Microlife TWIN200-1 WatchBP Professional User Manual
Microlife Corporation WatchBP Professional
User Manual
EN 1
Simultaneous dual-arm
measurement and
auscultatory function for
accurate office blood
pressure measurement.
Instruction Manual
WatchBP Office combines the advantages of an automatic blood pressure device and hybrid
sphygmomanometer and is designed to provide convenient, accurate and reliable in-office
blood pressure measurement following recommended guidelines of the European Society
of Hypertension (ESH) 1, American Heart Association (AHA), and World Health Organization
(WHO).
1 O’Brien E, Asmar R, Beilin L, Imai Y, et al. European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure
measurement. European Society of Hypertension Working Group on Blood Pressure Deviceing. J Hypertens 2003;21:821-848.
EN
Product description
Name of parts and display ................................. 3–4
Components and accessories ............................. 5–6
Before using WatchBP Office
Selecting and fitting the correct cuff .................. 7–8
Three operation modes
«AUSCULTATION» Mode .......................................... 9
«1st VISIT» Mode ............................................ 10–11
«FOLLOW-UP» Mode .............................................12
Taking measurements using WatchBP Office
«AUSCULTATION» Mode .................................. 13–14
«1st VISIT» Mode ............................................ 15–17
«FOLLOW-UP» Mode ....................................... 18–20
Special Functions
Hiding measurement values ................................ 21
Mean arterial pressure and pulse pressure ......... 22
Taking fewer than three measurements ............ 23
Skipping the 60-second countdown .................... 23
Setting inflation pressure ..................................... 24
Setting measurement intervals ........................... 24
Viewing Measurements
«AUSCULTATION» Mode ....................................... 25
«1st VISIT» and «FOLLOW-UP» Mode ............. 25–26
Controlling the WatchBP Office Wirelessly
Setting up the Bluetooth receiver .................. 27–28
Installation of the software program ................... 29
Controlling the device wirelessly .................. 29–30
Appendix
Rechargeable battery and power adaptor .......... 31
Troubleshooting ................................................... 32
Error messages ............................................... 33–34
Safety, care, accuracy test and disposal ......... 35–36
Technical specifications ....................................... 37
Table of Contents
3
L
R
mmH g
mmHg
SYS
mmHg
DIA
SYS
DIA
mmHg
PUL
/min
AUSCULT 1st FOLLOW UP
L/R
3 sec.
INTERVAL
3 sec.
7 sec.
HIDE
7 sec.
I/O switch
Bluetooth Switch
Main Adaptor Socket
Battery Compartment
Cuff Socket (L)
Cuff Socket (R)
Recharge Indicator
Display
M Button (Memory)
Mode Switch
Start/Stop Button
DC 6V 600mADC 6V 600mA
Product description
Name of parts
4EN
Mean Arterial Pressure (MAP)
Pulse Pressure (PP)
Mode Indication
External Power
Bluetooth
Rest and Arm Position Reminder
Battery Symbol
Hide Display
Memory Mode
Pulse Rate
Heart Beat
L/R Cuff
Value for Left Arm
Systolic Value
Diastolic Value
Value for Right Arm
Systolic Value
Diastolic Value
Display
5
System Requirements: 550MHz CPU. 256MB Memory, 1024x768
pixel resolution, 256 color, CD-ROM drive, 1 free USB port, 40MB
free hard disk space, Microsoft Windows 2000 / XP / Vista.
Install CD
©
2
0
0
8
M
i
c
r
o
l
i
f
e
C
o
r
p
o
r
a
t
i
o
n
.
A
l
l
R
i
g
h
t
s
R
e
s
e
r
v
e
d
.
WatchBP Analyzer Office
Version 1.0.2.4
WatchBP Office and components
WatchBP Office blood
pressure device (1 pcs.)
