Rayence RY1417WHD Medical Image Processing Unit User Manual 1417WCC WGC

Rayence Co., Ltd. Medical Image Processing Unit 1417WCC WGC

User Manual

Download: Rayence RY1417WHD Medical Image Processing Unit User Manual 1417WCC   WGC
Mirror Download [FCC.gov]Rayence RY1417WHD Medical Image Processing Unit User Manual 1417WCC   WGC
Document ID3524108
Application ID4vhq/fB05OgmlMeNhXswdg==
Document DescriptionUser Manual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize299.39kB (3742316 bits)
Date Submitted2017-08-23 00:00:00
Date Available2017-08-23 00:00:00
Creation Date2017-08-23 13:56:20
Producing Software3-Heights(TM) PDF Optimization Shell 4.8.25.2 (http://www.pdf-tools.com)
Document Lastmod2017-08-23 05:02:36
Document Title1417WCC / WGC
Document Author: Administrator

1417WGC/WCC
User & Installation Manual
R-USM-020/021
Date: 2017-04-19
Preface
Preface
Please note that this information is for proper use and safety of the equipment. The following
symbols may indicate a hazardous situation in which, if not heeded, may result in serious injury
or even death to the user or others, or damage to the equipment.
Used to emphasize essential information.
Be sure to read this information to avoid incorrect operation.
Indicates warning and safety instructions. If not adhered to, it could result in death or
serious injury to the user or others.
WARNING
Indicates a hazardous situation which, if not heeded, may result in minor or moderate
injury to the user or others, or damage to the equipment.
CAUTION
For users in the United States:

United State federal law restricts this equipment to be used by or on the order of a
physician.

Since the X-ray exposure condition can be changed depending on the age, gender and
bone density of the patient, in case of Pediatric, X-ray exposure condition can be changed
by expert’s judge. For further information, please refer to FDA Pediatric X-ray Imaging
webpage.
http://www.fda.gov/radiationemittingproducts/radiationemittingproductsandprocedures/medicalimaging/ucm298899.htm
For users in other countries:

This equipment is to be used by or on the order of a licensed person under the related laws
for each country.
Intended use:
Digital Flat Panel X-Ray Detector is indicated for digital imaging solution designed for general
radiographic system for human anatomy. It is intended to replace film or screen based
radiographic systems in all general purpose diagnostic procedures. Not to be used for
mammography.
VATECH Dental Manufacturing Ltd.
Chancery House, St. Nicholas Way, Sutton, SM1 1JB, United Kingdom
Tel : +44 208 652 1990, Fax : +44 208 652 1909
Rayence Co., Ltd.
14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea
www.rayence.com
1417WGC/WCC
Contents
Contents
PART I.
User & Installation Manual
1.
2.
3.
4.
5.
6.
Safety Information ......................................................................................... 6
1.1
Safety Standard .................................................................................... 6
1.2
Symbols ..............................................................................................12
1.3
Warning ..............................................................................................14
1.4
Caution ...............................................................................................16
1.5
Safety Information ...............................................................................20
1.6
Label and Location of Attachment .........................................................22
1.7
Summary of usability specifications .......................................................22
Product Introduction and Specification ........................................................... 23
2.1
Product Features ..................................................................................23
2.2
Product Components ............................................................................24
2.3
Part Names and Functions ....................................................................24
2.4
Part Specifications ................................................................................32
2.5
Environmental Requirements ................................................................36
2.6
Dimensions (Unit: mm) ........................................................................37
Installation and Calibration ............................................................................ 42
3.1
Installation ..........................................................................................42
3.2
Calibration ...........................................................................................91
Usage ........................................................................................................... 97
4.1
Set Up .................................................................................................97
4.2
Image Acquisition ................................................................................99
4.3
View Images ...................................................................................... 101
4.4
Additional Function............................................................................. 103
Maintenance ............................................................................................... 114
5.1
Cleaning ............................................................................................ 114
5.2
Inspection ......................................................................................... 114
5.3
Replaceable Parts and Instruction of Replacement ............................... 115
5.4
Disposal or Recycling .......................................................................... 116
Warranty .................................................................................................... 117
6.1
Warranty ........................................................................................... 117
1417WGC/WCC
PART II. Service Manual
PART I. User & Installation Manual
1417WGC/WCC
1.
Safety Information………….…………………….8
2.
Product Introduction and Specification……25
3.
Installation and Calibration…………………..44
4.
Usage……………………………...………………..99
5.
Maintenance………………………...…………..116
6.
Warranty………………………..…………..……119
PART II. Service Manual
1.
Safety Information
1.1
Safety Standard
1.1.1
Medical Device Classification
Item
1.1.2
Description
Classification by protection type against Electric
Shock
Class I or Internally Powered Equipment
Classification according to the degree of
protection against ingress of water
IP53
Mode of operation
Continuous Operation
Environment of Use
This equipment is not suitable for use in the
presence of flammable anesthetic or oxygen.
Regulations
1. Safety and Electromagnetic Compatibility Information
Item
Description
IEC/EN/UL 60601-1
Medical electrical equipment
Part 1: General requirements for safety
IEC/EN 60601-1-2
Medical electrical equipment
Part 2: Electromagnetic compatibility-requirements and tests
This equipment has been tested and found to comply with the limits for medical devices in IEC
60601-1-2. These limits are designed to provide reasonable protection against harmful
interference in a typical medical installation.
This equipment generates, uses and can radiate radio frequency energy. If not installed and
used in accordance with the instructions, it may cause harmful interference to other devices in
the vicinity. However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to other devices, which can be
determined by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures.

Reorient or relocate the equipment.

Increase the separation between the equipment.

Connect the equipment into an outlet on a circuit different from that to which the other
devices are connected.

Contact Rayence Customer Service team or authorized agent for help.
1417WGC/WCC
PART II. Service Manual
2. Radio Frequency compliance
FCC (For USA)
•
FCC ID: QIIRY1417WHD
•
5.15- 5.25 GHz band is restricted to indoor operations only.
•
Host device of the approved module shall be marked with the following item:
•
Compliance with FCC requirement 15.407(c)
Data transmission is always initiated by software, which is the passed down through the
MAC, through the digital and analog baseband, and finally to the RF chip. Several special
packets are initiated by the MAC. These are the only ways the digital baseband portion will
turn on the RF transmitter, which it then turns off at the end of the packet. Therefore, the
transmitter will be on only while one of the aforementioned packets is being transmitted. In
other words, this device automatically discontinues transmission in case of either absence
of information to transmit or operational failure.
This device complies with part 15 of the FCC Rules. Operation is subject to the following
two conditions: (1) This device may not cause harmful interference, and (2) this device
must accept any interference received, including interference that may cause undesired
operation.

FCC CAUTION
Changes or modifications not expressly approved by the party responsible for compliance
could void the user’s authority to operate the equipment. This transmitter must not be colocated or operated in conjunction with any other antenna or transmitter.
When installing it in a mobile equipment
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled
environment and meets the FCC radio frequency (RF) Exposure Guidelines in CFR
§2.1093.
When you use the detector with wire mode, the wireless function is automatically off.
•
5150-5250 MHz band is restricted to indoor operations only.
< Note >

The front with touch configuration was only tested since only the front is touched to human
body in normal operation of this device

The highest reported SAR values for head & body are: 1.18 W/kg & 1.06 W/kg respectively.
1417WGC/WCC
PART II. Service Manual
IC Notice (For CANADA)

IC: 10742A-1417WHD
This Class A digital apparatus complies with Canadian ICES-003
This device complies with Industry Canada licence-exempt RSS standard(s). Operation is
subject to the following two conditions: (1) this device may not cause interference, and (2)
this device must accept any interference, including interference that may cause undesired
operation of the device.
Le present appareil est conforme aux CNR d'Industrie Canada applicables aux appareils
radio exempts de licence. L'exploitation est autorisee aux deux conditions suivantes: (1)
l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur de l'appareil doit accepter
tout brouillage radioelectrique subi, meme si le brouillage est susceptible d'en
compromettre le fonctionnement.
R&TTE Notice (European Union)
The product compliance to the Directive 1999/5/EC as radio equipment and telecommunications
terminal equipment.
Authorized by Notified Body (code no. : 0678)

5150-5250 MHz band is restricted to indoor operations only.
Japan Ratio Raw (For Japan)

