Honeywell Analytics SPMFLEX RFID Chemcassette Recognition User Manual Manual 2
Honeywell Analytics Inc RFID Chemcassette Recognition Manual 2
Contents
- 1. Manual 1
- 2. Manual 2
Manual 2
SPM Flex Gas Detector Backward-compatible mounting bracket Figure 20 shows the detector’s backward-compatible mounting bracket (used when replacing an SPM detector with an SPM Flex detector). 13.0 in. (32.9 cm) 5.4 in. (13.7 cm) 13.0 in. (33.0 cm) 0.4 in. (1.1 cm) Figure 20. Backward-compatible mounting bracket At least 2 screws must be used when mounting the backward-compatible bracket to a wall. Use the bracket as a template for determining the location of the holes to be drilled in the wall. See the Specifications section on page 52 for a description of the appropriate screws. 29 SPM Flex Gas Detector In-line filters Use an external filter to protect the tubing and the detector from contamination. Use particulate filter part number 780248 for non-corrosive gases. Use filter 1991-0147 for corrosive gases. Refer to Detectable Gases on page 51 for specific gases. Filters should be replaced every 3 to 6 months of operation, depending on the cleanliness of the installation environment. Optional sampling wand The sampling wand is connected to the inlet port and used to detect toxic gas at specific locations2. An inlet filter is installed in the grip to prevent debris from entering the unit. All gas-wet surfaces are either Teflon-coated or made of Kynar to be compatible with sticky corrosive gases. An appropriate filter should be used for the gas type being monitored. The sampling wand is supplied with the mounting bracket and required hardware. filter wand back of detector bracket Figure 21. Attaching the sampling wand bracket to the detector 2 Do not use the sampling wand with diisocyanates. 30 SPM Flex Gas Detector filter sampling wand Figure 22. Wand with sampling filter Sampling filters are replaced by unscrewing the halves of the wand’s grip. They will fit in only one orientation. Attaching the shoulder strap If the optional shoulder strap is to be used with a portable detector, attach its spring clips to each of the rings on the detector’s handle. The strap must be laundered before use in clean-room applications. 31 SPM Flex Gas Detector Operation ! WARNING • Protection afforded by the SPM Flex gas detector may be impaired if it is not used according to the manufacturer’s instructions. • Do not operate two-way radios near the detector. Power Before operation, verify that external power is present (the blue LED will be on) or that the internal battery is charged (indicated by the icon shown in the illustration on the following page). The physical power switch, located under the Chemcassette access door, should be on. Typically, it is recommended to leave the physical switch in the on position unless transporting/shipping the unit where there is a possibility of the detector accidentally being turned on with the soft power switch. When the detector’s startup sequence ends, an “In Monitor,” or “Out of Monitor” display (depending on how it is configured in the Setup menu) will appear, depending on the mode the detector was in when it was turned off. The default startup mode can be configured under Power Options (Setup/General). Most options will require acknowledgment (i.e., pressing the Select button). Those that don’t will be displayed for 3 seconds. Pressing the Power button briefly will return the user to the previous screen. Do not power up the detector with a flash drive attached. To turn the unit off, hold the Power button for 5 seconds while on the main display. This will bring up the power off options. Alternately, the power off mode can be accessed via the menu. The user must have the appropriate security level to access. The detector can be used intermittently as long as the Chemcassette cartridge is removed and stored according to manufacturer’s guidelines. Controls The arrows are used to scroll up and down through lists of options, highlighting one at a time. The select button is used to select a highlighted selection. During operation, the power button will cancel a command or, when pressed for more than 3 seconds, will display the main menu. The Open/Close Gate function (see Maintenance on page 46) can be used to open the gate and remove the cartridge for storage. The detector provides tactile and visual feedback (clicks and the unit’s LCD display) to all key presses. The display reflects key presses by illuminating icons representing each of the buttons. The blinking green LED indicates that the unit is on. The green LED blink rate is faster during bootup. The yellow LED indicates a fault; it blinks during instrument 32 SPM Flex Gas Detector faults and is steady during maintenance faults. The red LED indicates alarms; a steady light indicates an Alarm 1, a blinking LED indicates and Alarm 2. The blue LED indicates that the unit is receiving external power. The concentration level of the gas is displayed with the name of the gas below the left side of the display and the units of concentration below its right side. ! CAUTION Do not leave the optics gate of fixed units open. Doing so may allow pressurized gases to escape through the tubing into the unit and then into the local environment. Time There is a real-time clock in the header bar. The real-time clock also verifies that the detector is active. Audible alarm The alarm can be configured for high, medium, and low sound levels. The sound can also be turned off. Low is approximately 75 dB at 1 meter (for office/lab use), Medium is approximately 85 dB at 1 meter (for light industrial use), and High is greater than 90 dB at 1 meter (for heavy industrial use). Monitoring icons A round green icon in the lower left of the display indicates that the detector is on and detecting. “In monitor” is shown next to the icon and the current gas concentration is displayed. The icon changes to indicate that the detector is out of monitor mode, or is in a fault or alarm state. Alarms and faults can be present simultaneously (the alarm will supercede fault notification on the main display and status bar color). When the detector is not detecting, “Out of monitor” is displayed and 4 dashes are shown instead of a gas concentration. When the detector is out of monitor mode, the status bar will change to blue. When a gas concentration exceeds the limit, “Over limit” is displayed and the gas concentration is preceded by the greater-than symbol (“>”). Battery The detector’s battery level is indicated from 0 (fully discharged) to 100% (fully charged). The battery icon displays the approximate battery level on the status bar, while a more accurate value can be found in the Review mode under Additional Status. If the battery’s charge is too low for safe operation, a “Critically low battery!” message will be displayed. Bar graph The bar graph displays the concentration reading up to double the Alarm 2 value for the gas being monitored. The Alarm 1 and Alarm 2 values are indicated numerically on the bar graph. 