IRay Technology 02110113 Wireless Digital Flat Panel Detector User Manual 1

iRay Technology (Shanghai) Ltd. Wireless Digital Flat Panel Detector Users Manual 1

Contents

Users Manual-1

Download: IRay Technology 02110113 Wireless Digital Flat Panel Detector User Manual 1
Mirror Download [FCC.gov]IRay Technology 02110113 Wireless Digital Flat Panel Detector User Manual 1
Document ID2816526
Application ID/r/QROIXSurrMonmHt7+oA==
Document DescriptionUsers Manual-1
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize281.81kB (3522623 bits)
Date Submitted2015-11-19 00:00:00
Date Available2015-11-23 00:00:00
Creation Date2015-11-06 14:46:50
Producing SoftwareAcrobat Distiller 9.3.2 (Windows)
Document Lastmod2015-11-19 10:52:08
Document TitleMicrosoft Word - 037-201-02 Mars1717V User Manual A1 20151106
Document CreatorPScript5.dll Version 5.2.2
Document Author: dell

iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
Mars1717V User’s Manual
Authors
Dong Zhu
Doc. No.
037-201-02
Date
2015.11.06
Version
A1
Status
Released
Project
Mars1717V /037
037‐201‐02
A1
2015.11.06
1 / 141
Mars1717V User’s Manual
Author
Dong Zhu
Approver
037‐201‐02
A1
2015.11.06
Doc N0.
Version
Date
Page
iRay Technology (Shanghai) Ltd.
Inspector
2/141
Yimin Huang
Zhiqiang Fang
Verified Record
Department
Name
Signature
Quality
Wei PAN
Test-verification
Long Yu
Date
Department
Name
Signature
Date
Yong WANG
Pre-production
Marketing
Yao LU
Product
Lin Liu
Delivery Department
Department
Department
Department
Department
Department
Others
Document History
Ver.
Date
A0
2015.08.26
Revised
Checked
Approved
by
by
by
Dong Zhu
Yimin.Huang
Zhiqiang Fang
Description
First Released
Update the
A1
2015.11.06
Dong Zhu
Yimin.Huang
Zhiqiang Fang
information to the
customers about
FCC
Page 2
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
3/141
To Customers
Congratulations on your purchase of the Mars1717V Wireless Digital Flat Panel
(hereinafter referred to as Mars1717V) which is manufactured by iRay Technology (Shanghai)
Ltd. (hereinafter referred to as iRay).
At iRay, we strive to not only make the world-class products that deliver the best value possible
to our customers but also offer the highest quality of service and customer care. Please take time
to read through this user guide in order to utilize the product effectively. We hope you enjoy the
experience with iRay Mars1717V.
If you have any questions or suggestions, please feel free to contact us.
Service Office
Tel: +86 21 50720560 - 8059
Fax: +86 21 50720561
E-mail: service@iraychina.com
Location: 2F, Building 7, No.590, Ruiqing Rd, Pudong, Shanghai, China PC: 201201
About FCC
FCC Compliance
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference;
(2) This device must accept any interference received, including interference that may cause undesired operation.
Caution: Any changes or modifications not expressly approved by the party responsible for compliance
could void the user's authority to operate this equipment.
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no guarantee that interference will not occur in
a particular installation. If this equipment does cause harmful interference to radio or television reception,
which can be determined by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:

Reorient or relocate the receiving antenna.

Increase the separation between the equipment and receiver.

Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected.
Page 3
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
4/141

Consult the dealer or an experienced radio/TV technician for help.

FCC requires this product to be used indoors for the frequency range 5.15 to 5.25 GHz to reduce
the potential for harmful interference to co-channel Mobile Satellite systems.
Federal Communication Commission (FCC) Radiation Exposure Statement.
This EUT is compliance with SAR for general population/uncontrolled exposure limits in ANSI/IEEE
C95.1-1999 and had been tested in accordance with the measurement methods and procedures
specified in OET Bulletin 65 Supplement C. This equipment should be installed and operated contact
with the radiator & your body.
Notes on usage and management of the equipment
 Read all of the instructions in the user guide before your operation. Give particular attention to all safety
precautions.
 Only a physician or a legally certified operator should use this product.
 The equipment should be maintained in a safe and operable condition by maintenance personnel.
 Use only computers and image display monitors complying with IEC 60601-1 or IEC 60950-1. For details,
consult our sales representative or local iRay dealer.
 Use only the dedicated cables. Do not use any cables other than those supplied with this product.
 Request your sales representative or local iRay dealer to install this product
Caring for your environment
This symbol indicates that this product is not to be disposed of with your residential or
commercial waste.
Recycling iRay Equipment
Please do not dispose of this product with your residential or commercial waste. Improper handling of this
type of waste could have a negative impact on health and on the environment. Some countries or regions,
such as the European Union, have set up systems to collect and recycle electrical or electronic waste items.
Contact your local authorities for information about practices established in your region. If collection
systems are not available, call iRay Customer Service for assistance.
Disclaimer
1. iRay shall not be liable to the purchaser of this product or third parties for any damage, losse, or injury
incurred by purchaser or third parties as a result of fire, earthquake, any accident, misuse or abuse of this
product.
Page 4
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
5/141
2. iRay shall not be liable to any damage, loss, or injury arising from unauthorized modifications, repairs, or
alterations to this product or failure to strictly comply with iRay’s operating and maintenance instructions.
3. iRay shall not be liable for any damage or loss arising from the use of any options or consumable products
other than those dedicated as Original iRay Products by iRay Technology.
4. It is the responsibilities of the user/attending physicians for maintaining the privacy of image data and
providing medical care services. iRay shall not be responsible for the legality of image processing , reading
and storage nor it shall be responsible for loss of image data for any reason.
5. Information regarding specification, compositions, and appearance of this product is subject to change
without prior notice.
Copyright

All rights reserved

No part of this publication may be reproduced in any form or by any means without the written permission
of iRay. The information contained herein is designed only for use with iRay Mars1717V.
Trademarks