WatchBP Office Software
CD X1
AC Adaptor X1 (Input:
100-240V~50/60Hz 0.7A
Output: +7.5V 2A)
M size (22cm~32cm) X2
L size (32cm~42cm) X2
Manual Instruction X1
Quick Start Guide X1
Bluetooth receiver X1
Install CD X1
mm Hg
mmHg
SYS
mmHg
DIA
SYS
DIA
mmHg
PUL
/min
AUSCULT 1st FOLLOW UP
Bluetooth Receiver
Driver
Install CD
How to take blood press sure measurement in three differe nt modes
Quick start guide
Flip over the quick start guide to see ‘Product description’.
1Plug cuff connector
in the cuff socket
Turn on the device
Switch the mode
«1st VISIT» Mode (Double arm)
«FOLLOW-UP» Mode (Single arm)
«AUSCULTA TION» Mode
(Single arm) Press Start/Stop
button to begin a
measurement
Note blood
presure using
auscultatory
method
Place the stethoscope on
the hollow of the elbow ,
over the brachial artery
Press Start/Stop
button to begin a
measurement
An averaged result will
be displayed after three
consecutive measurments
2
3
EN 1
Simultaneous dual-arm
measurement and
auscultatory function for
accur a te in-offi c e blood
pressure measurement.
Instruction Manual
EN
6
Selective Cuffs
M (Medium size)
22 - 32 cm (8.75 - 12.5 inches)
With air tube 130 cm
* Please contact Microlife or its distributor to purchase cuffs.
L (Large size)
32-42 cm (12.5-16.5 inches)
With air tube 130 cm
7
2–3cm
M (Medium size)
22 - 32 cm (8.75 - 12.5 inches)
M is the correct size for most people.
L (Large size)
32-42 cm (12.5-16.5 inches)
Before using WatchBP Office
Selecting the correct cuff
Two different cuff sizes are provided with the
WatchBP Office: Medium and Large. Select the cuff
size that best matches the circumference of the
patient’s upper arm.
Fitting the cuff properly
1) Place the cuff over the left (right) upper arm so
that the air tube and artery mark arrow point
toward the lower arm.
2) Lay the cuff on the arm. Make sure that the lower
edge of the cuff lies approximately ¾” to 1”(2 to
3cm) above the elbow.
Connect the cuff to the device by inserting the cuff
connector into the cuff connector socket.
R
R
EN
8
3) Wrap and tighten the cuff around the arm.
4) Leave a little free space between the arm and
the cuff. The patient should be able to fit 2 fingers
between the arm and the cuff. Clothing must not
restrict the arm. Remove all clothing covering or
constricting the measurement arm.
5) Cuffs that do not fit properly can result in
incorrect blood pressure measurements. Use a
different size cuff if the range Index at the end of
the cuff does not fall into the range specified by
the range stripes.
6) Apply the same steps to the other arm if a double
arm measurement is required.
* The range index of cuff should
be fallen into this range.
9
AUSCULT 1st FOLLOW UPAUSCULT 1st FOLLOW UP AUSCULT 1st FOLLOW UP
Three operation modes
mmHg
mmHg
SYS
mmHg
DIA
SYS
DIA
mmHg
PUL
/min
AUSCULT 1st FOLLOW UP
L/R
3 sec.
INTERVAL
3 sec.
7 sec.
HIDE
7 sec.
«AUSCULTATION» Mode
Select «AUSCULTATION» Mode for blood pressure
measurement of patients with arrhythmia, and,
if needed, to confirm if a patient is suitable for
the oscillometric method using a digital blood
pressure monitor.
Choose a desired operation mode
The WatchBP Office device includes three convenient
operation modes: «AUSCULTATION», «1st VISIT» and
«FOLLOW-UP» Mode.Use the Mode Switch on the
side of the device to select the desired mode.
AUSCULT 1st FOLLOW UPAUSCULT 1st FOLLOW UP AUSCULT 1st FOLLOW UP
EN
10
AUSCULT 1st FOLLOW UPAUSCULT 1st FOLLOW UP AUSCULT 1st FOLLOW UP
Simultaneous dual-arm measurements
In «1st VISIT» Mode, the WatchBP Office device
measures the patient’s blood pressure on both
arms simultaneously, helping to determine the
preferred measurement arm and reveal other
potential cardiovascular risks.