33 SPM Flex Gas Detector The display can be configured for different backlight and dimming options. By default the backlight will dim after a few minutes of no activity. The display can be configured to turn off the backlight entirely after a set period of time. Inhibit When the detector is in inhibit mode, it will show a bell icon with a red slash through it. 1. Use the [Up] or [Down] buttons to select the “Inhibit” on the Maintenance menu. 2. Press the [Accept] button. 3. The Inhibit Type menu or the Time Out menu can then be selected. Possible inhibit types are none, alarms only, faults only, alarms and faults, and all, as shown in the following table). The Time Out options (the time until the detector exits inhibit mode and returns to active monitoring) are from 1 to 60 minutes. 4. To take the unit out of inhibit, select “None” from the Inhibit Type menu and press the [Accept] button twice to return to the Maintenance menu. NOTE If the inhibit times out before the inhibit state is returned to “none,” maintenance fault code M17 will be displayed. 34 SPM Flex Gas Detector Inhibit Modes Other I/O Behavior 2 4-20 mA Operation 1 Parameter Inhibit Alarms Only Inhibit None Inhibit Faults Only Over-range Level Inhibit Alarms & Faults Concentration > Full Scale Over-range Level Over-range Level Concentration Above Alarm Threshold Concentration Level Concentration Concentration Concentration Inhibit Level Level Level Level Instrument Fault < 1 mA < 1 mA Non-Zero Concentration Concentration Concentration Concentration Concentration Inhibit Level Level Level Level Level Maintenance Fault Maint Fault Level Maint Fault Level #N/A #N/A #N/A Simulated Instrument Fault < 1 mA < 1 mA #N/A #N/A #N/A Simulated Maintenance Fault Maint Fault Level Maint Fault Level #N/A #N/A #N/A Zero Concentration 4 mA Inhibit Level Inhibit Level Inhibit Level Inhibit Level Out Of Monitor 4 mA Inhibit Level Inhibit Level Inhibit Level Inhibit Level Red LED Yes Yes Yes Yes Yes Yellow LED Yes Yes Yes Yes Yes Alarm 1 Relay Yes No Yes No No Alarm 2 Relay Yes No Yes No No Any Alarm Relay Yes No Yes No No Maintenance Fault Relay Yes Yes No No No Instrument Fault Relay Yes Yes No No No Any Fault Relay Yes Yes No No No No #N/A No Over-range Level Inhibit All #N/A Inhibit Level #N/A Buzzer Yes Yes LCD - Alarm Yes Yes Yes Yes Yes LCD - Fault Yes Yes Yes Yes Yes LCD - Concentration Yes Yes Yes Yes Yes Web - Alarm Yes Yes Yes Yes Yes Web - Fault Yes Yes Yes Yes Yes Web - Concentration Yes Yes Yes Yes Yes MODBUS/TCP - Alarm Yes No Yes No No MODBUS/TCP - Fault Yes Yes No No No MODBUS/TCP Concentration Yes Yes Yes Yes No Event History Yes Yes Yes Yes Yes Yes Conditions are listed with the highest priority at the top. The 4-20 mA output will take the state of the highest priority applicable condition that is present. Yes = operates normally Yes with conditions = active only under some conditions No = not active For real and simulated faults only For real and simulated alarms only 35 SPM Flex Gas Detector Review menu Event History The detector’s history can be reviewed from the Review menu. It can be searched by event (alarms, faults, alarms and faults, or all events) or by a range of dates. The results of a search can be displayed as a list of events. Results of searches can be sorted by oldest event first or most recent event first. The results can be exported to a USB flash drive. Web events (event type, event subtype, date/time stamp, concentration/data, or event descriptions) can be viewed. USB port Figure 23. Locations of the USB port and on/off rocker switch NOTE FAT32 is the default file system of the SPM Flex detector. The detector does not read/write NTFS disks. Chemcassette The Chemcassette review screen displays the Chemcassette type, its serial number, expiration date, and the days remaining before it must be changed. Gas Settings The gas being detected, its abbreviation, the CAS number, the LDL, the full scale, the Alarms 1 and 2 setpoints, and the 4-20 mA full scale can be seen on the Gas Settings review screen. 36 SPM Flex Gas Detector Network The Network Summary option displays the IP address, the Subnet mask, the Gateway, and the MAC address. Additional Status The flow rate of the gas, the battery level, the ambient temperature and relative humidity, and the stream temperature and relative humidity can be found on the Additional Status Summary screen. Output State The detector’s relay assignment (i.e., Alarm 1, Alarm 2, or Instrument Fault), the current relay states (e.g., off, off, on), the current mA output (driven value and measured value), and the current inhibit state are displayed in the Output State Summary. Trend/Plot The Trend/Plot review screen shows the date, time, and time range. A trend plot over the specified time range can be displayed. The data can be exported to a flash drive via the USB port as a csv file. Software The Software summary screen displays the number of the current version numbers of the software, the user interface, the gas detection algorithm, the optics algorithm, and the RFID algorithm. ! WARNING Calibration, set-up, and test modes are intended for use by trained personnel or service engineers only. Access to these modes can be passcode protected. Maintenance menu Inhibit Alarms or alarms and faults can be inhibited from the Maintenance menu. The timeout period (the length of time before inhibit mode is exited and monitoring resumes) can also be configured here. The range is 1 to 60 minutes. The default timeout period is 30 minutes. A timeout period of 0 minutes means that the feature is disabled. 4-20 mA current loop calibration This feature allows the detector’s output to match an external monitoring device. Flow characterization Flow characterization is an optional function. When utilized, it allows a flow 37 SPM Flex Gas Detector system to be optimized with a certain type of Chemcassette cartridge. This allows the SPM Flex detector to quickly reach the correct flow rate. Open/close gate The gate can be opened and closed manually. For intermittent use, this allows Chemcassette cartridges to be conveniently removed and stored. Update program Through the Update program option, the new firmware can be loaded via the USB port. Factory service mode For Honeywell Analytics service personnel only. Setup menu Many of the functions of the SPM Flex detector can be configured from the Setup menu. General The backlight intensity and timeout period are chosen from this menu as well. The power-up options are chosen from this menu. The detector can be programmed to be in monitoring mode or out of monitoring mode when powered up. It can also be programmed to power up in the state it was in when it was powered down. (This is the default state.) Finally, if languages other than English are loaded in the detector’s software, the language can be chosen from the General menu. The unit IDs, either short (up to 20 characters) or long (up to 35 characters) and an idle timeout period of from 1 to 60 minutes (20 minutes default) can also be entered from the General menu. An idle timeout period of 0 minutes will disable the idle timeout function. Monitoring The available gases are displayed and can be selected from the Gas menu. Alarms 1 and 2 can be enabled or disabled and their respective setpoints can be adjusted. The LDL can also be enabled or disabled and user-defined LDL limits can be entered. The minimum and maximum values for the 4-20 mA full scale can be entered. The TWA mode can be set for either a fixed start time or a floating start time at the TWA menu. The TWA start time can also be entered. The default is 08:00 but another time can be entered by the user. In this case, the detector will calculate the second time (8 hours later) and third time (16 hours later). Before storing the detector following a gas event, purge the flow system by operating the detector in clean air until it returns to zero. If the case must be wiped down, install push fittings in the Sample In/Out ports to prevent fluids from entering the detector. 