PHYSICIAN’S TECHNICAL MANUAL INGENIO™ PACEMAKER Model K172, K173, K174 CAUTION: Federal law (USA) restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. Table of Contents Additional Technical Information................................................................................. Device Description...................................................................................................... Related Information .................................................................................................... Indications and Usage ................................................................................................ Contraindications........................................................................................................ Warnings .................................................................................................................... Precautions................................................................................................................. Supplemental Precautionary Information ................................................................. Post-Therapy Pulse Generator Follow Up........................................................ Transcutaneous Electrical Nerve Stimulation (TENS)...................................... Electrocautery and Radio Frequency (RF) Ablation ......................................... Ionizing Radiation............................................................................................. Elevated Pressures .......................................................................................... Potential Adverse Events ......................................................................................... Mechanical Specifications ........................................................................................ Items Included in Package ....................................................................................... Symbols on Packaging ............................................................................................. Characteristics as Shipped....................................................................................... X-Ray Identifier......................................................................................................... Federal Communications Commission (FCC) .......................................................... Pulse Generator Longevity ....................................................................................... Warranty Information ................................................................................................ 27 27 29 31 33 34 37 39 41 42 45 47 48 49 55 Product Reliability..................................................................................................... Patient Counseling Information ................................................................................ Patient Handbook............................................................................................. Lead Connections..................................................................................................... Implanting the Pulse Generator................................................................................ Step A: Check Equipment.............................................................................. Step B: Interrogate and Check the Pulse Generator ..................................... Step C: Implant the Lead System .................................................................. Step D: Take Baseline Measurements........................................................... Step E: Form the Implantation Pocket ........................................................... Step F: Connect the Leads to the Pulse Generator....................................... Step G: Evaluate Lead Signals...................................................................... Step H: Program the Pulse Generator........................................................... Step I: Implant the Pulse Generator .............................................................. Step J: Complete and Return the Implantation Form .................................... Bidirectional Torque Wrench..................................................................................... Follow Up Testing ..................................................................................................... Predischarge Follow Up ................................................................................... Routine Follow Up ............................................................................................ Explantation.............................................................................................................. 55 56 57 58 59 60 60 61 62 64 66 71 72 72 73 74 76 77 77 79 ADDITIONAL TECHNICAL INFORMATION For additional technical reference guides, go to www.bostonscientific.com/ifu. Boston Scientific Corporation acquired Guidant Corporation in April 2006. During our transition period, you may see both the Boston Scientific and Guidant names on product and patient material. As we work through the transition, we will continue to offer doctors and their patients technologically advanced and high quality medical devices and therapies. DEVICE DESCRIPTION This manual contains information about the INGENIO family of implantable pacemakers, including the following types of pulse generators (specific models are listed in "Mechanical Specifications" on page 39): • SR—single chamber pacemaker providing ventricular or atrial pacing and sensing • DR—dual-chamber pacemaker providing ventricular and atrial pacing and sensing Therapies These pulse generators provide bradycardia pacing and adaptive rate pacing to detect and treat bradyarrhythmias. Leads The pulse generator has independently programmable outputs and accepts one or more of the following leads, depending on the model: • One IS-11 unipolar or bipolar atrial lead • One IS-1 unipolar or bipolar right ventricular lead NOTE: Single-chamber devices will accept either an IS-1 atrial or an IS-1 ventricular lead. The pulse generator and the leads constitute the implantable portion of the pulse generator system. PRM System These pulse generators can be used only with the ZOOM LATITUDE Programming System, which is the external portion of the pulse generator system and includes: • Model 3120 Programmer/Recorder/Monitor (PRM) • Model 2869 ZOOMVIEW Software Application • Model 6577 Accessory Telemetry Wand 1. IS-1 refers to the international standard ISO 5841-3:2000. You can use the PRM system to do the following: • Interrogate the pulse generator • Program the pulse generator to provide a variety of therapy options • Access the pulse generator’s diagnostic features • Perform noninvasive diagnostic testing • Access therapy history data RELATED INFORMATION Refer to the lead’s instruction manual for implant information, general warnings and precautions, indications, contraindications, and technical specifications. Read this material carefully for implant procedure instructions specific to the chosen lead configurations. Refer to the PRM system Operator’s Manual for specific information about the PRM such as setup, maintenance, and handling. INDICATIONS AND USAGE Boston Scientific pacemakers are indicated for treatment of the following conditions: • Symptomatic paroxysmal or permanent second- or third-degree AV block • Symptomatic bilateral bundle branch block • Symptomatic paroxysmal or transient sinus node dysfunction with or without associated AV conduction disorders (i.e., sinus bradycardia, sinus arrest, sinoatrial [SA] block • Bradycardia-tachycardia syndrome, to prevent symptomatic bradycardia or some forms of symptomatic tachyarrhythmias • Neurovascular (vaso-vagal) syndromes or hypersensitive carotid sinus syndromes Adaptive-rate pacing is indicated for patients exhibiting chronotropic incompetence and who may benefit from increased pacing rates concurrent with increases in minute ventilation and/or level of physical activity. Dual-chamber and atrial tracking modes are also indicated for patients who may benefit from maintenance of AV synchrony. Dual chamber modes are specifically indicated for treatment of the following: • Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block • VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm • Low cardiac output or congestive heart failure secondary to bradycardia CONTRAINDICATIONS These Boston Scientific pacemakers are contraindicated in patients who have a separate implanted cardioverter-defibrillator (ICD). Use of certain pacing modes and/or features available in these Boston Scientific pacemakers is contraindicated for the following patients under the circumstances listed: • Minute ventilation in patients with both unipolar atrial and ventricular leads • Single-chamber atrial pacing in patients with impaired AV nodal conduction • Atrial tracking modes for patients with chronic refractory atrial tachyarrhythmias (atrial fibrillation or flutter), which might trigger ventricular pacing • Dual-chamber and single-chamber atrial pacing in patients with chronic refractory atrial tachyarrhythmias • Asynchronous pacing in the presence (or likelihood) of competition between paced and intrinsic rhythms WARNINGS General • Labeling knowledge. Read this manual thoroughly before implantation to avoid damage to the pulse generator and/or lead. Such damage can result in patient injury or death. • For single patient use only. Do not reuse, reprocess, or resterilize. Reuse, reprocessing, or resterilization may compromise the structural integrity of the device and/or lead to device failure which, in turn, may result in patient injury, illness, or death. Reuse, reprocessing, or resterilization may also create a risk of contamination of the device and/or cause patient infection or cross-infection, including, but not limited to, the transmission of infectious disease(s) from one patient to another. Contamination of the device may lead to injury, illness, or death of the patient. • Backup defibrillation protection. Always have external defibrillation protection available during implant and electrophysiologic testing. If not terminated in a timely fashion, an induced ventricular tachyarrhythmia can result in the patient’s death. • Separate pulse generator. Do not use this pulse generator with another pulse generator. This combination could cause pulse generator interaction, resulting in patient injury or a lack of therapy delivery. • Safety Core operation. In response to applicable nonrecoverable or repeat fault conditions, the pulse generator will switch