Resmed Sensor Technologies BM08 Motion Detector Module User Manual Module Specifications

Resmed Sensor Technologies Ltd Motion Detector Module Module Specifications

Manual

Download: Resmed Sensor Technologies BM08 Motion Detector Module User Manual Module Specifications
Mirror Download [FCC.gov]Resmed Sensor Technologies BM08 Motion Detector Module User Manual Module Specifications
Document ID1424346
Application IDFLvuhmW6Uaa7tDofTWlVMA==
Document DescriptionManual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize47.66kB (595797 bits)
Date Submitted2011-03-02 00:00:00
Date Available2011-03-02 00:00:00
Creation Date2011-03-02 09:15:09
Producing SoftwareMicrosoft® Office Word 2007
Document Lastmod2011-03-02 09:15:09
Document TitleModule Specifications
Document CreatorMicrosoft® Office Word 2007
Document Author: rshouldice

User Manual
The BiancaMed BM08 Sensor is a 10.525GHz motion detector. It comprises a motherboard
PCBA, an integral RF PCBA, die cast RF metalwork and a custom plastic anti-tamper enclosure.
Physical Specifications
Parameter
Circuit board dimensions
Complete assembly (including
anti-tamper enclosure)
dimensions
Value
44.5mm by 77.5mm by 6 mm
63.75mm (x) by 79.3mm (y)
by 10mm (z) main body of
enclosure,
Remarks
See diagram below
13.8 mm (z) at enclosure
hump
22.5mm (z) max by horn, see
Figure 2 below
Figure 1: PCB Layout, showing the Antenna Horn on the top side and the connector at the
bottom
BM08_2DOC_001
Page 1
User Manual
z
Figure 2: 3-D Model of PCB Enclosed in Anti-Tamper Enclosure
Radio Frequency Sensor Specifications
Parameter
Centre frequency
RF t/x Pulse
PRF
RF Duty Cycle
Value
10.525 GHz
500nS
1MHz
50%
Remarks
Value
+4.5 to 12 VDC
25 mA
22mA
100 nVrms /(root(Hz))
Remarks
Electrical Specifications
Parameter
Input voltage
Maximum current draw
nominal current draw
Maximum voltage ripple
allowed
Maximum supply voltage
allowed
Minimum tolerable supply
12 VDC
+4.5 VDC
Page 2
Parameter
voltage
Reference Voltage
Reference Voltage line output
impedance
Sensor start time before data
is available
Connector
Value
Remarks
1.25 VDC
1 kOhm
40 s
Hirose 10-Way 2mm Pitch
Header Connector
Analogue Sensor Data
Where:
Input:
VDC = +4.5 to +12 VDC
Shutdown = active low, connect to
pin 2 to enable sensor
Output:
Vref = +1.25 VDC
M-I= Movement I channel data
R-I= Respiration I channel data
M-Q= Movement Q channel data
R-Q= Respiration Q channel data
Figure 3: Pin out of Connector
Environmental Specifications
Parameter
Value
Operating Temperature Range +5 to +35 deg C
Operating Humidity Range
30 to 85% RH (non
condensing)
Storage Temperature range
-20 to +60C
Storage Humidity Range
20 to 95 RH (non-condensing)
IP value
Not Applicable
Page 3
Remarks
IP value provided by OEM
enclosure
FCC Warning Statements
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates
uses and can radiate radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio communications. However, there is
no guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the interference by
one of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected.
- Consult the dealer or an experienced radio/TV technician for help.
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received, including interference
that may cause undesired operation.
FCC Caution: Any changes or modifications not expressly approved by BiancaMed could void
the user's authority to operate this equipment.
This transmitter module is authorized to be used in other devices only by OEM further
transmitter testing will not be required. However, the OEM integrator is still responsible for
testing their end-product for any additional compliance requirements required with this
module installed (for example, digital device emissions, PC peripheral requirements, etc.).
The OEM integrator has to be aware not to provide information to the end user regarding how
to install or remove this RF module in the user manual of the end product.
End Product Labeling
The final end product must be labeled in a visible area with the following:
“Contains TX FCC ID: YAKBM08” or “Contains FCC ID: YAKBM08”.
Page 4

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : No
Page Count                      : 4
Language                        : en-IE
Tagged PDF                      : Yes
Title                           : Module Specifications
Author                          : rshouldice
Creator                         : Microsoft® Office Word 2007
Create Date                     : 2011:03:02 09:15:09
Modify Date                     : 2011:03:02 09:15:09
Producer                        : Microsoft® Office Word 2007
EXIF Metadata provided by EXIF.tools
FCC ID Filing: YAKBM08

Navigation menu