Microlife TWIN200-1 WatchBP Professional User Manual

Microlife Corporation WatchBP Professional

User Manual

EN 1Simultaneous dual-arm measurement and  auscultatory function for accurate office blood pressure measurement.Instruction Manual
WatchBP Office combines the advantages of an automatic blood pressure device and hybrid sphygmomanometer and is designed to provide convenient, accurate and reliable in-office blood pressure measurement following recommended guidelines of the European Society of Hypertension (ESH) 1, American Heart Association (AHA), and World Health Organization (WHO). 1 O’Brien E, Asmar R, Beilin L, Imai Y, et al. European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure measurement. European Society of Hypertension Working Group on Blood Pressure Deviceing. J Hypertens 2003;21:821-848.
ENProduct descriptionName of parts and display ................................. 3–4 Components and accessories ............................. 5–6Before using WatchBP OfficeSelecting and fitting the correct cuff .................. 7–8Three operation modes«AUSCULTATION» Mode .......................................... 9 «1st VISIT» Mode ............................................ 10–11 «FOLLOW-UP» Mode .............................................12Taking measurements using WatchBP Office«AUSCULTATION» Mode .................................. 13–14 «1st VISIT» Mode ............................................ 15–17 «FOLLOW-UP» Mode ....................................... 18–20Special FunctionsHiding measurement values ................................ 21Mean arterial pressure and pulse pressure  ......... 22Taking fewer than three measurements  ............ 23Skipping the 60-second countdown .................... 23Setting inflation pressure ..................................... 24Setting measurement intervals ........................... 24Viewing Measurements«AUSCULTATION» Mode  ....................................... 25 «1st VISIT» and «FOLLOW-UP» Mode ............. 25–26Controlling the WatchBP Office WirelesslySetting up the Bluetooth receiver .................. 27–28 Installation of the software program ................... 29 Controlling the device wirelessly  .................. 29–30AppendixRechargeable battery and power adaptor  .......... 31 Troubleshooting ................................................... 32 Error messages ............................................... 33–34 Safety, care, accuracy test and disposal ......... 35–36 Technical specifications ....................................... 37Table of Contents
3LRmmH gmmHgSYSmmHgDIASYSDIAmmHgPUL/minAUSCULT  1st  FOLLOW UPL/R3 sec.INTERVAL3 sec.7 sec.HIDE7 sec.I/O switch Bluetooth Switch Main Adaptor Socket Battery Compartment Cuff Socket (L)Cuff Socket (R)Recharge Indicator Display M Button (Memory) Mode Switch Start/Stop ButtonDC 6V 600mADC 6V 600mAProduct descriptionName of parts
4ENMean Arterial Pressure (MAP)Pulse Pressure (PP)Mode IndicationExternal PowerBluetoothRest and Arm Position ReminderBattery SymbolHide DisplayMemory ModePulse RateHeart BeatL/R CuffValue for Left ArmSystolic ValueDiastolic ValueValue for Right ArmSystolic ValueDiastolic ValueDisplay
5System Requirements: 550MHz CPU. 256MB Memory, 1024x768 pixel resolution, 256 color, CD-ROM drive, 1 free USB port, 40MB free hard disk space, Microsoft Windows 2000 / XP / Vista.Install CD©2008 Microlife Corporation. All Rights Reserved.WatchBP Analyzer OfficeVersion 1.0.2.4WatchBP Office and componentsWatchBP Office blood pressure device (1 pcs.)WatchBP Office Software CD X1AC Adaptor X1 (Input: 100-240V~50/60Hz 0.7A Output: +7.5V 2A)M size (22cm~32cm) X2 L size (32cm~42cm) X2Manual Instruction X1 Quick Start Guide X1Bluetooth receiver X1Install CD X1mm HgmmHgSYS mmHgDIASYS DIAmmHgPUL/minAUSCULT  1st  FOLLOW UPBluetooth Receiver DriverInstall CDHow to take blood press sure measurement in three differe nt modesQuick start guide Flip over the quick start guide to see ‘Product description’.  1Plug cuff connector  in the cuff socketTurn on the deviceSwitch the mode«1st VISIT» Mode (Double arm)«FOLLOW-UP» Mode (Single arm)«AUSCULTA TION» Mode(Single arm) Press Start/Stop button to begin a measurementNote blood presure using auscultatory methodPlace the stethoscope on the hollow of the elbow ,  over the brachial arteryPress Start/Stop button to begin a measurementAn averaged result will be displayed after three consecutive measurments23EN 1Simultaneous dual-arm measurement and auscultatory function for accur a te in-offi c e  blood pressure measurement.Instruction Manual