The iRay name and iRay logo are registered trademarks of iRay Technology (Shanghai) Ltd.
Page 5
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
6/141
Symbols and Conventions
The following symbols and conventions are used throughout the user guide.
This symbol is used to identify conditions under which improper use of the
product may cause death or serious personal injury.
This notice is used to identify conditions under which improper use of the
product may cause minor personal injury.
This notice is used to identify conditions under which improper use of the
product may cause property damage.
This is used to indicate a prohibited operation.
This is used to indicate an action that must be performed.
This is used to indicate important operations and restrictions.
This is used to indicate operations for reference and complementary
information.
Page 6
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
037‐201‐02
A1
2015.11.06
Doc N0.
Version
Date
Page
7/141
Labels and markings on the equipment
The contents of the labels and markings on iRay Mars1717V product are indicated below:
Caution: please refer to the instructions in the user manual.
This symbol is used to indicate that the equipment has passed CE testing and it
is followed by the CE number.
This symbol is used to identify the manufactuer’s series number which is after,
below or adjacent to the symbol. The series number of iRay products is
usually made of thirteen digits as shown below:
A1A2A3A4
C1C2 M DD Y XXX
Numerical Order
Year
Date
Month
Version
Product Code
Page 7
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
8/141
This symbol is used to indicate the name and address of the manufacturer.
This symbol is used to indicate the name and address of iRay authorized
representative in the European region.
This symbol is used to indicate consultation of the user guide for general
information.
Safety Signs: please refer to the user guide for safety instructions
Safety Signs:Dangerous Voltage
Stand-by
Handled with care
FPD is allowed to withstand 100 kg on the surface
This symbol is used to indicate the operational temperature limits.
Page 8
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
9/141
This symbol is used to indicate the storage temperature limits.
non-ionizing radiation
FCC
Federal Communications Commission certificate
Package symbol, fragile.
Package symbol, keep away from sunlight
Package symbol, keep dry
Package symbol, this symbol is used to indicate the humidity limits.
Package symbol, keep the equipment up right
Package symbol, do not roll the transportation package.
Package symbol, this symbol is used to indicate stacking limit number.
Page 9
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
10/141
Contents
To Customers
Symbols and Conventions
Labels and markings on the equipment
Contents
10
Safety Information
14
1.1
Safety precautions
14
1.2
Notes for Using
20
General Description
22
2.1
Scope
22
2.2
Model
23
2.3
Characteristic
23
2.4
Intended use
23
2.5
Standard Product Components
24
2.6
Optional Product Component
26
2.7
Components Description
27
2.8
Product Specification
30
2.8.1
Detector
30
2.8.2
Battery
32
2.8.3
Battery Charger
33
2.8.4
Power supply
34
2.8.5
Infrared Device (Optional)
34
2.8.6
AP Router (Optional)
34
2.8.7
Wireless Communication
34
2.8.8
Recommended Applicance Condition
35
2.8.9
Mechanical Outlines
36
2.8.10 Use Environment
38
Installation
39
3.1
Detector Installation
39
3.2
Battery Chargering Installation
42
Page 10
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
11/141
3.3
Software Installation
43
3.4
Panel Infrastructure
43
Operation
64
4.1
Main Operation
64
4.1.1
Software Mode
64
4.1.2
Inner Mode
67
4.1.3
Isync Plus Mode
69
4.2
Connection Build
71
4.3
Panel Configuration
72
4.4
Correction and Calibration Template Generation
72
4.4.1
Pre-offset Template Generation
72
4.4.2
Gain Calibration Template Generation
73
4.4.3
Defect Correction Template Generation
75
4.4.4
Lag Template Generation
78
4.5
Image Acquisition Continually
79
4.6
Image save
79
4.6.1
Multiple images
79
4.6.2
Single image
80
4.7
Image Check and upload
81
4.7.1
Local Image Check
81
4.7.2
Panel Image Upload
82
4.7.3
Panel Image Check
85
4.8
Defect Template Check and Modification
85
4.8.1
Defect Template Check
85
4.8.2
Defect Template Modification
86
4.9
Correction and Calibration Management
88
4.9.1
Correction and Calibration template synchronization
88
4.9.2
Correction and Calibration management
90
4.10
Sleep and Wake Up
92
4.10.1 Sleep
93
4.10.2 Wake Up
95
4.11
95
Firmware Update
4.11.1 MCU Update
95
4.11.2 FTP Server
97
4.11.3 Firmware Update
99
4.12
Short cut
101
Page 11
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
4.13
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
12/141
Software
102
4.13.1 Main GUI
102
4.13.2 Message Box
103
4.13.3 Configuration GUI
105
4.13.4 Correction and Calibration
113
4.13.5 Image Check and upload
115
4.13.6 Infrared Registration
116
Regulatory Information
118
5.1
Medical equipment safety standards
118
5.2
Guidance and manufacture’s declaration for EMC
119
5.3
Radio Frequency Compliance Information
123
5.3.1
FCC Compliance
124
5.4
Battery Safety Standards
124
5.5
Product Label
125
5.5.1
Detector
125
5.5.2
Battery
127
5.5.3
Battery Charger
128
Trouble Shooting
129
6.1
Cable Inspection
129
6.1.1
Ethernet cable
129
6.1.2
Date cable inspection
130
6.2
Adapter Inspection
130
6.2.1
Adapter connector definition
130
6.2.2
Inspection method
131
6.3
Battery Inspection
131
6.3.1
Battery pin definition
131
6.3.2
Inspect Method
132
6.4
Dock Inspection
132
6.4.1
Dock pin definition
132
6.4.2
Inspection method
133
6.5
Detector Main Unit Inspection
133
6.5.1
Get SDK and detector log
133
6.6
Power up failure inspection
135
6.7
Connection inspection
135
6.7.1
Connection failed in wired mode
135
Page 12
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
13/141
6.7.2
Connection Failed in Client Mode
136
6.7.3
Disconnect in AP Mode
136
6.8
Dark Image Acquisition After Exposure in Inner Mode
136
6.9
No Image Acquire after Exposure
137
6.10
Image Inspection
137
6.10.1 Artifact on bright image
137
6.10.2 Horizontal Line on bright image
137
6.10.3 Defective line
139
6.10.4 Defective point
139
6.10.5 TFT broken
139
Service Information
140
7.1
Product lifetime
140
7.2
Regular inspection and Maintenance
140
7.3
Repair
141
7.4
Replacement parts support
141
Page 13
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
14/141
1 Safety Information
1.1 Safety precautions
Follow these safeguards and properly use the equipment to prevent injury and damage to any
equipment/data.
WARNING
Installation and environment of use

Do not use or store the equipment near flammable chemicals such as alcohol, thinner,
benzene, etc.
If chemicals are spilled or evaporate, it may result in fire or electric shock through contact
with electric parts inside the equipment. Also, some disinfectants are flammable. Be sure to
take care when using them.

