227A/227C MANp 10120.01 BARTON S 227A And 227C DP Indicators IOM

User Manual: BARTON s 227A and 227C DP Indicators IOM Resource Library

Open the PDF directly: View PDF PDF.
Page Count: 16

Download227A/227C MANp 10120.01 BARTON S 227A And 227C DP Indicators IOM Barton-dp-indicators-227a-227c-iom
Open PDF In BrowserView PDF
MODELS 227A/227C
DIFFERENTIAL PRESSURE INDICATOR
(Standard and Inverted Configurations)
227A uses 224 DPU
227C uses 224C DPU
Installation Manual
Part No. 10120, Rev. 01
July 2007
[This manual is for indicator only - see separate 224/224C DPU manual, per model
number listed above.]
CONTENTS
Prepages................................................................................
Safety Statement..............................................................
Product Warranty Statement..............................................
Record of Changes...........................................................
Section 1 - Introduction...........................................................
1-1. General....................................................................
1-2. Product Description...................................................
1-3. Theory of Operation..................................................
1-4. Specifications............................................................
Section 2 - Installation.............................................................
2-1. General....................................................................
2-2. Mounting..................................................................
2-3. DPU Installation........................................................
2-4. Zero Check/Adjustment.............................................
Section 3 - Maintenance and Calibration...................................
3-1. DPU Installation/Maintenance....................................
3-2. Tools Required for Maintenance..................................
3-3. Calibration Check......................................................
3-4. Pointer Installation and Removal.................................
3-5. Complete Calibration.................................................
3-6. Overrange Stop Adjustment........................................
3-7. Locking Drive Arm to Torque Tube.............................
3-8. Drive Arm Tightness Test...........................................
Section 4 - Parts Drawing/List.................................................
Figure 4-1. 227A/227C Parts Drawing..............................
Table 4-1. 227A/227C Parts List......................................
Section 5 - Install/Dimensional Drawings..................................
Figure 4-1. 227A/227C Parts Drawing..............................

Page 2
Page 2
Page 2
Page 2
Page 3
Page 3
Page 3
Page 3
Page 4
Page 5
Page 5
Page 5
Page 5
Page 6
Page 7
Page 7
Page 7
Page 7
Page 7/8
Page 8
Page 10
Page 10
Page 11
Page 12
Page 12
Page 13
Page 14
Page 14/15

SAFETY
Before installing this instrument, become familiar with the installation instructions
in Section 2 and in the separate 224/224C DPU manual.
DANGER notes indicate the presence of a hazard which will cause severe
personal injury, death, or substantial property damage if warning is ignored.
WARNING notes indicate the presence of a hazard which can cause severe
personal injury, death, or substantial property damage if warning is ignored.
CAUTION notes indicate the presence of a hazard which will or can cause
moderate personal injury or property damage if warning is ignored.
DANGER, WARNING, and/or CAUTION notes that appear on the following
pages of this manual should be reviewed before proceeding: None. (Important!
Before installing or operating this instrument, review all safety notices
contained in the separate 224/224C DPU manual, per models listed

PRODUCT WARRANTY STATEMENT
The product warranty applicable to this instrument is stated on the back cover of this
manual.

DPU "C" VERSION DESIGN CHANGE
The 227C utilizes a 224C DPU, which is a redesigned version of the 224 DPU. The "C"
version is identical in function, performance, installation, and operation to the previous
version  - redesign was for improved manufacturing only. This design change does not
affect the instrument being actuated. Selected configurations of 227A
(using a 224 [Non-C] DPU) are for Nuclear/Government only.

RECORD OF CHANGES
DATE

DESCRIPTION

00E2

5/00

Reduced Booklet 227C INV. w/224C APX  (99H4)

00K25c

11/00

Booklet Format 227C w/224 DPU information in body

03D77d

4/03

Combined 200/227A/227C Manual

03G76d

7/03

Combined 227A/227C (Standard and Inverted) Versions
(Replaces 227A (10121), 227C (10130) Inverted, and 227
portion of 200/227A/227C (10055) Manuals);
Rev. Co. Name/Logo to PRIME Measurement Products;.
Added para. 2-4, Zero Check/Adjustment procedure;.
Rev. Section 3 - Maintenance and Calibration;.
Rev. Section 4 - Parts Drawing and Parts List
Rev. Section 5 - Install/Dimensional Drawings

04L55a

12/04

Rev. warranty - made specific to M227A/227C units

01

7/07

Revised corporate name/logo/contact information to reflect
Cameron ownership.



