depicted to the left. See chapter 4.2.2 Sampling. Concentration 1 av. 65.50 % Conc av. Conc act 64.35% ESC PROBE ..▲▼.. ZOOM After completion of sampling, the individual samples will be enabled or disabled during input of the laboratory values, relative to the set split values. All samples smaller or equal to the split value will be assigned to the lower concentration range and all samples above to the upper concentration range. The split value to be set must correspond to the point of intersection of both calibration curves. 1| - |Advanced | 07.05 –13:25 Num. Cal. Sweeps 10 Process Type Split Conc. Split Value 75.00 % ⌂◄ ▲ ▼ 1| - | Advanced | ► 07.05 –13:25 Split Value Enter the split value and confirm with ..√... 75.00 | ESC SPLIT VALUE ..?.. DEL ..√.. Push the Home button .⌂◄. to return to the calibration page. 1 | - | Calibration | 07.05 – 13:25 System Adjust Calibrate Conc Calibrate Conc2 Advanced ⌂◄ ▲ ▼ CALIBRATE CONC ► 1 | - |Calibrate Conc | 07.05 –13:25 CALIBR. AUTOM. Sampling Calibr. autom. Calibr. manual Tuning View ⌂◄ 44 ▲ ▼ ► The lower concentration is now calibrated. Then select CONC2 and repeat the calibration process. Back to the main menu and start the measurement. Micro-Polar Brix LB 565 Chapter 4. Calibration 4.3.4 Calibration with Temperature Compensation Before running a sample measurement you have to enable the desired compensation input and check the calibration. If all inputs are enabled, the measured values of all inputs will be stored automatically in the sample table. 1 | - |Calibrate Conc| 07.05 –13:25 Sampling Calibr. autom. Calibr. manual Tuning View ⌂◄ ▲ From the main menu, you get in the Advanced mode to the display to the left via | SETUP | CALIBRATION | CALIBRATE CONC. | CALIBR. MANUAL ▼ ► 1 | - | Calibr. manual | 07.05 –13:25 Cal. Order Cal. Base Coefficients Compensation Start Calibr. ⌂◄ ▲ Linear Phase ▼ ► 1 | - |Compensation | 07.05 –13:25 Input None Mode Additive Order Linear Ref. Temp. 0.00 TC_Phi 1 0.00000 TC_dB 1 0.00000 ESC ..?.. COMPENSATION ▲▼ INPUT ..√.. If all inputs have been enabled during sample measurement, you have the option to select a compensation from the list, since all input values have been stored in the sample table. 1 | - |Compensation| 07.05 –13:25 Input None Current In 1 Current In 2 Pt 100 ⌂◄ ▲ ▼ Pt 100 ► Micro-Polar Brix LB 565 45 Chapter 4. Calibration You can select additive or multiplicative mode and set the order to linear or quadratic. If you select automatic calibration mode, the above modes will be calculated automatically. This is recommended for non-professional users. 1 | - |Compensation | 07.05 –13:25 Input Pt 100 Mode Additive Order Linear Ref. Temp. 0.00°C TC_Phi 1 0.00000 TC_dB 1 0.00000 ESC ..?.. ▲▼ REF TEMP ..√.. 1 | - |Compensation | 07.05 –13:25 Ref. Temp. 80.00 | ESC °C ..?.. DEL ..√.. 1 | - |Compensation | 07.05 –13:25 Input Pt 100 Mode Additive Order Linear Ref.Temp. 80.00°C TC_Phi 1 0.00000 TC_dB 1 0.00000 ESC ..?.. ▲▼ Enter the average operating temperature and confirm. The coefficients TC_Phi 1 and TC_dB 1 are automatically calculated during calibration. ..√.. If you have completed the entries described above and have carried out the steps described in chapters 4.1 System calibration 4.2.1 Tuning 4.2.2 Sampling and 4.2.3 Entering the Lab Values you may proceed with the calibration as described below. 