«1st VISIT» Mode
Select «1st VISIT» Mode to complete a fully-
automated triple measurement on both arms
according to recommended ESH/AHA blood
pressure measurement protocols for a patient’s
first office visit.
11
Automated triple measurement
In «1st VISIT» Mode, the WatchBP Office device
automatically takes three consecutive measurements
at one minute intervals by default. The result of these
three measurements are averaged to produce the
first visit blood pressure measurement.
Determine proper measurement arm
If the measurement result between the two arm
differs more than 20mmHg for Systolic or 10mmHg
for Diastolic for all three measurements, the device
will indicate the higher measurement arm by a
flashing L or R.
Three operation modes (cont.)
123
60 sec. 60 sec.
* The user can manually set up the measurement
interval of 15, 30, 45 or 60 seconds in 1st Visit and
Follow-up Mode. (Please refer to special functions
section page.24 “Set up measurement interval”).
The higher arm should be used as the preferred
measurement arm for future measurements.
EN
12
«FOLLOW-UP» Mode
Select «FOLLOW-UP» Mode to complete a fully-
automated triple measurement on the preferred
measurement arm. According to recommended
ESH/AHA blood pressure measurement protocols for
follow-up blood pressure measurement.
Automated triple measurement
In «FOLLOW-UP» Mode, the WatchBP Office
device automatically takes three consecutive
measurements at one minute intervals by default.
The result of these three measurements are
averaged to produce the follow-up visit blood
pressure measurement.
AUSCULT 1st FOLLOW UPAUSCULT 1st FOLLOW UP AUSCULT 1st FOLLOW UP
* The user can manually set up the measurement
interval of 15, 30, 45 or 60 seconds in 1st Visit and
Follow-up Mode. (Please refer to special functions
section page.24 “Set up measurement interval”).
123
60 sec. 60 sec.
13
AUSCULT 1st FOLLOW UPAUSCULT 1st FOLLOW UP AUSCULT 1st FOLLOW UP
Taking measurements using WatchBP Office
«AUSCULTATION» Mode
1) Turn on the power – Turn on the device by
switching the I/O switch at the back of the
device into the ON position. and make sure
Bluetooth switch is in off position.
3) Switch to «AUSCULTATION» Mode – Slide the
Mode Switch to «AUSCULTATION» Mode.
2) Apply the cuff – Properly fit one cuff to the
preferred measurement arm.
DC 6V 600mADC 6V 600mA
* Additional visual instruction can be found on the cuff
4) Adjust inflation pressure if necessary –
Inflation pressure can be set manually from 140
to 280 mmHg by pressing and holding the Start/
Stop Button for 3 seconds.
Press the M Button to adjust the pressure and
press Start/Stop to save the setting.
EN
14
5) Start inflation – Press the Start/Stop Button to
start inflation of the cuff.
6) Take note – Take note of the systolic and
diastolic values in the same manner used for
sphygmomanometer measurements.
7) Set device to Stand by Mode – The device can
be set to Stand by Mode by pressing the Start/
Stop Button. The device will automatically switch
to Idle Mode if left for five minutes.
Stand by mode
* The default inflation pressure is 180 mmHg.
* When the device is in standby mode, it will remind you
the current preset pressure setting by displaying the
current pressure setting on the screen.
Press and hold
for 3 seconds...
15
AUSCULT 1st FOLLOW UPAUSCULT 1st FOLLOW UP AUSCULT 1st FOLLOW UP
Taking measurements using WatchBP Office (cont.)
«1st VISIT» Mode
1) Turn on the power – Turn on the device by
switching the I/O switch at the back of the device
to the ON position. and make sure Bluetooth
switch is in off position.
3) Switch to «1st VISIT» Mode – Slide the Mode
Switch to «1st VISIT» Mode.
* Additional visual instruction can be found on the cuff
4) Adjust inflation pressure if necessary –
Inflation pressure can be set manually from 140
to 280 mmHg by pressing and holding the Start/
Stop Button for 3 seconds.
