SD Biosensor 01GM51 SD GlucoNavii Voice N Blood Glucose Meter User Manual pt2
SD Biosensor, Inc. SD GlucoNavii Voice N Blood Glucose Meter pt2
Contents
- 1. User Manual pt1
- 2. User Manual pt2
User Manual pt2
SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM 4) After cocking the lancing device back, hold the lancing device firmly against the side of finger and then press the release button. CAUTION t A lancet should only be used once. Do not share used lancets with another person. To prevent possible infection, a used lancet should not be touched by another person. Used lancets in the regular trash can be dangerous. We recommend that you throw out the used lancets in sharps containers or test strip vials. Please be sure to save the cap so that the lancets cannot spill out of the container into the trash. Testing Your Blood Sugar t 37 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM 2. Performing a Blood Glucose Test NOTE 1) 2) Always wash hands with warm, soapy water. Rinse and dry them before testing. Remove a new test strip from container. Be sure to tightly replace container cap after removing test strip. Make sure the meter turn off. And then, Insert the test strip into the meter. Testing Your Blood Sugar VOICE N 38 3) When the blood drop symbol flashes (Blood Stand-by Display), you are ready to perform a test. 4) Let your arm hang down at your side to allow blood to flow to your fingertips. Grasp your finger just below the joint closest to the fingertip. Obtain a drop of blood sample using the lancet and lancing device. Hold your finger to the tip of the strip until the yellow window is completely filled with blood. Do not place the blood drop on top of the strip. 5) 6) SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM The blood will be drawn into the strip automatically. If beeper is turned on, meter will beep to let you know the test is beginning. 8) When blood is applied to the strip, start the test. After 5 seconds, the test result appears on the screen. 9) You can set the pre-meal or post-meal test if you want. The symbol indicating pre-meal( ) or post-meal( ) appears on the screen. Choose what you want by pressing the left or right button. If you press the On/Off button, no select any features. 10) When the test is done, pull out the used test strip and discard it. t t NOTE t Remove the inserted lancet from the lancing device and dispose the used lancet with care. Always use fresh capillary or venous whole blood for tests. If you want to use NFC function, refer to âChapter 4. Using the NFC functionâ on page 42~46 . Testing Your Blood Sugar 7) 39 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM 3. Understanding Test Results Testing Your Blood Sugar Your test results 1. After 5 seconds testing time from applying a blood into strip, you will receive a normal result, 10mg/dL to 600mg/dL. 40 2. If your blood glucose is above 600 mg/dL, you will receive a âHIâ and is below 10 mg/dL, you will receive âLoâ. In these cases, repeat the test with new test strip. If this message show again, contact your healthcare professional immediately. 3. If you set the pre-meal mark after the test, you will receive a result with pre-meal mark. ( ) SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM CAUTION If you set the post-meal mark after the test, you will receive a result with post-meal mark. ( If you perform the pre-meal test while the postmeal alarm setting is on, then the post-meal mark will appear automatically on your LCD when you test within following period: from 30min to 130min after your Pre-meal test. Normal Blood Glucose Readings The normal fasting blood glucose range for an adult without diabetes is 74 - 106 mg/dL.2 Two hours after meals, the blood glucose range for an adult without diabetes is less than 140 mg/dL.1 t 'BTUJOHUPNHE- t IPVSTBGUFSNFBMTNHE- What This Means For You Frequent blood glucose testing is the best means to track how well you are doing with your diabetes management. It helps you track the effects of medications, diet, exercise, and stress management. Blood glucose test results can also tell you if your diabetes is changing. This may alert you to adjust your treatment plan. Always consult your healthcare professional before making any adjustments. Frequency of Testing Work with your healthcare professional to decide when and how often to test. This will depend on such things as age, type of diabetes, and medications. It is important to make testing part of your daily routine. Testing Your Blood Sugar 4. 