irreversibly to Safety Core operation. Safety Core pacing is unipolar, which is contraindicated for patients with an ICD. Handling • Do not kink leads. Do not kink, twist, or braid the lead with other leads as doing so could cause lead insulation abrasion damage or conductor damage. Programming and Device Operations • Atrial tracking modes. Do not use atrial tracking modes in patients with chronic refractory atrial tachyarrhythmias. Tracking of atrial arrhythmias could result in ventricular tachyarrhythmias. • Lead safety switch programmed to On. In devices with the lead safety switch programmed to On, the lead polarity will switch to unipolar in the presence of a lead impedance of ≤ 200 or ≥ 2000 Ω. Unipolar pacing is contraindicated for patients with an ICD. • Sensitivity settings and EMI. If programmed to a fixed atrial sensitivity value of 0.15 mV, the pulse generator may be more susceptible to electromagnetic interference. This increased susceptibility should be taken into consideration when determining the follow-up schedule for patients requiring such a setting. Post-Implant • Protected environments. Advise patients to seek medical guidance before entering environments that could adversely affect the operation of the active implantable medical device, including areas protected by a warning notice that prevents entry by patients who have a pulse generator. • Magnetic Resonance Imaging (MRI) exposure. Do not expose a patient to MRI scanning. Strong magnetic fields may damage the pulse generator and/or lead system, possibly resulting in injury to or death of the patient. • Diathermy. Do not subject a patient with an implanted pulse generator and/or lead to diathermy since diathermy may cause fibrillation, burning of the myocardium, and irreversible damage to the pulse generator because of induced currents. PRECAUTIONS Clinical Considerations • STAT PACE. STAT PACE will initiate unipolar pacing. • Pacemaker-mediated tachycardia (PMT). Programming minimum PVARP less than retrograde V–A conduction may increase the likelihood of a PMT. • Automatic Capture. Automatic Capture is intended for ventricular use only. Do not program Amplitude to Auto for single-chamber devices implanted in the atrium. • MV sensor modes. The safety and efficacy of the MV sensor modes have not been clinically established in patients with abdominal implant sites. • MV sensor mode performance. MV sensor performance may be adversely affected under transient conditions such as pneumothorax, pericardial effusion, or pleural effusion. Consider programming the MV sensor Off until these conditions are resolved. • Adaptive-rate modes. Adaptive-rate modes based completely or in part on MV might be inappropriate for patients who can achieve respiratory cycles shorter than one second (greater than 60 breaths per minute). Higher respiration rates attenuate the impedance signal, which diminishes the MV rate response (i.e., the pacing rate will drop toward the programmed LRL). Adaptive-rate modes based completely or in part on MV should not be used for patients with: 10 • An ICD • Unipolar leads—for MV detection, a bipolar lead is required in either the atrium or ventricle • Epicardial ventricular leads—MV measurement has only been tested with a bipolar transvenous lead • A mechanical ventilator—use of the ventilator might result in an inappropriate MV sensor-driven rate Sterilization and Storage • If package is damaged. The blister trays and contents are sterilized with ethylene oxide gas before final packaging. When the pulse generator and/or lead is received, it is sterile provided the container is intact. If the packaging is wet, punctured, opened, or otherwise damaged, return the pulse generator and/or lead to Boston Scientific. • Storage temperature and equilibration. Recommended storage temperatures are 0°C–50°C (32°F–122°F). Allow the device to reach a proper temperature before using telemetry communication capabilities, programming or implanting the device because temperature extremes may affect initial device function. • Device storage. Store the pulse generator in a clean area away from magnets, kits containing magnets, and sources of EMI to avoid device damage. • Use by date. Implant the pulse generator and/or lead before or on the USE BY date on the package label because this date reflects a validated shelf life. For example, if the date is January 1, do not implant on or after January 2. Implantation • Expected benefits. Determine whether the expected device benefits provided by programmable options outweigh the possibility of more rapid battery depletion. 11 • Evaluate patient for surgery. There may be additional factors regarding the patient’s overall health and medical condition that, while not related to device function or purpose, could render the patient a poor candidate for implantation of this system. Cardiac health advocacy groups may have published guidelines that may be helpful in conducting this evaluation. • Lead compatibility. Prior to implantation, confirm the lead-to-pulse generator compatibility. Using incompatible leads and pulse generators can damage the connector and/or result in potential adverse consequences, such as undersensing of cardiac activity or failure to deliver necessary therapy. • Telemetry wand. Make sure a sterile telemetry wand is available should loss of ZIP telemetry occur. Verify that the wand can easily be connected to the programmer and is within reach of the pulse generator. • Line-powered equipment. Exercise extreme caution if testing leads using line-powered equipment because leakage current exceeding 10 µA can induce ventricular fibrillation. Ensure that any line-powered equipment is within specifications. 12 • Replacement device. Implanting a replacement device in a subcutaneous pocket that previously housed a larger device may result in pocket air entrapment, migration, erosion, or insufficient grounding between the device and tissue. Irrigating the pocket with sterile saline solution decreases the possibility of pocket air entrapment and insufficient grounding. Suturing the device in place reduces the possibility of migration and erosion. • Do not bend the lead near the lead-header interface. Insert the lead terminal straight into the lead port. Do not bend the lead near the lead-header interface. Improper insertion can cause insulation or connector damage. • Absence of a lead. The absence of a lead or plug in a lead port may affect device performance. If a lead is not used, be sure to properly insert a plug in the unused port, and then tighten the setscrew onto the plug. • Dual chamber device without a functional RV lead. If a dual-chamber device is programmed to AAI(R), ensure that a functional RV lead is present. In the absence of a functional RV lead, programming to AAI(R) may result in undersensing or oversensing. 13 • Electrode connections. Do not insert a lead into the pulse generator connector without taking the following precautions to ensure proper lead insertion: • Insert the torque wrench into the preslit depression of the seal plug before inserting the lead into the port, to release any trapped fluid or air. • Visually verify that the setscrew is sufficiently retracted to allow insertion. Use the torque wrench to loosen the setscrew if necessary. • Fully insert each lead into its lead port and then tighten the setscrew onto the terminal pin. • Shunting energy. Do not allow any object that is electrically conductive to come into contact with the lead or device during induction because it may shunt energy, resulting in less energy getting to the patient, and may damage the implanted system. • Do not suture directly over lead. Do not suture directly over the lead body, as this may cause structural damage. Use the suture sleeve to secure the lead proximal to the venous entry site to prevent lead movement. • MV Sensor. Do not program the MV sensor to On until after the pulse generator has been implanted and system integrity has been tested and verified. 14 Device Programming • Device communication. Use only the designated PRM and software application to communicate with this pulse generator. • STAT PACE settings. When a pulse generator is programmed to STAT PACE settings, it will continue to pace at the high-energy STAT PACE values if it is not reprogrammed. The use of STAT PACE parameters will likely decrease device longevity. • Pacing and sensing margins. Consider lead maturation in your choice of pacing amplitude, pacing pulse width, and sensitivity settings. • • An acute pacing threshold greater than 1.5 V or a chronic pacing threshold greater than 3 V can result in loss of capture because thresholds may increase over time. • An R-wave amplitude less than 5 mV or a P-wave amplitude less than 2 mV can result in undersensing because the sensed amplitude may decrease after implantation. • Pacing lead impedance should be within the range of 200 Ω and 2000 Ω. Lead impedance values and Lead Safety Switch. If leads with measured impedance values approaching 200 or 2000 Ω are used, consider programming Lead Safety Switch Off. 15 • Proper programming of the lead configuration. If the Lead Configuration is programmed to Bipolar when a unipolar lead is implanted, pacing will not occur. • Programming for supraventricular tachyarrhythmias (SVTs). Determine if the device and programmable options are appropriate for patients with SVTs because SVTs can initiate unwanted device therapy. • Adaptive-rate pacing. Rate adaptive pacing should be used with care in patients who are unable to tolerate increased pacing rates. • Ventricular refractory periods (VRPs) in adaptive-rate pacing. Adaptive-rate pacing is not limited by refractory periods. A long refractory period programmed in combination with a high MSR can result in asynchronous pacing during refractory periods since the combination can cause a very small sensing window or none at all. Use dynamic AV Delay or dynamic PVARP to optimize sensing windows. If you are entering a fixed AV delay, consider the sensing outcomes. • Atrial oversensing. Take care to ensure that artifacts from the ventricles are not present on the atrial channel, or atrial oversensing may result. If ventricular artifacts are present in the atrial channel, the atrial lead may need to be repositioned to minimize its interaction. 