SECTION 1 - INTRODUCTION
1-1. General
The Model 227A/227C Differential Pressure Indicator is designed to measure
liquid flow or liquid level. In flow measurement applications, the instrument is
connected to the low and high pressure sides of a primary device (orifice plate,
a venturi tube, or a flow tube) located in the process run. In liquid level measurement applications, the instrument is connected to monitor changes in differential
pressure relative to variations of liquid height in a vessel.
1-2. Product Description
The indicator has a 6 inch dial. The indicating pointer traverses a 270° arc
(nominal) to measure differential pressure, flow, or liquid level. The indicator
movement has a micrometer screw for range adjustment. Zero is adjusted by
"slipping" pointer on hub, which can be done without removing the scaleplate
or pointer. The Range can also be adjusted without removing the scaleplate or
pointer. Linearity adjustments are accessible after removal of the scale plate.
The 227A is actuated by a Model 224 Differential Pressure Unit (DPU). The
227C is actuated by a Model 224C DPU. Refer to the separate DPU Manuals.
1-3. Theory of Operation
The 227A/227C consists of the actuating DPU and the indicator (case assembly).
A. Differential Pressure Unit (DPU)
The 224/224C DPU consists of a dual bellows assembly (one high pressure and
one low pressure bellows), which is enclosed by a set of two pressure housings
(pressure heads).
The dual bellows assembly consists of two internally-connected bellows, a center
plate, overrange valves, a temperature compensator, a torque tube assembly,
and range springs. The bellows and center plate are completely filled with a noncorrosive liquid and then sealed. The pressure housings, located on either side of
the center plate, are connected by pipe or tubing to the system primary device.
Variations in differential pressure within the pressure housing cause the bellows to expand/contract laterally, in the direction of lower pressure. The lateral
movement of the bellows is converted to angular rotation as transmitted to the
torque tube shaft by the drive arm. The torque tube assembly provides mechanical output in response to the bellows movement and actuates the mechanism of
the process monitoring instruments. For more information on the DPU, refer to
the separate 224/224C DPU manual.
B. Indicator
The indicator is attached directly to the actuating unit (DPU). The indicator
consists of a drive arm assembly, linkage, and a movement assembly to which
the indicating pointer is attached. Changes in differential pressure are sensed
by the DPU bellows and transmitted as rotary motion through the torque tube
to the drive arm assembly of the indicator. The linkage and movement assembly
respond to the action of the drive arm assembly to position the pointer in direct
relationship to the amount of bellows travel.


1-4. Specifications
Accuracy (DP Range):
0-30" w.c. to 0-50" w.c.
(0-75 mbar to 0-124 mbar).................
0-51" w.c. to 0-60 psi
(0-127 mbar to 0-4.1 bar)...................
0-61 psi to 0-150 psi
(0-4.2 bar to 0-10.3 bar)....................
0-151 psi to 0-400 psi
(0-10.4 bar to 0-27.6 bar)..................
0-401 psi to 0-600 psi
(0-27.7 bar to 0-41.3 bar)..................
0-601 psi to 0-1,000 psi
(0-41.4 bar to 0-69 bar).....................
Dial Size...............................................
Temperature Limits...............................
Indicator Case.......................................
Lens.....................................................
O-rings (Wetted)....................................
Overall Dimensions...............................
Weight.................................................

±0.75% of full scale
±0.5% of full scale
±0.75% of full scale
±1.0% of full scale
±1.5% of full scale
±4.0% of full scale
6" Diameter
-40 °F/°C to +180°F (+82°C).
(+40°F minimum for water filled
units)
Aluminum Alloy, Enamel or Epoxy
Finish or 316 SST
Polycarbonate (shatterproof glass for
SST Version)
Buna-N, Viton, or EPT
per configuration and 224/224C
DPU housing rating. See ODD
drawings in Section 5.
6 to 9.5 lbs (approx.) Weight
depends on DPU configuration/
materials.

Differential Pressure Unit (DPU):
For DPU specifications, refer to the separate 224/224C DPU manual.



