Nihon Kohden ZM-540PA Medical Telemetry Transmitter User Manual OM ZM 540 541PA
Nihon Kohden Corporation Medical Telemetry Transmitter OM ZM 540 541PA
Users Manual
Transmitter ZM-540PA/ZM-541PA If you have any comments or suggestions on this manual, please contact us at: www.nihonkohden.com 0614-902754 Copyright Notice The entire contents of this manual are copyrighted by Nihon Kohden. All rights are reserved. No part of this document may be reproduced, stored, or transmitted in any form or by any means (electronic, mechanical, photocopied, recorded, or otherwise) without the prior written permission of Nihon Kohden. GENERAL HANDLING PRECAUTIONS This device is intended for use only by qualified medical personnel. Use only Nihon Kohden approved products with this device. Use of non-approved products or in a non-approved manner may affect the performance specifications of the device. This includes, but is not limited to, batteries, recording paper, pens, extension cables, electrode leads, input boxes and AC power. Please read these precautions thoroughly before attempting to operate the instrument. 1. To safely and effectively use the instrument, its operation must be fully understood. 2. When installing or storing the instrument, take the following precautions: (1) Avoid moisture or contact with water, extreme atmospheric pressure, excessive humidity and temperatures, poorly ventilated areas, and dust, saline or sulphuric air. (2) Place the instrument on an even, level floor. Avoid vibration and mechanical shock, even during transport. (3) Avoid placing in an area where chemicals are stored or where there is danger of gas leakage. (4) The power line source to be applied to the instrument must correspond in frequency and voltage to product specifications, and have sufficient current capacity. (5) Choose a room where a proper grounding facility is available. 3. Before Operation (1) Check that the instrument is in perfect operating order. (2) Check that the instrument is grounded properly. (3) Check that all cords are connected properly. (4) Pay extra attention when the instrument is in combination with other instruments to avoid misdiagnosis or other problems. (5) All circuitry used for direct patient connection must be doubly checked. (6) Check that battery level is acceptable and battery condition is good when using batteryoperated models. 4. During Operation (1) Both the instrument and the patient must receive continual, careful attention. (2) Turn power off or remove electrodes and/or transducers when necessary to assure the patient’s safety. (3) Avoid direct contact between the instrument housing and the patient. 5. To Shutdown After Use (1) Turn power off with all controls returned to their original positions. (2) Remove the cords gently; do not use force to remove them. (3) Clean the instrument together with all accessories for their next use. Operator’s Manual ZM-540PA/541PA 6. The instrument must receive expert, professional attention for maintenance and repairs. When the instrument is not functioning properly, it should be clearly marked to avoid operation while it is out of order. 7. The instrument must not be altered or modified in any way. 8. Maintenance and Inspection (1) The instrument and parts must undergo regular maintenance inspection at least every 6 months. (2) If stored for extended periods without being used, make sure prior to operation that the instrument is in perfect operating condition. (3) Technical information such as parts list, descriptions, calibration instructions or other information is available for qualified user technical personnel upon request from your Nihon Kohden representative. 9. When the instrument is used with an electrosurgical instrument, pay careful attention to the application and/or location of electrodes and/or transducers to avoid possible burn to the patient. 10. When the instrument is used with a defibrillator, make sure that the instrument is protected against defibrillator discharge. If not, remove patient cables and/or transducers from the instrument to avoid possible damage. ii Operator’s Manual ZM-540PA/541PA WARRANTY POLICY Nihon Kohden Corporation (NKC) shall warrant its products against all defects in materials and workmanship for one year from the date of delivery. However, consumable materials such as recording paper, ink, stylus and battery are excluded from the warranty. NKC or its authorized agents will repair or replace any products which prove to be defective during the warranty period, provided these products are used as prescribed by the operating instructions given in the operator’s and service manuals. No other party is authorized to make any warranty or assume liability for NKC’s products. NKC will not recognize any other warranty, either implied or in writing. In addition, service, technical modification or any other product change performed by someone other than NKC or its authorized agents without prior consent of NKC may be cause for voiding this warranty. Defective products or parts must be returned to NKC or its authorized agents, along with an explanation of the failure. Shipping costs must be pre-paid. This warranty does not apply to products that have been modified, disassembled, reinstalled or repaired without Nihon Kohden approval or which have been subjected to neglect or accident, damage due to accident, fire, lightning, vandalism, water or other casualty, improper installation or application, or on which the original identification marks have been removed. In the USA and Canada other warranty policies may apply. CAUTION United States law restricts this product to sale by or on the order of a physician. Operator’s Manual ZM-540PA/541PA iii Equipment Authorization Requirement This device complies with Part 15 of the FCC (Federal Communications Commission) Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. This device complies with Part 95 Subpart H of the FCC Rules to be used in wireless medical telemetry service. Operation of this equipment requires the prior coordination with a frequency coordinator designated by FCC for the Wireless Medical Telemetry Service. CAUTION To comply with the FCC radio frequency (RF) exposure compliance requirements, no change to the antenna or the device is permitted. Any change to the antenna or the device could result in the device, exceeding the RF exposure requirements and void user’s authority to operate this device. NOTE • Use this device only indoors. • This device has been tested and complies with FCC radiation exposure limits set forth for an uncontrolled environment. The RF transmission power from the antenna conforms to the general public FCC RF Exposure Guidelines in Supplement C to OET65 limit of Specific Absorption Rate (SAR) 1.6 W/kg. The maximum SAR value measured from this device was extremely smaller than 1.6 W/kg. This device must not be located together with or operated in conjunction with any other unspecified antenna or transmitter. • The devices require registration and deployment by an authorized frequency coordinator. The ASHE (American Society for Healthcare Engineering) has been designated by the FCC to manage the WMTS frequencies. This device has frequency bands which may not be used in some areas. For details, contact your Nihon Kohden representative. For details on the guidelines, refer to the ASHE home page. iv Operator’s Manual ZM-540PA/541PA EMC RELATED CAUTION This equipment and/or system complies with IEC 60601-2 International Standard for electromagnetic compatibility for medical electrical equipment and/or system. However, an electromagnetic environment that exceeds the limits or levels stipulated in IEC 60601-1-2, can cause harmful interference to the equipment and/or system or cause the equipment and/or system to fail to perform its intended function or degrade its intended performance. Therefore, during the operation of the equipment and/or system, if there is any undesired deviation from its intended operational performance, you must avoid, identify and resolve the adverse electromagnetic effect before continuing to use the equipment and/or system. The following describes some common interference sources and remedial actions: 1. Strong electromagnetic interference from a nearby emitter source such as an authorized radio station or cellular phone: Install the equipment and/or system at another location. Keep the emitter source such as cellular phone away from the equipment and/or system, or turn off the cellular phone. 2. Radio-frequency interference from other equipment through the AC power supply of the equipment and/or system: Identify the cause of this interference and if possible remove this interference source. If this is not possible, use a different power supply. 3. Effect of direct or indirect electrostatic discharge: Make sure all users and patients in contact with the equipment and/or system are free from direct or indirect electrostatic energy before using it. A humid room can help lessen this problem. 4. Electromagnetic interference with any radio wave receiver such as radio or television: If the equipment and/or system interferes with any radio wave receiver, locate the equipment and/or system as far as possible from the radio wave receiver. 5. Interference of lightning: When lightning occurs near the location where the equipment and/or system is installed, it may induce an excessive voltage in the equipment and/or system. In such a case, disconnect the AC power cord from the equipment and/or system and operate the equipment and/or system by battery power, or use an uninterruptible power supply. Operator’s Manual ZM-540PA/541PA 6. Use with other equipment: When the equipment and/or system is adjacent to or stacked with other equipment, the equipment and/or system may affect the other equipment. Before use, check that the equipment and/or system operates normally with the other equipment. 7. Use of unspecified accessory, transducer and/or cable: When an unspecified accessory, transducer and/or cable is connected to this equipment and/or system, it may cause increased electromagnetic emission or decreased electromagnetic immunity. The specified configuration of this equipment and/or system complies with the electromagnetic requirements with the specified configuration. Only use this equipment and/or system with the specified configuration. 8. Use of unspecified configuration: When the equipment and/or system is used with the unspecified system configuration different than the configuration of EMC testing, it may cause increased electromagnetic emission or decreased electromagnetic immunity. Only use this equipment and/or system with the specified configuration. 9. Measurement with excessive sensitivity: The equipment and/or system is designed to measure bioelectrical signals with a specified sensitivity. If the equipment and/or system is used with excessive sensitivity, artifact may appear by electromagnetic interference and this may cause mis-diagnosis. When unexpected artifact appears, inspect the surrounding electromagnetic conditions and remove this artifact source. If the above suggested remedial actions do not solve the problem, consult your Nihon Kohden representative for additional suggestions. vi Operator’s Manual ZM-540PA/541PA Conventions Used in this Manual and Instrument Warnings, Cautions and Notes Warnings, cautions and notes are used in this manual to alert or signal the reader to specific information. WARNING A warning alerts the user to the possible injury or death associated with the use or misuse of the instrument. CAUTION A caution alerts the user to possible injury or problems with the instrument associated with its use or misuse such as instrument malfunction, instrument failure, damage to the instrument, or damage to other property. NOTE A note provides specific information, in the form of recommendations, prerequirements, alternative methods or supplemental information. Operator’s Manual ZM-540PA/541PA vii Explanations of the Symbols in this Manual and Instrument The following symbols found in this manual/instrument bear the respective descriptions as given. On Panel Symbol Description Symbol Description Change screen Defibrillation proof type BF applied part Attention, consult operator’s manual Defibrillation proof type CF applied part Moves cursor, scrolls data Serial number Direction for attaching battery cover Year of manufacture Direction for inserting battery RF transmitter Non-ionizing radiation Direct current CSA mark On LCD Symbol viii Description Symbol Description Full battery Battery very weak Replace battery Battery 2/3 full Alarm suspended Low battery NIBP cannot be measured QRS/pulse sync mark Operator’s Manual ZM-540PA/541PA Intended Use General The ZM-540PA and ZM-541PA transmitter transmits ECG, respiration and pulse waveforms, SpO2 and NIBP data from a patient to a Nihon Kohden monitor for continuous monitoring. The front LCD displays ECG (or pulse wave), numeric values of monitoring parameters, NIBP measuring mode and interval, messages and battery condition.* It also displays the compressed waveform and numeric data of the latest 10 minutes. * Essential performance of this transmitter. The difference between the ZM-540PA and ZM-541PA is the transmission frequency range. ZM-540PA: 608.0250 MHz (channel number 9002) to 613.9750 MHz (channel number 9478) ZM-541PA: 1395.0250 MHz (channel number E002) to 1399.9750 MHz (channel number E398) 1427.0250 MHz (channel number E502) to 1431.9750 MHz (channel number E898) The transmitter channel can be changed with a QI-901PK channel writer. Read the operator’s manual for the receiving monitor together with this manual before use. WARNING Do not use the same transmitter on more than one patient at the same time. Do not connect different sensors on different patients to the same transmitter. CAUTION • Do not use the same channel for different patients. If the same channel is used for two patients, the two patients’ data will be lost due to mutual modulation interference, or another patient’s data may appear on the receiving monitor screen. • Do not use two transmitters with adjacent channels in the same hospital. If transmitters with adjacent channels are used, their radio waves interfere with each other. CAUTION Be aware that signal loss and artifact may occur because of the multipath cancellation* when using a transmitter. * Multipath Cancellation (Standing Wave Interference) When a radio wave reflects off a surface, there may be some points in the room where the Operator’s Manual ZM-540PA/541PA reflected and direct waves are exactly out of phase. At these points in the room, the reflected and direct waves cancel each other out so that the signal strength is 0. This is called “multipath cancellation” or “standing wave interference”. Locations where signal loss occurs are called “null spots”. If the transmitter is moving or nearby people or objects are moving, null spots can occur anytime and anywhere. NOTE • To prevent interference between channels, assign a channel administrator in the hospital and only he or she should manage channel assignment. • Use Nihon Kohden parts and accessories to assure maximum performance from your instrument. • For stable signal reception, it is recommended to use a diversity antenna system on the receiving monitor. Otherwise, spike noise from transient fading of electric field strength (for example, people moving) may interfere with the transmitter signal and may be mistaken as an arrhythmia on the receiving monitor. • For details on antennas and antenna construction, contact your Nihon Kohden representative. You can also refer to the Telemetry System Installation Guide. • Do not diagnose a patient based on only part of the monitoring data on the transmitter or only on the data acquired by the transmitter. Overall judgment must be performed by a physician who understands the features, limitations and characteristics of the transmitter by reading this operator’s manual thoroughly and by reading the biomedical signals acquired by other instruments. Receiving Monitor Any Nihon Kohden receiving monitor (central monitor with multiple patient receiver) can receive signals from this transmitter as long as the protocol version and channel setting are the same on the receiving monitor and transmitter. NOTE • For details on the receiving monitor and upgrade information, contact your Nihon Kohden representative. • The transmitter does not give any patient alarm, only a “no battery” alarm. Patient alarms must be managed on the receiving monitor. Operator’s Manual ZM-540PA/541PA Panel Description Front Panel Refer to the WARNING on the next page. No. Name Description LCD Displays numeric values, ECG or pulse wave, NIBP measuring mode and interval, messages and battery status. For details, refer to the “Screen Descriptions” section. Infrared receiver Used for upgrading the transmitter software. Screen key Toggles the screen in the following order. After power on: Start up → Check electrodes → Numeric and waveform → Waveform review → Numeric review → Display off → Check electrodes … After auto display off: Numeric and waveform → Waveform review → Numeric review → Display off → Check electrodes → Numeric and waveform … On a SETUP or CHECK screen, this key cancels changing setting or exits the screen. Operator’s Manual ZM-540PA/541PA 4 Refer to the WARNING below. No. Name Description Function key Depending on the setting on the transmitter, this key suspends alarms, pauses monitoring on the receiving monitor or transmits “Patient confirmed” message. On a SETUP screen, this key registers the selected setting and moves the cursor to the next setting item. On a CHECK screen, this key starts or stops maintenance test. NIBP Start/Stop key Starts/stops NIBP measurement in selected mode. NIBP Interval key Selects NIBP measurement mode. Lead/Scroll keys On the numeric and waveform screen, these keys change the ECG lead. On the waveform review screen, these keys scroll data. On a SETUP screen, these keys move the cursor. Battery case Contains three 1.5 V dry cell batteries (AA TYPE). WARNING Close the battery case cover during operation. If the transmitter is used with the battery case cover open, anyone who touches the opened battery case may receive an electrical shock when defibrillation is performed. Touching the opened battery case may cause electrostatic discharge and intermittently interfere with the waveform or data. Operator’s Manual ZM-540PA/541PA Rear Panel Lock plate Fastens the transmitter to an NIBP cuff. Refer to the symbol page. Refer to the WARNING below. WARNING This transmitter is not waterproof. If detergent or liquid spills into the transmitter, stop using it and contact your Nihon Kohden representative. If a wet transmitter is used, the patient or operator may receive an electrical shock or injury. Operator’s Manual ZM-540PA/541PA Top Panel Refer to the WARNING on the next page. Refer to the WARNING on the next page. Refer to the symbol page. NIBP socket: Connects the cuff hose. Refer to the symbol page. ECG/impedance RESP socket: Connects the electrode lead for measuring ECG and/or respiration (impedance). Bottom Panel Channel number label: Indicates the channel number of the transmitter. Attach the channel number label to the panel of the receiving monitor. Refer to the symbol page. Refer to the WARNING on the next page. SpO2 socket: Connects the SpO2 probe. Operator’s Manual ZM-540PA/541PA WARNING WARNING Before defibrillation, all persons must keep clear of the bed and must not touch the patient or any equipment or cord connected to the patient. Failure to follow this warning may cause electrical shock or injury. When performing defibrillation, discharge as far as possible from electrodes, patches and any gel, cream or medicine on the chest of the patient. If there is a possibility that the defibrillator paddle could touch these materials, remove them from the patient. If the defibrillator paddle directly contacts these materials, the discharged energy may cause skin burn to the patient. WARNING CAUTION When the transmitter is used with an electrosurgical unit (ESU), firmly attach the entire area of the ESU return plate. Otherwise, the current from the ESU flows into the electrodes of the transmitter, causing electrical burn where the electrodes are attached. For details, refer to the ESU manual. Do not shake or swing the transmitter while holding the leads or cables connected to the transmitter. The transmitter may come off and injure someone or damage surrounding instruments. Operator’s Manual ZM-540PA/541PA Important Safety Information General WARNING WARNING Never use the transmitter in the presence of any flammable anesthetic gas or high concentration oxygen atmosphere. Failure to follow this warning may cause explosion or fire. Never use the transmitter in a hyperbaric oxygen chamber. Failure to follow this warning may cause explosion or fire. WARNING WARNING Do not take this transmitter into the MRI test room. This transmitter is not designed to be used during MRI tests. When performing MRI test, remove all electrodes, probe and cuff from the patient which are used with this transmitter. Failure to follow this warning may cause skin burn on the patient. For details, refer to the MRI manual. WARNING WARNING Before defibrillation, all persons must keep clear of the bed and must not touch the patient or any equipment or cord connected to the patient. Failure to follow this warning may cause electrical shock or injury. When performing defibrillation, discharge as far as possible from electrodes, patches and any gel, cream or medicine on the chest of the patient. If there is a possibility that the defibrillator paddle could touch these materials, remove them from the patient. If the defibrillator paddle directly contacts these materials, the discharged energy may cause skin burn to the patient. Operator’s Manual ZM-540PA/541PA WARNING WARNING When the transmitter is used with an electrosurgical unit (ESU), firmly attach the entire area of the ESU return plate. Otherwise, the current from the ESU flows into the electrodes of the transmitter, causing electrical burn where the electrodes are attached. For details, refer to the ESU manual. This transmitter is not waterproof. If detergent or liquid spills into the transmitter, stop using it and contact your Nihon Kohden representative. If a wet transmitter is used, the patient or operator may receive an electrical shock or injury. WARNING WARNING Close the battery case cover during operation. If the transmitter is used with the battery case cover open, anyone who touches the opened battery case may receive an electrical shock when defibrillation is performed. Touching the opened battery case may cause electrostatic discharge and intermittently interfere with the waveform or data. Do not use the same transmitter on more than one patient at the same time. Do not connect different sensors on different patients to the same transmitter. CAUTION Only use Nihon Kohden specified electrodes, electrode leads, SpO2 probes, and NIBP cuffs. Otherwise, the maximum performance from the transmitter cannot be guaranteed. CAUTION Do not reuse disposable parts and accessories. CAUTION CAUTION The measurement values and displayed waveforms on the transmitter and receiving monitor may be different due to timing delay of the display or difference in detection settings. Do not shake or swing the transmitter while holding the leads or cables connected to the transmitter. The transmitter may come off and injure someone or damage surrounding instruments. Operator’s Manual ZM-540PA/541PA CAUTION CAUTION Be aware that signal loss and artifact may occur because of the multipath cancellation* when using a transmitter. Turn off the power of mobile phones, small wireless devices and other devices which produce strong electromagnetic interference around a patient (except for devices allowed by the hospital administrator). Radio waves from devices such as mobile phones or small wireless devices may be mistaken as pulse waves and the displayed data may be incorrect. * Multipath Cancellation (Standing Wave Interference) When a radio wave reflects off a surface, there may be some points in the room where the reflected and direct waves are exactly out of phase. At these points in the room, the reflected and direct waves cancel each other out so that the signal strength is 0. This is called “multipath cancellation” or “standing wave interference”. Locations where signal loss occurs are called “null spots”. If the transmitter is moving or nearby people or objects are moving, null spots can occur anytime and anywhere. CAUTION • Do not use the same channel for different patients. If the same channel is used for two patients, the two patients’ data will be lost due to mutual modulation interference, or another patient’s data may appear on the receiving monitor screen. • Do not use two transmitters with adjacent channels in the same hospital. If transmitters with adjacent channels are used, their radio waves interfere with each other. Output Signal WARNING Do not use the output signal from the receiving monitor as the synchronization signal for other equipment such as IABP, MRI, echocardiography or defibrillator. There may be time delay between the monitor and the other equipment caused by waveform transmission delay and spike noise may interfere on the output signal and be mistaken as a trigger. 