1 | - |Calibrate Conc1|07.05 –13:25 Sampling Calibr. autom. Calibr. manual Tuning View ⌂◄ 46 ▲ ▼ CALIBR. AUTOM. ► Micro-Polar Brix LB 565 Chapter 4. Calibration 1 | - |Calibr. autom. |07.05 –13:25 Start Calibr. Compensation ⌂◄ ▲ ▼ Pt 100 START CALIBR. ► 1 | - | Calibr. autom. | 07.05 –13:25 Start calibration process. Calibrate now ? ..X.. ..√.. 1 | - | Calibr. autom. | 07.05 –13:25 Calibrating... ..OK.. 1 | - | Calibr. autom. | 07.05 –13:25 Calibrated ! ..OK.. Micro-Polar Brix LB 565 Push .O.K. to confirm the calibration. Calibration is finished. Push the Home button .⌂◄. four times to return to the main menu and to start a measurement. 47 Chapter 5. Password Chapter 5. Password The measuring system can be protected by passwords against unauthorized access. 5.1 Password The following access levels are available. Read only The measuring system cannot be started and stopped. You can only switch from the live display to Diagnostic and to Access Level. Basic On the Basic level you can make essential entries, and stop and start the system. Profi The Profi mode allows additional entries in the process type menu, calibration menu and opens the Service menu. Service The Service level is reserved for technical service. You have to enter a password to change from the access level „Read only“ to „Basic“ or „Profi“. At the time of delivery, this password is PASS1 Changing from Profi to Basic or vice versa is possible without password. You can change the password on the Profi or Basic level. 48 Micro-Polar Brix LB 565 Chapter 6. Inputs / Outputs Chapter 6. Inputs / Outputs The measuring system includes two separate floating current outputs. 6.1 Current Outputs Current outputs 1 and 2 can be assigned to the concentrations for calibration with two concentrations. You get to the setup display as follows. 1| - | LB 565 | 07.05 – 13:25 Live Diplay Diagnostic Setup Access Level Language RUN SETUP Basic English ▲ ▼ 1 | - |Setup | ► 07.05 – 13:25 Configuration Calibration Input / Output Service Product ⌂◄ ▲ INPUT/OUTPUT ▼ ► 1 | - |Input / Output | 07.05 –13:25 CURRENT OUTPUT Current Output Current Input Pt 100 Digital Output Digital Input ⌂◄ ▲ ▼ ► 1 | - |Current Output | 07.05 –13:25 Current Out 1 Current Out 2 ⌂◄ ▲ ▼ Now you can select the respective current output and assign it to the concentration after calibration. ► Micro-Polar Brix LB 565 49 Chapter 6. Inputs / Outputs 6.1.1 Current Output Setup 1 | - |Current Output | 07.05 –13:25 Current Out 1 Current Out 2 ⌂◄ ▲ CURRENT_OUT 1 ▼ ► 1 | - | Current Out 1 | 07.05 –13:25 Assignmet Upper Value Lower Value Test / Adjust Error current ⌂◄ ▲ Conc. 95.00 % 60.00 % ▼ ► 1 | - | Current Out 1|07.05 –13:25 Assignment Conc. Conc. 2 Current In 1 Current In 2 Pt 100 .ESC. ..?.. ASSIGNMENT ▲▼ CONC Select and confirm with ..√... ..√.. In case of two concentrations and outputs enabled, the assignment can be chosen as needed. 1 | - |Current Out 1|07.05 –13:25 Assignment Upper value Lower value Test / Adjust Error current ⌂◄ ▲ Conc. 95.00 % 60.00 % ▼ UPPER VALUE ► 1 | - |Current Out 1| 07.05 –13:25 Upper value 95.00 | .ESC. ..?.. DEL Delete the old value with DEL. Enter the new limit value and confirm with ..√... ..√.. Set the lower value also as described above. 50 Micro-Polar Brix LB 565 Chapter 6. Inputs / Outputs 6.1.2 Test and Adjustment 1 | - |Current Out 1|07.05 –13:25 Assignment Upper value Lower value Test / Adjust Error current ⌂◄ ▲ Conc. 95.00 % 60.00 % TEST / ADJUST ▼ ► 1 | - |Test / Adjust | 07.05 –13:25 Loop test DAC trim Live current ⌂◄ ▲ 4.00 mA ▼ ► DAC TRIM ! CAUTION: Current output 1 can only be set between 4 and 20 mA, since it is foreseen for a Hart® communicator. 