Elpas Solutions 5-STC00433 Temperature snesor tag User Manual

Elpas Solutions Ltd. Temperature snesor tag

USER MANUAL

Download: Elpas Solutions 5-STC00433 Temperature snesor tag User Manual
Mirror Download [FCC.gov]Elpas Solutions 5-STC00433 Temperature snesor tag User Manual
Document ID2009510
Application IDbp0NfozAJX9DV3PlZxbUCw==
Document DescriptionUSER MANUAL
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize29.72kB (371498 bits)
Date Submitted2013-07-05 00:00:00
Date Available2013-07-05 00:00:00
Creation Date2013-06-09 10:58:50
Producing SoftwareMicrosoft® Word 2010
Document Lastmod2013-06-09 14:01:25
Document TitleUSER MANUAL
Document CreatorMicrosoft® Word 2010
Document Author: barryi

Elpas Temperature Sensors
User Guide
P/Ns: 5-STE00433-1 and 5-STC00433-1
Introduction
Elpas Temperature Sensors are supervised Active RFID Tags that provide wireless remote
monitoring of environmental sensitive assets or environments that require strict temperature range
compliance.
The sensors report location based temperature monitoring data which is received and relayed over
the Elpas infrastructure to a host RTLS application or an integrated 3rd party platform for regulatory
compliance, temperature-range alerting or trend analysis.
Internal Thermistor Configuration
Sensors are available with an internal thermistor for ambient temperature monitoring or with an
external probe for applications requiring higher precision. Both sensors feature wireless call-button
functionality to report preconfigured maintenance messages plus remote device supervision to
ensure field reliability.
Elpas Temperature Sensors are water/dust resistant, cleanable with alcohol based disinfectants
and are powered by one field-replaceable lithium battery.
It is important that you understand, and follow the instructions in this document. Should you have
any questions, please contact your local ELPAS support representative before proceeding with the
installation, operation or maintenance of these products.
External Probe Configuration
Product Specifications
Internal Thermistor (P/N: 5-STE00433-1)
* External Probe (P/N: 5-STC00433-1)
Signaling Technology
RF (433MHz)
RF (433MHz)
Transmission
Supervision transmission duration - 4ms
Supervision transmission period - 127sec
Supervision transmission duration - 4ms
Supervision transmission period - 127sec
Sensor Range
-10°C to 70°C (14°F to 158°F)
-40°C to 80°C (-40°F to 176°F)
Sensor Accuracy
+/- 2.0°C (-10°C to 70°C)
+/- 0.5°C (0°C to 70°C)
+/- 1.0°C (-40°C to 0°C) & (70°C to 80°C)
Sensor Resolution
+/- 0.1°C
+/- 0.1°C
Sensor Cable Length
N/A
2.5m (8.0 FT)
Supervised Events
Button Press/Lo w Battery
6 RF transmissions (each transmission 4ms in
duration), total burst duration - 927.5ms
6 RF transmissions (each transmission 4ms in duration),
total burst duration - 927.5ms
Power Source
3V/270mAh lithium battery, Type CR 2430
3V/600mAh lithium battery, Type CR 2450
Average Batter y Life
Three years
Three years
Status Indicators
Red LED: Button Press, Low Battery
Green LED: Activation from sleep-mode
Red LED: Button Press, Low Battery
Green LED: Activation from sleep-mode
Badge ID
Unique, factory-programmed ID
Unique, factory-programmed ID
Construction
PA Plastic (Nylon)
PA Plastic (Nylon)
IP Rating
Sensor: IP64
Sensor: IP64.
Dimensions (H x W x D)
34 x 40 x 15 mm (1.3 x 1.6 x 0.6 inches)
34 x 40 x 15mm (1.3 x 1.6 x 0.6 inches)
Weight
20 grams (0.75 ounces)
20 grams (0.75 ounces)
Remote Management
Eiris Software V4.7.2 (or higher)
Eiris Software V4.7.2 (or higher)
Standards
IC, FCC & CE compliant
IC, FCC & CE compliant
Warranty
1 year limited warranty (excluding battery)
1 year limited warranty (excluding battery)
External Probe: IP67
*Note: P/N: 5-STC00433-1 includes external probe - temperature sensor, one glycol bottle, one tag mounting clip and one probe mounting clip.
Product offerings and specifications are subject to change without notice.