EN6Selective CuffsM (Medium size) 22 - 32 cm (8.75 - 12.5 inches) With air tube 130 cm* Please contact Microlife or its distributor to purchase cuffs. L (Large size)32-42 cm (12.5-16.5 inches) With air tube 130 cm
72–3cmM (Medium size) 22 - 32 cm (8.75 - 12.5 inches) M is the correct size for most people.L (Large size)32-42 cm (12.5-16.5 inches)Before using WatchBP OfficeSelecting the correct cuff Two different cuff sizes are provided with the WatchBP Office: Medium and Large. Select the cuff size that best matches the circumference of the patient’s upper arm. Fitting the cuff properly1)  Place the cuff over the left (right) upper arm so that the air tube and artery mark arrow point toward the lower arm.2)  Lay the cuff on the arm. Make sure that the lower edge of the cuff lies approximately ¾” to 1”(2 to 3cm) above the elbow. Connect the cuff to the device by inserting the cuff connector into the cuff connector socket.RR
EN83)  Wrap and tighten the cuff around the arm. 4)  Leave a little free space between the arm and the cuff. The patient should be able to fit 2 fingers between the arm and the cuff. Clothing must not restrict the arm. Remove all clothing covering or constricting the measurement arm.5)  Cuffs that do not fit properly can result in incorrect blood pressure measurements. Use a different size cuff if the range Index at the end of the cuff does not fall into the range specified by the range stripes.6)  Apply the same steps to the other arm if a double arm measurement is required.* The range index of cuff should be fallen into this range.
9AUSCULT  1st  FOLLOW UPAUSCULT  1st  FOLLOW UP AUSCULT  1st  FOLLOW UPThree operation modesmmHgmmHgSYSmmHgDIASYSDIAmmHgPUL/minAUSCULT  1st  FOLLOW UPL/R3 sec.INTERVAL3 sec.7 sec.HIDE7 sec.«AUSCULTATION» ModeSelect «AUSCULTATION» Mode for blood pressure measurement of patients with arrhythmia, and, if needed, to confirm if a patient is suitable for the oscillometric method using a digital blood pressure monitor. Choose a desired operation modeThe WatchBP Office device includes three convenient operation modes: «AUSCULTATION», «1st VISIT» and «FOLLOW-UP» Mode.Use the Mode Switch on the side of the device to select the desired mode. AUSCULT  1st  FOLLOW UPAUSCULT  1st  FOLLOW UP AUSCULT  1st  FOLLOW UP
EN10AUSCULT  1st  FOLLOW UPAUSCULT  1st  FOLLOW UP AUSCULT  1st  FOLLOW UPSimultaneous dual-arm measurements In «1st VISIT» Mode, the WatchBP Office device measures the patient’s blood pressure on both  arms simultaneously, helping to determine the preferred measurement arm and reveal other potential cardiovascular risks. «1st VISIT» ModeSelect «1st VISIT» Mode to complete a fully-automated triple measurement on both arms according to recommended ESH/AHA blood pressure measurement protocols for a patient’s first office visit.
11Automated triple measurementIn «1st VISIT» Mode, the WatchBP Office device automatically takes three consecutive measurements at one minute intervals by default. The result of these three measurements are averaged to produce the first visit blood pressure measurement.Determine proper measurement armIf the measurement result between the two arm differs more than 20mmHg for Systolic or 10mmHg for Diastolic for all three measurements, the device will indicate the higher measurement arm by a flashing L or R. Three operation modes (cont.)12360 sec. 60 sec.* The user can manually set up the measurement interval of 15, 30, 45 or 60 seconds in 1st Visit and Follow-up Mode. (Please refer to special functions section page.24 “Set up measurement interval”).The higher arm should be used as the preferred measurement arm for future measurements.