10 Operator’s Manual ZM-540PA/541PA Battery CAUTION CAUTION Battery replacement must be performed by the operator. When replacing the batteries of a transmitter that is currently used for a patient, disconnect the electrode leads from the transmitter before replacing batteries or do not touch the patient during replacement. Refer to the battery and battery charger manuals for details on handling the batteries. Transmitter Channel Management WARNING The following actions must be taken to properly receive the transmitter signal of the correct patient on the receiving instrument. Otherwise, there may be signal loss or signals may mix causing a serious accident, such as monitoring a different patient. • Assign a channel administrator in the hospital and only he or she should manage channel assignment. • The channel administrator must manage the channels in the facility so that there is no signal interference. • When the transmitter channel is changed, the channel administrator must check that the channel on the receiving monitor is also changed and the signal is properly received. • The channel administrator must replace the channel number label on the transmitter with the new one after changing the channel. Operator’s Manual ZM-540PA/541PA 11 For Patients Using Implantable Pacemaker WARNING Interaction Between Minute Ventilation Rate-Adaptive Pacemakers and Cardiac Monitoring and Diagnostic Equipment* The bioelectric impedance measurement sensor of a minute ventilation rate-adaptive implantable pacemaker may be affected by transmitter which is connected to the same patient. If this occurs, the pacemaker may pace at its maximum rate and the transmitter may give incorrect data to the monitor. If this occurs, disconnect the electrode leads from the patient or change the setting on the pacemaker by referring to the pacemaker’s manual. For more details, contact your pacemaker representative or Nihon Kohden representative. * Minute ventilation is sensed in rate-adaptive pacemakers by a technology known as bioelectric impedance measurement (BIM). Many medical devices in addition to pacemakers use this technology. When one of these devices is used on a patient with an active, minute ventilation rateadaptive pacemaker, the pacemaker may erroneously interpret the mixture of BIM signals created in the patient, resulting in an elevated pacing rate. For more information, see the FDA web site. http://www.fda.gov/cdrh/safety.html ECG Monitoring 12 CAUTION CAUTION Only use Nihon Kohden specified electrodes and electrode leads. When other electrodes or electrode leads are used, the “CHECK ELECTRODES” message may appear and monitoring may stop. When the “ELECTRODE OFF” or “CHECK ELECTRODES” message is displayed on the receiving monitor, ECG is not monitored properly and the ECG alarm does not function. Check the electrode, electrode leads, and if necessary, replace with new ones. Operator’s Manual ZM-540PA/541PA SpO2 Monitoring WARNING WARNING SpO2 measurement may be incorrect in the following cases. • When the patient’s carboxyhemoglobin or methemoglobin increases abnormally. • When dye is injected in the blood. • When using an electrosurgical unit. • During CPR. • When measuring at a site with venous pulse. • When there is body movement. • When the pulse wave is small (insufficient peripheral circulation). When not monitoring SpO2, disconnect the SpO2 cable from the transmitter. Otherwise, noise from the probe sensor may interfere and incorrect data is displayed on the screen. WARNING WARNING • When using the TL-201T finger probe, do not fasten the probe and cable to the finger by wrapping with tape. This may cause burn, congestion or pressure necrosis from poor blood circulation. • When using probes other than the TL-201T finger probe, to avoid poor circulation, do not wrap the tape too tight. Check the blood circulation condition by observing the skin color and congestion at the skin peripheral to the probe attachment site. Even for short-term monitoring, there may be burn or pressure necrosis from poor blood circulation, especially on neonates or low birth weight infants whose skin is delicate. Accurate measurement cannot be performed on a site with poor peripheral circulation. Check the circulation condition by observing the skin color at the measurement site and pulse waveform. Change the measurement site every 8 hours for disposable probes and every 4 hours for reusable probes (every 8 hours for TL-630T/TL631T series probe). The skin temperature may increase at the attached site by 2 or 3°C (4 or 5°F) and cause a burn or pressure necrosis. When using the probe on the following patients, take extreme care and change the measurement site more frequently according to symptoms and degree. • Patient with a fever • Patient with peripheral circulation insufficiency • Neonate or low birth weight infant with delicate skin Operator’s Manual ZM-540PA/541PA 13 WARNING When monitoring SpO2 of a patient who is receiving photodynamic therapy, the light from the finger probe sensor may cause a burn. Photodynamic therapy uses a photosensitizing agent that has a side effect of photosensitivity. CAUTION CAUTION Refer to the probe instruction manual for details. The disposable probe is not sterilized. Use the disposable probe only for a single patient. Never reuse the disposable probe for another patient because it causes cross infection. CAUTION CAUTION NIBP and SpO2 can be measured on the same limb, but the SpO2 monitoring might not be accurate during NIBP measurement. Be careful when reading the SpO2 values.* While a patient is on medication which causes vasodilation, the pulse waveform may change and in rare cases the SpO2 value might not be displayed. * Monitoring SpO2 during NIBP Measurement When the SpO2 probe is attached to the same limb as the NIBP cuff, the blood flow decreases during NIBP measurement and pulse wave cannot be detected and SpO2 cannot be monitored properly. When “INHIBIT SpO2 DURING NIBP” on the PARAMETER SETUP screen is set to ON (factory default setting), SpO2 monitoring is paused during NIBP measurement to avoid SpO2 alarm occurrence. However, when monitoring SpO2 on the same limb as the NIBP, be careful when reading SpO2 values. 14 CAUTION Normal external light does not affect monitoring but strong light such as a surgical light or sunlight may affect monitoring. If affected, cover the measuring site with a blanket. CAUTION When a message indicates a faulty probe, stop monitoring and replace the probe with a new one. Operator’s Manual ZM-540PA/541PA CAUTION Do not use a damaged or disassembled probe. It causes incorrect measurement and may injure the patient. CAUTION If the attachment site is dirty with blood or bodily fluids, clean the attachment site before attaching the probe. If there is nail polish on the attachment site, remove the polish. Otherwise, the amount of transmitted light decreases, and measured value may be incorrect or measurement cannot be performed. CAUTION Do not pull or bend the probe cable, and do not put caster feet on the probe cable. Do not immerse the disposable probe cable in chemical solutions or water. Failure to follow these instructions may cause cable discontinuity, short circuit, skin burn on the patient and incorrect measurement data. Replace any broken probe with a new one. Operator’s Manual ZM-540PA/541PA CAUTION Do not use a probe which is deteriorated by aging. Accurate measurement cannot be performed. CAUTION If the skin gets irritated or redness appears on the skin from the probe, change the attachment site or stop using the probe. Take extreme care for the patients with delicate skin. CAUTION When the probe is attached on an appropriate site with sufficient circulation and the error message confirming the probe attachment repeatedly appears, the probe may be deteriorated. Replace it with a new one. 15 NIBP Monitoring WARNING Be careful when measuring NIBP on a patient with known bleeding disorders or coagulation. After NIBP measurement, there may be dot hemorrhage, or circulatory disorder by thrombus where the cuff is attached. WARNING When performing NIBP measurements in STAT mode or 5 minute intervals, periodically remove the cuff from the patient for ventilation. The skin temperature may increase at the cuff attachment site by 2 or 3°C (4 or 5°F). When measuring a patient with a fever or peripheral circulation insufficiency, it may cause a burn. CAUTION CAUTION Do not wrap the cuff on an arm or thigh which is used for injection. NIBP measurement on an arm or thigh which is used for injection may cause reflux of blood and stop injection. Do not wrap the cuff too tight. It may cause poor blood circulation and congestion. If the cuff is wrapped too loosely, the NIBP value may increase. CAUTION Do not attach the cuff to the site where there is injury or inflammation. If the skin gets irritated or redness appears on the skin from the cuff, change the attachment site or stop using the cuff. Take extreme care on the patients with delicate skin. 16 WARNING NIBP measurement may be incorrect in the following cases. • When using an electrosurgical unit • When there is body movement • When the pulse wave is small (insufficient peripheral circulation) • Too many arrhythmias • When there is vibration • When there is a rapid blood pressure change • During CPR CAUTION When using an extension hose, check that the extension hose is not bent or squeezed. Otherwise, the cuff might not inflate or deflate. If the cuff cannot deflate, it may cause congestion on the patient at the cuff attachment site. Operator’s Manual ZM-540PA/541PA CAUTION When performing NIBP measurement repeatedly, have a rest between measurements to recover adequate circulation. Maintenance CAUTION This transmitter is not waterproof. If detergent or liquid spills into the transmitter, stop cleaning or disinfecting it and contact your Nihon Kohden representative. The transmitter needs to be checked for safety and function before use. Operator’s Manual ZM-540PA/541PA CAUTION Never disassemble or repair the transmitter. Disassembly and repair must be performed by qualified service personnel. 17 Preparation on Transmitter Batteries WARNING and CAUTION for Handling Batteries CAUTION CAUTION Refer to the battery and battery charger manuals for details on handling the batteries. When the transmitter is not in use, remove the batteries. When the batteries are installed, battery power is consumed even when measurement is not performed. When NiMH batteries are left in the transmitter when it is not used, the batteries may overdischarge and leak liquid which makes the batteries unusable and damages the transmitter. CAUTION Do not handle the batteries with wet hands. NOTE Remove the batteries from the transmitter before disposing of it. Battery Lifetime Use three AA type alkaline dry cell batteries. NiMH batteries can also be used but the lifetime of alkaline batteries is longer. The lifetime of NiMH batteries is about 1/2 of alkaline batteries (when fully charged). With new Nihon Kohden recommended alkaline batteries, the transmitter can continuously measure ECG, respiration, SpO2 and NIBP for approximately 1 day. The measurement is performed at room temperature, NIBP is measured in auto mode at 60 minute intervals and SpO2 is measured on an index finger of a male patient with weight 60 kg. Operation time depends on the thickness of the SpO2 probe attachment site. Recommended batteries NiMH secondary: SANYO HR-3UF (W) Battery charger: SANYO NC-M55 Alkaline primary: Nihon Kohden Medipower (equivalent to Panasonic LR6 (G)) 18 Operator’s Manual ZM-540PA/541PA Installing (Replacing) Batteries CAUTION Battery replacement must be performed by the operator. When replacing the batteries of a transmitter that is currently used for a patient, disconnect the electrode leads from the transmitter before replacing batteries or do not touch the patient during replacement. If electrode leads are attached to the patient and the person replacing batteries touches the patient during battery replacement, excess patient leakage current may flow into the patient. NOTE • • • • Replace all batteries at the same time. Do not use different types of batteries together. Insert the batteries with the correct polarity (+ and –). The capacity of rechargeable NiMH batteries is reduced if the batteries are recharged before they are fully discharged. For details, refer to the battery operator’s manual. Operator’s Manual ZM-540PA/541PA 1. Remove the battery case cover. 2. Insert three new or fully charged batteries into the battery case observing the correct polarity. 19 3. Close the cover. The transmitter is automatically turned on when the batteries are installed and the cover is closed. Situations Requiring Battery Replacement Replace the batteries when any of the following occurs: • The transmitter displays the “BATTERY WEAK” message or “ ” icon. • The transmitter generates a constant alarm (continuous “peep” sound). • The transmitter LCD does not display anything when the power is turned on. • The receiving monitor displays a battery replacement message. Battery Level Indication The following icons on the LCD indicate the battery level. When “PROTOCOL” on the SYSTEM SETUP screen is set to 57, the battery level indication is transmitted to the receiving monitor. Indication Battery Level Message on the Receiving Monitor Fully charged batteries Batteries are 2/3 full. There is no message on the monitor. Batteries are low. NIBP cannot be measured. Replace batteries. Batteries are very weak. No indication Dead batteries Message requiring battery replacement is displayed. No signal can be transmitted to the monitor. There is no indication on the monitor. Turning On the Transmitter When the batteries are installed correctly, the power is turned on. A one second “peep” sounds and the startup screen appears, then the check electrodes screen appears. (There is no “peep” sound when there is no battery power.) 20 Operator’s Manual ZM-540PA/541PA After checking that the ECG is stable on the check electrodes screen, press the Screen key to display the numeric and waveform screen. For details on the screen, refer to the “Screen Descriptions” section. Check Items Before Use Before turning on the transmitter power, check the following to confirm that the transmitter can be used in normal and safe condition. Appearance • There are no damaged or dirty parts on the outside of the transmitter (LCD, keys, sockets, battery case cover, battery case, lock plate, etc.). • The transmitter is completely dry. • The electrodes, electrode lead, SpO2 probe and NIBP cuff are not broken. Batteries • The battery polarity is correct. • The battery case spring is firmly attached and the battery is not loose. • The battery case cover is firmly closed. Channel Setting • The transmitter channel matches the receiving monitor channel. • There is no nearby transmitter with the same channel. Operator’s Manual ZM-540PA/541PA 21 Check Items After Power On After turning on the power, check the following. Power On • The transmitter generates a one second “peep” sound and the startup screen appears. • The transmitter displays the check electrodes screen. • The transmitter is not too hot. • The transmitter does not display the “BATTERY WEAK” message. • The transmitter does not interfere with the operation of other medical instruments. Daily Check • The “signal loss” message is not displayed on the receiving monitor when the transmitter is inside the receiving range of the monitor. • The battery replacement message is not displayed on the monitor. • The keys on the transmitter function properly. • The LCD brightness is appropriate. To adjust brightness, refer to the “Changing SYSTEM SETUP Settings” section. Check Items After Use To use the transmitter in safe and optimum condition for next time, check the following. Before Turning Power Off • Temporarily changed settings are changed back to the previous settings. • There was no malfunction on the transmitter. Storage • ECG electrode leads, SpO2 probe and NIBP cuff are cleaned and disinfected. • If the transmitter got wet, liquid is wiped off and the transmitter is thoroughly dried. • There are enough consumables, such as disposable electrodes. • The transmitter power is turned off by removing batteries from the transmitter. • Dead batteries are disposed of properly. Turning Off the Transmitter To turn off the power, remove batteries. When the power is turned off, the saved waveform and numeric data are deleted. 22 Operator’s Manual ZM-540PA/541PA Changing the Transmitter Channel The channel of the transmitter can be changed with an optional QI-901PK Channel Writer. WARNING The following actions must be taken to properly receive the transmitter signal of the correct patient on the receiving instrument. Otherwise, there may be signal loss or signals may mix causing a serious accident, such as monitoring a different patient. • Assign a channel administrator in the hospital and only he or she should manage channel assignment. • The channel administrator must manage the channels in the facility so that there is no signal interference. • When the transmitter channel is changed, the channel administrator must check that the channel on the receiving monitor is also changed and the signal is properly received. • The channel administrator must replace the channel number label on the transmitter with the new one after changing the channel. NOTE • The software version of the QI-901PK channel writer must be 02-01 or later to change the channel on the transmitter. • The channel writer must be used outside the patient environment. The channel is displayed in the upper left corner of the screen. Channel Operator’s Manual ZM-540PA/541PA 23 Changing Parameter and System Setup Settings The initial settings on the PARAMETER SETUP and SYSTEM SETUP screens can only be changed before monitoring. Changing these settings during monitoring interrupts monitoring. NOTE Changing Parameter and System Setup settings must be done by qualified personnel. Notes on Parameter Settings When monitoring NIBP and SpO2, the following setting must be set as indicated in the table to properly transmit the monitoring data to the receiving monitor. Otherwise, SpO2 cannot be monitored properly during NIBP measurement. Some receiving monitors require the software to be upgraded. For details, contact your Nihon Kohden representative. SpO2 probe attachment site INHIBIT SpO2 DURING NIBP setting Probe attached to the same limb as the cuff* ON Probe attached to the limb without cuff OFF * When the SpO2 probe is attached to the same limb as the NIBP cuff and the cuff is inflated, the SpO2 value becomes unstable and SpO2 or PR alarm may occur. 24 Operator’s Manual ZM-540PA/541PA Changing PARAMETER SETUP Settings Parameter Setup Setting List The factory default settings are underlined. Setting Item Description ECG ELECTRODES Select the electrode lead type. LEAD TYPE Select the type of ECG leads. Turn ECG monitoring on or off. When electrodes are attached to the patient and ECG leads are connected, ECG monitoring starts even when this setting is set to OFF. If this setting is set to OFF, the same setting on the receiving monitor must also be set to OFF. ECG MEASUREMENT NOTE When “PROTOCOL” on the transmitter is set to 57 and the receiving monitor is able to receive protocol 57, ECG measurement on the receiving monitor is automatically set to OFF when this setting is set to OFF on the transmitter. Turn respiration monitoring on or off. RESP When this setting is set to OFF, the same setting on the MEASUREMENT receiving monitor is automatically set to OFF. Select the SpO2 response mode. FAST: Select this for special applications that require a fast response. FAST is suitable for detecting short apnea. SpO2 RESPONSE NORMAL: For normal monitoring. SLOW: Select this when you need to suppress a rapid change in SpO2. INHIBIT SpO2 Turn SpO2 monitoring on or off during NIBP DURING NIBP measurement. SELECTABLE Select the NIBP measurement modes for the mode INTERVALS (min) selection. INITIAL INTERVAL Select the initial NIBP measurement mode at power (min) on. NIBP MODE AFTER Select the NIBP measurement mode after completing STAT (min) STAT measurement. START/FINISH SOUND Turn ON or OFF the sound for NIBP measurement start/finish. Operator’s Manual ZM-540PA/541PA Settings IEC, AHA AUTO, 6 LEADS ON, OFF ON, OFF FAST, NORMAL, SLOW ON, OFF STAT, 5, 10, 15, 30, 60, 120, 240 MANUAL, 5, 10, 15, 30, 60, 120, 240 MANUAL, 5, 10, 15, 30, 60, 120, 240 START: ON, OFF FINISH: ON, OFF 25 Setting Item Description Settings Select whether to hide or dim the NIBP data after DATA: HIDE, OLD NIBP DATA/ measurement and how long to wait after measurement DIM AFTER (min) to dim or hide it. AFTER: 5, 10, 30 INITIAL CUFF 120, 150, 180, Select the NIBP cuff inflation pressure. PRESSURE (mmHg) 210, 240 Displaying the PARAMETER SETUP Screen 1. Turn off the transmitter by removing one battery. 2. While pressing the Function key, turn on the transmitter (install the battery). The MENU screen appears. 3. Press the key to move the cursor to “PARAMETER SETUP”. 4. Press the Function key to enter PARAMETER SETUP. The current settings are highlighted. 5. Change settings: • To move the cursor and select the setting item, press the or key then press the Function key. • To select and register the setting, press the or key then press the Function key. • To cancel changing the setting of the selected item, press the Screen key. 6. When changing settings on the PARAMETER SETUP screen is complete, press the Screen key to return to the MENU screen. Cursor Setting item Setting ▼ ▼ 7. Press the 8. Press the Function key. The numeric and waveform screen appears. 26 or ▼ ▼ ▼ PARAMETER SETUP screen - page 1 ▼ MENU screen ▼ Cursor key to move the cursor to “EXIT”. Operator’s Manual ZM-540PA/541PA Changing Parameter Setup Settings ECG ELECTRODES Select the electrode lead type. Cursor On the PARAMETER SETUP screen, press the key to move the cursor to “ECG ELECTRODES”. 2. Press the Function key. The cursor moves to the selection item. 3. Press the 4. Press the Function key to register the selected setting. The cursor returns to “ECG ELECTRODES”. ▼ 1. Selected setting ▼ Cursor Setting item key to select “IEC” or “AHA”. ▼ LEAD TYPE Select the type of ECG leads. In normal use, select “AUTO”. When using DIN type lead with 6 electrodes, select “6 LEADS”. 1. On the PARAMETER SETUP screen, press the 2. Press the Function key. 3. Press the 4. Press the Function key to register the selected setting. The cursor returns to “LEAD TYPE”. ▼ key to move the cursor to “LEAD TYPE”. key to select “AUTO” or “6 LEADS”. Operator’s Manual ZM-540PA/541PA 27 ECG MEASUREMENT Turn ECG monitoring on or off. When electrodes are attached to the patient and ECG leads are connected, ECG monitoring starts even when this setting is set to OFF. If this setting is set to OFF, the same setting on the receiving monitor must also be set to OFF. NOTE ▼ When “PROTOCOL” on the transmitter is set to 57 and the receiving monitor is able to receive protocol 57, ECG measurement on the receiving monitor is automatically set to OFF when this setting is set to OFF on the transmitter. 1. On the PARAMETER SETUP screen, press the MEASUREMENT”. 2. Press the Function key. 3. Press the 4. Press the Function key to register the selected setting. The cursor returns to “ECG MEASUREMENT”. ▼ key to move the cursor to “ECG key to select “ON” or “OFF”. ▼ RESP MEASUREMENT Turn respiration monitoring on or off. When this setting is set to OFF, the same setting on the receiving monitor is automatically set to OFF. 1. On the PARAMETER SETUP screen, press the MEASUREMENT”. 2. Press the Function key. 3. Press the 4. Press the Function key to register the selected setting. The cursor returns to “RESP MEASUREMENT”. 28 ▼ key to move the cursor to “RESP key to select “ON” or “OFF”. Operator’s Manual ZM-540PA/541PA SpO2 RESPONSE Select the SpO2 response mode. FAST: Select this for special applications that require a fast response. FAST is suitable for detecting short apnea. NORMAL: For normal monitoring. SLOW: Select this when you need to suppress a rapid change in SpO2. On the PARAMETER SETUP screen, press the key to move the cursor to “SpO2 RESPONSE”. “SpO2 RESPONSE” is on the second page of the PARAMETER SETUP screen. 2. Press the Function key. 3. Press the key to select “FAST”, “NORMAL” or “SLOW”. 