1 | - | Trim Cout | 07.05 –13:25 Loop should be removed from automatic control ..X.. Push .OK. to confirm that the process is not affected by the measurement. .OK. 1 | - | Trim Cout | 07.05 –13:25 Connect reference meter! ..X.. Push .OK. to confirm that the measuring system is connected. .OK. 1 | - | TrimCout | 07.05 –13:25 Setting Current Output To 4 mA ..X.. Push .OK. to confirm. .OK. Micro-Polar Brix LB 565 51 Chapter 6. Inputs / Outputs 1 | - | Trim Cout | 07.05 –13:25 Enter meter value 3.80 .ESC. mA .DEL. Read off display of measuring system and enter value. ..√.. 1 | - | Trim Cout | 07.05 –13:25 4 mA adjusted 4 mA value adjusted. ..X.. .OK. 1 | - | Trim Cout | 07.05 –13:25 Setting Current Output to 20 mA ..X.. Push the .OK. button to confirm. .OK. 1 | - | Trim Cout | 07.05 –13:25 Enter meter value 19.80 | .ESC. mA .DEL. Read off display of measuring system and enter value. ..√.. 1 | - | Trim Cout | 07.05 –13:25 20 mA adjusted ..X.. Adjustment finished. .OK. 1 | - |Test / Adjust | 07.05 –13:25 Loop test DAC trim Live current ⌂◄ 52 ▲ 15.00 mA ▼ To check the current loop and possibly connected remote displays, you can set a current between 4 and 20 mA via the test function. If you quit the test function, the system automatically switches back to the live current. ► Micro-Polar Brix LB 565 Chapter 6. Inputs / Outputs 6.1.3 Error Current Different signal effects can be assigned to the output current. 1 | - |Current Out 1|07.05 –13:25 Assignment Upper value Lower value Test / Adjust Error current ⌂◄ ▲ Conc. 95.00 % 60.00 % ▼ ERROR CURRENT ► 1 | - | Error current | 07.05 –13:25 Alarm Mode Value 22.00 mA ALARM MODE ⌂◄ ▲ ▼ ► Fixed values, Hold or freely adjustable values between 0 and 24 mA can be assigned. 1 | - | Error current |07.05 –13:25 Alarm mode VALUE 22.00 mA 3.5 mA Hold Value ⌂◄ ▲ ▼ ► 1 | - |Error current| 07.05 –13:25 Value 3.50 ESC With this setting, you can default any current value for the error case. CAUTION: The current output 1 can only be set between 3.5 mA and 24 mA. [HART®] mA DEL ..√.. Enter value and confirm with ..√... Micro-Polar Brix LB 565 53 Chapter 6. Inputs / Outputs 6.1.4 Current Output 2 1 | - |Current Output | 07.05 –13:25 Current Out 1 Current Out 2 ⌂◄ ▲ CURRENT OUT 2 ▼ ► 1 | - |Current Out 2|07.05 –13:25 Assignment Conc.2 Upper value 95.00 % Lower value 60.00 % Range 4 - 20 mA Test / Adjust Error current ⌂◄ ▲ ▼ 0 – 20 mA 4 – 20 mA 54 ▲▼ RANGE ► 1 | - |Current Out 2|07.05 –13:25 Range .ESC. All settings for current output 2 have to be made in the same manner as for output 1, with the exception of the range setting. After selection of the required range, carry out all setting and calibration steps as described in chapter 6.1.2. ..√.. Micro-Polar Brix LB 565 Chapter 6. Inputs / Outputs 6.2 Current Inputs If the window below is not displayed, you can invoke it on the main menu via |SETUP |INPUT/OUTPUT | 1 | - |Input / Output | 07.05 –13:25 Current Output Current Input Pt 100 Digital Output Digital Input ⌂◄ ▲ CURRENT INPUT ▼ ► 6.2.1 Enabling the Current Input 1 | - |Current Input | 07.05 –13:25 Current In 1 Current In 2 ⌂◄ ▲ CURRENT IN 1 ▼ ► 1 | - |Current In 1 | 07.05 –13:25 Enabled DAC trim Live current ⌂◄ ▲ nein 4.00 mA ▼ ENABLED ► 1 | - |Current In 1 | 07.05 –13:25 Enabled no yes .ESC. ENABLING ..?.. ▲▼ ..√.. If a measurement is running, enabling an adjusted current input which is not used may cause an error. Micro-Polar Brix LB 565 55 Chapter 6. Inputs / Outputs 6.2.2 Range Setting and Adjustment 1 | - | Current In 1 | 07.05 –13:25 Enabled ja Range 0 – 20 mA Upper value 100.00 °C Lower value 0.00 ° C DAC trim Live current 4.00 mA ⌂◄ ▲ ▼ RANGE SETTING ► 1 | - |Current In 1| 07.05 –13:25 Range 0 – 20 mA 4 – 20 mA .ESC. 4 – 20 mA ▲▼ ..