Type Certification No. : 011-170018
3. Electro-Magnetic Compatibility Information
Electro-Magnetic Emissions
This 1417WCC/WGC is intended for use in the electromagnetic environment specified below.
The customer or the user of the 1417WCC/WGC should assure that it is used in such an
environment.
Immunity Test
Compliance
Electromagnetic Environment – Guidance
RF Emissions CISPR 11
Group 1
The 1417WCC/WGC uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF Emissions CISPR 11
Class A
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations / Flicker
emissions IEC 61000-3-3
1417WGC/WCC
Complies
The 1417WCC/WGC is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes. provided the following warning is
heeded:
Warning: This equitpment/system is intended for use
by healthcare professionals only, This
equipment/system may cause radio interference or
may disrupt the operation of nearby equipment. It
may be necessary to take mitigation measures, such
as re-orenting or relocating the 1717WCC/WGC or
shielding the location.
PART II. Service Manual
Electro-Magnetic Immunity
This 1417WCC/WGC is intended for use in the electromagnetic environment specified below.
The customer or the user of the 1417WCC/WGC should assure that it is used in such an
environment.
Immunity Test
IEC 60601-1-2 Test Level
Compliance Level
Electromagnetic environment
– guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
IEC 60601-1-2 Test
level
Floors should be wood,
concrete or ceramic tiles. If
floors are covered with synthetic
material, relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output lines
IEC 60601-1-2 Test
level
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to lines(s)
± 2 kV line(s) to earth
IEC 60601-1-2 Test
level
Mains power quality should be
that of a typical commercial or
hospital environment.
<5% Uт
for 0.5cycle
Voltage dips,
short
interruption, and
voltage
variations on
power supply
input lines
IEC 60601-4-11
40% Uт
for 5, 6 cycles
IEC 60601-1-2 Test
level
70% Uт
for 25, 30 cycles
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the 1417WCC/WGC requires
continued operation during
power mains interruptions, it is
recommended that the
1417WCC/WGC be powered
from an uninterruptible power
source or battery.
<5% Uт
for 5s
Power
frequency
(50/60 Hz)
IEC 61000-4-8
3.0 A/m
IEC 60601-1-2 Test
level
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE UT is the a.c mains voltage prior to application of the test level.
1417WGC/WCC
PART II. Service Manual
Immunity Test
Conducted RF
IEC61000-4-6
IEC 60601-1-2 Test Level
3 Vrms
150 kHz to 80MHz
Compliance Level
3 Vrms
Electromagnetic environment
– guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the 1417WCC/WGC,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
Recommended separation
distance
Radiated RF
IEC61000-4-3
3 V/m
80 MHz to 2.5GHz
3 V/m
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as deter-mined by
an electromagnetic site survey,
(a) Should be less than the
compliance level in each
frequency range (b).
Interference may occur in the
vicinity of equipment marked
with the
following symbol:
Note 1)
Uт is the A.C. mains voltage prior to application of the test level.
Note 2)
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 3) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength the location in which the
1417WCC/WGC is used exceeds the applicable RF compliance level above, the 1417WCC/WGC should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the 1417WCC/WGC.
(b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V / m.
1417WGC/WCC
10
PART II. Service Manual
Recommended separation distance between portable and mobile RF
communications equipment and the 1417WCC/WGC
The 1417WCC/WGC is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The user of the 1417WCC/WGC can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the 1417WCC/WGC as recommended below,
according to the maximum output power of the communications equipment.
Separation distance (m) according to frequency of transmitter
150kHz to 80MHz
Rated maximum
output power (W)
of transmitter
80MHz to 800MHz
800MHz to 2.5GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note 1) At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
1417WGC/WCC
11
PART II. Service Manual
1.2
Symbols
Symbols
Descriptions
Refer to instruction manual / booklet
Consult instructions for use
Alternate current
Protective earth (Ground)
Off (power : disconnect from the main switch)
On (power : connect from the main switch)
On / Off (button type)
Warning
WARNING
Caution
CAUTION
To indicate a reference to the X-ray tube, for example to identify the surface of a
component such as a focused ant scatter grid, that has to be oriented towards the
X-ray tube.
This is a Type B applied part according to UL 60601-1 and IEC 60601-1.
Handle with care
Non-ionizing radiation
Partial Pressure Limit
Overall Pressure
Read carefully
Manufacturer
Date of manufacture
1417WGC/WCC
12
PART II. Service Manual
Symbols
Descriptions
Serial number
WEEE : Waste Electrical and Electronic Equipment
Authorized representative in the European community.
Batteries Marking (Battery Directive 2006/66/EC)
CE symbol grants the equipment compliance to the European Directive for
Medical Devices 93/42/EEC as a class �a device and 1999/5/EC. Authorized by
Notified Body SGS (code no.:0120) of British
Recognized Component Mark for Canada and the United States
For Korea
Symbol for safety
For Japan
Product Safety of Electrical Appliance & Materials
For JAPAN
Japan Radio Law Certification
Keep dry
Fragile, handle with care
This side up
4 layer stacking
Temperature limit
1417WGC/WCC
13
PART II. Service Manual
1.3
Warning
Environment of Use and Storage
 Follow the specified process of operational instructions written in this manual for the
safety of the users and patients.
 Does not use or store the detector near any flammable chemicals such as thinner,
WARNING
benzene, etc. Also, this detector is not a category AP or APG equipment. If chemicals
are spilled or evaporate, it may result in fire or electric shock through contact with
electric parts inside the detector. Also, some disinfectants are flammable. Be sure to
take care when using them.
Connection
 Do not connect the detector with anything other than specified. Otherwise, it may result
in fire or electric shock.
 To avoid the risk of electric shock, this detector must only be connected to supply
mains with protective earth.
 Do not touch SIP/SOP and the patient simultaneously. There is a risk of electric shock
from current leakage.
 Additional equipment connected to medical electrical equipment must comply with the
WARNING
respective IEC or ISO standards (e.g. IEC 60950 for data processing equipment).
Furthermore all configurations shall comply with the requirements for medical electrical
systems (see IEC 60601-1-1 or clause 16 of the 3 Ed. of IEC 60601-1, respectively).
Anybody connecting additional equipment to medical electrical equipment configures a
medical system and is therefore responsible that the system complies with the
requirements for medical electrical systems. Attention is drawn to the fact that local
laws take priority over the above-mentioned requirements. If in doubt, contact Rayence
Customer Service team or authorized agent.
 Equipment connected to the detector and in the patient environment must be powered
from a medically-isolated power source or must be a medically-isolated device.
Equipment powered from a non-isolated source can result in chassis leakage currents
exceeding safe levels. Chassis leakage current created by an accessory or device
connected to a non-isolated outlet may add to the chassis leakage current of the
detector.
Handling
 Always be sure to keep checking the condition of the system and the patient to ensure
they are normal during the use of the detector. If any problem is found, take
appropriate measures, such as stopping the operation of the detector, as required.
 Never disassemble or modify the detector as it may result in fire or electric shock.
WARNING
Also, since the detector incorporates parts that may cause electric shocks and other
hazardous parts, touching them may cause death or serious injury.
 Do not hit or drop the detector. The detector may be damaged if it receives a strong
jolt, which may result in fire or electric shock if the detector is used without being
repaired.
When Problem Occurs
Should any of the following occur, immediately turn OFF the power of each detector,
unplug the power supply cord from the AC outlet, and contact Rayence Customer Service
team or authorized agent.
WARNING
 When there is smoke, odd smell or abnormal sound.
 When liquid has been spilled into the detector or a metal object has entered through an
opening.
 When the detector has been dropped and it is damaged.
1417WGC/WCC
14
PART II. Service Manual
Maintenance and Inspection
 For safety reasons, be sure to turn off the power of the detector when the following
inspections are going to be performed. Otherwise, it may result in electric shock.
 When the detector is going to be cleaned, be sure to turn off the power of each
detector, and unplug the power cable from the AC outlet.
WARNING
Do not use any type of solvent, such as benzene. Otherwise, fire or electric shock may
result.
Wear waterproof gloves to protect your hands from direct contact with IPA (Isopropylalcohol) or any other liquid.
 Maintenance of the detector should be done by an authorized service provider. If
problem still cannot be corrected, it may result in fire or electric shock.
Wireless Connection
 SSID & PSK value should match to Router’s setting. If these values are not matched
with Detector and Router, the connection is not allowed for security.
 Avoid using co-channel or adjacent-channel with other wireless devices for high quality
wireless performance.
 Transmitting omnidirectional radio wave and sending same information to different
WARNING
place at once is against the law.
 Any service related to the human life safety cannot be supported since this wireless
detector has potential electric wave interference.
 This wireless detector has potential electric wave interference during use.
1417WGC/WCC
15
PART II. Service Manual
1.4
Caution
Environment of Use and Storage
Do not install the detector in a location with the conditions listed below. Otherwise, it may
result in failure or malfunction, cause fire or injury.