Do not connect the equipment with anything other than specified.
Doing so may result in fire or electric shock.

All the patients with active implantable medical devices should be kept away from the
equipment.
Power supply

Do not operate the equipment using any type of power supply other than the one
indicated on the rating label.
Otherwise, it may result in fire or electric shock.

Do not handle the equipment with wet hands.
You may experience electric shock that could result in death or serious injury.

Do not place heavy object such as medical equipment on cables and cords. Do not pull,
bend, bundle, or step on them to prevent their sheath from being damaged, and do not
alter them neither.
Doing so may damage the cords which could result in fire or electric shock.

Do not supply power to more than one piece of equipment using the same AC outlet.
Doing so may result in fire or electric shock.

Do not turn ON the system power when condensation has formed on the equipment.
Doing so may result in fire or electric shock.
Page 14
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual

Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
15/141
Do not connect a multiple portable socket-outlet or extension cord to the system.
Doing so may result in fire or electric shock.

To avoid the risk of electric shock, this equipment must only be connected to power
supply with protective earth.
Not doing so may result in fire or electric shock.

Securely plug the power cord into the AC outlet.
If contact failure occurs, or if metal objects come into contact with the exposed metal prongs
of the plug, fire or electric shock may result.

Be sure to turn OFF the power to each piece of equipment before connecting or
disconnecting the cords.
Otherwise, you may get an electric shock that could result in death or serious injury.

Be sure to hold the plug or connector to disconnect the cord.
If you pull the cord, the core wire may be damaged, resulting in fire or electric shock.
WARNING
Handling

Never disassemble or modify the equipment. No modification of this equipment is
allowed. Parts of the Mars1717V that are not serviced or maintained while in use with
the patient
Doing so may result in fire or electric shock. Also, since the equipment incorporates parts
that may cause electric shock as well as other hazardous parts, touching them may cause
death or serious injury.

Do not place anything on top of the equipment.
The object may fall and cause an injury. Also, if metal objects such as needles or clips fall
into the equipment, or if liquid is spilled, it may result in fire or electric shock.

Do not hit or drop the equipment.
The equipment may be damaged if it receives a strong jolt, which may result in fire or
electric shock if the equipment is used without being repaired.

Do not put the equipment and pointed objects together.
The equipment may be damaged. If so, the equipment should be used in bucky.
Page 15
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual

Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
16/141
Have the patient take a fixed posture and do not let the patient touch parts unnecessarily.
If the patient touches connectors or switches, it may result in electric shock or malfunction
of the equipment.
When a problem occurs

Should any of the following occurs, immediately unplug the power cord of adaptor or
battery, and contact your sales representative or local iRay dealer:
When there is smoke, an odd smell or abnormal sound.
When liquid has been spilled into the equipment or a metal object has entered through an
opening.
When the equipment has been dropped and damaged.
Maintenance and inspection

Please turn OFF the power of the equipment and unplug the power cord of adaptor
before cleaning.

NEVER use alcohol, ether and other flammable cleaning agent for safety. NEVER use
methanol, benzene, acid and base because they will erode the equipment.

DON’T dip the equipment into the liquid.

Please make sure that the equipment’s surface & plugs are dry before turning ON.
Otherwise, it may result in fire or electric shock.

Clean the plug of the power cord periodically by unplugging it from the AC outlet and
removing dust or dirt from the plug, its periphery and AC outlet with a dry cloth.
If the cord is kept plugged in for a long time in a dusty, humid or sooty place, dust around
the plug will attract moisture; this could cause insulation failure that may result in a fire.

For safety reasons, be sure to turn OFF the power to each piece of equipment when
performing inspections indicated in this manual.
Otherwise, electric shocks may occur.
CAUTION
Page 16
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
17/141
Installation and environment of use

Do not install the equipment in any of the locations listed below. Doing so may result in
failure, malfunction, equipment falling, fire or injury.
Close to facilities where water is used
Where it will be exposed to direct sunlight
Close to the air outlet of an air-conditioner or ventilation equipment
Close to heat source such as a heater
Where the power supply is unstable
In a dusty environment
In a saline or sulfurous environment
Where temperature or humidity is high
Where there is freezing or condensation
In areas prone to vibration
On an incline or in an unstable area

Take care that cables do not become tangled during use. Also, be careful not to get your
feet caught by cable.
Otherwise, it may cause a malfunction of the equipment or injury of the user due to tripping
over the cable.

Non-medical equipment such as the battery charger, access point and IR data
communication unit cannot be used in patient’s vicinity.
Power supply

Always connect the three-core power cord plug to a grounded AC power outlet.

To make it easy to disconnect the plug at any time, avoid putting any obstacles near the
outlet. Otherwise, it may not be possible to disconnect the plug in an emergency.

Be sure to ground the equipment to an indoor grounded connector. Also, be sure to
connect all the grounds for the system to a common ground.

Do not use any power source other than the one provided with this equipment.
Page 17
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
18/141
Otherwise, fire or electric shock may be caused due to leakage.
Handling

Do not spill liquid or chemicals onto the equipment. In case the patient is injured, it is
not allowed to contact with blood or other body fluids.
Doing so may result in fire or electric shock.
In such a situation, protect the equipment with a disposable cover as necessary.

Turn OFF the power and pull out the plug to each piece of equipment for safety when
not used.
CAUTION
Handling

Handle the equipment carefully.

Do not submerge the equipment in water.

The internal image sensor may be damaged if something hits against it or it is dropped.
If the equipment is dropped, the drop sensor inside will turn red and the equipment will
not be warranted by iRay.

Do not place excessive weight on the equipment.
Otherwise, the internal image sensor may be damaged and image may be incorrect.
Page 18
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
19/141

Uniform load:100 kg over the
Local load:100 kg on an area
whole area of the detector
4 cm diameter.
surface.

