Pfeiffer, Dual Gauge TPG 262 Channel Control Unit, For Compact Gauges Pfeiffer TPG262 Operating Instructions

User Manual:

Open the PDF directly: View PDF PDF.
Page Count: 106

DownloadPfeiffer, Dual Gauge TPG 262 Channel Control Unit, For Compact Gauges Pfeiffer TPG262 Operating Instructions
Open PDF In BrowserView PDF
Operating Instructions

DualGauge™
Dual-Channel Measurement and
Control Unit for Compact Gauges

TPG 262
VACUUM

BG 805 196 BE / B

(2004-08)

1

Product Identification

In all communications with Pfeiffer Vacuum, please
specify the information on the product nameplate. For
convenient reference copy that information into the
space provided below.

Pfeiffer Vacuum, D-35614 Asslar
Typ:
No:
F-No:
V

Validity

Hz

VA

This document applies to products with part number
PTG28280.
The part number (No.) can be taken from the product
nameplate.
This manual is based on firmware version 302-510-A.
If your unit does not work as described in this document,
please check that it is equipped with the above firmware
version (→  60).
We reserve the right to make technical changes without
prior notice.
All dimensions are indicated in mm.

2

BG 805 196 BE / B

(2004-08)

TPG262.oi

Intended Use

The TPG 262 is used together with Pfeiffer Vacuum
Compact Gauges (in this document referred to as
gauges) for total pressure measurement. All products
must be operated in accordance with their respective
Operating Instructions.

Scope of Delivery

The scope of delivery consists of following parts:
1 TPG 262 Dual-Channel Measurement and Control
Unit
1 Power cord
1 Connector for control connection
4 Collar screws and plastic sleeves
2 Rubber feet
1 Rubber bar
1 Operating Instructions (this document)
1 Betriebsanleitung

Trademarks

DualGauge™
FullRange™

BG 805 196 BE / B

(2004-08)

TPG262.oi

INFICON AG
INFICON GmbH

3

Contents
Product Identification
Validity
Intended Use
Scope of Delivery
Trademarks

2
2
3
3
3

1 Safety
1.1 Symbols Used
1.2 Personnel Qualifications
1.3 General Safety Instructions
1.4 Liability and Warranty

6
6
7
7
8

2 Technical Data

4

9

3 Installation
3.1 Personnel
3.2 Installation, Setup
3.2.1 Rack Installation
3.2.2 Installation in a Control Panel
3.2.3 Use as Desk-Top Unit
3.3 Mains Power Connector
3.4 Gauge Connectors sensor 1, sensor 2
3.5 control Connector
3.6 relay Connector
3.7 Interface Connector RS232

14
14
14
14
17
18
19
20
21
22
23

4 Operation
4.1 Front Panel
4.2 Turning the TPG 262 On and Off
4.3 Operating Modes
4.4 Measurement mode
4.5 Parameter Mode
4.5.1 Switching Function Parameters
4.5.2 Gauge Parameters
4.5.3 Gauge Control
4.5.4 General Parameters
4.5.5 Test Parameters

24
24
25
26
27
31
33
38
47
54
58

BG 805 196 BE / B

(2004-08)

TPG262.oi

5 Communication (Serial Interface)
5.1 RS232C Interface
5.1.1 Data Transmission
5.1.2 Communication Protocol
5.2 Mnemonics
5.2.1 Measurement Mode
5.2.2 Parameter Mode
5.2.2.1 Switching Function Parameters
5.2.2.2 Gauge Parameters
5.2.2.3 Gauge Control
5.2.2.4 General Parameters
5.2.2.5 Test Parameters
5.2.3 Example

68
68
68
70
72
73
79
79
80
84
85
86
92

6 Maintenance

93

7 Troubleshooting

94

8 Repair

95

9 Storage

96

10 Disposal

96

Appendix
A: Conversion Tables
B: Default Parameter Settings
C: Firmware Update
D: Literature
E: Index

97
97
98
99
102
104

Declaration of Conformity

106

For cross-references within this document, the symbol
(→  XY) is used, for cross-references to further docu[Z]).
ments, listed under "Literature", the symbol (→

BG 805 196 BE / B

(2004-08)

TPG262.oi

5

1 Safety
1.1 Symbols Used
Symbols for residual
risks

DANGER
Information on preventing any kind of physical injury.

WARNING
Information on preventing extensive equipment and
environmental damage.

Caution
Information on correct handling or use. Disregard can
lead to malfunctions or minor equipment damage.

Further symbols
The lamp/display is lit.
The lamp/display flashes.
The lamp/display is dark.
Press the key (example: PARA key).
Do not press any key.

6

BG 805 196 BE / B

(2004-08)

TPG262.oi

1.2 Personnel
Qualifications

Skilled personnel
All work described in this document may only be carried out by persons who have suitable technical training and the necessary experience or who have been
instructed by the end-user of the product.

1.3 General Safety
Instructions

Adhere to the applicable regulations and take the necessary precautions for all work you are going to do and
consider the safety instructions in this document.

DANGER

Caution: mains voltage
Contact with live parts is extremely hazardous
when any objects are introduced or any liquids penetrate into the unit.
Make sure no objects enter through the louvers and no liquids penetrate into the equipment.

Communicate the safety instructions to all other users.

BG 805 196 BE / B

(2004-08)

TPG262.oi

7

1.4 Liability and
Warranty

8

Pfeiffer Vacuum assumes no liability and the warranty
becomes null and void if the end-user or third parties
• disregard the information in this document
• use the product in a non-conforming manner
• make any kind of interventions (modifications, alterations etc.) on the product
• use the product with accessories not listed in the corresponding product documentation.

BG 805 196 BE / B

(2004-08)

TPG262.oi

2 Technical Data
Mains specifications

Voltage
Frequency
Power consumption
Overvoltage category
Protection class
Connection

Ambiance

Temperature
storage
operation
Relative humidity
Use
Pollution degree
Protection type

Compatible gauges

Number
Compatible
Compact Gauges
Pirani
Pirani Capacitance
Cold Cathode
FullRange™ CC
Process Ion
FullRange™ BA
Capacitance
Piezo

Gauge connections

BG 805 196 BE / B

(2004-08)

TPG262.oi

Number
sensor connector

90 … 250 VAC
50 … 60 Hz
≤45 W
II
1
European appliance connector IEC 320 C14

–20 … +65 °C
+ 5 … +50 °C
≤80% up to +31 °C,
decreasing to 50% at +40 °C
indoors only
max. altitude 2000 m NN
II
IP30

2

TPR 261, TPR 265, TPR 280,
TPR 281
PCR 260
IKR 251, IKR 261, IKR 270
PKR 251, PKR 261
IMR 265
PBR 260
CMR 261 … CMR 275
APR 250 … APR 267

2 (1 per channel)
Amphenol C91B appliance
connector, female, 6-pole
(pin assignment →  20)

9

Gauge supply

Voltage
Current
Power
Fuse protection

+24 VDC ±5%
750 mA
18 W
900 mA with PTC element,
self-resetting after turning the
TPG 262 off or disconnecting
the gauge. The supply conforms to the requirements of a
grounded protective extra low
voltage (SELV-E according to
EN 61010).

Operation

Front panel
Remote control

via 4 keys
via RS232C interface

Measurement values

Measurement ranges

depending on gauges
(→
[1] … [14])

Measurement error
gain error
offset error
Measurement rate
Display rate
Filter time constant
slow
normal (nor)
fast
Measurement units
Offset correction
Calibration factor

A/D converter

10

≤0.01% F.S.
≤0.01% F.S.
50 / s
10 / s
1.2 s (fg = 0.13 Hz)
400 ms (fg = 0.4 Hz)
20 ms (fg = 8 Hz)
mbar, Pa, Torr
for linear gauges
–5 … 110% F.S.
for logarithmic gauges
0.10 … 9.99
for linear gauges
0.500 … 2.000
resolution 0.001% F.S.

BG 805 196 BE / B

(2004-08)

TPG262.oi

Switching functions

Number
Reaction delay

Adjustment range
Hysteresis

Switching function relays

Contact type
Load max.
Service life
mechanic
electric
Contact positions
Relay connector

4 (user-assignable)
≤20 ms if switching threshold
close to measurement value
(for larger differences consider filter time constant).
depending on gauge
(→
[1] … [14])
≥1% F.S. for linear gauges,
≥10% of measurement value
for logarithmic gauges

floating changeover contact
30 VAC, 30 W (ohmic)
60 VDC, 1 A, 30 W (ohmic)
5×107 cycles
1×105 cycles (at max. load)
→  22
D-Sub appliance connector,
female, 15-pole
(pin assignment →  22)

Error signal

Number
Reaction time

1
≤20 ms

Error signal relay

Contact type
Load max.

floating normally open contact
30 VAC, 30 W (ohmic)
60 VDC, 1 A, 30 W (ohmic)

Service life
mechanic
electric
Contact positions
Control connector

BG 805 196 BE / B

(2004-08)

TPG262.oi

5×107 cycles
1×105 cycles (at max. load)
→  21
Amphenol C91B appliance
connector, female, 7-pole
(pin assignment →  21)

11

Gauge control

Automatic
ON setpoint
OFF setpoint

adjustable (→  50)
adjustable (→  52)

Manual
via keys
activation/deactivation (→  28, 49, 51)
External
via control connector
ON condition
OFF condition

signal ≤ +0.8 VDC
signal +2.0 … 5 VDC or input
open

Hotstart
when mains power on (→  50)
Self control
deactivation when
pressure is rising
OFF threshold
Control connector

Analog outputs

Number
Voltage range
Internal resistance
Measuring signal vs.
pressure
Control connector

2 (1 per channel)
0 … +10 VDC
660 Ω
depending on gauge
(→
[1] … [14])
Amphenol C91B appliance
connector, female, 7-pole
(pin assignment →  21)

Interface

Standard
Protocol

RS232C
ACK/NAK, ASCII with
3-character mnemonics,
bi-directional data flow,
8 data bits, no parity bit,
1 stop bit
only TXD and RXD used
9600, 19200, 38400 baud
D-Sub appliance connector,
male, 9-pole
(pin assignment →  23)

RS232C
Transmission rate
RS232 connector

12

adjustable (→  52)
Amphenol C91B appliance
connector, female, 7-pole
(pin assignment →  21)

BG 805 196 BE / B

(2004-08)

TPG262.oi

Dimensions [mm]
70.8
55.9

≈310
228.5

ø3.5

2.5

68.1

111.6

122.5
128.5

VACUUM

Use

For incorporation into a rack or control panel or as desktop unit.