CAUTION
Close to facilities where water is used.
Where it will be exposed to direct sunlight.
Close to air-conditioner or ventilation equipment.
Close to heat source such as a heater.
Prone to vibration
Insecure place.
Dusty environment.
Saline or sulfurous environment.
High temperature or humidity.
Freezing or condensation.
Do not place the storage case in a location with the conditions listed below.
 Where the cable of the detector unit will be strongly pulled when the detector is put into
the case, otherwise, the cable may be damaged, resulting in fire or electric shock.
 Where someone might get their foot caught in the cable of the detector.
 Non-medical equipment such as the battery charger, access point cannot be used in
patient’s vicinity.
Handling
 Applied part: CFRP (Carbon Fiber Reinforced Plastic) Applied part having contact with
the patient for a time “t”: 1 min ≤ t < 10 min
 If contact over 24 hours on CFRP (Carbon Fiber Reinforced Plastic) of detector, it
could be cause skin irritation.
 In order to prevent infection, please wipe the CFRP (Carbon Fiber Reinforced Plastic)
with a soft cloth moistened with IPA (Isopropyl-alcohol) liquid.
CAUTION
Wear waterproof gloves to protect your hands from direct contact with IPA or any other
liquid.
 For safety reasons, be sure to turn off the power of each equipment when detector is
not used.
 This detector is contraindicated for pregnant woman.
1417WGC/WCC
16
PART II. Service Manual
Location of Cables
 Make sure all cables are located so that they cannot be stepped on, tripped over, or
CAUTION
otherwise subjected to damage or stress.
Maintenance and Inspection
 For safety reasons, be sure to inspect the detector before using it. In addition, carry out
a regular inspection at least once a year.
 If the detector is defective, do not disassemble the detector randomly. Maintenance of
CAUTION
the detector should be done by an authorized service provider. Please contact
Rayence Customer Service team or authorized agent.
 Be sure to check the user’s manual for replaceable components.
Modifications
 Any changes or modifications in construction of this detector which are not expressly
CAUTION
approved by the party responsible for compliance could void the user’s authority to
operate the detector.
Battery
 Do not let the detector or battery come in contact with liquids. Liquids can get into the
detector’s circuits, leading to corrosion. Even when the detector appears to be dry and
appears to operate normally, the circuitry could slowly corrode and pose a safety
hazard. If the battery gets wet, have them checked by authorized agent or contact
Rayence Customer Service team, even if they appear to be working properly.
 Do not place your battery near a heat source. Excessive heating can damage the
detector or the battery and could cause the detector or the battery to explode.
 Do not dry a wet or damp battery with an appliance or heat source such as a
microwave oven, hair dryer, iron, or radiator.
 Do not dispose of the detector or the battery in a fire. The detector or the battery may
explode when overheated.
CAUTION
 Use only Rayence-approved batteries and recharge your battery (Model name:
RB37WHA) only with Rayence-approved chargers (Model name: RC120WA) which are
specifically designed for your detector.
 Use of a non-Rayence-approved battery or charger may present a risk of fire,
explosion, leakage, or other hazard. Rayence’s warranty does not cover damage to the
detector caused by non-Rayence-approved batteries and/or chargers.
 Misuse or use of incompatible batteries and charging detectors could result in damage
to the detector and a possible risk of fire, explosion, or leakage, leading to serious
injuries, damages to your detector, or other serious hazard.
 Check the battery status frequently to avoid battery empty. When the low battery LED
of detector is turned on, change the battery or charge the battery using cable.
Recommendations to equipment manufacturers and battery assemblers
 The following represents a typical, but non-exhaustive, list of good advice to be
provided by the manufacturer of secondary cells and batteries to equipment
manufacturers and battery assemblers.
 Do not dismantle, open or shred cells. Batteries should be dismantled only by trained
CAUTION
personnel. Multicell battery cases should be designed so that they can be opened only
with the aid of a tool.
 Do not short-circuit a cell or battery. Do not store cells or batteries haphazardly in a box
or drawer where they may short-circuit each other or be short-circuited by conductive
materials.
 Do not remove a cell or battery from its original packaging until required for use.
1417WGC/WCC
17
PART II. Service Manual
 Do not expose cells or batteries to heat or fire. Avoid storage in direct sunlight.
 Do not subject cells or batteries to mechanical shock.
 In the event of a cell leaking, do not allow the liquid to come into contact with the skin
or eyes. If contact has been made, wash the affected area with copious amounts of
water and seek medical advice.
 Equipment should be designed to prohibit the incorrect insertion of cells or batteries
and should have clear polarity marks. Always observe the polarity marks on the cell,
battery and equipment and ensure correct use.
 Do not mix cells of different manufacture, capacity, size or type within a battery.
 Seek medical advice immediately if a cell or battery has been swallowed.
 Consult the cell/battery manufacturer on the maximum number of cells, which may be
assembled in a battery and on the safest way in which cells may be connected.
 A dedicated charger should be provided for each equipment. Complete charging
instructions should be provided for all secondary cells and batteries offered for sale.
 Keep cells and batteries clean and dry.
 Wipe the cell or battery terminals with a clean dry cloth if they become dirty.
 Secondary cells and batteries need to be charged before use. Always refer to the cell
or battery manufacturer’s instructions and use the correct charging procedure.
 Do not maintain secondary cells and batteries on charge when not in use.
 After extended periods of storage, it may be necessary to charge and discharge the
cells or batteries several times to obtain maximum performance.
 Secondary cells and batteries give their best performance when they are operated at
normal room temperature.
 Retain the original cell and battery literature for future reference.
 When disposing of secondary cells or batteries, keep cells or batteries of different
electrochemical systems separate from each other.
 Contact the Rayence Customer Service team to destroy a battery.
1417WGC/WCC
18
PART II. Service Manual
Recommendations to the end-users
 The following represents a typical, but not exhaustive list of good advice to be provided
by the equipment manufacturer to the end-user.
 Do not dismantle, open or shred secondary cells or batteries.
 Do not expose cells or batteries to heat or fire. Avoid storage in direct sunlight.
 Do not short-circuit a cell or a battery. Do not store cells or batteries haphazardly in a
box or drawer where they may short-circuit each other or be short-circuited by other
metal objects.
 Do not remove a cell or battery from its original packaging until required for use.
 Do not subject cells or batteries to mechanical shock.
 In the event of a cell leaking, do not allow the liquid to come in contact with the skin or
eyes. If contact has been made, wash the affected area with copious amounts of water
and seek medical advice.
 Do not use any charger other than that specifically provided for use with the
equipment.
 Observe the plus (+) and minus (–) marks on the cell, battery and equipment and
ensure correct use.
CAUTION








Do not use any cell or battery which is not designed for use with the equipment.
Do not mix cells of different manufacture, capacity, size or type within a device.
Keep cells and batteries out of the reach of children.
Seek medical advice immediately if a cell or a battery has been swallowed.
Always purchase the correct cell or battery for the equipment.
Keep cells and batteries clean and dry.
Wipe the cell or battery terminals with a clean dry cloth if they become dirty.
Secondary cells and batteries need to be charged before use. Always use the correct
charger and refer to the manufacturer’s instructions or equipment manual for proper
charging instructions.
 Do not leave a battery on prolonged charge when not in use.
 After extended periods of storage, it may be necessary to charge and discharge the
cells or batteries several times to obtain maximum performance.
 Secondary cells and batteries give their best performance when they are operated at
normal room temperature (20 °C ± 5 °C).





1417WGC/WCC
Retain the original product literature for future reference.
Use only the cell or battery in the application for which it was intended.
When possible, remove the battery from the equipment when not in use.
Dispose of properly.
Contact the Rayence Customer Service team to destroy a battery.
19
PART II. Service Manual
1.5
Safety Information
Preparation

Be sure to connect the cables to the proper connectors. Otherwise, the detector may
malfunction or may be damaged.

The power supply provided by Rayence is designed for the detector from Rayence. Please
contact Rayence, if any other type of power supply is needed to be used.

Be sure to fully charge the battery before use. Charge the battery on the day of examination
or on the previous day.

Battery slowly discharges even when not in use. The battery may have expired if it
discharges immediately after being fully charged. You can purchase an optional battery to
replace an exhausted one.

The battery charger provided by Rayence is designed for the dedicated battery.

When the detector will not be used for some time, remove the battery.
Handling

Handle the detector carefully, as it may become damaged if it is hit, dropped, or receives a
strong jolt.

Be sure to use the detector on a flat place so it will not bend. Otherwise, the detector may
be damaged.
1417WGC/WCC
20
PART II. Service Manual

Be sure to check the detector daily and confirm that it works properly. Sudden heating of
the room in cold areas will cause condensation to form on the detector. In this case, wait
until condensation disappears before performing exposure. If the detector is used with
condensation formed on it, problems may occur in the quality of the detector. When an airconditioner is going to be used, be sure to raise/lower the temperature gradually so that a
difference in temperature in the room and in the detector does not occur, to prevent forming
of condensation. Follow the recommended proper Room temp.

Do not use the detector near devices generating a strong magnetic field. Doing so may
produce image noise or artifacts.

Keep the connectors free from being in contact with the patient.

Connectors are intended to be connected to an external device and must follow IEC
standards.

Do not apply excessive weight to the detector. Otherwise, the detector may be damaged.
Overall Pressure: 300kg(661lb) over the whole area of detector window.
Partial Pressure: 150kg(330lb) on an area 40 mm in diameter.
Disinfection and Cleaning

Do not spray disinfectants or detergents on the detector.

When cleaning the detector, be sure to turn off the power, and unplug the power cable from
the AC outlet.

Do not use any flammable chemicals such as thinner, benzene for cleaning. Otherwise, fire
or electric shock may result.

Wear waterproof gloves to protect your hands from direct contact with disinfectants or
detergents.
1417WGC/WCC
21
PART II. Service Manual
1.6
Label and Location of Attachment
Refer to the back of the device for details.
1.7
Summary of usability specifications
Medical purposes

Provision and reading of disease and injury diagnostic images
Patient groups

No patient population exists who uses or is in contact with the device.

Patient population for the X-ray images read is not specified.
Parts of body or organizations to which the device is mounted or that interact
with the device

Detector contacts the body surface of a patient and an operator.
Significant physical and performance characteristics

Refer to 2.4 Part Specifications in this manual
Operating principles

Flat panel detector is a system that can acquire, save, process and transfer digital images
of an area of interest taken with X-ray. X-ray beam entering the X-ray imaging sensor is
converted into visible light by scintillation layer of the sensor. The amorphous silicon (a-Si)
and Photo Diode on TFT Array of the sensor further converts visible light into electric signal.
Electric signals are amplified and converted to digital signals to form image data.
Obtained image data is transferred to the computer via Ethernet or Wi-Fi interface and
visually displayed on the monitor screen.
Intended user profile