16 • ATR entry count. Exercise care when programming the Entry Count to low values in conjunction with a short ATR Duration. This combination allows mode switching with very few fast atrial beats. For example, if the Entry Count was programmed to 2 and the ATR Duration to 0, ATR mode switching could occur on 2 fast atrial intervals. In these instances, a short series of premature atrial events could cause the device to mode switch. • ATR exit count. Exercise care when programming the Exit Count to low values. For example, if the Exit Count was programmed to 2, a few cycles of atrial undersensing could cause termination of mode switching. • Proper programming without an atrial lead. If an atrial lead is not implanted (port is plugged instead), or an atrial lead is abandoned but remains connected to the header, device programming should be consistent with the number and type of leads actually in use. • MV Recalibration. To obtain an accurate MV baseline, the MV sensor will be calibrated automatically or can be calibrated manually. A new, manual calibration should be performed if the pulse generator is removed from the pocket following implant, such as during a lead repositioning procedure, or in cases where the MV baseline may have been affected by factors such as lead maturation, air entrapment in the pocket, pulse generator motion due to inadequate suturing, external defibrillation or cardioversion, or other patient complications (e.g., pneumothorax). 17 • Sensing adjustment. Following any Sensitivity parameter adjustment or any modification of the sensing lead, always verify appropriate sensing. Programming Sensitivity to the highest value (lowest sensitivity) may result in undersensing of cardiac activity. Likewise, programming to the lowest value (highest sensitivity) may result in oversensing of non-cardiac signals. • Sensitivity in unipolar lead configuraton. The amplitude and prevalence of myopotential noise is increased in unipolar lead configurations, as compared to bipolar lead configurations. For patients with a unipolar lead configuration and myopotential oversensing during activity involving the pectoral muscles, the programming of Fixed Sensitivity is recommended. Environmental and Medical Therapy Hazards • Avoid electromagnetic interference (EMI). Advise patients to avoid sources of EMI. The pulse generator may inhibit pacing due to oversensing, or may switch to asynchronous pacing at the programmed pacing rate or at the magnet rate in the presence of EMI. Moving away from the source of the EMI or turning off the source usually allows the pulse generator to return to normal operation. 18 Examples of potential EMI sources are: • Electrical power sources, arc welding or resistance welding equipment, and robotic jacks • High voltage power distribution lines • Electrical smelting furnaces • Large RF transmitters such as radar • Radio transmitters, including those used to control toys • Electronic surveillance (antitheft) devices • An alternator on a car that is running • Medical treatments and diagnostic tests in which an electrical current is passed through the body, such as TENS, electrocautery, electrolysis/thermolysis, electrodiagnostic testing, electromyography, or nerve conduction studies • Any externally applied device that uses an automatic lead detection alarm system (e.g., an EKG machine) 19 Hospital and Medical Environments • Mechanical ventilators. Program the MV Sensor to Off during mechanical ventilation. Otherwise, the following may occur: • • Inappropriate MV sensor-driven rate • Misleading respiration-based trending Conducted electrical current. Any medical equipment, treatment, therapy, or diagnostic test that introduces electrical current into the patient has the potential to interfere with pulse generator function. • 20 External patient monitors (e.g., respiratory monitors, surface ECG monitors, hemodynamic monitors) may interfere with the pulse generator’s impedance-based diagnostics (e.g., Respiratory Rate trend). This interference may also result in accelerated pacing, possibly up to the maximum sensor-driven rate, when MV is programmed to On. To resolve suspected interactions, deactivate the MV sensor either by programming it to Off (no MV rate driving or MV sensor-based trending will occur), or Passive (no MV rate driving will occur). Alternatively, program the Brady Mode to a non-rate responsive mode (no MV rate driving will occur). If a PRM is not available and the pulse generator is pacing at the sensor-driven rate, apply a magnet to the pulse generator to initiate temporary asynchronous, non-rate responsive pacing. • • Medical therapies, treatments, and diagnostic tests that use conducted electrical current (e.g., TENS, electrocautery, electrolysis/thermolysis, electrodiagnostic testing, electromyography, or nerve conduction studies) may interfere with or damage the pulse generator. Program the device to Electrocautery Mode prior to the treatment, and monitor device performance during the treatment. After the treatment, verify pulse generator function ("Post-Therapy Pulse Generator Follow Up" on page 27). Internal defibrillation. Do not use internal defibrillation paddles or catheters unless the pulse generator is disconnected from the leads because the leads may shunt energy. This could result in injury to the patient and damage to the implanted system. 21 • External defibrillation. It can take up to 15 seconds for sensing to recover after an external shock is delivered. In non-emergency situations, for pacemaker dependent patients, consider programming the pulse generator to an asynchronous pacing mode prior to performing external cardioversion or defibrillation. External defibrillation or cardioversion can damage the pulse generator. To help prevent damage to the pulse generator, consider the following: • Avoid placing a pad (or paddle) directly over the pulse generator. Position the pads (or paddles) as far from the pulse generator as possible. • Position the pads (or paddles) in a posterior-anterior orientation when the device is implanted in the right pectoral region or an anterior-apex orientation when the device is implanted in the left pectoral region. • Set energy output of external defibrillation equipment as low as clinically acceptable. • In non-emergency situations, prior to performing external cardioversion or defibrillation, program the MV sensor to Off. Following external cardioversion or defibrillation, verify pulse generator function ("Post-Therapy Pulse Generator Follow Up" on page 27). 22 • Lithotripsy. Extracorporeal shock wave lithotripsy (ESWL) may cause electromagnetic interference with or damage to the pulse generator. If ESWL is medically necessary, consider the following to minimize the potential for encountering interaction: • Focus the ESWL beam at least 15 cm (6 in) away from the pulse generator. • Depending on the pacing needs of the patient, program the Brady Mode to a non-rate-responsive VVI or VOO mode. • Ultrasound energy. Therapeutic ultrasound (e.g., lithotripsy) energy may damage the pulse generator. If therapeutic ultrasound energy must be used, avoid focusing near the pulse generator site. Diagnostic ultrasound (e.g., echocardiography) is not known to be harmful to the pulse generator. • Electrical interference. Electrical interference or “noise” from devices such as electrocautery and monitoring equipment may interfere with establishing or maintaining telemetry for interrogating or programming the device. In the presence of such interference, move the programmer away from electrical devices, and ensure that the wand cord and cables are not crossing one another. If telemetry is cancelled as a result of interference, the device should be re-interrogated prior to evaluating information from pulse generator memory. 23 • • Radio frequency (RF) interference. RF signals from devices that operate at frequencies near that of the pulse generator may interrupt ZIP telemetry while interrogating or programming the pulse generator. This RF interference can be reduced by increasing the distance between the interfering device and the PRM and pulse generator. Examples of devices that may cause interference include: • Cordless phone handsets or base stations • Certain patient monitoring systems Central line guidewire insertion. Use caution when inserting guidewires for placement of other types of central venous catheter systems such as PIC lines or Hickman catheters in locations where pulse generator leads may be encountered. Insertion of such guidewires into veins containing leads could result in the leads being damaged or dislodged. Home and Occupational Environments • 24 Home appliances. Home appliances that are in good working order and properly grounded do not usually produce enough EMI to interfere with pulse generator operation. There have been reports of pulse generator disturbances caused by electric hand tools or electric razors used directly over the pulse generator implant site. • Magnetic fields. Advise patients that extended exposure to strong (greater than 10 gauss or 1 mTesla) magnetic fields may trigger the magnet feature. Examples of magnetic sources include: • Industrial transformers and motors • MRI scanners • Large stereo speakers • Telephone receivers if held within 1.27 cm (0.5 inches) of the pulse generator • Magnetic wands such as those used for airport security and in the Bingo game • Electronic Article Surveillance (EAS). Advise patients to avoid lingering near antitheft devices such as those found in the entrances and exits of department stores and public libraries. Patients should walk through them at a normal pace because such devices may cause inappropriate pulse generator operation. • Cellular phones. Advise patients to hold cellular phones to the ear opposite the side of the implanted device. Patients should not carry a cellular phone that is turned on in a breast pocket or on a belt within 15 cm (6 inches) of the implanted device since some cellular phones may cause the pulse generator to deliver inappropriate therapy or inhibit appropriate therapy. 