SECTION 2 - INSTALLATION
2-1. General
The instrument should be inspected at time of unpacking to detect any damage
that may have occurred during shipment. Note: The DPU was checked for
accuracy at the factory — do not change any of the settings during examination or accuracy will be affected.
For applications requiring special cleaning/precautions, a polyethylene
bag is used to protect the instrument from contamination. This bag
should be removed only under conditions of extreme cleanliness.
2-2. Mounting
Per ODD drawings in Section 5.
2-3. DPU Installation
Refer to the separate 224/224C DPU manual, per models listed on front page,
for DPU installation and maintenance.



2-4. Zero Check/Adjustment
It may be necessary to reset the ZERO before placing the 227A/227C into
service. This procedure can be accomplished without removing the scaleplate.

NOTICE

Adjusting the instrument's ZERO in accordance with the following procedure does not invalidate the original factory calibration or warranty.
Range Adjust
Zero Adjust

Pointer

Scale Plate

Zero Line

Max. Line

Figure 2-2. Instrument Face
Pointer Stop
FRONT VIEW

1.

Fix the instrument in a stable position, with the top of the DPU level with
the horizontal plane.

2.

Using a flat-blade screwdriver, remove the three (3) or (9 on SST case) bezel
screws (see Figure 2-2).

3.

Lift the screws, bezel, lens, and lens gasket away from the case.

4.

Set the pointer to ZERO position, by "slipping" the pointer on the hub, per
the following (refer to Figure 2-3):
With a 1/4" open-end wrench (included in calibration kit 0288-1032B),
hold the hexagon pointer hub fixed
and rotate the pointer with fingers
until the pointer accurately indicates
ZERO on the scale.

5.

Reassemble the unit, attaching the lens
gasket, lens, and bezel to the case.
Secure the bezel with the three (3) (or
9 on SST case) bezel screws. Tighten
the screws until they are snug.

Pointer "slipped"
rotated on hub

HUB

Hub held in place
with 1/4" wrench

Figure 2-3. "Slipping" Pointer



SECTION 3 - MAINTENANCE AND CALIBRATION
3-1. DPU Installation/Maintenance/Repair
DPU inspection, cleaning, service, repair, range change, and BUA replacement
procedures (along with applicable WARNINGS, CAUTIONS, and NOTICES,
are presented in the separate 224/224C DPU manual.
3-2. Tools Required For Maintenance
Table 3-1. Tools
Equipment

Purpose

Pointer Puller

Pointer Removal

Small Screwdriver

Scaleplate Removal and Replacement

Medium Screwdriver

Bezel Removal and Replacement

1/4" & 1/8" Open-end Wrenches

Zero (1/4") /Range (1/8") Adjustment

For DPU specific tools, refer to separate DPU Manual.

3-3. Calibration Check
To ensure the unit calibration is within factory-set calibration tolerances, perform the following procedure. NOTICE: Review all procedures, WARNINGS/NOTICES in the separate 224/224C DPU manual BEFORE
performing this procedure.
1.

Mount instrument and connect to a standard pressure source (see separate
224/224C manual).

2.

If the zero indication is incorrect, remove bezel/lens assembly and re-adjust
zero, per Zero Check/Adjustment procedure, per paragraph 2-4.

3.

Apply 0, 50, and 100% of full scale pressure. If indication is within specified limits, no adjustments are necessary. If indication is not within specified
limits, perform a complete calibration, per paragraph 3-5.

3-4. Pointer Installation and Removal
During adjustment and calibration of the unit, it may be necessary to remove
and reinstall the pointer, per the following procedures:
A.	 Pointer Installation
1.

Position pointer on movement shaft with pointer set at zero scale.
It may be necessary to enlarge the hub hole, using a tapered broach
(included in the toolkit (p/n 0288-1032B).

2.

Lightly tap pointer hub with a hand-set or other flat-end tool. Use
perpendicular blows to avoid bending shaft.

3.

Check indicating switch for calibration over its entire range (refer
to Switch Calibration in this section). If indicating switch is correctly
calibrated, secure pointer to movement shaft by tapping hub with a
hand-set or other flat-end tool.

(Continued on next page...)


3-4. Pointer Installation and Removal (Continued)
B. Pointer Removal
Pointer is removed with Pointer Puller (p/n 0163-0005B), which is included in the toolkit (p/n 0288-1032B), see Figure 3-1.
1.

2.

Slide pointer puller
along pointer until
pin protruding from
tip of screw in pointer
puller is directly over
movement shaft and
arms of pointer puller
are directly under
pointer.