38 SPM Flex Gas Detector NOTE The detector’s continuous monitoring algorithm enables a fast response and high sensitivity. It also rapidly indicates trends. However, in some cases, this algorithm can cause the decay time to be substantial, especially with gases having slow stain development, specifically low level AsH3 and GeH4. In the absence of gas, it may be necessary to continue monitoring for several minutes before zero concentration is reported. Latching The SPM Flex detector’s alarms and faults can be either latching or nonlatching, depending on the settings on the Latching menu. The default for both is latching. Outputs The detector’s display characteristics (e.g., brightness, time to dim/time to off) are set at the Display menu. The volume of the alarms (silent, low, medium, or high) are chosen from the Audio menu options. Parameters set from the Relays menu are relay assignments, that alarm relay’s normal state (energized or de-energized), and the maintenance fault’s normal state (energized or deenergized). Three parameters can be set from the 4-20 mA Levels menu: the inhibit level from 1.5 mA to 3.5 mA (the default is 2.0 mA), the maintenance fault level from 1.5 mA to 3.5 mA (the default is 3 mA), and the overrange level from 21.0 mA to 22.0 mA (the default is 21.5 mA). The mA levels are adjusted in 0.5 mA increments. Network The Ethernet mode can be set for either auto or manual. In manual mode, the IP address, subnet mask, and default gateway can be specified. The Modbus TCP menu can be enabled or disabled (disabled is the default). The web server is not designed for more than ten simultaneous connections. (Responses will be slower with more connections.) 39 SPM Flex Gas Detector Security ! WARNING Follow local and site procedures when working with the SPM Flex gas detector. If needed, ensure that the associated control panel is inhibited in order to prevent false alarms. The following procedures must be followed carefully and performed only by suitably trained personnel. When enabled, the detector has five security levels, summarized in this table. Passcodes are entered at the login menu. Log in/log out Reset alarms and faults Review mode Limited maintenance access Change Chemcassette (to same type) Testing Maintenance Limited setup Full setup Change Chemcassette (to any type) Security Program update Factory service menu (Honeywell personnel only) Configuration Manager Access to options for up to five internal flash slots can be chosen from the Configuration Manager menu. Information can also be imported to or exported from the detector via the USB option. 40 5 Factory Service 4 Admin 3 Power User 2 Advanced Maintneance 1 Routine Maintneance Access Rights 0 Not Logged In Security Levels SPM Flex Gas Detector Test menu The detector’s optical system can be checked from the Optics Verification screen (using the verification card.) Alarms 1 and 2, maintenance faults, and instrument faults can be tested from the Simulate menu. The alarms and faults can also be reset from the Simulate menu. The Force Relays and Force 4-20 mA options are also available through the Test menu. Power off Select “Power off” to shut off the detector. Selecting this option has the same effect as pressing and holding the power button for 5 seconds. 41 SPM Flex Gas Detector Troubleshooting Symptom Cause No display Faulty connection at display No blue light No power to the detector Incorrect network connection Detector does not appear on the network Verify that network configuration is enabled Detector not configured for the network Use the “Auto...” setting No external power (blue light) SPM Flex won’t turn on No display SPM Flex is not on Ethernet network Battery too low Blue light present but won’t turn on Check physical power switch (located in Chemcassette cartridge bay) is in the on position. Press and hold the on switch on the top left of the display until you see activity. Faulty connection Check ribbon cable is properly seated Broken display or interface board Replace display/interface board. Contact HA Unit is not powered on See section for turning unit on and troubleshooting power issues Physical connection issue Check correct CAT5 cable is used. Check cable is properly connected at both ends. Confirm network port is active. Incorrect configuration If possible, use Auto detect configuration to automatically get IP configuration from network. Otherwise, check IP configuration is correct. Reboot. Detector does not comFaulty connection municate via USB Detector cannot provide Faulty connection a 4-20 mA output Relay not activated when unit states that it has Faulty connection been Real-time battery expired Shipping tab still in place Chemcassette cartridge Gate closed will not install properly Security lockout Unit will not read ChemDefective RFID chip cassette Chemcassette gate won’t open Chemcassette tape will not advance Chemcassette tape continues to advance 42 Check external connection Check power supply is plugged in Check circuit breaker Plug SPM Flex in to recharge the battery or use external power supply. Confirm that the blue light is present in the LED status bar. If the battery pack is overly discharged the unit will not turn on unless plugged in to the external power. When fully discharged, the battery pack will initially charge at a slower rate than normal until it reaches a minimum voltage. Cannot get Modbus TCP Incorrect configuration data over network No date or time Corrective Action Check the multi-pin connector on the back of the display (the display must removed from cabinet) Turn the circuit breaker on Check the power cable Check cable connection to Ethernet port System is for internal use only Confirm network configuration for SPM Flex and target device. Confirm target device is on the same network/subnet. Confirm SPM Flex IP address on Review/Network Summary. Check the cable connection Check the cable connection Check the cable connection Replace the real-time battery Contact Honeywell Analytics Confirm that the red shipping tab has been removed from the payout spool Use Change Chemcassette wizard to open gate and guide you through the process Confirm you are logged in with an account with appropriate security level to change Chemcassette cartridge Replace Chemcassette Gate open command not sent Use either Change Chemcassette wizard or Maintenance / Open/Close gate. Gate motor is broken Replace gate motor or contact HA No power Check the cable connection Encoder Check the cable connection Hall effect sensor Check the cable connection SPM Flex Gas Detector Symptom Cause Corrective Action Chemcassette gate will not open No power