Be sure to use the equipment on a flat surface so it will not bend. Otherwise, the internal
image sensor may be damaged. Be sure to securely hold the detector while using it in
upright positions. Otherwise, the detector may fall over, resulting in injury to the user
or patient, or may flip over, resulting in damage to the inner device.
Keep the same load (same pressure) on the detector when acquiring the image. Or the image
will be incorrect.
CAUTION

Do not close to fire, do not use in high temperature

Do not invert positve and negative pole

Do not contact with metal in case of short circuit

Do not insert sharp objects into battery

Do not beat battery

Do not stand on battery
Page 19
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
20/141

Do not use battery out of rules

Do not dispose battery or change internal structure

Do not submerge battery in water, please keep dry in storage and do not contact with water in use

Please charge battery with charger following IEC60601-1 & IEC62133 Standards provide by us

Do not mix battery with ones not provided by our company

Do not charge battery with broken charger.
1.2 Notes for Using
When using the equipment, take the following precautions. Otherwise, problems may occur and the
equipment may not function correctly.
Before exposure

Be sure to check the equipment daily and confirm that it works properly.

Be sure there be a battery installing on the Mars1717V to avoid the power off suddenly.

Sudden heating of the room in cold areas will cause condensation to form on the equipment. In
this case, wait until the condensation evaporates before performing an exposure. If the equipment
is used while condensation is formed on it, problems may occur in the quality of captured images.
When an air-conditioner is used, be sure to raise/lower the temperature gradually so that a
difference of temperature in the room and equipment does not occur, to prevent condensation.

The detector should warm up for 15 minutes before exposure or updating the gain map or defect
map.
During exposure

Do not move the power or Ethernet Cables during exposure, or it may cause image noise or artifacts,
even incorrect images.

Do not use the devices near the equipment generating a strong magnetic field. Otherwise, it may
cause image noise, artifacts or even incorrect images.
Disinfection and Cleaning

After every examination, wipe the patient contact surfaces of the detector using disinfectants such
as ethanol, to prevent the risk of infection. For details on how to sterilize, consult a specialist.

Do not spray the detector directly with disinfectants or detergents.

Wipe it with a cloth slightly damped with a neutral detergent. Do not use solvents such as alcohol,
thinner, benzene, acid and base. Doing so may damage the surface of the equipment.

It’s recommended to use a waterproof non-woven cover as the isolated layer between detector and
Page 20
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
the blooding patient.
Page 21
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
21/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
22/141
2 General Description
Mars1717V is a cassette-size wireless X-ray flat panel detector based on amorphous silicon thin-film
transistor technologies. It is developed to provide the highest quality of radiographic image, which contains
an active matrix of 3072×3072 with 139um pixel pitch. Detectors’ scitinator has two options which are
Standard GOS(Gadolinium Sulfoxylate) and CsI(CaesiumIodide). However the most great improvement is
Mars1717V supports wireless communication between detectors and PC. Mars1717V’s power supply
includes battery. Mars1717V can be used as a real portable panel.
2.1 Scope
This manual contains information about the iRay Mars1717V. All operators must read and understand this
manual before using equipment. All information in this manual, including the illustrations, is based on
equipment prototype. If configuration of your equipment does not have any of these items, information about
these items in the manual does not apply to your equipment.
Page 22
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
23/141
2.2 Model
Mars □ □ □
PSI- GOS scintillator screen, VSI- CsI scintillator screen
Product Application:cassette size detector
Product dimension:1717,17inch×17inch
Product serials:wireless digital flat panel detector serials
Product Type: Battery-KV---------Rechargeable lithium battery
Product Type: Charger-KV--------Battery charger
2.3 Characteristic

Wireless static Flat Panel Detector used for general radiography.

17 × 17 inch

Sync-Shot exposure trigger

GOS or CsI scintillation screen.

Easy to change the cable and upgrade firmware.

Battery recycling
2.4 Intended use
This equipment provides digital X-ray imaging for diagnosis of disease, injury, or any applicable health
problem. The image is obtained as the result of X-ray passing through the human body and detected by the
equipment. This device is intended to be used in the holder or bucky which is well insulated to panel. The
holder or the bucky is well grounded. This panel is not intended for directly contacting with patient
iRay will provide equipment and software support for integration of system. The length of both Ethernet
Cable and DC Power Cable cannot exceed 3.5 m. or the impedance of protective earth connections may
exceed the safety threshold.
This panel is not intended for mammography or dental applications,and prohibited for pregnant women and
children.
According to the Mars1717V intended use and the result of risk management, identifying and describing the
essential performance as the following:
a)
To get imaging of dark field, the Mars1717V shall be not influenced to the imaging acquisition
Page 23
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
b)
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
24/141
To keep the data transmission function, the Mars1717V shall be not influenced to the data and signal
transmission
2.5 Standard Product Components
Mars1717V comes with power supply both 24V DC and battery package. Once powered on, it would build a
connection with PC through Ethernet cable or Wireless connection.
Item
Description
Mars1717V Detector
1 pcs
Main Unit
Medical Adapter for