EN12«FOLLOW-UP» ModeSelect «FOLLOW-UP» Mode to complete a fully-automated triple measurement on the preferred measurement arm. According to recommended ESH/AHA blood pressure measurement protocols for follow-up blood pressure measurement.Automated triple measurementIn «FOLLOW-UP» Mode, the WatchBP Office device automatically takes three consecutive measurements at one minute intervals by default. The result of these three measurements are averaged to produce the follow-up visit blood pressure measurement.AUSCULT  1st  FOLLOW UPAUSCULT  1st  FOLLOW UP AUSCULT  1st  FOLLOW UP* The user can manually set up the measurement interval of 15, 30, 45 or 60 seconds in 1st Visit and Follow-up Mode. (Please refer to special functions section page.24 “Set up measurement interval”).12360 sec. 60 sec.
13AUSCULT  1st  FOLLOW UPAUSCULT  1st  FOLLOW UP AUSCULT  1st  FOLLOW UPTaking measurements using WatchBP Office«AUSCULTATION» Mode1)  Turn on the power – Turn on the device  by switching the I/O switch at the back of the device into the ON position. and make sure Bluetooth switch is in off position.3)  Switch to «AUSCULTATION» Mode – Slide the Mode Switch to «AUSCULTATION» Mode.2)  Apply the cuff – Properly fit one cuff to the preferred measurement arm.DC 6V 600mADC 6V 600mA* Additional visual instruction can be found on the cuff4)  Adjust inflation pressure if necessary – Inflation pressure can be set manually from 140 to 280 mmHg by pressing and holding the Start/Stop Button for 3 seconds.Press the M Button to adjust the pressure and press Start/Stop to save the setting.
EN145)  Start inflation – Press the Start/Stop Button to start inflation of the cuff.6)  Take note – Take note of the systolic and diastolic values in the same manner used for sphygmomanometer measurements.7)  Set device to Stand by Mode – The device can be set to Stand by Mode by pressing the Start/Stop Button. The device will automatically switch to Idle Mode if left for five minutes.Stand by mode* The default inflation pressure is 180 mmHg. * When the device is in standby mode, it will remind you the current preset pressure setting by displaying the current pressure setting on the screen. Press and hold for 3 seconds...
15AUSCULT  1st  FOLLOW UPAUSCULT  1st  FOLLOW UP AUSCULT  1st  FOLLOW UPTaking measurements using WatchBP Office (cont.)«1st VISIT» Mode1)  Turn on the power – Turn on the device by switching the I/O switch at the back of the device to the ON position. and make sure Bluetooth switch is in off position.3)  Switch to «1st VISIT» Mode – Slide the Mode Switch to «1st VISIT» Mode.* Additional visual instruction can be found on the cuff4)  Adjust inflation pressure if necessary – Inflation pressure can be set manually from 140 to 280 mmHg by pressing and holding the Start/Stop Button for 3 seconds.Press the M Button to adjust the pressure and press Start/Stop to save the setting.2)  Apply the cuff – Properly fit one cuff to each of the patient’s arms.DC 6V 600mADC 6V 600mA
EN16* The default inflation pressure is 180 mmHg. * The device will delate and re-pump more air into the cuff if the blood pressure of the patient is higher than that of the inflation setting of the device. * When the device is in standby mode, it will remind you the current preset pressure setting by displaying the current pressure setting on the screen. 5)  Start measurements – Press the Start/Stop Button to start the measurement sequence.6)  Three consecutive measurements – The device will take three consecutive measurements with one minute intervals in between measurements.60-second countdown...60 seconds60 seconds21 3* A 60-second countdown will proceed the first measurement .Press and hold for 3 seconds...