Check the cable connection Cannot remove Chemcassette cartridge Gate open command not sent Use either Change Chemcassette wizard or Maintenance / Open/Close gate. Chemcasette cartridge Door is not properly closed or door will not latch closed aligned Make sure Chemcassette cartridge is fully seated. Make sure USB stick is not interfering with door. Make sure door is fully seated on both sides – latches should hook under the tabs on the side of the case easily. Make sure all four latches are properly secured. Pump will not turn on No power Check the cable connection Inlet or outlet tube is blocked Confirm inlet and outlet tube are free of restrictions or kinks, and that all filters are clean. Gate is not sealing correctly Confirm Chemcassette cartridge is fully seated. Exit and enter monitor to pull a fresh part of the tape. Pump is old Replace pump. Contact HA. Filters are clogged Check filters (internal and external) and replace as needed Tubing connections are not properly seated Confirm all tubing connections are properly seated (internal and external). Cannot achieve target flow rate SPM Flex will not con- Incorrectly installed mounting nect properly or stay on plate mounting plate Mounting screws not installed Flow system issue Gas readings are not as expected Chemcassette cartridge issue Buzzer does not activate No power SPM Flex won’t turn off Make sure the mounting plate is installed the correct direction so that the mounting heads on the back of the SPM Flex can slot in and slide down to secure. Confirm the mounting screws have been installed on the back of the SPM Flex Confirm all tubing connections are properly inserted. Confirm correct filter type is used for the target gas. When in doubt use the corrosive filter. Confirm target flow rate is achieved. Confirm correct Chemcassette cartridge is selected for the target gas. Confirm Chemcassette cartridge is within operating age, sampling conditions, and has been correctly stored. If generating a gas bump test gas to confirm performance, refer to gas generation document located in the online High Tech Technical Library. Refer to individual Chemcassette type technical notes for specific performance information. Check the cable connection Power button on top left / menu Ensure that user is logged in with the correct security level option not available Physical power switch is on, SPM Flex unintentionally unintentionally pushing power turns on during transport switch by display Put the physical power switch located behind the Chemcassette door in the off position. Place switch back in the on position when ready to use again. Detector will not charge Faulty connection Check the cable connection *corrective actions vary with security levels 43 SPM Flex Gas Detector Instrument Fault/Maintenance Fault/Information Codes Display String Subtype Type Meaning of Parameter Technical Description Probable Cause Pump failure 101 Flow Failure Flow failure Flow, cc/min 103 Gate Motor Failure Gate timeout N/A -Flow line obstruction -Positive pressure Corrective Action Replace pump Check flow lines Check for positive pressure Gate motor failure -Gate motor sensor failure Replace Broken tape 104 Tape Advance Failure Tape advance timeout N/A -Tape drive motor failure Replace Instrument Faults -Encoder failure 105 106 107 108 High Background Gas Table Invalid Comm. Failure RFID. Chemcassette Read Failure optics counts too low after multiple pulls Gas table invalid comm. fail to RFID reader RFID read failure N/A error code N/A N/A 109 Temperature Too High overtemperature Temperature, C 110 Temperature Too Low undertemperature Temperature, C 111 Optics Error 112 Low Voltage Fault 113 Chemcassette Expired 114 Optics Subsystem Fail value of LED drive or reference detector signal out of range DC input voltage low Now>LifeInBag+Mfg or Now>LifeOutOfBag+FirstUse no comm. to optics block N/A Tape issue CC RFID issue RFID PCB failure CC RFID issue Environment is over recommended temperature Environment is below recommended temperature Check, replace Replace, contact HA Replace Replace, contact HA Optics PCB failure Replace Chemcassette cartridge has expired Replace Improve ventilation/ cooling Bring environment to correct temperature voltage in Volts 1.0 or 2.0 for two causes N/A Optics cable disconnected Reconnect, replace -Optics PCB failure 115 Comm. Failure GD no comm. to GD N/A Main PCB failure Contact HA Electrical Noise Optics signal noisy N/A External electrical noise Check environment 116 Contact HA Maintenance Faults PCB failure 44 PCB failure Replace N/A External connection issue Check external Ethernet connection Battery critically low, shutting down Battery level Main PCB failure Low battery Flow Warning Flow is unregulated Flow, cc/min Flow out of control range Replace Charge unit Check Chemcassette Chemcassette Low Tape will run out in 24 hrs N/A Low tape Idle Timeout out of monitor too long N/A Out of monitor 10 11 Low Voltage Warning Comm Failure SPI File System Error Comm. Invalid GD Warn CC Expiring Software Error Optics Adjust Fail Flow Cal Fail DC input voltage low no SPI comm. to UI File system error in NAND or USB. Invalid SPI parameter data from GD. Prediction tape will expire in 24 hours. catch all voltage in Volts N/A N/A N/A N/A error code error code error code 12 Inhibit Timeout N/A 13 Force mA Timeout N/A 117 118 NV Memory Corrupt mA Output Mismatch CRC error Feedback ADC mismatches DAC code to which test failed. Error magnitude in mA. 119 Ethernet Fail Ethernet hardware in UI failed to initialize 120 Battery Very Low 14 Force Relay Timeout 15 Battery Low N/A Battery Low Old tape Unit has been left in inhibit mode longer than timeout warning Unit has been in force mode too long Unit has been in force mode too long Battery is low Check sample tube Replace Put in monitor and reset fault Contact HA Contact HA Contact HA Replace Reset Take out of force mode, reset fault Take out of force mode, reset fault Charge battery table continued... SPM Flex Gas Detector Instrument Fault/Maintenance Fault/Information Codes Information Display String Subtype Type Meaning of Parameter Probable Cause Corrective Action N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A zero zero N/A User level which just logged in, 0 is lowest level User level which just logged out, 0 is lowest level N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A which type of inhibit. N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A LED drive counts 1 – first SLDE N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A 2 – second SLDE N/A N/A N/A N/A N/A N/A Technical Description SPM Energized Monitoring Started Monitoring Stopped The microprocessor booted Monitoring started Monitoring stopped A new windows was pulled. (remove for production) Tape Advanced Force mA Requested mA Output Forced mA Output Released current loop forcing started current loop forcing ended. User Login User logged-in successful. User Logged Out User logged out manually or by timeout. 