Detector and

Battery Charger
2 pcs
DC 24V
2 pcs
Battery-KV
Battery Pack
Page 24
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
25/141
Ethernet Cable
1 pcs
3.5 m
Gigabit Ethernet Cable
1 pcs
3m
AC Power Cable
2 pcs
DC Power cable
1 pcs
3.5 m
Battery Charger
1 pcs
Page 25
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
26/141
1 pcs
Gain correction data
Defect correction map
SDK
Manual
CD-Rom
2.6 Optional Product Component
Item
Description
Wireless AP Device
1 pcs
Infrared Device
1 pcs
Page 26
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
2.7 Components Description
2.7.1.1
Detector
External Signals Input
Control Panel
Control Panel
Item
Name
Description
DC Input Interface
24V DC input
Ethernet Interface
Gigabit Ethernet Wire
Detector Indicator
Detector indicator of control panel
Power Button
Power button of control panel
Page 27
037‐201‐02
A1
2015.11.06
27/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
Infrared Window
Infrared device window
Antenna
Antenna
Handle
For the panel carried
Maintenance Cover
For service engineer to maintenance
Battery Lock
The lock button for detaching battery
Detector Label
Product information.
Switch
AP/Client Mode Switch
Carbon Film
Panel Carbon Film, have biocompatibility
2.7.1.2
Battery
Item
Name
Description
Battery Label
Battery Interface
8 Pin Battery connector
The location pin
Indicator
Installation direction indicator
Page 28
037‐201‐02
A1
2015.11.06
28/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
2.7.1.3
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
29/141
Battery Charger
Item
Name
Description
Battery Interface
8 Pin Battery connector
Capacity Indicator
The indicator definition is as follow
Power Indicator
The indicator definition is as follow
Hand Pull Position
The limit ball plug
DC Jack
24V DC input
Power indicator definition:
Power Indicator
Lighting Status
Operating Status
OFF
No external DC adaptor input
GREEN ON
External DC adaptor input
The battery charging capacity indicator definition:
X Group Indicator
Lighting Status
Operating Status
I, II and III grid OFF
No battery Insert
Page 29
Mars1717V User’s Manual
I grid blinking
037‐201‐02
A1
2015.11.06
Doc N0.
Version
Date
Page
iRay Technology (Shanghai) Ltd.
30/141
Battery Insert with capacity ≤30%,charging
II and III grid OFF
II grid blinking
Battery Insert with capacity >30% and ≤60%,
I and III grid OFF
charging
III grid blinking
Battery Insert with capacity >60% and ≤95%,
I and II grid OFF
charging
Battery Insert with capacity >95% and
I and II grid OFF
charging, when capacity = 100%, charging can
III grid ON
2.7.1.4
stop automatically
Power Supply
Mars1717V supports both DC Power and Battery package input.
2.7.1.5
Infrared Device
Mars1717V does not include Infrared Device. Users can select by themselves, however some basic
requirements should be followed.
2.8 Product Specification
2.8.1 Detector
2.8.1.1
Basic
Item
Specification
Model
Mars1717V-PSI (GOS)
Mars1717V-VSI (CsI)
Image Sensor
a-Si (Amorphous Silicon) TFT
Pixel Size
139 μm
Effective Array
3072 x 3072
Effective Area (H x V)
427 x 427 mm
Fill Factor
60%
Greyscales
14bit
Spatial Resolution
2.8 Lp/mm (Standard GOS)
Page 30
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
31/141
3.1Lp/mm (CsI)
Image Acquisition Time (Wired)
Preview Acquisition Time : 6 sec
Processed Acquisition Time : 10 sec. (including Preview Time)
Image Acquisition Time (Wireless)
Preview Acquisition Time : 9 sec.
Both AP mode and Client mode
Processed Acquisition Time : 16 sec. (including Preview Time)
Cycle Time
Min. 13s @Wired;Min.16s @Wireless
Power Consumption
Max. 16W
Dimension (L × W × H)
460 x 460 x 15.2 mm
Weight (with one battery)
Mars1717V-PSI: 4.3 kg without battery, 4.6 kg with battery
Mars1717V-VSI: 4.5 kg without battery, 4.8 kg with battery
Image Transfer
Wired : Gigabit Ethernet(1000BASE-T)
Wireless : IEEE802.11a/b/g/n
Data Transmission Rate (Wireless)
802.11b : Max. 11Mbps
802.11a/g : Max. 54Mbps
802.11n : Max. 300Mbps (MIMO 2x2)
X-ray Energy
2.8.1.2
40kV to 150kV
MTF
The MTF with GOS should meet the following table
2.8.1.3
Spatial frequency (lp/mm)
MTF (GOS)
MTF(CsI)
1.0
0.39
0.60
2.0
0.12
0.31
3.0
0.04
0.15
DQE
The DQE with GOS should meet the following table
Page 31
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
037‐201‐02
A1
2015.11.06
Doc N0.
Version
Date
Page
32/141
Spatial frequency (lp/mm)@RQA5
DQE (GOS)@3.2uGy
DQE (CsI)@2.5uGy
0.30
0.52
1.0
0.15
0.39
2.0
0.05
0.24
3.0
0.01
0.16
2.8.2 Battery
Page 32
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Item
Model
Rated Capacity
Nominal Voltage
Charge Voltage
Battery-KV
Min. 3950mAh, Typ. 4180mAh @ Discharge 0.2C
10.8V
12.6±0.05V
8.25V
Charging Method
CC-CV
Charge 0℃-+45℃, Discharge -20℃-+60℃
1 month -20℃-+45℃
Storage Temperature
3 month 0℃-+30℃
6 month 5℃-+20℃
Relative Humidity