√.. 1 | - | Current In 1 | 07.05 –13:25 Enabled yes Range 4 – 20 mA Upper value 100.00 °C Lower value 0.00 ° C DAC trim Live current 4.00 mA ⌂◄ ▲ ▼ DAC trim ► 1 | - | Trim Cin | 07.05 –13:25 Loop should be removed from automatic control! ..X.. Push .OK. to confirm that the process is not affected by the measurement. .OK. 1 | - | Trim Cin | 07.05 –13:25 Connect current generator! ..X.. 56 Push .OK. to confirm that the current generator is connected. .OK. Micro-Polar Brix LB 565 Chapter 6. Inputs / Outputs 1 | - | Trim Cin | 07.05 –13:25 Set current generator to 4 mA ..X.. 1 | - | Trim Cin Set current generator to 4 mA. .OK. | 07.05 –13:25 Generator set to 4 mA? ..X.. 1 | - | Trim Cin .OK. | 07.05 –13:25 4 mA adjusted. Push .OK. to confirm adjustment of the lower value. ..X.. 1 | - | Trim_Cin .OK. | 07.05 –13:25 Set current generator to 20 mA ..X.. 1 | - | Trim Cin Set current generator to 20 mA. .OK. | 07.05 –13:25 Generator set to 20 mA? ..X.. 1 | - | Trim Cin Push the .OK. button to confirm. .OK. | 07.05 –13:25 Push .OK. to confirm adjustment of the lower value. 20 mA adjusted. ..X.. Push the .OK. button to confirm. .OK. Micro-Polar Brix LB 565 57 Chapter 6. Inputs / Outputs 1 | - |Current In 1 | 07.05 –13:25 Enabled ja Range 4 – 20 mA Upper value 100.00 °C Lower value 0.00 ° C DAC trim Live current 4.00 mA ⌂◄ ▲ ▼ Adjustment finished. The live current is displayed. ► If necessary, carry out range setting and calibration of current input 2 as described above. 6.3 Pt 100 Before enabling, you have to stop the measuring system, as the change in configuration may lead to errors. 6.3.1 Pt 100 Enabling 1 | - |Input / Output | 07.05 –13:25 Current Output Current Input Pt 100 Digital Output Digital Input ⌂◄ ▲ 1|-| Pt 100 PT 100 ▼ | 07.05 –13:25 Enabled Trim Pt 100 Pt 100 live ⌂◄ ▲ 1|-| Pt 100 ► no ENABLED ▼ ► | 07.05 –13:25 Enabled no yes ESC 58 Select „yes“ and confirm with ..√…. ▲▼ ..√.. Micro-Polar Brix LB 565 Chapter 6. Inputs / Outputs 6.3.2 Pt 100 Calibration 1|-| Pt 100 | 07.05 –13:25 Enabled Trim Pt 100 Pt 100 live ⌂◄ ▲ yes TRIM Pt 100 ▼ ► 1 | - |Trim Pt 100 | 07.05 –13:25 Connect 100 Ohm resistor to Pt 100 terminals ! (for 0°C / 31.4°F adjustment) ..X.. ..OK.. .OK. Connect 100 Ohm resistor to the Pt100 terminals [11] [23]. 1 | - |Trim Pt 100 | 07.05 –13:25 100 Ohm resistor connected? ..X.. Confirm once more with .O.K.. .OK. 1 | - |Trim Pt 100 | 07.05 –13:25 Lower point adjusted ..X.. Adjustment of lower point finished. .OK. 1 | - |Trim Pt 100 | 07.05 –13:25 Connect 138,5 Ohm resistor to Pt 100 terminals ! (for 100°C / 211.4°F adjustment) Micro-Polar Brix LB 565 After connection of the resistor, confirm with ..OK... 59 Chapter 6. Inputs / Outputs 1 | - |Trim Pt 100 | 07.05 –13:25 138.5 Ohm resistor connected? ..X.. Confirm prompt. .OK. 1 | - |Trim Pt 100 | 07.05 –13:25 Upper point adjusted. ..X.. Calibration is finished. .OK. 6.4 Digital Output The measuring system includes two changeover relay outputs which can be assigned to the respective application. Relay 1 is associated with LED signal 1 and relay 2 with signal 2. 6.4.1 Digital Output Assignment 1 | - |Digital Output | 07.05 –13:25 Relay 1 Relay 2 Test From the main menu you get to the display depicted to the left via | Setup | Input/Output | Digital Output RELAY 2 ⌂◄ 1|-| ▲ Relay 2 ▼ ► | 07.05 –13:25 Function No product Select the display with the arrow keys. ⌂◄ 60 ▲ ▼ ► Micro-Polar Brix LB 565 Chapter 6. Inputs / Outputs 1|-| Relay 2 Function None Error Hold meas. No product Alarm min Alarm max ESC ..?.. | 07.05 –13:25 Push ▲▼ and then ..