4. Press the Function key to register the selected setting. The cursor returns to “SpO2 RESPONSE”. PARAMETER SETUP screen - page 2 ▼ ▼ 1. INHIBIT SpO2 DURING NIBP Turn SpO2 monitoring on or off during NIBP measurement. When the SpO2 probe is attached to the same limb as the NIBP cuff and this setting is set to OFF, the pulse may become unstable and SpO2 or PR alarm may occur. It is recommended to set this setting to ON so that SpO2 is not measured during NIBP measurement. When the SpO2 probe is attached to the other limb from the NIBP cuff, this setting can be set to OFF. NOTE ▼ When this “INHIBIT SpO2 DURING NIBP” is set to OFF, refer to the “Monitoring SpO2 during NIBP Measurement” section. On the PARAMETER SETUP screen, press the key to move the cursor to “INHIBIT SpO2 DURING NIBP”. “INHIBIT SpO2 DURING NIBP” is on the second page of the PARAMETER SETUP screen. 2. Press the Function key. 3. Press the ▼ 1. key to select “ON” or “OFF”. Operator’s Manual ZM-540PA/541PA 29 4. Press the Function key to register the selected setting. The cursor returns to “INHIBIT SpO2 DURING NIBP”. SELECTABLE INTERVALS (min) When the NIBP INTERVAL key is pressed, the measurement mode changes according to the modes selected in this item. MANUAL mode is already selected for the mode selection. On the PARAMETER SETUP screen, press the key to move the cursor to “SELECTABLE INTERVALS”. “SELECTABLE INTERVALS” is on the third page of the PARAMETER SETUP screen. 2. Press the Function key. 3. Press the key to select or unselect the mode. The selected modes are highlighted. 4. Press the Function key to register the selected setting. The cursor returns to “SELECTABLE INTERVALS”. PARAMETER SETUP screen - page 3 ▼ ▼ 1. ▼ INITIAL INTERVAL (min) Select the initial NIBP measurement mode at power on. On the PARAMETER SETUP screen, press the key to move the cursor to “INITIAL INTERVAL”. “INITIAL INTERVAL” is on the third page of the PARAMETER SETUP screen. 2. Press the Function key. 3. Press the key to select the mode. Only the mode or interval selected for “SELECTABLE INTERVALS” are available. 4. Press the Function key to register the selected setting. The cursor returns to “INITIAL INTERVAL”. ▼ 1. 1. 30 ▼ NIBP MODE AFTER STAT (min) Select the NIBP measurement mode after completing the STAT measurement. On the PARAMETER SETUP screen, press the key to move the cursor to “NIBP MODE AFTER STAT”. “NIBP MODE AFTER STAT” is on the third page of the PARAMETER SETUP screen. Operator’s Manual ZM-540PA/541PA Press the Function key. 3. Press the key to select the mode. Only the mode or interval selected for “SELECTABLE INTERVALS” are available. 4. Press the Function key to register the selected setting. The cursor returns to “NIBP MODE AFTER STAT”. ▼ 2. START/FINISH SOUND Turn on or off the sound for NIBP measurement start and finish. On the PARAMETER SETUP screen, press the key to move the cursor to “START/FINISH SOUND”. “START/ FINISH SOUND” is on the fourth page of the PARAMETER SETUP screen. 2. Press the Function key. The cursor moves to “START”. 3. Press the key to turn “ON” or “OFF” the sound for NIBP measurement start. 4. Press the Function key to register the setting for “START”. The cursor moves to “FINISH”. 5. Press the key to turn “ON” or “OFF” the sound for NIBP measurement finish. 6. Press the Function key to register the selected setting. The cursor returns to “START/FINISH SOUND”. ▼ ▼ 1. ▼ PARAMETER SETUP screen - page 4 ▼ OLD NIBP DATA/AFTER (min) Select whether to dim or hide the NIBP data after measurement and how long to wait after NIBP measurement to dim or hide it. 1. On the PARAMETER SETUP screen, press the key to move the cursor to “OLD NIBP DATA/AFTER”. “OLD NIBP DATA/AFTER” is on the fourth page of the PARAMETER SETUP screen. 2. Press the Function key. The cursor moves to “DATA”. Operator’s Manual ZM-540PA/541PA 31 ▼ key to select “HIDE” or “DIM” the NIBP data. Press the 4. Press the Function key to register the setting for “DATA”. The cursor moves to “AFTER”. 5. Press the 6. Press the Function key to register the selected setting. The cursor returns to “OLD NIBP DATA/ AFTER”. ▼ 3. key to select the interval after NIBP measurement to dim or hide. ▼ INITIAL CUFF PRESSURE (mmHg) Select the NIBP cuff inflation pressure. On the PARAMETER SETUP screen, press the key to move the cursor to “INITIAL CUFF PRESSURE”. “INITIAL CUFF PRESSURE” is on the fourth page of the PARAMETER SETUP screen. 2. Press the Function key. 3. Press the 4. Press the Function key to register the selected setting. The cursor returns to “INITIAL CUFF PRESSURE”. ▼ 1. key to select the inflation pressure. Changing SYSTEM SETUP Settings System Setup Setting List The factory default settings are underlined. Setting Item PROTOCOL Description Select the transmitting protocol. 57: New protocol. A central monitor with ORG9100A/9110A/9700A multiple patient receiver whose software version 02-01 or later can receive this protocol. 42: Old protocol. A central monitor with ORG9100A/9110A/9700A multiple patient receiver can receive this protocol. Settings 57, 42 NOTE When 57 is set, the receiving monitor must be able to receive protocol 57. Otherwise, signals from the transmitter cannot be received. 32 Operator’s Manual ZM-540PA/541PA Setting Item Description BRIGHTNESS Select the screen brightness. PRESSURE UNIT Select the unit for NIBP. Select the function of the Function key. SUSPEND ALARM & PAUSE: Suspends alarm on the receiving monitor for 2 minutes. Pauses monitoring on the transmitter and receiving monitor. SUSPEND ALARM: Suspends alarm on the receiving monitor for 2 minutes. CONFIRM: Displays the “PATIENT CONFIRMED” message on the transmitter screen and transmits the message to the receiving monitor. OFF: No function. FUNCTION KEY Settings DARK, BRIGHT mmHg, kPa SUSPEND ALARM & PAUSE, SUSPEND ALARM, CONFIRM, OFF NOTE “SUSPEND ALARM & PAUSE” and “CONFIRM” can only be set when PROTOCOL is set to 57. AUTO RESUME AFTER PAUSE Select the interval to resume monitoring after PAUSE. Operator’s Manual ZM-540PA/541PA 10 s, 30 s, 1 min, 2 min, 3 min, OFF 33 Displaying the SYSTEM SETUP Screen 1. Turn off the transmitter by removing one battery. 2. While pressing the Function key, turn on the transmitter (install the battery). The MENU screen appears. 3. Press the key to move the cursor to “SYSTEM SETUP”. 4. Press the Function key to enter SYSTEM SETUP. The current settings are highlighted. 5. Change settings. • To move the cursor and select the setting item, press the or key then press the Function key. • To select and register the setting, press the or key then press the Function key. • To cancel changing the setting of the selected item, press the Screen key. ▼ Cursor Setting ▼ ▼ ▼ Setting item ▼ MENU screen ▼ The SYSTEM SETUP screen has two pages. To display the second page, press the key when the cursor is at “PRESSURE UNIT”. SYSTEM SETUP screen - page 1 7. Press the “EXIT”. 8. Press the Function key. The numeric and waveform screen appears. or ▼ When changing settings on the SYSTEM SETUP screen is complete, press the Screen key to return to the MENU screen. ▼ 34 6. key to move the cursor to Operator’s Manual ZM-540PA/541PA Changing System Setup Settings PROTOCOL Select the transmitting protocol. For differences between protocols, refer to the table below. 57: New protocol. A central monitor with ORG-9100A/9110A/9700A multiple patient receiver whose software version 02-01 or later can receive this protocol. 42: Old protocol. A central monitor with ORG-9100A/9110A/9700A multiple patient receiver can receive this protocol. NOTE When 57 is set, the receiving monitor must be able to receive protocol 57. Otherwise, signals from the transmitter cannot be received. Differences Between Protocols Cursor Setting Operator’s Manual ZM-540PA/541PA No Yes No Yes No Yes Some messages (refer to the “Indication and All messages Message List” section) ▼ Transmit SpO2 messages Protocol 42 Protocol 57 No (ECG measurement must be turned off on Yes the receiving monitor) 1. Press the key to move the cursor to “PROTOCOL”. 2. Press the Function key. 3. Press the ▼ Function Setting ECG MEASUREMENT to OFF on the transmitter automatically turns off the ECG measurement setting on the receiving monitor Pause monitoring on the receiving monitor from the transmitter Transmit “PATIENT CONFIRMED” message Display battery level of the transmitter on the receiving monitor key to select “57” or “42”. 35 NOTE FUNCTION KEY (on the second page of the SYSTEM SETUP screen) can be set to “SUSPEND ALARM & PAUSE” or “CONFIRM” only when PROTOCOL is “57”. If PROTOCOL is changed to “42”, FUNCTION KEY is automatically changed to “OFF”. 4. Press the Function key to register the selected setting. The cursor returns to “PROTOCOL”. ▼ BRIGHTNESS Select the screen brightness. 1. Press the 2. Press the Function key. 3. Press the 4. Press the Function key to register the selected setting. The cursor returns to “BRIGHTNESS”. ▼ key to move the cursor to “BRIGHTNESS”. key to select “DARK” or “BRIGHT”. ▼ PRESSURE UNIT Select the unit for NIBP. key to move the cursor to “PRESSURE UNIT”. Press the 2. Press the Function key. 3. Press the 4. Press the Function key to register the selected setting. The cursor returns to “PRESSURE UNIT”. 36 ▼ 1. key to select “mmHg” or “kPa”. Operator’s Manual ZM-540PA/541PA FUNCTION KEY Select the function of the Function key. SUSPEND ALARM & PAUSE: Suspends alarm on the receiving monitor for 2 minutes. Pauses monitoring on the transmitter and receiving monitor. SUSPEND ALARM: Suspends alarm on the receiving monitor for 2 minutes. CONFIRM: Displays the “PATIENT CONFIRMED” message on the transmitter screen and transmits the message to the receiving monitor. OFF: No function. NOTE “SUSPEND ALARM & PAUSE” and “CONFIRM” can only be set when PROTOCOL is set to 57. On the SYSTEM SETUP screen, press the key to move the cursor to “FUNCTION KEY”. “FUNCTION KEY” is on the second page of the SYSTEM SETUP screen. 2. Press the Function key. 3. Press the 4. Press the Function key to register the selected setting. The cursor returns “FUNCTION KEY”. SYSTEM SETUP screen - page 2 ▼ ▼ 1. key to select the function. ▼ AUTO RESUME AFTER PAUSE Select the interval to resume monitoring after PAUSE. Press the key to move the cursor to “AUTO RESUME AFTER PAUSE”. “AUTO RESUME AFTER PAUSE” is on the second page of the SYSTEM SETUP screen. 2. Press the Function key. 3. Press the 4. Press the Function key to register the selected setting. The cursor returns to “AUTO RESUME AFTER PAUSE”. ▼ 1. key to select interval. Operator’s Manual ZM-540PA/541PA 37 Initializing Settings Do the following procedure to initialize all settings, except for channel, to the factory default settings. 1. Turn off the transmitter by removing a battery. 2. While pressing the Function key, turn on the transmitter (put back the battery). The MENU screen appears. 3. Press the key to move the cursor to “SYSTEM INITIALIZE”. 4. Press the Function key to enter SYSTEM INITIALIZE screen. 5. Press the Function key to initialize settings. The confirmation message appears. ▼ Cursor To return to the MENU screen without initializing, press the Screen key. Press the Function key to initialize settings. ▼ 6. To cancel initializing, press the key. The screen returns to the MENU screen. 38 Operator’s Manual ZM-540PA/541PA Attaching Electrodes, SpO2 Probe and NIBP Cuff to the Patient The transmitter can be attached to an arm of the patient or placed on the bedside. The required length of the electrode leads and SpO2 probe cable depends on how the transmitter is to be attached to the patient. NOTE Monitoring SpO2 during NIBP Measurement When the SpO2 probe is attached to the same limb as the NIBP cuff, the blood flow decreases during NIBP measurement and pulse wave cannot be detected and SpO2 cannot be monitored properly. When “INHIBIT SpO2 DURING NIBP” on the PARAMETER SETUP screen is set to ON (factory default setting), SpO2 monitoring is paused during NIBP measurement to avoid SpO2 alarm occurrence. However, when monitoring SpO2 on the same limb as NIBP, be careful when reading SpO2 values. When monitoring SpO2 is important, attach the probe to the limb to which the NIBP cuff or catheter is not attached. Attachment Examples When transmitter is attached on an arm When transmitter is placed on a bedside NOTE When placing the transmitter on a bedside, place it on a stable and flat place. If the transmitter falls off, it may be damaged. Operator’s Manual ZM-540PA/541PA 39 Attaching Electrodes Selecting Electrode Lead CAUTION Only use Nihon Kohden specified electrodes and electrode leads. When other electrodes or electrode leads are used, the “CHECK ELECTRODES” message may appear and monitoring may stop. The following electrode leads can be used on the transmitter (option). BR-903PA, 3 electrodes, clip type BR-906PA, 6 electrodes, clip type Connecting the Electrode Lead to the Transmitter Connect the electrode lead to the ECG/RESP socket on the transmitter. When transmitter is attached on an arm 40 CAUTION CAUTION Do not shake or swing the transmitter while holding the leads or cables connected to the transmitter. The transmitter may come off and injure someone or damage surrounding instruments. Hold the connector of the electrode lead when connecting/disconnecting the electrode lead. If you disconnect the electrode lead by pulling the lead, it damages the electrode lead. Operator’s Manual ZM-540PA/541PA Electrode Position Follow the physician’s instructions for electrode placement when available. For ECG monitoring, electrodes are attached only on the chest to allow patient movement and obtain continuous stable ECG. The following leads are examples. When also monitoring respiration, refer to the “Electrode Position for Respiration Monitoring” section. NOTE The optimum electrode positions for ECG measurement are not always optimum for respiration measurement. Select positions that are suitable for both ECG and respiration measurement or positions which give priority to either ECG or respiration measurement. Electrode Positions for ECG Monitoring Six Electrodes The 6-electrode method with lead II and lead V5 is effective for monitoring myocardial ischemia. You can improve monitoring accuracy considerably by adding lead V4 to this combination. Va and Vb can be at any position of the standard 12 leads V1 to V6, but V4 and V5 are most appropriate for myocardial ischemic monitoring. LA/L RA/R Va/Ca Vb/Cb RL/RF LL/F Electrode Position Symbol AHA IEC Lead Color AHA IEC Left infraclavicular fossa LA Black Yellow Right infraclavicular fossa RA White Red Below lowest rib on the left anterior axillary line LL Red Green Right anterior axillary line at the same level as LL/F RL RF Green Black Va Ca Brown-blue Whitebrown Vb Cb Brownorange White-black Fifth intercostal space on the left midclavicular line. (V4 position of standard 12 leads) Left anterior axillary line at the same level as Va. (V5 position of standard 12 leads) Operator’s Manual ZM-540PA/541PA 41 Lead Position Standard limb leads Lead I RA Lead II LA Lead III RA LL RA LA LL LL N (RL) N (RL) Monopolar limb leads aVR lead RA N (RL) aVL lead aVF lead RA LA RA LA N (RL) LA LL LL LL LA N (RL) N (RL) Monopolar chest leads V1 to V6 leads to RA LA LL N (RL) Three Electrodes • Lead MII, which is similar to standard lead II, used when ECG measurement has priority Symbol AHA IEC Electrode Position Left infraclavicular fossa Right infraclavicular fossa Below lowest rib on the left anterior axillary line 42 Lead Color AHA IEC LA Black Yellow RA White Red LL Red Green Operator’s Manual ZM-540PA/541PA • Lead MI, which is similar to standard lead I Change F/LL and L/LA of lead MII. • Lead MIII, which is similar to standard lead III Change R/RA and L/LA of lead MII. If the electrode positions above are not available due to chest surgery, attach the electrodes to the root of the limbs or below the clavicles for stable ECG monitoring. Electrode Positions for Respiration Monitoring Place the R/RA and F/LL electrodes so that the lungs are between the electrodes. Position 1 In this position, respiration measurement is available; however, there is a difference in amplitude between different patients. R or RA Right infraclavicular fossa F or LL Fifth intercostal space on the left midclavicular line, V4 Position 2 In this position, the waveform amplitude is usually large and the ECG lead is similar to Lead MII. This position can be generally recommended. R or RA Right infraclavicular fossa Operator’s Manual ZM-540PA/541PA F or LL Fifth intercostal space on the left midaxillary line, V6 43 Position 3 In this position, the respiration waveform is optimum, but the ECG lead is unusual. R or RA Right midaxillary at the horizontal level of V6 F or LL Fifth intercostal space on the left midaxillary line, V6 Position 4 In this position, the respiration measurement is influenced by the impedance variation of the abdomen, so the cardiac pulse wave included in the respiration wave is reduced. Note that the waveform is inverted in phase compared with the chest movement (the waveform goes down during inspiration). It is difficult to measure the ECG at the same time. R or RA Lowest rib on the right anterior axillary line F or LL Lowest rib on the left anterior axillary line Attaching Electrodes to the Patient and Connecting the Electrode Leads to Disposable Electrodes Prepare the Patient Skin Shave off excessive body hair. To reduce skin impedance, clean the electrode site with cream or with a cotton pad moistened with alcohol. Thoroughly dry the skin with a clean cotton pad. NOTE • For a patient with frequent body movement, rub the sites with Skinpure skin preparation gel. However, do not use Skinpure skin preparation gel on sensitive skin. • Do not place electrodes on a wound or on an inflamed, wrinkled or uneven skin surface. Attaching Electrodes to the Patient CAUTION Do not reuse disposable parts and accessories. NOTE • To maintain good contact between the electrode and skin, check that the paste of the 44 Operator’s Manual ZM-540PA/541PA disposable electrode is not dry. • When contact between the disposable electrode and skin becomes poor, replace electrodes with new ones immediately. Otherwise, contact impedance between the skin and the electrode increases and accurate ECG cannot be obtained. Refer to the electrode operator’s manual for details. 1. Carefully remove the backing paper from the electrode. Avoid touching the adhesive surface. 2. Place the electrode on the previously cleaned skin. Pay attention to the electrode lead color and symbol. 3. Clip the electrode lead to the electrode. 4. Fasten the electrode lead wire with surgical tape with an extra length of wire between the tape and the electrode. This lessens the movement of electrode leads by body movement and helps stable monitoring. Checking ECG on the Transmitter Screen After attaching electrodes and connecting ECG leads, check that the electrodes are properly attached to the patient and the ECG waveform is acquired on the check electrodes screen. For details on the screen, refer to the “Screen Descriptions” section. Attaching the SpO2 Probe Selecting the SpO2 Probe Select an appropriate probe for the patient. CAUTION Only use Nihon Kohden specified electrodes, electrode leads, SpO2 probes, and NIBP cuffs. Otherwise, the maximum performance from the transmitter cannot be guaranteed. Operator’s Manual ZM-540PA/541PA CAUTION Do not use a damaged or disassembled probe. It causes incorrect measurement and may injure the patient. 45 Reusable Probes When using a TL-201T finger probe, choose the appropriate cable length for attachment. Probe Finger probe TL-201T Cable Length 0.6 m Patient Adult or child 20 kg or more Attachment Site Finger Adult or infant 3 kg or more Finger or toe Neonate 3 kg or less Instep and sole 1.6 m Multi-site probe TL-220T Attachment tape Finger probe TL-630T1, TL-630T3, TL-631T1, TL-631T3 TL-630T1, TL631T1: 0.6 m TL-630T1/630T3: Finger or toe Adult or child 50 kg or more Attachment tape TL-631T1/631T3: Adult or child 20 kg or more TL-630T3, TL631T3: 1.6 m Disposable Probes CAUTION The disposable probe is not sterilized. Use the disposable probe only for a single patient. Never reuse the disposable probe for another patient because it causes cross infection. 46 Operator’s Manual ZM-540PA/541PA Probe Patient Adult 30 kg or more Attachment Site Finger or toe TL-252T Child 3 to 40 kg Finger or toe TL-253T Neonate 3 kg or less Instep and sole TL-051S, TL-052S Adult 50 kg or more Finger Neonate 3 kg or less Instep and sole Adult or child 15 to 50 kg Finger Infant 3 to 15 kg Toe TL-251T 40 mm Cable length TL-051S: 0.8 m TL-052S: 1.6 m TL-061S, TL-062S 35 mm Cable length TL-061S: 0.8 m TL-062S: 1.6 m Operator’s Manual ZM-540PA/541PA 47 Probe TL-271T, TL-271T3 Cable length TL-271T: 0.8 m TL-271T3: 1.6 m TL-272T, TL-272T3 Cable length TL-272T: 0.8 m TL-272T3: 1.6 m TL-273T, TL-273T3 Patient Adult 30 kg or more Attachment Site Finger or toe Child 10 to 50 kg Neonate 3 kg or less Instep and sole Adult 40 kg or more Finger or toe Cable length TL-273T: 0.8 m TL-273T3: 1.6 m TL-274T, TL-274T3 Infant 3 to 20 kg Cable length TL-274T: 0.8 m TL-274T3: 1.6 m 48 Operator’s Manual ZM-540PA/541PA Connecting the SpO2 Probe to the Transmitter Connect the probe to the SpO2 socket on the transmitter. When transmitter is attached on an arm CAUTION Do not shake or swing the transmitter while holding the leads or cables connected to the transmitter. The transmitter may come off and injure someone or damage surrounding instruments. CAUTION Do not use a damaged or disassembled probe. It causes incorrect measurement and may injure the patient. Attaching the Probe to the Patient Attach the probe to the patient by referring to the probe’s manual. Make sure that the light emitter and photo detector of the probe face each other at the attachment site. Operator’s Manual ZM-540PA/541PA 49 WARNING WARNING • When using the TL-201T finger probe, do not fasten the probe and cable to the finger by wrapping with tape. This may cause burn, congestion or pressure necrosis from poor blood circulation. • When using probes other than the TL-201T finger probe, to avoid poor circulation, do not wrap the tape too tight. Check the blood circulation condition by observing the skin color and congestion at the skin peripheral to the probe attachment site. Even for short-term monitoring, there may be burn or pressure necrosis from poor blood circulation, especially on neonates or low birth weight infants whose skin is delicate. Accurate measurement cannot be performed on a site with poor peripheral circulation. Check the circulation condition by observing the skin color at the measurement site and pulse waveform. Change the measurement site every 8 hours for disposable probes and every 4 hours for reusable probes (every 8 hours for TL-630T/TL631T series probe). The skin temperature may increase at the attached site by 2 or 3°C (4 or 5°F) and cause a burn or pressure necrosis. When using the probe on the following patients, take extreme care and change the measurement site more frequently according to symptoms and degree. • Patient with a fever • Patient with peripheral circulation insufficiency • Neonate or low birth weight infant with delicate skin WARNING When monitoring SpO2 of a patient who is receiving photodynamic therapy, the light from the finger probe sensor may cause a burn. Photodynamic therapy uses a photosensitizing agent that has a side effect of photosensitivity. 