No special training is required to use this device. The intended users of this device are as
follows.
A professional in good health with specialist knowledge/ qualifications who has fully understood
the content of this document. (Such as a doctor or radiological technologist)
1417WGC/WCC
22
PART II. Service Manual
2.
Product Introduction and Specification
2.1
Product Features
The 1417WGC/WCC are wireless digital flat panel detectors that has been designed for a
faster, more streamlined approach to digital radiography systems.
The 1417WGC/WCC detector contains a built-in Access Point (AP) enabling images to be
directly sent to a Wi-Fi connected computer within seconds. Built-in image memory storage
permits taking images where a computer connection is not available and also prevents lost
images should there be an interruption of power. Whether an image was taken with the detector
in the portrait or landscape position, the auto image rotation function allows images to be
displayed in the correct orientation.
These features, coupled with an auto-trigger signal sensing technology that allows the detectors
to be used without generator integration, makes the 1417WGC/WCC the ideal flat panel
detector solutions for both fixed and portable applications.
1417WGC/WCC
23
PART II. Service Manual
2.2
Product Components
2.2.1
Basic Components
2.2.2
Detector 1EA
Battery 2EA
Charger Adaptor 1EA &
Mobile charger adaptor 1EA
AC power cord 1EA
(110V or 220V 1.8m)
Installation CD 1EA
Manual 1EA
Optional Components
Power
supply
package
Power supply1EA
AC power cord 1EA
(110V or 220V 1.8m)
LAN cable 1EA
(CAT 6, 10m)
(straight-through)
Link Cable 1EA
(7, 9m)
Trigger Cable 1EA
P-interface Cable
1EA
AGI Box 1EA
USB Cable 1EA
AGI
package
1417WGC/WCC
24
PART II. Service Manual
Interface
box
package
RAP001A
Link Cable 1EA
(7, 9m)
Trigger Cable 1EA
P-interface Cable
1EA
Battery Charger 1EA
Battery Charger
adapter 1EA
IrDA module 1EA
Micro USB cable
1EA
AC power cord 1EA
(110V or 220V 1.8m)
LAN cable 1EA
(CAT 6, 10m)
(straight-through)
AC power cord 1EA
(110V or 220V 1.8m)
Battery 2EA
Battery
Charger
Package
Sharing
Package
1417WGC/WCC
25
PART II. Service Manual
2.3
Part Names and Functions
2.3.1
Detector
1.
Wireless Antenna
2.
Battery Unlock button
This is an unlock button to remove battery
3.
IrDA & OLED window
IrDA module can communicate with infrared ray and inform the product status.
4.
Mode select button & LED indicator
Mode Change
Press the 1 second
Indication the status of detector
Name of LED
Status of LED
Status of product
Orange on
Station mode(wireless)
Green on
AP mode(wireless)
Off
Wired mode
Blinking Orange and Green
alternatively
Sleep mode
Mode
5.
Power button & LED indicator
Power On / Off
Press over than 3 seconds
Indicating the status of detector
Name of LED
Power
1417WGC/WCC
Status of LED
Status of product
Green
Power on
Blinking green
Power booting
Red
Low battery
(Battery remain 0~7%)
Orange
Low battery
(Battery remain 7~15%)
26
PART II. Service Manual
6.
Link cable connector
Use for data transfer and charging battery while wired mode is in use (Connect between
detector and power supply.)
2.3.2
Battery & Mobile charger
1.
Battery unlock level
This is an unlock-level to remove battery
2.
Power connector
Connects to the mobile charger adapter
3.
LED Indicator
Display battery charging status.
4.
LED Color
Battery Status
Green
Fully charged
Orange
Charging
Blinking orange and Green alternatively
Error
Battery : Rechargeable Lithium Ion battery(Charging Time-3 hrs)
•
In the diagram above, the box shows where the remaining battery percentage is
displayed.
Battery Remain Indicator
Battery Level
75~100 %
50~75 %
1417WGC/WCC
27
PART II. Service Manual
Battery Remain Indicator
Battery Level
25~50 %
0~25 %
•
2.3.3
Battery warranty period: 6 months
RAP001A (Optional)
1.
Switch
Power On/Off switch
2.
Fuse
T3.15 AL 250V
3.
Power plug connector
Connects to the AC power cord
4.
Trigger connector
This is a connector to synchronize the detector and generator. Connect the RAP001A to the
generator by using a P-interface cable or trigger cable.
5.
Link connector 1~3
Use for data transfer and charging battery while detector is in use (Connect between
detector and RAP001A), Up to three detector can be connected
6.
LAN connector
Ethernet port for transmitting an image/command between the detector and PC
7.
Reset button
Reset all settings to default values
8.
Indicating the status of RAP001A
LED
PWR/LAN
TRIGGER
1417WGC/WCC
LED Color
Status
Red
Turned on while booting after connecting with power supply.
Turned off upon the completion of booting.
Blinking Green
Connected with PC LAN
Red
Trigger READY DONE
Green
Trigger READY IN
28
PART II. Service Manual
LED
LED Color
Status
Red
Frequency of 2.4GHz
Green
Frequency of 5GHz
Yellow
Frequency of 2.4GHz and 5GHz
DET1
Green
Connected with detector of port1
DET2
Green
Connected with detector of port2
DET3
Green
Connected with detector of port3
5G/2.4G
2.3.4
Power Suppy (Optional)
1.
Power Plug Connector
Connects to the AC power cord
2.
LAN Connector
Ethernet port for transmitting an image/command between the detector and PC
3.
Link Connector
Used for charging the battery while the detector is in use (Connect the detector and power
supply)
4.
LED Indicator
Display status of the power supply.
5.
Color
Status
Power Status
Green
On
Power on
Fuse
T3.15 AL 250V
6.
Switch
Power On/Off switch
1417WGC/WCC
29
PART II. Service Manual
2.3.5
AGI Box (Optional)
7.
Trigger Connector
This is a connector to synchronize the detector and generator. Connect the AGI to the
generator by using a P-interface cable or trigger cable.
8.
USB Connector
This is a connector for communication between the AGI and PC. Connect the AGI to the PC
by using a USB cable.
2.3.6
IrDA module (Optional)
1.
IrDA window
Window for communicating intrared with Detector
2.
Micro USB connector
Connector for communication with IrDA and PC connect the PC through the Micro USB
cable.
1417WGC/WCC
30
PART II. Service Manual
2.3.7
Battery & Charger (Optional)
1.
Battery : Rechargeable Lithium Ion battery(Charging Time-3 hrs)
•
In the diagram above, the box shows where the remaining battery percentage is
displayed.
Battery Remain Indicator
Battery Level
75~100 %
50~75 %
25~50 %
0~25 %
•
Battery warranty period: 6 months
2.
Battery Charger : Two port cradle type
3.
LED Indicator
Display battery charging status.
1417WGC/WCC
LED Color
Battery Status
Green
Fully charged
Orange
Charging
Red
Error
31
PART II. Service Manual
2.4
Part Specifications
2.4.1
Detector
Parameter
Spec.
Unit
Sensor Type
Amorphous Silicon with TFT (Single panel)
Scintillator Type
1417WCC_127/140: CsI:Tl
1417WGC_127/140: Gd2O2S:Tb
Total Pixel Matrix
127type: 3328 X 2816
140type: 2500 X 3052
Pixels
Total Pixel Area
127type: 422.7 X 357.6
140type: 350.0 X 427.3
mm
Pixel Pitch
127type: 127
140type: 140
μm
Effective Pixel Matrix
127type: 3268 X 2756
140type: 2440 X 2992
A/D Conversion
14 / 16
Data Transfer
1Gbps Ethernet
802.11 a/g/n Wireless LAN, Wireless LAN up to
867Mbps
Preview time
≤2 (2x2 binning)
sec
Energy range
40 ~ 150
kVp
Limiting Resolution
127 type: Min.2.5/ Max. 3.93
140 type: Min.2.5/ Max. 3.57
Pixels
bits
lp/mm
1417WCC_127 - Min. 50 / Typ. 59
1417WCC_140 - Min. 50 / Typ. 57
MTF (@1lp/mm)
1417WGC_127 - Min. 50 / Typ. 57
1417WGC_140 - Min. 50 / Typ. 55
1417WCC_127 - Min 50 / Typ. 65
1417WCC_140 - Min. 50 / Typ. 65
DQE (@0.1lp/mm)
1417WGC_127 -Min. 36 / Typ. 45
1417WGC_140 - Min. 36 / Typ. 45
Dimension
460 x 384 x 15
mm
Weight
3 (incl. battery)
Kg
Sensor Protection Material
Carbon fiber plate
Trigger mode
Manual Mode
Auto Trigger Mode (Auto Exposure Detection)
Power consumption
Typ. 8.8
Wireless specifications
1417WGC/WCC
Standard
802.11 a/g/n compliance
Without DFS (5.25GH to 5.35GHz
and 5.47 to 5.725) Band
Peak Rate
1300Mbs
Frequency
2.4 GHz / 5GHz
Bandwidth
20MHz / 40MHz
32
PART II. Service Manual
Parameter
Spec.
MIMO
2X2

Maximum wireless signal rate derived from IEEE standard specifications. Actual data
throughput will vary. Network conditions and environmental factors, including volume of
network traffic, building materials and construction, and network overhead, lower actual
data throughput rate.

Recommended Maximum operable distance : 10m (From the Access Point)