17Taking measurements using WatchBP Office (cont.)7)  View the average measurement – Only the average measurement will be displayed once the triplicate measurement is completed.8)  View all measurements – Results of the three individual measurements can be accessed by pressing the M Button. Press the M Button once to review the first measurement. Continue to press the M Button to review the other measurements.* When the measurements are completed, the blood pressure measurement corresponding with the consistently higher arm will blink continuously for 5 seconds. If no significant difference (20mmHg for systolic blood pressure or 10 mmHg for diastolic blood pressure or both) exists between the two arms, neither measurement will blink.* WatchBP Office only stores the  last three measurements completed in «1st VISIT» and «FOLLOW-UP» Modes. To store patient blood pressure for later retrieval, please use the included WatchBP Office software.213
EN18AUSCULT  1st  FOLLOW UPAUSCULT  1st  FOLLOW UP AUSCULT  1st  FOLLOW UP«FOLLOW-UP» Mode1)  Turn on the power – Turn on the device by switching the I/O switch at the back of the device to the ON position and make sure Bluetooth switch is in off position..3)  Switch to «FOLLOW-UP» Mode – Slide the Mode Switch to «FOLLOW-UP» Mode.4)  Select the inflation cuff – Set the device to inflate only the designated (right or left) cuff. Press and hold the M Button for 3 seconds. 2)  Apply the cuff – Properly fit one cuff to the preferred measurement arm.Press and hold for 3 seconds...* Additional visual instruction can be found on the cuffDC 6V 600mADC 6V 600mA
195)  Press the M Button to toggle between the right(R) and left(L) cuff. Press the Start/Stop Button to save the setting. 6)  Change inflation pressure – Inflation pressure can be set manually from 140 to 280 mmHg by pressing and holding the Start/Stop Button for 7 seconds.Press M Button to adjust the pressure and press Start/Stop again to save the setting.Taking measurements using WatchBP Office (cont.)Press and hold for 7 seconds...7) Start measurements – Press the Start/Stop Button to start the measurement sequence.60-second countdown...
EN208)  Three consecutive measurements – The device will take three consecutive measurements with one minute intervals in between measurements.9)  View the average measurement – Only the average measurement will be displayed once the triplicate measurement is completed.60 seconds60 seconds21310) View all measurements – Results of the three individual measurements can be accessed by pressing the M Button. Press the M Button once to review the first measurement. Continue to press the M Button to review the other measurements.213* WatchBP Office only stores the  last three measurements completed in «1st VISIT» and «FOLLOW-UP» Modes. To store patient blood pressure for later retrieval, please use the included WatchBP Office software.
21Special FunctionsHiding measurement valuesThe WatchBP Office device features a Hide function in order to prevent unnecessary elevated blood pressure in patients due to nervousness triggered by visible blood pressure measurements. * This feature is only available in follow up mode1)  Activate the «Hide» Function – pressing and holding the M Button for 7 seconds. 2)  Select ON or OFF – Press the M Button again to turn the “Hide” function ON or OFF. 3)  Confirm – Press Start/Stop to confirm the setting.Press and hold for 7 seconds...* The «1st VISIT» Mode “Hide” default setting is set to ON and cannot be set to OFF.* When the «Hide» function is ON, the countdown between each measurement and the Hide icon will be displayed. * Please make sure the Bluetooth switch is in OFF position, otherwise the «Hide» function can not be set.
EN22MAP (Mean Arterial Pressure)The WatchBP Office device measures the true mean arterial pressured (MAP) of the patient. Each measurement includes the calculation of a single MAP value. The average measurement will display the average MAP value. When viewing in Memory mode, the MAP value will be displayed with the systolic/diastolic pressure once every 5 seconds. PP (Pulse Pressure)The WatchBP Office device measures the pulse pressure (PP) of the patient: pulse pressure = systolic - diastolic pressure. Each measurement includes the calculation of a single PP value. The average measurement will display the average PP value. When viewing in Memory mode, the PP value will be displayed with the systolic/diastolic pressure once every 5 seconds.