25 Follow-up Testing • Pacing threshold testing. If the patient’s condition or drug regimen has changed or device parameters have been reprogrammed, consider performing a pacing threshold test to confirm adequate margins for pace capture. • Follow-up considerations for patients leaving the country. Pulse generator follow-up considerations should be made in advance for patients who plan to travel or relocate post-implant to a country other than the country in which their device was implanted. Regulatory approval status for devices and associated programmer software configurations varies by country; certain countries may not have approval or capability to follow specific products. Contact Boston Scientific, using the information on the back cover, for help in determining feasibility of device follow-up in the patient’s destination country. Explant and Disposal • 26 Incineration. Be sure that the pulse generator is removed before cremation. Cremation and incineration temperatures might cause the pulse generator to explode. • Device handling. Before explanting, cleaning, or shipping the device, complete the following actions to prevent overwriting of important therapy history data: • Program the pulse generator Brady Mode to Off • Program Ventricular Tachy EGM Storage to Off Clean and disinfect the device using standard biohazard handling techniques. • Explanted devices. Return all explanted pulse generators and leads to Boston Scientific. Examination of explanted pulse generators and leads can provide information for continued improvement in system reliability and warranty considerations. Do not implant an explanted pulse generator and/or lead in another patient as sterility, functionality, and reliability cannot be ensured. SUPPLEMENTAL PRECAUTIONARY INFORMATION Post-Therapy Pulse Generator Follow Up Following any surgery or medical procedure with the potential to affect pulse generator function, you should perform a thorough follow-up, which may include the following: • Interrogating the pulse generator with a programmer • Reviewing clinical events and fault codes 27 • Reviewing the Arrhythmia Logbook, including stored electrograms (EGMs) • Reviewing real-time EGMs • Testing the leads (threshold, amplitude, and impedance) • Reviewing MV sensor-based diagnostics, MV sensor performance, and performing a manual MV sensor calibration if desired • Verifying battery status • Programming any permanent brady parameter to a new value and then reprogramming it back to the desired value • Saving all patient data • Verifying the appropriate final programming prior to allowing the patient to leave the clinic 28 Transcutaneous Electrical Nerve Stimulation (TENS) CAUTION: TENS involves passing electrical current through the body, and may interfere with pulse generator function. If TENS is medically necessary, evaluate the TENS therapy settings for compatibility with the pulse generator. The following guidelines may reduce the likelihood of interaction: • Place the TENS electrodes as close together and as far away from the pulse generator and leads as possible. • Use the lowest clinically-appropriate TENS energy output. • Consider cardiac monitoring during TENS use, especially for pacemaker-dependent patients. Additional steps can be taken to help reduce interference during in-clinic use of TENS: • If interference is suspected during in-clinic use, turn off the TENS unit. • If pacing inhibition is observed, use a magnet to pace asynchronously. • Do not change TENS settings until you have verified that the new settings do not interfere with pulse generator function. 29 If TENS is medically necessary outside the clinical setting (at-home use), provide patients with the following instructions: • Do not change the TENS settings or electrode positions unless instructed to do so. • End each TENS session by turning off the unit before removing the electrodes. • If the patient experiences symptoms of lightheadedness, dizziness, or loss of consciousness during TENS use, they should turn off the TENS unit and contact their physician. Follow these steps to use the PRM to evaluate pulse generator function during TENS use: 1. Observe real-time EGMs at prescribed TENS output settings, noting when appropriate sensing or interference occurs. NOTE: Patient triggered monitoring may be used as an additional method to confirm device function during TENS use. 2. When finished, turn off the TENS unit. You should also perform a thorough follow-up evaluation of the pulse generator following TENS, to ensure that device function has not been compromised ("Post-Therapy Pulse Generator Follow Up" on page 27). 30 For additional information, contact Boston Scientific using the information on the back cover. Electrocautery and Radio Frequency (RF) Ablation CAUTION: Electrocautery and RF ablation may induce ventricular arrhythmias and/or fibrillation, and may cause asynchronous pacing, inhibition of pacing and/or a reduction in pulse generator pacing output possibly leading to loss of capture. RF ablation may also cause ventricular pacing up to the MTR and/or changes in pacing thresholds. Additionally, exercise caution when performing any other type of cardiac ablation procedure in patients with implanted devices. If electrocautery or RF ablation is medically necessary, observe the following to minimize risk to the patient and device: • Depending on the pacing needs of the patient, enable the Electrocautery Protection Mode, program to an asynchronous pacing mode, or use a magnet to switch to asynchronous pacing. An option for patients with intrinsic rhythm is to program the Brady Mode to VVI at a rate below the intrinsic rate to avoid competitive pacing. • Have temporary pacing and external defibrillation equipment available. 31 • Avoid direct contact between the electrocautery equipment or ablation catheters and the pulse generator and leads. RF ablation close to the lead electrode may damage the lead-tissue interface. • Keep the path of the electrical current as far away as possible from the pulse generator and leads. • If RF ablation and/or electrocautery is performed on tissue near the device or leads, monitor pre- and post-measurements for sensing and pacing thresholds and impedances to determine the integrity and stability of the system. • For electrocautery, use a bipolar electrocautery system where possible and use short, intermittent, and irregular bursts at the lowest feasible energy levels. • RF ablation equipment may cause telemetry interference between the pulse generator and PRM. If device programming changes are necessary during an RF ablation procedure, turn off the RF ablation equipment before interrogation. When the procedure is finished, cancel the Electrocautery Protection Mode in order to reactivate the previously programmed therapy modes. 32 Ionizing Radiation CAUTION: It is not possible to specify a safe radiation dosage or guarantee proper pulse generator function following exposure to ionizing radiation. Multiple factors collectively determine the impact of radiation therapy on an implanted pulse generator, including proximity of the pulse generator to the radiation beam, type and energy level of the radiation beam, dose rate, total dose delivered over the life of the pulse generator, and shielding of the pulse generator. The impact of ionizing radiation will also vary from one pulse generator to another and may range from no changes in function to a loss of pacing. Sources of ionizing radiation vary significantly in their potential impact on an implanted pulse generator. Several therapeutic radiation sources are capable of interfering with or damaging an implanted pulse generator, including those used for the treatment of cancer, such as radioactive cobalt, linear accelerators, radioactive seeds, and betatrons. Prior to a course of therapeutic radiation treatment, the patient’s radiation oncologist and cardiologist or electrophysiologist should consider all patient management options, including increased follow-up and device replacement. Other considerations include: • Maximizing shielding of the pulse generator within the treatment field • Determining the appropriate level of patient monitoring during treatment 33 Evaluate pulse generator operation during and following the course of radiation treatment to exercise as much device functionality as possible ("Post-Therapy Pulse Generator Follow Up" on page 27). The extent, timing, and frequency of this evaluation relative to the radiation therapy regimen are dependent upon current patient health, and therefore should be determined by the attending cardiologist or electrophysiologist. Many pulse generator diagnostics are performed automatically once per hour, so pulse generator evaluation should not be concluded until pulse generator diagnostics have been updated and reviewed (at least one hour after radiation exposure). The effects of radiation exposure on the implanted pulse generator may remain undetected until some time following exposure. For this reason, continue to monitor pulse generator function closely and use caution when programming a feature in the weeks or months following radiation therapy. Elevated Pressures The International Standards Organization (ISO) has not approved a standardized pressure test for implantable pulse generators that experience hyperbaric oxygen therapy (HBOT) or SCUBA diving. However, Boston Scientific developed a test protocol to evaluate device performance upon exposure to elevated atmospheric pressures. The following summary of pressure testing should not be viewed as and is not an endorsement of HBOT or SCUBA diving. 34 CAUTION: Elevated pressures due to HBOT or SCUBA diving may damage the pulse generator. During laboratory testing, all pulse generators in the test sample functioned as designed when exposed to more than 1000 cycles at a pressure up to 5.0 ATA. Laboratory testing did not characterize the impact of elevated pressure on pulse generator performance or physiological response while implanted in a human body. Pressure for each test cycle began at ambient/room pressure, increased to a high pressure level, and then returned to ambient pressure. Although dwell time (the amount of time under elevated pressure) may have an impact on human physiology, testing indicated it did not impact pulse generator performance. Pressure value equivalencies are provided below (Table 1 on page 35). Table 1. Pressure Value Equivalencies Pressure value equivalencies Atmospheres Absolute 5.0 ATA Sea water deptha 40 m (130 ft) Pressure, absolute 72.8 psia Pressure, gaugeb 58.8 psig 35 Table 1. Pressure Value Equivalencies (continued) Pressure value equivalencies Bar 5.0 kPa Absolute 500 a. All pressures were derived assuming sea water density of 1030 kg/m3. b. Pressure as read on a gauge or dial (psia = psig + 14.7 psi). Prior to SCUBA diving or starting an HBOT program, the patient’s attending cardiologist or electrophysiologist should be consulted to fully understand the potential consequences relative to the patient’s specific health condition. A Dive Medicine Specialist may also be consulted prior to SCUBA diving. More frequent device follow-up may be warranted in conjunction with HBOT or SCUBA diving. Evaluate pulse generator operation following high pressure exposure ("Post-Therapy Pulse Generator Follow Up" on page 27). The extent, timing, and frequency of this evaluation relative to the high pressure exposure are dependent upon current patient health, and should be determined by the attending cardiologist or electrophysiologist. If you have additional questions, or would like more detail regarding the test protocol or test results specific to HBOT or SCUBA diving, contact Boston Scientific using the information on the back cover. 