POINTER PULLER
(0163.0005B)

HUB
POINTER

HUB
Gently turn knurled
MOVEMENT SHAFT
head of screw
Figure 3-1. Pointer Puller
clockwise, pushing pin against movement shaft and lifting pointer with arms. Finger
pressure should be sufficient to pull the pointer free. If more pressure
is required, an Allen wrench (inserted into head of the screw) can be
used. However, care should be exercised: too much pressure can cause
the pin to break.

3-5. Complete Calibration
(DP=Differential Pressure)
A complete calibration of instrument is required whenever the DPU assembly is
replaced. NOTICE: Review all procedures, WARNINGS/NOTICES in
the separate 224/224C DPU manual BEFORE performing this procedure.
1. 	 Securely mount instrument in an approximately level position and connect
DPU into the test setup, as described in the appropriate (separate) DPU
manual.
2.

Attach linkage between drive arm and movement. See Figure 3-2 (standard
227A/C) or Figure 3-3 (Inverted 227A/C) for linkage alignment at 50%
DP. Inspect parts for straightness and pivot-fit without binding.

Figure 3-2. (Standard) Range/Linearity Adj. (50% DP)


3-5. Complete Calibration (Continued)

(DP=Differential

Figure 3-3. (Inverted) Range/Linearity Adj. (50% DP)
Pressure)
3. Set pointer at zero on scale by slipping pointer on hub. Hold tip of pointer
and turn hub with wrench (refer to paragraph 2-4).
4.

Apply 100% DP. If pointer exceeds 100% on scale, lengthen range arm.
Remove pressure.

5.

Set zero and span, using hub for zero adjustment and range adjust screw on
the movement for span adjustments.

6. 	 Apply 50% DP. If pointer does not indicate 50% scale, a linearity adjustment is necessary. Loosen drive arm screw and move arm to shift pointer
in direction of error (approx. 10:1). Tighten drive arm screw.
7. 	 Release pressure and reset pointer at zero. Check the span. If gear in movement reaches limit of travel as a result of linearity adjustment (step 6), slip
range arm along gear approximately 5 degrees from normal 37.5 degree
angle to approximately 43 degrees. Range arm is slipped by applying pressure to range arm with thumb, while holding gear firmly in place. Retest
pointer response at 50% and 100% DP, and adjust linkage until readings
are acceptable.
8.

Apply 0%, 25%, 50%, 75%, 100%, 75%, 50%, 25%, and 0% of DP
consecutively to instrument without overshoot. Lightly tap indicator to overcome friction. Pointer should accurately indicate each applied pressure.

10. Test instrument repeatability by applying 0%, 50%, 0%, 50% DP. Indicator
should accurately indicate each applied pressure.
(Continued on next page...)


3-5. Complete Calibration (Continued) (DP=Differential Pressure)
11. Set overrange stops to prevent pointer from striking snubbers on scale (see
paragraph 3-6). Tighten all screws. Test setting by manually moving pointer
from zero position to 50%, then let the pointer return freely. An off-set
in zero reading indicates pointer slippage. If necessary, tap pointer hub to
tighten it to shaft.
12. Reassemble unit, making sure bezel screws are snug.
3-6. Overrange Stop Adjustment
1. Apply sufficient pressure to the high pressure housing to deflect the pointer
against the snubber on the scale plate. Slide the upper overrange stop
against the drive arm and tighten the overrange stop screw.
2.

Apply sufficient pressure to the low pressure housing to deflect the pointer
against the zero stop snubber on the scale plate. Slide the zero-stop against
the drive arm and tighten the zero-stop screw.

3.

Remove the pointer and calibration scale. Replace the pointer at zero (adjust zero as necessary). Replace the lens and bezel assembly.

3-7. Locking Drive Arm to Torque Tube

Figure 3-4. Locking Drive Arm to Torque Tube
1.

2.

Slip the drive arm over the torque tube shaft; clear the end of the torquetube housing by approximately 0.030-inches before securing to prevent
interference.
To tighten the drive arm assembly onto the torque-tube shaft:
a. While supporting the block and shaft, tighten the clamp screw until
snug to the shaft.
b. Still supporting block/shaft, tighten clamp screw an additional:
• Sintered: 1/3 to 1/2 turn (This screw can normally turn one full
revolution before breaking.)
• Slotted: 1/4 to 1/3 turn (The slot in the slotted clamp block should
still be open.)