Wireless Module and Wireless Antenna
•
2.4.2
Wireless antennas: The module adopts the latest 802.11n Dual-Band technology
(2.4Ghz and 5Ghz). The transmitter of the module is powered by host equipment
(Detector). The antennas are 2 printed-dipole antennas.
Battery [Model name: RB37WHA]
Parameter
2.4.3
Spec.
Unit
Size
236.2 x 127.8 x 6
mm
Weight
0.3
Kg
Input
12.6
VDC
Output
11.1
VDC
Cycle life
Max. 500
Operation temp. range
5~40
℃
Charging time
Typ. 3
hours
Capacity
Typ. 3400
mAh
Operating time
Typ. 4
hours
cycles
Mobile charger [Model name: RMC001A]
Parameter
2.4.4
Unit
Spec.
Unit
Dimension
280 X 160 X 17
mm
Weight
0.3
Kg
Input
18
VDC
Output
12.6
VDC
Mobile charger Adaptor [Model name: AFM60US18]
Parameter
Spec.
Unit
Dimension
125 X 62 X 34
mm
Weight
0.4
Kg
Input
80-264VAC, 47~63Hz, 1.5A
Output
18VDC, Max 3.34A
1417WGC/WCC
33
PART II. Service Manual
2.4.5
Interface box (Optional) [Model name: RAP001A]
Parameter
2.4.6
Spec.
Dimension
240 X 191.2 X 65.8 (not incl. antenna)
mm
Weight
1.85
Kg
Input rate
85 ~ 264VAC (50/60Hz)
Output
Typ. 24VDC (Max 5.4A)
Wireless communication
802.11 a/g/n Wireless LAN, up to 867Mbps
Wired communication
Ethernet 1Gbps
Detector Port
Trigger Port
Spec.
Unit
188 X 92 X 41.5
mm
Weight
0.5
Kg
Rated power supply(Input)
100-240VAC (50/60Hz)
Rated power supply(Output)
Typ. 24VDC (Max 1.6A)
AGI Box (Optional)
Spec.
Unit
Dimension
92.5 X 119 X 33.5
mm
Weight
0.3
Kg
IrDA module (Optional) [Model name – RI001A]
Parameter
2.4.9
Dimension
Parameter
2.4.8
Power Supply (Optional) [Model name -RP003A]
Parameter
2.4.7
Unit
Spec.
Unit
Dimension
66 X 46 X 9.5
mm
Weight
0.1
Kg
Battery Charger (Optional) [Model name: RC120WA]
Parameter
Spec.
Unit
Size
200 x 100 x 150
mm
Weight
0.9
Kg
Input
20
VDC
Output
12.6
VDC
2.4.10 Battery Charger Adapter (Optional) [Model name: PMP120-13-3]
Parameter
Size
1417WGC/WCC
Spec.
160 x 76 x 43.7
Unit
mm
34
PART II. Service Manual
Parameter
Spec.
Unit
Weight
0.8
Kg
Input
100-240VAC, 47~63Hz, 1.4~0.6A
Output
20VDC, Max 6.0A
2.4.11 Cable
Parameter
Length
Unit
7 up to 9
LAN cable
(CAT 6(straight-through), Optional)
10
Power cord (110V or 220V)
1.8
USB cable (Optional)
1.8
Trigger cable (Optional)
10
P-interface cable (Optional)
Micro USB cable (Optional)
Link cable (Optional)
1417WGC/WCC
35
PART II. Service Manual
2.5
Environmental Requirements
2.5.1
PC Requirement
Item
Detail
CPU
At least Intel Pentium IV HT with 2.8GHz, Intel Core Duo / Core 2 or
comparable AMD Dual Core processor
RAM
At least 3GB of RAM requirement
(4GB for 32 BITS OS and 8GB for 64 BITS OS recommended)
Capacity of Disk Drive
At least 500GB for application and archiving. Recommended 500GB for
applications and secondary drive of 1TB for image archiving.
 Dual 10/100/1000 network card system required. One for network
(Internet) and one for the DR Panel communication
Network Card
 802.11 a/g/n Wireless LAN card required (optional)
Graphic Card / Monitor
Graphics card / monitor: Resolution of at least 1,600 x 900 for desktop and
1366 x 768 for laptop. For diagnostics purpose we recommend 1920 x
1080 resolution (2 mega pixels) monitor
Operating System(OS)
MicrosoftⓇ Windows XP/VISTA/7/8/10 32BIT/64BIT
ETC
2.5.2
2.5.3
No antivirus except for MicrosoftⓇ Security Essentials.
Environmental Requirement
Environment
Min.
Temperature(Storage)
Typ.
Max.
Unit
-10
50
℃
Temperature(Operation)
35
℃
Humidity(Storage)
10
80
% H.R.
Humidity(Operation)
30
75
% H.R.
Pressure(Operation)
70
106
kPa
Note
Grid Requirement
127um
Item
Description
SID
100/130/180 cm
Ratio
8:1
Frequency
230 Line/inch
Item
Description
SID
100/130/180 cm
Ratio
8:1
Frequency
215 Line/inch
 139um
1417WGC/WCC
36
PART II. Service Manual
2.6
Dimensions (Unit: mm)
2.6.1
Detector
2.6.2
Battery
1417WGC/WCC
37
PART II. Service Manual
2.6.3
Mobile charger
2.6.4
Mobile charger adapter
1417WGC/WCC
38
PART II. Service Manual
2.6.5
Interface box (Optional)
2.6.6
Power Supply (Optional)
1417WGC/WCC
39
PART II. Service Manual
2.6.7
AGI Box (Optional)
2.6.8
IrDA module (Optional)
1417WGC/WCC
40
PART II. Service Manual
2.6.9
Battery Charger (Optional)
2.6.10 Battery Charger Adapter (Optional)
1417WGC/WCC
41
PART II. Service Manual
3.
Installation and Calibration
3.1
Installation
3.1.1
Software Installation
1.
Insert the CD that comes with the Detector.
2.
Install "setup.exe" from "\Release Davinci_version" and click "Next".
3.
Choose the model from the Detector Type list and click "Next".
1417WGC/WCC
42
PART II. Service Manual
4.
Click “Next”.
5.
Select “MultiByte-Character Set” and click “Next”.
 Choose UNICODE if console SW is supporting UNICODE.
 If Character set is not installed correctly, images will not be properly acquired.
1417WGC/WCC
43
PART II. Service Manual
6.
Click “Install”.
7.
Click “Finish”.
1417WGC/WCC
44
PART II. Service Manual
3.1.2
Install battery
Attach the battery (RB37WHA) to the detector as below.
•
Battery removal
Push the battery unlock button.
3.1.3
Mode Selection
Mode
Description

Automatically detects X-ray radiation without integration of the
generator and detector.

Detects X-ray radiation by sending and receiving sync signals through
the integration between generator and detector.

Communicates with the wireless AP, and the wireless AP
communicates with the PC through the LAN cable.
(Rayence does not provide wireless AP.)

Detector communicates with the PC without the wireless AP.
(PC must maintain wireless network card.)

Communicates with the power supply, and the power supply
communicates with the PC through the LAN cable.
1417WGC/WCC
45
PART II. Service Manual
Auto Trigger & Station Mode
Follow instructions from 3.1.4 Product Set Up 1. Auto Trigger & Station Mode.
Auto Trigger & AP Mode
Follow instructions from 3.1.4 Product Set Up 2. Auto Trigger & AP Mode.
1417WGC/WCC
46
PART II. Service Manual
Manual Trigger & Station mode
Follow instructions from 3.1.4 Product Set Up 3. Manual Trigger & Station Mode.
1417WGC/WCC
47
PART II. Service Manual
Manual Trigger & AP mode
Follow instructions from 3.1.4 Product Set Up 4. Manual Trigger & AP Mode.
1417WGC/WCC
48
PART II. Service Manual
Auto Trigger & Wired Mode
Follow instructions from 3.1.4 Product Set Up 5. Auto Trigger & Wired Mode.
1417WGC/WCC
49
PART II. Service Manual
Manual Trigger & Wired Mode
Follow instructions from 3.1.4 Product Set Up 6. Manual Trigger & Wired Mode.
1417WGC/WCC
50
PART II. Service Manual
3.1.4
Product Set Up
1. Auto Trigger & Station Mode
Product Set up
Connect the cable
1. Connect the wireless AP (or RAP001A) and PC with the LAN cable.
1417WGC/WCC
51
PART II. Service Manual
PC Set up
1.
2.
Set up the Network as below.
•
Desktop > Network Icon > Right click > Properties > Change Adaptor Settings
•
Control Panel > Network and Sharing Center > Change Adaptor Settings
To use station mode, right click "Local Area Connection" and click Properties.
1417WGC/WCC
52
PART II. Service Manual
3.
Double click “Internet Protocol Version 4 (TCP/IPv4)”.
4.
Select “Obtain an IP address automatically” and click "OK".
Wireless AP (Access Point) Set up
 Rayence does not provide wireless AP. Please use certified wireless AP and follow
each manufacture's setup manual.
(Recommended Model: Minimum ASUS RT-AC66U or RAP001A)
1.
Set up wireless AP as below.
•
SSID: Griffon
•
Internal network
•
•
•
IP address: 2.2.2.1
•
Subnet mask: 255.255.255.0
•
Dynamic IP allocation range: 2.2.2.2 ~ 2.2.2.254
Pre-Shared Key(PSK): project302
•
Authentication methods: WPAPSK or WPA2PSK
•
Password methods: TKIP/AES
Channel (Frequency)
Avoid the crowded channel option.
 Recommend to use "Auto-Channel selection" function if external AP has the
feature.
Part.2 Service Manual Supplement 1. Refer to Wireless AP Set Up Instruction
(WAP Model: ASUS RT-AC66U).
1417WGC/WCC
53
PART II. Service Manual
Set up SW
1.
Connect Detector and turn on the power.
2.
Choose Station Mode by pressing and holding the Mode Button.
LED
LED Color
Orange
Station Mode(Wireless)
Green
AP Mode(Wireless)
None
Wired Mode
Blinking Orange and Green
alternatively
Sleep Mode
MODE
3.
Mode
Open “_vadav.lnk” from “C:\davinci”.
Once the program is opened and the detector is connected, the LINK LED light from the
detector will blink and the Detector Status will display information of the detector as below.
 Once the correct Sensor IP is put into the Davinci, it will automatically pull the
parameter of the connected detector.
Model type
Parameter Selected
1417WCC _127A
[69]1417WCC_127A.par
1417WCC _140A
[71]1417WCC_140A.par
1417WGC _127A
[70]1417WGC_127A.par
1417WGC _140A
[72]1417WGC_140A.par
 Default IP address for wireless connection is 2.2.2.100 and for wired connection is
2.2.2.101. If the IP address needs to be changed, please refer to 2.1 Detector IP
Address Set Up in Part.2 Service Manual.
1417WGC/WCC
54
PART II. Service Manual
If the detector does not communicate with the PC, please check the connection of the
cable, PC set up and power of the detector.
4.
After checking connectivity, click the “Calibration & Acquisition” tab and click "Edit".
5.
Another window will now be opened as shown below. Select "Auto Trigger" from "Trigger
Mode". If the "Window time" needs to be changed, type the value at "Window Time" from
"Setting".
 In Auto trigger mode, be sure to set the "Window time" longer than an exposure
time. If the "Window time" is shorter than the exposure time, images will not be
properly acquired.
1417WGC/WCC
55
PART II. Service Manual
2. Auto Trigger & AP Mode
Product Set Up
PC Set up
1.
Set up the Network as below.
•
Desktop > Network Icon > Right click > Properties > Change Adaptor Settings
•
Control Panel > Network and Sharing Center > Change Adaptor Settings
1417WGC/WCC
56
PART II. Service Manual
2.
To use AP mode, right click "Wireless Network Connection" and click Properties.
3.
Double click “Internet Protocol Version 4 (TCP/IPv4)”.
4.
Select “Obtain an IP address automatically” and click "OK".
1417WGC/WCC
57
PART II. Service Manual
Set up SW
1.
Connect the Detector and turn on the power.
2.
Choose AP Mode by pressing and holding the Mode button
LED
MODE
3.
LED Color
Mode
Orange
Station Mode(Wireless)
Green
AP Mode(Wireless)
None
Wired Mode
Blinking Orange and Green
alternatively
Sleep Mode
Choose the SSID (detector's SN) from Wireless Network Connection list. (PW: project302)
1417WGC/WCC
58
PART II. Service Manual
4.
Open “_vadav.lnk” from “C:\davinci”.
Once the program is opened and the detector is connected, the LINK LED light will blink
and the Detector Status will display detector information as below.
 Once the correct Sensor IP is put into the Davinci, it will automatically pull the
parameter of the connected detector.
Model type
Parameter Selected
1417WCC _127A
[69]1417WCC_127A.par
1417WCC _140A
[71]1417WCC_140A.par
1417WGC _127A
[70]1417WGC_127A.par
1417WGC _140A
[72]1417WGC_140A.par
 Default IP address for wireless connection is 2.2.2.100 and for wired connection is
2.2.2.101. If the IP address needs to be changed, please refer to 2.1 Detector IP
Address Set Up in Part.2 Service Manual.
 In the AP mode, select "AP" from "Wireless Link Quality" and type the detector's
serial number at the "SSID".
If the detector does not communicate with the PC, please check the connection of the
cable, PC set up and power of the detector.
1417WGC/WCC
59
PART II. Service Manual
5.
After checking connection, click the “Calibration & Acquisition” tab and click "Edit".
6.
Another window will open as below once the “Edit” button is pressed. Select "Auto Trigger"
from "Trigger Mode". If "Window time" needs to be changed, type the value at "Window
Time" from "Setting".
 In Auto trigger mode, be sure to set the "Window time" longer than an exposure time. If
the "Window time" is shorter than the exposure time, images will not be properly
acquired.
1417WGC/WCC
60
PART II. Service Manual
3. Manual Trigger & Station Mode
Product Set up
Connect the cable (with AGI and AP)
1.
Connect the wireless AP and PC with the LAN cable.
2.
Connect the AGI box and PC with the USB cable.
1417WGC/WCC
61
PART II. Service Manual
3.
Connect the AGI box and generator with the P-interface or trigger cable.
Instruction of P-interface cable Integration