Weight

1.06 kg

BG 805 196 BE / B

(2004-08)

TPG262.oi

13

3 Installation
3.1 Personnel
Skilled personnel

The unit may only be installed by persons
who have suitable technical training and the
necessary experience.

3.2 Installation, Setup

The TPG 262 is suited for incorporation into a 19" rack
or a control panel or for use as desk-top unit.
DANGER

Caution: damaged product
Putting a damaged product into operation can
be extremely hazardous.
In case of visible damages, make sure the
product is not put into operation.

3.2.1 Rack Installation

The TPG 262 is designed for installation into a 19" rack
chassis adapter according to DIN 41 494. For this purpose, four collar screws and plastic sleeves are supplied
with it.
DANGER

Caution: protection class of the rack
If the product is installed in a rack, it is likely
to lower the protection class of the rack
(protection against foreign bodies and water)
e.g. according to the EN 60204-1 regulations
for switching cabinets.
Take appropriate measures for the rack to
meet the specifications of the protection
class.

14

BG 805 196 BE / B

(2004-08)

TPG262.oi

Guide rail

In order to reduce the mechanical strain on the front
panel of the TPG 262, preferably equip the rack chassis
adapter with a guide rail.

Slide rails

For safe and easy installation of heavy rack chassis
adapters, preferably equip the rack frame with slide rails.

BG 805 196 BE / B

(2004-08)

TPG262.oi

15

Height 3 U rack
chassis adapter

n

Secure the rack adapter in the rack frame.
The admissible maximum ambient temperature (→  9) must not be exceeded
neither the air circulation obstructed.
Rack chassis adapter
Height 3 U

o

Slide the TPG 262 into the rack chassis adapter …

… and fasten the adapter panel to the rack chassis adapter using the screws supplied with the
TPG 262.

16

BG 805 196 BE / B

(2004-08)

TPG262.oi

3.2.2 Installation in a
Control Panel

DANGER

Caution: protection class of the control panel
If the product is installed in a control panel, it
is likely to lower the protection class of the
control panel (protection against foreign
bodies and water) e.g. according to the
EN 60204-1 regulations for switching cabinets.
Take appropriate measures for the control
panel to meet the specifications of the protection class.
For mounting the TPG 262 into a control panel, the following cut-out is required:
70

The admissible maximum ambient temperature (→  9) must
not be exceeded neither the air circulation
obstructed.

113

122.5

55.9

M3 or ø 3.5

For reducing the mechanical strain on the front panel,
preferably support the unit.

BG 805 196 BE / B

(2004-08)

TPG262.oi

17

n

Slide the TPG 262 into the cut-out of the control
panel …

… and secure it with four M3 or equivalent screws.

3.2.3 Use as Desk-Top
Unit

The TPG 262 is also suited for use as desk-top unit. For
this purpose, two self-adhesive rubber feet as well as a
slip-on rubber bar are supplied with it.

n

Stick the two supplied rubber feet to the rear part
of the bottom plate …

… and slip the supplied rubber bar onto the bottom edge of the front panel.

18

BG 805 196 BE / B

(2004-08)

TPG262.oi

Select a location where the admissible maximum ambient temperature (→  9) is not exceeded (e.g. due to sun irradiation).

3.3 Mains Power
Connector

DANGER

Caution: line voltage
Incorrectly grounded products can be extremely hazardous in the event of a fault.
Use only a 3-conductor power cable with
protective ground. The power connector may
only be plugged into a socket with a protective ground. The protection must not be nullified by an extension cable without protective
ground.
The unit is supplied with a power cord. If the mains connector is not compatible with your system, use your own,
suitable cable with protective ground (3×1.5 mm3).

The socket must be
fuse-protected with
10 Amax

If the unit is installed in a switching cabinet, the mains
voltage should be supplied and turned on via a central
distributor.

BG 805 196 BE / B

(2004-08)

TPG262.oi

19

Do not unfasten
this screw
(internal ground
protection)

3.4 Gauge Connectors
sensor 1, sensor 2

For each measurement channel, there is a female appliance connector on the rear of the unit.
Connect the gauge to the sensor connector via
a sensor cable set available from us (→ sales
literature) or your own, screened (electromagnetic compatibility) sensor cable. Use compatible gauges only (→  9).

Pin assignment
sensor 1, sensor 2

2
1

2

1

3

4

5

3

6

6
5

4

Pin assignment of
the two female
6-pole Amphenol
C91B appliance
connectors:
Pin
1
6
2
3
4
5

20

Signal
Identification
Supply
Supply common
Signal input
Signal common
Screening

+24 VDC
GND
(measuring signal+)
(measuring signal–)

BG 805 196 BE / B

(2004-08)

TPG262.oi

3.5 control Connector

This connector allows to read the measuring signal, to
evaluate the state of the floating contacts of the error
relay, and to activate or deactivate the gauges (→  47).
Connect the peripheral components to the
control connector on the rear of the unit using
your own, screened (electromagnetic compatibility) cable.

Pin assignment
Contact positions
control

1

Pin
2
1
5
4
6
3
7

1
6

2

2

3

7

8

8

4

7

3

5

Pin assignment of
the female 7-pole
Amphenol C91B appliance connector:

4

6

5

Signal
Analog output gauge 1
Analog output gauge 2
Screening GND
Gauge 1 on
off
Gauge 2 on
off

0 … 10 VDC
0 … 10 VDC

signal ≤+0.8 VDC
signal +2.0 … 5 VDC or input open
signal ≤+0.8 VDC
signal +2.0 … 5 VDC or input open

No error

Error or power
supply turned off

A suitable connector is supplied with the TPG 262.

BG 805 196 BE / B

(2004-08)

TPG262.oi

21

3.6 relay Connector

This connector allows to use the floating switching contacts for an external control system.
Connect the peripheral components to the relay
connector on the rear of the unit using your
own, screened (electromagnetic compatibility)
cable.

Pin assignment
Contact positions
relay

8
15

9
1

Pin assignment of
the female 15-pole
D-Sub appliance
connector:

Pin

Signal
Switching function 1

4
3
2

Pressure below
threshold

Pressure above
threshold or power
supply turned off

Switching function 2
7
6
5

Pressure below
threshold

Pressure above
threshold or power
supply turned off

Switching function 3
11
10
9

Pressure below
threshold

Pressure above
threshold or power
supply turned off

Switching function 4
14
13
12

15
1
8

22

Pressure below
threshold

Pressure above
threshold or power
supply turned off

Supply for relays with higher switching power
Fuse-protected at 300 mA with PTC
element, self-resetting after power
+24 VDC, 200 mA
off or pulling the relay connector.
GND
Meets the requirements of a
GND
grounded protective extra low voltage (SELV-E according to
EN 61010).

BG 805 196 BE / B

(2004-08)

TPG262.oi

3.7 Interface Connector
RS232

The RS232C interface allows for operating the TPG 262
via a HOST or terminal (→  68). It can also be used for
updating the firmware (→  99).
Connect the serial interface to the RS232 connector on the rear of the unit using your own,
screened (electromagnetic compatibility) cable.
e.g. PC

TPG 262

Chassis

Chassis

RXD

2 RXD

TXD

3 TXD

GND

5 GND
(Minimum configuration)

Pin assignnment
RS232

Screening

Pin assignment of
the male 9-pole
D-Sub appliance
connector:

BG 805 196 BE / B

(2004-08)

TPG262.oi

9
5
1
6

Pin

Signal

Pin

2
3
5

RXD
TXD
GND

1
6
9

4
7
8

DTR
RTS
CTS

Signal
not connected
not connected
not connected

Casing = screening

23

4 Operation
4.1 Front Panel
Measurement value in floating point
or exponential format or status messages

Parameter mode
activated

VACUUM

Measurement
unit

Warning/error
(flashing)

Operator keys

Switching function status
Offset value ≠ 0
Correction factor ≠ 1
Degas activated
Gauge activated
Active measurement channel

24

BG 805 196 BE / B

(2004-08)

TPG262.oi

4.2 Turning the
TPG 262 On and Off

Turning the TPG 262
on

Make sure the TPG 262 is correctly installed and the
specifications in the Technical Data are met.

The power switch is on the
rear of the unit.
Turn the TPG 262 on with
the power switch (or centrally, via a switched power
distributor, if the unit is incorporated in a rack).

After power on, the TPG 262 …
• automatically performs a self-test
• identifies the connected gauge
• activates the parameters that were in effect before
the last power off
• switches to the Measurement mode
• adapts the parameters if required (if another gauge
was previously connected).

Turning the TPG 262
off

Turn the TPG 262 off with the power switch (or centrally,
via a switched power distributor, if the unit is incorporated in a rack).
Wait at least 10 s before turning the TPG 262
on again in order for it to correctly initialize itself.

BG 805 196 BE / B

(2004-08)

TPG262.oi

25

4.3 Operating Modes

The TPG 262 works in the following operating modes:
• Measurement mode
for displaying measurement values or statuses
(→  27)
• Parameter mode
for displaying or editing parameters (→  31)
−

Switching function parameter group
for entering or displaying thresholds (→  33)

−

Gauge parameter group
for entering or displaying gauge parameters
(→  38)

−

Gauge control group
for entering or displaying gauge control parameters (→  47)

−

General parameter group
for entering or displaying general parameters
(→  54)

−

Test program group
for running internal test programs (→  58)
• Program transfer mode
for updating the firmware (→  99)

26

BG 805 196 BE / B

(2004-08)

TPG262.oi

4.4 Measurement mode

The Measurement mode is the standard operating mode
of the TPG 262. Measurement values and statuses as
well as the gauge identification are displayed in this
mode.

Measurement mode
depending
on history

Parameter
mode

Measurement value

Status

Gauge identification

Power on
Measurement value

Status

Gauge identification

>10 s

BG 805 196 BE / B

(2004-08)

TPG262.oi

27

Selecting a measurement channel

Ö Channel 1
is activated

Ö Channel 2
is activated

Turning a gauge on
and off

Certain gauges can be turned on and off manually, if the
gauge control is set to
(→  51).
Available for:
† Pirani Gauge
† Pirani Capacitance Gauge
; Cold Cathode Gauge
; FullRange™ CC Gauge
; Process Ion Gauge
; FullRange™ BA Gauge
† Capacitance Gauge
† Piezo Gauge

(TPR)
(PCR)
(IKR)
(PKR)
(IMR)
(PBR)
(CMR)
(APR)

Ö Press key >1 s:
The gauge is
turned off.
is displayed instead of
the measurement
value.
Ö Press key >1 s:
The gauge is
turned on. A status
message may be
displayed instead
of the measurement value.