Not all products include all features
Compatible Accessories
Supported Button Press Events
Part Number
Description
5-BC012430
3V/270mAh Lithium Battery, CR2430 (25pcs)
5-PBA90003
3V/600mAh Lithium Battery, CR2450 (25pcs)
5-ETC09001
Mounting Clip (5 units)
5-STC90001
Glycol Bottle and Mounting Kit
Temperature Sensors support Button Press events (designated B1) to report
maintenance messages.
 Button Press: 3 messages are transmitted
(each @ 4ms in duration), 463.75ms apart. B1=1
 Button Release: 3 messages are transmitted
(each @ 4ms in duration), 463.75ms apart. B1=0
 Steady State (no button press/button release) B1=0
www.elpas.com
Page 1 of 4
V5/June 2013
Elpas Temperature Sensors – User Guide
Initial Activation
Battery Replacement
Temperature sensors come from the factory in Sleep Mode to conserve
Change the sensor’s status in the host RTLS application to
battery power during shipping.
‘Inactive’.
Place the sensor back cover side up. Unscrew the 4 screws to
remove the back cover
To Activate the sensor from sleep mode, remove the Activation Label.
Then press the sensor’s call button
for at least 7 seconds.
Slide the battery out from under the battery holder. Dispose of
used battery in accordance with local regulations.
If the activation process is successful the sensor’s
LED indicator will glow green for about 3 seconds.
The sensor is now ready to for enrollment in the host RTLS application.
Before inserting a new battery into place, position the battery
Isolator such that its holes align with the battery contact pins.
Glycol Bottle Usage Tips
When inserting the probe into the bottle, push the probe fully through the
screw cap so that a leak-proof seal is formed and then ensure that the
cap is properly screwed onto the bottle.
When removing the probe from the bottle, first unscrew the cap and only
then PUSH the probe out from the cap.
Replace the battery ensuring that the positive (+) side of the
Caution DO NOT pull the
battery faces up.
probe from the cap as this may
Close the back cover such that the screw holes are correctly
cause damage to the probe or
aligned. Tighten the 4 screws snugly into place. Do not over-
to the interface cable.
tighten as this may strip the case threads.
Change the sensor’s status in the host RTLS application back to
‘Active’.
Installation & Usage Considerations
1.
Temperature sensors are not compatible with Elpas1 Lon Readers.
2.
The transmission range of the temperature sensor may vary when mounted in different types of refrigeration units;
therefore a facility site survey to examine specific sensor mounting environments is required.
3.
Always use an Elpas Mounting Clip (P/N: 5-ETC09001) to mount the temperature sensor so that it can easily be removed
for service or replacement,
4.
Supervisory Alerts should be configured in the host RTLS application to notify staff members that the temperature sensor
is: 1) being tampered with, 2) is in a low voltage condition or 3) is not properly transmitting temperature data. Additionally,
the operational status of these system alerts should be triggered once a week to ensure system integrity.
www.elpas.com
Page 2 of 4
V5/June 2013
Elpas Temperature Sensors – User Guide
Sensor Mounting Guidelines - Refrigerator Applications
The Elpas Internal Thermistor - Temperature Sensor (P/N: 5-STE00433-1) is
suitable
for
general
purpose
refrigerator
applications
(such
as
food
and
pharmaceutical storage) where ambient temperature monitoring is sufficient.
Use an Elpas Mounting Clip (P/N: 5-ETC09001) with an ultra-high bound adhesive
to mount the temperature sensor on the side wall as close to the middle of the
refrigerator as possible. Avoid mounting the temperature sensor near or in the direct
air path of cold air intake since there will may be wide temperature swings at this
location. If the refrigerator is relatively empty consider adding bottles of water or other items to help stabilize the temperature
of the refrigerator.
The Elpas External Probe - Temperature Sensor (P/N: STC00433-1) is for refrigerator applications (such
as blood, vaccines and pathology storage) where it is necessary to immerse the probe in propylene glycol to
better match the thermal properties of the items being stored. Mount the probe and glycol bottle on the side
or back corner of the refrigerator where it will not be disturbed, while placing the sensor’s electronics either
inside or outside the refrigerator. Use tape or tie-wraps to secure the probe cable to the side wall or shelf.
Sensor Mounting - Freezer Applications
The Elpas External Probe - Temperature Sensor (P/N: STC00433-1) is also for
freezer applications below -20°C/4°F where placement is limited by the operating
limits of the sensor’s electronics and battery. The sensor’s remote probe enables
temperature measurements inside the freezer while its electronics and battery is
mounted externally on the outside of the freezer. The probe can be attached to a
shelf or other bracket using tie-wraps. Avoid the cold air intake and try not to mount
it on the top of the freezer as this area can become warm during defrost cycles, In
relatively empty freezers add bottles of frozen water or ice packs to help stabilize
the temperature of the freezer.