Instruction of Trigger cable Integration

1417WGC/WCC

62
PART II. Service Manual
Product Set up (with RAP001A)
Connect the cable (with RAP001A)
1. Connect the RAP001A and PC with the LAN cable.
1417WGC/WCC
63
PART II. Service Manual
2. Connect the RAP001A and generator with the P-interface or trigger cable.
Instruction of P-interface cable Integration


Instruction of Trigger cable Integration

1417WGC/WCC

64
PART II. Service Manual
PC Set up
1.
2.
Set up the Network as below.
•
Desktop > Network Icon > Right click > Properties > Change Adaptor Settings
•
Control Panel > Network and Sharing Center > Change Adaptor Settings
To use station mode, right click "Local Area Connection" and click Properties.
1417WGC/WCC
65
PART II. Service Manual
3.
Double click “Internet Protocol Version 4 (TCP/IPv4)”.
4.
Select “Obtain an IP address automatically” and click "OK".
Wireless AP (Access Point) Set up
 Rayence does not provide wireless AP. Please use certified wireless AP and follow
each manufacture's setup manual.
(Recommended Model: Minimum ASUS RT-AC66U or RAP001A)
1.
Set up wireless AP as below.
•
SSID: Griffon
•
Internal network
•
•
•
IP address: 2.2.2.1
•
Subnet mask: 255.255.255.0
•
Dynamic IP allocation range: 2.2.2.2 ~ 2.2.2.254
Pre-Shared Key(PSK): project302
•
Authentication methods: WPAPSK or WPA2PSK
•
Password methods: TKIP/AES
Channel (Frequency)
Avoid the Crowded channel option.
 Recommend to use "Auto-Channel selection" function if external AP has the
feature.
2.
Part.2 Service Manual Supplement 1. Refer to Wireless AP Set Up Instruction (WAP
Model: ASUS RT-AC66U
1417WGC/WCC
66
PART II. Service Manual
Set up SW
1.
Connect the Detector and turn on the power.
2.
Choose Station Mode by pressing and holding the Mode Button
LED
MODE
1417WGC/WCC
LED Color
Mode
Orange
Station Mode(Wireless)
Green
AP Mode(Wireless)
None
Wired Mode
Blinking Orange and Green
alternatively
Sleep Mode
67
PART II. Service Manual
3.
Open “_vadav.lnk” from “C:\davinci”.
Once the program is opened and the detector is connected, the LINK LED light will blink
and the Detector Status will display detector information as below.
 Once the correct Sensor IP is put into the Davinci, it will automatically pull the
parameter of the connected detector.
Model type
Parameter Selected
1417WCC _127A
[69]1417WCC_127A.par
1417WCC _140A
[71]1417WCC_140A.par
1417WGC _127A
[70]1417WGC_127A.par
1417WGC _140A
[72]1417WGC_140A.par
 Default IP address for wireless connection is 2.2.2.100 and for wired connection
is 2.2.2.101. If the IP address needs to be changed, please refer to 2.1 Detector
IP Address Set Up in Part.2 Service Manual.
If the detector does not communicate with the PC, please check the connection of the
cable, PC set up and power of the detector.
1417WGC/WCC
68
PART II. Service Manual
4.
After checking connectivity, click the “Calibration & Acquisition” tab and click "Edit".
5.
Another window will be opened as below once the “Edit” button is pressed. Select "Manual"
from "Trigger Mode". If "Window time" needs to be changed, type the value at "Window
Time" from "Setting".
1417WGC/WCC
69
PART II. Service Manual
4. Manual Trigger & AP Mode
Product Set Up
Connect the cable (with AGI)
1.
Connect the AGI box and PC with the USB cable.
1417WGC/WCC
70
PART II. Service Manual
2.
Connect the AGI box and generator with a P-interface or trigger cable.
Instruction of P-interface cable Integration


Instruction of Trigger cable Integration

1417WGC/WCC

71
PART II. Service Manual
Product Set Up (with RAP001A)
Connect the cable (with RAP001A)
1. Connect the RAP001A and PC with the LAN cable.
1417WGC/WCC
72
PART II. Service Manual
2. Connect the RAP001A and generator with the P-interface or trigger cable.
Instruction of P-interface cable Integration


Instruction of Trigger cable Integration

1417WGC/WCC

73
PART II. Service Manual
PC Set up
1.
2.
Set up the Network as below.
•
Desktop > Network Icon > Right click > Properties > Change Adaptor Settings
•
Control Panel > Network and Sharing Center > Change Adaptor Settings
To use AP mode, right click "Wireless Network Connection" and click Properties.
1417WGC/WCC
74
PART II. Service Manual
3.
Double click “Internet Protocol Version 4 (TCP/IPv4)”.
4.
Select “Obtain an IP address automatically” and click "OK".
Set up SW
1.
Connect the Detector and turn on the power.
2.
Choose AP Mode by pressing and holding the Mode
LED
MODE
3.
LED Color
Mode
Orange
Station Mode(Wireless)
Green
AP Mode(Wireless)
None
Wired Mode
Blinking Orange and Green
alternatively
Sleep Mode
Choose the SSID (detector's SN) from the Wireless Network Connection list. (PW:
project302)
1417WGC/WCC
75
PART II. Service Manual
4.
Open “_vadav.lnk” from “C:\davinci”.
Once the program is opened and the detector is connected, the LINK LED light from the
detector will blink and the Detector Status will display panel information as below.
 Once the correct Sensor IP is put into the Davinci, it will automatically pull the
parameter of the connected detector.
Model type
Parameter Selected
1417WCC _127A
[69]1417WCC_127A.par
1417WCC _140A
[71]1417WCC_140A.par
1417WGC _127A
[70]1417WGC_127A.par
1417WGC _140A
[72]1417WGC_140A.par
 Default IP address for wireless connection is 2.2.2.100 and for wired connection is
2.2.2.101. If the IP address needs to be changed, please refer to 2.1 Detector IP
Address Set Up in Part.2 Service Manual.
 In the AP mode, select "AP" from "Wireless Link Quality" and type the detector's
serial number at the "SSID".
If the detector does not communicate with PC, please check the connection of the cable,
PC set up and power of detector.
1417WGC/WCC
76
PART II. Service Manual
5.
After checking connectivity, click the “Calibration & Acquisition” tab and click "Edit".
6.
Another window will be opened as below once the “Edit” button is pressed. Select "Manual"
from "Trigger Mode". If the "Window time" needs to be changed, type the value at "Window
Time" from "Setting".
1417WGC/WCC
77
PART II. Service Manual
5. Auto Trigger & Wired Mode
Product Set Up
Connect the cable
1. Connect the power supply or RAP001A and PC with the LAN cable.
2. Connect the power supply and Detector with the Link cable.
1417WGC/WCC
78
PART II. Service Manual
PC Set up
1.
Set up the Network as below.
•
Desktop > Network Icon > Right click > Properties > Change Adaptor Settings
•
Control Panel > Network and Sharing Center > Change Adaptor Settings
2.
To use Wired mode, right click "Local Area Connection" and click Properties.
3.
Double click “Internet Protocol Version 4 (TCP/IPv4)”.
1417WGC/WCC
79
PART II. Service Manual
4.
Select “Use the following IP address” and type "2.2.2.20" at the "IP address" and type
"255.255.255.0" at the "Subnet mask".
Set up SW
1.
Turn on the power supply or RAP001A
2.
Connect the Detector to the power supply or RAP001A using the Link cable (Wired mode
will be selected automatically)
LED
MODE
1417WGC/WCC
LED Color
Mode
Orange
Station Mode(Wireless)
Green
AP Mode(Wireless)
None
Wired Mode
Blinking Orange and Green
alternatively
Sleep Mode
80
PART II. Service Manual
3.
Open “_vadav.lnk” from “C:\davinci”.
Once the program is opened and the detector is connected, the LINK LED light will blink
and the Detector Status will display detector information as below.
 Once the correct Sensor IP is put into the Davinci, it will automatically pull the
parameter of the connected detector.
Model type
Parameter Selected
1417WCC _127A
[69]1417WCC_127A.par
1417WCC _140A
[71]1417WCC_140A.par
1417WGC _127A
[70]1417WGC_127A.par
1417WGC _140A
[72]1417WGC_140A.par
 Default IP address for wireless connection is 2.2.2.100 and for wired connection is
2.2.2.101. If the IP address needs to be changed, please refer to 2.1 Detector IP
Address Set Up in Part.2 Service Manual.
If the detector does not communicate with the PC, please check the connection of the
cable, PC set up and power of the detector.
1417WGC/WCC
81
PART II. Service Manual
4.
After checking connectivity, click the “Calibration & Acquisition” tab and click "Edit".
5.
Another window will be opened as below once the “Edit” button is pressed. Select "Auto
Trigger" from "Trigger Mode". If "Window time" needs to be changed, type the value at
"Window Time" from "Setting".
 In Auto trigger mode, be sure to set the "Window time" longer than an exposure
time. If the "Window time" is shorter than the exposure time, images will not be
properly acquired.
1417WGC/WCC
82
PART II. Service Manual
6. Manual Trigger & Wired Mode
Product Set up (with Power supply and AGI)
Connect the cable (with Power supply and AGI)
1.
Connect the power supply and PC with the LAN cable.
2.
Connect the power supply and Detector with the Link cable.
1417WGC/WCC
83
PART II. Service Manual
3.
Connect the AGI box and PC with the USB cable.
4.
Connect the AGI box and generator with the P-interface or trigger cable.
Instruction of P-interface cable Integration