10 Alarm/Fault Reset Request 11 Silent Buzzer Request 12 13 14 15 16 17 18 Enter Monitor Request. Exit Monitor Request Change CC Started Inhibit Started Inhibit End Request 4-20mA Calibration Started Flow Characterization Started 19 Update Program Started 20 21 22 23 24 25 26 27 28 29 30 31 Update Program Failed Update Program Success Gas Related Configuration. Non Gas Related Configuration. Security Set Up Configuration Optics Verification Started Simulation Started Force Relay Started Force Relay End Request Time Changed Optics Auto Adjust Requested Optics Auto Adjust Success UI requests alarm/fault reset. UI requests to shut up buzzer through Alarm/ fault reset menu. UI requests to enter monitor UI requests to out of monitor UI initialize change CC sequence UI initializes inhibit. UI request to end inhibit. UI initializes 4-20 calibration. UI initializes flow char User chose an update file to perform program update. Update program failed Update program success Gas related set up changed by UI. Non-gas set up changed by UI. Security set up changed UI initialized optics verification sequence UI requests to start simulation. UI started force relay. UI exited force relay function. UI time set 32 Optics Corrected Optics Auto-Adjust Success. LED output dropped unexpectedly 3 – Reference Photodiode Shift 33 Mfg Service Mode 34 Electrical Noise UI has received commands to go into manufacturing service mode Optics block reports signal is noisy None 45 SPM Flex Gas Detector Maintenance Perform maintenance activities according this schedule. Use only Honeywell Analytics replacement parts. Use appropriate static discharge mitigation while servicing the interior of the detector to avoid damage. Maintenance Intervals (months) Description Replace the end-of-line filter Check pump Check for system leaks 12 Authorized Honeywell preventive maintenance Verify optic system response Replace internal filters Check stepper motor as needed Check gate motor as needed l* Clean the exterior surfaces as needed Replace real-time coin battery 3 years or as needed Replace main battery return the detector to Honeywell Analytics *or as needed Authorized Honeywell preventive maintenance In addition to the standard maintenance schedule performed by the end user, it is recommended that units are periodically returned to Honeywell for comprehensive inspection, cleaning, and systems tests. While not required, users are encouraged to have this service performed every one to three years depending on usage, site conditions, and gases monitored to help ensure years of continued smooth operation. To replace the end-of-the-line filter Hold down the locking ring on the push fitting, as shown in Figure 24, to release the filter. Gently press the new filter into the fitting until it locks. An arrow is printed on the filter to show the gas’s proper direction of flow. locking ring filter push fitting to inlet port Figure 24. Removing a filter from a push fitting 46 SPM Flex Gas Detector To check the pump The detector will display an instrument fault in the event of a pump malfunction. See pages 44-45. To check for system leaks 1. Place the detector in Inhibit mode. 2. In the Review menu, navigate to the Additional Status option. 3. Plug the inlet tube and observe the flow reading on the LCD display. If the flow stabilizes at more that maximum cc/m (shown for each gas in the following table), there is likely a system leak. Family Gas Hydrides Mineral Acids Mineral Acids (export unrestricted) Oxidizers Amines COCl2 Diisocyanates Hydrazines HCN SO2 O3 H2O2 AsH3 PH3 B2H6 SiH4 GeH4 H2Se H2S HF HCl HBr BF3 HF HCl HBr BF3 Cl2 F2 NO2 ClO2 NH3 DMA TDMAT TMA COCl2 TDI MDI HDI N2H4 MMH UDMH Maximum Allowable Leak Rate (cc/min) 30 30 30 30 30 30 50 50 50 50 50 50 50 50 50 50 30 30 50 30 30 30 TBD 50 TBD TBD TBD 50 50 50 30 30 30 TBD 47 SPM Flex Gas Detector To verify optic system response Using the optional optics test card, perform the optics verification test. From the Test menu, select the Optics verification option and follow the instructions on the display. If the “Optics verification success!” message is displayed, the optic system is functioning normally. If the “Optics verification failed! Please remove card and put Chemcassette back” message is displayed, return the detector the Honeywell Analytics for service. To replace internal filters Dust filter/acid scrubber assembly Dust filter/acid scrubber locking ring Manifold Area of detail Bypass filter bracket Bypass filter screw Bypass filter Figure 25. Removing the internal filters The dust filter and acid scrubber are attached to a single manifold. Press and hold the locking ring on the fitting firmly in the direction of the arrow and lift the filter assembly out together. Replace the filters in the manifold and reattach the assembly by gently pressing it into the Push fittings until it locks. To free the bypass filter, remove the bypass filter screw and pull out the bracketfilter assembly. Disconnect the black bypass tube. Attach the tube to the new filter-bracket assembly, insert it in the manifold, and tighten the bypass filter screw. Perform a leak test (see the previous page) after every filter replacement. To check the stepper motor The detector will display an instrument fault in the event of a stepper motor malfunction. See pages 44-45. To check the gate motor The detector will display an instrument fault in the event of a gate motor malfunction. See pages 44-45. 48 SPM Flex Gas Detector To replace the real-time coin battery The battery can be removed by gripping it firmly with needle-nose pliers and pulling straight out. Gently press the new battery with the positive (+) side facing out until it snaps into place. 24 VDC COM Fault NC NO Relay 2 COM NC NO Relay 1 4-20 mA 357/30 Ener0% Hg gize MA DE IN U SA Real-time coin battery Figure 26. The real-time coin battery To clean the exterior surfaces Exterior surfaces of the detector can be cleaned with a soft cloth moistened with water or with 6% IPA alcohol wipes. Storing the detector When storing the detector with the power adaptor plugged in, the rocker switch can be left in the on position; the battery will continue charging. The detector can also be charged with the rocker switch off. When storing the detector without being plugged into the power adaptor, turn the rocker switch off. This will minimize the power .. This is the recommended setting for long-term storage or for transporting the detector. Charge the detector at least four hours every three months (the rocker switch can remain off). Recycling At the end of its usable life, return the detector to Honeywell Analytics for recycling. Storing Chemcassette® cartridges Follow the instructions on the cartridge label for acceptable storage temperatures and expiration dates. 49 SPM Flex Gas Detector Stor.Temp.: Less than 0°C (32°F) Exp.Date: 04-Mar-2015 SFID# 1265-40129131026 (XP-MA) Made in USA Flex CC-S-U XP Mineral Acids (XP-MA) P/N 1265-4012 Flex CC-S-U XP Mineral Acids Labels Detect: HF, HCl, HBr, BF3 Packing label P/N 1265-4012 Flex CC-S-U XP Mineral Acids (XP-MA) P/N 1265-4012 Exp. Date: 04-Mar-2015 Stor. Temp.:Less than 0°C (32°F) SFID# 1265-40129131026 Cartridge top Made in USA Cartridge bottom Figure 27. Chemcassette packing and cartridge labels The cartridge’s bottom label contains the RFID tag. Figure 28. The detector’s nameplate 50 SPM Flex Gas Detector Detectable Gases Default Alarm Family Hydrides Mineral Acids Mineral Acids (export unrestricted) Oxidizers Amines COCl2 Gas Range TLV AsH3 0.5-500ppb 5 ppb PH3 3-3000 ppb 300 ppb B2H6 5-1000 ppb 100 ppb LAL 2014 NIC: 0.1ppmTWA; 0.5ppm STEL-C A1 A2 Max. Response time Sample (T50) at 2TLV Tubing gas conc.