Press the M Button to adjust the pressure and
press Start/Stop to save the setting.
2) Apply the cuff – Properly fit one cuff to each of
the patient’s arms.
DC 6V 600mADC 6V 600mA
EN
16
* The default inflation pressure is 180 mmHg.
* The device will delate and re-pump more air into the cuff
if the blood pressure of the patient is higher than that of
the inflation setting of the device.
* When the device is in standby mode, it will remind you
the current preset pressure setting by displaying the
current pressure setting on the screen.
5) Start measurements – Press the Start/Stop
Button to start the measurement sequence.
6) Three consecutive measurements – The device
will take three consecutive measurements with
one minute intervals in between measurements.
60-second
countdown...
60
seconds
60
seconds
21 3
* A 60-second countdown will proceed the first
measurement .
Press and hold
for 3 seconds...
17
Taking measurements using WatchBP Office (cont.)
7) View the average measurement – Only the
average measurement will be displayed once the
triplicate measurement is completed.
8) View all measurements – Results of the three
individual measurements can be accessed by
pressing the M Button. Press the M Button once to
review the first measurement. Continue to press
the M Button to review the other measurements.
* When the measurements are completed, the blood
pressure measurement corresponding with the
consistently higher arm will blink continuously for
5 seconds. If no significant difference (20mmHg for
systolic blood pressure or 10 mmHg for diastolic blood
pressure or both) exists between the two arms, neither
measurement will blink.
* WatchBP Office only stores the last three measurements
completed in «1st VISIT» and «FOLLOW-UP» Modes. To
store patient blood pressure for later retrieval, please use
the included WatchBP Office software.
2
1
3
EN
18
AUSCULT 1st FOLLOW UPAUSCULT 1st FOLLOW UP AUSCULT 1st FOLLOW UP
«FOLLOW-UP» Mode
1) Turn on the power – Turn on the device by
switching the I/O switch at the back of the device
to the ON position and make sure Bluetooth
switch is in off position..
3) Switch to «FOLLOW-UP» Mode – Slide the Mode
Switch to «FOLLOW-UP» Mode.
4) Select the inflation cuff – Set the device to
inflate only the designated (right or left) cuff.
Press and hold the M Button for 3 seconds.
2) Apply the cuff – Properly fit one cuff to the
preferred measurement arm.
Press and hold
for 3 seconds...
* Additional visual instruction can be found on the cuff
DC 6V 600mADC 6V 600mA
19
5) Press the M Button to toggle between the right(R)
and left(L) cuff. Press the Start/Stop Button to
save the setting.
6) Change inflation pressure – Inflation pressure
can be set manually from 140 to 280 mmHg by
pressing and holding the Start/Stop Button for
7 seconds.
Press M Button to adjust the pressure and press
Start/Stop again to save the setting.
Taking measurements using WatchBP Office (cont.)
Press and hold
for 7 seconds...
7) Start measurements – Press the Start/Stop
Button to start the measurement sequence.
60-second
countdown...
EN
20
8) Three consecutive measurements – The device
will take three consecutive measurements with
one minute intervals in between measurements.
9) View the average measurement – Only the
average measurement will be displayed once the
triplicate measurement is completed.
60
seconds
60
seconds
2
13
10) View all measurements – Results of the three
individual measurements can be accessed by
pressing the M Button. Press the M Button once to
review the first measurement. Continue to press
the M Button to review the other measurements.
2
1
3
* WatchBP Office only stores the last three measurements
completed in «1st VISIT» and «FOLLOW-UP» Modes. To
store patient blood pressure for later retrieval, please
use the included WatchBP Office software.
21
Special Functions
Hiding measurement values
The WatchBP Office device features a Hide function
in order to prevent unnecessary elevated blood
pressure in patients due to nervousness triggered by
visible blood pressure measurements.
* This feature is only available in follow up mode
1) Activate the «Hide» Function – pressing and
holding the M Button for 7 seconds.
2) Select ON or OFF – Press the M Button again to
turn the “Hide” function ON or OFF.