Instruction of Trigger cable Integration

1417WGC/WCC

84
PART II. Service Manual
Product Set up (with RAP001A)
Connect the cable (with RAP001A)
1.
Connect the RAP001A and PC with the LAN cable.
2.
Connect the RAP001A and Detector with the Link cable.
1417WGC/WCC
85
PART II. Service Manual
3.
Connect the RAP001A and generator with the P-interface or trigger cable.
Instruction of P-interface cable Integration


Instruction of Trigger cable Integration

1417WGC/WCC

86
PART II. Service Manual
PC Set up
1.
Set up the Network as below.
•
Desktop > Network Icon > Right click > Properties > Change Adaptor Settings
•
Control Panel > Network and Sharing Center > Change Adaptor Settings Set up SW
2.
To use Wired mode, right click "Local Area Connection" and click Properties.
1417WGC/WCC
87
PART II. Service Manual
3.
Double click “Internet Protocol Version 4 (TCP/IPv4)”.
4.
Select “Use the following IP address” and type "2.2.2.20" at the "IP address" and type
"255.255.255.0" at the "Subnet mask".
Set up SW
1.
Turn on the power supply or RAP001A
2.
Connect the Detector to the power supply or RAP001A using the Link cable (Wired mode
will be selected automatically)
LED
MODE
1417WGC/WCC
LED Color
Mode
Orange
Station Mode(Wireless)
Green
AP Mode(Wireless)
None
Wired Mode
Blinking Orange and Green
alternatively
Sleep Mode
88
PART II. Service Manual
3.
Open “_vadav.lnk” from “C:\davinci”.
Once the program is opened and the detector is connected, the LINK LED light will blink
and the Detector Status will display detector information as below.
 Once the correct Sensor IP is put into the Davinci, it will automatically pull the
parameter of the connected detector.
Model type
Parameter Selected
1417WCC _127A
[69]1417WCC_127A.par
1417WCC _140A
[71]1417WCC_140A.par
1417WGC _127A
[70]1417WGC_127A.par
1417WGC _140A
[72]1417WGC_140A.par
 Default IP address for wireless connection is 2.2.2.100 and for wired connection is
2.2.2.101. If the IP address needs to be changed, please refer to 2.1 Detector IP
Address Set Up in Part.2 Service Manual.
If the detector does not communicate with the PC, please check the connection of the
cable, PC set up and power of the detector.
1417WGC/WCC
89
PART II. Service Manual
4.
After checking connectivity, click the “Calibration & Acquisition” tab and click "Edit".
5.
Another window will be opened as below once the “Edit” button is pressed. Select "Manual"
from "Trigger Mode". If "Window time" needs to be changed, type the value at "Window
Time" from "Setting".
1417WGC/WCC
90
PART II. Service Manual
3.2
Calibration
In order to properly acquire images, calibration must be performed. Without calibration, optimum
images cannot be acquired.
 Rayence recommends to warm up the detector for 5 minutes after turning on the
power.
 Calibration should be performed at regular intervals, typically once every six (6)
months, or whenever the central beam of the X-ray source has been moved relative to
the Detector.
3.2.1
Auto Calibration Mode
1.
Connect the detector and turn the power on.
2.
Open “_vadav.lnk” from “C:\davinci”.
3.
Once the detector is connected, information of the detector is displayed in Detector Status
and Link & Ready are checked as below.
 If "Detector Status" does not show anything, please refer to 3.1Installation in
Part.1 User & Installation Manual to connect the detector properly.
1417WGC/WCC
91
PART II. Service Manual
4.
After checking connection, click “Calibration & Acquisition” tab.
5.
Click “Calibration”.
1417WGC/WCC
92
PART II. Service Manual
6.
Once "Start” is pressed, the program automatically gets a Dark frame and the acquired
Dark frame is stored in “C:\davinci\cal”. A Calibration Point file will be created automatically.
7.
After acquiring the Dark frame, shoot an X-ray when the “Acquiring bright frame” window
pops up.
1417WGC/WCC
93
PART II. Service Manual
8.
“Acquiring bright frame” is closed after radiation is detected, and the program will show if
the detected radiation is within acceptable range.
When PASS is displayed
Shoot X-ray with same technique when
“Acquiring bright frame” is popped up.
9.
When FAIL is displayed
Adjust technique to get acceptable value and
shoot again.
For each Calibration point, four images must successfully be acquired. After successfully
doing so for every point, the Calibration process is complete.
10. Click "OK" to move to the next step.
1417WGC/WCC
94
PART II. Service Manual
3.2.2
Manual Bad Pixel Map Set Up
1.
Click “Calibration”.
2.
Click "View Frames".
1417WGC/WCC
95
PART II. Service Manual
3.
Set Manual bad pixel map (BPMM) as below.
No.
Overview
At Figure 1, choose S from the list of M(❶) and click "BPM". Check if BPMM window is
popped up as Figure 3.
Choose "Bad" from Pixel list (Figure 3 - ❷ ).
Choose either “row” or “col” from Figure 3 - ❸.
Put the coordinate of pixel to set bad pixel at Figure 3 - ❹.
If bad pixel is a line, put the range as below at Figure 3 – ❺.
From
4.
To
Row
3327(127type)
2449(140type)
Col
2815(127type)
3051(140type)
If bad pixel is not a line but some pixels, put the rest coordinate at Figure 3 - ❺.
After completing step 5 , check if bad pixel has been changed to green as Figure 2 - ❻.
Click “Fill” at Figure 3 - ❼.
Click "Save" at Figure 3 - ❽.
Once setting BPMM is done, “BPMM.raw” file will be saved at C:\Davinci\CAL.
1417WGC/WCC
96
PART II. Service Manual
4.
Usage
4.1
Set Up
4.1.1
Product Connectivity
1.
Connect the detector and turn on the power.
2.
Open “_vadav.lnk” from “C:\davinci”.
3.
Once the detector is connected, detector information is displayed in Detector Status and
Link & Ready are checked as below.
 If "Detector Status" does not show anything, please refer to 3.1 Installation in
Part.1 User & Installation Manual to connect the detector properly.
1417WGC/WCC
97
PART II. Service Manual
4.1.2
4.1.3
Image Set Up
1.
In order to rotate or flip an image, use the option of "Flip and Rotate" as shown below.
2.
In order to change the size of an image, use "Crop Rows and Columns" as below.
3.
Click "Apply" to save.
Multi Detector Set Up
Refer to 3 Multi Detector Set Up in Part.2 Service Manual for Multi-Detector Setting.
1417WGC/WCC
98
PART II. Service Manual
4.2
Image Acquisition
4.2.1
Product Connection
1.
Connect the detector and turn on the power.
2.
Open “_vadav.lnk” from “C:\davinci”.
3.
Once the detector is connected, information of the detector is displayed in Detector Status
and Link & Ready & Calibration are checked as below.
 If "Detector Status" does not show anything, please refer to 3.1 Installation in
Part.1 User & Installation Manual to connect the detector properly.
 If Calibration is not checked along with black dots checking off "Link" and "Ready"
as above, please refer to 3.2 Calibration in Part.1 User & installation Manual and
perform calibration again.
1417WGC/WCC
99
PART II. Service Manual
4.2.2
Image Acquisition
1.
Click the "Calibration & Acquisition” tab and type the name of the image inside the box
below. After naming the image, click “Get Image”.
2.
Shoot an X-ray once the "Acquiring bright frame" window pops up.
3.
An acquired image will be stored in “C:\davinci\I.3268x2756 (127type) or 2440x2992
(140type)” and the name of the file will be "(typed name from Step 1).raw”.
4.
The format of the stored file is 16 bit little-endian order.
1417WGC/WCC
100
PART II. Service Manual
4.3
View Images
1. Click “View Images”
2. Another window will be popped up as below.
Histogram Set Up
1417WGC/WCC
101
PART II. Service Manual
Pixel value at certain level
Choose "S" from marked box.
Profile for horizontal line
Choose "R" from marked box.
1417WGC/WCC
102
PART II. Service Manual
Profile for vertical line
Choose "C" from the marked box.
4.4
Additional Function
4.4.1
Battery Remain
Once you click "Get Image" under the "Calibration & Acquisition” tab, the Status window will
show how much battery remains.
1417WGC/WCC
103
PART II. Service Manual
4.4.2
Wireless signal Strength
Once you click "Get Image" under the "Calibration & Acquisition” tab, the Status window will
show the strength of the wireless signal.
Before you check the wireless signal in the AP mode, the detector's serial number should
be entered at the "SSID".
1417WGC/WCC
104
PART II. Service Manual
4.4.3
Sleep Mode
1.
Click “Edit” under the Calibration & Acquisition tab.
2.
Under the “Sleep Timeout” setting, enter a designated time for the detector to go into Sleep
Mode.
 Sleep Mode does not apply when set to zero.
3.
To turn off Sleep Mode, attempt to acquire an image or press the power button on the
detector just once.
 A normal image can be acquired after 10 seconds Sleep Mode has been turned
off.
1417WGC/WCC
105
PART II. Service Manual
4.4.4
Preview
1.
Click “Edit” under the Calibration & Acquisition tab.
2.
After checking the Preview and 4x4 Binning, a 4x4 binned image appears which allows for
a quicker image preview.
 By unchecking 4x4 Binning, a normal image preview appears.
 By unchecking Preview, a full frame image appears.
1417WGC/WCC
106
PART II. Service Manual
4.4.5
Recent Frame
The last acquired image can be opened by clicking "Recent Frame" under the "Calibration &
Acquisition” tab.
4.4.6
Restore Connection
When the connection between the detector and PC is lost, the connection can be made again
by clicking "Restore Connection" under the "Calibration & Acquisition” tab.
1417WGC/WCC
107
PART II. Service Manual
4.4.7
Image Storage function
1.
Press the "Mode select button" twice (within 2 seconds) and click one more time to go to
"storage mode".
 Mode status can be checked by the OLED window of Detector