√.. to assign a function to the relay. ▲▼ ..√.. Function Description None Relay and LED function disabled Error In case of error, relay and LED will be set. Hold meas. If Hold function is enabled, relay and LED will be set. No product If Pause detection is enabled, this will be signaled via relay and LED. Alarm min. The relay switches if the value falls below the limit value to be set. Alarm max. The relay switches if the value exceeds the limit value to be set. 6.5 Digital Input Different functions can be assigned to the digital inputs. See table below: Function1 Function2 Function3 Terminals DI 1 None Start DI 2 None Hold Product 13/25 DI 3 None Sample Product 14/26 12/24 For external start function, the start function has to be set to external in the Measurement menu window. Hold means that averaging is stopped, but the measurement continues to run. Sample means that sampling is started by closing the contact. Micro-Polar Brix LB 565 61 Chapter 6. Inputs / Outputs 6.5.1 External Product Switchover 1 | - |Digital Input |07.05 –13:25 Status DI 1 function DI 2 function DI 3 function ⌂◄ ▲ None None None ▼ From the main menu you get to the window display depicted to the left via |SETUP |INPUT/OUTPUT |DIGITAL IINPUT. DI 2 FUNCTION ► 1 | - |Digital Input |07.05 –13:25 DI 2 function PRODUCT None Hold Product ESC DEL ▲▼ ..√.. DI 3 also has to be set to product. Please take the terminal assignment from the table below. 1 | - |Digital Input |07.05 –13:25 Status DI 1 function DI 2 function DI 3 function ⌂◄ ▲ None Produkt None ▼ ► Terminals DI 2 13 / 25 DI 3 14 / 26 Product 1 Open Open Product 2 Closed Open Product 3 Open Closed Product 4 Closed Closed CAUTION! If you change to another product, the product-specific data will be loaded automatically, including: Configurations data System calibration Calibration data Input/Output definitions 62 Micro-Polar Brix LB 565 Chapter 7. Factory Settings Chapter 7. Factory Settings 1| - | LB 565 | 07.05 – 13:25 Live Display Diagnostic Setup Access Level Language RUN ▲ 1 | - | Setup | SETUP ▼ ► 07.05 – 13:25 Configuration Calibration Input / Output Service Product Change password ⌂◄ 1| This function allows you to reset the measuring system to its original status. ▲ SERVICE ▼ ► - | Service | 07.05 – 13:25 Factory settings FACTORY SETTINGS ⌂◄ ▲ ▼ ► 1| - |Factory Settings| 07.05 – 13:25 If you confirm this prompt, all parameters and configuration settings will be reset to their original status, with the exception of the sampling table and the reference measurement. Restore now? ..X.. ..√.. Micro-Polar Brix LB 565 63 Chapter 8. Error Lists Chapter 8. Error Lists 8.1 Error Lists All errors will be signaled by a red LED and a flashing RUN LED. Error displayed on the LCD have to be confirmed. After the error has been remedied, the LED’s are reset to the normal status. 