23Special Function (cont.)Taking fewer than three measurementsIn «FOLLOW-UP» Mode only, the measurement sequence can be stopped at anytime by pushing the Start/Stop Button. By pushing the Start/Stop Button remaining the device enters stand-by, cancelling remaining measurements in the sequence. Data from the measurements completed can be by viewed by pushing the M Button.Skipping the 60-second countdownIn «1st VISIT» and «FOLLOW-UP» Mode, the 60-second pre-measurement countdown can be skipped by pushing the Start/Stop Button. Once the Start/Stop Button is pushed, the device will immediately begin the next measurement.* The device can put into Stand by Mode by pressing the Start/Stop Button after the completion of measurements. The device will automatically switch to Stand by Mode if left unattended for five minutes.Cancels remaining measurements at anytime during the measurement sequence.Skips the 60-second countdown and begins measurement.Stand by mode
EN24Setting inflation pressureInflation pressure can be set manually from 140 to 280 mmHg by pressing and holding the Start/Stop Button for 7 seconds.Setting measurement intervalsIt is concluded that a 15-second interval between blood pressure readings and these measurements are started after a 5 minute rest period.1)  Pressing and holding the Start/Stop Button for 3 seconds. Press and hold for 7 seconds... Press and hold for 3 seconds...Press the M Button to adjust the pressure and press Start/Stop again to save the setting.2)  Press the M Button to adjust the measurement interval, then press the Start/Stop Button to confirm, the meter will back to stand by Mode. 15453060
25AUSCULT  1st  FOLLOW UPAUSCULT  1st  FOLLOW UP AUSCULT  1st  FOLLOW UPViewing Measurements«AUSCULTATION» ModeThe WatchBP Office blood pressure measurement device in «AUSCULTATION» mode serves only as a pressure gauge. No oscillometric measurements will be taken and a mental note must be made of the systolic and diastolic values. «1st VISIT» and «FOLLOW-UP» Mode 1)   Switch the mode switch –The M Button can be pressed in either «1st VISIT» or the «FOLLOW-UP» mode to retrieve the last set of the triplicate measurements. 2)  Viewing the average of all measurements –  At first an “A” will be displayed, followed by an average of all measurements taken in the last sequence of measurements.MemoryMemory
EN263)  Viewing MAP and PP values –  Ten seconds after the display of the average blood pressure, the device will display values for Pulse Pressure (PP) and Mean Arterial Pressure (MAP). 5)   Values for the third measurement (including PP and MAP values) will be shown on the display in the order listed in step 3. 4)  Viewing individual measurements – Press the M Button once again to view individual measurements. The display will flash a number “3” indicating the third measurement.10  seconds...6)  Press the M Button to retrieve values for second and first measurements from the last completed set of measurements.10  seconds...321
27AUSCULT  1st  FOLLOW UPINTERVALPRESSURE7 sec.         Controlling the WatchBP Office wirelesslyThe WatchBP Office can be controlled wirelessly via Bluetooth with the included WatchBP Office software. Please follow the steps listed here to set up a wireless connection with a computer. Setting up the Bluetooth receiver1)  Place the included “Bluetooth USB Adaptor” CD in the CD-ROM of the computer. Click on «setup.exe» in the CD-ROM directory.2)  An installation window will appear. Follow the instructions on the screen.* Before installing the WatchBP Office software, the driver software for the Bluetooth receiver must be first installed on the computer.* The distance between WatchBP office and operating computer should be within 10m.Bluetooth Receiver DriverInstall CDInstruction
EN28         3)  Insert the Bluetooth receiver into a USB port of the computer after the installation program requests to do so. 4)  Restart the computer after installation is completed.* Select a single USB port and make sure to always use the same port in the future.* Do not use other Bluetooth receiver.5)  Turn on the Bluetooth function of the WatchBP Office device on by switching the Bluetooth switch at the back of the device to the “ON” position. The Bluetooth icon will illuminate and flash continuously to show a connection.*  It is noted that WatchBP office device will not respond to any button pressing or switching when the Bluetooth is on. It is to prevent accidental interruptions during the measurement. Restart your computerDC 6V 600mADC 6V 600mA
29Installation of the software program1)  Put the CD in the CD-ROM drive of your computer. Then click on «setup.exe» in the CD’s directory.2)  Follow the instructions provided in the installation window on the computer screen.3)  When the installation is complete, be sure restart the computer before using the program for the first time.Controlling the device wirelessly 1)  Turn on the Bluetooth switch located on the back of the device. 2)  Make sure the Bluetooth receiver is at right position and the light is on (if applicable). 3)  Launch WatchBP Office Software from the computer. Follow these steps to control the WatchBP Office from the computer.Controlling the WatchBP Office wirelessly (cont.)DC 6V 600mADC 6V 600mASystem Requirements: 550MHz CPU. 256MB Memory, 1024x768 pixel resolution, 256 color, CD-ROM drive, 1 free USB port, 40MB free hard disk space, Microsoft Windows 2000 / XP / Vista.Install CD©2008 Microlife Corporation. All Rights Reserved.WatchBP Analyzer OfficeVersion 1.0.2.4* System Requirements: 550MHz CPU. 256MB Memory, 1024x768 pixel resolution, 256 color, CD-ROM drive, 1 free USB port, 40MB free hard disk space, Microsoft Windows 2000 / XP / Vista.