NOTICE: For Seismic and High Shock Qualified Units, perform
Drive Arm Tightness Test, per paragraph 3-8.
10

3-8. Drive Arm Tightness Test
This procedure tests the drive arm to torque tube attachment for tightness, by
applying torque developed by the DPU onto a fixed drive arm. Care should be
taken to apply pressure slowly, as torque is being applied to the connection
through the torque tube drive shaft and not the torque tube itself.
With pointer at normal 0% torque tube rotation position (max. minimum scale
position or 0% on a normal 0 to 100% scale unit), adjust drive arm stop bracket
(or use alternate means) to prevent pointer from moving (stop bracket interferes
with drive arm movement). Note: On reverse acting/split range units, it will be
necessary to pressurize DPU to move pointer to max. minimum scale position,
and on suppressed units, it will be necessary to apply pressure to establish a
reference point to check for "zero" shift.
Pressurize DPU to full calibrated scale DP (100% of full scale range) to achieve
8-degrees of torque tube drive shaft equivalent torque onto the connection.
Observe shift in the unit "zero" following DPU depressurization (as required) and
drive arm stop bracket readjusting (to allow free movement of drive arm and
pointer). A downscale (counter-clockwise) shift in "zero" of greater than 1/2% is
indicative of drive arm slippage necessitating further clamp block tightening.

11

12

11

2

21

13

3

14

10

6

4

7

12

5

A

17 18

9

19 15

19 15 20

8

1/4

1
(Inverted
Configuration)

1
(Standard
Configuration)

PED LABEL - on DPU Housing
(PED complient units only)

SECTION 4 - PARTS DRAWING/LIST

Figure 4-1. 227A/227C Parts Drawing

Table 4-1. 227A/227C Parts List
Item
1

Description

Part No.

Per
Unit
1

Case:
Standard
0227-0014C
Inverted
0227-1028C
2
Bezel (Part Of Item #21)
0277-0029C
1
3* Lens, Cover  (Part Of Item #21)
0181-0038C
1
4* Gasket  (Part Of Item #21)
0277-0026C
1
5** Scale Plate
1
0227-0013C
White On Black Background
Black On White Background
0227-1003C
6* Pointer Assembly:
1
White
0288-0030B
Black
0288-0031B
7
Movement Assembly
0288-0035B
1
8
Drive Arm Assembly
0226-0023B
1
9
Link Assembly
0226-0020B
1
10 Stop Bracket
0288-0028C
1
11* Bezel Screws, SST
0181-0007C
3
12 Snubber
0226-0028C
2
13 Screw, Sl Fil. Hd. 4-40 X 3/16, SST
0114-0023J
4
14* Screw, Bd. Hd., 4-40 X 3/16, SST
0117-0012J
1
15* Washer, Split Lock, #4, SST
0003-0062K
2
16 Screw, Flat Hd., #10 X 1/2
0240-0019J
6
17 Screw, Slotted Set, 1/4-20 X 1
0340-0003J
4
18 Nut, 1/4-20, Hex, STL
0500-0010J
4
19 Screw, Rd. Hd., 4-40 X 1/4, SST
0111-0034J
2
20 Washer, Flat, #4, SST
0003-0096K
1
21 Bezel Assembly (Items 2, 3, & 4)
0277-0018B
1
22 Differential Pressure Unit (DPU).
SEE DPU
1
(Not Shown)
MANUAL
227A Uses 224 DPU
227C Uses 224C DPU
Inverted Units Use Inverted 224/224C DPUs
23 Calibration Kit (Not Shown)
0288-1032B
1
24 Plug, Restrictor Vent (Not Shown)
0080-1022T
1
Notes: * Indicates recommended spare part; ** SCALE PLATE INDENTIFICATION:
If the scale plate shows an SCR number, this will identify it. Otherwise, provide the
following information:
1. Square Root or Linear Graduations
2. Scale (e.g., 0-100, 25-0-100, etc.)
3. Number of Graduations (linear scales only)
4. Data (e.g., PSI (bar), inches of water column (meter), etc.)
5. Standard plates have black background
When ordering parts, specify serial number of instrument. Minimum parts order is
$100.00. For 224/224C parts, refer to separate 224/224C DPU manual.

13

SECTION 5 - INSTALL/DIMENSIONAL DRAWINGS
Note:  For other DPU dimensional information, refer to the separate DPU manual (as
listed on the front page).