41 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM Chapter 4: Using NFC function 1. Information on NFC function NFC(Near field communication) is a set of standards for smartphones and similar devices to establish radio communication with each other by touching them together or bringing them into close proximit. This function is comfortable for test results back-up at PC or smartphones. So, you can easily monitor your blood glucose results. SD GlucoNaviiÂŽ Voice N follows âFelicaâ standard which include ISO/IEC 18092. 2. Information on Caution for Voice N Using NFC function t 42 t t t Check your smatphone or PC OS version. For PC, It is available for Windows XP, Windows VISTA, Windows 7 and need to SD NFC reader writer purchased separately. For Smartphone, It is available for only smartphones with Android OS v.2.3.1 or higher. Check the list in http://www. nfcworld. com/nfc-data/android/. Before using this function, check the meter turn off. If turn the meter on, you cannot use NFC function. Check the location of chip for NFC in your smartphone. It is different by manufacturer. For more information, refer to manufacturer website of your smartphone. SD NFC software for PC is able to download at www. sdbiosensor.com . SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM t If the distance between the meter and a smartphone(or SD NFC reader writer) is far(about over 2cm/0.06ft), NFC function is not operated. 3. Using NFC function? 2) 3) 4) NOTE After testing or you want to test result back-up, turn the meter off. Check the location of Voice N at your smartphone. If you tap SD GlucoNaviiÂŽ Voice N to your Smart phone, two kinds of action are happened. First, start App. Seconds, Upload Data. (If App. is already running, please, wait a minute.) After finishing the communication, check that the test results are transferred well in your smartphone or PC. More information on App or software for PC, refer to each manual. Using NFC function 1) 43 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM 4. How to use GlucoNaviiApp 1) Download & Install t Search âGlucoNaviiâ in Google Play at your smartphone. You can search other keyword that are âSD Biosensorâ or âNFC glucose meterâ. t Install GlucoNavii App. in your phone. NFC enabled smart-phone with Android OS v.2.3.1 or higher. Check the list in http://www.nfcworld.com/nfc-data/ android/. NOTE t 2) Check the NFC function at your smart-phone. If it is the âOffâ setting, turn NFC function âOnâ. Communicate with SD GlucoNaviiÂŽ Voice N t Please power off SD GlucoNaviiÂŽ Voice N meter. t If you tap SD GlucoNaviiÂŽ Voice N to your smartphone, two kinds of action will be happened. First, if App. is not turning it on, start App. Seconds, if App. is already running, please, wait a minute to upload test results. Please remember the points of contact for both devices. Using NFC function NOTE 44 3) Utilize GlucoNaviiApp (1) Start App. (2) If you need to upload new data, just tap GlucoNaviiÂŽ Voice N to your smart-phone. SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM 5. How to use SD NFC Software for PC 1) 2) 3) 4) NOTE Download the software at www.sdbiosensor.com. And then, start the software. Turn the meter off. If not, you cannot start NFC function. Tap the meter with SD NFC reader writer plugged with PC. And then, Start the communication. After finishing the communication, you can look the log book(test results) or graph. More information on App or software for PC, refer to each manual. If you have any question or queries about SDGlucoNaviiApp, or SD NFC Software for PC , please email to (SD_Dev@sdbiosensor.com). European Union Directive Conformance Statement Hereby, SD Biosensor declares that SD GlucoNaviiÂŽ Voice N blood glucose meter is in compliance with the essential requirements and other relevant provisions of Directive 1999/5/EC. You can view your productâs Declaration of Conformity(DoC) to Derictives 1995/5/EC(R&TTE) Directive at www.sdbiosensor.com. CAUTION WARNING: Changes or modifications made to this equipment not expressly approved by SD Biosensor, Inc. for compliance could void the userâs authority to operate the equipment. Using NFC function NOTE 45 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM Using NFC function NOTE 46 This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: t Reorient or relocate the receiving antenna. t Increase the separation between the equipment and receiver. t Connect the equipment into an outlet on a circuit different. t from that to which the receiver is connected. t Consult the dealer or an experienced radio/TV technician for help. SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM CHAPTER 5 : Using the Meter Memory Using the Meter Memory The meter automatically stores about 500 glucose results, letting your review them in order from the most recent to the oldest. If you have set the time/date feature, the time and date of the results are also displayed. If the memory is full and a new result is added, the meter deletes the oldest result. The meter also calculates three kinds of 7, 14 and 30-day averages of test results stored in memory, 1)normal, 2)premeal and 3)post-meal state averages. You do not need to set the time and date for the meter to give you average calculations. HI/Lo result (results outside of the meterâs reading range) and results with control solution symbol are not included in averages. 