EN30Software commandsChoose measure blood pressure Click «1st VISIT», «FOLLOW-UP »Enter Patient ID Input Patient IDStart to perform Press the «Measure» button Store blood pressure measurements Click «Save». The data is stored in an Excel data table. The file name is generated automatically from the following format «Patient ID+ WatchBP Office +measurement mode + yyyymmdd.xls».View the data Go to the folder where the file is saved and open the corresponding excel file.Print the data sheet While the corresponding Excel file is open, click print in Excel.Disconnect the device Click <<Disconnect>> before existing the programClose the program Click «Close»* The software automatically searches for the nearest WatchBP device when the program is launched. If the device is not located, confirm that the Bluetooth switch on the WatchBP Office is set to the ON position.
31DC 6V 600mADC 6V 600mARechargeable BatteryWatchBP Office features a built-in, rechargeable Ni-MH battery pack that delivers up to 400~500 measurement cycles. The battery can be recharged between use with the provided power adaptor.Rechargeable battery and power adaptor         Using a power adaptorPlease use only the Microlife adaptor (DSA-20P-05 Adaptor) supplied with the WatchBP Office to recharge the device.1)  Plug the adapter cable into the Power Plug in the WatchBP Office device.2)  Plug the adaptor plug into the wall socket. When the power adaptor is connected, no battery power is consumed.* An orange recharge indicator means recharge in progress.* A green recharge indicator means recharge is completed.mmHgmmHgSYSmmHgDIASYSDIAmmHgPUL/minAUSCULT  1st  FOLLOW UPL/R3 sec.INTERVAL3 sec.7 sec.HIDE7 sec.DC 6V 600mADC 6V 600mA
EN32Troubleshooting         Problem Possible cause How to make correctionNo power(No LCD display) Power supply is not properly pluggedPlug power supply into wall socket.Battery is fully discharged Recharge the rechargeable battery by plugging in the power supply.Cuff does not inflate properlyLoose connection of the tube Make sure the tube of the cuff is securely connected to the device.Leakage of the tube / bladderCheck for cracks on the tube or the bladder. Please contact Microlife customer service for this issue.Hide, pressure, or pause interval can not be setWireless option is ON The device can not be operated by the Start/Stop and M Button when the Bluetooth switch is in thev ON position. Please switch off the Bluetooth or use the included software to control the device.Unable to take measurementNo result displayed between measurementsDevice is in AUSCULTATION mode or the Hide function is activatedDisable “hide” function, or use «1st VISIT» / «FOLLOW-UP» mode to measure the blood pressure.Can’t access device via computerBluetooth software not installedMake sure the Bluetooth software provided with the device is installed on the computer.
33If an error occurs during measurement, the measurement is interrupted and an error message «Err» is displayed.•  Please contact your local Microlife service center if the error persists.•  If you think the results are unusual, please read through the information in this instruction manual carefully.Error Description Potential cause and remedy«Err 1»Signal too weakThe pulse signals on the cuff are too weak. Re-position the cuff and repeat the measurement.«Err 2»Error signal During the measurement, error signals were detected by the cuff, caused for instance by movement or muscle tension. Repeat the measurement, keeping your arm still.Error messages
EN34«HI»Pulse or cuff pressure too highThe pressure in the cuff is too high (over 300 mmHg) OR the pulse is too high (over 200 beats per minute). Relax for 5 minutes and repeat the measurement.«LO»Pulse too lowThe pulse is too low (less than 40 beats per minute). Repeat the measurement.«Err 3»No pressure in the cuffAn adequate pressure cannot be generated in the cuff. A leak may have occurred. Replace the batteries if necessary. Repeat the measurement.«Err 5»Abnormal resultThe measuring signals are inaccurate and no result can therefore be displayed. Read through the checklist for performing reliable measurements and then repeat the measurement.