3) Confirm – Press Start/Stop to confirm the setting.
Press and hold
for 7 seconds...
* The «1st VISIT» Mode “Hide” default setting is set to ON
and cannot be set to OFF.
* When the «Hide»
function is ON, the
countdown between
each measurement and
the Hide icon will be
displayed.
* Please make sure the
Bluetooth switch is in OFF
position, otherwise the
«Hide» function can not
be set.
EN
22
MAP (Mean Arterial Pressure)
The WatchBP Office device measures the true
mean arterial pressured (MAP) of the patient. Each
measurement includes the calculation of a single
MAP value. The average measurement will display
the average MAP value. When viewing in Memory
mode, the MAP value will be displayed with the
systolic/diastolic pressure once every 5 seconds.
PP (Pulse Pressure)
The WatchBP Office device measures the pulse
pressure (PP) of the patient: pulse pressure = systolic
- diastolic pressure. Each measurement includes
the calculation of a single PP value. The average
measurement will display the average PP value.
When viewing in Memory mode, the PP value will be
displayed with the systolic/diastolic pressure once
every 5 seconds.
23
Special Function (cont.)
Taking fewer than three measurements
In «FOLLOW-UP» Mode only, the measurement
sequence can be stopped at anytime by pushing
the Start/Stop Button. By pushing the Start/
Stop Button remaining the device enters stand-
by, cancelling remaining measurements in the
sequence. Data from the measurements completed
can be by viewed by pushing the M Button.
Skipping the 60-second countdown
In «1st VISIT» and «FOLLOW-UP» Mode, the
60-second pre-measurement countdown can be
skipped by pushing the Start/Stop Button. Once
the Start/Stop Button is pushed, the device will
immediately begin the next measurement.
* The device can put into Stand by Mode by pressing
the Start/Stop Button after the completion of
measurements. The device will automatically switch to
Stand by Mode if left unattended for five minutes.
Cancels remaining
measurements at
anytime during
the measurement
sequence.
Skips the 60-second
countdown and begins
measurement.
Stand by mode
EN
24
Setting inflation pressure
Inflation pressure can be set manually from 140 to
280 mmHg by pressing and holding the Start/Stop
Button for 7 seconds.
Setting measurement intervals
It is concluded that a 15-second interval between
blood pressure readings and these measurements
are started after a 5 minute rest period.
1) Pressing and holding the Start/Stop Button for 3
seconds.
Press and hold
for 7 seconds... Press and hold
for 3 seconds...
Press the M Button to adjust the pressure and press
Start/Stop again to save the setting.
2) Press the M Button to adjust the measurement
interval, then press the Start/Stop Button to
confirm, the meter will back to stand by Mode.
15
45
30
60
25
AUSCULT 1st FOLLOW UPAUSCULT 1st FOLLOW UP AUSCULT 1st FOLLOW UP
Viewing Measurements
«AUSCULTATION» Mode
The WatchBP Office blood pressure measurement
device in «AUSCULTATION» mode serves only as a
pressure gauge. No oscillometric measurements will
be taken and a mental note must be made of the
systolic and diastolic values.
«1st VISIT» and «FOLLOW-UP» Mode
1) Switch the mode switch –The M Button can be
pressed in either «1st VISIT» or the «FOLLOW-
UP» mode to retrieve the last set of the triplicate
measurements.
2) Viewing the average of all measurements –
At first an “A” will be displayed, followed by an
average of all measurements taken in the last
sequence of measurements.
MemoryMemory
EN
26
3) Viewing MAP and PP values –
Ten seconds after the display of the average
blood pressure, the device will display values for
Pulse Pressure (PP) and Mean Arterial Pressure
(MAP).
5) Values for the third measurement (including PP
and MAP values) will be shown on the display
in the order listed in step 3.
4) Viewing individual measurements – Press
the M Button once again to view individual
measurements. The display will flash a number
“3” indicating the third measurement.
10
seconds...
6) Press the M Button to retrieve values for second
and first measurements from the last completed
set of measurements.
