Resmed Sensor Technologies BM08 Motion Detector Module User Manual Module Specifications

Resmed Sensor Technologies Ltd Motion Detector Module Module Specifications

Manual

BM08_2DOC_001  Page 1  User Manual    User Manual  The BiancaMed BM08 Sensor is a 10.525GHz motion detector. It comprises a motherboard PCBA, an integral RF PCBA, die cast RF metalwork and a custom plastic anti-tamper enclosure. Physical Specifications Parameter Value Remarks Circuit board dimensions 44.5mm by 77.5mm by 6 mm See diagram below Complete assembly (including anti-tamper enclosure) dimensions 63.75mm (x) by 79.3mm (y) by 10mm (z) main body of enclosure,   13.8 mm (z) at enclosure hump 22.5mm (z) max by horn, see Figure 2 below  Figure 1: PCB Layout, showing the Antenna Horn on the top side and the connector at the bottom
Page 2   Figure 2:  3-D Model of PCB Enclosed in Anti-Tamper Enclosure  Radio Frequency Sensor Specifications Parameter Value Remarks Centre frequency 10.525 GHz  RF t/x Pulse 500nS  PRF 1MHz  RF Duty Cycle 50%   Electrical Specifications Parameter Value Remarks Input voltage  +4.5 to 12 VDC  Maximum current draw 25 mA  nominal current draw 22mA  Maximum voltage ripple allowed 100 nVrms /(root(Hz))  Maximum supply voltage allowed 12 VDC  Minimum tolerable supply +4.5 VDC   x y z
Page 3  Parameter Value Remarks voltage Reference Voltage 1.25 VDC  Reference Voltage line output impedance 1 kOhm  Sensor start time before data is available 40 s   Connector Hirose 10-Way 2mm Pitch Header Connector  Analogue Sensor Data   Environmental Specifications Parameter Value Remarks Operating Temperature Range +5 to +35 deg C  Operating Humidity Range  30 to 85% RH (non condensing)  Storage Temperature range -20 to +60C  Storage Humidity Range 20 to 95 RH (non-condensing)  IP value Not Applicable IP value provided by OEM enclosure  Where: Input: VDC = +4.5 to +12 VDC Shutdown = active low, connect to pin 2 to enable sensor Output: Vref = +1.25 VDC M-I= Movement I channel data R-I= Respiration I channel data M-Q= Movement Q channel data R-Q= Respiration Q channel data Figure 3: Pin out of Connector
Page 4  FCC Warning Statements This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant  to  Part  15  of  the  FCC  Rules.  These  limits  are  designed  to  provide  reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the  instructions, may cause harmful interference to radio  communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause  harmful  interference  to  radio  or  television  reception,  which  can  be  determined  by turning the equipment off and on, the user is encouraged to try to correct the interference by one of the following measures:  - Reorient or relocate the receiving antenna. - Increase the separation between the equipment and receiver. - Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. - Consult the dealer or an experienced radio/TV technician for help.  This device complies with Part 15 of the FCC Rules.   Operation  is  subject  to  the  following  two  conditions:  (1)  This  device  may  not  cause  harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. FCC  Caution:  Any  changes  or  modifications  not  expressly  approved  by  BiancaMed  could  void the user's authority to operate this equipment. This  transmitter  module  is  authorized  to  be  used  in  other  devices  only  by  OEM  further transmitter  testing  will  not  be  required.  However,  the  OEM  integrator  is  still  responsible  for testing  their  end-product  for  any  additional  compliance  requirements  required  with  this module installed (for example, digital device emissions, PC peripheral requirements, etc.).  The OEM integrator has to be aware not to provide information to the end user regarding how to install or remove this RF module in the user manual of the end product.  End Product Labeling  The final end product must be labeled in a visible area with the following: “Contains TX FCC ID: YAKBM08” or “Contains FCC ID: YAKBM08”.

Navigation menu