(sec) length (m) ChemCassette Name P/N P/N (30-90d) (14d) Optimum Optimum %RH range6 Temp for +/-20% range accuracy (ºC) 1 ppb 2.5 ppb 5 ppb 55 30 10-70% RH 3, 5 5 ppb 150 ppb 300 ppb 30 30-70% RH 3, 5 10 ppb 50 ppb 100 ppb 14 30 Flex CC XP Hydrides 12653000 12654000 30-70% RH 4, 5 0-40 34-50% RH 3, 5 SiH4 0.03 - 50 ppm 5 ppm 0.05 ppm 2.5 ppm 5 ppb 13 30 GeH4 50-2000 ppb 200 ppb 100 ppb 100 ppb 200 ppb 245 30 40-50% RH 3, 5 H2Se 2-500 ppb 50 ppb 5 ppb 25 ppb 50 ppb 14 30 10-60% RH 3, 5 H2S 0.001-9.999 ppm 1 ppm 0.005 ppm 0.5 ppm 1 ppm 30 10-75% RH 3, 5 HF 0.02-20 ppm 0.5 ppm 2 ppm STEL-C 0.03 ppm 1 ppm 2 ppm 15-75% RH 4, 5 30-50% RH 4, 5 HCl 0.02-20 ppm 2 ppm STEL-C 0.03 ppm 1 ppm 2 ppm HBr 0.02-10 ppm 2 ppm STEL-C 0.03 ppm 1 ppm 2 ppm BF3 0.05-10 ppm 1 ppm STEL-C 0.1 ppm 0.5 ppm 1.0 ppm 15-60% RH 4, 5 HF 0.4-20 ppm 0.5 ppm 2 ppm STEL-C 0.4 ppm 1 ppm 2 ppm 15-75% RH 4, 5 HCl 0.02-20 ppm 2 ppm STEL-C 0.03 ppm 1 ppm 2 ppm HBr 0.02-10 ppm 2 ppm STEL-C 0.03 ppm 1 ppm 2 ppm BF3 0.05-10 ppm 1 ppm STEL-C 0.1 ppm 0.5 ppm 1.0 ppm Cl2 0.005 - 5 ppm 0.5 ppm 0.02 ppm 0.25 ppm 0.5 ppm 10 Cl2 0.01-5 ppm 0.5 ppm F2 0.01-10 ppm 1 ppm NO2 0.03-10 ppm ClO2 NH3 Flex CC XP Mineral Acids 12653001 12654001 0-35 Flex CC-U XP Mineral Acids 12653012 12654012 0-35 Flex CC XP Chlorine 12653002 12654002 0-40 15-60% RH 4, 5 30-50% RH 4, 5 15-60% RH 4, 5 15-60% RH 4, 5 30-55% RH 3, 5 0.05 ppm 0.25 ppm 0.5 ppm 10 0.05 ppm 0.5 ppm 1.0 ppm 10 0.2 ppm 0.05 ppm 0.1 ppm 0.2 ppm 56 30 20-1000 ppb 100 ppb 25 ppb 50 ppb 100 ppb 36 10 5-90% RH 0.01-150 ppm 25 ppm 25 ppm 30 0-90% RH 3 0.1 ppm OSHA PEL 0.05 ppm 12.5 ppm DMA 0.5-50 ppm 5 ppm 0.1 ppm 2.5 ppm 5 ppm 10 30 TDMAT 0.01 -20 ppm n/a 0.05 ppm 1 ppm 2 ppm 14 30 TMA 0.5-50 ppm 5 ppm 0.1 ppm 2.5 ppm 5 ppm 1 ppb MDI 2-60 ppb HDI 12654004 0-40 Flex CC XP Ammonia 12653003 12654003 Flex CC XP Phosgene 12653007 12654007 0-40 Flex CC Diisocyanates 12653006 12654006 0-40 0-35 0-85% RH 10-70% RH 4 ,5 5-90% RH 3 5-90% RH 3 0-90% RH 3 30 100 ppb 30 2 ppb 0.15 5 ppb 2.5 ppb 5 ppb 0.15 2-60 ppb 5 ppb 2.5 ppb 5 ppb 0.15 TBD N2H4 5-1000 ppb 10 ppb 5 ppb 10 ppb 0.15 10-70% RH 2 MMH 3-2000 ppb 10 ppb 5 ppb 10 ppb UDMH 3-5000 ppb 10 ppb 5 ppb 10 ppb HCN 0.5-30 ppm 4.7 ppm 2.4 ppm 4.7 ppm 10 SO2 5-2500 ppb 250 ppb 120 ppb 250 ppb 30 O3 10-1000 ppb 100 ppb 50 ppb 100 ppb Flex CC Ozone H2O2 0.1-3 ppm 100 ppb 100 ppb Flex CC Hydrogen Peroxide Hydrazines 100 ppb 0.3-150 ppb 12653004 1 ppb nates 7-4000 ppb TDI Flex CC Fluorine/Oxidizers 50 ppb Diisocya- COCl2 0-85% RH 2014 NIC (1 ppb TWA; 3 ppb STEL) 0.5 ppb 10-90% RH 25-65% RH Flex CC Hydrazines 12653008 12654008 n/a 12654009 12654005 12654011 12654010 0-40 TBD TBD TBD Flex CC Hydrogen Cyanaide Flex CC Sulfur Dioxide 12653005 12653011 12653010 0-30 30-75% RH 0-40 TBD 0-40 30-55% RH 0-40 TBD 1 Source: ACGIH 2014. Outside of RH range: 2 Tends to have lower response at higher humidities. 3 Tends to increase sensitivity at higher humidities (due to the chemistry of the reaction). 4 Tends to under-report at higher humidities (typically >75% RH) due to the gas characteristics to adhere or decompose on contact with water/moisture. The response seems to be lower but the actual gas concentration under these high humidity conditions will be lower than expected. 5 Tends to under-report in dry conditions (<25-30% RH). 6 Depending on the combination of temperature and humidity, even within the ranges specified above, a unit’s performance efficiency can be influenced due to condensation, physical tape material changes, or optical changes. Consult Honeywell Analytics’ Service Department. 51 SPM Flex Gas Detector Specifications Description Value Detection Technique Chemcassette-cartridge-based with advanced self monitoring optics design Dimensions see page 27 Weight 9.1 lbs. (4.1 kg) Operating Temperature 0C to 40C for most gases / applications Operating Humidity 0-100% RH for unit (Sample RH limited per tape/calibration). Sample line requires additional hardware to remove moisture in high RH conditions where condensing may occur. The sample must be non-condensing. Dry conditions may require humidification. Power supply Universal Line powered (90-260 VAC 50/60 Hz) for battery charger & non-classified use. Battery: 6+ hours under typical conditions – acts as battery back-up in fixed applications Power consumption Power adaptor ~1.9 A at 24 VDC (including battery-charging current) Manufacturer FSP Group Model FSP135-AAAN1 Input 100-240 VAC, 2 A, 50-60 Hz Output 24 VDC, 5.62 A CCN QQGQ (E190414) Mark of conformity UL listed Communications Relays: Alarm 1, Alarm 2, Fault (user configurable for normally open/closed) 4-20mA Ethernet (with Modbus TCP/IP and web server) USB port (for memory stick configuration/data transfer) Flow System Automatic flow control with bypass system, 250 or 500 cc/min at tape, higher flow at inlet to reduce sample time (internal bypass system); sample up to 100 ft Local Alarms/Status Visual: LEDs for alarm, normal condition and fault Audible: User selectable: Off, Low ~75 dB at 1 m, Medium ~85 dB at 1 m, High >90 dB at 1 m Interface 4 large buttons, 3.5” Color LCD TFT display, web server Data Logging Rolling 3 months (15 sec. with no gas reading, 1 sec. when reading gas), Event history (1500 events – approx. 1 year) 5/16 in x 2 in vibration-resistant stud anchor for concrete (McMaster-Carr 94475A185 or equivalent), add 0.25 in. to length when mounting bracket with sun shield Mounting screws concrete wood 5/16 in. x 2 in. flange head lag screw for wood (McMaster-Carr 95526A375 or equivalent), add 0.25 in. to length when mounting bracket with sun shield Battery type Battery life Maximum inlet/outlet pressure differential Lithium ion Approximately 70% of its original capacity after 300 full charge/discharge cycles The overall maximum load on the pump between the inlet and the exhaust should not exceed 10 inches H2O Communication connector, optional communications cable 60 V, 5 A maximum Relays 250 V, 6 A maximum Wire gauges Minimum 18 Maximum 12 USB 2.0 or later Indoor/outdoor use? Operating altitudes -1,000 to 3,000 ft. above sea level above 3,000 ft. to 6,000 ft. above sea level Ingress Protection rating both (the power supply is indoor only) standard requires factory adjustment, contact Honeywell Analytics IP65 External switch or circuit breaker requirement (description & location) meet or exceed all local codes and regulations Ventilation requirements mount with no obstructions within 4 in. (10 cm) of either side or within 2 in. (5 cm) above and below the detector 4-20 mA output defaults and ranges Storage conditions Inhibit 2 mA, programmable from 1.5-3.5 mA in 0.5 mA increments Maintenance 3 mA, programmable from 1.5-3.5 in 0.5 increments Instrument fault 1 mA or less, not programmable (will be driven under 1 mA) Over-scale 21.5 mA, programmable 21-22 mA 4-20 mA configurations sink, source, isolated Detector 32°F to 104°F (0°C to 40°C), 0-100% RH non-condensing Chemcassette cartridges less than 32°F (0°C) UL 61010-1, 3rd Edition, 2012-05 (ELECTRICAL EQUIPMENT FOR