36 POTENTIAL ADVERSE EVENTS Based on the literature and on pulse generator implant experience, the following alphabetical list includes the possible adverse events associated with implantation of a pulse generator and/or lead system: • Air embolism • Allergic reaction • Bleeding • Cardiac tamponade • Chronic nerve damage • Component failure • Conductor coil fracture • Death • Elevated thresholds • Erosion • Excessive fibrotic tissue growth • Extracardiac stimulation (muscle/nerve stimulation) • Foreign body rejection phenomena • Fluid accumulation • Formation of hematomas or seromas 37 • Heart block • Inability to pace • Inappropriate pacing • Incisional pain • Incomplete lead connection with pulse generator • Infection including endocarditis • Lead dislodgment • Lead fracture • Lead insulation breakage or abrasion • Lead perforation • Lead tip deformation and/or breakage • Local tissue reaction • Myocardial infarction (MI) • Myocardial necrosis • Myocardial trauma (e.g., tissue damage, valve damage) • Myopotential sensing • Oversensing/undersensing • Pacemaker-mediated tachycardia (PMT) (Applies to dual-chamber devices only.) • Pericardial rub, effusion 38 • Pneumothorax • Pulse generator migration • Shunting current during defibrillation with internal or external paddles • Tachyarrhythmias, which include acceleration of arrhythmias and early, recurrent atrial fibrillation • Thrombosis/thromboemboli • Valve damage • Venous occlusion • Venous trauma (e.g., perforation, dissection, erosion) • Worsening heart failure Patients may develop psychological intolerance to a pulse generator system and may experience the following: • Dependency • Depression • Fear of premature battery depletion • Fear of device malfunction MECHANICAL SPECIFICATIONS Mechanical and material specifications specific to each model are listed below. 39 Table 2. Mechanical and Material Specifications K172 K173 K174 SR DR DR RA/RV: IS-1 RA:IS-1 RV: IS-1 RA:IS-1 RV: IS-1 Dimensions W x H x D (cm) 4.45 x 4.57 x 0.75 4.45 x 4.70 x 0.75 4.45 x 5.56 x 0.75 Volume (cm3) 11.5 12.0 14.0 Mass (g) 23.5 24.5 32.0 Case Electrode Surface Area (cm2) 29.78 29.78 35.98 hermetically sealed titanium hermetically sealed titanium hermetically sealed titanium implantation-grade polymer implantation-grade polymer implantation-grade polymer Type Connector Case Material Header Material 40 Table 2. Mechanical and Material Specifications (continued) K172 K173 K174 Power Supply Lithium-carbon monofluoridesilver vanadium oxide cell Greatbatch 2808 Lithium-carbon monofluoridesilver vanadium oxide cell Greatbatch 2808 Lithiummanganese dioxide cell Boston Scientific 402125 Usable Battery Capacity (Ah) 1.05 1.05 1.47 Residual Usable Battery Capacity at Explant (Ah) 0.06 0.08 0.08 Models include ZIP telemetry operating with a transmit frequency of 916.5 MHz. ITEMS INCLUDED IN PACKAGE The following items are included with the pulse generator: • One torque wrench • Product literature 41 NOTE: Accessories (e.g., wrenches) are intended for one-time use only. They should not be resterilized or reused. SYMBOLS ON PACKAGING The following symbols may be used on packaging and labeling (Table 3 on page 42): Table 3. Symbol Symbols on packaging Description Reference number Package contents Pulse generator Torque wrench 42 Table 3. Symbol Symbols on packaging (continued) Description Literature enclosed Serial number Use by Lot number Date of manufacture Non-ionizing electromagnetic radiation Sterilized using ethylene oxide 43 Table 3. Symbol Symbols on packaging (continued) Description Do not resterilize Do not reuse Do not use if package is damaged Consult instructions for use Temperature limitation 44 Table 3. Symbol Symbols on packaging (continued) Description Wand placement indicator for interrogation Opening instruction Authorized Representative in the European Community Manufacturer CHARACTERISTICS AS SHIPPED Refer to the table for pulse generator settings at shipment (Table 4 on page 46). 45 Table 4. Characteristics as shipped Parameter Setting Pacing Mode Storage Pacing Therapy available DDDR (DR models) SSIR (SR models) Sensor Blend (Accel and MV) Pace/Sense Configuration RA: BI/BI (DR models) Pace/Sense Configuration RV: BI/BI Magnet Rate 100 ppm The pulse generator is shipped in a power-saving Storage mode to extend its shelf life. In Storage mode, all features are inactive except: • Telemetry support, which allows interrogation and programming • Real-time clock • STAT PACE command The device leaves Storage mode when one of the following actions occurs; however, programming other parameters will not affect the Storage mode: 46 • STAT PACE is commanded • The pulse generator automatically detects lead insertion (refer to "Implanting the Pulse Generator" on page 59) • Device Mode is programmed to Exit Storage Once you have programmed the pulse generator out of Storage mode, the device cannot be reprogrammed to that mode. X-RAY IDENTIFIER The pulse generator has an identifier that is visible on x-ray film or under fluoroscopy. This identifier provides noninvasive confirmation of the manufacturer and consists of the following: • The letters, BSC, to identify Boston Scientific as the manufacturer • The number, 011, to identify the Model 2869 PRM software application needed to communicate with the pulse generator The x-ray identifier is embedded in the header of the device at the approximate location shown (Figure 1 on page 48). 47 X-Ray Identifier Header Pulse Generator Case Figure 1. X-ray identifier For information on identifying the device via the PRM, refer to the PRM Operator’s Manual. The pulse generator model number is stored in device memory and is shown on the PRM summary screen once the pulse generator is interrogated. FEDERAL COMMUNICATIONS COMMISSION (FCC) This device complies with Title 47, Part 15 of the FCC rules. Operation is subject to the following two conditions: • This device may not cause harmful interference, and • This device must accept any interference received, including interference that may cause undesired operation. 48 CAUTION: Changes or modifications not expressly approved by Boston Scientific could void the user’s authority to operate the equipment. PULSE GENERATOR LONGEVITY Based on simulated studies, it is anticipated that these pulse generators have average longevity to explant as shown below. The longevity expectations, which account for the energy used during manufacture and storage, apply at the conditions shown in the table along with the following: • Assumes 60 ppm LRL, ventricular and atrial settings of 0.4 ms pacing pulse width; sensors On. • These calculations also assume EGM Onset is set to On. Table 5. INGENIO Longevity Projections (Years) Modelsa Amplitude and Pacing K172 SR K173 DR K174 DR EL A and V Amplitudes 3.5 V 100% Paced 49 Table 5. INGENIO Longevity Projections (Years) (continued) Modelsa Amplitude and Pacing 500 Ω 7.3 5.5 8.0 750 Ω 7.9 6.3 9.0 1000 Ω 8.4 6.8 9.6 500 Ω 8.5 7.0 9.9 750 Ω 9.0 7.5 10.7 1000 Ω 9.2 7.8 11.2 50% Paced A and V Amplitudes 2.5 V 100% Paced 500 Ω 8.5 6.9 9.8 750 Ω 8.9 7.5 10.7 1000 Ω 9.2 7.9 11.2 50 Table 5. INGENIO Longevity Projections (Years) (continued) Modelsa Amplitude and Pacing 50% Paced 500 Ω 9.3 7.9 11.3 750 Ω 9.5 8.4 11.8 1000 Ω 9.6 8.6 12.1 a. Assumes ZIP telemetry use for 1 hour at implant and for 20 minutes during each quarterly follow-up. Table 6. INGENIO Longevity Projections (Years) with Right Ventricular Automatic Capture Modelsa Amplitude and Pacing K173 DR K172 SR Automatic Capture On (A = 3.5 Vb , K174 DR EL V = 1.0 V [assuming a threshold of 0.5]) 100% Paced 51 Table 6. INGENIO Longevity Projections (Years) with Right Ventricular Automatic Capture (continued) Modelsa Amplitude and Pacing 500 Ω 9.7 6.8 9.6 750 Ω 9.8 7.3 10.5 1000 Ω 9.8 7.8 11.0 500 Ω 9.9 7.9 11.2 750 Ω 9.9 8.3 11.6 1000 Ω 9.9 8.5 11.9 50% Paced Automatic Capture On (A = 2.5 Vb , V = 1.0 V [assuming a threshold of 0.5]) 100% Paced 500 Ω 9.7 7.8 11.0 750 Ω 9.8 8.2 11.5 52 Table 6. INGENIO Longevity Projections (Years) with Right Ventricular Automatic Capture (continued) Modelsa Amplitude and Pacing 9.8 8.4 11.8 500 Ω 9.9 8.5 12.0 750 Ω 9.9 8.8 12.3 1000 Ω 9.9 8.9 12.4 1000 Ω 50% Paced a. Assumes ZIP telemetry use for 1 hour at implant and for 20 minutes during each quarterly follow-up. b. This value is not applicable in single chamber devices programmed to VVI(R). Longevities at “worst case” settings of 5.0 V, 500 Ω, 1.0 ms are: 3.2 years for K172; 1.7 years for K173; 2.7 years for K174 at 70 ppm, and 2.4 years for K172; 1.1 years for K173; 1.9 years for K174 at 100 ppm. Longevities at an LRL of 70 ppm, 500 Ω, 0.5 ms, 100% paced, sensors On, and pacing mode most comprehensive are: K172 at 2.5 V = 7.9 years, at 5.0 V = 4.7 years; K173 at 2.5 V = 6.3 years, at 5.0 V = 2.9 years; K174 at 2.5 V = 8.9 years, at 5.0 V = 4.3 years. 53 NOTE: The energy consumption in the longevity table is based upon theoretical electrical principles and verified via bench testing only. The pulse generator longevity may increase with a decrease in any of the following: • Pacing rate • Pacing pulse amplitude(s) • Pacing pulse width(s) • Percentage of paced to sensed events Longevity is also reduced in the following circumstances: • With a decrease in pacing impedance • When Patient Triggered Monitor is programmed to On for 60 days, longevity is reduced by approximately 5 days • One hour of additional ZIP wandless telemetry reduces longevity by approximately 9 days. Device longevity may also be affected by: • Tolerances of electronic components • Variations in programmed parameters • Variations in usage as a result of patient condition 54 Refer to the PRM Summary and Battery Detail Summary screens for an estimate of pulse generator longevity specific to the implanted device. WARRANTY INFORMATION A limited warranty certificate for the pulse generator is packaged with the device. For additional copies, contact Boston Scientific using the information on the back cover. PRODUCT RELIABILITY It is Boston Scientific’s intent to provide implantable devices of high quality and reliability. However, these devices may exhibit malfunctions that may result in lost or compromised ability to deliver therapy. These malfunctions may include the following: • Premature battery depletion • Sensing or pacing issues • Error codes • Loss of telemetry Refer to Boston Scientific’s CRM Product Performance Report on www.bostonscientific.com for more information about device performance, including the types and rates of malfunctions that these devices have experienced historically. While historical data may not be predictive of future device performance, such data can provide important context for understanding the overall reliability of these types of products. 