Dimension (L × W × H)
Weight
2.8.3
65±20%
210 x 115 x 7.5 mm
0.29 kg
Battery Charger
Item
Model
Simultaneous Charging
Full charging time
Rated power supply
Dimension (L × W × H)
Weight
037‐201‐02
A1
2015.11.06
Specifications
Discharged End Voltage
Operating Temperature
Doc N0.
Version
Date
Page
Specifications
Charger-KV
2 battery packs
2 hours
24V(DC)
300 x 263 x 42 mm
1.26 kg
Page 33
33/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
2.8.4
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
34/141
Power supply
Mars1717V supports both DC Power and Battery package input.
Item
DC Power
Battery Package
2.8.5
Specifications
24V(DC), 0.8A
10.8V(DC),1.78A
Infrared Device (Optional)
Mars1717V does not include Infrared Device. Users can select by themselves, however some basic
requirements should be followed.
Item
IRDA Protocol
Compliant with IrDA V1.0 and V1.1
USB
Compliant with USB V2.0 and V1.1
Data Rate
2.8.6
Specifications
Max. 4Mbps
AP Router (Optional)
Mars1717V do not include AP Router. Users can select AP Router as they wish, however specification below is a
requirements.
Item
Specifications
Wireless Standard
IEEE 802.11 a/b/g/n
Frequency Range
2.412 ~ 2.4835 GHz and 5.15 ~ 5.85 GHz
802.11b : Max. 11Mbps
Wireless Data Rate
Wired Data Rate
2.8.7
802.11a/g : Max. 54Mbps
802.11n : Max. 300Mbps (MIMO 2x2)
Max. 1Gbps
Wireless Communication
Item
Wireless Standard
Description
IEEE802.1a/b/g/n
2.4G: 2.412 ~ 2.4835 GHz
Frequency Range
11: (Ch. 1-11) – United States
13: (Ch. 1-13) – Europe
Page 34
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
35/141
14: (Ch. 1-14) – Japan
5G: 5.15 ~ 5.85 GHz
12: United States
19: Europe
8: Japan
802.11b : Max. 11Mbps
Data Transmission Rate
802.11a/g : Max. 54Mbps
802.11n : Max. 300Mbps (MIMO 2x2)
802.11b:
CCK, DQPSK, DBPSK
Modulation
802.11a/g:
64 QAM, 16 QAM, QPSK, BPSK
802.11n:
64 QAM, 16 QAM, QPSK, BPSK
Transmission Power
2.8.8
Max.17dBm
Security
WPA, WPA-PSK, WPA2, WPA2-PSK, WEP 64bit & 128bit
Antenna
2 Dual Band internal antenna
Recommended Applicance Condition
Item
Operating System
CPU
Memory
Hard Disk
LAN Card
Description
Windows XP/7 32/64bit
Intel Core i73.6G
4G DDR3
160 G
Intel Pro EXP9301CT PRO
Gigabit Network Adapter with PCIe interface
Page 35
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
2.8.9
Mechanical Outlines
Page 36
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
36/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Page 37
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
37/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
38/141
2.8.10 Use Environment
Operating
Storage
Temperature
Temperature
change
Humidity
AtmosphericPr
essure
Pressure Change
5~30℃
<1k/min
45~85% RH
700~1000hPa1
0~70% RH
<10kp/min
(1kp=1.0197E-5Pa)
-20~60℃
<1k/min
45~85%RH
700~1000hPa1
0~70% RH
The Mars1717V serial detectors shall operate at an altitude specified not more than 3000m.
Page 38
<10kp/min
(1kp=1.0197E-5Pa)
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
39/141
3 Installation
3.1 Detector Installation
3.1.1.1
Attach Battery Pack
Mars1717V can be powered on by both battery package and DC power. Once battery package is inserted or
DC power is on, Detectors would be activated immediately. If none of battery and DC power is on,
Mars1717V would power off. Please see below for battery installation.
Make sure that the connectors on the battery package
are pointed to the cave in battery compartment.
Slide battery package into battery compartment
(Make sure battery capacity overpass 10%).
Slide the battery lock lever.
3.1.1.2
Attach DC Power
Please see below for DC power installation.
Page 39
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Connect one end of DC Power Cable to the Medical
Adapter
As figure is power interface and Ethernet interface
Connect another end of DC Power Cable to the DC
input of the detector.
If user want to use wired Ethernet, connect the
Ethernet cable to Ethernet interface.
3.1.1.3
Booting Up
On the control panel, user can press power button to power on/off.
If detector is powered off, users can press the button
for 4 seconds to power on the detector when battery
is inserted and battery capacitor is not less than 10%,
or direct current power is connected.
If detector is powered on, users can press the button
for 4 seconds to shut down the detector. On the other
hand, it can also be used as reset internal control IC
when button is active for 8s.
Page 40
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
40/141
Doc N0.
Version
Date
Page
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
037‐201‐02
A1
2015.11.06
41/141
After booting up the detector, use can check the status LED indicator of detector.
Operating Status
Power Indicator
Lighting Status
Battery
Operating
Capacity
DC Input
OFF
Power OFF
Orange ON
Power ON
≤10%
NO