50 Operator’s Manual ZM-540PA/541PA CAUTION If the attachment site is dirty with blood or bodily fluids, clean the attachment site before attaching the probe. If there is nail polish on the attachment site, remove the polish. Otherwise, the amount of transmitted light decreases, and measured value may be incorrect or measurement cannot be performed. CAUTION If the skin gets irritated or redness appears on the skin from the probe, change the attachment site or stop using the probe. Take extreme care for the patients with delicate skin. CAUTION When the probe is attached on an appropriate site with sufficient circulation and the error message confirming the probe attachment repeatedly appears, the probe may be deteriorated. Replace it with a new one. CAUTION Do not use a probe which is deteriorated by aging. Accurate measurement cannot be performed. CAUTION Refer to the probe instruction manual for details. Operator’s Manual ZM-540PA/541PA 51 Attaching the NIBP Cuff Selecting the NIBP Cuff Select the NIBP cuff appropriate for the patient. NOTE NIBP cannot be measured on neonates using this transmitter. Reusable Cuffs When attaching the transmitter to the patient arm, a special NIBP cuff is required. An optional YN-990P extension hose (1.5 m) is available to extend the length between the NIBP socket on the transmitter and NIBP cuff (e.g. when not attaching the transmitter to the patient arm and placing the transmitter on a bedside). Reusable Cuff For adult Model Width (cm) Air Hose Length (cm) Standard YP-503P 13 15 Large YP-504P 15 15 Air hose Width When not attaching the transmitter to the patient arm, the following cuffs can be used. To use these cuffs, an optional YN-990P extension hose (1.5 m) is required. Reusable Cuff For infant For child For adult Model Width (cm) YP-960T Small YP-961T Standard YP-962T 10 Standard YP-963T 13 Large YP-964T 15 Air Hose Length (cm) 15 Width Air hose 52 Operator’s Manual ZM-540PA/541PA Disposable Cuffs CAUTION Disposable cuffs are not sterilized. If necessary, sterilize the cuff using glutaraldehyde solution. When not attaching the transmitter to the patient arm, the following disposable cuffs can be used. To use these cuffs, an optional YN-990P extension hose (1.5 m) is required. Model Width (cm) Air Hose Length (cm) For infant Reusable Cuff YP-810P 17 For child YP-811P 17 Small YP-812P 10 17 Standard YP-813P 14 20 Medium large YP-814P 15 20 Large YP-815P 17 20 For adult Width Air hose Extension Hose CAUTION When using an extension hose, check that the extension hose is not bent or squeezed. Otherwise, the cuff might not inflate or deflate. If the cuff cannot deflate, it may cause congestion on the patient at the cuff attachment site. YN-990P extension hose, 150 cm Operator’s Manual ZM-540PA/541PA 53 Reference for selecting a cuff The AHA (American Heart Association) recommends that the cuff width be 40% of the circumference of the upper arm. Refer to the following graph and select the cuff which suits the patient’s arm. NOTE • If a range of arm circumference appropriate for the cuff is prescribed, use a cuff within that range. • To obtain accurate measured values, select a wide cuff which can be attached to the upper arm. Measuring with a very narrow cuff may result in measured values higher than the actual values. • The YP-503P NIBP cuff is for standard size adult. Do not use this cuff when it does not fit the patient. Cuff Width and Arm Circumference Reusable Cuffs Adults large YP-504P (cuff for transmitter) YP-964T Cuff width (cm) 15 Adults standard YP-503P (cuff for transmitter) YP-963T 10 Children standard YP-962T Children small YP-961T Infants YP-960T 10 20 30 40 50 60 Arm circumference (cm) Disposable Cuffs Cuff width (cm) 20 Adults large YP-815P 15 Adults medium large YP-814P Adults standard YP-813P 10 Adults small YP-812P Children standard YP-811P Infants YP-810P 54 10 20 30 40 50 60 Arm circumference (cm) Operator’s Manual ZM-540PA/541PA Connecting the NIBP Cuff to the Transmitter When Using YP-503P/YP-504P NIBP Cuff To attach the YP-503P/YP-504P NIBP cuff to the transmitter, the lock plate is required. YP-503P/YP-504P NIBP cuff Air hose Front cover D ring Belt Belt for the strap Lock plate pocket Front cover open Lock plate Top tab For attaching the NIBP cuff to the transmitter NOTE Do not roll up or put weight on the cuff when the lock plate is attached to it. The lock plate may break if the cuff is rolled up or weight is put on it when the lock plate attached. Bottom tab Operator’s Manual ZM-540PA/541PA 55 1. Remove the lock plate from the transmitter. 2. Insert the lock plate into the lock plate pocket on the NIBP cuff. 3. Attach the transmitter to the lock plate by inserting the tabs on the lock plate into the slots on the transmitter. 4. Cover the transmitter with the front cover of the NIBP cuff. 5. Connect the air hose to the NIBP socket on the transmitter. Turn the cuff connector joint until it clicks. 56 Operator’s Manual ZM-540PA/541PA When Using Disposable Cuffs or YP-960T series Reusable Cuffs To use these NIBP cuffs, an optional YN-990P extension hose (1.5 m) is required. NOTE Connect the joints properly. If there is an air leak, NIBP cannot be measured properly. 1. Connect the NIBP cuff to the extension hose. 2. Connect the other end of the extension hose to the NIBP socket on the transmitter. Turn the joint clockwise until it clicks. To disconnect the cuff from the transmitter, turn the hose joint counterclockwise. Attaching the NIBP Cuff to the Patient WARNING Be careful when measuring NIBP on a patient with known bleeding disorders or coagulation. After NIBP measurement, there may be dot hemorrhage, or circulatory disorder by thrombus where the cuff is attached. CAUTION Do not wrap the cuff on an arm or thigh which is used for injection. NIBP measurement on an arm or thigh which is used for injection may cause reflux of blood and stop injection. CAUTION Do not wrap the cuff too tight. It may cause poor blood circulation and congestion. If the cuff is wrapped too loosely, the NIBP value may increase. Operator’s Manual ZM-540PA/541PA CAUTION Do not attach the cuff to the site where there is injury or inflammation. If the skin gets irritated or redness appears on the skin from the cuff, change the attachment site or stop using the cuff. Take extreme care on the patients with delicate skin. CAUTION Do not reuse disposable parts and accessories. 57 CAUTION NIBP and SpO2 can be measured on the same limb, but the SpO2 monitoring might not be accurate during NIBP measurement. Be careful when reading the SpO2 values.* * Monitoring SpO2 during NIBP Measurement When the SpO2 probe is attached to the same limb as the NIBP cuff, the blood flow decreases during NIBP measurement and pulse wave cannot be detected and SpO2 cannot be monitored properly. When “INHIBIT SpO2 DURING NIBP” on the PARAMETER SETUP screen is set to ON (factory default setting), SpO2 monitoring is paused during NIBP measurement to avoid SpO2 alarm occurrence. However, when monitoring SpO2 on the same limb as the NIBP, be careful when reading SpO2 values. NOTE • Measuring NIBP at a site other than the upper arm gives different values from those measured at the upper arm. When making diagnosis based on the NIBP values, measure NIBP on an upper arm. • To accurately detect the pulsatile flow of the artery, the cuff should be wrapped around a bare upper arm. • Do not use an abnormal cuff. The cuff deteriorates from use and cleaning. Before use, check the cuff and confirm that there is no flaw, crack or hole in it. Be careful not to damage the inflation bag. If the inflation bag has a hole or a flaw, it may burst during use. Dispose of an abnormal cuff and replace it with a new one. • Refer to the NIBP cuff manual for details. Cuff Position When placing the transmitter on a bed, make sure that the hose is not bent. Heart 58 Place the cuffed upper arm (brachium) at the same height as the patient’s heart. If the cuff is not at the same level as the heart, the weight of the blood affects the blood pressure reading. The pressure difference per unit height is 0.7 mmHg/cm. The blood pressure reading decreases when the arm is higher than the heart and increases when lower. The best measuring condition is when the patient is lying on his/her back with arms and legs relaxed. If the cuff position cannot be on the same level as the heart, the displayed blood pressure reading must be mathematically adjusted. Operator’s Manual ZM-540PA/541PA Attaching the Transmitter on an Arm (Using the YP-503P/504P NIBP Cuff) 1. Attach the NIBP cuff to the transmitter. Refer to the “Connecting the NIBP Cuff to the Transmitter” section. 2. Before putting the cuff over the arm, insert the end of the cuff into the belt and then through the D ring as shown at left. Belt D ring End of cuff 3. Put the patient arm through the cuff. Fold back the cuff at the D ring and fasten it using the velcro tape. Make sure that the cuff is not attached on a joint. NOTE The cuff must not wrap around the elbow. Operator’s Manual ZM-540PA/541PA 59 Attaching the Strap to the Transmitter NOTE • Use the strap to prevent the transmitter from falling. • Do not attach the clip to hard objects such as thick cloth or zipper. It will break the clip. Attach a strap provided with the transmitter to the NIBP cuff and patient clothes. To open the clip, firmly pull out the tab in direction of the arrow. To adjust the strap length, push down the tab on the adjuster and slide. 1. Adjust the length of the strap. 2. Clip one end of the strap to the belt for the strap on the NIBP cuff. 3. Clip the other end of the strap to the patient’s clothes as shown left. Belt for the strap on the NIBP cuff 60 Operator’s Manual ZM-540PA/541PA Placing the Transmitter on the Bed (Using the Disposable Cuffs or YP-960T series Reusable Cuffs) ARTERY ▼ 1. Put the cuff on the upper arm so that the ▼ mark of “ARTERY ▼” aligns with the artery of the patient. 2. Wrap the cuff so that “INDEX within the “ RANGE ”. RANGE Operator’s Manual ZM-540PA/541PA ” is not within the “ If “Index ”, change the cuff size. ” comes RANGE 61 Locking the Keys on the Transmitter ▼ Press the appears. and ▼ To prevent the patient from pressing the keys on the transmitter during monitoring, you can lock the keys. keys at the same time and hold for more than 3 seconds. The key lock screen When there is no key operation for one minute after locking the keys, the display turns off. 62 ▼ ▼ To unlock the keys: Press the and keys at the same time and hold for more than 3 seconds. Operator’s Manual ZM-540PA/541PA Monitoring CAUTION The measurement values and displayed waveforms on the transmitter and receiving monitor may be different due to timing delay of the display or difference in detection settings. Screen Descriptions When the transmitter is turned on, the startup screen appears, then the check electrodes screen appears to check the electrode attachment. The screen changes in the following order each time the Screen key is pressed. Check electrodes → numeric and waveform → waveform review → numeric review → display off → check electrodes . . . Screen key The screen automatically turns off when there is no key operation for 2 minutes on the check electrodes screen or 1 minute on other screens. When the display is off and the Screen key is pressed, the numeric and waveform screen appears. When the Screen key is pressed within 5 minutes after the display off, the screen before the display off appears. Operator’s Manual ZM-540PA/541PA 63 Check Electrodes Screen You can check whether the electrodes are properly attached to the patient and the ECG waveform is acquired. When 6 leads are used, I, II, Va and Vb lead waveforms are displayed. When 3 leads are used, only II lead waveform is displayed. Battery level Channel number Lead ECG waveform Filter: off Sweep speed: 12.5 mm/s Waveform sensitivity: 0.5 cm/1 mV Message Detached electrode position When electrodes are detached, the “CHECK ELECTRODES” message and detached electrode position appear on the screen. NOTE When ECG measurement is set to OFF on the PARAMETER SETUP screen, the check electrodes screen does not appear. 64 Operator’s Manual ZM-540PA/541PA Numeric and Waveform Screen ▼ ▼ Numeric values and waveforms of the monitoring parameters are displayed. You can change the ECG lead with the and keys. Respiration rate SpO2 Pulse bar graph QRS sync mark Heart rate NIBP measurement mode NIBP values ECG lead ECG waveform Filter: on Sweep speed: 12.5 mm/s Waveform sensitivity: auto Waveform sensitivity When ECG and respiration measurement is turned off Pulse rate SpO2 Pulse bar graph Pulse sync mark NIBP measurement mode NIBP values Pulse wave sensitivity Pulse waveform Sweep speed: 12.5 mm/s Waveform sensitivity: auto Operator’s Manual ZM-540PA/541PA 65 NOTE The pulse wave amplitude varies according to the ratio of the pulsation component to the entire transmitted IR signal. When the pulsation component ratio is 1%, the pulse wave amplitude is about 5 mm at ×1 sensitivity on the screen. Waveform Review Screen ECG full disclosure for up to 10 minutes can be saved and reviewed. When ECG measurement is turned off and SpO2 is monitored, the pulse waveform is saved. When ECG lead is changed on the numeric and waveform screen, the ECG full disclosure of the changed lead is saved. The saved data is deleted when the transmitter is turned off. Time range of the displayed waveform (time before the waveform review screen is displayed) Displayed page Older data ECG lead Newer data ▼ To scroll the waveform, press the 66 or ▼ Waveform sensitivity Compressed ECG waveform 7.5 s x 4 traces per page, total of 20 pages key. The waveform is scrolled by 30 seconds. Operator’s Manual ZM-540PA/541PA Numeric Review Screen Numeric data of heart rate (or pulse rate when ECG is turned off), SpO2 and respiration rate for up to 10 minutes are saved at 1 minute intervals. NOTE NIBP measured values are not saved. The saved data is deleted when the transmitter is turned off. Older data Time before the numeric review screen is displayed Newer data Display Off The display can be turned off any time. To turn off the display, press the Screen key several times until the following screen appears then wait 5 seconds. Operator’s Manual ZM-540PA/541PA 67 Basic Monitoring Operation Using the Function Key Function key One of the following functions can be assigned to the Function key on the SYSTEM SETUP screen. Refer to the “Changing SYSTEM SETUP Settings” section. SUSPEND ALARM: Suspends alarms on the receiving monitor before they occur for 2 minutes. PAUSE: Pauses monitoring on the transmitter and receiving monitor. CONFIRM: Transmits the signal that the patient is confirmed and displays the “PATIENT CONFIRMED” message on the transmitter. NOTE To use the Function key for PAUSE or CONFIRM, you must set “PROTOCOL” on the SYSTEM SETUP screen of the transmitter to 57 and the receiving monitor must be able to receive protocol 57. 68 Operator’s Manual ZM-540PA/541PA Suspending Alarms on the Receiving Monitor When the FUNCTION KEY is set to “SUSPEND ALARM” or “SUSPEND ALARM & PAUSE” on the SYSTEM SETUP screen, alarms can be suspended for 2 minutes on the receiving monitor before they occur. 2. ▼ To suspend alarms: 1. Press the Function key. The “Suspend alarms” confirmation screen appears. Press the key to suspend alarms. To cancel suspending alarms and return to the previous screen, press the Screen key. When the alarms are suspended, the “ALARMS SUSPENDED” message and alarm suspended icon with the remaining minutes in alarm suspension appear on the transmitter screen. Message Icon To cancel suspending alarms during 2 minute alarm suspension: 1. Press the Function key while the “ALARMS SUSPENDED” message is displayed. The confirmation screen appears. Operator’s Manual ZM-540PA/541PA 69 ▼ 2. Press the key to cancel alarm suspension. Press the Screen key to not cancel alarm suspension. Pause Monitoring When the FUNCTION KEY is set to “SUSPEND ALARM & PAUSE” on the SYSTEM SETUP screen, you can pause monitoring on the receiving monitor from the transmitter when the patient cannot be monitored, such as during X-ray examination. NOTE To use the Function key for PAUSE, you must set “PROTOCOL” on the SYSTEM SETUP screen of the transmitter to 57 and the receiving monitor must be able to receive protocol 57. To pause monitoring: 1. Press the Function key. The “Suspend alarms” confirmation screen appears. 2. 70 Press the Function key for 3 seconds to display the “Pause monitoring” confirmation screen. Operator’s Manual ZM-540PA/541PA ▼ 3. Press the key to pause monitoring. To cancel pause monitoring, press the Screen key. 4. Wait about 5 seconds until the “Turn power off” screen appears. 5. Turn off the transmitter. If the transmitter is not turned off and monitoring continues for the interval set for “AUTO RESUME AFTER PAUSE” on the SYSTEM SETUP screen, pause monitoring is cancelled and monitoring continues. Resume Monitoring after Pause To resume monitoring after pause, check that the electrodes, electrode leads and probe are attached to the patient then turn on the transmitter. Operator’s Manual ZM-540PA/541PA 71 Confirming Patient When the FUNCTION KEY is set to “CONFIRM” on the SYSTEM SETUP screen, you can transmit signal to the receiving monitor to indicate that the patient is confirmed by a medical staff by pressing the Function key. Message Turning the Display Off The display can be turned off any time. To turn off the display, press the Screen key several times until the following screen appears then wait 5 seconds. 5 second time bar until the display turns off Turning the Display On after It was Turned Off Press the Screen key. The numeric and waveform screen appears. If the screen is turned off on the “Keys locked” screen, the “Keys locked” screen appears. When the Screen key is pressed within 5 minutes after the display was turned off, the previous screen appears. 72 Operator’s Manual ZM-540PA/541PA ECG and Respiration Monitoring When the electrodes are attached and the ECG leads are connected, the heart rate, ECG waveform, respiration rate and respiration waveform appear on the receiving monitor. Refer to the operator’s manual of the receiving monitor for details. When 6 leads are used on this transmitter, up to 8 lead (I, II, III, aVR, aVL, aVF, Va and Vb) of ECG waveforms can be displayed on the receiving monitor. The heart rate is also measured. When 3 leads are used, one channel ECG waveform of lead II can be displayed on the receiving monitor. Refer to the operator’s manual of the monitor for details. QRS sync mark Respiration rate ▼ ECG lead You can change the ECG lead with the and keys. ▼ ECG waveform Heart rate Filter: on Sweep speed: 12.5 mm/s Waveform sensitivity: auto WARNING Interaction Between Minute Ventilation Rate-Adaptive Pacemakers and Cardiac Monitoring and Diagnostic Equipment* The bioelectric impedance measurement sensor of a minute ventilation rate-adaptive implantable pacemaker may be affected by transmitter which is connected to the same patient. If this occurs, the pacemaker may pace at its maximum rate and the transmitter may give incorrect data to the monitor. If this occurs, disconnect the electrode leads from the patient or change the setting on the pacemaker by referring to the pacemaker’s manual. For more details, contact your pacemaker representative or Nihon Kohden representative. * Minute ventilation is sensed in rate-adaptive pacemakers by a technology known as bioelectric Operator’s Manual ZM-540PA/541PA 73 impedance measurement (BIM). Many medical devices in addition to pacemakers use this technology. When one of these devices is used on a patient with an active, minute ventilation rateadaptive pacemaker, the pacemaker may erroneously interpret the mixture of BIM signals created in the patient, resulting in an elevated pacing rate. For more information, see the FDA web site. http://www.fda.gov/cdrh/safety.html WARNING When the transmitter is used with an electrosurgical unit (ESU), firmly attach the entire area of the ESU return plate. Otherwise, the current from the ESU flows into the electrodes of the transmitter, causing electrical burn where the electrodes are attached. For details, refer to the ESU manual. CAUTION Turn off the power of mobile phones, small wireless devices and other devices which produce strong electromagnetic interference around a patient (except for devices allowed by the hospital administrator). Radio waves from devices such as mobile phones or small wireless devices may be mistaken as pulse waves and the displayed data may be incorrect. NOTE • Noise generated from an electrosurgical unit may interfere on an ECG waveform, but will not damage it. • If an electric blanket is used and incorrect heart rate is displayed on the monitor, turn off the pacing spike detection on the monitor. • Turn the pacing spike detection to ON on the receiving monitor when monitoring a pacemaker patient. Pacing pulse is detected by the transmitter and transmitted to the monitor. If the pacing spike detection is turned OFF, QRS and pacemaker spike might not be distinguished and pacemaker failure might not be recognized. • ECG cannot be monitored on a neonate using this transmitter. 74 Operator’s Manual ZM-540PA/541PA Turning ECG Measurement On/Off ECG measurement can be turned on or off on the PARAMETER SETUP screen. When electrodes are attached to the patient and ECG leads are connected, ECG monitoring starts even when ECG is turned off. When PROTOCOL on the SYSTEM SETUP screen is set to 57: ECG measurement on the receiving monitor is automatically set to off. NOTE ECG measurement on the transmitter cannot be turned on or off from the receiving monitor. When PROTOCOL on the SYSTEM SETUP screen is set to 42: If ECG measurement is turned off on the transmitter, ECG measurement on the receiving monitor must also be turned off. Turning Respiration Measurement On/Off Respiration measurement can be turned on or off on the PARAMETER SETUP screen. If respiration measurement is turned off, respiration measurement on the receiving monitor is also turned off. Electrode Detachment In the following conditions, the “CHECK ELECTRODES” message is displayed on the transmitter and receiving monitor. • Electrode is detached from skin. • Electrode lead is disconnected from the electrode. • Polarization voltage between the electrode and skin is excessively high. In these cases, check the cause and if necessary, replace electrodes with new ones. CAUTION When the “ELECTRODE OFF” or “CHECK ELECTRODES” message is displayed on the receiving monitor, ECG is not monitored properly and the ECG alarm does not function. Check the electrode, electrode leads, and if necessary, replace with new ones. Operator’s Manual ZM-540PA/541PA 75 SpO2 Monitoring When monitoring starts, SpO2 and pulse waveform are sent to the monitor and SpO2 and pulse level bar graph are displayed on the transmitter screen. When ECG is not measured, pulse waveform and pulse rate are also displayed. SpO2 Pulse bar graph Filter: on Sweep speed: 12.5 mm/s Waveform sensitivity: auto 76 Operator’s Manual ZM-540PA/541PA WARNING WARNING SpO2 measurement may be incorrect in the following cases. • When the patient’s carboxyhemoglobin or methemoglobin increases abnormally. • When dye is injected in the blood. • When using an electrosurgical unit. • During CPR. • When measuring at a site with venous pulse. • When there is body movement. • When the pulse wave is small (insufficient peripheral circulation). Check the circulation condition by observing the skin color at the measurement site and pulse waveform. Change the measurement site every 8 hours for disposable probes and every 4 hours for reusable probes (every 8 hours for TL-630T/TL631T series probe). The skin temperature may increase at the attached site by 2 or 3°C (4 or 5°F) and cause a burn or pressure necrosis. When using the probe on the following patients, take extreme care and change the measurement site more frequently according to symptoms and degree. • Patient with a fever • Patient with peripheral circulation insufficiency • Neonate or low birth weight infant with delicate skin WARNING WARNING When monitoring SpO2 of a patient who is receiving photodynamic therapy, the light from the finger probe sensor may cause a burn. Photodynamic therapy uses a photosensitizing agent that has a side effect of photosensitivity. When not monitoring SpO2, disconnect the SpO2 cable from the transmitter. Otherwise, noise from the probe sensor may interfere and incorrect data is displayed on the screen. Operator’s Manual ZM-540PA/541PA 77 CAUTION CAUTION Turn off the power of mobile phones, small wireless devices and other devices which produce strong electromagnetic interference around a patient (except for devices allowed by the hospital administrator). Radio waves from devices such as mobile phones or small wireless devices may be mistaken as pulse waves and the displayed data may be incorrect. Do not pull or bend the probe cable, and do not put caster feet on the probe cable. Do not immerse the disposable probe cable in chemical solutions or water. Failure to follow these instructions may cause cable discontinuity, short circuit, skin burn on the patient and incorrect measurement data. Replace any broken probe with a new one. CAUTION CAUTION Normal external light does not affect monitoring but strong light such as a surgical light or sunlight may affect monitoring. If affected, cover the measuring site with a blanket. When the probe is attached on an appropriate site with sufficient circulation and the error message confirming the probe attachment repeatedly appears, the probe may be deteriorated. Replace it with a new one. CAUTION CAUTION When a message indicates a faulty probe, stop monitoring and replace the probe with a new one. While a patient is on medication which causes vasodilation, the pulse waveform may change and in rare cases the SpO2 value might not be displayed. NOTE In order to maintain sufficient blood circulation, keep the measurement site warm by covering it with a blanket or something similar. Warming the site is effective, especially for a patient with a small pulse amplitude. 