47 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM 1. Searching Test Results 1) In Strip Stand-by Display, press the left arrow button to review in sequence from the most recent test result to the last test results stored in memory. Using the Meter Memory [Normal result] 48 [Pre-meal result] [Post-meal result] SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM 2) After the result with date and time display for 1 second, the date will change into the appropriate memory number automatically. 3) If there arenât the stored test results, the following display appears for 1 second, and then the meter will display Strip Stand-by Display automatically. Using the Meter Memory [Control Solution result] 49 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM 4) In Strip Stand-by Display, press the right arrow button to review three kinds of 7, 14 and 30 day averages of test result stored in memory in sequence. (normal, pre-meal and post-meal state) You can also review the number of results at each average in the right bottom of the LCD window. If you press the right arrow button once more after displaying the 30 day average(with post-meal mark), the 7-day average result appear again. Using the Meter Memory [ Normal average ] [Pre-meal average] [Post-meal average] 50 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM NOTE If there arenât any stored 7, 14 and 30-day average of test results, following display will appear on the LCD. You cannot search the stored test results and average of results in the meter, if a test strip is inserted in the meter, Blood Stand-by Display. After removing the test strip from the meter, you can search the test results and average of results stored in memory by pressing the left or the right button. Using the Meter Memory 5) 51 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM 2. Downloading results to a computer You can use your meter with SD NFC Software to store your records and to help you spot patterns for planning meals, exercise, and medication. SD NFC Management Software puts information downloaded from the meter into charts, diagrams and graphs. 1) 2) Using the Meter Memory NOTE 52 Obtain the SD NFC Software and SD NFC R/W. Install the software on a personal computer. Please refer to Software Product Manual. t While the meter is connected to the PC, it is unable to perform a blood glucose test. t For downloading SD NFC Software and Software Product manual(both are free of charge), please visit www.sdbiosensor.com. t For ordering SD NFC Reader/Writer, please contact our representative. t For more information,please refer to Software Product Manual. If the cable port got ESD while downloading data, time delay, a few seconds, may be happened. After a while, the meter retransmit data automatically. SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM CHAPTER 6 : Maintenance and Troubleshooting 1. Performing Check strip Test When you do the check strip test? t When you want to easily check the performance of the meter. t Before using your meter for the first time. t Whenever your result does not agree with the level you feel. t If you have repeated a test and the blood glucose result is still lower or higher than expected. Check Strip test does not replace Control Solution test. How to Use the check strip 1. Insert SD Glucose Check strip (facing up âCheck stripâ printed in arming knob) into test strip slot. The meter turns on automatically. 2. If the check strip is inserted properly, the meter will start the check. VOICE N Check Strip Check Strip Maintenance and Troubleshooting CAUTION 53 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM 3. The check result appears on the screen in 5 seconds. âOKâ message appears on the screen if there isnât any problem for the meter. If not, âEEEâ message appears on the screen. [ 'OK' Message ] [ 'EEE' Message ] Maintenance and Troubleshooting 2. Cleaning the meter 54 Caring for SD GlucoNaviiÂŽ Voice N meter is easy. Just keep it free of dust. If you need to clean it, follow these guideline carefully to help you get the best performance possible. To prevent malfunction of the meter, keep the test strip port free of blood, moisture, dirt, or dust. Use a lint-free cloth dampened