35Protect the device from water and moistureProtect the device from extreme heat and coldProtect device from impact and dropsProtect the device from direct sunlightAvoid proximity to electromagnetic fields, such as those produced by mobile phonesNever open deviceSafety, care, accuracy test and disposalSafety and protectionThis device may be used only for the purpose described in this booklet. The device comprises of sensitive components and must be treated with caution. The manufacturer cannot be held liable for damage caused by incorrect application. Observe the storage and operating conditions described in the “Technical specifications” section of this manual.• Ensure that children do not use the device unsupervised; some parts are small enough to be swallowed.• Only activate the pump when cuff is installed.• Do not use the device if you think it is damaged or if anything appears unusual.• Read the further safety instructions in the individual sections of the instruction manual.• Do not connect the device to a computer  until prompted to do so by the computer software.
EN36AUSCULT  1st  FOLLOW UPINTERVALPRESSURE7 sec.Do not wash the cuff!Do not iron the cuff!Accuracy testWe recommend the WatchBP Office device be tested for accuracy every 2 years or after mechanical impact (e.g. being dropped). Please contact Microlife to arrange for an accuracy test.Cleaning the cuff DO NOT wash the cuff. DO NOT iron the cuff cover.Device care Clean the device with a soft, dry cloth.DisposalBatteries and electronic instruments must be disposed of in accordance with the locally applicable regulations, and not as domestic waste.
37Technical specifications Operation temperature/humidity:• 10 - 40 °C / 50 - 104 °F Storage temperature/humidity:• -20 - +50 °C / -4 - +122 °F• 15 - 90 % relative maximum humidityWeight: • 1100 g (including rechargeable battery pack)Dimensions: • 200 x 125 x 90 mmMeasuring method: • Oscillometric, corresponding to KorotkoffMeasurement range: • 30 - 280 mmHg – blood pressure• 40 - 200 beats per minute – pulseCuff pressure display:  • Range: 0 - 299 mmHg• Resolution: 1 mmHg• Static accuracy: pressure within ±3 mmHg• Pulse accuracy: ±5 % of the readout valueVoltage source: • Rechargeable battery pack;  4.8V C4000 mAh • Mains adapter DC 7.5V, 2 AReference to Standards: • EN 1060-1 /-3, NIBP-requirements• IEC60601-1 General requirement for safety• IEC60601-1-2 Requirements for EMC• EN1060-4, NIBP clinical investigation• ANSI/AAMI SP10, NIBP requirements• Bluetooth system to V2.0• FCC part15CFederal Communications Commission (FCC) Statement15.21You are cautioned that changes or modifications not expressly approved by the part responsible for compliance could void the user’s authority to operate the equipment.5.105(b)This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against
EN38harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:-Reorient or relocate the receiving antenna.-Increase the separation between the equipment and receiver.-Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.-Consult the dealer or an experienced radio/TV technician for help.Operation is subject to the following two conditions:1) this device may not cause interference and2) this device must accept any interference, including interference that may cause undesired operation of the device.FCC RF Radiation Exposure Statement:This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. End users must follow the specific operating instructions for satisfying RF exposure compliance. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter.
This device is covered by a two-year guarantee and accessories are cover by a one-year guarantee from the date of purchase. This guarantee is valid only on presentation of the guarantee card completed by the owner confirming date of purchase or purchase receipt.Product: WatchBP OfficeProduct Number: Twin200Date: Name: Address: Date: Telephone:Email:Guarantee card
IB WatchBP Office 2508Europe / Middle-East / AfricaMicrolife WatchBP AG Espenstrasse 139 9443 Widnau, Switzerland Tel +41 71 727 7000 Fax +41 71 727 7011 Email: watchbp@microlife.ch www.watchbp.com AsiaMicrolife Corporation  9F, 431, RuiGang Road, NeiHu  Taipei, 114, Taiwan, R.O.C. Tel +886 2 8797 1288 Fax+886 2 8797 1283 Email: watchbp@microlife.com.tw www.watchbp.com North / Central / South AmericaMicrolife Medical Home Solutions, Inc.  2801 Youngfield St., Suite 241 Golden, CO 80401, USA Tel +1 303 274 2277 Fax +1 303 274 2244 Email: watchbp@mimhs.com www.watchbp.com