28

BG 805 196 BE / B

(2004-08)

TPG262.oi

Measurement
range

Pressure p
BG 805 196 BE / B

(2004-08)

TPG262.oi

29

Ö Press keys >0.5 s:
The type of the connected
gauges is automatically identified and displayed for 4 s
(2 s per channel):

Displaying the gauge
identification

Pirani Gauge
(TPR 261, TPR 265, TPR 280,
TPR 281)

Pirani Capacitance Gauge

1)

(PCR 260)

Cold Cathode Gauge
(IKR251, IKR261)

Cold Cathode Gauge
(IKR270)

FullRange™ CC Gauge
(PKR251, PKR261)

Process Ion Gauge
(IMR265)

FullRange™ BA Gauge
(PBR260)

Capacitance Gauge
(CMR261 … CMR275)

Piezo Gauge
(APR250 … APR267)

No gauge connected
(no Sensor)

Connected gauge cannot be
identified (no Identifier)
1)

TPR and PCR have identical identifiers. In
the TPG 262, there is no distinction made on
the display and in data evaluation, since pressure ranges of these gauges are approximately
the same.

Getting to the
Parameter mode

30

→  31

BG 805 196 BE / B

(2004-08)

TPG262.oi

4.5 Parameter Mode

The Parameter mode is used for displaying, editing and
entering parameter values as well as for testing the
TPG 262. For ease of operation, the parameters are divided into groups.

Measurement
mode

Parameter mode
Switching function parameters
depending
on history

Gauge parameters

Gauge control

Power on

General parameters

>10 s
Test parameters

BG 805 196 BE / B

(2004-08)

TPG262.oi

31

Selecting a parameter
group

Ö Switching function parameters →  33
Gauge parameters →  38
Gauge control →  47
General parameters
→  54
Test parameters
→  58

Selecting a parameter
in a parameter group

Editing a parameter in
a parameters group

32

Modifications of parameters come into effect immediately and are stored automatically. Exceptions are mentioned under the corresponding parameters.

BG 805 196 BE / B

(2004-08)

TPG262.oi

4.5.1 Switching
Function
Parameters

The switching function parameter
group (setpoint parameters) is
used for displaying, entering and
editing threshold values and assigning the four switching functions to a measurement channel.

Parameter mode
depending
on history

Switching function 1 lower threshold

Measurement mode

Switching function 1 upper threshold

>10 s
Switching function 2 lower threshold

Swiching function 4 upper threshold

BG 805 196 BE / B

(2004-08)

TPG262.oi

33

The TPG 262 has four switching functions with two adjustable thresholds each. The status of the switching
function is displayed on the front panel (→  24, 21) and
can be evaluated via the floating contacts at the Control
connector.

Pressure p

ue
ent val
urem
s
a
Me

Switching
function

Display

Time t

Selecting a parameter

4
3
2

4
3
2

Off

4
3
2

On

Off

Ö The name of the parameter,
e.g.:
Switching function 1
lower setpoint

is displayed as long as the
key is pressed or at least for
1.5 s.
Afterwards, the currently
valid threshold value is displayed.

34

BG 805 196 BE / B

(2004-08)

TPG262.oi

Ö Press key <1 s:
The value is increased/
decreased by 1 increment.

Editing the threshold
value

Press key >1 s:
The value is increased/
decreased continuously.

Value

Limits of the lower
switching thresholds

The lower switching threshold
(Setpoint low) defines the pressure at which the switching function is activated when the pressure is dropping.
Ö gauge dependent
(→ table).
If another gauge type is connected, the TPG 262 automatically adjusts the switching
threshold if required.

upper
threshold
limit

lower
threshold
limit

e.g.:

5×10-4

1500

1×10-9

1×10-2

1×10-11

1×10-2

1×10-9

1000

1×10-6

1000

5×10

1000

-10

F.S. / 1000

F.S.

all values in mbar, CAL=1

BG 805 196 BE / B

(2004-08)

TPG262.oi

35

The minimum hysteresis between the upper
and lower switching threshold is at least 10% of
the lower threshold or 1% of the set full scale
value. If the value of the minimum hysteresis
drops below these values, the upper threshold
is automatically adjusted to a minimum hysteresis. This prevents unstable states.

Value

Limits of the upper
switching thresholds

The upper switching threshold
(Setpoint high) defines the pressure at which the switching function is deactivated when the pressure is rising.

lower
threshold
limit
lower threshold

upper
thershold
limit

Ö Gauge dependent
(→ table).
If another gauge type is connected, the TPG 262 automatically adjusts the threshold if
required.

e.g.:

+10% lower threshold

1500

+10% lower threshold

1×10-2

+10% lower threshold

1×10-2

+10% lower threshold

1000

+10% lower threshold

1000

+10% lower threshold

1000

+1% measurement
range (F.S.)

F.S.

all values in mbar, CAL=1

36

BG 805 196 BE / B

(2004-08)

TPG262.oi

The minimum hysteresis between the upper
and lower switching threshold is at least 10% of
the lower threshold or 1% of the set full scale
value. If the value of the minimum hysteresis
drops below these values, the upper threshold
is automatically adjusted to a minimum hysteresis. This prevents unstable states.

Ö Switching
function is
assigned to
channel 1.

Assigning a switching
function

Ö Switching
function is
assigned to
channel 2.
The lower

and the upper
switching threshold of a switching
function are always assigned to the same
channel. The last assignment is valid for both
thresholds.

BG 805 196 BE / B

(2004-08)

TPG262.oi

37

4.5.2 Gauge Parameters

The Gauge parameter group
(sensor parameters) is used for
displaying, entering and editing
parameters of the connected
gauges.
Parameter mode

depending
on history

Filter time constant

Measurement mode

Calibration factor

Measurement range

>10 s

Offset

Penning Underrange Control

Degas

38

BG 805 196 BE / B

(2004-08)

TPG262.oi

Ö The name of the parameter,

Selecting a parameter

e.g.:
Filter time constant

is displayed as long as the
key is pressed or at least for
1.5 s.
Afterwards, the currently
valid threshold value is displayed.
Some parameters are not available for all gauges and
thus not always displayed.

Available for

→  39

41

42

43

44

46

*)
*)

*) depending on pressure

Measurement value
filter

The measurement value filter permits a better evaluation
of unstable or disturbed measuring signals.
The measurement value filter does not affect
the analog output (→  21).

BG 805 196 BE / B

(2004-08)

TPG262.oi

39

Value

Ö Fast:

The TPG 262 responds quickly
to fluctuations of the measurement value. As a result, it will
respond faster to interference
in the measured values.
Pressure p

Time t

Ö Normal:

Good relationship between response and sensitivity of the
display and the switching function to changes in the measured values.
Pressure p

Time t

Ö Slow:

The TPG 262 does not respond to small changes in
measured values. As a result,
it will respond more slowly to
changes in the measured
values.
Pressure p

Time t

40

BG 805 196 BE / B

(2004-08)

TPG262.oi

Ö The value is increased/

decreased by the defined increments.

Calibration factor

The calibration factor allows the measured value to be
calibrated for other gases than N2 (→ characteristic
curves in
[1] … [12]).
Available for:
; Pirani Gauge
; Pirani Capacitance Gauge
; Cold Cathode Gauge
; FullRange™ CC Gauge
; Process Ion Gauge *)
; FullRange™ BA Gauge **)
; Capacitance Gauge
; Piezo Gauge
*)
**)

(TPR)
(PCR)
(IKR)
(PKR)
(IMR)
(PBR)
(CMR)
(APR)

only for pressures <1×10-2 mbar.
only for pressures <1×10-1 mbar.

Value

Ö No correction
e.g.:

e.g.:

BG 805 196 BE / B

(2004-08)

TPG262.oi

Ö Measurement value
corrected by a factor
of 0.10 … 9.99
(logarithmic gauges).
Measurement value
corrected by a factor
of 0.500 … 2.000
(linear gauges).

41

Ö Press key <1 s:
The value is increased/
decreased by 1 increment.
Press key >1 s:
The value is increased/
decreased continuously.

Measurement range
(F.S.) of linear
gauges

For linear gauges, the full scale (F.S.) value has to be
defined according to the connected gauge type. For
logarithmic gauges it is automatically recognized.
Available for:
† Pirani Gauge
† Pirani Capacitance Gauge
† Cold Cathode Gauge
† FullRange™ CC Gauge
† Process Ion Gauge
† FullRange™ BA Gauge
; Capacitance Gauge
; Piezo Gauge

(TPR)
(PCR)
(IKR)
(PKR)
(IMR)
(PBR)
(CMR)
(APR)

Value

e.g.:

Ö 0.01 mbar
0.1 mbar
1 mbar
10 mbar
100 mbar
1000 mbar
2 bar
5 bar
10 bar
50 bar
Conversion table
→ Appendix  97
Ö The value is increased/
decreased by the defined increments.

42

BG 805 196 BE / B

(2004-08)

TPG262.oi

Offset correction

The offset value is displayed and readjusted according
to the actual measurement value (in the range of
-5 … +110% of the set full scale value).
Available for:
† Pirani Gauge
† Pirani Capacitance Gauge
† Cold Cathode Gauge
† FullRange™ CC Gauge
† Process Ion Gauge
† FullRange™ BA Gauge
; Capacitance Gauge
; Piezo Gauge

(TPR)
(PCR)
(IKR)
(PKR)
(IMR)
(PBR)
(CMR)
(APR)

The offset correction affects:
; the displayed measurement value
† the displayed threshold value of the switching
functions
† the analog outputs at the control connector (→  21)
Value

Ö Offset correction
deactivated
e.g.:

Ö Offset correction
activated
Ö Press key >1.5 s:
The offset value is readjusted. The actual measurement value is accepted as
new offset value.
Ö Reset the offset value.

When the offset correction is activated, the saved offset
value is subtracted from the actual measurement value.
This allows measuring relative to a reference pressure.
When the zero of the gauge is readjusted, the
offset correction must be deactivated.

BG 805 196 BE / B

(2004-08)

TPG262.oi

43

Underrange control

Behavior in the event of an underrange with Cold Cathode Gauges (Penning underrange control).
Available for:
† Pirani Gauge
† Pirani Capacitance Gauge
; Cold Cathode Gauge
† FullRange™ CC Gauge
† Process Ion Gauge
† FullRange™ BA Gauge
† Capacitance Gauge
† Piezo Gauge

(TPR)
(PCR)
(IKR)
(PKR)
(IMR)
(PBR)
(CMR)
(APR)

There is a number of possible causes of an underrange:
• the pressure in the vacuum system is lower than the
measurement range
• the measurement element has not ignited (yet)
• the discharge has failed
• a defect has occurred

Caution

Caution: relay is switching
An underrange can lead to unintended reactions of the connected control system.
Prevent false control signals and messages
by disconnecting the sensor and control cables.
Value

Ö Underrange state is interpreted
as admissible measurement
value.
is displayed.
The switching function remains
ON.
Ö Underrange state is interpreted
as inadmissible measurement
value.
is displayed.
The switching function
changes to OFF.