2.
Press the mode select button for 3 seconds to check the status of storaged images.
1417WGC/WCC

When Storage mode is used if the connection between the detector and PC,
the image is stored the memory in the Detector internal

When Mobile sevice connects to Detector, can check correctly Image by
Storage Viewer Web Server


For Storage Viewer, must running DBMS and Web Server envirnoment

If enable ‘0x1e’ parameter by VADAV, running DBMS and Web Server
environment during enter Storage Mode menu


Parameter ‘0x1e’ is Storage Function Disable

Stored Image can be opened and deleted by referring SDK
If disable ‘0x1e’ parameter by VADAV, running DBMS and Web Server
environment durning Detector boot up
For running Storage Viewer environment, OLED is displayed “Check DB
Server”
108
PART II. Service Manual
A.
Mobile Device can connect to Detector’s AP Mode
B.
Connect to "http://192.168.1.80:3000". Sign in with password ‘1234’
C.
Ater enter to worklist menus, click ‘+’ icon and create worklist info.
D.
Input to space Patient ID/Name info that will be insert to image info.
E.
Can see all worklist for will acquire image. Click ‘Capture’ button.
1417WGC/WCC
109
PART II. Service Manual
F.
In detector ready status, can take X-ray image.
G. Shot X-ray  Show image and can check correct  Select Accept/Suspend button
H.
Show all acquisition image list on Storage Mode
 When worklist menu is empty, detector will be created woklist menu and matching
 Some guide image is created by web server demo mode
 After 180 days of acquisition, image is automatically deleted
1417WGC/WCC
110
PART II. Service Manual
4.4.8
Sharing function
1.
Connect the PC and USB IrDA Dongle by using the Micro USB cable.
2.
Set vadavas below
No.
1417WGC/WCC
Overview
❶
Click “sharing” tab
❸
Click “insert”
❺
Check “Enable check box”
❼
Confirm the connection port in device manager of OS
❷
Enter follow thing in order “serial number, IP, bucky number”
❹
Put the SSID and PW of AP
❻
Click “apply” to save
❽
Baud select 115200 and select confirmed port in device manager
111
PART II. Service Manual
3.
Press the "Mode select button" turn on the detector twice (within 2 seconds) and transfer to
"sharing mode".
 Mode status can be checked by the OLED window of Detector

4.
When approaching the detector with USB IrDA connecting to the PC, transfer the Serial
Number information of detector.
5.
Based on the transmitted Serial Number, the CAL folder at the PC is created and the
shooting condition is set then USB IrDA of PC transfers IP/SSID information with Detector.
6.
Detector sets the shooting conditions with the IP/ received SSID.
7.

STA Mode: Detector is set by information of wireless router.

AP Mode: PC connects to S/N AP Host name’s wireless router (Detector).
If the shooting condition setting is completed, completion status appears at the "OLED
window" and "Sharing mode" is turn off.
IrDA LED Color
Status
Green
Connected PC
Blinking Orange
Transfer the data
End Blinking
Turn on Green
Transfer success
End Blinking
Turn off Green
Transfer fail

1417WGC/WCC
112
PART II. Service Manual
4.4.9
OLED function
1.
OLED is operating when detector is turned on.
2.
Use the Mode select button and Power button can be used OLED function as follows.
1417WGC/WCC
 1 click
 Press the mode select button
 2 click
 Press the mode select button twice (within 2 seconds)
 10 sec
 Press the mode select button for 10 seconds
 1 click
 Press the power button
 Normal menu
 - Will Show the (Time / Battery remain / Wireless signal) of
 Sharing menu
 - Operation the “sharing function”
 Storage menu
 - Operation the “storage function”
 SSID menu
 - Will Show the SSID information of detector
 PSK menu
 - Will Show the PSK information of detector
 IP Menu
 - Will Show the IP information of detector
 Mask menu
 - Will Show the Mask information of detector
 Factory reset menu
 - Factory reset menu
detector
113
PART II. Service Manual
5.
Maintenance
5.1
Cleaning
5.2
1.
Clean the detector with IPA (Isopropyl-alcohol) when it is contaminated.
2.
Before cleaning the detector, turn off the power and separate the battery.
3.
Wear waterproof gloves to protect your hands from direct contact with IPA or any other
liquid.
4.
Do not pour or spray IPA directly on the detector. Use fabric or soft cloth moistened with IPA
to clean.
5.
Avoid getting IPA or any other liquid into the detector.
6.
After cleaning, wait until the IPA is dried completely.
Inspection
1.
In order to ensure that the detector is used safely and normally, please be sure to inspect
the product regularly before use. If any problem occurs, please contact Rayence Customer
Service team.
2.
Please perform inspections based on the check list below.
Inspection List
1417WGC/WCC
User
Vendor
Cycle
Check if cables are not damaged
Daily
Check if plugs and connectors are not loose or
damaged
Daily
Check if cover or part is not damaged
Daily
Check the LED indicator
Daily
Re-Calibration
Check the performance of the product by doing test
shots with Phantom or resolution chart
Half Year
Yearly
114
PART II. Service Manual
5.3
Replaceable Parts and Instruction of Replacement
5.3.1
Fuse: T3.15 AL 250V
Replacing the Fuse
1.
Press the fuse as below and pull the fuse box.
2.
Pull the fuse and replace with another fuse.
5.3.2
Power cord: H05VV-F 0.75SQ * 3C
5.3.3
Ethernet Cable: UTP 4PR 24AWG (CAT.6, straight-through)
1417WGC/WCC
115
PART II. Service Manual
5.4
Disposal or Recycling
Follow local governing ordinances and recycling plans regarding the disposal or recycling of
device components.
Disposal of old Electrical & Electronic Equipment
(Application in the European Union and other European countries with separate
collection system.) This symbol indicates that this product shall not be treated as
household waste. Instead, it shall be handed over to the applicable collection point
for the recycling of electrical and electronic equipment. By ensuring this product is disposed of
correctly, you will help prevent potential negative consequences for the environment and human
health, which could otherwise be caused by inappropriate waste handling of this product. For
more detailed information about recycling this product, please refer to local governing
ordinances and recycling plans.
1417WGC/WCC
116
PART II. Service Manual
6.
Warranty
6.1
Warranty
If Buyer promptly notifies RAYENCE or Seller regarding any parts that fail to perform as
specified under normal usage during the Warranty Period and RAYENCE determines that such
failure resulted from a defect in materials or workmanship during the Warranty Period, then
RAYENCE, at its option, shall repair, rebuild or adjust the affected parts.
RAYENCE shall have no obligation for any defects to the extent that such defect arises out of (i)
normal and fair wear and tear or Product which has been modified without RAYENCE's
approval, (ii) Product which has not been installed in strict conformity to the RAYENCE's
directions or which have been subjected to electrical or other abuse, or damaged by improper
handling, storage or use by Buyer or a third party, (iii) use of Product in combination with
devices or products not purchased from RAYENCE; (iv) use or application of Product in a field
or in an environment for which such Product was not designed or contemplated; (v) use of any
parts or material not provided by RAYENCE for warranty service; or (vi) the third party’s
maintenance not certified by RAYENCE; or (vii) force majeure such as natural disaster.
The remedies contained in this warranty are Buyer’s exclusive remedies. RAYENCE shall not,
in any event or under any circumstances, be responsible for damages or other sums in excess
of the total purchase price actually paid by Buyer to Seller i.e., RAYENCE or RAYENCE’s
authorized agent. Without limiting the generality of the foregoing under no circumstance shall
RAYENCE be responsible or liable in any regard with respect to damages from loss of use, loss
of time, loss of data, inconvenience, commercial loss, lost profits or savings, or other incidental,
special or consequential damages claimed by Buyer to arise out of the use or inability to use the
Product, even if Buyer has been advised of the possibility of such damages.
In the event that the product is returned to RAYENCE after the warranty has expired, RAYENCE
reserves the right to invoice a reasonable fee for the repair services provided to Buyer.
RAYENCE shall make the sole final determination about whether the fail to perform occurred in
normal usage (under warranty) or not (excluded from warranty). If the authorized agent or the
Buyer doesn’t accept the result of RAYENCE’s investigation, the burden of proof is on them.
Warranty Procedure
If Buyer needs to make a claim based on this Warranty, Buyer should advise Seller in writing
immediately at the following address:
RAYENCE Co., Ltd.
14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea

Tel: +82-31-8015-6245

Fax: +82-31-8015-6300

E-mail: marketing@rayence.com

www.rayence.com
1417WGC/WCC
117
14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea
Tel: +82-31-8015-6245
Fax: +82-31-8015-6300
E-mail: marketing@rayence.com
www.rayence.com

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : Yes
Author                          : Administrator
Comments                        : 
Company                         : 
Create Date                     : 2017:08:23 13:56:20+09:00
Creator                         :  Word용 Acrobat PDFMaker 15
Source Modified                 : D:20170816060533
Subject                         : 
Title                           : 1417WCC / WGC
Modify Date                     : 2017:08:23 05:02:36Z
Producer                        : 3-Heights(TM) PDF Optimization Shell 4.8.25.2 (http://www.pdf-tools.com)
Page Count                      : 118
Tagged PDF                      : Yes
Page Layout                     : OneColumn
EXIF Metadata provided by EXIF.tools
FCC ID Filing: QIIRY1417WHD

Navigation menu