MEASUREMENT, CONTROL, AND LABORATORY USE Part 1: General Requirements Detector CAN/CSA-C22.2 No. 61010-1, 3rd Edition, 2012-05, (ELECTRICAL EQUIPMENT FOR MEASUREMENT, CONTROL, AND LABORATORY USE - Part 1: General Requirements) IEC 61010-1:2010, 3rd Edition Certifications FCC approval for RFID board + Canadian and European UL/cUL Recognition to UL 2054 + 60950-1 Battery IEC 62133 1st Edition CB Certification UN Test Report to UN 38.3 Self-declared European CE Mark on detector for: 52 EMC, LVD, ROHS, WEEE SPM Flex Gas Detector Accessories and Parts ! WARNING Use only accessories and parts meeting or exceeding Honeywell Analytics’ specifications. Description SPM Flex base units SPM Flex fixed units1 SPM Flex portable units2 Accessories Part No. SPM Flex base unit 1265-0500 Australia SPMF-F1AU Switzerland SPMF-F1CH Denmark SPMF-F1DK Europe SPMF-F1EU Great Britain SPMF-F1GB Japan SPMF-F1JP North America SPMF-F1US Australia SPMF-P1AU Switzerland SPMF-P1CH Denmark SPMF-P1DK Europe SPMF-P1EU Great Britain SPMF-P1GB Japan SPMF-P1JP North America SPMF-P1US Shoulder strap for portable base unit (made from clean-room-compatible materials; the strap must be properly laundered and stored for clean room use) SPMF-STRP Inlet sample wand 1265-0171 Optics packaged test card assembly 1265-2014 Power battery Energizer CR2032 coin battery factory-replaceable only 0140-0013 continued... 53 SPM Flex Gas Detector Description Power Cables and Connectors Universal power supply, for indoor use only (no cord) SPMF-PWRS North America (120VAC) power cord 874333 Australia 874557 Great Britain 874558 Denmark 874559 Switzerland 874560 Europe 874561 Japan (100VAC) 1874-0112 Ethernet cable with weatherproof connector SPMF-ECON Communication cable with weatherproof connector SPMF-CCON Spare handle for portable base unit Parts Dust covers Tubing Mounting brackets Filters3 Part No. SPMF-HNDL For Ethernet port SPMF-DCET For communication port SPMF-DCCO For power port SPMF-DCPO Tubing 1/4” O.D x 1/8” I.D. FEP - for sample inlet, price per foot 102599 Tubing 1/4” O.D x 1/8” I.D. FEP - for sample inlet, price per foot 100440 Union fitting - use to connect disposible end of line filter to tubing 0235-0095 Standard mounting bracket for fixed unit (for SPM Flex only) SPMF-MBST Retro-fit mounting bracket for fixed unit (compatible with original SPM) SPMF-MBRF Filter particulate manifold assy 1265A0115 Bypass filter 871134 Disposable microfibre dust filter 780248 Acid scrubber filter 710235 Line filter for corrosive gases 1991-0147 continued... 54 SPM Flex Gas Detector Description Long --up to 90 days (XP) or 30 days (standard) of continuous monitoring Chemcassette cartridges Short --up to 2 weeks of continuous monitoring Part No. SPM Flex CC XP Hydrides 1265-3000 SPM Flex CC XP Mineral Acids (may require an export license) 1265-3001 SPM Flex CC XP Chlorine 1265-3002 SPM Flex CC XP Ammonia 1265-3003 SPM Flex CC Fluorine Oxidizers 1265-3004 SPM Flex CC Sulfur Dioxide 1265-3005 SPM Flex CC Diisocyanates 1265-3006 SPM Flex CC XP Phosgene 1265-3007 SPM Flex CC Hydrazine 1265-3008 SPM Flex CC Hydrogen Peroxide 1265-3010 SPM Flex CC Ozone 1265-3011 SPM Flex CC-U XP Mineral Acids (No export restriction - LDL limited above 366 ppb for HF) 1265-3012 SPM Flex CC-S XP Hydrides 1265-4000 SPM Flex CC-S XP Mineral Acids (may require an export license) 1265-4001 SPM Flex CC-S XP Chlorine 1265-4002 SPM Flex CC-S XP Ammonia 1265-4003 SPM Flex CC-S Fluorine Oxidizers 1265-4004 SPM Flex CC-S Sulfur Dioxide 1265-4005 SPM Flex CC-S Diisocyanates 1265-4006 SPM Flex CC-S XP Phosgene 1265-4007 SPM Flex CC-S Hydrazine 1265-4008 SPM Flex CC-S Hydrogen Cyanide 1265-4009 SPM Flex CC-S Hydrogen Peroxide 1265-4010 SPM Flex CC-S Ozone 1265-4011 SPM Flex CC-S-U XP Mineral Acids (No export restriction - LDL limited above 366 ppb for HF) 1265-4012 1 Includes a standard wall mounting bracket, a battery, a power supply (for indoor use only), a manual on CD, a printed quick start guide, and a power cord appropriate for the region. 2 Includes an Ethernet connector, a power connector, a handle accessory kit, a clean-room-safe shoulder strap, a user manual on CD, a printed quick start guide, and a power adaptor with a plug and cable appropriate for the region. (The shoulder strap must be properly laundered and stored for clean room use.) 3 Use an external filter to protect the tubing from contamination (the particulate filter for non-corrosive gases, the corrosive filter for corrosive gases). For multiple-gas applications or if the correct filter is not known, use the corrosive filter. Replace the filter every 3 months. Refer to Detectable Gases for information on specific gases. 55 SPM Flex Gas Detector Certifications Figure 29. Battery certification 56 SPM Flex Gas Detector Figure 30. IEC certification 57 SPM Flex Gas Detector ce ld Figure 31. XXX certification 58 SPM Flex Gas Detector ld ce pla Figure 32. YYY certification 59 SPM Flex Gas Detector ce ld pla Figure 33. ZZZ certification 60 SPM Flex Gas Detector 61 SPM Flex Gas Detector Warranties SPM Flex warranty All products are designed and manufactured to the latest internationally recognized standards by Honeywell Analytics under a Quality Management System that is certified to ISO 9001. As such, this instrument (including the pump) is warranted under proper use, to the original enduser purchaser, against any defects in materials or workmanship related failures for a period of 12 months from the date of first turn-on or 18 months from delivery from Honeywell Analytics to the customer, whichever is less. Separate warranty conditions apply to the sensor cartridges limited as indicated below. During this period, Honeywell Analytics will repair or replace defective parts on an exchange basis, F.O.B. to approved service centers on a global basis. This warranty does not cover damage caused by accident, abuse, or abnormal operating conditions. Defective equipment must be returned to Honeywell Analytics for repair. Before returning materials for repair or replacement, the Customer must obtain a Service Event Number (SE#) by contacting Honeywell Analytics Service in advance; include a detailed report stating the nature of the defect and ship the equipment prepaid to Honeywell Analytics’ factory. If no detail report is included, Honeywell Analytics reserves the right to charge an investigative fee (prices available upon request) before any repair or replacement is performed. Returned goods must detail the Service Event Number (SE#) clearly on the package. Service in the field or at the customer’s premises is not covered under these warranty terms. Time and travel expenses for on-site warranty services will be charged at Honeywell Analytics’ normal billing rates. Contact your Honeywell Analytics representative for information on available Service Contracts. Honeywell Analytics shall not be liable for any loss or damage whatsoever or howsoever occasioned which may be a direct or indirect result of the use or operation of the Contract Goods by the Buyer or any Party. This warranty covers the gas detector and parts sold to the Buyer only by authorized distributors, dealers and representatives as appointed by Honeywell Analytics. This warranty does not cover defects attributable to