Cleaning Procedures
Use an appropriate antibacterial disinfectant such as Dispatch® Hospital Cleaner Disinfectant with Bleach from Caltech
Industries (http://www.caltechind.com) to clean the Elpas Temperature Sensors.
Since ‘Cleaning Procedures’ may vary according to facility guidelines, thus the procedures given below are for illustrative
purposes only:
Option 1 – Using Dispatch Disinfectant Spray
Option 2 – Using Dispatch Disinfectant Towels
1.
Lightly wet a disposable towel with Dispatch spray
1.
Open a new Dispatch pre-moistened towel
2.
Do not saturate the towel
2.
Wipe the outer surfaces of the sensor
3.
Wipe the outer surfaces of the sensor
3.
Next wipe the sensor with a dry disposable
4.
Next wipe the sensor with a dry disposable towel
5.
Allow the sensor to air dry
4.
Allow the sensor to air dry
6.
Return the clean sensor to inventory or usage
5.
Return the clean sensor to inventory or usage
7.
Dispose of used towels per facility policies
6.
Dispose of used towels per facility policies
www.elpas.com
towel
Page 3 of 4
V5/June 2013
Elpas Temperature Sensors – User Guide
Compliance Standards
T h e se d e vi ce s ( F C C # : 0 4 X 5 - RL E0 0 1 2 5 a n d I C # : 1 4 6 7 G - R L E 0 0 1 2 5 ) co mp l i e s
However, there is no guarantee that interference will not occur in a particular
wi t h F C C Ru l e s Pa r t 1 5 a n d wi t h i n d u st r y C a n a d a l i ce n se e xe mp t R SS .
installation. If this equipment does cause harmful interference to radio or television
O p e r a t i o n i s s u b je ct t o t h e f o l l o wi n g t wo co n d i t i o n s:
reception, which can be determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of the following measures:
( 1 ) T h i s d e vi ce ma y n o t ca u se h a r mf u l i n t e r f e r e n ce , a n d
( 2 ) T h i s d e vi ce mu st a cc e p t a n y i n t e r f e r e n ce r e ce i ve d , o r t h a t ma y ca u se
u n d e si r e d o p e r a t i o n .
F o r i n d u st r y C a n a d a : L e p r é se n t a p p a r e i l e st co n f o r me a u x C NR d ' I n d u st r i e


Increase the separation between the equipment and receiver.

Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
Ca n a d a a p p l i ca b l e s a u x a p p a r e i l s r a d i o e xe mp t s d e l i ce n ce . L ' e xp l o i t a t i o n
e st a u t o r i s é e a u x d e u x c o n d i t i o n s su i va n t e s : ( 1 ) l ' a p p a r e i l n e d o i t p a s
p r o d u i r e d e b r o u i l l a g e , e t ( 2 ) l ' u t i l i sa t e u r d e l ' a p p a r e i l d o i t a cce p t e r t o u t
b r o u i l l a g e r a d i o é l e ct r i q u e su b i , mê me si l e b r o u i l l a g e e st su s ce p t i b l e d ' e n
co mp r o me t t r e l e f o n ct i o n n e me n t .
Reorient or relocate the receiving antenna.

Consult the dealer or an experienced radio/TV technician for help.
Changes or modifications to this equipment not expressly approved by the party
responsible for compliance (Elpas Solutions Ltd.) could void the user’s authority to
operate the equipment.
No t e T h i s e q u i p me n t h a s b e e n t e st e d a n d f o u n d t o co mp l y wi t h t h e l i mi t s f o r
a Cl a ss B d i g i t a l d e vi ce , p u r su a n t t o p a r t 1 5 o f t h e F CC Ru l e s. T h e se l i mi t s
a r e d e si g n e d t o p r o vi d e r e a so n a b l e p r o t e c t i o n a g a i n s t h a r mf u l i n t e r f a c e i n a
r e si d e n t i a l i n st a l l a t i o n . T h i s e q u i p me n t g e n e r a t e s; u se s a n d ca n r a d i a t e r a d i o
Company Contact
Elpas, Inc.
Westford, Massachusetts (USA) - Tel: 1-800-223-0020
f r e q u e n cy e n e r g y a n d , i f n o t i n st a l l e d a n d u se d i n a cco r d a n c e wi t h t h e
EU Declaration of Conformity
i n st r u c t i o n s, ma y ca u se h a r mf u l i n t e r f e r e n c e t o r a d i o co m mu n i ca t i o n s.