with water to clean meter. Thoroughly wring out cloth before use. Do not use an abrasive cloth or antiseptic solution, as these may damage the display screen. SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM 3. Maintenance, Testing and Transportation Precautions for Maintenance, Testing and Transportation 1) Meter t Keep the test strip slot free of dust. t Protect the internal meter from humidity. t The carrying case is designed to let you store a variety of supplies you may need and helps to protect your meter. t If you keep the meter with the battery inserted, then keep it in a low humidity environment. t Do not modify the meter, such as resembling or remodeling, NFC function is not operated well. If any function is not perated, please contact SD Biosensor, Inc. 2) Test strip t The test strip is sensitive to humidity, keep it in a dry and cool environment, and do not store in direct sunlight. t After pulling out the test strip from its container, close a container cap of the test strip immediately. t The test strip container closes tightly and can protect the test strips, so you should keep the unused test strips in the container in which they came. Maintenance and Troubleshooting The meter needs little or no maintenance with normal use. It automatically tests its own systems every time you turn it on and lets you know if something is wrong. If you drop the meter or think it is not giving accurate results, make sure that your test strips and control solution havenât expired, and then run a control solution test. 55 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM Maintenance and Troubleshooting 3) Lancet and Lancing device t The needle of lancet is sharp, keep the lancet away from children. t Keep the lancet and lancing device dry and do not store in direct sunlight, or high heat and humidity locations. t A lancet should not use for the other intended use except sampling blood. t A lancet is for single use only. Do not reuse. t A lancet provided with SD GlucoNaviiÂŽ Voice N Blood glucose meter is able to use every lancing devices manufactured by other company. But if at all possible, use the lancing device manufactured by SD Biosensor, Inc. t Before using, check a packaging condition, if there is any problem, you should not use it. t If a lancet protective disk is loosed or needle of a lancet is exposed, you should not use it. t To reduce the chance of infection for the used lancet, discard it. 56 4) Control solution t Keep SD GlucoNaviiÂŽ control solution in 8-30°C (4686°F) environment. t Do not refrigerate or freeze. t Do not use SD GlucoNaviiÂŽ control solution that has passed the expiration date. t SD GlucoNaviiÂŽ control solution can be used for 3 months after opening the container. Write the opened date on the container when you first opened. t No reconstitution or dilution is necessary. t Wipe the container tip clean and reseal the container tightly after each use. 4. Cleaning the Lancing Device Clean the outside of SD lancing device regularly with 70% isopropyl (rubbing) alcohol. Do not place the entire device under water. Do not use bleach. At least once a week, disinfect the cap after cleaning by placing it in 70% rubbing alcohol for 10 minutes. Allow the cap to air-dry after disinfecting. SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM 5. Screen Messages and Troubleshooting Message Description t The meter turns on normally. [Blood Stand-by Display] t The meter is ready for a drop of blood. t The meter is ready for a drop of blood with pre-meal mark. Maintenance and Troubleshooting [Strip Stand-by Display] t The meter is ready for you to insert a test strip. 57 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM t The meter is ready for a drop of blood with post-meal mark. t The meter shows the test result. t The meter shows three kinds of average of results. Maintenance and Troubleshooting [Normal Results Average] [Pre-meal Results Average] 58 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM t The meter shows the saved results of blood glucose. t The meter is ready for a drop of control solution. Maintenance and Troubleshooting [Post-meal Results Average] 59 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM t Maintenance and Troubleshooting t 60 t [Low battery] At this time, battery is getting low but you can still perform about 50 tests. Replace the battery soon. See Chapter 1 â9. Changing the Batteryâ. [Replace battery] Battery power is low. Replace the battery immediately. See Chapter 1 â9. Changing the Batteryâ. If you press the ON/OFF button after discharging of the battery, the battery icon will flash and then after ten seconds the meter will turn off automatically. [HI message] Blood glucose