10
seconds...
3
21
27
AUSCULT 1st FOLLOW UP
INTERVAL
PRESSURE
7 sec.
Controlling the WatchBP Office wirelessly
The WatchBP Office can be controlled wirelessly via
Bluetooth with the included WatchBP Office software.
Please follow the steps listed here to set up a wireless
connection with a computer.
Setting up the Bluetooth receiver
1) Place the included “Bluetooth USB Adaptor” CD
in the CD-ROM of the computer. Click on «setup.
exe» in the CD-ROM directory.
2) An installation window will appear. Follow the
instructions on the screen.
* Before installing the WatchBP Office software, the
driver software for the Bluetooth receiver must be first
installed on the computer.
* The distance between WatchBP office and operating
computer should be within 10m.
Bluetooth Receiver
Driver
Install CD
Instruction
EN
28
3) Insert the Bluetooth receiver into a USB port
of the computer after the installation program
requests to do so.
4) Restart the computer after installation is
completed.
* Select a single USB port and make sure to always use
the same port in the future.
* Do not use other Bluetooth receiver.
5) Turn on the Bluetooth function of the WatchBP
Office device on by switching the Bluetooth
switch at the back of the device to the “ON”
position. The Bluetooth icon will illuminate and
flash continuously to show a connection.
* It is noted that WatchBP office device will not
respond to any button pressing or switching when
the Bluetooth is on. It is to prevent accidental
interruptions during the measurement.
Restart
your
computer
DC 6V 600mADC 6V 600mA
29
Installation of the software program
1) Put the CD in the CD-ROM drive of your computer.
Then click on «setup.exe» in the CD’s directory.
2) Follow the instructions provided in the
installation window on the computer screen.
3) When the installation is complete, be sure restart
the computer before using the program for the
first time.
Controlling the device wirelessly
1) Turn on the Bluetooth switch located on the back
of the device.
2) Make sure the Bluetooth receiver is at right
position and the light is on (if applicable).
3) Launch WatchBP Office Software from the
computer. Follow these steps to control the
WatchBP Office from the computer.
Controlling the WatchBP Office wirelessly (cont.)
DC 6V 600mADC 6V 600mA
System Requirements: 550MHz CPU. 256MB Memory, 1024x768
pixel resolution, 256 color, CD-ROM drive, 1 free USB port, 40MB
free hard disk space, Microsoft Windows 2000 / XP / Vista.
Install CD
©
2
0
0
8
M
i
c
r
o
l
i
f
e
C
o
r
p
o
r
a
t
i
o
n
.
A
l
l
R
i
g
h
t
s
R
e
s
e
r
v
e
d
.
WatchBP Analyzer Office
Version 1.0.2.4
* System Requirements:
550MHz CPU. 256MB
Memory, 1024x768 pixel
resolution, 256 color,
CD-ROM drive, 1 free USB
port, 40MB free hard disk
space, Microsoft Windows
2000 / XP / Vista.
EN
30
Software commands
Choose measure blood pressure Click «1st VISIT», «FOLLOW-UP »
Enter Patient ID Input Patient ID
Start to perform Press the «Measure» button
Store blood pressure measurements Click «Save». The data is stored in an Excel data table. The file
name is generated automatically from the following format
«Patient ID+ WatchBP Office +measurement mode +
yyyymmdd.xls».
View the data Go to the folder where the file is saved and open the
corresponding excel file.
Print the data sheet While the corresponding Excel file is open, click print in Excel.
Disconnect the device Click <<Disconnect>> before existing the program
Close the program Click «Close»
* The software automatically searches for the nearest WatchBP device when the program is launched. If the device is
not located, confirm that the Bluetooth switch on the WatchBP Office is set to the ON position.
31
DC 6V 600mADC 6V 600mA
Rechargeable Battery
WatchBP Office features a built-in, rechargeable
Ni-MH battery pack that delivers up to 400~500
measurement cycles. The battery can be recharged
between use with the provided power adaptor.