improper installation, repair by an unauthorized person or the use of unauthorized accessories/parts on the product. A warranty claim will only be accepted if a proof of purchase is submitted and all conditions obtained within this Warranty are met. Honeywell Analytics reserves the right to validate any warranty claim prior to processing. Upon acceptance of a warranty claim, Honeywell Analytics will repair or replace the defective product free of charge. The initial warranty period is not extended by virtue of any works carried out there after. Instruments which have been repaired or replaced during the warranty period are warranted for the remainder of the unexpired portion of the original warranty period. Honeywell Analytics is released from all obligations under its warranty in the event repairs or modifications are made by persons other than its own authorized personnel, unless such work is authorized in writing by Honeywell Analytics. Honeywell Analytics is released from all obligations under its warranty in the event that detection substrates other than Honeywell Analytics’ Chemcassettes® have been installed and used in Honeywell Analytics’ instruments. Honeywell Analytics reserves the right to change this policy at any time. Contact Honeywell Analytics for the most current warranty information. Chemcassette® cartridge warranty All Chemcassette cartridges® are warranted for a period not to exceed the Chemcassette® cartridge expiration date printed on each package and tape reel. 62 SPM Flex Gas Detector 63 SPM Flex Gas Detector Index accessories 53 alarms 33, 37, 38 configuration 33 levels 16 relay’s normal state 39 Allen wrench 22 altitude 5 default gateway 39 display 16 back light 34 characteristics 39 dimming options 34 LCD 32 system, LCD and LED 16 backlight intensity 38 timeout period 38 bar graph 16, 33 battery 10, 30, 32 charge, low 33 level 33, 37 life 10, 30 pack, lithium-ion 10, 30 real-time coin 49 replacement 10, 30 bootup 32 buttons 15 Accept/Select 16, 32 arrows 16, 32 Power/Cancel 16 error messages 18 Ethernet 19, 39 event history 36 exporting 36 searching 36 sorting 36 web events 36 gas concentration level 33 reading 33 gas detection algorithm version number 37 gases, detectable (table) 51 sticky corrosive 31 gas-wet surfaces 31 gate, manual operation 38 gate motor, checking 48 gateway 37 certifications 56 Chemcassette cartridges 33 access door 32 bottom label 50 labels 50 RFID tag 50 expiration dates 49 labels 49 loading 17 storage 17, 32, 38, 49 takeup spindle 17 packing labels 50 cleaning 49 clock 33 conduit 19 configuration 32 Configuration Manager 40 connections configurable 22 relay terminal 20 three-wire 22 connectors 19 communications 19, 20 Ethernet 19 power 19 weatherproof 19 controls 13, 14 cover 11 64 handle 11, 17 hinges 17 Factory service mode 38 false alarms, preventing 40 fasteners 26 icons 32, 33 idle timeout period 38 IDs 38 inhibit level 39 modes 34, 35 state 37 inlet port 25, 31 In Monitor mode 32 IP address 37, 39 faults 32, 37 instrument 32 maintenance 33 notifications 33 tactile 32 visual 32 FEP, see Teflon fluorinated ethylene polymer 25 file system, default (FAT32) 36 NTFS 36 filters, acid scrubber 48 dust 48 end-of-the-line, replacing 46 external 30 for non-corrosive gases 30 inlet 31 inline 25, 30 internal, replacing 48 particulate 30 firmware 38 flash drive 36 flash slots, internal 40 Flow characterization 37 flow rate 25, 37 flow system optimization 37 4-20 mA 19, 38 current loop calibration 37 isolated configuration 22 output 22 sink configuration 22 source configuration 22 J-K Kynar 31 languages 38 Latching 11, 17, 39 LDL 38 LEDs 32 blue 33 green 32 red 33 yellow 32 local codes 19 local detector option 25 SPM Flex Gas Detector MAC address 37 main display mode 16 maintenance 46 fault level 39 faults 34 intervals 46 Maintenance menu 37 Maintenance menu, Inhibit 37 mA output 37 menu navigation 15 Modbus register definitions 32 TCP menu 39 Monitoring 38 mounting brackets, 26 backward-compatible 29 standard 26 recharging 10, 30 recycling 49 relative humidity 37 relays 19, 37 assignments 39 configuration 23 menu 39 states 19,37 review menu Additional Status 37 Chemcassette 36 Gas Settings 36 Network 37 Output State 37 Software 37 rend/Plot 37 RFID algorithm version number 37 tag 50 Teflon 31 fluorinated ethylene polymer 25 temperature ambient 37 stream 37 terminal block 19 terminal module 20 terminals 22 Test menu 41 timeout period 37 transporting 32 transport time 25 Troubleshooting 42 tubing 25 TWA mode 38 network 39 operation 32 optical system, checking 41 optics algorithm version number 37 verification screen 41 system response, verifying 48 Out of Monitor mode 32 outputs 39 overrange level 39 parts, replacement 46, 53 passcodes 37, 40 ports 19 power 41 adaptor 19 button 13, 41 cable 10, 30 external 32 options 32 supply/charger 10, 30 switch 32 switch, soft 32 power-up options 38 pressure, maximum 25 pump checking 47 malfunction 47 push fittings 25, 46, 48 sampling wand 31 security levels 32, 40 Admin 40 Advanced Maintneance 40 Factory Service 40 Not Logged In 40 Power User 40 Routine Maintneance 40 Select button 32 Setup menu 38 shipping 32 software 23 version number 37 sound levels 33 specifications 52 start time fixed 38 floating 38 startup mode, default 32 startup sequence 13 static discharge 46 status bar colors 16 stepper motor, checking 48 subnet mask 37, 39 system leaks, checking for 47 unit IDs 38 update program option 38 USB port 36, 38 user interface version number 37 users, unqualified 18 vacuum 25 verification card 41 wall outlet 10, 30 warranties 62 web server response time 39 simultaneous connections 39 wiring cable bundles 20, 23 routing 10 4-20 mA 20 AWG 19 configurable, isolated 22 configurable, sink 22 configurable, source 22 configuration, default 20 cover 22 diagram, 4-20 mA isolated 21 diagram, 4-20 mA sink 21 diagram, 4-20 mA source 21 normally closed 23 normally de-energized 23 normally energized 23 normally open 23 65 SPM Flex Gas Detector 1998M0845 Revision 0.74 (will be released as Revision 1) July 2014 ©2014 Honeywell Analytics
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.7 Linearized : Yes Create Date : 2014:07:17 18:20:44-05:00 Creator : Adobe InDesign CC (Windows) Modify Date : 2014:07:30 14:47:34-05:00 XMP Toolkit : Adobe XMP Core 5.2-c001 63.139439, 2010/09/27-13:37:26 Metadata Date : 2014:07:30 14:47:34-05:00 Creator Tool : Adobe InDesign CC (Windows) Instance ID : uuid:075681d7-9c7a-4917-9439-bcaeb08532b8 Original Document ID : xmp.did:6D9690E42B98E011A2F9E9A72743D35B Document ID : xmp.id:3da3fa51-8d2f-4f4a-add9-6c2655d0de2c Rendition Class : proof:pdf Derived From Instance ID : xmp.iid:426285fd-7d97-514b-b7d7-7cd84ccb2027 Derived From Document ID : xmp.did:6904d152-635d-364b-9a56-3349d5023744 Derived From Original Document ID: xmp.did:6D9690E42B98E011A2F9E9A72743D35B Derived From Rendition Class : default History Action : converted History Parameters : from application/x-indesign to application/pdf History Software Agent : Adobe InDesign CC (Windows) History Changed : / History When : 2014:07:17 18:20:44-05:00 Format : application/pdf Producer : Adobe PDF Library 10.0.1 Trapped : False Page Count : 40EXIF Metadata provided by EXIF.tools