78 Operator’s Manual ZM-540PA/541PA Monitoring SpO2 during NIBP Measurement CAUTION NIBP and SpO2 can be measured on the same limb, but the SpO2 monitoring might not be accurate during NIBP measurement. Be careful when reading the SpO2 values. When the SpO2 probe is attached to the same limb as the NIBP cuff, the blood flow decreases during NIBP measurement and pulse wave cannot be detected and SpO2 cannot be monitored properly. When “INHIBIT SpO2 DURING NIBP” on the PARAMETER SETUP screen is set to ON (factory default setting), SpO2 monitoring is paused during NIBP measurement to avoid SpO2 alarm occurrence. However, when monitoring SpO2 on the same limb as NIBP, be careful when reading SpO2 values. When monitoring SpO2 is important, attach the probe to the limb to which the NIBP cuff or catheter is not attached. When SpO2 monitoring is paused during NIBP measurement, the SpO2 value just before the start of NIBP measurement and an mark are displayed on the transmitter for 30 seconds. When NIBP measurement is not completed after 30 seconds, “– – –” is displayed for the SpO2 value. The same data also appears on the monitor screen. NOTE • When continuous SpO2 monitoring is necessary, attach the probe to the limb to which the NIBP cuff is not attached and set “INHIBIT SpO2 DURING NIBP” on the PARAMETER SETUP screen to OFF. • When the probe is attached to the same limb as the NIBP cuff, set the sync source to a parameter other than SpO2 on the receiving monitor. • When monitoring SpO2 during STAT NIBP measurement, attach the probe to the limb to which the NIBP cuff is not attached. Operator’s Manual ZM-540PA/541PA 79 NIBP Monitoring Selecting the Initial Cuff Inflation Pressure The initial cuff inflation pressure can be changed on the PARAMETER SETUP screen. The default setting is 180 mmHg. To change the setting, refer to the “Changing PARAMETER SETUP Settings” section. Selecting the Measurement Mode and Interval Measurement Modes There are three measurement modes: manual, auto and STAT. The selected mode or interval is displayed on the screen. ▼ ▼ The measurement mode and interval can be changed by pressing the NIBP Interval key. When the key is pressed, the NIBP mode setting screen appears. The measurement modes selected at “SELECTABLE INTERVALS” on the PARAMETER SETUP screen are displayed (key color: white). Select the measurement mode with the and keys or NIBP Interval key and press the Function key. Selected mode (key color: blue) Modes selected at “SELECTABLE INTERVALS” on the PARAMETERS SETUP screen (key color: white) To select the modes to be displayed on the NIBP mode setting screen, refer to the “Changing PARAMETER SETUP Settings” section. Manual Measurement In Manual mode, a single NIBP measurement is performed when the NIBP Start/Stop key is pressed. STAT (Continuous) Measurement In STAT mode, measurement is continuously repeated for 15 minutes after the NIBP Start/Stop key is pressed. When the STAT measurement for 15 minutes is completed, the measurement mode automatically changes to the Manual mode or Auto mode of selected interval depending on the “NIBP MODE 80 Operator’s Manual ZM-540PA/541PA AFTER STAT” setting on the PARAMETER SETUP screen. The default setting is Manual mode. Refer to the “Changing Parameter Setup Settings” section. The STAT measurement completes within 15 minutes. When more than 12 minutes elapse from the start of measurement, there will be no more measurement performed and the measurement mode changes to the mode selected for “NIBP MODE AFTER STAT” on the PARAMETER SETUP screen. Auto Measurement In Auto mode, measurement is performed automatically at the preset time intervals. In Auto mode, a single measurement can be performed by pressing the NIBP Start/Stop key between auto measurements. Measuring NIBP WARNING Be careful when measuring NIBP on a patient with known bleeding disorders or coagulation. After NIBP measurement, there may be dot hemorrhage, or circulatory disorder by thrombus where the cuff is attached. WARNING When performing NIBP measurements in STAT mode or 5 minute intervals, periodically remove the cuff from the patient for ventilation. The skin temperature may increase at the cuff attachment site by 2 or 3°C (4 or 5°F). When measuring a patient with a fever or peripheral circulation insufficiency, it may cause a burn. WARNING NIBP measurement may be incorrect in the following cases. • When using an electrosurgical unit • When there is body movement • When the pulse wave is small (insufficient peripheral circulation) • Too many arrhythmias • When there is vibration • When there is a rapid blood pressure change • During CPR CAUTION When performing NIBP measurement repeatedly, have a rest between measurements to recover adequate circulation. Operator’s Manual ZM-540PA/541PA 81 NOTE • When measuring patients who are conscious, help the patient to relax. Measurement may not be accurate if the patient’s arm is tense or if the patient talks. • The data for measurement on a leg tends to be higher than measurement on the arm. When making diagnosis based on the NIBP values, measure NIBP on an upper arm. • Do not apply pressure to the cuff or air hose. NIBP may not be measured correctly because of noise or NIBP measurement may stop due to the NIBP safety circuit. • When the transmitter is attached to the patient arm and the NIBP measurement is performed when moving, tell the patient to relax and keep quiet. Otherwise, measurement may be stopped or remeasurement is repeated due to body movement. • If there is an abnormal noise generated during measurement, stop using the transmitter and contact your Nihon Kohden representative. • Do not measure NIBP of a patient on whom an IABP is being used. Measurement may be incorrect due to the mixing of the patient’s own pulse and IABP pulse. • NIBP cannot be measured on a neonate using this transmitter. 1. Press the NIBP Interval key to display the NIBP mode setting screen. ▼ 2. Select the measurement mode by pressing the NIBP Interval key or and ▼ NIBP Interval key keys. Selected mode (key color: blue) 3. Press the Function key. 4. Press the NIBP Start/Stop key to perform measurement. 82 Operator’s Manual ZM-540PA/541PA NIBP Start/Stop key The cuff is inflated and the inflation pressure is displayed on the screen. Inflation pressure In manual mode: Measurement is performed once. In STAT mode: Measurement is performed repeatedly for 15 minutes. In auto mode: The first measurement is performed when the NIBP Start/Stop key is pressed. The second measurement is performed when the current time in the transmitter reaches the selected time interval. To stop measurement during measurement, press the NIBP Start/Stop key again. In STAT mode, after completing the STAT measurement, the measurement mode changes to the mode set for “NIBP MODE AFTER STAT” on the PARAMETER SETUP screen. In auto mode, to stop measurement in auto mode, change the mode to manual. To cancel one measurement, press the NIBP Start/Stop key during measurement. After the measurement is complete, the measured data is displayed on the screen and is transmitted to the monitor. Operator’s Manual ZM-540PA/541PA 83 Systolic value Diastolic value NIBP measurement mode MAP value When ECG and SpO2 are not monitored (ECG measurement is turned off and SpO2 probe is not connected to the transmitter), the pulse rate at the end of NIBP measurement is displayed. A buzzer can be set to sound at the start and end of NIBP measurement. Refer to the “Changing PARAMETER SETUP Settings” section. Auto Mode Measurement When auto mode measurement is selected, “STANDBY” message is displayed on the screen until the NIBP Start key is pressed for the first time. STANDBY message 84 Operator’s Manual ZM-540PA/541PA A time bar appears to indicate the interval between auto mode measurements. Time bar indicating the interval between measurements During auto mode measurement, the measurement mode can be changed. During the interval, press the NIBP Interval key to change the mode. When “MANUAL” is displayed for more than one second, the measurement in auto mode is stopped. Data Display After NIBP Measurement When the time set at “OLD NIBP DATA” on the PARAMETER SETUP screen elapses after the last measurement, the NIBP data is dimmed or hidden. Whether to dim or hide the old data can also be selected at “OLD NIBP DATA”. Refer to the “Changing PARAMETER SETUP Settings” section. Data Display on the Receiving Monitor When the “BATTERY” message is displayed on the receiving monitor, NIBP might not have been measured according to the NIBP interval setting. Therefore, the NIBP data displayed on the receiving monitor might not be updated. In this case, check the measurement time of the NIBP data displayed on the receiving monitor. Monitoring SpO2 during NIBP Measurement When the SpO2 probe is attached to the same limb as the NIBP cuff, the blood flow decreases during NIBP measurement and pulse wave cannot be detected and SpO2 cannot be monitored properly. When “INHIBIT SpO2 DURING NIBP” on the PARAMETER SETUP screen is set to ON (factory default setting), SpO2 monitoring is paused during NIBP measurement to avoid SpO2 alarm occurrence. However, when monitoring SpO2 on the same limb as the NIBP, be careful when reading SpO2 values. Operator’s Manual ZM-540PA/541PA 85 Indication and Message List Indication Indication Cause Countermeasure Fully charged batteries —— Batteries are 2/3 full. Batteries are low. NIBP cannot be measured. Replace batteries. Batteries are very weak. Alarms on the receiving monitor were suspended by pressing the Function key on the transmitter. Alarms resume when the suspend interval elapses. To cancel alarm suspension, press the Function key again. Messages When PROTOCOL on the SYSTEM SETUP screen is set to 57, all messages are transmitted. When PROTOCOL is set to 42, the messages marked with * are not transmitted. Message AIR LEAK Cause The cuff and extension hose are not properly connected. The cuff hose (or extension hose) is not properly connected to the NIBP socket. Countermeasure Connect them properly. The cuff or extension hose is damaged. Replace with a new one. ALARMS SUSPENDED Alarms on the receiving monitor is suspended by pressing the Function key on the transmitter. Alarms resume when the 2 minute suspend interval elapses. To cancel alarm suspension, press the Function key again. BATTERY WEAK Dead batteries Replace batteries. 86 Operator’s Manual ZM-540PA/541PA Message Cause Poor blood circulation for measuring the SpO2 value. The probe is attached too tightly and is CANNOT obstructing the blood circulation. DETECT The probe is not attached to the patient PULSE* properly. (displayed in blue) “LIGHT INTERFERENCE”, “CHECK PROBE SITE” or “DETECTING PULSE” message is displayed for more than 30 seconds. The patient’s pulse wave is too small CANNOT to measure NIBP. DETECT PULSE The cuff is not wrapped on the patient (displayed in pink) properly. Electrode lead is disconnected from the electrode. Electrode lead is disconnected from the transmitter. CHECK Electrode lead discontinuity. ELECTRODES Electrode is not firmly attached to the skin. Polarization voltage is abnormally high. The probe is not attached to the patient properly. The probe is not attached at the CHECK PROBE appropriate site. The probe is disconnected from the transmitter. CHECK PROBE SITE* Reattach the probe. Attach the probe to the patient properly. Refer to the cause and countermeasure for each message in this Messages table and remove the cause. Measure by palpation or auscultation. Wrap the cuff on the patient properly. Firmly connect the electrode lead to the electrode. Firmly connect the electrode lead to the transmitter. Replace the electrode lead with a new one. Replace the electrode with a new one. Attach the probe to the patient properly. Attach the probe to an appropriate site indicated in the probe manual. Connect the probe cable to the transmitter. The probe is past its expiration date. Replace the probe with a new one. The probe is not attached at the appropriate site. Attach the probe to an appropriate site indicated in the probe manual. The probe is deteriorated. The probe is past its expiration date. CUFF OCCLUSION Countermeasure Check the patient condition, probe attachment or change the attachment site. Transmitter malfunction. Operator’s Manual ZM-540PA/541PA Replace the probe with a new one. Immediately remove the cuff from the patient and contact your Nihon Kohden representative. 87 Message DETECTING PULSE HIGH CUFF PRESS INFLATION PRESS LOW Cause Searching for the correct pulse wave for SpO2 monitoring. The SpO2 value cannot be obtained because the waveform is unstable. The probe is not attached to the patient properly. Enormous pressure was applied by the pressure of the cuff. Insufficient cuff inflation pressure. The SpO2 measurement site is under LIGHT fluorescent light, surgical light, INTERFERENCE sunlight, or other strong light. Countermeasure Wait until the pulse wave is detected. Attach the probe to the patient properly. Remove the cause. Wait for the remeasurement to be performed with increased cuff inflation pressure. Cover the measurement site with a blanket or cloth. When the message is displayed frequently, check the patient condition The probe is not attached to the patient and, if necessary, change the properly. attachment site. SpO2 monitoring is paused for NIBP Wait for NIBP measurement to finish. measurement. The NIBP measuring time exceeded Remove the cause if the cause the specified time due to arrhythmia, is body movement, vibration or body movement, vibration or, cuff or squeezing of cuff or hose. air hose being squeezed. Contact your Nihon Kohden Module malfunction. representative. Considerable body movement. MEAS TIMEOUT NIBP MODULE ERROR NO NIBP CHANGE BATTERIES PATIENT CONFIRMED* PROBE FAILURE* 88 NIBP cannot be measured due to low battery. Function key is pressed and the “PATIENT CONFIRMED” signal is transmitted to the receiving monitor. (When “PATIENT” is assigned as the function for the Function key on the SYSTEM SETUP screen.) The probe is past its expiration date. Probe is damaged or short-circuited. Replace batteries with new ones. —— Replace the probe with a new one. Operator’s Manual ZM-540PA/541PA Message Cause Countermeasure If the message still appears after remeasurement, remove the cause NIBP is being remeasured due to REMEASURING arrhythmia, body movement, vibration if the cause is body movement, vibration or squeezing of cuff or or, cuff or air hose being squeezed. hose. Immediately remove the cuff from SAFETY The NIBP safety circuit error. the patient and contact your Nihon CIRCUIT ERROR Kohden representative. Check that the hose is not bent or SAFETY squeezed. CIRCUIT RUNNING (When this NIBP measurement stopped by the message is Wait 40 seconds, then perform safety circuit. displayed, remeasurement. If the message still measurement appears, contact your Nihon Kohden cannot be representative. performed for 40 seconds.) SpO2 MODULE Contact your Nihon Kohden Transmitter failure. representative. ERROR* The maximum blood pressure cannot be measured even when the cuff SYS OUT OF Measure by palpation or auscultation. inflation pressure exceeded 280 mmHg RANGE when using adult cuff. Check the patient condition and Poor peripheral circulation. change the probe attachment site. WEAK PULSE* Check the probe attachment (displayed in blue) The probe is attached too tightly and is condition and if necessary, reattach obstructing the blood circulation. the probe. The patient’s pulse wave is too small to measure NIBP. WEAK PULSE (displayed in pink) The cuff is wrapped too loosely. ZEROING Measure NIBP by palpation or auscultation. Wrap the cuff properly. The cuff size is not appropriate. Use the appropriate cuff. NIBP zero balance is being adjusted. Do not touch the cuff during zeroing. Wait for the message to disappear. Operator’s Manual ZM-540PA/541PA 89 Message Display Priority When more than one message condition occurs on the transmitter, only the message with the highest priority is displayed. Priority Highest Message PATIENT CONFIRMED SAFETY CIRCUIT RUNNING CUFF OCCLUSION PROBE FAILURE CHECK ELECTRODES NIBP MODULE ERROR SYS OUT OF RANGE HIGH CUFF PRESS AIR LEAK MEAS TIME OUT CANNOT DETECT PULSE (NIBP) SpO2 MODULE ERROR CHECK PROBE CHECK PROBE SITE CANNOT DETECT PULSE (SpO2) LIGHT INTERFERENCE REMEASURING INFLATION PRESS LOW WEAK PULSE (NIBP) ZEROING NO NIBP CHANGE BATTERIES DETECTING PULSE WEAK PULSE (SpO2) ALARMS SUSPENDED Lowest 90 BATTERY WEAK Operator’s Manual ZM-540PA/541PA Troubleshooting If a problem occurs, use the following to find and fix it. If the problem still remains after checking the following, contact your Nihon Kohden representative. Transmitter Problem Nothing is displayed on the LCD after turning the power on. LCD is difficult to see (too dark or too light). Cause Batteries are not installed correctly. The battery polarity is wrong. Batteries are completely discharged. LCD brightness is not appropriate. Nothing is displayed on the receiving monitor after turning the transmitter power on. The channel of the transmitter and monitor does not match. The software version of the multiple patient receiver or central monitor is old. Signal receiving condition is poor. Protocol on the transmitter and monitor does not match. Protocol on the transmitter is set to 57 but the monitor cannot receive protocol 57. Another transmitter with the same channel is used nearby. Signals of another patient are mixing. The transmitter is damaged. Operator’s Manual ZM-540PA/541PA Countermeasure Install the batteries correctly. Replace the batteries with new ones. Change the LCD brightness on the SYSTEM SETUP screen. Refer to the “Changing SYSTEM SETUP Settings” section. Set the correct channel on the monitor. Upgrade the multiple patient receiver or central monitor software to receive signal from the transmitter. The software version must be 01-09 or later. Set the same protocol on the transmitter and monitor. Set the protocol on the transmitter to 42. Refer to the “System Setup Setting List” section. Turn the transmitter power off. If the monitor still receives a signal, there is a high probability that another transmitter of the same channel is used nearby. Follow the instruction of your channel administrator and use another transmitter with a different channel. Follow the instructions of your channel administrator and use another transmitter of a different channel. Contact your Nihon Kohden representative. 91 ECG/Respiration Problem Cause The heart rate is unstable. Pacing detection setting on the monitor is not correct. No respiration rate is displayed. “RESP MEASUREMENT” on the PARAMETER SETUP screen is set to OFF. Countermeasure Turn off the pacing detection setting on the monitor. When monitoring a pacemaker patient, turn on pacing detection. The “CHECK Electrode lead is disconnected Firmly connect the electrode lead to the ELECTRODES” from the electrode. electrode. message appears Electrode lead discontinuity Replace the electrode lead with a new one. on the receiving Electrode is not firmly attached Replace the electrode with a new one. monitor. to the skin. Polarization voltage is Use Nihon Kohden specified electrodes. abnormally high. ECG baseline is The gel on the electrode is dried Replace the electrode with a new one. thick (AC hum) out. The gel on the electrode is coming off. An electric blanket is used. Cover the blanket with a shield cover. The hum filter is set to OFF on Set the filter to ON. the monitor The heart rate of a Pacing pulse detection is turned Turn OFF the pacing pulse detection on the receiving monitor. patient who is using ON on the receiving monitor. an electric blanket is unstable on the receiving monitor. No heart rate or “ECG MEASUREMENT” If ECG monitoring is necessary, set “ECG ECG is displayed. on the PARAMETER SETUP MEASUREMENT” to ON. screen is set to OFF. The gel on the electrode is dried Replace the electrode with a new one. Respiration out. waveform measurement is The gel on the electrode is unstable. coming off. 92 If respiration monitoring is necessary, set “RESP MEASUREMENT” to ON. Operator’s Manual ZM-540PA/541PA SpO2 Problem SpO2 data is unstable and not reliable. Cause Countermeasure The probe size is not appropriate for the patient. Probe attachment condition is poor. The probe is partly detached from the skin. External light is entering the probe. Measurement site is dirty. Patient is wearing nail polish. The probe is attached to the same limb that is used for NIBP measurement. Use the appropriate probe for the patient. Firmly attach the probe according to the procedure in the probe operator’s manual. Problem Cause Cuff inflation pressure The cuff hose is not connected is less than 10 mmHg. to the NIBP socket properly. The cuff is not wrapped around the arm or is wrapped too loosely. The cuff hose is not connected The cuff does not inflate when the NIBP to the NIBP socket. Start/Stop key is The cuff hose or extension hose pressed. may be folded or squeezed when the cuff pressure display on the screen increases quickly but the actual cuff does not inflate. The cuff size is not correct. Abnormal measurement results are displayed. The cuff is not wrapped around the arm correctly. NIBP data is not correct because of body movement. Vibration on the cuff. Countermeasure Connect the cuff hose to the socket properly. Wrap the cuff around the upper arm. Remove dirt and nail polish. Attach the probe to the opposite limb. Avoid a site where blood circulation condition changes greatly. NIBP Operator’s Manual ZM-540PA/541PA Connect the cuff hose to the socket firmly. Check the cuff hose and air hose. Select the cuff which fits the patient’s limb circumference. Wrap the cuff around the upper arm, not too tightly or too loosely. Prevent the patient from moving during measurement. Check that nothing is touching the cuff during measurement. Change the measuring site. 93 Problem The cuff is suddenly deflated during inflation. Auto mode measurement does not start even when the time interval has passed. The cuff suddenly inflates. Cannot connect cuff to the air hose. Cause The NIBP Start/Stop key is pressed during inflation. The NIBP Interval key is pressed and the measurement mode is changed. Countermeasure — Check the measurement mode and interval. The measurement mode is set to Check the time interval. If necessary, auto mode. stop measurement. Unspecified cuff is used. Use a cuff specified by Nihon Kohden. Cannot measure NIBP. Vibration on the cuff. Check that nothing is touching the cuff during measurement. The cuff hose or extension hose Remove the cause. is bent or squeezed. The cuff has worn out. Use a new cuff. Blood congestion Measuring over a long period of Increase the measuring interval. occurs. time at short intervals. Do not measure NIBP over a long time. Do not perform NIBP measurement on Thrombus occurs. Measuring on a patient with such a patient. known bleeding disorders or coagulation. When NIBP is measured again, the data NIBP data on the The time set for “OLD NIBP is displayed in normal brightness. screen is --- or dark. DATA” on the PARAMETER SETUP screen elapsed from the last measurement. Three loud pip sounds The cuff is not deflated enough Wait 30 seconds and measure again. to start another measurement. indicating NIBP measurement cannot be started. 94 Operator’s Manual ZM-540PA/541PA Maintenance To use the transmitter in safe and optimum condition, perform maintenance check every six months. The following units are necessary for some checking items. • AX-400G Vital Sign Simulator • AX-300T SpO2 Checker • Electric or mercury manometer • 700 mL dummy cuff • Receiving monitor CAUTION Never disassemble or repair the transmitter. Disassembly and repair must be performed by qualified service personnel. NOTE • The measurement accuracy of the above units must be managed to perform accurate maintenance check. • For details on the operation of the above units, refer to the manuals provided with these units. A maintenance check sheet is provided at the end of this section. Make a copy of this check sheet before performing maintenance check. 1. External Check • There is no damage or dirt on the outside of the transmitter. • The battery case cover is not damaged, the spring is firmly attached and the battery case cover can be closed firmly. • No keys are damaged. • NIBP socket is not damaged. • No electrode leads are damaged. • There are no blood or chemicals on the transmitter. 2. Transmitter Channel • The channel number label on the transmitter is not torn or removed. • The channel of the transmitter matches the label. The transmitter channel is displayed in the upper left corner of the screen. The channel number also appears on the startup screen. Operator’s Manual ZM-540PA/541PA 95 Channel number Startup screen Numeric and waveform screen 3. Transmitting/Receiving Signal Use the AX-400G vital sign simulator and receiving monitor. 1. Connect the vital sign simulator to the transmitter. AX-400G 2. Place the transmitter 2 to 3 m from the receiving monitor. 3. Set the channel on the receiving monitor to the channel of the transmitter. 4. Turn on the transmitter and vital sign simulator. 5. Check that the ECG of the transmitter appears on the receiving monitor. 6. Turn off the transmitter. 7. Check that the ECG disappears from the receiving monitor. 4. Display Check that there are no dots missing on the screen. 1. Turn off the transmitter. 2. While pressing the Function key, turn on the transmitter. The MENU screen appears. 96 Operator’s Manual ZM-540PA/541PA Press the ▼ 3. key to move the cursor to “MANUAL CHECK” and press the Function key. Press the or 5. Each time the key to move the cursor to “LCD TEST” and press the Function key. ▼ ▼ 4. ▼ Cursor key is pressed, the screen changes as below. Check that no dots are missing. → red → → green → blue 6. Press the Screen key to return to the MANUAL CHECK screen. 