44

BG 805 196 BE / B

(2004-08)

TPG262.oi

Ö Activate/deactivate the
underrange control.

If chances are that the pressure in the vacuum
system drops below the measurement range of
the gauge, it is advisable to select
.
If
is selected, the evaluation of the
switching function is suppressed for approx.
10 seconds when the gauge is turned on and
each time after an underrange has occurred.
During this time, the switching function remains
OFF.

BG 805 196 BE / B

(2004-08)

TPG262.oi

45

Degas

Contamination deposits on the electrode system of hot
cathode gauges may cause instabilities of the measurement values. The Degas function allows to clean the
electrode system.
Available for:
(TPR)
† Pirani Gauge
† Pirani Capacitance Gauge (PCR)
† Cold Cathode Gauge
(IKR)
† FullRange™ CC Gauge
(PKR)
† Process Ion Gauge
(IMR)
; FullRange™ BA Gauge
(PBR )
† Capacitance Gauge
(CMR)
† Piezo Gauge
(APR)
Value

Ö Normal operation.
Ö Degas: The electron
collection grid is
heated to ≈700 °C by
electron bombardment and the electrode system is thus
cleaned.
Ö Start Degas.
Duration of the Degas function 3 min. (can be aborted).
Ö Abort Degas.

46

BG 805 196 BE / B

(2004-08)

TPG262.oi

4.5.3 Gauge Control

The Gauge control group (control
parameters) is used for displaying, entering and editing parameters which define how the connected gauges are activated/ deactivated.
If the connected gauges
cannot be controlled
(→  49), this group is
not available.
Parameter mode

depending
on history

Measurement mode

Gauge activation

ON threshold

Gauge deactivation

>10 s

OFF threshold

BG 805 196 BE / B

(2004-08)

TPG262.oi

47

Ö The name of the parameter,

Selecting a parameter

e.g.:
Gauge activation

is displayed as long as the
key is pressed or at least for
1.5 s.
Afterwards, the currently
valid threshold value is displayed.
Some parameters are not available for all gauges and
thus not always displayed.
50

51

52

Available for

→  49

48

BG 805 196 BE / B

(2004-08)

TPG262.oi

Gauge activation

Certain gauges can be activated by different means.
The following gauges can be controlled:
(TPR)
† Pirani Gauge
† Pirani Capacitance Gauge (PCR)
; Cold Cathode Gauge
(IKR)
; FullRange™ CC Gauge *) (PKR)
; Process Ion Gauge
(IMR)
; FullRange™ BA Gauge
(PBR)
† Capacitance Gauge
(CMR)
† Piezo Gauge
(APR)
*)

except by a gauge connected to the other measurement
channel

Value

Ö Automatic activation:
The gauge is activated by one
of the following gauges connected to the other measurement channel.
;
;
†
;
;
;
;
;
*)

Pirani Gauge
Pirani Capacitance Gauge
Cold Cathode Gauge
FullRange™ CC Gauge
Process Ion Gauge
FullRange™ BA Gauge
Capacitance Gauge *)
Piezo Gauge *)

(TPR)
(PCR)
(IKR)
(PKR)
(IMR)
(PBR)
(CMR)
(APR)

only gauges with F.S.
1, 10 or 100 mbar

Ö Manual activation:
The gauge is activated by
pressing the
key.
Ö External activation:
The gauge is activated by an
input signal fed via the control
connector (→  21).
Ö Hot start:
The gauge is automatically activated when the TPG 262 is
turned on. Measurement is
thus automatically resumed
after a power failure. Gauge
deactivation →  51.

BG 805 196 BE / B

(2004-08)

TPG262.oi

49

Ö Increase/decrease the value
by the defined increments.

ON threshold

Definition of the ON threshold for the gauge to be activated by a gauge connected to the other measurement
channel.
Available for:
† Pirani Gauge
† Pirani Capacitance Gauge
; Cold Cathode Gauge
† FullRange™ CC Gauge
; Process Ion Gauge
; FullRange™ BA Gauge
† Capacitance Gauge
† Piezo Gauge

(TPR)
(PCR)
(IKR)
(PKR)
(IMR)
(PBR)
(CMR)
(APR)

Adjustment range

→ table below

e.g.:

PKR
IMR
PBR

TPR
PCR
-3

-2

-5

CMR, APR
F.S.=1
-2

-3

F.S.=10
-2

F.S.=100

IKR

10 …10

10 …10

10 …10

IMR

-3

10 …1

-5

10 …1

-3

—

—

-2

-1

10 …1

10 …1

10 …1

PBR

10-3…1

10-5…1

10-3…1

10-2…1

10-1…1

all values in mbar, CAL=1

Value

50

must be ≥

BG 805 196 BE / B

.

(2004-08)

TPG262.oi

Ö Press key <1 s:
The value is increased/
decreased by 1 increment.
Press key >1 s:
The value is increased/
decreased continuously.

Gauge deactivation

Certain gauges can be deactivated by different means.
The following gauges can be controlled:
(TPR)
† Pirani Gauge
† Pirani Capacitance Gauge (PCR)
; Cold Cathode Gauge
(IKR)
; FullRange™ CC Gauge *, **) (PKRx)
; Process Ion Gauge *)
(IMR)
; FullRange™ BA Gauge *) (PBR)
† Capacitance Gauge
(CMRx)
† Piezo Gauge
(APR)
*)
**)

except for self control
except by a gauge connected to the other measurement
channel

Value

Ö Automatic deactivation:
The gauge is deactivated by
one of the following gauges
connected to the other measurement channel.
;
;
†
;
;
;
;
;
*)

Pirani Gauge
Pirani Capacitance Gauge
Cold Cathode Gauge
FullRange™ CC Gauge
Process Ion Gauge
FullRange™ BA Gauge
Capacitance Gauge *)
Piezo Gauge *)

(TPR)
(PCR)
(IKR)
(PKR)
(IMR)
(PBR)
(CMR)
(APR)

only for gauges with F.S. 1, 10, or
100 mbar

Ö Manual deactivation:
The gauge is deactivated by
pressing the
key.
Ö External deactivation:
The gauge is deactivated by
an input signal via the control
BG 805 196 BE / B

(2004-08)

TPG262.oi

51

connector (→  21).
Additionally for Cold
Cathode Gauge:

Ö Self control:
The gauge deactivates itself
when the pressure rises
(→  52).
Ö Increase/decrease the value
by the defined increments.

OFF threshold

Definition of the OFF threshold for the gauge to be deactivated by a gauge connected to the other measurement channel or by itself.
Available for:
† Pirani Gauge
† Pirani Capacitance Gauge
; Cold Cathode Gauge
† FullRange™ CC Gauge
; Process Ion Gauge
; FullRange™ BA Gauge
† Capacitance Gauge
† Piezo Gauge

(TPR)
(PCR)
(IKRx)
(PKR)
(IMR)
(PBR)
(CMR)
(APR)

Adjustment range

→ table below

e.g.:

52

BG 805 196 BE / B

(2004-08)

TPG262.oi

IKR
IMR
PBR

TPR
PCR

PKR
IMR
PBR

10-3…10-2

10-5…10-2

-3

10 …1
-3

10 …1

CMR, APR
F.S.=1

-5

10 …1
-5

10 …1

10-3…10-2
-3

10 …1
-3

10 …1

F.S.=10

F.S.=100

—

—

-2

-1

10 …1

-2

10-1…1

10 …1
10 …1

all values in mbar, CAL=1

Value

must be ≥

.

Ö Press key <1 s:
The value is increased/
decreased by 1 increment.
Press key >1 s:
The value is increased/
decreased continuously.

BG 805 196 BE / B

(2004-08)

TPG262.oi

53

4.5.4 General
Parameters

The General parameters group
(general parameters) is used for
displaying, entering and editing
generally applicable system parameters.
Parameter mode

depending
on history

Measurement unit

Measurement mode

Transmission rate

Display resolution

>10 s
Default settings

Display changeover

54

BG 805 196 BE / B

(2004-08)

TPG262.oi

Ö The name of the parameter,

Selecting a parameter

e.g.:
Measurement unit

is displayed as long as the
key is pressed or at least for
1.5 s.
Afterwards, the currently
valid threshold value is displayed.
The parameters are available for all gauge types and
thus always displayed.

Available for

→  56

Editing a parameter

BG 805 196 BE / B

(2004-08)

TPG262.oi

56

56

57

57

all gauges

Ö Increase/decrease the value
by the defined increments.

55

Measurement unit

Unit of measured values, thresholds etc. See Appendix
(→  97) for conversion.
Value

Ö mbar/bar

Ö Torr (only available if
Torr lock is not activated
i.e. Torr is not suppressed → 63)

Ö Pascal

Transmission rate

Transmission rate of the RS232C interface.
Value

e.g.:

Display resolution
(digits)

Ö 9600 baud
19200 baud
38400 baud

Display resolution of measured values.
Value

Ö Display
• rounded to one
decimal digit
• or two integrals
Ö Display
• rounded to two
decimal digits
• or three integrals
56

BG 805 196 BE / B

(2004-08)

TPG262.oi

Default values

All user parameter settings are replaced by the factory
settings.
Loading of the default parameter settings is irreversible.
Value

Ö The default values are loaded
(→  98).

Display changeover

Definition of the measurement display behavior when a
Pirani gauge or a Pirani Capacitance Gauge is combined with a linear gauge with F.S. 1000 mbar.
10 mbar
1000 mbar
5×10-4 mbar
Pirani gauge
Linear gauge
or
Pirani Capacitance
Gauge

Automatic display changeover is available for
this gauge combination only.
Value

Ö Manual change of measurement value display
Ö Automatic change of measurement value display when the
measured value of the linear
gauge drops below or rises
above 10 mbar

BG 805 196 BE / B

(2004-08)

TPG262.oi

57

4.5.5 Test Parameters

The Test parameter group is used
for displaying the firmware version, entering and editing special
parameter values, and for running
test programs.
This group is only available if the
key was
pressed while the
TPG 262 was turned on.
Parameter mode

depending
on history

Power
on

Firmware version

Measurement mode

Watchdog control

Keylock

RAM test program

RS232C test program

58

BG 805 196 BE / B

(2004-08)

TPG262.oi

Ö The name of the parameter,

Selecting a parameter

e.g.:
Firmware version

is displayed.
The parameters are available for all gauge types and
thus always displayed.
→  60

61

61

Available for
all gauges

61

The name of the parameter is displayed as long as the key is pressed
or at least for 1.5 s.
The firmware version is continuously displayed.
62

63

63

64

64

64

65

65

67

Available for
all gauges

→  62

The name of the test program is displayed until
it is started.