55 Sometimes device malfunctions result in the issuance of product advisories. Boston Scientific determines the need to issue product advisories based on the estimated malfunction rate and the clinical implication of the malfunction. When Boston Scientific communicates product advisory information, the decision whether to replace a device should take into account the risks of the malfunction, the risks of the replacement procedure, and the performance to date of the replacement device. PATIENT COUNSELING INFORMATION The following topics should be discussed with the patient prior to discharge. • External defibrillation—the patient should contact their physician to have their pulse generator system evaluated if they receive external defibrillation • Signs and symptoms of infection • Symptoms that should be reported (e.g., sustained high-rate pacing requiring reprogramming) • Protected environments—the patient should seek medical guidance before entering areas protected by a warning notice that prevents entry by patients who have a pulse generator • Avoiding potential sources of EMI in home, work, and medical environments • Reliability of their pulse generator ("Product Reliability" on page 55) 56 • Activity restrictions (if applicable) • Minimum heart rate (lower rate limit of the pulse generator) • Frequency of follow up • Travel or relocation—Follow-up arrangements should be made in advance if the patient is leaving the country of implant • Patient ID card—the patient should be advised to carry their patient ID card at all times (a temporary patient ID card is provided with the device, and a permanent ID card will be sent to the patient 4 to 6 weeks after the implant form is received by Boston Scientific) Patient Handbook The Patient Handbook is provided for each device. It is recommended that you discuss the information in the Patient Handbook with concerned individuals both before and after implantation so they are fully familiar with pulse generator operation. For additional copies, contact Boston Scientific using the information on the back cover. 57 LEAD CONNECTIONS Lead connections are illustrated below. CAUTION: Prior to implantation, confirm the lead-to-pulse generator compatibility. Using incompatible leads and pulse generators can damage the connector and/or result in potential adverse consequences, such as undersensing of cardiac activity or failure to deliver necessary therapy. CAUTION: If the Lead Configuration is programmed to Bipolar when a unipolar lead is implanted, pacing will not occur. IS-1 UNI/BI [1] RA/RV [2] Suture Hole Figure 2. 58 Lead connections and setscrew locations, RA/RV: IS1 IS-1 RA IS-1 RV UNI/BI UNI/BI [1] RA [2] RV [3] Suture Hole Figure 3. Lead connections and setscrew locations, RA: IS-1, RV: IS-1 NOTE: The pulse generator case is used as a pace electrode when the pulse generator has been programmed to a unipolar lead setting. IMPLANTING THE PULSE GENERATOR Implant the pulse generator by performing the following steps in the sequence provided. Some patients may require pacing therapies immediately upon connecting the leads to the pulse generator. If modifications to the nominal settings are needed, consider programming the pulse generator before or in parallel with implanting the lead system and forming the implantation pocket. 59 Step A: Check Equipment It is recommended that instrumentation for cardiac monitoring, defibrillation, and lead signal measurement should be available during the implant procedure. This includes the PRM system with its related accessories and the software application. Before beginning the implantation procedure, become completely familiar with the operation of all the equipment and the information in the respective operator’s and user’s manuals. Verify the operational status of all equipment that may be used during the procedure. In case of accidental damage or contamination, the following should be available: • Sterile duplicates of all implantable items • Sterile wand • Sterile PSA cables • Torque and non-torque wrenches • External defibrillation equipment Step B: Interrogate and Check the Pulse Generator To maintain sterility, test the pulse generator as described below before opening the sterile blister tray. The pulse generator should be at room temperature to ensure accurately measured parameters. 60 1. Interrogate the pulse generator using the PRM. Verify that the pulse generator’s device mode is programmed to Storage. If otherwise, contact Boston Scientific using the information on the back cover. 2. Review the pulse generator’s current battery status. Counters should be at zero. If the pulse generator battery status is not at full capacity, do not implant the pulse generator. Contact Boston Scientific using the information on the back cover. 3. If a unipolar pacing configuration is required at implant, program the lead configuration to unipolar before implant. Step C: Implant the Lead System The pulse generator requires a lead system for pacing and sensing. Selection of lead configuration and specific surgical procedures is a matter of professional judgment. The following leads are available for use with the pulse generator depending on the device model. • Unipolar or bipolar atrial lead • Unipolar or bipolar right ventricular lead. NOTE: Single-chamber devices can be used with either an atrial or a ventricular lead. NOTE: Using bipolar pacing leads will reduce the chance of myopotential sensing. 61 CAUTION: The absence of a lead or plug in a lead port may affect device performance. If a lead is not used, be sure to properly insert a plug in the unused port, and then tighten the setscrew onto the plug. CAUTION: If a dual-chamber device is programmed to AAI(R), ensure that a functional RV lead is present. In the absence of a functional RV lead, programming to AAI(R) may result in undersensing or oversensing. CAUTION: Do not suture directly over the lead body, as this may cause structural damage. Use the lead stabilizer to secure the lead proximal to the venous entry site to prevent lead movement. Implant the leads via the surgical approach chosen. When replacing a previously implanted pulse generator, it may be necessary to use an adapter to enable the new pulse generator to be connected to the existing leads. When using an adapter, follow the connection procedure described in the applicable adapter product data sheet. Always connect the adapter to the lead and repeat threshold and sensing measurements before connecting the adapter to the pulse generator. Step D: Take Baseline Measurements Once the leads are implanted, take baseline measurements. Evaluate the lead signals. If performing a pulse generator replacement procedure, existing leads should be reevaluated, (e.g., signal amplitudes, pacing thresholds, and impedance). The 62 use of radiography may help ensure lead position and integrity. If testing results are unsatisfactory, lead system repositioning or replacement may be required. • Connect the pace/sense lead(s) to a pacing system analyzer (PSA). • Pace/sense lead measurements, measured approximately 10 minutes after placement, are listed below. Note that the pulse generator measurements may not exactly correlate to the PSA measurements due to signal filtering. Table 7. Lead measurements Pace/ sense lead (acute) Pace/ sense lead (chronic) R-wave amplitudea > 5 mV > 5 mV P-wave amplitudea > 1.5 mV > 1.5 mV durationb c d < 100 ms < 100 ms Pacing threshold (right ventricle) < 1.5 V endocardial < 2.0 V epicardial < 3.0 V endocardial < 3.5 V epicardial Pacing threshold (atrium) < 1.5 V endocardial < 3.0 V endocardial R-wave 63 Table 7. Lead measurements (continued) Lead impedance (at 5.0 V and 0.5 ms atrium and right ventricle) Pace/ sense lead (acute) Pace/ sense lead (chronic) 200–2000 Ω 200–2000 Ω a. Amplitudes less than 2 mV cause inaccurate rate counting in the chronic state, and result in inability to sense a tachyarrhythmia or the misinterpretation of a normal rhythm as abnormal. b. Lower R-wave amplitudes and longer duration may be associated with placement in ischemic or scarred tissues. Since signal quality may deteriorate chronically, efforts should be made to meet the above criteria by repositioning the leads to obtain signals with the largest possible amplitude and shortest duration. c. Durations longer than 135 ms (the pulse generator’s refractory period) may result in inaccurate cardiac rate determination, inability to sense a tachyarrhythmia, or in the misinterpretation of a normal rhythm as abnormal. d. This measurement is not inclusive of current of injury. Step E: Form the Implantation Pocket Using standard operating procedures to prepare an implantation pocket, choose the position of the pocket based on the implanted lead configuration and the patient’s body habitus. Giving consideration to patient anatomy and pulse generator size and motion, gently coil any excess lead and place adjacent to the pulse generator. It is 64 important to place the lead into the pocket in a manner that minimizes lead tension, twisting, sharp angles, and/or pressure. Pulse generators are typically implanted subcutaneously in order to minimize tissue trauma and facilitate explant. However, deeper implantation (e.g., subpectoral) may help avoid erosion or extrusion in some patients. Verify magnet function and wanded telemetry to ensure the pulse generator is within acceptable range. If an abdominal implant is suitable, it is recommended that implantation occur on the left abdominal side. If it is necessary to tunnel the lead, consider the following: • If a Guidant tunneler is not used, cap the lead terminal pins. A Penrose drain, large chest tube, or tunneling tool may be used to tunnel the leads. • Gently tunnel the leads subcutaneously to the implantation pocket, if necessary. • Reevaluate all lead signals to determine if any of the leads have been damaged during the tunneling procedure. If the leads are not connected to a pulse generator at the time of lead implantation, they must be capped before closing the incision. 