7. Press the Screen key again to return to the MENU screen. Operator’s Manual ZM-540PA/541PA → → → →... 97 5. Key Operation Turn off the transmitter. 2. While pressing the Function key, turn on the transmitter. The MENU screen appears. 3. Press the ▼ 1. key to move the cursor to “MANUAL CHECK” and press the Function key. ▼ or ▼ Cursor 4. Press the key to move the cursor to “KEY CHECK” and press the Function key. 5. Press each key one at a time and check that the pressed key is highlighted on the screen. Example: when the Function key is pressed, the key name is highlighted 6. 98 After checking, press and hold the Screen key to return to the MANUAL CHECK screen. Operator’s Manual ZM-540PA/541PA 6. ECG Check 1. Connect the vital sign simulator to the transmitter. AX-400G 2. Place the transmitter 1 m from the receiving monitor. 3. Turn on the transmitter and vital sign simulator. 4. Check that the ECG of the transmitter appears on the receiving monitor. 7. Respiration Check 1. Connect the vital sign simulator to the transmitter. AX-400G 2. Place the transmitter 1 m from the receiving monitor. 3. Turn on the transmitter and vital sign simulator. 4. Check that the respiration waveform of the transmitter appears on the receiving monitor. Operator’s Manual ZM-540PA/541PA 99 8. SpO2 Check 1. Connect the SpO2 checker to the transmitter. AX-300T 2. Place the transmitter 1 m from the receiving monitor. 3. Turn on the transmitter and SpO2 checker. 4. Check that the pulse bar graph appears on the transmitter screen. 5. Check that SpO2 and pulse rate on the transmitter is within the following range. SpO2 on the SpO2 Checker SpO2 Pulse rate 97% 80% 70% 60 beats/min 120 beats/min Range 95 to 99%SpO2 (±2 digit) 78 to 82%SpO2 (±2 digit) 67 to 73%SpO2 (±3 digit) 57 to 62 beats/min (±3%/±1 beat/min) 115 to 124 beats/min (±3%/±1 beat/min) NOTE The above range includes the error margin of the SpO2 checker. Therefore, the range is ±1 digit outside the SpO2 measuring accuracy of the transmitter in the “Specifications” section. 6. 100 Check that the SpO2 and pulse waveform of the transmitter appear on the receiving monitor. Operator’s Manual ZM-540PA/541PA 9. NIBP Check Check that the transmitter displays the correct cuff pressure and that there is no air leak. The following procedure uses an electric manometer. 1. Connect the electric manometer and dummy cuff to the transmitter. NOTE Air leaks from the rubber ball during inflation. Use a clip on the air hose of the rubber ball to stop air leaking. Clip Rubber ball Electric manometer Dummy cuff (700 cc) Turn on the electric manometer. 3. While pressing the Function key, turn on the transmitter. The MENU screen appears. 4. Press the ▼ 2. key to move the cursor to “MANUAL CHECK” and press the Function key. Cursor Operator’s Manual ZM-540PA/541PA 101 Press the ▼ 7. or or ▼ Press the ▼ 6. key to move the cursor to “NIBP HARDWARE TEST” and press the ▼ Press the or Function key. ▼ ▼ 5. key to move the cursor to “AIR LEAK TEST” and press the Function key. key to move the cursor to “AIR LEAK (AUTO)” and press the Function key. The transmitter inflates the cuff up to about 250 mmHg and measures air leakage from 60 seconds to 120 seconds after inflation. 102 Operator’s Manual ZM-540PA/541PA 8. Check the following. • The value for “AIR LEAK (AUTO)” is below 10 mmHg. • The difference between the pressure value displayed on the manometer and transmitter is within ±6 mmHg. 10. NIBP Cuff for Attaching Transmitter to Patient Arm The NIBP cuff is a consumable. Check the following and when necessary, replace it with a new one. Appearance • There are no dirty parts. • There are no broken stitches on the cuff. • The label on the cuff is readable. • The velcro tape on the cuff is not removed and there are no broken stitches. • The lock plate is not damaged and functions properly. Inflation bag • The inflation bag is not torn or damaged. • There is no water inside the inflation bag. • The connector on the inflation bag is not damaged. Operator’s Manual ZM-540PA/541PA 103 Maintenance Check Sheet Hospital/Organization: Service Personnel: Instrument Name: Transmitter Instrument Model: ZM-540PA/ZM-541PA Instrument Serial Number: Hardware Revision Number: Software Revision Number: 1. External Check OK No 2. Transmitter Channel OK No 3. Transmitting/Receiving Signal OK No 4. Display OK No 5. Key Operation OK No 6. ECG Check OK No 7. Respiration Check OK No 8. SpO2 Check OK No 9. NIBP Check OK No 10. NIBP Cuff for Attaching Transmitter to Patient Arm OK No Overall Judgement □ OK □ Can be used but needs maintenance □ Maintenance required. Cannot be used. 104 Operator’s Manual ZM-540PA/541PA Lifetime and Disposal CAUTION Dispose of the transmitter, options and accessories as specified by Nihon Kohden. Otherwise, it causes infection or environmental contamination. Disposing of Used Batteries Battery Lifetime Replace the batteries when the battery replacement indication appears on the transmitter. When using rechargeable batteries, recharge them. Disposal NOTE Remove the batteries before disposing of the transmitter. Before disposing of the batteries, check with your local solid waste officials for details in your area for proper disposal. It may be illegal to dispose of these batteries in the municipal waste stream. Disposing of Electrodes, SpO2 Probes and NIBP Cuffs Refer to the manual for each item. Disposing of Transmitter Remove the batteries from the transmitter and dispose of the transmitter following your local laws for disposal. Operator’s Manual ZM-540PA/541PA 105 Cleaning, Disinfection and Sterilization Transmitter and Electrode Leads CAUTION This transmitter is not waterproof. If detergent or liquid spills into the transmitter, stop cleaning or disinfecting it and contact your Nihon Kohden representative. The transmitter needs to be checked for safety and function before use. NOTE The transmitter cannot be sterilized. Sterilizing the transmitter may damage it. Before cleaning or disinfecting, remove the batteries from the transmitter. Be careful not to let any liquid get inside the transmitter. Cleaning Wipe the transmitter and electrode leads with a soft cloth moistened with disinfecting alcohol or neutral detergent diluted with water. After cleaning, dry them completely. Use cotton swab moistened with neutral detergent diluted with water to clean inside the battery compartment. Disinfection CAUTION Do not immerse the electrode lead connector in liquid. Wipe the outside surface of the transmitter and electrode lead with a non-abrasive cloth moistened with any of the disinfectants listed below. For details on the disinfectants, refer to the instruction provided with the disinfectants. Use the recommended concentration. Disinfectant Glutaraldehyde solution Alkyldiaminoethylglycine hydrochloride Benzalkonium chloride Benzethonium chloride solution Chlorhexidine gluconate solution 106 Concentration (%) 2.0 0.5 0.2 0.2 0.5 Operator’s Manual ZM-540PA/541PA SpO2 Probe Refer to the probe manual. YP-503P/YP-504P NIBP Cuffs NOTE • • • • Do not autoclave the cuff. Use only glutaraldehyde solution. Never allow liquid to get inside the inflation bag. Do not sterilize or disinfect the cuff with ultraviolet light or ozone. Cleaning To clean the cuff, remove the lock plate and carefully pull out the inflation bag from the cloth cover. Cloth cover: Wash with neutral detergent and water. Thoroughly dry it. When washing in a washing machine, put it in a net. Inflation bag: Wipe with a soft cloth or cotton moistened with isopropyl alcohol. Thoroughly dry it. Disinfection To disinfect the cuff, use glutaraldehyde solution. Use the recommended concentration of the disinfectant. Refer to the disinfectant manual for details. After disinfection, clean the cuff as described above. Repair Parts Availability Policy Nihon Kohden Corporation (NKC) shall stock repair parts (parts necessary to maintain the performance of the instrument) for a period of 8 years from the date of delivery. In that period NKC or its authorized agents will repair the instrument. This period may be shorter than 8 years if a board or part necessary for the faulty section is not available. Operator’s Manual ZM-540PA/541PA 107 Specifications ZM-540PA Measuring Parameters Measuring waveforms: Measuring numeric data: ECG, respiration in impedance method, pulse Heart rate, respiration rate, SpO2, NIBP, pulse rate Transmitting Data Waveform data: Numeric data: Status information: ECG, respiration, pulse wave SpO2, NIBP, pulse rate Battery replacement, battery level*, alarm suspended, pause monitoring*, patient confirmed*, ECG lead, pacing detection, electrode detachment, electrode impedance*, ECG off*, respiration method (impedance)*, SpO2 status, NIBP status, channel ID, time constant (3.2 s), type of transmitter, transmitter code number*, transmitter serial number* * The items marked with * are transmitted only when the protocol is “57”. Display Display size: Viewing area: Resolution: 2.2 inch TFT color LCD 44.16 (H) × 33.12 (V) mm 320 (H) × 240 (V) dots Displayed Data Numeric and waveform screen: Waveform review screen: Numeric review screen: Check electrodes screen: ECG (one waveform from lead I, II, III, Va or Vb), heart rate, pulse rate, respiration rate, SpO2, NIBP (systolic, diastolic, MAP), message, battery level, QRS/pulse sync mark, pulse bar graph, NIBP measurement mode and status information, ECG lead ECG or pulse wave of past 10 minutes Heart rate or pulse rate, respiration rate and SpO2 at 1 minute interval for past 10 minutes ECG for checking electrode attachment ECG ECG measurement Channels: 108 Operator’s Manual ZM-540PA/541PA Input dynamic range: Electrode offset potential tolerance: Input impedance: Common mode rejection ratio: ±10 mV or more ±500 mV or more 5 MΩ or more 95 dB or more IEC 60601-2-27 50.102.10 complied Pacing pulse detection: amplitude ±2 to 700 mV, duration 0.1 to 2 ms IEC 60601-2-27: 2005 complied Based upon pacemaker pulse rejection capability Defibrillation-proof: ECG input protected against 400 Ws/DC 5 kV IEC 60601-2-27 17.101 complied ECG recovery time after defibrillation: within 10 s Electrode condition: Displays CHECK ELECTRODES message Tall T-wave rejection capability: Complies with the heights of T-waves from 0 to 1.6 mV IEC 60601-2-27: 2005 50.102.17 complied Pacemaker pulse rejection capability, without overshoot: Complies with the amplitudes of pacemaker pulses ±2 to ±700 mV and widths 0.1 to 2 ms (As defined by method B of IEC 60601-2-27: 2005 50.102.13, this corresponds to the pacemaker pulses amplitudes and widths of ±2.8 mV/2 ms to amplitudes ±45.6 mV/0.1 ms.) Pacemaker pulse rejection capability, with overshoot: Overshoot amplitudes and time constants of ±0.12 mV/100 ms to ±2 mV/4 ms (As defined by method B of IEC 60601-2-27: 2005 50.102.13, this corresponds to the pacemaker pulses amplitudes and widths of ±2.8 mV/2 ms to amplitudes ±45.6 mV/0.1 ms.) ECG display and heart rate count Frequency characteristic: Heart rate detection method: QRS detection: filter on: 1 to 18 Hz, filter off: 0.05 to 60 Hz Average 70 to 120 ms: amplitude ≥ 0.5 mV, rate 30 to 200 beats/min 40 to 120 ms: amplitude ≥ 0.5 mV, rate 30 to 250 beats/min Heart rate counting range: 0, 15 to 300 beats/min Heart rate counting accuracy*: ±2 beats/min, (0, 15 to 300 beats/min) * Essential performance of this transmitter Respiration Measurement Measuring method: Impedance method Measuring lead: Between R and F Impedance range: 2 kΩ or less Respiration rate measuring accuracy*: ±2 counts/min (at 0 to 150 counts/min) Respiration rate counting range: 0 to 150 counts/min * Essential performance of this transmitter Operator’s Manual ZM-540PA/541PA 109 SpO2 Measurement (ISO 9919: 2005 complied) Measuring range: Declared range: Minimum display range: Display update cycle: Measuring accuracy (rms)*: Total accuracy including probe: 0 to 100%SpO2 70 to 100%SpO2 1%SpO2 Every 3 seconds Accuracy assurance temperature: 18 to 40ºC 80%SpO2 ≤ %SpO2 ≤ 100%SpO2: ±2%SpO2 70%SpO2 ≤ %SpO2 < 80%SpO2: ±3%SpO2 under 70%SpO2: not specified Accuracy of the transmitter: 80%SpO2 ≤ %SpO2 ≤ 100%SpO2: ±1%SpO2 50%SpO2 ≤ %SpO2 < 80%SpO2: ±2%SpO2 under 50%SpO2: not specified Pulse rate measuring range: 30 to 300 bpm Pulse rate display range: 30 to 300 bpm Pulse rate accuracy (rms)*: ±3% ±1 bpm * Essential performance of this transmitter NOTE • The SpO2 measuring accuracy was tested using the TL-201T, TL-260T, TL-271T and TL-631T SpO2 probes. The testing was performed during induced hypoxia on healthy volunteers (Ethnicity: 10 Caucasians, 2 Africans, 1 Asian and 3 Indians), (Skin: 8 light, 4 medium, 4 dark), (Age: 21 to 34), (5 women and 11 men) under the condition of no motion. Arterial blood was sampled and measured by a CO-oximeter. The difference between SpO2 measured by the SpO2 probe and functional SaO2 measured by a COoximeter was calculated using the root-mean-square (rms) method according to ISO 9919: 2005. This measurement accuracy figure represents 2/3 of all test measurements. • A pulse oximeter tester that generates simulated signals can be used to check the difference from the design specification, but it cannot be used as a replacement for human signals for testings accuracy. Response time: Selectable from “Slow”, “Normal” and “Fast”. The following graphs show the response time example when SpO2 changes 0.6%/s. 110 Operator’s Manual ZM-540PA/541PA Pulse Rate = 70 bpm SpO2 Reference SpO2 Fast SpO2 Normal SpO2 Slow 100 95 SpO2 (%) 90 85 80 75 70 65 60 10 20 30 40 50 60 70 80 90 100 110 120 130 140 Time (seconds) Pulse Rate = 140 bpm SpO2 Reference SpO2 Fast SpO2 Normal SpO2 Slow 100 95 SpO2 (%) 90 85 80 75 70 65 60 10 20 30 40 50 60 70 80 90 100 110 120 130 140 Time (seconds) The following graph shows the response time example when pulse rate changes 10 bpm/s. SpO2 = 97 PR Reference PR 150 140 PR (bpm) 130 120 110 100 90 80 70 60 10 20 30 40 50 60 70 80 90 100 110 120 Time (seconds) Operator’s Manual ZM-540PA/541PA 111 Non Invasive Blood Pressure, NIBP (IEC 60601-2-30: 1999 complied) Measuring method: Measurement mode: Intended patient type: Measuring range: Pressure display range: Measuring accuracy*: Cuff inflation time: Oscillometric Manual, STAT (≤ 15 min), Periodic Adult, child 0 to 300 mmHg 0 to 300 mmHg ±3 mmHg (0 mmHg ≤ NIBP ≤ 300 mmHg) AAMI SP-10: 2002 complied ≤ 20 s (700 cc), 0 to 200 mmHg ≤ 15 s (70 cc), 0 to 200 mmHg ≤ 5 mmHg (250 cc at 250 mmHg inflation for 10 seconds) ≤ 3 mmHg/min (700 cc at 300 mmHg inflation) Deflate immediately after power down Pressure retention: Air leakage: Power discontinuity: Safety Maximum pressurization value cuff inflation limiter: 300 to 330 mmHg Cuff inflation time limiter: ≤ 180 s Interval time limiter: ≤ 30 s * Essential performance of this transmitter Transmitter FCC regulation: Field strength limits: Undesired emissions: Antenna: Transmission channel: Transmission frequency range: Channel spacing: Modulation: Type of emission: Occupied bandwidth: Effective radiated power: FCC part 95 Subpart H Wireless Medical Telemetry Service (WMTS) < 200 mV/m (at 3 m) below 960 MHz: < 200 μV/m (at 3 m) above 960 MHz: < 500 μV/m (at 3 m) Internal Indicated on the transmitter 608.0250 to 613.9750 MHz 50 kHz or 37.5 kHz (12.5 kHz when interleaved) FSK (frequency shift keying) F1D < 20 kHz 1.0 mW Power Requirements Rated voltage: Operating voltage: Battery type: 3.6 V 3.2 to 4.8 V Three AA (R6) type NiMH secondary batteries Three AA (R6) type alkaline dry cell primary batteries Battery lifetime (with alkaline batteries, at room temperature): approximately 1 day (measuring ECG, respiration, SpO2 of approximately 60 kg weight adult male patient at the index finger, NIBP at 60 minute intervals) 112 Operator’s Manual ZM-540PA/541PA Dimension and Weight Dimension: Weight: 114 W × 125 H × 63 D (mm) about 340 g (excluding batteries and other accessories) about 410 g (including batteries, excluding other accessories) Environment Operating environment Temperature: Humidity: Atmospheric pressure: 5 to 40°C, 41 to 104°F 30 to 85% (noncondensing) 700 to 1060 hPa Storage and transport environment Temperature: Humidity: Atmospheric pressure: −20 to +65°C, −4 to +149°F 10 to 95% 700 to 1060 hPa Safety Standards Safety standard: CAN/CSA-C22.2 No. 601-1 M90 CAN/CSA-C22.2 No. 601-1. 1S1-94 CAN/CSA-C22.2 No. 601-1. 1B-90 CAN/CSA-C22.2 No. 60601-2-49-04 CAN/CSA-C22.2 No. 60601-2-27-06 CAN/CSA-C22.2 No. 60601-2-30-02 IEC 60601-1:1988 IEC 60601-1 Amendment 1: 1991 IEC 60601-1 Amendment 2: 1995 IEC 60601-2-27: 2005 IEC 60601-2-30: 1999 IEC 60601-2-49: 2001 ISO 9919: 2005 Type of protection against electrical shock: INTERNALLY POWERED EQUIPMENT Degree of protection against electrical shock: ECG and impedance method respiration: DEFIBRILLATION-PROOF TYPE CF APPLIED PART SpO2 and NIBP: DEFIBRILLATION-PROOF TYPE BF APPLIED PART Degree of protection against harmful ingress of water: IPX0 (Ordinary equipment) Degree of safety of application in the presence of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE: Equipment not suitable for use in the presence of FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE Operator’s Manual ZM-540PA/541PA 113 Mode of operation: CONTINUOUS OPERATION Electromagnetic Compatibility IEC 60601-1-2: 2001 IEC 60601-1-2 Amendment 1: 2004 Electromagnetic Emissions This Model ZM-540PA is intended for use in the electromagnetic environment specified below. The customer or the user of the ZM-540PA should assure that it is used in such an environment. Emissions test RF emissions CISPR 11 RF emissions CISPR 11 Harmonic emissions IEC 61000-3-2 Voltage fluctuations/ flicker emissions IEC 61000-3-3 114 Compliance Electromagnetic environment - guidance Group 1 The ZM-540PA uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. Class B The ZM-540PA is suitable for use in all establishments, including domestic establishments. Not applicable Not applicable Operator’s Manual ZM-540PA/541PA Electromagnetic Immunity This Model ZM-540PA is intended for use in the electromagnetic environment specified below. The customer or the user of the ZM-540PA should assure that it is used in such an environment. Immunity test Electrostatic discharge (ESD) IEC 61000-4-2 IEC 60601 test level ±6 kV contact ±8 kV air Compliance level ±6 kV contact ±8 kV air Not applicable ±2 kV for power supply lines ±1 kV for input/output lines Not applicable Surge ±1 kV differential IEC 61000-4-5 mode ±2 kV common mode Voltage dips, short <5% UT (>95% dip in Not applicable interruptions and UT) for 0.5 cycle voltage variations on power supply 40% UT (60% dip in input lines UT) for 5 cycles IEC 61000-4-11 70% UT (30% dip in UT) for 25 cycles Electrical fast transient/ burst IEC 61000-4-4 <5% UT (>95% dip in UT) for 5 s 3 A/m Electromagnetic environment - guidance Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. — — — Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. NOTE: UT is the AC mains voltage prior to application of the test level Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 3 A/m Avoiding Electromagnetic Interference (Impedance Respiration) Impedance respiration measurement is very sensitive and affected by electromagnetic interference. Technological limitations do not allow immunity levels higher than 1 V/m for radiated RF electromagnetic fields. Electromagnetic fields with field strengths above 1 V/m may cause measurement error. Do not use electrically radiating equipment near the impedance respiration measurements. Operator’s Manual ZM-540PA/541PA 115 Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Portable and mobile RF communications equipment should be used no closer to any part of the ZM-540PA, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance Conducted RF IEC 61000-4-6 3 Vrms 3 Vrms 150 kHz to 80 MHz d = 1.2㺕P Radiated RF IEC 61000-4-3 3 V/m 3 V/m 80 MHz to 80 MHz to 2.5 GHz 2.5 GHz d = 1.2㺕P 80 MHz to 800 MHz d = 2.3㺕P 800 MHz to 2.5 GHz (1 V/m 80 MHz to 2.5 GHz for respiration) (d = 3.5㺕P 80 MHz to 800 MHz for respiration d = 7.0㺕P 800 MHz to 2.5 GHz for respiration) where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as deter mined by an electromagnetic site survey*1, should be less than the compliance level in each frequency range*2. Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. *1 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the ZM-540PA is used exceeds the applicable RF compliance level above, the ZM-540PA should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the ZM-540PA. *2 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m for respiration and 3 V/m for all other functions. 116 Operator’s Manual ZM-540PA/541PA Recommended Separation Distances between Portable and Mobile RF Communications Equipment The ZM-540PA is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the ZM-540PA can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the ZM-540PA as recommended below, according to the maximum output power of the communications equipment. Separation distance according to frequency of transmitter (m) Rated maximum output power of transmitter (W) 150 kHz to 80 MHz d = 1.2㺕P 80 MHz to 800 MHz d = 1.2㺕P 800 MHz to 2.5 GHz d = 2.3㺕P (For respiration: d = 3.5㺕P) (For respiration: d = 7.0㺕P) 0.01 0.12 0.12 (0.35*) 0.23 (0.7*) 0.1 0.38 0.38 (1.1*) 0.73 (2.2*) 1.2 1.2 (3.5*) 2.3 (7.0*) 10 3.8 3.8 (11*) 7.3 (22*) 100 12 12 (35*) 23 (70*) For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. (* For respiration) NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Recovery Time after Defibrillation The transmitter returns to the normal operating mode within 10 seconds after defibrillation. The stored settings are not affected. System Composition for EMC Test The ZM-540PA bedside monitor is tested to comply with IEC 60601-1-2: 2001 with the following composition. Units ZM-540PA transmitter YP-503P NIBP cuff BR-906P ECG electrode lead TL-201T finger probe Operator’s Manual ZM-540PA/541PA Cable length — 0.15 m 0.8 m 1.6 m 117 ZM-541PA Measuring Parameters Measuring waveforms: Measuring numeric data: ECG, respiration in impedance method, pulse Heart rate, respiration rate, SpO2, NIBP, pulse rate Transmitting Data Waveform data: Numeric data: Status information: ECG, respiration, pulse wave SpO2, NIBP, pulse rate Battery replacement, battery level*, alarm suspended, pause monitoring*, patient confirmed*, ECG lead, pacing detection, electrode detachment, electrode impedance*, ECG off*, respiration method (impedance)*, SpO2 status, NIBP status, channel ID, time constant (3.2 s), type of transmitter, transmitter code number*, transmitter serial number* * The items marked with * are transmitted only when the protocol is “57”. Display Display size: Viewing area: Resolution: 2.2 inch TFT color LCD 44.16 (H) × 33.12 (V) mm 320 (H) × 240 (V) dots Displayed Data Numeric and waveform screen: Waveform review screen: Numeric review screen: CHECK ELECTRODE screen: ECG (one waveform from lead I, II, III, Va or Vb), heart rate, pulse rate, respiration rate, SpO2, NIBP (systolic, diastolic, MAP), message, battery level, QRS/pulse sync mark, pulse bar graph, NIBP measurement mode and status information, ECG lead ECG or pulse wave of past 10 minutes Heart rate or pulse rate, respiration rate and SpO2 at 1 minute interval for past 10 minutes ECG for checking electrode attachment ECG ECG measurement Channels: Input dynamic range: Electrode offset potential tolerance: Input impedance: Common mode rejection ratio: 118 ±10 mV or more ±500 mV or more 5 MΩ or more 95 dB or more IEC 60601-2-27 50.102.10 complied Operator’s Manual ZM-540PA/541PA Pacing pulse detection: amplitude ±2 to 700 mV, duration 0.1 to 2 ms IEC 60601-2-27: 2005 complied Based upon pacemaker pulse rejection capability Defibrillation-proof: ECG input protected against 400 Ws/DC 5 kV IEC 60601-2-27 17.101 complied ECG recovery time after defibrillation: within 10 s Electrode condition: Displays CHECK ELECTRODES message Tall T-wave rejection capability: Complies with the heights of T-waves from 0 to 1.6 mV IEC 60601-2-27: 2005 50.102.17 complied Pacemaker pulse rejection capability, without overshoot: Complies with the amplitudes of pacemaker pulses ±2 to ±700 mV and widths 0.1 to 2 ms (As defined by method B of IEC 60601-2-27: 2005 50.102.13, this corresponds to the pacemaker pulses amplitudes and widths of ±2.8 mV/2 ms to amplitudes ±45.6 mV/0.1 ms.) Pacemaker pulse rejection capability, with overshoot: Overshoot amplitudes and time constants of ±0.12 mV/100 ms to ±2 mV/4 ms (As defined by method B of IEC 60601-2-27: 2005 50.102.13, this corresponds to the pacemaker pulses amplitudes and widths of ±2.8 mV/2 ms to amplitudes ±45.6 mV/0.1 ms.) ECG display and heart rate count Frequency characteristic: Heart rate detection method: QRS detection: filter on: 1 to 18 Hz, filter off: 0.05 to 60 Hz Average 70 to 120 ms: amplitude ≥ 0.5 mV, rate 30 to 200 beats/min 40 to 120 ms: amplitude ≥ 0.5 mV, rate 30 to 250 beats/min Heart rate counting range: 0, 15 to 300 beats/min Heart rate counting accuracy*: ±2 beats/min, (0, 15 to 300 beats/min) * Essential performance of this