BG 805 196 BE / B

(2004-08)

TPG262.oi

59

Editing a parameter

Ö Increase/decrease the value
by the defined increments.

Starting the test
program

Ö Start test program.

Firmware version

The firmware version (program version) is displayed.
Version

Ö The two parts of the firmware
number are displayed alternately.

The last character indicates the modification index (-, A … Z). Please mention
this index when contacting
Pfeiffer Vacuum in the event of a problem.

60

BG 805 196 BE / B

(2004-08)

TPG262.oi

Watchdog control

Behavior of the system control (watchdog) in the event
of an error.
Setting

Ö The system automatically acknowledges error messages of
the watchdog after 2 s.
Ö Error messages of the watchdog have to be acknowledged
by the operator.

Torr lock

The measurement unit
can be suppressed in the
corresponding parameter setting
(→  56).
Setting

Ö Measurement unit
available.
Ö Measurement unit
available.

Keylock

not

The keylock function prevents inadvertent entries in the
Parameter mode and thus malfunctions.
Setting

Ö Keylock function disabled.
Ö Keylock function enabled.
is displayed when
the user attempts to edit a setting in the Parameter mode.

BG 805 196 BE / B

(2004-08)

TPG262.oi

61

RAM test

Test of the main memory.
Test sequence
The test runs automatically one
time:
Ö Test in process (very briefly).
Ö Test finished, no error found.
Ö Test finished, error(s) found.
The
lamp flashes.
If the error message persists
after several test sequences
have been run, please contact
your local Pfeiffer Vacuum
service center.

EPROM test

Test of the program memory.
Test sequence
The test runs automatically one
time:
Ö Test in process
Ö Test finished, no error found.
After the test, a four-digit
checksum (hexadecimal format) is displayed.
Ö Test finished, error(s) found.
After the test, a four-digit
checksum (hexadecimal format) is displayed. The
lamp flashes.
If the error message persists
after several test sequences
have been run, please contact
your local Pfeiffer Vacuum
service center.

62

BG 805 196 BE / B

(2004-08)

TPG262.oi

EEPROM test

Test of the parameter memory.
Test sequence
The test runs automatically one
time:
Ö Test in process (very briefly).
Ö Test finished, no error found.
Ö Test finished, error(s) found.
The
lamp flashes.
If the error message persists
after several test sequences
have been run, please contact
your local Pfeiffer Vacuum
service center.

Display test

Test of the display.
Test sequence
The test runs automatically one time*):
Ö First, all display elements are lit at the
same time, ...

Ö ... and then, each element is lit individually.
:

*)

BG 805 196 BE / B

(2004-08)

TPG262.oi

Ö Stop the test sequence
and activate one element
after another by pressing
the key once per element.

63

A/D converter test 0

Test of channel 0 of the analog/digital converter (with a
reference voltage at the signal input of the sensor connector (→  20)).
If the signal input is open, the TPG 262 displays a default value that may easily fluctuate
because of the high sensitivity of the open
measurement circuit.
Test sequence

Ö Measuring signal CH1 in Volt.
e.g.:

A/D converter test 1

Test of channel 1 of the analog/digital converter (with a
reference voltage at the signal input of the sensor connector (→  20)).
If the signal input is open, the TPG 262 displays a default value that may easily fluctuate
because of the high sensitivity of the open
measurement circuit.
Test sequence

Ö Measuring signal CH2 in Volt.
e.g.:

A/D converter test 2

Test of channel 2 of the analog/digital converter (with a
reference voltage at the identification input of the sensor
connector (→  20).
If the signal input is open, the TPG 262 displays a default value that may easily fluctuate
because of the high sensitivity of the open
measurement circuit.

64

BG 805 196 BE / B

(2004-08)

TPG262.oi

Test sequence

e.g.:

Ö Gauge identification voltage
CH1
Ö No gauge connected

A/D converter test 3

Test of channel 3 of the analog/digital converter (with a
reference voltage at the identification input of the sensor
connector (→  20)).
If the signal input is open, the TPG 262 displays a default value that may easily fluctuate
because of the high sensitivity of the open
measurement circuit.
Test sequence

e.g.:

Ö Gauge identification voltage
CH2
Ö No gauge connected

I/O test

Test of the relays of the TPG 262. The program tests
their switching function.
Caution

Caution: The relays switch irrespective of the
pressure
Starting a test program may cause unwanted
effects in connected control systems.
Disconnect all sensor cables and control
system lines to ensure that no control commands or messages are triggered by mistake.

BG 805 196 BE / B

(2004-08)

TPG262.oi

65

The relays switch on and off cyclically. The switching
operations are indicated optically and can be heard.
The contacts of switching functions 1 … 4 are connected
to the relay connector (→  22), the contacts of the error
relay to the control connector (→  21) on the rear of the
housing. Check their function with an ohmmeter.
Test sequence
The test runs automatically one
time:
Ö All relays deactivated
Ö Switching function relay 1

Ö Switching function relay 2

:
Ö Switching function relay 4

Ö Gauge relay CH1

Ö Gauge relay CH2

Ö Error relay

66

BG 805 196 BE / B

(2004-08)

TPG262.oi

RS232C test

Test of the RS232C interface. The TPG 262 repeats
each sign transmitted by the communicating HOST.
The data transferred from/to the TPG 262 can
be displayed by the computer only (→  68).
Test sequence
The test runs automatically.

BG 805 196 BE / B

(2004-08)

TPG262.oi

67

5 Communication (Serial Interface)
5.1 RS232C Interface

The serial interface is used for communication between
the TPG 26x 1) and a computer. A terminal can be connected for test purposes.
When the TPG 26x is put into operation, it starts transmitting measured values in intervals of 1 s. As soon as
the first character is transferred to the TPG 26x, the
automatic transmission of measured values stops. After
the necessary inquiries or parameter modifications have
been made, the transmission of measured values can be
started again with the COM command (→  75).

Connection diagram
connection cable

5.1.1 Data Transmission

Data format

Pin assignment of the 9-pole D-Sub connector and
RS232 interface cable →  23.

The data transmission is bi-directional, i.e. data and
control commands can be transmitted in either direction.

1 start bit
8 data bits
No parity bit
1 stop bit
No hardware handshake

1)

68

Communication structure and procedures are identical for both controllers TPG 261 and TPG 262.
Therefore the term TPG 26x is used in this chapter.

BG 805 196 BE / B

(2004-08)

TPG262.oi

Definitions

The following abbreviations and symbols are used:
Symbol
HOST
[...]
ASCII

Meaning
Computer or terminal
Optional elements
American Standard Code for Information
Interchange
Dec. Hex.
 END OF TEXT (CTRL C)
3
03
Reset the interface

CARRIAGE RETURN
13
0D
Go to beginning of line

LINE FEED
10
0A
Advance by one line

Request for data transmission
 ACKNOWLEDGE
6
06
Positive report signal
 NEGATIVE ACKNOWLEDGE 21
15
Negative report signal
"Transmit":
"Receive":

Flow Control

BG 805 196 BE / B

(2004-08)

TPG262.oi

Data transfer from HOST to TPG 26x
Data transfer from TPG 26x to HOST

After each ASCII string, the HOST must wait for a report
signal ( or  ).
The input buffer of the HOST must have a capacity of at
least 32 bytes.

69

5.1.2 Communication
Protocol
Transmission format

Messages are transmitted to the TPG 26x as ASCII
strings in the form of mnemonic operating codes and
parameters. All mnemonics comprise three ASCII characters.
Spaces are ignored.  (CTRL C) clears the input
buffer in the TPG 26x.

Transmission
protocol

HOST
TPG 26x
Mnemonics
[and parameters]
>
[]
>
<


Reception format

When requested with a mnemonic instruction, the
TPG 26x transmits the measurement data or parameters
as ASCII strings to the HOST.

Explanation
Receives message
with "end of message"
Positive acknowledgment of a received
message

 must be transmitted to request the transmission
of an ASCII string. Additional strings, according to the
last selected mnemonic, are read out by repetitive
transmission of .
If  is received without a valid request, the
ERROR word is transmitted.

70

BG 805 196 BE / B

(2004-08)

TPG262.oi

Reception protocol

HOST
TPG 26x
Mnemonics
[and parameters]
>
[]
>
<



<
<

<
<

Explanation
Receives message
with "end of message"
Positive acknowledgment of a received
message
> Requests to transmit
data
Measurement values Transmits data with
or parameters "end of message"

:
:
> Requests to transmit
data
Measurement values Transmits data with
or parameters "end of message"


Error processing

The strings received are verified in the TPG 26x. If an
error is detected, a negative acknowledgment  is
output.