This equipment is in compliance with the essential requirements and other relevant
provisions of Directive 1999/5/EC.
Product Warranty
E l p a s S o l u t i o n s L t d . ( E L P A S o r t h e C o m p a n y ) , a n d i t s a f f i l i a t e s , w a r r a n t s i t s p r o d u c t s ( h e r e i n a f t e r r e f e r r e d t o a s " t h e P r o d u c t ” ) t o b e f r e e o f d ef e c t s i n m a t e r i a l s a n d
workmansh ip under normal operat ing conditions and use f or a pe riod of one year f rom the date of ship ment by ELPAS. The Company’s ob ligat ions shall be limited within
the warranty pe riod, at its option, to re pair o r to rep lace the def ective Product or any def ective co mponent or part the reof . To exercise th is warranty, the p roduct must be
returned to the manuf acture r f reight p re paid and insu red.
This warranty does not app ly to repairs or rep lacement caused by imp roper installat ion, P roduct misu se, f ailu re to f ollow insta llation o r operat ing instructions, alt erat ion,
abuse, accident, tampe ring, repair by anyone other than ELPAS, exte rnal ca uses, and f ailure to pe rf orm required p reventive ma intenance. Th is warranty also does not
apply to an y p roducts, acce ssories, or attachments used in con junct ion with the Product, in clud ing batteries, wh ich sh all be cove red so lely by their o wn warra nties, if any.
ELPAS shall not be liable f or any da mage or loss what soever, whethe r d ire ctly, indirect ly, incidentally, consequent ia lly o r othe rwise, resu lt ing f rom a malf unction of the
Product due to p roducts, accessories, o r attach ments of others, includ ing batte rie s, used in conjun ction with the Produ ct.
E L P A S MA K E S N O E X P R E S S W A R R A N T IE S E X C E P T T H O S E S T A T E D IN T H IS S T A T E ME N T . E L P A S D IS C L A I M S A L L O T H E R W A R R A N T IE S , E X P R E S S O R I M P L IE D ,
I N C L U D I N G W IT H O U T L I M I T A T IO N I M P L IE D W A R R A N T IE S O F ME R C H A N T A B IL I T Y A N D F IT N E S S F O R A P A R T IC U L A R P U R P O S E . E L P A S ’ S S O L E R E S P O N S IB IL IT Y
F O R W A R R A N T Y C L A I M S IS L I M I T E D T O R E P A IR O R T O R E P L A C E A S S E T F O R T H IN T H I S S T A T E ME N T .
ELPAS shall have no liability f or an y d eath, persona l inju ry, prope rty da mage, or othe r loss whethe r d ire ct, indirect , incidenta l, consequentia l, o r othe rwise, b ased on a
claim that the Product f ailed to f unctio n. Howeve r, if ELPAS is held liab le, whether direct ly o r indirect ly, f or any lo ss or damage a rising under this limite d warranty o r
o t h e r w i s e , r e g a r d l e s s o f c a u s e o r o r i g i n , E L P A S ' s m a x i m u m l i a b i l i t y s h a l l b e l i m i t e d t o t h e p u r c h a s e p r i c e o f t h e P r o du c t , w h i c h s h a l l b e f i x e d a s l i q u i d a t e d d a m a g e s a n d
not as a penalty, and shall be the co mp lete and exclusive liab ility of ELPAS.
E L P A S s h a l l n o t , u n d e r a n y c i r c u m s t a n c e s w h a t s o e v e r , b e l i a b l e f o r a n y i n a c c u r a c y , e r r o r o f j u d g m e n t , d e f a u l t , o r n eg l i g e n c e o f E L P A S , i t s e m p l o y e e s , o f f i c e r s , a g e n t s ,
o r a n y o t h e r p a r t y , o r o f t h e p u r c h a s e r o r u s e r , a r i s i n g f r o m a n y a s s i s t a n c e o r c o m m u n i c a t i o n o f a n y k i n d r e g a r d i n g t h e c o n f i g u r a t i o n , d e s i g n , i n s t a l l a t i o n , o r c r e a t i o n of
security system in volving the Product, that being the responsibilit y of the p urchase r o r use r. f ELPAS is unable t o make such repair or rep lacement , E LPAS’s ent ire
liab ility sha ll be limited to the cost of a reasonable sub stitute product. ELPAS shall not be responsib le f or any dismantling, in stallat ion, re insta llat ion, purchasing,
shipping, insu rance, o r any simila r cha rges.