transmitter Respiration Measurement Measuring method: Impedance method Measuring lead: Between R and F Impedance range: 2 kΩ or less Respiration rate measuring accuracy*: ±2 counts/min (at 0 to 150 counts/min) Respiration rate counting range: 0 to 150 counts/min * Essential performance of this transmitter SpO2 Measurement (ISO 9919: 2005 complied) Measuring range: Declared range: Minimum display range: 0 to 100%SpO2 70 to 100%SpO2 1%SpO2 Operator’s Manual ZM-540PA/541PA 119 Display update cycle: Every 3 seconds Measuring accuracy (rms)*: Accuracy assurance temperature: 18 to 40ºC Total accuracy including probe: 80%SpO2 ≤ %SpO2 ≤ 100%SpO2: ±2%SpO2 70%SpO2 ≤ %SpO2 < 80%SpO2: ±3%SpO2 under 70%SpO2: not specified Accuracy of the transmitter: 80%SpO2 ≤ %SpO2 ≤ 100%SpO2: ±1%SpO2 50%SpO2 ≤ %SpO2 < 80%SpO2: ±2%SpO2 under 50%SpO2: not specified Pulse rate measuring range: 30 to 300 bpm Pulse rate display range: 30 to 300 bpm Pulse rate accuracy (rms)*: ±3% ±1 bpm * Essential performance of this transmitter NOTE • The SpO2 measuring accuracy was tested using the TL-201T, TL-260T, TL-271T and TL-631T SpO2 probes. The testing was performed during induced hypoxia on healthy volunteers (Ethnicity: 10 Caucasians, 2 Africans, 1 Asian and 3 Indians), (Skin: 8 light, 4 medium, 4 dark), (Age: 21 to 34), (5 women and 11 men) under the condition of no motion. Arterial blood was sampled and measured by a CO-oximeter. The difference between SpO2 measured by the SpO2 probe and functional SaO2 measured by a COoximeter was calculated using the root-mean-square (rms) method according to ISO 9919: 2005. This measurement accuracy figure represents 2/3 of all test measurements. • A pulse oximeter tester that generates simulated signals can be used to check the difference from the design specification, but it cannot be used as a replacement for human signals for testings accuracy. Response time: Selectable from “Slow”, “Normal” and “Fast”. The following graphs show the response time example when SpO2 changes 0.6%/s. Pulse Rate = 70 bpm SpO2 Reference SpO2 Fast SpO2 Normal SpO2 Slow 100 95 SpO2 (%) 90 85 80 75 70 65 60 10 20 30 40 50 60 70 80 90 100 110 120 130 140 Time (seconds) 120 Operator’s Manual ZM-540PA/541PA Pulse Rate = 140 bpm SpO2 Reference SpO2 Fast SpO2 Normal SpO2 Slow 100 95 SpO2 (%) 90 85 80 75 70 65 60 10 20 30 40 50 60 70 80 90 100 110 120 130 140 Time (seconds) The following graph shows the response time example when pulse rate changes 10 bpm/s. SpO2 = 97 PR Reference PR 150 140 PR (bpm) 130 120 110 100 90 80 70 60 10 20 30 40 50 60 70 80 90 100 110 120 Time (seconds) Non Invasive Blood Pressure, NIBP (IEC 60601-2-30: 1999 complied) Measuring method: Measurement mode: Intended patient type: Measuring range: Pressure display range: Measuring accuracy*: Cuff inflation time: Pressure retention: Air leakage: Power discontinuity: Oscillometric Manual, STAT (≤ 15 min), Periodic Adult, child 0 to 300 mmHg 0 to 300 mmHg ±3 mmHg (0 mmHg ≤ NIBP ≤ 300 mmHg) AAMI SP-10: 2002 complied ≤ 20 s (700 cc), 0 to 200 mmHg ≤ 15 s (70 cc), 0 to 200 mmHg ≤ 5 mmHg (250 cc at 250 mmHg inflation for 10 seconds) ≤ 3 mmHg/min (700 cc at 300 mmHg inflation) Deflate immediately after power down Operator’s Manual ZM-540PA/541PA 121 Safety Maximum pressurization value cuff inflation limiter: 300 to 330 mmHg Cuff inflation time limiter: ≤ 180 s Interval time limiter: ≤ 30 s * Essential performance of this transmitter Transmitter FCC regulation: Field strength limits: Undesired emissions: Antenna: Transmission channel: Transmission frequency range: Channel spacing: Modulation: Type of emission: Occupied bandwidth: Effective radiated power: FCC part 95 Subpart H Wireless Medical Telemetry Service (WMTS) < 740 mV/m (at 3 m) below 960 MHz: < 200 μV/m (at 3 m) above 960 MHz: < 500 μV/m (at 3 m) Internal Indicated on the transmitter 1395.0250 to 1399.9750 MHz 1427.0250 to 1431.9750 MHz 50 kHz or 37.5 kHz (12.5 kHz when interleaved) FSK (frequency shift keying) F1D < 20 kHz 5.0 mW Can be changed to 1.0 mW if required Power Requirements Rated voltage: Operating voltage: Battery type: 3.6 V 3.2 to 4.8 V Three AA (R6) type NiMH secondary batteries Three AA (R6) type alkaline dry cell primary batteries Battery lifetime (with alkaline batteries, at room temperature): approximately 1 day (measuring ECG, respiration, SpO2 of approximately 60 kg weight adult male patient at the index finger, NIBP at 60 minute intervals) Dimension and Weight Dimension: Weight: 122 114 W × 125 H × 63 D (mm) about 340 g (excluding batteries and other accessories) about 410 g (including batteries, excluding other accessories) Operator’s Manual ZM-540PA/541PA Environment Operating environment Temperature: Humidity: Atmospheric pressure: 5 to 40°C, 41 to 104°F 30 to 85% (noncondensing) 700 to 1060 hPa Storage and transport environment Temperature: Humidity: Atmospheric pressure: −20 to +65°C, −4 to +149°F 10 to 95% 700 to 1060 hPa Safety Standards Safety standard: CAN/CSA-C22.2 No. 601-1 M90 CAN/CSA-C22.2 No. 601-1. 1S1-94 CAN/CSA-C22.2 No. 601-1. 1B-90 CAN/CSA-C22.2 No. 60601-2-49-04 CAN/CSA-C22.2 No. 60601-2-27-06 CAN/CSA-C22.2 No. 60601-2-30-02 IEC 60601-1:1988 IEC 60601-1 Amendment 1: 1991 IEC 60601-1 Amendment 2: 1995 IEC 60601-2-27: 2005 IEC 60601-2-30: 1999 IEC 60601-2-49: 2001 ISO 9919: 2005 Type of protection against electrical shock: INTERNALLY POWERED EQUIPMENT Degree of protection against electrical shock: ECG and impedance method respiration: DEFIBRILLATION-PROOF TYPE CF APPLIED PART SpO2 and NIBP: DEFIBRILLATION-PROOF TYPE BF APPLIED PART Degree of protection against harmful ingress of water: IPX0 (Ordinary equipment) Degree of safety of application in the presence of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE: Equipment not suitable for use in the presence of FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE Mode of operation: CONTINUOUS OPERATION Electromagnetic Compatibility IEC 60601-1-2: 2001 IEC 60601-1-2 Amendment 1: 2004 Operator’s Manual ZM-540PA/541PA 123 Electromagnetic Emissions This Model ZM-541PA is intended for use in the electromagnetic environment specified below. The customer or the user of the ZM-541PA should assure that it is used in such an environment. Emissions test RF emissions CISPR 11 RF emissions CISPR 11 Harmonic emissions IEC 61000-3-2 Voltage fluctuations/ flicker emissions IEC 61000-3-3 124 Compliance Electromagnetic environment - guidance Group 1 The ZM-541PA uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. Class B The ZM-541PA is suitable for use in all establishments, including domestic establishments. Not applicable Not applicable Operator’s Manual ZM-540PA/541PA Electromagnetic Immunity This Model ZM-541PA is intended for use in the electromagnetic environment specified below. The customer or the user of the ZM-541PA should assure that it is used in such an environment. Immunity test Electrostatic discharge (ESD) IEC 61000-4-2 IEC 60601 test level ±6 kV contact ±8 kV air Compliance level ±6 kV contact ±8 kV air Not applicable ±2 kV for power supply lines ±1 kV for input/output lines Not applicable Surge ±1 kV differential IEC 61000-4-5 mode ±2 kV common mode Voltage dips, short <5% UT (>95% dip in Not applicable interruptions and UT) for 0.5 cycle voltage variations on power supply 40% UT (60% dip in input lines UT) for 5 cycles IEC 61000-4-11 70% UT (30% dip in UT) for 25 cycles Electrical fast transient/ burst IEC 61000-4-4 <5% UT (>95% dip in UT) for 5 s 3 A/m Electromagnetic environment - guidance Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. — — — Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. NOTE: UT is the AC mains voltage prior to application of the test level Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 3 A/m Avoiding Electromagnetic Interference (Impedance Respiration) Impedance respiration measurement is very sensitive and affected by electromagnetic interference. Technological limitations do not allow immunity levels higher than 1 V/m for radiated RF electromagnetic fields. Electromagnetic fields with field strengths above 1 V/m may cause measurement error. Do not use electrically radiating equipment near the impedance respiration measurements. Operator’s Manual ZM-540PA/541PA 125 Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Portable and mobile RF communications equipment should be used no closer to any part of the ZM-541PA, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance Conducted RF IEC 61000-4-6 3 Vrms 3 Vrms 150 kHz to 80 MHz d = 1.2㺕P Radiated RF IEC 61000-4-3 3 V/m 3 V/m 80 MHz to 80 MHz to 2.5 GHz 2.5 GHz d = 1.2㺕P 80 MHz to 800 MHz d = 2.3㺕P 800 MHz to 2.5 GHz (1 V/m 80 MHz to 2.5 GHz for respiration) (d = 3.5㺕P 80 MHz to 800 MHz for respiration d = 7.0㺕P 800 MHz to 2.5 GHz for respiration) where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as deter mined by an electromagnetic site survey*1, should be less than the compliance level in each frequency range*2. Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. *1 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the ZM-541PA is used exceeds the applicable RF compliance level above, the ZM-541PA should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the ZM-541PA. *2 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m for respiration and 3 V/m for all other functions. 126 Operator’s Manual ZM-540PA/541PA Recommended Separation Distances between Portable and Mobile RF Communications Equipment The ZM-541PA is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the ZM-541PA can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the ZM-541PA as recommended below, according to the maximum output power of the communications equipment. Separation distance according to frequency of transmitter (m) Rated maximum output power of transmitter (W) 150 kHz to 80 MHz d = 1.2㺕P 80 MHz to 800 MHz d = 1.2㺕P 800 MHz to 2.5 GHz d = 2.3㺕P (For respiration: d = 3.5㺕P) (For respiration: d = 7.0㺕P) 0.01 0.12 0.12 (0.35*) 0.23 (0.7*) 0.1 0.38 0.38 (1.1*) 0.73 (2.2*) 1.2 1.2 (3.5*) 2.3 (7.0*) 10 3.8 3.8 (11*) 7.3 (22*) 100 12 12 (35*) 23 (70*) For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. (* For respiration) NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Recovery Time after Defibrillation The transmitter returns to the normal operating mode within 10 seconds after defibrillation. The stored settings are not affected. System Composition for EMC Test The ZM-541PA bedside monitor is tested to comply with IEC 60601-1-2: 2001 with the following composition. Units ZM-541PA transmitter YP-503P NIBP cuff BR-906P ECG electrode lead TL-201T finger probe Operator’s Manual ZM-540PA/541PA Cable length — 0.15 m 0.8 m 1.6 m 127 Standard Accessories No. Name 1 NIBP cuff for adult, standard 2 Strap Model YP-503P — Q’ty Supply Code No. S937C Y236 The following parts are available for replacement. Lock plate is a standard accessory of the YP-503P and YP504P NIBP cuff. No. Name 3 Battery case cover 4 Lock plate 128 Model — — Q’ty Supply Code No. 6143-901517A 6113-049585 Operator’s Manual ZM-540PA/541PA Options CAUTION Only use Nihon Kohden specified electrodes, electrode leads, SpO2 probes, and NIBP cuffs. Otherwise, the maximum performance from the transmitter cannot be guaranteed. Transmitter • Channel writer, QI-901PK • NIBP cuff for adult, large (width 15 cm) (for attaching transmitter to patient arm), YP-504P, supply code no. S937D ECG/RESP Name Electrode lead Application 3 electrodes, clip type, lead length 80 cm 6 electrodes, clip type, lead length 80 cm Operator’s Manual ZM-540PA/541PA Model Q’ty Supply Code No. BR-903PA K911A BR-906PA K912A 129 SpO2 Name Finger probe (reusable) Multi-site probe (reusable) Finger probe (reusable) SpO2 probe (for adult, disposable) SpO2 probe (for child, disposable) SpO2 probe (for neonate, disposable) SpO2 probe (for adult, disposable) SpO2 probe (for child, disposable) SpO2 probe (for neonate/adult, disposable) SpO2 probe (for child/infant, disposable) SpO2 probe (for adult/neonate, disposable) SpO2 probe (for child/infant, disposable) COTTONY tape Foam tape for TL-051S/052S/ 061S/062S Attachment tape for TL-220T/ 251T/252T/253T/630T/631T Probe fastener 130 Cable length 0.6 m 1.6 m 1.6 m 0.6 m 1.6 m 0.6 m 1.6 m 1.6 m 0.8 m 1.6 m 0.8 m 1.6 m 0.8 m 1.6 m 0.8 m 1.6 m 0.8 m 1.6 m 0.8 m 1.6 m — Model/ Code No. 20 Supply Code No. P225H P225F P225G P310A P310C P311A P311C P201A P201B P201C P203A P203E P203B P203F P203C P203G P203D P203H P228A P228B P229A P229B P259 4 × 25 packages P260 3 × 30 packages P263 30 P267 Q’ty TL-201T TL-220T TL-630T1 TL-630T3 TL-631T1 TL-631T3 TL-251T TL-252T TL-253T TL-271T TL-271T3 TL-272T TL-272T3 TL-273T TL-273T3 TL-274T TL-274T3 TL-051S TL-052S TL-061S TL-062S 340703 24 — YS-093P2 Operator’s Manual ZM-540PA/541PA NIBP Width (cm) Name Cuff for adult, for attaching transmitter to patient arm Cuff for infant Cuff for child Cuff for adult Standard Large Small Standard Standard Large Disposable cuff for infant Disposable cuff for child Disposable cuff for adult Small Standard Medium large Large Extension hose Air Hose Length (cm) 13 15 Model Q’ty YP-503P* 15 YP-504P* 10 13 15 10 14 17 17 17 20 YP-960T YP-961T YP-962T YP-963T YP-964T YP-810P YP-811P YP-812P YP-813P 15 20 YP-814P 17 — 20 150 YP-815P YN-990P 15 Supply Code No. S937C 20 S937D S943A S943B S943C S944B S944C S945C S945D S946E S946F S946G S946H S903 * The lock plate is provided with these NIBP cuffs. Operator’s Manual ZM-540PA/541PA 131 Transmission Frequencies Channel: 9002 to 9478 Transmission Transmission Transmission frequency Channel No. frequency Channel No. frequency Channel No. (MHz) (MHz) (MHz) 608.0250 608.0375 608.0500 608.0625 608.0750 608.0875 608.1000 608.1125 608.1250 608.1375 608.1500 608.1625 608.1750 608.1875 608.2000 608.2125 608.2250 608.2375 608.2500 608.2625 608.2750 608.2875 608.3000 608.3125 608.3250 608.3375 608.3500 608.3625 608.3750 608.3875 608.4000 608.4125 608.4250 132 9002 9003 9004 9005 9006 9007 9008 9009 9010 9011 9012 9013 9014 9015 9016 9017 9018 9019 9020 9021 9022 9023 9024 9025 9026 9027 9028 9029 9030 9031 9032 9033 9034 608.4375 608.4500 608.4625 608.4750 608.4875 608.5000 608.5125 608.5250 608.5375 608.5500 608.5625 608.5750 608.5875 608.6000 608.6125 608.6250 608.6375 608.6500 608.6625 608.6750 608.6875 608.7000 608.7125 608.7250 608.7375 608.7500 608.7625 608.7750 608.7875 608.8000 608.8125 608.8250 608.8375 9035 9036 9037 9038 9039 9040 9041 9042 9043 9044 9045 9046 9047 9048 9049 9050 9051 9052 9053 9054 9055 9056 9057 9058 9059 9060 9061 9062 9063 9064 9065 9066 9067 608.8500 608.8625 608.8750 608.8875 608.9000 608.9125 608.9250 608.9375 608.9500 608.9625 608.9750 608.9875 609.0000 609.0125 609.0250 609.0375 609.0500 609.0625 609.0750 609.0875 609.1000 609.1125 609.1250 609.1375 609.1500 609.1625 609.1750 609.1875 609.2000 609.2125 609.2250 609.2375 609.2500 9068 9069 9070 9071 9072 9073 9074 9075 9076 9077 9078 9079 9080 9081 9082 9083 9084 9085 9086 9087 9088 9089 9090 9091 9092 9093 9094 9095 9096 9097 9098 9099 9100 Operator’s Manual ZM-540PA/541PA Transmission Transmission Transmission frequency Channel No. frequency Channel No. frequency Channel No. (MHz) (MHz) (MHz) 609.2625 609.2750 609.2875 609.3000 609.3125 609.3250 609.3375 609.3500 609.3625 609.3750 609.3875 609.4000 609.4125 609.4250 609.4375 609.4500 609.4625 609.4750 609.4875 609.5000 609.5125 609.5250 609.5375 609.5500 609.5625 609.5750 609.5875 609.6000 609.6125 609.6250 9101 9102 9103 9104 9105 9106 9107 9108 9109 9110 9111 9112 9113 9114 9115 9116 9117 9118 9119 9120 9121 9122 9123 9124 9125 9126 9127 9128 9129 9130 609.7125 609.7250 609.7375 609.7500 609.7625 609.7750 609.7875 609.8000 609.8125 609.8250 609.8375 609.8500 609.8625 609.8750 609.8875 609.9000 609.9125 609.9250 609.9375 609.9500 609.9625 609.9750 609.9875 610.0000 610.0125 610.0250 610.0375 610.0500 610.0625 610.0750 9137 9138 9139 9140 9141 9142 9143 9144 9145 9146 9147 9148 9149 9150 9151 9152 9153 9154 9155 9156 9157 9158 9159 9160 9161 9162 9163 9164 9165 9166 610.1625 610.1750 610.1875 610.2000 610.2125 610.2250 610.2375 610.2500 610.2625 610.2750 610.2875 610.3000 610.3125 610.3250 610.3375 610.3500 610.3625 610.3750 610.3875 610.4000 610.4125 610.4250 610.4375 610.4500 610.4625 610.4750 610.4875 610.5000 610.5125 610.5250 9173 9174 9175 9176 9177 9178 9179 9180 9181 9182 9183 9184 9185 9186 9187 9188 9189 9190 9191 9192 9193 9194 9195 9196 9197 9198 9199 9200 9201 9202 609.6375 9131 610.0875 9167 610.5375 9203 609.6500 9132 610.1000 9168 610.5500 9204 9205 609.6625 9133 610.1125 9169 610.5625 609.6750 9134 610.1250 9170 610.5750 9206 9207 9208 609.6875 9135 610.1375 9171 610.5875 609.7000 9136 610.1500 9172 610.6000 Operator’s Manual ZM-540PA/541PA 133 Transmission Transmission Transmission frequency Channel No. frequency Channel No. frequency Channel No. (MHz) (MHz) (MHz) 610.6125 9209 610.6250 9210 611.0750 9246 611.5250 9282 610.6375 610.6500 610.6625 610.6750 610.6875 610.7000 610.7125 610.7250 610.7375 610.7500 610.7625 610.7750 610.7875 610.8000 610.8125 610.8250 610.8375 610.8500 610.8625 610.8750 610.8875 610.9000 610.9125 610.9250 610.9375 610.9500 610.9625 610.9750 610.9875 611.0000 611.0125 611.0250 611.0375 611.0500 9211 9212 9213 9214 9215 9216 9217 9218 9219 9220 9221 9222 9223 9224 9225 9226 9227 9228 9229 9230 9231 9232 9233 9234 9235 9236 9237 9238 9239 9240 9241 9242 9243 9244 611.0875 611.1000 611.1125 611.1250 611.1375 611.1500 611.1625 611.1750 611.1875 611.2000 611.2125 611.2250 611.2375 611.2500 611.2625 611.2750 611.2875 611.3000 611.3125 611.3250 611.3375 611.3500 611.3625 611.3750 611.3875 611.4000 611.4125 611.4250 611.4375 611.4500 611.4625 611.4750 611.4875 611.5000 9247 9248 9249 9250 9251 9252 9253 9254 9255 9256 9257 9258 9259 9260 9261 9262 9263 9264 9265 9266 9267 9268 9269 9270 9271 9272 9273 9274 9275 9276 9277 9278 9279 9280 611.5375 611.5500 611.5625 611.5750 611.5875 611.6000 611.6125 611.6250 611.6375 611.6500 611.6625 611.6750 611.6875 611.7000 611.7125 611.7250 611.7375 611.7500 611.7625 611.7750 611.7875 611.8000 611.8125 611.8250 611.8375 611.8500 611.8625 611.8750 611.8875 611.9000 611.9125 611.9250 611.9375 611.9500 9283 9284 9285 9286 9287 9288 9289 9290 9291 9292 9293 9294 9295 9296 9297 9298 9299 9300 9301 9302 9303 9304 9305 9306 9307 9308 9309 9310 9311 9312 9313 9314 9315 9316 134 611.0625 9245 611.5125 9281 Operator’s Manual ZM-540PA/541PA Transmission Transmission Transmission frequency Channel No. frequency Channel No. frequency Channel No. (MHz) (MHz) (MHz) 611.9625 611.9750 611.9875 612.0000 612.0125 612.0250 612.0375 612.0500 612.0625 612.0750 612.0875 612.1000 612.1125 612.1250 612.1375 612.1500 612.1625 612.1750 612.1875 612.2000 612.2125 612.2250 612.2375 612.2500 612.2625 612.2750 612.2875 612.3000 612.3125 612.3250 612.3375 612.3500 612.3625 612.3750 612.3875 612.4000 9317 9318 9319 9320 9321 9322 9323 9324 9325 9326 9327 9328 9329 9330 9331 9332 9333 9334 9335 9336 9337 9338 9339 9340 9341 9342 9343 9344 9345 9346 9347 9348 9349 9350 9351 9352 612.4125 612.4250 612.4375 612.4500 612.4625 612.4750 612.4875 612.5000 612.5125 612.5250 612.5375 612.5500 612.5625 612.5750 612.5875 612.6000 612.6125 612.6250 612.6375 612.6500 612.6625 612.6750 612.6875 612.7000 612.7125 612.7250 612.7375 612.7500 612.7625 612.7750 612.7875 612.8000 612.8125 612.8250 612.8375 612.8500 Operator’s Manual ZM-540PA/541PA 9353 9354 9355 9356 9357 9358 9359 9360 9361 9362 9363 9364 9365 9366 9367 9368 9369 9370 9371 9372 9373 9374 9375 9376 9377 9378 9379 9380 9381 9382 9383 9384 9385 9386 9387 9388 612.8625 612.8750 612.8875 612.9000 612.9125 612.9250 612.9375 612.9500 612.9625 612.9750 612.9875 613.0000 613.0125 613.0250 613.0375 613.0500 613.0625 613.0750 613.0875 613.1000 613.1125 613.1250 613.1375 613.1500 613.1625 613.1750 613.1875 613.2000 613.2125 613.2250 613.2375 613.2500 613.2625 613.2750 613.2875 613.3000 9389 9390 9391 9392 9393 9394 9395 9396 9397 9398 9399 9400 9401 9402 9403 9404 9405 9406 9407 9408 9409 9410 9411 9412 9413 9414 9415 9416 9417 9418 9419 9420 9421 9422 9423 9424 135 Transmission Transmission Transmission frequency Channel No. frequency Channel No. frequency Channel No. (MHz) (MHz) (MHz) 136 613.3125 9425 613.5375 9443 613.7625 9461 613.3250 9426 613.5500 9444 613.7750 9462 613.3375 9427 613.5625 9445 613.7875 9463 613.3500 9428 613.5750 9446 613.8000 9464 613.3625 9429 613.5875 9447 613.8125 9465 613.3750 9430 613.6000 9448 613.8250 9466 613.3875 9431 613.6125 9449 613.8375 9467 613.4000 9432 613.6250 9450 613.8500 9468 613.4125 9433 613.6375 9451 613.8625 9469 613.4250 9434 613.6500 9452 613.8750 9470 613.4375 9435 613.6625 9453 613.8875 9471 613.4500 9436 613.6750 9454 613.9000 9472 613.4625 9437 613.6875 9455 613.9125 9473 613.4750 9438 613.7000 9456 613.9250 9474 613.4875 9439 613.7125 9457 613.9375 9475 613.5000 9440 613.7250 9458 613.9500 9476 613.5125 9441 613.7375 9459 613.9625 9477 613.5250 9442 613.7500 9460 613.9750 9478 Operator’s Manual ZM-540PA/541PA Channel: E002 to E398 Transmission frequency (MHz) 1395.0250 1395.0375 1395.0500 1395.0625 1395.0750 1395.0875 1395.1000 1395.1125 1395.1250 1395.1375 1395.1500 1395.1625 1395.1750 1395.1875 1395.2000 1395.2125 1395.2250 1395.2375 1395.2500 1395.2625 1395.2750 1395.2875 1395.3000 1395.3125 1395.3250 1395.3375 1395.3500 1395.3625 1395.3750 1395.3875 1395.4000 1395.4125 1395.4250 1395.4375 1395.4500 Channel No. E002 E003 E004 E005 E006 E007 E008 E009 E010 E011 E012 E013 E014 E015 E016 E017 E018 E019 E020 E021 E022 E023 E024 E025 E026 E027 E028 E029 E030 E031 E032 E033 E034 E035 E036 Transmission frequency (MHz) 1395.4625 1395.4750 1395.4875 1395.5000 1395.5125 1395.5250 1395.5375 1395.5500 1395.5625 1395.5750 1395.5875 1395.6000 1395.6125 1395.6250 1395.6375 1395.6500 1395.6625 1395.6750 1395.6875 1395.7000 1395.7125 1395.7250 1395.7375 1395.7500 1395.7625 1395.7750 1395.7875 1395.8000 1395.8125 1395.8250 1395.8375 1395.8500 1395.8625 1395.8750 1395.8875 Operator’s Manual ZM-540PA/541PA Channel No. E037 E038 E039 E040 E041 E042 E043 E044 E045 E046 E047 E048 E049 E050 E051 E052 E053 E054 E055 E056 E057 E058 E059 E060 E061 E062 E063 E064 E065 E066 E067 E068 E069 E070 E071 Transmission frequency (MHz) Channel No. 1395.9000 1395.9125 1395.9250 1395.9375 1395.9500 1395.9625 1395.9750 1395.9875 1396.0000 1396.0125 1396.0250 1396.0375 1396.0500 1396.0625 1396.0750 1396.0875 1396.1000 1396.1125 1396.1250 1396.1375 1396.1500 1396.1625 1396.1750 1396.1875 1396.2000 1396.2125 1396.2250 1396.2375 1396.2500 1396.2625 1396.2750 1396.2875 1396.3000 1396.3125 1396.3250 E072 E073 E074 E075 E076 E077 E078 E079 E080 E081 E082 E083 E084 E085 E086 E087 E088 E089 E090 E091 E092 E093 E094 E095 E096 E097 E098 E099 E100 E101 E102 E103 E104 E105 E106 137 Transmission frequency (MHz) 1396.3375 1396.3500 1396.3625 1396.3750 1396.3875 1396.4000 1396.4125 1396.4250 1396.4375 1396.4500 1396.4625 1396.4750 1396.4875 1396.5000 1396.5125 1396.5250 1396.5375 1396.5500 1396.5625 1396.5750 1396.5875 1396.6000 1396.6125 1396.6250 1396.6375 1396.6500 1396.6625 1396.6750 1396.6875 1396.7000 1396.7125 1396.7250 1396.7375 1396.7500 1396.7625 1396.7750 138 Channel No. E107 E108 E109 E110 E111 E112 E113 E114 E115 E116 E117 E118 E119 E120 E121 E122 E123 E124 E125 E126 E127 E128 E129 E130 E131 E132 E133 E134 E135 E136 E137 E138 E139 E140 E141 E142 Transmission frequency (MHz) 1396.7875 1396.8000 1396.8125 1396.8250 1396.8375 1396.8500 1396.8625 1396.8750 1396.8875 1396.9000 1396.9125 1396.9250 1396.9375 1396.9500 1396.9625 1396.9750 1396.9875 1397.0000 1397.0125 1397.0250 1397.0375 1397.0500 1397.0625 1397.0750 1397.0875 1397.1000 1397.1125 1397.1250 1397.1375 1397.1500 1397.1625 1397.1750 1397.1875 1397.2000 1397.2125 1397.2250 Channel No. E143 E144 E145 E146 E147 E148 E149 E150 E151 E152 E153 E154 E155 E156 E157 E158 E159 E160 E161 E162 E163 E164 E165 E166 E167 E168 E169 E170 E171 E172 E173 E174 E175 E176 E177 E178 Transmission frequency (MHz) 1397.2375 1397.2500 1397.2625 1397.2750 1397.2875 1397.3000 1397.3125 1397.3250 1397.3375 1397.3500 1397.3625 1397.3750 1397.3875 1397.4000 1397.4125 1397.4250 1397.4375 1397.4500 1397.4625 1397.4750 1397.4875 1397.5000 1397.5125 1397.5250 1397.5375 1397.5500 1397.5625 1397.5750 1397.5875 1397.6000 1397.6125 1397.6250 1397.6375 1397.6500 1397.6625 1397.6750 Channel No. E179 E180 E181 E182 E183 E184 E185 E186 E187 E188 E189 E190 E191 E192 E193 E194 E195 E196 E197 E198 E199 E200 E201 E202 E203 E204 E205 E206 E207 E208 E209 E210 E211 E212 E213 E214 Operator’s Manual ZM-540PA/541PA Transmission frequency (MHz) Channel No. Transmission frequency (MHz) Channel No. Transmission frequency (MHz) Channel No. 1397.6875 1397.7000 1397.7125 1397.7250 1397.7375 1397.7500 1397.7625 1397.7750 1397.7875 1397.8000 E215 E216 E217 E218 E219 E220 E221 E222 E223 E224 1398.1375 1398.1500 1398.1625 1398.1750 1398.1875 1398.2000 1398.2125 1398.2250 1398.2375 1398.2500 E251 E252 E253 E254 E255 E256 E257 E258 E259 E260 1398.5875 1398.6000 1398.6125 1398.6250 1398.6375 1398.6500 1398.6625 1398.6750 1398.6875 1398.7000 E287 E288 E289 E290 E291 E292 E293 E294 E295 E296 1397.8125 1397.8250 1397.8375 1397.8500 1397.8625 1397.8750 1397.8875 1397.9000 