Green ON
Power ON
Input

Orange Fast
Blinking
Orange Slow
Blinking
Green Fast
Blinking
Green Slow
Blinking
OFF after Green
ON with 1 sec.
OFF after Orange
ON with 1 sec.
Battery Capacity >10%, NO DC
No Battery with DC Input
Power OFF
≤10%
YES
Power ON
≤10%
YES
Power OFF
>10%
YES
Power ON
>10%
YES
Power OFF
>10%
NO
Power OFF
≤10%
NO
Link indicator is as table:
Link Indicator
Lighting Status
Description
OFF

Panel shut down

wired connection broken and wireless
connection not ready

Blue blinking
Client mode, wireless connection is ready
for connection, but not connected

Blue ON
Page 41
Client mode, wireless connection is built
Doc N0.
Version
Date
Page
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual

037‐201‐02
A1
2015.11.06
42/141
AP mode, wireless AP is ready for
connecting
Green ON
Green blinking

Wired Connection is built

Panel Initialization

Infrared configuration
Status indicator is as table:
Status Indicator
Lighting Status
Description

Panel shut down

Panel is idle
Green ON

Data Transmission
Orange blinking

Fatal Error
Orange ON

Panel Initialization
OFF
AP/Client indicator is as table:
AP/Client Indicator
Lighting Status
Description
1.
2.
OFF
Wireless connection ok,connection mode is
Green ON
AP
Wireless connection ok,connection mode is
Blue ON
Client
3.2 Battery Chargering Installation
Operating
Network connection error;
Network connection is wired connection;
Figure
Page 42
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
Unload Battery from battery charger.
Insert battery into battery charger. Note align the
interface position as figure.
Press the battery to the bottom of battery
compartment.
3.3 Software Installation
In the case of IDemo doesn’t work, please install following VC redistribute package.
3.4 Panel Infrastructure
Mars1717V supports three connection modes as follows:
1) Wired connection mode
Page 43
037‐201‐02
A1
2015.11.06
43/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
2) Wireless Client Mode
3) Wireless AP Mode
To build connection between workstation and Panel, User should follow steps below.
3.4.1.1
Wired Mode
To complete Wired connection configuration, users have to finish actions listed below.
Configuration of Ethernet Card
To begin configuration of Ethernet Card, users should finish 31.1.2
Page 44
037‐201‐02
A1
2015.11.06
44/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Connect detector to PC with Ethernet
Cable
Open Network configuration
Open Local network configuration
Page 45
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
45/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
open IPV4 setting
IP setting
Network mask setting
IP address: 192.168.8.188
Subnet mask: 255.255.255.0
Open SDK and start connection
IP and port setting
3.4.1.2
IP: 192.168.8.188
Port: 28000
Wireless Client mode
To complete Wireless Client mode configuration, users have to finish actions listed below.
Configuration of External wireless AP
Connect one end of Gigabit Ethernet
Cable to PC,Connect another end to
LAN port of External wireless AP
Page 46
037‐201‐02
A1
2015.11.06
46/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Open Network configuration
Open Local network configuration
Page 47
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
47/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
open IPV4 setting
IP setting
Network mask setting
Select “Obtain an IP address automatically”
Open browser and type 192.168.1.1
Log into external wireless AP
Page 48
037‐201‐02
A1
2015.11.06
48/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
49/141
Wireless setup
Configure 2.4GHz wireless network
SSID: NETGEAR_BIG_24
Security: WPA2-PSK
Password: 12345678
Channel: [Please check the current Wi-Fi environment, and choose a
relatively clean channel]
Configure 5GHz wireless network
SSID: NETGEAR_BIG_50
Security: WPA2-PSK
Password: 12345678
Channel: [Please check the current Wi-Fi environment, and choose a
relatively clean channel]
Page 49
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
LAN setup
Configure LAN IP address
IP address: 192.168.8.1
Subnet Mask: 255.255.255.0
External Wireless AP Reboot
Apply above settings and reboot your wireless router.
Recover Local Network IPv4 setting of
PC wired Ethernet interface
IP setting
IP address: 192.168.8.188
Network mask setting
Subnet mask: 255.255.255.0
Configuration of detector
Either Wired Cable or Infrared device can be used to configure detector in wireless client mode.
Page 50
037‐201‐02
A1
2015.11.06
50/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
51/141
a.To start configuration with wired cable. It is necessary to finish 3.2.2.1, then proceed to the steps below.
Click “Configure” in IDemo
Click “Network setting”
Click “Add”
Page 51
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
52/141
Change SSID and Key,
Click “Apply”
Click “write Config to FPD”
Do not remove wired cable until “write
to FPD” button recovers
Since we have chosen default SSID and password, it would connect to wireless AP immediately after powered on
next time.
B. To start Infrared configuration. Please see below
Connect Infrared device with
Workstation
Page 52
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Start IrDARegister.exe
Click “
” to open wifi setting
Page 53
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
53/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Change SSID and password,do not
select AP mode
Click “Apply”
Click”
”
Page 54
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
54/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Click”Start”
Point Infrared device to detector’s
infrared interface
Page 55
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
55/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
Do not click”Exit” until success
Disconnect Infrared device from PC
3.4.1.3
Wireless AP mode
To complete Wired connection configuration, user has to finish actions listed below.
Configuration of detector
Either Wired cable or Infrared device can be used to configure detector wireless AP mode.
a.To start wired cable configuration, users should finish 3.4.1.1, then proceed to the steps below.
Click “Configure” in IDemo
Page 56
037‐201‐02
A1
2015.11.06
56/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Click “Network setting”
Select AP mode
Change SSID and password setting
Click “write Config to FPD”
Do not remove wired cable until “write
to FPD” button recovers
Page 57
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
57/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
58/141
Since we have chosen default SSID and password, it would connect to wireless AP immediately after powered on
next time.
b.To start Infrared configuration, please see below
Connect Infrared device with PC
Start IrDARegister.exe
Page 58
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Click “
” to open wifi setting
Select “AP mode”
Page 59
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
59/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Change SSID and password and other
parameter
Click “Apply”
Click”
”
Page 60
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
60/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Click”Start”
Do not click”Exit” until success
Page 61
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
61/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Configuration of external wireless card
Open Wireless signal List
Select SSID which belongs to
Detectors, input password and log into
system
Open wireless card configuration
Page 62
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
62/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
open IPV4 setting
IP setting
IP address: 192.168.8.188
Network mask setting
Subnet mask: 255.255.255.0
Open SDK and start connection
IP and port setting
IP: 192.168.8.188
Port: 28000
Page 63
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
63/141
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
64/141
4 Operation
Mars1717V provides SDK for users to integrate detector into their DR system. Additionally, it also provides
an application for demonstration, i.e. Idemo. User can use Idemo to control detector without DR system.
4.1 Main Operation
To Acquire X-ray image is the main operation of Mars1717V. Most importantly, detector should build
synchronization with X-ray generator.Mars1717V is born with two ways to acquire X-ray image, that is
Software Mode, Inner Mode and Isync Plus Mode.
4.1.1
Software Mode
4.1.1.1
Block Diagram
Software mode is the basic way to acquire X-ray image. Please see figure below for general feature
Workstation is a host PC device installed with idemo and SDK. Chapter 3.1 has described how to establish
connection between detectors and workstation. In software mode, workstation does not have to control X-ray
generator. Users would decide when to shoot X-ray.
Page 64
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
4.1.1.2
037‐201‐02
A1