Error recognition
protocol

HOST
Mnemonics
[and parameters]
[]

TPG 26x Explanation
Receives message
> with "end of mes> sage"

***** Transmission or programming error *****
<

 Negative acknowledgment of a received message
Mnemonics
Receives message
[and parameters]
> with "end of mes[]
> sage"
<
 Positive acknowledgment of a received message

BG 805 196 BE / B

(2004-08)

TPG262.oi

71

5.2 Mnemonics
→
ADC
BAU
COM
CAL
DCD
DGS
DIC
DIS
EEP
EPR
ERR
FIL
FSR
IOT
LOC
OFC
OFD
PNR
PR1
PR2
PRX
PUC
RAM
RES
RST
SAV
SC1
SC2
SCT
SEN
SP1
SP2
SP3
SP4
SPS
TID
TKB
TLC
UNI
WDT

72

A/D converter test
Baud rate (transmission rate)
Continuous mode
Calibration factor
Display control digits (display resolution)
Degas
Display control (display changeover)
Display test
EEPROM test
EPROM test
Error status
Filter time constant (measurement value filter)
Full scale range (measurement range of linear gauges)
I/O test
Keylock
Offset correction (linear gauges)
Offset display (linear gauges)
Program number (firmware version)
Pressure measurement (measurement data) gauge 1
Pressure measurement (measurement data) gauge 2
Pressure measurement (measurement data) gauge 1 and 2
Penning underrange control (underrange control)
RAM test
Reset
RS232 test
Save parameters to EEPROM
Sensor control 1 (gauge control 1)
Sensor control 2 (gauge control 2)
Sensor channel change (measurement channel change)
Sensors on/off
Setpoint 1 (switching function 1)
Setpoint 2 (switching function 2)
Setpoint 3 (switching function 3)
Setpoint 4 (switching function 4)
Setpoint status (switching function status)
Transmitter identification (gauge identification)
Keyboard test (operator key test)
Torr lock
Pressure unit
Watchdog control
BG 805 196 BE / B

89
85
75
81
85
83
86
88
88
88
77
80
81
90
87
82
82
86
73
73
74
83
88
78
91
86
84
84
77
76
79
79
79
79
80
76
91
87
85
87
(2004-08)

TPG262.oi

5.2.1 Measurement
Mode
Measurement data
gauge 1 or 2

Transmit:

PRx []
Measurement value x = 1 –> Gauge 1
2 –> Gauge 2

Receive:
Transmit:




Receive:

x,sx.xxxxEsxx 
1)

Measurement value
[in current pressure unit]
Status, x =
0 –> Measurement data okay
1 –> Underrange
2 –> Overrange
3 –> Sensor error
4 –> Sensor off (IKR, PKR, IMR, PBR)
5 –> No sensor
(output: 5,2.0000E-2 [mbar])
6 –> Identification error
1)

Values always in exponential format.
For logarithmic gauges, the 3rd and 4th decimal
are always 0.

BG 805 196 BE / B

(2004-08)

TPG262.oi

73

Measurement data
gauges 1 and 2

Transmit:

PRX []

Receive:
Transmit:




Receive:

x,sx.xxxxEsxx,y,sy.yyyyEsyy 
Measurement
1)
value gauge 2
[in current
pressure unit]
Status gauge 2
Measurement value gauge 1
[in current pressure unit]

1)

Status gauge 1, x =
0 –> Measurement data okay
1 –> Underrange
2 –> Overrange
3 –> Sensor error
4 –> Sensor off (IKR, PKR, IMR, PBR)
5 –> No sensor
(output: 5,2.0000E-2 [mbar])
6 –> Identification error
1)

Values always in exponential format.
For logarithmic gauges, the 3rd and 4th decimal
are always 0.

74

BG 805 196 BE / B

(2004-08)

TPG262.oi

Continuous output of
measurement values
(RS232)

Transmit:

COM [,x] []
Mode x = 0 –> 100 ms
1 –> 1 s (default)
2 –> 1 min.

Receive:


 is immediately followed by the continuous output of the measurement value in
the desired interval.

Receive:

x,sx.xxxxEsxx,y,sy.yyyyEsyy 
Measurement
1)
value gauge 2
[in current
pressure unit]
Status gauge 2
Measurement value gauge 1
[in current pressure unit]

1)

Status gauge 1, x =
0 –> Measurement data okay
1 –> Underrange
2 –> Overrange
3 –> Sensor error
4 –> Sensor off (IKR, PKR, IMR, PBR)
5 –> No sensor
(output: 5,2.0000E-2 [mbar])
6 –> Identification error
1)

Values always in exponential format.
For logarithmic gauges, the 3rd and 4th decimal
are always 0.

BG 805 196 BE / B

(2004-08)

TPG262.oi

75

Turning a gauge
on/off

Transmit:

SEN [,x,x] []
Gauge 2, x =
0 –> No status change
1 –> Turn gauge off
2 –> Turn gauge on
Gauge 1

Receive:
Transmit:




Receive:

x,x 
Status gauge 2, x =
0 –> Gauge cannot be turned on/off
1 –> Gauge turned off
2 –> Gauge turned on
Status gauge 1

Gauge identification

Transmit:

TID []

Receive:
Transmit:




Receive:

x,x 
Identification gauge 2, x =
TPR
(Pirani Gauge or
1)
Pirani Capacitive gauge )
-9
IKR9
(Cold Cathode Gauge 10 )
-11
IKR11 (Cold Cathode Gauge 10 )
PKR
(FullRange CC Gauge)
PBR
(FullRange BA Gauge)
IMR
(Pirani / High Pressure Gauge)
CMR
(Linear gauge)
noSEn (no SEnsor)
noid
(no identifier)
Identification gauge 1
1)

TPR and PCR have identical identifiers.
There is no distinction made in communication
and in data evaluation, since pressure ranges
of these gauges are approximately the same.

76

BG 805 196 BE / B

(2004-08)

TPG262.oi

Measurement
channel change

Transmit:

SCT [,x] []
Display channel, x =
0 –> Gauge 1
1 –> Gauge 2

Receive:
Transmit:




Receive:

x 
Display channel

Error status

Transmit:

ERR []

Receive:
Transmit:




Receive:

xxxx 
xxxx =
0000 –> No error
1000 –> Error

0100 –> NO HWR
0010 –> PAR
0001 –> SYN

Controller error
(See display
on front panel)
No hardware
Inadmissible
parameter
Syntax error

The ERROR word is cancelled when read out.
If the error persists, it is immediately set again.

BG 805 196 BE / B

(2004-08)

TPG262.oi

77

Reset

Transmit:

RES [,x] []
x = 1 –> Cancels currently active
error and returns to
measurement mode

Receive:
Transmit:




Receive:

[x]x,[x]x,... 
List of all present error messages,
xx =
0 –> No error
1 –> Watchdog has responded
2 –> Task fail error
3 –> EPROM error
4 –> RAM error
5 –> EEPROM error
6 –> DISPLAY error
7 –> A/D converter error
9 –> Gauge 1 error (e.g. filament
rupture, no supply)
10 –> Gauge 1 identification error
11 –> Gauge 2 error (e.g. filament
rupture, no supply)
12 –> Gauge 2 identification error

78

BG 805 196 BE / B

(2004-08)

TPG262.oi

5.2.2 Parameter Mode
5.2.2.1 Switching
Function
Parameters
Threshold value
setting, allocation

Transmit: SPx [,y,x.xxxxEsxx,x.xxxxEsxx] []
Upper threshold
[in current
pressure unit]
(default =
depending on
gauge)

1)

1)

Lower threshold
[in current pressure unit]
(default = depending on
gauge)
Switching function assignment, y =
0 –> Meas. channel 1
1 –> Meas. channel 2
1 –> Switching function 1
2 –> Switching function 2
3 –> Switching function 3
4 –> Switching function 4
1)

Values can be entered in any format. They
are internally converted into the floating point
format.

Receive:
Transmit:




Receive:

y,x.xxxxEsxx,x.xxxxEsxx 
Upper threshold
[in current
pressure unit]
Lower threshold
[in current pressure unit]
Switching function assignment

BG 805 196 BE / B

(2004-08)

TPG262.oi

79

Switching function
status

Transmit:

SPS []

Receive:
Transmit:




Receive:

x,x,x,x 
1)

Status switching function 4
1)

Status switching function 3
1)

Status switching function 2
1)

Status switching function 1
1)

x=

0 –> off
1 –> on

5.2.2.2 Gauge Parameters
Measurement value
filter

Transmit:

FIL [,x,x] []
Gauge 2 x = 0 –> fast
1 –> medium
(default)
2 –> slow
Gauge 1

Receive:
Transmit:




Receive:

x,x 
Filter time constant gauge 2
Filter time constant gauge 1

80

BG 805 196 BE / B

(2004-08)

TPG262.oi

Calibration factor

Transmit:

CAL [,x.xxx,x.xxx] [] (

)

Gauge 2
log. 0.100 ... 9.990
(default = 1.000)
lin. 0.500 ... 2.000
(default = 1.000)
Gauge 1

Receive:
Transmit:




Receive:

x.xxx,x.xxx 
Calibration factor gauge 2
Calibration factor gauge 1

Measurement range
(F.S.) of
linear gauges

The full scale value of the measurement range
(Full Scale) of linear gauges has to be defined
by the user; the full scale value of logarithmic
gauges is automatically recognized.
Transmit:

FSR [,x,x] []
Gauge 2, x =
0 –> 0.01 mbar
1 –> 0.1 mbar
2 –> 1 mbar
3 –> 10 mbar
4 –> 100 mbar
5 –> 1000 mbar (default)
6 –> 2 bar
7 –> 5 bar
8 –> 10 bar
9 –> 50 bar
Gauge 1

Receive:
Transmit:




Receive:

x,x 
Measurement range gauge 2
Measurement range gauge 1

BG 805 196 BE / B

(2004-08)

TPG262.oi

81

Offset correction
(linear gauges)

Transmit:

OFC [,x,x] [] (

)

Gauge 2, x =
0 –> off (default)
1 –> on
2 –> auto (offset measurement)
Gauge 1

Receive:
Transmit:




Receive:

x,x 
Gauge 2
Gauge 1

Offset display
(linear gauges)

Transmit: OFD [,sx.xxxxEsxx,sx.xxxxEsxx] []
1)

Gauge 2 Offset
[in current
pressure unit]
(default = 0.0000)
Gauge 1
1)

Values can be entered in any format. They
areinternally converted into the floating point
format.

Receive:
Transmit:




Receive:

sx.xxxxEsxx,sx.xxxxEsxx 
Gauge 2
Gauge 1

82

BG 805 196 BE / B

(2004-08)

TPG262.oi

Underrange control

Transmit:

PUC [,x,x] []
Gauge 2, x = 0 –> off (default)
1 –> on
Gauge 1

Receive:
Transmit:




Receive:

x,x 
Gauge 2
Gauge 1

Degas

Transmit:

DGS [,x,x] [] (

.)