may be higher than the measuring range of the system. See chapter 3 â4. Understanding Test Resultsâ. SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM t t [Lo message] Blood glucose may be lower than the measuring range of the system. See chapter 3 â4. Understanding Test Resultsâ. [Internal Error Message for a meter] Turn off a meter. Then turn on the meter again. If there is still error message, please contact SD Biosensor, Inc. TEL: +82-31-300-0400. [Strip Error] Defective test strip or the test strip is damaged or inserted improperly. Discard this test strip and test again using new test strip. See Chapter 3 â2. Performing a Blood Glucose Testâ. Maintenance and Troubleshooting t 61 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM t Maintenance and Troubleshooting t 62 t [Blood Sample Error] An insufficient amount of blood was applied. Discard this test strip and test again using new test strip and a larger sample, making sure blood is placed to the narrow channel in the top edge of the test strip. See Chapter 3 â2. Performing a Blood Glucose Testâ. [Temperature Error] If the environmental temperature is above or below the operating range of a meter, a thermometer icon will appear on the display. Move to an area between 8-45°C (46-113°F), wait for 30 minutes, and perform a test. Do not artificially heat or cool the meter. See Chapter 7. âProduct Technical Informationâ. [Communication Error] The communication between meter and computer is failed. Connect again between meter and PC. SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM t t t t t t Never make significant changes to your diabetes control program or ignore physical symptom without consulting with your healthcare professional. Severe dehydration (excessive water loss) may cause false low results. If you believe you are suffering from dehydration, consult your healthcare professional right away. Extremes in hematocrit may affect test results. Hematocrit levels greater than 70% may cause falsely low readings. Inaccurate results may occur in severely hypotensive individuals or patients in shock. Inaccurate low results may occur for individuals experiencing a hyperglycemic-hyperosmolar state, with or without ketosis. Critically ill patients should not be tested with blood glucose meters. Normal endogenous (within body) natural levels of uric acid, ascorbic acid (vitamin C), bilirubin, triglycerides, and hemoglobin do not interfere with your blood glucose results obtained. Interferences : The following compounds, elevated levels of ascorbic acid, uric acid, acetaminophen, total bilirubin, triglycerides may affect results. Compound Limitation Ascorbic acid > 4 mg/dL Uric acid > 9 mg/dL Acetaminophen > 6 mg/dL Total bilirubin > 40 mg/dL Triglycerides > 1800 mg/dL Maintenance and Troubleshooting 6. Warnings, Precautions and Limitations 63 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM t t t t Maintenance and Troubleshooting t 64 Do not use during or soon after xylose absorption testing. Xylose in the blood will cause an interference. SD GlucoNaviiÂŽ Voice N System is not designed to be a substitute for pathology laboratory equipment and should not be used for the diagnosis of diabetes. Do not use SD GlucoNaviiÂŽ Voice N meter to test neonates. It has not been validated for neonatal use. Preservatives: When using the venous blood foe sample, blood may be collected by healthcare professional into test tubes the anticoagulants(EDTA, lithium heparin or sodium heparin). Do not use other anticoagulants or preservatives. Peritoneal dialysis solution: Icodextrin does not interfere with SD GlucoNaviiÂŽ Voice N test strip. SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM CHAPTER 7 : Product Technical Information 1. System Specifications Result Range 10 - 600 mg/dL, (0.6 - 33.3 mmol/L) Calibration Plasma-equivalent Sample Fresh capillary or venous whole blood Sample Size 0.5 microliter Test Time 5 seconds Glucose Dehydrogenase Biosensor Two replaceable 1.5V Alkaline Batteries type AAA Battery Life Around 1,000 tests Glucose Unit mg/dL , mmol/L Display LCD (Customized) Controls 3 Buttons Size 52 mm x 95 mm x 19 mm Weight 150g (with batteries) t Automatic Shutoff t 1 minutes after last user action without inserting test strip into the meter 3 minutes after last user action when inserting test strip into the meter Product Technical Information Assay Method ON/OFF Source 65 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM Memory 500 blood glucose tests Function - Hypo warning : 60, 70, 80 mg/dL(3.3, 3.9, 4.4 mmol/L) - Pre-meal and post-meal mark - Alarm setting (up to 4 times) - Post-meal Alarm - 7-, 14- and 30-day Averages of the following results 1)Normal Results 