65 Step F: Connect the Leads to the Pulse Generator To connect leads to the pulse generator, use only the tools provided in the pulse generator sterile tray or accessory kit. Failure to use the supplied torque wrench may result in damage to the setscrews, seal plugs, or connector threads. Do not implant the pulse generator if the seal plugs appear to be damaged. Retain the tools until all testing procedures are complete and the pulse generator is implanted. Automatic Lead Detection Until a right ventricular lead is detected (or any appropriate lead in a single chamber device), the lead impedance is measured in both unipolar and bipolar configurations. Upon insertion of the lead into the header the impedance measurement circuit will detect an impedance which indicates that the device is implanted (automatic lead detection). If the impedance is in range (200 – 2000 Ω, inclusive) the pulse generator will automatically switch to the nominal parameters and start sensing and delivering therapy. The pulse generator can also be programmed out of the Storage mode prior to implant using the PRM. NOTE: If the lead being used for automatic lead detection is unipolar, an in-range impedance will not be obtained until the pulse generator is in stable contact with the subcutaneous tissue of the pocket. 66 NOTE: Arrhythmia Logbook and stored EGM data will not be stored for the first two hours after the lead is detected except for PaceSafe and patient triggered episodes. If the device is programmed out of Storage, asynchronous pacing spikes could be observed on intracardiac EGMs before bipolar RV lead insertion or before placing the pulse generator into the subcutaneous pocket if a unipolar RV lead is present. These subthreshold spikes will not occur once a bipolar RV lead is detected in the header or when contact between the pacemaker case and subcutaneous tissue completes the normal pacing circuit for a unipolar RV lead. If the device exits Storage as the result of automatic lead detection, the pulse generator may take up to 2 seconds plus one LRL interval before pacing begins as a result of lead detection. Leads should be connected to the pulse generator in the following sequence (for pulse generator header and setscrew location illustrations, refer to "Lead Connections" on page 58): NOTE: a. For single-chamber devices, use an RA or RV lead as appropriate. Right ventricle. Connect the RV lead first because it is required to establish RV-based timing cycles that yield appropriate sensing and pacing in all chambers, regardless of the programmed configuration. NOTE: Tightening the RV setscrew is not required for automatic lead detection to occur but should be done to ensure full electrical contact. 67 • b. In models with an IS-1 RV lead port, insert and secure the terminal pin of an IS-1 RV pace/sense lead. Right atrium. • In models with an IS-1 RA lead port, insert and secure the terminal pin of an IS-1 atrial pace/sense lead. Connect each lead to the pulse generator by following these steps (for additional information about the torque wrench, refer to "Bidirectional Torque Wrench" on page 74): a. Check for the presence of any blood or other body fluids in the lead ports on the pulse generator header. If fluid inadvertently enters the ports, clean them thoroughly with sterile water. b. If applicable, remove and discard the tip protection before using the torque wrench. c. Gently insert the torque wrench blade into the setscrew by passing it through the preslit, center depression of the seal plug at a 90° angle (Figure 4 on page 69). This will open up the seal plug, relieving any potential pressure build-up from the lead port by providing a pathway to release trapped fluid or air. NOTE: Failure to properly insert the torque wrench in the preslit depression of the seal plug may result in damage to the plug and its sealing properties. 68 CAUTION: Do not insert a lead into the pulse generator connector without taking the following precautions to ensure proper lead insertion: • Insert the torque wrench into the preslit depression of the seal plug before inserting the lead into the port, to release any trapped fluid or air. • Visually verify that the setscrew is sufficiently retracted to allow insertion. Use the torque wrench to loosen the setscrew if necessary. • Fully insert each lead into its lead port and then tighten the setscrew onto the terminal pin. Figure 4. Inserting the torque wrench 69 d. With the torque wrench in place, fully insert the lead terminal into the lead port. The lead terminal pin should be clearly visible beyond the connector block when viewed through the side of the pulse generator header. Place pressure on the lead to maintain its position and ensure that it remains fully inserted in the lead port. CAUTION: Insert the lead terminal straight into the lead port. Do not bend the lead near the lead-header interface. Improper insertion can cause insulation or connector damage. NOTE: If necessary, lubricate the lead connectors sparingly with sterile water to make insertion easier. NOTE: For IS-1 leads, be certain that the terminal pin visibly extends beyond the connector block at least 1 mm. e. Apply gentle downward pressure on the torque wrench until the blade is fully engaged within the setscrew cavity, taking care to avoid damage to the seal plug. Tighten the setscrew by slowly turning the torque wrench clockwise, until it ratchets once. The torque wrench is preset to apply the proper amount of force to the captive setscrew; additional rotation and force is unnecessary. f. Remove the torque wrench. g. Apply gentle traction to the lead to ensure a secure connection. 70 h. If the lead terminal is not secure, attempt to reseat the setscrew. Reinsert the torque wrench as described above, and loosen the setscrew by slowly turning the wrench counterclockwise, until the lead is loose. Then repeat the sequence above. i. If a lead port is not used, insert a plug into the unused port and tighten the setscrew. CAUTION: The absence of a lead or plug in a lead port may affect device performance. If a lead is not used, be sure to properly insert a plug in the unused port, and then tighten the setscrew onto the plug. Step G: 1. Evaluate Lead Signals Evaluate the pace/sense lead signals by viewing the real-time EGMs and markers. Lead measurements should reflect those above (Table 7 on page 63) Depending on the patient’s intrinsic rhythm, it may be necessary to temporarily adjust pacing parameters to allow assessment of pacing and sensing. If proper pacing and/or sensing are not demonstrated, disconnect the lead from the pulse generator and visually inspect the connector and leads. If necessary, retest the lead. CAUTION: Take care to ensure that artifacts from the ventricles are not present on the atrial channel, or atrial oversensing may result. If ventricular artifacts are present in the atrial channel, the atrial lead may need to be repositioned to minimize its interaction. 71 2. Evaluate all lead impedances. Step H: Program the Pulse Generator 1. Check the programmer clock and set and synchronize the pulse generator as necessary so that the proper time appears on printed reports and PRM strip chart recordings. 2. Program the pulse generator appropriately if a lead port(s) is not used. Step I: 1. Implant the Pulse Generator Ensure that the pulse generator has good contact with surrounding tissue of the implantation pocket, and then suture it in place to minimize device migration (for suture hole location illustrations, refer to "Lead Connections" on page 58). Gently coil excess lead and place adjacent to the pulse generator. Flush the pocket with saline solution, if necessary, to avoid a dry pocket. WARNING: Do not kink, twist, or braid the lead with other leads as doing so could cause lead insulation abrasion damage or conductor damage. 2. 72 Close the implantation pocket. Consideration should be given to place the leads in a manner to prevent contact with suture materials. It is recommended that absorbable sutures be used for closure of tissue layers. 3. If Electrocautery mode was used during the implant procedure, cancel it when done. 4. Confirm final programmed parameters. CAUTION: Following any Sensitivity parameter adjustment or any modification of the sensing lead, always verify appropriate sensing. Programming Sensitivity to the highest value (lowest sensitivity) may result in undersensing of cardiac activity. Likewise, programming to the lowest value (highest sensitivity) may result in oversensing of non-cardiac signals. 5. Use the PRM to print out parameter reports and save all patient data. Step J: Complete and Return the Implantation Form Within ten days of implantation, complete the Warranty Validation and Lead Registration form and return the original to Boston Scientific along with a copy of the patient data saved from the PRM. This information enables Boston Scientific to register each implanted pulse generator and set of leads, initiate the warranty period, and provide clinical data on the performance of the implanted system. Keep a copy of the Warranty Validation and Lead Registration form and programmer printouts, and the original patient data for the patient’s file. 73 Complete the temporary patient identification card and give it to the patient. After receiving the validation form, Boston Scientific sends the patient a permanent identification card. NOTE: A registration form is packaged with each pulse generator lead. If completing the pulse generator Warranty Validation and Lead Registration form for the pulse generator, completing separate validation forms for each lead is not necessary. BIDIRECTIONAL TORQUE WRENCH A torque wrench (model 6628) is included in the sterile tray with the pulse generator, and is designed for tightening and loosening #2-56 setscrews, captured setscrews, and setscrews on this and other Boston Scientific pulse generators and lead accessories that have setscrews that spin freely when fully retracted (these setscrews typically have white seal plugs). This torque wrench is bidirectional, and is preset to apply adequate torque to the setscrew and will ratchet when the setscrew is secure. The ratchet release mechanism prevents overtightening that could result in device damage. To facilitate the loosening of tight extended setscrews, this wrench applies more torque in the counterclockwise direction than in the clockwise direction. NOTE: As an additional safeguard, the tip of the torque wrench is designed to break off if used to overtighten beyond preset torque levels. If this occurs, the broken tip must be extracted from the setscrew using forceps. 74 This torque wrench may also be used for loosening setscrews on other Boston Scientific pulse generators and lead accessories that have setscrews that tighten against a stop when fully retracted (these setscrews typically have clear seal plugs). However, when retracting these setscrews, stop turning the torque wrench when the setscrew has come in contact with the stop. The additional counterclockwise torque of this wrench may cause these setscrews to become stuck if tightened against the stop. Loosening Stuck Setscrews Follow these steps to loosen stuck setscrews: 1. From a perpendicular position, tilt the torque wrench to the side 20º to 30º from the vertical center axis of the setscrew (Figure 5 on page 76). 