Rechargeable battery and power adaptor
Using a power adaptor
Please use only the Microlife adaptor (DSA-20P-05
Adaptor) supplied with the WatchBP Office to
recharge the device.
1) Plug the adapter cable into the Power Plug in the
WatchBP Office device.
2) Plug the adaptor plug into the wall socket. When
the power adaptor is connected, no battery
power is consumed.
* An orange recharge
indicator means recharge
in progress.
* A green recharge
indicator means recharge
is completed.
mmHg
mmHg
SYS
mmHg
DIA
SYS
DIA
mmHg
PUL
/min
AUSCULT 1st FOLLOW UP
L/R
3 sec.
INTERVAL
3 sec.
7 sec.
HIDE
7 sec.
DC 6V 600mADC 6V 600mA
EN
32
Troubleshooting
Problem Possible cause How to make correction
No power(No LCD display) Power supply is not
properly plugged
Plug power supply into wall socket.
Battery is fully discharged Recharge the rechargeable battery by plugging in
the power supply.
Cuff does not inflate
properly
Loose connection of the
tube
Make sure the tube of the cuff is securely
connected to the device.
Leakage of the tube /
bladder
Check for cracks on the tube or the bladder. Please
contact Microlife customer service for this issue.
Hide, pressure, or pause
interval can not be set
Wireless option is ON The device can not be operated by the Start/Stop
and M Button when the Bluetooth switch is in thev
ON position. Please switch off the Bluetooth or use
the included software to control the device.
Unable to take
measurement
No result displayed
between measurements
Device is in AUSCULTATION
mode or the Hide function
is activated
Disable “hide” function, or use «1st VISIT» /
«FOLLOW-UP» mode to measure the blood
pressure.
Can’t access device via
computer
Bluetooth software not
installed
Make sure the Bluetooth software provided with
the device is installed on the computer.
33
If an error occurs during measurement, the
measurement is interrupted and an error
message «Err» is displayed.
• Please contact your local
Microlife service center if
the error persists.
• If you think the results
are unusual, please read
through the information
in this instruction manual
carefully.
Error Description Potential cause and
remedy
«Err 1»Signal too
weak
The pulse signals
on the cuff are too
weak. Re-position the
cuff and repeat the
measurement.
«Err 2»Error signal During the
measurement,
error signals were
detected by the cuff,
caused for instance by
movement or muscle
tension. Repeat the
measurement, keeping
your arm still.
Error messages
EN
34
«HI»Pulse or cuff
pressure too
high
The pressure in the
cuff is too high (over
300 mmHg) OR the
pulse is too high
(over 200 beats per
minute). Relax for 5
minutes and repeat the
measurement.
«LO»Pulse too
low
The pulse is too low
(less than 40 beats per
minute). Repeat the
measurement.
«Err 3»No pressure
in the cuff
An adequate pressure
cannot be generated
in the cuff. A leak
may have occurred.
Replace the batteries if
necessary. Repeat the
measurement.
«Err 5»Abnormal
result
The measuring signals
are inaccurate and no
result can therefore
be displayed. Read
through the checklist
for performing reliable
measurements and
then repeat the
measurement.
35
Protect the device from water and moisture
Protect the device from extreme heat and cold
Protect device from impact and drops
Protect the device from direct sunlight
Avoid proximity to electromagnetic fields, such
as those produced by mobile phones
Never open device
Safety, care, accuracy test and disposal
Safety and protection
This device may be used only for the purpose
described in this booklet. The device comprises of
sensitive components and must be treated with
caution. The manufacturer cannot be held liable for
damage caused by incorrect application.
Observe the storage and operating conditions
described in the “Technical specifications” section
of this manual.
• Ensure that children do not use the device
unsupervised; some parts are small enough to be
swallowed.
• Only activate the pump when cuff is installed.
• Do not use the device if you think it is damaged or
if anything appears unusual.
• Read the further safety instructions in the
individual sections of the instruction manual.
• Do not connect the device to a computer
until prompted to do so by the computer software.
EN
36
AUSCULT 1st FOLLOW UP
INTERVAL
PRESSURE
7 sec.