Gauge 2, x =
0 –> Degas off (default)
1 –> Degas on (3 min.)
Gauge 1

Receive:
Transmit:




Receive:

x,x 
Degas status gauge 2
Degas status gauge 1

BG 805 196 BE / B

(2004-08)

TPG262.oi

83

5.2.2.3 Gauge Control
Gauge control

Transmit:

SCx [,x,y,x.xxEsxx,y.yyEsyy] []
OFF threshold
ON threshold
Controlling source for gauge
deactivation, x =
0 –> no control
1 –> automatic deactivation
2 –> manual deactivation
(default)
3 –> external deactivation
4 –> self control
Controlling source for gauge
activation, x =
0 –> no control
1 –> automatic activation
2 –> manual activation (default)
3 –> external activation
4 –> hot start
Controlled gauge, x =
1 –> Gauge 1
2 –> Gauge 2

Receive:
Transmit:




Receive:

x,y,x.xxEsxx,y.yyEsyy 
OFF threshold
ON threshold
Controlling source for
deactivating the gauge
Controlling source for activating the
gauge

84

BG 805 196 BE / B

(2004-08)

TPG262.oi

5.2.2.4 General
Parameters
Pressure unit

Transmit:

UNI [,x] []
Pressure unit, x =
0 –> mbar/bar (default)
1 –> Torr
2 –> Pascal

Receive:
Transmit:




Receive:

x 
Pressure unit

Transmission rate

Transmit:

BAU [,x] []
Transmission rate, x =
0 –> 9600 baud (default)
1 –> 19200 baud
2 –> 38400 baud

As soon as the new baud rate has been entered, the report signal is transmitted at the new
transmission rate.
Receive:
Transmit:




Receive:

x 
Transmission rate

Display resolution

Transmit:

DCD [,x] []
Resolution, x =
2 –> Display x.x (2 digits)
(default)
3 –> Display x.xx (3 digits)

Receive:
Transmit:




Receive:

x 
Resolution

BG 805 196 BE / B

(2004-08)

TPG262.oi

85

Save parameters to
EEPROM

Transmit:

SAV [,x] []
x = 0 –> Save default
parameters
1 –> Save user parameters

Display changeover

Receive:



Transmit:

DIC [,x] []
Measurement display behavior
when a Pirani gauge or a Pirani
Capacitance gauge is combined
with a linear gauge with
1000 mbar F.S., x =
0 –>manual (default)
1 –>automatic

Receive:
Transmit:




Receive:

x 
Measurement display behavior

5.2.2.5 Test Parameters
Firmware version

(For service personnel)
Transmit:

PNR []

Receive:
Transmit:




Receive:

302-510-x 
-x = Modification index
(-- = original version)
Firmware number

86

BG 805 196 BE / B

(2004-08)

TPG262.oi

Watchdog control

Transmit:

WDT [,x] []
x = 0 –> Manual error
acknowledgement
1 –> Automatic error
1)
acknowledgement
(default)

1)

If the watchdog has responded, the error is
automatically acknowledged and cancelled after 2 s.

Receive:
Transmit:




Receive:

x 
Watchdog control

Torr lock

Transmit:

TLC [,x] []
x = 0 –> off (default)
1 –> on

Receive:
Transmit:




Receive:

x 
Torr lock status

Keylock

Transmit:

LOC [,x] []
x = 0 –> off (default)
1 –> on

Receive:
Transmit:




Receive:

x 
Keylock status

BG 805 196 BE / B

(2004-08)

TPG262.oi

87

RAM test

Transmit:

RAM []

Receive:
Transmit:


 Starts the test (duration <1 s)

Receive:

xxxx 
ERROR word

EPROM test

Transmit:

EPR []

Receive:
Transmit:


 Starts the test (duration ≈5 s)

Receive:

xxxx,yyyy 
Check sum (hex)
ERROR word

EEPROM test

Transmit:

EEP []

Receive:
Transmit:


 Starts the test (duration <1 s)

Do not keep repeating the test (EEPROM life).
Receive:

xxxx 
ERROR word

Display test

Transmit:

DIS [,x] []
x = 0 –> Stops the test – display
according to current
operating mode (default)
1 –> Starts the test –
all LEDs on

Receive:
Transmit:




Receive:

x 
Display test status

88

BG 805 196 BE / B

(2004-08)

TPG262.oi

ADC test

Transmit:

ADC []

Receive:
Transmit:




Receive:

[x]x.xxxx,[x]x.xxxx,x.xxxx,x.xxxx 
ADC channel 4
identification
Gauge 2
[0.0000 ...
5.0000 V]
ADC channel 3
Gauge 1 identification
[0.0000 ... 5.0000 V]
ADC channel 2
Measurement signal
gauge 2
[0.0000 ... 11.0000 V]
ADC channel 1
Measurement signal gauge 1
[0.0000 ... 11.0000 V]

BG 805 196 BE / B

(2004-08)

TPG262.oi

89

I/O test
Caution

Caution: The relays switch irrespective of the
pressure.
Starting a test program may cause unwanted
effects in connected control systems.
Disconnect all sensor cables and control
system lines to ensure that no control commands or messages are triggered by mistake.

Transmit:

IOT [,x,yy] []
Relay status (in hex format), yy =
00 –> All relays deactivated
01 –> Switching function relay 1
activated
02 –> Switching function relay 2
activated
04 –> Switching function relay 3
activated
08 –> Switching function relay 4
activated
10 –> Gauge relay CH1 activated
20 –> Gauge relay CH2 activated
40 –> Error relay activated
7F –> All relays activated
x=

0 –> Test stopped
1 –> Test runs

Receive:
Transmit:




Receive:

x,yy 
Relay status
I/O test status

90

BG 805 196 BE / B

(2004-08)

TPG262.oi

Operator key test

Transmit:

TKB []

Receive:
Transmit:




Receive:

xxxx 
Key 4

x = 0 –> Not pushed
1 –> Pushed

Key 3
Key 2
Key 1

RS232 test

Transmit:

RST []

Receive:
Transmit:


Starts the test (repeats each

character, test is interrupted
with  C)

BG 805 196 BE / B

(2004-08)

TPG262.oi

91

5.2.3 Example
"Transmit (T)" and "Receive (R)" are related to Host.
S:
E:
S:
E:

TID  []
  

TPR,CMR  

Request for gauge identification
Positive acknowledgement
Request for data transmission
Gauge identifications

S:
E:
S:
E:

SEN  []
  

0,0  

Request for gauge statuses
Positive acknowledgement
Request for data transmission
Gauge statuses

S: SP1  []
E:   
S: 
E: 0,1.0000E-09,9.0000E-07  
S: SP1,1,6.80E-3,9.80E-3  []
E:   
S: FOL,1,2  []
E:
S:
E:
S:
E:
S:
E:

92

  

0001  
FIL,1,2  []
  

1,2  

Request for parameters of
switching function 1 (setpoint 1)
Positive acknowledgement
Request for data transmission
Thresholds
Modification of parameters of
switching function 1 (setpoint 1)
Positive acknowledgement
Modification of filter time constant
(syntax error)
Negative acknowledgement
Request for data transmission
ERROR word
Modification of filter time constant
Positive acknowledgement
Request for data transmission
Filter time constants

BG 805 196 BE / B

(2004-08)

TPG262.oi

6 Maintenance
The product requires no maintenance.

Cleaning the TPG 262

For cleaning the outside of the TPG 262, a slightly moist
cloth will usually do. Do not use any aggressive or scouring cleaning agents.

DANGER

Caution: mains voltage
Contact with live parts is extremely hazardous
when liquids penetrate into the unit.
Make sure no liquids penetrate into the equipment.

BG 805 196 BE / B

(2004-08)

TPG262.oi

93

7 Troubleshooting
Signalization of errors
and the error relay opens (→  21).

Error messages

Possible cause and remedy/
acknowledgement
Interruption or instability in sensor line
or connector (Sensor error).
Ö Acknowledge with the . key
If the problem persists,
is displayed.

or

Possible cause and remedy/
acknowledgement
The TPG 262 has been turned on too
fast after power off.
Ö Acknowledge with the
key.
If the watchdog is set to
the TPG 262 acknowledges the
message automatically after 2 s
(→  61).

,

The watchdog has tripped because of
a severe electric disturbance or an operating system error.
Ö Acknowledge with the
key.
If the watchdog is set to
the TPG 262 acknowledges the
message automatically after 2 s
(→  61).

,

Possible cause and remedy/
acknowledgement
Main memory (RAM) error.
Ö Acknowledge with the

key.

Possible cause and remedy/
acknowledgement
Program memory (EPROM) error.
Ö Acknowledge with the

94

BG 805 196 BE / B

key.

(2004-08)

TPG262.oi

Possible cause and remedy/
acknowledgement
Parameter memory (EEPROM) error.
Ö Acknowledge with the

key.

Possible cause and remedy/
acknowledgement
Display driver error.
Ö Acknowledge with the

key.

Possible cause and remedy/
acknowledgement
A/D converter error.
Ö Acknowledge with the

key.

Possible cause and remedy/
acknowledgement
Operating system (Task Fail) error.
Ö Acknowledge with the

key.

Technical support
If the problem persists after the message has been
acknowledged for several times and/or the gauge
has been exchanged, please contact you local
Pfeiffer Vacuum service center.

8 Repair
Return defective products to your nearest
Pfeiffer Vacuum service center for repair.
Pfeiffer Vacuum assumes no liability and the warranty
becomes null and void if repair work is carried out by the
end-user or third parties.

BG 805 196 BE / B

(2004-08)

TPG262.oi

95

9 Storage

Caution

Caution: electronic component
Inappropriate storage (static electricity, humidity etc.) can damage electronic components.
Store the product in an antistatic bag or container. Observe the corresponding specifications in the technical data (→  9).

10 Disposal
WARNING
N

Separating the
components

96

Caution: substances detrimental to the environment
Products or parts thereof (mechanical and
electric components, operating fluids etc.)
can be detrimental to the environment.
Dispose of such substances in accordance
with the relevant local regulations.

After disassembling the product, separate its components according to the following criteria:

Non-electronic
components

Such components must be separated according to their
materials and recycled.

Electronic
components

Such components must be separated according to their
materials and recycled.

BG 805 196 BE / B

(2004-08)

TPG262.oi

Appendix
A:

Conversion Tables
Weights
kg

kg
1

lb
2.205

slug
68.522×10-3

oz
35.274

lb
slug

0.454

1

31.081×10-3

16

14.594

32.174

1

oz

514.785

28.349×10-3

62.5×10-3

1.943×10-3

1

bar
10×10-6

mbar
10×10-3

Torr
7.5×10-3

at
9.869×10-6

1

103

750.062

0.987

1

750.062×10-3

0.987×10-3

1.333

1

1.316×10-3

Pressures
N/m2, Pa

N/m2, Pa
1

bar

100×10

mbar

100

10

Torr

133.322

1.333×10

at

101.325×103

1.013

1.013×103

760

1

psi
14.504×10-3

3

-3
-3

Pressure units used in
the vacuum technology
mbar

mbar
1

Pascal
100

Torr
750.062×10 -3

mmWs
10.2

Pascal

10×10-3

1

7.5×10-3

0.102

0.145×10-3

Torr

1.333

133.322

1

13.595

19.337×10-3

mmWs

9.81×10-2

9.81

7.356×10-2

1

1.422×10-3

psi

68.948

6.895×103

51.715

703

1

mm
1
103
25.4
304.8

m
10-3
1
25.4×10-3
0.305

inch
39.37×10-3
39.37
1
12

ft
3.281×10-3
3.281
8.333×10-2
1

Kelvin

Kelvin
1

Celsius
°C+273.15

Fahrenheit
(°F+459.67)×5/9

Celsius

K-273.15

1

5/9×°F-17.778

Fahrenheit

9/5×K-459.67

9/5×(°C+17.778)

1

Linear measures
mm
m
inch
ft

Temperature

BG 805 196 BE / B

(2004-08)

TPG262.oi

97

B:

Default Parameter
Settings

The following values are activated when the default parameter settings are loaded (→  57):
Default
-11

1×10

User
mbar

9×10-11 mbar
normal
1.00 (log)
1.000 (lin)
1000 mbar
off
0×10-2 mbar
off
mbar
9600
2 Digit
Hand
Auto
off
off

98

BG 805 196 BE / B

(2004-08)

TPG262.oi

C:

Firmware Update

If your TPG 262 firmware needs updating, e.g. for
implementing a new gauge type, please contact
your local Pfeiffer Vacuum service center.