2. Rotate the wrench clockwise (for retracted setscrew) or counterclockwise (for extended setscrew) around the axis three times, such that the handle of the wrench orbits the centerline of the screw (Figure 5 on page 76). The torque wrench handle should not turn or twist during this rotation. 3. As needed, you may attempt this up to four times with slightly more angle each time. If you cannot fully loosen the setscrew, use the #2 torque wrench from Wrench Kit Model 6501. 4. Once the setscrew has been freed, it may be extended or retracted as appropriate. 5. Discard the torque wrench upon completion of this procedure. 75 20°–30° [1] Clockwise rotation to free setscrews stuck in the retracted position [2] Counterclockwise rotation to free setscrews stuck in the extended position Figure 5. Rotating the torque wrench to loosen a stuck setscrew FOLLOW UP TESTING It is recommended that device functions be evaluated with periodic follow-up testing by trained personnel. Follow up guidance below will enable thorough review of device performance and associated patient health status throughout the life of the device. 76 Predischarge Follow Up The following procedures are typically performed during the predischarge follow up test using PRM telemetry: 1. Interrogate the pulse generator and review the Summary screen. 2. Verify pacing thresholds, lead impedance, and amplitude of intrinsic signals. 3. Review counters and histograms. 4. When all testing is complete, perform a final interrogation and save all the patient data. 5. Print the Quick Notes and Patient Data reports to retain in your files for future reference. 6. Clear the counters and histograms so that the most recent data will be displayed at the next follow up session. Counters and histograms can be cleared by pressing Reset on the Histogram screen, Tachy Counters screen, or Brady Counters screen. Routine Follow Up During early and middle life of the device, monitor performance by routine follow up one month after the predischarge check and at least annually thereafter. Office visits 77 may be supplemented by remote monitoring where available. As always, the physician should evaluate the patient’s current health status, device status and parameter values, and local medical guidelines to determine the most appropriate follow up schedule. When the device reaches One Year Remaining status and/or a magnet rate of 90 ppm is observed, follow up at least every three months to facilitate timely detection of replacement indicators. NOTE: Because the duration of the device replacement timer is three months (starting when Explant status is reached), three month follow up frequency is particularly important after the One Year Remaining status is reached. Consider performing the following procedures during a routine follow-up test: 1. Interrogate the pulse generator and review the Summary screen. 2. Verify pacing thresholds, lead impedance, and amplitude of intrinsic signals. 3. Print the Quick Notes and Patient Data reports to retain in your files for future reference. 4. Review the Arrhythmia Logbook screen and for episodes of interest, print episode details and stored electrogram information. 5. Clear the counters and histograms so that the most recent episode data will be displayed at the next follow-up session. 78 6. Verify that important programmed parameter values (e.g., lower rate limit, AV Delay, rate adaptive pacing, output amplitude, pulse width, sensitivity) are optimal for current patient status. NOTE: Echo-Doppler studies may be used to non-invasively evaluate AV Delay and other programming options post-implant. EXPLANTATION NOTE: Return all explanted pulse generators and leads to Boston Scientific. Examination of explanted pulse generators and leads can provide information for continued improvement in system reliability and warranty considerations. WARNING: Do not reuse, reprocess, or resterilize. Reuse, reprocessing, or resterilization may compromise the structural integrity of the device and/or lead to device failure which, in turn, may result in patient injury, illness, or death. Reuse, reprocessing, or resterilization may also create a risk of contamination of the device and/or cause patient infection or cross-infection, including, but not limited to, the transmission of infectious disease(s) from one patient to another. Contamination of the device may lead to injury, illness, or death of the patient. Complete and send an Observation/Complication/Out-of-Service Reporting form to Boston Scientific when any of the following occur: • When a product is removed from service. 79 • In the event of patient death (regardless of cause), along with an autopsy report, if performed. • For other observation or complications reasons. NOTE: Disposal of explanted pulse generators and/or leads is subject to applicable laws and regulations. For a Returned Product Kit, contact Boston Scientific using the information on the back cover. NOTE: Discoloration of the pulse generator may have occurred due to a normal process of anodization, and has no effect on the pulse generator function. CAUTION: Be sure that the pulse generator is removed before cremation. Cremation and incineration temperatures might cause the pulse generator to explode. CAUTION: Before explanting, cleaning, or shipping the device, complete the following actions to prevent overwriting of important therapy history data: • Program the pulse generator Brady Mode to Off • Program Ventricular Tachy EGM Storage to Off Clean and disinfect the device using standard biohazard handling techniques. Consider the following items when explanting and returning the pulse generator and/or lead: 80 • Interrogate the pulse generator and print a Combined Follow-up report. • Deactivate the pulse generator before explantation. • Disconnect the leads from the pulse generator. • If leads are explanted, attempt to remove them intact, and return them regardless of condition. Do not remove leads with hemostats or any other clamping tool that may damage the leads. Resort to tools only if manual manipulation cannot free the lead. • Wash, but do not submerge, the pulse generator and leads to remove body fluids and debris using a disinfectant solution. Do not allow fluids to enter the pulse generator’s lead ports. • Use a Boston Scientific Returned Product Kit to properly package the pulse generator and send it to Boston Scientific . 81 82 For additional technical reference guides, go to www.bostonscientific.com/ifu. Boston Scientific 4100 Hamline Avenue North St. Paul, MN 55112–5798 USA www.bostonscientific.com 1.800.CARDIAC (227.3422) +1.651.582.4000 © 2011 Boston Scientific or its affiliates. All rights reserved. 358566-004 EN US 6/11 FCC ID: ESCCRMV17311 *358566-004*
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.4 Linearized : Yes Page Count : 88 XMP Toolkit : XMP toolkit 2.9.1-13, framework 1.6 About : uuid:d826733c-e8b2-4de4-8bf8-446ad975bfd6 Producer : Acrobat Distiller 6.0.1 (Windows) Creator Tool : Adobe Illustrator CS3 Create Date : 2011:06:28 17:18:30-05:00 Modify Date : 2011:06:28 17:18:30-05:00 Metadata Date : 2010:01:20 18:27:04-06:00 Thumbnail Width : 256 Thumbnail Height : 128 Thumbnail Format : JPEG Thumbnail Image : (Binary data 9155 bytes, use -b option to extract) Document ID : uuid:8C415927077711DF8C3DD080D375B605 Instance ID : uuid:8C415928077711DF8C3DD080D375B605 Derived From Instance ID : uuid:7D1DA116E9CE11DEAA08F32E60B0114D Derived From Document ID : uuid:7D1DA115E9CE11DEAA08F32E60B0114D N Pages : 1 Has Visible Transparency : False Has Visible Overprint : False Max Page Size W : 8.500000 Max Page Size H : 11.000000 Max Page Size Unit : Inches Font Name : ArialMT Font Family : Arial Font Face : Regular Font Type : TrueType Font Version : Version 2.60 Font Composite : False Font File Name : Arial Plate Names : Cyan, Magenta, Yellow, Black Swatch Group Name : Default Swatch Group Swatch Group Type : 0 Swatch Colorant Swatch Name : White, Black, Charcoal, Graphite, Ash, Smoke, Latte, Capuccino, Mochaccino, Mocha, Mars Red, Ruby, Red, Pumpkin, Squash, Sunshine, Pure Yellow, Peridot, Little Sprout Green, Jade, Mint Julep, Emerald, Sea Green, Caribbean Blue, Pure Cyan, Hawaiian Blue, Twillight Blue, Starry Night Blue, Deep Sea Blue, Fresh Lavender, Purple, Amethyst, Rasberry, Pure Magenta, Global Red, Global Squash, Global Pure Yellow, Global Green, Global Pure Cyan, Global Deep Sea Blue Swatch Colorant Mode : CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK, CMYK Swatch Colorant Type : PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS, PROCESS Swatch Colorant Cyan : 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 10.000002, 15.999996, 32.000000, 42.999996, 25.000000, 25.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 25.000000, 50.000000, 50.000000, 75.000000, 100.000000, 100.000000, 100.000000, 100.000000, 100.000000, 100.000000, 100.000000, 100.000000, 75.000000, 75.000000, 50.000000, 25.000000, 0.000000, 0.000000, 0.000000, 0.000000, 80.000000, 100.000000, 100.000000 Swatch Colorant Magenta : 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 23.000002, 37.000000, 48.000000, 52.000000, 100.000000, 100.000000, 100.000000, 75.000000, 50.000000, 25.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 25.000000, 25.000000, 25.000000, 0.000000, 25.000000, 50.000000, 75.000000, 100.000000, 75.000000, 100.000000, 100.000000, 100.000000, 100.000000, 100.000000, 50.000000, 0.000000, 0.000000, 0.000000, 100.000000 Swatch Colorant Yellow : 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 33.000000, 58.000000, 76.000000, 80.000000, 100.000000, 100.000000, 100.000000, 100.000000, 100.000000, 100.000000, 100.000000, 100.000000, 100.000000, 100.000000, 100.000000, 100.000000, 50.000000, 25.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 100.000000, 100.000000, 100.000000, 100.000000, 0.000000, 0.000000 Swatch Colorant Black : 0.000000, 100.000000, 75.000000, 60.000004, 45.000000, 30.000002, 0.000000, 0.000000, 19.999998, 38.000000, 25.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 25.000000, 25.000000, 25.000000, 25.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000, 0.000000 Swatch Colorant Tint : 100.000000, 100.000000, 100.000000, 100.000000, 100.000000, 100.000000 Format : application/pdf Title : wrench_rotation_side Creator : SYSTEM Author : SYSTEMEXIF Metadata provided by EXIF.tools