2)Pre-meal Results 3)Post-meal Results - Automatic shutoff - Voice announce (3 Language) Product Technical Information - Meter 66 Operation Temperature 8 - 45°C (46 - 113°F) Hematocrit 0 - 70% Altitude Up to 11,549 feet. (3,520 meters) - Test strip Storage Temperature 2 â 32°C (36 â 90°F) SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM Annex 1 : Information for Healthcare Professionals Decisions about whether to recommend alternative site testing (AST) should take into account the motivation and knowledge level of the patient and his or her ability to understand the considerations relative to diabetes and AST. If you are considering recommending AST for your patients, you need to understand that there is a potential for a significant difference between fingertip and alternative site blood glucose test results. The difference in capillary bed concentration and blood perfusion throughout the body can lead to sample site-to-site differences in glucose results. These physiological effects vary between individuals and can vary within a single individual based upon his or her behavior and relative physical condition. Our studies involving AST of adults with diabetes show that most persons will find their glucose level changes more quickly in the fingersâ blood than the alternative sitesâ blood. This is especially important when glucose levels are falling or rising rapidly. If your patient is used to making treatment decisions based upon fingertip readings, he or she should consider the delay or lag-time, affecting the reading obtained from an alternative site. Product Technical Information Healthcare professionals : Follow the infection control procedures appropriate for your facility 67 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM Annex 2 : Symbol The following list describes all symbols used on SD GlucoNaviiÂŽ Voice N Blood glucose monitoring (BGM) system. Symbol Description Caution, consult accompanying documents Consult instructions for use Product Technical Information In-vitro diagnostic medical device: This system is intended to use outside the body (in vitro diagnostic use). 68 Serial number for this meter. Date of manufacture : To indicate the date of manufacture for this meter SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM Symbol Description Lot No. or Batch Code To indicate the lot number for this system Expiry date: This system should be used by the given date. To indicate the temperature limitations in which the transport package has to be kept and handled Catalogue number: To indicate the catalogue number for this system To indicate the manufacturer To use for 6 months after first opening the container. DO NOT REUSE: To warn the user of a piece of equipment that it is for single use only and that it must not therefore be used more than once Product Technical Information $POUBJOT4VĂłDJFOUGPSO5FTUT 69 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM Annex 3 : References 1. 2. Product Technical Information 3. 70 American Diabetes Association, Clinical Practice Recommendation Guidelines 2003, Diabetes care, Vol. 26. Supplement 1. p.22 Stedman, TL. Stedmanâs Medical Dictionary, 27th Edition, 1999, p. 2082. Ellen T. Chen, James H. Nichols, Show-Hong Duh, Glen Hortin, MD: Diabetes Technology & Therapeutics, Performance Evaluation of Blood Glucose Monitoring Devices, Oct 2003, Vol. 5, No. 5 : 749 -768 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM Return You must contact SD Biosensor Customer Service at +8231-300-0400 before returning your meter. You will be instructed how to return the meter to SD Biosensor, Inc. Returned meters without this authorization will not be accepted. Product Technical Information This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. 71 SD GlucoNaviiÂŽ Voice N BLOOD GLUCOSE MONITORING SYSTEM Manufactured by SD BIOSENSOR, Inc. C-4th&5th Floor Digital Empire Building 980-3, Yeongtong-dong, Yeongtong-gu, Suwon-si, Kyonggi-do, Korea Product Technical Information EC-Representative 72 Voice N FCC ID : RPJ01GM51 0168 0123
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.7 Linearized : No Encryption : Standard V4.4 (128-bit) User Access : Print, Print high-res Page Count : 37 XMP Toolkit : XMP toolkit 2.9.1-13, framework 1.6 About : ea9344a9-97a4-11e2-0000-3654ea62d053 Producer : GPL Ghostscript 9.02; modified using iTextSharp 5.0.5 (c) 1T3XT BVBA Keywords : () Modify Date : 2013:03:25 12:42:25+00:00 Create Date : 2013:03:25 12:41:48+00:00 Creator Tool : PDFCreator Version 1.2.2 Document ID : ea9344a9-97a4-11e2-0000-3654ea62d053 Format : application/pdf Title : User Manual pt2.pdf Creator : emery-na Description : () Author : emery-na Subject :EXIF Metadata provided by EXIF.tools