Do not wash the cuff!
Do not iron the cuff!
Accuracy test
We recommend the WatchBP Office device
be tested for accuracy every 2 years or after
mechanical impact (e.g. being dropped). Please
contact Microlife to arrange for an accuracy test.
Cleaning the cuff
DO NOT wash the cuff. DO NOT iron the cuff cover.
Device care
Clean the device with a soft, dry cloth.
Disposal
Batteries and electronic instruments must be
disposed of in accordance with the locally applicable
regulations, and not as domestic waste.
37
Technical specifications
Operation
temperature/humidity:•
10 - 40 °C / 50 - 104 °F
Storage temperature/
humidity:
•
-20 - +50 °C / -4 - +122 °F
•
15 - 90 % relative maximum humidity
Weight: •
1100 g (including rechargeable
battery pack)
Dimensions: •
200 x 125 x 90 mm
Measuring method: •
Oscillometric, corresponding to
Korotkoff
Measurement range: •
30 - 280 mmHg – blood pressure
•
40 - 200 beats per minute – pulse
Cuff pressure display: •
Range: 0 - 299 mmHg
•
Resolution: 1 mmHg
•
Static accuracy: pressure within ±3
mmHg
•
Pulse accuracy: ±5 % of the readout
value
Voltage source: •
Rechargeable battery pack;
4.8V C4000 mAh
•
Mains adapter DC 7.5V, 2 A
Reference to Standards: •
EN 1060-1 /-3, NIBP-requirements
•
IEC60601-1 General requirement for
safety
•
IEC60601-1-2 Requirements for EMC
•
EN1060-4, NIBP clinical investigation
•
ANSI/AAMI SP10, NIBP requirements
•
Bluetooth system to V2.0
•
FCC part15C
Federal Communications Commission (FCC) Statement
15.21
You are cautioned that changes or modifications not expressly
approved by the part responsible for compliance could void the
user’s authority to operate the equipment.
5.105(b)
This equipment has been tested and found to comply with the
limits for a Class B digital device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against
EN
38
harmful interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy and, if
not installed and used in accordance with the instructions, may
cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to
radio or television reception, which can be determined by turning
the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
-Reorient or relocate the receiving antenna.
-Increase the separation between the equipment and receiver.
-Connect the equipment into an outlet on a circuit different from that
to which the receiver is connected.
-Consult the dealer or an experienced radio/TV technician for help.
Operation is subject to the following two conditions:
1) this device may not cause interference and
2) this device must accept any interference, including interference
that may cause undesired operation of the device.
FCC RF Radiation Exposure Statement:
This equipment complies with FCC radiation exposure limits set forth
for an uncontrolled environment. End users must follow the specific
operating instructions for satisfying RF exposure compliance. This
transmitter must not be co-located or operating in conjunction with any
other antenna or transmitter.
This device is covered by a two-year
guarantee and accessories are cover by
a one-year guarantee from the date of
purchase. This guarantee is valid only
on presentation of the guarantee card
completed by the owner confirming date of
purchase or purchase receipt.
Product: WatchBP Office
Product Number: Twin200
Date:
Name:
Address:
Date:
Telephone:
Email:
Guarantee card
IB WatchBP Office 2508
Europe / Middle-East / Africa
Microlife WatchBP AG
Espenstrasse 139
9443 Widnau, Switzerland
Tel +41 71 727 7000
Fax +41 71 727 7011
Email: watchbp@microlife.ch
www.watchbp.com
Asia
Microlife Corporation
9F, 431, RuiGang Road, NeiHu
Taipei, 114, Taiwan, R.O.C.
Tel +886 2 8797 1288
Fax+886 2 8797 1283
Email: watchbp@microlife.com.tw
www.watchbp.com
North / Central / South America
Microlife Medical Home Solutions, Inc.
2801 Youngfield St., Suite 241
Golden, CO 80401, USA
Tel +1 303 274 2277
Fax +1 303 274 2244
Email: watchbp@mimhs.com
www.watchbp.com