ELPAS shall have no liabilit y f or any damages, includin g without limitation, a ny dire ct, ind ire ct, in cidenta l, special, or consequentia l damages, expense s, costs, prof its,
lost sa vings or ea rning s, or other da mages arising out of the use of the Prod uct or the re moval, in stallation, re insta llat ion , repair or rep lacement of the Product or any
r e l a t e d e v e n t s . In t h e e v e n t t h a t t h e r e i s a n y l i a b i l i t y a g a i n s t E L P A S , s u c h l i a b i l i t y s h a l l b e l i m i t e d t o t h e p u r c h a s e p r i c e o f t h e P r o d u c t w h i c h a m o u n t s h a l l b e f i x e d as
liqu idated damages.
The purchase r and use r understan d th at this P roduct may be comp romised or circumvented by intentional acts; th at the Product will not in all ca ses p revent de ath,
personal inju ry, p roperty damage, o r ot her loss re sult ing f rom bu rgla ry, robbery, f ire o r other causes; and that the P roduct will not in a ll cases p rovide adequate warn ing
or protection. The pu rchase r and user also understand that a prope rly installe d and maintained ala rm may reduce th e risk of e vents such as burgla ry, robb ery, and f ire
without warn ing, but it is not in surance or a guarantee that such events will n ot occur o r that there will be no death, personal inju ry, p ropert y damage, or ot her loss a s a
result of such events.
By purcha sing the P roduct , the purcha ser and use r shall def end, indemnif y a nd hold ELPAS, its of f icers, directo rs, aff iliates, subsidia ries, agents, se rvant s, emp loyees,
and authorized rep resentat ives ha rmle ss f rom and against any and all cla ims, suits, costs, da mages, and judgments incurred, c laimed, or su stained wheth er f or death,
personal inju ry, p ropert y damage, o r ot herwise, because of or in any wa y re la ted to the conf iguration, design, installation, or creat ion of a securit y syste m involving the
Product, and the use, sale, d ist ribution, and installation of the Product, including payment of any and all attorney’s f ees, costs, and expenses incu rred as a result of any
such events.
The purchase r o r use r should f ollo w th e Product insta llat ion and ope ration instruct ions and te st the P roduct and the entire s ystem at least once each week. For various
reasons, in clud ing but not limited to changes in environmenta l condit ions, ele ctric, elect ronic, or electro magnetic d isruptions, and ta mpering, the Product may not perf orm
as expected. The pu rchase r and use r a re advised to ta ke all necessary pre cautions f or the prot ect ion and saf ety of persons and prope rty.
This statement pro vides ce rta in legal rights. Other rights may va ry by state or country. Under ce rtain circu mstances, some st ates or count ries may not allow exclusion or
limitation of incidental o r consequent ial d amages o r imp lied warrantie s, so the above exclu sions may not apply under those circumstances and in tho se states o r
countrie s.
ELPAS rese rves the right to mod if y this statement at any t ime, in its so le d iscretion without not ice to any purchase r or user. H owe ver, this sta tement shall not be modif ied
or
va ried
except
by
ELPAS
in
writing,
and
ELPAS
does
Any questions about this state ment sho uld be directed to ELPAS.
Page 4 of 4
V5/June 2013
not
authorize
any
sing le
individual
to
act
on
3/0
W.E.E.E. Product Recycling Declaration
For information regarding the recycling of this product you must contact the company from which you orignially purchased it.
If you are discarding this product and not returning it for repair then you must ensure that it is returned as identified by your supplier.
This product is not to be thrown away with everyday waste - Directive 2002/96/EC Waste Electrical and Electronic Equipment.
its
behalf
to
mod if y
or
va ry
this
statement.

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : Yes
Author                          : barryi
Create Date                     : 2013:06:09 10:58:50+03:00
Modify Date                     : 2013:06:09 14:01:25+03:00
Language                        : en-US
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 5.2-c001 63.139439, 2010/09/27-13:37:26
Format                          : application/pdf
Creator                         : barryi
Creator Tool                    : Microsoft® Word 2010
Metadata Date                   : 2013:06:09 14:01:25+03:00
Producer                        : Microsoft® Word 2010
Document ID                     : uuid:b78ea703-9d99-433c-a778-9f305ec08929
Instance ID                     : uuid:da5009a3-9e04-43b6-85e7-47e4d18b6237
Page Count                      : 4
EXIF Metadata provided by EXIF.tools
FCC ID Filing: O4X5-STC00433