2015.11.06
Doc N0.
Version
Date
Page
65/141
Work flow
Click “Prep”
Panel preparation,Wait until the warning message change from “Exposure
Prohibit” to “Exposure Enable”
Shoot X-ray at any time. However the longer time you wait. The worse
image would be. So please shoot once X-ray generator is ready.
After shooting, Click”Acquire”
Wait for image uploaded
4.1.1.3
Timing Setting
To set a clear scenario for program, see diagram below for details
Clear
Process
Detector
Workstation
Xray
Generator
1.
Exposure
Prohibited
Exposure
Acquisition
10
Anode
Rotate
11
Exposure Enable
X ray
shooting
Workstation receives “prep” request, send command “Clear” to panel.
Page 65
12
Exposure
Prohibited
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
2.
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
66/141
Panel receives “clear” from workstation, start detector internal clear cycle. At the same time, detector
would tell workstation “Exposure Prohibited”.
3.
Detector finished ”Clear” action and send a message reminding “Exposure Enable”
4.
Workstation shows “Exposure Enable” on the iDemo’s message bar to tell user shoot X-ray now.
5.
User triggers X-ray generator to initialize and do anode rotation to prepare for X-ray shooting.
6.
X-ray generator finishes preparation for X-ray shooting and reminds user to shoot.
7.
X-ray generator starts releasing X-ray
8.
X-ray generator finishes X-ray shooting.
9.
Workstation receives “Acquire” request, send command “Data Acquisition” to panel.
10. Panel receives “Data Acquisition” from workstation, start data acquisition operation.
11. Panel completes image acquisition and begins to send data to workstation.
12. Workstation receives all image data from panel.
If Hardware Pre-offset and Hardware calibration is selected, image is the final image.
If Software Pre-offset and Software Calibration is selected, image would be the raw image, workstation
would finish image processing work and show on screen.
If Hardware Post offset and Hardware calibration is selected, image got would be preview image (2x2
binning). After step12, panel would do another dark image acquisition. With both light image and dark
image, panel completes all the correction and calibration process. Finally, panel uploads processed image to
workstation.
If Software Post offset and Software calibration is selected, image got would be preview image (No binning).
After step12, Workstation sends another “clear Acquire” command to panel, panel finishes a dark image
acquisition and uploads dark raw image to workstation. With both light image and dark image, workstation
completes all the correction and calibration process. Finally, corrected image shows on screen.
Page 66
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
67/141
4.1.2 Inner Mode
4.1.2.1
Block Diagram
Workstation (Software )
HVG
Tube
Workstation is a host PC device installed with iDemo and SDK. Chapter 3 has described how to establish
connection between panels and workstation. In inner mode, workstation does not control X-ray generator.
Users would decide when to shoot X-ray.
4.1.2.2
Work Flow
Click ”Prepare”
Panel preparation,Wait unitl the warning message change from “Exposure
Prohibit” to “Exposure Enable”
Shoot X‐ray in exposure window. However the longer time user waits, the
worse image would be. So please shooting once X‐ray generator is ready.
After shooting , X‐ray in panel would trigger panel to start image.
Wait for image uploaded
Page 67
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
4.1.2.3
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
68/141
Timing Setting
To set a clear scenario for program, see diagram below for details
1. Workstation receives “prep” request and sends “Clear” to panels.
2. Panel receives “clear” from Workstation, start clear operation. Meanwhile, panel would send “Exposure
Prohibited” to Workstation.
3. Panel finishes “Clear” operation and send “Exposure Enable” to Workstation.
4. Workstation shows “Exposure Enable” on the iDemo’s message bar to tell user shoot X-ray.
5. User triggers X-ray generator to initialize and do anode rotation to prepare for X-ray shooting
6. X-ray generator finishes preparation and reminds users.
7. X-ray generator begins releasing X-ray
8. X-ray generator finishes X-ray shooting.
9. X-ray sensor in panel triggers panel to start image acquisition operation.
10. Panel completes image acquisition and begins to send data to Workstation.
11. Workstation starts receiving image data from panel.
12. Workstation receives all image data from panel.
If Hardware Pre-offset and Hardware calibration is selected, image got is the final image.
If Software Pre-offet and Software Calibration is selected, image got would be raw image, workstation
would finish image processing work and show on screen.
If Hardware Post offset and Hardware calibration is selected, image got from detector would be preview
image (2x2 binning). After step12, Detector would do another dark image acquisition. With both light image
Page 68
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
69/141
and dark image, detector completes all the correction process. Finally, detector uploads corrected image and
workstation shows on screen.
If Software Post offset and Software calibration is selected, image got from panel would be preview image
(No binning). After step12, Workstation sends another “clear Acquire” to panel, panel would do dark image
acquisition and uploads dark raw image to workstation. With both light image and dark image, workstation
completes all the correction process. Finally, corrected image shows on screen.
4.1.2.4
Abnormal Action
Action1: after Step4, if user wants to cancel this exposure cycle, Idemo provides an “Abort Exp” function to
close exposure windows. However, Idemo allows user to click “Abort Exp” until Workstation receives first
image.
Action2: after Step4, if user does not shoot X-ray in exposure windows, panel would close exposure
windows automatically and send a message to workstation that waiting for X-ray shooting is overtime.
Meanwhile, panel would also start image acquisition. After image acquisition, panel sends image to
workstation.
4.1.3 Isync Plus Mode
4.1.3.1
Block Diagram
Page 69
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
70/141
Workstation is a host PC device installed with iDemo and SDK. Chapter 3 has described how to establish
connection between panel and Workstation. In Isync Plus mode, User doesn’t interact with Workstation.
After shooting, images would be shown on screen immediately.
4.1.3.2
Work Flow
Check SDK to make sure panel is ready
Shoot X-ray
X-ray sensor in panel triggers panel to stop flushing the panel
After fixed time, panel starts image acquisition
Wait for image uploaded
4.1.3.3
Timing Setting
Workstation
Panel
X ray
Generator
Anode Rotate
Ray
1. X-ray generator is ready for X-ray shooting and begins to release X-ray.
2. Workstation receives “Exposure Prohibited” from Panel.
3. Panel starts uploading Pre-dark image and Light image to Workstation for preview. If hardware offset is
selected, panel would do offset first, and then upload preview image (2X2 binning).
Page 70
iRay Technology (Shanghai) Ltd.
Mars1717V User’s Manual
Doc N0.
Version
Date
Page
037‐201‐02
A1
2015.11.06
71/141
4. Panel starts uploading Post-dark image to Workstation. If hardware offset is chosen, panel would do
correction and calibration first, then upload processed image to Workstation.
5. Workstation receives “Exposure Enable” from Panel.
4.2 Connection Build
Click “Start”
Input IP address and port number. The IP
address should be the same as the IP
address of the network card connected
with panel. The port should use the
default value of 28000;
Click “Connect”;
Note:
1. Once changing connection from different network card, user must re-connect panel with different IP
address.
2. Switching between wired and wireless connection does not need any extra operation.
Page 71

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : Yes
Author                          : dell
Create Date                     : 2015:11:06 14:46:50+08:00
Modify Date                     : 2015:11:19 10:52:08+08:00
Has XFA                         : No
XMP Toolkit                     : Adobe XMP Core 4.2.1-c043 52.372728, 2009/01/18-15:08:04
Format                          : application/pdf
Title                           : Microsoft Word - 037-201-02 Mars1717V User Manual A1 20151106
Creator                         : dell
Creator Tool                    : PScript5.dll Version 5.2.2
Metadata Date                   : 2015:11:19 10:52:08+08:00
Producer                        : Acrobat Distiller 9.3.2 (Windows)
Document ID                     : uuid:bba11b2b-a51d-431f-aa4c-70d961dc05cd
Instance ID                     : uuid:f1c7bb73-03e8-4694-9c6e-8144f4ae044f
Page Count                      : 71
EXIF Metadata provided by EXIF.tools
FCC ID Filing: 2ACHK-02110113

Navigation menu