8.1.1 Hardware Error Battery voltage Program memory faulty Data memory faulty I/O Communication trouble 8.1.2 Input Error Error Value too high Value too low Table is empty Chart data faulty No chart data available Sampling full 64 Cause Input value is too high Input value is too low Sampling has been selected without previous sample measurement The measuring system has determined faulty chart data during calibration. The calculates chart data have been deleted or calibration has not been completed. You have tried to measure more then 10 samples. Micro-Polar Brix LB 565 Chapter 8. Error Lists 8.1.3 Measurement Error Error Variance of phase is too large Attenuation too high Current input 1 out of range Current input 2 out of range Pt 100 temperature out of range Concentration out of range Concentration 2 out of range Possible cause The measured phase exceeds the permissible limit value. The measured attenuation exceeds the permissible maximum value. The enabled current input has not yet been calibrated or is not occupied. The enabled current input has not yet been calibrated or is not occupied. The enabled Pt 100 input has not yet been calibrated or is not occupied. The concentration calculated on the basis of the raw data lies outside the valid measuring range. The concentration calculated on the basis of the raw data lies outside the valid measuring range. All remedied measurement errors have to be confirmed with O.K. so that the measuring system can return to the live display. Micro-Polar Brix LB 565 65 Chapter 9. Calibration Data Sheet Chapter 9. Calibration Data Sheet 9.1 Configuration 9.1.1 General General Date Time Tag Default 01.06.04 13:00 Setting Default continuous keyboard 20 no Mark1 Setting Default 60.0 – 100.0 Setting 9.1.2 Measurement Measurement Meas. mode Start mode Averaging Reset averaging Marker Units 9.1.3 Plausibility Plausibility Process limits Phase measurement Variance Ratio Phi/dB Pause detection 100.00 6.0 yes -15.0 dB 9.1.4 Microwave Microwave Cable Ref. cable length Signal cable length 66 Default Setting 4.00 m 4.00 m Micro-Polar Brix LB 565 Chapter 9. Calibration Data Sheet 9.2 Product Product Product Default Setting 9.3 Input / Output 9.3.1 Current Output Current_out 1 Assignment Upper value Lower value Test/Calibration Error current Default Conc. 95.00 60.00 O.K. Hold Setup Current_out 2 Assignment Upper value Lower value Range Test/Calibration Error current Default conc. 2 95.00 60.00 4 – 20 mA O.K. Hold Setting Current_in 1 Enabled Range Upper value Lower value Calibration Live current Default no 4 – 20 mA 100.00 0.00 O.K. 17.05 mA Setting Current_in 2 Enabled Range Upper value Lower value Calibration Live current Default No 4 – 20 mA 100.00 0.00 O.K. 17.05 mA Setting 9.3.2 Current Input Micro-Polar Brix LB 565 67 Chapter 9. Calibration Data Sheet 9.3.3 Pt 100 Input Pt 100 Enabled Pt 100 Calibration Pt 100 Live Default no O.K. Setting 99.99 °C 9.3.4 Digital Output Digital output Relay 1 Relay 2 Default Error No Product Setting Default none none none Setting Default linear phase Setting 9.3.5 Digital Input Digital input DI 1 function DI 2 function DI 3 function 9.4 Calibration Data 9.4.1 Calibration Coefficients Calibration Calibration order Calibration Basic Coefficients A1 A2 B1 B2 Compensation Mode Order Ref. temp. TC_Phi 1 TC_Phi 2 TC_Attn 1 TC_Attn 2 68 -0.19 0.0 0.0 0.0 75.00 none additive linear 0.00 0.00000 0.00000 0.00000 0.00000 Micro-Polar Brix LB 565 Active Meas. value Micro-Polar Brix LB 565 10 No. 9.4.2 Sampling Table Lab value Current_in 1 Current _in 2 Pt 100 Phi (m) Attenuation 69 Calibration Data Sheet Notes 70 Micro-Polar Brix LB 565 Notes Micro-Polar Brix LB 565 71
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.3 Linearized : No Producer : Acrobat Distiller 4.05 for Windows Creator : Modify Date : 2004:11:03 13:25:41+01:00 Title : . Create Date : 2004:11:03 13:18:09 Page Count : 137 Page Mode : UseNoneEXIF Metadata provided by EXIF.tools