1397.9125 1397.9250 1397.9375 1397.9500 1397.9625 1397.9750 1397.9875 1398.0000 1398.0125 1398.0250 1398.0375 1398.0500 1398.0625 1398.0750 1398.0875 1398.1000 1398.1125 1398.1250 E225 E226 E227 E228 E229 E230 E231 E232 E233 E234 E235 E236 E237 E238 E239 E240 E241 E242 E243 E244 E245 E246 E247 E248 E249 E250 1398.2625 1398.2750 1398.2875 1398.3000 1398.3125 1398.3250 1398.3375 1398.3500 1398.3625 1398.3750 1398.3875 1398.4000 1398.4125 1398.4250 1398.4375 1398.4500 1398.4625 1398.4750 1398.4875 1398.5000 1398.5125 1398.5250 1398.5375 1398.5500 1398.5625 1398.5750 E261 E262 E263 E264 E265 E266 E267 E268 E269 E270 E271 E272 E273 E274 E275 E276 E277 E278 E279 E280 E281 E282 E283 E284 E285 E286 1398.7125 1398.7250 1398.7375 1398.7500 1398.7625 1398.7750 1398.7875 1398.8000 1398.8125 1398.8250 1398.8375 1398.8500 1398.8625 1398.8750 1398.8875 1398.9000 1398.9125 1398.9250 1398.9375 1398.9500 1398.9625 1398.9750 1398.9875 1399.0000 1399.0125 1399.0250 E297 E298 E299 E300 E301 E302 E303 E304 E305 E306 E307 E308 E309 E310 E311 E312 E313 E314 E315 E316 E317 E318 E319 E320 E321 E322 Operator’s Manual ZM-540PA/541PA 139 Transmission frequency (MHz) 1399.0375 1399.0500 1399.0625 1399.0750 1399.0875 1399.1000 1399.1125 1399.1250 1399.1375 1399.1500 1399.1625 1399.1750 1399.1875 1399.2000 1399.2125 1399.2250 1399.2375 1399.2500 1399.2625 1399.2750 1399.2875 1399.3000 1399.3125 1399.3250 1399.3375 1399.3500 1399.3625 1399.3750 1399.3875 1399.4000 1399.4125 1399.4250 1399.4375 1399.4500 140 Channel No. E323 E324 E325 E326 E327 E328 E329 E330 E331 E332 E333 E334 E335 E336 E337 E338 E339 E340 E341 E342 E343 E344 E345 E346 E347 E348 E349 E350 E351 E352 E353 E354 E355 E356 Transmission frequency (MHz) 1399.4625 1399.4750 1399.4875 1399.5000 1399.5125 1399.5250 1399.5375 1399.5500 1399.5625 1399.5750 1399.5875 1399.6000 1399.6125 1399.6250 1399.6375 1399.6500 1399.6625 1399.6750 1399.6875 1399.7000 1399.7125 1399.7250 1399.7375 1399.7500 1399.7625 1399.7750 1399.7875 1399.8000 1399.8125 1399.8250 1399.8375 1399.8500 1399.8625 1399.8750 Channel No. E357 E358 E359 E360 E361 E362 E363 E364 E365 E366 E367 E368 E369 E370 E371 E372 E373 E374 E375 E376 E377 E378 E379 E380 E381 E382 E383 E384 E385 E386 E387 E388 E389 E390 Transmission frequency (MHz) 1399.8875 1399.9000 1399.9125 1399.9250 1399.9375 1399.9500 1399.9625 1399.9750 Channel No. E391 E392 E393 E394 E395 E396 E397 E398 Operator’s Manual ZM-540PA/541PA Channel: E502 to E898 Transmission frequency (MHz) 1427.0250 1427.0375 1427.0500 1427.0625 1427.0750 1427.0875 1427.1000 1427.1125 1427.1250 1427.1375 1427.1500 1427.1625 1427.1750 1427.1875 1427.2000 1427.2125 1427.2250 1427.2375 1427.2500 1427.2625 1427.2750 1427.2875 1427.3000 1427.3125 1427.3250 1427.3375 1427.3500 1427.3625 1427.3750 1427.3875 1427.4000 1427.4125 1427.4250 1427.4375 E502 E503 E504 E505 E506 E507 E508 E509 E510 E511 E512 E513 E514 E515 E516 E517 E518 E519 E520 E521 E522 E523 E524 E525 E526 E527 E528 E529 E530 E531 E532 E533 E534 Transmission frequency (MHz) 1427.4500 1427.4625 1427.4750 1427.4875 1427.5000 1427.5125 1427.5250 1427.5375 1427.5500 1427.5625 1427.5750 1427.5875 1427.6000 1427.6125 1427.6250 1427.6375 1427.6500 1427.6625 1427.6750 1427.6875 1427.7000 1427.7125 1427.7250 1427.7375 1427.7500 1427.7625 1427.7750 1427.7875 1427.8000 1427.8125 1427.8250 1427.8375 1427.8500 E535 1427.8625 Channel No. Operator’s Manual ZM-540PA/541PA E536 E537 E538 E539 E540 E541 E542 E543 E544 E545 E546 E547 E548 E549 E550 E551 E552 E553 E554 E555 E556 E557 E558 E559 E560 E561 E562 E563 E564 E565 E566 E567 E568 Transmission frequency (MHz) 1427.8750 1427.8875 1427.9000 1427.9125 1427.9250 1427.9375 1427.9500 1427.9625 1427.9750 1427.9875 1428.0000 1428.0125 1428.0250 1428.0375 1428.0500 1428.0625 1428.0750 1428.0875 1428.1000 1428.1125 1428.1250 1428.1375 1428.1500 1428.1625 1428.1750 1428.1875 1428.2000 1428.2125 1428.2250 1428.2375 1428.2500 1428.2625 1428.2750 E569 1428.2875 Channel No. Channel No. E570 E571 E572 E573 E574 E575 E576 E577 E578 E579 E580 E581 E582 E583 E584 E585 E586 E587 E588 E589 E590 E591 E592 E593 E594 E595 E596 E597 E598 E599 E600 E601 E602 E603 141 Transmission frequency (MHz) 1428.3000 1428.3125 1428.3250 1428.3375 1428.3500 1428.3625 1428.3750 1428.3875 1428.4000 1428.4125 1428.4250 1428.4375 1428.4500 1428.4625 1428.4750 1428.4875 1428.5000 1428.5125 1428.5250 1428.5375 1428.5500 1428.5625 1428.5750 1428.5875 1428.6000 1428.6125 1428.6250 1428.6375 1428.6500 1428.6625 1428.6750 1428.6875 1428.7000 1428.7125 1428.7250 1428.7375 142 Channel No. E604 E605 E606 E607 E608 E609 E610 E611 E612 E613 E614 E615 E616 E617 E618 E619 E620 E621 E622 E623 E624 E625 E626 E627 E628 E629 E630 E631 E632 E633 E634 E635 E636 E637 E638 E639 Transmission frequency (MHz) 1428.7500 1428.7625 1428.7750 1428.7875 1428.8000 1428.8125 1428.8250 1428.8375 1428.8500 1428.8625 1428.8750 1428.8875 1428.9000 1428.9125 1428.9250 1428.9375 1428.9500 1428.9625 1428.9750 1428.9875 1429.0000 1429.0125 1429.0250 1429.0375 1429.0500 1429.0625 1429.0750 1429.0875 1429.1000 1429.1125 1429.1250 1429.1375 1429.1500 1429.1625 1429.1750 1429.1875 Channel No. E640 E641 E642 E643 E644 E645 E646 E647 E648 E649 E650 E651 E652 E653 E654 E655 E656 E657 E658 E659 E660 E661 E662 E663 E664 E665 E666 E667 E668 E669 E670 E671 E672 E673 E674 E675 Transmission frequency (MHz) 1429.2000 1429.2125 1429.2250 1429.2375 1429.2500 1429.2625 1429.2750 1429.2875 1429.3000 1429.3125 1429.3250 1429.3375 1429.3500 1429.3625 1429.3750 1429.3875 1429.4000 1429.4125 1429.4250 1429.4375 1429.4500 1429.4625 1429.4750 1429.4875 1429.5000 1429.5125 1429.5250 1429.5375 1429.5500 1429.5625 1429.5750 1429.5875 1429.6000 1429.6125 1429.6250 1429.6375 Channel No. E676 E677 E678 E679 E680 E681 E682 E683 E684 E685 E686 E687 E688 E689 E690 E691 E692 E693 E694 E695 E696 E697 E698 E699 E700 E701 E702 E703 E704 E705 E706 E707 E708 E709 E710 E711 Operator’s Manual ZM-540PA/541PA Transmission frequency (MHz) 1429.6500 1429.6625 1429.6750 1429.6875 1429.7000 1429.7125 1429.7250 1429.7375 1429.7500 1429.7625 1429.7750 1429.7875 1429.8000 1429.8125 1429.8250 1429.8375 1429.8500 1429.8625 1429.8750 1429.8875 1429.9000 1429.9125 1429.9250 1429.9375 1429.9500 1429.9625 1429.9750 1429.9875 1430.0000 1430.0125 1430.0250 1430.0375 1430.0500 1430.0625 1430.0750 1430.0875 Channel No. E712 E713 E714 E715 E716 E717 E718 E719 E720 E721 E722 E723 E724 E725 E726 E727 E728 E729 E730 E731 E732 E733 E734 E735 E736 E737 E738 E739 E740 E741 E742 E743 E744 E745 E746 E747 Transmission frequency (MHz) 1430.1000 1430.1125 1430.1250 1430.1375 1430.1500 1430.1625 1430.1750 1430.1875 1430.2000 1430.2125 1430.2250 1430.2375 1430.2500 1430.2625 1430.2750 1430.2875 1430.3000 1430.3125 1430.3250 1430.3375 1430.3500 1430.3625 1430.3750 1430.3875 1430.4000 1430.4125 1430.4250 1430.4375 1430.4500 1430.4625 1430.4750 1430.4875 1430.5000 1430.5125 1430.5250 1430.5375 Operator’s Manual ZM-540PA/541PA Channel No. E748 E749 E750 E751 E752 E753 E754 E755 E756 E757 E758 E759 E760 E761 E762 E763 E764 E765 E766 E767 E768 E769 E770 E771 E772 E773 E774 E775 E776 E777 E778 E779 E780 E781 E782 E783 Transmission frequency (MHz) 1430.5500 1430.5625 1430.5750 1430.5875 1430.6000 1430.6125 1430.6250 1430.6375 1430.6500 1430.6625 1430.6750 1430.6875 1430.7000 1430.7125 1430.7250 1430.7375 1430.7500 1430.7625 1430.7750 1430.7875 1430.8000 1430.8125 1430.8250 1430.8375 1430.8500 1430.8625 1430.8750 1430.8875 1430.9000 1430.9125 1430.9250 1430.9375 1430.9500 1430.9625 1430.9750 1430.9875 Channel No. E784 E785 E786 E787 E788 E789 E790 E791 E792 E793 E794 E795 E796 E797 E798 E799 E800 E801 E802 E803 E804 E805 E806 E807 E808 E809 E810 E811 E812 E813 E814 E815 E816 E817 E818 E819 143 Transmission frequency (MHz) 1431.0000 1431.0125 1431.0250 1431.0375 1431.0500 1431.0625 1431.0750 1431.0875 1431.1000 1431.1125 1431.1250 1431.1375 1431.1500 1431.1625 1431.1750 1431.1875 1431.2000 1431.2125 1431.2250 1431.2375 1431.2500 1431.2625 1431.2750 1431.2875 1431.3000 1431.3125 1431.3250 1431.3375 1431.3500 1431.3625 1431.3750 1431.3875 1431.4000 1431.4125 1431.4250 144 Channel No. E820 E821 E822 E823 E824 E825 E826 E827 E828 E829 E830 E831 E832 E833 E834 E835 E836 E837 E838 E839 E840 E841 E842 E843 E844 E845 E846 E847 E848 E849 E850 E851 E852 E853 E854 Transmission frequency (MHz) 1431.4375 1431.4500 1431.4625 1431.4750 1431.4875 1431.5000 1431.5125 1431.5250 1431.5375 1431.5500 1431.5625 1431.5750 1431.5875 1431.6000 1431.6125 1431.6250 1431.6375 1431.6500 1431.6625 1431.6750 1431.6875 1431.7000 1431.7125 1431.7250 1431.7375 1431.7500 1431.7625 1431.7750 1431.7875 1431.8000 1431.8125 1431.8250 1431.8375 1431.8500 1431.8625 Channel No. E855 E856 E857 E858 E859 E860 E861 E862 E863 E864 E865 E866 E867 E868 E869 E870 E871 E872 E873 E874 E875 E876 E877 E878 E879 E880 E881 E882 E883 E884 E885 E886 E887 E888 E889 Transmission frequency (MHz) 1431.8750 1431.8875 1431.9000 1431.9125 1431.9250 1431.9375 1431.9500 1431.9625 1431.9750 Channel No. E890 E891 E892 E893 E894 E895 E896 E897 E898 Operator’s Manual ZM-540PA/541PA
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.4 Linearized : No XMP Toolkit : Adobe XMP Core 4.0-c316 44.253921, Sun Oct 01 2006 17:14:39 Instance ID : uuid:e102f189-7017-4b76-82dc-7ef7c48643dc Document ID : adobe:docid:indd:0be4451c-499f-11de-b1a1-b5ef563dca21 Rendition Class : proof:pdf Derived From Instance ID : 0be4451b-499f-11de-b1a1-b5ef563dca21 Derived From Document ID : adobe:docid:indd:0be4451a-499f-11de-b1a1-b5ef563dca21 Manifest Link Form : ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream Manifest Placed X Resolution : 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00 Manifest Placed Y Resolution : 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00 Manifest Placed Resolution Unit : Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches Manifest Reference Instance ID : uuid:b77c2a50-4a1c-49e8-870b-bafb74e85fd2, uuid:b77c2a50-4a1c-49e8-870b-bafb74e85fd2, uuid:e88451f9-d989-48e5-8d4b-af774243178a, uuid:7a9a7801-103a-4f57-8ef9-f91f7ef190c0, uuid:f4927192-0845-488b-bcff-f7397508fa23, uuid:01adecfd-286d-46ab-afce-5e0bba0d0620, uuid:71c31040-d088-484c-b775-592dcf56617b, uuid:71c31040-d088-484c-b775-592dcf56617b, uuid:ef8babeb-0117-4dc8-b519-225ef6421da2, uuid:ef8babeb-0117-4dc8-b519-225ef6421da2, uuid:2a7d0c18-7780-40c3-8756-562fe0301ea1, uuid:dc7bc059-3ee9-4a76-aa45-a745b2812f0a, uuid:90619dff-4e12-4d29-ad12-e00a7d8b8047, uuid:cde80b63-5647-4972-a7c1-f5d19350de8d, uuid:2a7d0c18-7780-40c3-8756-562fe0301ea1, uuid:e1988716-be11-4934-8e61-3cb9f4b08061, uuid:e1988716-be11-4934-8e61-3cb9f4b08061, uuid:0ADDB70EE448DE11851BE0B103435EBC, uuid:f081a1b8-a235-4f6b-920d-f9baceaa0cd9, uuid:25F8CC709449DE1194E4E1F5385B0D25, uuid:3A2ACE77CF48DE11851BE0B103435EBC, uuid:3A2ACE77CF48DE11851BE0B103435EBC, uuid:9C3458254843DE118C08FB0039C43290, uuid:A53458254843DE118C08FB0039C43290, uuid:06DDB70EE448DE11851BE0B103435EBC, uuid:A53458254843DE118C08FB0039C43290, uuid:967FC0B64843DE118C08FB0039C43290, uuid:7FBCC0C94943DE118C08FB0039C43290, uuid:88BCC0C94943DE118C08FB0039C43290, uuid:9C3458254843DE118C08FB0039C43290, uuid:E1BFB77C4B43DE118C08FB0039C43290, uuid:41DF6E62E448DE11851BE0B103435EBC, uuid:44DF6E62E448DE11851BE0B103435EBC, uuid:BE33D2224D43DE118C08FB0039C43290, uuid:9C3458254843DE118C08FB0039C43290, uuid:C233D2224D43DE118C08FB0039C43290, uuid:C633D2224D43DE118C08FB0039C43290, uuid:0fe5f344-9997-45d9-b4fd-a3be94671f14, uuid:bad16a36-b567-437a-b488-6b939bd195d4, uuid:48ce1f7d-05b9-4780-a88a-f2de0647f132, uuid:074a781f-f323-4942-9281-32f0f79f6f9d, uuid:f33f7008-c1a6-42b9-a864-84b77ea36c08, uuid:5fedc2a8-2307-4505-b46d-95d98e9dcd04, uuid:49c02f07-0863-4747-beca-6a2a4ab97440, uuid:577de756-59ec-4598-8b00-158e57afd233, uuid:3fe32c3c-7ffe-4b2b-b3b2-fb06ea33a59e, uuid:3fe32c3c-7ffe-4b2b-b3b2-fb06ea33a59e, uuid:36b324f5-7bf6-4ba1-b946-44205de9662b, uuid:48603d8d-2502-4db5-ae95-e81d0ee1561d, uuid:89919db1-801f-4cd7-a9a7-9cb31ecf7075, uuid:89919db1-801f-4cd7-a9a7-9cb31ecf7075, uuid:7441dbca-2539-4dae-b5ff-c8d2382a851f, uuid:7441dbca-2539-4dae-b5ff-c8d2382a851f, uuid:550f8dfe-81be-4e3f-8f60-565038f0ccd7, uuid:550f8dfe-81be-4e3f-8f60-565038f0ccd7, uuid:ffa85218-128c-45d8-ba5e-1cb8b1c04097, uuid:ffa85218-128c-45d8-ba5e-1cb8b1c04097, uuid:f68d18d0-b555-4b86-9740-86130437c3d2, uuid:020a20f7-77c7-4bc2-b41d-47b9e8c15159, uuid:f68d18d0-b555-4b86-9740-86130437c3d2, uuid:020a20f7-77c7-4bc2-b41d-47b9e8c15159, uuid:0e49f1c4-00b5-42d5-871b-544d5e7745ae, uuid:e5cad8b4-d087-49a7-8800-4d3bf9633064, uuid:18916a9d-78b0-4688-a104-b721d9cc478a, uuid:de780b3e-fca1-4ce3-8c54-5a21e53e80e8, uuid:eb3ce270-0176-407f-8d52-e18174cfee5c, uuid:1b992720-8932-4e9a-a951-eec0eaf40f3b, uuid:dabba8ab-5e83-47b5-8f96-c3a975593dc0, uuid:d06163cc-4243-4c23-9871-4b088a4cea24, uuid:ec6aef5d-f2e3-4407-9357-f9267405726d, uuid:7aef1364-bbd0-48d4-becd-0fc1e1381595, uuid:5bda642e-99be-4fae-aa5d-65ddceaf6958, uuid:f2344d5b-7d92-4d41-ad52-098201488766, uuid:3ab9cdb5-2e9d-4148-a7a2-36bc9d393a88, uuid:db4122d3-7de5-470f-bc7b-1bc478e842d5, uuid:bdfd58d9-68bc-4333-8619-714dc67b65b2, uuid:86fcbfe3-c420-4a86-850f-c58c94bac675, uuid:d8024c41-3a9b-4848-84ea-bcb3a9adbbdf, uuid:e6862862-edb5-4056-a62a-ee442f6d93bc, uuid:fa0c13fc-7dba-4d60-93e4-56fc453aa5e3, uuid:73908542-533c-47c7-9433-866580378ee7, uuid:7a9a7801-103a-4f57-8ef9-f91f7ef190c0, uuid:7a9a7801-103a-4f57-8ef9-f91f7ef190c0, uuid:97a34b76-81e3-4780-be9c-62ee2949e7be, uuid:6e55aca3-ebdd-4ad0-abd0-b586ad19bb57, uuid:06d11ac2-8159-4187-92b4-a571bb8e8ec9, uuid:98367ea7-dcdc-41d6-afa7-8da2a473be7f, uuid:879b39bc-69cd-4fcf-991a-acd546e24f0f, uuid:97a34b76-81e3-4780-be9c-62ee2949e7be, uuid:6e55aca3-ebdd-4ad0-abd0-b586ad19bb57, uuid:06d11ac2-8159-4187-92b4-a571bb8e8ec9, uuid:98367ea7-dcdc-41d6-afa7-8da2a473be7f, uuid:879b39bc-69cd-4fcf-991a-acd546e24f0f, uuid:3862c2ef-4fdd-4cec-97b9-76aad10057d1, uuid:3862c2ef-4fdd-4cec-97b9-76aad10057d1, uuid:d4d1977f-72cb-439c-997a-74b03bfdd39d, uuid:d4d1977f-72cb-439c-997a-74b03bfdd39d, uuid:df355c3b-0a2f-4521-8c94-682c776f63a2, uuid:df355c3b-0a2f-4521-8c94-682c776f63a2, uuid:d88919d1-d79b-4e28-84e1-43b8876d83c1, uuid:d88919d1-d79b-4e28-84e1-43b8876d83c1, uuid:62ca9703-a362-4a46-90b4-375612514e66, uuid:62ca9703-a362-4a46-90b4-375612514e66, uuid:02ef6c0c-a8d1-40d5-a00b-fe612c3d960e, uuid:27063f93-1319-4c25-813e-b8543ea00227, uuid:f1739e35-4ede-4f54-9185-7e22e9bd37db, uuid:2baabe61-8e34-4ff9-82ec-c9b2e1ed9d81, uuid:2baabe61-8e34-4ff9-82ec-c9b2e1ed9d81, uuid:2baabe61-8e34-4ff9-82ec-c9b2e1ed9d81, uuid:2baabe61-8e34-4ff9-82ec-c9b2e1ed9d81, uuid:2baabe61-8e34-4ff9-82ec-c9b2e1ed9d81, uuid:2baabe61-8e34-4ff9-82ec-c9b2e1ed9d81, uuid:df9e97f6-0ac3-4f34-9b22-dce473b418eb, uuid:2baabe61-8e34-4ff9-82ec-c9b2e1ed9d81, uuid:2baabe61-8e34-4ff9-82ec-c9b2e1ed9d81, uuid:df9e97f6-0ac3-4f34-9b22-dce473b418eb, uuid:2baabe61-8e34-4ff9-82ec-c9b2e1ed9d81, uuid:2baabe61-8e34-4ff9-82ec-c9b2e1ed9d81, uuid:48DF6E62E448DE11851BE0B103435EBC, uuid:b77c2a50-4a1c-49e8-870b-bafb74e85fd2, uuid:25F8CC709449DE1194E4E1F5385B0D25, uuid:25F8CC709449DE1194E4E1F5385B0D25, uuid:2AF8CC709449DE1194E4E1F5385B0D25, uuid:3A2ACE77CF48DE11851BE0B103435EBC, uuid:6CEF551AE848DE11851BE0B103435EBC, uuid:470BED2F22E8DD11B250A216EA750288, uuid:2AA16DD6E848DE11851BE0B103435EBC, uuid:2EA16DD6E848DE11851BE0B103435EBC, uuid:b77c2a50-4a1c-49e8-870b-bafb74e85fd2, uuid:0C607061E948DE11851BE0B103435EBC, uuid:8DED3885EA48DE11851BE0B103435EBC, uuid:A7EECD302449DE118C329EDAD159CF80, uuid:0C607061E948DE11851BE0B103435EBC, uuid:ABEECD302449DE118C329EDAD159CF80, uuid:AFEECD302449DE118C329EDAD159CF80, uuid:1F873F322549DE118C329EDAD159CF80, uuid:2EA16DD6E848DE11851BE0B103435EBC, uuid:3A2ACE77CF48DE11851BE0B103435EBC, uuid:3A2ACE77CF48DE11851BE0B103435EBC, uuid:a31a21be-ba1d-4684-9aa5-998fb1f43fec, uuid:3166AAA29549DE1194E4E1F5385B0D25, uuid:b77c2a50-4a1c-49e8-870b-bafb74e85fd2, uuid:3166AAA29549DE1194E4E1F5385B0D25, uuid:b77c2a50-4a1c-49e8-870b-bafb74e85fd2, uuid:2B66AAA29549DE1194E4E1F5385B0D25, uuid:3A2ACE77CF48DE11851BE0B103435EBC, uuid:33EB7FF62749DE118C329EDAD159CF80, uuid:98B9D9CB2A49DE118C329EDAD159CF80, uuid:dc7bc059-3ee9-4a76-aa45-a745b2812f0a, uuid:90619dff-4e12-4d29-ad12-e00a7d8b8047, uuid:cde80b63-5647-4972-a7c1-f5d19350de8d, uuid:2a7d0c18-7780-40c3-8756-562fe0301ea1, uuid:0AC403612DE8DD11B250A216EA750288, uuid:a31a21be-ba1d-4684-9aa5-998fb1f43fec, uuid:0ADDB70EE448DE11851BE0B103435EBC, uuid:3A2ACE77CF48DE11851BE0B103435EBC, uuid:8a1cb3cf-2dd5-41f2-ab60-7e8066468555, uuid:9C3458254843DE118C08FB0039C43290, uuid:1D697F724E43DE118C08FB0039C43290, uuid:59E17026912ADE119BD09A2D75DB870E, uuid:0A067C1B3FE8DD11B250A216EA750288, uuid:3543B0BB3FE8DD11B250A216EA750288, uuid:3943B0BB3FE8DD11B250A216EA750288, uuid:0A067C1B3FE8DD11B250A216EA750288, uuid:0A067C1B3FE8DD11B250A216EA750288, uuid:0A067C1B3FE8DD11B250A216EA750288, uuid:59E17026912ADE119BD09A2D75DB870E, uuid:9C3458254843DE118C08FB0039C43290, uuid:1D697F724E43DE118C08FB0039C43290, uuid:9EB9D9CB2A49DE118C329EDAD159CF80, uuid:8a1cb3cf-2dd5-41f2-ab60-7e8066468555, uuid:8a1cb3cf-2dd5-41f2-ab60-7e8066468555, uuid:d6cf6ab5-a842-406b-9516-bcc48f1cef6c, uuid:a847bf35-1a0a-47ac-ad74-9ec3465d05c5, uuid:9C3458254843DE118C08FB0039C43290, uuid:1D697F724E43DE118C08FB0039C43290, uuid:EC9887815343DE118C08FB0039C43290, uuid:F09887815343DE118C08FB0039C43290, uuid:01e60357-b49c-4e3a-9fec-a175750e5d56, uuid:01e60357-b49c-4e3a-9fec-a175750e5d56, uuid:bdfd58d9-68bc-4333-8619-714dc67b65b2, uuid:1038ac5b-2258-4a83-855d-6591371fba56, uuid:7a9a7801-103a-4f57-8ef9-f91f7ef190c0, uuid:b4afa466-df8e-4239-ba8b-c7cf81282500 Manifest Reference Document ID : uuid:BFDEC2E54643DE11B24780B72FA1A0BD, uuid:BFDEC2E54643DE11B24780B72FA1A0BD, uuid:C53BA17E112ADE1194BDFEAF2413E22B, uuid:5A954661CD14DD11947795C8998DF36F, uuid:C73BA17E112ADE1194BDFEAF2413E22B, uuid:C93BA17E112ADE1194BDFEAF2413E22B, uuid:C1DEC2E54643DE11B24780B72FA1A0BD, uuid:C1DEC2E54643DE11B24780B72FA1A0BD, uuid:C3DEC2E54643DE11B24780B72FA1A0BD, uuid:C3DEC2E54643DE11B24780B72FA1A0BD, uuid:D172999FC648DE1194EECA9F7BD7368D, uuid:B64DD33CBE14DD11A65ABAC307FA1A18, uuid:1A7145A7E122DD118B75A1EA7098731F, uuid:1C7145A7E122DD118B75A1EA7098731F, uuid:D172999FC648DE1194EECA9F7BD7368D, uuid:C5DEC2E54643DE11B24780B72FA1A0BD, uuid:C5DEC2E54643DE11B24780B72FA1A0BD, uuid:E5C8513ACB48DE11851BE0B103435EBC, uuid:EF9866B8C42DDC11979FED628EE1FD8A, uuid:24F8CC709449DE1194E4E1F5385B0D25, uuid:721F7398CC48DE11851BE0B103435EBC, uuid:721F7398CC48DE11851BE0B103435EBC, uuid:6E23E3764643DE118C08FB0039C43290, uuid:A43458254843DE118C08FB0039C43290, uuid:285CC7D6CD48DE11851BE0B103435EBC, uuid:A43458254843DE118C08FB0039C43290, uuid:957FC0B64843DE118C08FB0039C43290, uuid:7EBCC0C94943DE118C08FB0039C43290, uuid:87BCC0C94943DE118C08FB0039C43290, uuid:6E23E3764643DE118C08FB0039C43290, uuid:E0BFB77C4B43DE118C08FB0039C43290, uuid:2C5CC7D6CD48DE11851BE0B103435EBC, uuid:7D745900CF48DE11851BE0B103435EBC, uuid:BD33D2224D43DE118C08FB0039C43290, uuid:6E23E3764643DE118C08FB0039C43290, uuid:C133D2224D43DE118C08FB0039C43290, uuid:C533D2224D43DE118C08FB0039C43290, uuid:308A1A81142ADE1194BDFEAF2413E22B, uuid:C7DEC2E54643DE11B24780B72FA1A0BD, uuid:057A21A1C114DD11A65ABAC307FA1A18, uuid:731AFE49CA14DD11947795C8998DF36F, uuid:7CBFEF024F43DE11B24780B72FA1A0BD, uuid:1929EB4E162ADE1194BDFEAF2413E22B, uuid:791AFE49CA14DD11947795C8998DF36F, uuid:866C17B7DA47DD11A748873697FBB31A, uuid:B500FA9D8420DD11BCE48BBEBD155BA1, uuid:B500FA9D8420DD11BCE48BBEBD155BA1, uuid:B300FA9D8420DD11BCE48BBEBD155BA1, uuid:BB8572208520DD11B6F1C255342FADC9, uuid:79646D33BFE6DD1181FED50D97FF8277, uuid:79646D33BFE6DD1181FED50D97FF8277, uuid:77646D33BFE6DD1181FED50D97FF8277, uuid:77646D33BFE6DD1181FED50D97FF8277, uuid:7B646D33BFE6DD1181FED50D97FF8277, uuid:7B646D33BFE6DD1181FED50D97FF8277, uuid:7D646D33BFE6DD1181FED50D97FF8277, uuid:7D646D33BFE6DD1181FED50D97FF8277, uuid:828BBFBFC314DD11A65ABAC307FA1A18, uuid:848BBFBFC314DD11A65ABAC307FA1A18, uuid:828BBFBFC314DD11A65ABAC307FA1A18, uuid:848BBFBFC314DD11A65ABAC307FA1A18, uuid:E3495543C914DD11947795C8998DF36F, uuid:E5495543C914DD11947795C8998DF36F, uuid:886B77A80FEFDC119595EDD5F250AAE8, uuid:E7495543C914DD11947795C8998DF36F, uuid:E9495543C914DD11947795C8998DF36F, uuid:EB495543C914DD11947795C8998DF36F, uuid:1E9E6B85C814DD11947795C8998DF36F, uuid:209E6B85C814DD11947795C8998DF36F, uuid:03315393C914DD11947795C8998DF36F, uuid:05315393C914DD11947795C8998DF36F, uuid:07315393C914DD11947795C8998DF36F, uuid:09315393C914DD11947795C8998DF36F, uuid:7EBFEF024F43DE11B24780B72FA1A0BD, uuid:1B29EB4E162ADE1194BDFEAF2413E22B, uuid:A0625022090EDE119180C840008F9A82, uuid:9A3E6B7BD31BDD11A3A692EDA8AF9917, uuid:426E9D68D61BDD118DDBEE8D9B52EB4D, uuid:8C6D3F2BCE14DD11947795C8998DF36F, uuid:D372999FC648DE1194EECA9F7BD7368D, uuid:A2625022090EDE119180C840008F9A82, uuid:5A954661CD14DD11947795C8998DF36F, uuid:5A954661CD14DD11947795C8998DF36F, uuid:80BFEF024F43DE11B24780B72FA1A0BD, uuid:A8625022090EDE119180C840008F9A82, uuid:82BFEF024F43DE11B24780B72FA1A0BD, uuid:84BFEF024F43DE11B24780B72FA1A0BD, uuid:75D97D7C0A0EDE119180C840008F9A82, uuid:80BFEF024F43DE11B24780B72FA1A0BD, uuid:A8625022090EDE119180C840008F9A82, uuid:82BFEF024F43DE11B24780B72FA1A0BD, uuid:84BFEF024F43DE11B24780B72FA1A0BD, uuid:75D97D7C0A0EDE119180C840008F9A82, uuid:86BFEF024F43DE11B24780B72FA1A0BD, uuid:86BFEF024F43DE11B24780B72FA1A0BD, uuid:449C5333C214DD11A65ABAC307FA1A18, uuid:449C5333C214DD11A65ABAC307FA1A18, uuid:77D97D7C0A0EDE119180C840008F9A82, uuid:77D97D7C0A0EDE119180C840008F9A82, uuid:79D97D7C0A0EDE119180C840008F9A82, uuid:79D97D7C0A0EDE119180C840008F9A82, uuid:7BD97D7C0A0EDE119180C840008F9A82, uuid:7BD97D7C0A0EDE119180C840008F9A82, uuid:C0344A92CA14DD11947795C8998DF36F, uuid:57086D2F0C0EDE119180C840008F9A82, uuid:AB5BBDC8CC40DD119114C7C22FF0D170, uuid:017A21A1C114DD11A65ABAC307FA1A18, uuid:017A21A1C114DD11A65ABAC307FA1A18, uuid:017A21A1C114DD11A65ABAC307FA1A18, uuid:017A21A1C114DD11A65ABAC307FA1A18, uuid:017A21A1C114DD11A65ABAC307FA1A18, uuid:017A21A1C114DD11A65ABAC307FA1A18, uuid:45B287A9C71FDD1194EBB62A2B9F6EA5, uuid:017A21A1C114DD11A65ABAC307FA1A18, uuid:017A21A1C114DD11A65ABAC307FA1A18, uuid:45B287A9C71FDD1194EBB62A2B9F6EA5, uuid:017A21A1C114DD11A65ABAC307FA1A18, uuid:017A21A1C114DD11A65ABAC307FA1A18, uuid:47DF6E62E448DE11851BE0B103435EBC, uuid:BFDEC2E54643DE11B24780B72FA1A0BD, uuid:24F8CC709449DE1194E4E1F5385B0D25, uuid:24F8CC709449DE1194E4E1F5385B0D25, uuid:29F8CC709449DE1194E4E1F5385B0D25, uuid:721F7398CC48DE11851BE0B103435EBC, uuid:9FAFCE7DE548DE11851BE0B103435EBC, uuid:420BED2F22E8DD11B250A216EA750288, uuid:70EF551AE848DE11851BE0B103435EBC, uuid:2DA16DD6E848DE11851BE0B103435EBC, uuid:BFDEC2E54643DE11B24780B72FA1A0BD, uuid:0B607061E948DE11851BE0B103435EBC, uuid:8CED3885EA48DE11851BE0B103435EBC, uuid:E32640DF2349DE118C329EDAD159CF80, uuid:0B607061E948DE11851BE0B103435EBC, uuid:AAEECD302449DE118C329EDAD159CF80, uuid:AEEECD302449DE118C329EDAD159CF80, uuid:1E873F322549DE118C329EDAD159CF80, uuid:2DA16DD6E848DE11851BE0B103435EBC, uuid:721F7398CC48DE11851BE0B103435EBC, uuid:721F7398CC48DE11851BE0B103435EBC, uuid:2C5AD82CC814DD11947795C8998DF36F, uuid:3066AAA29549DE1194E4E1F5385B0D25, uuid:BFDEC2E54643DE11B24780B72FA1A0BD, uuid:3066AAA29549DE1194E4E1F5385B0D25, uuid:BFDEC2E54643DE11B24780B72FA1A0BD, uuid:2CEB7FF62749DE118C329EDAD159CF80, uuid:721F7398CC48DE11851BE0B103435EBC, uuid:32EB7FF62749DE118C329EDAD159CF80, uuid:97B9D9CB2A49DE118C329EDAD159CF80, uuid:B64DD33CBE14DD11A65ABAC307FA1A18, uuid:1A7145A7E122DD118B75A1EA7098731F, uuid:1C7145A7E122DD118B75A1EA7098731F, uuid:D172999FC648DE1194EECA9F7BD7368D, uuid:09C403612DE8DD11B250A216EA750288, uuid:2C5AD82CC814DD11947795C8998DF36F, uuid:E5C8513ACB48DE11851BE0B103435EBC, uuid:721F7398CC48DE11851BE0B103435EBC, uuid:C99109975043DE11B24780B72FA1A0BD, uuid:6E23E3764643DE118C08FB0039C43290, uuid:1C697F724E43DE118C08FB0039C43290, uuid:58E17026912ADE119BD09A2D75DB870E, uuid:09067C1B3FE8DD11B250A216EA750288, uuid:0F067C1B3FE8DD11B250A216EA750288, uuid:3843B0BB3FE8DD11B250A216EA750288, uuid:09067C1B3FE8DD11B250A216EA750288, uuid:09067C1B3FE8DD11B250A216EA750288, uuid:09067C1B3FE8DD11B250A216EA750288, uuid:58E17026912ADE119BD09A2D75DB870E, uuid:6E23E3764643DE118C08FB0039C43290, uuid:1C697F724E43DE118C08FB0039C43290, uuid:9DB9D9CB2A49DE118C329EDAD159CF80, uuid:C99109975043DE11B24780B72FA1A0BD, uuid:C99109975043DE11B24780B72FA1A0BD, uuid:CB9109975043DE11B24780B72FA1A0BD, uuid:CD9109975043DE11B24780B72FA1A0BD, uuid:6E23E3764643DE118C08FB0039C43290, uuid:1C697F724E43DE118C08FB0039C43290, uuid:24697F724E43DE118C08FB0039C43290, uuid:EF9887815343DE118C08FB0039C43290, uuid:2D01A37DDF6FDC118BF9A43F8C0B68AC, uuid:2D01A37DDF6FDC118BF9A43F8C0B68AC, uuid:A0625022090EDE119180C840008F9A82, uuid:CF9109975043DE11B24780B72FA1A0BD, uuid:5A954661CD14DD11947795C8998DF36F, uuid:BC344A92CA14DD11947795C8998DF36F Create Date : 2009:06:19 17:16+09:00 Modify Date : 2009:06:19 17:17:06+09:00 Metadata Date : 2009:06:19 17:17:06+09:00 Creator Tool : Adobe InDesign CS2_J (4.0.5) Thumbnail Format : JPEG Thumbnail Width : 256 Thumbnail Height : 256 Thumbnail Image : (Binary data 9211 bytes, use -b option to extract) Format : application/pdf Title : OM_ZM-540_541PA.pdf Creator : Mika Producer : Acrobat Distiller 8.1.0 (Windows) Trapped : False Page Count : 154 Author : MikaEXIF Metadata provided by EXIF.tools