Figure 5-1. 227A/227C Dimensions (Based on 0227C-10013, Rev.00)
14

Figure 5-1. 227A/227C Dimensions (Continued)
15

Product Warranty
A.

Warranty
Cameron International Corporation ("Cameron") warrants that at the time of shipment, the
products manufactured by Cameron and sold hereunder will be free from defects in material and
workmanship, and will conform to the specifications furnished by or approved by Cameron.

B.

Warranty Adjustment
(1)

If any defect within this warranty appears, Buyer shall notify Cameron immediately.

(2)

Cameron agrees to repair or furnish a replacement for, but not install, any product which
within one (1) year from the date of shipment by Cameron shall, upon test and examination
by Cameron, prove defective within the above warranty.

(3)

No product will be accepted for return or replacement without the written authorization of
Cameron. Upon such authorization, and in accordance with instructions by Cameron, the
product will be returned shipping charges prepaid by Buyer. Replacements made under this
warranty will be shipped prepaid.

C. Exclusions from Warranty
(1)

THE FOREGOING WARRANTY IS IN LIEU OF AND EXCLUDES ALL OTHER EXPRESSED
OR IMPLIED WARRANTIES OF MERCHANTABILITY, OR FITNESS FOR .
A PARTICULAR PURPOSE, OR OTHERWISE.

(2)

Components manufactured by any supplier other than Cameron shall bear only the warranty
made by the manufacturer of that product, and Cameron assumes no responsibility for the performance or reliability of the unit as a whole.

(3)

"In no event shall Cameron be liable for indirect, incidental, or consequential damages nor shall
the liability of Cameron arising in connection with any products sold hereunder (whether such
liability arises from a claim based on contract, warranty, tort, or otherwise) exceed the actual
amount paid by Buyer to Cameron for the products delivered hereunder."

(4)

The warranty does not extend to any product manufactured by Cameron which has been
subjected to misuse, neglect, accident, improper installation or to use in violation of instructions
furnished by Cameron.

(5)

The warranty does not extend to or apply to any unit which has been repaired or altered at any
place other than at Cameron's factory or service locations by persons not expressly approved by
Cameron.

Product Brand
Barton® is a registered trademark of Cameron International Corporation ("Cameron").

MEASUREMENT SYSTEMS
Formerly: NuFlo Measurement Systems • Barton Instrument Systems • Caldon, Inc.
HOUSTON
HEAD OFFICE
ASIA
PACIFIC

281.582.9500

603.2287.1039
ms-asiapacific@c-a-m.com

NORTH
AMERICA
EUROPE,
MIDDLE EAST
& AFRICA

1.800.654.3760
ms-us@c-a-m.com

44.1243.826741
ms-uk@c-a-m.com

USA • CANADA • UK • SCOTLAND • CHINA • UAE
ALGERIA • MALAYSIA • SINGAPORE • www.c-a-m.com/flo



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : Yes
Encryption                      : Standard V4.4 (128-bit)
User Access                     : Print, Copy, Extract, Print high-res
Page Mode                       : UseOutlines
XMP Toolkit                     : 3.1-702
About                           : uuid:9661e21a-4df2-11d9-8b97-000a958d622e
Create Date                     : 2003:08:28 09:01:44Z
Modify Date                     : 2007:07:25 13:50:06-05:00
Metadata Date                   : 2007:07:25 13:50:06-05:00
Creator Tool                    : Adobe InDesign 2.0.2
Version ID                      : 1
Rendition Class                 : proof:pdf
Document ID                     : adobe:docid:indd:93e08a67-dac8-11d7-895a-9e4c662392c2
Instance ID                     : uuid:62e7548e-b1fd-4b46-98f4-0827e1f6d8f2
Rendition Of Document ID        : adobe:docid:indd:93e08a67-dac8-11d7-895a-9e4c662392c2
Rendition Of Version ID         : 1
Rendition Of Rendition Class    : default
Format                          : application/pdf
Title                           : 227A/227C MANp 10120.01
Description                     : 227A/227C DP Indicator Manual
Creator                         : .
Subject                         : 10120, 227, 227A, 227C, DP, Differential Pressure, Indicator, 224, 224C, DPU
Has XFA                         : No
Page Count                      : 16
Keywords                        : "10120, 227, 227A, 227C, DP, Differential Pressure, Indicator, 224, 224C, DPU"
EXIF Metadata provided by EXIF.tools

Navigation menu