User parameters

Most of the settings you may have defined in the Parameter and Test mode will not be affected by a firmware update. To be sure, note your parameter settings
before upgrading the firmware (→  98).

Preparing the TPG 262
for a program transfer

n
o
p

Turn the TPG 262 off.
Connect the TPG 262 with the serial COM1
(COM2) interface of your PC via a 9-pole D-Sub
extension cable (→  23) (the firmware of the
TPG 262 cannot be loaded from a Mac).
With a pin (ø<2 mm) depress the switch on the top
of the unit, under the housing, and turn the
TPG 262 on.

After power on, the display remains dark.

BG 805 196 BE / B

(2004-08)

TPG262.oi

99

Program transfer

In the following instructions, the index
of the actual index.

n

o

is used instead

Unpack the self extracting file
.

If you have not connected the TPG 262 to the
COM1 interface:
• Open the batch file

…

• … edit the interface …

• … and save the new setting.

p

Start batch file

.

Ö The new firmware is transmitted to the TPG 262.

100

BG 805 196 BE / B

(2004-08)

TPG262.oi

Starting the TPG 262
with the updated firmware

If the program transfer was successful, quit the Update
mode by turning the TPG 262 off.

Wait at least 10 s before turning the TPG 262
on again in order for it to correctly initialize itself.
The TPG 262 is now ready for operation. To be
sure, check that the current parameter settings are
identical with the previously defined settings
(→  98).

BG 805 196 BE / B

(2004-08)

TPG262.oi

101

D:

102

Literature

[1]

www.pfeiffer-vacuum.de
Instruction Sheet
Compact Pirani Gauge TPR 261
BG 805 105 BE
Pfeiffer Vacuum GmbH, D–35614 Asslar,
Deutschland

[2]

www.pfeiffer-vacuum.de
Operating Instructions
Compact Pirani Gauge TPR 265
BG 805 177 BE
Pfeiffer Vacuum GmbH, D–35614 Asslar,
Deutschland

[3]

www.pfeiffer-vacuum.de
Operating Instructions
Pirani-Messröhre TPR 280
BG 805 178 BE
Pfeiffer Vacuum GmbH, D–35614 Asslar,
Deutschland

[4]

www.pfeiffer-vacuum.de
Operating Instructions
Pirani-Messröhre TPR 281
BG 5179 BE
Pfeiffer Vacuum GmbH, D–35614 Asslar,
Deutschland

[5]

www.pfeiffer-vacuum.de
Operating Instructions
Compact Pirani Capacitance Gauge PCR 260
BG 805 180 BE
Pfeiffer Vacuum GmbH, D–35614 Asslar,
Deutschland

[6]

www.pfeiffer-vacuum.de
Instruction Sheet
Compact Cold Cathode Gauge IKR 251
BG 805 110 BN
Pfeiffer Vacuum GmbH, D–35614 Asslar,
Deutschland

[7]

www.pfeiffer-vacuum.de
Instruction Sheet
Compact Cold Cathode Gauge IKR 261
BG 805 113 BN
Pfeiffer Vacuum GmbH, D–35614 Asslar,
Deutschland

BG 805 196 BE / B

(2004-08)

TPG262.oi

[8]

www.pfeiffer-vacuum.de
Instruction Sheet
Compact Cold Cathode Gauge IKR 270
BG 805 115 BE / A
Pfeiffer Vacuum GmbH, D–35614 Asslar,
Deutschland

[9]

www.pfeiffer-vacuum.de
Instruction Sheet
Compact FullRange™ Gauge PKR 251
BG 805 119 BN
Pfeiffer Vacuum GmbH, D–35614 Asslar,
Deutschland

[10] www.pfeiffer-vacuum.de
Instruction Sheet
Compact FullRange™ Gauge PKR 261
BG 805 122 BN
Pfeiffer Vacuum GmbH, D–35614 Asslar,
Deutschland
[11] www.pfeiffer-vacuum.de
Instruction Sheet
Compact Process Ion Gauge IMR 265
BG 805 132 BE
Pfeiffer Vacuum GmbH, D–35614 Asslar,
Deutschland
[12] www.pfeiffer-vacuum.de
Instruction Sheet
Compact FullRange™ BA Gauge PBR 260
BG 805 131 BE
Pfeiffer Vacuum GmbH, D–35614 Asslar,
Deutschland
[13] www.pfeiffer-vacuum.de
Instruction Sheet
Compact Capacitance Gauge
CMR 261 … CMR275
BG 805 133 BE
Pfeiffer Vacuum GmbH, D–35614 Asslar,
Deutschland
[14] www.pfeiffer-vacuum.de
Instruction Sheet
Compact Piezo Gauge APR 250 … APR 267
BG 805 127 BN
Pfeiffer Vacuum GmbH, D–35614 Asslar,
Deutschland

BG 805 196 BE / B

(2004-08)

TPG262.oi

103

E:

Index

–A–
A/D converter test

–G–
64

–B–
Baud rate

56

–C–
Calibration factor
Cleaning
Communication
interface
Connectors
control
mains power
relay
RS232
sensor
Contact positions
control
relay
Contents
control connector
Conversion

41
93
68
68
21
19
22
23
20
21
22
4
21
97

–D–
Default values
Degas
Display
changeover
resolution
test
Disposal

104

I/O test
Identification of gauges
Intended use
Interface
connector
test

–K–

57
56
63
96

–L–

Keylock

65
30
3
23
67

61

Liability
Literature

8
102

–M–
63
62
94
92

–F–
Factory settings
Filter
Firmware
update
version

–I–

57; 98
46

–E–
EEPROM test
EPROM test
Error messages
Example

Gauge connector
20
Gauge control
12; 47
gauge activation
49
gauge deactivation
51
OFF threshold
52
ON threshold
50
Gauge identification
30
Gauge parameters
38
calibration factor
41
measurement value filter 39
General parameters
54
default values
57
display changeover
57
display resolution
56
measurement unit
56
transmission rate
56

57; 98
39
99
2; 60

Mains power connector
19
Maintenance
93
Measurement mode
27; 73
gauge identification
30
selecting a measurement
channel
28
status messages
28
turning a gauge on and off 28
Measurement range
42
Measurement units
conversion
97
selection
56
Measurement value filter
39
Mnemonics
72

BG 805 196 BE / B

(2004-08)

TPG262.oi

–O–
OFF threshold
Offset
ON threshold
Operating modes
Measurement mode
overview
Parameter mode
Program transfer mode
Operation
power off
power on

–S–
52
43
50
27
26
31
99
25
25

–P–
Parameter mode
79
gauge control
47
gauge parameters
38
general parameters
54
measurement value filter 39
switching function
parameters
33
switching thresholds
35; 36
test parameters
58
Parameters
Degas
46
measurement range
42
Offset
43
underrange control
44
Pin assignment
control
21
relay
22
RS232
23
sensor
20
Power connector
19
Power off
25
Power on
25
Pressure units
conversion
97
selection
56
Program
update
99
version
2; 60

–R–
RAM test
relay connector
Repair
RS232 connector
RS232C test

BG 805 196 BE / B

(2004-08)

TPG262.oi

62
22
95
23
67

Safety
Scope of delivery
sensor connector
Serial interface
connector
test
Status messages
Storage
Switching function
parameters
Switching functions
thresholds
Symbols

6
3
20
23
67
28
96
33
11; 22
35; 36
6

–T–
Test parameters
Test program
A/D converter test
display test
EEPROM test
EPROM test
firmware version
I/O test
keylock
RAM test
RS232C test
Torr lock
watchdog
Thresholds
OFF
ON
Torr lock
Transmission rate
Troubleshooting

58
58
64
63
63
62
60
65
61
62
67
61
61
35; 36
52
50
61
56
94

–U–
Underrange control
Units
conversion
selection
Update

44
97
56
99

–V–
Validity

2

–W–
Warranty
Watchdog

8
61

105

Original: German BG 805 196 BD / B (2004-08)

bg805196be/ b

Berliner Strasse 43
D–35614 Asslar
Deutschland
Tel +49 (0) 6441 802-0
Fax +49 (0) 6441 802-202
info@pfeiffer-vacuum.de
www.pfeiffer-vacuum.de



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : Yes
Encryption                      : Standard V4.4 (128-bit)
User Access                     : Print, Copy, Extract, Print high-res
Create Date                     : 2010:11:10 14:56:57-07:00
Keywords                        : Pfeiffer, Dual Gauge, TPG 262, Dual Channel Control Unit, Compact Gauges
Modify Date                     : 2012:01:19 07:41:16-07:00
Has XFA                         : No
XMP Toolkit                     : Adobe XMP Core 4.2.1-c043 52.372728, 2009/01/18-15:08:04
Producer                        : Acrobat Distiller 6.0.1 (Windows)
Metadata Date                   : 2012:01:19 07:41:16-07:00
Document ID                     : uuid:1f359092-c052-4658-bf06-e1e3ad7a2dfb
Instance ID                     : uuid:2d94853d-295a-4a57-acec-df14d4ea7818
Format                          : application/pdf
Title                           : Pfeiffer, Dual Gauge TPG 262 Dual Channel Control Unit, for Compact Gauges
Description                     : Pfeiffer, Dual Gauge TPG 262 Dual Channel Control Unit, for Compact Gauges
Creator                         : 
Subject                         : Pfeiffer, Dual Gauge, TPG 262, Dual Channel Control Unit, Compact Gauges
Page Count                      : 106
EXIF Metadata provided by EXIF.tools

Navigation menu